Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
pH is a measure of the acidity or basicity of a solution. It is defined as the co
logarithm of the activity of dissolved hydrogen ions (H+). Hydrogen ion activity
coefficients cannot be measured experimentally, so they are based on theoretical
calculations. The pH scale is not an absolute scale; it is relative to a set of standard
solutions whose pH is established by international agreement.
The lower‐case letter "p" in pH stands for the negative common (base ten)
logarithm, while the upper‐case letter "H" stands for the element hydrogen. Thus,
pH is a logarithmic measurement of the number of moles of hydrogen ions (H+) per
litre of solution. Incidentally, the "p" prefix is also used with other types of chemical
measurements where a logarithmic scale is desired, pCO2 (Carbon Dioxide) and pO2
(Oxygen) being two such examples.
The logarithmic pH scale works like this: a solution with 10‐12 moles of H+ ions per
liter has a pH of 12; a solution with 10‐3 moles of H+ ions per liter has a pH of 3.
While very uncommon, there is such a thing as an acid with a pH measurement
below 0 and a caustic with a pH above 14. Such solutions, understandably, are quite
concentrated and extremely reactive. [1]
History
The concept of pH was first introduced by Danish chemist Søren Peder Lauritz
Sørensen at the Carlsberg Laboratory in 1909. It is unknown what the exact
definition of p is. Some references suggest the p stands for “Power”, others refer to
the German word “Potenz” (meaning power in German), and still others refer to
“potential”. Jens Norby published a paper in 2000 arguing that p is a constant and
stands for “negative logarithm”; which has also been used in other works. H stands
for Hydrogen. Sørensen suggested the notation "PH" for convenience, standing for
"power of hydrogen", using the cologarithm of the concentration of hydrogen ions in
solution, p[H] Although this definition has been superseded p[H] can be measured if
an electrode is calibrated with solution of known hydrogen ion concentration. [1]
1
Measurement methods
There are two methods for measuring pH; a probe and meter, and litmus paper. The
most accurate and reliable method is the probe and meter. This method is no less
convenient than the other method, but requires a more expensive piece of
equipment.
1. Litmus Paper:
The term litmus comes from an Old Norse word meaning
“to dye or color”. This is fitting since the lichens used to
make litmus have also been used to dye cloth for
hundreds of years. Very little information is available
about the beginnings of litmus. There is some data that
suggest that litmus paper was developed by J.L. Gay‐
Lussac, a French chemist during the early 1800s.
Figure(1).Litmus paper.
Litmus paper is the most recognized member of chemical indicators. Litmus changes
color when exposed to an acidic or basic solution. The simple pH scale ranges from 0‐
14 with 0 being the most acidic, 7 being neutral, and 14 being the most basic or
alkaline. Litmus paper is commonly used in educational science classes. Because it
has such wide recognition, it has become a cultural reference in our society as well.
It is common to use the term litmus test when referring to a test in which a single
factor determines the outcome.
While litmus paper is effective at indicating whether a substance is acidic or basic, it
cannot report an exact numerical pH value.
The Future
Litmus paper will most certainly continue to be used extensively in education due to
its reasonable cost and ease of use. However, some varieties of lichens are becoming
extinct. As a result, it is possible that manufacturers of litmus paper may switch to
synthetic materials in the future. This is already being done by manufacturers of
other types of pH papers. Additionally, because litmus cannot give quantitative
results, it cannot replace other pH papers and pH meters. In fact, the trend is to
make pH indicators that are even more accurate and less subjective. One such trend
is to utilize fiber optic probes in pH meters in order to make them even more
sensitive. [3]
2
2. Potentiometry method:
All potentiometry does is to measure (meter) the voltage (potential) caused by the
hydronium ion: H3O+. This new method gave accuracy, reliability and fast results.
Tools for measuring pH:
1. A pH meter (to accurately measure and transform the voltage caused by
hydronium ion into a pH value);
2. A pH electrode (to sense all the hydronium ions and to produce a potential).
3. A reference electrode (to give a constant potential no matter what the
concentration of hydronium ion is).
The pH meter
Basically, a pH meter measures the potential between the pH electrode (which is
sensitive to the hydronium ions) and the reference electrode (which doesn't care
what's in the solution).
The pH electrode
The pH electrode's potential changes with the H3O+ ion
concentration in the solution.
The clever bit is that the pH electrode only senses the
hydronium ions. This means that any voltage produced is from
hydronium ions only. This way we can relate the potential
directly to the hydronium concentration.
Figure(2). pH electrode
The reference electrode
The reference electrode supplies a “constant” value against
which we measure the potential of the pH electrode. That's the
thing about potentials, they have to be in Pairs to produce a
voltage. [2]
The function of the diaphragm on a pH electrode: Figure(3).reference electrode
The diaphragm on a pH electrode allows a flow of electrolyte from the reference
solution into the solution being measured.
