Technical Developments

David F. Kallmes, MD Ferdinand K. Hui, MD John P. Mugler III, PhD

Index terms: Artifact Brain, MR, 10.12149 Cerebrospinal uid, 167.12149 Magnetic resonance (MR), pulse sequences, 10.12149 Published online: September 18, 2001 10.1148/radiol.2211001712 Radiology 2001; 221:251255 Abbreviations: CSF cerebrospinal uid FLAIR uid-attenuated inversion recovery SAS subarachnoid space 3D three-dimensional 2D two-dimensional
1

Suppression of Cerebrospinal Fluid and Blood Flow Artifacts in FLAIR MR Imaging with a Single-Slab Three-dimensional Pulse Sequence: Initial Experience1
The authors compared high-signalintensity ow-related artifacts present with a conventional two-dimensional (2D) uid-attenuated inversion recovery (FLAIR) sequence with those seen with a single-slab, three-dimensional (3D) FLAIR sequence. Four readers graded the subarachnoid space and intraventricular artifacts, the pulsation artifacts, and the conspicuity of cranial nerves in the posterior fossa. For all comparisons, differences between 2D and 3D images were highly statistically signicant, with 3D imaging being superior in all cases.
resulting from CSF-related artifacts with high signal intensity. Virtually all investigators who evaluated the accuracy of FLAIR imaging have noted these highsignal-intensity artifacts, which are present within both the SAS and ventricles (1,6,7). These artifacts, which are most pronounced in the posterior fossa, almost certainly result from the inow of CSF that was originally outside the region affected by the inversion pulse. Although several investigators have attempted to diminish these artifacts by increasing the width of the inversion slab, none were successful in their eradication (1,7,8). Image quality in the posterior fossa is often further degraded by pulsatile motion of nonsuppressed CSF or blood in the major vessels, which results in artifactual signal intensities overlying the anatomy of interest. Our group has developed a three-dimensional (3D), single-slab, FLAIR pulse sequence with high spatial resolution. Because this 3D pulse sequence uses nonspatially selective radio-frequency pulses, including the inversion pulse that affects the entire imaging slab, we hypothesized that high-signal-intensity artifacts from inow of noninverted CSF, as well as artifacts from pulsatile motion, would be diminished or eradicated. The purpose of our study was to test this hypothesis.

From the Department of Radiology, University of Virginia Health System, East Hospital, Rm 1063, Lee St, Box 800170, Charlottesville, VA 22908. Received October 25, 2000; revision requested December 15; revision received March 7, 2001; accepted March 29. Supported in part by grant NS35142 from the National Institute of Neurological Disorders and Stroke and by Siemens Medical Systems. Address correspondence to D.F.K. (e-mail: dfk3b @virginia.edu). RSNA, 2001

Author contributions: Guarantor of integrity of entire study, D.F.K.; study concepts, D.F.K., J.P.M.; study design, D.F.K.; literature research, D.F.K.; clinical studies, D.F.K., F.K.H.; data acquisition, D.F.K., F.K.H.; data analysis/interpretation, D.F.K.; statistical analysis, D.F.K., F.K.H.; manuscript preparation, D.F.K., J.P.M.; manuscript denition of intellectual content, D.F.K.; manuscript editing, J.P.M.; manuscript revision/review, D.F.K., J.P.M.; manuscript nal version approval, D.F.K.

Magnetic resonance (MR) imaging with a uid-attenuated inversion recovery (FLAIR) sequence demonstrates outstanding lesion conspicuity in numerous central nervous system disorders (1). Heavy T2 weighting combined with nulled cerebrospinal uid (CSF) signal results in high contrast-tonoise ratios for long-T2 lesions in the brain parenchyma (2). FLAIR imaging offers diagnostic capabilities beyond the evaluation of the brain parenchyma, including detailed information about the subarachnoid space (SAS). Whereas standard spinecho sequences show poor sensitivity for SAS processes, FLAIR imaging has proved highly sensitive not only for detecting subarachnoid hemorrhage but also for detecting inltrative processes of the leptomeninges (35). The promise of high sensitivity in the detection of SAS lesions with FLAIR has been mitigated by suboptimal specicity

Materials and Methods


Image Acquisition With institutional review board approval, seven healthy subjects (age range, 3139 years; mean age, 36 years) underwent brain MR imaging with a 1.5-T unit (Magnetom Symphony; Siemens Medical
251

Figure 1. Subject 1. (a) Transverse 2D FLAIR image (9,000/105/2,450) at the level of the fourth ventricle demonstrates severe inow-related high signal intensity within the fourth ventricle (short straight arrow) and completely lling the prepontine cistern (curved arrow). Pulsation artifact is present adjacent to the fourth ventricle (long straight arrows). (b) Transverse 3D FLAIR image (9,000/328/2,450) obtained at the same level as a. Inow-related artifact is absent in the fourth ventricle, and pulsation artifact is not seen. The absence of CSF-inow artifact in the prepontine cistern permits clear identication of the left cranial nerve V (arrow).

