Developing methods for LC/MS Getting the most out of each ionization technique Practical considerations with LC/MS
CID: Use of voltage and gas pressure in the first vacuum region to break ions into smaller fragments. Also called In-source fragmentation.
HPLC versus MS
SHIMADZU LCMS Training
HPLC Prefers neutral species Requires dissolved molecule Separates based on solubility (ie, retention time)
Mass Spectrometer Prefers ionized species Requires dry or desolvated molecule Separates based on mass
HPLC versus MS
SHIMADZU LCMS Training
Problem
HPLC prefers neutral compounds in order to achieve a good separation LCMS requires ionic species in order to obtain good signal levels
Solutions
Pick a compromise pH that gives some separation and some ionization Change pH post column Use HILIC (hydrophilic interaction chrom) (normal phase column with reverse phase solvents) Use APCI and do ionization in gas phase Consider the use of volatile ion-pairing reagents to create neutral species for separation and ionic species for detection
Mass: A quantitative measure of matter resistance to being accelerated. Nominal (or unit) mass: The mass based on nominal atomic
weights: I.e. Carbon = 12; Hydrogen = 1; Oxygen = 16, etc.
Average Mass: The average weighted mass based on all the isotopes
naturally occurring. Average mass of carbon = 12.011. Hydrogen = 1.00794. Chlorine35 = 34.968852 [it is a mixture of the isotope at 35 (76% of the total) and 37 (24% of the total)].
Molecular Weight: The sum of all the average masses of each of the
elements present in a compound.
He
Li
Be
Ne
Na
Mg
Al
Si
Cl
Ar
Ca
Sc
Ti
Cr
Mn
Fe
Co
Ni
Cu
Zn
Ga
Ge
As
Se
Br
Kr
Rb
Sr
Zr
Nb
Mo
Tc
Ru
Rh
Pd
Ag
Cd
In
Sn
Sb
Te
Xe
Cs
Ba
La
Hf
Ta
Re
Os
Ir
Pt
Au
Hg
Tl
Pb
Bi
Po
At
Rn
# of natural isotope:
8+
Clomethacillin
Trichlordiuride
Mercurichrome
Improving Quantitation
SHIMADZU LCMS Training
Reduce number of isotopic elements to reduce number of ions diluting the signal levels
Eg, for Cisplatin, hydrolyze two chlorine atoms to hydroxyls to gain ~70 more signal intensity. Cisplatin Cisplatin after hydrolysis
Lysine Ionization
SHIMADZU LCMS Training
m/z=145 m/z=147
O
NH3
+
NH3 OH H2N O
OH
H3N
Lysine
O
1
Relative ionization
[M+H] +1
[M-H] -1
NH2 O H2N O
-
[M+2H] +2
10
12
14 pH
Amino Acids
SHIMADZU LCMS Training
How do you analyze an amino acid if it is ionic at all pH values? If each amino acid has a unique pKa, how to analyze a mixture of them without having to change chemistry for each one? How to avoid doing derivitization? How to separate them without having to do ion exchange separations? Can they still be detected by LC/MS? As positive or negative ions?
2750e3
2500e3
2250e3
2000e3
76.00(44.67) 90.00(13.60) 118.00(4.13) 132.00(2.43) 106.00(9.01) 120.00(6.17) 150.00(2.45) 166.00(1.41) 182.00(6.64) 116.00(2.35) 134.00(8.59) 148.00(3.08) 147.00(1.00) 175.10(1.00) 156.00(1.00)
NH3
OH H2N O
1750e3
1500e3
1250e3
1000e3
750e3
500e3
250e3
2.5
5.0
7.5
10.0
12.5
15.0
17.5
20.0
22.5
min
Conditions: Column : Shim-pack VP-ODS (2.0 mm I.D. x 150 mm L) Mobile phase A : 5 mM HFBA-water, Mobile phase B : acetonitrile Gradient program : 15%B (0 min) 35%B (12 - 25 min) Ionization mode : ESI positive
ion-pair formation:
Analyte and the ion-pair reagent form a neutral pair that partitions between the mobile phase and the stationary phase.
dynamic ion-exchange:
The hydrophobic portion of the ion-pair reagent adsorbs onto the hydrophobic stationary phase to form an active ion-exchange surface that separates ionic species just like classical ion exchange.
ion-interaction:
An electrical double layer is formed when the ion-pair reagent permeates the stationary phase, carrying with it an associated counter ion. The separation occurs as a combination of factors.