This completes the electrical circuit between the reference and sensing electrode
and gives the potential difference which is measured by the pH meter. The type of
diaphragm used will influence the outflow of electrolyte into the sample, which will
in turn affect the response speed of the electrode. The common diaphragm types
are listed in table(B.1) in appendix(B) below along with their inherent properties. For
most common applications a platinum diaphragm is the one of choice.
3
Glass bulb at the bottom of electrodes
The glass membrane on pH electrodes comes in a variety of different shapes based
on the fitness for purpose of the electrode. The most common membrane shapes in
use today are outlined in table (B.2) in appendix (B).
In practice there is a trade‐off between the speed of response of the electrode and
its ruggedness, thus the application the electrode is being selected for will govern
the choice of membrane design. The classical sphere shaped membrane offers low
electrical resistance due to its large surface area but is relatively fragile. The cone
shaped membrane is seen as a universal design as it is robust, of medium resistance
and is easy to clean.
The difference between the various types of membrane glass for pH electrodes
As pH is measured in a wide variety of solutions at varying temperatures it is not
possible to formulate a single pH membrane glass which can give accurate and
reliable pH measurements in all conditions. For this reason different types of
membrane glass have been developed to give good performance in a variety of
measurement conditions. The most common membrane glass types found in Schott
electrodes are as outlined in table (B.3) in appendix (B). [5]
The classical set‐up
The classical set‐up for measuring pH consisted of a pH meter;
a pH electrode and a reference electrode. [2]
The modern set‐up
Figure(4).Classical setup
Although you can perfectly measure pH using the "classical"
set‐up, it was soon realized that the two electrodes could be
built into the same probe (although there still are two totally
separate electrodes). This is nowadays called the combined pH
electrode, which is much more practical.
The formula for pH calculation Figure(5).Modern setup
From the above we know that in order to measure pH we measure potential
differences. Nernst studied the relation between the mV (millivolts) reading
produced by a pH electrode against a standard hydrogen electrode and the pH of the
measuring solutions. Now, a standard hydrogen electrode is a rather complicated
piece of equipment. Suffice to say that it gives very well defined and reproducible
4
mV readings over a wide range of pH values and strictly responds only to hydronium
ions.
Nernst equation:
U =U 0 + (2.303RT/ziF) log(H3O+)
Where:
‐ U: is the potential measured between indicator and reference electrode.
‐ U0: is the standard potential of the electrode assembly, depends on its
construction (well made electrodes give 0 mV with pH 7 – neutral solution).
‐ R: is the gas constant (8.31441 J K‐1 mol‐1).
‐ T: is the absolute temperature in K (273.15 + t in °C).
‐ Zi: is the charge of the hydronium ion (1+) (each ion carries a single positive
charge with it).
‐ F: is Faraday constant (96 484.56 C * mol‐1).
The pH meter basically uses this equation to calculate the pH value. However, all the
individual parts within the equation have to be known (not everything is constant).
The potential is measured by the electrodes and the temperature (T) needs to be
measured as well and if there isn't any temperature sensor connected to the meter,
the temperature (T) value has to be entered. [2]
pH measurement in theory and in practice
In theory a pH electrode is so constructed that its potential in buffer pH = 7 equals 0
mV. (A buffer is a solution which keeps a given concentration of hydronium ions and
therefore a given pH). This potential is called the zero potential of a pH electrode.
This is one of the two main characteristics of a pH electrode. The second
characteristic is the electrode slope. The slope of an electrode gives the change of
potential measured between two buffers with a pH difference of 1 in comparison to
what is expected by theory (from Nernst). Nernst calculated this value as 59.16
mV/pH at 25 degrees Celsius. But this is just theory and as we all know the practice
always looks a little different. In practice the zero point as well as the slope of the
electrode differ from what they should be. This difference has to be known so that
the meter can compensate for any changes in electrode response. This way you get
“true” results. Determining the deviation of the zero point and the slope from their
theoretical values and saving these values into the meter (it works automatically) is
called calibration. [5]
How to calibrate your pH meter
1. Take the pH electrode and connect it to the meter. Remove the storage vessel and
rinse the sensing area thoroughly with distilled water. Dab the electrode with a
soft paper tissue to remove the rest of water (don't rub the electrode surface).
5
2. Dip the electrode into buffer pH = 7, so that the diaphragm of the electrode is well
immersed. If the pH electrode doesn't have an integrated temperature sensor,
connect a separate temperature sensor to the meter and dip it also into the
buffer solution. If you don't have a temperature sensor which you can connect to
the meter, use a conventional sensor and enter the temperature into the meter
when required.
3. Start the calibration on the meter (check instructions for use of the meter). You
will notice that the temperature is measured first and then a mV reading is taken,
which should be around 0 mV. The difference to 0 mV will be registered by the
meter and taken into account during later measurements.
4. Rinse the sensors thoroughly with distilled water and dab the electrode with a soft
paper tissue to remove the rest of water (don't rub the electrode surface). Dip in
buffer pH = 4 and repeat the procedure.
