Manual for Good Bioavailability and

Bioequivalence Practices

Volume II

Braslia 2002

Rights reserved for Brazilian Sanitary Surveillance Agency SEPN 515, Edifcio mega, Bloco B, Braslia (DF), CEP 70770-502. Internet: www.anvisa.gov.br Copyright 2002. Brazilian Sanitary Surveillance Agency. The total or partial reproduction of this work is permitted provided that the source is mentioned. First edition - 2002 ISBN: 85-88233-06-1 Brazilian Sanitary Surveillance Agency Performed by: Inspection Coordination in Bioequivalence Centers General Office of Inspection and Control of Drugs and Products General Coordination: Cludia Franklin de Oliveira / Inspection Coordination in Bioequivalence Centers Revision: Karla de Arajo Ferreira / Inspection Coordination in Bioequivalence Centers Divulgation: Divulgation Unit Covers: Joo Carlos de Souza Machado / Multimedia Communication Office Graphic Art, composition and printing: Dupligrfica Editora Ltda./DF Printed in Brazil

Manual de boas prticas em biodisponibilidade: bioequivalncia/Agncia Nacional de Vigilncia Sanitria. Gerncia-Geral de Inspeo e Controle de Medicamentos e Produtos. Braslia: ANVISA, 2002. 2 v. QV38 1. Equivalncia teraputica. 2. Bioequivalncia. 3. Disponibilidade biolgica. 4 . Medicamentos. I. Agncia Nacional de Vigilncia Sanitria. Gerncia-Geral de Inspeo e Controle de Medicamentos e Produtos.

PRESENTATION

The Brazilian Sanitary Surveillance Agency, embedded of its institucional mission, elaborated the present work with the intention to foment discussion, education and the conduction of clinical studies in Brazil. Concerning the bioavailability/bioequivalence studies, this compendium is part of a bigger context which is intended to enclose the clinical research in its amplest aspect: non clinical studies and clinical studies of phase I, phase II, phase III and IV. In presenting this work , The Brazilian Sanitary Surveillance Agency would thank the contribution of all the professionals directly and indirectly involved in its consolidation, and it is available for critics and suggestions that may contribute for future editions, as well as, for the subsequent modules.

Dr. Gonzalo Vecina Neto

PREFACE

A number of clinical trials are being presently conducted in Brazil aiming to evaluate the Bioavailability/Bioequivalence of pharmaceutical products. As of June 2001, Anvisa, through the Inspection Coordination in Bioequivalence Centers bound to the General Office of Inspection and Control of Drugs and Products started evaluating such centers by means of periodical inspections in order to assure the quality of the trials. During the inspection activities, originally for orientation purposes, the coordination observed the need to clarify some aspects that still posed technical questions to the centers, especially concerning the standardization of analytical methods, statistical analysis of the trials, storage of biological samples, confinement of volunteers and stability studies of drugs and others. When such requirement was identified and in an attempt to avoid impairing the quality of the works carried out, the coordination stimulated the creation of discussion nuclei intended to wear out the explanation of all aspects related to the conduction of the studies and the integration of their phases. Such nuclei counted on the participation of 40 experts from the Pharmaceuticals, Medicine, Statistics and Chemistry fields. This Manual for Good Bioavailability/Bioequivalence Practices was created under such context and comprises six main topics presented in a didactic manner, trying to overwhelm the centers difficulties and, as a result, complement the guidelines provided for in the Brazilian health legislation for conducting trials. The working nuclei started the discussions in September 2001. Each nucleus was dedicated to one of the three phases of the process Clinical, Analytical and Statistical Phases under the coordination of Dra. Cludia Franklin de Oliveira and the inestimable cooperation of professor Slvia Storpirts. During the subsequent months, several meetings were held so as to promote technical debates and reach a consensus concerning theses questions. Consequently, each group prepared a set of relevant topics to be included in the manual. All such topics were then subjected to careful researches and the results reported so as to allow a good understanding by the target population. The complete manual took 11 months to be finished. This process involved the participation of about 50 professionals in total from several fields of knowledge, including researchers from public universities, technicians from Anvisa and manufacturers of laboratory instruments and equipment. Anvisa is thankful to all such associates for having contributed to such an excellent job. The final structure comprises two volumes with three modules each. The first volume technically details each phase of the Bioavailability/Bioequivalence studies in the natural sequence of conduction: Module 1: Clinical Phase, Module 2: Analytical Phase and Module 3: Statistical Phase. The second volume covers the important aspects concerning laboratory instrumentation and equipment used to carry out the analytical phase and regarded as critical for the process. In that volume, Module 1 relates to the Principles and Operation of Micropipettes, Module 2 refers to Water for Chemical

Instrumental Analyses and the contents of Module 3 relates to the Ultraviolet Visible Spectrophotometry, Liquid Chromatography (LC), Gas Chromatography (GC), Chromatography Systems Connected to Mass Detectors and Checking of Analytical Instruments Performance. It is worth emphasizing the international novelty character of gathering information from several subjects in a single compendium aiming at synthesizing all the aspects involving the Good Bioavailability/Bioequivalence Practices. It is evident that the major purpose of this work is to improve the quality of the Bioavailability/ Bioequivalence assays carried out in Brazil and, as a result, contribute to the quality of generic drugs available in the market by providing widely studied and carefully prepared technical subsides. With this in view, we hope to contribute to capacitating the Bioavailability/Bioequivalence Centers, in addition to develop study monitors for the domestic pharmaceutical industry and help in the formation of knowledge multiplier agents at the Brazilian universities. The preparation of this manual was only possible thanks to the crucial work of several persons. I would like to apologize in advance for possibly forgetting any name. The editors Jos Pedrazzoli Jnior (USF/Unifag) and Joo Antnio Saraiva Fittipaldi (Pfizer) were essential, as well as the cooperators Fernanda Maria Villaa Boueri (Anvisa), Eliana Regina Marques Zlochevsky (Anvisa), Cludia Simone Costa da Cunha (Ministry of Health) and Beatriz Helena Carvalho Tess (Ministry of Health), for the module Clinical Phase; the editors Cludia Franklin de Oliveira (GGIMP/Anvisa), Rui Oliveira Macedo (UFPA), Flvio Leite (T&E Analtica) and Pedro Eduardo Froehlich (UFRGS), and the cooperators Pedro de Lima Filho (GGMEG/SP), Davi Pereira de Santana (UFPE), Rafael Eliseo Barrientos Astigarraga (Cartesius), Silvana Calafatti de Castro (Unifag), Thas Reis Machado, Jaime Oliveira Ilha (Cartesius), Itapuan Abimael Silva (Anvisa), Karen Noffs Brisolla (Anvisa), Marcelo Cludio Pereira (Anvisa), for the module Analytical Phase; the editors Arminda Lucia Siqueira (UFMG), Chang Chiann (GGMEG/SP), Cicilia Yuko Wada (Unicamp), Karla de Arajo Ferreira (Anvisa) and Gilberto Bernasconi (USF/Unifag), and the cooperators Reinaldo Charnet (Unicamp) and Renato Almeida Lopes (Anvisa), for the module Statistical Phase; the editors Melissa M. Silva (Nova Analtica) and Walter Pereira (Nova Analtica), Principles and Operation of Micropipettes; the editor Jos Muradian Filho (Millipore), Water for Clinical Analyses; the editors Ivan Jonaitis (Agilent), Renato Garcia Peres (Flowscience), Ricardo Lira (Flowscience), Renato Gouveia, Jos Aparecido Soares (Varian), Josu D.M. Neto (Sync Brazil) Juarez Arajo Filho (Sync Brazil), Alexandre Rosolia (Waters), Adauto Silva (Varian) and Reinaldo Castanheira (Agilent), Analytical Instrumentation; and the coordination team composed by Cludia Franklin de Oliveira (GGIMP/ Anvisa), Marcelo Cludio Pereira (GGIMP/Anvisa), Max Weber Marques Pereira (GGIMP/Anvisa), Karla de Arajo Ferreira (GGIMP/Anvisa), Karen Noffs Brisolla (GGIMP/Anvisa), Itapuan Abimael da Silva (GGIMP/ Anvisa) and Renato Almeida Lopes (GGIMP/Anvisa). Dr. Gonzalo Vecina Neto

TECHNICAL STAFF

Volume II Module 1 Micropipettes: Principles and Operation Editors: Melissa M. Silva Nova Analtica Valter A. Pereira - Nova Analtica Coordination: Cludia Franklin de Oliveira ANVISA Itapuan Abimael da Silva - ANVISA Karen de Aquino Noffs Brisolla - ANVISA Karla de Arajo Ferreira - ANVISA Marcelo Cludio Pereira - ANVISA Max Weber Marques Pereira - ANVISA Renato Almeida Lopes ANVISA

SUMMARY

1 . DEFINITION ................................................................................................................................... 5 1.1. Types of pipettes ........................................................................................................................... 5 2. OPERATION PRINCIPLE .............................................................................................................. 6 2.1. Air displacement pipettes ............................................................................................................. 6 2.2. Positive displacement pipettes ..................................................................................................... 8 3. OPERATION TECHNIQUE ......................................................................................................... 11 3.1. How to store the pipette ........................................................................................................... 11 3.2. Volume adjustment ..................................................................................................................... 11 3.3. Tip-fitting ..................................................................................................................................... 12 3.4. Pre-rinsing of the tip .................................................................................................................. 13 3.5. Aspiration ..................................................................................................................................... 13 3.6. Dispensing.................................................................................................................................... 14 3.7. Tips ejection ................................................................................................................................. 14 4. TIPS ...................................................................................................................................................... 15 4.1. Tips for air displacement pipettes ............................................................................................. 15 4.2. Tips for positive displacement pipettes.................................................................................... 15 5. MAINTENANCE ............................................................................................................................. 16 5.1. Cleaning ........................................................................................................................................ 16 5.2. Parts replacement ........................................................................................................................ 17 5.3. Performance checking ................................................................................................................ 20 6. SUGGESTIONS OF PROCEDURE ............................................................................................ 21 7. ATTACHMENT I: EXAMPLE OF A CLEANING PROCEDURE ...................................... 22 7.1. Disassembling .............................................................................................................................. 22 7.2. Cleaning ........................................................................................................................................ 23 7.3. Assembling ................................................................................................................................... 23 8. ATTACHMENT II: PERFORMANCE STANDARD VERIFICATION .............................. 24 8.1. General ......................................................................................................................................... 24 8.2. Testing room conditions ............................................................................................................ 24 8.3. Operator ....................................................................................................................................... 25 8.4. Tips ............................................................................................................................................... 25 8.5. Equipment used in the test ........................................................................................................ 25 8.6. The checking procedure ............................................................................................................. 27

9. ATTACHMENT III: DECONTAMINATION PROCEDURES ............................................ 38 9.1. Chemistry ..................................................................................................................................... 39 9.2. Microbiology and cell culture .................................................................................................... 40 9.3. Molecular biology ....................................................................................................................... 41 9.4. Methods spectrum of action ..................................................................................................... 44 9.5. Advantages and disadvantages .................................................................................................. 45 9.6. Autoclaving .................................................................................................................................. 46 9.7. UV irradiation .............................................................................................................................. 46 9.8. Chemical solutions ...................................................................................................................... 46 10. ATTACHMENT IV: DEFINITIONS ........................................................................................... 50 11. GLOSSARY AND ABBREVIATIONS ........................................................................................ 51 12. REFERENCES .................................................................................................................................. 52

Volume II / Module 1 - Micropipettes: Principles and Operation

1. DEFINITION
Piston-operated pipettes are equipment used to aspirate and dispense specific volumes of liquids. The single-channel pipettes have only one piston. The multichannel pipettes have simultaneously several receptacles. The pipettes can be factory-set to dispense a certain dado volume or volumes selected by the user within a specific volume range, for example, between 10 L and 100 L. Pistonoperated pipettes can be of two types: air displacement and positive displacement.

1.1. Types of pipettes The pipettes can be designed as follows: Concerning the volume: Fixed volume, designed by the manufacturer to dispense only its nominal volume, for example, 100L. Varying volume, designed by the manufacturer to dispense volumes selected by the user within a specific volume range. Concerning the piston: There may be an air layer between the piston and the liquid surface (air displacement pipettes). The piston is in direct contact with the liquid (positive displacement or direct displacement pipettes). The capillary and the piston may be reusable or disposable.

5 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

2. OPERATION PRINCIPLE
The plastic or glass tip is coupled to the pipette. With the piston at the lower limit (lower aspiration position), dip the tip into the liquid to be transferred. When the piston moves up to the upper aspiration position, the liquid is drawn. The volume of liquid is dispensed when the piston is pushed again to the lower position. The operation principle is detailed below according to the type of pipette (air displacement or positive displacement).

2.1. Air displacement pipettes When the button of an air displacement pipette is pressed, the piston located inside the equipment moves down and displaces the air in contact with it out of the pipette (the air volume expelled is equal to the volume set in the pipette). The air volume remaining inside the pipette is inversely proportional to the sample volume: the lower is the air volume inside the pipette, the greater is the volume of liquid to be drawn. a) Volume adjustment (valid for the varying volume pipettes only) The user adjusts the desired volume. The piston moves to the appropriate position. Remarks: In fixed volume pipettes, this positioning is provided at the factory.

Selected volume, for example, 100 mL

Piston Piston position after volume adjustment Cone holder

Zone filled with air Tip

6 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

b) Preparation for aspiration The button is pressed. The piston move down and expels an air volume equal to the selected volume (in fixed volume pipettes, the volume of air expelled is equal to the nominal volume of the pipette).

The piston moves to the lower limit

Volume of air expelled (equal to the selected volume)

c) Liquid aspiration Dip the tip into the liquid until the bore is submerged. When releasing the button, the piston returns to the initial position and a partial vacuum is formed inside the pipette. The atmospheric pressure forces the entrance of liquid through the tip bore.

The piston returns to the initial position Partial vacuum is formed

Tip bore submerged into the liquid

Volume of liquid drawn

7 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

d) Liquid dispensing The button is pressed again. The piston moves down displacing the air and increases the pipette inner pressure. The compressed air expels the liquid out of the tip.

Destination container

e) Second stage In some air displacement pipettes, the piston can be displaced just below the lower limit (thus displacing an additional amount of air), allowing the expulsion of the last drops of liquid. The most common term employed by users for this additional displacement is second stage.

2.2. Positive displacement pipettes The positive displacement pipettes work as a syringe. There is no air volume between the piston and the liquid. Since there is no air to contract or expand, the aspiration force is always constant e remains unchanged by the physical properties of the liquid to be handled. This kind of equipment is ideal for handling viscous or high-density liquids.

8 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

a) Volume adjustment The user adjusts the desired volume. The piston moves down to the initial position.

Selected volume, for example, 100 L

Cone holder

Capillary Piston

Initial position

b) Preparation for aspiration The button is pressed. The piston moves down to the end of the capillary.

9 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

c) Liquid aspiration The capillary bore is submerged into the liquid. When releasing the button, the piston moves up and the room pressure forces the entrance of the liquid through the capillary bore.

Piston at the initial position

Volume drawn Submerged capillary

d) Liquid dispensing The button is pressed again. The piston moves down and expels the liquid out of the capillary.

Destination container Dispensed volume

10 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

3. OPERATION TECHNIQUE
A good pipetting technique is essential to assure good results. A script on some aspects of the technique to be considered during the operation is showed next. 3.1. How to store the pipette The pipette should be stored in the upright position for two reasons: Prevents it from contacting the surfaces that can be contaminated, for example, the working bench (when laying the pipette on the bench or in a chest of drawers, contaminants can be transferred to the body of the instrument and subsequently transferred to the experiment). If the tips are not ejected and if a residual amount of liquid remains inside the tips, this liquid may flow into the instrument when the pipette is laid down, thus contaminating and damaging the inner parts and/or the cone holder. Examples of supports:

3.2. Volume adjustment In order to avoid a parallax error, the volume should be adjusted with the pipette in the horizontal position. If the operator prefers to adjust the volume with the pipette in the upright position, one of the eyes should be closed. When reducing the volume, carefully reach the desired amount and do not surpass the mark. When increasing the volume, surpass the desired amount by 1/3 of a cycle and then carefully reduced the volume until the desired amount is reached. Do not surpass the mark.

11 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

3.3. Tip-fitting The tip is fitted according to the manufacturers instructions, thus assuring an adequate seal between the pipette tip and the cone holder. Remark: Seal means an opaque ring formed around the entire tip collar, indicating a perfect connection between the cone holder and the tip. An incomplete seal will lead to imprecise results. Some suggestions for fitting the tips are given below: Loose tips: Hold the tip by its collar (be careful not to touch the end) and fit it to the cone holder. Press the tip against the cone holder by following a rotational movement, as indicated in the figure below. This assures a perfect connection between the tip and the cone holder (the above mentioned seal is formed).

Tips on racks: Fit the cone holder on the tip and press the pipette downwards by following a rotational movement as indicated in the figure.

12 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Multichannel fitting: For a simultaneous fit on all channels, position the cone holders on the tip collars. Make a toand-from movement as showed in the figure, thus assuring an adequate connection of all tips on all channels.

3.4. Pre-rinsing of the tip When a new tip is placed (or when increasing the volume to be drawn), it is necessary to prerinse the tip. To do so, just drawn and dispense the liquid a couple of times. The pre-rinsing of the new tip assures accuracy and precision of the volume to be subsequently transferred. This is because a film is formed on the tip inner wall when drawing a liquid. The nature of this film, that causes an error in the first measurement, depends on the liquid being transferred. However, this film remains relatively constant after some pipetting operations using the same tip. It is necessary to pre-rinse the tip in order to maximize the pipette performance.

3.5. Aspiration The aspiration should be slow and constant, always keeping the same immersion depth of the tip (during the aspiration and between the samples). It is recommended that the immersion depth is around 2 to 3 mm below the liquid surface. Some manufacturers suggest that the immersion depth should vary according to the working volume, as described in the table below:

13 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

The pipette should be kept in the upright position during the complete aspiration. Wait for 1 to 2 seconds before removing the tip from the liquid. If liquid droplets still remain outside the tip, clean them using soft paper (if the methodology allows). Be careful not to touch the tip bore.

3.6. Dispensing Touch the tip end on the inner wall of the container and tilt the pipette approximately 30 to 45; Continuously press the button until the end of the first stage. Wait some seconds (from 1 to 3 depending on the liquid viscosity) and then press the button until the second stage (purging) to eliminate droplets that may have remained on the tip; Keep the button pressed until the end. Remove the pipette from the container keeping the tip in touch with the container inner wall (scratch the tip end on the inner wall of the container).

3.7. Tips ejection Dispose the tip by pressing the tips ejector button; Take all required precautions when working with contaminated and/or radioactive materials. The appropriate procedure should be defined by the laboratory according to the kind of material pipetted, taking the safety of the users and the working place into consideration.

14 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

4. TIPS
The quality of the results is directly affected by the quality of the tip used. Good results require not only good quality tips, but also tips that are more appropriate to the model and brand of the pipette used.

4.1. Tips for air displacement pipettes The manufacturer should indicate which brand of tip the user should employ to comply with the specifications indicated in the pipette operation manual. The plastic tips for air displacement pipettes should be used disposed after use. Such tips should not be cleaned or reused, since their metrological features cannot be assured.

4.2. Tips for positive displacement pipettes The manufacturer should indicate which brand of tip the user should employ to comply with the specifications indicated in the pipette operation manual. The fitting between the piston and the capillary (perfect sealing between the capillary and the piston), as well as the piston displacement inside the capillary, should be optimum to assure an even dispensation of the liquid drawn. These tips can be reused or disposed (when changing the tip, the capillary and the piston should be simultaneously changed).

15 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

5. MAINTENANCE
This chapter will cover three essential steps of the maintenance procedure: cleaning, change of parts and performance checking. REMARK It is necessary to adapt the information of this chapter to the manufacturers instructions manual. Particularly, the following aspects should be regarded: Checkout whether the parts listed herein can be washed; Assembling and disassembling procedure of the pipette; How to change damaged parts; Checkout whether any of these procedures alter the pipette calibration.

If they do change, it is necessary to verify the instrument performance (see attachment II) e occasionally adjust the equipment (according to the manufacturers instructions). 5.1. Cleaning It is necessary to create a cleaning and inspection routine of the pipette parts. So that the best results are always reached, the pipette parts have to be in good operation conditions. There are two different situations requiring different cleaning procedures: a) Standard cleaning The aerosols formed during pipetting settle inside the pipette, thus impairing its normal operation. In addition to clog the cone holder bore and impair the air displacement inside the pipette, this settlement may damage internal parts, for example, the piston (leading to scratches or corrosion). In both cases, the pipette will not work in compliance with the specifications. In order to avoid this situation, the pipette internal parts (cone holder, ejector, piston, seal and oring) should be cleaned. Follow the procedure recommended by the manufacturer. Attachment I describes an example of a cleaning procedure.

Cleaning is the main tool to allow users to extent the pipette life and assure reliable results. There is no cleaning schedule to be adopted by the laboratory (this will depend on the type of material handled, the routine to which the pipette is subjected, number of users of the same pipette and acceptable errors for the procedure). So as to assure reliable results, the pipette should be subjected to performance testing in addition to be frequently cleaned (attachment II).

16 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

b) Pipette used for handling acid or corrosive materials The piston conditions directly interfere with the pipette specifications. If the piston is scratched or corroded, the accuracy and precision standards are not met. Acid materials (for example, sulfuric acid, hydrochloric acid, nitric acid, phenol) or corrosive materials (for example sodium hydroxide, chlorinated solvents) may damage the piston. A damaged piston leads to unreliable results. In order to avoid piston damage, it is recommended to clean the pipette parts right after handling this kind of liquid. Contrary to the procedure described in item a, this cleaning does not require the use of a chemical agent. Distilled water should be enough. Disassemble the pipette according to the manufacturers instructions. Pour distilled water on the following parts: ejector, cone holder, piston and piston seals. Let it dry at room temperature or in an oven up to 40. Assemble the pipette according to the manufacturers instructions.

REMARKS Certify that said parts can be washed and whether this procedure will affect its calibration (that is, if it is necessary to follow the performance checking procedure right after cleaning). The procedure described in item b above is particularly indicated for pipettes with metallic piston. For pistons made of other materials (glass, plastic, ceramic) the use of silicone grease for sealing purposes is common. Thus, it is recommended to verify and follow the specific procedures contained in the manual.

5.2. Parts replacement Some parts of the pipette have to be periodically replaced, since they wear out and directly interfere with the pipette performance. 5.2.1. Piston seals This part or parts provide a sealing between the piston and the cone holder (upper portion). With use, microvilli are formed on the surface of these parts and the sealing becomes imperfect, thus impairing the instrument precision. Such parts should be replaced once a year (see the replacement procedure in the manufacturers manual). 5.2.2. Cone holder Another important sealing point is the contact zone between the cone holder and the tip. With the use, the cone holder surface wears out on the tip-fitting portion and the contact between them becomes inadequate and impairs the measurement (loss of precision). Replacement is required
17 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

whenever the tip-fitting portion is worn out or in case of part damage (broken, scraped). Verify in the manufacturers manual whether the procedure affects the instrument calibration. 5.2.3. Other parts Other parts interfere with the performance pipette. A very simple checking procedure is described below to help recognizing the damaged parts. Some procedures should be followed according to the manufacturers manual (for example: disassembling, assembling). We emphasize again the importance of certifying whether it is necessary to recalibrate the pipette after one of these procedures (this information can be found in the manufacturers manual). 1st STEP: GENERAL ASPECT

Operations bar

Button Ejector button

Volumeter Tips ejector

Cone holder

Tip

Verify that there is no apparent defect.

18 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

2nd STEP: FUNCTIONS TESTING Turn the volumeter fully so as to go through the complete volumetric range of the pipette. Verify whether the minimum and maximum adjustable volumes match with the actual volumetric range of the pipette.

Problems and causes: Adjustment is not possible Incorrect volume adjustment Autoclaved pipette * Incorrect adjustment or assemble of the volumeter

* Some pipettes can be autoclaved, others cant. See the manufacturers manual.

With the volumeter set at the maximum value specified for the pipette model, press the button fully to verify the easiness of movement, feel occasional obstructions or friction variations that may cause damages, such as piston corrosion or crooked operations bar. Listen to the spring noise that may indicate its incorrect positioning.

Problems and causes: Uneven movement Lack of displacement Inappropriate movement Damage of the friction ring The operations bar is crooked Corroded, dirty or scratched piston

Fit one tip on the pipette and press the tips ejector button to verify the ejection efficiency. Lack of movement Undue fit of the tip Difficult removal Damage of the ejection mechanism Ejector misplacement Corrosion

3rd STEP: LEAKAGE TESTING For pipettes having a maximum volume above 200 l, set the volume at the maximum value and draw some water to fill the pipette tip. Observe the tip for 20 seconds. A drop at the end of the tip means the occurrence of leakage. This is also true for pipettes with volume ranges below 200 L; for this volume range, however, the following procedure should be adopted: after observing the presence of a drop for 20 seconds, dip the tip end into the water in the container. The level of water inside the tip should not be changed. The presence of leakages is confirmed if the level decreases. The cause of leakages can be: Damaged piston seals; Scratched or corroded piston;

19 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Chemical or mechanical damages on the cone holder at the tip-fitting portion; Use of tips not recommended by the manufacturer (see chapter 4).

4th STEP: DISASSEMBLING This process should be adopted in case of any leakage in step 3. When disassembling the pipette, one can evaluate the internal parts and find out the cause of the problem (damaged piston, lack or impairment of the seals, etc.) and define which parts should be replaced. An example of disassembling routine adopted for certain models pipettes is showed below. Be sure to follow the manufacturers instructions to disassemble and assemble the pipette. Pull down and remove the tips ejector (in some models, the ejector is not detachable); Remove the connection nut; Carefully remove the piston; Verify the piston surface (whether it is corroded and/or scratched) as well as the seals.

5th STEP: ASSEMBLING Assemble the pipette according to the manufacturers instructions. REMARK We recommend steps 1 to 3 to be carried out daily before starting the work and recorded for the laboratory control. If abnormalities are found, carry out steps 4 and 5. Some parts can be changed by the user itself without compromising the instrument calibration. After the change of some other parts, it is necessary to check the instrument calibration and occasionally adjust it. This information is contained in the pipette instructions manual.

