Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
GUIDE
to the implementation of DIRECTIVE
98/79/EC
on
IVD
In Vitro Diagnostic Devices
EuropeAid /114385/D/SV/CY
This Guide was printed with funds of the EuropeAid /114385/D/SV/CY project: Assessment and administration capacity building for the harmonisation with the New Approach directives The views expressed in this publication do not necessarily reflect the views of the European Commission or the Cyprus Ministry of Commerce, Industry and Tourism. Republic of Cyprus, Nicosia, May 2004
EuropeAid /114385/D/SV/CY
INTRODUCTION
This information is intended as general guidance and should not be regarded as an authoritative statement of the law. Manufacturers and others should not rely on this information but should consult the legislation referred to, making their own decisions on matters affecting them in conjunction with their lawyers and other professional advisers.
THE DIRECTIVES
THE ACTIVE IMPLANTABLE MEDICAL DEVICES DIRECTIVE
This first Directive covers all powered implants or partial implants that are left in the human body. Heart pacemakers are the most common example of powered implants.
EuropeAid /114385/D/SV/CY
CLASSIFICATION
The Medical Devices Directive and the In Vitro Diagnostic Medical Devices Directive include a classification system whereby the level of regulatory control applied to devices is proportionate to the degree of risk associated with the device. The strictest controls therefore apply only to high-risk products.
NOTIFIED BODIES
The Competent Authority is also responsible for designating the independent certification organisations (Notified Bodies) that check that manufacturers of medium and high risk medical devices have followed the requirements. Once they are satisfied, manufacturers may apply the CE mark to their products and place them on the market.
CLINICAL INVESTIGATIONS
The Directives require all devices intended for clinical investigation in the EU to be formally assessed on the risks such investigations might pose to the health and safety of patients. For most medical devices (except in-vitro diagnostic devices) manufacturers have to inform the relevant Competent Authority if they
EuropeAid /114385/D/SV/CY
intend to start a clinical investigation at least 60 days before it is due to begin. The investigation may start after the 60-day period unless the Competent Authority notifies the manufacturer of a decision to the contrary on grounds of public health.
EuropeAid /114385/D/SV/CY
diagnosis, prevention, monitoring, treatment or alleviation of disease, diagnosis, monitoring, treatment, alleviation or compensation for an injury or handicap, investigation, replacement or modification of the anatomy or of a physiological process, control of conception, and which does not achieve its principal intended action in or on the human body by pharmacological, immunological or metabolic means, but which may be assisted in its function by such means. In order for a product to come within the scope of the Medical Device Directives (including the IVDD) it must be intended by the manufacturer to be used for a medical purpose. If no medical purpose is intended by the manufacturer then the product is not a medical device. Thus the intended purpose for a product will be key in determining whether or not the product is a (IVD) medical device. It is to be noted that Article 1.2 (h) of the IVDD provides that the intended purpose is the use for which a device is intended according to the data (presumably, the information) supplied by the manufacturer on the labelling, in the instructions for use and / or in promotional material. Examples: Devices for detection of agents of biological or chemical warfare in the environment are not IVDs because such products have no medical purpose. On the other hand a device intended to be used on human specimens in the detection of biological or chemical warfare agents with medical purpose would fall within the scope of the IVDD. Devices intended to be used only in the course of law enforcement or other non-medical purposes, for example for detecting drugs of abuse/alcohol, are outside the scope of the IVDD. If however, the in- vitro examination of human specimens with a medical purpose is one of the intended uses of a specific product, the IVD Directive will apply. A product for research use only which has no medical purpose, cannot be a medical device and, therefore, cannot be an IVD medical device. By definition, these products fall outside the scope of the IVDD and the other medical devices directives.
SPECIMEN RECEPTACLES
Article 1.2(b) of the IVD Directive clearly states that specimen receptacles are considered as IVDs. It defines specimen receptacles for this purpose as those devices that are specifically intended by the manufacturer to be for the primary containment and preservation of specimens derived from the human body for the purposes of in vitro diagnostic examination. This applies whether the product is vacuum type or not. This specific intended use should be included on the labelling and any associated promotional literature for the product. The manufacturer must also have evidence and technical documentation to support this use for the product. A manufacturer cannot place the CE mark on a piece of general laboratory equipment as a marketing claim, without ensuring that it complies with all the relevant essential and other requirements of the directive and have evidence to substantiate this. It is possible for more that one specimen receptacle to be involved in the collection, transport and storage of an individual specimen. In such cases the manufacturer of each receptacle must have evidence of compliance with the directive as above.
