Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Although generally considered safe, magnetic resonance imaging (MRI) has a number of safety issues, including the effects of high magnetic fields and radiofrequency pulses on the body, and on implanted devices, the side effects of contrast agents, toxicity during pregnancy, claustrophobia, and hearing loss. (J Neuro-Ophthalmol 2002;22: 3539)
of implantation, the mechanism by which the device is secured, and the strength of the static and gradient magnetic fields are also important parameters in determining clip or implant motion in MRI (5). Pacemakers are a strict contraindication for MRI. Dislodgement, disruption of the electrical system of the pacemaker, and asynchronous pulsing may result from the magnetic and gradient fields imposed on the pacer (6). Complications include rapid cardiac stimulation, arrhythmias, fibrillation, and burns (7,8). Much anxiety and controversy surrounds the safety of MRI in patients with aneurysm clips. Although no metal is entirely nonferromagnetic, clips may be classified nonferromagnetic for clinical purposes if they pass the widely accepted deflection test (9). On the basis of these studies, New et al. (9) recommended that new implants be made of alloys of titanium, stellite, and Elgiloy. Stellite and Elgiloy are cobaltnickel alloys. Clips made of titanium and its alloys are nonferromagnetic; alloys containing at least 10% to 14% nickel are weakly ferromagnetic but safe for MRI because nickel is able to stabilize the iron into a form that renders it less magnetic. Nearly all aneurysm clips made today are nonferromagnetic and have been deemed safe in the MRI environment (5,1016). In vitro studies of titanium alloy, titanium, austenitic stainless steel, Elgiloy, and Phynox aneurysm clips undergoing long-term or multiple exposures to the strong magnetic fields of a 1.5 Tesla magnetic resonance system have shown less than 2 degrees of deflection, rendering them safe for MRI environments (11, 12,14,17,18). Human studies confirm the safety of these clips (12,13,15,17). However, after the fatality of a patient with a misidentified ferromagnetic clip in 1993, MRI centers are highly sensitized to the appropriate identification of aneurysm clips (19). There are numerous tables identifying aneurysm clips and their MRI compatibility. To insure the patients safety, the written operative note describing the implant specifications should be verified before the MRI procedure. Many other implantable devices must be reviewed for safety before MRI. Vascular implants are both nonferromagnetic and ferromagnetic. However, these implants are
35
S. M. Chung
believed to have become firmly incorporated into the vessel wall within 6 weeks of implantation. Therefore, intravascular coils, filters, and stents are MRI safe (7,20). Intraocular metallic foreign bodies create potentially serious risks for visual loss and are therefore considered a relative contraindication to MRI (21). Prosthetic valves have been shown to be safe (22). Although many such valves do show deflection forces (2.53 degrees in a 1.5 T magnet), these forces are far less than those exerted by the beating heart itself (7,22). Orthopedic implants are nonferromagnetic and therefore compatible with safe MRI scanning. However, the Perfix (Instrument Makar, Okemos, MI) interference screw used in repair of the anterior cruciate ligament is ferromagnetic. The force keeping this screw in place prevents dislodgement and pain but creates artifactual changes precluding interpretation (7). Pellets, bullets, and shrapnel may be ferromagnetic, depending on the country of origin and their intended use. Because of the uncertainty, these objects are considered relative contraindications to MRI scanning. A number of magnetically activated implants, such as magnetic stoma plugs, magnetic sphincters, ocular implants, and certain cochlear implants, are hazardous in the MRI environment and should be removed before imaging (7,23,24). Time-varying gradient magnetic fields create voltages and electrical currents. The currents created in MRI systems are considered insufficient to cause biologic effects except for peripheral nerve stimulation and flashes of light (magnetophosphenes), the latter believed to be generated by rotation and alignment of retinal photoreceptors. Such currents are generated particularly with 4 T MRI units (25,26). Therefore, the Food and Drug Administration (FDA) limits the switching rates (necessary to generate a time-varying gradient magnetic fields) to a factor of three below the mean threshold for peripheral nerve stimulation (4,8). Pulsed radiofrequency fields are used to create MRI signals from tissue. These magnetic fields can cause skin burns when closed conducting loops are created by skin-toskin contact between extremities (27), as by pulse oximeters (28) or electrocardiogram leads (8). Patients may also receive burns at the site of tattoos (29). Metallic prostheses such as artificial hips can absorb heat but do not appear to cause injury (16,30). Gold weight lid implants used to treat impaired eyelid closure after seventh nerve palsies have been shown to be safe for MRI (31). The electromagnetic fields created for MRI render scanning unsafe in patients with neurostimulators, bone growth stimulators, and drug infusion pumps. Local pain and malfunction of the units have been described (24,32 34). Cochlear implants are dangerous because some are controlled electronically, and some have magnets within the implants (35). Dental brace wires cause significant imaging artifacts but are not hazardous.
