Sei sulla pagina 1di 8

ARTICLE2

RADIOGRAPHICEXAMINATION
FIELD T-210 applicable material specification or
The radiographic method described in this referenced code section, with
article for the examination of materials, additionalconditioning, if necessary, by any
including castings and gaskets is used in suitable procedure to such a degree that the
conjunction with Article 1, General resulting radiographic image due to surface
Requirements. Definitions of the terms used irregularities cannot mask or be confused with
in this article are in Appendix V Mandatory the image of any discontinuities.
of this article. Certain product-specific, -
specific technical and application-specific T-222.2 welds. Weld waves or irregularities
requirements are also given in other of the weld surface, both inside (where
mandatory appendices to this article, which accessible) and outside the regulation shall
are listed in the table of contents. These be eliminated by any suitable procedure to
additional requirements must also be met such a degree that the resulting
when an appendix is applicable to the radiographic image due to surface
radiographic or fluoroscopic examination irregularities cannot mask or be confused
being performed. with the image of any discontinuities. The
finished surface of all butt welded joints
T-220 GENERAL REQUIREMENTS may be flush with the base material or may
have reasonably uniform crowns, with the
T-221 procedure requirements reinforcement not to exceed the amount
indicated in the referenced Code Section.
T-221.1 radiographic examination shall be
performed in accordance with a written T-223 radiation back scattering
procedure. Each procedure should include A lead symbol "B", with minimum
at least the following information, as dimensions of 1/2 in. (13 mm) high and
appropriate: 1/16 in. (1.6 mm) thick, shall be attached to
(A) the type of material and thickness the back of each film holder during each
range, exposure to determine if the backscatter
(B) of isotopes or maximum X-ray voltage radiation is the exposure of the film.
used, T-224 identification system
(C) a minimum source-object distance,
(D) the side-to-side distance from the object A system must be used to produce a
source to the film at the minimum source- permanent identification on the radiograph
to-object distance, traceable to the contract, components, weld
(E) the maximum font size, or weld seam, or part numbers, as
(F) of the mark and name of the film, and appropriate. In addition, the manufacturer's
(G) The screens used. symbol or name and date of the radiograph
shall be clearly legible and indelibly
T-221.2 Demonstration Procedure. included on the radiograph. This
Demonstration of the density and ICI image identification system does not necessarily
requirements of the written phase on require that the information appears as
production or technical radiographs shall be radiographic images. In any case, such
considered as satisfactory evidence of information should not obscure the area of
compliance with this procedure. interest.
T-225 Limitations of Density Monitoring
T-222 Surface preparation Radiographs
Either a densitometer or stepped wedge
T-222.1 Materials including cast iron. film comparison is used to judge the density
Surfaces shall meet the requirements of the of the film.
SECTION V
The extent of the radiographic
T - 226 2001 examination shall be as specified by the
T-226 scope of review referenced Code Section.
with the requirements or alternates
T-230 EQUIPMENT AND MATERIALS permitted in SE-747, except that the larger
wire number or identity number may be
T-231 Film omitted. ASME standard IQIs shall consist
T - 231.1 Selection. The radiographs of those in Table T-233.1 for hole type and
will be taken using industrial those in Table T-233.2 for wire type.
radiographic film.
T-233.2 Alternative ICI Design. IQIs
T-231.2 processing. Standard Guide designed and manufactured in accordance
for Quality Control of Industrial with other national or international
Radiographic Film Processing, SE-999, standards may be used provided that the
or in paragraphs 23 to 26 of the requirements of either (a) or (b) below are
Standard Guide for Radiographic met, and if the material requirements of T-
Examination SE-94 is used as a guide 276.1 are met.
for film processing. (A) Hole Type IQIs. The calculated
equivalent sensitivity ICI (EPS), per SE-
T-232 intensifying screens 1025, Appendix X1, is equal to or better
intensifying screens may be used when than the required standard hole type ICI.
performing radiographic examination in (B) Type of cable IQIs. The diameter of the
accordance with this article. alternative ICI core wire is equal to or
smaller than the required standard ICI core
wire.

