Sei sulla pagina 1di 408

Agilent 7500

Inductively Coupled Plasma


Mass Spectrometry
Course Number H8974A
ChemStation Revision 01.XX
NT Operating System

Student Manual
Revision 1

Mass Spectrometry Data Systems

Liquid Chromatography

Capillary
Electrophoresis
Gas Chromatography
Agilent 7500
Inductively Coupled Plasma
Mass Spectrometry
Course Number H8974A
ChemStation Revision 01.XX
NT Operating System

Student Manual
Revision 1

Manual Part Number H8974-90000


Printed in the USA January, 2001
Notice
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to the implied warranties of merchantability and fitness
for a particular purpose.
Agilent Technologies shall not be liable for errors contained herein or for
incidental, or consequential damages in connection with the furnishing,
performance, or use of this material.
No part of this document may be photocopied or reproduced, or translated to
another program language without the prior written consent of Agilent
Technologies, Inc.

Agilent Technologies, Inc


11575 Great Oaks Way
Suite 100, MS 304B
Alpharetta, GA 30319

 2000 by Agilent Technologies, Inc.


All rights reserved
Printed in the United States of America

ii
Table Of Contents
INTRODUCTION: ELEMENTAL ANALYSIS ..........................................................................1
ATOMIC SPECTROMETRY ..............................................................................................................2
ATOMIC MASS AND WEIGHT .........................................................................................................3
ISOTOPES AND ISOBARS ................................................................................................................4
ANALYTICAL TECHNIQUES FOR ELEMENTAL ANALYSIS ...............................................................5
ELEMENTAL ANALYSIS: FAAS.....................................................................................................6
ELEMENTAL ANALYSIS: GFAAS ..................................................................................................7
ELEMENTAL ANALYSIS: ICP-OES ................................................................................................8
ELEMENTAL ANALYSIS: ICP-MS..................................................................................................9
COMPARISON OF ELEMENTAL TECHNIQUES ................................................................................10
GRAPHICAL COMPARISON OF ELEMENTAL TECHNIQUES ............................................................11
COMPARISON OF THE COMPLEXITY OF MULTI-ELEMENTAL TECHNIQUES ...................................12
USERS/APPLICATIONS OF ICP-MS ..............................................................................................13
MULTI-ELEMENTAL ANALYSIS OF METALS ................................................................................14
INTRODUCTION: INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY .....15
WHAT IS ICP-MS? ......................................................................................................................16
ADVANTAGES OF ICP-MS ..........................................................................................................17
AGILENT TECHNOLOGIES AND ICP-MS ......................................................................................18
PROCESSES IN ICP-MS................................................................................................................19
OVERVIEW OF AGILENT 7500 FEATURES ....................................................................................20
SCHEMATIC DIAGRAM OF AGILENT 7500A .................................................................................21
SCHEMATIC DIAGRAM OF AGILENT 7500S ..................................................................................23
ISIS FOR APPLICATION FLEXIBILITY ...........................................................................................24
SAMPLE INTRODUCTION ..............................................................................................................25
AGILENT 7500 SAMPLE INTRODUCTION......................................................................................26
AUTOSAMPLERS ..........................................................................................................................27
TYPICAL NEBULIZER ...................................................................................................................28
SPECIALIZED SAMPLE INTRODUCTION SYSTEMS .........................................................................29
TYPICAL SPRAY CHAMBER – DOUBLE PASS ...............................................................................30
DROPLET DISTRIBUTION WITH AND WITHOUT SPRAY CHAMBER ...............................................31
NEW DESIGN AGILENT ICP TORCH BOX ....................................................................................32
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY ...........................................................33
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY (CONTINUED) .....................................34
WHY ARGON?.............................................................................................................................35
DISTRIBUTION OF IONS IN THE PLASMA ......................................................................................36
SAMPLE IONIZATION IN THE PLASMA ..........................................................................................37
FULL MASS CONTROL OF ALL GAS FLOWS .................................................................................38
INTERFACE ..................................................................................................................................39
AGILENT 7500 ION LENS SYSTEM...............................................................................................40
DISTRIBUTION OF IONS AND ELECTRONS AROUND THE INTERFACE ............................................41
ION ENERGY DISTRIBUTION IN THE INTERFACE ..........................................................................42
THE ELECTROSTATIC LENSES .....................................................................................................43
WHY “OFF-AXIS”?......................................................................................................................44
LOW TRANSMISSION PHOTON STOP SYSTEM ..............................................................................45
AGILENT HIGH TRANSMISSION OFF-AXIS SYSTEM .....................................................................46
ION FOCUSING – NEW OMEGA II LENS .......................................................................................47
FLAT RESPONSE CURVE – HIGH SENSITIVITY AT ALL MASSES...................................................48
AGILENT 7500 QUADRUPOLE .....................................................................................................49

iii
RESOLUTION AND ABUNDANCE SENSITIVITY..............................................................................50
NEW SIMULTANEOUS DUAL MODE DETECTOR & HIGH SPEED LOG AMPLIFIER – TRUE 9 ORDER
DYNAMIC RANGE ........................................................................................................................51
THE DETECTOR ...........................................................................................................................52
INTERFERENCES IN ICP-MS ..................................................................................................53
INTERFERENCES IN ICP-MS ........................................................................................................54
MASS SPECTROSCOPIC INTERFERENCES ......................................................................................55
ISOBARIC INTERFERENCES ..........................................................................................................56
POLYATOMIC INTERFERENCES ....................................................................................................57
MASS SPECTROSCOPIC INTERFERENCES ......................................................................................58
OPTIMIZING TO MINIMIZE INTERFERENCE FORMATION IN THE PLASMA [1]................................59
OPTIMIZING TO MINIMIZE INTERFERENCE FORMATION IN THE PLASMA [2]................................60
OPTIMIZING TO MINIMIZE INTERFERENCE FORMATION IN THE PLASMA [3]................................61
EFFECT OF PLASMA TEMPERATURE ON DEGREE OF IONIZATION .................................................62
EFFICIENT AEROSOL DECOMPOSITION ........................................................................................63
OXIDES AND DOUBLY CHARGED IONS ........................................................................................64
DEALING WITH MASS SPECTROSCOPIC INTERFERENCES .............................................................65
INTERFERENCE EQUATIONS ........................................................................................................66
AS INTERFERENCE CORRECTION .................................................................................................67
INTERFERENCE CORRECTION EQUATIONS - AGILENT 7500.........................................................68
NON-SPECTROSCOPIC INTERFERENCES .......................................................................................69
EFFECT OF HIGH DISSOLVED SOLIDS ..........................................................................................70
FIRST IONIZATION POTENTIAL ....................................................................................................71
IONIZATION EFFICIENCY .............................................................................................................72
SIGNAL SUPPRESSION .................................................................................................................73
MATRIX EFFECTS – ON LOW MASS ANALYTE ............................................................................74
MATRIX EFFECTS – ON MEDIUM MASS ANALYTE ......................................................................75
MATRIX EFFECTS – ON HIGH MASS ANALYTE ...........................................................................76
SPACE CHARGE INTERFACE AND LENS REGION ..........................................................................77
IONIZATION SUPPRESSION PLASMA REGION ...............................................................................78
WHAT CAN BE DONE ABOUT MATRIX EFFECTS .........................................................................79
TUNING THE AGILENT 7500...................................................................................................81
WHY TUNE THE ICP-MS?...........................................................................................................82
TUNING PROCEDURE OVERVIEW.................................................................................................83
AGILENT 7500 ICP-MS MANUAL TUNE CHECKLIST [1].............................................................84
AGILENT 7500 ICP-MS MANUAL TUNE CHECKLIST [2].............................................................85
AUTOTUNE SCREEN ....................................................................................................................86
AUTOTUNING OF ICP TORCH POSITION AND NEW TARGET TUNE ..............................................87
FEATURES OF AUTOTUNE (1) ......................................................................................................88
FEATURES OF AUTOTUNE (2) ......................................................................................................89
CHOOSING THE AUTOTUNE MODE ..............................................................................................90
BASICS OF THE SOFT EXTRACTION MODE ...................................................................................91
COMPARISON OF EXTRACTION MODES SETTINGS .......................................................................92
AUTOTUNE - TARGET SETTING ...................................................................................................93
TARGET SETTING - RANGE SETTING ...........................................................................................94
SENSITIVITY TUNING ..................................................................................................................95
PEAK SHAPE AND RESOLUTION ...................................................................................................96
ABUNDANCE SENSITIVITY ..........................................................................................................97
QUADRUPOLE MASS FILTER - SCAN LINE ...................................................................................98
DETECTION LIMITS IN NORMAL MODE .......................................................................................99
DETECTION LIMITS IN SOFT EXTRACTION MODE ......................................................................100
LOW BECS IN SOFT EXTRACTION MODE ..................................................................................101
PULSE/ANALOG (P/A) TUNING .................................................................................................102

iv
MAINTENANCE OF THE AGILENT 7500............................................................................103
MAINTENANCE SCHEDULE ........................................................................................................104
RUNNING TIME MAINTENANCE SCREEN ...................................................................................105
EARLY MAINTENANCE FEEDBACK (EMF) ................................................................................106
NORMAL MAINTENANCE OF THE SAMPLE INTRODUCTION SYSTEM ..........................................107
OVERNIGHT CLEANING OF THE SAMPLE INTRODUCTION SYSTEM.............................................108
SAMPLE INTRODUCTION MAINTENANCE ...................................................................................109
NEBULIZER CONNECTIONS ........................................................................................................110
MAINTENANCE OF A BABINGTON NEBULIZER ...........................................................................111
TORCH MAINTENANCE..............................................................................................................112
INTERFACE MAINTENANCE .......................................................................................................113
MAINTENANCE OF THE CONES ..................................................................................................114
EXTRACTION LENSES MAINTENANCE .......................................................................................115
EXTRACTION LENSES ................................................................................................................116
CLEANING OF THE EINZEL LENS AND OMEGA LENS ASSEMBLY ...............................................117
INSTRUMENT SHUTDOWN .........................................................................................................118
REMOVAL OF THE EINZEL LENS - OMEGA LENS ASSEMBLY .....................................................119
EXPANDED VIEW OF EINZEL LENS - OMEGA LENS ASSEMBLY .................................................120
PLATE BIAS LENS......................................................................................................................121
PENNING GAUGE .......................................................................................................................122
ROTARY PUMP MAINTENANCE .................................................................................................123
CHANGING ROTARY PUMP OIL .................................................................................................124
MAINTENANCE LOGBOOK SETTING ..........................................................................................125
MAINTENANCE LOGBOOK .........................................................................................................126
SAMPLE INTRODUCTION MAINTENANCE ...................................................................................127
AIR FILTERS MAINTENANCE .....................................................................................................128
INSTRUMENT START-UP ............................................................................................................129
INTERNAL STANDARDIZATION IN ICP-MS.....................................................................131
THE ROLE OF INTERNAL STANDARDS .......................................................................................132
HOW THE INTERNAL STANDARDS WORK - 1 .............................................................................133
HOW THE INTERNAL STANDARDS WORK - 2 .............................................................................134
CHOICE OF THE INTERNAL STANDARD ......................................................................................135
CONCENTRATION OF INTERNAL STANDARDS ............................................................................136
ON-LINE ADDITION OF INTERNAL STANDARDS .........................................................................137
SAMPLE PREPARATION TECHNIQUES FOR ICP-MS....................................................139
CONTAMINATION ......................................................................................................................140
TYPES OF CONTAMINATION ......................................................................................................141
CHALLENGES OF TRACE ANALYSIS ...........................................................................................142
WHEN A CONTAMINATION CAN OCCUR....................................................................................143
REAGENTS.................................................................................................................................144
WATER - MILLIPORE .................................................................................................................145
NITRIC ACID .............................................................................................................................146
SELECTED METHODS OF SAMPLE PREPARATION .......................................................................147
COMMONLY USED REAGENTS (1) .............................................................................................148
COMMONLY USED REAGENTS (2) .............................................................................................149
COMMONLY USED REAGENTS (3) .............................................................................................150
SEMI-QUANTITATIVE ANALYSIS OF SAMPLES.............................................................151
SEMI-QUANTITATIVE ANALYSIS................................................................................................152
WHAT IS SEMI-QUANTITATIVE ANALYSIS? ...............................................................................153
DATA ACQUISITION ..................................................................................................................154
METHOD SET-UP FOR SEMI-QUANTITATIVE ANALYSIS .............................................................155
PARAMETERS SELECTION - SPECTRUM ACQUISITION................................................................156

v
PARAMETERS SELECTION - SELECTION OF MASSES ..................................................................157
MORE ACQUISITION PARAMETERS ............................................................................................158
REPORT GENERATION ...............................................................................................................159
SEMI-QUANT PARAMETERS .......................................................................................................160
SEMI-QUANTITATIVE DATA ANALYSIS .....................................................................................161
EDITING PARAMETERS ..............................................................................................................162
DAILY UPDATE OF THE SEMI-QUANT PARAMETERS .................................................................163
INTERNAL STANDARD CORRECTION FOR OFF-LINE INTERNAL STANDARD ADDITION ..............164
INTERNAL STANDARD CORRECTION FOR ON-LINE INTERNAL STANDARD ADDITION................165
EXAMPLE OF SEMI-QUANT REPORT [1].....................................................................................166
EXAMPLE OF SEMI-QUANT REPORT [2].....................................................................................167
GENERATING A SEMI-QUANT REPORT .......................................................................................168
MANUAL VERIFICATION OF THE DATA .....................................................................................169
QUANTITATIVE ANALYSIS OF SAMPLES ........................................................................171
WHAT IS QUANTITATIVE ANALYSIS? ........................................................................................172
METHOD SET-UP FOR QUANTITATIVE ANALYSIS ......................................................................173
STEP ONE: EDITING THE AMU SELECT FILE ............................................................................174
EDITING A METHOD FOR QUANTITATIVE ANALYSIS .................................................................175
METHOD INFORMATION ............................................................................................................176
ACQUISITION MODES ................................................................................................................177
ACQUISITION PARAMETERS - MULTITUNE METHOD .................................................................179
PERIODIC TABLE .......................................................................................................................180
MASS TABLE .............................................................................................................................181
PERISTALTIC PUMP PROGRAM ..................................................................................................182
RAW DATA CORRECTIONS ........................................................................................................183
CONFIGURE REPORTS ................................................................................................................184
CALIBRATION............................................................................................................................185
CALIBRATION TABLE ................................................................................................................186
SAVE THE CALIBRATION AND THE METHOD .............................................................................187
QUANTITATIVE DATA ANALYSIS ..............................................................................................188
STANDARD DATA FILES ............................................................................................................189
CALIBRATION CURVES ..............................................................................................................190
EXAMPLES OF THE CALIBRATION CURVES FOR “EXCLUDED” ...................................................191
SIMPLE SEQUENCING (INTELLIGENT SEQUENCING DISABLED) ...........................193
SEQUENCING .............................................................................................................................194
ASX-500 VIAL POSITION NOMENCLATURE ..............................................................................195
SEQUENCING .............................................................................................................................196
SAMPLE LOG TABLE - SEQUENCE FLOW AND PERIODIC BLOCK ...............................................197
SAMPLE LOG TABLE .................................................................................................................198
SPECIAL FEATURES - KEYWORDS .............................................................................................199
RUNNING A SEQUENCE..............................................................................................................200
CHAINED SEQUENCE .................................................................................................................201
CHAINED SEQUENCE .................................................................................................................202
METHOD OF STANDARD ADDITIONS (MSA)...................................................................203
EXTERNAL CALIBRATION..........................................................................................................204
PROS AND CONS OF EXTERNAL CALIBRATION ..........................................................................205
METHOD OF STANDARD ADDITION (MSA) ...............................................................................206
PROS AND CONS OF METHOD OF STANDARD ADDITIONS ..........................................................207
DETERMINATION OF URANIUM IN URINE BY MSA....................................................................208
CONVERTING FROM MSA TO EXTERNAL CALIBRATION ...........................................................209
MATRIX-MATCHED URANIUM IN URINE EXTERNAL CALIBRATION ...........................................210
OFF-LINE DATA ANALYSIS AND SEQUENCE REPROCESSING .................................211

vi
OFF-LINE DATA ANALYSIS .......................................................................................................212
PROCEDURE FOR OFF-LINE DATA ANALYSIS ............................................................................213
OFF-LINE CALIBRATION REVIEW OF CURRENTLY RUNNING METHOD ......................................214
USING DOLIST FOR OFF-LINE DATA REPROCESSING.................................................................215
HOW TO USE DOLIST ................................................................................................................216
SELECTING FILES USING DOLIST ..............................................................................................217
SEQUENCE - REPROCESSING DATA BATCH ...............................................................................218
SEQUENCE REPROCESSING ........................................................................................................219
CUSTOM REPORTS AND DATABASES...............................................................................221
WHAT YOU WILL LEARN ..........................................................................................................222
CUSTOM REPORTS AND DATABASES .........................................................................................223
CREATING AND EDITING A REPORT TEMPLATE .........................................................................224
CUSTOM REPORTS - REPORT WIZARD.......................................................................................225
CUSTOM REPORTS - DRAG AND DROP (1) .................................................................................227
CUSTOM REPORTS - DRAG AND DROP (2) .................................................................................228
FORMATTING CUSTOM REPORTS...............................................................................................229
CUSTOM REPORTS - PRINTING SET-UP ......................................................................................230
CUSTOM REPORTS - SAVING THE TEMPLATE ............................................................................231
PRINTING CUSTOM REPORTS - INTERACTIVELY ........................................................................232
PRINTING CUSTOM REPORTS - PRINTING MULTIPLE FILES [1]..................................................233
PRINTING CUSTOM REPORTS - PRINTING MULTIPLE FILES [2]..................................................234
DATABASES ..............................................................................................................................235
DATABASE WIZARD ..................................................................................................................236
DATABASE - DRAG AND DROP ..................................................................................................237
DATABASE - FORMATTING ........................................................................................................238
DATABASE - CHARTS ................................................................................................................239
GLOBAL CHART OPTIONS .........................................................................................................240
DATABASE - SAVING.................................................................................................................242
UPDATING THE DATABASE - INTERACTIVELY ...........................................................................243
UPDATE THE DATABASE - MULTIPLE FILES [1].........................................................................244
UPDATE THE DATABASE - MULTIPLE FILES [2].........................................................................245
ISOTOPE RATIO MEASUREMENTS....................................................................................247
EDITING A METHOD FOR QUANTITATIVE ANALYSIS .................................................................248
ACQUISITION MODES ................................................................................................................249
ACQUISITION PARAMETERS FOR ISOTOPIC RATIO MEASUREMENTS..........................................251
REPORT SELECTION ..................................................................................................................252
SETTING PARAMETERS FOR ISOTOPIC RATIOS ...........................................................................253
EXAMPLE OF THE ISOTOPIC RATIO REPORT ..............................................................................254
AGILENT ICP-MS CHEMSTATION AND WINDOWS OVERVIEW ...............................255
THE WINDOWS INTERFACE .......................................................................................................256
WINDOWS MENUS.....................................................................................................................257
USEFUL WINDOWS TIPS ............................................................................................................258
MAINTAINING THE COMPUTER SYSTEM ....................................................................................259
WINDOWS NT EXPLORER - ENHANCED FILE MANAGEMENT ....................................................260
DIRECTORY STRUCTURE OF THE AGILENT CHEMSTATION ........................................................261
FILE NAMING ............................................................................................................................262
CHEMSTATION FILE EXTENSIONS .............................................................................................263
AN OVERVIEW OF ICP-MS ENVIRONMENTAL APPLICATIONS ...............................265
OPTIMIZING AGILENT 7500 FOR ENVIRONMENTAL SAMPLES ANALYSIS ..................................266
ENVIRONMENTAL TUNING ........................................................................................................267
THREE GOALS OF ENVIRONMENTAL TUNING ............................................................................269
TUNING FLOW CHART ...............................................................................................................271

vii
RECOMMENDATIONS ON INTERFERENCE EQUATIONS ...............................................................272
MORE INTERFERENCE CORRECTIONS ........................................................................................274
CALIBRATION STANDARDS .......................................................................................................276
LINEAR RANGE DETERMINATION..............................................................................................277
INTERFERENCE CHECK SAMPLES ..............................................................................................278
TROUBLESHOOTING ENVIRONMENTAL APPLICATIONS [1] ........................................................279
TROUBLESHOOTING ENVIRONMENTAL APPLICATIONS [2] ........................................................281
TROUBLESHOOTING ENVIRONMENTAL APPLICATIONS [3] ........................................................283
TROUBLESHOOTING ENVIRONMENTAL APPLICATIONS [4] ........................................................285
SEMICONDUCTOR APPLICATIONS OF ICP-MS AND ADVANTAGES OF AGILENT
7500S SYSTEM ...........................................................................................................................287
CHEMICALS AND MATERIALS USED IN SEMICONDUCTOR INDUSTRY ........................................288
METALS ANALYSIS IN THE SEMICONDUCTOR INDUSTRY - CUSTOMER GROUPS AND
REQUIREMENTS.........................................................................................................................289
SHIELDTORCH INTERFACE ........................................................................................................290
SHIELDTORCH INTERFACE ........................................................................................................291
NORMAL AND “COOL” PLASMAS ..............................................................................................292
SHIELD TORCH “COOL PLASMA” ..............................................................................................293
SHIELD TORCH INSTALLATION ..................................................................................................294
COOL PLASMA TUNING .............................................................................................................295
ADVANTAGES OF COOL PLASMA AT HIGHER POWER (900 - 1100 W) ......................................296
ADVANTAGES OF COOL PLASMA AT LOWER POWER (700-800 W) ...........................................297
DETECTION LIMITS STUDY [1] ..................................................................................................298
DETECTION LIMITS STUDY [2] ..................................................................................................299
AUTOMATIC SWITCHING BETWEEN NORMAL AND COOL PLASMA............................................300
INTELLIGENT SEQUENCE TRAINING TEXT...................................................................301
WHAT IS INTELLIGENT SEQUENCE? ..........................................................................................302
TYPICAL ANALYTICAL FLOW ....................................................................................................303
USING INTELLIGENT SEQUENCING ............................................................................................304
SETTING UP A QC CONFIGURATION..........................................................................................319
LABORATORY 1: AGILENT 7500 CONFIGURATION, STARTUP AND TUNING.......327
CONFIGURATION .......................................................................................................................328
STARTUP AND TUNING ..............................................................................................................329
AGILENT 7500 STARTUP CHECKLIST ........................................................................................330
SHUTDOWN CHECKLIST ............................................................................................................331
LABORATORY 2: AGILENT 7500 ROUTINE MAINTENANCE ......................................333
GENERAL ..................................................................................................................................334
SAMPLE INTRODUCTION ............................................................................................................335
INTERFACE ................................................................................................................................336
NEBULIZER, SPRAY CHAMBER AND TORCH...............................................................................337
RE-IGNITE THE PLASMA AND CHECK THE TUNE .........................................................................338
LABORATORY 3: SEMI-QUANTITATIVE ANALYSIS .....................................................339
SEMI-QUANTITATIVE ANALYSIS ...............................................................................................340
LABORATORY 4: QUANTITATIVE ANALYSIS OF UNKNOWN SAMPLE..................341
QUANTITATIVE ANALYSIS ........................................................................................................342
APPENDIX 1 – GENERAL INFORMATION.........................................................................343
PROFESSIONAL ORGANIZATIONS...............................................................................................344
JOURNALS .................................................................................................................................345
SELECTED WEB SITES (1)..........................................................................................................346

viii
SELECTED WEB SITES (2)..........................................................................................................347
APPENDIX 2 – FLOW CHATS ................................................................................................349
MANUAL TUNE TROUBLESHOOTING FLOWCHART [1]...............................................................350
MANUAL TUNE TROUBLESHOOTING FLOWCHART [2]...............................................................351
APPENDIX 3 – DEALING WITH POLYATOMICS .............................................................353
THE PROBLEM ...........................................................................................................................354
STRATEGY #1: (HIGH POWER) COOL PLASMA ANALYSIS .........................................................355
COMMERCIALIZATION OF COOL PLASMA ANALYSIS .................................................................356
SCHEMATIC OF AGILENT SHIELDTORCH ...................................................................................357
NOT ALL COOL PLASMAS* ARE THE SAME! [1] .......................................................................358
NOT ALL COOL PLASMAS* ARE THE SAME! [2] .......................................................................359
FE IN 31% H2O2 - 5 PPT SPIKE RECOVERY...............................................................................360
SHIELDTORCH TECHNOLOGY ELIMINATES INTERFERENCES BEFORE THEY FORM! ..................361
CAN HEAVY MATRICES BE ANALYZED? ...................................................................................362
CR IN UNDILUTED METHANOL ..................................................................................................363
EXAMPLE OF HEAVY MATRIX ANALYSIS ..................................................................................364
CALIBRATION FOR 56FE IN 1000 PPM PT ...................................................................................365
CALIBRATION FOR 66ZN IN 1000 PPM PT ..................................................................................366
DETERMINATION OF SE BY HIGH POWER COOL PLASMA ..........................................................367
SPECTRUM OF 10 PPB SE AND BLANK .......................................................................................368
CALIBRATION FOR 80SE ............................................................................................................369
DETECTION LIMITS FOR SE BY COOL PLASMA ..........................................................................370
CURRENT RESEARCH DEVELOPMENTS USING THE SHIELDTORCH ............................................371
AS CALIBRATION IN 10% HCL .................................................................................................372
LOW LEVEL P CALIBRATION.....................................................................................................373
LOW LEVEL S CALIBRATION.....................................................................................................374
LOW LEVEL SI CALIBRATION ....................................................................................................375
STRATEGY #2: RESOLVE THE INTERFERENCES ..........................................................................376
LIMITATIONS OF HR-ICP-MS ...................................................................................................377
RESOLUTION VS. SENSITIVITY ..................................................................................................378
OTHER FACTS ABOUT HR-ICP-MS [1].....................................................................................379
OTHER FACTS ABOUT HR-ICP-MS [2].....................................................................................380
OTHER FACTS ABOUT HR-ICP-MS [3].....................................................................................381
STRATEGY #3: DISSOCIATE INTERFERENCES WITHIN THE SPECTROMETER .............................382
PRINCIPLE OF COLLISION TECHNOLOGY ...................................................................................383
SELECTING A GAS PHASE REAGENT ..........................................................................................384
OPTIMIZING THE GAS PHASE REAGENT ....................................................................................385
SIDE REACTIONS ARE INEVITABLE!!.........................................................................................386
SIDE REACTIONS CREATE NEW INTERFERENCES ......................................................................387
HYDROCARBONS ARE PARTICULARLY PRONE TO COMPLEX CHEMISTRIES EVEN AT TRACE
LEVELS .....................................................................................................................................388
EFFECTS OF SAMPLE MATRIX ...................................................................................................389
STRATEGIES TO OVERCOME THE PROBLEM OF SIDE REACTIONS ..............................................390
LIMITATION OF SCANNING THE ANALYZER QUAD ....................................................................391
COLLISION CELLS CAN CREATE INTERFERENCES .....................................................................392
IN SUMMARY ............................................................................................................................393

ix
x
Introduction: Elemental Analysis
Introduction: Elemental Analysis
Atomic Spectrometry

Atomic Spectrometry

Atomic Spectrometry

Excited Light of specific characteristic wavelength


Atomic Absorption State
-
is absorbed by promoting an electron to a
higher energy level (excitation)
Light of specific wavelength
Light absorption is proportional to
from Hollow Cathode Lamp Ground -
State elemental concentration
(HCL)

High energy (light and heat) promotes an


Atomic Emission -
electron to a higher energy level
(excitation). Electron falls back and emits
Light and heat energy from high
light at characteristic wavelength
intensity source (flame or plasma) - - Light emission is proportional to
elemental concentration
-
High energy (light and heat) ejects
Mass Spectrometry electron from shell (ionization). Result is
Light and heat energy from high free electron and atom with positive
intensity source (plasma) charge (Ion)
- Ions are extracted and measured directly
in mass spectrometer

Figure 1

2
Introduction: Elemental Analysis
Atomic Mass and Weight

Atomic Mass and Weight

Atomic Mass and Weight

1 Proton 1 Neutron
Nucleus
Electron shell +
or cloud

Electron
- + + -
1840
Proton Electrons
Atomic number of an - - -
1 Proton - - - -
element is the number of - - - - -
- - - - - -
Neutron Protons in its nucleus - - - - -
+
- - - -
An atom has an equal number of Protons
(1 +ve charge) and electrons (1 -ve
charge) and so is electrically neutral.

