Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
(Submitted: September 29, 2008 ; Revised: November 12, 2008 ; Accepted: December 9, 2008)
Abstract - This paper is an overview of the recent developments in immunosensors, which have attracted
considerable attention. Immunosensors can play an important role in the improvement of public health by
providing applications for which rapid detection, high sensitivity, and specificity are important, in areas such
as clinical chemistry, food quality, and environmental monitoring. This review focuses on the current research
in immunoassay methods based on electrochemical detection for the analysis of environmental samples or
medical diagnostic methods with emphasis on recent advances, challenges and trends. Technological aspects
in the development of immunosensors such as kinetics of biomolecular interaction, techniques of
immobilization, simplification of assay procedures, immunointeration and catalytic studies and system
miniaturization are presented
Keywords: Immunoassay; Protein; Immunosensor; Bioengineering approach.
10-9mol.L-1) of hormones, enzymes, virus, tumor automated and miniaturized. Their potential of
antigens, and bacterial antigens (He et al., 2009; commercialization is increasing, because they allow
Campàs et al., 2008; Wang et al., 2008). solving great scale problems, such as problems in the
The immune system is a theme of great interest in areas of health, pharmacy and environment. A
investigations due to its powerful capacity of comparison is also presented of the analytic
information processing. The main objective of the capacities of several kinds of immunosensors, which
immune system is to recognize all the cells or constitute a particular interest area of biochemical
molecules in the system and to classify those cells as engineering.
self or not-self defensive mechanisms. In these The biological materials can be selected in order
assays not only the sensibility is considered, but also to satisfy analytic needs, operating in several
the specificity. The immunoassays are widely used in specificity levels. They can be highly selective,
clinical analysis. However, other applications of specific for a narrow margin of compound, or to
immunoassays of increasing importance have been show a wide specificity spectrum such as a sensitive
observed in other areas, such as in environmental biosensor, for example, only to one antibiotic (for
control (Michal et al., 2007; Rodriguez-Mozaz et al., instance, the gentamicin) or then to all the
2005; Velasco-Garcia, 2003) and in the quality antibiotics. This flexibility in selecting the biological
control of food (Nandakumar et al., 2008; Skottrup et materials allows the user to adapt the biosensor to his
al., 2008; Choi, 2005; Sadik et al., 2004; Gaag et al., or her needs. Biosensors based on the antigen –
2003; Shan et al., 2002). antibody interactions or haptens as biosensitive
The use of biosensors along the last 20 years has elements are called immunosensors, which combine
been an approach for immunoassays which has had the sensitivity and specificity of immunoassays with
important and interesting results (Skottrup et al., physical signal transduction.
2008; Zhu et al., 2005; Wang et al., 1998; Bergveld, Transducers convert the physical-chemical
1991). The production of analytic systems that model interaction between biomolecules and their specific
and simulate living organisms for detecting the analytes into a signal, which is amplified and
presence of certain kinds of substances or organisms registered as an analytical result. Immunosensors are
is an area with a fast development; it is receiving built by means of the appropriate combination of the
great attention in the scientific community in the last biomolecule (antibody polyclonal, antigen, hapten)
two decades. In this new field, the development of with the transducer (electrochemical, amperometric,
biosensors involves the identification and the potentiometric, piezoelectric, optical, etc.); used
optimization of the analytic system, composed by together, they can be applied in specific analytical
immobilized biological material, which interacts in a situations. In the classical biosensor, the receptor is
specific way with one or more analytes; in the usually immobilized onto the transducer surface,
interface with bio-electronics, this material is which enables it to detect interaction with analyte
coupled to electrodes, which translate a specific molecules. In contrast to immunoassays, immunosensors
interaction by the generation of a signal typically commonly rely on the reuse of the same receptor
detected through electrochemical, piezoelectric or surface for many measurements. It has already been
optical means (Xu et al. 2008; Hirst et al., 2008; shown by various authors that the antibody layer was
D’Amico et al., 2005; Turner et al., 1987). However, largely secured during sensor reuse, which might
biosensors are different from the existing techniques imply an economic advantage of the immunosensor
in at least two very useful and fundamental aspects: compared to commercial kit assays like ELISA The
the first of them is the intimate contact of the reusability is evaluated as an important feature of
biological material (whole cells, tissue, antibodies or biosensor. Between assays, the regeneration of the
enzymes) with a transducer that converts the used immunosensor has been carried out by stirring
biological signal into a measurable signal (Theâvenot in basic solution (NaOH/NaCl) or more commonly
et al., 1999); and the second aspect is its functional by use of glycine/HCl buffer solution (pH 2 - 3) for
size. The sensitive portion of a biosensor is usually few min, and then by washing with distilled water
small and it allows small sample volumes and a several times to desorb the binding antigens. (Wang
minimum interference with the existent processes Z. et al., 2008; Liu et al. 2008; Wang S. et al., 2008).
after the implementation. Direct signal generation potentially enables real-time
This review aims to explore key characteristic in monitoring of analytes, thus making immunosensors
the design and development of immunosensors, with suitable tools for continuous monitoring. Also
emphasis in the amperometric ones, which allow a factors such as progresses in microelectronics,
fast and sensitive detection in a system that can be electrochemistry and production of optical fibers and
nanotechnology have contributed to the development marked for destruction or elimination by some other
of several detecting elements. Thus, the integration method. The constant region determines the
of these technologies makes possible the production mechanism used to destroy antigen. The antibodies
of immunosensors applicable to a wide variety of are divided in five main classes, IgM, IgG, IgA, IgD
detection and monitoring problems. and IgE, according to the structure of the constant
In October of 2008 the Global Industry Analysts, area and its immune function. The literature offers a
Inc (http://www.strategyr.com/), reported that great amount of references with detailed information
several factors has motivated the use of biosensors in on antibody structures (Briand et al., 2006;
industrial, environmental, and especially medical Subramanian et al., 2004; Bao et al., 2002; Schuetz
diagnostic applications. Thus, the world market for et al., 1999) and there are also structural databases of
biosensors is estimated to reach $6.1 billion by 2012. proteins. It is also possible to find tutorials that
The growing population, rising incidences of chronic explore many current purification methods as new
diseases, such as, diabetes, and the growing need for emergent technologies. Subramanian et al. (2004)
environmental monitoring, are all factors expected to show an evaluation of the current progresses applied
prop up growth in the upcoming years. to the production of monoclonal antibodies,
industrial strategies, importance of antibody
fragments, application of chromatographic methods,
BIORECEPTOR MOLECULES AND quality control, virus removal and bio-security.
IMMUNOASSAYS In the tests of an immunoassay, in order to detect
the interaction between the antibody and the antigen,
The bioreceptor molecules of an immunosensor one of the immunoagents must be conjugated or
are the antibodies. The antibodies are also called modified by joining it to specific molecules or
immunoglobulins, because they are proteins related biological markers, to facilitate either the capture or
to the immunological system. The immunoglobulin the detection of the analyte. Several biological labels
G (IgG), the main antibody in the serum, consists of are commercially available in a wide range of styles
four polypeptides: two heavy chains and two light (See Table 1), of which the radioactive ones were
ones, joined to form a "Y" shaped molecule. The initially used, due to their inherent sensibility. Later,
amino acid sequence in the tips of the "Y" varies the restriction in relation to the use of radioisotopes
greatly among different antibodies. This variable led to the use of other markers, such as
region, composed of 110-130 amino acids, give the chemiluminescent compositions and enzymes (for
antibody its specificity for binding the antigen. The instance, alkaline phosphatases, horseradish
variable region includes the ends of the light and peroxidases), which convert the substrate of the
heavy chains. Each Ab has a unique structure that enzyme into a measurable product (Bratov et al., 2008;
attaches to an antigen in a lock-and-key fit. When the González-Martínez et al., 2006; Hoefelschweiger et al.,
Ab is attached to the Ag, the antigen is destroyed or 2005; Gosling et al., 1997).
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
230 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
Figure 1: Generally used formats: (a) a homogeneous competitive immunoassay, (b) a heterogeneous
non-competitive immunoassay, (c) a heterogeneous competitive immunoassay and (d) a heterogeneous
competitive immunometric assay.