5.3. Performance checking In addition to the parts cleaning and replacement procedures, the user has to verify the pipette performance in order to assure reliable measurements. Such checking analyses both the accuracy (systematic error) and the precision (random error). The performance checking, in fact, evaluates the set pipette + tip + operator (pipetting technique). So as to reach reliable testing results, the operator should be trained and skilled (pipetting technique described in chapter 3) and the tips should be those recommended by the pipette manufacturer. The user should define a schedule for testing the pipettes based on: requirements in terms of accuracy and precision, frequency of use, number of users of a same pipette, number of cycles in which the pipette is used and the nature of the liquids pipetted. In attachment II, we described the gravimetric method for evaluating the pipettes performance (reference: rule ISO/FDIS 8655- Pistonoperated volumetric apparatus - Part 6 - Gravimetric test methods).
20 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

6. SUGGESTIONS OF PROCEDURE
Piston-operated pipettes are equipment for drawing and dispensing specific volumes of liquids. In order to assure the reliability of the results, it is necessary to implement the following routine: Checking of the general aspect, functions testing and leakage testing (steps 1, 2 and 3) described in item 5.2.3. herein. If abnormalities are found out, follow steps 4 and 5 of the same procedure. Periodic cleaning of the pipette parts The cleaning periodicity should be established according to the frequency of use, number of users of a same pipette, number of cycles in which the pipette is used and the nature of the liquids pipetted. We recommend laboratories to follow the cleaning procedure every three months (if dirt accumulates within these three months, reduce the intervals between each cleaning procedure). It is very important for the laboratory to record the cleaning procedures; said records should contain the pipette serial number, cleaning date and procedure used. Parts replacement The piston seals should be replaced annually (if the number of pipetting cycles exceed 100,000, the seals should be replaced within a shorter period of time to be estimated by the laboratory routine). The cone holder should be replaced every 2 years or when the tipfitting portion is worn out. Such replacements should be recorded (recording should contain the pipette serial number, replacement date and information on the new part - lot or bill of sale number). Parts affecting the pipette performance (for example: piston, volumeter, rod) requiring replacement or repair should be forwarded to specialized technical assistance; the repair work should be duly recorded, in addition to the new calibration of the pipette. Periodically check the pipette performance (accuracy and precision) as described in attachment II. Such verification should be made every 3 months and duly recorded (recording example in attachment II item 8.6.5. herein). The maximum errors found can be up to twice higher than the errors defined in tables 1 and 2 herein (attachment II, item 8.6.4.). When replacing parts that affect the pipette calibration, the checking procedure should be followed (see item parts replacement above). Operator testing Since the operation technique is crucial for obtaining good results, it is necessary to test and record the operators pipetting technique. For that, using an equipment operating within the manufacturers specifications, carry out 10 measurements with the maximum e minimum volume specified for that pipette model. Assess the accuracy and precision of the measurements obtained. The maximum errors found should be up to twice the values specified in tables 1 and 2 herein (attachment II, item 8.6.4.). Record data in a report (an example is provided in attachment II item 8.6.5.) identifying the operator tested.
21 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

7. ATTACHMENT I: EXAMPLE OF A CLEANING PROCEDURE


The example below can be used for various pipette models, particularly those having a metallic piston with no grease; for other models, an adaptation is required. Follow the manufacturers instructions contained in the users manual. Be sure to certify that the steps described herein do not impair the pipette calibration.

Warning! Do not mix parts of a pipette with parts of another pipette

7.1. Disassembling It is necessary to disassemble the pipette for cleaning (wear gloves when contamination risk is involved). A general scheme of the pipettes parts is showed below:

LIST OF PARTS Figures 1 and 2 A B C D E F G and H Button Handle Cone holder Tips ejector Tip Connection nut Piston seals

Remarks: This scheme may have variations according to the model and brand of the pipette. For further details, see the manufacturers manual.
22 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Remove the tips ejector (D) by pulling it; Disconnect the connection nut (F) thus separating the pipette body (B) from the cone holder (C); Remove the piston set carefully and separate the piston seals (G and H), keeping the cone holder with the finest portion downwards. The piston seals are frequently captured inside the cone holder. To remove them, place the piston again and press it against the cone holder as if you are pipetting. When removing the piston, the seals will be attached to it.

7.2. Cleaning Place the cone holder, the piston set, the tips ejector, the seals (do not place the handle) in a beaker and complete to volume with 4% to 8% neutral detergent solution with warm water ( 50C); Place the beaker in an ultrasound bath for about 15 minutes (if an ultrasound is not available, keep the beaker in the solution for 40 minutes); Dispose the detergent solution used; Wash each part with running water; Place the parts in a clean container and wash for the last time with distilled water; Dry it in an oven at approximately 50C (do not exceed this temperature) for at most 2 hours, or allow it to dry at room temperature. REMARKS Use a 4% neutral detergent solution for cleaning slightly dirty pipettes and a 8% solution for heavily dirty pipettes. For those pipettes still dirty after washing, use cotton wetted with isopropyl alcohol. Be careful not to damage the piston. After using the pipette in acid and/or corrosive solutions, always wash all parts with distilled water or ethanol.

7.3. Assembling Place the seal and the oring on the piston by first fitting the seal; Using one of the hands, hold the pipette body in the upright position and place the piston; Fit the cone holder on the pipette body; Next, place and thread the connection nut; Place the tips ejector; The pipette is ready for use.

23 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

8. ATTACHMENT II: PERFORMANCE STANDARD VERIFICATION


8.1. General This attachment describes the pipettes performance checking procedure using the gravimetric method (according to rule ISO/ FDIS 8655 - Piston-operated volumetric apparatus - Part 6: Gravimetric tests). The use of the procedures described in this method assures compliance with the international specifications for accuracy and precision. The compliance test for checking accuracy and precision is applicable to any pipette (definitions in chapter 1). This test is in according to international standards, analyzes the pipetting system as a whole: pipette, tip and operator. The method described herein includes a procedure for correcting the evaporation losses for small volumes (below 50 L). Further, when converting to the volume of masses obtained in the balance, corrections are made concerning the temperature and pressure at the time of testing. The user should define a routine for testing the pipettes based on: requirements in terms of accuracy and precision, frequency of use, number of users of a same pipette, number of cycles in which the pipette is used and the nature of the liquids pipetted. There is no standardized interval between each checking. It is the user responsibility to define such an interval. Rule ISO/ FDIS 8655 defines a minimum periodicity of once a year. The pipette should be disassembled and assembled (according to the manufacturers manual instructions) at least once a year before the checking test. The pipette should be handled according to the manufacturers manual.

8.2. Testing room conditions The testing room (laboratory) should be cleaned and the temperature and humidity controlled so that the conditions in the room where verification will take place and the equipment temperature are stable and homogeneous both before and throughout the procedure. The pipette and water used in the gravimetric test should be stabilized in the checking room temperature at least 2 hours before starting the procedure. Ideally, the verification should take place under the following conditions: 1) Stable temperature (water, pipette and room) between 15 and 30C; 2) Relative Humidity above 50%.

24 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

8.3. Operator 8.3.1. Pipette operation An adequate pipetting procedure significantly contributes for the reproducibility of the verification test results. Inexperienced operators may have significant fluctuations in the results. For reliable results, the operator should be duly trained and qualified. The pipetting technique aspects were discussed in chapter 3. 8.3.2. Training A qualified operator should take the compliance test. If required, contact a specialized company to request the operators training program.

8.4. Tips The pipette to be tested should be handled according to the instructions provided in the operations manual. Again, based on the pipettes operation manual, certify that the pipette is clean, correctly assembled and that the tips recommended by the manufacturer are used before starting the test. See chapter 4 for further details.

8.5. Equipment used in the test To assure the checking procedure integrity, all equipment used during the procedure should be regularly verified. 8.5.1. Balance The balances should be calibrated and certified by skilled personnel using weights certified by domestic authorities (Companies/Institutions accredited by INMETRO and members of Rede Brasileira de Calibrao - RBC).

The balance legibility should be chosen according to the volume selected in the pipette being tested.

25 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

In practice, the nominal volume (highest volume selectable by the user) can be used for choosing the balance. 8.5.2. Other equipment Thermometer: maximum measurement uncertainty of 0.2C Hygrometer: maximum measurement uncertainty of 10% Barometer: maximum measurement uncertainty of 0.5 kPa 8.5.3. Containers Containers used during the procedure: Origin container The container holding the water to be used in measurements (the volume of such a container should be enough for ALL measurements). Weighing container The container to be used for measuring (to be placed on the balance weighing plate). Disposition container A third container for collecting the aliquots not weighted. Especially for testing the lowest volume of the pipette, we recommend an average ratio of 3:1 to the container diameter and height for weighing purposes or that the container is provided with a lid. Some companies supply container kits appropriate for the gravimetric test. 8.5.4. Water The test requires distilled or deionized water. In both cases, the water should be degasified (grade 3 water according to Rule ISO 3696) at the temperature of the room where verification will take place. In order to avoid water temperature fluctuations, use an origin container big enough to hold the water for all measurements. 8.6. The checking procedure The performance checking analyzes both the accuracy (systematic error) and precision (random error). The conditions, procedures and qualifications previously described herein should be implemented in order to assure the validity of the tests results. After rinsing the tip, perform 10 individual weighing runs* for each selected volume. For the varying volume pipettes, three different volumes should be selected according to the volume range of the model being evaluated. Such
* O user can select the number of measurements according to the accuracy and precision specifications accepted by its methodology.

26 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

volumes should be the nominal volume (maximum), 50% of the nominal volume (intermediate volume) and the minimum volume of the model volume range (these volumes are defined in the pipettes manual). For the fixed volume pipettes, only the nominal volume is employed. 1. 2. 3. 4. 5. Adjust the testing volume in the pipette (this volume should be kept throughout the procedure). Estimate the evaporation rate (for small volumes, below 50L). See the procedure below. Run the checking test: record the masses on the Performance Checking Report. Calculate: record the results on the Performance Checking Report. Compare the results with the accuracy and precision specifications showed in tables 1 and 2 (item 8.6.4).

These steps will be detailed later. 8.6.1. Estimating the evaporation rate (average mass evaporated per weighing cycle) For volumes below 50L, it is necessary to use a lidded container. The purpose is to minimize, control and quantify the evaporation loss during the weighing cycle. The use of clamps for handling this container is also advisable. The evaporation can be estimated by a sequence of four weighing simulations where the weighing cycle is repeated without dispensing water into the weighing container. The total difference attributed to the evaporation is calculated and divided by 4 so as to obtain the average. The range is expressed in mg/cycle (in case of a single cycle, it should be expressed in mg). Excessive handling of the container should be avoided. The use of clamps for handling the weighing container is advisable. The evaporation rates are usually between 0.010 mg and 0.025 mg per weighing cycle. Recalculate the evaporation rate whenever the conditions are changed (temperature, pressure e humidity). 1. 2. 3. 4. 5. 6. 7. 8. 9. Add water to the weighing container up to 1/3 of its volume. Place the capped container on the balance plate. Using the pipette, draw the aliquot from the origin container. Tare the balance and remove the weighing container from the plate (the use of clamps for removal is advisable). Dispense the aliquot into the disposal container or return it to the origin container. Do not dispense into the weighing container. Record the resulted e1. Repeat steps 3 to 8 three times so as to obtain e2, e3 and e4. Calculate the loss/cycle: e = (e1 + e2 + e3 + e4)/4 (mg). The evaporation per cycle e (mg) should be added to the average weight before calculating the average volume.

* O usurio pode selecionar o nmero de medidas de acordo com as especificaes de exatido e preciso aceitas pela sua metodologia.

27 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

An example calculation of the evaporation rate is showed at the end of this chapter. 8.6.2. Gravimetric test The aspiration and dispensing techniques are described in chapter 3. Although several alternative pipetting methods are possible, always follow the normal mode (direct mode, do not use the reverse mode). For all pipettes, avoid holding or heating the cone holder using your hand during the test. For multichannel pipettes, fit the tips in every channel; however, test each channel individually. Record the results for each channel. 1. Preparation 1.1. Transfer the water from the origin container to the weighing container at a depth of at least 3 mm. 1.2. Measure and record the origin container water temperature (t1), the air pressure and relative humidity of the room. If the weighing container has a lid, place it. 1.3. Fit the new tip. 1.4. Select and adjust the volume to be tested (this volume cannot be changed during the test). 1.5. Pre-rinse the tip by drawing an aliquot from the origin container and dispensing it into the disposition container 5 times to balance the air humidity inside the pipette atmosphere. 1.6. Place the weighing container with water on the balance plate and tare (m0 = 0). 2. Testing cycle Each testing cycle should take less than 1 minute. However, a constant rhythm during the weighting operation should be kept (the cycle rhythm and the rhythm between the cycles). 2.1. Place a new tip in the pipette. 2.2. Pre-rinse the tip by drawing and dispensing into the disposition container once. 2.3. Draw the volume to be tested as specified in chapter 3. 2.4. If the weighing container has a lid, remove it. 2.5. Dispense the volume tested in the weighing container and replace the lid with the aid of the clamps. 2.6. Record the mass m1 of the volume tested. 2.7. Tare the balance. 3. Repeat the test described above (from item 2.1 to 2.7) up to 10 measurements (m1 to m10). 4. After the last measurement, check and record the origin container water temperature (t2), the atmospheric pressure and the relative humidity of the room.

28 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

8.6.3. Calculations 1. Average temperature calculation Calculate the average temperature (t) of the distilled water (maximum fluctuation between t1 and t2: 0.2C) 2. Average pressure calculation Use the average barometric pressure (B) and the average temperature (t) to determine the corresponding Z factor in the table. 3. Volume calculation from the mass After the evaporation losses corrections (for volumes below 50 L), multiply the masses (mg) by the Z factor to obtain the volumes (L). Vi= Z (mi + e) Vi mi e Z 4. Average volume calculation After calculating the weightings individual volume, calculate the average volume (result expressed in L). = individual volumes (L) = individual masses (mg) = evaporation rate (mg) = Z factor (L/ mg) B = (B1 + B2)/ 2 t = (t1 + t2)/ 2

Vi= individual volumes V = average volume n= number of weightings 5. Accuracy (systematic error) The difference between the test average volume and the volume set in the pipette. For fixed volume pipettes, replace Vs with Vo = nominal volume. The accuracy can be expressed in mL or ... ...as percentage: es = V - Vs es = average error V = average volume Vs = adjusted volume es = 100% (V - Vs)/Vs

29 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

6. Precision (random error) Dispersion of the volumes dispensed around the dispensed volumes average. Also known (according to the context) as standard deviation, reproducibility or repeatability.

SD

Vi = individual volumes V = average volume (calculated as in the previous item) n = number of measurements SD = standard deviation As percentage, also known as coefficient of variation (CV). CV = (SD / V)x 100%

30 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Calculations: Z factor Z = conversion factor (mL/ mg), t (C) = temperature average * , B = atmospheric pressure (hPa)

* this average is obtained from the origin container water temperature measured at the beginning of the test (t1) and the origin container water temperature measured at the end of the test (t2).

31 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

8.6.4. Specifications TABLE 1 This specifications table applies to the air displacement fixed volume pipettes and positive displacement pipettes using reusable capillaries and pistons. This table can be also used for the air displacement varying volume pipettes. The nominal volume in these pipettes is the greatest volume selectable by the user and specified by the manufacturer. For example: a pipette with a specified volume range from 10 L to 100 L, the nominal volume is 100 L. The maximum acceptable errors for the nominal volume apply to all the volumes within the specified range: if we consider a pipette with a volume range from 10 L to 100 L, the maximum acceptable systematic error for any selected volume is 0,8 L and the maximum acceptable random error for any selected volume is 0,3 L.

For intermediate volumes to those listed in the table, the immediately greater absolute value of the volume maximum acceptable errors should apply. For example: to determine the maximum acceptable errors for 25 L, the values specified for 50 L are used.

32 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

TABLE 2 This specifications table applies to the positive displacement pipettes using reusable capillaries and pistons and/or fixed volume pipettes.

For intermediate volumes to those listed in the table, the immediately greater absolute value of the volume maximum acceptable errors should apply. REMARKS The user may define the maximum acceptable errors (systematic and random) based on its methodology. The user may regularly test the equipment, for example, every three months. Other time intervals between the tests can be established (provided that they do not exceed one year) based on the following aspects: frequency of use, number of users of the pipette, aggressive nature of the transferred material, maximum errors accepted by the user. When the test is carried out after the equipment maintenance or repair, the maximum acceptable errors (systematic and random) should follow the tables listed herein (tables 1 and 2). 8.6.5. Checking procedure report After the test, information has to be recorded. We attached a report template, however, the user may prepare his own template provided that it contains the following information: Pipette identification; Date; Test conditions (temperature, pressure, relative humidity); Measurements obtained for each volume tested; The calculations results; Result from the comparison with the specification table.

33 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

TEMPLATE:
Pipette identification Model Technician identification Company Technician Conditions Date/ Test time Temperature (C) Balance Model Humidity (%)

Serial number

Tips used in the no test

Name Name

Phone/Email Phone/Email

Location Pressure (hPa)

Serial number

Sensitivity

Test results

COMPARING TO THE SPECIFICATIONS:

APPROVED FAILED

34 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

8.6.6. Performance checking numeric example We will now give an example of the calculations for the performance checking. We will be based on a hypothetic verification of 10 L in an air displacement varying volume pipette. 1. Since the tested volume is lower than 50 L, the evaporation rate should be determined (item 8.6.1.). Number of measurements: 04 (e1, e2, e3, e4) Average evaporation rate: e e1 = 0.016 e2= 0.018 e3 = 0.021 e4 = 0.017

e = (e1 + e2 + e3 + e4)/4 (mg) e = (0.016 + 0.018 + 0.021 + 0.017)/4 e = 0.018 mg/ per cycle 2. The test initial temperature and pressure are recorded: Ti = 21.5C Pi = 1013 3. Ten measurements (item 8.6.2.) are performed and the results recorded. W1 = 9,84 W2 = 9,90 W3 = 9,91 W4 = 9,86 W5 = 9,87 W6 = 9,90 W7 = 9,92 W8 = 9,93 W9 = 9,95 W10 = 9,92

4. The test final temperature and pressure are recorded: Tf = 21,5C Pf = 1013 5. Average pressure and temperature calculation: T = (Ti + Tf) / 2 T = (21,5 + 21,5)/ 2 T = 21,5C P = (Pi + Pf) /2 P = (1013 + 1013) / 2 P = 1013

These temperature and pressure average results will be used for determining the conversion factor Z.

35 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

6. Conversion of mass to volume: Vi = (Wi +) x Z For a temperature of 21.5C and an atmospheric pressure of 1013 hPa, the Z factor is 1.0032 L/ mg.

7. Average volume calculation

V = (9.89 + 9.95 + 9.96 + 9.91 + 9.92 + 9.95 + 9.97 + 9.98 + 10.00 + 9.97)/10 V = 9.95 L 8. Accuracy analysis Systematic error - E E = V - Vo Vo is the value set in the pipette (nominal volume). In this example, it is 10L. E = 9.95 - 10 E = - 0.05 L Relative error - E% E% = (V - Vo) x 100/ Vo E% = (-0.05*100)/ 10 E% = - 0.50 % 9. Precision analysis (repeatability - random error) Standard deviation - SD

SD

SD

36 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

SD2v = 1/9 x [(9,89 - 9,95)2 + (9,95- 9,95)2 + (9,96 - 9,95)2 + (9,91 - 9,95)2 + (9,92 - 9,95)2 + (9,95 - 9,95)2 + (9,97 - 9,95)2 + (9,98 - 9,95)2 + (10,00 - 9,95)2 + (9,97 - 9,95)2 +] SDv = 0,03 L Coefficient of variation CV CV = (DP/ V)x 100% CV= (0,03/9,95) x 100% CV = 0,34 % 10. Results are transferred to the report (item 8.6.5.)

The following specifications for 10 L are showed in table 1 (item 8.6.4.):

Simply compare the results obtained (report) with those specified in the table. In this case, the tested pipette is approved.

37 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

9. ATTACHMENT III: DECONTAMINATION PROCEDURES


The procedures described next can be used for some pipette models, particularly those having a metallic piston. Thus, it is essential to read the manufacturers instructions before applying any of the methods described herein. INTRODUCTION This document describes a number of methods for pipette decontamination, especially those having a metallic piston. Such methods can be physical (autoclaving or UV irradiation) or chemical. According to the type of application, the most efficient decontamination method should be chosen. However, considering the action spectrum as well as the advantages and disadvantages of these methods, we conclude that the chemical is the most effective. Commercial chemical solutions containing detergents such as disinfectants, assure an efficient decontamination. The chemical decontamination method is recommend for routine decontamination procedures since it is effective, fast and easy.

Decontamination procedure
The types of contamination include: from sample to operator, from one sample to another or from the sample to the pipette. This can affect the safety of the operator, the sample and consequently the experiment result or eventually the pipette. There are several methods for eliminating contaminants; however, there are some doubts concerning its efficiency and compatibility with the pipettes. This document intends to solve doubt related to the decontamination process, taking the various applications of the pipettes into consideration. Thus, different fields of application will be commented (chemistry, microbiology and molecular biology) and the chemical or physical methods discussed. The efficiency of a method can be explained by means of its mechanism of action. When more than one method is proposed, there will be a comparison to help choosing the most appropriate. Once the decontamination method is chosen, read the protocols at the end of this chapter.

38 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Field of application

9.1. Chemistry By simply washing the pipettes with distilled water eliminates contaminations by acids, alkalis, buffer solutions or organic solvents. It may be necessary to remove other products such as oils and greases. Then, the use of detergents is required. Detergents (anionic surfactants) are organic molecules having a polar hydrophilic end and a non-polar hydrophobic end. Due to its structure, they convert insoluble residues (grease, for example) into soluble residues, therefore, becoming effective cleaning agents. 9.1.1. Standard cleaning procedure (for more details, see attachment I) This procedure should be used to eliminate contaminants resulting from the use of the pipette (for example: proteins and buffer solutions) and that do not require special treatment for elimination, such as microorganisms, radioactivity and other that will be discussed later. Remove the tips ejector by pulling it; Disconnect the connection nut thus separating the pipette body from the cone holder; Remove the piston set carefully and separate the seal(s), keeping the cone holder with the finest portion downwards; Place the cone holder, the piston set, the tips ejector, the seal(s) in a beaker (do not place the pipette body) and complete to volume with 4% to 8% detergent solution in warm water ( 50C); Place the beaker in an ultrasound bath for about 15 minutes (if an ultrasound is not available, keep the beaker in the solution for 40 minutes); Dispose the detergent solution used; The seals are frequently captured inside the cone holder. To remove them, place the piston again and press it against the cone holder as if you are pipetting. When removing the piston, the seals will be attached to it. Place the parts in a clean container and wash for the last time with distilled water; Dry it in an oven at approximately 50C (do not exceed this temperature) for at most 2 hours, or allow it to dry at room temperature. REMARKS Use a 4% detergent solution for cleaning slightly dirty pipettes and a 8% solution for heavily dirty pipettes. For those pipettes still dirty after washing, use cotton wetted with isopropyl alcohol. Be careful not to damage the piston.
39 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

After using the pipette in acid and/or corrosive solutions, always wash all parts with distilled water or ethanol. Assemble the pipette.

9.2. Microbiology and cell culture In these fields it is essential to eliminate several microorganisms (for example: virus, bacteria and/or fungi, such as yeasts). Several techniques with different spectra and mechanisms of action can be used. 9.2.1. Autoclaving The characteristics of the sterilization cycle depend on the initial amount of microorganisms (the so-called microbial load). It is generally accepted that the possibility of a viable organism to be present in an autoclaved item is below 1 in 1 million (Sterility Assurance Level 10-6). Aims Action Any vegetative cell or endospore 19. The kinetics leading to virus inactivation is still unclear 9. Heat degrades nucleic acids, enzymes and other essential proteins. The cell membrane may be damaged.

9.2.2. UV irradiation The UV wavelength is between 100 nm and 400 nm. There are 3 types of UV radiation, according to the wavelength: UVC (200 nm to 290 nm), UVB (290 nm to 320 nm) and UVA (320 nm to 400 nm). Aims Action Lethal for most of the microorganisms 19 (Table 2.1) Photochemical changes of cell enzymes (UVA) and DNA (UVC) - see details in table 1.3.2.

9.2.3. Chemical agents A chemical solution should be chosen taking into the consideration the type of microorganism to be eliminated.

40 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

The following table lists the main microorganisms and some common chemical agents. When special strains are handled, consideration should be taken regarding the specific commercial products used to destroy them and also certify the chemical compatibility of the pipette and the corresponding chemical (see 3.3).

Level of activity: +++: strong, ++: medium, +: weak, -: no activity. The mechanisms of action of such chemicals are described in the following table:

9.3. Molecular biology Biochemists and molecular biologists are familiar with the contamination sources such as contaminating proteases, nucleic acid (DNA, RNA), nucleases (RNases, DNases) and radioactivity. 9.3.1. Proteases Proteases are enzymes that degrade proteins. Several decontamination methods and their mechanisms of action are detailed in the table below:

41 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

9.3.2. Nucleic acids (DNA and RNA) The specificity and sensitivity of the PCR and RT-PCR techniques involve a disadvantage: easy contamination of reagents and samples, thus leading to systematic errors 7.

Other method: Glycine/HCl buffer (pH=2). The action of this solution is not documented (see the protocol later) Note: all such methods degrade the nucleic acids into molecules having lower molecular weights that are not normally amplified; however, such methods do not clean.

9.3.3. Nucleases Nucleases are enzymes that degrade nucleic acids. They have a wide distribution and are found in prokaryotic and eukaryotic cells, vegetal and animals (skin, water). DNases are not very resistant and are not essential for most of the applications in molecular biology, while the RNases are more durable and resistant to heat. DNases need metal ions for its activity and can be easily inactivated using some appropriate buffer solutions (such as chelating agents, for example, EDTA). RNases do not require metal ions, as they use the 2-hydroxyl group as reagent; therefore, they are not easy to inactivate.

42 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Method/Chemical agent Autoclaving

Action Not always enough to inactivate RNases that may recover their activity after several treatment methods (boiling, for example) Lack of data when reading. Alkylates residues of histidine from the catalytic sites of RNases A, thus inactivating them. This is a strong agent; however, it is not an absolute inhibitor.