Figure 2

3
Introduction: Elemental Analysis
Isotopes and Isobars

Isotopes and Isobars

Isotopes and Isobars

Isotopes:
Isotopes:Atomic
Atomicnumber
number(number
(numberof
of Isobars:
Isobars:Atomic
Atomicnumber
numberisisdifferent,
different,but
but
protons)
protons)isisthe
thesame,
same,but
butnumber
numberof of atomic
atomicweight
weightisisalmost
almostidentical
identicalso
sospecies
species
neutrons
neutronsisisdifferent
different(e.g.
(e.g.Pb204
Pb204&&PbPb208)
208) appear
appearatatsame
samemass
mass(e.g.
(e.g.Pb204
Pb204& &Hg204)
Hg204)
Chemical
Chemicalcharacteristics
characteristicsarearesame,
same,but
but Chemical
Chemicalcharacteristics
characteristicsare
aredifferent,
different,but
but
physical
physicalproperties
propertiesarearedifferent.
different. physical properties are similar
physical properties are similar. .

Isotopes
Isotopes Isobars
Isobars

Figure 3

4
Introduction: Elemental Analysis
Analytical Techniques for Elemental Analysis

Analytical Techniques for Elemental Analysis

Analytical Techniques for Elemental Analysis

FAAS - Flame Atomic Absorption Spectrometry


GFAAS - Graphite Furnace Atomic Absorption Spectrometry
ICP-OES - Inductively Coupled Plasma Optical Emission
Spectrometry = Inductively Coupled Plasma Atomic Emission
Spectrometry (ICP-AES)
ICP-MS - Inductively Coupled Plasma Mass Spectrometry

Figure 4

5
Introduction: Elemental Analysis
Elemental Analysis: FAAS

Elemental Analysis: FAAS

Elemental Analysis: FAAS

Advantages:
Inexpensive
Rapid for few selected elements
Limited use for organic solvents
Disadvantages
Poor sensitivity (high detection limits)
Single element determination at-the-time
Requires large amount of sample
Narrow linear range

Figure 5

6
Introduction: Elemental Analysis
Elemental Analysis: GFAAS

Elemental Analysis: GFAAS

Elemental Analysis: GFAAS

Advantages:
Relatively inexpensive
Requires small sample volume
Excellent sensitivity (low detection limits)
Disadvantages
Single element determination at-the-time
High operating costs (consumables)
Very narrow linear range
Cumbersome and time-consuming technique
Not suited for organic solvents
Requires matrix modifiers

Figure 6

7
Introduction: Elemental Analysis
Elemental Analysis: ICP-OES

Elemental Analysis: ICP-OES

Elemental Analysis: ICP-OES

Advantages:
Good general-purpose technique
Good dynamic range
Accommodates organic solvents
Multi-elemental technique
Disadvantages
Cost of the instrument
Limits of detection
Sample volume requirements
Spectral interferences for unknown/complicated matrices

Figure 7

8
Introduction: Elemental Analysis
Elemental Analysis: ICP-MS

Elemental Analysis: ICP-MS

Elemental Analysis: ICP-MS

Advantages:
Requires small amount of sample
Excellent dynamic range
Accommodates organic solvents
Multi-elemental technique
Isotope differentiation and determination
Scanning (semi-quant) capabilities
Superior limits of detection
Limited and well defined interferences
Disadvantages
Cost of the instrument

Figure 8

9
Introduction: Elemental Analysis
Comparison of Elemental Techniques

Comparison of Elemental Techniques

Comparison of Elemental Techniques

Sequential Simultaneous
Criteria1 GFAAS ICP-OES ICP-OES ICP-MS
Detection Limits ppt ppb ppb ppq-ppt
Linear Range 2-3 4-6 4-6 9*
Interferences Moderate Many Many Few
Speed Slow Slow Fast Fast
Elemental Coverage Poor Good Good Excellent
Multi-element No Yes Yes Yes
Simultaneous No No Yes Yes
Sample Size uL mL mL uL or mL
Capital Cost $ $ $$ $$
Operating Cost $$$ $$ $$ $

Figure 9

10
Introduction: Elemental Analysis
Graphical Comparison of Elemental Techniques

Graphical Comparison of Elemental Techniques

Graphical Comparison of Elemental Techniques

high
ICP-MS
GFAA
low sensitivity

ICP OES
ICP-OES
Flame AA

low number of analyses high

Figure 10

11
Introduction: Elemental Analysis
Comparison of the Complexity of Multi-elemental Techniques

Comparison of the Complexity of Multi-elemental


Techniques

Comparison of the Complexity of Multielemental


Techniques

# emission lines # (natural) isotopes


alkali metals
lithium 30 2
cesium 645 1
alkali earths
magnesium 173 3
calcium 662 6
transition metals
chromium 2277 4
iron 4757 4
cerium 5755 4

Figure 11

12
Introduction: Elemental Analysis
Users/Applications of ICP-MS

Users/Applications of ICP-MS

Users/Applications of ICP-MS

• Environmental
• Semiconductor
• Nuclear
• Clinical/Pharmaceutical
• Petrochemical
• Geological
• Forensic
• Academia

Figure 12

13
Introduction: Elemental Analysis
Multi-elemental Analysis of Metals

Multi-elemental Analysis of Metals

Multi-elemental Analysis of Metals

GFAAS, ICP-OES & CVAA ICP-OES


Analysis
Sample
Digestion
Data
GFAAS Data
Sample Compilation
Reporting Archival
Analysis & Review
Logging

Hg
Preparation CVAAS
Analysis

ICP-MS

Sample Sample ICP-MS Data Review


Archival
Logging Digestion Analysis & Reporting

Figure 13

14
Introduction: Inductively Coupled
Plasma Mass Spectrometry
Introduction: Inductively Coupled Plasma Mass Spectrometry
What is ICP-MS?

What is ICP-MS?

What is ICP-MS?

● Inorganic (elemental) analysis technique.

● ICP - Inductively Coupled Plasma


high temperature ion source
● MS - Mass Spectrometer
– quadrupole scanning spectrometer
– mass range from 7 to 250 amu (Li to U...)
– separates all elements in rapid sequential scan
– ions measured using dual mode detector
➨ ppt to ppm levels
➨ isotopic information available

Figure 14

16
Introduction: Inductively Coupled Plasma Mass Spectrometry
Advantages of ICP-MS

Advantages of ICP-MS

Advantages of ICP-MS

• Trace and ultratrace measurement of >70 elements - from Li to U


– Agilent 7500 can measure from <1ppt to >500ppm (9 orders linear range)
• Spectral simplicity
– Every element (except In) has an isotope which is free from direct overlap
• Speed of multi-element analysis
– Typical multi-element acquisition in 1-2 min (~4 min including rinse)
• Flexibility to optimize for specific applications
– Automated set-up and autotuning give improved ease of use
• Fast semi-quantitative analysis - accurate data without calibration
– measurement is based on comparison of relative isotope sensitivity
• Isotope ratio measurements
– nuclear, geological, environmental and nutrition studies

Figure 15

17
Introduction: Inductively Coupled Plasma Mass Spectrometry
Agilent Technologies and ICP-MS

Agilent Technologies and ICP-MS

Agilent Technologies and ICP-MS

1987 - PMS 100 - first computer controlled ICP-MS


1988 - PMS 200 - 2nd generation ICP-MS
1990 - PMS 2000 - featuring Omega lens system - lowest random
background in ICP-QMS
1992 - ShieldTorch interface developed - interferences
fundamentally reduced for the first time in ICP-QMS -
enables analysis of K, Ca, Fe by ICP-QMS
1994 - HP 4500 introduced - World's first benchtop system
1998 - Over 500 systems installed
1999 - HP 4500 Series 100, 200, 300 introduced
2000 - Agilent 7500 series. 7500a, 7500i and 7500s - the next
generation in ICP-MS instrumentation

Figure 16

18
Introduction: Inductively Coupled Plasma Mass Spectrometry
Processes in ICP-MS

Processes in ICP-MS

Processes in ICP-MS

Nebulization Desolvation Vaporization Atomization Ionization

Aerosol
Absorption process
liquid sample Desolvation
Particle Atomization
Nebulization
Molecule Atom Ion
Ionization
Vaporization
solid sample
Emission process
Mass analyzer

Figure 17

19
Introduction: Inductively Coupled Plasma Mass Spectrometry
Overview of Agilent 7500 Features

Overview of Agilent 7500 Features

Overview of Agilent 7500 Features


Exclusive ShieldTorch Omega II off axis lens
technology for unrivaled system providing superior
True hyperbolic profile
sensitivity and ion transmission and
solid molybdenum
elimination of Ar -based exceptionally low
Open architecture quadrupole operating at
interferences backgrounds
sample introduction 3.0MHz, providing
system for ease of excellent peak shape and
access and connectivity abundance sensitivty

Peltier cooled spray Simultaneous dual mode


chamber for superior detector with high speed
stability, low oxide amplifier providing 9
interferences and orders of dynamic range
analysis of organics
4 (or 5) mass flow controllers for
Low pulsation peristaltic improved signal stability and
pump located close to the analysis of organics
nebulizer for rapid sample Computer controlled torch
introduction & washout positioning in 3 planes with Maintenance-free solid state RF
Autotune for effortless, generator. 27.12MHz frequency
Durable stainless steel consistent torch alignment generates highest temperature plasma
chassis. Benchtop design after maintenance for reduced matrix effects
minimizes the need for
laboratory space

Figure 18

20
Introduction: Inductively Coupled Plasma Mass Spectrometry
Schematic Diagram of Agilent 7500a

Schematic Diagram of Agilent 7500a

Schematic Diagram of Agilent 7500a

Robust Omega II Hyperbolic Rod Fast


Interface Off Axis Lens Quadrupole Simultaneous
Dual Mode
Detector

Peltier
Cooled
Spray
Chamber

Peri
Pump

Sample Turbo Turbo


pump pump

Novel High
Capacity Vacuum
System Design

Figure 19

• Sample solution is pumped into the nebulizer. The sample stream is


nebulized with argon gas and forms an aerosol of fine droplets.
• The argon gas carries the finest droplets through the turns of the spray
chamber and into the plasma where the sample is atomized and ionized.
• Ions are extracted from the atmospheric pressure plasma into the high
vacuum region of the mass analyzer via the interface. The interface
consists of two water-cooled orifices called cones.
• A three-stage vacuum system provides pressures of 1 Torr between the
cones, 10-4 Torr in the lens chamber and 10-6 Torr in the analyzer
chamber.
• The ion lens system focuses ions into the analyzer. Light is excluded from
the analyzer and detector regions by the Omega lens, which reduces
background noise.
• The quadrupole mass filter allows only ions of a specific mass to charge
ratio to pass through to the detector at any point in time.

21
Introduction: Inductively Coupled Plasma Mass Spectrometry
Schematic Diagram of Agilent 7500a

• The EM detector measures the ion signal at each mass and stores it in the
MCA. Data is expressed as counts per second, which is directly
proportional to the concentration of the element at that mass.

22
Introduction: Inductively Coupled Plasma Mass Spectrometry
Schematic Diagram of Agilent 7500s

Schematic Diagram of Agilent 7500s

Schematic Diagram of Agilent 7500s

Robust Omega II
Interface Off Axis Lens Hyperbolic Rod
Fast
Quadrupole
Simultaneous
Dual Mode
Detector
Peltier
Cooled
Spray
Chamber

Peri
Pump
Sample Turbo
Valves Turbo
pump pump

Novel High Capacity


Vacuum System Design

Figure 20

23
Introduction: Inductively Coupled Plasma Mass Spectrometry
ISIS for Application Flexibility

ISIS for Application Flexibility

ISIS for Application Flexibility

Figure 21

24
Introduction: Inductively Coupled Plasma Mass Spectrometry
Sample Introduction

Sample Introduction

Sample Introduction

Plasma Gas ICP Torch


Auxiliary Gas Plasma

Peltier Cooled
Spray Chamber
RF Coil
Blend Gas
Carrier Gas Nebulizer

Peristaltic
Pumps
Internal Standard/
Diluent
Sample

Figure 22

The ease of removal of our torch is a big point:


• 1 minute with Agilent
• 5 minutes with VG
• 10 – 15 minutes with PE
Especially with gloved hands, as in a cleanroom.
We are the only company to offer Pt injector torches. This is in response to
demand from Japanese semiconductor users. All other vendors use Al2O3 or
sapphire, which give high Al background.
Also, we are the only ones to use a polypropylene spray chamber:
• VG use Teflon (poor wetting - bad stability and washout)
• PE use Ryton, which is impure - high Ba, etc. from filler, and also it is not
resistant to H2O4

25
Introduction: Inductively Coupled Plasma Mass Spectrometry
Agilent 7500 Sample Introduction

Agilent 7500 Sample Introduction

Agilent 7500 Sample Introduction

Externally mounted spray


chamber with new Peltier
cooling system

New, low-pulsation 3-channel


sample introduction pump -
close-coupled to spray chamber
to reduce uptake time and dead
volume

Open sample area protected with


sealed polymer tray - easy access
to sample intro components and


connection of external devices -


laser ablation


LC


GC
CE

Figure 23

26
Introduction: Inductively Coupled Plasma Mass Spectrometry
Autosamplers

Autosamplers

Autosamplers

ASX -100

ASX -500

Figure 24

27
Introduction: Inductively Coupled Plasma Mass Spectrometry
Typical Nebulizer

Typical Nebulizer

Typical Nebulizer

Concentric
ConcentricNebulizer
Nebulizer
Sample in Cross-flow
Cross-flowNebulizer
Nebulizer

Fine capillary -prone A rgon in


Argon in to blockages

Sample
High
Highsolids
solidsNebulizer
Nebulizer out
(Babbington
(Babbingtondesign)
design) Pt/R h
Sample in capillary
Sam ple in
Argon in
Ar gas outlet

Figure 25

28
Introduction: Inductively Coupled Plasma Mass Spectrometry
Specialized Sample Introduction Systems

Specialized Sample Introduction Systems

Specialized Sample Introduction Systems

Organic analysis kit including


exclusive oxygen inlet
connector for safe addition of
oxygen for organics analysis

Inert sample kit with unique


polypropylene spray chamber

Exclusive Agilent Micro Flow


Nebulizer for trouble-free analysis
of microvolume samples

Widest range of ICP torches


including exclusive platinum
injector torch for HF
and unique photoresist torch for
photoresist matrices

Figure 26

29
Introduction: Inductively Coupled Plasma Mass Spectrometry
Typical Spray Chamber – Double Pass

Typical Spray Chamber – Double Pass

Typical Spray Chamber - Double Pass


Scott-Type

Small Droplets to ICP


Nebulizer
Sample solution (High solids type)
Aerosol

Ar carrier gas

Sample drain
Large Droplets
to Waste

Figure 27

30
Introduction: Inductively Coupled Plasma Mass Spectrometry
Droplet Distribution With and Without Spray Chamber

Droplet Distribution With and Without Spray


Chamber

Droplet Distribution With and Without Spray


Chamber

(%)
No Spray Chamber With Spray Chamber
(%)
50 30

40 25

20
30
15
20
10
10
5

0 8 14 20 26 32 38 44 50 56 62 68 74 80
0
2 3 4 5 6 7 8 9 10 11
Particle Size ( um ) Particle Size ( um)

Figure 28

31
Introduction: Inductively Coupled Plasma Mass Spectrometry
New Design Agilent ICP Torch Box

New Design Agilent ICP Torch Box

New Design Agilent ICP Torch Box

New torchbox position control


stepper motors (x-, y- and z-
adjustment) are fast and precise.

Quick release torch mounting allows


for easy torch removal and
replacement for cleaning.

Plasma compartment is separated


from the main cabinet, and plasma
gases vented separately direct to the
exhaust duct.

Figure 29

32
Introduction: Inductively Coupled Plasma Mass Spectrometry
Inductively Coupled Plasma Mass Spectrometry

Inductively Coupled Plasma Mass Spectrometry

Inductively Coupled Plasma Mass Spectrometry

RF Load Coil
Radio Frequency voltage induces
Quartz "torch" made
rapid oscillation of Ar ions and
of concentric tubes electrons -> HEAT (~10,000 K)

Auxiliary Gas

Carrier or
Injector or
Nebulizer Gas

Sample aerosol is carried


through center of plasma
-> dried, dissociated,
Coolant or atomized, ionized
Plasma Gas ~6500 K.

Figure 30

33
Introduction: Inductively Coupled Plasma Mass Spectrometry
Inductively Coupled Plasma Mass Spectrometry (continued)

Inductively Coupled Plasma Mass Spectrometry


(continued)

Inductively Coupled Plasma Mass Spectrometry

● Plasma is electrical discharge, not chemical flame


Î Ar gas used
Î plasma at atmospheric pressure -> very high temperature
Î(a low pressure plasma is a fluorescent lamp)
Î plasma is generated through inductive coupling of free electrons with
rapidly oscillating magnetic field (27 MHz)
Î Energy is transferred collisionally to argon molecules
Î plasma is contained in gas flow in a quartz tube (torch)
Î sample aerosol is carried through the center of the plasma
Î proximity to 10,000 C plasma causes dissociation, atomization and
ionization
Î ions are extracted into the spectrometer

Figure 31

34
Introduction: Inductively Coupled Plasma Mass Spectrometry
Why Argon?

Why Argon?

Why Argon?

● Ar is inert
● Ar is relatively inexpensive!
● Ar is easily obtained at very high purity

Most importantly -
● Ar has a 1st ionization potential of 15.75 electron volts (eV)
– higher than the 1st ionization potential of most other elements (except
He, F, Ne) and
– lower than the 2nd ionization potential of most other elements (except
Ca, Sr, Ba,etc)
● Since the plasma ionization environment is defined by the
Ar, most analyte elements are efficiently singly charged

Figure 32

35
Introduction: Inductively Coupled Plasma Mass Spectrometry
Distribution of Ions in the Plasma

Distribution of Ions in the Plasma

Distribution of Ions in the Plasma

(%)

100
Ar

80 Co Distance from the work coil


Relative Ion intensity

mm 5 10 15 20 25 30
60 9
6
3
0
40 3
6
9
mm Cool Plasma
20
Sampling Depth

0 Load coil Normal Plasma


-8 -6 -4 -2 0 2 4 6 8 Sampling Depth
Distance from the center (mm)

Figure 33

36
Introduction: Inductively Coupled Plasma Mass Spectrometry
Sample Ionization in the Plasma

Sample Ionization in the Plasma

Sample Ionization in the Plasma

Hottest part of Residence time is a


plasma ~ 8000K few milliseconds

By sample cone, analytes


present as M+ ions

+ Highest M+ population
should correspond to lowest
polyatomic population

Atoms are formed


Aerosol is Dried and then ionized
Particles are decomposed
and dissociated

Figure 34

37
Introduction: Inductively Coupled Plasma Mass Spectrometry
Full Mass Control of All Gas Flows

Full Mass Control of All Gas Flows

Full Mass Control of All Gas Flows

Nebulizer gas flow is an important parameter


to tune for optimizing signal - separate
control of nebulizer gas and total injector
flow (by varying make-up gas) is essential for
optimum performance
Mass flow control (MFC) has the benefits of
superior stability - better short and long
term signal precision
more reproducible set-up and
optimization
electronic control via the PC

• 4500 Series - 2 MFCs - nebulizer and blend (make-up)


– blend gas is required for optimum ShieldTorch analysis, or for organics analysis
• 7500a, 7500i - 4 MFCs - plasma, auxiliary, nebulizer, blend
• 7500s - 5 MFCs - plasma, auxiliary, nebulizer, blend, option

Figure 35

38
Introduction: Inductively Coupled Plasma Mass Spectrometry
Interface

Interface

Interface
■ Sampling cone
■ Skimmer cone
Allows introduction of ions into the vacuum chamber
Material : Nickel
Platinum
Interface
1.0 E-02 torr

Mass Spectrometer
Plasma
1.0 E-05 torr
1 torr

Sampler Cone
1 mm orifice
To pumps
Skimmer Cone
0.4 mm orifice

Figure 36

39
Introduction: Inductively Coupled Plasma Mass Spectrometry
Agilent 7500 Ion Lens System

Agilent 7500 Ion Lens System

Agilent 7500 Ion Lens System

Serves to focus ions coming from the skimmer into the mass filter. Rejects
neutral atoms and minimizes the passage of any photons from ICP.
■ Extraction - Extract and accelerate ions from the plasma
■ Einzel - Collimate and focus ion beam

■ Omega - Bend ion beam to eliminate photons and neutrals


■ QP focus - Refocus ion beam
10 -2 Torr 10 -5 Torr

QP- Focus
(+) (-)

(-) (+)

Figure 37

40
Introduction: Inductively Coupled Plasma Mass Spectrometry
Distribution of Ions and Electrons Around the Interface

Distribution of Ions and Electrons Around the


Interface

Distribution of Ions and Electrons Around the


Interface

Neutral Plasma
equal numbers of electrons and Sheath

positive ions at high temp Ar


Ar
Ar Ar+ + +
Cooler Interface e Ar
Ar e Ar Ar
does not support ion stability e e + e Ar+
Ar Ar Ar
neutral Ar sheath forms acting as a
condensor preventing the plasma
from grounding on the cones

Figure 38

41
Introduction: Inductively Coupled Plasma Mass Spectrometry
Ion Energy Distribution in the Interface

Ion Energy Distribution in the Interface

Ion Energy Distribution in the Interface

• Ion lenses are optimized for a Ion lens setting


particular range of ion energies
(potential + kinetic). Low mass Y
ions have lower kinetic energy. Li Tl
• Cooling the plasma increases the
thickness of the sheath,

Sensitivity
increasing the plasma potential
and the energy of the ions.
– Shifts the energy distribution
profile to the right -
Low High
increasing low mass
sensitivity. Ion energy

Figure 39

42
Introduction: Inductively Coupled Plasma Mass Spectrometry
The Electrostatic Lenses

The Electrostatic Lenses

The Electrostatic Lenses

● Ions, photons and neutrals all enter the spectrometer through the
interface
➨ the detector is sensitive to photons/neutrals, as well as ions
● Ions are charged particles
➨ can be deflected using electric fields
● Photons travel in straight lines
● If ions can be deflected off-axis, they will be separated from non-
charged species (photons/neutrals)
➨ must ensure that mass bias is not introduced when ions are deflected

Figure 40

43
Introduction: Inductively Coupled Plasma Mass Spectrometry
Why “Off-Axis”?

Why “Off-Axis”?

Why “Off-Axis”?

● Detector must be screened from Plasma


– Plasma is an intense source of photons and neutrals
– Electron Multiplier is photon/neutral sensitive

● Common approach is to place a metal disc in the light path


– "Photon Stop"
– "Shadow Stop"

● BUT -With the "Photon Stop" or "Shadow Stop" ions must be


defocused around the disc and then re-focused on the other side
– This is very inefficient and will introduce mass bias

Figure 41

44
Introduction: Inductively Coupled Plasma Mass Spectrometry
Low Transmission Photon Stop System

Low Transmission Photon Stop System

Low Transmission Photon Stop System

Higher sample uptake rate (0.7-1.0mL/min), Ions re-focused after


small injector tube (2mm), lower temp plasma photon stop,
(40MHz), fixed sample depth & large cone causing mass bias - loss
orifices (1.1/0.9mm) - inefficient matrix of low mass ions
deposition

Ions must be defocused around photon stop Simple ion lens - inefficient
- loss of ion transmission. Matrix deposition focusing - must use voltage scan on
on photon stop and lens, causing drift lens to reduce loss of low mass ions

Low transmission - higher sample uptake, large interface orifices and small
torch injector must be used to compensate.
Higher matrix loading on the system - more frequent ion lens cleaning, and
faster degradation of interface rotary pump oil

Figure 42

45
Introduction: Inductively Coupled Plasma Mass Spectrometry
Agilent High Transmission Off-Axis System

Agilent High Transmission Off-Axis System

Agilent High Transmission Off-Axis System

Lower sample uptake rate (0.3mL/min), Compound ion lens -


larger injector tube (2.5mm), higher temp efficient focusing, high
plasma (27MHz), variable sample depth & transmission across the
small cone orifices (1.0/0.4mm) - efficient mass range
matrix deposition

Dual extraction lenses prevent loss of


low mass ion on exit from interface. Photons and neutrals removed -
Also serve to protect main ion lenses ions are deflected off axis into
by trapping sample matrix. quadrupole with minimal mass bias
High transmission - sensitivity maintained with less sample loading on system -
lower sample uptake, small interface orifices and larger diameter torch injector.
Results in much less frequent ion lens cleaning and extended interface rotary pump
oil lifetime.