EuropeAid /114385/D/SV/CY
During the actual analytical process the receptacles into which the specimen is placed (by aliquoting or otherwise), may be glass or plastic tubes, cups, cuvetttes or other receptacles. These are unlikely to be specimen receptacles as defined in the directive. They are usually considered to be general laboratory equipment, however in some cases they could be considered accessories to an IVD. In the context of the definition of specimen receptacle in Art. 1.2 (b), of the IVD Directive it should be noted that: a. the word primary does not necessarily refer to the initial or first container of the specimen in point of time, but rather to a container that is intended by its manufacturer to mainly come into direct contact with the specimen and which could therefore affect the specimen, and b. the word preservation does not imply that the receptacle has to contain a specimen preservative, but that the receptacle is one intended to protect the specimen, for example, from temperature fluctuations, from light, from physical breakage, etc. Example: Sample tubes, microbial transportation devices (tubes). Article 1 (2) b applies.
EuropeAid /114385/D/SV/CY
ACCESSORIES TO IVDS
Accessory means an article which, whilst not being an in vitro diagnostic medical device, is intended specifically by its manufacturer to be used together with a device to enable that device to be used in accordance with its intended purpose. (Article 1.2(c)). For the purposes of the Directive accessories are treated as IVDs in their own right. However "invasive sampling devices" or those which are directly applied to the human body for the purpose of obtaining a specimen, are not considered to be accessories to IVDs. Generally, such devices will be regulated by Directive 93/42/EEC (Medical Devices). However, where a diagnostic device incorporates an invasive element and a diagnostic element and is sold as a single integrated unit (rather than two separate products within the same pack) the general view is that such a device will generally be treated as an IVD, rather than a general device regulated by Directive 93/42/EEC.
EuropeAid /114385/D/SV/CY
specimens derived from the human body for the purposes of providing information the IVD directive would apply. (Art 2b). Devices which during their measuring function involve contact with the human system (as defined in 93/42/EEC Annex IX) in order to obtain a continuous sample are not considered to be IVD devices. Example: A device involving the vacuum suction of saliva into the integrated handle of a device which contains reagent material (e.g. for the detection of HIV). The use of such a device involves the penetration of the device into a body orifice for the collection of the specimen and this may appear to make it a medical device within the scope of the MDD. However, its principal intended purpose is the provision of relevant information by the in vitro examination of the specimen derived from the patient. The devices brief contact with the patient or penetration into the patients body to collect the specimen is subsidiary and incidental to its principal intended purpose. Mouth and other swabs having integrated reagents or reagent areas are IVDs because their principal intended purpose is to provide information relevant to the medical purposes specified in Article 1.2 (b). The intended purpose, invasiveness and continuity of sampling are important criteria in deciding the correct regulatory route for this kind of product. Example: A Holter blood glucose monitoring system that includes a subcutaneous catheter to provide a continuous supply of the patients specimen to an in vitro analysing instrument is a medical device not an IVD because, during the in vitro measuring function, surgically invasive contact with the patient is necessary in order to obtain a continuous specimen flow. In this case, the analytical function is carried out at the same time as the continuous specimen collection process is still going on. There is no dissociation of the specimen from the patient and therefore the analytical function cannot properly be regarded as being in vitro. Such a device would therefore be a medical device within the scope of the MDD. Other examples are continuous pH measurements during haemodialysis and oxygen saturation devices.
