Sei sulla pagina 1di 36

I.

Introduction to Biomedical Engineering


• Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 1
Electrical • Industry Regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 3

Safety • BMET Responsibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4

• History of Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . .page 4

Made Easy
II. Fundamental Concepts of Electrical Safety . . . .page 6

III. Physiological Effects of Electricity


Table Of • Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 7

Contents • Macroshock and Microshock . . . . . . . . . . . . . . . . . . . . . . . . .page 7

• High Frequency Effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8

• Potential Equipment Safety Hazards . . . . . . . . . . . . . . . . . . .page 8

• Equipment Safety Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8

• Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 8

• The Electrical Power System . . . . . . . . . . . . . . . . . . . . . . . . . .page 8

• Contact with a Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9

• Skin Resistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 9

• Leakage Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10

• Electrical safety Power System Devices . . . . . . . . . . . . . . . .page 10

• Ground Fault Circuit Interrupter . . . . . . . . . . . . . . . . . . . . .page 10

• Isolation Transformer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 10

• Equipotential Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11

• Codes and Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 11-12

IV. The AAMI Standard . . . . . . . . . . . . . . . . . . . . . . . . . .page 13-32

Bibliography . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 32

Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .page 33

www.bapcoinfo.com
T he field of medical
instrumentation is by no
means new. Many
consequently, much research and
development money went into this
area. The aerospace medicine
instruments were developed programs were expanded
as early as the nineteenth century –
for example, the electro-
cardiograph, first used by
considerably, both within NASA
facilities and through grants to
Universities and Hospital research
Electrical
Einthoven at the end of that
century. Progress was rather slow,
however, until after World War II
units. Some of the concepts ad
features of patient monitoring
systems presently in use in
Safety
when a surplus of electronic hospitals throughout the world
equipment such as amplifiers and evolved from the base of astronaut Made Easy
recorders became available. At that monitoring. The use of adjunct
time, many technicians and fields, such as biotelemetry, also
engineers, both within industry found some basis in the NASA
and on their own, started to programs.
experiment with and modify
existing equipment for medical Also, in the 1960’s, an awareness of
use. This process occurred during the need for engineers and
the 1950’s and the results were
often disappointing. For the
experimenters soon learned that
technicians to work with the
medical profession developed. All
the major engineering technical
I.
physiological parameters are not societies recognized this need by
measured in quite the same way as forming "Engineering Medicine
physical parameters. They also and Biology" subgroups and new
encountered a severe societies were organized, such as
communication problem with the the Biomedical Engineering
medical profession. Society. Along with the medical
research programs at the
During the next decade many
instrument manufacturers entered
universities, a need developed for
courses and curricula in
Introduction
the field of medical instrumenta-
tion, but development costs were
biomedical engineering and today
almost every major university has to Biomedical
high, and the medical profession some type of biomedical
and hospital staffs were suspicious
of new equipment and often
engineering program. However,
much of this effort is not
Engineering
uncooperative. Many develop- concerned with biomedical
ments with excellent potential instrumentation per se.
seemed to have become lost
causes. It was during this period One of the problems of
that some progressive companies "biomedical engineering" is
decided that rather than modify defining it. The prefix bio- of
existing hardware, they would course, denotes something
design medical instrumentation connected with life. Biophysics and
specifically designed for medical biochemistry are relatively old
use. Although it is true that many disciplines in which basic sciences
of the same components were have been applied to living things.
used, the philosophy was changed; One school of thought subdivides
equipment analysis and design bioengineering into different
were applied directly to medical engineering areas – for example,
problems. biomechanics, and bioelectronics.
These categories usually indicate
A large measure of help was the use of that area of engineering
provided by the U.S. government, applied to living rather than to
in particular NASA (National physical components.
Aeronautics and Space Bioinstrumentation implies
Administration). The Mercury, measurement of biological
Gemini and Apollo programs variables, and this field of
needed accurate physiological measurement is often referred to as
monitoring for the astronauts; biometrics, although the latter -1-
term is also used for mathematical responsibly, effectively, and safely
and statistical methods applied to manage and interface with medical
biology. devices, instruments and systems
and the use thereof during patient

Electrical Naturally, committees have been


formed to define these terms; the
professional societies have become
care, and who can, because of this
level of competence, responsibly and
directly serve the patient and

Safety involved. The latter includes the


IEEE Engineering in medicine and
biology group, the ASME
physician, nurse and other health
care professionals relative to their
use of and other contact with
Biomechanical and Human Factors medical instrumentation."
Made Easy division, the Instrument society of
America and the American Institute Most engineers go into the
of aeronautics and Astronautics. profession through the engineering
Many new cross-disciplinary degree route, but many start out as
societies have also been formed. physicists or physiologists. They
must have at least a B.S. degree
Several years ago an engineering and many of them have M.S. or
committee was formed to define Ph.D. degrees.

