Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
1.INTRODUCTION TO MICROSCOPE.
History of microscope.
What is a microscope?
Why use a microscope?
Interaction of light with matter.
Image formation
Lens aberrations :Chromatic aberration
Spherical aberration
Coma
Astigmatism
Field curvature
Distortion
HISTORY OF MICROSCOPE
BIRTH OF THE LIGHT MICROSCOPE
1590 Zacharias Janssen and his son Hans, while experimenting with several lenses in a tube,
discovered that nearby objects appeared greatly enlarged that was the forerunner of the
compound microscope and of the telescope.
1609, Galileo father of modern physics and astronomy, heard of these early experiments,
worked out the principles of lenses, and made a much better instrument with a focusing device.
4 centuries of light
microscopy
400
years
Compound microscope
ROBERT HOOKE: The English father of microscopy, re-confirmed Anton van Leeuwenhoek's
discoveries of the existence of tiny living organisms in a drop of water. Hooke made a copy of
Leeuwenhoek's light microscope
ERNST ABBE
WHAT IS A MICROSCOPE?
Micro=small, Scope=to view
Microscope is an instrument used to produce an enlarged, well defined image
of the objects too small to be observed with the naked eyes.
It is built based on properties of light. It is most commonly used instrument in
the lab, yet least known about it by its users.
The term microscope strictly speaking includes not only instrument fitted with
separate objectives and eyepiece lenses, but also those composed of a single
lens.
HAND LENS
REPRESENTATION OF LIGHT
RAY SHOWING WAVELENGTH
AND AMPLITUDE
RETARDATION AND REFRACTION:Media through which light is able to pass will slow down or retard the speed of the light in
proportion to density of medium.
Higher the density ,greater the degree of retardation .
Rays of light entering sheet of glass at right angles are retarted in speed but there direction is
unchanged.
If light enters glass at any other angle deviation of direction will occur in addition
to the retardation this is called refraction.
A curve lens exhibits both retardation and refraction ,the extent of which is
governed by :a)Angle of incidence -angle at which light strikes the lens.
b)Refractive index-density of glass.
c)Curvature of lens.
Rays passing through a curved lens exhibit both retardation and refraction
Angle to which rays are deviated within the glass or other transparent medium is called
angle of refraction and ratio of sine values of the angle of incidence (i) and refraction
(r)is refractive index(R.I)
Greater the R.I ,higher the density of the medium
R.I of air 1.00
R.I of water 1.30 .
R.I of glass 1.5.
Refraction of light
Total internal
reflection
TYPES
OF
MICROSCOPE
There are two main types of microscopy:1.Light microscopy(0.2 resolution)
- Visible light Routine microscopy
-Ultraviolet Fluorescence microscopy
2.Electron microscopy(0.2nm resolution)
-Beam of electrons in an electromagnetic field are used to visualize the specimen
TYPES OF MICROSCOPE:The following are the types of microscopes employed :1.Optical or light Microscope
2.Compound Microscope
3.Phase contrast Microscope
4.Dark field /dark ground Microscope
5.Interference Microscope.
7.Polarisation Microscope.
8.Electron Microscope.
TYPES OF MICROSCOPE
IMAGE FORMATION:-
Parallel rays of light entering a simple lens are brought together by refraction to a single point,
the principal focus or focal point, where a clear image of an object will be formed. The
distance between the optical center of the lens and the principal focus is the focal length.
In addition to the principal focus ,the lens also has other pairs of points ,on either side of the
lens called conjugate foci such that the object placed at one will form a clear image on a screen
placed at the other. The conjugate foci vary in the position ,and as the object is moved nearer to
the lens the image will be formed further away ,at a greater magnification and inverted. This is
the real image and is formed by the objective lens of the microscope.
If the object is placed yet nearer the lens within the principal focus the image is formed on
the same side of the object and is enlarged the right way up and cannot be projected on to
the screen. This is the virtual image and is that formed by the eye piece of the microscope
level of the real image projected from the objective.This appears to be at a distance of
approximately 25cm from the eye around the object stage level.
LENS:-Lens is the name given to the piece of glass or other transparent material
,usually circular having the 2 surfaces ground and polished in a specific form in order that
rays of light passing through it shall either converge or diverge.
A lens is called positive when it causes light rays to concentrate or converge to form a
real image.
It is called negative ,in which case light rays passing through will diverge or scatter
and positive or real image will not be seen.
These two types can easily be differentiated since positive lenses are thicker at the
centre than at the periphery, whereas negative lenses are thicker at centre .
TYPES OF LENS
POSITIVE LENS
NEGATIVE LENS
Plano-concave lens:
Plano-concave glass lenses are manufactured with one concave
surface and one plane surface. These lenses have negative focal
lengths.