5. Having two mV readings, this and the previous one, the meter can establish a
linear function U = f (pH) and also calculate the slope of the electrode. The real
value differs again from the theory of 59.16 mV/pH. The meter saves the
difference and takes it into account during later measurement. (see Appendix(A)).
[2]
Temperature compensation
Temperature plays a key role in pH measurement. There are two reasons for this:
1. The pH value of the solution changes with the temperature.
2. The slope (theoretical slope) of the electrode changes with the temperature.
pH measurement should always be performed together with temperature
measurement because only pH values measured at the same temperature can be
compared.
Of course there is one exception which confirms the rule above. This is when you
calibrate. In this case both problems mentioned above are eliminated. [2]
The response time of an electrode
The response time of an electrode is the length of time necessary to get a stable
reading when the electrode is moved from one solution to another of different pH or
temperature.
Response time is dependent on the electrode type, the measuring sample,
temperature, the magnitude and direction of concentration change and the
presence of interfering substances. A slow response time can be indicative of the
incorrect selection of electrode for the intended application. [5]
6
Paper vs. Meter comparison
Litmus paper
• Advantages:
1. Easy for young children to use.
2. Does not need calibration.
• Disadvantages:
1. Resolution is not as good as meters (reads in 0.5 pH unit increments).
2. It is not temperature compensated.
pH meter
• Advantages:
1. Measures to 0.1 pH units.
2. Can be temperature compensated.
Note: avoid using meters with one point calibration.
• Disadvantages:
1. The meter must be calibrated with buffer solutions before each use.
2. More expensive than litmus paper.
3. Performance deteriorates over time.
4. Better meters have at least a two point calibration and have an automatic
temperature compensation (ATC). Buffer solutions may be ordered in liquid
or powdered form. The liquid is more expensive and has a shorter shelf life,
but may be more convenient than mixing the powdered buffers. Most
meters require the small, flat ‘watch‐type’ batteries. Although the batteries
last a long time, if the meter is turned off when not in use, it is a good idea to
have an extra set of batteries on hand. [6]
7
Some application areas:
1. pH measurement in food industry
pH clearly plays an important role in food processing. Among the reasons for
measuring pH in food processing include:
• To produce products with consistent well defined properties.
• To efficiently produce products at optimal cost.
• To avoid causing health problems to consumers.
• To meet regulatory requirements.
The pH range of foods varies considerably with some typical values of food and
related solutions. Due to the logarithmic nature of the measurement, even small
changes in pH are significant. The difference between pH 6 and pH 5 represents a
ten‐fold increase in acid concentration; a change of just 0.3 represents a doubling of
acid concentration. Variations of pH can impact flavor, consistency, and shelf‐life. [4]
Examples:
• pH Measurement in Corn Wet Milling
BACKGROUND
Corn wet‐milling is the most common method used to process corn. In this process,
corn kernels are separated into their component fractions and then further treated
to yield cornstarch, corn sweeteners, corn oil, and animal feed.
pH measurements are made throughout the milling process to optimize the product
yield. The bulk of the water used for rinsing and washing usually needs pH
adjustment to prevent altering the nominally acidic pH present during the various
steps. Rinse waters are generally recycled upstream, but must be eventually treated
(and neutralized) before disposal.
Specific pH applications include the following:
1. The pH in the steeping tank is used to control the addition of the sulfur dioxide (or
other acid) that begins to release the starch from the corn. Too much acid may
release the starch prematurely and causes corrosion of stainless steel process
lines. Not adding enough will not prepare the corn kernels adequately.
2. Starch modification is used to lower the viscosity of the starch product. This
process must be con‐ducted under controlled pH to meet product specifications.
3. The enzymes used in starch conversion are expensive and function best at well
defined pH levels (typically in the 3‐5 range depending on the enzyme used). pH
levels higher than normal will not utilize the enzymes efficiently, while operation
at lower pH may allow the reaction to proceed too far and cause plugging of
8
downstream filters. High pH values cause the product to set into a viscous gel that
is difficult to process. [7]
• pH of Cheese
pH, together with temperature, rank as the most important indicators of food quality
and safety. The pH of cheese is measured to ensure that quality standards have been
properly met. pH is also monitored at different stages of cheese processing and
transformation to guarantee safety and improve production. Along with
temperature and water activity, pH is an important determinant in the shelf life of
foods. The pH value of cheese varies not only among types of cheese, but also
between batches of the same variety. There is also variation throughout the
maturation process, decreasing early on and increasing again as the cheese matures.
Typically, the pH of cheese ranges from 5.1 to 5.7 with a few exceptions. Increasingly
pressure is being put on cheese makers to supply end product which has been pH
tested. [8]
2. Ph measurements in agriculture:
Example:
• Soil pH Measurement with a Portable Meter
Background:
Soil pH is a very useful measurement, indicative of soil chemical and biological
properties important to plant growth. The availability of plant nutrients, activities
and nature of microbial populations, solubility toxic substances and activities of
certain pesticides are all influenced by soil pH.
