Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Atomic line absorption spectra are used for analyzing various metals. Atomic absorption
spectrophotometry ("AAS") offers sensitivity, selectivity and simplicity and was developed
by CSIRO Melbourne scientists around 1955. "AAS" is an important tool in analytical
laboratories. Using this technique, students will be determine Pb and Zn in water samples by
flame atomic absorbance spectroscopy. Students will also be familiar with QA/QC for the
analytical procedure through the example of analyzing Pb and Zn in soluble species and total
concentrations in water samples.
II. Introduction
Flame atomic absorption spectroscopy (F-AAS) is a very common technique for detecting
metals and metalloids in solid and aqueous samples. It is very reliable and simple to use. The
technique is based on the fact that ground state metals absorb light at specific wavelengths.
Metal ions in a solution are converted to atomic state by means of a flame. Light of the
appropriate wavelength is supplied and the amount of light absorbed can be measured against
a standard curve.
The sample may be an aqueous or organic solution, indeed it may even be solid provided it
can be dissolved successfully. The liquid is drawn in to a flame where it is ionised in the gas
phase.
Light of a specific wavelength appropriate to the element being analyzed is shone through the
flame, the absorption is proportional to the concentration of the element. Quantification is
achieved by preparing standards of the element.
The technique of (FAAS) requires a liquid sample to be aspirated, aerosolized, and mixed
with combustible gases, such as acetylene and air or acetylene and nitrous oxide. The mixture
is ignited in a flame whose temperature ranges from 2100 to 2800 oC. During combustion,
atoms of the element of interest in the sample are reduced to free, unexcited ground state
atoms, which absorb light at characteristic wavelengths. The characteristic wavelengths are
element specific and accurate to 0.01-0.1nm. To provide element specific wavelengths, a light
beam from a lamp whose cathode is made of the element being determined is passed through
the flame. A device such as a photonmultiplier can detect the amount of reduction of the light
intensity due to absorption by the analyte, and this can be directly related to the amount of the
element in the sample.
1
Flame atomic absorption hardware is divided into six fundamental groups that have two major
functions: generating atomic signals and signal processing. Signal processing is a growing
additional feature to be integrated or externally fitted to the instrument. The instrument parts
are shown in figure 1 and 2.
A cathode lamp (1), shown in figure 3, is a stable light source, which is necessary to emit the
sharp characteristic spectrum of the element to be determined. A different cathode lamp is
needed for each element, although there are some lamps that can be used to determine three or
four different elements if the cathode contains all of them. Each time a lamp is changed,
proper alignment is needed in order to get as much light as possible through the flame, where
the analyte is being atomized, and into the monochromator.
2
Fig.3: Hollow cathode lamp
The atom cell (2), shown in figure 4, is the part with two major functions: nebulization of
sample solution into a fine aerosol solution, and dissociation of the analyte elements into free
gaseous ground state form. Not all the analyte goes through the flame, part of it is disposed as
shown in the figure.
As the sample passes through the flame, the beam of light passes through it into the
monochromator (3). The monochromator isolates the specific spectrum line emitted by the
3
light source through spectral dispersion, and focuses it upon a photomultiplier detector (4),
whose function is to convert the light signal into an electrical signal.
The processing of electrical signal is fulfilled by a signal amplifier (5). The signal could be
displayed for readout (6), or further fed into a data station (7) for printout by the requested
format.
F-AAS is also based on transmitting light through the sample and measuring the fraction of
light absorbed. Mathematically, the measurement can be expressed as:
In Eq. 1, T is termed the transmittance, P denotes the light intensity transmitted through the
sample, and P0 is the light intensity incident on the sample. The P0 value is measured by use
of a blank solution that contains the sample matrix but contains no analyte. The values T, P,
and P0 are all a function of λ, the wavelength of the light. The measured transmittance can be
In Eq. 2, A is termed the absorbance, c is the analyte concentration, and ε (absorptivity) and b
(optical path length) are constants that define the linear relationship between A and c.
Absorptivity is an intrinsic property of the analyte, and the path length is a characteristic of
the instrumental measurement. The value of εb is determined by a calibration procedure that
employs a set of known standards that bracket the desired concentration range.
Apparatus: - All sample plastic containers must be prewashed with detergents, acids, and
water. Both plastic and glass containers are suitable.
Provided Reagents: - Reagent Water shall be interference free. All references to water in the
method refer to reagent water unless otherwise specified.
- Nitric acid (concentrated), HNO3. Acid should be analyzed to determine level of impurities.
If method blank is < MDL, then acid can be used.
- Water Sampling:
+Total recoverable metals - All samples must be acidified at the time of collection with
HNO3 (5 mL/L).
+ Dissolved metals - All samples must be filtered through a 0.45-m filter and then
acidified at the time of collection with HNO3 (5 mL/L).
4
To be prepared:
Concentrations of the metal ion standard solutions need to cover a range of 0 to 25 mg/L (ie. 0
to 25 ppm). Some metal ion stock solutions (250 mg/L or 1000 mg/L) will already have been
prepared for you and simply require dilution with 0.002M to the appropriate final
concentrations.
Note: High levels of silicon, copper, or phosphate may interfere to the determination of Zn.
Addition of strontium (1,500 mg/L) removes the copper and phosphate interference. Zinc is a
universal contaminant, and great care should be taken to avoid contamination.