Among the different immunoassay formats limited amount of antibodies (Figure 1d). In the case
described in the literature (Stolpera et al., 2007; of the immunosensors, direct assays have been more
Tschmelaka et al., 2004; Matsunaga et al., 2003; frequently applied. The most common formats in this
Sadana et al., 2002; Warsinke et al., 2000), four of field, of fast detection, are the competitive assays
them are more commonly used (see Figure 1). In a and the sandwich assays.
homogeneous immunoassay (Figure 1a), the The limitation of sandwich analysis is that this
antibodies, antigens and labeled antigens are mixed. cannot be used for hapten determination (analytes of
The antigens freely marked and those which are low molecular weight). The small size of those
linked to the antibodies can be distinguished by a molecules just allows the immunointeraction with an
change of activity of the marker when coupled. The antibody molecule. In a general sense, sandwich
immunoassays are usually heterogeneous, which analysis schemes give the lower detection limit,
means that the antibody or the antigen is resulting in analyte concentrations analyzed in
immobilized in a solid support and an picomolar ranges.
immunocomplex is formed upon entering into The conventional immunoassay techniques are
contact with a solution containing the other convenient for analytical practice with a great
immunoagent. The non-linked proteins are removed number of analyses (frequently identical) and are
by washing and the answer obtained from the labels commonly used. However, these techniques can only
is proportional to the amount of linked protein. be used in hospitals and laboratories especially
The more common kind of enzymatic equipped with personnel with technical training. The
immunoassay used in clinical analyses is known as automation of the measures of this multi-step
Enzyme Linked Immune Sorbent Assay, or ELISA. procedure is difficult (Farré et al.2007; Ghindilis et
There exist different schemes of enzymatic al., 1998, 1997). Consequently, the time for analysis
immunoassays (of competitive and non-competitive through conventional immunoassays usually goes
type) and, in the clinical analyses practice, two of the from one to several hours; it makes this technique
most popular methods are the sandwich method and inadequate for the fast determination of analytes.
the competitive immunoassay method (Harlow and The basic principles of the alternative
Lane, 1988). immunoassay methods are similar to the
In a non-competitive sandwich immunoassay, the conventional immunoassay techniques. In the
antibodies are immobilized and, after the addition of alternative assays, also based on the discovery of
the sample which contains the antigen, a conjugated interaction of the antigen-antibody, several
or secondary labeled antibody is added (Figure 1b). approaches are used, such as the development of: (i)
In a competitive assay, the competition happens discovery methods highly sensitive for the label; (ii)
between the free and the linked antigen for a limited improved immunointeraction schemes; (iii) kinetic
amount of labeled antibody (Figure 1c) or between studies of these immunointeractions, (iv) automated
the antigen (sample) and the labeled antigen for a immunoassay schemes and, (v) miniaturization.
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
232 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
surfaces effects, regeneration effects, temperature, entrapment within a thin Nafion film. In other words,
influence of the flow rate etc. an ion exchange is performed in a very porous
In this same context, on kinetic studies used for polymer material that exhibits a good attachment to
biosensor technology, a recent work by Nordin et al. gold. In a similar way, Nafion has also been widely
(2005) presents a study of eight different kinds of used in amperometric immunosensors (Agui et al.,
representative kinases of a great family of proteins. 2008). For example, a novel immunosensor was
Through the kinetic study, the authors found the recently proposed by Wu et al. (2008) that is based
critical conditions of immobilization, which allowed on gold nanoparticles assembled onto the
them to create strategies to preserve the coupling TMB/Nafion film modified electrode to provide
capacity for the inibitors and for the developed ATP. active sites for the immobilization of antibody (anti-
The assays presented by the authors also include MIgG) molecules. Another amperometric
kinetic characterization of inhibitors that couple to immunosensor is based on gold nanoparticles/
kinases and coupling characteristics in the presence thionine/Nafion-membrane-modified gold electrode
of ATP to identify competitive ATP binders. for determination of α-1-fetoprotein (Zhuo et al.,
2005), Zhou et al. (2003b) described another
ii) Techniques for Biomolecule Immobilization amperometric immunosensor for the assay of the
antibody of Schistosoma japonicum, where the
Techniques for biomolecule immobilization polyanionic perfluorosulfonated Nafion polymer was
should allow a stable bond between the sensorial used to modify the glassy carbon electrode as a
surface and the bioreceptor, without interfering with platform for the immobilization of S. japonicum
the biological activity of the biomolecule. This is a antigen. In this same line, our research group
key aspect in immunosensor assembly. In proposed an immunosensor for detecting the
heterogeneous assays, immobilizing the antibody or antibody anti-apyrase of S. mansoni based on rigid
the antigen in the solid support frequently requires composite materials, containing graphite powder and
several wash steps and blocking in order to remove epoxy resins, A surface modification strategy for the
the excess reagent and to cover new immobilization use of oxidized graphite in the detection of antibody–
sites. General strategies for immobilizing the antigen interaction was developed. This modification
immune reagent on the electrode surface include: strategy is based on silanization of a conductive
physical adsorption, entrapment in a polymeric composite (Bojorge et al., 2007).
matrix and covalent attachment. Among these Another well-known immobilization method is the
methods, physical adsorption has exhibited capacity covalent attachment of a protein to an inorganic or
for assuring the protein activity, but the forces organic surface (Quan et al., 2004a, 2004; Divya et al.,
involved can produce weak interactions with the 1998). This method is potentially more aggressive, but
surface and can suffer desorption (Zhou et al., it can produce an almost irreversible immobilization of
2003a,2003b,2002; Walcarius et al.1998). the protein onto the sensor surface, which allows
Another method currently used in some reusing the sensor after washing procedures or
applications is the method of entrapping the protein regeneration. Tedeschi et al. (2003) established new
on the polymeric membrane surface (Jiang X. et al., immobilization methods applied to covalent TIRF
2008; Darain et al., 2003; Liu et al., 2003; Naqvi et sensors such as: (a) GOPS-dextran method, which
al., 2002; Rabinovich et al., 2001). In the case of consists in tying the antibody in a covalently bonded
optical detection, for instance when this method is activated dextran matrix to the surface through GOPS
applied in total internal reflection fluorescence (glycidyloxypropyl-trimethoxysilane); (b) NaIO4/AADH
immunosensors (TIRF), it is necessary to ensure that method: sacharideus oxidation in the fragment Fc of the
the antibodies in the evanescent wave zone, very antibody using periodate (NaIO4) and its connection to
close to the optical surface, are very well trapped in surfaces activated with hydrazide; (c) GOPS/F(ab´)
the polymer so as to be optically transparent. The use method, where the antibody fragments F(ab´) are
of Nafion for antibody entrapment was explored in linked to the surface activated with GOPS; (d)
order to immobilize antibodies or antigens. Susmel et APTS/Sulfo-SMCC method: fragments of the
al. (2005) studied the performance of a piezoelectric antibody, linked to an APTS (3-(2-aminoethylamino)
immunosensor prototype in which the propyl-trimethoxysilane).
immobilization of the Bacillus cereus antibody on There are several studies on protein
the crystalline surface was accomplished by simple immobilization and, particularly, several procedures
selection of individual clones for specific linking of applied to immunoassays were developed (Wang H.
the antibody to the hapten; and (d) selection of et al., 2008; Dong et al., 2007; Tang et al., 2007).
antibodies which exhibit the wanted catalytic Bange et al. (2005) indicate and discuss in full detail
activity. A mathematical analysis to describe a other fundamental aspects that should be considered
distribution of Michaelis-Menten antibody catalysts for all the devices accomplishing immunoassays,
could to be performed to help to interpret the results. such as: (a) type of micro-fluid; (b) electrode surface
However, it is essential that the catalytic antibodies modification in order to prevent the adsorption of the
are completely purified from possible endogenous sample and the key reagents, which can degrade the
enzyme contamination, which can catalyze the same assay performance; (c) detection device adaptation:
kind of reaction in study. Catalytic antibodies have the systems of detection of assays for different signal
great potential in the pharmaceutical industry (Feng transducers, which recognize the event of the Ab-Ag
linking, should be adapted to very small volumes.
et al. 2008; Khorasani-Motlagh et al., 2004; Schuetz
All these aspects show that immunosensor
et. al., 1999).
improvement is a topic of optimization applied to
Biochemical Engineering, whose variables are:
vi) System Miniaturization
sensitivity enhancement and decrease of the smallest
detection limit, time of analysis decrease,
Another aspect is the system miniaturization of
simplification of the analysis procedure (fewer
sensors and, in some cases, the simplification of assay
stages), miniaturization of the equipments and
procedures. All these advances have generated an
automation of the measurement procedures.