These lenses diverge collimated incident light.
They form virtual images which are seen through the lens.
They are often used to expand light or to increase focal lengths in
existing systems.
Bi-convex lens:
Bi-convex lenses are manufactured with an identical convex surface on both sides
of the lens.
Bi-convex lenses have positive focal lengths and form both real and virtual
images.
Aberrations such as coma, distortion, and chromatic aberration almost exactly
cancel out at unit conjugate ratios due to the symmetry.
Recommended for virtual imaging of real objects, and for positive conjugate
ratios from approximately 0.2 to 5. Values are wavelength sensitive.
Antireflection coatings can be ordered as an option.
LENS ABERRATIONS:These are the defects which impair the ability of a lens to produce an image
which is an exact copy of an object.
There are six basic types of aberrations:1) Spherical aberrations
2) Chromatic aberration
3) Coma
4) Astigmatism
5) Curvature of field
6) Distortion
In the objective lenses used in the modern microscopes only spherical aberration is
significant, since it is often introduced into the image by the incorrect use of the or by using
unsatisfactory preparation.
CHROMATIC ABERRATION:-When light is spilt into its component parts ,each part
vibrates to a different degree producing different colour to the eye. These
colours(vibgyor)are known as primary spectrum and are seen in the rainbow, or through a
spectroscope.
Red has the longest wavelength with vibration of 0.7,blue 0.45,and violet 0.35.It will
be seen that vibration of red is twice the length of those of violet. (results in unsharp image
and colored fringes)
Since light rays cannot vibrate as easily in a dense medium as in a
rare medium, It follows that various colours will be affected by lens to different degrees the
colours with shorter wavelength such as blue,violet being affected to a greater degree
than those having longer wavelength such as red and orange.
A positive lens is combined with a negative lens made of glass producing a greater chromatic
aberration,but with a same refractive index.
Negative lens corrects the chromatic aberration in the positive lens and only partially
neutralises its magnifying power.This type of lens is known as an achromatic lens.
If flourspar is incoporated in the glass of the achromatic lens, three colours can
be bought to a single focal point,and amount of chromatic aberration visible in
the image will be negative.such lenses are known as apochromatic lens.
This fault could be minimised by using only central area of the lens but since microscope objectiv
must have a shorter working distance and high magnification,
a large angle of light is required from each point of object. Hence it can be corrected by using
powerful positive lens and partially neutralising its magnifying power
with a negative lens made of glass having a greater relative aberration.
COMA: This is an aberration associated with points which lie off the axis and
which as the consequence of the coma are imaged as a conical or comet shaped blur.
Coma is the most objectable aberration but is not found in todays objective lenses.
A lens corrected for the coma has equal magnification in its different zones for points in
the field which are off axis.
When a lens is corrected for coma throughout ,say about two-third of its diameter and is
nearly free of coma for reminder ,and in addition is free of spherical aberration ,then it is said to
be aplanatic.
ASTIGMATISM:-This aberration often affects the oblique rays passing through the
lens. The effect is that the image of an off axis point is not reproduced as such but its
appearance differs according to the focal point chosen.
FIELD CURVATURE:- Here instead of image lying parallel with that of the lens
it falls upon a surface which has a spherical curvature. This defect if present will
be troublesome with high aperture objectives of short focal length, when only
either the center of the field of the view or peripheral areas will be in sharp focus.
Curvature of the field may be partially corrected for the medium power
objectives by use of a special eye piece.
Simple microscope
Invented by leeuwenhoek.
It consists of a single lens or a
magnifying glass.
PRINCIPAL:-a lens of short
focal length is used to produce
an enlarged image of an
illuminated object at a short
distance. the lens fixed in a
frame is adjustable to view the
object. shorter the focal length
larger is the image.
Source of power is sunlight.
Compound
microscope.
Invented by hooke.
It consists of two or more
lenses.
PRINCIPAL:- if a short focal
length is used to produce an
enlarged image of an
illuminated object at a short
distance, then another lens can
be so fixed that it would
produce a further enlargement
of that image.
Source of power is electricity.
COMPONENT PARTS OF A
MICROSCOPE
A.
(i)
(ii)
(iii)
SUPPORT SYSTEM
Base
Pillars
Handle/Limb
B. FOCUSSING SYSTEM
(i) Course adjustment screw
(ii) Fine adjustment screw
C. THE STAGE
(i) Fixed Stage
(ii) Mechanical stage
D. OPTICAL SYSTEM
(i) Body Tube
(ii) Nose piecea.Fixed
b. Revolving
(iii) Objective lenses
(iv) Eye piece
E. ILLUMINATION SYSTEM
(i) Source of light
(ii) The mirror
(iii) The condenser
TYPES OF MICROSCOPE
1. SIMPLE MICROSCOPE [LEEUWENHOEK] [250X]
>Hand held magnifiers[5x-20x]
>Supported magnifiers{watch makers eye glass, dissecting
microscope, binocular headband
magnifier , spectacle supported lenses}
>micro film readers{for large histo specimens }
COMPOUND MICROSCOPE
Two types:monocular
binocular
COMPONENTS OF MICROSCOPE
[MONOCULAR]
A.SUPPORT SYSTEM
Pillars
.