UV irradiation DEPC (diethyl pyrocarbonate)

9.3.4. Comments on nucleic acids and nucleases All methods described partially degrade nucleic acids and nucleases, however, none of them showed 100% effectiveness. Precaution is the best way to avoid contamination in molecular biology: Separate the areas and sets of pipettes to prepare the samples and carry out the assays. Use tips provided with filters so as to avoid contamination by aerosols. Or use positive displacement pipettes provided with disposable capillaries and pistons (total protection against cross contamination). For critical PCR protocols, the use of positive displacement pipettes is required when no contamination is involved.

9.3.5. Radioactivity In case of weak radioactive contamination, specific detergents containing complexing agents can be used. Comparing the methods The methods are compared by first considering its efficiency in terms of contaminants and then the advantages and disadvantages.

9.4. Methods spectrum of action Consider two approaches: 1. I have the materials required for this method in my lab. How effective are they (read the table vertically)? 2. I need to eliminate this contaminant. What can I do (read the table horizontally)?

43 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Level of activity: +++: strong, ++: medium, +: weak, -: no activity, SI: not informed. *: specific detergents (with complexing agents). For UV irradiation: +: some activity, -: no activity. The UV irradiation activity limit is difficult to evaluate since its action do not depend on the wavelength only (between 100 nm and 400 nm), but also other factors such as: time of exposure, distance of the lamp, emission density, angle of exposure and atmospheric temperature.

9.5. Advantages and disadvantages See the table below. The method should comply with the following requirements: Compatibility with the pipette manufacturing materials. Fast action. Easy and safe to apply.

44 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

* Those made of PVDF. UV radiation resistant cone holders (made of PBT) are available for P200 and P1000. Thus, the chemical procedures can be considered the most appropriate for pipettes decontamination. Protocols According to the method used, different parts of the pipette can be decontaminated. Method Autoclaving UV irradiation Chemical solutions Portion of the pipette Cone holder, connection nut, tips ejector External portions Different levels, depending on the solution used (external portion, complete immersion)

Note: Wear disposable gloves throughout the decontamination procedure.

9.6. Autoclaving Remove the tips ejector and unthread the connection nut. Clean the cone holder, the connection nut and the tips ejector using detergent and then autoclave for 20 minutes at 121C, 0.1 Mpa. Allow the parts to dry or place them in an oven at 50 - 60C for about 30 minutes. Note: Do not autoclave such parts at 134C as they will be damaged. Do not autoclave parts other than those listed above.

45 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

9.7. UV irradiation UV lamps can be installed on the top of the bench or in a biosafety hood where handling is allowed. The combination of 254 nm and 300 nm bulbs for 20 minutes can destroy double strand DNA (long sequences are easier to inactivate) 7.

9.8. Chemical solutions Before using a chemical solution, be sure of the compatibility of the materials used to manufacture the pipettes. The manufacturer should provide a list of the materials used to manufacture the pipette. Some examples are showed below: PVDF, PBT, PC, PE, POM, nitrile rubber, stainless steel, ABS, PEI and PMP. See Glossary and Abbreviations.

Note: it is not advisable to soak the seal and the oring.


9.8.1. Cone holder, tips ejector and connection nut decontamination Two examples of chemical solutions and their protocols are showed below. Sodium hypochlorite Sodium hypochlorite has a wide antimicrobial activity, fast antibacterial action e also denatures proteases, DNA and RNA. In addition, it is an easy-to-use and low-cost chemical agent atoxic at the concentrations used. It is very important to clean the pipette (using a detergent, for example, Mucapur) before disinfecting using sodium hypochlorite, since the method may lose efficiency in the presence of high concentrations of organic material. Cleaning Remove the tips ejector, unthread the connection nut and remove the cone holder piston. Dilute the detergent with warm water at 50-60C in: a) A ultrasound bath or in b) A beaker. Place the cone holder, the tips ejector and the connection nut: a) In an ultrasound bath for 15 minutes. or b) In the beaker the parts should be brushed. Remove the parts and rinse well.

46 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Before disposing the cleaning solution, add 10% sodium hypochlorite to the solution and allow it to react for 10 minutes. This procedure will decontaminate the cleaning solution before entering the local water network. Disinfection Dilute the sodium hypochlorite in distilled water at a concentration of 10% and place it in a large beaker. Place the cone holder, the tips ejector and the connection nut in the beaker and allow it to react for 30 minutes. Rinse well with tap running water first and then with distilled water. Allow them to dry at 50C - 60C for about 30 minutes. Keep them at room temperature for 15 minutes before assembling the pipette (so that the parts cool down before assembling the pipette).

Note: the hypochlorite solution should be discarded after 3 days.

Glycine /HCl Buffer (pH2=2)


Prepare the solution as follows: 10x buffer: NaCl 30.6 g; glycine 39.2 g, complete with distilled water to the volume of 523 ml. Add 1N HCl in a sufficient amount to reach a final volume of 1000 mL. Remove o tips ejector, unthread the connection nut and remove the cone holder piston. Place the cone holder, the connection nut and the tips ejector into the 1x diluted buffer (1/10 dilution of the previously mentioned buffer) at 95C for 30 minutes. Rinse well with tap running water first and then with distilled water. Allow them to dry at 50C - 60C for 30 minutes. Keep them at room temperature for 15 minutes before assembling the pipette. 9.8.2. Decontamination of the pipette lower portion and body For most of the applications, decontamination of the pipette lower portion and body suffices. A mixture of chemical agents is recommended (detergent and disinfectant in a single product). The pipette body should be cleaned using a compatible solution (see item 3.3) on the surface only (do not immerse the pipette body on the decontaminating solution). The lower portion of the pipette, including the piston, the tips ejector, the cone holder and the connection nut (except for the piston seals) should be also cleaned (using small brushes to clean the cone holder interior) or can be immersed as per the following procedures:

47 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Remove the tips ejector, unthread the connection nut and remove the cone holder piston. Dilute the chemical agent chosen (according to the manufacturers instructions) in a beaker. When disassembled, place the lower portions (piston, tips ejector, cone holder and connection nut) in the beaker. Rinse well with tap running water first and then with distilled water. Dry at 50 C - 60 C for 30 minutes. Keep the parts at room temperature for 15 minutes before assembling the pipette. Conclusion The pipette decontamination method is selected according to the type of application. For most of the applications, just thoroughly decontaminate the lower portions and clean the pipette body using a chosen solution (see 3.3.2).

48 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Appendix 1
What about the disposable materials (tips, capillaries, pistons and Distritips)? Methods used for the sterilization of consumable plastics of the pipettes At the laboratory, autoclaving is the most common method for sterilizing the tips. However, this process takes time: the autoclaving time itself, the drying cycle and stabilization at room temperature for some hours.

Note: the autoclaving process should be carried out at 121C (20 minutes, 1 MPa). It is not recommended to autoclave at 134 C as plastics may be damaged.
Depending on the type of material, other methods are used by the manufacturer to sterilize consumables: ethylene oxide or more often irradiation (gamma rays: 60Co, or beta rays: electrons beam). As this manufacturing process is automated, including the final step, clean tips are generally produced with no manual contact. Method Ethylene oxide Irradiation Action Alkylates the amino terminus of amino acids, thus resulting in the cellular death 11. Ionization of important cell components (including amino acids) 11. Inactivation of DNA templates through free radicals.

Comments Ethylene oxide is highly toxic and mutagenic. Therefore, health and environmental control authorities in several countries have questioned its use. The irradiation used in the industrial sterilization processes does not render materials radioactive. The dosage is chosen based on the regular experimental determinations.

Appendix 2
Pyrogenic materials Contamination by pyrogenic materials can be troublesome in some pharmaceutical procedures in which medical devices are used in parenteral preparations (mentioned in various methods of the Pharmacopeia) 4. Pyrogens are endotoxins produced by Gram negative bacteria and are chemically and physically stable. Very little has been published on how to eliminate them. The use of gamma rays at high levels of energy (minimum of 25 kGy) seems to eliminate the risks of pyrogen contamination. In other words, a product has to be manufactured and controlled as a medical device so as to be certified as pyrogen-free.
49 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

10. ATTACHMENT IV: DEFINITIONS

Accuracy (average error or systematic error): the difference between the volume dispensed and the nominal volume or the volume selected in the pipette. Calibration: set of operations that establish the relation between the volume dispensed and the nominal volume or the corresponding selected volume in the pipette. Dead air volume: in air displacement pipettes, the volume of air between the lower portion of the piston and the liquid surface. Evaporation rate: estimated water loss caused by evaporation during the weighting process. Gravimetric analysis: general procedure based on the determination of the mass of the water aliquots dispensed by a pipette. The values are corrected concerning the evaporation loss; then, the actual mass and volume are calculated based on the water density at specific temperatures with corrections to the room atmospheric pressure (Z factor). Nominal volume: the greatest volume selectable by the user and specified by the manufacturer. Note: therefore, for a varying volume pipette with a useful volume range from 10 l to 100 l, its nominal volume is 100l. Selected volume: the volume set by the user for dispensing the chosen volume within the useful volume range. Note: for the fixed volume pipettes, the selected volume corresponds to the nominal volume. Precision (random error): dispersion of the volumes dispensed around the dispensed volumes average. Also known (according to the context) as standard deviation, reproducibility or repeatability. Repeatability: the dispersion between the results of successive measurements (carried out within a short period of time) with no change of parameters or conditions. Reproducibility: the dispersion between the results of measurements made under different conditions (different location or operator), the parameters being constant. Volumetric specifications (maximum acceptable errors): the allowed limits (upper and lower) for deviation between the dispensed volume and the nominal volume or the selected volume. Useful volume range: part of the nominal volume that can be dispensed within the maximum error specified in the International Rule ISO/DIS 8655. The upper value of the useful volume range is always the nominal volume. The lower valor is 10% of the nominal volume if not otherwise specified by the manufacturer. Recalibration (operator-adjusted): a procedure defined by the manufacturer that can be followed by the final user to assure that the pipette works according to the published specifications. Weighting: weigh means to determine the mass of an aliquot. Z factor: conversion factor (L/mg) considering the density of the water in contact with the air as a function of temperature and pressure. Ex: Designation (abbreviation) of an instrument designed to dispense volumes. In: Designation (abbreviation) of an instrument designed to draw volumes.
50 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

11. GLOSSARY AND ABREVIATIONS


Glossary Detergent: organic molecules that convert insoluble residues into soluble residues (oils and fats). They are effective cleaning agents (most of them do not kill microorganisms). Decontamination: disinfection of infected articles, thus allowing their handling /use (reduction of microorganisms to an acceptable level).

Note: chemical decontamination requires a previous cleaning process using a detergent so that the decontamination process can be effective.
Disinfection: selective elimination of some kinds of microorganisms so as to avoid transmission (reduction of the number of infecting microorganisms below the level required to cause an infection). Sterilization: complete elimination of all organisms (for example: cells, spores and virus). Abbreviations DNA: Deoxyribonucleic Acid EDTA: Ethylene Diamine Tetraacetic Acid RNA: Ribonucleic Acid Materials ABS: Styrene Butadiene Acrylonitrile PBT: PolyButylene Terephthalate PC: PolyCarbonate PE: PolyEthylene PEI: PolyEther Imide PMP: PolyMethylPentene POM: PolyOxyMethylene PP: PolyPropylene PVDF: PolyVinyliDene Fluoride

51 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

12. REFERENCES

Cimino, G.D., Metchette, K.C., Tessman, J.W., Hearst, J.E. and Isaacs, S.T. 1990. Post PCR sterilization: a method to control carryover contamination for the polymerase chain reaction. Nucleic Acids Research, 19 (1): 99-107. Deragon, J.M., Sinnet, D., Mitchell, G., Potier, M., Labuda, D. 1990. Use of irradiation to eliminate DNA contamination for PCR. Nucleic acids research, 18 (20): 6149. Ferron, A.. 1992. Spectre dactivit des antiseptiques et dsinfectants. Bactriologie mdicale ( lusage des tudiants en mdecine), Edition C et R. Gilson Guide to Pipetting prEN ISO/ FDIS 8655-1 - Piston-operated volumetric apparatus Guyomard, S., Goury, V., Laizier, J., Darbord, J.C. Defining of the pyrogenic assurances level (PAL) of irradiated medical devices. 1987. International Journal of Pharmaceutics, 40: 173-174. Hanne, A., Krupp, G.. Removing DNA contamination from pipettes. Bionews, Eppendorf, No.8. Hayatsu, H., Pan, S.K, and Ukita, T. 1971. Reaction of sodium hypochlorite with nucleic acids and their constituents. Chem. Pharm. Bull, 19 (10) 2189-2192. Heinrich, M. 1991. PCR carry-over. BFE, 8 (10): 594-597. Jette, L.P., Ringuette, L., Ishak, M., Limmer, M. and Saint-Antoine, P. 1995. Evaluation of three glutaraldehyde-based disinfectants used in endoscopy. J Hosp Infect, 30 (4): 295-303. Jrgen, H. and Kaiser, K. 1996. Avoiding viral contamination in biotechnological and pharmaceutical processes. Nature Biotechnology, 16: 1077-1079. Kwok, S. and Higuchi, R. 1989. Avoiding false positives with PCR. Nature, 339: 237-238. Leuci, C. 1998. Intrt des EVC dans le secteur biomdical. Caoutchoucs et plastiques. 769. Manual de operao da Pipetman P - marca Gilson Mifftin, T.E. Control of contamination. Molecular Bio-Products. Ou, Chin-Yih, Moore, J.L, and Schochetman, G. 1991. Use of UV irradiation to reduce false positivity in Polymerase Chain Reaction. BioTechniques, 10 (4): 442. Part 1: terminology, general prEN ISO/ FDIS 8655-2 - Piston-operated volumetric apparatus Part 2: pipettes prEN ISO/ FDIS 8655-6 - Piston-operated volumetric apparatus Part 6: Gravimetric test methods * Prince, A.M. and Andrus, L. 1992. PCR: how to kill unwanted DNA. BioTechniques, 12 (3): 358360. * * LT801117 800353A

52 of 53

Volume II / Module 1 - Micropipettes: Principles and Operation

Sambrook, Fritsch, and Maniatis. 1989. In Vitro Amplification of DNA by the Polymerase Chain Reaction. In Molecular Cloning: A Laboratory Manual, 2nd ed. 14:14. Taibi, C. Linfection existe, sa prvention aussi. Guide pratique dhygine hospitalire. Tritt, C.S. 1997. Sterilization methods. http://www.msoe.edu/~tritt/be4xx/stermeth.html. Van Bueren, J., Simpson, R.A., Salman, H., Farelly, H.D. and Cookson, B.D. 1995. Inactivation of HIV-1 by chemical disinfectants: sodium hypochlorite. Epidemiol Infect, 115 (3): 567-579. Le contrle des micro-organismes par les agents physiques et chimiques. Partie 4, chapitre 5: le contrle des micro-organismes. Extraction and purification of RNA. Extraction, purification, and analysis of messenger RNA from eukaryotic cells (7.3-7.5). Sterilization and sterility assurance of compendial articles, USPXXI. Sterilization and sterility assurance, general information Avoiding Ribonuclease contamination. neb.com.

53 of 53

TECHNICAL STAFF

Volume II Module 2 Water for Instrumental Chemical Analyses Editor: Jos Muradian Filho - Millipore

Coordination: Cludia Franklin de Oliveira ANVISA Itapuan Abimael da Silva ANVISA Karen de Aquino Noffs Brisolla ANVISA Karla de Arajo Ferreira ANVISA Marcelo Cludio Pereira ANVISA Max Weber Marques Pereira ANVISA Renato Almeida Lopes - ANVISA

SUMMARY
1. WATER FOR INSTRUMENTAL CHEMICAL ANALYSES ..................................................... 5 1.1. Introduction ................................................................................................................................... 5 1.2. Water contaminants ...................................................................................................................... 5 1.2.1. Dissolved inorganic compounds ...................................................................................... 6 1.2.2. Dissolved organic compounds ......................................................................................... 6 1.2.3. Particles and colloids .......................................................................................................... 6 1.2.4. Microorganisms .................................................................................................................. 7 1.2.5. Dissolved gases ................................................................................................................... 7 1.3. Monitoring and control of the purified water quality.............................................................. 7 1.3.1. Monitoring of inorganic ionic contaminants conductivity and resistivity .............. 8 1.3.2. Monitoring of organic contaminants TOC total oxidizable carbon .................. 10 1.4. Quality specifications .................................................................................................................. 11 1.5. Water purification technologies for laboratories .................................................................... 12 1.5.1. Distillation ......................................................................................................................... 12 1.5.2. Deionization ...................................................................................................................... 13 1.5.3. Reverse osmosis ................................................................................................................ 14 1.5.4. Continuous electrodeionization ...................................................................................... 16 1.5.5. Ultrafiltration ..................................................................................................................... 18 1.5.6. Membrane microfiltration ............................................................................................... 19 1.5.7. Activated charcoal ............................................................................................................ 20 1.5.8. Ultraviolet radiation (UV) ............................................................................................... 20 2. PURIFIED WATER SPECIFICATIONS ...................................................................................... 23 2.1. Purpose ......................................................................................................................................... 23 2.2. Monitoring of the quality of several types of water .............................................................. 24 2.3. Calibration and qualification...................................................................................................... 24 2.4. Storage recommendations ......................................................................................................... 25 2.5. Water purification equipment maintenance ............................................................................ 25 2.6. Control forms .............................................................................................................................. 25 2.7. Reference documents ................................................................................................................. 25 3. REFERENCES ................................................................................................................................... 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1. WATER FOR INSTRUMENTAL CHEMICAL ANALYSES

1.1. Introduction Water, as found in nature, contains several substances other than the H2O molecule: salts, organic compounds, microorganisms, particles, dissolved gases. This is basically attributed to a very known feature: water is a universal solvent. At the laboratory, the several usages of water require a higher or lower grade of purity. This grade is defined as a function of the application, analysis method, interferers and its sensitivity and limit of detection. That is why the purified water and its quality are essential factors for the preparation of solutions in general, buffer solutions, mobile phases in chromatography, blanks e also other common, but not less important applications, such as glassware washing (final rinsing). Defining, knowing, removing and controlling water contaminants are critical steps to effectively and economically obtain purified water in the laboratory, particularly complying with the application requirements. Information and concepts will be presented next in four topics to help attaining such purposes and decide what kind of water is to be used in a specific instrumental analysis: Water contaminants Monitoring and control of the purified water quality Quality specifications Technologies that can be used for purifying water in laboratories

1.2. Water contaminants As mentioned, contaminants have a direct relationship with the analysis in progress, as well as the sensitivity and limit of detection of the method used. For example, in HPLC High Performance Liquid Chromatography the analyses using ultraviolet detector are sensitive to organic compounds present in the water (mobile phase component) and that absorb in the wavelength used. This compromises the results and the calibration curve itself. Further, organic compounds not detected in this wavelength can be retained and accumulate in the HPLC column, thus impairing its efficiency and functionality, leading to the so-called ghost peaks in a chromatogram. For didactics purposes, we divided the water contaminants in five categories:

5 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1.2.1. Dissolved inorganic compounds Basically consist in water-dissolved salts, the inorganic ions present. Can be present as cations or anions. Anions: Chloride ion (Cl-), hypochlorite ion (HClO-), Nitrates, nitrites, carbonates, Sulfates, Silicates, etc... Cations: Sodium (Na+) , Calcium and Magnesium (Ca++, Mg++, Iron (Fe++), Aluminum (Al+++), Heavy metals (Pb, Ni, Cr, Hg), etc... 1.2.2. Dissolved organic compounds This class of contaminants can be found in nature or result from pollution or the water potability process itself for human consumption.

Natural origin
Include organic materials resulting from vegetal and animal decomposition. Vegetal decomposition is very common in open fountains (dams, reservoirs) where the water is captured for subsequent treatment. Among other examples, we can mention tannins, lignins, phenols, humic acids (complex mixture of macromolecules having phenolic polymer structure). Also include proteins, enzymes (nucleases, for example), amino acids e their derivatives.

Unnatural origin
Includes organic compounds originated from human activities. It can be emphasized all kinds of water-soluble pesticides (fungicides, insecticides, herbicides), polyaromatic hydrocarbons (PAHs), polychlorinated biphenyls (PCBs), EDTA (chelating agent always present in soap, detergents and cosmetics formulations), citrates (in regions next to citric crops). 1.2.3. Particles and colloids Particles can be rigid (sand, stones, earth) or deformable (vegetal wastes). It is characterized as a contaminant for being a protector shield for microorganisms against the action of UV rays and disinfecting agents. They are also a vehicle for such microorganisms, since bacteria, for example, may adhere to the particles. In instrumental analyses, the particles may also cause direct serious problems. In HPLC, for example, there is always a risk of column clogging and damage to the injection pump piston. Further, particles immobilized at the inlet of the column may become another stationary phase, thus unexpectedly impairing the column retention features and the separation itself.
6 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Colloids are stable suspensions of organic or inorganic particles with size ranging from 0,1 to 0,001 mm. All particles have the same charge and signal. The resulting repulsion keeps the suspension stable. They are difficult to remove by microporous filtration, but can be easily retained by reverse osmosis or ultrafiltration. 1.2.4. Microorganisms Microorganisms bacteria, fungi (mold and yeasts) and virus are commonly present in drinking water distributed in a city. Although treatment companies process the water so as to remove microorganisms hazardous to the human health, the supplied water is not sterile at all. The other microorganisms remaining after water drinkability process may impair the analyses. 1.2.5. Dissolved gases We can find all gases existing in the atmosphere dissolved in water. The most important gas dissolved in water is carbon dioxide (CO2). It is balanced with carbonic acid that, in turn, is balanced with the bicarbonate ion:

This balance generates an important ion, bicarbonate, which may interfere with the analyses and impair the purification processes. CO2 dissolution rate and its concentration in water depend on the temperature (the lower is the temperature, the higher is the dissolution and absorption rate) and on the CO2 partial pressure in the environment to which the water is exposed. Further effects of other gases in the air: Oxygen: causes corrosion in tanks, pipings, etc. Ammonia: contaminant resulting from agricultural fertilization. SO2 contained in automobiles escape gases and industrial emissions.

1.3. Monitoring and control of the purified water quality As mentioned above, the contaminants contained in water may significantly interfere with analytical determinations. Therefore, it is important to quantify them and be sure they comply with the quality specifications. Besides, this quantification is a measure of the efficiency of the water purification process.

7 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

This topic will basically consider the detection of inorganic (ions) and organic contaminants. There is a wide range of inorganic contaminants and analyzing them individually would be extremely hard and laborious. It is possible to determine the total ionic contamination total of an aqueous solution using only one parameter: conductivity that is the specific conductance or its reciprocal, the resistivity (specific resistance). Equally, there are several organic contaminants contained in the purified water. Here, also, one single parameter can be used to quantify them: Total Organic / Oxidizable Carbon (or TOC Total Oxidizable Carbon). Bacteria and other microorganisms are quantified using methods employing filtration in microporous membranes and subsequent incubation using culture mediums or, more recently, chemiluminescence. The results are expressed in cfu/mL (colonies forming units per mL). The amount of particles can be controlled by means of specific removal technologies (microfiltration, ultrafiltration and reverse osmosis). In water purification processes, this is enough and generally there are no concerns on the quantification of particles using appropriate technologies and assuring that all particles above a certain size are removed. 1.3.1. Monitoring of inorganic ionic contaminants conductivity and resistivity The conduction of electric current depends on the ions contained in water. The purer is the water, the lower is the concentration of such ions and, therefore, the lower is conductivity. For low conductivity, the reciprocal applies, high resistivity, as they are different expressions for the same phenomena. Therefore: C= 1/R or R= 1/C Measurement units: Conductivity microhm/cm or microSiemens/cm Resistivity Megohm.cm The water conductivity at the purification theoretical limit, that is, when almost all ions have been removed and only H+ and OH- remains, is 0.055 microSiemens/cm or 18.2 Megohm.cm, at 25 C. This is called type I ultrapure water (ASTM). Graph 1 shows that temperature is an important aspect when measuring conductivity. The lower is the temperature, the higher is the resistivity (therefore, the lower is conductivity). This is because at low temperatures, a lower mobility of ions in solution takes place and such mobility is inversely proportional to the resistivity or directly proportional to conductivity.

8 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Graph 1. Resistivity and Conductivity Variation as a function of temperature Resistivity (Megohm.cm)

At 25C, the ultrapure water resistivity is 18.2 M.cm

Temperature (degrees C) That is why water temperature should be known when comparing these measurements. Conductivity meter and resistivity meters installed in water purification equipment generally have a temperature offset circuit for 25C, thus allowing a direct reading. As water becomes purer, that is, as its conductivity decreases or resistivity increases, it is more prone and faster to aggregate contaminants from the environment, thus returning to its natural state. For that reason, it is crucial that ultrapure water is not stored and obtained right before usage. Graph 2 shows the highly purified water tendency to reuptake contaminants, evidenced by a sharp decrease in resistivity some minutes after purification. This decrease in resistivity is due to the atmospheric carbon dioxide contamination that is balanced with the bicarbonate ion, the presence of which leads to an increase in conductivity (or decrease in resistivity). Graph 2. Contamination of ultrapure water by atmospheric gases - CO2 (20 C)
Resistivity (M .cm)

9 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1.3.2. Monitoring of organic contaminants TOC total oxidizable carbon There are currently two kinds of TOC analyzers. Analyzers using physicochemical oxidation and analyzers using a technology patented by Anatel Corporation. TOC determination by physicochemical methods first involve the removal of CO2 from the water to be tested by purging with nitrogen. This leads to a problem: it is difficult to remove all the CO2. The second step involves the addition of reagents (peroxides) and the action of a catalyst (UV or heat) to start the oxidation reaction. Once this oxidation is complete, the carbon dioxide produced is removed from the water sample by bubbling with nitrogen and collected by adsorption in a column. Next, the column is desorbed by increasing the temperature and carryover using a flow of pure nitrogen. The presence of CO2 is detected by infrared spectrometry and concentration determined by peak integration. This method has several limitations when measuring TOC in ultrapure water: First, there is a risk of contamination during sampling and the method cannot be easily adapted to on-line measurements. As mentioned above, it is difficult to remover the last traces of CO2 in the purging stage. Volatile organic compounds can be also carried over by nitrogen and, therefore, are not accounted in the total TOC. The oxidation reagents used are kept in a normal room and, therefore, subject to CO2 contamination. Organic materials that are adsorbed and uncontrollably released slowly and cumulatively contaminated the instrument piping and the column. Finally, the technique requires the use of a great amount of high purity nitrogen, what is extremely expensive. The second technique, developed by Anatel Corporation, is simpler and involves the organic compounds oxidized by UV radiation. The UV radiation at a wavelength of 185 nm converts O2 into ozone (O3) that is a strong oxidizing agent. With the aid of UV radiation at 254 nm, ozone, in turn, will react with water to form free radicals hydroxyl that will then react with organic compounds, thus oxidizing them into carbon dioxide. The resulting decrease in resistivity (or an increase in conductivity) is then measured and correlated with the TOC reading. The main advantage is that such a method allows quick e automatic on-line measurements, being very sensitive (< 1 ppb TOC), reproducible, does not use reagents and define all kinds of organic compounds, inclusive the volatile ones. However, this method is appropriate for waters having a resistivity above 3 Megohm.cm. The TOC concentration is expressed as ppb parts per billion.

