NEW RED CELL PARAMETERS ON THE SYSMEX XE-2100 AS POTENTIAL MARKERS OF FUNCTIONAL IRON XEDEFICIENCY.

C. Briggs,* R. Rogers, and S.J. Machin,*


*Department

of Haematology, University College Hospital, Grafton Way, London, U.K.

Department of Haematology, Royal Berkshire Hospital, Reading, U.K.

INTRODUCTION
Early detection of iron deficiency is vital to optimise erythropoietin treatment of chronic anaemia associated with renal failure and malignancy. Failure to supply iron at sufficient rate adversely affects patient outcome. Standard biochemical measurement of iron status in these patients may be misleading. Ferritin is an indirect marker of storage iron but not iron supply. It is also an acute phase reactant so may be normal or high despite evolving iron deficiency. Impedance blood cell counters simply provide the traditional red cell parameters and indices that have little bearing on iron availability to erythroid cells. The Technicon series of cell counters (Bayer Diagnostics) measure the volume and haemoglobin content of individual cells and defines the number of cells with a concentration of less than 28g/dl as being hypochromic (% Hypo). On the Advia 120 (Bayer Diagnostics) in addition to % Hypo a more acute evaluation of bone marrow activity is achieved by measuring the reticulocyte haemoglobin content (CHr). In the reticulocyte channel of the Sysmex XE-2100 two new parameters are measured by fluorescence flow cytometry (Fig 1). These are the mean value of the forward scatter light histograms of the red cells (RBC-Y) and reticulocytes (RET-Y). These values seem to equate with the red cell/reticulocyte haemoglobin content With sTfR levels and ferritin there is some overlap between the normal and iron deficient range. 10 out of 70 apparently iron deficient patients had a normal sTfR (6 of these also had a normal ferritin) however all the samples had microcytic and hypochromic indices, low CHr, RBC-Y and RET-Y and high % Hypo so it was decided to include this data. 22 samples from the iron deficient group had normal or in one case high ferritin level (6 had a normal sTfR). Again these all had microcytic and hypochromic indices, low CHr, RBC-Y and RET-Y and high % Hypo. One sample in this group had a ferritin level of 804 g/l with an iron deficient sTfR of 4.83 mg/l. Figure 1. Determination of RBC-Y and RET-Y on the Sysmex XE-2100. RBCRETXETable 1. Normal reference ranges (n=40) and iron deficient ranges ranges (n=70)
FWSc

The soluble transferrin assay (sTfR) was performed using the Bio-stat diagnostics kit. This is a particle enhanced immunoturbidmetric assay based on the detection of an immunoreaction between sTfR and TfRspecific antibodies. This has been shown to be a more sensitive and less variable index of iron status than the conventional measurement of ferritin. Ferritin was measured by an immunoassay technique using an Abbott Architect analyser and Abbott reagents in the routine chemical pathology service.

There was reasonable correlation between sTfR and RBC-Y and RET-Y (both r = 0.68) and sTfR and % Hypo (r = 0.66). Three was slightly less good correlation with CHr (r = 0.55).

Figure 2. Correlation of % Hypo with RBC-Y. Samples broken down into RBCgroups and normal values marked.
Normal % HYPO Normal RBC-Y

Figure 3. Correlation of CHr with RET-Y. Samples broken down into RETgroups and normal values marked.
CHr v RET-Y
Normal RET-Y

% HYPO v RBC-Y
Abnormal % HYPO Normal RBC-Y

220 200

220 200 180 RET-Y


Abnormal % HYPO Abnormal RBC-Y

Normal RET-Y Abnormal CHr

Normal CHr

RBC-Y

RESULTS
Clearly defined ranges for CHr, % Hypo, RBC-Y and RET-Y have been defined in health and untreated simple iron deficiency (Table 1).

180 160 140 120 100 80 0


Normal % HYPO Abnormal RBC-Y

160 140 120


Normal CHr

100 80

Abnormal RET-Y

10 20 30 40 50 60 70 80 90

15 17 19 21 23 25 27 29 31 33 35 37 39
Abnormal RET-Y Abnormal CHr

% HYPO

CHr pg

Yellow = Normal Purple = Pre Dialysis/CAPD Blue = Iron Deficient Green = Haemodialysis

A normal or abnormal sTfR does not always give a corresponding normal or abnormal red cell or reticulocyte parameter by the Advia or XE-2100. If iron deficiency is defined as an sTfR level of greater than 2.29mg/l (the upper limit of our normal range) the ability of the four different cell counter parameters to correctly identify iron deficient samples can be calculated (Table 2). For CHr and RET-Y in every case of false negative or positive results the

Normal Ferritin g/L MEAN MINIMUM MAXIMUM 101.4 14 260 Ferritin g/L MEAN MINIMUM MAXIMUM 42 4 804 sTfR mg/L 1.4 O.93 2.29 CHr pg 31.7 28 36.3 %Hypo 1.0 0.1 4.8 RBC-Y 175.6 168.7 183.8 RET-Y 184.4 171 196.3

RET-Y RBC-Y

samples were from the same patients

Table 2. Ability of CHr, % Hypo, RET-Y and RBC-Y to identify iron deficient RETRBCpatients, defined as having a sTfR level of greater than 2.29 mg/l.