.
.
.
.
.
.
C. THE STAGE
1.Fixed stage
The stage is provided with simple metal springs to hold the object
or is fitted with the mechanical stage
2. Mechanical stage
It is a calibrated metal frame fitted on the right edge of the fixed stage
Two screw heads move the slide from side to side and forwards and
backwards
The standard mechanical stage takes a 3x1inch slide and moves over
an area of ~3 x 1 1/4 inches so that the whole slide may be
examined
Special stages are available to take very large slide and petridishes
D.OPTICAL SYSTEM
2. Nose piece
It is fitted at the lower end of the body tube.
It rotates on a central pillar and is designated by the number of
objectives it carries , for example , double,triple or quadruple
nosepiece.
It should bring each objective into center of the optical axis and at
correct tube length which is indicated by a click.
It is 18mm in depth.
Magnification can be increased by rotating the nosepiece.
3.Objective lenses
It is attached to the lower end of the body tube.
Three spring loaded objectives of varying magnifying powers are
commonly used.
Within the objective there may be lenses and elements from 5 15 in
number ,depending on image ratio,type and quality.
The main task of the objective is to collect maximum amount of light
possible from the object, unite it and form a high quality magnified
real image, some distance above.
Object to image ratios of objectives are from 1:1 to 1:100 in normal biological
instruments.
NUMERICAL APERTURE:-[NA]
The ability of an objective to resolve detail is indicated by its NA and not by its
magnifying power.
It allows the cone of light that can be collected from the object.
It is expressed as the product of two factors by a formula
NA= n x sin u
Where
n refractive index of the medium between the coverglass over the object
and front lens of the objective. For example , air, water or immersion oil.
u angle between the optical axis of the lens and the outermost ray which
can enter the front lens.
Greatest angle will be formed when the surface of front lens coincides with the
specimen ie u= 90 degrees.
Dry objective NA= 0.95
Oil immersion NA= 1.5
Air R.I =1.0
Water R.I = 1.3
EFFECTS OF HIGH NA:(i) it reduces the depth of focus ie the ability to focus on more than one
an object at same time.
(ii) it reduces the flatness of the field , so that the edges are out of
layer of
focus.
RESOLUTION
It is the smallest distance between two dots or lines that can be seen as
separate entities.
The angle formed by the images of these two points or dots at the eye
is called as the VIEWING ANGLE.
RESOLVING POWER
TOTAL MAGNIFICATION
----------------------------- x
brightness = (NA)(NA)
-------------------------------------(magnification) (magnification)
TYPES OF OBJECTIVES
1.
2.
3.
4.
5.
6.
COVERGLASS THICKNESS
is not neccesary for oil immersion lenses because of same R.I.
For dry lenses No.1 coverglass is used[0.12-0.22mm]
And then the collar setting is done to change the contrast.
PARFOCAL SYSTEM
Switching from LP to HP requires only a little turn of fine adjustment to get a sharp image.
4. Eye piece
Function is to magnify the image formed by the objective within the body
tube and present the eye with a virtual image apparently in the plane of the
object being observed, usually this is an optical distance of 250mm from
the eye.
Field lens collects the divergent rays of the primary image and passes these
to eye lens which further magnifies the image
Pointer eyepiece has a small pin mounted in it which is used to point out a
particular cell or object in the field.
TYPES OF EYEPIECES
1.
2.
3.
4.
5.
Wide field eyepieces has large field of view for use in biological lab.
6.
7.
Periplan eyepiece corrected the chromatic abberation and gave a flat field
of view. It has six lens elements.
8.
ILLUMINATION SYSTEM
1.
.
.
.
.
.
.
.
.
CONDENSER.
It is a system of lenses mounted in a short cylinder and lies beneath the stage
hence comprises of the substage.
It can be raised or lowered by a rack and pinion arrangement.
It concentrates light onto the object with uniform intensity over the whole
field.
It controls the aperture of the illuminating cone of light and matches it to the
NA of the objective.
TYPES.
Low power NA= 0.25
Medium power dry - NA=0.95
High power oil immersion-NA=1.4
OTHER TYES.
Abbe three lens or aplanatic- corrected for spherical abberation but not for
colour.
NA=0.65
when used with oil NA=1.2
It is commonly used but expensive
IRIS DIAPHRAGM
It is employed to limit the angle of cone of light passing through the object so
that it will just fill the front of the objective.