T-233 Image Quality Indicator (IQI)


Design

T-233.1 IQI Design Standard. IQIs will


be either the type of hole or the type of
wire. Hole type IQIs shall be manufactured
and identified in accordance with the
requirements or alternates permitted in SE-
1025. Type of wire IQIs shall be
manufactured and identified in accordance
Measurement of Focal Points of Industrial X-
Ray Tubes by Image Hole. (B) Iridium-192
Sources. For iridium-192, the source size can be
determined in accordance with SE-1114,
Standard Test Method for Determining the
Focal Size of Industrial Iridium-192
Radiographic Sources.

T - 234 Facilities for the visualization of T-262 densitometer and wedge Step Film
X-rays comparison
Viewing facilities shall provide subdued
background illumination of an intensity T-262,1densitometers.
such that it does not cause distracting glare, Densitometers shall be calibrated at least every
shadows, reflections or reflections on the 90 days during use as follows: (A) A national
radiograph. The equipment used to view standard-pitch tablet or a wedge-pitch
calibration film, which is traced back to a
radiographs for interpretation shall provide
national standard-pitch tablet and which has at
a variable source of light sufficient for the least 5 pitches with neutral densities of at least
essential ICI hole or designated wire to be 1.0 to 4.0, shall be used. The calibration film of
visible for the specified density range. The the stepped wedge has been verified in the last
viewing conditions should be such that light year, compared to a national standard step tablet
from the entire outer edge of the radiograph unless, prior to first use, it is kept in a sealed
or coming through low-density portions of light- and water-tight package supplied by the
the radiograph does not interfere with the manufacturer. Wedge calibration films can be
interpretation. used without verification for one year after
opening, as long as it is within the shelf life
stated by the manufacturer. (B) will be followed
T-260 CALIBRATION
by the densitometer manufacturer's step-by-step
instructions for densitometer operation. (C) The
T-261 Font size density of the steps closest to 1.0, 2.0, 3.0, and
4.0 in the tablet stage or wedge stage national
T - 261.1 Verification of font size. standard calibration film will be read. (D) The
Supplier publications, such as technical densitometer is acceptable if the density
manuals, drop curves, or written statements readings do not vary by more than ± 0.05
documenting the actual or maximum font density units from the actual density shown on
size or equipment manufacturer's focal the step tablet film or national standard step
point, shall be acceptable as verification of wedge calibration. T-262.2 Film comparison
wedge step. Step wedge comparison films shall
font size.
be verified prior to first use, unless otherwise
performed by the manufacturer, as follows: (A)
T - 261.2 Determination of size of origin. The density of the steps in a stepped wedge
When manufacturer or distributor comparison film shall be verified by means of a
publications are not available, the font size can calibrated densitometer. (B) The step wedge
be determined as follows: (A) X-Ray machines. comparison film is acceptable if the density
For X-ray machines operating at 500 kV and readings do not vary by more than ± 0.1 density
below, the focal spot size can be determined by units from the density indicated on the step
the pin-hole method, 1 or in accordance with wedge comparison film.
SE- 1165, Standard Test Method for the