increasing demand for miniaturized and portable
solutions. The miniaturization of immunosensors
allows the application in medical diagnoses at home, in
CONFIGURATIONS OF TRANSDUCERS
the field of environmental monitoring, in the
scientifical detection of crimes, in the supervision of
The transducer selection depends on the
quality in small food industries etc. Suzuki et al. (2001) physical-chemical changes of the specific reaction
from the Department of Biochemical Engineering of that happens in the bio-layer. According to the
Kyushu Institute of Technology, together with transducer used, immunosensors can be classified
researchers from the Department of Electric and as electrochemical (potentiometric, amperometric
Electronic Engineering of the University of Toyama, and conductometric), optical (luminescence
developed a chip and a miniaturized SPR measurement, fluorescence, ellipsiometry etc.) and
immunosensor for detecting human IgG and molecules piezoelectric - mass detectors (they relate the
such as pesticides and dinitrophenol. The size of the oscillation frequency of piezoelectric crystals with
sensor is just 22 mm width x 30 mm length. Zhou et al. mass variation). Each immunosensor is designed
(2003b) developed an electrochemical immunosensor and optimized for a function under defined
of carbon paste of 6.00 mm id. And Yoon et al. (2004) conditions, related to a specific problem, which
developed an immunosensor system for ferritin involves considerations related to sensitivity, speed,
analysis, consisting of two rectangular gold and efficiency and simplicity of the assay procedures.
titanium electrodes of 10 μm x 500 μm and circular The greatest problem is to reproduce the obtained
electrodes with 50 μm radius. results, especially depending on the technique and
When designing an immunosensor, besides on the analytical range.
determining the assay format, it is necessary to Several immunosensor configurations are presented
consider that this device should have the smallest in Figure 3. Note that the bioreceptor molecules are
possible size and, depending on the scale, in some immobilized on an appropriate matrix in order to form
cases the accomplishment of studies of microfluidies a bio-layer in the exposed surface of the transducer.
is demanded. Transducers that use ion-selective electrodes and Field
The technology of micro-fluids aims, in a general Effect Transistors (FET) belong to the category of
sense, to improve the analytical performance of potentiometric transducers; surface plasma detectors
sensors through the reduction of reagent and surface acoustic wave detectors belong to the
consumption, the decrease of the time of analysis category of piezoelectric transducers. Potentiometric
and the increase of the reliability and of the transducers, together with conductive and
sensibility, all this by means of the automation and amperometric sensors, belong to the category of
integration of multiple processes in only one device. electrochemical transducers. The construction materials
Several designs using technology of microfluids for transducers are also presented in Figure 3.
Immunosensors of wider use are currently those This means that the biological recognition is measured
which contain electrochemical detection elements, by means of an electrochemical signal. The electrodes
such as physical transducers. These detecting employed can be easily miniaturized and, due to the
elements can be subdivided into potentiometric (field advanced technology of semiconductors and
effect transistors and ion-selective electrodes) and serigraphy, they can be mass-produced. These
amperometric detectors (graphite electrodes or noble characteristics have made possible great development
metal electrodes, such as gold or platinum); the last in the area of biosensors, as corroborated by the
ones measure the current for a certain potential - number of publications and patents. Inside the group of
constant or variable. The popularity of these electrochemical immunosensors, it is possible to
electrochemical immunosensors is due not only to classify them in three groups: Amperometric,
their high sensitivity and selectivity, appropriate for Potentiometric and Impedimetric and Conductometric.
immunoassays, but also because, in comparative The amperometric ones are the most used (Mehrvar et
terms, these instruments are indeed cheaper and it is al., 2004).
possible to miniaturize them. Moreover, the Electrochemical transducers have a very
continuous response of the sensor system allows important role in environmental protection.
computerized control, simplifying the Particularly, electrochemical sensors and detectors
electrochemical detection, bio-security and lower are interesting because of the advantages of their
cost than other analytical techniques. technology, which include economy, portability and
Piezoelectric immunosensors and Quartz Crystal the possibility of directly identifying and quantifying
Microgravimetric (QCM) immunosensors have also specific compounds in complex mixtures, commonly
been object of research. These immunosensors found in air, soil and water and in biological samples
determine the mass variation due to the formation of (Stradiotto et al., 2003; Erdem et al., 2000). Such
an immune Ab-Ag complex on the surface of a devices satisfy many of the demands for
piezoelectric crystal, used as detector. environmental analysis, and they are inherently
The optical immunosensors are also objects of sensitive and selective for electroactive species.
study, although their practical use is presently They are also fast, precise and have a low cost (Chen
limited due to the high costs of instrumentation, the et al., 2006; Mehrvar et al., 2004; Richards et al.,
sophistication of preparation procedures and the 2002).
necessity of highly qualified personnel.
Amperometric Immunosensors
Electrochemical Immunosensors
The interconnection of voltammetric principles
This kind of sensor uses the bio-recognition with immunologic reactions makes it possible to
element linked with electrochemical transducers. develop low cost, highly sensitive, and selective
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
236 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
Schematic representation of components in the Schematic layout of the SPCE. Adapted from
amperometric cell. Adapted from Salinas et al., 2005. Darain et al., 2003.
Figure 4: Different prototype design of amperometric immunosensors proposed in the literature
Another immunosensor modality considers the use Avidin/Biotin coupling. Stiene et al. (2002) described a
of screen-printed carbon electrode (SPCE), Two screen-printed flow-through cell for immunoanalysis,
techniques borrowed from the electronics industry have which allows determining the activity of the peroxidase
proved particularly important--screen printing (ink is by means of the electrochemical reduction of p-
pressed through a mask to form a pattern on a ceramic benzoquinone. The advantage of using SPCE is that it
or plastic base) and photolithography (a photoresistant offers the possibility of low cost production, without
material is exposed to ultraviolet light passed through a needing any additional mechanical part. It facilitates
mask and then the silicon is chemically etched). miniaturizing, since the detector works as a fine layer
Reproducible manufacture of biosensors with screen detector with very small internal volumes (usually <1
printing techniques means that each instrument does mL). The carbon or graphite paste or screen printing
not have to be calibrated before use. In this modality, techniques are frequently used in preparing electrodes
Sarkar et al. (1999) developed an electrode for immunosensors.
immunosensor of the SPCE type without mediator, Several methodologies conventionally applied in
covered by a conductive polymer (5.2´:5´2´´- immunoassays have been used in amperometric
terthiophene-3-carboxylic acid) for the detection of immunosensors, although the heterogeneous
rabbit IgG as analyte. Horseradish peroxidase (HRP) methods stand out because of their widest
and streptavidin were covalently linked with the application. They are characterized by the need of
polymer in the electrode and the biotinylated antibody accomplishing a physical separation of the formed
was immobilized on the electrode surface using immunocomplex and the non-linked species.
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
238 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
The use of highly dispersed carbon material for size of protein molecules. This results in a pseudo-
developing amperometric immunoassays by flow has homogeneous immunointeraction among the free and
also been reported. Abdel-Hamid et al. (1998) immobilized immunespecies. The system demonstrated
developed an immunoassay system for flow based on a low detection limit ranging from 10-11M (for Rabbit
an amperometric immunosensor which uses a highly IgG) and the total assay time was only 17 min. Table 2
dispersed conductive immunosorbent. The size of presents some analytical characteristics, usually used in
carbon particles (thousands of Å) is comparable to the amperometric immunosensor systems.