The size and volume of the cone of light is altered by this diaphragm.
If closed too much image becomes too contrasting and if it is wide open image
will suffer GLARE.
FILTERS
FILTER CARRIERS
MIRROR.
SOURCE OF LIGHT.
All the light that reaches the eye should come from the specimen under study.
Light from any other part of the slide will tend to obscure the details, this is
called GLARE.
So only a small area of the slide is illuminated.
Daylight may be diffuse, reflected and scattered by atmosphere.
TYPES OF ILLUMINATION:-
Kohler illumination.
TRANSMITTED ILLUMINATION
BINOCULAR MICROSCOPE.
Eyepiecies can be easily moved together or apart and the inter pupillary
distance can be adjusted to suit individual requirements.
ADVANTAGES
Minimum amount of eye fatigue.
TECHNIQUE
The lamp should be positioned opposite the microscope and a blue day light
filter inserted to absorb the excess yellow light given by artificial light.
Position the lamp so that light strikes the center of the mirror and adjust mirror
so that light is directed upward into the condenser .modern microscopes have
in-built lamps, condensing lens and mirror.
Focus on an object on the stage and ensure that the field is evenly illuminated.
With the object in focus , rack the condenser up and down until a sharp image
of lamp iris appears.
Center the image of field iris using substage centering controls.
Open the field iris until its circle of light is just larger than the
field of view, this reduces the GLARE.
Remove an eyepiece and adjust the iris until twothird of the
back focal plane of the objective is illuminated. Replace the
eyepiece. The microscope is ready to use.
MICROMETRY:
Assuming that 100 eyepiece divisions were equal to 10 small stage divisions
and object had covered 12 eyepiece division
100 stage divisions = 1mm = 1000 um
100 eyepiece division = 10 stage division
therefore 100 eyepiece division = 100 um
1 eyepiece division = 1um
Direct light is not allowed to take part in the formation of the image, this is done by
either blocking the zero-order light in back focal plane of the objective or more
usually by illuminating the specimen with a hollow cone of light of such obliquity
that all the light falls outside the acceptable angle of objective.
A central OPAQUE PATCH STOP of the correct size is fitted in the front focal plane of
the condenser , this ensures that all the light falling on the object is too oblique to enter the
objective.
Light diffracted by the specimen, emerges in a cone over 180 degrees , some of this will enter
the objective and form an image in which features scattering light will appear bright.
For high NA objectives, reflecting condensers are used instead , oiled to the slide.
Even with such condensers it may not be able to obtain an adequate darkgrounds when a
high-aperture, oil-immersion objective is used .This problem may be solved by restricting the
NA by a small iris diaphragm built into its back focal plane.
Darkgrounds reflecting condensers need care in their centration and focusing so that the
specimen is located exactly at the focal point of rays.
Objects examined by dark ground illumination give a misleading impression of size ; fine
particle appear to be much larger than they are , owing to their light scattering properties
After use,the oil should be carefully cleaned off both the condenser and objective
1.
2.
3.
4.
PHASE
CONTRAST
MICROSCOPE
Phase is only useful on specimens that are colorless and transparent and usually
difficult to distinguish from their surroundings. We call these specimens "phase
objects".
Examples of phase objects include cell parts in protozoans, bacteria, sperm tails
and other types of unstained cells. This phase contrast technique proved to be such
an advancement in microscopy that Zernike was awarded the Nobel prize (physics) in
1953
Two rays of light from the same source having the same frequency , are said to be
COHORENT , and when recombined will double in amplitude or brightness.
If they are in phase with each other
INTERFERENCE
it is called CONSTRUCTIVE
If however they are out of the phase with each other , DESTRUCTIVE
INTERFERENCE takes place .
The following figure represents the wave form of a light ray.
TECHNICAL DETAILS :To achieve phase contrast the microscope requires modified objective and
condenser and relies on the specimen retarding the light by between 1/8 th and
1/4th lambda.
An intense light source is required to be set up this microscope.
The microscope condenser usually carries a series of annular diaphragms made
of opaque glass, with a clear narrow ring to produce a controlled hollow cone
of light.
Each objective requires different size of annulus , an image of which is formed by
the condenser in the BFP of the objective as a bright ring of light.
THE POLARIZING
MICROSCOPE
produced by birefringent
object. these devices are
placed in the stage.
NATURAL LIGHT
POLARISATION OF LIGHT
A light beam normally contains waves of
Nicol devised a prism from which light rays, having passed through ,
would emerge vibrating in a single plane , that is , as a POLARIZED
LIGHT.
The single direction in which the light is vibrating when it emerges is
known as OPTICAL PATH of the prism.
HOW IS THE PRISM MADE ?