Systems, Iselin, NJ). Each subject underwent axial two-dimensional (2D) FLAIR imaging with a pulse sequence supplied by the manufacturer and axial 3D FLAIR imaging with our custom pulse sequence. The subjects gave oral informed consent. The order of the pulse sequences was varied randomly among subjects. Imaging parameters for the 2D FLAIR pulse sequence were as follows: 9,000/105/2,450 (repetition time msec/effective echo time msec/inversion time msec), 17.3 23-cm eld of view, 161 256 matrix, 5-mmthick sections, and an echo train length of seven. Two concatenated sequences were performed, which yielded 37 contiguous sections that covered the entire brain in 7.2 minutes. Imaging parameters for the single-slab 3D FLAIR pulse sequence were as follows: 9,000/328/2,450, in-plane eld of view of 17.3 23 cm, in-plane matrix of 160 256, 3-mmthick sections, echo train length of 160, 40% oversampling in the section direction (to suppress aliasing), half-Fourier acquisition in the section direction, and an acquisition time of 8.0 minutes. Conventional sinc interpolation was used
252 Radiology October 2001

in the section direction, which yielded 132 sections at 1.5-mm intervals. The 3D FLAIR pulse sequence included a very long spin-echo train with variable-ipangle refocusing radio-frequency pulses that permitted a decreased acquisition time while providing image contrast similar to that with conventional T2weighted spin-echo imaging. The structure of the single-slab 3D pulse sequence and the very long spin-echo train are described in detail elsewhere (9,10). The images were printed on lm for review. Filming was performed by an experienced technologist, who chose window and level parameters appropriate for FLAIR imaging in routine clinical practice. Image Interpretation Four readers interpreted the images. Three readers had subspecialty training in neuroradiology, and one was a radiology resident. Images were interpreted independently and in random order. One of the subspecialty-trained neuroradiologists (D.F.K.) was aware of the study de-

sign but remained blinded to sequence type. The other three readers, to avoid bias, were told only that the pulse sequences were intended to yield dark CSF; they were blinded to pulse sequence type, hypothesis of the study, number of pulse sequences being compared, number of study subjects, and all identifying subject information. With a ve-point graded response scale, the readers rated the degree of high-signal-intensity CSF artifacts in the fourth ventricle and at four locations within the SAS (upper cervical, prepontine, perimesencephalic, and suprasellar cisterns) as follows: grade 1, diffuse high signal intensity; grade 2, majority ( 50%) of cistern lled with high signal intensity; grade 3, minority ( 50%) of cistern lled with high signal intensity; grade 4, equivocal areas of high signal intensity; and grade 5, absence of high signal intensity. In addition, readers identied individual cranial nerves where possible. Last, the presence and severity of artifacts from vessels and/or CSF pulsation were graded by using an analogous scale (1 worst, 5 best). The ratings of high-signal-intensity CSF artiKallmes et al

Figure 2. Subject 3. (a) Transverse 2D FLAIR image (9,000/105/2,450) at the level of the cerebral aqueduct demonstrates inow-related high-signal-intensity artifact within and around the aqueduct (curved arrow). The cistern ventral to the mesencephalon is lled with inow-related artifact (straight arrow). (b) Transverse 3D FLAIR image (9,000/328/2,450) obtained at the same level as a. Inow-related artifacts are absent in the cerebral aqueduct and in the cistern ventral to the mesencephalon, which enables clear identication of cranial nerve III bilaterally (arrows).

facts and pulsation artifacts were compared between the 2D and 3D images by using Wilcoxon rank-sum analysis. The numbers of identiable cranial nerves were also compared between the two groups by using similar statistical tests. Interreader variability was evaluated by using a statistic, where agreement was dened as readings between two readers within one category on the ve-point graded response curve. The degree of agreement was considered to be almost perfect (0.8 1.0), substantial (0.6 0.79), moderate (0.4 0.59), fair (0.2 0.39), slight (0 0.19), or poor ( 0) (11).

