Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Figure:
(a) Normal waveform.
(b) Ventricular fibrillation.
(c) Ventricular tachycardia.
Defibrillators
The defibrillator is an electrical device that delivers a pulse of therapeutic
current intended to reverse a ventricular fibrillation (VF) or a life-threatening
ventricular tachycardia (VT) in the heart of a patient.
where
U is the energy in joules (J)
C is the capacitance of C1, in farads (F)
V is the voltage across C1, (V)
Lown Defibrillator Waveform
Example: Calculate the energy stored in a 16-F capacitor that is charged
to a potential of 5000 V dc.
Solution
Monopulse Defibrillator
Waveform
The monopulse waveform shown in Figure is a modified Lawn waveform
and is commonly found in certain portable defibrillator.
It is created by a Lown waveform defibrillator circuit but without inductor L,
which create the negative second pulse.
Consequently, the waveform decays to zero in the exponential manner
expected of an R-C network.
Tapered DC Delay Defibrillator
Waveform
This waveform differs from the others in that it uses a lower amplitude and
longer duration to achieve the energy level.
The energy transferred is proportional to the area under the square of the
curve, so we may attain the same energy as in other waveforms.
The double-humped waveform characteristic of tapered delay machines is
achieved by placing two L-C sections, such as in Lown waveform defibrillator
circuit, in cascade with each other.
Trapezoidal Defibrillator Waveform
The trapezoidal waveform shown in figure is another low-voltage, long
duration shape.
The initial output potential is about 800 V, which drops continuously for
about 20 ms until it reaches 500 V, where it is terminated .
Standard Anterior Paddle
The energy from a defibrillator is delivered through a set of high-voltage
paddle electrodes.
The type shown in Figure is called an anterior paddle.
Standard Anterior Paddle
A thumb switch to control the discharge is mounted at the top of the grip.
A defibrillator paddle and cable set using two of these electrodes is called
an anterior-anterior set.
To defibrillate, one electrode is placed on the chest directly over the heart
while the second electrode is placed on the left side of the patient's chest.
These arrhythmias are also correctable by electrical shack to the heart, but
it is necessary to avoid delivering the shock during the ventricles' refractory
period (the T wave of the ECG waveform).
Sometime the shock intended to correct the problem will create a much
more serious arrhythmia such as ventricular fibrillation.
The shock is usually timed to occur approximately 30 sec after the R wave
peak.
Cardioverter
Human operators cannot be trusted to time the ECG waveform properly to
avoid this problem, so an automatic electronic circuit is used.
Asynchronous (does not synchronize with intrinsic (natural) heart rate (HR)).
Both types of pacemakers sense the presence (or absence) of intrinsic R-wave.
The output of the R-wave inhibited unit is suppressed as long as natural R-
waves are present. Thus its output is held back or inhibited when the heart is able
to pace itself.
On the other hand, R-wave triggered pace emits an impulse with the occurrence
of each sensed R wave.
Pacing Modes
In cases of complete heart block where the atria are able to depolarize but the
impulse fails to depolarize the ventricle, then atrial synchronous pacing may
be used.
Atrial programmed pacers are always synchronized with the P wave of the
ECG.
Here the pulse generator is connected through wires and electrodes to both
the atria and the ventricles.
The atrial electrode couples atrial impulses to the pulse generator which then
emits impulses to stimulate the ventricles via the ventricular electrodes.
In this way the heart is paced at the same rate as the natural pacemaker.
When the SA node rate changes because of sympathetic neuronal control, the
ventricle will change its rate accordingly but not above some maximum rate
(about 125 per minute) .
Questions?