Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Foto : Cahaya
Meter : Pengukuran
X solution
I A I
o
b
An incident light beam with intensity Io passing through a
square curette containing a solution of a compound that
absorbs light of a certain wavelength ( ).
The intensity of the transmitted light beam Is will be less than
Io and we defined the transmittance ( T ) of light as Is / Io.
The absorbance light of a certain wavelength is related to the
concentration of the substance being measured.
A
A max
A1 Figure I
Max = 500nm
A2
A1=<<
= 400nm
At
A2 = >>
100 500
nm
( max)
A Max
Figure II
max = 490 -510
490 510
490 510 nm
A
Figure III
min = 550 nm
A min
Filter :
340, 405, 492. 546, 578
m
Spectrum
Light source :
* Tungsten lamp Cuvette :
* Halogen lamp * Round / square
* Mercury lamp * Material : glass, silica / quartz, plastic
* Hydrogen lamp
Detectors :
Monochromator : * Barrier layer cell
* Filter * Photomultipleir tube
* Grating
* Prism
Meter = readout devices
* Direct reading : output photocell, directly to the meter
* Digital readout devices, absorbance concentration
Hg - line spectrum
Energy
Energy
Io = incident light
I = transmitted light
b =the distance of the light
passed through
Io A I
A = a.b.c = log 10 / % T =
2 log % T
B
A = absorbance
a = absorptivity
b = distance of the light passed through = diameter of the cuvette
c = concentration of the subtance in a solution
%T = percent transmittance ( T = I / Io )
Filter photometer:
Light source:
Lamp/ halogen lamp
Source lamp
Cuvette adaptor
I II
%T A
Conc. Conc.
Absorbance
2.0 1.0 .80 .70 .60 .50 .40 .30 .20 .10
0 10 20 30 40 50 60 70 80 90 100
Transmmitance
Figure I : %T - concentration
Figure II : A - concentration
Figure III : scale of the relation A -%T
Beers law application
Solution I
Subtance = p
Concentration = Cl mg / 100 ml
Reading of absorbance = A1
Absorptivity of max = a
cuvette = b cm
I
A1 = a.b.cl
Solution II
Subtance = p
Concentration = C2 mg / 100 ml
Reading of absorbance = A2
Absorptivity of max = a
cuvette = b cm
II A2 = a.b.c2
I / II = A1 / A2 = a.b.c1 / a.b.c2 A1 / A2 = c1 / c2
Requirement :
1. Max
2. Standard solution
3. Blank solution
Photometric measurement
I. End point reaction
Color appeared = calorimetric measurement
Turbidity appeared = turbidimetric measurement
II. Kinetic reaction
* the measurement is conducted during the reaction occurs
* the absorbance changing in a define time unit, is the
function of a substance concentration in the solution
Calculation
* Beers law : A1 / A2 = C1 / C2
* Transmission - concentration calculation
* Calibration curve
Calibration curve
A 5
4
3
2
1
0
50 100 Cx C = mg / 100 ml
Beers law requirement :
1. Monochromatic light : max
2. Reading at the 20 - 70%T
3. No fluorescent subtance
4. The standard concentration being used is in the range
concentration of linearity with the concentration of the subtance
being subtance being measured.
Serum/Standard Reagensia
Baca Absorbans