Figure 43

46
Introduction: Inductively Coupled Plasma Mass Spectrometry
Ion Focusing – New Omega II Lens

Ion Focusing – New Omega II Lens

Ion Focusing - New Omega II lens

Integrated 1 piece design


for easy cleaning (when
required)

No wires to attach,
makes replacement fast
Ions enter and easy
here
Gives very high
sensitivity and low
The second part is background performance
First 3 lenses are called the Omega II. This
an “Einzel” lens. These is where the ions
focus the ions are sent “off axis”

Figure 44

47
Introduction: Inductively Coupled Plasma Mass Spectrometry
Flat Response Curve – High Sensitivity at All Masses

Flat Response Curve – High Sensitivity at All


Masses

Flat Response Curve - High Sensitivity at All


Masses

70

60

50

40
Mcps/ppm

30

20

10

0
0 50 100 150 200 250

Mass

Figure 45

Photon stop systems suffer from significant mass bias against low masses due to
space charge effects.

48
Introduction: Inductively Coupled Plasma Mass Spectrometry
Agilent 7500 Quadrupole

Agilent 7500 Quadrupole

Agilent 7500 Quadrupole

TRUE hyperbolic rods - precision ground


from solid Molybdenum.
Novel digitally synthesized 3.0MHz RF
generator - produce excellent
transmission, peak shape and abundance
sensitivity
[1] Spectrum No.1 [ 70.409 sec]:YCS8.D# [CPS] [Log]
1.0E7

1.0E6

1.0E5

1.0E4 Very low


contributions
Log scale plot of 1ppm Y solution 1000 to adjacent
showing excellent peak shape and masses
100

abundance sensitivity
- note no tailing at low or high mass 10

m/z-> 82 84 86 88 90 92 94 96 98 100

Figure 46

49
Introduction: Inductively Coupled Plasma Mass Spectrometry
Resolution and Abundance Sensitivity

Resolution and Abundance Sensitivity

Resolution and Abundance Sensitivity

Resolution Abundance Sensitivity


Abundance
Peak Sensitivity is
Heigh ratio of peak
t height M to 10% Peak
M-1 & M+1 Height
Peak Width (amu)
at 50% Peak Height
(typically 0.5 - 0.6) Poor Abundance
Sensitivity. Good Abundance
Peak tails into Sensitivity.
Peak Width (amu) neighboring peaks No contribution
at 10% Peak Height to neighboring
typically 0.65 - 0.75) peaks

M-1 M M+1 M-1 M

Figure 47

50
Introduction: Inductively Coupled Plasma Mass Spectrometry
NEW Simultaneous Dual Mode Detector & High Speed Log Amplifier
– True 9 Order Dynamic Range

NEW Simultaneous Dual Mode Detector & High


Speed Log Amplifier – True 9 Order Dynamic
Range

NEW Simultaneous Dual Mode Detector & High


Speed Log Amplifier - True 9 Order Dynamic Range

New true simultaneous detector - with extended 9


order dynamic range - largest in ICP-MS!
Agilent’s unique new detection circuit means
acquisition speed is not compromised when
analyzing in analog mode
Pulse counting mode - min dwell time - 100usec
Analog mode - min dwell time - 100usec!
Transient signals such as those from a laser
ablation pulse or chromatography can be measured
over a wide dynamic range

Figure 48

51
Introduction: Inductively Coupled Plasma Mass Spectrometry
The Detector

The Detector

The Detector

● Electron multiplier
➨ discrete dynode detector (ETP)

■ each dynode gives "cascade" of electrons

■ -> signal is multiplied

Amp
M+

M+
Ion Electrons
-
e -
e

Dynode

Figure 49

52
Interferences in ICP-MS
Interferences in ICP-MS
Interferences in ICP-MS

Interferences in ICP-
MS

Interferences in ICP-MS

y Mass Spectroscopic Interferences


y Inability to resolve same nominal masses

y Non-spectroscopic Interferences
y Result from sample matrix

Figure 50

54
Interferences in ICP-MS
Mass Spectroscopic Interferences

Mass Spectroscopic
Interferences

Mass Spectroscopic Interferences

z Isobaric
z Polyatomic
z Argides
z Oxides
z Other (i.e. Chlorides, Hydrides, etc.)
z Doubly-charged

Figure 51

55
Interferences in ICP-MS
Isobaric Interferences

Isobaric
Interferences

Isobaric Interferences

Isotopes AMU % Abundance


V 50 0.25
Ti 50 5.4
Cr 50 4.35

Zr 96 2.8
Ru 96 16.68
Mo 96 5.52

Ba 138 71.7
La 138 0.09
Ce 138 0.25

Figure 52

56
Interferences in ICP-MS
Polyatomic Interferences

Polyatomic
Interferences

Polyatomic Interferences

Interferent m/z Overlaps with


+
N2 28 Si
+
NO 30 Si
+
O2 32 S
34 S
+
Ar 40 Ca
+
Ar0 56 Fe
+
Ar2 80 Se
78 Se
76 Se

Figure 53

57
Interferences in ICP-MS
Mass Spectroscopic Interferences

Mass Spectroscopic
Interferences

Mass Spectroscopic Interferences

y Choose an isotope free of interferences


¾ 137Ba instead of 138 Ba
y Optimize instrument to minimize interference
¾ Oxides, Doubly-charged ions
y ShieldTorch
¾ Reduces polyatomic ions with high ionization potential
¾ Removes ArO
¾ Removes ArH

Figure 54

58
Interferences in ICP-MS
Optimizing to Minimize Interference Formation in the Plasma [1]

Optimizing to Minimize Interference Formation in


the Plasma
[1]

Optimizing to Minimize Interference Formation


in the Plasma [1]

Minimize ‘matrix’ loading


low sample uptake rate
reduce water loading
cooled spray chamber
desolvation
Maximize residence time in plasma
maximum sampling depth
large diameter torch injector for lower aerosol velocity

Figure 55

59
Interferences in ICP-MS
Optimizing to Minimize Interference Formation in the Plasma [2]

Optimizing to Minimize Interference Formation in


the Plasma
[2]

Optimizing to Minimize Interference Formation


in the Plasma [2]

Maximize available energy for ionization


high forward power
reduce sample and carrier flow
eliminate/reduce matrix easily ionizable elements where practical
dilute if necessary

Figure 56

60
Interferences in ICP-MS
Optimizing to Minimize Interference Formation in the Plasma [3]

Optimizing to Minimize Interference Formation in


the Plasma
[3]

Optimizing to Minimize Interference Formation


in the Plasma [3]

Residence time is on the order of milliseconds. It is essential to


optimize plasma energy input ensure sample matrix breakdown!

By sample cone, analytes


present as M+ ions

+
Atoms are formed
and then ionized
Aerosol is Dried Particles are decomposed
and dissociated

Figure 57

61
Interferences in ICP-MS
Effect of Plasma Temperature on Degree of Ionization

Effect of Plasma Temperature on Degree of


Ionization

Effect of Plasma Temperature on Degree of


Ionization

100%
90%
degree of ioniztaion

80%
70% 5000 K
60%
6000 K
50%
40% 7000 K
30% 8000 K
20%
10%
0%
0 5 10 15
Ionization potential

Figure 58

62
Interferences in ICP-MS
Efficient Aerosol Decomposition

Efficient Aerosol
Decomposition

Efficient Aerosol Decomposition

Wide bore torch injector results in a diffuse aerosol to minimize


localized cooling as the aerosol droplets are dried and minimizes
potential sample deposition on the inner surface

Figure 59

63
Interferences in ICP-MS
Oxides and Doubly Charged Ions

Oxides and Doubly Charged


Ions

Oxides and Doubly Charged Ions

4 Ce2+
Ratio (%)

Ba2+
3
CeO
2 BaO

0
-5 0 5 10 15 20
Temp. (degreeC)

Figure 60

64
Interferences in ICP-MS
Dealing with Mass Spectroscopic Interferences

Dealing with Mass Spectroscopic


Interferences

Dealing with Mass Spectroscopic Interferences

y Matrix Elimination
y Chelation
y Chromatography
y ETV
y Desolvation
y membrane
y thermal
y Interference correction equations

Figure 61

65
Interferences in ICP-MS
Interference Equations

Interference
Equations

Interference Equations

Mathematical equations used to minimize the effect of elemental,


doubly-charged and polyatomic isobaric interferences in ICP-MS analysis.

Isobaric
204 Hg on 204 Pb

Polyatomic
75ArCl on 75As

Doubly charged ions


88Sr++ on 44 Ca

Figure 62

66
Interferences in ICP-MS
As Interference Correction

As Interference
Correction

As interference Correction

75As = 75M - { 77ArCl (35Cl abundance / 37Cl abundance) }


75As = 75M - { 77ArCl ( 3.127 ) } (1)
But there is Se at m/z 77...
77ArCl = 77M - { 82Se ( 77Se abundance / 82Se abundance) }
77ArCl = 77M - { 82Se ( 0.874 ) } (2)
1.2
Relative signal

1
0.8 Kr So equation 1 and 2 becomes:
0.6 Se 75As = 75M - {[77M - { 82Se (0.874)}] (3.127)}
0.4 ArCl
0.2 As 75 75 77 82
0 As = M - M(3.127) + Se(2.733) (3)
75

76

77

78

79

80

81

82

83

m/z But, there is Krypton at 82…


82Se = 82M - { 83Kr ( 82Kr abundance / abundance 83Kr ) }
82Se = 82M - { 83Kr ( 1.009 ) } (4)
So equation 3 and 4 becomes:
75 As = 75M - 77M(3.127) + {[82M - {83Kr (1.009)}](2.733)}
75As = 75M - 77M(3.127) + 82M(2.733) - 83M(2.757)

Figure 63

67
Interferences in ICP-MS
Interference Correction Equations - Agilent 7500

Interference Correction Equations - Agilent


7500

Interference Correction Equations - Agilent 7500

Figure 64

Interference equations are edited from Top >> Methods >> Edit Interference
Equation… or from Edit Entire Method.
Equations must be simplified and terms combined before entering them into the
Edit Interference Equation dialog box.
The actual values are stored within the method folder as ‘correct.icp’, a text file
which can be directly edited if desired.

68
Interferences in ICP-MS
Non-Spectroscopic Interferences

Non-Spectroscopic
Interferences

Non-Spectroscopic Interferences

■ Total Dissolved Solids


■ High Mass Elements
y High mass elements affect the signal of low mass
element. (Space Charge)
■ Easily Ionized Elements
y Limited ionization energy is consumed by easily ionized
elements such as Na and K.

Result from sample matrix

Figure 65

69
Interferences in ICP-MS
Effect of High Dissolved Solids

Effect of High Dissolved


Solids

Effect of High Dissolved Solids

y Signal suppression

y Deposits on sampler and skimmer cones

y Deposits on ion optics

Figure 66

70
Interferences in ICP-MS
First Ionization Potential

First Ionization
Potential

First Ionization Potential

First ionization potential


(eV) He 24.58eV
He
25
Ne

20
Ionisation potential

F Ar 15.75eV
Ar
15 N Kr
Cl Xe
O Br Au
C Rn
P I Ir Hg
Be Fe Zn As Ru Cd
10 Mg S Mn Co Se Mo Pd Sb Os Po
Te Ta
B Si Ca
Ti Ge Y
Zr Lanthanides Pt Pb Ac
Cu Ag Sn W
Cr Ni Nb Rh Hf Bi
Al Ga Re
5 Li Sc Sr Tc In Ba
Tl
Ra
Na K V Rb Cs

0
0 10 20 30 40 50 60 70 80 90
Atomic number

Figure 67

71
Interferences in ICP-MS
Ionization Efficiency

Ionization
Efficiency

Ionization Efficiency

Ru Lanthanides
Li Na Al K ScV MnCo Ga Rb Y Nb Pd In Cs Hf Tl Pb Ra
100 Sn Ba Ac
Mg Ca Cu Sr Zr Tc Re Bi
Ge Ta
Ti Fe Mo RhAg
Si W Po
Cr Ni
80 Cd Sb Os
Be
Ionization efficiency (%)

Zn
Te Pt
B
60
As
Au

40
P Se Rn
Hg
I
20
Ir
S
C O Xe
Ne Cl Ar Br
0
0 He F10 20 30 Kr 40 50 60 70 80 90
N

Figure 68

72
Interferences in ICP-MS
Signal Suppression

Signal
Suppression

Signal Suppression

10ppb Co in NaCl Matrix

1.2

1
Sensitivity

0.8

0.6

0.4

0.2

0
0.00 0.05 0.10 0.50 1.00 2.00 5.00

% NaCl

Figure 69

73
Interferences in ICP-MS
Matrix Effects – On Low Mass Analyte

Matrix Effects – On Low Mass


Analyte

Matrix Effects - on Low Mass Analyte

Lithium 7

B Zn Cd
1 Au
Signal

Na Low eV
Rb
0.5 Cs High eV
U
Tl
0
0 50 100 150 200 250

Mass of Matrix Element

Molar Ratio of Matrix Element to Analyte = 1000:1

Figure 70

74
Interferences in ICP-MS
Matrix Effects – On Medium Mass Analyte

Matrix Effects – On Medium Mass


Analyte

Matrix Effects - on Medium Mass Analyte

Rhodium 103

B Cd
Zn
1
Signal

Na Au
Rb Low eV
0.5 Cs Tl High eV
U
0
0 50 100 150 200 250

Mass of Matrix Element

Molar Ratio of Matrix Element to Analyte = 1000:1

Figure 71

75
Interferences in ICP-MS
Matrix Effects – On High Mass Analyte

Matrix Effects – On High Mass


Analyte

Matrix Effects - on High Mass Analyte

Thorium 232

B Zn Cd Au
1
Cs Tl Low eV
Na Rb
Signal

0.5 U High eV

0
0 50 100 150 200 250
Mass of Matrix Element
Molar Ratio of Matrix Element to Analyte = 1000:1

Figure 72

76
Interferences in ICP-MS
Space Charge Interface and Lens Region

Space Charge Interface and Lens


Region

Space Charge
Interface and Lens Region

+ +
+ +
+ + + ++ + ++
+ + +
++
+ +
+ + +
+ + +
+

Figure 73

After Extraction Lenses, Ion beam is predominantly positive charged.


Strong repulsive forces exist within the ion beam which affect low mass ions
much more than high mass ions tending to disperse the low mass portion of the
ion beam. Uncontrolled, space charge results in loss of low mass sensitivity,
especially in the presence of high mass matrix. Complex, multi-element ion
optics can compensate for this effect.

77
Interferences in ICP-MS
Ionization Suppression Plasma Region

Ionization Suppression Plasma


Region

Ionization Suppression
Plasma Region

Na+
- - -
Zn Na+ Na+ - Zn
-
- -
Na+ - - - Na+
- Na+
- - - --Na+-
Na+ Na+ - - -
Na+ - Na+
Zn Zn
- - - -

Na Na+ + e-
Zn+ + e- Zn

Figure 74

78
Interferences in ICP-MS
What Can Be Done About Matrix Effects

What Can Be Done About Matrix


Effects

What Can Be Done About Matrix Effects

y Dilution of Sample
y Internal Standardization
y Standard Additions
y Matrix Elimination
y Chromatography
y ETV
y Membrane desolvation

Figure 75

79
Interferences in ICP-MS
What Can Be Done About Matrix Effects

80
Tuning the Agilent 7500
Tuning the Agilent 7500
Why Tune the ICP-MS?

Why Tune the ICP-


MS?

Why Tune the ICP-MS?

■ Optimize Sensitivity
y Maximize Signal
y Minimize Noise
■ Verify Correct Mass Calibration
■ Verify Correct Ion Ratio Response
■ Minimize Interferences
y Oxides
y Doubly-Charged Ions
y Argides

Figure 76

82
Tuning the Agilent 7500
Tuning Procedure Overview

Tuning Procedure
Overview

Tuning Procedure Overview

■ Tune Plasma Parameters


RF Power
Gas Flows
Peristaltic Pump Flow
Torch Position
■ Tune Ion Optics
Extraction and Ion focusing lenses
Omega Lenses
■ Tune Quadrupole Mass Analyzer
Optimum Mass Resolution and Response
Correct Mass Assignments
■ Tune Detector
Optimum Sensitivity
Optimum Dual Mode (Pulse and Analog) linearity
■ Save Tune Conditions
■ Generate Tune Report

Figure 77

83
Tuning the Agilent 7500
Agilent 7500 ICP-MS Manual Tune Checklist [1]

Agilent 7500 ICP-MS Manual Tune Checklist


[1]

Agilent 7500 ICP-MS Manual Tune Checklist [1]

See “Manually Tuning the HP-4500 ICP-MS” for detailed instructions


I. Verify Hardware
■ vacuum, gas pressures and flows, peri-pump tubes and connections, error log
■ examine cones with magnifier
II. Verify Plasma Parameters
■ Aspirate tune solution #1, Warm up for 15-30 min
■ Check sensitivity, and precision
■ Fine tune carrier and/or blend gas flows for maximum signal, minimum RSDs (high
Li RSDs are usually related to worn or damaged cones)
■ Verify torch position and run torch position autotune if in doubt
■ Check oxides (<0.8% is fine). If high: 1. Decrease carrier and/or blend gas flow. 2.
Decrease peri-pump flow. 3. Increase sampling depth. 4. Increase RF power
III. Ion Lenses
■ Adjust Ion lenses for maximum (or desired) signal and minimum noise and RSDs in this
order :
Extract 1 and Extract 2 simultaneously (maintain ~ 50 V difference)
Einzel 2
Omega Bias, Omega + and Omega - in that order
QP focus if necessary

Figure 78

84
Tuning the Agilent 7500
Agilent 7500 ICP-MS Manual Tune Checklist [2]

Agilent 7500 ICP-MS Manual Tune Checklist


[2]

Agilent 7500 ICP-MS Manual Tune Checklist [2]

IV. Quadrupole Parameters


■ Select Resolution/Axis
■ Observe peak shapes, optimize by increasing Plate Bias and Pole Bias together if
necessary
■ Adjust peak widths (0.7-0.75 AMU at 10%) if necessary with AMU offset for low mass
and AMU gain for mid- and high-mass
■ Adjust mass calibration (nominal mass +/- 0.05 AMU) with Mass Offset for low mass and
Mass Gain for mid and high mass

V. Detector Parameters
■ Automatically: Select SetEM from Tune menu
■ Run P/A factor Autotune from Tune while aspirating 100 ppb standard

VI. Save Tune and Print Tune Report

Figure 79

85
Tuning the Agilent 7500
Autotune Screen

Autotune
Screen

Autotune Screen

Figure 80

Full Autotune should normally only be used when manually tuning the instrument
is unsuccessful. Most adjustments can be made more easily and quickly
manually. Exceptions are Torch Position, SetEM, Axis and Resolution, and PA
factor setting.
Setting realistic tune targets will increase the probability of a successful autotune
and speed up the process.
Setting appropriate and relatively narrow parameter ranges will result in faster
and more consistent autotunes

86
Tuning the Agilent 7500
Autotuning of ICP Torch Position and New Target Tune

Autotuning of ICP Torch Position and New Target


Tune

Autotuning of ICP Torch Position and


New Target Tune
• Autotuning is used for consistent optimization, against a pre-
defined set of tuning criteria (sensitivity, background, oxides, etc.)

• ICP torch position is critical for obtaining best sensitivity and


lowest molecular interferences - manual ICP torch adjustment is
imprecise and highly dependent on operator skill and experience.
Can also be VERY time consuming.

• Target tune - user can specify the required sensitivity - even at


different parts of the mass range. Target tune gives the novice
user the expertise of an ICP-MS expert!

• Not limited to pre-defined suite of tune elements.

Figure 81

87
Tuning the Agilent 7500
Features of Autotune (1)

Features of Autotune
(1)

Features of Autotune (1)

z Customizing
Ö Tune mass selection
Ö Target tune set up for sensitivities, etc.(*)
Ö Tune parameter range set up, fixed tune parameter set up, etc. (*)
z Speed
Ö Quick Mode Option
Ö Fix parameter set up (*)
Ö Quick measurement
z Visualizing
Ö Real time display of the Indicator
(*) These items were realized in the Agilent 4500, but enforced in the Agilent
7500

Figure 82

88
Tuning the Agilent 7500
Features of Autotune (2)

Features of Autotune
(2)

Features of Autotune (2)

• Multiple Functions and Abilities


¾ 4 tuning modes
♦ Extraction,
♦ Soft Extraction,
♦ Cool Plasma
♦ Hot ShieldTorch
¾ Target Tune
¾ RSD Option
¾ Saving function of Autotune’s target set up file
¾ Skip function
¾ Intelligence of Autotune sequence

Figure 83

89
Tuning the Agilent 7500
Choosing the Autotune Mode

Choosing the Autotune


Mode

Choosing the Autotune Mode

Figure 84

90
Tuning the Agilent 7500
Basics of the Soft Extraction Mode

Basics of the Soft Extraction


Mode

Basics of the Soft Extraction Mode


Extraction Soft Extraction
Plasma Plasma

Skimmer 2nd Extraction Lens Skimmer 2nd Extraction Lens


1st Extraction Lens 1st Extraction Lens

• Extraction: Charge Separation between the skimmer and the 1st extraction lens
• Soft Extraction: Charge Separation between the 1st extraction lens and the 2nd
extraction lens

Figure 85

91
Tuning the Agilent 7500
Comparison of Extraction Modes Settings

Comparison of Extraction Modes


Settings

Comparison of Extraction Modes Settings

Extraction Mode Soft Extraction Mode


Cool Hot Cool Hot
Shield Li: 20000 Li: 200 Li: 6
Torch (Co: 1000) -------
Y : --- Y : 700 Y : 300
Tl: 300 Tl: 300 Tl: 200
BKG: < 1cps BKG: <20cps BKG: < 1cps
CeO: >>> CeO: 30% CeO: 8%
Non-Shield Li: 15
Torch ---------- -------- ----------
Y : 30
Tl: 15 Typical Sensitivity for
BKG: < 5cps Each Mode By Agilent
CeO: 0.5 % 7500s (Mcps/ppm)

Figure 86

92
Tuning the Agilent 7500
Autotune - Target Setting

Autotune - Target
Setting

Autotune - Target Setting

Figure 87

93
Tuning the Agilent 7500
Target Setting - Range Setting

Target Setting - Range


Setting

Target Setting - Range Setting

Figure 88

94
Tuning the Agilent 7500
Sensitivity Tuning

Sensitivity
Tuning

Sensitivity Tuning

Figure 89

95
Tuning the Agilent 7500
Peak Shape and Resolution

Peak Shape and


Resolution

Peak Shape and Resolution

Quadrupole peaks Increasing the Resolution


tail slightly to the (narrower peaks) affects
low mass side the low mass side more.

Peak height is reduced


and peak center shifts to
high mass side.

