PRODUCT DATA PIXEL HF 100 KHz

Main features
PIXEL HF 100 KHz is a modern high frequency x-ray generator. PIXEL HF 100 KHz is the latest revision of the PIXEL family born in 1984. The 100 KHz model has been introduced in the market in December 2000. The microprocessor control unit provides ease of use and high reliability. PIXEL HF is available in many kind of models, with power from 50 to 80 KW, 1 or 2 tubes, for radiography only or radiofluoroscopy purposes. The high versatility interface of PIXEL HF allows an kind of connections of X-ray tubes or radiological devices (old or new). PIXEL HF offers variety of applications: from basic radiography up to digital cardio/angiographic xray diagnostic applications. Models offering economy benefits combined with appropriate performances are available. Main features: The x-ray generator PIXEL HF can control up to six different devices and feed on two x-ray tubes at the same time. The power supply, part of the 100 KHz technology, has a smaller dimension and weight than the 20 KHz generator (dimension : p 500 mm x l 457 mm x h 1400 mm, weight 135 Kg). The highly ergonomic control panel offers excellent supervision of all parameters and possible regulations (more then 500 anatomical technical program). The wide high contrast Liquid crystal display (295 mm x 250 mm) keeps a constant flow of main information in real time and allows the user to be updated on all exposure data. Precise and accurate setting of the radiographic factors KV/mA/s (by three points technique) or KV/mAs (by the two points technique). Technologically advanced Automatic Exposure Control for connections of maximum four different ultra-thin ionization chambers. High speed starter for 9000 rpm supplied as standard for the 80 KW unit, available as optional for the other models. DIGI-AEC, automatic control device for selection of radiography KV during fluoroscopy examinations. Especially designed to work with digital video fluorography system. Automatic setting of the x-ray dose both during continuous and pulsed fluoroscopy (R/F models). Possibility to adjust frequency and period of single impulse in pulsed fluoroscopy and HCF, this allows interface with many kind of digital sistem. X-ray tube safety protection and real time heat unit calculator, temperature overload signal. Visual and acoustic signalling of: doors open, failed anode rotation and wrong exposure. Very short switching time from Fluoroscopy into Radiography modality. Multilanguage software installed. Wider list of the present x-ray tubes.

August 02

page 2 of 7

PRODUCT DATA PIXEL HF 100 KHz

Certification
The PIXEL HF x-ray generator is a II b class device, according to 93/42 CEE European directive. Det Norske Veritas Italia (certification corporate body number 0434) provides CE certification of Italray PIXEL HF, certificate number 98-OSL-SD-0197. The device is made according to directives 89/336/EEC and the following rules:

CEI EN 60601-1 CEI EN 60601-1-2 CEI EN 60601-1-4 CEI EN 60601-2-7 EN475 CEI 62-2 IEC 878

Model and configurations


PIXEL HF is manufactured in 13 different models, belong of 3 groups of same power: PIXEL HF 650: 50 KW 600 mA PIXEL HF 850: 65 KW 800 mA PIXEL HF 1050: 80 KW 1000 mA Every tree models are available in many different version in operation of the final use, to power supply one or two tubes, only radiography function or radiofluoroscopy and with or without High Speed Starter. Following are report all models and radiographic characteristic. PRODUCT CODE GEN+IR201/A PIXEL HF 650 2RF GEN+IR201/B PIXEL HF 650 2RF
PIXEL HF 650 2 RF HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES RADIO/FLUOROSCOPY, 150KV 600 mA 50 KW PIXEL HF 650 2 RF HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES RADIO/FLUOROSCOPY, 150KV 600 mA 50 KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

NAME

DESCRIPTIONS

NOTE

WITH HIGH SPEED STARTER

GEN+IR201/C PIXEL HF 650 1RF GEN+IR201/D PIXEL HF 650 1RF

PIXEL HF 650 1 RF HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBE RADIO/FLUOROSCOPY, 150KV 600 mA 50 KW PIXEL HF 650 1 RF HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBE RADIOF/LUOROSCOPY, 150KV 600 mA 50 KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