36
be readily appreciated on manual injection. Furthermore, pain and tissue necrosis from extravasation are important adverse effects demonstrated both clinically and experimentally (4547). Magnevist was originally shown to cause a transient rise in bilirubin and serum iron. However, after the reformulation of Magnevist, no changes in levels of bilirubin and iron were detected (37). ProHance also causes nausea (1%1.5%), and headache (0.6%), and can rarely cause altered taste sensation (0.9%1.4%) (37). Ten anaphylactoid reactions have occurred in an estimated 150,000 administrations (48). Omniscan and Dotarem have similar profiles to the other gadolinium agents with respect to side effects. There are no known contraindications to use of any of the MRI contrast agents. In vitro studies of sickle erythrocytes demonstrated alignment of deoxygenated cells perpendicular to the external magnetic field (49,50). However, there have been no reports of sickle crisis precipitated by MRI contrast agents.
pregnant patients unless the benefits clearly outweigh the risks (37,38). As a matter of interest, pregnant health care personnel exposed to chronic low levels of static magnetic fields (but not to active gradient or radiofrequency electromagnetic fields, which are found in MRI) do not appear to have a higher rate of spontaneous abortion, preterm delivery, low birth weight, or infertility (54).
CLAUSTROPHOBIA
Claustrophobia, anxiety, and panic attacks are the most disabling psychologic phenomena experienced by patients. The proximity of the patients face to the inner wall of the gantry, the duration of the study, loud noises, restricted body movements, and the temperature and humidity contribute to the patients distress (55). Comforting measures include educating the patient about the specifics related to the procedure, maintaining physical and auditory communication with the patient, and providing ample flow of air and lights to minimize the feeling of close surroundings (56,57). Sedation is many times necessary for adequate relaxation. With the development of wider gantries, these problems are expected to diminish.
HEARING LOSS
The noise created during MRI can be a risk to the patient. As many as 43% of patients without appropriate ear protection report temporary hearing loss (58). Cases of permanent hearing loss have also been reported (59). Duration of exposure, frequency of exposure, and intensity of noise are the most important factors inducing hearing loss. The noise arises primarily from the rapid alterations in currents within the gradient coils. The force causes motion or vibration of the coils which impact against the mountings. The banging that results is typically between 65 and 96 dB (60). As the section thickness and field of view decreases and as the repetition time and echo time become shorter, the noise becomes louder. But as acquisition time is reduced, hearing loss is less likely. Earplugs are the safest and least costly means to prevent hearing loss. They reduce noise by 10 to 20 dB (58). Other options include MRI-compatible headphones and antinoise techniques (61). Fourier analysis of the MRI noise can also be used to create a signal of opposite phase to create a 50% to 70% reduction in perceived noise. REFERENCES
1. Schenck JF, Dumoulin CL, Redington RW. Human exposure to 4.0Tesla magnetic fields in a whole-body scanner. Med Phys 1992;19: 108998. 2. Robitaille PML, Abduljalil AM, Kangarlu A, et al. Human magnetic resonance imaging at 8.0 T. NMR Biomed 1998;11:2635. 3. Beischer DE, Knepton JC Jr. Influence of strong magnetic fields on
37
S. M. Chung
4. 5.
6.
7.
8.
9.
15. 16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26.