FWSc

FWSc Fl

Iron deficient sTfR mg/L 4.9 1.56 19.1 CHr pg 24.2 19.3 28.8 %Hypo 38.3 4.3 80.8 RBC-Y 137.6 93.6 161.7 RET-Y 141.6 86.4 171.1
CHr TRUE POS OR TRUE NEG = 69.6% FALSE NEG = 22.2% FALSE POS = 8.2% INCORRECT RESULT = 30.4% RET-Y TRUE POS OR TRUE NEG = 70% FALSE NEG = 21.5% FALSE POS = 8.5% INCORRECT RESULT = 30% % HYPO TRUE POS OR TRUE NEG = 73.2% FALSE NEG = 7.7% FALSE POS = 19.1% INCORRECT RESULT = 26.8% RBC-Y TRUE POS OR TRUE NEG = 68% FALSE NEG = 19.4% FALSE POS = 12.6% INCORRECT RESULT = 32%

RBCRBC-Y: the mean value of the forward scatter histogram within the mature RBC population mature RETRET-Y: the mean value of the forward scatter histogram within the reticulocyte population reticulocyte

These new parameters could potentially provide additional information on iron status and erythroid response following recombinant For all samples RET-Y and RBC-Y are closely correlated to CHr, r = 0.94 and 0.91 % Hypo correlated to a better extent with RBC-Y and RET-Y (r = 0.84 and 0.81) than CHr (r = 0.75). Figure 2 demonstrates the correlation of % Hypo with RBC-Y and figure 3 CHr with RET-Y. The normal range cut of values for all parameters are marked on the graphs. This enables truth tables of agreement between each method to be visualized. When comparing % Hypo to RBC-Y there are a significant number of samples with a normal RBC-Y and abnormal % Hypo, these are all from renal patients, either pre dialysis or haemodialysis. There are also some The samples studied were from 40 normal healthy males, 50 untreated iron deficiency samples (selected by indices), 90 from patients with chronic renal disease before dialysis or on continuous ambulatory peritoneal dialysis and 65 from patients receiving long term heamodialysis. 128 of the renal patients were on rHuEPO therapy and 64 on regular iron therapy. An additional 70 sample, 20 from normal healthy males and 50 chronic renal disease patients were collected at the Royal Berkshire hospital and analysed on the XE-2100 and the Technicon H2 for % Hypo. There is no reticulocyte analysis on the H2, CHr was unavailable. In total 315 samples were analysed. After blood count analysis the plasma was frozen for the soluble transferrin assay and ferritin levels. There are only 3 samples with a normal CHr and abnormal RET-Y. Two of these renal patients had other haematological disease, one myeloma and the other acute myeloid leukaemia. There is only one sample with an abnormal CHr and normal RET-Y. There was very weak correlation between ferritin and any of the red cell or reticulocyte parameters. There was no correlation between ferritin and sTfR. In patients with renal disease the mean ferritin level was high at 544g/l in contrast to a level of 2.49 mg/l for the sTfR which would indicate developing iron deficiency. normal % Hypo with an abnormal RBC-Y.

erythropoietin (rHuEPO) therapy in chronic renal failure. The aim of this pilot study was to compare biochemical markers of iron status, ferritin and soluble transferrin receptor levels, with % Hypo, CHr and the two new parameters, RBC-Y and RET-Y and assess their potential predictive value in diagnosing iron deficiency.

CONCLUSIONS
We have clearly defined ranges for CHr, % Hypo, RBC-Y and RET-Y in health and untreated simple iron deficiency. The primary purpose of this study has been achieved. Despite the fact that CHr is a measurement of mean reticulocyte haemoglobin content and RET-Y a measurement of reticulocyte size the correlation between the two in all patient groups was excellent and effectively interchangeable for clinical diagnostic purposes. There is reasonably good correlation between % Hypo and RBC-Y. No correlation was found between sTfR levels and ferritin and only weak correlation between ferritin and the red cell and reticulocyte parameters. This was not unexpected as ferritin is an acute phase reactant. Reasonable correlation was a found between sTfR and RBC-Y, RET-Y and % Hypo, slightly less good correlation with CHr. Where CHr may be used for monitoring functional iron deficiency RET-Y can be used in the same way. Future studies are needed whereby iron therapy is regularly monitored by standard haematological parameters in comparison with the new parameters, RBC-Y and RET-Y, to determine their true predictive value for clinical diagnostic purposes.

MATERIALS AND METHODS


245 fresh peripheral blood samples collected into K3EDTA (Beckton Dickinson) were analysed at University college Hospital London using the XE-2100 for RET-Y and RBC-Y. The same samples were analysed within 4 hours on the Advia 120 for % Hypo and CHr.