EuropeAid /114385/D/SV/CY
also applies to any MDD components. I.e. each set of regulatory requirements needs to be met for each of the individual medicinal products, medical device and/or IVD medical device components. Example: A Helicobacter pylori breath test kit containing labelled urea (a licensed pharmaceutical product) to be ingested prior to sampling for analysis, a straw (a medical device) and a sample container (an IVD). Note: So far as the medicinal product is concerned, it must have been granted a marketing authorisation covering the actual use for which it is being included in the IVD kit and it must be labelled in accordance with the regulations relating to medicinal products. The IVD component of the kit must comply with all the applicable essential and other requirements of the IVDD (including the labelling requirements) and it must bear the CE marking. It is important to note that, while the CE marking of the IVD content of the kit entitles it to free movement within the European Economic Area (subject to language requirements), that is not so as regards the medicinal product. That aspect must be considered separately under the relevant medicinal product legislation as must the extent to which the label of the kit must provide information about the medicinal product over and above a statement that the kit contains the medicinal product.
EuropeAid /114385/D/SV/CY
EuropeAid /114385/D/SV/CY
placed on the market under his own name. This does not apply to a person who assembles or adapts devices already on the market to their intended purpose for an individual patient. AUTHORISED REPRESENTATIVE Manufacturers who do not have a registered place of business in the EC must designate an authorised representative to perform certain obligations (e.g. to make certain documentation available on request) and may designate an authorised representative to perform other substantive obligations. Additionally, such a manufacturer must also designate an authorised representative as the person responsible for marketing the IVD in the EC and for registering that device with the appropriate Competent Authority.
NOTIFIED BODY
A "Notified Body" is a certification organisation which the Competent Authority designates to carry out certain tasks in respect of the conformity assessment procedures described in the Annexes to the Directive. A notified body must be qualified to perform all the functions set out in any Annex for which it is designated. The tasks which a notified body can carry out may be restricted by the Competent Authority. The activities of notified bodies are regularly monitored. Manufacturers are free to apply to any notified body in the EU designated to carry out the desired conformity assessment procedure, regardless of which Member State that notified body is designated in. Manufacturers are required to inform their notified body of changes to their product ranges and quality system. In cases where design or type examination has been carried out by the notified body the manufacturer is required to notify them of changes to the design, as well as any information they have on changes to the pathogen and markers of infection to be tested. All such changes need to be approved by the notified body prior to implementation.
ESSENTIAL REQUIREMENTS
Annex I of the Directive lists various "essential requirements" with which IVDs must comply before being placed on the market/put into service. These aim to ensure that IVDs do not compromise the health and safety of patients, users, and others, and are designed and manufactured so that they are suitable for the relevant purpose specified by the manufacturer and achieve the performances stated by the manufacturer. Not all the essential requirements will apply to all devices and it is for the manufacturer of the device to assess which are appropriate for his particular product. In determining this, account must betaken of the intended purpose of the device.
EuropeAid /114385/D/SV/CY
GENERAL IVDS
The manufacturer must fulfil the applicable obligations imposed by sections 1 to 5 of Annex III of the Directive and must declare and ensure that the device meets the provisions of the Directive which apply (EC declaration of conformity). No notified body involvement is required.
ANNEX II IVDS
For Annex II List B devices, the manufacturer must follow the applicable obligations imposed either by Annex IV of the Directive (EC declaration of conformity, full quality assurance system), or by Annexes V (EC type-examination) and VI (EC verification), or alternatively by Annexes V (EC type-examination) and VII (EC declaration of conformity, production quality assurance) and must declare and ensure that the device meets the provisions of the Directive which apply. For List A devices, the manufacturer must follow either Annex IV, or alternatively Annexes V and VII (i.e. it cannot follow Annexes V and VI). All Annex II IVDs require the intervention of a notified body before a declaration of conformity with the Directive can be made.
EuropeAid /114385/D/SV/CY
either carry out an audit of the full quality assurance system or carry out type examination plus verification of each batch or product or carry out type examination plus audit of the production quality assurance system.
EuropeAid /114385/D/SV/CY
EuropeAid /114385/D/SV/CY
Diagram 2:
EuropeAid /114385/D/SV/CY
DOCUMENTATION
The declaration of conformity, the technical documentation and the decision, reports and certificates of notified bodies must be kept available for inspection for a period of five years after manufacture of the last device.