Figure 90

96
Tuning the Agilent 7500
Abundance Sensitivity

Abundance
Sensitivity

Abundance Sensitivity

Spectrum.1 [10.409 sec]:YCS8.D#[CPS][Log] Spectrum.2 [10.409 sec]:YCS9.D#[CPS][Log]


1.0E7 1.0E7

8
1.0E6 1.0E6

1.0E5

1.0E4
9 good
1.0E5

1.0E4
bad
1000 1000

100 100

10 10

m/z->

m/z-> 82 84 86 88 90 92 m/z-> 82 84 86 88 90 92

Figure 91

97
Tuning the Agilent 7500
Quadrupole Mass Filter - Scan Line

Quadrupole Mass Filter - Scan


Line

Quadrupole Mass Filter - Scan Line

U
Mass Resolution dependent on: Tl
513 V
RF frequency
Length of quadrupole
Ion speed Y

Resolution ∝ f2 L2 / V
Li
f : Frequency
L: Length of Q-pole
v: Speed of ion V

3056 V

Figure 92

98
Tuning the Agilent 7500
Detection Limits in Normal Mode

Detection Limits in Normal


Mode

Detection Limits in Normal Mode

Unit : ng/L(ppt)
Li Be 3 sigma B C N O F Ne
5 2.8 Integration Time :3sec. 11 5000
51 0.9 93 82000
Na Mg Upper Value : Detection Limit Al Si P S Cl Ar
100 40 Lower Value : BEC 10 700 500 3000 6000
730 110 64 16000 5200 44000 38000
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
3000 1300 10 2 3 15 2 900 1 4 3 22 3 1 8 80 600
34000 14000 120 9 8 65 30 19000 3.2 19 15 260 6.2 6 35 460 2300
Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
0.8 1 0.2 0.3 0.2 0.5 0.8 7 1 0.7 0.7 0.1 0.6 0.7 7 70
3.4 2 0.2 0.3 0.2 0.5 0.8 100 2 1.4 1.7 0.2 2 1 7 230
Cs Ba * Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
0.5 2.5 30 0.08 0.3 0.3 0.2 18 0.8 1.6 1 1 0.2
2.7 3.5 4 0.1 0.5 0.3 0.2 310 2.3 1.2 1.8 6 0.3
Ra **

* La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
0.1 0.1 0.08 1 0.7 0.1 0.4 0.3 0.3 0.08 0.3 0.07 0.2 9
0.1 0.1 0.09 0.6 0.7 0.1 0.5 0.9 0.4 0.09 0.2 0.09 0.3 1
** Ac Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
0.2 0.6
0.2 0.8

Figure 93

99
Tuning the Agilent 7500
Detection Limits in Soft Extraction Mode

Detection Limits in Soft Extraction


Mode

Detection Limits in Soft Extraction Mode

Unit : ng/L(ppt)
Li Be 3 sigma B C N O F Ne
66 0.5 Integration Time :3sec. 6
800 1.1 56
Na Mg Upper Value : Detection Limit Al Si P S Cl Ar
200 0.7 Lower Value : BEC 2 800 1000 10000 3000
2200 10 17 19000 13000 100000 120000
K Ca Sc Ti V Cr Mn Fe Co Ni Cu Zn Ga Ge As Se Br Kr
2000 90 0.9 0.5 0.1 4.2 0.3 200 0.2 0.1 0.2 0.6 0.08 5 0.4 8 20
14000 2700 23 3.5 1.2 74 8 7500 3.1 0.8 1.7 2.5 0.8 47 5.2 160 830
Rb Sr Y Zr Nb Mo Tc Ru Rh Pd Ag Cd In Sn Sb Te I Xe
0.05 0.02 0.01 0.01 0.02 0.1 0.04 0.04 0.05 0.1 0.04 0.01 0.1 0.04 0.3 1
0.8 0.03 0.02 0.02 0.1 0.8 0.08 0.8 0.1 0.2 0.1 0.02 0.9 0.2 0.7 40
Cs Ba * Hf Ta W Re Os Ir Pt Au Hg Tl Pb Bi Po At Rn
0.8 0.1 0.1 0.1 0.3 0.05 0.05 0.08 0.3 0.8 0.2 0.1 0.03
23 0.2 0.1 0.3 1.7 0.07 0.07 0.4 1.4 8.4 0.8 0.4 0.07
Ra **

* La Ce Pr Nd Pm Sm Eu Gd Tb Dy Ho Er Tm Yb Lu
0.01 0.01 0.008 0.03 0.07 0.02 0.03 0.01 0.09 0.02 0.08 0.01 0.06 0.02
0.02 0.02 0.01 0.08 0.1 0.03 0.06 0.05 0.07 0.02 0.06 0.02 0.08 0.02
** Ac Th Pa U Np Pu Am Cm Bk Cf Es Fm Md No Lr
0.07 0.05
0.1 0.08

Figure 94

100
Tuning the Agilent 7500
Low BECs in Soft Extraction Mode

Low BECs in Soft Extraction


Mode

Low BECs in Soft Extraction Mode

Background Equivalent Concentration (ppt)


Elements Soft-extraction Extraction
Li 4 800

Na 190 2200

Y 0.005 0.01

Ce 0.004 0.02

Th 0.005 0.1

Figure 95

101
Tuning the Agilent 7500
Pulse/Analog (P/A) Tuning

Pulse/Analog (P/A)
Tuning

Pulse/Analog (P/A) Tuning

Figure 96

102
Maintenance of the Agilent 7500
Maintenance of the Agilent 7500
Maintenance Schedule

Maintenance
Schedule

Maintenance Schedule

• Daily
– Lab conditions, Argon, drain, peristaltic pump tubing, cones
• When Needed
– Cones, nebulizer, peristaltic pump tubing, torch, water filter,
electron multiplier
• Weekly
– tuning solution preparation, torch, spray chamber, nebulizer,
carrier gas line, cooling system
• Monthly
– check rotary pump, oil mist filter, check extraction lens
• 6 months
– clean lenses, change rotary pump oil, replace gas tubing
• Yearly
– replace o-rings, clean penning gauge, check/replace mist filter

Figure 97

104
Maintenance of the Agilent 7500
Running Time Maintenance Screen

Running Time Maintenance


Screen

Running Time Maintenance Screen

Figure 98

105
Maintenance of the Agilent 7500
Early Maintenance Feedback (EMF)

Early Maintenance Feedback


(EMF)

Early Maintenance Feedback (EMF)

EMF monitors the


operating hours of all
major system
components and
proactively informs the
user when:
maintenance is
required
consumable items
should be replaced

Users can define their


own maintenance
schedule

Figure 99

106
Maintenance of the Agilent 7500
Normal Maintenance of the Sample Introduction System

Normal Maintenance of the Sample Introduction


System

Normal Maintenance of the Sample Introduction


System

Non-Glassware Components
- Sample tubing
- Peristaltic pump tubing 1. Soak in 1% to 5% nitric acid (5 min.)
- Babington nebulizer 2. Clean in ultrasonic bath (5 min.)
- Crossflow nebulizer 3. Rinse with DI water
- Nebulizer end caps
- O-rings
Glassware
-Concentric nebulizer
- Spray chamber 1. Soak in 1% to 5% nitric acid (5 min.)
or sonicate in 10% Citranox
- Ball joint connector 2. Rinse with DI water
- Torch

Figure 100

107
Maintenance of the Agilent 7500
Overnight Cleaning of the Sample Introduction System

Overnight Cleaning of the Sample Introduction


System

Overnight Cleaning of the Sample Introduction


System

For severely contaminated glassware and non-glassware


components

Procedure
1. Soak in 5% nitric acid overnight or boil in 10%
Citranox® for 1 hour, rinse in 5% nitric acid
2. Rinse with DI water

Figure 101

108
Maintenance of the Agilent 7500
Sample Introduction Maintenance

Sample Introduction
Maintenance

Sample Introduction Maintenance

Top Cover

Sample Transfer Line


Spray Chamber

Ball Joint Clamp

Nebulizer and End Cap

Spray Chamber Chiller

Figure 102

109
Maintenance of the Agilent 7500
Nebulizer Connections

Nebulizer
Connections

Nebulizer Connections

Blend Gas Carrier Gas

Concentric
Nebulizer Sample Uptake

Blend Gas Sample Uptake

Babington Carrier Gas


Nebulizer

Blend Gas

Crossflow
Nebulizer
Carrier Gas
Sample Uptake

Figure 103

110
Maintenance of the Agilent 7500
Maintenance of a Babington Nebulizer

Maintenance of a Babington
Nebulizer

Maintenance of a Babington Nebulizer

Thumb Screws Teflon Tubing

Tube Connector
Attachment Tool
Removal Tool

Babington Nebulizer
Sample Tubing

Disconnecting the tubing connector


Plate
using the attachment tool

Babington Nebulizer

End Cap Wide Groove Narrow Grove

Connection Port for


Make-up Gas

Figure 104

111
Maintenance of the Agilent 7500
Torch Maintenance

Torch
Maintenance

Torch Maintenance

Auxiliary Gas
Torch

Plasma Gas

Torch Mounting Assembly

Figure 105

112
Maintenance of the Agilent 7500
Interface Maintenance

Interface
Maintenance

Interface Maintenance

Routine Maintenance
Components
- Sampling cone
- Skimmer cone
Procedure
1. Soak in 1-5% nitric acid (<10 min.)
2. Rinse with DI water
Removing Severe Deposits
Components
- Sampling cone
- Skimmer cone
Procedure
1. Polish with waterproof abrasive paper
- be careful not to damage the orifice
2. Rinse with DI water

Figure 106

113
Maintenance of the Agilent 7500
Maintenance of the Cones

Maintenance of the
Cones

Maintenance of the Cones

Removal of the Skimmer Cone

Removal of the Sampling Cone

Figure 107

114
Maintenance of the Agilent 7500
Extraction Lenses Maintenance

Extraction Lenses
Maintenance

Extraction Lenses Maintenance

● Remove skimmer base from vacuum manifold


● Disassemble extraction lenses, screws and spacers
● Polish extraction lenses with waterproof abrasive paper
● Wash extraction lenses in DI water
● Sonicate lenses, spacers and screws in DI water for 5 minutes
● Sonicate lenses, spacers and screws in acetone for 5 minutes
● Reassemble lenses, spacers and screws on skimmer base
● Install skimmer base on vacuum manifold

Figure 108

115
Maintenance of the Agilent 7500
Extraction Lenses

Extraction
Lenses

Extraction Lenses

Screw 1
Skimmer Base

Extraction Lens 1
Extraction Lens 2
Spacer 2

Screw 2
Spacer 1

Spacer 1

Figure 109

116
Maintenance of the Agilent 7500
Cleaning of the Einzel Lens and Omega Lens Assembly

Cleaning of the Einzel Lens and Omega Lens


Assembly

Cleaning of the Einzel Lens and Omega Lens


Assembly

●Disassemble lenses, remove screws and spacers


●Polish each lens, lens orifice and the curved surfaces of the Omega lenses
using waterproof abrasive paper
●Wash lenses in DI water
●Sonicate lenses, spacers and screws in DI water for 5 minutes
●Sonicate lenses, spacers and screws in acetone or alcohol for 5 minutes
●Reassemble and install the lens assemblies and install them in the
vacuum manifold
●Replace top cover and vent nut and restart system.

Figure 110

117
Maintenance of the Agilent 7500
Instrument Shutdown

Instrument
Shutdown

Instrument Shutdown

Shutdown

Standby

Figure 111

118
Maintenance of the Agilent 7500
Removal of the Einzel Lens - Omega Lens Assembly

Removal of the Einzel Lens - Omega Lens


Assembly

Removal of the Einzel Lens - Omega Lens


Assembly

Screw 1 Einzel Lenses - Omega


Lens Assembly
Screw 2

Figure 112

119
Maintenance of the Agilent 7500
Expanded View of Einzel Lens - Omega Lens Assembly

Expanded View of Einzel Lens - Omega Lens


Assembly

Expanded View of Einzel Lens - Omega Lens


Assembly
Screw 2
Spacer 1

Spacer 1

Omega (-)
Spacer 1 Shaft
Spacer 1 QP Focus (back)
QP Focus (front)
Omega Bias (back)
Spacer 1
Omega (+)
Spacer 1
Spacer 2
Spacer 1 Omega Contact
Omega
Spacer 1 Bias (front)
Screw 1 Spacer 3
Base Plate
Screw 3
Einzel 3
Einzel 2
Einzel 1

Figure 113

120
Maintenance of the Agilent 7500
Plate Bias Lens

Plate Bias
Lens

Plate Bias Lens

Plate Bias Lens

QP Focus Guide

Screw 1

Figure 114

121
Maintenance of the Agilent 7500
Penning Gauge

Penning
Gauge

Penning Gauge

1. Magnet Housing
2. Collar
3. Anode Assembly
4. O-ring
5. Body Tube
6. Cathode Plate
7. Cathode Tube
8. Circlip

Figure 115

122
Maintenance of the Agilent 7500
Rotary Pump Maintenance

Rotary Pump
Maintenance

Rotary Pump Maintenance

● Put system in Shutdown mode (Turn vacuum off)


● Turn off pump circuit breakers on front panel
● Remove oil inlet cap on top of pump
● Remove oil drain plug using a flat blade screwdriver and allow oil to
drain into a waste container
● Replace oil drain plug and add new oil through the oil inlet until the
oil level window is 80% full
● Replace oil inlet cap, turn on pump breakers on front panel, and start
system (Turn vacuum on)

Figure 116

123
Maintenance of the Agilent 7500
Changing Rotary Pump Oil

Changing Rotary Pump


Oil

Changing Rotary Pump Oil

Figure 117

124
Maintenance of the Agilent 7500
Maintenance Logbook Setting

Maintenance Logbook
Setting

Maintenance Logbook Setting

Figure 118

125
Maintenance of the Agilent 7500
Maintenance Logbook

Maintenance
Logbook

Maintenance Logbook

Figure 119

126
Maintenance of the Agilent 7500
Sample Introduction Maintenance

Sample Introduction
Maintenance

Sample Introduction Maintenance

Figure 120

127
Maintenance of the Agilent 7500
Air Filters Maintenance

Air Filters
Maintenance

Air Filters Maintenance

■ Remove air filters from the instrument

■ Remove dust from the filters using a vacuum cleaner

■ Wash filters with water if necessary and allow to dry

■ Return filters to the instrument

Figure 121

128
Maintenance of the Agilent 7500
Instrument Start-up

Instrument Start-
up

Instrument Start-up

Shutdown

Analysis

Standby

Figure 122

129
Maintenance of the Agilent 7500
Instrument Start-up

130
Internal Standardization in ICP-MS
Internal Standardization in ICP-MS
The Role of Internal Standards

The Role of Internal


Standards

The Role of Internal Standards

Correct for variations in response due to:


Matrix Effects
Transport effects
Nebulization effects
Ionization effects
Space-charge effects
Instrument Drift

Figure 123

132
Internal Standardization in ICP-MS
How the Internal Standards Work - 1

How the Internal Standards Work -


1

How the Internal Standards Work - 1

• Added to each sample, standard and blank at identical


concentration
• System therefore expects identical response from ISTDs in each
solution
• Ratio of measured ISTD response to expected ISTD response is
used to correct the response of the non-internal standard
elements accordingly.

Figure 124

133
Internal Standardization in ICP-MS
How the Internal Standards Work - 2

How the Internal Standards Work -


2

How the Internal Standards Work - 2

■ In all cases where potential matrix suppression exists, the use of internal
standards is necessary.
■ The calibration curve is plotted using the ratio of the analyte signal to IS signal

(i.e. ax/isx), thereby canceling the effect of matrix suppression.


■ Each analyte in the sample will then be quantitated using the ratio of the IS and

the analyte in the sample.

Figure 125

134
Internal Standardization in ICP-MS
Choice of the Internal Standard

Choice of the Internal


Standard

Choice of the Internal Standard

It is assumed that the IS elements behave in the same way that the
analytes do in the plasma when using this correction. Therefore, selecting
the appropriate IS element is very important.

These are the things to take into consideration:


■ The element is not contained in the sample solution.

■ The mass number is close to that of the analyte.

■ The ionization potential is similar to that of the analyte.

■ Chemical characteristics

The ionization potential matching is extremely important for analytes


with high ionization potentials.

Elements that are commonly used as IS are:


6Li, Sc, Ge, Y, Rh, In, Tb, Ho, Bi

Figure 126

135
Internal Standardization in ICP-MS
Concentration of Internal Standards

Concentration of Internal
Standards

Concentration of Internal Standards

The IS elements can be added to the sample in 2 ways:

■ On-line addition by peristaltic pump


■ Spike in each standard and sample

In either case, it is recommended to add IS in concentration


levels around mid-calibration range (exact concentration is
application-specific)

For environmental applications 50 ppb concentration is usually applied.


For on-line IS addition use 1 ppm IS stock solution, and use the
peristaltic pumps as recommended.

Figure 127

136
Internal Standardization in ICP-MS
On-line Addition of Internal Standards

On-line Addition of Internal


Standards

On-line Addition of Internal Standards

■ On-line Addition
The IS solution is introduced by a narrow
tubing (f = 0.19 mm), and is mixed with
the sample at the Y connector. The
dilution factor of the IS by the sample is
about 1/20. Therefore, a 1 ppm solution
in 5 % HNO3 would yield approximately
50 ppb in the sample. At least 5 %
HNO3 is needed to avoid absorption of
elements to the tubing, as absorption is
more severe using narrow tubing.

Figure 128

137
Internal Standardization in ICP-MS
On-line Addition of Internal Standards

138
Sample Preparation Techniques for ICP-
MS
Sample Preparation Techniques for ICP-MS
Contamination

Contamination

Contamination

■ "Contamination is the introduction of any component which


affects the numerical value finally attributed to a constituent
relative to the amount present prior to sampling"
■ Types of contamination:

–positive
–negative
–pseudocontamination

D.E. Robertson "Ultrapurity, Methods and Techniques",


M. Zeif and R. Speights, Eds., Marcel Dekker, NY, 1972

Figure 129

140
Sample Preparation Techniques for ICP-MS
Types of Contamination

Types of
Contamination

Types of Contamination

Positive contamination - results in additive errors. Caused by


➤ impurities in reagents
➤ lab environment
➤ desorption from container walls
Negative contamination - results in subtractive errors. Caused by
➤ losses in handling
➤ adsorption to container walls
Pseudocontamination - results in either positive or negative errors.
Caused by
➤ irreproducibility of experimental conditions

Figure 130

141
Sample Preparation Techniques for ICP-MS
Challenges of Trace Analysis

Challenges of Trace
Analysis

Challenges of Trace Analysis

Environment

Sample Methodology

Analyst Reagents

Standards

Figure 131

142
Sample Preparation Techniques for ICP-MS
When a Contamination Can Occur

When a Contamination Can


Occur

When a Contamination Can Occur

■ Sample collection
– collection techniques
– collection devices
■ Sample storage
– prevention of positive contamination
– prevention of negative contamination
■ Sample preparation
– reagents
– lab environment
– apparatus
■ Sample measurement
– instrument sample introduction system
– standards

Figure 132

143
Sample Preparation Techniques for ICP-MS
Reagents

Reagents

Reagents

■ Water
■ Nitric acid
■ Hydrochloric acid
■ Sulfuric acid
■ Hydrofluoric acid
■ Other inorganic acids
■ Hydrogen peroxide
■ Alkaline solutions
■ Organic solvents

Figure 133

144
Sample Preparation Techniques for ICP-MS
Water - Millipore

Water -
Millipore

Water - Millipore

Millipore Corporation
80 Ashby Road
P.O. Box 9125
Bedford, MA 01730-9903
Tel 1-800-MILLIPORE (1-800-645-5476)
fax 781-533-8873
Internet: http://www.millipore.com/H2O

Figure 134

145
Sample Preparation Techniques for ICP-MS
Nitric Acid

Nitric
Acid

Nitric Acid
From Fisher Scientific (1-800-766-7000)

“TraceMetal” - for environmental analysis -


500 mL in glass, catalog # A509-500,
Certificate of Lot Analysis included with each shipment

“Optima” - for semiconductor and clinical applications -


catalog # A467-250, 250 mL in Teflon™,
catalog # A467-500, 500 mL in Teflon™,
Certificate of Lot Analysis included with each bottle

from Mallinkrodt-Baker (1-800-444-0880)

“INSTRA-ANALYZED” (equivalent of “TraceMetal”),


500 mL in poly coated glass, catalog # 9598-00

“ULTREX II” (equivalent of “Optima”)


500 mL in Teflon™, catalog # 6901-05

Figure 135

146
Sample Preparation Techniques for ICP-MS
Selected Methods of Sample Preparation

Selected Methods of Sample


Preparation

Selected Methods of Sample Preparation

Ö Dilution
Ö Preconcentration
Ö Filtration
Ö Acidification
Ö Digestion
Ö Open vessel digestion
Ö Closed vessel digestion
Ö Microwave digestion
➱ Fusion
➱ Matrix separation
➱ Chromatography
➱ Electrothermal Vaporization

Figure 136

147
Sample Preparation Techniques for ICP-MS
Commonly Used Reagents (1)

Commonly Used Reagents


(1)

Commonly Used Reagents (1)


Nitric Acid (HNO3)
Used to dissolve a variety of materials such as metals, alloys,
biological samples.
Available in very high purity form.
Most preferable acid for ICP-MS work, as polyatomic ions are not
increased.
Hydrochloric Acid (HCl)
Used commonly for sample digestion.
Generates Cl derived polyatomic ions.
Can be evaporated to dryness and reconstituted in HNO3.
Hydrofluoric Acid (HF)
Used to dissolve silica-based materials and geological samples.
HF attacks glass, therefore the inert sample introduction system must
be used.
Teflon containers are often used.
Extreme health hazard.

Figure 137

148
Sample Preparation Techniques for ICP-MS
Commonly Used Reagents (2)

Commonly Used Reagents


(2)

Commonly Used Reagents (2)

Hydrogen Peroxide (H2O2)


Strong oxidizing agent used with other acids for digestion.
Useful for ICP-MS work, as backgrounds are similar to water (H2O).
Sulfuric Acid (H2SO4)
Oxidizing agent used with other acids for digestion.
Causes numerous polyatomic ion interferences.
Difficult to decompose in plasma due to high boiling point and viscosity.
Deteriorates Ni and Cu interface material.
Not recommended for use if it can be avoided.
Phosphoric Acid (H3PO4)
Used to buffer other acids.
Causes numerous polyatomic ion interferences.
Difficult to decompose in plasma due to high boiling point and viscosity.
Deteriorates Ni interface material.
Should be avoided.

Figure 138

149
Sample Preparation Techniques for ICP-MS
Commonly Used Reagents (3)

Commonly Used Reagents


(3)

Commonly Used Reagents (3)

Perchloric Acid (HClO4)


Strong oxidizing agent used with other acids for digestion of
organic materials.
Generates Cl derived polyatomic ions.
More difficult to evaporate than HCl.
Not recommended for use if it can be avoided.
Handle with care as many solid perchlorates are explosive.

Aqua Regia
(1 part conc. HNO3 + 3 parts conc. HCl)
Used for metal digestion, especially precious metals.
Generates Cl derived polyatomic ions.
Cl matrix can be removed by evaporation.

Figure 139

150
Semi-quantitative Analysis of Samples
Semi-quantitative Analysis of Samples
Semi-quantitative Analysis

Semi-quantitative
Analysis

Semi-quantitative Analysis

• What is Semi-quantitative Analysis


• Setting Semi-quantitative acquisition parameters
• Analyzing the samples
• Semi-quantitative data analysis

Figure 140

152
Semi-quantitative Analysis of Samples
What is Semi-quantitative Analysis?

What is Semi-quantitative
Analysis?

What is Semi-quantitative Analysis?

y Is an analytical procedure used to calculate concentrations


of all elements present in an unknown sample
y Is useful as a screen prior to Quantitative Analysis
y concentration ranges of analytes
y selection of internal standards
y Does not require an external calibration
y Is generally accurate to +/- 30% or better on completely
unknown samples

Figure 141

153
Semi-quantitative Analysis of Samples
Data Acquisition

Data
Acquisition

Data Acquisition

Figure 142

154
Semi-quantitative Analysis of Samples
Method Set-up for Semi-quantitative Analysis

Method Set-up for Semi-quantitative


Analysis

Method Set-up for Semi-quantitative Analysis

To write comments & record updates

Do not check

Figure 143

155
Semi-quantitative Analysis of Samples
Parameters Selection - Spectrum Acquisition

Parameters Selection - Spectrum


Acquisition

Parameters Selection - Spectrum Acquisition

[2] Select
Mass Scale
[1] Select Semi-Quant peak pattern,
0.1 s/point, 1 repetition

Figure 144

156
Semi-quantitative Analysis of Samples
Parameters Selection - Selection of Masses

Parameters Selection - Selection of


Masses

Parameters Selection - Selection of Masses

[2] Exclude unlike, impossible, and interfered with masses

[1]Select mass ranges

Figure 145

157
Semi-quantitative Analysis of Samples
More Acquisition Parameters

More Acquisition
Parameters

More Acquisition parameters

If checked, use the


browse button to select
the appropriate file

Do not check

When background is
subtracted, and the
internal standards are
used, exclude up to 4
masses from
subtraction (IS masses)

Figure 146

158
Semi-quantitative Analysis of Samples
Report Generation

Report
Generation

Report Generation

Select a
report

Figure 147

159
Semi-quantitative Analysis of Samples
Semi-quant Parameters

Semi-quant
Parameters

Semi-quant Parameters

Threshold value
(tune-dependent)

Select ppb as the


units (or ppm)

Not to be
reported
(output = OFF)

Figure 148

Set the Minimum Peak threshold to reject results based on noise. The default is
50, but remember, typical response in tuning is 20 million cps/ppm which is 20K
cps/ppb. Therefore at least 200 cps represents a reporting threshold of
approximately tens of ppt, a reasonable value.
Output Mode is either ON, OFF or AUTO. ON - this element will always be
reported; OFF - this element will never be reported; AUTO - this element will
be reported IF no significant interference is detected. The acceptable level of
interference is stored in WIN.INI and can be edited there.
Concentration Units is either user selectable or when set to AUTO, the
ChemStation will select the appropriate units based on the estimated
concentration. Enter the concentration in ppb (or ug/L) of the elements in the cal
standard. Leave the other concentration fields blank.

160
Semi-quantitative Analysis of Samples
Semi-quantitative Data Analysis

Semi-quantitative Data
Analysis

Semi-quantitative Data Analysis

Figure 149

Semiquant analysis can be used to estimate the concentration of any element for
which a precise measurement can be made by ICP-MS (> 70 elements).
Typically, semiquantitation is accurate to within +/- 30 percent on completely
unknown samples. However, semiquantitation is subject to the same interferences
as quantitation. Possible interferences due to oxides, hydrides, argides, dimers
and doubly-charged ions are checked and flagged on the report. Interference
correction equations can be used to minimize these effects where applicable. The
use of internal standards can help correct for matrix differences. Blank
subtraction can be used to eliminate contributions from laboratory reagents and
sample preparation.
The ChemStation comes configured with default SemiQuant response factors.
These factors are based on relative ionization potentials and numbers of isotopes
for each element. These factors can be updated to reflect the tune state of the HP-
4500 by analyzing a calibration mix. At least 3 elements should be used, though
more is better. The ChemStation will then interpolate between analyzed masses
to update all SemiQuant response factors.

161
Semi-quantitative Analysis of Samples
Editing Parameters

Editing
Parameters

Editing Parameters

Holding <ctrl>
select elements
to be reported

Change default
mass for Cu
from 63 to 65

Enter the
concentration
of calibration
standard

Figure 150

162
Semi-quantitative Analysis of Samples
Daily Update of the Semi-Quant Parameters

Daily Update of the Semi-Quant


Parameters

Daily Update of the Semi-Quant Parameters

Load a previously
acquired file and
select “Correct by
Current Data”

Figure 151

Correct by Current Data recalculates all semi-quant response factors by first


dividing the supplied concentrations by the responses for those elements. Other,
non-calibrated element response factors are estimated by interpolation.

163
Semi-quantitative Analysis of Samples
Internal Standard Correction for Off-line Internal Standard Addition

Internal Standard Correction for Off-line Internal


Standard
Addition

Internal Standard Correction for Off-line Internal


Standard Addition

Activating internal Off-line addition


standard correction

IS concentration

Selecting internal standard

Figure 152

Internal Standardization is recommended since it corrects for changes in


instrument sensitivity due to matrix and other effects.
To configure internal standard correction:
Data Analysis >> SemiQuant >> Internal Standard Correction…
Internal standard correction can be applied in two modes:
‘Normal Mode’ assumes that internal standards are added to the samples only and
no ISTD reference data file is required. This can be used for analyses such as
Laser Ablation, where a matrix element is used as the internal standard. The
ISTD factor is calculated from the supplied ISTD concentration and the ISTD
response. Up to four internal standard elements can be selected.