10 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1.4. Quality specifications A number of standardization authorities have quality specifications intended to cover specific activities to which they are dedicated. The ASTM specifications American Society for Testing and Materials are intended to cover analytical instrumentation and the electronics industry. The United States Pharmacopeia (USP) specifications cover the production of drugs and ISO International Organization for Standardization. Further, the NCCLS National Committee for Clinical Laboratory Standards, in liaison with CAP College of American Patologists, specifies the kinds of water used in clinical laboratories. ASTM rule, reference D-1193-99, is the most appropriate for the specification of the kinds of water used in chemistry laboratories e, especially, instrumental analysis. Historically, ASTM was the first entity to propose rules for laboratory water, dividing water into four grades, according to the applications:

Type I -

as known as ultrapure water or reagent grade water, used in critical laboratory applications, including the HPLC instrumental analyses. for less demanding applications, such as glassware rinsing, qualitative analysis or organic synthesis.

Type II -

Type III - for general laboratory applications: preparation of culture mediums and final glassware rinsing in non-critical applications. Type IV - supply of reagent grade water producing systems.

Microbiologic Contamination when bacteria level requires control, the type of grade should be provided as follows:

11 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1.5. Water purification technologies for laboratories In order to purify water to the levels specified in ASTM and other authorities, several technologies are available: Distillation Deionization Reverse osmosis Continuous Electrodeionization (EDI) Ultrafiltration Membrane microfiltration Activated Charcoal Ultraviolet Radiation

1.5.1. Distillation Distillation is a classic water purification process that applies water change from the liquid to the gas state, with liquid phase condensation. By using the distillation process, several contaminants are removed, however it is still not possible to obtain type I ultrapure water. Organic substances having a low boiling point are also converted into distilled water. Azeotropic mixtures are also formed and even high molecular weight compounds are carried over with the vapor. Additionally, when chlorine reacts with natural organic compounds at high temperatures, organochlorinated compounds are formed and also carried over by the vapor. Silica is extracted from glass distillers, as well as other ions in metal distillers. Further, the distillation process consumes a great amount of electricity and cooling water. DISTILLATION Advantages Removes a great percentage of all types of contaminants Produces water with resistivity between 0.2 and 1 Megohm.cm. Medium investment Commonly known and easily operated process Disadvantages Not all contaminants are removed and many of them are produced during the process. The water quality is not controlled. High operation cost: electricity and water. Regular and pretreatment maintenance are essential to assure performance.

12 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

1.5.2. Deionization

Figure 1. Scheme of the mixed bed deionization process The deionization process is performed using ion exchange porous globular resins having diameters between 0.3 and 0.8 mm. Such resins are manufactured with polymeric chains containing styrene and divinylbenzene cross-links in which charged chemical groups are bonded. In anionic resins, that is, resins capturing anions, these groups contain quaternary ammonium. Sulfonic groups are used in cationic resins resins capturing cations. At the beginning of a deionization process, when the resin is still unused, ions bonded to negative groups of cationic resins are H+ protons and ions bonded to anionic resins are hydroxyls OH-. During the process, cations contained in water (Na+, Ca++, etc.) will bind to the cationic resin e displace the H+ proton, while anions (Cl-, NO3 -) will bind to the anionic resin and displace hydroxyls. Deionization is a very efficient process to remove ions and even some ionized organic compounds. However, resins are an excellent support for bacterial growth. As feeding water is not sterile, the result is that water bacteriological contamination after a deionization column can be 1000 times greater than the inlet water.

13 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

DEIONIZATION Advantages Effective removal of ions (Resistivity: 1 - 10 Megohm.cm) Simple installation. Low investment. Regenerable. Disadvantages Does not remove particles, organic material or microorganisms. Regenerated resins can produce particles, organic materials or promote the bacterial growth. Standard ion exchange: the resin origin is unknown. High operation costs: regeneration/ transportation. Water quality is variable; damaged globules.

Resins can be regenerated in the laboratory or in contract companies that provided this service by changing a worn-out column by a regenerated one. This can be a source of residual ionic contamination as a result of the application of the inlet or environmental water to which these resin are exposed. Single-use, therefore, non-regenerable, resins are the most indicated so as to avoid all such problems. 1.5.3. Reverse osmosis

Osmosis

Reverse Osmosis

Osmotic Pressure

Reverse Osmosis Membrane (RO)


Figure 2. Scheme of the osmosis and reverse osmosis principle Supposed we have a U-shaped tube with both branches separated by an osmosis membrane. In the left branch, the water molecules are dissolved (represented by dots). According to the laws of thermodynamics, the molecules would diffuse from the left to the right until their concentrations are equivalent. However, these molecules cannot cross the membrane, but the water molecules can. Thus, water will cross the membrane from the right toward the left branch to dilute the dissolved
14 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

molecules. This process will continue until a pressure differential (osmotic pressure) is created between the branches, thus balancing the difference in concentrations. A reverse osmosis is exactly the opposite of this process: water containing ions or other contaminants is pressurized against a reverse osmosis membrane; the pure water is obtained on the other side of the membrane. The pressure exerted should be higher than the osmotic pressure. Reverse osmosis membranes typically rejected 90% of the monovalent, 95% of the bivalent e 99% of the polyvalent ions. 99% of the organic compounds having a molecular weight above 300, virus and bacteria are also rejected. In order to obtain a continuous flow of water molecules on this membrane, contaminants have to be regularly removed. This can be achieved by the so-called tangential flow (figure 3) that allows part of the water flow to go across the membrane surface, thus promoting an authentic scanning and preventing the accumulation of contaminants.
Reverse Osmosis Membrane Water Feeding Permeated

Rejected

Figure 3. Scheme of the tangential flow in a reverse osmosis system On equipment, this setting is provided as helically wound membranes, thus forming a cartridge where permeate and refuse channels are perfectly separated and tight. This creates a great separation area in a relatively small volume, thus allowing the construction of compact systems. Reverse osmosis systems remove a reasonable percentage of all kinds of contaminants; however, since this is a percentile removal, it is not enough to reach even the type II water purity levels required for most of the situations. The table below shows the advantages and disadvantages of this process. REVERSE OSMOSIS Advantages Removes a reasonable percentage of all kinds of water contaminants (ions, organic matter, pyrogens, virus, bacteria, particles and colloids). Low operation costs as a result of the low consumption of electricity. Minimum maintenance Good control of operation parameters. Disadvantages The contaminants are not sufficiently removed so as to comply with the type II water requirements. The Reverse Osmosis Membranes are subject to long-term encrusting and clogging (if not adequately protected). Until recently, the consumption of water was similar to distillers.

15 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Electricity is employed in this process to operate the pressurization pump only; therefore, a very low consumption if compared to the distillation process. The osmosis membranes can be also incrusted and clogged as a result of salts precipitation, mainly calcium carbonate, thus reducing the outflow and damaging the membrane structure. For this reason, controlling the incrusting ions is required so as to avoid this kind of problem. 1.5.4. Continuous electrodeionization If we submerge two electrodes - anode and cathode into a NaCl solution, for example, and apply a potential difference between the two electrodes, the Na+ and Cl- ions will migrate towards the cathode and the anode respectively. There are membranes called ion exchange membranes selectively permeable to cations or anions and that allow such migration to be controlled. These membranes are made of ion exchange resin fragments included in a polyethylene matrix. Such fragments of cationic or anionic resins respectively allow cations or anions to pass. If these membranes are introduced in the NaCl and electrodes system described, we are able to control the migration of Na+ and Cl- ions as showed in the figure below.
Anionic Membrane Cationic Membrane

Anode

Cathode

Anode

Cathode

Figure 4. Control of the migration of ions through ion exchange membranes In our model, only the chloride ions can cross the anionic membrane. Conversely, only sodium ions can cross the cationic membrane. The so-called electrodialysis system is created in a system where both membranes are alternated and the electrodes maintained (figure 5a).

Anode

Cathode

A - Anionic Membrane C - Cationic Membrane

Product

Figure 5. Electrodialysis system


16 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

In the next cell, continuing on the right, both the chloride and sodium ions can leave. We can clearly notice a depletion of ions; electroneutrality is also maintained since for each chloride ion there is one sodium leaving the same cell. Conversely, the ions are cumulating in the next cell, as no sodium or chloride can leave. As a result, there is a system where the ions are concentrated in some cells (disposal flow) and practically absent in others (product flow). This process can be used for purifying water; however, it is extremely slow once the ions would have to move from a cell to a membrane and towards the electrode. In order to improve the systems performance, the product cells are filled with mixed bed ion exchange resins (anionic + cationic). This will allow an ionic transference from the cell center towards the semipermeable membrane instead of moving at a low speed in water where collisions caused by the Brownian movements would decrease their advance. Thus, the ions jump from one resin active site to another towards the electrode having an opposite signal. This setting also allows the capture of weakly charged organic substances. The resins are continuously regenerated as H+ and OH- are generated in the electric field created, thus composing microenvironments around the resins. Nevertheless, the continuous electrodeionization purification process has to be fed with water previously purified by reverse osmosis and having an attenuated concentration of salts.

Anode

Cathode

A - Anionic Membrane C - Cationic Membrane

Product

Figure 6. Ion exchange resins filling the product cells

17 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

CONTINUOUS ELECTRODEIONIZATION (EDI) Advantages Effectively removes the dissolved organic material. Requires minimum maintenance. Requires no regeneration of the resins. Low operation cost. Excellent pretreatment for water ultrapurification systems Disadvantages Requires pre-purification

1.5.5. Ultrafiltration

Figure 7. Representation of the schematic section of a ultrafilter showing the ratio between the active separation layer (1) and the support layer (100). Ultrafilters are polymeric asymmetric membranes with a very fine active layer (1 micrometer thick) on the top and a thicker support layer (100 micrometers). The ultrafiltration membranes (UF) operate under pressure. Under these conditions, small molecules will be able to cross the active layer, while the bigger molecules are retained. The cut-off is called Nominal Molecular Weight Limit NMWL and expressed in daltons. This value is used to characterize the UF membranes once the molecular weight is easier to found in literature than the molecules sizes. The molecular weight is directly related to the molecule size, however other factors also have to be regarded: shape, stereochemistry and pH of the dispersion in which they are. This is essential for proteins. The table below shows the advantages and disadvantages of UF.

18 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

ULTRAFILTRATION - UF Advantages Effective removal (>99%) of all organic molecules having a molecular weight above the NMWL. Very effective removal of pyrogens and virus, as well as particles. No risk of incrustation. Low consumption of water and electricity. Low maintenance; well documented/ accepted procedures Disadvantages Almost no removal of ions, gases and low molecular weight organic matter (UF membranes provided with a narrower mesh have a cut-off of 1.000 dalton)

As noticed, these membranes are very efficient when removing high molecular weight organic compounds. 1.5.6. Membrane microfiltration

Surface Filter

Durapore Membrane Microphotograph

Figure 8. Cross-section representation of a surface filter and microphotograph of the Durapore membrane (Photo: courtesy of Millipore Ind. e Com. Ltda.) There are three main types of filters: depth, surface and membrane filters. Depth filters are made of agglomerated fibers and retain contaminants throughout its thickness. Although having high capacity, their limit of detection is not very clear. Their efficiency is around 95% and varies as a function of the outflow. Besides, they may release contaminants in the filtrate, generally fibers or retained material. Surface filters are generally made of multiple non-fibrous material layers, capturing contaminants mainly on their surface and having an intermediate retention and efficiency capacity (98%). Screen or membrane filters retain contaminants on their surface by means of a sifting effect. In spite of their low capacity, such filters retain 100% of the contaminants having a size above their well-defined cut-off limit. Typical screen filters are the membrane filters which integrity can be verified by means of specific tests, for example, bubble point or diffusion.
19 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

MEMBRANE MICROFILTRATION Advantages Total (100%) removal of all contaminants (particles, bacteria) greater than the pore size. Integrity test available. Sterilizing filtration (0.22 mm membrane). Minimum maintenance: simply replace as required. High outflows are obtained at low pressures. Outflow independent efficiency. Disadvantages Minimum effect on other contaminants.

Surface retention: may be subject to clogging or obstruction.

The main advantage of the membrane filters is the total (100%) removal of all contaminants having a size greater than their pore diameter. Membranes with a pore diameter of 0.22 m have been used in the pharmaceutical industry for several years in the sterilizing filtration of solutions. 1.5.7. Activated charcoal The activated charcoal is mainly used due to capacity to adsorb organic materials as a result of its wide surface area up to 1000 m2/g. Another function is to reduce oxidizers, such as free chlorine, contained in the water and that could affect reverse osmosis membranes or ion exchange resins. Therefore, this is a technology mainly aimed at the pretreatment and protection of other steps. ACTIVATED CHARCOAL Advantages Effective removal of a wide range of organic substances (even those with low molecular weight) by means of non-specific bonds (Van der Waals force). Large capacity as a result of the great surface area. 1.5.8. Ultraviolet radiation (UV) UV radiations have wavelengths between 100 and 400 nm, divided into four fields: ultrashort, short (UV-C), medium (UV-B) and long (UV-A) waves. These radiations can be easily produced by low mercury vapor pressure lamps emitting radiations at two wavelengths: 185 and 254 nm. Disadvantages Very weak effect on other contaminants (except for some particles removed by depth filtration). When all sites are busy, balance is reached and the organic compounds released. Bacteria may develop after some time. Efficiency depends on the outflow.

20 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Radiations with a wavelength between 200 and 300 nm destroy microorganisms by breaking the DNA chain. This is more intense at 260 nm. Therefore, the wavelength of 254 nm has an efficiency that is very close (80%) to that regarded as optimum for this purpose. However, a very ascertained design is required so that such efficiency is really reached, mainly taking the UV radiation penetration capacity (in the order of 1 cm) into consideration. It is necessary to provide a sufficient outflow to allow this penetration, respecting the dwelling time of water in the exposure chamber. Another very critical application of the UV radiation is the reduction of the Total Oxidizable Carbon (TOC) levels of the purified water. In fact, the UV radiation does not directly destroy organic compounds, but generates ozone that will oxidize the substances contained in water. As it can be seen in the figure below, on the left, two different wavelengths are necessary to generate the free radicals hydroxyl (OH) that will subsequently oxidize organic compounds.

Figure 9. On the left: sequence of reactions to form the free radical hydroxyl. On the right: oxidation reactions of organic compounds. Example: methanol. Both mechanisms described in the literature (Norrish et al., 1965 and Banford et al. 1967) require the action of the wavelengths 185 and 254 nm and the presence of oxygen in the water. Both mechanisms have the same stoichiometry and result in the formation of the same number of hydroxyl radicals. On the right side of the figure, an example of oxidation of an organic compound with no charge (methanol) is showed. This is one of the simplest organic molecules, having only one carbon atom. Under the action of the free radicals hydroxyl, this molecule is successively oxidized in an increasing oxidation scale: starting with the alcohol, and then aldehyde, acid and finally carbon dioxide and water.
21 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

As previously discussed, carbon dioxide and water are balanced with carbonic acid that, in turn, is balanced with the bicarbonate and H+ ions. Therefore, the ion exchange resins remove the carbon. Note that by converting organic carbon into an ion that is easy to remove by deionization, it was possible to reduce the purified water TOC level. UV RADIATION (185 + 254 nm) Advantages Conversion of traces of organic contaminants into charged specimens and, at the end, into CO2 (185 + 254) Limited destruction of microorganisms and virus (254). Low electricity consumption. Easy operation. Disadvantages Polishing technique only: can be impaired if the concentration of organic matter in the feeding water is very high. Organic compounds are converted and not removed. Limited effect on other contaminants. The design has to be appropriate to assure efficiency.

22 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

2. PURIFIED WATER SPECIFICATIONS


2.1. Purpose This specification concerns appropriate requirements for water to be used in chemical analysis methods and physical tests. It has been based on ASTM D1193-99 specifications, as amended. Four categories were specified:

Measurement of pH in type I, II and III reagent grade water is not included in the specifications, since water at these purification grades contain no components in amounts sufficient to lead to a pH change. Methods for the preparation of several types of water determine the impurity limits and should comply with the following specifications: 2.1.1. Type I reagent grade water should be prepared by an appropriate pre-purification process, followed by polishing using a ultrapurification system comprising a mixed bed ion exchange resin and a 0.22 m final membrane. The preparation should be always carried out at the time of use. It is recommended, but not compulsory, to provide the ultrapurification system (final polishing) with a UV lamp with emission at 185 and 254 nm so as to maximally reduce the TOC levels. Additionally, it is also recommended, but not compulsory, to provide the ultrapurification equipment (final polishing) with a TOC on-line meter for constant monitoring of organic contaminants. Supply (pretreatment) of the water final step should have a conductivity of at least 10 S/ cm at 25C. 2.1.2. Type II reagent grade water should be prepared by a reverse osmosis process combined with electrodeionization, distillation or ion exchange having a conductivity below 1.0 S/cm at 25C. Ion exchange, or reverse osmosis and organic adsorption, can be required and associated with distillation if the purity is not reached by a single distillation step. 2.1.3. Type III reagent grade water should be prepared by reverse osmosis followed by continuous electrodeionization, distillation, ion exchange or a combination thereof, followed by a polisher with a 0.45 m membrane filter. The use of this filter is required when microbiologic contamination is required.

23 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

2.1.4. Type IV reagent grade water should be prepared by reverse osmosis, distillation, ion exchange, reverse osmosis followed by continuous electrodeionization, electrolysis or a combination thereof. The several types above can be chosen by method or investigation. This specification is not intended to cover safety concerns associated with the application. It is users responsibility to establish the appropriate safety standard and healthy practices and define the applicability of the priority use regulation limits.

2.2. Monitoring of the quality of several types of water 2.2.1. Conductivity and resistivity: For measuring the conductivity or resistivity, the conductivity or resistivity meter of the water purification equipment should be used provided that they are duly calibrated. If a distiller having no meter coupled is used, measurement should be carried out using a duly calibrated portable conductivity meter. 2.2.2. TOC Total oxidizable (organic) carbon Although the TOC values for all types of water are specified, only Type I grade water is critical, since this water will be in contact with the analytical equipment and high TOC levels may lead to impaired results. Therefore, this water grade requires a systematic monitoring of the TOC levels. We recommend that water ultrapurification equipment (final polishers) is provided with a duly calibrated TOC on-line meter. If this is not possible, measurement should be made on-line using equipment applying the photooxidation technique for resistivities above 3 megohm.cm. 2.2.3. Measurements frequency: 2.2.3.1. Resistivity and conductivity: 2.2.3.1.1. Daily for type I, II, III or IV water. 2.2.3.1.2. Upon removing water for type I water. 2.2.3.2. TOC total oxidizable carbon: 2.2.3.2.1. When the equipment is not provided with a meter, the TOC levels should be measured at least once every fifteen days. 2.2.3.2.2. When the ultrapurification equipment is provided with a TOC meter, measurement should be made upon the water removal.

2.3. Calibration and qualification 2.3.1. Each water purification system should have their meters calibrated and qualified according to the Installation, Operation and Performance aspects.

24 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

2.4. Storage recommendations 2.4.1. Type I grade water Should not stored due to fast degradation. It should be prepared at the moment of use. 2.4.2. Type II, III and IV grade water can be stored in a reservoir protected against external contaminations. It is strongly recommended to store such purified water for a period not exceeding twenty-four hours as of its preparation.

2.5. Water purification equipment maintenance All the manufacturers indications should to be followed mainly concerning the change of occasionally required consumable and cleaning material. It is recommended to establish a preventive maintenance program for each equipment and keep records for consumable material replacements and maintenances carried out.

2.6. Control forms Attachments 1 and 2 show control form templates for recording the resistivity or conductivity measurements , as well as TOC.

2.7. Reference documents D1193 D1125 D1129 D1293 D4453 D4779 Standard Specifications for Reagent Water Test Methods for Electrical Conductivity and Resistivity of Water Termninology Relating to Water Test Methods for pH of Water Practice for Handling of Ultra-Pure Water Samples Test Method for Total, Organic, and Inorganic Carbon in High Purity Water by Ultraviolet (UV) or Persulfate Oxidation, or Both, and Infrared Detection D5391 Test Method for Electrical Conductivity and Resistivity of a Flowing High Purity Water Sample D5542 Test Method for Trace Anions in High Purity Water by Ion Chromatography D5997 Test Method for On-Line Monitoring of Total Carbon, Inorganic Carbon in Water by Ultraviolet, Persulfate Oxidation and Membrane Conductivity Detection

25 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Attachment 1. Control form template for water ultrapurification equipment (final polishing) provided with a TOC meter.

26 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Attachment 2. Control form template for water ultrapurification equipment (final polishing) not provided with a TOC meter.

27 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

Attachment 3. Control form template for water purification equipment for obtaining type II, III or IV grade water.

28 of 29

Volume II / Module 2 - Water for Instrumental Chemical Analyses

3. REFERENCES
American Society for Testing and Materials D1193-99, Standard Specification for Reagent Water (1999). Millipore, Internet site: http//millipore.com/H2O. Millipore Indstria e Comrcio Ltda. Seminrios de Purificao de gua para Laboratrios. Norrish et al., Proc.Roy.Soc.Ser. A. 288: 316 (1965). Banford et al., Photochemistry and Reaction Kinetics. P.G. Ashmore et al., Ed. Cambridge (1967).

29 of 29

TECHNICAL STAFF
Volume II Module 3 Analytical Instrumentation Editors and Cooperators: Adauto da Silva VARIAN Celso Ricardo Camargo - SINC BRAZIL Demian Rocha Ifa SINC BRAZIL Ivan Jonaitis AGILENTTECHNOLOGIES Jos Apareido Soares - VARIAN Luiz Rinalo Bizaio - VARIAN Renato Guva - AGILENT TECHNLOGIES Renato eres - FLOWSERVICE Ricardo Lira - FLOWSERVICE Robson Sanches Bizi - VARIAN Coordination: Cludia Franklin de Oliveira ANVISA Itapuan Abimael da Silva - ANVISA Karen de Aquino Noffs Brisolla - ANVISA Karla de Arajo Ferreira - ANVISA Marcelo Cludio Pereira - ANVISA Max Weber Marques Pereira - ANVISA Renato Almeida Lopes - ANVISA

SUMMARY

1 . INTRODUCTION... ...................................................................................................................... 7 2. ULTRA VIOLET-VISIBLE (UV-VIS) SPECTROPHOTOMETRY. ............................ 8 2.1. Introduction ................................................................................................................................ 8 2.2. Technique principle ................................................................................................................. 10 2.2.1. Light spectrum ............................................................................................................ 10 2.2.2. Absorption processes ................................................................................................. 10 2.2.3. UV - Vis absorption spectrum .................................................................................. 12 2.2.4. Effect of solvents ....................................................................................................... 14 2.3. Basic description of the system ............................................................................................. 15 2.3.1. Sources of radiation ................................................................................................... 15 2.3.2. Monochromator (optical)........................................................................................... 16 2.3.2.1. Single-beam ................................................................................................... 17 2.3.2.2. Diodes arrangement .................................................................................... 18 2.3.2.3. Double-beam ................................................................................................ 18 2.3.3. Sample compartments ................................................................................................ 19 2.3.3.1. Optical fiber coupler .................................................................................... 19 2.3.3.2. Sipper ............................................................................................................. 19 2.3.3.3. Support for solid samples ........................................................................... 19 2.3.3.4. Reflectance accessories ................................................................................ 20 2.3.4. Data acquisition ........................................................................................................... 20 2.4. Minimum installation and operation requirements ............................................................. 20 2.5. Basic precautions ..................................................................................................................... 21 3. GAS CHROMATOGRAPHY (GC). ..................................................................................... 22 3.1. Introduction .............................................................................................................................. 22 3.2. Technique principle ................................................................................................................. 22 3.3. Basic description of the system ............................................................................................. 23 3.3.1. Gases used.................................................................................................................... 24 3.3.2. Flow and pressure controllers ................................................................................... 25 3.3.3. Injectors ........................................................................................................................ 25 3.3.3.1. Packed injectors ............................................................................................ 25 3.3.3.2. Capillary injectors ......................................................................................... 26 3.3.4. Columns ....................................................................................................................... 27 3.3.4.1. Columns oven ............................................................................................... 28 3.3.5. Detectors ...................................................................................................................... 28 3.3.5.1. Flame ionization detector (FID) ................................................................ 28 3.3.5.2. Nitrogen and phosphorus (NPD) or specific thermoionic detectors ...... (TSD) ............................................................................................................ 29 3.3.5.3. Electrons capture detector ECD ............................................................ 30 3.3.6. Data acquisition and processing ............................................................................... 31