Considerations Involved in Measuring Soil pH:
1. All meters should be calibrated once each day before use with standard pH buffer
solutions at two pH points. Standardize the pH meter until the instrument reading
agrees with both buffer pH’s.
2. It is best to make pH measurements at approximately room temperature because
pH reading sere temperature sensitive. If possible, adjust the instrument to the
temperature of the solution being measured.
3. the most common procedure is to place the soil into a paper cup and add one
tablespoon of distilled water. Stir the sample for 15 seconds and let stand for 30
minutes.
9
4. The electrode tip must be placed in the soil slurry and not in the overlying
solution. Stir the slurry carefully and read the pH immediately.
5. To avoid abrasion of the sensing surface, prevent the electrode tip from touching
the bottom of the sample cup. Some instruments have electrode holders to
prevent abrasion. Rinse the electrodes after each soil sample. [9]
Future research and development in pH sensors
The selection of the optimum sensor(s) and analyzer for a control system is as
important as its proper configuration.
pH measurement is not simple and furthering the problem is the fact that many, if
not most, applications requiring pH measurement are challenging environments
involving high temperatures and/or process elements that can foul, coat, or poison a
pH sensor.
In the past, this has resulted in the need for frequent cleaning and replacement of
the pH sensor, sometimes once a day in very harsh circumstances, making the
measurement costly in both equipment and time.
While there is no perfect pH sensor yet today, the technology is advancing to the
point that the pH sensor is a relatively trouble‐free part of the process, so long as
one selects carefully and performs appropriate maintenance too.
The two major areas of advance have been extending the life of the pH sensors and
extending their range of applications.
The most obvious area of advance in pH technology is in the area of glass durability.
New glass formulations and low‐stress handling techniques provide exceptional
resistance to thermal and caustic degradation. This translates to less breakage from
thermal stress or shock and improved speed of response at near theoretical levels
and minimal hysteresis for fast accurate calibrations even after months of service.
In addition to the more durable glass, new sensor designs also mount the glass bulbs
in protected tips that shield the glass from direct impact while in service or during
calibration. For applications where glass‐tipped sensors cannot serve, ion selective
field effect transistor (ISFET) technology may be the best choice.
The response time of these sensors can be as much as 10 times faster than
traditional glass electrodes. This response combined with stability make ISFET
sensors appropriate for use in cold processes like brine or water for cooling.
10
Aging effects caused by temperature fluctuations are also less with ISFET sensors
providing longer intervals between maintenance and calibration. However, the ISFET
sensor does not typically survive in harsh alkaline or strongly acidic solutions as well
as glass does. It is non‐linear at the ends of the pH spectrum; hence, the ISFET sensor
has a range of two to 12 pH versus zero to 14 pH for the glass electrode sensor. If the
application requires fast response in a relatively benign environment, or does not
allow the use of glass, then an ISFET sensor is the answer.
Despite the best technology developments, sensors still need cleaning in dirty
applications. Spray cleaners attach to the sensor and provide a periodic burst of
cleaner onto the sensor electrode, removing the coating that can degrade
performance and cause dead time.
If cleaning is critical and the application is sufficiently benign to not require frequent
sensor replacement then a mechanical or pneumatically activated retraction system
may be used. These systems easily and automatically remove the sensor from the
process for cleaning and then return it smoothly, reducing impact on employee time
and reducing downtime. The systems are costly, so a cost/benefit analysis is
appropriate; a long‐life sensor that is reliable and easy to install may prove to be a
better choice.
Many of these improvements significantly forward the operation of pH sensors in a
wide range of applications. Over the last decade, more and more steps to build in
sensor diagnostics that alert the operator before a critical failure occurs have come
about. With advance notice of sensor degradation, orderly maintenance or
shutdowns are now possible.
Most analyzers today are modular in design, allowing the same analyzer to be used
with a simple change of a circuit board for multiple applications (pH, ORP,
conductivity, chlorine, turbidity, and others), which can significantly speed training
and operation.
In addition, many analyzers allow multiple inputs, permitting a single analyzer to
control more than one sensor.
Finally, pH analyzers can be equipped with advanced digital communications (HART,
Profibus, Foundation fieldbus, and others) to allow integration with central
databases eliminating “islands of automation,” speeding record keeping and trouble‐
shooting, and complying with requirements in many industries. [10]
11
References:
[1] http://en.wikipedia.org/wiki/PH
[2] http://www.brinkmann.com/utilities/download_pdf.asp?DR=downloads
%2Fpdf%2Ffaqs%2Felectrodes&FN=pH-background.pdf.