Improved separations for highly ionic species Acid and basic compounds can be analyzed on reversed-phase columns More reproducible results Sharper peaks with less tailing Simultaneous separation of ionic and non-ionic compounds in one run More chromatographic options for increasing resolution
1:Diquat/Paraquat
5:Iminoctazine
14 10 11 9
8:Asulam
7 8 6 12 13
7:Carbendazim
Change Analytical Conditions for Pesticides Analysis using Ion Pair Reagents
Instrument: HPLC Column: Mobile Phase A: Time Program: Flow Rate: Column Temperature: Injection Volume: MS Ionization Mode: Nebulizer gas Flow: CDL Temperature: CDL Voltage: ESI(+) 1.5L/min 200 S-mode Applied Voltage: Drying gas Pressure: Heat Block Temperature: 4.5kV 0.2MPa 200 S-mode L-column ODS (2.1mm I.D.150mm L) 5mM HFBA Mobile Phase B: acetonitrile Shimadzu LCMS-2010A
Q-array Voltage:
Comparison of Retention Time between Ion Pair Reagent Conditions and 0.1% Formic acid
Name Acephate Asulam Diquat Oxamyl Paraquat Methomyl Oxine-copper Trichlorfon Benomyl(MBC) Dimethoate Iminoctadine triacetate
Retention Time (a); Retention Time (b);
N+ N+
Retention Time(a) (min) Ion pair 9.926 12.787 13.224 13.343 13.700 13.820 14.773 16.362 17.157 18.110 19.541
Retention Time(b) (min) Formic 10.162 13.156 1.583 14.201 10.106 11.752 10.964.
2Br-
Diquat dibromide
Paraquat
Retention Time and Detected Ions of Pesticides with HFBA Ion Pair Reagent
Name RetentionTime (min) 9.926 12.787 Monoisotopic mass 183.01 230.04 Detected ions (m/z) 229(M+AcN+Na) 184 483(2M+Na) 253(M+Na) 231 294(M+AcN+Na) 276(M+46) 183 92(M+2H)2+ 265(M+46) 293 93(M+2H)2+ 185 163 208(M+46) 146 257 298(M+46) 192 230 275(M+46) 179(M+2H)2+ 356 570 Acephate Asulam SIM Monitor Ion (m/z) 184 231
Diquat Oxamyl Paraquat Methomyl Oxine-copper Trichlorfon Benomyl(MBC) Dimethoate iminoctadine triacetate
183 265 185 163 146 257 192 230 178.8 356
xxx xxx
8.0 184.00
(14.86) 231.00 (36.62) 265.10 (1.75) 7.5 163.00 (5.77) 256.90 (25.62) 146.00 (26.40) 7.0 192.05 (1.51) 178.65 (2.28) 6.5 183.00 (9.16) 185.10 (9.64)
6.0 5.5 5.0 4.5 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5 0.0 5.0 6.0 7.0 8.0
Methomyl Oxine-Cu
Diquat Paraquat
9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 21.0 22.0 23.0 24.0
Oxine-copper
(x100,000) 4.5
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
500 750
(x100,000)
Zx
0.0
1000
2000
3000
4000
Zx
F(x)=431.1706+-354.7959 R2=0.99950
F(x)=88.15862+-1704.467 R2=0.99999
Concentration(ppb) CV(%) 100 3.83 200 2.01 500 2.58 1000 1.56
Concentration(ppb) CV(%) 100 11.45 500 9.34 1000 6.20 5000 3.16
Scope of API-MS
SHIMADZU LCMS Training
Molecular Weight
100,000
Electrospray
1000
ApPI
ApCI
Non-Polar
Polar
Ion generation
SHIMADZU LCMS Training
Atmospheric pressure ionization (API)
Sample ionization takes place outside the vacuum system of the MS at atmospheric pressure. Source sample molecules are ionized to produce an ion beam of pseudo-molecular ions.