3.4. Minimum installation and operation requirements ............................................................. 31 3.5. Basic precautions ..................................................................................................................... 32 4. LIQUID CHROMATOGRAPHY (LC). .............................................................................. 33 4.1. Introduction .............................................................................................................................. 33 4.2. Technique principle ................................................................................................................. 33 4.2.1. Separation modes: ....................................................................................................... 34 4.2.1.1. Normal phase chromatography ................................................................... 34 4.2.1.2. Reverse phase chromatography ................................................................... 35 4.3. Basic description of the system ............................................................................................. 36 4.3.1. Mobile phase ................................................................................................................ 37 4.3.2. Solvents pumping system ........................................................................................... 38 4.3.3. Sample introduction injector .................................................................................. 38 4.3.3.1. Manual injector .............................................................................................. 38 4.3.3.2. Automatic injector ......................................................................................... 39 4.3.4. Columns ....................................................................................................................... 39 4.3.4.1. Silica-based stationary phases ...................................................................... 39 4.3.4.2. Chemically modified phases......................................................................... 40 4.3.5. Detectors ...................................................................................................................... 40 4.3.5.1. UV-VIS detectors .......................................................................................... 41 4.3.5.2. Fluorescence detector ................................................................................... 43 4.3.5.3. Electrochemical detectors ............................................................................ 43 4.3.6. Data acquisition and processing ............................................................................... 44 4.3.6.1. Area percentages ........................................................................................... 45 4.3.6.2. Area normalization ........................................................................................ 45 4.3.6.3. External standard .......................................................................................... 46 4.3.6.4. Internal standard............................................................................................ 46 4.4. Minimum installation and operation requirements ............................................................. 47 4.4.1. Bench requirements .................................................................................................... 47 4.4.2. Electric network .......................................................................................................... 47 4.4.3. Environmental conditions ......................................................................................... 48 4.5. Basic precautions ..................................................................................................................... 48 5. CHROMATOGRAPHY SYSTEMS CONNECTED TO MASS DETECTORS .. 50 5.1. Introduction .............................................................................................................................. 50 5.2. Technique principle ................................................................................................................. 50 5.3. Basic description of the system ............................................................................................. 51 5.3.1. Ionization source ......................................................................................................... 51 5.3.1.1. Electrons impact (EI) ................................................................................... 51 5.3.1.2. Chemical ionization (CI) .............................................................................. 51 5.3.1.3. Electrospray (ES)........................................................................................... 52 5.3.2. Mass analyzers ............................................................................................................. 54 5.3.2.1. Quadrupole mass analyzers.......................................................................... 54 5.3.2.2. Ion trap quadrupole mass analyzers ....................................................... 55

5.3.2.3. Tandem mass spectrometry. ......................................................................... 57 5.3.3. Detectors ...................................................................................................................... 58 5.3.3.1. Electrons multipliers ..................................................................................... 58 5.3.3.2. Microchannel plates ...................................................................................... 59 5.3.3.3. Photomultiplier .............................................................................................. 60 5.3.4. Data acquisition and processing ............................................................................... 60 5.4. Minimum installation and operation requirements ............................................................. 61 5.4.1. Bench requirements .................................................................................................... 61 5.4.2. Electric network .......................................................................................................... 61 5.4.3. Environmental conditions ......................................................................................... 62 5.5. Basic precautions ..................................................................................................................... 62 5.5.1. Vacuum system ............................................................................................................ 62 5.5.2. Ions source ................................................................................................................... 63 5.5.3. Nitrogen system .......................................................................................................... 63 5.5.4. Basic precautions training .......................................................................................... 63 5.5.5. Autotune files .............................................................................................................. 63 6. ANALYTICAL INSTRUMENTS PERFORMANCE CHECKING.. ....................... 64 6.1. Introduction .............................................................................................................................. 64 6.1.1. DQ Design Qualification ....................................................................................... 65 6.1.2. IQ Installation Qualification ................................................................................... 65 6.1.3. OQ Operation Qualification ................................................................................... 65 6.1.4. PQ Performance Qualification ............................................................................... 66 6.2. Ultra violet visible (UV-VIS) spectrophotometry ............................................................ 66 6.2.1. Preventive maintenance ............................................................................................. 66 6.2.2. Operation Qualification (OQ) .................................................................................. 67 6.2.3. Performance Qualification (PQ) ............................................................................... 67 6.2.3.1. European pharmacopoeia ............................................................................ 67 6.2.3.2. United States Pharmacopoeia ...................................................................... 68 6.3. Gas chromatography GC .................................................................................................... 68 6.3.1. GC preventive maintenance ...................................................................................... 68 6.3.2. Operation qualification .............................................................................................. 68 6.3.2.1. Flows control ................................................................................................. 68 6.3.2.2. Temperatures control .................................................................................... 69 6.3.2.3. Detector(s) signal precision.......................................................................... 69 6.3.2.4. Automatic sampler precision ....................................................................... 70 6.3.2.5. Calculations by the data system ................................................................... 70 6.4. Liquid chromatography - HPLC............................................................................................ 70 6.4.1. HPLC Preventive maintenance ................................................................................. 70 6.4.2. Operation qualification .............................................................................................. 71 6.4.2.1. Detector qualification ................................................................................... 72 6.4.2.2. Pump qualification ........................................................................................ 72 6.4.2.3. Columns oven qualification ......................................................................... 73 6.4.2.4. Automatic sampler qualification.................................................................. 73

6.4.2.5. Data system .................................................................................................... 73 6.5. Chromatography systems connected to mass detectors ..................................................... 73 6.5.1. Mass detector systems preventive maintenance...................................................... 73 6.5.1.1. Vacuum systems ............................................................................................ 73 6.5.1.2. Mass detector ................................................................................................. 74 6.5.1.3. Chromatographer .......................................................................................... 74 6.5.2. Operation qualification .............................................................................................. 74 6.5.2.1. Injector precision ........................................................................................... 75 6.5.2.2. Injector linearity ............................................................................................. 75 6.5.2.3. Carry-over ....................................................................................................... 75 6.5.2.4. Detector linearity ........................................................................................... 75 6.5.2.5. Mass accuracy ................................................................................................ 75 6.5.2.6. Sensitivity ........................................................................................................ 75 7. REFERENCES. ................................................................................................................................. 76

Volume II / Module 3 - Analytical Instrumentation

1. INTRODUCTION
The use of analytical instrumentation techniques in laboratories means a requirement not only to characterize a substance or assure the quality of a product, but also to increase the analyses productivity; therefore, there is an increasing number of high automation grade instruments that can be remotely operated. However, we have to be aware that for a satisfactory use and performance of such instruments, some basic rules are to be regarded, such as: the usage purpose, adequate installation, performance evidences, trained personnel to operate, compliance with the manufacturers instructions concerning basic precautions to be considered by the operator and routines for replacing consumable materials, personnel with technique reliable knowledge for an occasional implementation of an analytical method, preventive maintenance and performance check programs compatible with the routine used, parameters evaluation to guarantee a continuous quality assurance of results and, finally, a detailed planning always including evidences of the current activities. This chapter encompasses the main analytical instrumentation techniques used in a bioequivalence laboratory so that the professional in this field can start his/her studies and use this material as a reference. Another purpose was to discuss the performance checking of each instrument and validation concepts applied to analytical instrumentation.

7 of 76

Volume II / Module 3 - Analytical Instrumentation

2. ULTRA VIOLET-VISIBLE (UV-VIS) SPECTROPHOTOMETRY

2.1. Introduction The Ultraviolet Visible spectrophotometry is an analytical technique based on the property of several ionic or molecular specimens to absorb ultraviolet and visible radiations in solution. Radiations at these regions involve photons with sufficient energy to cause the transition of valence electrons, as a function of disturbances that start occurring. The disturbance caused by electric and magnetic fields travels through space, thus creating the expression electromagnetic radiation. We should have in mind that the electric and magnetic fields are perpendicular fields between themselves and that the propagation values where the fields maximums and minimums are always coincident, that is, the fields are in phase, such as in a wave. We can see in Figure 1 that while propagating to the right at a velocity V, any point fixed on the Xaxis successively starts to have wave peaks and troughs. Then, the wavelength (l) is defined as the distance traveled by the wave so that both maximums reach a fixed observation point. The wavelength is usually expressed in nanometers (nm). Figure 1. Wavelength

The Ultraviolet - Visible (UV-Vis) radiations constitute a relatively small portion of the electromagnetic spectrum including other forms of radiations as radio waves, according to figure 2. The limits of such regions are determined by the practical limits of experimental methods for the production and detection of radiations. The spectral regions differentiation has an additional meaning for the chemist, once physical interactions follow different mechanisms and provide different kinds of information. When a radiation arrives a semitransparent substance, the radiation is only partially transmitted. The remaining radiation is reflected or absorbed at several angles (figure 3), depending on the substance and the radiation wavelength. The type and amount of absorbed radiation is the most important for analytical purposes. Unfortunately, there is no direct method to determine the absorbed radiation; however, this information can be indirectly obtained by measuring the transmitted radiation.

8 of 76

Volume II / Module 3 - Analytical Instrumentation

Figure 2. Electromagnetic spectrum

Absorbed Radiation

Incident Radiation

Transmitted Radiation

Reflected Radiation

Figure 3. Absorbed, transmitted and reflected radiation If the intensity of a transmitted light is graphically represented as a function of the wavelength, an absorption spectrum of the substance is obtained. This is the radiation selective absorption that provides the base for the quantitative and qualitative analysis by molecular absorption spectrophotometer.

9 of 76

Volume II / Module 3 - Analytical Instrumentation

2.2. Technique principle 2.2.1. Light spectrum The visible light is a region of the electromagnetic spectrum, the radiations of which are able to sensitize the retina. The visible radiation comprises the radiations with wavelengths located between 400 and 700 nm, although such limits are not fixed as the perception capacity varies according to the observer. Within this wavelengths range, subgroups can be separated according to the color produced in a small spectrum range, as the table below. Table 1. Wavelengths ranges in the visible region

2.2.2. Absorption processes Considering that light is a form of energy, the absorption of light photon by a molecule results in an energy increase, E, of the energy contained in the molecule. The amount of increase is exactly equal to the photon energy, that is E = hn. The absorption process is schematically represented in the energy levels diagram simplified in figure 4.

Figure 4. Energy diagram


10 of 76

Volume II / Module 3 - Analytical Instrumentation

If the molecule is initially in a normal or fundamental state, E0, before the interaction, the absorption process increases the energy contained in a higher level or in an excited state, E1. The energy exchange process by the absorbed light is not attenuated, by occurs as multiple energy units called quantum. The quantum (hen) is characteristic for each specimen absorbed. To be absorbed by a molecule, the energy photon should precisely correspond to the difference, ?E, between the two states of energy. The total energetic potential total of a molecule, except the nucleus energy, can be considered as the sum of the electronic, vibrational and rotational energies. The electronic energies are associated with the transitions of electrons inside the atom or molecule. This sum of energies is represented in figure 5 as a change in orbitals.

Figure 5. Electronic energy The vibrational and rotational energies are associated with molecular vibrations and rotations. In figure 6 a simple diatomic molecule is showed following a compression and elongation movement (vibration). A molecular rotation is exemplified in figure 7.

Figure 6. Vibrational energy

Figure 7. Rotational energy


11 of 76

Volume II / Module 3 - Analytical Instrumentation

The difference between the rotational energy states is relatively small, much smaller than the differences between the electronic energy states. The difference between the vibrational energy states is intermediate if compared to previous states; then: Absorptions associated with transitions among rotational energy levels are generally found in the low energy or high wavelength region of the electromagnetic spectrum, that is, the remote infrared region. Absorptions associated with transitions among electronic energy levels are found in the high energy or low wavelength region, that is, the ultraviolet - visible region of electromagnetic spectrum. Absorptions associated with transitions among vibrational energy levels are found between the two levels mentioned above, in the infrared region.

The several kinds of energy transitions are not independent, but interconnected. Rotational energy states are superposed on the vibrational states, and both are superposed on the electronic states. 2.2.3. UV - Vis absorption spectrum The presence of unsaturations or multiple bonds are widely known as a characteristic process of ultraviolet absorption while the saturated compound is transparent. According to the theory of molecular orbitals, the electrons involved in single bonds, such as the C H bond, are called sigma electrons (). Double bonds, such as C = O, involve pi electrons ( ). In the close infrared region, transitions involving pi electrons allow a better observation of the absorption bands. Electrons not bonded to the molecule, contained in atoms such as oxygen or nitrogen, are called n electrons, and the interactions between pi and n electrons are responsible for a great number of characteristic ultraviolet absorption bands. A chromophore is a group that, when having a saturated hydrocarbon introduced, produces a compound having and absorption band between 180 and 1000 nm. For example, n-octane is a saturated hydrocarbon transparent in this region; however, if a nitrile group is introduced in the octyl radical, the octyl-nitrile compound is produced. The nitrile group is classified as a chromophore and would show light absorption signs in the ultraviolet range. Table 2 shows a small group of chromophores simple with their maximum wavelengths and molar absorptivity. The molar absorptivity is the measure of the absorption band intensity. As it can be seen, the absorption intensity of several chromophores notably varies from one group to another. However, all members of a class usually have equal intensity absorption bands and occur in a relatively narrow spectral range. For example, data showed for acetic acid are typical of saturated carboxylic acids resulting from formic acid to stearic acid, C1 to C18, all of them having absorption bands in the region of 204 to 210 nm and molar absortivities from 40 to 75.

12 of 76

Volume II / Module 3 - Analytical Instrumentation

Table 2. Chromophores Group

If one or more chromophores occur in a molecule, their relative positions determine the effect produced in the spectrum. 1-hexane has a molar absorptivity of 10000 at 180 nm and 1,5-hexadiene has a molar absorptivity of 20000 at the same wavelength. 2,4-hexadiene, with band in another region, has a molar absorptivity of 25500 at 227 nm. The position and the intensity of the absorption band of 1,5-hexadiene are approximately what we can expect from two individual propylene molecules. Obviously, the two chromophore groups are widely spaced and do not interact. However, when both chromophores are conjugated as in 2,4hexadiene, the effect is not the expected from two individual molecules. Similar observations of a great number of compounds have construed the basic argumentation for some general rules: When two chromophores in a same molecule are separated by more than one carbon atom, the absorption spectrum is a simple sum of the spectrum of each chromophore. When two chromophores in a same molecule are adjacent, the maximum absorption is observed at high wavelengths and the intensity is increased. When two chromophores in a same molecule are bonded to a same carbon atom, the result is intermediate between those above.

In the ultraviolet-visible spectrophotometry, the most interesting compounds are usually those having more than one chromophore, especially when the chromophores are conjugated (conjugated systems). Compounds with at least two conjugated chromophores absorb within the visible range. Saturated carbohydrates, fats, oils, simple ethers, acids, most of the carbohydrates and proteins do not absorb light in the visible range, since their structures have no chromophores. Molecules with the same chromophore and similar electronic structure have similar absorption spectra. The presence of several neighbor chromophores groups changes the molecular absorption spectrum character. The possible resultants can be divided into 4 groups: Bathochromic Effect. Displacement of the absorption bands towards the red range of the spectrum, thus intensifying the color. Hypsochromic Effect. Displacement of the absorption bands towards the violet range of the spectrum (of a lower wavelength). Hyperchromic Effect. Increase of the absorption bands intensity. Hypochromic Effect. Decrease of the absorption bands intensity.
13 of 76

Volume II / Module 3 - Analytical Instrumentation

The auxochromes are groups that, when introduced in a system as a chromophore, increase the wavelength of the maximum absorption band, the so-called bathochromic effect. The auxochromes have no own absorption band at low wavelengths. For example, the hydroxyl group is an auxochrome. Alcohols are transparent and used as solvents at low wavelengths, however, the introduction of a hydroxyl group in a system having a chromophore will lead to a bathochromic effect. The typical auxochromes are the amine groups and their substituted derivatives, halogens, alkaline group and others. 2.2.4. Effect of solvents The spectrophotometric analyses in the UV region are generally performed in solutions. Low concentrations of the interest analyte are usually employed; high purity grade solvents have to be used since impurities may impair the results due to its high concentration in relation to the analyte. Further, solvents for UV spectrophotometry should have a high optical stability grade; therefore, it is not recommended to use commercial purity grade solvents that will hardly comply with the abovementioned requirements. Saturated carbohydrates, such as n-hexane, are the most common solvents in UV spectrophotometry and should be specifically prepared for this purpose. Their electronic absorptions caused by the electrons transitions occur out of the analyzed range. Solvents with heteroatoms, for example, H3COH-, are also used. Due to their electronic transitions, such solvents have weak absorption bands. This issue should be also focused based on the solvent degree of polarity. The absorption bands of several substances are clearer and may have fine structures when measurement is made in low bipolar moment solvents. The solvent interactions solutes are stronger when strong bipolar forces are involved. Figure 8 shows two spectra obtained using phenol in iso-octane and ethanol. One can observe the influence of the organic solvent on the determination of substances.

Figure 8. Phenol in iso-octane (____) and ethanol () A difference in pH may also lead to fluctuations in case of this wavelength displacement.
14 of 76

Volume II / Module 3 - Analytical Instrumentation

2.3. Basic description of the system The system, figure 9, including a UV-VIS spectrophotometer, comprises: 1. 2. 3. 4. 5. Light source Monochromator Sample compartment Detector Reading system

Input Slot Dispersion Element Output Slot

Figure 9. UV-Vis spectrophotometer 2.3.1. Sources of radiation The UV-Vis spectrophotometry requires sources that can produce continuous radiation in the spectral regions of interest. The radiation energy sources consist in material that can be excited by electric heating or high voltage discharge. When returning to energetic levels, the excited materials emit photons having energies that correspond to the differences between the energies from excited states and energies from fundamental states. Some materials have so numerous and close energetic levels that the emitted frequencies form a relatively wide range of continuous radiation. The sources of radiation used in the production of UV-Vis spectrophotometry equipment should meet some requirements: The source should provide a beam with sufficient radiating power to allow the detection by appropriate means. The source should generate continuous radiation, that is, comprising all wavelengths within the spectral region the equipment will operate.

15 of 76

Volume II / Module 3 - Analytical Instrumentation

The source should be stable, that is, the light beam power generated should remain constant during the determinations. In single-beam equipment, this condition is critical so that the absorbance measurements can be reproduced. In double-beam equipment, they are simultaneously measured; therefore, high stability of the source is not so critical.

The most appropriate source for obtaining continuous radiation in the visible region is the glass bulb tungsten incandescent lamp. The filament operates at 2600-3000K, supplying continuous radiation from 350 to 2500 nm. The energy emitted in the visible region approximately varies with the fourth potential of the operation voltage; therefore, the source requires a strict voltage control to generate radiation or electronic voltage regulators. The most used ultraviolet radiation sources are the hydrogen or deuterium discharge lamps with quartz window. When at low pressure and subject to electric discharge, hydrogen produces a continuous beam in the ultraviolet region. The deuterium lamp produces a continuous spectrum from 180 to 380 nm. When operating, the high frequency xenon lamps produce radiation pulses covering the spectrum from 180 to 1100 nm. 2.3.2. Monochromator (optical) Monochromators are devices based on diffraction networks that disperse complex radiation in their component wavelengths and then isolate the desired spectral range. Monochromators allow the free isolation of spectrally pure and very narrow wavelength ranges along the close ultraviolet, visible and infrared regions. The photometric performance of a spectrophotometer is directly related to the monochromator quality (optical system). The monochromator setting should provide high resolution of the analyzed sample wavelength and, at the same time, a greater amount of transmitted light and a low amount of spurious light (amount of light that was not duly isolated by the optical system). The optical system setting has to allow: High transference efficiency between the source of light and the monochromator. Adequate monochromator output slot to avoid excessive light loss. Greater focusing of the light source beam on the sample.

16 of 76

Volume II / Module 3 - Analytical Instrumentation

Figure 10 below shows a monochromator normally used in the manufacture of UV-Vis spectrophotometers.

Figure 10. Czerny Turner Monochromator There are additionally other usual settings, such as: Littrow, Ebert and Concave Grating. These monochromators settings can be seen in the following types of UV-VIS spectrophotometers: Single-beam UV-VIS Spectrophotometers Diodes Arrangement UV-VIS Spectrophotometers Double-beam UV-VIS Spectrophotometers

2.3.2.1. Single-beam In this type of spectrophotometer, the monochromators light beam dispersion is totally focused on the sample compartment and, as a result, the unabsorbed light is directed to the detection system comprising a detector; it is usually of the photodiode type. The manufacture of single-beam devices requires high quality stable precision components. The direct reading spectrophotometers operate with moderate precision around 1% transmittance; these are relatively cheap and easy-and-fast-to-operate devices with easy maintenance. The singlebeam device requires the incorporation of zero circuit to measure transmittance in order to improve the work accuracy.

17 of 76

Volume II / Module 3 - Analytical Instrumentation

2.3.2.2. Diodes arrangement This kind of spectrophotometer has a simple monochromator configuration, figure 11, using a single diffraction grid and a detection system comprising several photodiodes; a polychromatic light beam initially reaches the sample compartment and right after the transmitted light is directed to the monochromator; after dispersion, the light reaches the detection system.

Figure 11. Polychromatic monochromator 2.3.2.3. Double-beam The double-beam spectrophotometers separate the original radiation in space (by means of a mirror) or in time (by means of a rotary sectorial mirror). This kind of spectrophotometer is typically used in analyses requiring greater analytical precision. The light beam dispersed on the monochromator (figure 12) is synchronized by an optical device that will conduct the beam both to a sample compartment (also called analytical beam) and another reference compartment (also called reference beam). The light transmitted on both compartments is synchronized again with the detection system; as a result, interferences can be eliminated, such as: energy source fluctuations, electric voltage variations, as well as the solvent used in preparing the sample. Thus, the source power fluctuations, the detector response and the amplifier gain are offset by the difference in signs. The double-beam spectrophotometers are mechanically and electronically more complicated than the single-beam devices and, as a result, construction and maintenance are more expensive. Diffraction Grid Input Slot Synchronism Optical Device Sample Compartment Figure 12. Double-beam monochromator
18 of 76

Volume II / Module 3 - Analytical Instrumentation

2.3.3. Sample compartments The UV-Vis spectrophotometer samples compartment has a wide range of connections and accessories that allow us to expand the instruments possible applications. The basic samples compartment is equipped with a rectangular sample support for 10 mm quartz cuvettes. According to the application, it is possible to introduce sample supports for cuvettes varying from 1 to 100 mm long. 2.3.3.1. Optical fiber coupler In several instances, the spectrophotometer is classically used to measure solutions, that is, using cuvettes. However, some samples are difficult to analyze due to their type or size. In order to facilitate the analysis of such samples, an optical fiber coupler system can be used; some benefits are reached: Handling of samples that will be at high temperatures and/or pressures or radiated is avoided. Measurement of very small samples, such as crystal, or very large samples, such as metal structure windows that, otherwise, could never be made using devices with normal settings. Monitoring of chemical processes. Determinations in closed systems (reactors).

A series of optical fiber sensors can be used to reach the objective above, such as reflectance, absorbance, glass transmittance and fluorescence sensors. 2.3.3.2. Sipper A system comprising a peristaltic pump can be used to pump the sample solution to a flow cuvette without requiring manual transference of the sample to the cuvette, thus reducing the sample handling. Such flow cuvettes can be positioned on special supports that can heat the sample. 2.3.3.3. Support for solid samples This kind of support is intended to provide analytical results in transparent solid samples, such as glasses, plastics and lenses. In addition to the easy instrument connection, this support can also have a version in which the transparent sample travels in front of the optical beam so as to obtain the sample homogeneity profile. A rotary device can be used to check the behavior of this solid and transparent sample, but having different incident radiation angles.

19 of 76

Volume II / Module 3 - Analytical Instrumentation

2.3.3.4. Reflectance accessories Solid non-transparent samples can be analyzed and the sample profile (reflectance spectrum) can be obtained. There are several kinds of reflectance accessories available and that can be coupled to the spectrophotometers, for example: mirror reflectance, diffused reflectance, controlled temperature and pressure reflectance, powders and creams reflectance and solid sample color comparisons. 2.3.4. Data acquisition The photosensitive devices used in transmittance measuring devices convert the radiating energy into an electric signal. The photosensitive detectors should respond to the radiating energy on a wide spectral range. They should be sensitive at low light levels e quickly respond to the incident radiation. It is crucial that the electric signal generated is directly proportional to the incident beam power, that is, R = kP + k Wherein R is the detector electric response in current units, resistance or e.m.f. The constant k is a measure of the detector sensitivity in terms of electric response per unit of radiating power. Certain detectors display a small k response called dark current, even when they are not receiving incident radiation. The radiating energy detectors currently used in the manufacture of measurement equipment are photodiodes or photomultipliers. The signs detected by the photomultiplier or photodiodes after amplification generate a wavelength X transmittance or absorbance intensity graph called ultraviolet-visible spectrum. The identification is possible by comparing this spectrum to the spectrum of a standard substance. The quantification is also possible by fixing a wavelength and obtaining the corresponding transmittance or absorbance value. This value will be proportional to the sample component concentration. The actual concentration is determined by using a standard with a known concentration. 2.4. Minimum installation and operation requirements In order to assure reliable results in compliance with the Installation Qualification (IQ) requirements, the installation and operation requirements have to be met; in this case, the following can be mentioned: Electrical network: that meets the manufacturers specifications, with special concern to the electrical network stabilization. Instrument located under appropriate humidity and temperature conditions. Operational knowledge to use the equipment correctly.

20 of 76

Volume II / Module 3 - Analytical Instrumentation

2.5. Basic precautions Some precautions should be taken to keep the equipment life and performance. Such precautions are related to two kinds of action: user actions and preventive maintenance actions usually taken by the manufacturer. This section discusses the user actions. The manufacturer actions are described in the performance checking section. Daily actions to be performed by the user are crucial to reduce damages that can be caused to the instrument. Such actions are usually simple and can be summarized as follows: Equipment external cleaning. Sample compartments cleaning. Cuvettes washing. Consult the manuals / equipment dealer to clarify operation/maintenance doubts. Cleaning of the room windows interfacing the light beams and the sample compartments. The use of safety material when handling samples is critical for the operators safety. The personal protection equipment required includes: safety apron, surgical gloves and wide-view goggles.

The equipment should have consumables so that, is required, the operator himself can take actions such as changing lamps. The main consumable items required include: Deuterium and tungsten lamps, or xenon lamp. Quartz cuvette.

21 of 76

Volume II / Module 3 - Analytical Instrumentation

3. GAS CHROMATOGRAPHY (GC)

3.1. Introduction Gas chromatography is an analytical technique used to separate mixtures of chemical substances. The purpose of the technique is to allow a separation that helps in identifying and quantifying the substances contained in the mixture. As the main element used in the gas chromatography (GC) technique, the first chromatographic columns were packed in a typical tube between 1 to 5 m long and 2 to 4 mm diameter. As the analytical instrumentation advanced, open tubes of melted silica (capillaries) are currently used, the length of which range from 10 to 100 m and the internal diameters from 0,1 to 0,8 mm , thus providing better results. 3.2. Technique principle The sample is introduced in the system using a syringe or an injection valve. The sample is volatilized right after being introduced in the chromatography system and dissolved in an inert carrier gas (mobile phase). When dissolved in the gas, this sample will be carried to the chromatographic column containing the liquid or solid phase as a stationary phase, thus separating the compounds in this column. Separation is accomplished because the sample components have different affinities to the stationary phase, resulting in different elution speeds of the components by the column, that is, the higher is the affinity of the component to the stationary phase, the slower it will travel through the column and vice-versa.