2001 Section V
T-262.3 Periodic Verification readings are within ± 0.05 of the calibration
readings determined in T-262.1 C.
(a) Densitometers. Periodic calibration (B) Step of the film comparison wedge.
verification checks shall be conducted as Verification checks are performed annually by
described in T-262.1 at the beginning of each T-262.2.
shift, after 8 hours of continuous use, or after a T-262.4 Documentation
change of openings, whichever occurs first. The
densitometer is acceptable if the density (a) Densitometers. Densitometer calibrations
required by T-262.1 are documented, but the
actual readings of each step do not have to be so that there is no overlap of the areas to be
recorded. Periodic densitometer verification interpreted. When full coverage is required, a
checks required by T-262.3 (a) do not have to minimum of two exposures taken 90 deg. The
be documented. one to the other will be made for each joint.
(B) the staggered wedge of calibration films. (2) Alternatively, the weld can be radiographed
Step wedge calibration film checks required by with the radiation beam positioned so that the
images of the two walls overlap. When full
T 262.1(a), will be documented, but the actual
coverage is required, a minimum of three
readings of each step do not have to be
exposures taken at any one of 60 degrees. or
recorded. 120 degrees. + to each other will be made for
(C) the wedge step of the film comparison. each joint. (3) Additional exposures are entered
Wedge step film comparison checks required by if the required radiographic coverage cannot be
T-262.2 and T-262.3(b) shall be documented, obtained using the minimum number of
but the actual readings for each step need not be exposures indicated in (b) (1) or (b) (2) above.
recorded. T-272 Radiation Energy
The radiation energy used by any radiographic
T-270 EXAMINATION technique shall meet the ICI image and density
requirements of this article.
T-271 RadiographicTechnique2 A single-wall
exposure technique should be used for
radiography whenever practical. When it is
impractical to use a single-wall technique, a T-273 radiation direction
double-wall technique is used. An adequate The direction of the central beam of the
number of exposures were made to demonstrate radiation should be centered on the area of
that the required coverage has been obtained. interest whenever practical.
T-274 geometric lack of sharpness
T-271.1 One-wall technique. In the one-wall Geometric unsharpness of the radiograph shall
technique, radiation passes through only one be determined according to: Ug = Fd / D where
wall of the weld (material), which is considered Ug=geometrical unsharpness size
for acceptance in the radiograph. F = source size: the maximum projected
dimension of the radiation source (or effective
T-271.2 Double wall technique. When it is focal point) in the plane perpendicular to the
impractical to use a single-wall technique, one distance D from the weld or radiographed
of the following double-wall techniques will be object, at. D = distance from the radiation
used. source for welding or radiographed object, in. D
= distance from the side of the weld source or
(A) Visualization of a wall. For materials and radiographed object to the film NOTE: Refer to
component welds, a technique can be used in the Standard Guide for Radiographic Testing
which the radiation passes through two walls SE-94 for a method of determining geometric
and only the weld (material) on the film side unsharpness. Alternatively, a nomogram as
wall is seen for acceptance on the radiograph. shown in the Standard Guide for Radiographic
When full coverage of circumferential welds Testing SE-94 can be used.
(materials) is required, a minimum of three T-275 Location Location markers Location
exposures taken at 120 degrees. each other will markers (see Fig. T-275), which are to appear as
be made. radiographic images on the film, shall be placed
(B) Make double display wall. For materials and on the part, not on the exposure/cassette holder.
welds on 31/2 in. components. (89 mm) or less Their locations shall be indelibly marked on the
nominal outside diameter, a technique can be surface of the part being radiographed when
used in which radiation passes through two permitted, or on a map, in a manner that allows
walls and the weld (material) on both walls is the area of interest on a radiograph to be
seen for acceptance on the same radiograph. For accurately traced to its location on the part, for
a double wall display, only one ICI on the the required retention period of the radiograph.
source side will be used. Care must be taken to Tests will also be provided on the X-ray that the
ensure that the required geometric unsharpness required coverage of the region being examined
is not exceeded. If the geometric unsharpness has been obtained. Location markers will be
requirement cannot be met, then single wall placed as follows.
display is used. (1) For welds, the radiation T-275.1 Display of a wall
beam can be offset from the plane of the weld at (A) Source-secondary markers. Location
an angle sufficient to separate the images of the markers shall be placed on the source side when
source-side and film-side portions of the weld radiographing the following: (1) flat
components or in longitudinal joints cylindrical (B) Weld Without reinforcement. The thickness
or tapered components; (2) curved or spherical on which the ICI is based is the nominal single
components whose concave side is toward the wall thickness. Back-up rings or strips will not
source and when the "source-to-materials" be considered as part of the weld thickness in
distance is less than the inside radius of the the selection of ICI.
component; (3) curved or spherical components T-276.3 Welds joining dissimilar materials or
whose convex side is toward the source. (B) welds with dissimilar filler metal