HARMONISED STANDARDS
European Standards that have been harmonised under the IVD Directive may be used to show conformity with the relevant essential requirements. Compliance with an appropriate harmonised European Standard gives a presumption of conformity with the essential requirements to which the standard relates. The use of harmonised standards is not mandatory, and other standards exist that are not harmonised which may be used to assist in showing conformity. However, unlike harmonised standards they offer no presumption of conformity.
THE CE MARKING
WHAT THE CE MARK MEANS
IVDs must bear the CE marking when they are placed on the market. Devices for performance evaluation do not need to be CE marked. A manufacturer must not apply the CE mark unless he has fulfilled the applicable obligations of the Directive. The CE mark is therefore seen as a declaration by the manufacturer that the product meets all of the appropriate provisions of the relevant legislation, including those relating to safety. A device bearing a CE mark can be freely marketed anywhere in the EU without further control, except that competent authorities can take action to prevent the supply of a device in certain circumstances. The CE marking is as illustrated in diagram 3 below. It may not be smaller than 5 mm in its vertical height, and the proportions must be maintained whatever its size. The grid does not form part of the marking and is for information only. Diagram 3
EuropeAid /114385/D/SV/CY
It should be noted that the C and E are not formed by perfect semi-circles, i.e. the top and bottom arms extend one square beyond the semi-circles, and the middle arm of the E stops one square short.
REGISTRATION
A manufacturer who places a relevant device on the market (remember that for these purposes, market means Community market) or who makes available a device for performance evaluation under his own name must register with the Competent Authority in the Member State where he, or his authorised representative is established.
EUROPEAN DATABANK
A European Databank is being set up, accessible only by Member States and the European Commission. It will contain data relating to all devices available on the territory of the Community. Member States will load the information they receive from manufacturers and Notified Bodies onto the Databank. The Databank will include; Data relating to registration of manufacturers and devices; Data relating to certificates issued, modified; complemented, suspended, withdrawn or refused by notified bodies; Data obtained in accordance with the vigilance system (see below).
EuropeAid /114385/D/SV/CY
C 182 of 2002-07-31
EN 376: 2002
C 182 of 2002-07-31
EN 455-1: 2000
C 319 of 2001-11-14
C 182 of 2002-07-31
C 182 of 2002-07-31
EuropeAid /114385/D/SV/CY
EN 980: 1996/ A2:2001 EN 1280-1: 1997/ A1:2000 EN 1658: 1996 EN ISO 4074: 2002 EN ISO 4135: 2001
Graphical symbols for use in the labelling of medical devices Agent specific filling systems for anaesthetic vaporisers - Part 1: Rectangular keyed filling systems Requirements for marking of in vitro diagnostic instruments Natural latex rubber condoms Requirements and test methods (ISO 4074:2002) Anaesthetic and respiratory equipment - Vocabulary (ISO 4135:2001) Anaesthesic and respiratory equipment Heat and moisture
C 182 of 2002-07-31
C 182 of 2002-07-31
exhangers (HMEs) for humidifying respired gases in humans Part 1: HMEs for use with minimum tidal volumes of 250 ml (ISO 93601:2000)
C 319 of 2001-11-14
Biological evaluation of medical devices Part 8: Selection and qualification of reference materials for biological tests (ISO 109938:2000) Biological evaluation of medical devices - Part 14: Identification and quantification of degradation products from ceramics (ISO 1099314:2001) Biological evaluation of medical devices Part 15: Identification and quantification of degradation products from metals and alloys (ISO 10993-15:2000) Non-active surgical implants Body contouring implants Specific C 182 of 2002-07-31 C 319 of 2001-11-14 C 182 of 2002-07-31 C 182 of 2002-07-31
EN 12180: 2000