2) BIREFRINGENCE :If a dot is drawn on the sheet of the white card is viewed through a block of glass
laid on top , only one dot will be visible from above.
If the block of glass is replaced by a polished block of crystal ,two dot will be
visible .such a crystal is described as being BI-REFRINGENT or ANISOTROPHIC.
IT has split light ray from the dot into two rays which emerge from the crystal at
different points.
TYPES OF BIREFRINGENCE :Certain crystals or tissue structure show more than one index of
refraction for a given wavelength of light and are therefore said to be
doubly refracting or birefringent.
This is due to some sort of asymmetrically oriented spatial
arrangement of particles .
These particles carry resonating chargers capable of interacting with
the oscillations of light wave .
Birefringent material may show one or more than one type of such
arrangement , the more common types may be summarized as
follows :1.Intrinsic or crystalline birefringence.
2.Form birefrigence.
3.Strain birefringence.
4.Positive and negative birefringence.
POSITIVE AND NEGATIVE BIREFRINGENCE:In a polarizer , slow and fast rays are separated by
birefringence material .
If the slow ray (high R.I ) is parallel to the length of the
crystal or fiber, the birefringence is positive.
If slow ray is perpendicular to the long axis of the structure ,
the birefringence is negative.
In a collagen fiber the slow axis of transmission is parallel to
the long axis of the fiber , the fiber is thus said to show
positive birefringence with respect to its long axis .
A chromosome shows negative birefringence with respect to
its long axis.
THE USE OF POLARIZING MICROSCOPE :It has been shown that certain crystal have power to convert a
single ray into two rays of light , which are vibrating in a single
plane at right angles to each other .
If such a crystal is placed on the stage of the microscope having a
nicol prism in the substage , the effect will be seen as below.
The light ray will be vibrating in the optical path of the nicol prism
(AB) ,except those that pass through the crystal ,which will be
vibrating from C to D or E to F.
PARTS OF POLARIZING MICROSCOPE :Compared to a typical microscope, a polarizing microscope has a new
construction with the following added units:
1.A polarizing condenser that includes a polarizer,
2.A rotating stage that allows the positioning of the specimen .
3.A strain-free objective for polarized light,
4.a centerable revolving nosepiece ,
5.An analyzer,
6.A Bertrand lens for observing the pupil of the objective,
7.A test plate,
8.A compensator, and
9.A eyepiece with crosshair.
Polarizing condenser
A polarizing condenser has the following two
characteristics:
1)Built-in rotatable polarizer,
2) Strainfree optical system, like the objectives.
Bertrand lens
A Bertrand lens projects an interference image
of the specimen, formed in the objective pupil,
onto the objective image position (back focal
length). It is located between the analyzer and
eyepiece for easy in and out of the light path.
THE
INTERFERENCE
MICROSCOPE
The two rays which eventually combine to produce a final image are formed by a
plate of birefringent material at the face of objective, each point in the final image
is a compound made up of two superimposed mutually different views of the same
point of the object .
By using a polarized light and an analyser the phase relationship between two
rays can be adjusted and measured.
Whenever light passes across the edge of an opaque object the rays close to the edge
are diffracted , or bent away from the normal path.
If instead of a single edge, the rays pass through a narrow slit , then the rays the
edge of the beam will fan out on either side to quit wide angles.
Two slits closely side by side form two fan of rays which will cross and if cohorent
,will observably interfere.
Types of interference
microscopes
There are two types currently used in great Britan :1) The BAKERS interference microscope
2) DYSON interference microscope.
If the two paths are equal and in same phase, the interference bands can be seen
running straight and parallel across the field.
If into one beam path an object is introduced that causes some shift in the
phase , this will be seen as a displacement of the interference bands .
When using monochromatic light , each interval consisting of one dark and one
light band is one wavelength wide, and thus distance in nanometer is known.
Displacement of the bands is measured in micrometer eyepiece and this
information , coupled either with R.I or object thickness , the measurement
referred to earlier can be determined.
Two types of double beam system have been used .
1) Focusing the reference below the object the double focus system
2) Involved a lateral displacement of the reference beam called SHEARING
where the separation of the beam is very small .
USES :As an infinitely variable phase contrast microscope with which individual parts
of the cells may be studied with great detail; for this purpose little is to be
gained over a good phase contrast microscope.
As a highly accurate optical balance , it may be used for estimation of dry mass
down to 110-14g.
The discovery that an increase in R.I of 0.0018 is a result of 1 percent increase
in the solid substances contained in the cell , and the fact that the refractive
index of the cell components can be estimated from the phase differences
between them and the reference area (usually the fluid in which they are
suspended ) has made this possible. It is this aspect of interference microscope
that is at present being developed by many research workers.