High-Signal-Intensity Artifacts in the SAS The mean ratings for artifacts on 2D images were 1.2, 1.3, 2.1, and 2.9 in the upper cervical, prepontine, suprasellar, and perimesencephalic cisterns, respectively (Figs 13). The mean ratings on 3D images for these same cisterns were 4.8, 4.4, 4.3, and 4.7, respectively. In all cisterns for all readers, ratings for 3D FLAIR were superior or equivalent to those for 2D FLAIR (P .005 for all locations). Vascular and CSF Pulsation Artifacts The mean ratings for pulsation artifacts were 1.8 and 4.7 for 2D and 3D imaging, respectively (Fig 1). In all subjects for all readers, ratings for 3D FLAIR were superior or equivalent to those for 2D FLAIR (P .005). Cranial Nerve Identication The mean numbers of cranial nerves condently identied were 1.3 and 3.8 for 2D and 3D imaging, respectively (P .006) (Figs 1, 2). The third, fth, and VII/

VIII complex were the most frequently identied cranial nerves. Interreader Variability statistics ranged from 0.67 to 0.74 between various readers, which is indicative of substantial interreader agreement (11).

Discussion
Our data indicate that high-signal-intensity CSF-related artifacts in the posterior fossa within the SAS and fourth ventricle, as well as pulsation-induced artifacts, are signicantly diminished when using a single-slab 3D, rather than conventional 2D, FLAIR pulse sequence. Indeed, such artifacts were essentially eradicated with the 3D technique in our healthy volunteers. The exquisite CSF suppression throughout the posterior fossa, along with the absence of pulsation artifacts, enabled identication of individual cranial nerves in most subjects, which offers further evidence as to the imaging detail provided with the 3D FLAIR pulse se253

Results
High-Signal-Intensity Intraventricular Artifacts The mean ratings for the high-signalintensity CSF artifacts in the fourth ventricle were 2.1 and 4.7 for 2D and 3D imaging, respectively (Fig 1). In all subjects for all readers, ratings for 3D FLAIR were superior to those for 2D FLAIR (P .005).
Volume 221 Number 1

Suppression of Cerebrospinal Fluid and Blood Flow Artifacts

Figure 3. Subject 6. (a) Transverse 2D FLAIR image (9,000/105/2,450) at the level of C1. The SAS surrounding the upper cervical spinal cord is completely lled with inow-related artifact (arrows). (b) Transverse 3D FLAIR image (9,000/328/2,450) obtained at the same level as a. Flow-related artifact is absent in the SAS surrounding the upper cervical spinal cord. A dorsal rootlet (arrowhead) is clearly identied traversing the SAS.

quence. Our results suggest that this 3D FLAIR technique may help overcome the most persistent pitfall limiting the wider applicability of FLAIR imaging for SAS disease, that of CSF-related artifacts. More generally, the single-slab 3D FLAIR technique offers the advantages commonly associated with true 3D acquisition: thin, contiguous sections that result in reduced partial volume averaging of small structures compared with the thicker sections typically used in 2D acquisitions; the capability, with appropriate acquisition parameters, to obtain high-spatialresolution images in arbitrary planes; and potentially higher signal-to-noise performance per unit time than that with a comparable 2D pulse sequence. Other investigators have attempted modications to the 2D FLAIR pulse sequence in hopes of reducing CSF-related artifacts. The most common technique reported includes widening of the inversion region, on the order of 50% thicker than the imaging section, to diminish inow of noninverted magnetization (1,7,8). The use of wider inversion regions, however, does not necessarily eradicate the artifact. Bakshi et al (1), using a 2D FLAIR pulse sequence optimized for lesion conspicuity, imaging time, and artifact suppression, reported CSF-related artifacts with high signal intensity in the fourth ventricle in 50% of subjects. Conversely, in our study, high signal inten254 Radiology October 2001

sity within the fourth ventricle was absent in all subjects with our single-slab 3D FLAIR pulse sequence. Note that the use of a wider inversion pulse is, in effect, similar to the application of spatial presaturation slabs, which are commonly used for suppressing the signal from blood ow, on both sides of the image section. The usefulness of this approach is inherently limited by the effect of the wide inversion pulse on adjacent image sections. Tanaka et al (12) recently reported on the use of a nonspatially selective inversion pulse with 2D FLAIR. With this approach, the inversion time is different for each section. To address this issue, all sections were acquired twice, once in ascending order and once in descending order, and then averaged to achieve approximately the same contrast properties for each section. Because a nonselective inversion pulse is used, the degree of artifact suppression should be comparable to that with our single-slab 3D technique. With the 2D technique, however, the image contrast varies somewhat between sections. In addition, the total time over which sections can be acquired following each inversion pulse must be limited to maintain acceptable contrast in the averaged images. This results in a relatively low anatomic coverage per unit time; in the study by Tanaka et al (12),