Range/Limit of Time of
Analyte Assay Electrode (E0) Immobilization Marker Sample Reference
detection detection
Clenbuterol ELISA GCE MWCNTs Covalent Redox (free) 0.32 ng mL−1 40 min Buffer He P. et al., 2009
Lactoferrin Competitive Iodide Adsortion BAC-SulfoNHS 35 nM 5 min Milk Campanella et al.2008
Okadaic acid ciELISA SPEs Adsorption ALP/ HRP 1.07 and 1.98μgL−1 30 min Shellfish Campàs et al.2008
1.3×103 - 7.6×104
Staphylococcus Au SPEs: 0.00V Escamilla-Gómez et
Competitive Covalent HRP-antiRbIgG cellsmL−1 30 min Milk
aureus (vs. Ag/AgCl) al. 2008
(1.4×104 cellsmL−1)
Pt nanoparticle: 10pgmL−1 - 1.0 PBS/
hIgG Sandwich Adsorption ALP 10 min Huang et al., 2008
−1.2 - 0 V μgmL−1 Serum
GCE-DNA–PDDA: 0.5–120.0 ngml-1
CEA --- Adsorption --- -- Buffer Li et al., 2008
−0.6 to 0.2 (vs. SCE) (0.3 ng ml-1)
α-fetoprotein Competitive Au-wires Covalent HRP-anti-CEA --- 35 min. Serum Wang et al., 2008
Tyr–Aucoll–
Progesterone Competitive Adsorption ALP–prog 0.43 ngmL−1 --- Milk Carralero et al., 2007
graphite–Teflon
Pt((−300mV vs 10−3–102 µgmL−1
Probarley lectin ELISA Covalent mAb–HRP --- Buffer Sadik et al., 2007
Ag/Agul) (10−3 μgmL−1)
6.4 ng.ml-1- 100
SjAb Indirect GCE Adsorption Colloidal gold 60 min Buffer Chu et al., 2005
µg.ml-1 (3.0 ng.ml1)
HRP- rabbit -1
anti-CT Sandwich Pt or GCE Covalent 50 ng.ml --- Buffer Ionescu et al., 2005
IgG
Ab Salmonella SPCE ( 0.40 V vs
Indirect Adsorption Apt --- --- Serum Rao et al., 2005
typhi Ag/AgCl)
Salivary
amylase ELISA Au-electrode Covalent --- 1.57 pg.ml-1 --- Buffer Aluoch et al., 2005
proteins
HCG Competitive GCE Adsorption HRP-anti-HCG 1.4 mIU.ml-1 --- Serum Chen et al., 2004
HCS Competitive SPCE Adsorption GOx e HRP 0.01-1 ng ml-1 15 min Buffer Dzantiev et al., 2004
GCE(−0.2 & +0.4V
rabbit IgG Ag --- Entrapment label-free --- 200 ms Buffer Gooding et al., 2004
vs. Ag/AgCl)
0.11-22.4 mg ml-1
SjAg Competitive CPE Entrapment SjAg –HRP --- Buffer Lei et al., 2004
(0.06 mg ml-1)
mouse IgG Sandwich GCE Adsorption Apt 0.01-100 mg ml-1 --- Buffer Másson et al., 2004
0.5 - 2 mg.ml-1
rabbit IgG Competitive SPCE Covalent Gox --- Buffer Darain et al.2003
(0.33 mg.ml-1)
GCE (100 mV,
C3 Competitive Adsorption HRP-C3 2.4 ng.mL-1 30 min Serum Zhou et al., 2003a
Ag/AgCl)
GCE -250 mV -1
Sj Ab Competitive Adsorption HRP-SjAb 0.05 - 4,8mg.ml ---- Buffer Zhou et al., 2003b
(Ag/AgCl)
SPCE (+300 mV vs.
Sea food toxins Competitive --- Apt 10 - 160 ppb (2 ppb) 35 min Buffer Kreuzer et al., 2002
Ag/AgCl)
GCE - SjAb: -1
Sj Ag Competitive Adsorption HRP-SjAg 0.5 - 30mg.ml 30 min Buffer Zhou et al., 2002
0.1 V (Ag/AgCl)
Legends
Ab, antibody C3: complement III GOx: Glucose oxidase IgG: Immunoglobulin G
Ag, antigen E0: Potential CPE : Carbon paste Electrode CEA : carcinoembryonic antigen
HCS: herbicide HCG: chorionic gonadotrophin ELISA: enzyme-linked SPCE : screen-printed carbon
chlorsulfuron immunosorbent assay electrodes
Pt : platinum electrode FIA: Flow Injection Analyze MWCNTs: multi-wall carbon SPEs: screen-printed electrodes
nanotubes
GCE: Glassy Carbon PDDA::Poly - HRP: horseradish peroxidase
Electrode diallyldimethylammonium
chloride
GCE: Glassy Carbon PDDA::Poly - ALP: alkaline phosphatase
Electrode diallyldimethylammonium
chloride
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
240 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
Some comparison parameters in the development Amperometric immunosensors are based on the
of immunosensors are sensibility, detection level, determination of the Ab-Ag interaction by use of an
specificity and calibration. enzyme as marker. These enzymatic markers,
The required sensitivity for a particular analyte is peroxidase, alkaline phosphatase, laccase or glucose
determined by the concentration levels found in the oxidase are commonly attached to a secondary
measurement medium of interest. For enzymatic antibody. After their immobilization, the enzymes
biosensor measurements of glutamate, the difference catalyze the formation of electroactive species,
between the extracellular milieus of 10 μM and 10–10 which are amperometrically detected at the electrode
mM for intracellular measurement is significant surface, thus allowing the quantification of the
(Kahlert et al., 2004). The discrepancy between immunoreactions. The sensitivity of such an
intra- and extra-cellular concentrations is not approach obviously depends on the specific activity
confined to analytes measured with amperometric of the enzyme as well as on the electrochemical
immunosensors, but also to other important species properties of the products of the enzyme reaction. In
such as oxygen with intra- and extra-cellular addition, the immunosensor performances are mainly
concentrations given respectively by (0.032 a function of the accessibility to the electrode surface
mM/0.24 mM), Na+ (20 mM/140 mM), and K+ (5 for the generated electroactive species. The latter
mM/130 mM) (Kahlert et al., 2004). These point is controlled by the steric constraints due to the
differences create concentration gradients across the immobilization procedure of the immunoreagent
cell membrane, which may be changing rapidly in used as probe and, hence, by the bulkiness of the
the second to millisecond time domain. diffusing species. Ionescu et al. (2005) compared the
Immunosensor sensitivity can be enhanced by performances of amperometric immunosensors for
changing the surface of the electrode. For oxidation cholera antitoxin based on three enzyme markers
of peroxidase, which is most rapid on Pt, carbon (HRP, biotinylated polyphenol-oxidase (PPO-B) and
fibers can be platinized (Clark et al., 1998). biotinylated glucose-oxidase (GOX-B)), which
Deposition of platinum black on carbon or Pt present an average standard deviation of the
electrodes also increases the active surface area, sensitivity around 450 nA μg−1 ml cm−2.
further enhancing sensitivity. However, this strategy A key problem in the use of immunosensors is the
also increases the electrode sensitivity to reproducibility of certain results. Depending on the
electroactive interferences and makes the sensor detection technique and the analytical range, the
vulnerable to decreased response resulting from determination of the error can vary over a wide
adsorption of surface-active species. Special films range. Specifically, the use of optical
can serve to concentrate analytes by adsorption, such immunosensors, for instance, based on SPR, is
as Nafion. The sensitivity of a SPR immunosensor usually characterized by the low detection level and
for the detection of antigens with a very low the smallest determination errors (2-8% level)
concentration can be increased by control of the through the whole analytical range. However, these
orientation of antibodies immobilized on the SPR immunosensors exhibit a narrower analytical range
sensor surface. When antibodies are immobilized on and they require the use of sophisticated and
a solid surface, their activity is usually less than that expensive instruments. Therefore, they are less
of dispersed antibodies. The main reason for activity common than the amperometric immunosensors.
reduction is due to the random oriented array of the Shen et al. (2005) investigated the reproducibility of
antibody molecules on the solid surface. Therefore, the piezoelectric quartz crystal immunosensor; 100
the development of the immobilization method for mU/ml of antisperm antibody concentration was
antibodies is strongly required to make an oriented determined repeatedly under the optimized
layer. experiment conditions. The average frequency
Liang et al. (2005) developed a hepatitis B change of three parallel experiments is 170 ± 11 Hz,
surface antigen (HBsAg) immunosensor by self- and the relative standard deviation (R.S.D.) is 6.65%.
assembling gold nanoparticles to a thiol-containing At the same time, complex systems for amplification
sol–gel network. The amount of HBsAb immobilized of signals (e.g., with use of enzymatic reactions) are
on the electrode surface determines the sensitivity more frequently used in amperometric
and the linear range of the immunosensor. The larger immunosensors; they require the use of additional
amount of HBsAb immobilized on the electrode reagents and the course of the immunological
surface can provide more immunoactive sites for the reaction cannot be monitored directly. Moreover,
immunoreactions of antigen. amperometric immunosensors are characterized by a
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
242 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
wide range of error determination, depending on the Cl-, can create problems due to their presence in the
label used and the format of the immunoassay. More buffer composition and in the electrolytic solution.