STEREOMICROSCOPE
When one looks at any object , the image recorded by the optical system , in
conjunction with the co-ordinating facility of the brain, has the THREE
DIMENSIONAL EFFECT.
It is designed specially for LOW POWER works due to long working distance.
TYPES OF STEREOMICROSCOPE
IMAGE FORMATION
DISADVANTAGE
Because of the inclination of the tubes , there is a
tendency for the image to be slightly out of focus
towards the edge on each side of the centre line of the
field of view.
IMAGE FORMATION
For transmitted light photomicrography integral low wattage bulbs are used.
For transmitted light direct visual observation a large source like an OPAL
MAINS LAMP , below a pair of photographic enlarger condensers are sufficient.
For reflected light works ordinary FOCUSABLE TUNGSTEN LAMPS are
sufficient.
The lamps are adjusted at varied angles and distances so as not to impede access to
the specimen.
FIBRE-OPTIC LAMPS are most versatile for lowest magnifications it is easy to
adjust and conveys little heat to specimens.
Filters of different colors can be used with fibre-optic lamps to get a 3-D detail.
Magnification is 5- 50 times as compared to compound microscopes with 50- 1200
times magnification.
USES:
MACROSCOPE
It has a large stage and long working distance for the objectives.
It has a zoom lens sometimes fitted, to allow a magnification of range 4X to 80X
{MAKROZOOM LEITZ LENS}
A macroscope should allow the use of both oblique and coaxial incident light and
the base should have an aperture to allow transmitted light to be used.
Fluorescence Microscopy
INTRODUCTION
A fluorescence microscope (epifluorescent
microscope) is a light microscope used to
study properties of organic or inorganic
substances using the phenomena of
fluorescence and phosphorescence instead
of, or in addition to, reflection and
absorption.
Primary or Autofluorescence
It is the ability of some naturally occuring
compounds to fluoresce.
A substance which possesses a fluorophore
will fluoresce naturally. This is known as
primary fluorescence or autofluorescence.
Ultraviolet excitation is required for
optimum results with substances such as
vitamin A, porphyrins and chlorophyll.
Secondary Fluorescence
Demonstration of tissue components stained by
fluorescent dyes.
Dyes, chemicals and antibiotics added to tissues
produce secondary fluorescence, of structures and
are called fluorochromes.
This is the most common use of fluorescence
microscopy and the majority of fluorochromes
require only blue light excitation.
Induced Fluorescence:
Is the term applied to substances such as
catecholamines, which after treatment with
formaldehyde vapor are converted to
fluorescent quinoline compounds.
Pretreatment of tissues causes a reaction
with component under investigation,
producing a reaction which is fluorescent.
Principle of Fluorescence
When a quantum of
light is absorbed by
an atom or
molecule, an
electron is boosted
to a higher energy
level.
When this displaced
electron returns to
its original ground
state, it may emit a
quantum of light.
Principle of Fluorescence
1.
2.
3.
Energy is absorbed by
the atom which becomes
excited.
The electron jumps to a
higher energy level.
Soon, the electron drops
back to the ground state,
emitting a photon (or a
packet of light) - the
atom is fluorescing.
Thus an object is
illuminated with a
black light and,
when fluorescent,
appears as a bright
object against a
dark background.
COMPONENTS
Illuminating system
Filter system
Exciter filter
Contrast or barrier filter
Dichromatic filter attachment
Condenser
Objective
TYPES
TRANSMITTED LIGHT
FLUORESCENCE.
INCIDENT LIGHT FLUORESCENCE.
DIFFERENCES
TRANSMITTED
There is no use of
dichromatic mirror.
INCIDENT
There is use of
dichromatic mirror.
The fluorescence is
produced towards the
light source, and
viewed on the opposite
side.
Fluorescence is
produced on the
observers side, and
therefore more brilliant.
DIFFERENCES
TRANSMITTED
INCIDENT
APPLICATIONS:
INDUCED FLUORESCENCE: pretreatment of
tissues causes a reaction with component under
investigation, producing a reaction which is
fluorescent.
Eg: 5-hydroxytryptamine fluorescence induced by
formalin fixed tissue.
SECONDARY FLUORESCENCE: demonstration
of tissue components stained by fluorescent dyes.
CONFOCAL
MICROSCOPY
PRINCIPLE OF CONFOCAL
MICROSCOPY
TYPES
Three types of confocal microscopes are
commercially available:
Confocal laser scanning microscopes
Spinning-disk (Nipkow disk) confocal
microscopes
Programmable Array Microscopes (PAM)
DISADVANTAGE
However as much of the light from sample
fluorescence is blocked at the pinhole this
increased resolution is at the cost of
decreased signal intensity so long
exposures are often required.
USES
This technique has gained popularity in the
scientific and industrial communities.