only 11 sections were acquired in 4.5 minutes. Other researchers have investigated the use of a multislab 3D FLAIR pulse sequence for central nervous system imaging (2,13,14). Similar to our single-slab technique, the multislab implementation offers the advantage, compared with 2D techniques, of thin contiguous sections with a clinically acceptable signal-to-noise ratio. Nonetheless, substantial artifacts may still occur in regions of large CSF pulsation such as the posterior fossa (14). Furthermore, multislab imaging possesses several important limitations compared with a single-slab approach. Because of section prole effects, some of the outer sections in each slab are usually discarded, thus decreasing the efciency. Power deposition is relatively high and may compromise the coverage obtained per unit time, particularly at elds of 1.5 T and above. Each slab also experiences unwanted magnetization-transfer effects from the large number of high-ip-angle off-resonance radio-frequency pulses applied to the other slabs during the acquisition (15). Suppression of the CSF signal on T2weighted images may also be achieved without the use of the inversion recovery based approach of FLAIR. Essig et al (16) recently demonstrated a technique in which CSF-suppressed images are obtained by subtracting an image with a
Kallmes et al

very long echo time from an image with a moderate echo time; both images were obtained within the same acquisition by using a fast spin-echo pulse sequence. Because an inversion pulse is not used, this technique may provide increased contrast for certain lesions, such as those in multiple sclerosis, for which relatively poor conspicuity in the posterior fossa has been reported with FLAIR imaging (17,18). Nonetheless, to our knowledge, an investigation of the artifact characteristics for this dual-echo technique in the posterior fossa has yet to be presented. Thus, we cannot judge how it is likely to compare with our 3D approach for those processes that are well seen with FLAIR imaging. Our series, although useful for demonstrating the potential benets of singleslab 3D FLAIR imaging for suppressing high-signal-intensity CSF and pulsation artifacts, has several shortcomings. Only a small number of healthy volunteers underwent imaging. Even the small number of subjects, however, was sufcient for demonstrating a statistically signicant reduction in artifacts with 3D versus 2D imaging. In addition, our volunteers were relatively young, and intraventricular artifact may increase with increasing age as CSF ow rates increase. Relatively severe artifacts, however, were seen even in the young volunteers in our series with 2D imaging, and these artifacts were markedly reduced with 3D imaging. The 8-minute acquisition time used with our single-slab 3D FLAIR technique is several minutes longer than that for many pulse sequences routinely used for clinical MR imaging at 1.5 T. Thus, although our method appears to provide excellent suppression of high-signal-intensity CSF and pulsation artifacts, patient movement and subsequent image degradation during the relatively long imaging time is a concern. With our current pulse sequence, we use nonspatially selective radio-frequency pulses for inversion, excitation, and refocusing of the magnetization. These nonselective pulses, combined with radio-frequency transmission with the body coil, which is standard for head imaging with our MR system, resulted in aliasing (wraparound artifact) in the section direction unless the imaging slab covered the complete sensitive volume of the head coil. Therefore, for this study, the total imaging slab thickness was approximately 50% greater than that required to cover the anatomy of interest. We are developing an implementation of 3D FLAIR that uses a selective excitation radio-frequency pulse, which would enable the slab thickness to be

matched to the volume of interest. This would permit the acquisition time to be decreased from 8 to approximately 5 minutes. Nonselective pulses would still be used for inversion and refocusing, thereby maintaining artifact-suppression and sampling-efciency characteristics equivalent to those of our current technique. Another potential criticism of our study is that we acquired 5-mm-thick sections with the 2D pulse sequence and 3-mm-thick sections with the 3D technique. The 5-mm section thickness used for the 2D pulse sequence was chosen to match that routinely used at our institution in the clinical environment. The 3-mm section thickness chosen for the 3D technique represented a tradeoff between shortening the acquisition time, which favors thicker sections, and limiting truncation artifacts in the section direction to an acceptable level, which favors thinner sections. In 3D acquisitions, truncation artifacts in the section direction become pronounced when the section thickness approaches the characteristic dimension of structures along that direction (19). We acknowledge that the thinner sections used for 3D versus 2D images biased our cranial nerve evaluation in favor of the 3D pulse sequence, although the capability for obtaining thin, contiguous sections with clinically adequate signal-to-noise ratios is an advantage for the 3D method. Finally, although we have shown that the single-slab 3D FLAIR technique is useful for reducing artifacts, we have not demonstrated its ability to reveal abnormalities in the SAS or ventricles. It is possible that the 3D FLAIR pulse sequence may be less sensitive to these abnormalities than is the 2D FLAIR technique. Further studies are required to address this issue.
Acknowledgments: We thank John M. Christopher, RTR(MR), for valuable assistance with the data collection, and Clarence B. Yates, MD, Stephen F. Futterer, MD, and Robb E. Hoehlein, MD, for image interpretation.