164
Semi-quantitative Analysis of Samples
Internal Standard Correction for On-line Internal Standard Addition

Internal Standard Correction for On-line Internal


Standard
Addition

Internal Standard Correction for On-line Internal


Standard Addition
Activating internal On-line addition
standard correction

Selecting internal standard

Figure 153

‘Auto Add Mode’ assumes that the online internal standard addition configuration
is used. In this case, the exact concentration of the ISTD elements need not be
known since an ISTD reference file used. Only the internal standard masses need
be selected. In this case, the only requirement is that the ISTD concentration in
all samples be identical to the ISTD reference file.
The ISTD reference file can be either a blank or a calibration standard containing
online added internal standards.

165
Semi-quantitative Analysis of Samples
Example of Semi-Quant Report [1]

Example of Semi-Quant Report


[1]

Example of Semi-Quant Report [1]

For signal <200


cps (threshold), a
threshold
concentration
(corresponding to
200 cps) is report

For signal >200


cps (threshold),
the concentration
of the analyte is
reported

Figure 154

166
Semi-quantitative Analysis of Samples
Example of Semi-Quant Report [2]

Example of Semi-Quant Report


[2]

Example of Semi-Quant Report [2]

Software indicates
possible interferences

Figure 155

167
Semi-quantitative Analysis of Samples
Generating a Semi-quant Report

Generating a Semi-quant
Report

Generating a Semi-quant Report

Î Automatically as part of Run Method or Run Sequence


Î Manually from Data Analysis
Î SemiQuant also allows Custom Reports and Custom Databases
Î Use of DoList for multiple reports

Figure 156

168
Semi-quantitative Analysis of Samples
Manual Verification of the Data

Manual Verification of the


Data

Manual Verification of the Data

Figure 157

169
Semi-quantitative Analysis of Samples
Manual Verification of the Data

170
Quantitative Analysis of Samples
Quantitative Analysis of Samples
What is Quantitative Analysis?

What is Quantitative
Analysis?

What is Quantitative Analysis?

Quantitative Analysis:

■ An analytical procedure used to calculate concentrations of specific

elements in unknown samples

■ Uses calibration curves based on the response of one or more levels of

standards to calculate unknown concentrations

■ Allows the use of internal standards to correct for instrument drift

and matrix differences between standards and samples

Figure 158

172
Quantitative Analysis of Samples
Method Set-up for Quantitative Analysis

Method Set-up for Quantitative


Analysis

Method Set-up for Quantitative Analysis

Steps in Setting up a Quantitative Analysis:

■ Editing the AMU Select file, if necessary


■ Interference correction equations
■ Spectrum acquisition parameters
■ Peripump program
■ Calibration table
■ Acquiring calibration standards and updating the calibration table
■ Analyzing unknown samples

Figure 159

173
Quantitative Analysis of Samples
Step One: Editing the AMU Select File

Step One: Editing the AMU Select


File

Step One: Editing the AMU Select File

From Top >> AcquireData >> Edit AMU Select File (amu)...

Figure 160

The AMU select file is the database from which default isotopes are selected
when elements are selected from the Periodic Table in Edit Entire Method.
Multiple AMU select files can be created for different applications. For example
AMU select files can be created which automatically select the EPA specified
isotopes when running EPA methods. AMU select files can also be created for
specific matrices in order to avoid known isobaric or polyatomic interferences.
The element needs to have at least one isotope selected, in order to be accessible
in the method setting. If needed, select isotopes for P, and Si.

174
Quantitative Analysis of Samples
Editing a Method for Quantitative Analysis

Editing a Method for Quantitative


Analysis

Editing a Method for Quantitative Analysis

Select ‘Edit Entire Method’

QC Parameters will only be


available if Intelligent
Sequencing is enabled

Figure 161

175
Quantitative Analysis of Samples
Method Information

Method
Information

Method Information

Figure 162

176
Quantitative Analysis of Samples
Acquisition Modes

Acquisition
Modes

Acquisition Modes

Figure 163

Spectrum mode is the most common acquisition mode for standard applications:
• Quant
• Semiquant
Time Resolved Analysis (TRA) and Time Program (more sophisticated than
TRA) are used when a transient signal is measured:
• Electrothermal Vaporization (ETV)
• Laser Ablation (LA)
• Discrete Sampling Analysis (using ISIS)
• Chromatographic analysis (LC, GC, IC, CE)
Isotope Analysis mode is used when additional precision is needed for isotope
ratio measurements. It is similar to spectrum mode, but with 10X higher
sampling frequency.

177
Quantitative Analysis of Samples
Acquisition Modes

Multitune mode is used when during a single acquisition more than one tuning
parameters are needed to accomplish the optimum performance.

178
Quantitative Analysis of Samples
Acquisition Parameters - Multitune Method

Acquisition Parameters - Multitune


Method

Acquisition Parameters - Multitune Method

Figure 164

179
Quantitative Analysis of Samples
Periodic Table

Periodic
Table

Periodic Table

Figure 165

180
Quantitative Analysis of Samples
Mass Table

Mass
Table

Mass Table

Figure 166

181
Quantitative Analysis of Samples
Peristaltic Pump Program

Peristaltic Pump
Program

Peristaltic Pump Program

Maximum Speed should not exceed


0.30 rps with online ISTDs addition
Typical Stabilization Time is 50-60 sec
Probe Rinse should be very short (~1
sec)
Rinse time is sample/matrix dependent
(30-90 sec)

Figure 167

182
Quantitative Analysis of Samples
Raw Data Corrections

Raw Data
Corrections

Raw Data Corrections

Subtracts Background
spectrum (counts from
counts) before
Quantitation

Enables Interference
Correction Equations

Figure 168

183
Quantitative Analysis of Samples
Configure Reports

Configure
Reports

Configure Reports

Enabled with
Intelligent
Sequencing

Figure 169

184
Quantitative Analysis of Samples
Calibration

Calibration

Calibration

• Now independent of the method


• Multiple methods can share the same calibration
• Current calibration is displayed on the TOP and Data Analysis title bar

• Link between calibration and method is established when method is saved

Figure 170

185
Quantitative Analysis of Samples
Calibration Table

Calibration
Table

Calibration Table

Figure 171

Shortcuts:
• Double click any column selects entire column.
• Fill Across is useful for copying Internal Standard Concentrations to all
levels.
• Multiple entries can be selected using <Ctrl> plus left click or Shift plus
left click.
• Min. Conc. is the lower reporting limit, to disable it replace it with ‘---’,
often the reporting limits are entered here.

186
Quantitative Analysis of Samples
Save the Calibration and the Method

Save the Calibration and the


Method

Save the Calibration and the Method

Figure 172

187
Quantitative Analysis of Samples
Quantitative Data Analysis

Quantitative Data
Analysis

Quantitative Data Analysis

Figure 173

188
Quantitative Analysis of Samples
Standard Data Files

Standard Data
Files

Standard Data Files

Figure 174

189
Quantitative Analysis of Samples
Calibration Curves

Calibration
Curves

Calibration Curves

All measurements in
Pulse counting mode

Some measurements in Pulse counting


mode, some in Analog mode

Figure 175

190
Quantitative Analysis of Samples
Examples of the Calibration Curves for “Excluded”

Examples of the Calibration Curves for


“Excluded”

Examples of the Calibration Curves for


“Excluded”
Mass used for correction
- real element

Internal Standard

Mass used for correction

Figure 176

191
Quantitative Analysis of Samples
Examples of the Calibration Curves for “Excluded”

192
Simple Sequencing (Intelligent
Sequencing Disabled)
Simple Sequencing (Intelligent Sequencing Disabled)
Sequencing

Sequencing

Sequencing

Î Associates a list of samples with ALS positions and


analytical methods
Î Allows unattended analysis of multiple samples using
multiple acquisition and reporting methods including all
calibration updates.
Î Designed to be used with an Autosampler (ASX-500 or
ASX-100), can be used in a manual mode
Î Allows automatic shutdown upon completion of the
sequence

Figure 177

194
Simple Sequencing (Intelligent Sequencing Disabled)
ASX-500 Vial Position Nomenclature

ASX-500 Vial Position


Nomenclature

ASX-500 Vial Position Nomenclature

Large bottles (tuning solution,


rinse solutions, blank)

Vial 1304 =
rack 1, column 3,
Example of 21-40-40-40 raw 04 (two digits)
configuration

Figure 178

195
Simple Sequencing (Intelligent Sequencing Disabled)
Sequencing

Sequencing

Sequencing

Î NT ChemStation Sequencing is Spreadsheet based


Î Allows shortcuts such as
Îcut,
Îcopy,
Îpaste,
Îrepeat,
Îfill down
Î Allows sample list to be inserted from other applications by
Îimporting .csv file
Îcopy and paste from other application

Figure 179

196
Simple Sequencing (Intelligent Sequencing Disabled)
Sample Log Table - Sequence Flow and Periodic Block

Sample Log Table - Sequence Flow and Periodic


Block

Sample Log Table - Sequence Flow and Periodic


Block

Figure 180

Sequencing is Modular, each functional part of the sequence is created as a


separate block such as calibration block, sample block etc. This is used more
fully by Intelligent Sequencing.

197
Simple Sequencing (Intelligent Sequencing Disabled)
Sample Log Table

Sample Log
Table

Sample Log Table

Sequence block
selection menu

Shortcut options box

Figure 181

Right Click selects shortcut options box.


Left Click or Double Click selects options for Type, Method, Dil/Lvl, Action on
Failure (Intelligent Sequencing Only), and Skip.
Vial positions increment correctly by reading ALS rack configuration when
using fill down.
Any numeric characters in Sample Name, Data File Name, or Comment Fields
will be incremented by using fill down.
To avoid auto-incrementation, use copy and paste instead of fill down.
“List Method-Cal” displays the complete path for the method on each sequence
line with it’s associated calibration file.

198
Simple Sequencing (Intelligent Sequencing Disabled)
Special Features - Keywords

Special Features -
Keywords

Special Features - Keywords

Command - a macro program


Methpath - specifies a method path (different than the normal
ICPCHEM pathway)
Overwrit - overwrites a data file without asking for confirmation
Pause - pauses a sequence
Lotsep - separates sample batches (used mostly in intelligent
sequencing
StdToExt - converts MSA calibration to external calibration
Standby - puts the instrument in Standby mode

Figure 182

Keywords are enabled by selecting ‘Keyword’ Under ‘Type’ and then selecting
the desired keyword in the Method Column.
If Keyword Command is selected, the Command or Runstring is entered in the
‘Sample’ Column.
For Example to incorporate the shutdown macro into a Sample Log Table:
Keyword Command Tune “Macro `shutdown’, go”

199
Simple Sequencing (Intelligent Sequencing Disabled)
Running a Sequence

Running a
Sequence

Running a Sequence

Data batch directory

Figure 183

Upon selecting Run Sequence, a data batch directory is automatically created in


the form YYMDDHHx.b, where:
• YY is 2 digit year
• M is month, A=Jan, B=Feb etc
• DD is day HH is hour (24 hour clock)
• x is a letter from a to z for the sequence number within a given hour.
This can be appended, deleted, or modified as needed.

200
Simple Sequencing (Intelligent Sequencing Disabled)
Chained Sequence

Chained
Sequence

Chained Sequence

Î Allows Multiple Sequences to be Run in Series


Î Allows Different Tune Conditions to be Used for Each Sequence
Îfor example Hot and Cool Plasma
Îhigh and low sensitivity etc.
Tunes to be used in a chained sequence must be created (verified)
prior to running the sequence

Figure 184

201
Simple Sequencing (Intelligent Sequencing Disabled)
Chained Sequence

Chained
Sequence

Chained Sequence

Figure 185

202
Method of Standard Additions (MSA)
Method of Standard Additions (MSA)
External Calibration

External
Calibration

External Calibration

● All Standards are prepared in one matrix


● A calibration curve is generated by plotting the intensity vs.
concentration
● Sample concentration is calculated using the slope of the
calibration curve
● Internal standards are used to correct for matrix-related
signal changes
➨ IS can be added on-line

Figure 186

204
Method of Standard Additions (MSA)
Pros and Cons of External Calibration

Pros and Cons of External


Calibration

Pros and Cons of External Calibration

Pros
Minimal sample preparation
Good accuracy for simple matrices
Cons
IS additions increases contamination potential
Limited with difficult matrices (Photoresist)
It doesn't work well with cool plasma

Figure 187

205
Method of Standard Additions (MSA)
Method of Standard Addition (MSA)

Method of Standard Addition


(MSA)

Method of Standard Addition (MSA)

● Standards are prepared in the sample matrix


● A calibration curve is generated by plotting the
intensity vs. spiked concentration
● Sample concentration is calculated by extrapolating
to the Y intercept
● Calibration can be converted into an external to run
subsequent samples of the same matrix.

Figure 188

206
Method of Standard Additions (MSA)
Pros and Cons of Method of Standard Additions

Pros and Cons of Method of Standard


Additions

Pros and Cons of Method of Standard Additions

Pros
Method of choice for ultra-trace levels
Good accuracy for all matrices
No need for internal standards
Compensate for differences in sample nebulization
and transport efficiency
Cons
Different calibrations for each matrix
All signal is accounted for
can't distinguish between true signal or
background
BEC for Ca, Fe and K can be affected

Figure 189

207
Method of Standard Additions (MSA)
Determination of Uranium in Urine by MSA

Determination of Uranium in Urine by


MSA

Determination of Uranium in Urine by MSA

Reported
concentration

Figure 190

208
Method of Standard Additions (MSA)
Converting from MSA to External Calibration

Converting from MSA to External


Calibration

Converting from MSA to External Calibration

Figure 191

209
Method of Standard Additions (MSA)
Matrix-matched Uranium in Urine External Calibration

Matrix-matched Uranium in Urine External


Calibration

Matrix-matched Uranium in Urine External


Calibration

Note corrected
concentrations

Figure 192

210
Off-line Data Analysis and Sequence
Reprocessing
Off-line Data Analysis and Sequence Reprocessing
Off-line Data Analysis

Off-line Data
Analysis

Off-line Data Analysis

Î Reviews calibration and data for currently running


sequence
Î Edits currently running data analysis method
Î Saves modifications to currently running data analysis
method (if needed)
Î Reprocesses data for currently running sequence
Î Reviews/Reprocesses previously acquired data

Figure 193

212
Off-line Data Analysis and Sequence Reprocessing
Procedure for Off-line Data Analysis

Procedure for Off-line Data


Analysis

Procedure for Off-line Data Analysis

•Start Off-line Data Analysis


•Load appropriate method
•Verify calibration data files
•Make corrections/modifications (if needed)
•Reprocess data

Figure 194

213
Off-line Data Analysis and Sequence Reprocessing
Off-line Calibration Review of Currently Running Method

Off-line Calibration Review of Currently Running


Method

Off-line Calibration Review of Currently Running


Method

Review calibration graphs:


examine linearity,
RSDs,
intercepts,
internal standard reproducibility etc.
make any corrections or changes
Select Method >> Save to Online

This procedure updates currently running method

Figure 195

214
Off-line Data Analysis and Sequence Reprocessing
Using DoList for Off-line Data Reprocessing

Using DoList for Off-line Data


Reprocessing

Using DoList for Off-line Data Reprocessing

Dolist automatically reprocesses a batch or list of data files


according to the options selected using to the currently
loaded method and calibration.

DoList does NOT update the method or calibration

Figure 196

Dolist always uses the currently loaded method, not necessarily the method
originally used to acquire the data.
Dolist does not load the method from disk or resave the method to disk when
finished. Therefore, it is possible to make temporary changes to the method for
reprocessing only (such as different report destination etc.) and not save the
changes permanently to the disk.

215
Off-line Data Analysis and Sequence Reprocessing
How to Use DoList

How to Use
DoList

How to Use DoList

Data Analysis >> Tools >> Configure DoList


Data Analysis >> Tools >> DoList

Figure 197

216
Off-line Data Analysis and Sequence Reprocessing
Selecting Files Using DoList

Selecting Files Using


DoList

Selecting Files Using DoList

Figure 198

217
Off-line Data Analysis and Sequence Reprocessing
Sequence - Reprocessing Data Batch

Sequence - Reprocessing Data


Batch

Sequence - Reprocessing Data Batch

Used to reprocess entire (or partial) sequence


Does
Load methods from disk as specified
Update calibrations in order acquired
Recalculate (Requant) data and regenerate reports
Save updated calibrations to disk

Figure 199

218
Off-line Data Analysis and Sequence Reprocessing
Sequence Reprocessing

Sequence
Reprocessing

Sequence Reprocessing

Î Top Menu>> Sequence >> Reprocess Data Batch


Î Select Data Batch
Î Reprocess Data Batch (click OK)

Figure 200

Reprocess Data Batch uses the sequence stored within the data batch directory for
reprocessing. This sequence is created at the time of analysis and is named
according to the date and time of acquisition. It is possible to modify the
sequence before reprocessing by removing undesired data files. However, care
must be taken not o remove necessary files such as calibration files or reference
files for sample types such as spikes or duplicates.

219
Off-line Data Analysis and Sequence Reprocessing
Sequence Reprocessing

220
Custom Reports and Databases
Custom Reports and Databases
What You Will Learn

What You Will


Learn

What You Will Learn

• How to create a template and generate reports


• how to create a database and update it

Figure 201

This section will introduce you to the Custom Reports package included in the
Agilent 7500 Series ChemStation Software.
Custom Reports is a windows application with three basic functions:
‰ spreadsheet
‰ database
‰ charts
Spreadsheet functions allow to easily design a report template and produce a
report for a single sample.
Custom databases contain information from many samples.
Charting of the database is useful for trend analysis and/or monitoring QA/QC
samples.
A ChemStation method can have one report template and/or one database
assigned for FullQuant analysis and one report template and/or one database
assigned for SemiQuant analysis.

222
Custom Reports and Databases
Custom Reports and Databases

Custom Reports and


Databases

Custom Reports and Databases

• Based on Visual Basic runtime


program
• When installed, creates
\icpchem\custrpt directory
• Also creates \icpchem\1\rpttmp
• Accessible through FullQuant
and SemiQuant menus
• Custom report templates are
*.fqt or *.sqt files
• Custom report databases are
*.fqd or *.sqd files

Figure 202

The objective of Custom reports is to provide an interface between the


quantitative features of the Agilent 7500 Series ChemStation and the Visual Basic
Custom Reports. The link between the two programs is provided through a
feature of the Windows environment known as Dynamic Data Exchange (DDE).
This link allows easy transfer of information from one program to another.

223
Custom Reports and Databases
Creating and Editing a Report Template

Creating and Editing a Report


Template

Creating and Editing a Report Template

• Create new report template


• Edit the current report template
• Specify new report template

Figure 203

Selecting the “New …” menu item from the Template menu will bring up the
Custom Report / Database Wizard.

224
Custom Reports and Databases
Custom Reports - Report Wizard

Custom Reports - Report


Wizard

Custom Reports - Report Wizard

• Select a Report Contents section from the list on the right.


• Select an item from the Possible Items for Report list on the left.
• Click the Add button.
• Repeat until all items and sections are added.
• Click OK.

Figure 204

The Report Wizard dialog box is used to build a report template with up to two
sections. The header section contains general information about the sample. The
All Elements section contains element specific information arranged into tables.
A plus sign next to an item indicates there are sub-items available. Double-click
on the plus sign to open the sub-item listing. The plus sign becomes a minus sign.
Double-click on the minus sign to close the sub-item listing.
Spectral graphics as well as calibration curves can be added to a custom report
template. Graphics can NOT be added to a database.
The Graphic section of the possible report items has two main subsections: Draw
Spectrum and Graph of Each Element. Items from the Graphics section can only
be added to the Header section of the Report Contents. The items from the Graph
of Each Element can be added to either section of the report.
The order of the graphics in the Report Contents listbox of the Report Wizard
determines the position in the report for the graphics. The text items are always

225
Custom Reports and Databases
Custom Reports - Report Wizard

drawn together and cannot be interspersed with the graphics through the Report
Wizard.
Press the Ctrl key and click the graphics to resize or reposition the graphics.

226
Custom Reports and Databases
Custom Reports - Drag and Drop (1)

Custom Reports - Drag and Drop


(1)

Custom Reports - Drag and Drop (1)

Figure 205

227
Custom Reports and Databases
Custom Reports - Drag and Drop (2)

Custom Reports - Drag and Drop


(2)

Custom Reports - Drag and Drop (2)

• Accessed by selecting View / Edit Box or clicking Edit Box button of


the toolbar.
• Select an item and drag it to any cell on the spreadsheet.
• Use the Next button or select from the list to view other elements.
• The current value for the highlighted item is displayed.

Figure 206

228
Custom Reports and Databases
Formatting Custom Reports

Formatting Custom
Reports

Formatting Custom Reports

• The Report Wizard formats automatically but you can


customize the format using the Format menu items.
• Highlight the cell(s) you want to format.
• Choose a menu item or click a format button on the toolbar.
• Repeat as necessary.
• Save the template

Figure 207

Column width and row height can be controlled form the Format Menu or by
using the Mouse.
Other mouse actions:
• select a group of cells
• select a row or column
• select multiple rows or columns
• select multiple, non-contiguous, single cells
• select multiple, non-contiguous, rows and columns
• select multiple, contiguous items

229
Custom Reports and Databases
Custom Reports - Printing Set-up

Custom Reports - Printing Set-


up

Custom Reports - Printing Set-up

• Controls how pages are printed.


• Controls paper size.

Figure 208

The following “Print Options”are available:


• Grid Lines - lets you print grid lines (otherwise they are visible on the
screen only)
• Black &White - prints color in black and white on a color printer
If selected, “fit to Page(s) scales the document to print a document to print on a
single or on the number of pages specified in Pages Wide, Page High.
Scale sets the percentage to reduce or enlarge the document when printed.

230
Custom Reports and Databases
Custom Reports - Saving the Template

Custom Reports - Saving the


Template

Custom Reports - Saving the Template

• Select File / Save or File / Save As


• Default file name is <method
name.fqt>; choose this or enter
alternate file name and click OK.

• Click Yes to link the template with


the current method

Figure 209

For the report template name, any legal DOS name is OK.
The default file name will have the same prefix as the currently loaded method.
Notice that all report templates end with either FQT or SQT extension.
After saving the report template the “Link With Method Dialog Box” will appear.

231
Custom Reports and Databases
Printing Custom Reports - Interactively

Printing Custom Reports -


Interactively

Printing Custom Reports - Interactively

• Load a data file.


• Select FullQuant / Print Custom Report.

Figure 210

A custom report may be interactively printed at any time using a two step process.
• First, load the data file.
• Second, select FullQuant / Print Custom Report.

232
Custom Reports and Databases
Printing Custom Reports - Printing Multiple Files [1]

Printing Custom Reports - Printing Multiple Files


[1]

Printing Custom Reports - Printing Multiple Files


[1]

Figure 211

233
Custom Reports and Databases
Printing Custom Reports - Printing Multiple Files [2]

Printing Custom Reports - Printing Multiple Files


[2]

Printing Custom Reports - Printing Multiple Files


[2]

Selected data files


will be listed in this
box.

Select files to be processed

Figure 212

Select files from the Data File listing:


• multiple continuous
• multiple discontinuous
• a single file can be removed from the “Files Selected for Processing” by
double clicking on it.
Click Add to insert those names into the “Files Selected for” Section of the panel.
Click the Process button.

234
Custom Reports and Databases
Databases

Databases

Databases

• Create new database


• Edit the current database and/or create charts
• Specify new database

Figure 213

Selecting the “New…” menu item from the Template menu will bring up the
Custom Report / Database Wizard.

235
Custom Reports and Databases
Database Wizard

Database
Wizard

Database Wizard

• Select a Database section from the list of the right.


• Select an item from the Possible Items for Database list on the left.
• Click the Add button.
• Repeat until all items and sections are added.
• Click OK.

Figure 214

Clicking the Add button will add the selected item to the Database Content list on
the right.
Clicking the Remove button will remove the selected item from the Database
Content list.
Graphics can NOT be added to a database

236
Custom Reports and Databases
Database - Drag and Drop

Database - Drag and


Drop

Database - Drag and Drop

Figure 215

• Accessed by selecting View / Edit Box or clicking Edit Box button of the
toolbar.
• Select an item and drag it to any cell in row 3.
• Use the Next button or select from the list to view other elements.
• The current value for the highlighted item is displayed.
Be aware that if you are editing a database, you can only put items from the Edit
Box into row 3. This row contains all the information that you want to keep for
each data file.
“Next” accesses the next element in the quantitation results. Elements can also be
selected by using the element selection box.

237
Custom Reports and Databases
Database - Formatting

Database -
Formatting

Database - Formatting

Figure 216

238
Custom Reports and Databases
Database - Charts

Database -
Charts

Database - Charts

• The Database Wizard formats automatically but you


can customize the format using the Chart Options.
• Save the database.

Figure 217

“X-Axis” lets you choose which items to chart on the X-Axis. Column B and C
of the database determine the X-Axis items. The default item for Column B is
Date Acquired and for Column C is Data File Name.
“Y-Axis” lets you choose which items to chart on the Y-Axis. Only numerical
values (such as mass or counts) can be used for the Y-Axis.
The Print Option lets you print the current chart or print all charts in the database.

239
Custom Reports and Databases
Global Chart Options

Global Chart
Options

Global Chart Options

Figure 218

By default, all rows in database are charted. If you enter a number N (other than
0) in the “Data Points from End to Chart” field, then only the last N rows will be
charted. This feature is useful when you have added many rows to your database,
but are only interested in the last N files. You can override this value for a single
chart using the Individual Charts Options dialog box.
Automatic Scaling - The software will automatically choose a range which allows
all of the data to be seen. The data is shown in absolute units. The range chosen
can be slightly larger than the actual data range.
Manual Scale - The manual scale allows you to control exactly what the scale on
the Y axis will be.
Normalize to Maximum - This scale allows you to chart the data as a percentage
of the maximum value (set to be 100%).
Set the First Data Point to - This scale allows you to chart all points as a
percentage of the first row of data.

240
Custom Reports and Databases
Global Chart Options

You can draw up to four control/limits lines on a chart. These lines can be
relative to the Mean or to the first value charted.
You can draw the percentage lines or Standard Deviation lines

241
Custom Reports and Databases
Database - Saving

Database -
Saving

Database - Saving

• Select File / Save or File / Save As


• Default file name is <method
name.fqd>; choose this or enter
alternate file name and click OK.