Its typical applications are in life sciences
and semiconductor inspection.
ELECTRON MICROSCOPE
INTRODUCTION
The EM has gained its fundamental superiority
over light microscope because of its high
resolving power, and thereby producing extreme
fine details.
In light microscope the highest resolution
possible theoretically is half the wavelength of
light. (thus the limit of resolution of light
microscope is 0.2 microns = 2000A0)
HISTORY
o The first electron microscope prototype was built
in 1931 by Max knoll and Ernst Ruska.
o This initial instrument was capable of magnifying
objects by only four hundred times, although it
demonstrated the principles of an electron
microscope.
o Two years later, Ruska constructed an electron
microscope that exceeded the resolution possible
with an optical microscope.
PRINCIPLE
An electron microscope is a type of
microscope that uses a particle beam of
electrons to illuminate a specimen and
create a highly-magnified image.
The electron microscope is constructed
based on the same optical principle as
the light microscope.
Convergence of a light beam by a
convex glass lens has its counterpart
in the convergence of an electron beam
as it passes through the core of a
circular magnetic field.
R = 0.61 lambda
NA
Where: R, the RESOLUTION = represents the capacity
of the optical system to produce separate images of
objects very close together.
Lambda is the wavelength of the incident illumination.
NA is the numerical aperture of the lens.
Thus for any given lens, resolution is directly related to
the wavelength of the source radiation.
Parts of an EM
The electron beam is obtained from
a heated tungsten filament(cathode)
which is surrounded by a metal
cylinder known as the whenalt cap.
This cap serves to shape the
electron beam.
Just beyond the whenalt cap is the
anode which has an aperture
through which the electron beam
passes.
A large voltage is passed between
the cathode and anode, which gives
the electron their high velocity.
They pass through the rest of the
microscope without any
acceleration.
ELECTRON GUN:
the device responsible
for generating the
beam of electron is
the electron gun.
The most important
components of the
gun are tungsten
filament, wehnelt
sheild and anode.
4. ELECTRON LENS:
The lenses used to refract electrons are electromagnetic
coils or solenoids.
When energized these generate a magnetic field that
forces electrons into a spiral trajectory and towards a focal
point on the opposite side of the lens.
All lenses of TEM operate in the same way the
focussing effect of each lens can be changed simply by
varying the power and hence the strength of the magnetic
field.
In practice, electronic lenses are difficult to manufacture
and typically display spherical, and chromatic aberrations
as well as astigmatism.
CONDENSER:
The electron beam first passes through the condenser
lens as in light microscope.
This lens serves to focus the beam on the object, and so
provide illumination.
The magnetic lenses of an EM can have different
powers, depending on the amount of current flowing in
the electric coils. (In light microscope, the lens power is
fixed, but the lens can be moved so that the image can be
focused.
In EM all lenses are fixed rigidly but their focal points
are variable by adjusting the lens current.
Thus the illumination of the object is achieved by
varying the current to the condenser lens.
IMAGING SYSTEM:
The imaging system consists of three
lenses;
* The objective
* the intermediate
* the projector lenses
This gives three stages of
magnification and makes it possible
to achieve high magnification in a
reasonable amount of space.
The objective lens is placed with its
focal point close to the object.
Intermediate images are formed
between each lens.
The projector lens throws its image
on to a fluorescent screen, which may
be substituted by a photographic plate
to make a permanent record
Types of EM
Transmission EM
The original form of electron
microscope, the
transmission electron microscope (TEM)
uses a high voltage electron beam to
create an image.
When the electron beam emerges from
the specimen, the electron beam
carries information about the structure
of the specimen that is magnified by
the objective lens system of the
Scanning EM
Unlike the TEM, where electrons of the high
voltage beam carry the image of the specimen,
the electron beam of the
Scanning Electron Microscope does not at any
time carry a complete image of the specimen.
The SEM produces images by probing the
specimen with a focused electron beam that is
scanned across a rectangular area of the specimen
(raster scanning).
SEM
Electron beam scans over
surface of sample.
Sample can be of any
thickness and is mounted
on an aluminum stub.
Image shown on
fluorescent screen.
Image is a two
dimensional projection of
the sample.
RESOLUTION OF MICROSCOPES
MICROSCOPE
RESOLUTION
MAGNIFICATIO
N
1000X
OPTICAL
200nm
TEM
0.2nm
500 000X
SEM
2nm
200 000X
REFLECTING EM
In the Reflection Electron Microscope (REM) as in
the TEM, an electron beam is incident on a surface,
but instead of using the transmission (TEM) or
secondary electrons (SEM), the reflected beam of
elastically scattered electrons is detected.