frequently, the error determination may vary from 2 It is worthwhile to mention that the use of
to 20%; however, it can be larger in specific cases. chemiometry is the key in order to minimize the
As a consequence of the direct monitoring of the uncertainty in data processing in all the analytical
interaction between antigens and antibodies, methods. With the growing sophistication of the
piezoelectric immunosensors impose strict instrumental techniques, more complex data
requirements on the antibodies’ specificity, because treatments became necessary from a mathematical
any cross-linking can produce significant and statistical point of view, in order to relate the
interferences. However, if non-specific linkings are obtained signals (intensities for instance) with the
taken into account, the determination errors with desired results (concentrations). The statistical
piezoelectric immunosensors are relatively low and evaluation of analytic data plays a very important
they vary from 3 - 5 to 10 -15%. In addition, some role in the minimization of the uncertainty of
electronic filtering of background noise can be electrochemical sensors in clinical analyses. A good
effective. Extending the linear range by creating quantity of electrochemical sensors was proposed for
diffusion-limiting barriers with permselective clinical analysis. Lower uncertainties for the
polymer membranes is a widely employed approach. analytical information when an arrangement of
The sensor response is no longer controlled by the electrochemical sensor is used in vitro and in vivo
kinetics of the enzyme reaction, but by mass transfer. are obtained through the use of the best
This has the advantage of minimizing temperature chemiometric methods for data processing. If the
effects (approximately 2.5%/°C increase in the rate uncertainty of the simultaneous measurements of the
of mass transfer versus ~10%/°C for enzyme compounds that use the electrochemical sensor
catalysis). The disadvantage is a decrease in arrangement is still big, it is possible to think of the
sensitivity and response time and, under some application of electrochemical sensors/flow systems
conditions, complicated sensor response (Jablecki (e.g., flow injection analysis or sequential injection
and Gough, 2000). analysis) for a simultaneous assay of different
Assessment of sensor performance is critically components. This way the data processing can be
dependent on a reliable procedure for calibration. It made using simple statistical evaluation.
is frequently not possible to rely on in vitro According to Kissinger (2005), in research
calibrations as the basis for the in vivo performance. applications, the economy in the sensor itself is less
Much of the effort on calibration has been directed important than its application. For instance, if SPR
toward the performance of glucose sensors, where helps in the development of a drug that will treat
the basis for calibration is the assumption that the millions of people annually, the cost of the sensor
blood glucose concentration and the subcutaneous will be very small compared to the benefit that can
interstitial fluid yield the same value. This approach be reached. There are many examples, but it is
has been complicated by the fact that the ratio of necessary to be careful. Sensors are extraordinarily
blood/tissue glucose is not constant, but depends on expensive when the cost of developing them exceeds
a number of factors related to physiology of glucose the annual size of the market for the analyzed
and insulin (Wilson and Gifford et al., 2005). One of analyte. Cromatography, mass spectrometry and the
the disadvantages of the use of electrochemical immunoassay methods give better precision with
sensors is the need of calibrating them before each lower limits of quantification for a wide variety of
measurement. In order to minimize the measurement analytes. However, they present larger times of
uncertainty, it is necessary to have the same medium analysis and they don't make use of the advantages
(for instance, pH, ionic strength etc.) for the standard of miniaturization and portability of the instruments.
solution used for calibrating and for the sample. This
is possible only for analysis in vitro. For analyses in
vivo, the calibration method of the electrochemical CONCLUSIONS
sensor can significantly increase the measurement
uncertainty, because the medium to be simulated Immunosensors play an important role in the
consists of the organs or biological fluids which have areas of health and environment because they
to be measured (e.g., pH, buffer type, acidity through provide powerful analytical tools for diagnosis,
the addition of HCl etc.) (Stefan et al., 2003). The particularly when fast, low cost, and high sensitivity
determination of some ions, for instance Na+, K+, and specificity measurements are necessary.
Advances in the area of immunosensors in system for rapid detection of bacteria, Anal.
commercial contexts can be accelerated by using Chim. Acta, 399, 99–108 (1999).
modern instruments, besides signal processing Adányi, N., Levkovets, I. A., Rodriguez-Gil, S.,
methods such as chemiometry. It is possible to Ronald, A., Váradi, M., Szendrő, I., Development
observe that bioengineering is paying increasing of immunosensor based on OWLS technique for
attention to the development of basic components determining Aflatoxin B1 and Ochratoxin A,
and devices as a whole. Bioengineering will have a Biosens. Bioelectron, 22, (6), 797-802 (2007).
key role in this area in applying knowledge in order Agui, L., Yanez-Sedeno P., Pingarron, J. M., Role of
to improve sample data analysis and calibration. carbon nanotubes in electroanalytical chemistry:
Immunosensor R&D is highly multidisciplinary. A review, Anal. Chim. Acta, 622, (1-2), 11-47
A range of molecules with bio-recognition (2008).
properties can be used as bio-sensitive elements. A Alocilja, E. C., Radke, S. M., Market analysis of
wide range of transducers is also available for biosensors for food safety, Biosens. Bioelectron.
creating new immunosensitive devices. 18, (5-6), 841-846 (2003).
Nowadays, there is a great demand for the Aluoch, A. O., Sadik, O. A., Bedi, G., Development
production of antibodies with association of desired of an oral biosensor for salivary amylase using a
antigen. However, it is necessary to overcome monodispersed silver for signal amplification,
several challenges. The in vivo use of antibodies Anal. Biochem., 340, (1), 136-144 (2005).
requires protein stability, as well as antigen affinity. Babacan, S., Pivarnik, P., Letcher, S., Rand, A. G.,
The industrial scale production of antibodies is not Evaluation of antibody immobilization methods
yet massively accessible; additional advances in for piezoelectric biosensor application,
biochemical engineering technologies will be Biosens.Bioelectron., 15, 615–621 (2000).
necessary. It is necessary that alternative Bange, A., Halsall, H. B., Heineman, W. R.,
immunoassays - i.e., using immunosensors - be Microfluidic immunosensor systems, Biosens.
easily adapted for the analysis of a wide analyte Bioelectron., 20, (12), 2488-2503 (2005).
range, through the modification of antibodies used in Bao, W., Strömblad, S., Use of an Immobilized
the assay or by means of the employment of specific Monoclonal Antibody to Examine Integrin α5β1
labels. However, micro and nanotechnology not only Signaling Independent of Cell Spreading, Biol.
contribute to device miniaturization, but also Proced., 4, (1), 81-87 (2002).
contribute fundamentally to the development of Bergveld, P., A critical evaluation of direct electrical
modern principles of biosensitivity. protein detection methods, Biosens. Bioelectron.,
6, (1), 55-72 (1991).
Bojorge-Ramirez, N. I., Medeiros Salgado, A.,
ACKNOWLEDGEMENTS Valdman, B., Amperometric Immunosensor for
Detecting Schistosoma mansoni Antibody, Assay
Financial support from Conselho Nacional de and Drug Development Technologies, 5, (5), 673-
Desenvolvimento Científico e Tecnológico, CNPq 682 (2007).
(Proc. No. Proc: 473233/2007-0) and Fundação de Brahim, S., Wilson, A. M., Narinesingh, D., Iwuoha
Amparo à Pesquisa do Estado do Rio de Janeiro E., Guiseppi-Elie, A., Chemical and Biological
(Proc. No. E26/150.997/2006) are gratefully Sensors Based on Electrochemical Detection
acknowledged. Also, we are thankful to Prof. Eveline Using Conducting Electroactive Polymers,
Gomes Vasconcelos from the Department of Microchim. Acta, 143, 123–137 (2003).
Biochemical/ICB/UFJF for helpful discussion and Bratov, A., Ramón-Azcón, J., Abramova, N.,
cooperation to the immunoassays knowledge. Merlos, A., Adrian, J., Sánchez-Baeza, F., Marco,
M. P., Domínguez, C., Three-dimensional
interdigitated electrode array as a transducer for
REFERENCES label-free biosensors, Biosens. Bioelectron., 24,
(4), 729-735 (2008).