[3] http://www.madehow.com/Volume-6/Litmus-Paper.html
[4] http://www.foodmanufacturing.com/scripts/Products‐The‐Importance‐of‐pH‐
Measurement.asp
[5] http://www.reagecon.com/TechPapers/phfaqv4.pdf
[6] http://www.globe.gov/tctg/hydro_prot_ph.pdf?sectionId=152
[7] http://www.emersonprocess.com/raihome/documents/Liq_AppData_2000‐
84.pdf
[8] http://www.qclscientific.co.uk/ph‐measurement‐150‐c.asp
[9] www.rce.rutgers.edu
[10]http://www.isa.org/InTechTemplate.cfm?Section=Article_Index1&template=/Content
Management/ContentDisplay.cfm&ContentID=67497
12
Appendices
APPENDIX (A): Calibration of pH meter
Zero point adjustment
Here is a typical calibration report:
13
APPENDIX (B): diaphragm types and membrane shapes
• Table(B.1): common diaphragm types
• Table(B.2): The most common membrane shapes in use today
14
• Table(B.3): membrane glass types
15
APPENDIX (C): Data sheets
16
pH Measurement
Electrode Basics
pH electrode technology hasn’t changed therefore subject to breakage. Electrodes pH electrodes should always be stored in
much in the past 50 to 60 years. With all are designed to measure mostly aqueous a moistened condition. When not in use,
the technological advancements of the last media. They are not designed to be used it is best to store the electrode in either
30 to 40 years, pH electrode manufacturing in solvents, such as CCI4, which does not buffer 4.0 or buffer 7.0. Never store an
remains an art. The special glass body of have any free hydrogen ions. electrode in distilled or deionized water as
the electrode is blown to its configuration this will cause migration of the fill solution
by glass blowers. Not a terribly advanced The pH electrode due to the nature of its from the electrode.
nor “high tech” process but a very critical construction needs to be kept moist at all
and important step in the electrode times. In order to operate properly the
manufacturing. In fact, the thickness glass needs to be hydrated. Hydration is
of the glass determines its resistance required for the ion exchange process to
and affects its output. occur. If an electrode should become dry,
it is best to place it in some tap water for
a half hour to condition the glass.
MDS37, $750, see page B-29. pH electrodes are like batteries; they run
down with time and use. As an electrode
ages, its glass changes resistance. This
resistance change alters the electrode
potential. For this reason, electrodes
need to be calibrated on a regular basis.
Calibration in a pH buffer solution corrects
for this change. Calibration of any pH
equipment should always begin with buffer
7.0 as this is the “zero point.” The pH scale
has an equivalent mV scale. The mV scale
ranges from +420 to -420 mV. At a pH of
7.0 the mV value is 0. Each pH change
corresponds to a change of ±60 mV. As pH
values become more acidic, the mV
pH electrodes are constructed from a values become greater. For example a pH pH electrodes have a finite lifespan due
special composition glass which senses of 4.0 corresponds to a value of 180 mV. to its inherent properties. How long a pH
the hydrogen ion concentration. This glass electrode will last will depend on how it is
is typically composed of alkali metal ions. cared for and the solutions it is used to
The alkali metal ions of the glass and the measure. Typically, a gel-filled combination
hydrogen ions in solution undergo an ion pH electrode will last 6 months to 1 year
exchange reaction generating a potential PHB-212, $420, depending on the care and application.
difference. In a combination pH electrode, see page B-23. Even if an electrode is not used it still
the most widely used variety, there are ages. On the shelf, the electrode should
actually two electrodes in one body. One last approximately a year if kept in a
portion is called the measuring electrode, moistened condition. Electrode demise
the other the reference electrode. The can usually be characterized by a sluggish
potential that is generated at the junction response, erratic readings or a reading
site of the measuring portion is due to which will not change. When this occurs,
the free hydrogen ions present in solution. an electrode can no longer be calibrated.
The potential of the reference portion is
produced by the internal element in pH electrodes are fragile and have a
contact with the reference fill solution. limited lifespan. How long an electrode
This potential is always constant. In will last is determined by how well the
summary the measuring electrode probe is maintained and the pH
delivers a varying voltage and the application. The harsher the system,
reference electrode delivers a constant As pH values become more basic, the mV the shorter the lifespan. For this reason
voltage to the meter. values become more negative; pH=9 it is always a good idea to have a back-up
corresponds to -120 mV. Dual calibration electrode on hand to avoid any system
The voltage signal produced by the pH using buffers 4.0 or 10.0 provide greater down time. Calibration is also an important
electrode is a very small, high impedance system accuracy. part of electrode maintenance. This
signal. The input impedance requires assures not only that the electrode is
that it only be interfaced with equipment pH electrodes have junctions which allow behaving properly, but that the system
with high impedance circuits. The input the internal fill solution of the measuring is operating correctly. In summary these
impedance required is greater than electrode to leak out into the solution are the “electrode facts of life.”
1013 ohms. This is the reason why pH being measured. This junction can
electrodes do not interface directly will become clogged by particulates in the
all equipment. solution and can also facilitate poisoning
by metal ions present in the solution. If a
pH electrodes are available in a variety of clogged junction is suspected, it is best to
styles for both laboratory and industrial soak the electrode in some warm tap
applications. No matter their status, they water to dissolve the material and clear
are all composed of glass and are the junction.
For Sales
Our pH Field and Lab Instruments product line
B-3 continues to expand, visit omega.com for new details!