Injection

Migration

Elution

Figure 1. Basic scheme of a chromatographic separation process

22 of 76

Volume II / Module 3 - Analytical Instrumentation

After separation, the sample components reach the detector that generates an electrical signal by means of a physicochemical phenomena (for example, difference in thermal conductivity, flame ionization). These detectors can be selective to a certain chemical group (for example, organochlorinated compounds) and more sensitive for the analysis. As the signal is usually at a very low intensity, it is amplified and transmitted to a data station where it can be viewed in a signal intensity X time graph. This graph is called a chromatogram. The signal is proportional to the concentration or mass of each compound; there should be different responses from a compound to another (the most abundant compound not always generate the higher signal). For this reason, the sample can be quantified. The GC allows the analysis of several substances, including some inorganic gases such as O2, CO, CO2, NO2, etc., as well as thousands of organic substances of several functional groups, such as, alcohols, ketones, amines, aromatics and others. By using highly selective detectors, it is possible to determine very small amounts of components contained in the sample, for example, in the order of picograms (traces); as well as greater amounts, for example, in the order of percentage.

3.3. Basic description of the system A basic system includes the following modules: 1. System of gases: The most common gases used as mobile phase (carrier gas) are helium, hydrogen, argon and nitrogen. Other gases can be also used. The use of auxiliary gases is required depending of the detector to be employed. 2. Flow and pressure controllers: Used to maintain the mobile phase outflow uniformity. They are further used to measure the sample splitting ratio (splitter) when injectors for capillary columns are used and also to control the outflow of auxiliary gases of the detectors. 3. Injectors: Place where a sample is effectively introduced in the system. The injectors can be designed for packed or capillary columns; there are different techniques for using them (oncolumn, flash vaporization, split, splitless, etc). The sample can be also introduced using sampling valves, mainly for gases. Automation can be used to increase productivity and reproducibility. 4. Columns: The substances contained in the sample are separated inside the columns. The columns can be made of glass, stainless steel, nickel, teflon or melted silica; they can be packed or capillary. The capillary columns allow better separations. 5. Detectors: Devices that monitor the exit of eluted substances from the column, generating an electric signal that is proportional to the substance mass or the concentration. Detectors can be of the universal- or selective-type. These detectors are connected to signal amplifiers, called electrometers, which amplify the signal and transmit it to the output data devices. 6. Data acquisition and processing system: A system that collects data generated in the chromatographer, performs the necessary calculations for quantifying the substances and generates the chromatogram. This system can be integrators or integration software (PC).

23 of 76

Volume II / Module 3 - Analytical Instrumentation

Injector Pressure Controller Detector GC Pneumatics Chromatogram

Carrier Gas Column Workstation

Column Oven

Figure 2. Gas chromatography (GC) system 3.3.1. Gases used Two classes of gas can be used in the chromatographic system: carrier gases (mobile phase) and auxiliary gases. The most common carrier gases are nitrogen, argon, hydrogen and helium. The most indicated gas is helium, due to the better chromatographic performance when compared to other gases; however, some decisive aspects when choosing the carrier gas are to be considered: Detector compatibility: it is critical to determine that the carrier gas is compatible with the detector and lead to a loss of sensitivity or excessive noise. For example, when analyzing hydrogen gas in TCD, nitrogen and argon impart greater sensitivity than helium, as the thermal conductivity difference between the two first gases and hydrogen is greater than the thermal conductivity difference between hydrogen and helium; Costs; Availability; Performance; Purity: The purity of the gases should be respected, depending on the application. The manufacturers manuals contain all the required information.
24 of 76

Volume II / Module 3 - Analytical Instrumentation

3.3.2. Flow and pressure controllers The carrier gas flow or pressure is an important parameter that depends on the column diameter and length. It should be controlled during the analysis. This control is performed using flow or pressure controlling valves. This parameter can be adjusted manually, with the aid of a flow meter or a bubble meter, or electronically using a software; the purpose is to obtain a better chromatographic efficiency. For auxiliary gases, control is required since the flow adjustment is critical to reach the best performance. 3.3.3. Injectors The basic function of an injector is to introduce the sample into the chromatographic system. At the injector, the sample contacts the carrier gas and is dissolved within it. Therefore, in case of liquid samples, the injector has to be heated so as to assure a total volatilization of the sample. There are two categories of injectors: injectors for packed columns and injectors for capillary columns. Gaseous samples can be injected with the aid of a gas sampling valve provided or not with an injector in series with the valve. For each class of injectors, some options are available depending on the application. 3.3.3.1. Packed injectors These injectors are used with packed columns. They can be used in megabore-type capillary columns (internal diameter of column: > 0,53 mm). All the sample is directly injected on the head of the chromatographic column. This is to eliminate any and all loss when transferring the sample to the column. This injector is indicated to clean and diluted samples, in addition to samples having a great volatility variation of compounds.

Septum

Injector Body Gas Inlet

Column

Figure 3. Packed injector


25 of 76

Volume II / Module 3 - Analytical Instrumentation

3.3.3.2. Capillary injectors Injectors designed for use in capillary columns. The main injectors include: Split/Splitless Injector: The most commonly used injectors. Allows samples to be divided (split) so as to allow the use of capillary columns (as they are able to receive a small amount of sample). There are two operation modes: Split mode The injected sample is split between the column and the injector disposal outlet (vent). The valve in charge of this splitting action is called splitter and is used with concentrated samples (in the order of mg/mL).

Splitless mode Used with diluted samples. In this case, the splitter will remain closed during the solute transference to the column; next, the splitter is opened to dispose the solvent.
Carrier Gas Inlet Septum Purging Injector Disposal

Injector Disposal Column

Figure 4. Split/splitless injector On-column and/or programmable temperature injector: Allows two operation modes: On-column injection - For samples with wide molecular weight range, eliminating the mass discrimination effect. On-column injection with programmable temperature For low boiling temperature or low concentration thermolabile samples. Known in the market as PTV or SPI Injector.
Septum Septum Purging Glass Linner Cooling Inlet

Cooling Outlet

Injector Body

Heater

Carrier Gas Inlet Nut Washer Column

Figure 5. Programmable temperature injector


26 of 76

Volume II / Module 3 - Analytical Instrumentation

3.3.4. Columns The column is the main part of the chromatographic system where the separation of the sample components takes place as a result of the different affinities of the substances with the stationary phase, migrating at different speeds through the column. The columns can be classified into two groups: Packed columns: The oldest chromatographic columns. Usually made of stainless steel or glass tubes, however, other materials can be used. A stuffing (an adsorbent or a solid support impregnated with a film of a substance having a low vapor pressure) is introduced in this tube. Presently, these columns are not frequently used due to the lower performance when compared to capillary columns. Packed columns are more commonly used in the analysis of inorganic gases.

Capillary Columns: The most commonly used columns presently. They are usually made of melted silica tubes externally coated with a polyamide film, thus making the column very flexible. The stationary phase (a film or a solid adsorbent) is deposited on the inner wall. This kind of column has great separation capacity and a wide range of substances can be separated by each kind of column. The phase choice determines the column selectivity. Hundreds of stationary phases are available in the market with several trade names. They are usually chosen according to the polarity of the substances to be analyzed. Polar stationary phases are typically more reactive, thus having a working limit temperature lower than the non-polar columns.

POLYAMIDE FILM MELTED SILICA TUBE STATIONARY PHASE

WCOT
WALL FILM OPEN TUBE

PLOT
POROUS BED OPEN TUBE

SCOT
WALL SUPPORT OPEN TUBE

Figure 6. Types of capillary columns To choose the most appropriate column, the following parameters should be assessed: Columns internal length and diameter Film thickness (capillary); Liquid stationary phase; Solid Supports; Solid Stationary Phases.

27 of 76

Volume II / Module 3 - Analytical Instrumentation

3.3.4.1. Columns oven The chromatographic column is stored in the oven. For each kind of analysis, the oven should be set at a working temperature; at this temperature, the sample components of interest are separated and having the shortest elution time. A good oven has measurement precision and a wide ramp programming range. The use of cryogenics allows the analysis of more volatile compounds. 3.3.5. Detectors There are several kinds of detectors: universal detectors (Thermal Conductivity (TCD), Mass Spectrometry (MS)), and selective or specific detectors (Flame Ionization (FID), Electrons Capture Table 1. Types of detectors and their use

The most commonly used detectors in gas chromatography (GC) will be discussed. 3.3.5.1. Flame ionization detector (FID) The most used detector. Some features include: easy operation, high sensitivity and incomparable linearity. The figure below shows a scheme of the FID detector.

28 of 76

Volume II / Module 3 - Analytical Instrumentation

Collector

Ceramic insulator

Signal probe

Ignition probe Ignition spring Flame tip

Figure 7. FID Detector The detector operation involves a mixture of gases, air and hydrogen, thus forming a flame having a temperature of about 2000C. The sample is burned on this flame. A pair of electrodes is positioned close to the flame to capture the signal generated by ionized specimens and electrons resulting from the substances burned on the flame. This detector generates a signal corresponding to the compounds having carbon and hydrogen in their molecule. Rare exceptions include CS2. The detector should be operated at a temperature of about 400 C. However, the chosen temperature should assure that substances eluted in the column do not condense. The typical concentrations to be analyzed include the range from low (ppb) to high values (%). 3.3.5.2. Nitrogen and phosphorus (NPD) or specific thermoionic detectors (TSD) The detector has selective response to compounds containing nitrogen or phosphorus atoms. Internally, the detector has a rubidium salt incorporated to a refractory pearl subject to the passage of gases, air and hydrogen. When electrically heated, the pearl produces a gaseous plasma at 600 to 900C. The plasma thus produced is enough to make the N- and P-containing substances respond to the detector.

29 of 76

Volume II / Module 3 - Analytical Instrumentation

Ceramic Insulator Collector Signal Probe Ceramic Pearl Flame Tip H2 and Make-up Air Column Figure 8. TSD or NPD detector 3.3.5.3. Electrons capture detector ECD This detector is sensitive to molecules having high electronic affinity atoms and has inside a radioactive source, typically 63Ni, which emits beta particles. These particles ionize the carrier gas, usually nitrogen, and generate a cloud of electrons inside the ECD cell. The electrons produce a stable background current through the ECD cell electrodes. This signal is amplified by the detector electrometer. When an electrons-absorbent specimen passes through the cell, the current is decreased as a result of the electrons capture by the absorbent specimen, thus decreasing the number of electrons in the cell. Insulation Nickel Source Cloud of Electrons Cell Collector Electrode Ceramic Insulators Cell Probe

Pearl Probe

Signal Probe

Make-up Column Figure 9. ECD detector


30 of 76

Volume II / Module 3 - Analytical Instrumentation

3.3.6. Data acquisition and processing Data acquisition and processing were discussed in item 4.6 of HPLC technique. 3.4. Minimum installation and operation requirements In order to assure reliable and reproducible results in compliance with the Installation Qualification (IQ) requirements, the conditions of the installation room should be verified. For that purpose, the following conditions are to be mentioned: Electrical network: it is necessary to check that electrical network voltage is in accordance with the equipment specification. Network stabilization and the outlet plugs position (inverted phase) should have special attention, as they can interfere with the equipment operation and life. Environmental conditions: The instrument should be installed only if local humidity and temperature conditions are appropriate for the operation. In addition, safety conditions, such as the proximity to flammable and corrosive agents should be assessed. Purity of Gases: Should be in accordance with the system chromatographic requirements, mainly concerning the detectors. Gases Filter: Should be used to assure the quality of the gases used and the system gas lines. Should be replaced on a periodic basis. Pressure Regulators: Should assure the minimum working pressures specified for the equipment. Should be also provided with safety devices, such as double-stage and metal diaphragms. Operational knowledge of the equipment: For the correct use of the equipment.

Table 2. Purity of gases

31 of 76

Volume II / Module 3 - Analytical Instrumentation

3.5. Basic precautions In order to assure good results, some precautions are required when working with GC. These are usually simple, but important procedures so that the equipment can operate properly. One of the most common precautions is changing the injector septum typically between 30 and 100 injections depending on the injection technique (manual or automatic). The injector glass inserter is also a very important item for the good performance of the equipment. Thus, replacement is required according to the application and usage. Another relevant precaution concerns the use of the columns that should be conditioned and subject to thermal cleaning whenever required. When heating the column, care should be taken not to exceed the stationary phase thermal limit. Concerning the gases, in addition to the purity precautions discussed above, it is important to replace the gases filters when required and also regularly check the cylinder pressure to avoid complete use, thus preventing occasional contamination by the environmental air admitted by diffusion. It is recommended to clean the detectors every six months or whenever performance decrease occurs. The cleaning procedures are described in the equipment operation manuals.

32 of 76

Volume II / Module 3 - Analytical Instrumentation

4. LIQUID CHROMATOGRAPHY (LC)


4.1. Introduction High performance liquid chromatography (HPLC) was created when liquid chromatography was applied to theories and instrumentations originally developed for gas chromatography. The indispensable high performance liquid chromatography (HPLC) is currently highlighted in terms of modern separation methods. The difference between liquid chromatography and high performance liquid chromatography (HPLC) rests on the use of stationary phases preferably with spheric microparticles (10.5 or 3 mm). As they are much less permeable, such phases are required for the use of pumps to elute the mobile phase. 4.2. Technique principle The chromatographic separation is based on the differential migration of components from a mixture due to the differences between two immiscible phases, the mobile and stationary phases, and when widening the bands depending on physical processes and not the difference in balance. These interactions can take place by hydrogen bond interactions, electrostatic interactions and hydrophobic and Van der Waals forces, and others. The differential migration results from the analytes balance difference between both immiscible phases, being determined by factors affecting this balance: the mobile and stationary phases composition and the separation temperature. Changes of any of these factors lead to changes in the differential migration. The liquid chromatography classification according to the stationary phase led to a wide variety. The first great division was the adsorption chromatography and the partition chromatography, referring to the solid and liquid stationary phases, respectively. Taking the nature of the interactions and the above mentioned phenomena into consideration, the separation modes can be classified as follows: Reverse phase chromatography, normal phase chromatography, ion pairing chromatography (ion exchange) and exclusion chromatography. If liquid, the stationary phases can be simply adsorbed on a solid support or immobilized on it. In the first case, chromatography is referred as partition chromatography. The partition chromatography was replaced by the chemically bonded-phases chromatography by virtue of the greater stability imparted when compared to the adsorbed liquid phases. These phases, having modified supports, are considered separately by differing from the other two modes in terms of separation mechanism. The great chromatographic development obtained from the chemically bounded liquid phases led these phases to be mainly used in analytical HPLC.
33 of 76

Volume II / Module 3 - Analytical Instrumentation

4.2.1. Separation modes: There is always dependence between the solute-mobile phase, solute-stationary phase and mobile phase-stationary phase interactions. Then, the choice of the separation mode depends on the choice of the stationary and mobile phases for each class of solute. Two retention modes in liquid chromatography were proposed. The first by Scott and Kucera, interaction-solvent, and the second by Snyder, competition-solvent. The two models are equivalent, since both consider that in a given separation, the interaction of the solute with the stationary phase remains constant. Therefore, retention is determined by the mobile phase composition. 4.2.1.1. Normal phase chromatography The stationary phase is more polar than the mobile phase; the opposite occurs in reverse mode chromatography. The solvents used are usually a mixture of organic solvents without the addition of water. The stationary phases are organic adsorbents (silica, alumina) or chemically bounded polar phases (cyano, diol, phenyl, amino). Both retention models, interaction-solvent and competition-solvent have been successfully used to describe the effect of the mobile phase in normal mode liquid chromatography. Regardless the model used, retention in normal phase increases as the mobile phase polarity decreases. This chromatography mode is mainly applied to neutral molecules, although it can be also used to separate ionizing or ionic molecules. The order of elution respects the following sequence: hydrophobic molecules (less polar) are eluted first, while the hydrophilic molecules (more polar) are retained. When the sample dissolution has problems in polar solvents, the reverse phase injection is difficult; the normal phase separation is then recommended. The elution solvent in the normal mode is selected by choosing a weak solvent and mixing it with a strong solvent so as to reach the desired strength. The presence of water traces in the mobile phase is the main cause of poor retention reproducibility when working on a normal phase, especially when unmodified silica is used as stationary phase. This problem has been solved by the use of anhydrous solvents with a known volume of water, methanol or acetic acid to deactivate the most reactive silanol groups of the stationary phase. Besides improving reproducibility, it also improves the peak shape. This same effect can be obtained by adding triethylamine, an essential compound in the separation of amines in silica gel. The use chemically modified silicas in normal mode elution has been preferred, since they offer interaction sites with the solute, in addition to a more homogenous surface when compared to silica gel that has a variety of silanol groups with different polarities. The chemically bounded phases are
34 of 76

Volume II / Module 3 - Analytical Instrumentation

useful for the chromatography of moderately polar compounds. However, these solutes can be also efficiently resolved in reverse mode elution. The choice between normal or reverse mode elution is usually more dependent of the matrix than the solute. 4.2.1.2. Reverse phase chromatography While in the normal phase chromatography a stationary phase is more polar than the mobile phase, in the reverse mode the mobile phase is more polar that the stationary. The reverse phase chromatography is more commonly used in HPLC, since it allows the separation of a wide variety of solutes and the use of aqueous mobile phases. The most used mobile phase is a mixture of acetonitrile/water; when required, acetonitrile is replaced with methanol and tetrahydrofuran (THF). The use of only three solvents is a result of the small amount of water-miscible organic solvents. In the normal mode, a wider range of solvents is available. The principle of the reverse phase separation is the hydrophobia, being mainly attributed to the interactions between the non-polar portion of the solute and the stationary phase, that is, the repulsion of this portion of the solute by the aqueous mobile phase. It is worth emphasizing that the solvent strength in reverse phase increases as the solvent polarity decreases. Thus, the water strength (weaker solvent) is lower than the methanol, acetonitrile, tetrahydrofuran and dichloromethane strength, considering the crescent order (see table 1). As a water-insoluble substance, dichloromethane is not used in the reverse phase; however, as dichloromethane is a very strong solvent, it is sometimes used to clean the reverse phase columns contaminated by strongly retained solutes. Table 1. Comparison among strengths of different mobile phases

35 of 76

Volume II / Module 3 - Analytical Instrumentation

Besides being used at a low absorption range in the ultraviolet, acetonitrile results in aqueous solutions with low viscosity (as desired). Therefore, together with methanol and THF, these are the most commonly used solvents to control the selectivity and the elution reverse mode separation. When for some reason water is not used in the mobile phase and apolar stationary phases are used, the chromatography is called non-aqueous reverse phase chromatography. This chromatography will be used when working with highly hydrophobic solutes, such as lipids and polymers; the solvent usually consists in a mixture of polar solvents, such as acetonitrile or methanol (solvent A), with a weaker solvent (B), such as THF, chloroform, dichloromethane, acetone, methyl-t-butyl ether. In this case, retention is also changed by the percentage of solvent B. There are other modalities of ionic compounds chromatography or size exclusion chromatography (gel permeation and filtration, applied to the molecules size separations, for example, polymers, biopolymers, proteins, peptides, etc). As they are barely applied in the separation of drugs and medicines, these modalities will not be discussed herein. Another important aspect involving the HPLC chromatographic separations is the separation mode: isocratic or gradient. In the isocratic mode, the separation conditions remain unchanged throughout the chromatographic run, being restricted to a single solvent composition. In the gradient mode, there is a mobile phase and/or flow composition variation during the chromatographic run. There is a number of relevant parameters to be considered during the development of a chromatographic method, such as the chromatographic resolution, selectivity, retention factor, etc. These parameters are extremely important when developing, validating and using HPLC chromatographic methods.

4.3. Basic description of the system A HPLC system basically consists in modules having specific functions and carefully designed to provide intended analyses with versatility, fastness, reproducibility and high sensitivity. Current HPLC equipment in the market may range from very simple, where samples are manually injected, to more complex systems provided with an automatic sampling module and controlled by computer software able to control the systems functions, acquire, process and print data, store and organize them for future reference.

36 of 76

Volume II / Module 3 - Analytical Instrumentation

For a better understanding, the HPLC system can be divided into six main modules:

Figure 1: Schematic representation of a HPLC system: 1. 2. 3. 4. 5. 6. Mobile phase reservoir Solvents pumping system (Pump) Sample introduction (Injector) Column compartment Detection system Data system

4.3.1. Mobile phase The mobile phase used in liquid chromatography is extremely important for obtaining the desired separation, as it contacts the analyte being separated, as well as a chromatographic column and the system as a whole. A wide range of solvents is available; however, there are some desirable properties for their use in HPLC: High purity; Non-decomposition of the analyte and the stationary phase; Compatibility with the detection system; Low viscosity; Sample dissolution; Low cost.

Among the solvents used in liquid chromatography, it is worthy emphasizing the characteristics and purity of water; the ultrapure water it is strongly recommended as highly reliable results are obtained. Its resistivity (typically 18.2 MO/cm at 25C) is an excellent quality indication. However, the organic compounds should be monitored since they can interfere with the sensitivity of UV detectors
37 of 76

Volume II / Module 3 - Analytical Instrumentation

commonly used in HPLC analyses. It is important to remember that the ultrapure water should be always immediately used as the storage modifies its conditions.

4.3.2. Solvents pumping system The main function of the Pump is to push the mobile phase through the column. Once they have compacted padding consisting in very small diameter particles, in the order of 3 to 10 um, the columns highly resist to the passage of the mobile phase; therefore, the pumping system should be able to overpass this barrier and provide a constant, reproductive flow with no pulsations. Most of the pumps used are of the reciprocating type, also called piston or diaphragm pumps. The operation is based on an electric motor connected to gears that move a pistons system. As this kind of pump produce pulsating flows due to the to-and-forward movement of the piston(s), some resources were developed to overcome this problem ; these mechanisms are called pulse mufflers. Analytical pumps are designed to operate at high pressures and flow rates varying from 0.01 to 10 mL/min. They are usually made of inert material, such as stainless steel (pumping heads, piping and connections), sapphire, quartz, ceramics and titanium (pistons), ruby (check valves) and inert materials (seals). 4.3.3. Sample introduction injector The injector is the module in which the samples are introduced in the HPLC system so as to allow separation in the column. The injector can be manual (the user introduces the sample with the aid of a straight tip microsyringe) or automatic, also known as auto-injector or automatic sampler, that is able to automatically inject a great number of samples and even perform the dilution, derivation or reagents addition operations. 4.3.3.1. Manual injector The most commonly used manual injectors are basically the valve-type injectors having an external sampling loop, defined and precise volume piping that can be replaced to allow the injection of different sample volumes. Figure 2 shows the valve schematic diagram. The six small circles represent the valve internal bores, while the large circle represents the injection syringe needle inlet.

38 of 76

Volume II / Module 3 - Analytical Instrumentation

Pump

Pump

Column

Column

Syringe

Figure 2. Schematic diagram of an injection valve The Valve has two positions: LOAD and INJECT. At the LOAD position, the pump flow is deviated directly to the column (enters at 2 and exits at 3). Meanwhile, the sample is introduced in the sampling loop (LOOP) through the needle bore. The excess is immediately disposed through exit 6, that is, the injection precision is determined by the volume of sample at the LOOP. At the INJECT position, the mobile phase now travels through the LOOP dragging the sample to the column (travel 2-1-4-3). 4.3.3.2. Automatic injector The automatic injectors are devices widely used in laboratories aiming high productivity, as they operate in an unassisted mode. They further provide better reproducibility when compared to the manual injectors. Some models are able to work as a thermostat for the samples, thus increasing the chromatographic system versatility. 4.3.4. Columns 4.3.4.1. Silica-based stationary phases Silica is undoubtedly the most important material used in stationary phases for HPLC. Silica is a versatile material and its surface can be modified by chemical derivation, then giving raise to several interesting materials such as stationary phases for liquid chromatography. Silica also allows work with different separation mechanisms. Silica is an amorphous, highly porous and partially hydrated material usually prepared by acid hydrolysis of sodium silicate, followed by emulsification in a mixture of alcohol/water and subsequent condensation, it is then washed and dried for use.
39 of 76

Volume II / Module 3 - Analytical Instrumentation

Silica surface results from its preparation conditions; silicas having a high amount of free silanols are more acid than silicas with bounded hydroxylated groups. Silica is widely used in normal phase chromatography; it is not recommended in reverse phase chromatography. 4.3.4.2. Chemically modified phases Currently, the most commonly used phases. Their application to separate polar compounds was widely disseminated as a result of the difficulties when using silica-based columns. With the advent of the chemically bounded phases, this gap was filled up and we can currently state that about 90% of the chromatographic separation is performed using such phases. They have an interesting mechanism of solutes retention by means of interactions not exclusively base on polarity. The use of octadecyl groups lead to the formation of the octadecylsilane phase, known as ODS or C18. This group imparts the phase a polar character compared to the unmodified silica. Retention on these phases depends on the amount of carbon present, usually expressed as percentage. The quality of the separation to which they are applied will depend on this percentage and the amount of residual silanols. For sterical reasons, this derivation does not affect all silanol groups. In some instances, the remaining groups have a problem with the peaks tail when interacting with the solute. This problem can be solved by reacting silica with trimethylchlorosilane after the derivation; as the trimethylchlorosilane is smaller, it can reach some of these groups, then forming trimethylsilanes. Although it is not possible to derive all the silanol groups, this process is called end capping. Some analyses require the column temperature to be stable and, for that purpose, there are devices available in the market; such devices are known as oven or column compartment. 4.3.5. Detectors Detectors are transducing devices connected right after the column outlet. They are responsible for generating electrical signals proportional to the compounds passing through them. Several kinds of detectors have been used in HPLC depending on the physical or physicochemical features of the sample and the mobile phase. An ideal detector should have the following characteristics: High sensitivity and low limit of detection; Fast response to all solutes; Insensitivity to temperature and mobile phase outflow changes; Independent response of the mobile phase; Small contribution for the peak widening by the extra volume of the detector cell; Response increasing linearly with the amount of solute;
40 of 76

Volume II / Module 3 - Analytical Instrumentation

Non destruction of the solute; Safety and usage convenience; Qualitative information of the desired peak; Low cost.