• Click Yes to link the template


with the current method

Figure 219

242
Custom Reports and Databases
Updating the Database - Interactively

Updating the Database -


Interactively

Updating the Database - Interactively

• Load a data file.


• Select FullQuant / Update Database.

Figure 220

243
Custom Reports and Databases
Update the Database - Multiple Files [1]

Update the Database - Multiple Files


[1]

Update the Database - Multiple Files [1]

Figure 221

244
Custom Reports and Databases
Update the Database - Multiple Files [2]

Update the Database - Multiple Files


[2]

Update the Database - Multiple Files [2]

Selected data files


will be listed in this
box.

• Select Tools, DoList...


• Select FullQ -- Update Database
• Select files to be processed

Figure 222

245
Custom Reports and Databases
Update the Database - Multiple Files [2]

246
Isotope Ratio Measurements
Isotope Ratio Measurements
Editing a Method for Quantitative Analysis

Editing a Method for Quantitative


Analysis

Editing a Method for Quantitative Analysis

Select ‘Edit Entire Method’

QC Parameters will only be


available if Intelligent
Sequencing is enabled

Figure 223

248
Isotope Ratio Measurements
Acquisition Modes

Acquisition
Modes

Acquisition Modes

Figure 224

Spectrum mode is the most common acquisition mode for standard applications.
• Quant
• Semiquant
Time Resolved Analysis (TRA) and Time Program (more sophisticated than
TRA) are used when a transient signal is measured.
• Electrothermal Vaporization (ETV)
• Laser Ablation (LA)
• Discrete Sampling Analysis (using ISIS)
• Chromatographic analysis (LC, GC, IC, CE)
Isotope Analysis mode is used when additional precision is needed for isotope
ratio measurements. It is similar to spectrum mode, but with 10X higher
sampling frequency.

249
Isotope Ratio Measurements
Acquisition Modes

Multitune mode is used when during a single acquisition more than one tuning
parameters are needed to accomplish the optimum performance.

250
Isotope Ratio Measurements
Acquisition Parameters for Isotopic Ratio Measurements

Acquisition Parameters for Isotopic Ratio


Measurements

Acquisition Parameters for Isotopic Ratio


Measurements

Longer acquisition time


for less abundant isotope

Usually 5 or 10 repetitions

Figure 225

251
Isotope Ratio Measurements
Report Selection

Report
Selection

Report Selection

Figure 226

252
Isotope Ratio Measurements
Setting Parameters for Isotopic Ratios

Setting Parameters for Isotopic


Ratios

Setting Parameters for Isotopic Ratios

Figure 227

253
Isotope Ratio Measurements
Example of the Isotopic Ratio Report

Example of the Isotopic Ratio


Report

Example of the Isotopic Ratio Report

Figure 228

254
Agilent ICP-MS ChemStation and
Windows Overview
Agilent ICP-MS ChemStation and Windows Overview
The Windows Interface

The Windows
Interface

The Windows Interface

Access to disk drives and folders

Folder to temporarily hold files, folders,


or objects which have been deleted.

Start button to open programs

Task Bar

Figure 229

256
Agilent ICP-MS ChemStation and Windows Overview
Windows Menus

Windows
Menus

Windows Menus

Programs - Windows NT Explorer, WordPad,


Paint, Agilent 7500 ChemStation
Documents - The last 15 documents
you have worked on
Settings - Control Panel, Printers,
Taskbar
Shutdown - Select before turning
off your computer

Figure 230

257
Agilent ICP-MS ChemStation and Windows Overview
Useful Windows Tips

Useful Windows
Tips

Useful Windows Tips

●Add Programs to the first-level


menu instead of cascading
through menus.
●Create Shortcuts by dragging

a document to the desktop with


a right-click of the mouse
button.
●Change the desktop properties

with a right-click of the mouse


button.
●Use Ctrl-Alt-Del to end a task

that is not responding.

Figure 231

258
Agilent ICP-MS ChemStation and Windows Overview
Maintaining the Computer System

Maintaining the Computer


System

Maintaining the Computer System

• Delete temporary files on a regular basis


• Use Checkdisk to check for errors on the disk
• Defragment the hard drive using a WinNT utility
• Use Virus detection software
• Create a Windows NT Emergency Repair disk

Figure 232

259
Agilent ICP-MS ChemStation and Windows Overview
Windows NT Explorer - Enhanced File Management

Windows NT Explorer - Enhanced File


Management

Windows NT Explorer - Enhanced File


Management

Click on + sign to open


sublevels in the left pane.

Click on the item to


display the contents in the
right pane as folders.

Figure 233

260
Agilent ICP-MS ChemStation and Windows Overview
Directory Structure of the Agilent ChemStation

Directory Structure of the Agilent


ChemStation

Directory Structure of the Agilent ChemStation

D:

Siclnt Icpchem Winnt

1 Custrpt Database Icpexe Icpsetup

Calib Chnseq Data 7500 Methods Qcmode Rpttemp Sequence

Icpdemo.d Std-05.d Std-10.d

Figure 234

261
Agilent ICP-MS ChemStation and Windows Overview
File Naming

File
Naming

File Naming

Files have the form:


<filename>. <extension>
<up to 8 characters> . <up to 3 characters>
Period Separator

Forbidden characters for filenames:


. / \ : <space> []| + ; , ? = * < >

Figure 235

262
Agilent ICP-MS ChemStation and Windows Overview
ChemStation File Extensions

ChemStation File
Extensions

ChemStation File extensions

.d data "file” .qcc QC mode


.m method "file” .qct QC template
.s sequence file .fqt full quant template
.chs chained sequence .fqd database template
.mac macro file .u tune file
.exe program .pa P/A file
.txt text file .prm tune parameters
.db database
.log logfile

Figure 236

263
Agilent ICP-MS ChemStation and Windows Overview
ChemStation File Extensions

264
An Overview of ICP-MS Environmental
Applications
An Overview of ICP-MS Environmental Applications
Optimizing Agilent 7500 for Environmental Samples Analysis

Optimizing Agilent 7500 for Environmental


Samples
Analysis

Optimizing Agilent 7500 for Environmental


Samples Analysis

ÎTuning
Match instrument dynamic range to analyte concentration range
ÎDetune sensitivity, especially low mass
ÎDilute samples as necessary (off-line or auto-dilute with ISIS)
ÎMinimize Effects of Sample Matrix
Î High RF power
Î Longer sampling depth ==> longer residence time
Î Appropriate use/selection of internal standards
Î Appropriate selection of isotopes

Figure 237

266
An Overview of ICP-MS Environmental Applications
Environmental Tuning

Environmental
Tuning

Environmental Tuning

Three Steps:
Initial Setup and Hardware Checkout
Optimize Physical & Plasma Parameters
Detune Sensitivity via Ion Lenses

Figure 238

Initial setup/hardware checkout


When any maintenance is executed, it is important to verify that the instrument
can meet specifications for sensitivity, precision, mass calibration, oxides and all
other parameters. This will verify that the hardware is operating correctly.
Physical/Plasma Parameters
Since typical environmental samples can contain relatively high matrix, it is
important to make tuning adjustments that improve the decomposition of the
matrix, such as increasing the plasma temperature or increasing the residence time
of the sample in the plasma.
Detune Sensitivity via ion lenses
Typical compositions of environmental samples include higher concentrations of
low mass elements such as Na, K, Ca and Mg, and lower concentrations of mid
and high mass elements such as Se and Hg.

267
An Overview of ICP-MS Environmental Applications
Environmental Tuning

To analyze all the elements at the same time, detuning the low mass only can be
an effective technique. However, care must be taken in this process if analysis of
Be is required.
Tips
Conditioning of the interface is recommended to improve signal stability, after
cleaning the cones or extraction lenses. For conditioning, run tap water through
the system for 1/2 hour during tune before finalizing tune conditions and prior to
calibration and sample analysis.

268
An Overview of ICP-MS Environmental Applications
Three Goals of Environmental Tuning

Three Goals of Environmental


Tuning

Three Goals of Environmental Tuning

Optimize aerosol formation


Maximize analyte exposure to plasma temperature
Minimize matrix exposure to mass spectrometer components

Figure 239

Increasing sampling depth increases the sample residence time in the plasma. The
effect is to allow more time for decomposition (atomization and ionization) of the
analyte elements. Each nebulizer has optimal values for carrier gas flow rate and
sample flow rate. In general, higher carrier gas flow rates create higher carrier
gas pressures thereby generating finer droplets, which lead to better instrument
sensitivity. However, excessive carrier gas flows cool the plasma, decreasing the
sensitivity and increasing the ratios of oxides and doubly-charged ions
significantly. Excessively high carrier gas flows cool the plasma which increases
low mass (e.g. Li) sensitivity and noise to a non-acceptable level. Lithium signal
should not exceed Yttrium signal in a well tuned system.
Oxides are almost completely controlled by the interaction of four parameters,
spray chamber temperature, sample depth, carrier gas flow and peri-pump flow.
Since we have chosen to maximize sample depth for other reasons and spray
chamber temp should normally be set to 2 C. we must control oxides with carrier
gas flow and peri-pump flow. The goal here is to maximize the efficiency of the
particular nebulizer being used (smallest droplet size and size distribution),
without increasing either flow to the point of over-cooling the plasma.

269
An Overview of ICP-MS Environmental Applications
Three Goals of Environmental Tuning

Decreasing the negative voltage on the two extract lenses decreases the number of
ions drawn into the mass analyzer. This decreases the sensitivity of the
instrument. It also decreases the ion load on the rest of the mass analyzer which
minimizes the need to clean the other lenses and components. It is important that
the correct voltage gradient between the two lenses be maintained. As such, Ext1
should be set and then Ext2 fine tuned to give the best results. If large changes in
Ext1 voltage have little to no effect on sensitivity, this is an indication that the
extract lenses need to be cleaned.

270
An Overview of ICP-MS Environmental Applications
Tuning Flow Chart

Tuning Flow
Chart

Tuning Flow Chart

Tuning the Agilent 7500 ICP-MS


for Environmental Applications
Initial Setup/Hardware Checkout

Initial targets: Li > 8k


Initial Setup Initially, all tune parameters Y > 12 k
should be set according to Tl > 8 k
New Instrument or after Major System
the recommended values FAIL
Maintenance all RSDs < 10%
on table 1 (shaded) oxides < 1%

pass
Then...
begin the process of
Physical/Plasma Parameters environmental tuning

Set RF Power Set Carrier Gas Flow


1350-1500 Watts 1.15-1.2 L/min.
Adjust Sampling Depth Set Peri-pump
(monitor reflected power Monitor oxides; it should
8-9 mm (1) 0.08-0.1 RPS
and adjust if necessary with

See Appendix 4 RF Matching


not exceed 0.5% (2)

1. Decrease carrier (or


Oxides < 0.5 %
blend) gas flow
Li counts > 5000 No 2. Increase or decrease
(int. time = 0.1 sec)
peri-pump flow (3)

Yes

Run “Set_EM!”" Check Axis/Resolution


Run Torch Position from the Tune Menu Peak widths
Autotune to adjust all detector pass 0.75 at 10 % +/-0.05
parameters mass calib.:
nominal mass +/- 0.05
pass

FAIL FAIL

Detune Sensitivity via ion lenses--- target: low mass sensitivity is reduced in sensitivity by 10x
compared to mid and high mass.
. DO NOT ADJUST:
Increase both Omega +
Set Ext1 and 2 Tune Einzel 1/3, and 2 and QP Focus. Typical Any AMU settings
at low voltage, especially No Typical values are -60V values are 6 V and 9V, Any Axis settings
No
Ext2, which improves and 14V, respectively. respectively. Discriminator
stability below -35V.(4) Adjustments of these EM voltage
lenses improve signal Last Dynode
stability P/A factors

Achieve Goals Achieve Goals

Figure 240

271
An Overview of ICP-MS Environmental Applications
Recommendations on Interference Equations

Recommendations on Interference
Equations

Recommendations on Interference Equations

Generally Use Equations Specified in EPA Method 200.8

May require slight adjustment based on measured values

Figure 241

Arsenic
Since there are lots of polyatomic ions are generated around the mass, it is very
difficult to apply universal equation.
When the contribution of 82Kr is considered, the equation should be changed as
follows:
As(75) = (1.000)(75C) - (3.127)(77C) + (2.736)(82C) - (2.760)(83C)
However, an unknown peak sometimes appears at mass 83. BrH is also generated
at mass 82. As a result, the contribution of Kr is overcorrected, and As might
show a negative result.
Selenium
Generally, the use of 82 rather than 77 or 78 is recommended. 77 is interfered
with by 40Ar37Cl, and 78 is interfered with by 38Ar40Ar dimer. The 82 isotope
needs to be corrected for the possible presence of 82Kr in the Argon gas supply
using the following equation.

272
An Overview of ICP-MS Environmental Applications
Recommendations on Interference Equations

Se(82) = M(82) - 11.6/11.5 * Kr(83)


= M(82) - 1.0087 * Kr(83)
However, as described the above, the unknown peak might appear at mass 83,
thereby a portion of the signal at mass 83 is used practically. An example is as
follows:
Se(82) = M(82) - 0.6 * Kr(83)
The monitoring of mass 77 and 78 is also recommended just in case.

273
An Overview of ICP-MS Environmental Applications
More Interference Corrections

More Interference
Corrections

More Interference Corrections

Some correction factors are based on formation of doubly


charged species or oxides rather than on isotope ratios and
need to be updated periodically.

• Iron-54
• Ca-44

Figure 242

Vanadium
The interference correction might not be useful. Because it is almost redundant
(1% NaCl gives about 1ppb ClO equivalent) and can lead to problems when high
Cr is present.
Iron
The use of 54 rather than 56 or 57 is recommended. Since mass 54 is interfered
with ArN, the concentration of HNO3 should be the same. If the concentration of
HNO3 cannot be controlled, the following equation would be useful:
Fe(54) = M(54) - ratio of 54/15 * M(15)
In this case, a blank solution must be analyzed at first, and mass 15 and 54 should
be measured. The ratio of 54/15 will be calculated, and this ratio will be entered
into the equation.
CaN is unlikely to give an interference at 200 ppm Ca carbonate.
Calcium

274
An Overview of ICP-MS Environmental Applications
More Interference Corrections

Ca is normally included at very high concentration; therefore, there is no


interference on it.
However, when the mixed standard solution which contains the same
concentration of Ca and Sr, the apparent Ca44 will be almost double if the system
was calibrated in the absence of Sr. The same problem occurs with Ca43. In this
case, the following equation is useful:
Ca(44) = M(44) - ratio of Sr2+/Sr+ * M(88)
Sr solution must be analyzed at first to get the ratio of Sr2+/Sr+.

275
An Overview of ICP-MS Environmental Applications
Calibration Standards

Calibration
Standards

Calibration Standards

Calibrate no higher than necessary


Minimize memory effects and the detector wear
• typical ranges for ‘trace elements’: 0, 1, 10, 100 ppb
• major elements such as Na, K, Ca, Mg, Fe: 0, 100,
1000, 10,000 ppb
• Hg: 0, 0.1, 0.5, 1, 5 ppb
Major and trace elements are available in a single solution
with their concentration in 100:1 ratios, respectively
Mercury supplied in a separate solution
Addition of 100 ppb gold to ALL solutions required for
mercury determination

Figure 243

276
An Overview of ICP-MS Environmental Applications
Linear Range Determination

Linear Range
Determination

Linear Range Determination

‘Practical Linear Range’


Reasonable upper range of analyte concentrations after
sample dilution to avoid excessive carryover after
running LRS

Typically:
z ~1 ppm for most elements,
z 10-50 ppm for ‘minerals’,
z < 500 ppb for Ag, Mo, Sb, Tl, Pb and other
memory prone elements,
z < 20 ppb for Hg!

Much lower than actual Instrument Linear Range

Figure 244

277
An Overview of ICP-MS Environmental Applications
Interference Check Samples

Interference Check
Samples

Interference Check Samples

Purpose is to test effectiveness of plasma conditions and


correction equations:

High concentrations of ‘Interfering Elements’


Trace concentrations (~20 ppb) of analyte elements
Only the analyte elements need be measured

Figure 245

278
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [1]

Troubleshooting Environmental Applications


[1]

Troubleshooting Environmental Applications [1]

High Tune % RSDs

Non-linear Calibrations

High Analysis % RSDs

Poor Agreement Between Calibration and ICV

Figure 246

High RSDs During Tune


Incorrectly tuned plasma parameters such as carrier gas or blend gas flow,
peri-pump speed or sample depth.
Dirty cones
Worn peri-pump tubing. Incorrect shoe pressure on peri-pump. Should be
just tight enough to insure a smooth flow of sample (aspirate a bubble and
watch its progress through the line).
Non-Linear Calibrations
ICP-MS is a linear technique. Pulse mode calibrations should always be
linear, if not, suspect standard preparation errors or possible incompatibilities
among elements in multi-element standards.
Non-linearity between pulse and analog mode indicates incorrectly set P/A
factors or possible worn out detector.
High %RSDs During Analysis

279
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [1]

Insufficient uptake, rinse-out or stabilization time. Use ‘Edit Average File’ >
‘Tabulate’ to examine individual replicates for upward or downward trends.
Insufficient signal counts.
May be also be caused by worn peri-pump tubing or bubbles in either the
sample uptake or internal standard uptake tubing. Check the connections at
the ISTD addition “Y” and replace the peri-pump tubing.

280
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [2]

Troubleshooting Environmental Applications


[2]

Troubleshooting Environmental Applications [2]

Poor Analyte Recovery in Spiked Samples for Selected Analytes

Carry-over or Memory Effects for Certain Elements

Figure 247

Poor recovery for selected analytes in spikes.


Several conditions can cause poor recovery of certain analytes in spiked
samples. Ag is especially insoluble in the presence of even trace levels of Cl-
, therefore the use of HCl should be avoided whenever possible. Several
elements (Zn, As, Se, Cd) have relatively high first ionization potentials and
may not be as effectively ionized in samples with high concentrations of
easily-ionizable elements such as Na and K. Diluting the sample if possible,
or selecting an alternative internal standard with a higher ionization potential
may help. Possible alternative internal standards include Ge, Te, and Au.
Carryover or memory interference.
Several elements are prone to memory effects for various reasons. Ag, Mo
and Tl tend to stick to surfaces in the sample introduction system and slowly
rinse into subsequent samples. Keeping the sample introduction system
(sample tubing, peri pump tubing, nebulizer, spray chamber, torch and cones)
clean will help minimize carry-over. Also, rinsing between samples with
relatively high acid concentration rinse blanks (ca. 5% HNO3) will help. If

281
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [2]

possible, avoid introducing samples or standards with concentrations of these


elements above a few hundred ppb. Use of the Babington nebulizer should
also reduce carryover of these elements. Li, when analyzed for extended
periods of time or in very high concentrations tends to accumulate on the
back sides of the interface cones. Cleaning the cones will usually reduce Li
background and carryover. Volatile elements, or elements with volatile
hydrides such as Hg and Sb can also carryover due to off-gassing from
droplets on the spray chamber walls. Steps to reduce the volatility of these
species are helpful.

282
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [3]

Troubleshooting Environmental Applications


[3]

Troubleshooting Environmental Applications [3]

Calibration Drift Over Time

Poor Internal Standard Recoveries in Certain Samples

Figure 248

Calibration Drift over time.


Insure that the instrument is adequately warmed up before initial calibration
(warm-up, while scanning in tune for 15 minutes).
Insure that laboratory temperature does not vary by more than 3 degrees C.
per hour.
Check cones for signs of sample deposits which may be affecting the size
and shape of the cone orifices. Clean if necessary.
Check peri-pump tubing for signs of excessive wear or flattening.
Run SetEM. An electron multiplier which is near the end of its useful life
may be changing in response over short periods of time. If running SetEM
significantly changes the EM voltage from day to day, the EM should be
replaced.
Poor internal standard recoveries in samples.

283
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [3]

Reduction in internal standard signal is usually caused by high matrix


concentration in samples (especially Na, and K). Dilute the samples. It may
also be desirable to tune the instrument with a matrix matched tune solution
containing appropriate levels of the matrix elements to minimize the effect of
any matrix induced suppression of ionization.

284
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [4]

Troubleshooting Environmental Applications


[4]

Troubleshooting Environmental Applications [4]

High Relative Standard Deviations (%RSDs) in


Samples
Standards

High %RSDs in Tune

Figure 249

High relative standard deviations (RSDs) for analyte or internal standard


elements during sample analysis.
Usually caused by insufficient sample uptake or stabilization time. May be
also be caused by worn peri-pump tubing or bubbles in either the sample
uptake or internal standard uptake tubing. Check the connections at the
ISTD addition “Y” and replace the peri-pump tubing. Shoe pressure on the
peri-pump should be just tight enough to insure a smooth flow of sample
(aspirate a bubble and watch its progress through the line).
High RSDs during tune.
Incorrectly tuned plasma parameters such as carrier gas or blend gas flow,
peri-pump speed or sample depth.
Dirty cones.Worn peri-pump tubing.
Incorrect shoe pressure on peri-pump. Should be just tight enough to insure a
smooth flow of sample (aspirate a bubble and watch its progress through the
line).

285
An Overview of ICP-MS Environmental Applications
Troubleshooting Environmental Applications [4]

Omega lens settings can also affect tune precision as can parameters which
affect mass peak shape such as plate and pole bias.

286
Semiconductor Applications of ICP-MS
and Advantages of Agilent 7500s System
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Chemicals and Materials Used in Semiconductor Industry

Chemicals and Materials Used in Semiconductor


Industry

Chemicals and Materials Used in Semiconductor


Industry

HCl
HF
Etching
H2SO4 IPA
Si, GaAs
Wafer substrate
Washing
HNO3 Cleaning TMAH SiH4, TEOS, NF3, N2
Gases

Photoresists - Lithography
H2O2 Methanol
Pure Metals - Sputter Targets

Figure 250

BPSG = boron phosphorus silicon glass used for doping.


TEOS = tetra ethoxy silane (used for depositing SiO2 layers.
TMAH = tetra methyl ammonium hydroxide.
Multielement analysis and ultra low detection limits have made ICP-MS the
technique of choice for the determination of metallic impurities in the industry.

288
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Metals Analysis in the Semiconductor Industry - Customer Groups
and Requirements

Metals Analysis in the Semiconductor Industry -


Customer Groups and
Requirements

Metals Analysis in the Semiconductor Industry -


Customer Groups and Requirements

● Chemical manufacturers
➨ need for QC analysis of products

■ inorganic chemicals
■ organic chemicals
● Wafer manufacturers
➨ characterize bulk polysilicon

➨ monitor contamination control in wet stations

➨ analysis of chemicals and various mixes

➨ wafer surface characterization

● Device manufacturers

➨ monitor contamination control in wet stations

■ analysis of chemicals and various mixes


■ photoresist\stripper analysis

Figure 251

289
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
ShieldTorch Interface

ShieldTorch
Interface

ShieldTorch Interface

● Developed by HP/Yokogawa in 1992


➨ led to widespread use of "cool" plasmas
● Prevents secondary discharge in the ICP-MS interface
➨ virtually eliminates many polyatomic interferences
➨ Shield plate design assures complete discharge removal
● Enables the determination of Fe, Ca and K at low ppt levels
➨ 3 orders of magnitude improvement in detection limits
● Improved DLs for most other SEMI elements
➨ dramatically reduced interface background
■ sub-ppt DLs for Na
● Operates at higher power levels - up to 1100W
➨ other systems must be operated at 500-600W
● No consumables
➨ Shield plate lasts indefinitely - never has to be removed

Figure 252

The Agilent 7500 has no loss in sensitivity when switching to cool plasma - Other
instruments lose up to 95% of their sensitivity for all Transition metals.
Other Instruments have to run at 600W to reduce ArO - Agilent 7500 can run at
1100W, which means higher sensitivity, lower oxides, and lower matrix effects,
and also analyze As, Se.

290
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
ShieldTorch Interface

ShieldTorch
Interface

ShieldTorch Interface

Figure 253

291
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Normal and “Cool” Plasmas

Normal and “Cool”


Plasmas

Normal and “cool” Plasmas

Normal Plasma Cool Plasma


Forward Power at 1300W Forward Power at 900W
Sampling Depth at 7mm Sampling Depth at 13mm
Carrier Gas Flow at 1.2lpm Carrier Gas Flow at 1.7lpm

Figure 254

Shield Plate removes potential difference between plasma and interface, so no


polyatomic ions form behind the sample cone. Cool central channel of plasma
gives low Ar and Ar-based ion populations. Shield Plate can be used at high
powers.
High temperature gives good ionization and matrix tolerance, but high population
of Ar and Ar-based polyatomic species form in the plasma and behind the sample
cone, due to potential difference between plasma and interface.