Abdel-Hamid, I., Atanasov, P., Ghindilis, A. L., Briand, E., Salmain, M., Herry J-M, Perrot, H.,
Wilkins, E., Development of a flow-through Compère, C., Pradier, C. M., Building of an
immunoassay system, Sens. Actuators B, 49, immunosensor: How can the composition and
202–210 (1998). structure of the thiol attachment layer affect the
Abdel-Hamid, I., Ivnitski, D., Atanasov, P., Wilkins, immunosensor efficiency?, Biosens. Bioelectron.,
E., Highly sensitive flow-injection immunoassay 22, (3), 440-448 (2006).
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
244 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
Campanella, L., Martini, E., Tomassetti, M., New Curulli, A., Valentini, F., Orlanducci, S., Terranova,
immunosensor for Lactoferrin determination in M. L., Paoletti, C., Palleschi, G., Electrosynthesis
human milk and several pharmaceutical dairy of non conventional-polymer nanotubules: a new
milk products recommended for the unweaned nanostructured material for analytical
diet, Journal of Pharmaceutical and Biomedical applications, Sens. Actuators, B: Chemical, 100,
Analysis 48, 278–287 (2008). (1-2), 65-71 (2004).
Campàs, M., Iglesia, P., Berred, M., Kaned, M., D’Amico, A., Di Natale, C., Martinelli, E., Sandro,
Diogène, J., Martya, J. L., Enzymatic recycling L., Baccarani, G., Sensors small and numerous:
based amperometric immunosensor for the always a winning strategy? Sens. Actuators, B:
ultrasensitive detection of okadaic acid in Chemical, 106, (1), 29, 144-152 (2005).
shellfish, Biosens. Bioelectron. 24, 716–722 Darain, F., Park, S. U., Shim, Y. B., Disposable
(2008). amperometric immunosensor system for rabbit
Cantarero, L. A., Butler, J. E., Osborne, J. W., The IgG using a conducting polymer modified
adsorptive characteristics of proteins for screenprinted electrode, Biosens. Bioelectron.,
polystyrene and their significance in solid-phase 18, 773–780 (2003).
immunoassays, Anal. Biochem., 105, (1), 375- Divya, P. S., Savitri, D., Mitra, C. K., Covalent
382 (1980). enzyme immobilization onto glassy carbon
Carnes, E., Wilkins, E., The Development of A New, matrix-implication in biosensor design, Biosci.,
Rapid, Amperometric Immunosensor for the 23, (2), 131-136 (1998).
Detection of Low Concentrations of Bacteria Part Dong, H., Li, C. M., Zhang, Y. F., Cao, X. D., and
I: Design of the Detection System and Ye Gan, Screen-printed microfluidic device for
Applications, American J. Applied Sci., 2, (3): electrochemical immunoassay, Lab Chip, 7, 1752
597-606, (2005). – 1758 (2007).
Chen, G., Lin, Y. and Wang, J., Monitoring D'Orazio, P., Biosensors in clinical chemistry, Clin.
environmental pollutants by microchip capillary Chim. Acta, 334, (1-2), 41-69 (2003).
electrophoresis with electrochemical detection, Dursun, Z., Sahbaz, Ì., Ertas, F. N., Nisli, G.,
Talanta, 68, 497–503 (2006). Voltammetric and Flow Injection Amperometric
Chen, J., Yan, F., Dai, Z., Ju, H., Reagentless Determination of Cysteine at a Glassy Carbon
amperometric immunosensor for human chorionic Electrode in the Presence of Copper(II) Ions,
gonadotrophin based on direct electrochemistry of Turk J Chem., 27, 513-519 (2003).
horseradish peroxidase, Biosens. Bioelectron., 21, 2, Dzantiev, B. B., Yazynina, E. V., Zherdev, A. V.,
(15), 330-336 (2005). Plekhanova, Yu. V., Reshetilov, A. N., Chang, S.
Choi, M. M. F., Application of a long shelf-life C., McNeil, C. J., Determination of the herbicide
biosensor for the analysis of l-lactate in dairy chlorsulfuron by amperometric sensor based on
products and serum samples, Food Chemistry separation-free bienzyme immunoassay, Sens.
92(3), 575-581 (2005). Actuators, B: Chemical, 98, (2-3), 15, 254-261
Chu, X, Xiang Z. F, Fu, X., Wang, S. P., Shen, G. L., (2004).
Yu, R. Q., Silver-enhanced colloidal gold Erdem, A., Pabuccuoúlu, A., Merüc, B., Kerman, K.,
metalloimmunoassay for Schistosoma japonicum Özsöz, M., Electrochemical Biosensor Based on
antibody detection, J. Immunol. Methods, 301, Horseradish Peroxidase for the Determination of
(1-2), 77-88 (2005). Oxidizable Drugs, Turk J Med Sci, 30, 349-354
Clark, C., Jr., Lyons, C., Electrode System for (2000).
continuous monitoring in cardiovascular surgery, Escamilla-Gómez, V., Campuzano, S., Pedrero, M.,
Ann. N. Y. Acad. Sci., 102, 29 (1962). Pingarrón, J. M., Electrochemical immunosensor
Clark, R. A., Zerby, S. E., Ewing, A. G., designs for the determination of Staphylococcus
“Electrochemistry in Neuronal Microenvironments”, aureus using 3,3-dithiodipropionic acid
In: Electroanalytical Chemistry, A. J. Bard and I. di(Nsuccinimidyl ester)-modified gold electrodes,
Rubinstein (Eds.), Vol. 20, pp. 227–294,Marcel Talanta 77, 876–881 (2008).
Dekker, New York (1998). Farré, M., Kantiani L., Barcelo, D., Advances in
Cooreman, P., Thoelen, R., Manca, J., vandeVen, immunochemical, technologies for analysis of
M., Vermeeren, V., Michiels, L., Ameloot, M., organic pollutants in the environment, Trends in
Wagner, P., Impedimetric immunosensors based Analytical Chemistry, 26, 7, 1100-1112 (2007).
on the conjugated polymer PPV, Biosens. Feng, K., Kang, Y. Zhao, J.-J., Liu Y-L, Jiang, J-H,
Bioelectron., 20, (10), 2151-2156 (2005). Shen, G-L, Yu, R-Q, Electrochemical
immunosensor with aptamer-based enzymatic probe, Analytical Biochemistry, 344, (1), 122-129
amplification, Anal. Biochem. 378, 38–42 (2008). (2005).
Fernández-Sánchez, C., González-Garcia, A., Costa- Huang Y., Wen, Q., Jiang, J.-H, Shen, G.-L, Yu, R-Q,
Garcia, A., AC voltammetric carbon paste A novel electrochemical immunosensor based on
enzyme immunosensors, Biosens. Bioelectron, hydrogen evolution inhibition by enzymatic copper
14, 917–924 (2000). deposition on platinum nanoparticlemodified
Fowler, J. M., Wong, D. K. Y., Halsall, H. B., electrode. Biosens. Bioelectron. 24, 600–605
Heineman, W. R., Recent developments in (2008).
electrochemical immunoassays and immunosensors, Ionescu, R. E., Gondran, C., Cosnier, S., Gheber, L.
in “Electrochemical Sensors, Biosensors and their A., Marks, R. S., Comparison between the
Biomedical Applications”, Xueji Zhang, performances of amperometric immunosensors
Huangxian Ju, Joseph Wang (eds.), 115-143 for cholera antitoxin based on three enzyme
(2008). markers, Talanta, 66, (1), 15-20 (2005).
Gaag, van der B., Spath, S., Dietrich, H., Stigter, E., Iyer, P. V., Ananthanarayan, L., Enzyme stability
Boonzaaijer, G., Osenbruggen, T., Koopal, K., and stabilization-Aqueous and non-aqueous
Biosensors and multiple mycotoxin analysis, environment, Process Biochemistry, 43, (10),
Food Control, 14, 251–254 (2003). 1019-1032 (2008).
Ghindilis, A. L., Atanasov, P.,Wilkinst, M., Wilkins, Jablecki, M., Gough, D. A., Simulations of the
E., Immunosensors: electrochemical sensing and frequency response of implantable glucose
other engineering approaches, Biosens. sensors, Anal Chem., 72, (8), 1853-1859 (2000).
Bioelectron., 13, (1), 113-131 (1998). Jacobson, K. B., Biosensors and Other Medical and
Ghindilis, A. L, Atanasov, P., Wilkins, E., Environmental Probes, Home page:
Enzymecatalyzed direct electron transfer: http://www.ornl.gov/info/ornlreview/rev29_3/text
fundamentals and analytical applications, /contents.htm. (Consulted in July 2008).