I. bioengineering. This was sub-


committee B (Instrumentation) of
the Engineers Joint Council
Another popular term, also coined
in recent years, the biomedical
Committee on Engineering on equipment maintenance
engineering Interactions with technician (BMET) is defined
biology and Medicine. Their as follows:
recommendation was that
bioengineering be defined as "A biomedical equipment
application of the knowledge maintenance technician (BMET) is
gained by a cross fertilization of an individual who is knowledgeable
engineering and the biological about the theory of operation, the
sciences so that both will be more underlying physiologic principles
fully utilized for the benefit of man. and the practical, safe clinical
application of biomedical
More recently, as new applications equipment. His capabilities may
have emerged, the field has include installation, calibration,
produced definitions describing inspection, preventive maintenance
the personnel who work in it. A and repair of general biomedical
tendency has risen to define the and related technical equipment as
biomedical engineer as a person well as operation or supervision of
working in research or develop- equipment control, safety and
ment in the interface area of maintenance programs and
medicine and engineering, whereas systems."
the practitioner working with the
physicians and patients is called a This was also an AAMI definition.
clinical engineer. Typically, the BMET has two years
of training at community college.
One of the societies that has This person is not to be confused
emerged in this interface area is the with a medical technologist. The
Association for the Advancement of latter is usually used in an
medical Instrumentation (AAMI). operative sense, for example in
This association consists of both blood chemistry and in the taking
engineers and physicians. In late of electrocardiograms. The level of
1974, they developed a definition sophistication of the BMET is
that is widely accepted: usually higher than that of the
technologist in terms of
"A clinical engineer is a equipment, but possibly lower in
professional who brings to health terms of the life sciences.
care facilities a level of education,
experience and accomplishment In addition, other titles have been
-2- which will enable him to used, such as hospital engineer
and medical engineer. In one hospital equipment, as well as the
hospital the title biophysicist is maintenance and repair of such.
preferred for their biomedical Secondly, other agencies, under
engineers, for reasons best known which hospital equipment was
to themselves.

These definitions are all


monitored and controlled, invoked
requirements such as the release of
service literature on any new
Electrical
noteworthy, but whatever the
name, this age of the marriage of
engineering to medicine and
equipment purchased or manu-
factured to the hospital. This
literature, the law stated, must
Safety
biology is destined to benefit all contain the procedures for
concerned. Improved com- calibration and alignment of said Made Easy
munication among engineers, equipment. Thirdly, the training of
technicians and doctors, better and biomedical equipment technicians
more accurate instrumentation to began to catch up with the
measure vital physiological advancing technology so that the
parameters and the development level of competence was signifi-
of interdisciplinary tools to help cantly improved. Finally, through
fight the effects of body malfunc- the feeling that they provided the
tions and diseases are all a part of
this field. The name itself is actually
not all too important; however,
only service available, the OEMs
began to charge significantly more
for their service contracts each
I.
what the field can accomplish is year. Surprisingly, this was not a
important. With this point in mind, major issue to the hospital
we will be using the term community until the middle of
biomedical engineering for 1980 when the government, in an
describing the field in general and effort to control the rising cost of
biomedical instrumentation for medical treatment, and the sub-
the methods of measurement sequent rise in medical insurance
within the field. claims, imposed major restrictions
upon hospitals. These restrictions,
INDUSTRY REGULATION known as Diagnostic related
Groups (DRGs) limited the amount
In it’s infancy, biomedical equip- that medical insurance groups
ment was best serviced and would reimburse both hospitals
maintained by the original and physicians for each type of
equipment manufacturer as there illness for which a patient might
were few, if any, qualified enter the hospital. Without going
technicians who could do so into great detail about DRGs,
adequately. A lack of standards suffice it to say that their intro-
linking the technology, an absence duction brought with it a need for
of service literature, and the hospitals to lower their operating
general feeling from hospital costs. One area in which these
personnel that, unless trained by costs could be lowered was in the
these OEMs, no one would be able area of biomedical instrumetation.
to provide a suitable level of Where once a hospital would make
services. Thus the early BMET was a major equipment expenditure,
relegated to maintain and repair such as buying a new chemistry
only the simplest of clinical analyzer, every five or so years, they
equipment such as centrifuges, began keeping sophisticated
suction pumps and other such equipment such as this for longer
equipment. Several factors soon periods. So doing, the rising OEM
arose to alter this conception. First, service contract prices suddenly
agencies such as the Food and became an issue and the respon-
Drug Administration (FDA) sibility for these equipment fell
developed subdivisions under upon the BMET as a lower cost
which fell the area of medical alternative.
instrumentation. This brought
about a set of standards and As more and more responsibility
practices for manufacturing of fell upon the BMET within the -3-
hospital for larger and more be employed on sensitive, life-
sophisticated instrumentation, new saving medical equipment.
problems arose for the hospitals. In
order for the equipment to be Preventative
Electrical adequately maintained, more
BMETs needed to be employed,
each having sufficient training in
Maintenance Inspections
Periodic testing and evaluation of
patient and non-patient related