APCI or ESI?
SHIMADZU LCMS Training
Which is better?
For very polar compounds, proteins and peptides, ESI works better For non-polar compounds such as steroids and carotenes, APCI works better For most everything else, both will work
APCI will have fewer alkali metal adducts, but may have more solvent based adducts or thermal decomposition products ESI will usually require additives to foster ionization and will generally be more sensitive for polar compounds
1 2
NH2
4 3
N
R
1
APCI TIC m/z 200-800
X N
2
X
OH
ESI [M+Na]+
I nt .
275e3
250e3
400e3
350e3
225e3
200e3
175e3
150e3
300e3
250e3
200e3
125e3
100e3
75e3
50e3
25e3
0e3
7.5
150e3
100e3
50e3
10.0
12.5
15.0
mn i
7.5
10.0
12.5
15.0
mn i
Aflatoxins
Advantages of ESI
SHIMADZU LCMS Training
High sensitivity to polar compounds Produces multiply charged ions Concentration dependent technique Well suited to reverse phase solvents
Sensitivity
SHIMADZU LCMS Training
Varies greatly depending on compound, interface, and chemistry Can range by several orders of magnitude depending on structure, matrix, mobile phase, additives Varies with pH, temperature, ionic strength, etc Typically LC/MS is performed depending on the matrix and type of compound from low picograms up to high nanograms. Higher levels generally require dilution or splitting
ESI Sensitivity
SHIMADZU LCMS Training
1 pg Injections of Reserpine in SIM Mode
Analysis Results
The SIM mode used (M+H)+ as the monitored ion. Five consecutive analyses at 1 pg/L concentration and 1 L injected volume achieved results of 73.25 (rms) min. S/N ratio
Int.
S/N 75.91(rms) S/N 89.7(rms) S/N 82.76(rms) S/N 79.05(rms) S/N 73.25(rms)
4000
3500
3000
Analysis Conditions
HPLC
Column : 2 mmI.D. x 30 mm L(SBC18) Mobile phase : 5 mM ammonium formate/MeOH (25/75) Mobile phase flowrate: 0.4 mL/min Area 4000e3
7.5
10.0
12.5
15.0
17.5
20.0
22.5
min
3000e3
2000e3
MS
Ionization mode Block heater temp. CDL temp. Nebulizer gas flowrate : : : : ESI(+) 200C 250C 1.5 L/min 0 250 1000e3
500 Conc(pg/L)
750
1000
mAU Ch1-240nm,1nm (1.00) 8.0 Ch2-250nm,1nm (1.00) Ch3-260nm,1nm (1.00) Ch4-270nm,1nm (1.00) Ch5-280nm,1nm (1.00) 7.0
6.0 5.0
4.0 3.0
2.0
1.0
0.0 0.0 1.0 (x10,000) 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 min
6.0 5.5 5.0 4.5 4.0 3.5 3.0 2.5 2.0 1.5 1.0 0.5
484.30 (1.00)
CH3 H3C H3C O O H3C O
H O
OH O O CH3
MS detection
CH3
T-2 Toxin (Fusaritoxin) : C24H34O9 Exact Mass: 466.22 Mol. Wt.: 466.52
78 PPB
0.0 0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
11.0
12.0
13.0
14.0
15.0
16.0
17.0
18.0
19.0
Adduct ion generation must also be accounted for in interpretation and quantitation. Proton affinity of solvent or sample impurities can become a factor for difficult to ionize compounds Some compounds may ionize as multiply charged species and show up at lower masses than expected Keep contaminants out of spray chamber
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
6.591 7.0
7.5
8.0
8.5
min
[M+H]+
212.1
2500e3 153.0 170.0 0e3 150 Int. 152.9 10.0e6 5.0e6 0.0e6 150 171.0 194.0 200 230.1244.1 250 272.1 303.1 300 194.0 200
[M+H]+
212.0
[M+H+MeOH]+
253.1
When concentration is too high, adduct ions and peak saturation occur more.
