Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Sheet
5.1
We reserve the right to continuously improve our products and make any change in the stated specifications and dimensions without prior notice.
www.mjk.com
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Data Sheet
General
Applications
Inlet
Clarifier
Effluent outlet
Return to inlet
Dewatering process
Sludge thickner
The sensor provides a full range from 0.001 to 9999 FNU/NTU for turbidity and 0.001 to 400 g/l (SiO2) for suspended solids measurements. The turbidity measurement complies with ISO 7027. The SuSix is equipped with beam forming optics for multi-angle detection of turbidity and suspended solids. This advanced optical system is combined with a progressive algorithm using neural logic to generate a reliable high quality measurement. The sensor compensates for errors due to fouling or aging of the optical array and compensates for gas bubbles in the sample.
Calibration
The SuSix sensor is provided with factory calibration for turbidity, and zero point factory calibration for suspended solids. A single point in-situ calibration can be conducted for many applications. For more rigorous or difficult applications a two- or three-point calibration is also possible.
www.mjk.com
2
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Data Sheet
High Quality Materials
Flexible Installation
MJKs modular design allows up to 300 meters (950ft.) between the sensor and the converter, and the Display Unit can be mounted up to 1000 m (3000 ft.) from the signal converter with ordinary twisted wires. One Display Unit can control up to four MagFlux converters and/or SuSix signal converters for greater economy, space savings and an improved overview of the multiple measurement values.
1 pc. SuSix with display (206305) 3 pcs. SuSix without display (206308)
ll RS485
ll RS485
ll RS485
Simple to Operate
The SuSix Display Unit has a mobile phone-like menu structure and can display text in several selectable languages.
PC Connection
SuSix allows downloading setup configurations, uploading new instrument software updates and display customization. The onboard data logger captures 20.000 readings which are displayed as an electronic graph and can be retrieved as a CSV file on a PC. This connectivity is achieved with a common USB port and the free MJK Field Link software.
The SuSix converter has one 4-20 mA analogue output, two digital outputs for alarms or control, and one digital input for resetting alarms, batches, etc.
User-definable Text
Up to five lines of text and readings can be configured by the user. The graphic display is automatically adjusted to show the largest characters possible. Alarms can be displayed as pop-up alarms until they are reset.
Modbus Communication
The Display Unit uses the Modbus RTU communication protocol to connect to the SuSix Converter. The converter can in turn connect to SCADA systems with its Modbus communication protocol. Register lists are available upon request by our customers.
www.mjk.com
3
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Data Sheet
Specifications
Converter and Display Unit
Cabinet material Enclosure rating Temperature range Weight CE approvals Display Unit Enclosure rating Housing material Protection lid Display Indication Clock Communication Interface Memory Interface Temperature range
Dust and waterproof IP 67, NEMA 6 (when mounted on Converter) Polycarbonate, glass reinforced Transparent polycarbonate White backlit LCD-display (64 x 128 pixels) with softkeys Indication measurement, configuration and graph Real-time clock with built-in battery backup MODBUS RTU-mode, 9600 baud, 2-wire RS 485, master-mode RS 485 256 Kb Flash memory, 20.000 entries with date, time and value USB 1,1 type mini B, Female - 20 60 C
Brown
+DC
GND
White B
Grey
Pink
display unit
display unit
Shield
GND
www.mjk.com
4
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Line
Data Sheet
Specifications
Sensor
Cable Cable length Response time Flow velocity Power supply Current consumption Output Accuracy (turbidity) Accuracy (susp. solids) Temperature range Enclosure rating Weight Approvals
1)
3 x 2 x 0.34 mm2, outer diameter 8.3 mm 10 m (customized lengths, optional 2) 1 second No limit 24 V DC (15-30 V DC) Approx. 45 mA RS 485, 9600 baud, 2-wire Better than 3% of actual concentration Better than 5% of actual concentration (depends on calibration and media) 0 - 60 C short term to 80 C with wiper 0 to 50 C IP 68 to IEC 529 (10 m) / 10 bar pipe mounted, NEMA 6x 1,3 kg CE: EN61010-1, EN61326-1 ATEX: Ex II 3G
4000 upwards FNU internal manufacturer norm. 2 ) Can be delivered with other cable lengths on request.
Mechanical Dimensions
Immersion Fitting
SuSix Immersion fitting order no. 155215 Wall bracket for extension Extension for universal bracket Universal bracket
1000 mm 240 mm
Sensor
1 ISO RG
40 mm
475 mm
SuSix sensor
49 mm
Insertion Fitting
380 mm
SuSix sensor
Service position
www.mjk.com
5
DK: +45 45 56 06 56 NO: +47 69 20 60 70 SE: +46 53 31 77 50 NL: +31 251 672171 USA: +1 847 482 8655 AUS: +61 3 9758 8533
Data Sheet
Mechanical Dimensions
Converter and Display Unit
Order Numbers
display unit
display unit
150 144
M20
Max. four MJK converters with Modbus can be connected to one display unit:
Max. four MJK converters with Modbus can be connected to one display unit:
Converter #1
Converter #4
Display unit
521268
Converter #1
Converter #4
Display unit
521268
Fuse 24V AC: 500mA T Fuse 115V AC: 125mA T Fuse 230V AC: 63mA T
Grey
Brown
Pink
Shield
+DC GND A
Brown Grey Pink
B
Shield
display unit
Line
display unit
521269
NO C1
-DI-
N01 C2 N02
-D02-
PE
N
Mains supply
-Display-
-mA-
-D01-
521269
NO C1
-DI-
N01 C2 N02
-D02-
PE
N
Mains supply
-Display-
-mA-
-D01-
Line
Fuse 24V AC: 500mA T Fuse 115V AC: 125mA T Fuse 230V AC: 63mA T
SuSix Converters 206304 206305 206306 206307 206308 206309 207930 207935 SuSix Converter w/ display, 10 - 30 V DC SuSix Converter w/ display, 115 / 230 V AC SuSix Converter w/ display, 24 V AC SuSix Converter without display, 10 - 30 V DC SuSix Converter without display, 115 / 230 V AC SuSix Converter without display, 24 VA C Wall mounting kit / Junction box for sensor cable Panel mounting bracket
Accessories for SuSix Sensors 206350 206355 206360 691120 Wiper kit (5 pcs.) Wiper repair kit Gasket pulling/insertion tool Cable for SuSix sensor (PUR)
Fittings for SuSix Sensors 155210 155215 SuSix Insertion fitting SuSix Immersion fitting
Accessories for SuSix Sensor Fittings 200205 200210 200215 Universal bracket Wall bracket for extension Extension for universal bracket
Distributed by:
MJK offices
The Netherlands North America www.mjk.com www.mjk.com www.mjk.com DK: +45 45 56 06 56 mjknl@mjk.com mjkusa@mjk.com NO: +31 251 672171 +47 69 20 60 70 +1 847 482 8655 SE: +46 53 31 77 50 Sweden Australia NL: +31 251 672171 www.mjk.se www.mjk.com USA: +1 847 482 8655 kontoret@mjk.se mjk@mjk.com AUS: +61 3 9758 8533 +46 53 31 77 50 +61 3 9758 8533