292
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Shield Torch “Cool Plasma”

Shield Torch “Cool


Plasma”

Shield Torch “Cool Plasma”

Poorly Grounded Shield Plate

No polyatomic ion
formation

No secondary Potential difference between plasma


discharge and interface gives pinch discharge.
Plasma must be operated at
600W to remove polyatomics.
N o Shield Plate
Pseudo Cool Plasma achieved using
alternative load coil geometry

Figure 255

293
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Shield Torch Installation

Shield Torch
Installation

Shield Torch Installation

The shield should cover the coil


and the bonnet should cover the
shield
Bonnet
C oil

The coil should not be exposed


to the torch (plasma)

Shield

The shield should not be


exposed to the coil
Figure 256

294
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Cool Plasma Tuning

Cool Plasma
Tuning

Cool Plasma Tuning

● Reduce RF Power to 900W


● Increase blend gas to 0.5L/min
● Increase sampling depth to 13mm
● Minimize the ArO using x,y torch position
● Minimize the ArH using the sampling depth
● Optimize using the blend gas

Figure 257

295
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Advantages of Cool Plasma at Higher Power (900 - 1100 W)

Advantages of Cool Plasma at Higher Power (900 -


1100
W)

Advantages of Cool Plasma at Higher Power


(900 - 1100 W)

● No loss of sensitivity compared to normal plasma conditions


➨ Li - 150 Mcps/ppm

➨ Fe - 40 Mcps/ppm
● Higher ionizing power - greatly expanded analyte range

➨ many analytical requirements can be performed in one run

● Ability to analyze high matrix samples

➨ organics, including Photoresists

➨ H PO
3 4
➨ Si matrices

●Meet/exceed HR-ICP-MS performance with new applications

➨ ppt determination of As and V in 5% HCl

➨ low ppt determination of Cr, Mg in organics

Figure 258

296
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Advantages of Cool Plasma at Lower Power (700-800 W)

Advantages of Cool Plasma at Lower Power (700-


800
W)

Advantages of Cool Plasma at Lower Power


(700-800 W)

● Minimal loss of sensitivity compared to "hotter" plasma conditions


➨ Li -100 Mcps/ppm

➨ Fe - 20 Mcps/ppm

● Greatly improves the BEC for Ca, Fe and K

➨ Specially for the analysis of DI Water, H O , and diluted acids


2 2

Figure 259

297
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Detection Limits Study [1]

Detection Limits Study


[1]

Detection Limits Study [1]

50
Average = 10.50 StdDev = 4.74
40

30
3 sigm a
20 2 sigm a
1 sigm a
10 A verage

0
1 2 3 4 5 6 7 8 9 10

D L = 3*StdD ev blank

Figure 260

298
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Detection Limits Study [2]

Detection Limits Study


[2]

Detection Limits Study [2]

50
(X2,Y2)

40
In te n sity
30

20

10
(X1,Y1)

0
0 5 10
C o n ce n tra tion

Figure 261

299
Semiconductor Applications of ICP-MS and Advantages of Agilent 7500s System
Automatic Switching Between Normal and Cool Plasma

Automatic Switching Between Normal and Cool


Plasma

Automatic Switching Between Normal and Cool


Plasma

Chained Sequencing allows the user to run a set of samples


under one set of conditions such as cool plasma, then
automatically switch the conditions and rerun the same samples
under hot plasma.

Multitune Mode allows the user to run the same sample under
multiple sets of conditions, automatically switching from one
mode to another and then moving to the next sample

Figure 262

300
Intelligent Sequence Training Text
Intelligent Sequence Training Text
What is Intelligent Sequence?

What is Intelligent Sequence?


Intelligent Sequence is designed to dramatically improve sample throughput and
reporting accuracy while decreasing the time spent by the operator and data
reviewer. Intelligent Sequence automates a variety of QA/QC control procedures
as well as all the EPA-mandated quality control procedures (EPA Methods 200.8,
6020 and 6020-CLP) from analysis through reporting.

Features
Intelligent sequence has several unique features.

Smart Sequencing
Intelligent Sequence recognizes all EPA designated QA/QC sample types and,
when used with an autosampler, allows automatic, unattended analysis of batches
of samples with all necessary calibrations, checks and controls. During
sequencing, sample results are evaluated for pass/fail against a user-editable
database of QC criteria. If a QC parameter is out of range, sequencing
automatically performs a user-selectable action to attempt to remedy the problem.
All sample results and QC actions are logged and a QC exception report is
created.

QC Reporting
During sequencing, sample-type specific reports are generated and stored for all
runs. A QC summary report is also generated. This report can be viewed at any
time during or after the sequence. The summary report includes, in an easily
reviewed format, a list of samples run and any QC failures which may have
occurred. All batch or Sample Delivery Group relevant data are automatically
stored together in system-generated “batch directories” for convenient archival
and retrieval.

Tune Compliance Checking


A simple, graphical user interface allows user selectable tune compliance criteria
to be stored with each method. Evaluating a tune sample can be automatic or
manual via a simple pull-down menu item.

302
Intelligent Sequence Training Text
Typical Analytical Flow

Typical Analytical Flow


1) Select appropriate QC configuration Configuration >> QC Mode
2) Load ICP-MS method Methods >> Load
3) Edit method and QC parameters Methods >> Edit Entire Method
• Method Information
• Interference Equation
• Acquisition
• Data Analysis
‰ Report… Select QC report
‰ Calibration Table… Select Load Mass from Current Acq. in New
• QC Parameters
‰ QC Database
4) Set unreported elements Methods >> Set Unreported Elements
5) Save the method and calibration Methods >> Save
• Verify QC Method
6) Load the tuning method Methods >> Load
7) Edit method and QC parameters Methods >> Edit Entire Method
• Method Information
• Acquisition
• Data Analysis
‰ Report… Select QC report
• QC Parameters
• QC Tune Criteria
8) Save the method and calibration Methods >> Save
9) Edit the sample log table Sequence >> Edit Sample Log Table
10) Simulate the sequence Sequence >> Simulate Sequence
11) Save the sequence Sequence >> Save
12) Run the sequence Sequence >> Run

303
Intelligent Sequence Training Text
Using Intelligent Sequencing

Using Intelligent Sequencing

QC Configuration
To use Intelligent Sequencing, the appropriate QC configuration must be selected
in Configuration.

The following QC modes are provided as default.


• EPA2008.QCC --- intelligent functions defined by EPA Method 200.8.
• EPA6020.QCC --- intelligent functions defined by EPA Method 6020.
• EPA6020C.QCC --- intelligent functions defined by EPA Method 6020
CLP.
• EPAGEN.QCC --- Intelligent functions designed to meet both EPA 200.8
and 6020 requirements
• GENERAL.QCC --- Intelligent sequencing disabled (no expected values
or failure actions defined)
When selecting GENERAL.QCC, the sequencing mode is changed to the general
sequencing mode. When selecting other QC modes, the sequencing mode is
changed to intelligent sequencing mode.

304
Intelligent Sequence Training Text
Using Intelligent Sequencing

For more information about sample types included in each QC mode, see
Appendix of “AGILENT 7500 ChemStation Intelligent Sequence Manual”.

ICP-MS Method
When intelligent sequencing mode is selected, additional menus become available
in the Methods menu.

• Edit QC Database… Enables editing of QC related items on a sample


type basis such as high limit values, low limit values and error action.
• Edit QC Tune Criteria… Enables editing of tune compliance criteria such
as sensitivity, mass resolution and %RSD.
• Verify QC Method… Checks the QC database and QC tune criteria for
configuration errors.
• QC Information… Indicates the QC mode name which was used to make
the current method.
• Set Unreported Elements… Allows the user to select the elements which
are not to be reported on QC custom reports.

Default Methods Provided

305
Intelligent Sequence Training Text
Using Intelligent Sequencing

EPA Method 200.8 6020 CLP Remark

QC Mode (.qcc) 200_8 6020 6020CLP

Method (.m) EPA20_8 EPA6020 EPACLP For QC Database

Tuning Method (.m) TN200-8 TN6020 TNCLP For QC Tune Criteria

Edit Entire Method


Use Edit Entire Method to make a complete method including QC parameters.

• Method Information… Same settings as general use.


• Interference Equation… Same settings as general use.
• Acquisition… Same settings as general use.

306
Intelligent Sequence Training Text
Using Intelligent Sequencing

• Data Analysis… Same except Select Reports.

Select “QC Report”, and


then “Recalculate then
Print”

• QC Parameters… New menu


(A) QC Database
(B) QC Tune Criteria

307
Intelligent Sequence Training Text
Using Intelligent Sequencing

(A) QC Database
For information about how to read comparative expressions (definition of
criteria), see “Setting Up a Method” of “AGILENT 7500 ChemStation Intelligent
Sequence Manual”.

QC Database settings are


saved in cal file

Elements referred
from cal curve Comparative expressions
defined in QC configuration

Sample type
opened currently

Expected Value and


Low & High Limits

Error Actions and #Allowed


Failures

Click here

Off(x) is marked for the elements


for which Excluded is selected as
a curve fit in cal, or used for
interference correction

<Error Actions supported>


• NextSmpl… Next sample
• Abort… Abort the run

308
Intelligent Sequence Training Text
Using Intelligent Sequencing

• Blk (Abort) – NextSmpl… Run the blank block then continue (Abort if
all of the samples in the blank block fail).
• Blk (Cont.) – NextSmpl… Run the blank block then continue (Continue
even if all of the samples in the blank block fail)
• Blk (Abort) – SameSmpl… Run the blank block and re-run same sample
(Abort if all of the samples in the blank block fail).
• Blk (Cont.) – SameSmpl… Run the blank block and re-run same sample
(Continue even if all of the samples in the blank block fail).
• Cal – SameSmpl… Recalibrate and re-run same sample.
• Cal – AllSmpls… Recalibrate and re-run all samples since last CCV
block.
• NextLot… Run next lot of samples
• Run User Macro… Run the user macro which must be placed under the
method currently running, and named “QCUSER.MAC”.

<How proceed Criteria>


Up to four criteria for each sample type (except ISTD).

Area for 1st step Criteria 1

Area for 1st step Criteria 2

Area for 2nd step Criteria 1

Area for 2nd step Criteria 2

309
Intelligent Sequence Training Text
Using Intelligent Sequencing
Fail
st Error action set for
1 step, Criteria 1
1st step Criteria 1

Pass Check to select flag

Fail
st Error action set for
1 step, Criteria 2
1st step Criteria 2

Pass

FILTER

Fail
nd Error action set for
2 step, Criteria 1
2nd step Criteria 1

Pass Check to select flag

Fail
nd Error action set for
2 step, Criteria 2
2nd step Criteria 2

Pass

Next sample

Normally just one Criteria (1st step Criteria 1), or two Criteria (1st step Criteria 1
& 2nd step Criteria 1, or 1st step Criteria 1&1st step Criteria 2) is used.
<Examples>
• 1st step Criteria 1
“Check whether the analytical concentrations of certified reference material
(CRM) are 90-110% of the certified values (expected values).”

1st step Criteria 1


Low limit: CRM conc >= 0.9 x CRM expected values
High limit: CRM conc =< 1.1 x CRM expected values

• 1st step Criteria 1 & 2nd step Criteria 1

310
Intelligent Sequence Training Text
Using Intelligent Sequencing

“Check whether the analytical concentrations of CRM are 90-110% of the


certified values for elements whose counts are equal or more than 1000 cps”

1st step Criteria 1


Low limit: CRM cps >= 1000
High limit: none

2nd step Criteria 1


Low limit: CRM conc >= 0.9 x CRM expected values
High limit: CRM conc =< 1.1 x CRM expected values
• 1st step Criteria 1 & 1st step Criteria 2
“Check whether the analytical concentrations of certified reference material
(CRM) are 85-115% of the certified values. If not, recalibrate and then
analyze again. Also check whether concentrations are 90-110% of the
certified values. If not, have a error flag on a report.”

1st step Criteria 1


Low limit: CRM conc >= 0.85 x CRM expected values
High limit: CRM conc =< 1.15 x CRM expected values

1st step Criteria 2


Low limit: CRM conc >= 0.9 x CRM expected values
High limit: CRM conc =< 1.1 x CRM expected values

311
Intelligent Sequence Training Text
Using Intelligent Sequencing

(B) QC Tune Criteria

QC Tune Criteria settings are


saved in method file

Masses set in data


acq parameters

Set “Ref” for Fixed to 1 for Resp


Response Ratio Ratio Low&High Limits
check *1

Set “Bkg” for Only Max. Bkg


Bkg check *2 Count is effective

*1
: “Ref” appears in the Mode list when Response Ratio check is enabled.
*2
: “Bkg” appears in the Mode list when Max Bkg Count check is enabled.

312
Intelligent Sequence Training Text
Using Intelligent Sequencing

Notes on Setting a Method


• Use Auto Configuration first, when there are more Off (x) elements other
than the elements for which Excluded is selected as a curve fit in the
calibration table and the elements with parentheses. When Auto
Configuration is used, all Off column settings are always reinitialized.
• In the event of a simultaneous QC failure and ISTD failure, the action on
QC failure will take precedence.
• The error action for third failure is set in QC configuration on a sample
type basis.
• Set equal or more number of acquired elements for #Allowed QC Failure
when the Second Step is used. Otherwise, the First Step does not work as
a filter, and QC check won’t go to the Second Step.
• Expected values for Reference part (right side of comparative expressions)
always refer to the values set in the First Step Criteria 1. Expected values
for Measured part (left side of comparative expressions) refer to the
values set in each criteria.
• Use Set Unreported Elements, when there are elements which are not to
be reported on QC custom reports.

Before Saving the Method


• Use Verify QC Method to check whether there is any error on settings.
(Click this in Method Save Options when method is saved.)

Setting Up a Sequence
Structure of Sample Log Table
Sample Log Table is arranged in subroutines or “Blocks”.
The Sample Log Table is composed of four kinds of sheets as follows:
• Sequence Flow sheet
• Periodic Block sheet
• Individual Block sheet
• Whole List sheet

313
Intelligent Sequence Training Text
Using Intelligent Sequencing
Periodic Block

Analyzed every
10 samples

Analyzed every
8 hours (480
Main flow

TUNE Block
CALIB Block Individual Block – CALIB Block
SMPL Block
TERM Block

Whole List Keywords which


separate each block are
automativcally attached

Notes on Setting Each Sheet


<Periodic Block>
The upper block setting has priority when there is a conflict with multiple blocks.
<Individual Block>

314
Intelligent Sequence Training Text
Using Intelligent Sequencing

• Blanks which are analyzed when the error actions “Blk…….” are taken
should be set in BLANK Block.
• Samples which are analyzed when the sequence is aborted should be set in
ERRTERM Block.
• When ISTD check is set, the CalBlk should be set at first except Tune
since the ISTD counts from the CalBlk are used to establish the reference
values.
• When recovery, dilution or duplication check is set, a reference sample
type should be set prior to recovery, dilution or duplication sample.

Before Running the Sequence


• Use Simulate Sequence to check whether there is any error on settings.

Running a Sequence

The methods & calibrations used, actual


sample log table and sequence logs as well
as all the data in one sequence are saved in
this directory.

Note on Running a Sequence


• Insure the same QC configuration as the one used when the methods and
sequence to be executed were made.
• After editing the sample log table (online) during sequence run, close the
table immediately. The sequence is paused while the sample log table is
opened.

315
Intelligent Sequence Training Text
Using Intelligent Sequencing

• If the data analysis parameters, QC Database or QC Tune Criteria in the


method currently used needs to be changed instantly, load the method
using Offline Data Analysis, and change it. The modified method can be
saved to online using Offline Data Analysis >> Method >> Save to
Online.

Sequence Reprocessing of Data


Two ways for batch reprocessing of data:
A. Reprocessing Data Batch Directory
B. Running a Sequence with Reprocessing Only

(A) Reprocess Data Batch Directory


<Files to be selected and used>
1. Select data batch directory to be
reprocessed in Reprocess Data Batch
Directory

icpchem\1\ data\ 98k1109a.b\ 001tune.d

002calb.d Data files

050smpl.d

meth1.m
(copied automatically)
Used for data
meth1.c
reprocessing
(copied automatically)

result.s
(actual sequence*)

methods\ meth1.m
(original)

calib\ meth1.c
(original)

• The name for actual sequence is always result.s.


(B) Running a Sequence with Reprocessing Only
<Files to be selected and used>

316
Intelligent Sequence Training Text
Using Intelligent Sequencing

3. Select data batch directory to be


reprocessed in Start Sequence Dialog
Box

icpchem\1\ data\ 98k1109a.b\ 001tune.d

002calb.d Data files

050smpl.d

meth1.m
(copied automatically)

meth1.c
(copied automatically)

result.s 1. Load this sequence


(actual sequence*)

methods\ meth1.m
(original)
Used for data
calib\ meth1.c reprocessing
(original)

2. Insure the path for currently loaded method is the


same as the one used for data reprocessing

* The name for actual sequence is always result.s.

Notes on Reprocessing Data


• Insure the correct method and calibration file are selected when changing
parameters for reprocessing; change the copied method and calibration in
the data batch directory when using Reprocess Data Batch Directory, and
change the original method and calibration when running the sequence
with reprocessing only.
• The header information updated using Data Analysis >> File >> Edit
Header is not reflected when using batch reprocessing (both ways).
Change the Sample Log Table of result.s if needed. However, the

317
Intelligent Sequence Training Text
Using Intelligent Sequencing

header information to be printed out is not changed until the header


information is updated using Data Analysis >> File >> Edit Header.

318
Intelligent Sequence Training Text
Setting Up a QC Configuration

Setting Up a QC Configuration
QC configuration defines the QC sample type set which contains…
• Sample type name
• Data name suffix
• QC report template
• Type category
• QC Item name to be used for ISTD check
• Error counting way
• Action on 3rd failure
• Comparative expressions
• Error flags for QC report
The Changes in QC Configuration must be implemented with CAUTION as it
affects the settings in the QC Database, QC Tune Criteria, or Sample Log Table
directly.

319
Intelligent Sequence Training Text
Setting Up a QC Configuration

Configuring QC Items

320
Intelligent Sequence Training Text
Setting Up a QC Configuration

How to Create or Change a QC Item

Select “List” to
load QC Item
to be changed

Screen for selected


QC Item is opened

Select “New” Select “Delete” to delete the


to make new QC Item currently opened
CAUTION: once deleted, it
cannot be recovered

321
Intelligent Sequence Training Text
Setting Up a QC Configuration

Special Screens

Screen is completely
changed

Only QC Item
becomes
effective

322
Intelligent Sequence Training Text
Setting Up a QC Configuration

Notes on Setting QC Items


• Preservation of the original QC mode files is recommended.
• Once a QC Item is deleted, it will change the setting of related items in the
QC Database, QC Tune Criteria, or Sample Log Table even when the
same QC Item in the QC mode is restored.
• Also, once a Category of a QC Item is changed, it will change the setting
of related items in the QC Database, QC Configuration, or Sample Log
Table even when the same Category for the QC Item is restored.
• Basically the AGILENT 7500 ChemStation does a mass defect correction.
Therefore, Adopt exact mass should be OFF if not necessary.

323
Intelligent Sequence Training Text
Setting Up a QC Configuration

Creating a QC Report Template

Select QC Item whose


category is Tune to edit
QC report template for
Tune sample

When Report File is not selected

Select items
under an
adequate place

When Report
File is selected

324
Intelligent Sequence Training Text
Setting Up a QC Configuration

Notes on Editing QC Report Template

The very right cell


which contains
values is
recognized as the
right end of the
printed area

Repeated width should be


the same for each area

Concentration Mean

Dilution Factor

Equation using the


reference place

• The raw concentrations (not taken into account dilution factor) are always
indicated. To get the corrected concentration, set Concentration Mean
and Dilution Factor as printed Items. And then set the equation which
expresses Concentraion Mean multiplied by Dilution Factor using the
reference place.

325
Intelligent Sequence Training Text
Setting Up a QC Configuration

• There is no function to select the printed area. When there is a column


you don’t want to print out, select the column, and then select Format >>
Column Width >> Hide.
• When adding graph (spectrum) on a Tune type template, the repeat setting
cannot be applied. Individual setting is required.

326
Laboratory 1: Agilent 7500
Configuration, Startup and Tuning

Reference:
• Agilent 7500 ChemStation Operators Manual
• Agilent 7500 Customer Training Class, Module 4
• Agilent 7500 Customer Training Class, Appendix 2
Laboratory 1: Agilent 7500 Configuration, Startup and Tuning
Configuration

Configuration
1) Close ChemStation if open and open ‘Configuration’ under Agilent 7500
Program Group
2) Check: Offline Instrument – NOT CHECKED
3) Check: Remote Start – Don’t Use except for synchronization with external
sampling devices
4) Check: Sample Introduction as appropriate including peristaltic pump and
autosampler
5) Check: EM Protection – select Auto setting of integration time in analog
mode
6) Check: QC Mode – GENERAL.QCC
7) Save and Exit

328
Laboratory 1: Agilent 7500 Configuration, Startup and Tuning
Startup and Tuning

Startup and Tuning


1) Review Startup Checklist
2) Initiate Plasma and Warm-up Instrument
Position ALS in distilled water
Go to Tune
Select Tune >> PeriPump program for Autotune, review the values here
Select Sensitivity and Start
(Allow instrument to scan for 10-15 minutes to warm up the quadrupole)
Place the ALS probe in position 3 (10 ppb tune solution, Li, Y, Ce, Tl)
(Allow time for uptake, 1-2 min)
3) Select Sensitivity >> Start
Watch for awhile
Select File >> Print, (keep this for reference as your starting point)
4) Select Tune >> Autotune…
Select Torch Position, Sensitivity, Resolution/Axis and Tune Report
Run Autotune (watch what happens)
Compare the new tune report with the original one which you printed
5) Load “poor_s.u” (This has been deliberately mis-tuned for poor sensitivity)
Using the ChemStation Manual, Student Manual and Tune Flowcharts, fix
this tune
6) Load “high-ox.u”
Try to reduce the oxides in this tune
7) Ask the instructor to “Fix” the system so that it generates high %RSDs. Fix
this!
8) Save your best tune as best.u
9) Place the internal standard solution on line
10) Under Acq. Parms select masses 6, 140 and 209 (what are these elements
and where do they come from, e.g. why would we be interested in
monitoring them?)
11) Allow time for ISTD uptake and monitor the ISTD counts and %RSDs, do
they make sense?
12) Discuss your results with the instructor.

329
Laboratory 1: Agilent 7500 Configuration, Startup and Tuning
Agilent 7500 Startup Checklist

Agilent 7500 Startup Checklist


CHILLER
ARGON
PERIPUMP TUBES CLAMPED
INTERNAL STANDARD MIX
Out for tune
In for analysis
TUNE SOLUTION FULL
BLANK RINSE SOLUTION FULL
DRAIN RESERVOIRS NOT FULL
ALS RINSE PORT RESERVOIR OK
SHIELD IN or OUT (application dependent)
PLASMA ON

330
Laboratory 1: Agilent 7500 Configuration, Startup and Tuning
Shutdown Checklist

Shutdown Checklist
ALS IN RINSE SOLUTION, WAIT 1 MINUTE
PLASMA OFF
PERIPUMP TUBES UNCLAMPED
CHILLER

331
Laboratory 1: Agilent 7500 Configuration, Startup and Tuning
Shutdown Checklist

(ll) (III)
(l)
Poor Precision High Oxides and/or
Poor Sensitivity
(high RSDs) doubly-charged ions

Check Physical Adjust peri-pump Decrease carrier/blend


Parameters: shoe pressure gas flow
(decrease till no signal,
Sample Uptake rate then slowly increase Increase Sample depth
(bubble) just enough for stable
signal) Decrease peri-pump
Nebulization flow
(-visually) Adjust carrier gas flow
(-sodium bullet) (carrier + blend gas <=
1.4 - 1.5 ml/min) very
Torch Position slowly
(x,y visually)
Fine tune Extract Save Tune File
Cone orifices lenses and Omega
lenses
Torch Position
(7-8 mm) Clean or replace cones

Verify contents of tune


Sensitivity Check mass axis and solution
No No
Improved? resolution. Peaks?
Run SetEM

Yes Yes

Set:
Optimize Sample Carrier Gas: 1.2 - 1.5
Depth and X-Y position Peripump : .1 - .15
Ext. 1: -150 to -180
Ext. 2: -75 to -100
Omega Bias : -40
No Omega (+) : optimize
Sensitivity between 0 and 10 V
OK? QP Focus : optimize
between -10 and 10
Yes

RSDs OK? No Go to (II) HP 4500 ICP-MS


Manual Tune
Yes
Troubleshooting
Oxides/Doubly
Flowchart
Charged OK?
No GO to (III)

Yes
Save Tune File

332
Laboratory 2: Agilent 7500 Routine
Maintenance

Reference:
• Agilent 7500 Administration and Maintenance Manual
• Agilent 7500 Customer Training Class – Module 5
• Maintenance Log Table
Laboratory 2: Agilent 7500 Routine Maintenance
General

General
1) Remove and examine air filters
2) Examine level and color of oil in Rough Pumps
3) Check all fluids, belts and hoses (just kidding)
4) Check for corrosion and wipe down cabinet as necessary with damp cloth

334
Laboratory 2: Agilent 7500 Routine Maintenance
Sample Introduction

Sample Introduction
Remove:
• Peri-pump tubes
• Nebulizer
• Spray Chamber
• Torch
Clean or replace as necessary (refer to maintenance manual)

335
Laboratory 2: Agilent 7500 Routine Maintenance
Interface

Interface
1) Remove Sampler and Skimmer Cones and Extraction Lens assembly
2) Sonicate cones in 10% Citranox with occasional careful wiping until visibly
clean (10-30 minutes)
3) Rinse cones with water, then DI water, blow dry and set aside.
4) Disassemble extract lens assembly, examine lenses and insulators for
discoloration.
5) Sonicate in 10% Citranox and rinse thoroughly as above. Do not sonicate
the insulators unless obviously discolored as they take longer to dry.
6) Reassemble and reinstall extraction lenses and interface cones

336
Laboratory 2: Agilent 7500 Routine Maintenance
Nebulizer, Spray chamber and Torch

Nebulizer, Spray chamber and Torch


1) Babbington and Crossflow nebulizers can be sonicated in either dilute
Citranox or, 5-10% nitric acid as needed. Babbington nebulizers can be
unclogged (argon line) with a tiny GC syringe cleaning wire if needed.
Rinse well after cleaning.
2) GLASSWARE SHOULD NOT BE SONICATED! However, torches and
spray chambers can be boiled in hot 10% Citranox and then rinsed well.
This greatly improves the wetability of the spray chamber. Soaking in
strong (10-50%) nitric acid solution overnight may also be necessary for
extremely dirty or contaminated glassware.
3) Remove the glassware, examine carefully and replace. Be sure all gas and
spray chamber drain connections are leak free.

