Sei sulla pagina 1di 2

PRODUCT CODE PL.

080 (100 tests)


PRO-LAB PROLEX ™ STAPH LATEX KIT PRODUCT CODE PL.081 (300 tests)
DIAGNOSTICS For Identification of Staphylococcus aureus
FOR IN VITRO DIAGNOSTIC USE

INTENDED USE
4. Human source materials used in the manufacture of the reagent INTERPRETATION OF RESULTS
Prolex™ Staph Latex Kit provides a rapid method to distinguish have been tested and found negative for antibody to HIV and HBs
staphylococci (particularly Staphylococcus aureus) which possess coag- Ag. Although, the concentration of such materials in the reagent Positive results: A significantly rapid strong clumping (within 20
ulase (clumping factor) and / or protein A from other species of are very low, no known test method can offer complete assurance seconds) with the Latex Test Reagent and no agglutination with
staphylococci in cultured specimens. that infectious agents are absent. Therefore, test latex reagent Negative Control Latex Reagent . Reaction occurring after the 20 sec-
should be handled using the same safety precautions employed onds should be ignored.
SUMMARY AND EXPLANATION when handling any potentially infectious material. Negative results: A milky appearance without visible agglutination
5. The kit is intended for in vitro diagnostic use only. of the latex particles.
Although most Staphylococcus species are common inhabitants of the 6. Do not freeze the latex reagents.
skin and mucous membranes, certain species have been found fre- 7. The procedures, storage conditions, precautions and limitations
quently as etiological agents of a variety of human and animal infec- specified in these directions must be adhered to in order to obtain LIMITATION OF THE PROCEDURE
tions. valid test results.
Superficial supportive infections caused by S. aureus are the most 1. False negative or false positive results can occur if inadequate
common human staphylococcal infections.1 Food poisoning and toxic amounts of culture or reagent are used.
shock syndrome also have been attributed to infection with S. aureus2. STABILITY AND STORAGE 2. Some rare isolates of staphylococci, notably S. hyicus and S. inter-
The coagulase tube test was previously the most widely test used for medius, may agglutinate the latex reagent4.
the identification of S. aureus, however, false positives have led to Shelf life of kit is 12 months from the date of manufacture. The expi- 3. Some streptococci and possibly other organisms that possess
criticism of this method 3. Essers and Radebold 3 described a rapid ration date is stated on the outer label and the vial labels. All kit com- immunoglobulin binding factors and some species such as
slide agglutination test which has been shown to be a reliable method ponents should be stored at 2-8°C. Escherichia coli may also agglutinate latex reagents non-
for identification of S. aureus in the routine bacteriological laboratory. specifically5.

PRINCIPLE OF THE TEST SPECIMEN COLLECTION AND PREPARATION OF CULTURES PERFORMANCE CHARACTERISTICS

The Prolex™ Staph Latex Kit utilizes polystyrene latex particles For specific procedures regarding specimen collection and prepara- The Prolex™ Staph Latex Kit was evaluated in a study which includ-
which have been sensitized with fibrinogen and IgG. When tion of primary cultures refer to a standard microbiology textbook. In ed a total of 376 primary isolates. Genus and species were originally
Staphylococcal colonies which possess clumping factor and / or pro- general, a fresh ( 18-36 hour incubation) Gram positive isolate grown identified by PCR and Staphylococcus aureus isolates were confirmed
tein A are mixed with the latex reagent, the latex particles agglutinate on non-selective media such as blood agar should be used. by tube coagulase tests. The 376 isolates used in this study include 85
strongly within 20 seconds. methicillin susceptible S. aureus, 104 Methicillin resistant S. aureus
and 187 other staphylococci. (1 S. oxford; 35 S. epidermidis; 4 S. sciuri;
MATERIALS SUPPLIED 2 S. cohnii; 2 MRSA-H; 13 S. simulans; 25 S. CNS, 10 S. capitis; 4 S.
REAGENTS xylosus; 25 S. saprophyticus; 10 S. hominis; 6 S. species; 33 S.
Staph Test Latex Reagent haemolyticus; 10 S. warneri; 6 S. auricularis and 1 S. axium.) In this
Staph Test Latex Reagent: Two vials each containing 2.5 ml (100 Negative Control Latex Reagent study Prolex™ Staph Latex Kit correctly identified 100% of the iso-
test/kit - PL.080) or 7.5 ml (300 test/kit - PL.081) of latex particles Disposable cards with eight test circles labelled 1 through 8. lates which were originally identified by PCR as Methicillin suscepti-
coated with IgG and human fibrinogen. The latex particles are sus- Disposable mixing sticks. ble or Methicillin resistant S. aureus. However, the kit falsely identi-
pended in buffer containing 0.1% sodium azide as preservative. fied 6 strains out of 187 strains of "other staphylococci". The Prolex™
Negative Control Latex Reagent: One vial containing 2.5 ml (PL.080) Staph Latex Kit was shown by this study to have a sensitivity of 100%
or 7.5 ml (PL.081) of unsensitized latex particles suspended in buffer TEST PROTOCOL and a specificity of 96.8% for identification of S. aureus.
containing 0.1% sodium azide as preservative.
1. Bring the latex reagent to room temperature (22-28°C) for about REFERENCES
10 minutes prior to use.
PRECAUTIONS 2. Resuspend the latex reagent by shaking prior to use. 1. Schleifer, K.H., and Kloos, W.E. (1975). Int. J. Syst. Bacteriol. 25:
3. Dispense 1 drop of Staph Test latex reagent onto separate circles 50-61.
1. Do not use reagents after expiry date shown on product label. on the test card. 2. Schlievert, P.M., Shands, K.N., Dan, B.B., Schmid, G.P. and
2. Reagents contain sodium azide. Sodium azide can react explo- 4. Take a mixing stick and transfer two suspect colonies onto a circle. Nishimura, R.D. (1981). J. Infect. Dis. 143: 509-516.
sively with copper or lead if allowed to accumulate. Although the Mix this into the test latex reagent and spread to cover the com- 3. Essers, L. and Radebold, K. (1980). J. Clin. Microbiol. 12: 641-643.
amount of sodium azide in the reagents is minimal, large quanti- plete area of the circle. 4. Phillips, W. and Kloos, W. (1981). J. Clin. Microbiol. 14: 671-673.
ties of water should be used when flushing used reagents down 5. Gently rock the card allowing the mixture to flow slowly over the 5. Myhre, E.B. and Kuusela, P. (1983). Inf. Imm. 40: 29-34.
the sink. entire test ring area.
3. Safety precautions should be taken in handling, processing and 6. At 20 seconds, under normal lighting conditions, observe for
discarding all clinical specimens as a pathogenic organism may be agglutination.
present. 7. If the result is positive, repeat steps 2 to 6 in the same way, using
the negative control latex reagent.
Manufactured in Canada by Pro-Lab, Inc. (2003 12)
Canada ☎ 905 731 0300 Fax 905 731 0206 U.S.A.☎ 800 522 7740 Fax 800 332 0450 U.K.☎ 0151 353 1613 Fax 0151 353 1614
20 Mural Street, Unit #4, Richmond Hill Ont. L4B 1K3 Austin, Texas, 78754 Neston, Wirral, CH64 3UJ
= Use by

= Lot number

= Attention, see instructions for use

REF = Catalogue number

= Manufacturer

= Authorized Representative in the European Community.

= Contains sufficient for <n> tests

= in vitro diagnostic medical device.

= Temperature limitation

= Consult instructions for use.

Potrebbero piacerti anche