Scanning transmission EM
o The STEM rasters a focused incident probe across a specimen
that (as with the TEM) has been thinned to facilitate detection of
electrons scattered through the specimen.
o The high resolution of the TEM is thus possible in STEM. The
focusing action (and aberrations) occur before the electrons hit
the specimen in the STEM, but afterward in the TEM.
o The STEM's use of SEM-like beam rastering simplifies
annular dark-field imaging, and other analytical techniques, but
also means that image data is acquired in serial rather than in
parallel fashion.
USES
o Transmission electron microscopy has been widely used in routine
histopathology in recent years particularly in the field of renal
disease, tumor pathology and viral infections.
o The electron microscope is an essential item of equipment in many
laboratories.
o Researchers use them to examine biological materials (such as
microorganisms and cells), a variety of large molecules, medical
biopsy samples, metals and crystalline structures and the
characteristics of various surfaces.
o The electron microscope is also used extensively for inspection,
quality assurance and failure analysis applications in industry,
including, in particular, semiconductor device fabrication .
ELECTRON
MICROSCOPE
PRINCIPLE
Formation of an image is
due to scattering of
electrons by molecules of
the specimen and
scattering depends
solely on mass densities.
Elements of high
molecular weight such as
lead and uranium cause
marked electron scatterused for back ground
staining
Depends on wavelength
RESOLUTIO of moving electrons ;
N
resolution of 2 A0 can be
obtained.
VOLTAGE
KV
SOURCE
ELECTRONS
LIGHT MICROSCOPE
Image formation is due to
absorption of light which
depends more on
molecular structure than
on atomic weights C, n2
and H atoms- low
molecular wt .
Histological stains
depends more on certain
wavelenght of light, due to
molecular structure.
Limited to about 200 nm
LIGHT
Microsc
ope
image
formatio
n
Convergence of electron
beam is by passing
through circular
magnetic field - EM
lenses.
Beam is obtained from
heated tungsten
filament.
Lenses used are
condenser, intermediate
and projector lenses.
Convergence of a light
beam is by a convex
glass system
Light rays - are
obtained by a lamp
source/ natural light.
Lenses - condenser,
objective and eyepeice.
System
Column is rigidly
constructed and
maintained at high
vacuum since air
molecules would
deflect electron beam.
Specimen must be
placed inside
microscope living
materials cant be
examined.
Illuminating condenser
unit are placed below
the object and lens are
Constructed upside
placed above and
down is electron gun
image can be
and condenser lens are visualized.
placed above and
image formed below.
Unstained sections are
visualized and contrast
is increased by
staining with lead /
Tissue
processing
SPECIMEN
HANDLING
Fixatives are
buffered within ph
range of 7.2 7.6
and osmolarity,
ionic composition
is maintained.
Washing after
fixation - tissue is
rinsed in buffer
after post fixation
(immersed in 2%
aqueous
uranylacetate
enbloc staining)
DEHYDRATION
ETHANOL, ACETONE
are used.
Acetone is
inflammable, hence
avoided.
Propylene oxide is
used with ethanol to
ensure complete
To maintain the
morphology of
tissues and for
stabilizing proteins
it is fixed usually in
fixative (10%
formalin routinely
used )
Throughly washed
after fixation in
running water.
Increasing
concentration of any
organic solvent like
ethanol , acetone is
used
Not as sensitive as in
EM.
CLEARING
AND
EMBEDDING
MICROTOMY
KNIVES AND
SECTION
THICKNESS
Resin Embedding
for ultra thin
sections . Rigid
material
necessary,
withstand
vacuum and heat
generated,
preserve tissue
structure.
Epoxy resin or
acrylic resin is
used
ULTRAMICROTOMY
Glass knives or
diamond knives
used.
0.5 micron
2 step procedure
Prepare and examining semi
thin sections 1 micro m.
Trough fluid is used for
collection of sections(10 to
15% ethanol or acetone is
used.
Sections are collected onto
grids using an eye lash probe
or fine point forceps.
For drying grids should be
placed on filter paper in lidded
container.
Standard method for staining
sections.
Grids are floated .
Section is placed side down
and drops of staining solution
is added for required time.
It is stained with uranyl
acetate (2% aqueous) or
reynolds lead citrate stain.
1 step procedure
Placed in water
bath, camel brush is
used to fix it on
slightly along with
egg albumin and
thymol crystal
adhesive.
To fix the section it
is placed on slide
warmer and heated
at 600 C for 10 mins
It is then stained
routinely for various
histological stains.
Glass slides are
placed on stage and
viewed.
REFERENCES
BRADBURY and B.BRACEGIRDLEINTRODUCTION TO LIGHT
MICROSCOPY.
BANCROFT,HISTOLOGICAL
TECHNIQUES,5th edition.
ROYAL MICROSCOPICAL SOCIETY.
INTERNET SOURCES
THANK
YOU