3nethra aberro

Product
Description and
Special
Instructions
Product Description
Handheld Auto-Refractor – Fast & Convenient

Automatic capture within a few seconds

Based on advanced wavefront aberrometry technology

Works for all ages, from children to adults

Simple and intuitive UI

Capture in normal ambient lighting conditions

Lightweight and easily transportable

2
Product specification
Spherical Measurement Range -14D to +14D, increments (0.25D)
Cylindrical Measurement Range -7D to 0D, increments ( 0.25D)
Axis Measurement Range 0 to 180 degree, 1 degree increment
Min Pupil Diameter 2.5 mm Automatic capture within a few seconds
Vertex Distance 19.5 mm
Based on advanced wavefront aberrometry technology
Patient Age Range Above 5 years
Accommodation control Low spatial frequency sinusoidal grating with Gaussian envelop
Works for all ages, from children to adults
Technology Wave-front Sensing
Measurement Time <5 seconds per eye
Battery Life 12 hours, fully charged Capture in normal ambient lighting conditions
Power source 5V DC Micro USB A ,Li-ion battery
Lightweight and easily transportable
Display / Connectivity 3.5” Color touch screen / USB and Bluetooth
Weight / Dimension 650 g / 8.9” X 7.5” X 2.9”
Internal Storage Memory 400 measurements

3
Package contents

Note: Target chart included in the packaging


4
Device Feature's

5
Advantages 3nethra aberro vs AR
 Easy to operate, Optometrist not required, can be used with
less training
 No moving parts so, less maintenance cost.
Automatic capture within a few seconds
 Overall cost of ownership will be less
 Easy to take measurements once the device aligned with pupil
 In aberro each eye 5 seconds to take the readings, but AR
more than 20 seconds
 Portable, easy to carry - Home screening possible.
 AR cannot take the measurement for cataract patients , aberro
can read for mild and bit cataract patients.
 Aberro can take measurement in supine position, Axis will be
compensated by 90deg.
 Cylindrical accuracy is higher than AR.
6
Accuracy based on clinical evaluation

Automatic capture within a few seconds


 3nethra aberro achieves 95% accuracy compared with
desktop auto refractor.
Works for all ages, from children to adults
 3nethra aberro achieves 86% accuracy compared with
subjective refraction.

7
Instruction to the Operator
 Ensure the below points before taking readings
 Place the target(provided with the package) at 6 meter
distance.
 If there is a space constraint use the mirror at 3 meter
distance(target will be behind patient max possible
distance)
 Avoid bright light above patient.

8
Instruction to the Patient
Pl say it every time for each eye
Non-Dilated Eye
 Don’t look into the device
 Look straight
 Look far to see the target
 Wide open the eye
 This instruction is very important to get patient cooperation and accuracy of reading
 Especially for Patient below 20 years, their eye can accommodate easily.
 For patient with small pupil – wide opening of eye will help.

9
Dilated Eye
 It will be better to Enable internal fixation for the dilated Eye’s
before taking reading. ( Pl make sure to un check this while in the normal use)

 Inform the patient to look inside the internal fixation target and
Close the other eye using hand as per the below image.

 Look into the device


 Look into the device red dot
 Look inside the internal fixation
of the centre
 Wide open the eye

10
Do’s and Don’t – Operator Position

Automatic capture within a few seconds

✔️
 Operator should not block the long
distance view of patients.
11
Do’s and Don’t – Device position

✔️
 Operator should position the device straight to the
patient axis. 12
Do’s and Don’t – Device position

Automatic capture within a few seconds

Capture in normal ambient lighting conditions

✔️
 Operator should hold the device straight.
13
Do’s and Don’t – Align the Pupil

 Align the pupil to the centre of the white circle


✔️
Patient not cooperating – How to find it 14
Special Note
 If the message appears as the circled are ,for the patient with
pupil diameter less than 3mm dia results may have reduced
accuracy of 0.5D to 1.5D – Spherical readings

 During this time take the readings again by dimming the lights
and ask patient to close the eye for sometimes (or using single
dilation drop or short dilation mode, if needed )

 For diseased eye and very low pupil eyes – in case of not able to
take readings, Pl use override mode with best pupil alignment
position without shaking the hand. But the readings will be taken
with errors ( +/- 1D) and the pupil diameter reading will be Zero

15
Special Note

 For diseased eye or very low pupil eyes or some eye even after
aligned properly not able to take readings, Pl use override mode
with best pupil alignment position without shaking the hand. But
the readings will be taken with errors ( +/- 1D) and the pupil
diameter reading will be Zero.

16
Cleaning of Eye Cup
Use cotton or tissue paper dipped in IPA for cleaning the eye cup.
Clean the Eye cup on both inner and outer side edge of the eye cup.

17
Printer Rolls

One roll of thermal printing paper for Bluetooth printer has been provided with the Bluetooth
printer.
For further use please inform customer to buy the Thermal paper Roll of dimension
57mm*13mm.

Please refer the below amazon link where they can buy the printer rolls :

https://www.amazon.in/gp/product/B0148WE1Y6/ref=ppx_yo_dt_b_asin_title_o04_s00?ie=UTF8
&psc=1

18
Training - UI Features

Start Refractor
Measurement
Current s/w version in device
Interchange the OS/OD option
Enable/Disable accommodation
target
Restart the device
Update new released software
Export saved reading to USB
To view captured
Delete all the data
and saved readings
Pair new BT printer/ View
paired BT printer
Back to Home page

19
Training - How to take, save and print the readings?
STEP-1 STEP-3 STEP-5

STEP-2 STEP-4
20
STEP-1 Training – Update Software

STEP-2 Alert

STEP-3

21
STEP-1 Training – Export Tests/Saved Data

STEP-2 Alert

STEP-3 STEP-4

22
Training – Export Tests/Saved Data
STEP-1

STEP-2

Note: The device will capture five images for each eye. Device shall be considered best three images
out of five images for the refractive error calculation
23
Training – Exported Tests
STEP-1 STEP-2

STEP-4

Keratometer

Wavefront patern
24
STEP-1 Training – Pair BT Printer

STEP-2 STEP-4

STEP-3 STEP-5

25
Training – View Readings
STEP-1

STEP-2
26
Thank You