Film-side markers (1) placeholders are placed When the weld metal is of an alloy group or
on the film side when radiographing either grade that has a radiation attenuation that differs
curved or spherical components whose concave from the base material, the selection of the ICI
side is toward the source and when the "source- material is based on the weld metal and be in
to-material" distance is greater than the inner accordance with T-276.1. When the density
radius. (2) As an alternative to the source-side limits of T-282.2 cannot be satisfied with a
placement of T-275.1 (a) (1), location markers single ICI, and the area of exceptional density(s)
may be placed on the film side when the is at the interface of the weld metal and the base
radiograph shows coverage beyond the location metal, the selection of materials for additional
markers to the extent shown in the figure. T- IQIs shall be based on the base material and be
275, sketch (e), and when this alternative is in accordance with T-276.1.
documented in accordance with T-291. T-277 Use of IQIs for follow-up radiographic
Markers (c) both sides. Location markers can be examination
placed either on the source side or film side T-277.1 Placement of IQIs
when radiographing either curved or spherical (A) ICI source side(s). The ICI(s) shall be
components whose concave side is toward the placed on the source side of the part being
source and the "-source-to-material" distance is examined, except for the condition described in
equal to the inner radius of the component. T-277.1(b). When, due to part or weld
T-275.2 double wall display. For double-wall configuration or size, it is not practical to place
visualization, at least one placeholder is placed the ICI(s) in the part or weld, the ICI(s) can be
adjacent to the weld (or on the material in the placed in a separate block. Separate blocks are
area of interest) for each radiograph. made from the same or radiographically similar
T-275.3 Marker placement mapping materials (as defined in SE-1025) and can be
Location. used to facilitate positioning of the ICI. There is
When inaccessibility or other limitations no restriction on separate block thickness,
prevent placement of markers as stipulated in T- provided that the IQI/area-of-interest density-
275.1 and T-275.2, a dimensional map of the interest tolerance requirements of T-282.2 are
actual marker location and shall be met.
accompanied by radiographs to demonstrate that (1) The IQI on the supply side of the separate
full coverage has been obtained. The block is placed closer to the film than the
T-276 ICI Selection source side of the part being radiographed.
T-276.1 materials. IQIs are selected from the (2) The separated block shall be placed as close
same group of alloy material of grade o as as possible to the part being radiographed.
identified in SE-1025, or SE-747, as the case (3) When using hole-type IQIs, the dimensions
may be, or from a group of alloy material of of the blocks shall exceed the ICI dimensions
grade o with less radiation absorption than the such that the outline of at least three sides of the
material being radiographed. ICI image shall be visible on the radiograph.
T-276.2 size. The designated ICI hole or (B) IQI Film-side(s). When inaccessibility
essential wire shall be as specified in Table T- prevents placement of the IQI(s) on the source
276. A thin or thicker ICI-type hole can be side, the IQI(s) is placed on the side of the film
substituted for any section thickness listed in in contact with the part being examined. An "F"
Table T-276, provided an equivalent ICI lead card will be placed next to or in the ICI(s),
sensitivity is maintained. See T-283.2. but may not hide the essential hole when using
(A) welds with reinforcements. The thickness hole IQIs.
on which the ICI is based is the nominal single (C) Placement of IQI Welds - Hole IQIs. The
wall thickness plus estimated weld ICI(s) can be placed adjacent to or in the weld.
reinforcement not to exceed the maximum The identification number(s) and, when used,
allowed by the referenced Code Section. Back- the lead chart "F" will not be applied in the area
up rings or strips will not be considered as part of interest, except when geometric configuration
of the thickness in the selection of the ICI. makes it impractical.
Actual measurement of the weld armor is not (D) The placement of IQI wire welds IQIs. The
required. IQI(s) shall be placed in the weld so that the
length of the wires is perpendicular to the length
of the weld. Identification numbers and, when single exposure, at least three IQIs, spaced
used, the lead chart "F" will not be applied in approximately 120 degrees apart. In addition,
the area of interest, except when geometric they are required under the following
configuration makes it impractical. conditions: (A) When a complete circumference
(E) IQI placement of materials other than welds. is X-rayed using one or more film holders, or;
The IQI(s) with the IQI(s) identification (B) When a section or sections of a
number, and when used, the lead letter "F", can circumference, where the length between the
be placed in the area of interest. ends of the outermost sections spans 240
T-277.2 IQIS number. When one or more film degrees or more, is X-rayed using one or more
holders are used for an exposure, at least one film holders. Additional film locations may be
ICI image shall appear on each radiograph, necessary to obtain the required ICI spacing. (5)
except as indicated in (b) below. For spherical components where the source is
(A) Multiple IQIs. If the requirements of T 282 placed at the center of the component for a
are met by the use of more than one ICI, one single exposure, at least three IQIs, with one
shall be representative of the lightest area of placed at each end of the radiographed span of
interest and the other the darkest area of the radiographed circumference and one at the
