Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Report
1) TABLE OF RAW DATA
At 1500 rpm the discharge and suction pressure doesn’t equal to the differential pressure reading but as
the rpm was increased to 2000 rpm the sum of suction and discharge pressure was very close to
differential pressure reading.
Comments:
Reading from the flow meter and calculated flow by timing were very close to each other. You can see
the comparison in Table 12.
1. Pump Performance
Flow of Water (lb/s) Calculated Flow of Water (gpm) Actual Flow of Water (gpm)
6.55 47.08 49
5.6 40.28 40.5
3.56 25.58 29
2.67 19.18 20
1.4 10.07 10
Comment: Calculated flow of water was very close to the actual measured
flow of water.
Differential Pressure
Suction (inHg) Discharge ( lbs/in2) Calculated ( lbs/in2)
(lbs/in2)
4.5 0 5 5
4.7 0 5.5 5.5
4.7 0 5.8 5.8
4.9 0 6.2 6.2
5.1 0 6.5 6.5
Differential Pressure
Suction (inHg) Discharge ( lbs/in2) Calculated ( lbs/in2)
(lbs/in2)
7.5 0.5 8.2 8.45
8 1.1 10 9.46
8.5 0 11 11
8.9 0 12 12
9.5 0 13 9.5
Table 15: Calculated Differential Pressure at 1500rpm
Differential Pressure
Suction (inHg) Discharge ( lbs/in2) Calculated ( lbs/in2)
(lbs/in2)
11.9 3 13.5 12.03
14.3 0 21 21
A pump starting out full or air might not have enough pressure to pump put
the air, therefore it will never achieve the flow of liquid. To prime the pump, an
operator should fill the pump up with the liquid to remove all the air. Usually there is
a check valve fitted in the suction line so that the liquid will not drain out if the
pump is stopped. The reason to do pump prim is that if there were vapours present
in the pump while the motor is on, the impeller of the pump becomes gas-bound
and the pump will be powerless.
8) CAVITATION
At high flow rate and high rpm we noticed a white strip. The pump sounded
different and we could hear growling sounds and vibration. As the suction valve was
closed the liquid inside the pump started vaporizing due to pressure drop to vapor
pressure limit. Some bubbles could also be seen in the pump.
Conclusion
• When operating the pump, we need to first prime the pump therefore the
data won’t have a high % error in calculations
• When cavitation occurs within the pump, to correct it you must change
operating conditions such as increasing suction pressure and maintaining
pipe lines.
• The pump has a better efficiency when running at higher rpm’s
• The measured flow rate and the calculated flow rate were not the same due
to equipment.
• The differential pressure measured and the differential pressure calculated
was not equal due to lack of pressure sensitivity in pressure gauges.
Recommendations
Using gauges which sense lower pressure and this will provide more
accurate readings.
Torque arm sometimes got stuck while weights were placed on it, a
digital weighing machine will provide more accurate results.
References