Main Factor
Effect on ionization
Improve method Adjust chemistry to induce dipole in nonpolar compounds Use higher % organic phase Use drying gas or heat Use high efficiency HPLC column Use low ST solvent (like CH3CN)
Polarity of compound The less polar, the more difficult Solvent B.P. Size of droplet Number of ions in droplets Surface tension Organic easy to be evaporated A small size is critical Depend on polarity of compound Lower ST, easy for ion emitting
Sensitivity
SHIMADZU LCMS Training
Most LCMS instruments are sensitive to low picograms of target compounds. Instruments stay cleaner longer when operated in this range. Most operators do not use appropriate techniques to deal with such low concentrations!
What is a Picogram?
SHIMADZU LCMS Training
Sensitivity of quadrupole LCMS = 0.1 picograms (for ideal compounds by electrospray) Average amount of erythromycin in surface water (106 river streams according to USGS) = 1 picogram/microliter Weight of 100 E. Coli bacteria cells = 3 picograms Total organic carbon in B&J HPLC grade water = up to 50 pg/microliter Total non-volatile residue in Fisher brand HPLC grade water = up to 3 nanograms/microliter One pollen grain (Alder tree) = 9,700 picograms Solubility of fused quartz in high purity 18 megohm water = 11,000 picograms/microliter/4 weeks One typically messy fingerprint = 20,000,000,000 picograms
Contamination Sources
SHIMADZU LCMS Training
Dirty containers (detergents and soaps) Dirty wooden or plastic drying racks (detergents, dust, salts) Dirty tools, wrenches & screwdrivers (oils) Vials, pipettes, glassware (Na+ from soda-lime glass) Paper (detergents, salts) Adhesives from labels (phthalates, solvents) Bottle liners (phthalates, silicones, antioxidants) Plastic tubing (phthalates, silicones) Plastic trays (amide slip agents, antioxidants, UV stabilizers) Fingerprints (every chemical in the universe)
Suggestions
SHIMADZU LCMS Training
Dedicate and mark glassware for LCMS Clean and dry glassware without using detergents or chromate cleaners
Use furnace to burn off any organic residue
Use a glass plate or tray to dry optical parts (never use paper) Tools should be cleaner than surgical tools (not just sterile, but free of anything organic) Clean the containers used to hold water daily. Use an in-line filter between column and LCMS source
Filter all samples Condition all new HPLC columns BEFORE you connect to the LCMS unit Clean up samples to remove DMSO, DMAc or DMF
Or use a bypass valve to prevent these from getting into the source
Use clean tools (not the hardware you use to work on your car) Dont put MS optical parts on paper towels to dry
TIC(1.00) 854.00(1.00)
TIC(1.00) 854.00(1.00)
3750
3000
3500
3250
2750
3000
2500
2750
2500
2250
2250
2000
2000
min
1.50
1.75
2.00
2.25
2.50
2.75
3.00
3.25
3.50
3.75
4.00
4.25
4.50
4.75
5.00
5.25
min
Low solubility ethoxylated soybean oil emulsifier. Valve should be used before AND after peaks of interest.
Valve used to eliminate polar impurities at V0, and ethoxylated soybean oil at end of chromatogram. (Note injection size has been increased to show additional peaks)
TIC(1.00) 391.05(1.00)
7.5
10.0
12.5
15.0
17.5
20.0
22.5
25.0
27.5
min
Red trace shows m/z 391 (diethylhexyl phthalate) from adhesive tape label on outside of container used for methanol (e.g., the B solvent). Phthalates will dissolve into organic solvents from room air.
10
20
30
40
50
60
Many compounds such as sulfur containing peptides can react with metals in solution, or with metal surfaces. The result is a loss of a target analyte and inconsistent results The cure: use a metal free system (plastics, ceramics, etc), and remove metals from water and reagents. In the following two slides, note the change in the peaks indicated by the arrows. These show a loss of specific target peptides in a metallic system due to reaction of the analyte and metal.
MS-TIC Trace of Tryptic Peptides from -Casein RP with the Shimadzu LC-10i Bio-Inert HPLC System
3000e3
2500e3
M+ H 369.00
2000e3
M+ H+CH3CN 2M+ H
737.05 401.00
1500e3
1000e3
500e3
256.10 278.90 319.95 344.95 0e3 350 300 250
386.05
452.00
482.00
400
450
500
540.10 550
600.05 600
646.95 650
684.15
700
794.05 m/z
Tricresyl phosphate
Int.