GEN+IR251/A PIXEL HF 650 2R GEN+IR251/B PIXEL HF 650 2R

PIXEL HF 650 2R HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES ONLY RADIOGRAPHY 150KV 600 mA 50 KW PIXEL HF 650 2R HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES ONLY RADIOGRAPHY 150KV 600 mA 50 KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

August 02

page 3 of 7

PRODUCT DATA PIXEL HF 100 KHz

PRODUCT CODE

NAME

DESCRIPTION
PIXEL HF 650 1R HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBE ONLY RADIOGRAPHY 150KV 600 mA 50 KW PIXEL HF 650 1R HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBE ONLY RADIOGRAPHY 150KV 600 mA 50 KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

NOTE

GEN+IR251/C PIXEL HF 650 1R GEN+IR251/D PIXEL HF 650 1R

WITH HIGH SPEED STARTER

GEN+IR281/A PIXEL HF 850 2RF GEN+IR281/B PIXEL HF 850 2RF

PIXEL HF 850 2 RF HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES RADIO/FLUOROSCOPY, 150KV 800mA 65KW PIXEL HF 850 2 RF HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES RADIO/FLUOROSCOPY, 150KV 800mA 65KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

GEN+IR281/C PIXEL HF 850 1RF GEN+IR281/D PIXEL HF 850 1RF

PIXEL HF 850 1 RF HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBES RADIO/FLUOROSCOPY 150KV 800mA 65KW PIXEL HF 850 1 RF HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBE RADIO/FLUOROSCOPY 150KV 800mA 65KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

GEN+IR301/A PIXEL HF 1050 2RF

PIXEL HF 1050 2 RF HIGH FREQUENCY X-RAY GENERATOR FOR 2 TUBES RADIO/FLUOROSCOPY 150KV 1000mA 80KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

GEN+IR301/B PIXEL HF 1050 1RF

PIXEL HF 1050 1 RF HIGH FREQUENCY X-RAY GENERATOR FOR 1 TUBES RADIO/FLUOROSCOPY 150KV 1000mA 80KW WITH HIGH SPEED STARTER FOR 9000 rpm TUBE

WITH HIGH SPEED STARTER

Available optional.
PRODUCT CODE GEN+IR900/A CONSOLLE
PEDESTAL
PIXEL HF CONSOLLE PEDESTAL HIGH 80 cm, WEIGHT 15 Kg WALL SUPPORT BRACKET FOR PIXEL CONSOLLE WEIGT 3 Kg EXPOMAT INTERFACE BOARD FOR CONNECTION OF IONIZATION CHAMBERS ULTRA THIN IONIZATION CHAMBER, FOR AUTOMATIC EXPOSURE CONTROL FREQUENCY TRIPLICATOR FOR 9000 rpm TUBE SUPPLY.

NAME

DESCRIPTION

NOTE

OPTIONAL OPTIONAL OPTIONAL OPTIONAL OPTIONAL when


is not include

GEN+IR901/A CONSOLLE WALL


BRACKET

GEN+IR902/A EXPOMAT
INTERFACE

GEN+IR903/A IONIZATION
CHAMBER

GEN+IR904/A HIGH SPEED


STARTER

August 02

page 4 of 7

PRODUCT DATA PIXEL HF 100 KHz

Technical features
Electric features PIXEL HF 650 Max Power Power supply Frequency line Instant current Standby current Instant max power Line resistance X- Ray features PIXEL HF 650 Radiography KV Radiography KV adjustment High tension precision Radiography mA Radiography mA adjustment Radiography times Radiography times adjustment Fluoroscopy KV Fluoroscopy KV adjustment Fluoroscopy mA Fluoroscopy mA adjustment Pulse fluoroscopy and HCF KV Frame per second Radiographic mAs, 2 point technique 40 150 KV 111 steps of 1 KV +/- 3% + 1 KVp 25 600 mA 9 steps 1 5000 mSec 28 valori 40 125 KV 85 steps di 1 KV 0,5 6 mA 0.1 mA Steps 40 125 KV 30 esp/s
0.90 800 mAs @ 40 to 80 KV 0.90 600 mAs @ 81 to 120 KV 0.90 400 mAs @ 121 to 150 KV