& Service
U.S.A. and Canada
Each electrode
supplied with
either BNC
(shown) or US
standard PHE-1304
connector. economy,
$32.
PHE-1335, extra long test tubes (PHE-1335-D detachable style shown), $89 + 5 = $94. See C-13 of the GREENBOOK®
All electrodes shown smaller than actual size. for all electrodes shown here.
Compound pH Value
Sulfuric Acid 0.3
Limes 2.8-3.8
Wines 3.0
Oranges 3.0-4.0
Beer 4.0-5.0
Automatic Temperature Blood, Human 7.3-7.5
59 mV per decade at 25°C (77°F) Compensation becomes
52 mV per decade at 0°C (32°F) more critical as the temperature Egg whites 7.6-8.0
74 mV per decade at 100°C (212°F) changes from 25°C (77°F), Sodium Bicarbonate 8.4
or the pH from 7.0 Ammonia 11.6
Sodium Hydroxide 14.0
+
H (Hydrogen Ions) OH– (Hydroxyl Ions)
Acid Neutral Alkaline
pH 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14
mV @ + + + + + + + – – – – – – –
25°C 414 355 296 237 177 118 59 00 59 118 117 237 296 355 414
(77°F)
Order Online
omega.com
Over 100,000 Products Available!
® To download information and to order pH Field and
Lab Instrument products online, visit omega.com
B-4
Temperature Compensation
pH Measurement Error
The millivolt output of all pH electrodes At a pH of 7 and a temperature of 25°C temperature is other than 25°C (77°F)
varies with temperature in a manner (77°F), temperature error is approximately and the pH other than 7, the temperature
predicted by theory. The magnitude of this zero. So, at any temperature when the pH error is:
variation is a function of temperature and is about 7, there is no temperature error. 0.03 pH error/pH unit/10°C (18°F).
of the pH value of the system being And, at any pH when the temperature is at The following table illustrates this
measured. 25°C (77°F), there is no error. When the combined effect:
Add the appropriate error factor to correct pH accuracy. For example, if the accuracy Any of the above devices–automatic
uncompensated readings. Correct requirement is ±0.1 pH, at a pH of 6 and compensation, manual compensation or
uncompensated readings as follows a temperature of 45°C (113°F), the error fixed resistor–operate as a function of the
(the factors from the table assume that is 0.06, well within the accuracy pH meter’s electronic circuit. As such,
the electrodes were calibrated in buffer requirements. On the other hand, with information and parts should be obtained
at 25°C/77°F): the same ±0.1 pH accuracy requirement, from the meter manufacturer.
operating at pH 10 and 55°C (131°F)
Effects of Temperature on pH would give an error of 0.27 pH and If automatic compensators are used, they
compensation should be used. should always be at the same location as
Temperatures Above 25°C: temperature the pH electrode. When electrodes are
compensation lowers high pH and raises calibrated in buffer, the temperature
low pH, resulting in value closer to neutral. When compensation is required, it can be
done in one of two ways. If the compensator also should be in the buffer.
Temperatures Below 25°C: temperature temperature fluctuates, then an automatic In a similar way, a manual temperature
compensation raises high pH (more basic) compensator should be used. If the compensator should be adjusted to reflect
and lowers low pH (more acidic), resulting temperature is constant within several the temperature to which the pH electrode
in values further away from neutral. degrees C, then a manual compensator is exposed during both calibration and
can be used. If no compensator is needed, operation.
Whether or not temperature compensation a fixed resistor can be installed across the
need be used is a matter of the needed temperature compensator terminals.
Reproduced with permission of Sensorex
For Sales
Our pH Field and Lab Instruments product line
B-5 continues to expand, visit omega.com for new details!
& Service
U.S.A. and Canada
pH Measurement Tips
The measurement of pH is very
important in many laboratories
and industries. Below are a few
guidelines to aid in making PHB-600R, $570, with pH electrode,
accurate and precise temperature probe, 2 buffer solutions
pH readings. and electrode holder (all included),
shown smaller than actual size.
See page B-25 for more information.
Meter Calibration
Frequency
For accurate results a
pH meter should be
calibrated at least once
per 8-hour shift.
Buffers
Standard buffers should
always be used for meter
calibrations. Buffers can
be purchased already
prepared or in capsule form.
Standard buffers usually
are available in three
pH values—pH 4.00,
pH 7.00, and pH 10.00.
Buffers should be stored
away from heat and in tightly
sealed containers.
Always use freshly poured
buffers for meter calibration.
All buffers should be used at room
temperature, 25°C (77°F).
Types of Calibration
5. Rinse the electrode with distilled or
One-Point Calibration deionized water. (This would be the
Meter calibration using only one buffer. procedure for a one-point calibration. Use and Maintenance
The value of buffer used should be that Continue through step 8 for a two-point Electrodes should be used in a
closest to value anticipated for sample. calibration.) vertical position.
