Sei sulla pagina 1di 4

SI- Units

Parameters that we calibrate at non-SI units also.


Parameter SI Unit
Pressure pascal (Pa)
Torque newton-meter (Nm)
Force newton (N)
Mass kilogram (kg)
Volume cubic meter (m³)
Length meter (m)
Area square meter (m²)
Temperature kelvin (K)
Speed meter per second (m/s)
Illuminance lux (lx)

Parameters that do not have non-SI units.


Parameter SI Unit
Current ampere (A)
Voltage volt (V)
Resistance ohm (Ω)
Capacitance farad (F)
Inductance henry (H)
Frequency hertz (Hz)
Energy joule (J)
Power watt (W)
Power Factor PF
Phase Angle °
Relative Humidity %RH
Mass flow rate kilogram per second (kg/s)
Volume flow rate cubic meter per second (m³/s)
Acceleration meter per second squared (m/s²)
Sound decibel (dB)
RPM Revolution per minute (rpm)
Time Second (s)
pH % Hydrogen (pH)
Conductivity siemens per meter (S/m)
Note: There is no conversion factor required to be mentioned in the certificate for
the metric prefixes (powers of 10).
Example: The SI unit of Mass is kilogram (kg); if you calibrate in milligram (mg) or
gram (g), then you don't need a conversion factor since these are metric prefixes
(powers of 10).

Possible non-SI units with conversion factor

Pressure
Unit Certificate Remark
Calibration is carried out using SI standards, and bar values are reported using a
bar
conversion factor of 1 bar = 100000 Pa.
Calibration is carried out using SI standards, and psi values are reported using a
psi
conversion factor of 1 psi = 6894.757 Pa.
Calibration is carried out using SI standards, and kg/cm² values are reported using a
kg/cm²
conversion factor of 1 kg/cm² = 98066.5 Pa.
Calibration is carried out using SI standards, and mbar values are reported using a
mbar
conversion factor of 1 mbar = 100 Pa.
Calibration is carried out using SI standards, and mmH2O values are reported using a
mmH2O
conversion factor of 1 mmH2O = 9.80665 Pa.
Calibration is carried out using SI standards, and inH2O values are reported using a
inH2O
conversion factor of 1 inH2O = 248.84 Pa.
Calibration is carried out using SI standards, and mmHg values are reported using a
mmHg
conversion factor of 1 mmHg = 133.3224 Pa.
Calibration is carried out using SI standards, and inHg values are reported using a
inHg
conversion factor of 1 inHg = 3386.388 Pa.

Torque
Unit Certificate Remark
lb.in Calibration is carried out using SI standards, and lb.in values are reported using a conversion
factor of 1 lb.in = 0.112985 N.m.
lb.ft Calibration is carried out using SI standards, and lb.ft values are reported using a conversion
factor of 1 lb.ft = 1.355818 N.m.
Force
Unit Certificate Remark
kgf Calibration is carried out using SI standards, and kgf values are reported using a conversion
factor of 1 kgf = 9.80665 N.
lbf Calibration is carried out using SI standards, and lbf values are reported using a conversion
factor of 1 lbf = 4.44822 N.

Mass
Unit Certificate Remark
lb Calibration is carried out using SI standards, and lb values are reported using a conversion
factor of 1 lb = 0.453592 kg.

Volume
Unit Certificate Remark
ml Calibration is carried out using SI standards, and ml values are reported using a conversion
factor of 1 ml = 1 cm³.

Length
Unit Certificate Remark
inch Calibration is carried out using SI standards, and inch values are reported using a conversion
factor of 1 inch = 25.4 mm.

Area
Unit Certificate Remark
ft2 Calibration is carried out using SI standards, and ft2 values are reported using a conversion
factor of 1 ft2 = 0.092903 m².

Temperature
Unit Certificate Remark
°C Calibration is carried out using SI standards, and °C values are reported using a conversion
factor of T(K) = T(°C) + 273.15.
°F Calibration is carried out using SI standards, and °F values are reported using a conversion
factor of (T)K = 5/9 [(T)F + 459.67].
Speed
Unit Certificate Remark
ft/min Calibration is carried out using SI standards, and ft/min values are reported using a conversion
factor of 1 ft/min = 0.00508 m/s.
mph Calibration is carried out using SI standards, and mph values are reported using a conversion
factor of 1 mph = 0.44704 m/s.

Illuminance
Unit Certificate Remark
Fc Calibration is carried out using SI standards, and Fc values are reported using a conversion
factor of 1 Fc = 10.7639 lx.

Potrebbero piacerti anche