Ozone
Choices of hardware:
APM411 Attachment with RDS
Timer
Time Totalizer
Important Changes:
Pump provides enough suction pressure to allow
unit.
Impinger designed for Ozone monitoring provided
as an Optional Accessory
UV ABSORPTION METHOD for
Measurement of Ozone
Ozone is continuously measured by ultraviolet absorption
photometry.
In the ultraviolet analyzer, a mercury vapor lamp is used to
produce ultraviolet radiation at 254 nm which is
absorbed by the ozone in the air sample. The ozone
signal is determined by the difference between
ambient air containing ozone and ambient air with
the ozone removed or scrubbed.
The ultraviolet analyzer is calibrated by comparison with an
ozone photometer which is certified as a transfer standard.
The transfer standard is certified against absolute ozone
photometers located at the California Air Resources Board
test and laboratory facilities or NIST. The minimum
detectable level of UV monitors is about 2-5 ppbv.
UV ABSORPTION METHOD
Interference's
Water vapor may also interfere with the UV method when water
vapor concentrations are high and variable. These interference's
appear to be due to the condensation of water vapor on imperfect
absorption cell windows and Ozone scrubber.
In tropical area we recommend the installation of heated “ Ozone
Scrubber” to reduce interference's from water vapour.
EC9810 Principle
Sample Microprocessor
Inlet
O3 Output
Particulate
Filter
Detector
UV source Absorption
(Measurement Cell)
exhaust
EC9810 Operation
Agra
§ Sri Ramchandra Institute, Chennai
§ NIT, Trichy
Siting of Monitoring Station