QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
Institution Profile
Phthalate
Gas Chromatography
Mass Spectrometer
Control Chart
Analytical Method
Result Discussion
Conclusion
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
Instrument Laboratory
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
PHTHALATE
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
History
1910
Mikhail S. Tswett
1930
Schuftan and Eucken
1941
Martin and Synge
1952
Gas Chromatography
1955
PerkinElmer
1959
GOW-MAC
1973
HawlettPackard (HP)
2007
Agilent Technologies
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Main Definition
Gas Chromatography is a separating method by using Sample for GC liquid as stationary phase and gas as mobile pha of This method can analyze gases, of Good Factorsse. Separating Process and Indicationvolatile compounds, or easy to be volatilized.
Separation
The Advantages of GC
The Disadvantages of GC
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Basic Principal
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Instrumentation
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Carrier Gas
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Sample Injection
Direct Injection
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Sample Injection
Split Injection
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Sample Injection
Splitless Injection
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Sample Injection
On Column Injection
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Oven
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Oven
Isothermal Temperature Programming
Low Temp.
High Temp.
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Column
Packed Column
Capillary Column
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
GAS CHROMATOGRAPHY
Detector
An electronic sensor to convert the signal of carrier gas and the components within it into electronic signals.
MASS SPECTROMETER
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
MASS SPECTROMETER
-- - Ion Source
sample
+ + + ++ + ++
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
CONTROL CHART
Statistical Process Control
A powerful problem-solving tools and useful in achieving process stability and improving capability through the reduction of variability. Histogram Check Sheet Pareto Chart Cause and Effect Diagram Defect Concentration Diagram Scatter Diagram
Control Chart
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
CONTROL CHART
Control Chart
A graphical design of a quality characteristic that has been measured from a sample versus the sample number or time.
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
CONTROL CHART
The Function of Control Chart IMPROVE THE PROCESS
Input
Output
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
ANALYTICAL METHOD
Vial 2 mL
GC/MS
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
RESULT | DISCUSSION
Phthalate Standards : 1. Diisobutyl Phthalate 2. Dibutyl Phthalate 3. Benzyl Butyl Phthalate 4. Di (2-ethylhexyl) Phthalate 5. Diisononyl Phthalate Start Date : January 2, 2012 End Date : January 12, 2012
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
RESULT | DISCUSSION
PARAMETER CASNUMBER STANDARD A-1 A-2 A-3 A-4 A-1 A-2 A-3 A-4 A-1 A-2 A-3 A-4 A-1 A-2 A-3 A-4 A-1 A-2 A-3 A-4 CONCENTRATION (ppm) 2.394 4.787 11.968 23.936 1.998 3.996 9.990 19.976 2.319 4.639 11.600 23.194 2.308 4.617 11.540 23.084 20.056 40.112 100.280 200.560 AVERAGE A STD/A ISTD 0.6518 1.4317 4.0382 8.4110 0.0564 0.1039 0.2853 0.6179 0.3115 0.7478 2.1429 4.8262 0.3382 0.8291 0.3726 5.5787 2.1688 5.3151 15.6758 40.1682 UCL 0.6978 1.5325 4.2832 8.9353 0.0700 0.1167 0.3120 0.6504 0.4140 0.8294 2.5221 5.3746 0.3984 0.9126 2.4775 6.1701 2.9409 6.6765 16.6987 43.6265 LCL 0.6058 1.3309 3.7932 7.8867 0.0428 0.0912 0.2587 0.5854 0.2089 0.6662 1.7639 4.2779 0.2780 0.7456 2.2677 4.9873 1.3967 3.9536 14.6530 36.7098 MAX 0.6737 1.4850 4.1099 8.5794 0.0623 0.1087 0.2964 0.6366 0.3744 0.7916 2.4189 5.1656 0.3672 0.8632 2.4204 5.8061 2.4684 5.9787 16.2198 41.2930 MIN 0.6384 1.3918 3.9064 8.1178 0.0490 0.0970 0.2695 0.6053 0.2720 0.7140 2.0403 4.5749 0.3112 0.7954 2.3099 5.1803 1.7021 4.7935 15.3427 37.9405 %RSD 2.35% 2.35% 2.02% 2.08% 8.02% 4.09% 3.11% 1.75% 10.98% 3.64% 5.90% 3.79% 5.94% 3.36% 1.47% 3.53% 11.87% 8.54% 2.17% 2.87%
Diisobutyl Phthalate
84-69-5
Dibutyl Phthalate
84-74-2
85-68-7
Di (2-ethylhexyl) Phthalate
117-81-7
Diisononyl Phthalate
28553-12-0
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
RESULT | DISCUSSION
Parameter Inject 1 Inject 2 Inject 3 Inject 4 Inject 5 Inject 6 Inject 7 DIBP 0.9999 0.9999 0.9997 0.9998 0.9999 0.9989 0.9999 DBP 1.0000 0.9999 0.9999 0.9998 0.9987 0.9991 1.0000 BBP 0.9999 0.9997 1.0000 0.9999 0.9997 0.9999 0.9999 DEHP 1.0000 0.9998 0.9998 0.9994 0.9999 0.9972 0.9997 DNIP 1.0000 1.0000 0.9994 0.9996 1.0000 0.9996 0.9996 Parameter Inject 1 DIBP 0.9927 0.9893 0.9934 0.9876 0.9817 0.9777 0.9753 DBP 0.9899 0.9861 0.9895 0.9811 0.9772 0.9731 0.9691 BBP 0.9905 0.9856 0.9906 0.9843 0.9843 0.9746 0.9707 DEHP 0.9897 0.9841 0.9891 0.9829 0.9818 0.9725 0.9690 DNIP 0.9881 0.9816 0.9858 0.9818 0.9753 0.9697 0.9645
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
CONCLUSION
Based on the calculation, it can be concluded that the performance of GC/MS Agilent 5975 Series is still in good condition and the addition of internal standards is important in phthalate analysis by the method of GC / MS.
QUALITY CONTROL PERFORMANCE FOR PHTHALATE ANALYSIS BY GAS CHROMATOGRAPHY MASS SPECTRA
SUGGESTION
The maintenance needs to be done regularly to achieve the optimum conditions, such as cleaning the ion source, checking the vacuum pressure, interface, and column. Limit of Quantitation (LoQ) and Limit of Detection (LoD) of the instrument needs to be verified again so that the reading process of the analyte is stable and under control.