4.

5.

6.

7.

8.

9.

10.

11. 12.

13. 14.

15. 16.

References
1. Bakshi R, Caruthers SD, Janardhan V, Wasay M. Intraventricular CSF pulsation artifact on fast uid-attenuated inversion-recovery MR images: analysis of 100 consecutive normal studies. AJNR Am J Neuroradiol 2000; 21: 503508. 2. Wieshmann UC, Barker GJ, Symms MR, Bartlett PA, Stevens JM, Shorvon SD. Fast uidattenuated inversion-recovery imaging: rst experience with a 3D version in epilepsy. Neuroradiology 1998; 40:483 489. 3. Noguchi K, Ogawa T, Inugami A, et al. Acute subarachnoid hemorrhage: MR imaging with

17.

18.

19.

uid-attenuated inversion recovery pulse sequences. Radiology 1995; 196:773777. Noguchi K, Ogawa T, Seto H, et al. Subacute and chronic subarachnoid hemorrhage: diagnosis with uid-attenuated inversion-recovery MR imaging. Radiology 1997; 203:257 262. Singer MB, Atlas SW, Drayer BP. Subarachnoid space disease: diagnosis with uid-attenuated inversion-recovery MR imaging and comparison with gadolinium-enhanced spin-echo MR imaging blinded reader study. Radiology 1998; 208:417 422. Bakshi R, Kamran S, Kinkel PR, et al. Fluidattenuated inversion-recovery MR imaging in acute and subacute cerebral intraventricular hemorrhage. AJNR Am J Neuroradiol 1999; 20:629 636. Jack CRJ, Rydberg CH, Krecke KN, et al. Mesial temporal sclerosis: diagnosis with uidattenuated inversion-recovery versus spinecho MR imaging. Radiology 1996; 199:367 373. Hashemi RH, Bradley WGJ, Chen DY, et al. Suspected multiple sclerosis: MR imaging with a thin-section fast FLAIR pulse sequence. Radiology 1995; 196:505510. Mugler JP III, Bao S, Mulkern RV, et al. Optimized single-slab three-dimensional spinecho MR imaging of the brain. Radiology 2000; 216:891 899. Mugler JP III, Kiefer B, Brookeman JR. Threedimensional T2-weighted imaging of the brain using very long spin-echo trains (abstr). In: Proceedings of the Eighth Meeting of the International Society for Magnetic Resonance in Medicine. Berkeley, Calif: International Society for Magnetic Resonance in Medicine, 2000; 1630. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977; 33:159 174. Tanaka N, Abe T, Kojima K, Nishimura H, Hayabuchi N. Applicability and advantages of ow artifactinsensitive uid-attenuated inversion-recovery MR sequences for imaging the posterior fossa. AJNR Am J Neuroradiol 2000; 21:10951098. Barker GJB. 3D fast FLAIR: a CSF-nulled 3D fast spin-echo pulse sequence. Magn Reson Imaging 1998; 16:715720. Tubridy N, Barker GJ, Macmanus DG, Moseley IF, Miller DH. Three-dimensional fast uid attenuated inversion recovery (3D fast FLAIR): a new MRI sequence which increases the detectable cerebral lesion load in multiple sclerosis. Br J Radiol 1998; 71:840 845. Melki PS, Mulkern RV. Magnetization transfer effects in multislice RARE sequences. Magn Reson Med 1992; 24:189 195. Essig M, Deimling M, Hawighorst H, Debus J, van Kaick G. Assessment of cerebral gliomas by a new dark uid sequence, high intensity reduction (HIRE): a preliminary study. J Magn Reson Imaging 2000; 11:506 517. Tubridy N, Barker GJ, MacManus DG, Moseley IF, Miller DH. Optimisation of unenhanced MRI for detection of lesions in multiple sclerosis: a comparison of ve pulse sequences with variable slice thickness. Neuroradiology 1998; 40:293297. Okuda T, Korogi Y, Shigematsu Y, et al. Brain lesions: when should uid-attenuated inversion-recovery sequences be used in MR evaluation? Radiology 1999; 212:793798. Carlson J, Crooks L, Ortendahl D, Kramer DM, Kaufman L. Signal-to-noise ratio and section thickness in two-dimensional versus three-dimensional Fourier transform MR imaging. Radiology 1988; 166:266 270.

Volume 221

Number 1

Suppression of Cerebrospinal Fluid and Blood Flow Artifacts

255