interest; the densities involved in the radiograph approximate center of the duration, are
shall be considered as having acceptable necessary, under the following conditions: (A)
density. When a section of a circumference, the length of
(B) Special cases3 which is greater than 120 degrees. . and less
(1) For cylindrical components where the source than 240 degrees tan, is radiographed using only
is positioned on the axis of the component for a a film holder, or; (B) When a section or sections
single exposure, at least three IQIs, spaced of a circumference, where the length between
approximately 120 deg. In addition, they are the ends of the outermost sections span less than
required under the following conditions: 240 degrees. More than one film holder is used
(A) When the entire circumference is for radiography. (6) In (4) and (5) above, where
radiographed using one or more film holders, other welds are radiographed simultaneously
or; with the circumferential weld, an additional ICI
(B) When a section or sections of the should be placed over the other weld. (7) When
circumference, where the length between the radiographing a series of components in a circle,
ends of the outermost sections spans 240 at least one ICI should be shown in each
degrees or more, is X-rayed using one or more component image. (8) In order to maintain
film holders. Additional film locations may be continuity of records relating to subsequent
necessary to obtain the required ICI spacing. exposures, all radiographs showing IQIs that
(2) For cylindrical components where the qualify the techniques permitted in accordance
source is placed on the axis of the component with (1) through (6) shall be retained above.
for a single exposure at least three IQIs, with T-277.3 wedges below hole IQIs. For welds, a
one placed at each end of the radiographed wedge of material radiographically similar to
circumference section and one at the that of the weld metal is placed between the part
approximate center of the light, are required and the IQI, if necessary, so that the
under the following conditions: radiographic density across the area of interest
(A) When a section of the circumference, the is no more than less than 15% from (lighter
length of which is greater than 120 degrees. . than) the radiographic density across the ICI.
And less than 240 degree tan, it is x-rayed using The wedge dimensions shall exceed the ICI
only one film holder, or; dimensions such that the outline of at least three
(B) When a section or sections of the sides of the ICI image will be visible on the
circumference, where the length between the radiograph.
ends of the outermost sections span less than T-280 EVALUATION
240 g, is X-rayed using more than one film T-281 X-Ray Quality
holder. All radiographs shall be free of mechanical,
(3) In (1) and (2) above, where sections of chemical, or other imperfections to the extent
longitudinal welds adjacent to the that they do not mask and are not confused with
circumferential weld were radiographed the image of any discontinuities in the area of
simultaneously with the circumferential weld, interest of the object being radiographed. Such
an additional ICI shall be placed in each defects include, but are not limited to:
longitudinal weld at the end of the section (A) fogging;
farthest from the junction with the (B) processing defects such as scratches,
circumferential weld to be radiographed. (4) For watermarks or chemical stains;
spherical components where the source is (C) scratches, finger marks, splices, dirt, static
placed in the center of the component for a marks, smudges, stains, or tears;
(D) false indications due to defective screens. do not apply for the IQI, provided the required
T-282 Radiographic density IQI sensitivity of T-283.1 is met.
T-282.1 Density limitations. The film density T-283 ICI sensitivity
transmitted through the radiographic image of T-283.1 Required Sensitivity. Radiography is
the IQI body of the appropriate hole or adjacent performed with a technique of sufficient
to the designated wire of a wire ICI and the area sensitivity to show the image designated IQI
of interest shall be 1.8 minimum to view film hole and 2T hole, or the essential wire of a wire
only for radiographs taken with an X-ray source ICI. The radiographs will also show the IQI
and 2.0 minimum for radiographs taken with a identifying numbers and letters. If the image of
gamma source. For a composite display of the hole and the designated IQI 2T hole, or
multiple exposures of the film, each film in the essential wire, are not seen on any film in a film
composite set shall have a minimum density of technique, but do show up in the composite film
1.3. The maximum density shall be 4.0, either view, the interpretation will only be allowed in
for simple or composite display. A tolerance of the composite film view.
0.05 is allowed for density variations between T-283.2 Equivalent Hole-Type sensitivity. If a
densitometer readings. thinner or thicker hole type IQI listed in Table
T-282.2 Density variation T-276 has been substituted, an equivalent
(A) general. If the density of the radiograph sensitivity of the ICI, as specified in Table T-
anywhere across the area of interest varies by 283, must have been maintained in addition to
more than minus 15% or more than 30% of the all other requirements for the radiograph having
density across the body of the IQI hole or been met.
adjacent to the designated wire of an IQI wire, T -284 excessive backscatter If an image of the
within the minimum permitted and/or maximum " B " light, as described in T -223, appears on a
dark background of the radiograph, the
backscatter protection is insufficient and the
radiograph shall be considered unacceptable. A
dark image of the "B" on a lighter background is
not a reason for rejection.
T - 285 geometric unsharpness Limitations
geometric unsharpness of the radiograph shall
not exceed the following limits:

