Sei sulla pagina 1di 4

Calibration technology

Standard thermometer Models CTP1000, CTP2000 and CTP9000


WIKAdatasheetCT61.10

Applications
Comparativecalibrationindrywellcalibrators,tubefurna-

cesandliquidbaths

Special features
Highstability Lowdrift,longservicelife Widetemperaturerange

Platinum resistance thermometer model CTP2000

Description
Calibration by means of an external standard thermometer TheWIKAstandardthermometersareparticularlysuited forapplicationsinindustriallaboratories.Theyenableeasy comparativecalibrationinourbaths,intubefurnaceandin drywellcalibrators. Theuseofanexternalreferencethermometerisrecommended,particularlyforthecalibrationofshorttemperature sensors.Thustheerrorsduetotheradialandaxialtemperaturedistributioninthetemperatureconditioningunitare considerablyreduced.

Forcalibration,thetestitemsandthestandardthermometer arebroughttothesametemperatureinatemperatureconditioningunit. Assoonasastabletemperatureisreached,thetestitems arereadortheiroutputsignalsaremeasured(resistance, thermoelectricvoltage,standardsignal)andcomparedwith thestandardthermometer. Usingthiscomparisonmethod,themeasuringuncertainty canbeconsiderablyreducedbecausenotonlythedisplayof thetemperatureconditioningunitistakenintoconsideration.

WIKAdatasheetCT61.1002/2012

Page1of4

Models Platinum resistance thermometers models CTP1000 and CTP2000


Features Themeasuringresistorconsistsofaplatinumwindingof highestpurity. Allpartsarepre-agedinordertoremovecontamination anddistortions. Measuring technology The4-wiredesignoffersanoptimumconnectionsolution forresistancethermometers.Themeasuringresultisaffectedneitherbytheleadresistancesnorthetemperaturedependantfluctuations. Theelectricalconnectionsareweldedinordertominimise thetransitionresistance.Theconnectingwiresareboundin a2mlong,shieldedconnectioncable.

Principle of a platinum winding

Thermocouple model CTP9000


Features ThestandardthermometerisatypeSelementwhose nominalcompositionconsistsofPlatinumRhodium10% (positiveleg)againstPlatinum(negativeleg)andbelongs tothegroupofnoblethermocouples. Itischaracterisedbyitshighstability. Thequalityofthethermowellusedisessentialforstability athightemperatures.Forthisreason,thehigh-purityaluminiumoxideceramicC799isused. ThetypeSthermocouple,besidesthelowageingdrift,also offerstheadvantageofalowbasictolerance. Measuring technology Duringmeasurementitmustbeensuredthatthecompensatingleadsfromthemeasuringpointtothecoldjunction consistofsubstitutematerialswhichhave,inalimited temperaturerange,thesamethermoelectricpropertiesas thematerialsofthethermocouple.Therefore,atthistransition,thereisnothermoelectricvoltage.Thisvoltageisonly generatedatthepointwherethecompensatingleadsare connectedtonormalcopperleads.

Platinum resistance thermometer model CTP1000

Calibration
Yourstandardthermometershouldbecalibratedoncea year. Ifitissubjecttohighmechanicalstresses,itshouldbe calibratedimmediatelytoguaranteethemeasuringuncertainty.
Thermocouple model CTP9000

Page2of4

WIKAdatasheetCT61.1002/2012

Platinum resistance thermometer


Specifications
Temperaturerange Nominalresistance Stability Temperaturecoefficient C 1/K

Model CTP1000
-100+670 100 <40mKafter100hat670C 25 6x480 stainlesssteel 2mPTFEcable strippedandtin-coated 4-wireconnection =0.003850

Model CTP2000
-200+450 100 <50mKafter100hat450C <20mKafter100hat300C =0.003850

Sensitivelength Dimensions,dxl Immersiontubematerial Measuringline Connector Sensorconnection

mm mm

notspecified 4x500

stainlesssteel 2mcable strippedandtin-coated 4-wireconnection

onrequest

4mmbananaplug

Thermocouple
Specifications
Temperaturerange Thermocouple Tolerance Stability Wiresize,dxl Outerdimensions,dxl Immersiontubematerial Measuringline mm mm C

Model CTP9000
0...1,300 TypeSperIEC584,PtRh10%Pt Class1 <0.5after250hat1,300C 0.5x1,500 7x600 CeramicC799 800mmwith4mmbananaplug

Scope of delivery
Thermometer Transportcase

Optional:DKD/DAkkScalibrationfor6temperatures

CalculationofthecharacteristicconstantandtableR(t90) Coldjunctioninthemetaltube,outerdimensions:dxl=6

degreebydegree

mmx250mm,measuringline:2mPVCcable,with4mm bananaplug Optional:bananaplugformodelCTP1000

WIKAdatasheetCT61.1002/2012

Page3of4

Schematic representation of the connections


Resistance thermometer
RL1 RL2 Pt100 RL3 RL4

TC

1 2

3 4 ext. Ref.

(Integrated measuring device CTD9300)

Thermocouples Contactpoint,t2 Thermocouple Coldjunction,t3

Measuring pointt1

CTD9100

CTD9100-ZERO

Ordering information Model/calibration/calculationofconstant/transportcase/coldjunction/additionalorderdetails


2008WIKAAlexanderWiegandSE&Co.KG,allrightsreserved. Thespecificationsgiveninthisdocumentrepresentthestateofengineeringatthetimeofpublishing. Wereservetherighttomakemodificationstothespecificationsandmaterials. 02/2012GB

Page4of4

WIKAdatasheetCT61.1002/2012

WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Strae30 63911Klingenberg/Germany Tel. (+49)9372/132-0 Fax (+49)9372/132-406 E-mail info@wika.de www.wika.de