Two-Point Calibration 6. Place electrode into the second buffer, Electrodes should be rinsed between
Meter calibration using two buffers, either pH 4.00 or pH 10.00. samples with distilled or deionized water.
one of which should always be 7.00. NEVER wipe an electrode to remove
The second buffer used would depend 7. Adjust the pH reading to display the excess water. Just blot the end of the
on the application. correct value using the SLOPE, electrode with a lint-free paper. Wiping
CALIBRATE, or GAIN controls the electrode can cause spurious readings
Method (coarse ajust). due to static charges.
Here is a general method for most pH
meters. Some pH meters require slightly 8. Adjust the pH reading to read the The level of filling solution in refillable
different techniques. Please read the correct value using the SLOPE knob electrodes should be kept at least 2/3 full.
instructions for their particular procedures. (fine ajust). The filling hole should be open during
use. pH electrodes are fragile. A proper
1. The temperature setting on the meter The pH meter is now calibrated and electrode holder should be used to provide
must correspond to the temperature ready to use. support and aid in raising and lowering the
of the buffers used, or an automatic probe into solutions.
temperature compensator must be Electrode Care
employed. Over 80% of pH measurement difficulties
are due to electrode problems. Proper
2. Turn pH meter to “pH” or “ATC” if storage, use, and maintenance increases
automatic temperature compensation accuracy.
is used.
Storage
3. Place clean electrode into fresh, Electrodes should be stored in an
room temperature pH 7.00 buffer. acidic solution with a low salt content.
Commercial soaking solutions are
4. Adjust the pH reading to exactly available or you can make your own
7.00 using the ZERO OFFSET, by mixing a 1M KCI solution adjusted
STANDARDIZED or SET knob. to pH 4.0.
Order Online
omega.com
Over 100,000 Products Available!
® To download information and to order pH Field and
Lab Instrument products online, visit omega.com
B-6
One Omega Drive | Stamford, CT 06907 | 1-888-TC-OMEGA (1-888-826-6342) | info@omega.com
www.omega.com
UNITED KINGDOM
www. omega.co.uk
Manchester, England
0800-488-488
GERMANY BENELUX
www.omega.de www.omega.nl
Deckenpfronn, Germany Amstelveen, NL
0800-8266342 0800-099-33-44
applications
Process Water
description
The Kuntze AH-300-K-1-1 is a combination electrode with a 1mm ceramic junction for pH measurement.
The sensor is liquid filled, thus blocking resistant and particulary suitable for lightly polluted media.
technical data
characteristic
measuring parameter pH-value
measuring range 0 .. 14 pH
ambient conditions
max. pressure 6 bar
adjustable by internal pressure
min. condutctivity > 50 µS
temperature -5 .. +100°C
Version 02/2009
AH-300-K-1-1
pH sensor
mechanical construction
junction Keramik
shaft material glass
electrode material AH glass ball
reference system Ag/AgCl/3M KCl
internal buffer pH 7
mechanical connection S7 plug
electrical connection 2-pole screw connection
Article Numbers
Article
Name Description
Number
pH sensor: KCl liquid filled, 1 mm ceramic
AH-300-K-1-1 24132040K
junction, S7 plug
accessories
accessories
pH-buffer solutions The slope of pH sensors change over the time (depending on the
measuring media), we recomment a regulary calibration with our
buffer solutions. We offer in a
1000 ml package pH 2, pH 3,56, pH 4, pH 7 and pH 9,22 and in
50 ml package pH 4 and pH 7.
Version 02/2009
Value-Added Solutions
www.abb.com/instrumentation
process strips away water and residual rangement allows insertion or removal of
Success Stories
organic material that is not part of the the sensor without disturbing the process.
monomer product. The plant recycles ABB supplies the sensor complete with
the excess organic materials. During a ball valve assembly that bolts to a
this phase, the reactor operates under standard flanged port on the reactor.
vacuum at elevated temperature, and The sensor has a patented reference
with a tight tolerance on alkalinity comprised of a matrix of immobilized
(9 – 10 pH). KCI salt fronted by a porous Teflon®
The pH value is used to control the rate of junction. This “solid state” reference is
caustic added to the reactor to maintain virtually immune to poisoning, plugging
the high alkalinity required for stripping. and pumping problems that plague
Salts formed during this phase account conventional liquid, slurry and gel-based
for the abrasive nature of the pH sensor’s references. Without this reference
environment. At the beginning of this technology the sensor would be quickly
phase, the sensor must respond virtually destroyed during the vacuum and
across its full range within 30 seconds – extreme high pH process phases.
from high-acidic to highly alkaline.