T-286 Manufacturer's Evaluation


The manufacturer shall be responsible for the
review, interpretation, evaluation, and
acceptance of the completed radiographs to
ensure compliance with the requirements of
Article 2 and the referenced instruction area.
Density ranges specified in T-282.1, then an
As an aid to examination and evaluation, the
additional ICI will be used for each area or areas
radiographic technical documentation required
of exceptional and retaken radiograph. In
by T-291 should be completed prior to
calculating the allowable variation in density,
evaluation. The X-ray review form required by
the calculation may be rounded to the nearest
T-292 will be performed during the evaluation.
0.1 within the range specified in T-282.1.
Details of the radiographic technique and form
(B) With supplements. When wedges are used
the radiographic opinion documentation will
with hole-type IQIs, the 30% density plus sign
accompany the radiographs. Acceptance was
restriction of (a) may be exceeded above, and
completed prior to submission of the
the minimum density requirements of T-282.1
radiographs and accompanying documentation
to the inspector.
T-290 DOCUMENTATION
T-291 radiographic technique documentation
details
The manufacturer shall prepare and document
the details of the radiographic technique. At a
minimum, the following information must be
provided.
(A) identification, e.g., job/contract number and
heat number (if applicable)
(B) the three-dimensional map (if used) of
marker location in accordance with T-275.3
(C) number of radiographs (exposures)
(D) the X-ray voltage or isotope type used
(E) X-ray machine focal spot size or isotope
size of physical sources
(F) Type of base material and thickness,
thickness of weld, thickness of weld
reinforcement, as the case may be.
(G) minimum from source - to - object at
distance
(H) the side-to-side distance from the object
source to the film at the minimum source-to-
subject distance
(I) film manufacturer and type / Manufacturer's
designation
(J) Film number on each film/cassette holder
(K) single- or double-wall exposure
(L) single or double wall display
T -292 Radiographic Review Form
The manufacturer will prepare a radiographic
review form. At a minimum, the following
information must be provided.
(A) a list of each radiography site
(B) the information required in T-291, by
inclusion or by reference
(C) Evaluation and disposition of material(s) or
weld(s) examined
(D) the identification (name) of the
manufacturer's representative who made the
final receipt of the radiographs
(E) the date of the Manufacturer's assessment

Potrebbero piacerti anche