Safety the sophisticated equipment


involved. Hospitals were faced with
weighing the additional cost of
equipment to insure that they
operate within the guidelines set
fourth by federal, state and
salaries, benefits and the costs of
Made Easy attending additional manufacturers
manufacturer guidelines. This
includes calibration and alignment
schools and seminars with the cost of said equipment along with the
of the OEM contracts. In very large repair and/or replacement of
hospital facilities, this was not such component parts in order to bring
a significant problem as it was with the equipment into compliance. It
smaller facilities. In many smaller is the belief of most BMETs that
facilities, typically those under good, quality preventive mainten-
200 patient beds, it was simply not

I.
ance inspections further reduce
feasible to consider an in-house hospital costs by reducing equip-
BMET program and, thus, these ment down-time for repairs.
facilities were forced to pay for the
OEM contracts.
Inservice & Education
Again, the government became In an effort to satisfy applicable
involved. Federal and state federal and state requirements as
agencies began to require all well as to insure that clinical
hospital facilities to comply with equipment are utilized both safely
stringent safety guidelines for and to their optimum, some BMET
patient related equipment. This shops provide periodic in-service
included a routine electrical safety and education services. This often
program under which all hospital ranges from simple electrical safety
equipment would be tested, lectures to the proper use of a
evaluated and forced to comply newly acquired piece of equipment
with. Typically, the BMET provides for which hospital personnel might
one or more of the following not yet be familiar.
services to hospital facilities:
Electrical Safety Inspections
BMET RESPONSIBILITIES This brings us to the main purpose
of this document. Periodic testing
Corrective Maintenance of patient and non-patient hospital
Non-scheduled repair of hospital equipment to insure that they meet
equipment other than during the safety requirements and guide-
Preventive Maintenance inspec- lines as set fourth by applicable
tions. These repairs range from federal and state regulations. This
replacement of minor parts or in no way measures the equipment
components to total equipment ability to perform the task for
overhauls as may be required. In an which is was designed.
effort to further provide a cost
savings to the hospital facility, HISTORY OF
some BMETs attempt to repair ELECTRICAL SAFETY
equipment to the smallest
component level, rather than The introduction of electricity into
simply replacing circuit boards or commerce at the close of the 19th
sub-assemblies which may be very century carried with it the need to
costly. Other BMETs follow a know how to deal with it safely. For
philosophy that component-level the hospital environment, much of
repair entails far too many risks of the electrical safety programs
repeated failure or improper started after World war II. Shortly
-4- component soldering techniques to thereafter, federal funds via the
Hill-Burton Act increased the hospitals" consumer critic Ralph
number of hospital facilities. Nader said yesterday.
Grounding for electrical safety was
implemented because most During the 1970’s, several
electrical accidents in the home
and industry occurred because
exposed metal was energized. The
proposals and regulations were
introduced to manage this
suspected problem in hospitals. In
Electrical
National Electrical Code stated that
"non-current carrying" metal parts
of the electrical apparatus could
1971, the National Fire Protection
Association published a recom-
mended standard (76BM) to help
Safety
prevent such accidents. hospital engineers understand the
Unfortunately, by the end of the principles of electrical safety and Made Easy
1950’s, electrical power in the coordinate a program of medical
hospitals was supplied in a equipment electrical testing in
haphazard, eclectic fashion. their facilities. In the spring of
1972, Underwriters laboratories
In the hospital environment of the issued document UL544 Medical
sixties, electricity came to be used and Dental Equipment, which was
more often on, in and around intended to serve as a guideline for
patients to a degree beyond
conception only a few years
previously. It was early in 1961 that
medical equipment manufacturers.
In 1970, AAMI (the Association for
the Advancement of medical
I.
there appeared the first news that Instrumentation) published a first-
"microshock" (small electric draft standard for electrical leakage
currents applied to a conductor current standards that was adopted
near the heart) was happening in as an American national standard
the medical field. in 1978.