All characteristics can be hardly found in a single detector; therefore, the most appropriate detector should be chosen according to the sample and having as many features as possible. The most commonly used detectors available in the market include: Ultraviolet and visible (UV-VIS) absorbance detectors; Fluorescence detectors; Electrochemical detectors; Mass detectors.

Other detectors: Refractive index detectors; Electric conductivity detectors.

4.3.5.1. UV-VIS detectors These detectors operate based on the amount of light absorbed by the solute at a certain characteristic wavelength. They can be basically divided into three types: Fixed wavelength; Variable wavelength; Diodes arrangement.

4.3.5.1.1. Fixed wavelength detectors The simplest of all detectors. It is not very common nowadays, as they are sensitive to variations of the mobile phase composition, thus impairing its use in gradient systems. The wavelength is selected through specific wavelength optical filters. 4.3.5.1.2. Variable wavelength detectors The most commonly used detectors nowadays, as they can be used both in isocratic and gradient systems. Used by most of the laboratories in which the sample and its corresponding wavelength are known. Figure 3 shows the schematic diagram of a variable wavelength detector.

41 of 76

Volume II / Module 3 - Analytical Instrumentation

Diffraction net

Lamp

Cell Photodiode

Figure 3. Schematic diagram of a variable wavelength detector 4.3.5.1.3. Diodes arrangement detectors Detectors having the capacity to generate absorbance spectra at a speed compatible with the eluant flow. Provides tridimensional data: absorbance x time x wavelength. They are used in research and development when the best wavelength is not known, when the spectrum is desired for characterization purposes (library) and to obtain the purity percentage of the peaks. The schematic diagram is showed in figure 4.

Diffraction net (fixed)

Cell Lamp

Diodes arrangement

Figure 4. Schematic diagram of a diodes arrangement detector

42 of 76

Volume II / Module 3 - Analytical Instrumentation

4.3.5.2. Fluorescence detector The fluorescence detectors are specific, as the solute has to fluoresce. It is one of the most sensitive among the most used detectors in HPLC. Figure 5 shows a schematic diagram of this kind of detector.
Light source Sample cell Detector Absorption

Detector Emission

Figure 5. Scheme of the fluorescence detector Some fluorescence detectors in the market have the capacity to generate excitation and emission fluorescence spectra during the chromatographic run. 4.3.5.3. Electrochemical detectors The most sensitive of all detectors and also the most delicate to work with. The molecules to be analyzed should have characteristics that allow oxidation or reduction. Detection is performed by means of electrodes as showed in figure 6.

Auxiliary electrode Mobile phase Working electrode

Potentiometer

Recorder

Figure 6. Schematic diagram of an electrochemical detector


43 of 76

Volume II / Module 3 - Analytical Instrumentation

4.3.6. Data acquisition and processing As part of the chromatographer, the data acquisition and processing system will allow the user to verify and document the obtained results. The system will basically receive the electric signal sent by the detector(s), convert them into graphics called chromatograms, integrate the peaks using the predetermined parameters and, from a calibration curve, quantify the samples. This signal x retention time graph has two important features. First, the retention time is characteristic and repetitive for each substance; second, the peak area or height is differently proportional to the substance concentration or mass. See the figure below:

Retention Time Peak Area Peak Height

Baseline

Minutes

A file is initially created and contains the information required for the analysis of the interest substances, such as: Text describing the chromatographic conditions, such as, the mobile phase and column used; Programming of parameters for the chromatograms integration.

When the system stabilizes, standards are injected in order to check out the performance. When all parameters are set and correct, a standard containing all components to be quantified is injected so that a table is created; in this the table, the system is informed about the peaks to be analyzed through the corresponding retention times. This table is required to create a calibration curve.
44 of 76

Volume II / Module 3 - Analytical Instrumentation

Next, the standards for creating this curve are injected and stored, ending the creation of this analysis file. When injecting the samples, care should be taken to inject the control standards at regular intervals, thus allowing the system stability verification. Next, the four methods that may be used for calibration purposes will be discussed: Area Percentage. Area Normalization. External Standard. Internal Standard.

4.3.6.1. Area percentages The area percentage is the simplest calculation method, as no standard is required. It is based on the principle that the detector response is proportional to the amount of substance passing through its cell. Can be used for substances having the same response to the detector, that is, if equal amounts are injected, equal areas should be obtained for all peaks. The sample calculation is:

Component A concentration (%) =

Component A area x 100 Areas of all components

4.3.6.2. Area normalization The principle is the same of the percentage area, however the detector response for each substance is corrected using one of the peaks as a calculation reference. For this method, it is necessary to inject a standard for calculating the response factor of each peak, as showed in the formula below: Component A = Component A area x response factor Component A concentration Reference component concentration Reference component area

When the response factor of each component is available, one can analyze the sample using the following equation: Component A = Component A area x Component A response factor concentration (%) (Areas of all components x Response factor of all components)

45 of 76

Volume II / Module 3 - Analytical Instrumentation

4.3.6.3. External standard The external standard is the most common calibration method. It consists in quantifying the sample by injecting standards with known concentrations. The standard area is corresponded to the concentration and the response factor is obtained for each compound, according to the equation below: Component A = Component A area response factor Component A concentration With this response factor generated after injecting the standard, the sample is injected. The area generated in the sample is divided by the response factor generated by the standard, the sample concentration is obtained by the equation: Component A = concentration 4.3.6.4. Internal standard The internal standard is the most precise calibration method, since it corrects volume fluctuations by using a reference standard added in the same amount both to the standard and the sample. The choice of this reference standard should meet the following conditions: Should not be present in the sample; The sample should not have any component with a retention time very close to the standard; It should be pure, non-reactive and preferably of the same functional group of the component to be determined. Component A area Component A response factor

Many times this method cannot be used as a result of such conditions; however, when feasible, this method provides very precise results even when using manual injection. The calculation consists in determining the response factor related to each component by injecting the standard and using the equation below: Component A = Component A area x response factor Component A concentration Internal standard concentration Internal standard area

46 of 76

Volume II / Module 3 - Analytical Instrumentation

When the response factors are calculated, the sample is injected with the internal standard and, with the areas so obtained, the component concentration is calculated using the equation: Component A = Component A Internal standard concentration area x concentration x Internal standard area Component A response factor

There are presently two options for acquiring and processing data: the integrator and the personal computer with a dedicated software. Both allow the user to perform the same tasks. The basic difference between them is the operating system. It is worth emphasizing the importance of transferring obtained data (chromatograms) from a platform to another; for interlaboratory interchange purposes, this cannot be a drawback for results comparison and analysis.

4.4. Minimum installation and operation requirements In order to assure reliable and reproducible results in compliance with the Installation Qualification (IQ) requirements, the conditions of the installation room should be verified. For that purpose, the following conditions are to be mentioned: 4.4.1. Bench requirements The bench should be sized so as to properly accommodate the HPLC equipment and make the user feel comfortable during work. The size of the HPLC and the data system (integrator or computer), usually two linear meters, is enough to accommodate the complete system. The current HPLC equipment is usually modular, their modules remaining over each other forming a tower; therefore, the total height (bench plus tower) should be safe so as to avoid accidents mainly when the operator is handling the mobile phase. The structure should be able to firmly bear about 80 kg without vibrations or swings. The equipment manufacturers provide detailed descriptions of the required bench on the preinstallation manuals. 4.4.2. Electric network The electric network should be duly grounded and stabilized. The use of a magnetic key is recommended in case of constant electric power supply failures.

47 of 76

Volume II / Module 3 - Analytical Instrumentation

The number of outlets should be adequate and in compliance with the system installation standards. One outlet is required for each module, in addition to the outlets for the integrator. If a computer is employed, outlets are required for the CPU, monitor and printer. For a traditional HPLC, having a pump, automatic injector, oven, detector and computer, eight outlets are enough. There should be a spare outlet for use in technical assistance and/or validations. Figure 7 shows the standard outlet used in HPLC equipment.

Standard outlet for liquid chromatographer

Figure 7. Standard outlet for HPLC The HPLC manufacturer should inform the total power consumption as well as the power consumption per module (expressed in Watts or VA or KVA). For the sizing and distribution of the electric system capacity, an electric professional should be contacted. HPLC equipment does not tolerate electric network fluctuations; therefore, it is critical to know the voltage stability and determine that voltage is compatible with the required specifications of each model. 4.4.3. Environmental conditions Temperature and humidity are the basic environmental conditions to be regarded. The temperature of the room where the HPLC equipment will be installed should be controlled with no wide fluctuations during use. The typical temperature required by manufacturers is 25C, fluctuations of +/ 2C are allowed. Humidity should be less than 95%; no condensation is allowed.

4.5. Basic precautions Maintenance by the user is critical for the good operation and durability of a chromatographer. After using the equipment, a cleaning procedure should be followed (that will be dependent on the kind of mobile phase used); this is also true for the chromatographic column; as the main component of the system, the chromatographic column should be given special attention. For cleaning the chromatographic column, carefully follow the manuals instructions.

48 of 76

Volume II / Module 3 - Analytical Instrumentation

Systems working with buffer solution as the mobile phase require special care. After use, it is essential to use water (100%) for the complete removal of the buffer solution and then a mixture, for example, 70% of methanol (or acetonitrile) and 30% of water. This mixture is required as fungi can grow if the system is left in water only. The user should be familiar with the replacement of consumable parts of the system. Any maintenance intervention should be reported in a logbook. The main consumable parts for each module are listed below: Pump Seals, pistons, filters. Manual injector Rotor seal, sample loop, connections. Automatic injector Filters, septa, rotor seal, syringe unit parts (if any), vials for samples and reagents, needle, sample loop. UV-Visible and fluorescence detectors Lamps, lenses, o-rings. Electrochemical detector Electrodes, KCl solution, o-rings.

In addition to the equipment operation, therefore, the users training should include these basic precautions. Due professional qualification is critical for this learning level. Reading the manuals is also an excellent practice. In addition to provide detailed information on the equipment operation, the manual clarifies several doubts of the user. It is important to observe the safety rules when handling the equipment. Wearing apron, gloves and goggles avoids severe injuries such as blindness or sample contamination, for example, plasma. The system decontamination is always mandatory in order to protect the health of the person in charge of maintenance.

49 of 76

Volume II / Module 3 - Analytical Instrumentation

5. CHROMATOGRAPHY SYSTEMS CONNECTED TO MASS DETECTORS


5.1. Introduction Chromatography is an essential separation technique in life sciences and related chemistry fields. Traditional detectors such as ultraviolet-visible, electrochemical, refractive index (HPLC), flame ionization, thermal conductivity, etc (GC) are widely used to quantify compounds as bidimensional data are produced (response x time). The mass detectors are characterized by generating tridimensional data (response x time x ionic specimen), that is, mass spectrum that can provide very important information on the sample molecular weight, its molecular structure, identity, quantity and purity. Data from the mass spectra add specificity to both quantitative and qualitative analyses. For most of the compounds, the mass detectors are more sensitive and much more specific than traditional detectors. They can analyze several compounds and identify components in non-separated chromatograms, thus reducing the need of a perfect chromatography. The mass spectral data may supplement data from other detectors. Although two compounds can have similar UV or Mass spectra, such as in LC-MS, this phenomenon is hardly simultaneous; therefore, both types of data together can be used to identify, confirm and quantify compounds with highly correct results. Some mass spectrometers have the characteristic of performing multiple mass spectrometry steps in a single sample. They can generate a mass spectrum, select a specific ion and then generate a new spectrum. Some of them are able to repeat this cycle several times until the structure is determined (MS/MS or MSn). 5.2. Technique principle The mass spectrometer basically operates through the ionization and fragmentation of molecules. Afterwards, the resulting ions are identified according to their mass/charge ratio. The three key components of the process include: ions source, analyzer and detector. The purpose of the ions source is to generate ions. There are several kinds of ions source in the chromatography connected to mass detectors. Each type is specific for a given class of compounds. There are also several kinas of analyzers for the separation of ions and detectors for the generation of measurable signals. Each one has advantages and drawbacks depending of the type of information searched. Details concerning the most common sources, analyzers and detectors will be discussed next:

50 of 76

Volume II / Module 3 - Analytical Instrumentation

5.3. Basic description of the system

SAMPLING SYSTEM

IONS SOURCE

MASSES FILTER

DETECTOR

IS

MF

EM

DATA SYSTEM

DS

5.3.1. Ionization source The molecules ionization and fragmentation will take place in the source. Several ionization techniques are available, including: Atom bombing - FAB, Laser Desorption - LD, Thermospray - TS, Particles beam, etc. However, the most commonly used sources are: 5.3.1.1. Electrons impact (EI) The Electrons Impact (EI) source uses a filament in charge of emitting electrons with a defined energy of 70 eV. Once the electrons beam energy is much higher than the first ionization potential in most of the compounds of the sample, this energy is ionized and then fragmented. This kind of ionization is related to the GC. 5.3.1.2. Chemical ionization (CI) The Chemical Ionization (CI) source uses liquid or gas agents to react with molecules. Ionization usually takes place by means of the transference of a proton to the molecule, thus forming specimens called molecular pseudo-ions. As this ionization is much smoother than the electrons impact, the
51 of 76

Volume II / Module 3 - Analytical Instrumentation

spectrum produced contains a few fragments and almost exclusively the molecular pseudo-ion; therefore, it is employed in the determination of the molecular weight and/or quantitative analyses. This kind of ionization is related to GC. 5.3.1.3. Electrospray (ES) The electrospray ionization has a great impact on the use of mass spectrometry applied to biological researches in the last years. It was the first method to expand the instruments useful mass range to above 50,000 Da. However introduced in its present model in 1984, the technique returns to the investigations of electrically assisted liquid dispersion in the beginning of this century. In fact, the main discovery was almost accidental in 1968 when Malcolm Dole and cooperators were able to bring macromolecules to the gas phase at atmospheric pressure. It was possible by spraying a sample solution to a small tube with a strong electric field in the presence of a warm nitrogen flow to help in the dissolvation and then measuring the ions formed. Later, innovative experiences in this field led to the introduction of an ES ionization source. Since then, a wide range of biomolecules was investigated by ES. The sample is usually dissolved in a mixture of water and organic solvent, typically methanol, isopropanol and acetonitrile: it can be directly infused or injected in continuous flow, that is, contained in the eluant of a HPLC column and CE capillary column. The ES source is simples, forming a spray occurring in a high voltage field as showed in Figure 5. In a proposed mechanism, it is believed that the ion formation is the result of an ionic evaporation process, first proposed in 1976. A droplets spray is generated by the electrostatic dispersion of the liquid applied by the capillary end. Favored by a heated gas (usually nitrogen), the droplets are disaggregated, lose solvent molecules in the process and occasionally produce individual ions. In another proposed mechanism, the droplets dissolvation lead to an increasing charge density on the droplet surface that will cause a coulomb explosion eventually producing individual ions.

Pump

Drop

Atmospheric Pressure

Droplet

Pump Clusters and ionic specimens Ions

Regardless the proposed mechanism, the ions are formed at atmospheric pressure and enter into a bore located in the vortex of a cone acting as the first barrier to the vacuum phase. A skimmer collects the ions and guides them to the mass spectrometer.

52 of 76

Volume II / Module 3 - Analytical Instrumentation

The spray formation is the most import part of the ES technique. It is usually advised to filter all solvents; high electrolyte concentrations should be avoided as they can cause ionization suppressions and unstable operating conditions. High flows compatible to those used in HPLC, can be now used through a heated nebulizing gas to help in producing the spray. For macromolecules, each ion that usually enters the mass spectrometer has a high charge number. As the mass spectrometers measure the mass/charge ratio instead of the mass, it is possible that high molecular mass molecules have enough charge to fall within the m/z range of a linear quadrupole, typically m/z 20-4000. As showed in Figure 6A, high molecular mass ions often have a wide distribution of charge status. The figure shows the mass spectrum of horse myoglobin (molecular mass: 16951.5) at low resolution having observed charges of +12 to +21. This ionic distribution allows the calculation of the original analyte molecular mass by means of neighbor ions with m/z values m1 and m2 , with n1 and n2 charges, respectively. If m1 < m2, and n2 = n1-1, then, M = n1(m1-mA) = n2 (m2-mA) (1) n2 = (m1-mA)/(m2-m1) (2) wherein M represents the molecular mass of the uncharged molecule and mA is the mass of the charged adduct A (for example H+, Na+, NH4+). The equations can be then solved to yield n2 and M. The measure of the charges distribution in macromoleules is not always easy or reproducible once changes, such as relatively small changes of the analysis conditions can occur, for example: pH; addition of solvents or salts; protein partial denaturation; breakdown of disulfide bonds, etc.

The electrospray source is used in connection to all single mass analyzers.


53 of 76

Volume II / Module 3 - Analytical Instrumentation

5.3.2. Mass analyzers After the admittance of the molecules into the ions source and subsequent ionization, it is necessary to determine the corresponding masses of ions formed so as to obtain the mass spectrum. The function of the mass analyzer is to separate ions according to their mass/charge ratios and transmit them to the detector. There are several kinds of mass analyzers. The most common and widely used are the Quadrupole and Ion Traps analyzers. The quadrupole are scanning analyzers, that is, after the admittance of a mixture of ions with different mass/charge ratios (m/z) and different abundances, electric fields are applied; at a given time, only ions with a specific mass can leave intact. By varying the electric field applied, one can select and record different ions. The Ion Trap analyzers, however, are not regarded as pure scanning devices, as the ions are stored before the scanning itself. 5.3.2.1. Quadrupole mass analyzers The instrument is based on four parallel bars in a quadrangular area where the ions beam is focused on the central axis of these bars, a fixed electric potential (DC) and a radio frequency potential (RF) are applied to these diagonal and opposite bars. For a given RF and DC combination, ions of a specific mass range m/z have their path changed from the central axis. The mass spectrum is obtained from the DC voltage and RF components in a synchronized fashion, that is, keeping the RF/DC ratio constant. The o potential applied to the opposite bar pairs is determined as follows: o = U + V cos t wherein U is a DC voltage and V cos t the time-dependent voltage in which V the RF amplitude and the RF frequency.
Detector output

Ions beam

Ions source

Quadrupole Architecture

Figure 2. Schematic representation of a linear quadrupole analyzer


54 of 76

Volume II / Module 3 - Analytical Instrumentation

The quadrupole operation (as well as the ion trap to be discussed later) can be qualitatively discussed through the stability diagram corresponding the DC potential amplitude, the RF potential amplitude with the path of a stable ion (that is, an ion that can remain intact after passing the quadrupole). This will be represented in the equations and in graph 1. The motion equation of a charged particle can be expressed as a Mathieu equation in which the au and qu parameters can be defined. au = ax = -ay = 4zU / m 2ro2 qu = qx = -qy = 2zV / m 2ro2 wherein m/z is the ion mass/charge ration, and ro is half of the distance between the two opposite bars. There is no parameter for z, since the RF field acts on the x/y plane (z is the main axis of the linear quadrupole).

Graph 1. Stability diagram in the analyzer In this case, the ion m2 is the only ion that remains stable (observe it is within the stability region of the graph), while m3 and m1 cannot reach the detector. The R=100 2e R=10 straight lines represent two distinct combinations of the DC/RF ratio; by changing these ratios, it is possible to change the mass filter resolution, that is, the capacity to differentiate or filter masses very close to each other. The a and q parameters are respectively proportional to the DC and RF values. 5.3.2.2. Ion trap quadrupole mass analyzers This mass analyzer has the same mathematic operation principles of the conventional quadrupole, that is, ions stability within a specific path.

55 of 76

Volume II / Module 3 - Analytical Instrumentation

However, the major difference is that in the ion trap the ions do not follow a single path towards the detector, but are trapped in orbits within the trap structure, thus giving raise to the name ion trap. While in the quadrupole the ions are formed in the ions source and then expelled towards the mass analyzer, in the ion trap the ions are formed ions and the mass analyzed in the same space region. This space region where the trapped ions are found corresponds to approximately the volume of 1 cm-side cube. The figure below shows a schematic representation of the ion trap analyzer.

Figure 3. Schematic representation of the ion trap analyzer As the quadrupole analyzers, the ion traps analyzers also use electric fields; these electric fields are intended to keep the ions confined and separate the masses (mass analysis). In this kind of mass analyzer, the electric field used is purely RF (radio frequency consisting in a sine wave with a frequency of about 1 Mhz) applied directly on the central annular electrode. Thus, depending on the RF amplitude applied, the ions can remain stable inside the trap. By increasing this amplitude, the ions with greater masses are ejected from the confinement region and then reach the detector. Another peculiarity of the ion trap analyzers is the need to control the number of ions within the structure, aiming to avoid reactions of these ions with molecules still present. These interactions can lead to a slight change in the final spectrum of some compounds, thus making interpretation/ identification difficult. This problem is eliminated in the quadrupole analyzers, as the ions are almost immediately thrown from the ions source after the formation and do not have the chance to interact with the molecules present. On the other hand, the final sensitivity in the full scanning mode is greater in trap analyzers since all ions formed are necessarily detected once they are confined.
56 of 76

Volume II / Module 3 - Analytical Instrumentation

This is not the situation for the quadrupole analyzers. That is, while the quadrupole is canning the mass range aiming to produce a spectrum, the unstable ions get lost and, therefore, do not generate a detectable signal. It is possible, however, to significantly increase the quadrupole sensitivity by defining a fixed mass (or a few masses) of interest; this technique is called SIM (Single Ion Monitoring). 5.3.2.3. Tandem mass spectrometry. The tandem mass spectrometry, MS/MS or MSn wherein n=2, 3..., uses two or more mass analysis steps, one to preselect an ion and the others to analyze the induced fragments, for example, by collision (CID) with an inert gas, such as argon or helium. This can be a tandem-in-space or tandemin-time analysis. Tandem-in-space means several mass analyzers in series. Various combinations are possible, the most common include: triple quadrupole (Q1qQ2), four sectors and hybrid instruments. Q represents a quadrupole mass filter and q a RF quadrupole only (collision chamber). In the case of the triple quadrupole, an ion of interest generated in the ionization source is selected with the first quadrupole Q1, dissociated in the collision chamber q with energies up to 300 eV; the fragmentation products are analyzed with the second quadrupole Q2.
Collision gas molecule

Ions source

Detector Ions-product formation

Total mass spectrum

Mass spectrum by C.I.D.

Thus, it is possible to obtain information on the sequence of a peptide by selecting the ion corresponding to the protoned peptide (called precursor ion) and analyzing the fragments of its structure using Q2. This process is called ions-product scanning. Several other types of scanning or analytical experiments can be performed. For example, the search of all precursors of a given fragment is called ions-precursors scanning. This can be reached by keeping Q2 steady at themass/ charge ratio of the ion concerned and scanning all ions present at Q1, while the collision dissociations at q take place. In another scanning mode, ions that lose a specific fragment can be identified by scanning Q1 and Q2 simultaneously, keeping the mass difference between the quadrupole analyzers equal to the mass of the neutral lost fragment.
57 of 76

Volume II / Module 3 - Analytical Instrumentation

Tandem-in-time can be obtained using ion trap devices and ICR mass spectrometers (also called FTMS). In fact, these devices are not limited to MS/MS experiments, but can reach multiple stages (MS/MS/MS/MS or MSn). In such devices, the ions are selected by applying specific voltage pulses and dissociations normally occur by collisions with other gases. 5.3.3. Detectors After selection by the analyzer, the ions are guided towards the detector where they will be converted into a measurable signal. Detectors can be divided in three groups. Photosensitive plates and Faraday cages are included in the first group and directly correlate the measured signal to the amount of the analyzed ion. The second group includes the electrons multipliers, photomultipliers and microchannel plates that amplify the intensity of the received signal. These are the most commonly used detectors and will be discussed herein. The third group is used in ICR devices (FTMS) and consist in a radio frequency detector applied to the trapped ions. 5.3.3.1. Electrons multipliers When reaching the conversion dinode where a negative potential is applied, positive or negative ions emit several secondary electrons. These electrons are accelerated towards an electrons multiplier and hit the walls with sufficient energy to remove some electrons. These electrons will hit the other side of the wall, thus releasing more electrons. This cascade effect continues until a measurable current is finally created at the end of the electrons multiplier.

Electron multiplier Ions

Secondary electrons

Electrons outlet

58 of 76

Volume II / Module 3 - Analytical Instrumentation

5.3.3.2. Microchannel plates The microchannel plates consist in a plate containing parallel cylindrical microchannels. The inlet side of these microchannels is kept with a negative potential of approximately 1kV when compared to the outlet side. The electrons multiplication, started with the collision of an ion in these channels, occurs through a semiconductive substance that coats each channel and generates secondary electrons. Curved channels prevent the acceleration of positive ions towards the inlet. The cascade effect inside the channels can multiply the number of electrons in the order of 105 and the use of several coupled plates allow an amplification that can reach 108. At the outlet of each channel, a metal anode collects the electrons current and the signal is transmitted to the processor. Another characteristic of these microchannel detectors is an extremely low signals multiplication time, making them inadequate for detection in devices such as time-of-flight analyzer.

Electron multiplier Secondary electrons

Ion

Electrons outlet

Inlet of an electron multiplier

Ions with different mass/charge rations separated in the space

59 of 76

Volume II / Module 3 - Analytical Instrumentation

5.3.3.3. Photomultiplier This kind of detector comprises two conversion dinodes; a phosphorescent screen and a photomultiplier. This detector, as the electrons multiplier and the microchannel plates, allows the detection of positive and negative ions. Upon detection, the ions are accelerated towards the dinode having a reverse inverse polarity of that of the ion; the electrons are then released and accelerated towards the phosphorescent screen where they are converted into photons. The photons are then detected by the photomultiplier. The phosphorescent screen surface is coated with a fine conductive aluminum layer so as to avoid the formation of charges that could refrain new electrons from reaching it. The amplification reaches values from 104 to 105.