Electroanalysis, 9, 661–674 (1997). Jamal, M., Sarac, A. S., Magner, E., Conductive
González-Martínez, M. A., Penalva, J., Rodríguez- copolymer-modified carbon fibre
Urbis, J. C., Brunet, E., Maquieira A., Puchades microelectrodes: electrode characterisation and
R., Immunosensors for pollutants working in electrochemical detection of p-aminophenol,
organic media. Study of performances of different Sens. Actuators B: Chemical, 97, (1), 59-66
tracers with luminescent detection, Anal. Bioanal. (2004).
Chem. 384, 1540–1547 (2006). Jiang, X., Li, D., Xu, X., Ying, Y., Li, Y., Ye, Z.,
Gooding, J. J., Wasiowych, C., Barnett, D., Hibbert, D. Wang, J., Immunosensors for detection of
B., Barisci, J. N., Wallace, G. G., Electrochemical pesticide residues, Biosens. Bioelectron., 23, (11),
modulation of antigen–antibody binding, Biosens. 1577-1587 (2008).
Bioelectron. 20, 260–268 (2004). Kahlert, S., Reiser, G., Glial Perspectives of
Gosling, J. P., Enzyme immunoassay: with and metabolic states during cerebral hypoxia –
without separation, in Principles and practice of calcium regulation and metabolic energy, Cell
immunoassay, 2nd [rev.] Ed., C. P. Price and D. Calcium, 36, (3-4), 295-302 (2004).
J. Newman, Eds., Stockton Press:London. 349, Kandimalla, V. B., Neeta, N. S., Karanth, N. G.,
351-388 (1997). Thakur, M. S., Roshini, K. R., Rani, B. E. A.,
Harlow, E., Lane, D., In: Antibodies: A laboratory Pasha, A., Karanth, N. G. K., Regeneration of
Manual. CRC Press, Cold Spring Harbor, New ethyl parathion antibodies for repeated use in
York, 553 (1988). immunosensor: a study on dissociation of
He, P., Wang, Z., Zhang, L., Yang, W., Development antigens from antibodies, Biosens. Bioelectron.
of a label-free electrochemical immunosensor based 20, (4), 903-906 (2004).
on carbon nanotube for rapid determination of Khorasani-Motlagh, M., Noroozifar, M.,
clenbuterol, Food Chemistry, 112, 707–714 (2009). Electrocatalytic Determination of Ascorbic Acid
Hirst, E. R., Yuan, Y. J., Xu, W. L., Bronlund, J. E., Using Glassy Carbon Modied with Nickel(II)
Bond-rupture immunosensors – A review, Biosens. Macrocycle Containing Dianionic Tetraazaannulene
Bioelectron., 23, 12, 1759-1768 (2008). Ligand, Turk J Chem, 28, 369–378 (2004).
Hoefelschweiger, B. K., Duerkop, A., Wolfbeis, O. Kissinger, P. T., Biosensors - A perspective,
S., Novel type of general protein assay using a Biosens. Bioelectron., 20, (12), 2512-2516
chromogenic and fluorogenic amine-reactive (2005).
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
246 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
interactions with protein kinases using biosensor application to bioaffinity separation. Ann N Y
technology. Anal. Biochem., 340, (2), 15, 359- Acad Sci. PMID: 2076021 PubMed-indexed for
368 (2005). Medline,(http://www2.kobeu.ac.jp/~katsuda/KRes
Ordóñez, S. S., Fàabregas, E., New antibodies _area/Kaffinity_puriE.html, consulted in May/
immobilization system into a graphite– 2007) (1990).
polysulfone membrane for amperometric Sadana, A., Engineering Biosensors: Kinetics and
immunosensors Biosens. Bioelectron. 22, 965– Design Applications, Academic Press, Chs 1, 4
972 (2007). and 13 (2002).
Pandey, P. C., Upadhyay, S., Shukla, N. K., Sharma, Sadik, O. A., Wanekaya, A. K., Andreescu, S.,
S., Studies on the electrochemical performance of Advances in analytical technologies for
glucose biosensor based on ferrocene environmental protection and public safety, J.
encapsulated ORMOSIL and glucose oxidase Environ. Monit. 6, (6), 513 – 522 (2004).
modified graphite paste electrode, Biosens. Sadik, O. A., Yan, F., Electrochemical biosensors for
Bioelectron., 18, 1257-1268 (2003). monitoring the recognition of glycoprotein–lectin
Popescu, C. I., Zetterberg, G., Lo Gorton, Influence interactions Analytica Chimica Acta, 588, 292–
of graphite powder, additives and enzyme 296 (2007).
immobilization procedures on a mediatorless Salinas, E., Torriero, A. A. J., Sanz, M. I., Battaglini, F.,
HRP modified carbon paste electrode for Raba, J., Continuous-flow system for horseradish
amperometric flow-injection detection of H2O2, peroxidase enzyme assay comprising a packed-
Biosens. Bioelectron., 10, 443-461 (1995). column, an amperometric detector and a rotating
Prieto-Simón, B., M. Campàs, 1, Andreescu, S. and bioreactor, Talanta 66, (1), 92-102 (2005).
Marty, J. L., Trends in Flow-based Biosensing Sarkar, P., Turner, A. P. F., Application of dual-step
Systems for Pesticide Assessment, Sensors, 6, potential on single screen-printed modified
1161-1186 (2006). carbon paste electrodes for detection of amino
Protein structure database: http://www.rcsb.org/pdb/ acids and proteins, Fresenius J. Anal Chem 364,
and http://www.biology.arizona.edu/immunology 154–159 (1999).
/tutorials/ antibody/structure.html Schmidt, H.-L., Biosensors and flow injection
Quan, D., Kim, Y., Shin, W., Characterization of an analysis in bioprocess control, J. Biotechnol., 31,
amperometric laccase electrode covalently (3), v-vi. (1993).
immobilized on platinum surface, J. Electroanal. Schuetz, A. J., Winklmair, M., Weller, M. G.,
Chem. 561, 181–189 (2004a). Niessner, R., Selection of hapten structures for
Quan, D., Kim, Y., Yoon, K. B., Shin, W., Assembly indirect immunosensor arrays, Fresenius J Anal
of Laccase over Platinum Oxide Surface and Chem, 363, 625–631 (1999).
Application as an Amperometric Biosensor, Bull. Schugerl, K., Progress in monitoring, modeling and
Korean Chem. Soc., 23, (3) 385 – 390 (2002). control of bioprocesses during the last 20 years,
Quan, D., Shin, W., Modification of electrode Journal of Biotechnology, 85, (2) 149-173 (2001).
surface for covalent immobilization of laccase, Shan, G., Lipton, C., Gee, S. J., Bruce, D.,
Mater. Sci. Eng. C, 24, 113–115 (2004b). Immunoassay, biosensors and other
Rabinovich, L., Lev, O., Sol-Gel Derived Composite nonchromatographic methods Hammock In:
Ceramic Carbon Electrodes, Electroanalysis, 13, Handbook of Residue Analytical Methods for
(4), 265 – 274 (2001). Agrochemicals p. 623–679, John Wiley & Sons,
Rao, V. K., Rai, G. P., Agarwal, G. S., Suresh, S., Ltd, Chichester (2002).
Amperometric immunosensor for detection of Shen, G., Wang, H., Tan, S., Li, J., Shen, G., Yu, R.,
antibodies of Salmonella typhi in patient serum, Detection of antisperm antibody in human serum
Anal. Chim. Acta, 531, (2), 28, 173–177 (2005). using a piezoelectric immunosensor based on
Richards, E., Bessant, C., Saini, S., Multivariate Data mixed self-assembled monolayers, Anal. Chim.
Analysis in Electroanalytical Chemistry, Acta, 540, (2), 279-284 (2005).
Electroanalysis 14, (22), 1533-1542 (2002). Shi, J., Zhu, Y., Zhang, X., Baeyens, W. R. G.,
Rodriguez-Mozaz, S., López de Alda, M. J., Marco, García-Campaña, A M., Recent developments in
M. P., Barceló, D., Biosensors for environmental nanomaterial optical sensors, TrAC, Trends Anal.
monitoring: A global perspective, Talanta, 65, Chem, 23, (5), 351-360 (2004).
291–297 (2005). Shipway, P. H., Ngao, N. K., Microscale abrasive
Sada, E., Katoh, S., Kondo, A., Sohma, Y., wear of polymeric materials Wear 255, (1-6),
Characterization of the immunointeraction and its 742-750 (2003).