369.00
4.0e6
M+ H
3.0e6
Backflow gas on
2.0e6
1.0e6
278.90 355.05 350 410.05 442.05 450 482.05 504.00 500 538.10 550 586.00 600 631.05 662.80 650 707.00 700 737.10 740.05 765.15 750 793.10 m/z
0.0e6
250
300
400
H3C N
N N
Cl
Drug
ODS surface->
Load
SHIMADZU LCMS Training
Eluent A
Pump
HPLC-column
3 4 2 1 6
Detector
Eluent B
MAYI column
Eluent A
Waste
Eluent A
Pump
HPLC-column
3 4 2 1 6
Detector
Eluent B
MAYI column
Eluent A
Waste
Recovery
Standard
Int.
Plasma spiked
Metoprolol
2500e3 2000e3 1500e3 1000e3 500e3
Int.
2.5
min
2.5
min
17
Plasma spiked isopropylantipyrine 100L 0.1%phosphate/Acetonitrile (95/5) Isopropylantipyrine 275nm Plasma matrix280nm
17
17
Sample: Indomethacin spiked plasma (1mg/mL) Solvent for sample loading : 0.1% phosphoric acid in Water/ AcCN (95:5) Injection volume: 500 L
Linearity
Plasma spiked, 10L injected 5ng/mL, 10ng/mL, 50ng/mL, 100ng/mL
350000 300000 250000
Area
Metoprolol
R = 0.999851
Propranolol
R = 0.999842
200000 150000
80
100
Lidocaine
R = 0.999930
Bupivacaine
R = 0.999904
20
40
60 Conc.(ng/mL)
80
100
APCI
SHIMADZU LCMS Training
Advantages
Works with many compounds, especially non-polar Rugged Handles HPLC flow rates up to 2 mL/min Only produces singly charged ions
Disadvantages
Compound must have some vapor pressure Limited to around masses under 1000 AMU Thermal degradation possible
Buffers must be volatile but are usually not needed Column and flow rate generally not critical Detection limits very compound dependent Ideal performance depends on gas phase chemistry
For + ions, solute should have higher proton affinity than mobile phase for - ions, solute should have higher acidity than mobile phase
2100
N H O O H3C
Vincristine Vincristine
N
2050
Vinblastine
2000
H3C
O H
1950
1900
1850
min
Negative mode
Oxygen or chlorine can increase signal intensity Carbon builds up on APCI electrode with CH3CN Carbon forms fastest at high organic % Air or oxygenated solvents can be used to burn off carbon
Antifungal agent used to treat complications of AIDS Relatively water insoluble drug Forms adducts in ESI Protonates easily at neutral pH
pKa = 5.28 (+), 4.47 (++)
N N N N F
N N OH F
MW=306
morpholine
96 28 16 30 47
3 14 1 1 1
Perception: There are few solvents for LCMS Reality: There are more solvents suitable for LCMS than for UV based HPLC Reality: Most solvents and water are not clean enough to take advantage of the full sensitivity of most commercial LCMS instruments (They are too dirty, too contaminated, and have too much nonvolatile residue)
Water Methanol Acetonitrile Ethanol Propanol / Isopropanol Butanol Acetone CH2Cl2 or CHCl3 Hexane/Cyclohexane Toluene CS2 Ethyl Acetate HFIP Ammonia THF
C46 C48
12.5
C50
30e6
20e6
10e6
0e6
2.5
I nt . 3000e3
5.0
7.5
10.0
15.0
C52
17.5
mn i
2500e3
2000e3
1500e3
1000e3
500e3
439.5
549.5
0e3 400
450
500
550
650
700
750
800
850
mz /
: Shim-pack VP-ODS (2.0 mm I.D. x 150 mm) : water Mobile phase B : acetone : 80%B (0 min) - 100%B (5 - 20 min) : 0.2 mL/min : +4.5 kV (APCI-positive mode)
Bottled HPLC grades usually contain 1-2 PPM of sodium, up to 50 PPM of organics, and dissolved silica TOC grade water is available with < 100 PPB organics
Available in 1 quart bottles
Install a C-18 column after the aqueous pump, but before the mixer. The column will trap any low polarity organic compounds soluble in water and remove them. Column can be regenerated by pumping methanol or acetonitrile through it. This approach is better than trying to batch purify water because water will immediately begin growing algae and bacteria. Cleaning up immediately prior to use is preferred.