Sample preparation:
* Dissolved metals - The sample is filtered through a 0.45-m filter at the time of collection
and the liquid phase is then acidified at the time of collection with nitric acid. Samples for
dissolved metals do not need to be digested as long as the acid concentrations have been
adjusted to the same concentration as in the standards.
* Total recoverable metals: The entire sample is acidified at the time of collection with nitric
acid. At the time of analysis the sample is heated with acid and substantially reduced in
volume. The digestate is filtered and diluted to volume, and is then ready for analysis.
Interferences: The analyst should be cautioned that this digestion procedure may not be
sufficiently vigorous to destroy some metal complexes.
Precipitation will cause a lowering of the silver concentration and therefore an inaccurate
analysis.
Together with unknown sample, add the amount of several ppm of mixture containing heavy
metals to the samples and digest follow the above procedure to evaluate the recovery rate.
CAUTION: Do not boil. Antimony is easily lost by volatilization from hydrochloric acid
media.
- Remove the beaker and allow to cool. Wash down the beaker walls and watch glass with
water and, when necessary, filter or centrifuge the sample to remove silicates and other
insoluble material that could clog the nebulizer.
Filtration should be done only if there is concern that insoluble materials may clog the
nebulizer; this additional step is liable to cause sample contamination unless the filter and
filtering apparatus are thoroughly cleaned and pre-rinsed with dilute HNO3.
5
2. Quality control
- For each analytical batch of samples processed, blanks should be carried throughout the
entire sample preparation and analytical process. These blanks will be useful in determining if
samples are being contaminated.
Calibration blank is used for establishing the calibration curve. The calibration blank
consists of the same concentration(s) of the acid(s) used to prepare the final dilution of
the calibration standards. In addition, an appropriate concentration of internal standard
is added.
6
Procedural blank (or reagent blank) is used to monitor for possible contaminations
resulting from the sample preparation procedure. The procedural blank must be carried
through the same procedure and contain the same volume of reagents as the sample
solution. In addition an appropriate concentration of internal standard is added.
Rinse blank is consist of 1-2 % (v/v) HNO3 and is used to flush the sample
introduction system between standards and samples.
The operating procedure will vary between instrument brands, so the instrument manual
should be followed carefully. The position of observation and the fuel:oxidant ratio must be
optimised. Some general guidelines are outlined below:
Light the hollow cathode lamp or electrode discharge lamp and D2-lamp if such
background correction is used. Set the lamp current to the value specified by the
manufacturer.
Position the monocromator at wavelength 283.3 for Pb and 213.9 for Zn and choose
slit with 0.7 and slit height “high”.
Carefully balance the intensity of the hollow cathode lamp and the D2-lamp if such
background correction is used.
Align the burner head to assure that the centre of the light beam passes over the burner
slot.
Light the flame and regulate the flow of fuel and oxidant to produce an oxidising
flame (lean blue).
Aspirate samples.
Knowing the “characteristic concentration” allows the analyst to check if the instrument is
correctly optimised and performing up to specifications.
Aspirate a quality control standard for every 10th sample to check for drift.
Samples that are found to have concentration higher than the highest standard should
be diluted and reanalysed.
1. Report the experimental results. Summarize the calculations that produced your calibration
design.
2. Calculate the exact concentrations of your calibration standards. Be sure to use the
appropriate number of significant figures. data. Report values to the correct number of
significant figures.
4. Make calibration models for each set of data. Use least-squares procedures to determine the
slope and intercept of each calibration model. Report the value of r2 and the standard error of
estimate.
4. Include separate plots of absorbance (with error bars) vs. concentration (ppm) for each
calibration model.
5. Use your calibration model together with your dilution scheme to compute the weight
percent of iron in your original solid sample. Report as weight percent Fe2O3. Predict each
replicate of the unknown separately, then compute a mean, standard deviation, and 95%
confidence interval for the predicted amount. Repeat this calculation for both calibration
models.
6. The limit of detection associated with each calibration can be estimated by use of the
following procedure. The intercept of the calibration model is an estimate of the absorbance
when no metal is present in the sample (i.e., the blank signal). The standard error of estimate
is an estimate of the measurement noise (i.e., in the absence of systematic error, the variation
about the least-squares line should be due solely to instrumental noise). A widely used
definition of limit of detection is the concentration corresponding to an absorbance value
equal to the blank signal plus three times the noise level. Thus, ALOD = b + 3s (3)
where ALOD is the absorbance corresponding to the limit of detection, b is the intercept of the
calibration model, and s is the computed standard error of estimate. The limit of detection,
CLOD, can then be estimated as CLOD = (ALOD - b)/m (4)
where ALOD are as defined above and m is the computed least-squares slope of the calibration
model. Equations 3 and 4 can be combined to yield
8
CLOD = 3s/m (5)
Use your calibration data to estimate the limit of detection for each method.
7. In your discussion section, compare the atomic and molecular spectroscopic methods
8. Tabulate the mean, standard deviation, and 95% confidence interval for your absorbance-
Report the data of the calibrations with linear regression.
- Calculate the Pb, Zn content in samples by using the calibration regression. Also report
uncertainty of the sample’s data from calibration curve.
- Discuss the major factors involved in the detection to use flame of electrothermal
atomization in AAS.
- F-AAS is known as a technique with few problems related to interference effects. Do you
know what are they?
- Why do hydride techniques for As and Se and the cold vapor techniques for Hg yield better
AA detection limit than conventional flame AAS with solution nebulization?