Ions

Secondary ions

Emitted photons

Photomultiplier

5.3.4. Data acquisition and processing Specific computer programs integrally perform data acquisition and processing. These programs are in charge of several tasks, ranging from the control of the monitoring time of an ion to the construction of calibration curves where the areas (or heights) of unknown sample peaks are interpolated, thus producing the desired quantitative datum. Each manufacturer has a data acquisition and treatment program with different resources and limitations. Therefore, the comprehension of these programs and the future preparation of standard operating procedures (SOP) are crucial for conducting any study. Details on the compounds quantification process are provided in item 4.6 of the HPLC technique.

60 of 76

Volume II / Module 3 - Analytical Instrumentation

5.4. Minimum installation and operation requirements The purpose of the minimum installation requirements is to assure that the installation room is duly evaluated and prepared with the consumables and supply items required for a perfect installation. The manufacturers will provide a checklist to be used before the installation date. This checklist will be specific for the instrument purchased; however, what should be generally noted is described next. 5.4.1. Bench requirements The bench should be sized so as to properly accommodate the LC/MS equipment and make the user feel comfortable during work. The bench size should be compatible with the LC/MS size plus the data system (computer and printer). Usually, three linear meters, is enough to accommodate the complete system. The structure should be able to safely bear the equipment weight; therefore, a margin of 20% above the equipment weight should be considered. The structure has to be firm, with no vibration or swings. Some free space behind the bench should be also allowed (do not place the bench touching the wall) so as to accommodate the items required for the equipment operation, for example, vacuum pumps and water cooling and circulation systems. The equipment manufacturers provide detailed descriptions of the required bench on the preinstallation manuals. 5.4.2. Electric network The electric network should be duly grounded and stabilized. The use of a magnetic key is recommended in case of constant electric power supply failures. The number of outlets should be adequate and in compliance with the system installation standards. In addition to the LC/MS module, outlets for the system HPLC and the data system, including the printer, are to be considered. The standard outlet for LC/MS will depend on the required operation voltage. Due to the high consumption, these systems generally require 220 VAC feeding with capacity for 1.5 to 2 KVA (for the MS module only). It is worth noticing that the equipment requires 220 VAC in single or split phase. The LC/MS manufacturer should inform the total power consumption as well as the power consumption per module (expressed in Watts or VA or KVA). For the sizing and distribution of the electric system capacity, an electric professional should be contacted.

61 of 76

Volume II / Module 3 - Analytical Instrumentation

The equipment does not tolerate electric network fluctuations; therefore, it is critical to know the voltage stability and determine that voltage is compatible with the required specifications of each model. 5.4.3. Environmental conditions Environmental conditions for LC/MS systems do not include temperature and humidity only. This kind of equipment also requires gas feeding and some of them require the circulation of cooled water, in addition to exhaustion. The temperature of the room where the LC/MS will be installed should be controlled with no major fluctuations during use. The typical temperature required by the manufacturers is 25C, with fluctuations of +/- 2C. A LC/MS system dissipates heat from 6800 to 8000 BTU; this heat should be taken into account when sizing the air conditioner. Humidity should be below 95%; no condensation is allowed. Gas (nitrogen) can be fed from cylinders, liquid nitrogen compartments or nitrogen generators (this latter being most indicated). Typical consumption is in the order of 15 L/min and purity should be 99.5% for cylinders and 98.0% for other nitrogen sources. Working pressure should be 80 to 100 psi. The room should be also provided with an exhaustion system with a minimum capacity of 15 L/ min. This system should be connected to the vacuum pump and the ionization source outlet. 5.5. Basic precautions Maintenance by the user is critical for the good operation and durability of any equipment. In addition to the basic precautions of the HPLC system described in chapter 4 above, some essential items should be regarded for a LC/MS system: 5.5.1. Vacuum system High vacuum systems usually have three pumps, one auxiliary (mechanical) and two turbomolecular pumps. The mechanical pump requires fluid replacement at least every six months, whether it is used or not. Some of them also have filters that have to be replaced and/or verified at intervals set by the manufacturer (typically six months). The turbomolecular pumps (high vacuum) are more specific and depend on the model and recommendations of each manufacturer. Some of them have permanent lubrication, but others need lubrication at regular intervals. Some others, yet, require the bearings to be replaced at predetermined intervals. Therefore, it is critical that the maintenance instructions are obtained from
62 of 76

Volume II / Module 3 - Analytical Instrumentation

the supplier and that the process is documented and followed, since a failure of such systems may definitely impair the equipment as a whole. The vacuum (pre and high vacuum) should be read daily and their values weekly recorded in a logbook. 5.5.2. Ions source The ions source is a part of the equipment most subject to dirt and contaminations since it receives the mobile phase with the sample at atmospheric pressure and converts them into ions for the vacuum system. Maintenance and cleaning should be made based on the usage background of each equipment and will depend on the type, concentration and number of samples injected. Once established, the cleaning interval should be followed and documented. The user should be familiar with the replacement of consumable parts of the system. Any maintenance intervention should be reported in a logbook. 5.5.3. Nitrogen system The nitrogen system should be checked at predetermined intervals. A nitrogen generator is normally chosen due to the cost/benefit ratio. Such devices are provided with filters, compressors, etc. that also require periodic maintenance. 5.5.4. Basic precautions training In addition to the equipment operation, therefore, the users training should include these basic precautions. Due professional qualification is critical for this learning level. It is important to observe the safety rules when handling the equipment. Wearing apron, gloves and goggles avoids severe injuries such as blindness or sample contamination, for example, plasma. The system decontamination is always mandatory in order to protect the health of the person in charge of maintenance. 5.5.5. Autotune files The LC/MS systems are provided with autotune programs to be performed at regular intervals. These programs generate reports that should be filed for the good operation background and to help with repair interventions. Basic equipment precautions, such as performance checking, range from equipment to equipment. Thus, the best procedures to perform such tasks are those suggested by the manufacturer. These procedures should be read and a standard operating procedure (SOP) created to define the frequency and kind of task to be performed. Any non-compliance with this SOP should be documented and justified.
63 of 76

Volume II / Module 3 - Analytical Instrumentation

6. ANALYTICAL INSTRUMENTS PERFORMANCE CHECKING

6.1. Introduction By the middle of 1990s, instruments performance checking started to be have a major importance and the first focus was on the HPLC systems as a result of its wide use in pharmaceutical industries; however, this concern covers all kinds of instruments used in a laboratory. As in data management systems, these services are usually offered by analytical instrument manufacturers. There are different terminologies among them sometimes referring to the same services rendered. For example, the Performance Checking of an analytical instrument can be also called Qualification, Operation Qualification (OQ) or Recertification. It is important to note that, regardless of the terminology used by manufacturers, it should correspond to the analytical instrumentation system performance checking following a documented procedure so as to meet the requirements of a quality system and in compliance with the GLP (Good Laboratory Practices) at appropriate intervals defined for each laboratory according to the usage and/or work performed. The instrumentation companies recommend this checking at regular periods every 6 months or 1 year and whenever the instrument is subject to service or repair that may directly interfere with its performance. The Validation terminology exists for a long time and has some variations according to the country and company. The best definition for validation is: the set of documented activities and/or information evidencing a high safety degree of the process/system employed, meeting the predetermined specifications with the proper quality attributes. Therefore, validation is not a performance checking, qualification, operation qualification (OQ) or recertification process only, but a process that should include the following phases: DQ - Design Qualification; IQ Installation Qualification; OQ Operation Qualification; PQ Performance Qualification.

6.1.1. DQ Design Qualification Within the instrument design qualification, access should be allowed to the instrument development details, such as:

64 of 76

Volume II / Module 3 - Analytical Instrumentation

Use of strict specification methods and design during the system development. Complete documentation of the quality assurance and quality control procedures. Use of experienced and qualified personnel. Comprehensive instrument test plan. Control data, error reports and corrective procedures applied during the system development and production. System development background.

Such information should be provided by the instrument manufacturers, however, aiming to check the specifications and proposed use adequacy. 6.1.2. IQ Installation Qualification The installation qualification means: checking all documentation and information for installing the instrument and software according to the system specifications described in the installation prerequisites and local regulations. The IQ defines that the instrument is received as specified, properly installed and under operation conditions. This activity should include: Checking the items received according to the appropriate documentation. Filing of all description details and identification of the system components, including the instrument connection diagrams. Copies of the users test certificate and declaration of compliance.

The Installation Qualification (IQ) has been discussed in every analytical instrumentation technique of this chapter as Minimum Installation Requirements. 6.1.3. OQ Operation Qualification The Operation Qualification should be documented, checking that the system meets the operating specifications. A follow-up system should be used to verify that: Operational training is given after the installation and before starting the routine operation of a certain instrument. User commitment to follow all steps required. Performance tests are conducted according to the instrument manufacturer in order to assure the instrument is under appropriate operation conditions.

6.1.4. PQ Performance Qualification The PQ consists in conducting activities to certify that all the instrument specifications are according to the usage requirements. This qualification should be strict and involve the tests needed to demonstrate the instrument functionality. It should include details such as:

65 of 76

Volume II / Module 3 - Analytical Instrumentation

Specific tests of individual components or the system as a whole. List of tests. Frequency of tests. Expected results. Acceptance criterion. How tests are documented. Operators qualification required. Actions to be taken in case of test failure.

For some analytical instrumentation techniques, there are routines defined in internationally accredited literature, such as the EP/BP (European Pharmacopoeia), USP (United States Pharmacopoeia) and others, setting parameters to be determined to validate an analytical system to be used by the user. They are many times performed by the analytical instruments manufacturers. We cover procedures normally used for checking the performance of the main analytical instrumentation techniques and, when applicable, the main Performance Qualification (PQ) routines described in the international literature.

6.2. Ultra violet visible (UV-VIS) spectrophotometry When checking the performance (Operation Qualification) of an analytical instrument, the first step to be taken is a Preventive Maintenance aiming to check the instrument operation status in advance, prevent and repair any possible failure. 6.2.1. Preventive maintenance Preventive maintenance of a UV-Vis spectrophotometer is crucial for the instrument life and should be conducted at 6-month intervals. This maintenance has to include specific activities carried out by an expert. A preventive maintenance routine should include: Checking the instrument safety items. Inspection of the sample compartments. Checking the lids and protection sensors. Checking the optical system. Checking and cleaning external optics. Conducting performance tests: wavelength accuracy and precision, baseline correction and photometric noise. Checking the lamps conditions and alignment. Electronic checking. Checking the instrument internal diagnostics. Feeding source and electric network checking. Checking cables and connectors. Cleaning electronic circuits and wearing, corrosion and failures.
66 of 76

Volume II / Module 3 - Analytical Instrumentation

Checking mechanical parts. Testing motors inside the instrument. Checking operation and communication with accessories.

6.2.2. Operation Qualification (OQ) The OQ of an Ultraviolet-Visible spectrophotometer consisting in checking the following: Wavelength accuracy; Wavelength precision; Baseline deviation; Photometric noise.

6.2.3. Performance Qualification (PQ) Some of the PQ routines that can be used with UV-Vis spectrophotometers and defined at the EP and USP are described below. 6.2.3.1. European pharmacopoeia The tests below are listed in the European Pharmacopoeia and British Pharmacopoeia (EP/BP) for the validation process of instruments, being divided into: Wavelengths accuracy. Tests aiming to check the wavelength positioning accuracy in order to assure low deviation of the optical system positioning. EP/BP recommends two routines for checking this item: Deuterium, xenon and mercury-line emission method. Holmium perchlorate method. Monochromator resolution. Test to verify whether the instrument monochromator separates the spectrum light and sends a small amount of this light through the sample compartments. This test is conducted using the Toluene / Hexane method. Stray Light. Stray Light is defined as the amount of light reaching the detector when no wavelength has been selected. Stray light causes deviations in the Beer-Lambert Law, thus reducing the system linearity. This test is performed by measuring the transmittance passing through a solution; next, the same reading is made with the source turned off. The method used is KCl at 200 nm. Photometric accuracy. The photometric accuracy is determined by measuring known absorbance values and calculating the difference between the expected value and the actual value measured by the instrument. The method used is Potassium Dichromate.

6.2.3.2. United States Pharmacopoeia The tests below are listed in the United States Pharmacopoeia (USP) for the validation process of instruments, being divided into:

67 of 76

Volume II / Module 3 - Analytical Instrumentation

Tests aiming to check the wavelength positioning accuracy in order to assure low deviation of the optical system positioning. USP recommends two routines for checking this item: Deuterium, xenon and mercury-line emission method. Holmium perchlorate method. Photometric accuracy. The photometric accuracy is determined by measuring known absorbance values and calculating the difference between the expected value and the actual value measured by the instrument. The methods used include: Potassium Dichromate method. NIST filter method.

6.3. Gas chromatography GC 6.3.1. GC preventive maintenance Before the Operation Qualification (OQ) step of a GC, it is necessary to follow the Preventive Maintenance procedures suggested by the manufacturer. In a Gas Chromatography System, the Preventive Maintenance routine include: Pneumatic System cleaning and checking leakages; Electronic circuits cleaning and checking the voltages used; Venting Systems cleaning and checking functionality; Injector(s) and detector(s) cleaning, checking leakages and voltages used; Detectors response checking test.

6.3.2. Operation qualification The Operation Qualification of a gas chromatography system consisting in verifying the following: Control of carrier gas and auxiliary gas flows; Control of the Temperature of Injector(s), Detector(s) and Column Oven; Detector(s) signal repeatability; Automatic sampler repeatability, when applicable; Calculations made by the Data System.

6.3.2.1. Flows control The purpose of this checking is the repeatability of the analyzed compounds retention times, repeatability of the volume of sample introduced in the chromatographic column, in case of split injectors and repeatability of the detector response. Usually, chromatographers can be provided with electronic flow controllers (EFC) or manual controllers.

68 of 76

Volume II / Module 3 - Analytical Instrumentation

The materials required for this verification in both cases include: Calibrated Digital Flow Meter; Reference or test-specific chromatographic column.

In addition to the carrier gas flow passing through the column, the flow control checking should also include the septum purging flows, the division rate in case of a split/splitless injector, the make-up flow and auxiliary flows of the detector, hydrogen and air in case of a FID detector. These tests are usually conducted under certain conditions previously defined in documented procedures. 6.3.2.2. Temperatures control The columns oven is the most important part and interferes with a chromatographic analysis; small fluctuations of the oven temperature may result in significant variations of the retention time and, as a result, affect the analytical repeatability. The materials required for this verification include: Calibrated Temperature Meters: Reference or test-specific chromatographic column; Reference standard sample.

The Injector(s) and Detector(s) operating temperatures are checked using temperature meters; in addition to the temperature checking, the columns oven are subject to retention time repeatability tests under previously set conditions and reference sample. 6.3.2.3. Detector(s) signal precision This checking aims to evaluate the detector repeatability; however, in most of the procedures followed, this can be used as the repeatability check for the system as a whole. The materials required for this verification include: Reference or test-specific chromatographic column; Reference standard sample.

This checking usually takes place under a certain analytical condition by means of consecutive injections of a reference standard sample. The purpose is to calculate the relative standard deviation as percentage of this standard sample areas. 6.3.2.4. Automatic sampler precision Automatic Samplers are checked under the same conditions of the detectors signal repeatability checking; thus, the results obtained as RSD% of the area values are used twice.
69 of 76

Volume II / Module 3 - Analytical Instrumentation

6.3.2.5. Calculations by the data system Checkings usually made in Data Systems refer to the calculations used in the workstation and include: Retention Times Calculation; Areas Count Calculation; Time Events Calculation; Calculation of the Calibration Curves Coefficients for External and Internal Standardization; Corrected Area Normalizations Calculation.

These checkings are usually made by protected files supplied by the data system manufacturer; when used, these files allow to check whether calculations are in accordance with the results defined in the reference files. Once the Operation Qualification of a Gas Chromatography system is completed, all such records should be stored at an easily accessible location. Manufacturers usually provide Manuals with all the compliance documentation of the chromatography system, including information for the Design and System Development Qualification (DQ), Installation Qualification (IQ) that should contain information on which system was installed and the Operation Qualification (OQ), including all procedures to be followed and storage of checking records. Therefore, all interventions should be evidenced in this Manual, used as a system background. 6.4. Liquid chromatography - HPLC 6.4.1. HPLC Preventive maintenance Before starting the HPLC system Operation Qualification (OQ) process, it is necessary to follow the preventive maintenance procedures suggested by the manufacturer. In a HPLC System, the preventive maintenance routine include: Electronics: cleaning of electronic parts and checking feeding and internal voltages, check the detectors lamps functionality and evaluate their life. Hydraulics: Cleaning of the pipings and pneumatic system components, check and eliminate any possible microleakage and occasional replacement of the seals and filters of the mobile phase and pump, check the pressure transducer functionality and the pulses muffling system when applicable, cleaning and checking the sample introduction system functionality. Mechanics: Cleaning and lubrication of mechanical parts of the automatic sampler pumping system.

70 of 76

Volume II / Module 3 - Analytical Instrumentation

6.4.2. Operation qualification The Operation Qualification process showed below is schematically represented in Figure 1: A Detector Wavelength accuracy; Wavelength precision; Noise Level and Baseline Deviation; Detector Linearity (optional); Detector Signal Precision (optional). B Pump Qualification Flow accuracy; Ramp and pressure drop test; Gradient Mixture Accuracy; Gradient Mixture Linearity (optional). C Columns Oven Qualification Temperature Accuracy D Automatic Injector Qualification Injection Volume Precision; Injection Volume Accuracy (optional); Injection Linearity (optional).

E Data System Qualification Calculation of the retention times, areas count and calibration curves coefficients. Figure 1. Schematic view of the Performance checking procedure for a HPLC System.

71 of 76

Volume II / Module 3 - Analytical Instrumentation

As a HPLC System is usually made of several modules, each module should be subject to the Operation Qualification. Thus, the Validation process will be completed when the Performance Qualification (PQ) is conducted. PQ should be always conducted using the complete HPLC System in activities confirming that all the instrument specifications meet the usage requirements. 6.4.2.1. Detector qualification The most important parameters to be verified in an Absorbance detector are: Wavelength accuracy; The purpose of this test is to check the accuracy when selecting a wavelength so as to assure lowpositioning deviation of the optical system diffraction grid; Wavelength precision; The deuterium or xenon-line emission method can be used (both regarded as a natural standard) and using a certified chromophoric substance standard; Noise Level and Baseline Deviation; This test is intended to assure the detector sensitivity specification, since these factors are directly related to this measurement. The baseline monitoring and measurement method can be used at established time intervals and standard values verification.

6.4.2.2. Pump qualification 6.4.2.2.1. Mobile phase flow accuracy This test is intended to assure that the programmed flow is the actual pump operation flow. It can be determined by measuring (volumetrically) the pumped flow at a certain period of time. 6.4.2.2.2. Ramp and pressure drop test The ramp test checks the inlet and outlet valves, connection leakages and the pump capacity to operate at high pressure. 6.4.2.2.3. Gradient mixture accuracy The purpose of this test is to check the accuracy of the static and dynamic composition of mobile phase mixtures during an analysis in gradient systems. The test of pumping two mobile phases at different ratios through tempo can be used and check the actual composition established.

72 of 76

Volume II / Module 3 - Analytical Instrumentation

6.4.2.3. Columns oven qualification 6.4.2.3.1. Oven temperature accuracy With the purpose of checking the oven temperature accuracy, one can use the method of programming a temperature of the oven and measure it using a calibrated electronic thermometer having a flexible tip that can be adapted to the oven internal length. 6.4.2.4. Automatic sampler qualification 6.4.2.4.1. Automatic sampler precision In order to assure the specified precision of the automatic sampler, one can use the test of injecting successive samples of a standard solution, thus obtaining the RSD% between the peaks areas; a value below 1% is recommended. 6.4.2.5. Data system The Data System Qualification for HPLC should be conducted according to the same GC procedures. Once the Operation Qualification of a Gas Chromatography system is completed, all such records should stores in an easily accessible location.

6.5. Chromatography systems connected to mass detectors 6.5.1. Mass detector systems preventive maintenance As the previous systems, preventive maintenance of mass detector systems should be made before the Operation Qualification (OQ) process. Each manufacturer provides a checklist of the equipment concerned. This checklist should be requested and completed in every preventive maintenance intervention. The items to be checked in a preventive maintenance are basically divided into three modules: Vacuum system; Mass detector; Chromatographer.

6.5.1.1. Vacuum systems The vacuum system is critical for the good operation of any mass spectrometry system and, therefore, regular maintenances are required. These maintenances should be recorded in the equipment operation background.
73 of 76

Volume II / Module 3 - Analytical Instrumentation

The items checked in a vacuum system include: Checking and replacing the pumps lubricant fluids (auxiliary and high vacuum pumps); Checking hoses and vacuum connections; Measuring and recording high-vacuum and pre-vacuum readings on a regular basis.

6.5.1.2. Mass detector The mass detectors are provided with an autotune or auto-calibration program to be used for evaluating the equipment operation conditions. This program usually produces a reported (to be filed); this report will be used to follow-up the performance as well as prematurely identify failures, for example, the need of cleaning the ions source. It is recommended that these reports are filed at least once a week, being kept for the period of the performance checking. The items observed in this report are: mass accuracy, peaks resolution and response as a function of the mass. The ions source should be cleaned whenever required; however, a minimum cleaning frequency should be adopted and will depend on how the equipment is being used (sample type, number of samples, etc). 6.5.1.3. Chromatographer The chromatographer preventive maintenance should be simultaneous with the mass detector. This procedure was described in the corresponding chapters (GC or HPLC). 6.5.2. Operation qualification The Operation Qualification is critical to assure the equipment is operating according to the manufacturers specifications and, therefore, at regular intervals. Manufacturers recommend the operation qualification at least once a year or after a major repair intervention. The expert in charge of that intervention will have means to recommend whether an Operation Qualification after the repair is required. This evaluation should be documented. The items evaluated in an Operation Qualification include: Injector precision; Injector linearity; Contamination test from sample to sample (carry-over); Detector linearity (optional); Mass accuracy; Sensitivity.

74 of 76

Volume II / Module 3 - Analytical Instrumentation

6.5.2.1. Injector precision This test is conducted by injecting several aliquots of a known standard and by verifying the area reproducibility and/or height. Six to ten measurements are made and the relative standard deviation calculated (it should be within the predetermined standards). 6.5.2.2. Injector linearity During this test, several aliquots of a standard with a known concentration will be injected; the injected volume is also ranged. Then, the linearity as a function of the injected volume will be analyzed using a calibration curve. A correlation factor is used to evaluate the test. The results are then compared to the predetermined acceptance values. 6.5.2.3. Carry-over The carry-over test or contamination from sample to sample is conducted by injecting a high concentration standard and then solvent only. The result is optimum when no residue of the previously injected standard is detected during the solvent injection. The limits are usually established as lower or equal to a certain area or height. 6.5.2.4. Detector linearity A mass detector becomes non-linear when the sample concentration increase does not lead to a proportional increase in the formation and detection of ions. The test is conducted by drawing different concentrations and plotting the result in a calibration curve. The correlation factor is evaluated by comparison with the predetermined acceptance results. 6.5.2.5. Mass accuracy The mass accuracy is evaluated by injecting a known standard either using the auto-calibration system or the samples injector. The results are evaluated by comparing the mass obtained using the equipment to the selected standard theoretical value. Each manufacturer has recommended standards and acceptance criteria for this test. 6.5.2.6. Sensitivity The test is conducted by injecting a known standard and checking the signal-noise ratio of the peak specified in the chromatogram thus obtained. Each equipment has different approval values for this test.

75 of 76

Volume II / Module 3 - Analytical Instrumentation

7. REFERENCES
Apostila do Curso de Espectrofotometria de UV-Vis, Varian Brasil (1992). Apostila do Curso de Espectrofotometria de UV-Vis, Varian Brasil (1998). Apostila do Curso de Treinamento de Espectroscopia, Parte A . Testes de performance, Varian Brasil (1999). Manual de Pr-Instalao de Espectrofotmetros UV-Vis, Varian Brasil (1998). Instruo de Trabalho . Manuteno Preventiva - Cary 1/100/3/300, ITST 02 AT Otto Alcides Ohlweiler - Qumica Analtica Quantitativa, Livros Tcnicos e Cientficos Editora, 2a.Edio, RJ (1980). Cienfuegos, F., Vaitsman - , D. Anlise Instrumental, Intercincia, RJ. (2000). Carol II, Collins, Gilberto L. Braga e Pierina S. Bonato - Introduo a Mtodos Cromatogrficos, Editora da UNICAMP, 6a ed., Campinas- SP (1995). Validation Overview , 85-101741-00, Varian (1999) . Back to Basics. Micromass UK (1996). Varian CP-3800 Gas Chromatograph Regulatory Compliance Documentation, Varian, Walnut Creek. (2000). Techniques of Gas Chromatography, Varian, Walnut Creek (1995). E. D. Hoffmann, J. Charette, V. Stroobant - Mass spectrometry - Principles and Applications Wiley, New York (1996). W. Paul, H. Steinwedel, - A new mass spectrometer without a magnetic field, Z. Naturforsch., 8a, 448-450, (1953). E. Fischer, - Three-dimensional stabilization of charge carriers in a quadrupole field, Z. Phys., 156, 1-26, (1959). Snyder, L.R.; Kirkland, J.J. - Introduction to Modern Liquid Chomatography, 2.ed., New York, John Wiley and Sons (1979). Snyder, L.R.; Kirkland, J.J.; Glajch, J.L - Practical HPLC Method Development. 2.ed. New York, John Wiley and Sons (1997). Riley, C.M. - Efficiency, retention, selectivy and resolution in chromatography. In WAINER, I.W.; LOUGH, W.J. eds. High Performance Liquid Chomatography Fundamental Principles and Practice. Glasgow, Blackie Academic & Professional. (1995) Scott, R.P.W - Silica Gel and Bonded Phases: Their Production, Properties and Use in LC. New York, John Wiley and Sons. (1993) Grant, D. W. Capilary Gas Chromatography. West Sussex. Jonh Wiley & Sons. (1996). Scott, R. P. W. Introduction to Analytical Gas Chromatography. New York. Marcel Dekker. (1981). McNair, H. M., Bonelli, E. J. Basic Gas Chomatography. Walnut Creek. Varian. (1969). Zweig, G., Sherma, J. CRC Handbook of Chromatography . Volume II. Cleveland. (1977). Walker, John Q.Chromatographic Systems 2nd ed.Academic Press, INC, San Diego, California USA (1977)
76 of 76