Signals from the pH sensor’s transmit-
PH Sensor Model TB
ter run to a distributed control system 557 Characteristics
(DCS). The DCS employs PID control The ABB sensor’s pH-sensitive hemi-
algorithms, developing signals for the spherical glass electrode is rugged
control valves that set the rate of raw enough to withstand an abrasive and
material and caustic additions. The end- strong caustic environment. It is rated
point of a monomer batch occurs when for a true 0 – 14 pH range at tempera-
the analysis of the stripped extraction tures to 284 degrees F. The body of
samples reaches a predetermined value. the sensor is Kynar® polyvinylidene
fluoride thermoplastic strengthened by a
The Search for the titanium sheath; both materials hold up
Right Sensor well to the harsh process conditions. An
Initially, Polychemie tried several types integral cable and temperature element
of pH sensors in its reactors, but none complete this rugged pH sensor.
could last long in the harsh environ- The Polychemie plant has been using
ment. Shortly after installation, sensor ABB TB557 pH sensors for three years.
response times slowed to a point that Polychemie calibrates the pH sensors
rendered the pH reading useless. Even and transmitters for its reactors on a
flowing reference sensors with both weekly schedule. To do so, technicians
conventional glass and unique glass/steel simply retract the sensor, close the ball
measuring electrodes quickly degraded. valve, and then remove the sensor for
Eventually, the plant found success field calibration. Technicians check both
with the TB557 pH sensor from ABB the sensor and transmitter as well as
Instrumentation. The TB 557 Model the sensor response time. Because pH
consists of a hot tap retractable sensor measurement is so critical to successful
that is conveniently inserted into and out batch production of this monomer
of the reactor through a ball valve. This ar- process, Polychemie replaces sensors
every six months as a matter of policy.
Copyright © 2005 ABB 3BUS340006R0001 Teflon® is a registered trademark of DuPont. Kynar® is a registered trademark of Arkema.
pH Sensor DT016A
Typical Experiments
• Acid – Base titration
• Investigation of water quality
• Acidification of milk and others foods
• pH measurements in tissue extracts
• Alcoholic fermentation in yeast
How it Works
The pH electrode contains two half-cells. One contains a reference element of known
H+-concentration. The other, at the bottom of the electrode, is an H+- sensitive glass
membrane. The adaptor measures and amplifies the potential difference between the
1
two half-cells. The pH level (pH= – log (H+)) is calculated from the potential
difference.
Sensor Specification
Range: 0 - 14 pH
Accuracy:
± 2% over entire range, after
temperature compensation
Operating Temperature: 0 °C – 50 °C
Contents
• Adaptor & electrode set DT016A
• Adaptor only DT017
• Electrode only with a bottle containing pH 4.01 buffer DT018
Adaptor & electrode set DT-016A Adaptor only DT-017 Electrode only DT-018
2
Equipment Setup
1. Connect the electrode to the adaptor.
2. Connect the adaptor to the data logger's input.
Required Equipment
• Nova5000, MultiLogPRO or TriLink data loggers
• Wash bottle with distilled or deionized water
• Several clean beakers
• Lab wipes
Technical Notes
• If other electrochemical type sensors (Oxygen and Conductivity) are placed in
the same solution at the same time and connected to the same data logger,
they can interfere with each other’s signals. Keep the sensors as far apart as
possible - the distance required will depend on the conductivity of the
solution. If there is still a problem, try connecting the sensors to different data
loggers or take readings using one sensor at a time.
• In order to use temperature compensation a Temperature sensor must be
connected to the data logger along with the pH sensor (the Temperature
sensor must be plugged into Input 1).
Calibration
The pH sensor is shipped fully calibrated.
For experiments that require very accurate calibration, however, the pH sensor is
equipped with an offset calibration screw. The screw is located at the back of the
sensor case. Place the electrode in a reference solution (buffer of pH 7) and start
recording. Insert a flat screwdriver to the calibration hole and slowly turn the
calibration screw until the reference value is reached.
3
Using the pH Sensor with Fourier Data Loggers and MultiLab Software
1. Rinse the electrode with distilled water and blot with a lab wipe.
3. Connect the Temperature sensor (DT029) to the data logger’s first sensor input
I/O-1.
Maintenance
• At the end of the measurement, remove the electrode from the sample; rinse the
electrode with distilled water over the waste beaker. Blot the electrode dry with a
lab wipe. The electrode is now ready to read the pH of other samples.
• When not in use, store the pH electrode in the supplied bottle containing the
storage solution.
• The recommended storage solution is comprised of 50% ph 4 buffer and 50% 4M
KCl salt. If this storage solution isn't available then a fresh ph 4 buffer can be
used as well.
4
2. If the electrode has been exposed to protein or similar materials, soak in acidic
pepsin, ASI part number CS 0003, for 5 minutes. Rinse thoroughly with distilled
water. Soak in storage solution for 30 minutes prior to recalibration.
3. If the previous cleaning procedures fail to restore response, soak the electrode in
0.1 N HCI for 30 minutes. Rinse thoroughly with distilled water. Recalibrate
before use.
4. If electrode response is still not restored, replace the electrode.
Troubleshooting
If the pH displays values which are out of the sensor range, verify that the sensor
cables are properly connected.
5
Technical Support
Please contact Fourier technical support as follows:
Web: http://www.fourier-sys.com/support_support.html
Email: support@fourier-sys.com