the electrocardiograph of squirrel monkeys (Saimiri sciureus). Aerospace Med 1964;35:93944. Price RR. The AAPM/RSNA physics tutorial for residents: MR imaging safety considerations. RadioGraphics 1999;19:164151. Shellock FG, Crues JV. High-field-strength MR imaging and metallic biomedical implants: an ex vivo evaluation of deflection forces. AJR Am J Roentgenol 1988;151:38992. Hayes DL, Holmes DR Jr., Gray JE. Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakers. J Am Coll Cardiol 1987;10:7826. Shellock FG, Kanal E. MR procedures and patients with electrically, magnetically, and mechanically activated implants and devices. Magnetic Resonance: Bioeffects, Safety, and Patient Management. New York: Raven Press, 1994:12331. Center for Devices and Radiological Health Magnetic Resonance Working Group. A primer on medical device interactions with magnetic resonance imaging systems 1997;115. Available at: http://www.fda.gov/cdrh/ode/primerf6.html. Accessed December 30, 2000. New PF, Rosen BR, Brady TJ, et al. Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging. Radiology 1983;147:14953. Barrafato D, Henkelman RM. Magnetic resonance imaging and surgical clips. Can J Sur 1984;27:50912. Brothers MF, Fox AJ, Lee DH, et al. MR imaging after surgery for vertebrobasilar aneurysm. AJNR Am J Neuroradiol 1990;11: 14961. Pride GL Jr., Kowal J, Mendelsohn DB, et al. Safety of MR scanning in patients with nonferromagnetic aneurysm clips. J Magn Reson Imaging 2000;12:198200. Dujovny M, Gundamjaj NR, Misra M, et al. Aneurysm clips. Crit Rev Neurosurg 1997;7:16975. Dujovny M, Kossovsky N, Kossovsky R, et al. Aneurysm clip motion during magnetic resonance imaging: in vivo experimental study with metallurgical factor analysis. Neurosurgery 1985;17: 5438. Holtas S, Olsson M, Romner B, et al. Comparison of MR imaging and CT in patients with intracranial aneurysm clips. AJNR Am J Neuroradiol 1988;9:8917. Shellock FG. MR imaging of metallic implants and materials: a compilation of the literature. AJR Am J Roentgenol 1988;151: 8114. Kanal E, Shellock FG. Aneurysm clips: effects of long-term and multiple exposures to a 1.5-T MR system. Radiology 1999;210: 5635. Shellock FG, Kanal E. Yasargil aneurysm clips: evaluation of interactions with a 1.5-T MR system. Radiology 1998;207:58791. Klucznik RP, Carrier DA, Ryka R, et al. Placement of a ferromagnetic intracerebral aneurysm clip in a magnetic field with a fatal outcome. Radiology 1993;187:8556. Teitelbaum GP, Bradley WG Jr., Klein BD. MR imaging artifacts, ferromagnetism, and magnetic torque of intravascular filters, stents, and coils. Radiology 1988;166:65764. Kelly WM, Paglen PG, Pearson JA, et al. Ferromagnetism of intraocular foreign body causes unilateral blindness after MR study. AJNR Am J Neuroradiol 1986;7:2435. Soulen RL, Budinger TF, Higgins CB. Magnetic resonance imaging of prosthetic heart valves. Radiology 1985;154:7057. Hunyadi S Jr., Werning JW, Lewin JS, et al. Effect of magnetic resonance imaging on a new electromagnetic implantable middle ear hearing device. Am J Otol 1997;18:32831. Shellock FG, Kanal E. Policies, guidelines, and recommendations for MR imaging safety and patient management. J Magn Reson Imag 1991;1:97101. Cohen MS, Weisskoff RM, Rzedzian RR, et al. Sensory Stimulation by Time-Varying Magnetic Fields. Magn Reson Med 1990;14: 40914. Starr SJ, Groch MW. Magnetochemical stress in the human retina and the safety of MR imaging [abstract]. Radiology 1988;169:276.
27. Knopp MV, Essig M, Debus J, et al. Unusual burns of the lower extremities caused by a closed conducting loop in a patient at MR imaging. Radiology 1996;200:5725. 28. Brown TR, Goldstein B, Little J. Severe burns resulting from magnetic resonance imaging with cardiopulmonary monitoring. Risks and relevant safety precautions. Am J Phys Med Rehab 1993;72:1667. 29. Wagle WA, Smith M. Tattoo-Induced Skin burn during MR Imaging. AJR Am J Roentgenol 2000;174:1795. 30. Mechlin M, Thickman D, Kressel HY, et al. Magnetic resonance imaging of postoperative patients with metallic implants. AJR Am J Roentgenol 1984;143:12814. 