337
Laboratory 2: Agilent 7500 Routine Maintenance
Re-ignite the plasma and check the tune

Re-ignite the plasma and check the tune

338
Laboratory 3: Semi-Quantitative Analysis

Reference:
• Agilent 7500 ChemStation Operators Manual
• Agilent 7500 Customer Training Class Manual, Module 08
Laboratory 3: Semi-Quantitative Analysis
Semi-Quantitative Analysis

Semi-Quantitative Analysis
1) From Top Level, load default.m
2) Select ‘Edit Entire Method’
3) Edit all method sections to create a semiquant method for unknown sample
screening
Do not waste time acquiring nonsense elements such as inert gasses, air, carbon,
halogens etc. Also exclude the transuranic elements.
4) Use 0.1 second integration for all elements
5) Use 60 second uptake, 5 second optional rinse, and 60 second stabilization.
6) Do not configure the use of internal standards, since we will be using this
method to screen for the presence if internal standard elements in the
unknown sample.
7) Examine your tune report and estimate the semiquant response factor
threshold (Minimum Peak in cps) necessary to exclude results lower than
~0.1 ppb from the report.
8) Save the method as a unique name.
9) Analyze a blank and a 10 ppb (or 100 ppb) multielement calibration standard
with your method.
10) Using the blank and 10 ppb (or 100 ppb) standard, enable blank subtraction
and reset the semiquant response factors.
11) Analyze the unknown sample to screen for the presence of the internal
standard elements, as well as the presence and approximate concentrations
of other analyte elements. This information will be used to develop a
quantitative method for analysis of the unknown sample(s).
12) Compare the results with the certified values.
Are semiquant results subject to interferences?
Can they be corrected?

340
Laboratory 4: Quantitative Analysis of
Unknown Sample

Reference:
• Agilent 7500 ChemStation Operators Manual
• Agilent 7500 Customer Training Class, Module 09
• HP/Agilent Standard Operating Procedure, EPA Method 200.8
Laboratory 4: Quantitative Analysis of Unknown Sample
Quantitative Analysis

Quantitative Analysis
1) Using your SemiQuant results for the Unknown sample and the EPA 200.8
SOP as guidelines, build a quantitative method for analysis of Ag, Al, As,
Ba, Be, Cd, Co, Cr, Cu, Mn, Mo, Ni, Pb, Sb, Se, Tl, V and Zn.
2) Include at least two calibration levels plus a blank for each element. If the
element is likely to trigger analog mode, include a calibration point which
will also be acquired in analog mode. (Why is this?)
Multi-element calibration standards will be available in 1; 10; 100; 500;
1,000 ppb concentrations.
3) Build a simple sequence to update your calibration and analyze your
unknown sample at two dilutions.
4) Compare your results with the certified values.
5) Discuss your results with the instructor.
Hints:
Always run at least 2 calibration blanks at the beginning of a sequence to insure
adequate flush-out of previously run samples.
Always run a blank after the high calibration standard and before any samples to
detect possible memory effects.
Always analyze a blank and mid-point calibration as samples at the end of a
sequence and every 10 samples to verify that the system is under control.

342
Appendix 1 – General Information
Appendix 1 – General Information
Professional Organizations

Professional
Organizations

Professional Organizations

★American Chemical Society (ACS) (800) 227-5558


http://www,acs.org
★Environmental Protection Agency (EPA) http://www.epa.gov
★American Association of Clinical Chemists (AACC) (800) 892-1400
★American Board of Clinical Chemistry, Inc.(ABCC) (202) 835-8727
★American Society of Clinical Pathologists (ASCP) (312) 738-1336 x.158
★Clinical Laboratory Management Association (CLMA) (610) 647-8970
★College of American Pathologists (CAP) (800) 323-4040
★National Committee for Clinical Laboratory Standards (NCCLS) (610) 688-0100
★Society for Applied spectroscopy (301)694-8122
★SEMI International Standards (650) 964-5111
http://www.semi.org

Figure 263

344
Appendix 1 – General Information
Journals

Journals

Journals

■ Analytical Chemistry
■ Journal of Analytical Atomic Spectrometry (JAAS)
■ Applied Spectroscopy (free with SAS membership)
■ Spectroscopy (free)
■ Spectrochimica Acta, Part B
■ Analyst
■ American Lab (free)
■ American Clinical Lab (free)

Figure 264

345
Appendix 1 – General Information
Selected Web Sites (1)

Selected Web Sites


(1)

Selected Web Sites (1)

• Agilent Technologies
http://www.agilent.com

• Sample Preparation (Duquesne University)


http://www.sampleprep.duq.edu/sampleprep/
• Eastern Analytical Symposium
http://www.eas.org/
• FACSS
http://facss.org/info.html
• Pittcon
http://www.pittcon.org/
• NIST Standard reference Program

Figure 265

346
Appendix 1 – General Information
Selected Web Sites (2)

Selected Web Sites


(2)

Selected Web Sites (2)

• Spex catalog
http://www.spexcsp.com/crmmain/cat_f.htm
• Spectron
http://www.vcnet.com/spectron/
• Inorganic Ventures
http://www.ivstandards.com/
• Michael Cheatcham’s Instrument pages
http://www.geochemistry.syr.edu/cheatham/InstrPages.html
• Plasmachem-L BB
http://www.geochemistry.syr.edu/cheatham/icpmsins.html
• US Pharmacopeia
http://www.usp.org/

Figure 266

347
Appendix 1 – General Information
Selected Web Sites (2)

348
Appendix 2 – Flow Chats
Appendix 2 – Flow Chats
Manual Tune Troubleshooting Flowchart [1]

Manual Tune Troubleshooting Flowchart [1]

Agilent 7500 ICP-MS Startup


Manual Tune
Troubleshooting
10 minutes warm-up
Flowchart

Torch Position Too Low Sensitivity?

OK?
Yes
No In Spec.
Sample Uptake

OK?
Yes Precision ? Poor Sample Uptake
No
Carrier Gas Flow OK?
Yes
In Spec. No
OK?
Yes Carrier Gas Flow
No
Extraction 1, 2 Oxides and OK?
Doubly Yes
Charge? No
OK?
Yes Extraction 1, 2
No
Omega Lens High OK?
Yes
Sample Depth No
OK?
Yes Einzel 2
No OK?
Yes
Instrument Maintenance No OK?
Yes
Carrier Gas Flow No
In Spec. Instrument Maintenance
OK?
Yes
No
Pump Speed

OK?
Yes
No
Instrument
Maintenance

Resolution / Axis

Set EM Voltage

Set P/A Factor

Save Tune File

Print Tune Report.


Ready for Analysis

Figure 267

350
Appendix 2 – Flow Chats
Manual Tune Troubleshooting Flowchart [2]

Manual Tune Troubleshooting Flowchart [2]

(ll) (III)
(l) Poor Precision High Oxides and/or
Poor Sensitivity (high RSDs) doubly-charged ions

C heck Physical Parameters: Adjust peri-pump shoe Decrease carrier/blend


pressure gas flow
Sample Uptake rate (bubble) (decrease till no signal, then
slowly increase just enough Increase Sample
Nebulization for stable signal) depth
(-visually)
(-sodium bullet) Adjust carrier gas flow Decrease peri-pump
(carrier + blend gas <= 1.4 - flow
Torch Position 1.5 ml/min) very slowly
(x,y visually)
Fine tune Extract lenses Save Tune File
Cone orifices and Omega lenses

Torch Position Clean or replace cones


(7-8 mm)

Verify contents of
Check mass axis
Sensitivity tune solution
No and resolution. No
Improved?
Peaks?
Run SetEM

Yes Yes
Optimize Sample Set:
Depth and X-Y Carrier Gas : 1.2 - 1.5
position Peripump : 0.1 - 0.15
Ext. 1: -150 to -180
Ext. 2 : -75 to -100
No Omega Bias : -40
Sensitivity
OK? Omega (+) : optimize between 0 and 10 V
QP Focus : optimize between -10 and 10
Yes

RSDs OK? No Go to (II)

Yes Agilent 7500


ICP-MS
Oxides/Doubly
Charged OK?
No GO to (III) Manual Tune
Troubleshooting
Yes Flowchart
Save Tune File
Η

Figure 268

351
Appendix 2 – Flow Chats
Manual Tune Troubleshooting Flowchart [2]

352
Appendix 3 – Dealing with Polyatomics
Appendix 3 – Dealing with Polyatomics
The Problem

The
Problem

The Problem...

Bulk argon from the plasma can combine with entrained


atmospheric gases and matrix constituents to form argon
polyatomic ions.
The resulting polyatomic ions can overlap analyte masses of
interest compromising detection limits
Common examples are:
ArO+ interferes with 56Fe
ArH+ interferes with 39K
Ar+ interferes with 40Ca
ArCl+ interferes with 75As

Figure 269

354
Appendix 3 – Dealing with Polyatomics
Strategy #1: (High Power) Cool Plasma Analysis

Strategy #1: (High Power) Cool Plasma


Analysis

Strategy #1: (High Power) Cool Plasma Analysis

• Technique first reported in 1988*:


• Reduction of ArH+ interference on Ca simply by
modifying ICP operating conditions
• lower plasma power
• increased carrier gas flow
• longer sampling depth

* Jaing, Houk, and Stevens, Anal. Chem., 1988, 60, 1217

Figure 270

355
Appendix 3 – Dealing with Polyatomics
Commercialization of Cool Plasma Analysis

Commercialization of Cool Plasma


Analysis

Commercialization of Cool Plasma Analysis

• Developed commercially by HP/Yokogawa in 1992*


•featuring patented ShieldTorch Interface
• Cool plasma ICP-MS (using the Agilent ShieldTorch) now
the method of choice for ultratrace metals analysis in the
semiconductor industry
• Over 100 HP 4500's and Agilent 7500’s performing
routine cool plasma analysis in the semiconductor industry
replacing GFAA, ETV-ICP-MS, HR-ICP-MS

* Sakata K et al., Spectrochim. Acta, 1994, 49B, 1027

Figure 271

356
Appendix 3 – Dealing with Polyatomics
Schematic of Agilent ShieldTorch

Schematic of Agilent
ShieldTorch

Schematic of Agilent ShieldTorch

• Shield plate removes


capacitive coupling of the
coil to the plasma
Load coil
• Plasma is at true ground
Plasm a Torch potential
• Cooler central channel
• No re-ionization of
polyatomic species
Shield plate • Background spectrum is
virtually free from plasma-
based peaks.

Figure 272

357
Appendix 3 – Dealing with Polyatomics
Not All Cool Plasmas* Are the Same! [1]

Not All Cool Plasmas* Are the Same!


[1]

Not All Cool Plasmas* Are the Same! [1]

• The ShieldTorch interface is unique to Agilent


• Reduces interferences such as ArH, Ar, ArO, C2, ArC to low
ppt levels, enabling sub-ppt level DLs for K, Ca, Fe etc
• The key feature of the ShieldTorch is that it can achieve this at
900-1000W forward power
• other systems can only remove interferences at 600-650W -
interfering peaks reappear at higher power settings
• operating at 650W gives rise to severe matrix effects, and
reduced analyte range

* Analysis at 600-650W is often referred to as cold plasma analysis

Figure 273

358
Appendix 3 – Dealing with Polyatomics
Not All Cool Plasmas* Are the Same! [2]

Not All Cool Plasmas* Are the Same!


[2]

Not All Cool Plasmas* Are the Same! [2]

• Advantages of operation at 900 W - 1000 W (High Power


Cool Plasma)
• minimal matrix effects - similar to normal plasma
operation (1200-1300W)
• higher ionizing power - wider analyte range - including
even Zn and B
• complete sample matrix decomposition - greatly
reduced possibility of interface and spectrometer
contamination

Figure 274

359
Appendix 3 – Dealing with Polyatomics
Fe in 31% H2O2 - 5 ppt Spike Recovery

Fe in 31% H2O2 - 5 ppt Spike


Recovery

Fe in 31% H2O2 - 5 ppt Spike Recovery

Excellent linearity even


at 5ppt using high
power cool plasma

DL - 0.3ppt in the
H2O2 matrix

No collision cell or HR-


ICP-MS data reported
to date showing Fe
linearity at 5ppt

Figure 275

360
Appendix 3 – Dealing with Polyatomics
ShieldTorch Technology Eliminates Interferences Before They Form!

ShieldTorch Technology Eliminates Interferences


Before They
Form!

ShieldTorch Technology Eliminates


Interferences Before They Form!

?
C ollision C ell

= A nalyte

Shield Torch Technology

Figure 276

361
Appendix 3 – Dealing with Polyatomics
Can Heavy Matrices be Analyzed?

Can Heavy Matrices be


Analyzed?

Can Heavy Matrices be Analyzed?

Example Organics analysis


•Until now, carbon-based interferences have prevented the
trace analysis of Mg and Cr in organic samples
•C2 interferes with Mg 24, 25, 26
•ArC interferences with Cr 52, 53
Îhigh power cool plasma (900W) removes these intereferences
down to single figure ppt level
ÎAND is stable in all types of undiluted organic sample,
including xylene, toluene and NMP

Figure 277

362
Appendix 3 – Dealing with Polyatomics
Cr in Undiluted Methanol

Cr in Undiluted
Methanol

Cr in Undiluted Methanol

Excellent linearity at ppt


level demonstrates
removal of ArC
interference using
ShieldTorch system

The concentration of Cr in
the sample is calculated at
7 ppt

Figure 278

363
Appendix 3 – Dealing with Polyatomics
Example of Heavy Matrix Analysis

Example of Heavy Matrix


Analysis

Example of Heavy Matrix Analysis

Example Analysis of trace impurities in Metal Alloys


Levels required have traditionally been 10’s ppm or % in
the solid, but requirements for lower level characterization
are increasing

1000x dilution typically used for ICP-MS, so


Î1ppm in solid requires measurement of 1ppb in
solution
Îeasily achieved for most elements by ICP-MS, except
K, Ca, Fe
Îrobust high power cool plasma could be used

Figure 279

364
Appendix 3 – Dealing with Polyatomics
Calibration for 56Fe in 1000 ppm Pt

Calibration for 56Fe in 1000 ppm


Pt

Calibration for 56Fe in 1000 ppm Pt

High power cool plasma


(900W)

Fe at mass 56 in matrix
of 1000ppm Pt

Standard Addition
calibration at 50, 100,
200 and 500ppt

Figure 280

365
Appendix 3 – Dealing with Polyatomics
Calibration for 66Zn in 1000 ppm Pt

Calibration for 66Zn in 1000 ppm


Pt

Calibration for 66Zn in 1000 ppm Pt

High power cool plasma


(900W)

Zn at mass 66 in matrix
of 1000ppm Pt.

Standard Addition
calibration at 50, 100,
200 and 500ppt

Zn cannot be measured
at 600W

Figure 281

366
Appendix 3 – Dealing with Polyatomics
Determination of Se by High Power Cool Plasma

Determination of Se by High Power Cool


Plasma

Determination of Se by High Power Cool Plasma

Agilent 7500 Operating Parameters


ShieldTorch interface
Higher forward power (1000W)
Higher gas flows
1.2 Lpm carrier gas
0.8 Lpm blend gas
Ar2 ionization minimized, but plasma has power to ionize Se

Figure 282

367
Appendix 3 – Dealing with Polyatomics
Spectrum of 10 ppb Se and Blank

Spectrum of 10 ppb Se and


Blank

Spectrum of 10 ppb Se and Blank

Figure 283

368
Appendix 3 – Dealing with Polyatomics
Calibration for 80Se

Calibration for
80Se

Calibration for 80Se

• Se at 0, 1, 2, 5 and
10ppb

• No blank subtraction

• No interference
correction

• No internal standard

• Precision and linearity


good

Figure 284

369
Appendix 3 – Dealing with Polyatomics
Detection Limits for Se by Cool Plasma

Detection Limits for Se by Cool


Plasma

Detection Limits for Se by Cool Plasma

M ass Blk Blk SD 1ppb 1ppb-Blk 3s D L (ppt)


76 121 8.91 786 665 40.23
77 1796 57.43 2234 438 393.05
78 33 6.04 375 342 52.93
80 113 7.63 478 365 62.72
82 211 13.48 873 662 61.06

Detection Limits (3 sigma, n=10) in 4% HCl


Note - Integration times varied for different isotopes

Figure 285

370
Appendix 3 – Dealing with Polyatomics
Current Research Developments Using the ShieldTorch

Current Research Developments Using the


ShieldTorch

Current Research Developments Using the


ShieldTorch

• Trace elements in organics


• Se isotope ratios
• Trace As in chloride matrices
• Trace analysis of "difficult" ICP-MS elements
– e.g. S, Si, P
• Removal of isobaric interferences (not possible by
collision cells or HR-ICP-MS)
– removal of Hg from Pb at mass 204
– removal of Zr from Sr at mass 90

Figure 286

371
Appendix 3 – Dealing with Polyatomics
As Calibration in 10% HCl

As Calibration in 10%
HCl

As Calibration in 10% HCl

• As at 0, 100, 250, 500


and 1000ppt
• ShieldTorch cool
plasma
– ArCl removed but As
not ionised
– measure AsO at mass
91
• As DL - 40ppt (in 10%
HCl)

Figure 287

372
Appendix 3 – Dealing with Polyatomics
Low Level P Calibration

Low Level P
Calibration

Low Level P Calibration

• P at 0, 5, 10, 20 ppb
• ShieldTorch cool
plasma
– measure PO at mass
47
• P DL - 30ppt

Figure 288

373
Appendix 3 – Dealing with Polyatomics
Low Level S Calibration

Low Level S
Calibration

Low Level S Calibration

• S at 0, 5, 10, 20 ppb
• ShieldTorch cool
plasma
– measure SO at
mass 48
• S DL - 0.6 ppb

Figure 289

374
Appendix 3 – Dealing with Polyatomics
Low Level Si Calibration

Low Level Si
Calibration

Low Level Si Calibration

• Si at 0, 5, 10, 20 ppb
• ShieldTorch cool
plasma
– measure SiO at
mass 44
• Si DL - 1.2 ppb

Figure 290

375
Appendix 3 – Dealing with Polyatomics
Strategy #2: Resolve the Interferences

Strategy #2: Resolve the


Interferences

Strategy #2: Resolve the Interferences

High resolution ICP-MS can be used to separate the analyte and


interferent peak
Increase mass resolution and separate the analyte and interferent
peaks, then make the measurement

Figure 291

376
Appendix 3 – Dealing with Polyatomics
Limitations of HR-ICP-MS

Limitations of HR-ICP-
MS

Limitations of HR-ICP-MS

• As in optical spectrometry, there is a trade off between


resolution and sensitivity
– increasing resolution reduces transmission
– the net result is an inability to determine the analyte at trace
levels.
• A resolution of up to 7500 is required for some
interferences
– at 7500R, signal transmission is only 1% of that at unit mass
resolution (300R)

Figure 292

377
Appendix 3 – Dealing with Polyatomics
Resolution vs. Sensitivity

Resolution vs.
Sensitivity

Increased Reso
Resolution vs, Sensitivity

Reduced
Sensitivity
Sensitivity

Resolution

Figure 293

378
Appendix 3 – Dealing with Polyatomics
Other Facts About HR-ICP-MS [1]

Other Facts About HR-ICP-MS


[1]

Other Facts About HR-ICP-MS [1]

• Abundance sensitivity (tailing) is


40Ar
much worse in a magnetic sector
than a quadrupole
– if the analyte peak is adjacent
to a large interferent peak, eg
39K next to 40Ar, the tail from

Intensity
the Ar overwhelms the K
39K
signal
– in these cases, increasing the
resolution does not help

Mass

Figure 294

379
Appendix 3 – Dealing with Polyatomics
Other Facts About HR-ICP-MS [2]

Other Facts About HR-ICP-MS


[2]

Other Facts About HR-ICP-MS [2]

• Mass calibration is critical to obtaining accurate


analytical results, however, it is a time consuming
process for HR-ICP-MS instruments
– resolution is defined by changing slit widths, and these require
calibration/characterization
– since the peak is so narrow at high resolution, the peak
maximum cannot be reproducibly located
– the analyser must be scanned across the whole peak, which
decreases S/N significantly

Figure 295

380
Appendix 3 – Dealing with Polyatomics
Other Facts About HR-ICP-MS [3]

Other Facts About HR-ICP-MS


[3]

Other Facts About HR-ICP-MS [3]

• High resolution
instruments cost >$350k
• Not perceived to be
appropriate for routine
analysis
• Require a higher level of
operator skill than ICP-
QMS

Figure 296

381
Appendix 3 – Dealing with Polyatomics
Strategy #3: Dissociate Interferences Within the Spectrometer

Strategy #3: Dissociate Interferences Within the


Spectrometer

Strategy #3: Dissociate Interferences Within the


Spectrometer

• Utilize collision/dynamic reaction cell technology


• Insert a collision cell/dynamic reaction cell within the
spectrometer between the main ion lens and quadrupole
analyser
• Control gas phase chemistry within the collision cell to
dissociate polyatomic ions thereby eliminating the
interferences

Figure 297

382
Appendix 3 – Dealing with Polyatomics
Principle of Collision Technology

Principle of Collision
Technology

Principle of Collision Technology

Collision Cell

Analyser Quadrupole
Gas inlet/outlet

Photon Stop Multipole Ion Guide

For Analyte and Interferent ions occurring at the same nominal mass:
I nterferent + Reaction Gas IR New Mass

A nalyte + Reaction Gas No Reaction; Analyte at Same Mass

Figure 298

383
Appendix 3 – Dealing with Polyatomics
Selecting a Gas Phase Reagent

Selecting a Gas Phase


Reagent

Selecting a Gas Phase Reagent

• The ideal reagent should:


– have high reactivity with the interferent ion
– result in 100% conversion of the interferent ion into
its constituent products
– result in a final product that is stable
• Unfortunately, no single gas has been found to be
applicable for a complete multielement analysis
• Collision cell chemistries must be tailored for each
individual element and sample matrix

Figure 299

384
Appendix 3 – Dealing with Polyatomics
Optimizing the Gas Phase Reagent

Optimizing the Gas Phase


Reagent

Optimizing the Gas Phase Reagent

• Often, more than one reaction gas is required to analyse a


series of different elements
• The need to fill the collision cell with gas for the analysis
and then vent the cell and refill with another gas, and re-
measure the sample significantly increases analysis time
• Not only must the type of gas used be optimized, but the
flow rate of gas through the cell must also be optimized
for best S/N
• Few elements optimize using the same type of gas at the
same flow rate. Therefore, compromise conditions must
be used

Figure 300

385
Appendix 3 – Dealing with Polyatomics
Side Reactions Are Inevitable!!

Side Reactions Are


Inevitable!!

Side Reactions are Inevitable!!

• Side reactions i.e. unexpected reactions of the collision gas with


other matrix components, have always been reported when using
this technology
• For example, when using NH3 as a collision gas when Ni is
present in the sample:
Reaction Product
„ Ni+ + NH3 Ni(NH3)+
„ Ni+ + Ni(NH3)+ Ni(NH3) 2+
„ Ni+ + Ni(NH3) 2+ Ni(NH3) 3+
„ Ni+ + Ni(NH3) 3+ Ni(NH3) 4+

Figure 301

386
Appendix 3 – Dealing with Polyatomics
Side Reactions Create New Interferences

Side Reactions Create New


Interferences

Side Reactions Create New Interferences

Product Interferes with


Ni(NH3)+ 75As, 77Se

Ni(NH3) 2+ 92Zr, 92Mo, 94Zr, 94Mo, 95Mo, 96Zr,


96Mo, 96Ru, 98Mo, 98Ru

Ni(NH3) 3+ 109Ag, 111Cd, 112Cd, 112Sn, 113Cd,


113In, 115In, 115Sn

Ni(NH3) 4+ 126Te, 128Te, 130Te, 130Ba, 132Ba

Figure 302

387
Appendix 3 – Dealing with Polyatomics
Hydrocarbons Are Particularly Prone To Complex Chemistries Even
at Trace Levels

Hydrocarbons Are Particularly Prone To Complex


Chemistries Even at Trace
Levels

Hydrocarbons Are Particularly Prone To


Complex Chemistries Even at Trace Levels

Even for simple hydrocarbons:

Ar+ + CH4 CH3+ + Ar + H +


Ar+ + CH3+ CH2+ + Ar + H2
Ar+ + CH2+ CH+ + Ar + H3 +

Figure 303

388
Appendix 3 – Dealing with Polyatomics
Effects of Sample Matrix

Effects of Sample
Matrix

Effects of Sample Matrix

• “Multipole Ion Optics In ICP-MS”, Jonathan Batey, 25th


Annual Conference of the Federation of Analytical Chemistry
and Spectroscopy Societies, Paper #661.
• Aspirated 400ppm Ca into a collision cell device
• Ca could be measured at 40Ca, however, strong peaks were
observed throughout the mass spectrum due to CaOH+ and
other Ca molecular ions
• Conclusion - collisions cells were not appropriate for samples
containing significant matrix components

Figure 304

389
Appendix 3 – Dealing with Polyatomics
Strategies to Overcome the Problem of Side Reactions

Strategies to Overcome the Problem of Side


Reactions

Strategies to Overcome the Problem of Side


Reactions

• The collision cell quad can be scanned in concert with


the analyser quad*
– enables the collision cell quad to act like a notch filter
– can prevent side reaction ions from entering the analyser quad
* Tanner S.D., Baranov V.I., At. Spectroscopy, 20 (2) 3-4/99

Figure 305

390
Appendix 3 – Dealing with Polyatomics
Limitation of Scanning the Analyzer Quad

Limitation of Scanning the Analyzer


Quad

Limitation of Scanning the Analyzer Quad

• When the collision cell quad becomes contaminated


with sample matrix, it will charge up, and be unable to
follow the scan speed of the analyser quad
• Analyte transmission will fall dramatically
• Quadrupoles are difficult to clean
• Correct realignment is critical to achieve maximum
transmission

Figure 306

391
Appendix 3 – Dealing with Polyatomics
Collision Cells Can Create Interferences

Collision Cells Can Create


Interferences

Collision Cells Can Create Interferences


Through...

• Interferring ions combining with the reaction gas to


shift to a new mass
– This mass cannot always be anticipated due to ion clustering
• Reaction gas combining with analytes and matrix
components not previously interferred with
• Reaction gas combining with ultra-trace contaminants
in the cell
• The presence of any type of matrix significantly
complicates data interpretation

Figure 307

392
Appendix 3 – Dealing with Polyatomics
In Summary

In
Summary

In Summary...

• High resolution ICP-MS can resolve the interferences, but


is not easy to use, and in complex matrices, poor
abundance sensitivity gives rise to severe spectral overlaps
• Collision cells must use different gases/gas mixtures and
gas flow rates to meet all analyte requirements
• Collision cells not only reduce targeted interferences but
create new unpredictable molecular ion clusters in every
individual matrix
• For almost all applications, high power cool plasma has
superior performance

Figure 308

393
Appendix 3 – Dealing with Polyatomics
In Summary

394

Potrebbero piacerti anche