In 1969, Carl Walter, M.D., who was Probably the most dramatic
at the time a well-known surgeon, proposal was the Joint
stated that, "1,200 patients were Commission on the Accreditation
being accidentally electrocuted in of Hospitals (now the Joint
U.S. hospitals each year." Although Commission on the Accreditation
many engineers and health care of Healthcare Organizations –
professionals believed that Dr. JCAHO) 1976 recommendation that
Walter’s estimates were unreal- hospitals maintain equipment
istically high, the concept of control programs to provide for
microshock suddenly became electrical safety training, create a
publicized. documented preventive main-
tenance program, and perform
Then, on June 16th, 1970, Ralph semi-annual safety and perfor-
Nader (a household word because mance equipment inspections and
of his 1965 book, Unsafe at Any annual inspections of electrical
Speed) gave a speech in Detroit. In receptacles.
this speech, he stated that, "1200
annual electrocutions in US Today, health care institutions in
hospitals was a very least figure" the United States support clinical
and quoted other experts engineering programs that provide
indicating that the real number ongoing electrical safety and
might be significantly higher. The performance testing as well as
event was picked up by a wire preventive maintenance and repair
service and run under an arresting of medical equipment. Typically,
headline the next day: Hospital these programs use the most recent
electrocutions cited Detroit (UPI) – editions of NFPA 70, NFPA 99,
"Accidental electrocutions claim NFPA 101, AAMI Recommended
5,000 lives in American Hospitals Standards, and Joint Commission
every year but seldom get reported accreditation manuals for their
due to the "close nature of reference standards.

-5-
S ome of the most basic
concepts of electricity must
be understood in order to
Where "I" is the current in amperes
that flows through a device or
grasp the potential hazards of conductor, "V" is the magnitude in

Electrical electrical devices as they are used


in a clinical setting such as a
hospital. Some of these important
volts of the applied electric field,
and "R" is the resistance that the
device or conductor has, measured

Safety ideas concerning electrical safety in


a medical environment are
explained below.
in Ohms.

Thus, if an electromedical device


such as an aspirator is plugged into
Made Easy a 120 volt receptacle and the unit
The Nature of Electricity requires 3.0 amperes, then the
Electrical current flowing through a resistance that the aspirator
conductor is the result of electrons provides to the voltage is:
moving from the outer shell of
atoms induced by an electric field
that is imposed on a conductor.
120 volts = 40 Ohms
This field can be caused by any 3 amperes
II. voltage generating source, such as
a local utility company, a battery, or
a chemical reaction. In the
a measure of the amount of power
that an electrical device consumes
hospital, this voltage source is during operation is defined as the
provided by the local electrical product of voltage and current and
company and is redirected through is recorded in Watts.
a series of transformers to 240, 208,
or 120 volts alternating current From the previous example:
(AC). The amount of current that
flows through an electrical device is Power (watts) =
Fundamental determined by the resistance that
the device is designed to provide to (V) (I) = (120)
Concepts Of the applied field. This relationship
is called "Ohm’s Law" and is
(3) = 360 Watts
written as:
Electrical
I=V
Safety R

BAPCO’S SA2115
“Safety Certifier”
is the most complete
safety analyzer
available that provides
everything you need to
safety test:
• Patient to non-
patient, medical to
commercial devices

