Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
CHAPTER 1 CHAPTER 2 CHAPTER 3 CHAPTER 4 CHAPTER 5 CHAPTER 6 CHAPTER 7 CHAPTER 8 CHAPTER 9 CHAPTER 10
INTRODUCTION CENTRIFUGATION CHROMATOGRAPHY ELECTROCHEMISTRY MICROSCOPE SPECTROMETRY ELECTROPHORESIS HEMATOLOGY ANALYSIS IMMUNOCHEMISTRY ANALYSIS FORENSIC ANALYSIS
CHAPTER 1 : INTRODUCTION BIOMEDICAL INSTRUMENTATION Devices that can be used to make measurements of biologic or medical quantities and give quantitative results
Minimum disturbance to physiologic Sensor must be at physiologic variable value Maintain simplicity
IMPORTANT INSTRUMENTATION TERMS SENSITIVITY - Change in output as a function of a change in input STABILITY Consistency in output for a constant input SPECIFICITY Ability to distinguish desired variable from other competing variables ACCURACY the closeness of the instrument output to the true value of the measured quantity
PRECISION The smallest change in a variables that can be correctly measured RESOLUTION The degree of fineness with which a measurement can be made
REPRODUCIBILITY Same output for the same input
TRANSFER CHARACTERISTIC
SENSITIVITY- change in output as a function of a change in input
Instrument Output
Variable Measured
Variable
Time
Instrument Output
Variable Measured
ACCURACY the closeness of the instrument output to the true value of the measured quantity Instrument Output
Variable Measured
ACCURACY
The accuracy of the instruments can be measured by using this calculation:
Percentage of true value = measured value true value x 100 true value
ACCURACY vs PRECISION
POSSIBLE TYPES OF TRANSDUCERS To convert the input signal to a form which can be conveniently processed Mechanical Electrical Mechanical Thermal Acoustic Hydraulic
TYPES OF SENSOR
PHYSICAL -THERMOMETER
CLARK ELECTRODE
ELECTRICAL CURRENT
BODY
THERMISTOR
SENSOR
DISPLACEMENT SENSOR
SERUM GLUCOSE
HYDROGEN PEROXIDE
H2O2 SENSOR
ELECTRICAL SIGNAL
STANDARDS OF MEASUREMENTS
-Defined as the physical representation of the unit of
measurement (S.I. unit) -For example : kilogram was originally defined as the mass of a cubic decimetre of water at its temperature of maximum density at 4 C
CALIBRATION
-Is the act or result of quantitative comparison between
a known standard and the output of the measuring system measuring the same quantity
-Calibration curve is a references standards with
ERROR
-Defined as the differences between the measured and the
true value
ERROR
Random error - arises from our ability to make physical measurement - always present and cannot be corrected
Systematic/cumulative Error
Instrument Error
Zero error
Instrument error
-Poor design of instrument -Can be avoided by selecting a suitable instrument for a given application and calibrating instrument using a suitable standard
Environmental error
-The condition of environment such as windy, rainy, hot . -Can be avoided by doing the measurement in close room.
CHAPTER 2 : CENTRIFUGATION
CENTRIFUGATION -Device for separating particles from a solution according to their size, shape, density,viscosity of the medium and rotor speed -Centrifuge have been used since middle 1800 -Hand-driven (only after 1912, it is electrically driven) THEORY OF CENTRIFUGATION -Sedimentation of a molecule is influenced by : a) properties of the molecule (size,shape,density) b) properties of the solvent (density, temperature) c) interactions between solute molecules & solvent material
FORCES IN CENTRIFUGATION As the rotor spins in a centrifuge, centrifugal force is applied to each molecule in the sample : CF = Mw2r
M = mass (particle weight/molecular weight) w = angular velocity r = distance from the axis of rotation
-The larger the molecule or the faster the centrifugation or the longer the axis of rotation, the greater the centrifugal force and the rate of sedimentation
-Two forces act to counteract the centrifugal force are bouyant force and frictional force
-A sedimenting molecule moves faster and faster in a
TYPES OF CENTRIFUGES
Desktop clinical centrifuges -Simplest & least expensive -Max speed of most desktop centrifuges is below 3000 rpm -Most operate at all temperature -Typical application : rapid sedimentation of blood samples Highspeed centrifuges -Operating up to speeds of 20,000 to 25,000 rpm -Typical application : collect larger organelles
the major
-4 principles : a)Drive and speed control : flexible drive shaft system, an overspeed system b)Temperature control : infrared temp. sensor c)Vacuum system d)Rotor : swinging bucket rotor & fixed angle rotor
TYPES OF CENTRIFUGATION
-Analytical -Preparative
ANALYTICAL -measuring physical properties of sedimenting particle such as molecular weight PREPARATIVE -To isolate specific particles which can be reused -Three types : rate zonal, differential, isopycnic
Analytical Uses small sample (less than 1 ml) Built in optical system to analyze progress of molecules during centrifugation Uses relatively pure sample Used to precisely determine sedimentation coefficient and MW of molecules
Preparative Larger sample size can be used No optical read out collect fractions and analyze them after the run Less pure sample can be used Can be used to estimate sedimentation coefficient and MW Generally used to separate organelles and molecules. Most centrifugation work done using preparative ultracentrifuge
Density gradient centrifugation 1) Rate zonal centrifugation -Separates molecules based on molecular weight -Used to separate various types of macromolecules such as different types of RNA, DNA 2) Isopycnic centrifugation -separate molecules based on density -used to separate cell organelles
Differential centrifugation
-Start with a uniform mixture of sample -Centrifuge to obtain pellet of heavier molecules and supernatant -Take supernatant & recentrifuge to pellet the next largest particle -Commonly used for separation of organelles
CLASSIFICATION OF CENTRIFUGATION BASED ON RATE/SPEED Ultracentrifugation - x> 20,000 rpm Super speed ultracentrifugation 10,000 <x<20,000 rpm Low speed centrifugation x<10,000 rpm
ROTOR -Two types of rotor : swinging-bucket rotor and fixed-angle rotor SWINGING-BUCKET ROTOR Consist of a rotor from which hang three to six free moving bucket that hang vertically when the rotor is at rest
ASSIGNMENT 1
TITLE
: INTRODUCTION TO LABORATORY INSTRUMENTS : 11, Arial, 1.5 spacing : Maximum 10 : Green (Name, ID Num., Lecturers name, Group, Dateline) : 06 JULY 2011
SAMPLE