In vitro diagnostic medical devices Measurement of quantities in EN 12286: 1998 samples of biological origin Presentation of reference measurement procedures EN 12286: 1998/ A1:2000 In vitro diagnostic medical devices Measurement of quantities in samples of biological origin Presentation of reference measurement procedures In vitro diagnostic medical devices Measurement of quantities in samples of biological origin Description of reference materials In vitro diagnostic medical devices Culture media for microbiology Performance criteria for culture media In vitro diagnostic medical devices Culture media for microbiology Performance criteria for culture media C 293 of 2000-10-14 C 319 of 2001-11-14 C 227 of 1999-08-10
EN 12287: 1999
EN 12322: 1999 EN 12322: 1999/ A1:2001 EN 12718: 2001 corrigendum ENV 12719: 2001 corrigendum EN 13328-1: 2001
C 288 of 1999-10-09
C 182 of 2002-07-31
C 182 of 2002-07-31
EuropeAid /114385/D/SV/CY
Test method to assess filtration performance Quality systems - Medical devices - Particular requirements for the EN 13485: 2000 application of EN ISO 9001:1994 (revision of EN 46001:1996) C 182 of 2002-07-31
Quality systems Medical devices - Particular requirements for the EN 13488: 2000 application of EN ISO 9002:1994 (revision of EN 46002 :1996)
C 182 of 2002-07-31
EN 13532: 2002
General requirements for in vitro diagnostic medical devices for selftesting Respiratory therapy equipment - Part 1: Nebulising systems and their components Respiratory therapy equipment - Part 3: Air entrainment devices Performance evaluation of in vitro diagnostic medical devices Stability testing of in vitro diagnostic medical devices Elimination or reduction of risk of infection related to in vitro diagnostic reagents Sampling procedures used for acceptance testing of in vitro diagnostic medical devices - Statistical aspects Sterilisation of health care products - General requirements for
C 314 of 2002-12-17
EN 13544-1: 2001 EN 13544-3: 2001 EN 13612: 2002 EN 13640: 2002 EN 13641: 2002
C 182 of 2002-07-31 C 182 of 2002-07-31 C 314 of 2002-12-17 C 314 of 2002-12-17 C 314 of 2002-12-17
EN 13975:2003
C 280 of 2003-11-21
characterisation of a sterilising agent and the development, validation and routine control of a sterilisation process for medical devices (ISO 14937:2000) Medical devices Application of risk management to medical
C 182 of 2002-07-31
C 182 of 2002-07-31
In vitro diagnostic medical devices - Measurement of quantities in EN ISO 18153:2003 biological samples - Metrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials (ISO 18153:2003) EN ISO 15225: 2000 Nomenclature - Specification for a nomenclature system for medical devices for the purpose of regulatory data exchange (ISO 15225:2000) Safety requirements for electrical equipment for measurement, EN 61010-2-101: 2002 control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment Reference document: IEC 61010-2-101:2002 (Modified) C 182 of 2002-07-31 C 280 of 2003-11-21
EuropeAid /114385/D/SV/CY
Note 1: Generally the date of cessation of presumption of conformity will be the date of withdrawal ("dow"), set by the European Standardisation Organisation, but attention of users of these standards is drawn to the fact that in certain exceptional cases this can be otherwise.
FURTHER INFORMATION
Information on medical devices directives, harmonised standards, competent authorities notified bodies, etc. may be obtained from the web site of the European Commission at; http://europa.eu.int/comm/enterprise/medical_devices/index.htm
STANDARDS
Further information on standards and guidance information can be obtained from the Standards Body of Cyprus: CYS Cyprus Organisation for the Promotion of Quality At the Ministry of Commerce, Industry and Tourism 13-15, Andreas Araouzou Street 1421 Nicosia http://www.cys.mcit.gov.cy
USEFUL LINKS
Further information and links to all relevant European websites are available at
www.cys.mcit.gov.cy
under the button named:
NEW APPROACH
EuropeAid /114385/D/SV/CY
CYPRUS LEGISLATION
General requirements and stipulation are layed down in the Cyprus Framework Law N. 30(I)/2002 (published 5.4.2002) as amended by Law N. 29(I)/2003 (published 28.3.2003). The Framework Law represents the legal basis for the Cyprus Regulations addressing respective Directives. The Cyprus Regulation addressing the IVDD (In vitro diagnostic medical devices) directive is implemented as Regulation P.I. 597/2003 (published 18.7.2003). Furthermore, the MDD directive is Regulation P.I. 598/2003 (published 18.7.2003) and the AIMD directive is Regulation P.I. 599/2003 (published 18.7.2003).