31. Marra S, Konior R, Leonetti JP, et al. Effect of magnetic resonance imaging on implantable eyelid weights. Ann Otol Rhinol Laryngol 1995;104:44852. 32. DeLuca SA, Castronovo FP Jr. Hazards of magnetic resonance imaging. Am Fam Physician 1990;41:1456. 33. Gleason CA, Kaula NF, Hricak H, et al. The effect of magnetic resonance imagers on implanted neurostimulators. Pacing Clin Electrophysiol 1992;15:8194. 34. Von Roemeling R, Lanning RM, Eames FA. MR imaging of patients with implanted drug infusion pumps. J Magn Res Imag 1991;1:7781. 35. Teissl C, Kremser C, Hochmair ES, Hochmair-Desoyer IJ. Magnetic resonance imaging and cochlear implants: compatibility and safety aspects. J Magn Reson Imag 1999;9:2638. 36. Oksendal AN, and Hals PA. Biodistribution and toxicity of MR imaging contrast media. J Magn Reson Imagin 1993;3:157. 37. Runge VM. Safety of approved MR contrast media for intravenous injection. J Magn Reson Imaging 2000;12:20513. 38. Shellock FG, Kanal E. Safety of Magn Reson Imaging Contrast Agents. J Magn Reson Imaging 1999;10:47784. 39. Rofsky NM, Weinreb JC, Bosniak MA, Libes RB, Birnbaum BA. Renal lesion characterization with gadolinium-enhanced MR imaging: efficacy and safety in patients with renal insufficiency. Radiology 1991;180:8589. 40. Yoshikawa K, Davies A. Safety of ProHance in special populations. Eur Radiol 1997;7(suppl 5):246250. 41. Krahe T, Landwehr P, Gotz R, et al. The dialyzability of gadolinium-DTPA [abstract]. Radiology 1990;177:167. 42. Carr JJ. Magnetic resonance contrast agents for neuroimaging. Safety issues. Neuroimaging Clin North Am 1994;4:4354. 43. Niendorf HP, Dinger JC, Haustein J, et al. Tolerance data of GdDTPA: a review. Eur J Radiol 1991;13:1520. 44. Nelson KL, Gifford LM, Lauber-Huber C, et al. Clinical safety of gadopentetate dimeglumine. Radiology 1995;196:43943. 45. Cohan RH, Leder RA, Herzberg AJ, et al. Extravascular toxicity of two magnetic resonance contrast agents. Preliminary experience in the rat. Invest Radiol 1991;26:2246. 46. McAlister WH, McAlister VI, Kissane JM. The effect of Gddimeglumine on subcutaneous tissues: a study with rats. AJNR Am J Neuroradiol 1990;11:3257. 47. Vehmas T, Markkola AT. Gd-DTPA as an alternative contrast agent in conventional and interventional radiology. Acta Radiol 1998;39: 2236. 48. Shellock FG, Kanal E. Bioeffects and safety. In: Stark DD, Bradley WG Jr., eds. Magnetic Resonance Imaging. St. Louis: Mosby, Inc., 1999:291306. 49. Brody A, Sorette M, Gooding C, et al. Induced alignment of flowing sickle erythrocytes in a magnetic field. Invest Radiol 1985;20:5606. 50. Brody A, Embury S, Mentzer W, et al. Preservation of sickle cell blood flow patterns during MR imaging: an in vivo study. AJR Am J Roentgenol 1988;151:239241. 51. Shellock FG, Kanal E. Magnetic resonance procedures and pregnancy In: Magnetic Resonance: Bioeffects, Safety, and Patient Management. New York, NY: Raven Press, 1994:417. 52. Clements H, Duncan KR, Fielding K, et al. Infants exposed to MRI in utero have a normal paediatric assessment at 9 months of age. Br J Rad 2000;73:1904. 53. Weinmann H, Brasch RL, Press W, et al: Characteristics of gadolinium-DTPA complex: a potential NMR contrast agent. AJR Am J Roentgenol 1984;142:619.
38
54. Kanal E, Gillen J, Evans J, et al. Survey of reproductive health among female MR workers. Radiology 1993;187:3959. 55. Quirk ME, Letendre AJ, Ciottone RA, Lingley JF. Anxiety in patients undergoing magnetic resonance imaging. Radiology 1989; 170:4636. 56. Flaherty JA, Hoskinson K. Emotional distress during magnetic resonance imaging. N Engl J Med 1989;320:4678. 57. Granet RB, Gelber LJ. Claustrophobia during MR imaging. N J Med 1990;87:47982.
58. Brummett RE, Talbot JM, Charuhas P. Potential hearing loss resulting from MR imaging. Radiology 1988;169:53940. 59. Kanal E, Shellock FG, Sonnenblick D. MRI clinical site safety survey: phase I results and preliminary data [abstract]. Magn Reson Imaging 1988;7:106. 60. Kanal E, Shellock FG, Talagala L. Safety considerations in MR imaging. Radiology 1990;176:593606. 61. Goldman AM, Gossman WE, Friedander PC. Reduction of sound levels with antinoise in MR imaging. Radiology 1989;173:54950.
39