PIXEL HF 850 65 KW

PIXEL HF 1050 80 KW

50 KW

380-400 Vac +/- 10%, 3 +3N 50/60 Hz 120 A / fase 5 A su 1 fase 85 KVA for exposition at 50 KW 0,15 ohm 150 A / fase 5 A su 1 fase 100 KVA for exposition at 65 KW 0,12 ohm 200 A / fase 5 A su 1 fase 145 KVA for exposition at 80KVA 0,10 ohm

PIXEL HF 850 40 150 KV 111 steps of 1 KV +/- 3% + 1 KVp 25 800 mA 10 steps 1 5000 mSec 28 valori 40 125 KV 85 steps di 1 KV 0,5 6 mA 0.1 mA Steps 40 125 KV 30 esp/s
0.90 800 mAs @ 40 to 80 KV 0.90 600 mAs @ 81 to 120 KV 0.90 400 mAs @ 121 to 150 KV 28 steps @ 40 to 80 KV

PIXEL HF 1050 40 150 KV 111 steps of 1 KV +/- 3% + 1 KVp 25 1000 mA 11 steps 1 5000 mSec 28 valori 40 125 KV 85 steps di 1 KV 0,5 6 mA 0.1 mA Steps 40 125 KV 30 esp/s
0.90 800 mAs @ 40 to 80 KV 0.90 600 mAs @ 81 to 120 KV 0.90 400 mAs @ 121 to 150 KV 28 steps @ 40 to 80 KV

Radiographic mAs adjustment 2 point technique

28 steps @ 40 to 80 KV

27 steps @ 81 to 120 KV 27 steps @ 81 to 120 KV 27 steps @ 81 to 120 KV 26 steps @ 121 to 150 KV 26 steps @ 121 to 150 KV 26 steps @ 121 to 150 KV

August 02

page 5 of 7

PRODUCT DATA PIXEL HF 100 KHz

Operating modes of PIXEL HF Device interface User interface Up to 6 different accessory powered up by PIXEL HF, identifiable by unequivocal different graphic symbols, separate for tube 1 or tube 2. Radiography: setting of the radiographic factors KV/mA/s and x-ray command by consolle. Fluoroscopy: setting of the radiographic factors KV/mA from consolle or accessory, x-ray command from accessory. 3 Point technique: are variable KV, mA and msec 2 Point technique: are variable KV, mAs More than 500 examination selection program The symbols (usually present in normal use on display) change in reverse mode, in the following event. RX tube maximum load limitation. Delivered power limitation. Max 600 mAs limitation. Anode rotation safety. Filament safety.

Use technique Organ/anatomic programming mode Consolle signal

The RX tube sheath has a high temperature.

Ray emission break, one of the diagnostic area doors is open.

UTH
100%
KV set in reverse mode: KV

Thermal capacity X Ray tube limit.

High voltage discharge

Installation and warranty


PIXEL HF X ray generator can be installed in preexisting diagnostic room or in a new room, by our technician staff or customer technician taught by Italray. The Italray technical staff can provide the best position plan of PIXEL HF and other device in the customer room. In the standard configuration we supply the generator with 14 meters cable. Italray production staff can provide a preinstallation of generator and all different accessory. All cable can be personalised for the customer exigency. All ITALRAY device has one year factory warranty, from delivery date. The customer can extend the warranty by subscription of technical service agreement.

August 02

page 6 of 7

PRODUCT DATA PIXEL HF 100 KHz

Dimension and encumbrance


Encumbrance of PIXEL HF: Length 500 x Width 457 x Height1400 mm The weight can change by the different configurations from 135Kg to 140 Kg Consolle dimension : Length 290 x Widht 360 x Height 80 mm Consolle weight, 5 Kg The consolle pedestal height is 800 mm and weight, 15 Kg, Its use to stable floor fixing. Wall consolle support: It has the same consolle dimension, its used in very small room to fix the consolle on the wall.

Packing and transport


The X ray generator can be packed in two different ways. For sea transport: wooden case with barrier bag. For normal transport: wooden case. Wooden case dimension: 68 x 72 x 176 cm.

Via Del Parlamento Europeo 9/D 50010 Scandicci Firenze ITALY tel. +39/055/721083 fax. +39/055/7310177 www.italray.it email info@italray.it