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009
248 N. Bojorge Ramírez, A. M. Salgado and B. Valdman
Shriver-Lake, L. C., Donner, B., Edelstein, R., Definitions and Classification, Pure Appl. Chem.
Breslin, K., Bhatia, S. K., Ligler, F. S., Antibody 71 (12), 2333 -2348 (1999).
immobilization using heterobifunctional Tijssen, P., Practice and theory of enzyme
crosslinkers, Biosens. Bioelectron., 12, (11), immunoassays. In Laboratory Techniques in
1101-1106 (1997). Biochemistry and Molecular Biology, vol. 15.
Skládal, P., Piezoelectric Quartz Crystal Sensors Elsevier, Amsterdam, (1985).
Applied for Bioanalytical Assays and Tschmelak, J., Proll, G., Riedt, J., Kaiser, J.,
Characterization of Affinity Interactions, J. Braz. Kraemmer, P., Barzaga, L., Wilkinson, J. S., Ping
Chem. Soc. 14, (4), 491-502 (2003). Hua, Hole, J. P., Nudd, R., Jackson, M.,
Skottrup, P. D., Nicolaisen, M., Justesen, A. F., Abuknesha, R., Barcelo, D., Rodriguez-Mozaz,
Towards on-site pathogen detection using S., Lopez de Alda, M. J., Sacher, F., Stien, J.,
antibody-based sensors, Biosens. Bioelectron, 24, Slobodnik, J., Oswald, P., Kozmenko, H.,
(3), 339-348 (2008). Korenkova, E., Tothova, L., Krascsenits, Z.,
Stefan, R. I., Staden, J. F., Aboul-Enein, H. Y., Gauglitz, G., Automated Water Analyser
Estimation of uncertainties for the application of Computer Supported System (AWACSS) Part I:
electrochemical sensors in clinical analysis, Project objectives, basic technology,
Accred Qual Assur, 8, 86–89 (2003). immunoassay development, software design and
Stiene, M., Bilitewski, U., Electrochemical networking, Biosens. Bioelectron. 20, (8), 1499-
characterization of screen-printed carbonaceous 1508, (2005).
electrodes for the determination of peroxidase Turner, A. P. F., Karube, I., Wilson, G. S.,
activity in novel screen-printed flow-through Biosensors: Fundamentals and Applications, eds.,
modules, Anal Bioanal Chem 372, 240–247 (2002). Oxford University Press: Oxford, chap. 2 (1987).
Stolper, S. I., Mark, S. S., Park, J. Y. and Kricka, L. Velasco-Garcia, M. N., Mottram, T., Biosensor
J., Nanotechnology and Immunoassay, Clinical Technology addressing Agricultural Problems,
Chemistry 53, 1874 (2007). Biosystems Engineering, 84, (1), 1–12 (2003).
Stradiotto, N. R., Yamanaka, H., Valnice, M., Walcarius, A., Analytical Applications of Silica-
Zanoni, B., Electrochemical Sensors: A Powerful Modified Electrodes – A Comprehensive Review,
Tool in Analytical Chemistry, J. Braz. Chem. Electroanalysis, 10, (18), 1217-235 (1998).
Soc. 14, (2), 159-173 (2003). Wang, C., Wang, H., Wu, Z., Shen, G., Yu R., A
Su, X., Chew, F. T., Li, S. F. Y., Design and piezoelectric immunoassay based on
Application of Piezoelectric Quart Crystal–based selfassembled monolayers of cystamine and
Immunoassay, Anal. Sci., 16, 107-114 (2000). polystyrene sulfonate for determination of
Subramanian, G., In: Antibodies Production and Schistosoma japonicum antibodies, Anal Bioanal
Purification: Production and Purification, Vol 1 Chem, 373: 803–809 (2002).
G. Subramanian (Editor) (2004). Wang, H., Meng, S., Guo, K., Liu, Y., Yang, P., Zhong,
Susmel S., Toniolo R., Pizzariello A., Dossi N., W., Liu, B., Microfluidic immunosensor based on
Bontempell G., A piezoelectric immunosensor stable antibody-patterned surface in PMMA
based on antibody entrapment within a non-totally microchip, Electrochemistry Communications, 10, 3,
rigid polymeric film Sensors and Actuators B: 447-450 (2008).
Chemical, 111-112, (11), 331-338 (2005). Wang, J., Tian, B. M., Rogers, K. R., Thick film
Suzuki, M., Ozawa, F., Sugimoto, W., Aso, S., electrochemical immunosensor based on stripping
Miniaturization of SPR Immunosensors, Anal. potentiometric detection of a metal ion label,
Sci. 17 Supplement i265 - i267 (2001). Anal. Chem., 70, 1682-1685 (1998).
Tang, D., Yuan, R. and Chai, Y., Magnetic Control Wang, S., Wu, Z., Zhang, F. Q. S., Shen, G., Yu, R.,
of an Electrochemical Microfluidic Device with A novel electrochemical immunosensor based on
an Arrayed Immunosensor for Simultaneous ordered Au nano-prickle clusters, Biosens.
Multiple Immunoassays, Clinical Chemistry 53, Bioelectron., 24, (4), 1026-1032 (2008).
(7), 1323–1329 (2007). Wang, Z., Tua, Y., Liu, S., Electrochemical
Tedeschi, L., Domenici, C., Ahluwalia, A., Baldini, immunoassay for α-fetoprotein through a
F., Mencagli, A., Antibody immobilisation on phenylboronic acid monolayer on gold, Talanta,
fibre optic TIRF sensors, Biosens. Bioelectron. 77, 815–821 (2008).
19, 85-93 (2003). Warsinke, A., Benkert, A., Scheller, F. W.,
Theâvenot, D. R., Toth, K., Durst, R. A., Wilson, G. Electrochemical immunoassays, Fresenius J.
S., Electrochemical Biosensors: Recommended Anal. Chem., 366, 622–634 (2000).
Wightman, R. M., Runnels, P., Troyer, K., Analysis Zhou, Y.-M., Wu, Z.-Y., Shen, G.-L., Yu Ru-Qin,
of chemical dynamics in microenvironments, An amperometric immunosensor based on
Anal. Chim. Acta, 400, (1-3), 5-12 (1999). Nafionmodified electrode for the determination of
Wilson, G. S., Gifford, R., Biosensors for real-time Schistosoma japonicum antibody, Sens. Actuators
in vivo measurements, Biosens. Bioelectron. 20, B: Chem. 89, 292–298 (2003b).
(2), 2388-2403 (2005). Zhou, Y.-M, Wu Zhao-Yang, Shen G.-L., Yu, Ru-Qin,
Wu, Y., Zheng, J., Li, Z., Zhao, Y., Zhang, Y., A An amperometric immunosensor based on a
novel reagentless amperometric immunosensor Conducting Electrode for the Determination of
based on gold nanoparticles/TMB/Nafion- Schistosoma japonicum antigen. Anal. Sci. 18, 155-
modified electrode, Biosens. Bioelectron, 24, (5), 159 (2002).
1389-1393, (2009). Zhu, P., Shelton, D. R., Karns, J. S., Sundaram, A.,
Yoon, H. C., Yang, H., Byun, S. Y., Ferritin Li S., Amstutz, P., Tang, C, Detection of water-
Immunosensing on Microfabricated Electrodes borne E. coli O157 using the integrating
Based on the integration of immunoprecipitation waveguide biosensor, Biosens. Bioelectron. 21,
and electronical signaling reactions. Anal. Sci. (4), 678-683 (2005).
20, 1249–1253 (2004). Zhuo, Y., Yuan, R., Chai, Y., Tang, D., Zhang Y, Wang
Zhou, Y.-M, Hu, S. Q., Shen, G. L, Qin, Y. R., An N., Li X., Zhu Q., A reagentless amperometric
amperometric immunosensor based on an immunosensor based on gold nanoparticles/
electrochemically pretreated carbon-paraffin thionine/Nafion-membrane modified gold electrode
electrode for complement III (C3) assay, Biosens. for determination of α-1-fetoprotein, Electrochemistry
Bioelectron. 18, 473-481 (2003a). Communications, 7 (4), 355-360 (2005).
Brazilian Journal of Chemical Engineering Vol. 26, No. 02, pp. 227 - 249, April - June, 2009