Phosphates
SHIMADZU LCMS Training
Can lead to artifact peaks:
175
214
96
88
[Na3HPO4]+ [K3HPO4]+
80
72
64
56
[Na3HPO4]+ [K2H2PO4]+
263
48
40
[Na4H3P2O8]+
285
32
[K4H4P2O8]+
311
24
16
[Na2H2PO4]+
143
[K4H3P2O8]+
313
355
160
180
200
220
240
m/z
260
280
300
320
340
Common Problems
SHIMADZU LCMS Training
Dissolved gasses - Carbon Dioxide
CO2 will dissolve in water lowering pH
Purity of modifiers
Methanol and fluorinated acids often contain considerable impurities that give high noise levels.
Detergents
SHIMADZU LCMS Training
Detergents in biological samples will suppress ionization significantly 0.01% SDS can reduce protein signals by 20x in ESI Even trace levels of soaps can affect quantitation of small molecules Detergents may exist in samples as buffers and solubilizing agents or may originate as contamination from cleaning.
Reference: R.R. Ogorzalek Loo, N. Dales and P.C. Andrews, Protein Science 3(1994)1975-1983.
Selection of Additives
SHIMADZU LCMS Training
Comparison of SIM chromatograms
1.Metoprolol 2.Propranolol 3.Lidocaine 4.Bupivacaine
Int. 3 4 350e3 Int. 268.00(1.00) 260.00(1.00) 235.00(1.00) 289.00(1.00) 268.00(1.00) 260.00(1.00) 235.00(1.00) 289.00(1.00)
350e3
300e3
CH3COONH4
300e3
TFA
4
250e3
250e3
200e3
200e3 3
150e3
150e3 2 100e3
min
min
Use clean solvents Dedicate glassware just for LCMS use Do not dry glassware on a rying rack or allow inside surface to touch anything other than solvents Avoid non-volatile solvents
Eliminate DMSO, DMF, DMAc, etc.
Make sure your HPLC solvents are strong enough to elute your target molecules Eliminate salts from sample
Either desalt (SPE) or valve out
1 PPM of salt is approximately 1 grain in 100 mL If you dissolve 1 grain (size, not weight) in 2 mL of solvent (typical vial), the resulting solution equals about 50 ng/microliter. So 1 microliter of the sample solution should be the maximum injection amount for LCMS! Less is better.
Water
Use fresh water in small containers Use up quickly Never allow 100% water to stand around
Add some organic to A solvent Even the best water will cause growth of mold/fungus/algae on check valves, rotor seals, etc
Refrigerate bottles of water not being used Check each new bottle of water for background ions (they can vary considerably from bottle to bottle Use up solvents, especially water quickly and discard anything after a few days of use. Dont clean glassware with anything other than clean solvents
Solvent issues
SHIMADZU LCMS Training
HPLC grade solvents vs MS grade
NVR specification - should be PPB or lower
Most are PPM levels. Some are 0.02% or worse Most are ~0.05% or 100 to 1000 times worse than your samples
Degas mobile phase daily or continuously Each batch may have unique low MW impurity ions
Evaluate before using
Sample solubility
Samples that are difficult to dissolve will precipitate fastest during evaporation of droplets in source Depending on volatility of sample, a coating of dust or oil can show up on metal surfaces and affect results Running stronger, slightly less volatile solvents is often an advantage (ie, substitute ethanol or some IPA for methanol) Adding a make-up flow of additional solvent post column can keep samples from precipitating too early during the evaporation process and improve quantitative results
AVOID PLASTICS!
SHIMADZU LCMS Training
All plastics contain plasticizers and oligomeric species, like PEG (polyethylene glycols) or phthalates Use all-glass solvent filtering devices (ground glass connections only) Store samples in glass vials with Teflon liners Use glass autosampler vials with Teflon liners Rinse plastic syringe filters with solvent before filtering samples