UNICORN Basics
UNICORN is divided into four main sections: UNICORN Manager: Used to organise data (result files, user set up etc) and configure the system for different optional components Method Editor: Create and edit methods to control the instrument automatically System Control: View run data and control the instrument manually (in real time) Evaluation: Present the chromatographic data post acquisition in flexible, user defined formats
2/ GE / 02/12/2007
UNICORN Manager
Used to: Create new users and define their level of access to the instrument. Once created, users will have their own folder where results and methods are automatically saved (highly recommended) Configure the software to recognise new hardware components Run Method Queues (methods run one after the other or to start at a certain time) Perform system diagnostics
3/ GE / 02/12/2007
5/ GE / 02/12/2007
Create a new folder Create a new folder where the methods where the methods and results for that and results for that user will be saved user will be saved
6/ GE / 02/12/2007
Change the Group to Change the Group to the desired level of the desired level of access to the access to the instrument (see instrument (see below) * below) *
* An Administrator has full access to instrument including the creation of new users. To see the different levels of access select Access Groups
Select which folders Select which folders will be shared with will be shared with other users and which other users and which System the user is System the user is allowed to operate allowed to operate
7/ GE / 02/12/2007
Handle files
Curves Curves
Logbook Logbook
9/ GE / 02/12/2007
Manual instructions
The system is receiving instructions from UNICORN. These instructions are either : automatically sent (by a method) manually sent (by the user) Menu 'Manual'
10 / GE / 02/12/2007
11 / GE / 02/12/2007
12 / GE / 02/12/2007
13 / GE / 02/12/2007
14 / GE / 02/12/2007
15 / GE / 02/12/2007
16 / GE / 02/12/2007
Method Wizard
The Method Wizard is the easiest way to create chromatographic methods The Wizard is interactive and flexible Most of the parameters are defaulted to tried and tested values The Wizard steps you through method creation in the same order as you would perform a run (technique selection, equilibration, sample injection, fraction collection, buffer gradient etc
17 / GE / 02/12/2007
18 / GE / 02/12/2007
Wizard Continued Method and Column Select the type of chromatography Selection Ion exchange, gel filtration etc
Select the column from the list (default parameters such as flow rate and pressure limits are set for that particular column)
Select the column position on the column Valve (if applicable) Flexible flow rates = flows that can be modified at the beginning of each run Flow Regulation = flows that will automatically adjust downward if the column pressure exceeds pre-set level Buffer Prep = adding conc. buffer, salt, pH modifier and water to four different buffer lines allowing the system to mix the correct proportion to each to attain a desired pH (if 19 / GE / applicable) 02/12/2007
Select the pump inlets for buffer A and buffer B (only applicable to AKTA units with buffer selection or buffer prep valves)
20 / GE / 02/12/2007
Select the length of the equilibration (N.B. This value is set as Column Volumes (CV) i.e if you use a 1mL HiTrap column then it will equilibrate for 5mL) Instead of just using a Default value, you can ask the system to Watch Equilibration. Select the type of watch equilibration function from the drop down menu (e.g. continue with the method if the system reaches stable conductivity) A Watch Function can be set on many different parameters it is simply defined as a conditional control that if a parameter is met then the system proceeds to the next step
21 / GE / 02/12/2007
Select how much buffer will be used to empty (flush) the loop onto the column. Good practice is to enter a value here twice as much as the loop volume attached to the injection valve. For example if you have a 1mL loop, enter a value of 2mL here.
Remember to keep the syringe in place in the injection valve until after the system has injected the sample on the column (avoids sample loss or air getting in the system)
22 / GE / 02/12/2007
Tell the system which port your sample is located (8-port valve designated line S1 to S8)
These two tick boxes tell the system to flush the sample line choose either to flush the lines to waste using (say) 20% ethanol in sample line S8 or to flush the remainder of the sample onto the column with 10mL of buffer A located in the S8 line
23 / GE / 02/12/2007
The system remembers where the last fraction was collected from the previous run (if the same method is used). Tell the system if you want to continue collecting fractions in the next available tube or recycle back to the First Tube.
24 / GE / 02/12/2007
Tell the system to collect this bound fraction, i.e. what volume to be collected in each fraction and the tube type. If you select 18mm tube it will collect this fraction in 15mL Falcon tubes in the centre of the yellow rack.
The system remembers where the last fraction was collected from the previous run (if the same method is used). Tell the system if you want to continue collecting fractions in the next available tube or recycle back to the First Tube.
25 / GE / 02/12/2007
Select to Clean After Elution with 100% B (5CV) to remove any strongly bound contaminant from the column. Typically do not select Reequilibration as this is performed at the beginning of the next chromatography run
26 / GE / 02/12/2007
27 / GE / 02/12/2007
equilibration equilibration
Wash 1 Wash 1
elution elution
Wash 2 Wash 2
reequilibration reequilibration
Methods are divided into logical parts Useful to import blocks from other methods
28 / GE / 02/12/2007
Blocks
A group of instructions A method contains a main method and a set of blocks To use a block, it must Execute be : Block wash created called
Main Method Block equilibration
Main Method
29 / GE / 02/12/2007
Flexibility using Variables Flexibility using Variables Adapt value during Start Up Protocol
30 / GE / 02/12/2007
Conditional Control using Watch commands IF a certain condition is met THEN an action will follow
31 / GE / 02/12/2007
Start a method
Start from Main Menu, Select Methods, Select the appropriate method using right mouse button, Select Run Follow the Start Up Protocol and fill in as appropriate Start from System Control Menu Select File Select Run Select appropriate method Follow the Start Up Protocol and fill in as appropriate
32 / GE / 02/12/2007
33 / GE / 02/12/2007
HELP
On using KTA and UNICORN
US
ASK
34 / GE / 02/12/2007