• 110 V @ 200amp
to 240V @ 15 amp

• Automatically
to AAMI standards

-6-
B ecause the amount of
electricity that will flow
through any medium
MACROSHOCK AND
MICROSHOCK
depends on the resistance that
it encounters, current can flow
through the human body and
cause various effects. If direct
The effect of electric shock on the
human body can be anything from
barely perceptible tinges, to muscle
Electrical
current (polarized, nonchanging)
or high frequency alternating
current passes through the body,
spasms, to death. Each can occur
from small or large currents,
depending on how the currents are
Safety
heating effects and ultimately introduced into the body. Large
burns will occur. It is this effect that currents (milliamperes or larger) Made Easy
is intentionally created when that are introduced into the body
electrosurgical generators are used from one external point to another
to cut tissue and coagulate fluids. If (arm to leg, for example) can result
low frequency alternating current is in macroshock. If small currents (as
applied to the body, muscular low as 10 microamperes) are
polarization and depolarization introduced into the body from an
take place that can ultimately external source such as a catheter
create a "circus movement" in the
heart muscle, resulting in
fibrillation and death. It is this
or cardiac pacing wires, the
resistance to the heart muscle can
be very low, and electrocution can
III.
effect that normally accounts for occur from microshock.
death due to electric shock.
The chart below describes the
Unfortunately, the typical effect of different levels of current
resistance of the human body in that are introduced into the human
combination with the frequency of body by creating a voltage across
commercially generated electricity each arm. For microshock, the
(60 hertz) can create a potentially
hazardous situation in the hospital
same effect can be produced with
current levels that are only
Physiological
1/10,000 as great as those listed.
environment.
Effects Of
Electricity
CURRENT EFFECT

.001 Ampere (1 Milliampere) "Tingling", threshold of perception

.020 Amps (20Ma) Muscle Spasms, hard to release grip

.050 amps (50 Ma) Pain, possible fainting, transient


interruption of respiration

.100 Amps (100 Ma) Ventricular fibrillation

>5 Amps Sustained myocardial contraction,


possible burns, temporary
respiratory paralysis

-7-
HIGH FREQUENCY EQUIPMENT
EFFECTS SAFETY TESTING

Electrical As mentioned previously, the


primary effect of high frequency
current (500 kilohertz to 2 mega-
Tests to determine the electrical
safety of medical devices include a
measurement of the continuity as

Safety hertz) is to heat tissue as it is


concentrated in a certain area. The
amount of heat that is generated
well as the leakage current between
the chassis and the cord grounding
conductor. Equipment that has
depends on the amount of current been designed with patient leads or
Made Easy applied and the area that the contact points is also measured for
current passes through. current leakage from these points.
Figures 1 and 2 show typical
This relationship is : circuits that may be used to
Heat = I x I measure this leakage current.

A
Where "I" is the applied current ELECTRICAL SHOCK
III. and "A" is the area that the current
flows through. The three ingredients in the
scenario of electrocution are :

POTENTIAL 1. contact with the live


conductor of a grounded
EQUIPMENT electrical system
SAFETY HAZARDS 2. contact with a ground
3. diminished skin resistance
Because current flow through the
body can be hazardous if it is of a
certain amplitude and frequency, THE ELECTRICAL
stray currents must be eliminated POWER SYSTEM
from medical equipment. The best
method to prevent leakage currents Domestic voltages in the United
is to ensure that all conductive States are from 120V for lighting
parts of the equipment are con- and small appliances, and 240V for
nected to the hospital grounding electrical ranges and dryers. In the
system through the power plug. hospital, outlets are 120V while
Also, to minimize patient contact 277V and 208V are often used for
with current leakage on ECG signal fixed lighting and special
leads, electrically isolated receptacles.
amplifiers should be used.
Typically, a voltage of 240V is
In theory, the use of line isolated provided by the service drop to the
power systems helps to minimize power meter of the hospital. This
safety hazards by isolating the line consists of a bare cable plus
neutral power line from earth two wires having black insulation.
ground. A line isolation monitor is The bare cable is connected to
also installed with these systems to "earth" ground via a water pipe or
identify visually and aurally the grounding rod. It is this connection
presence of leakage current that makes the electrical service a
between the isolated power line grounded system.
and the grounded conductor. Other
power system devices that help to NOTE : The national electric Code defines
identify or disconnect power ground as "a conducting connection,
sources when leakage currents are whether intentional or accidental, between
present include ground fault an electrical circuit or equipment and the
detectors (GFDs) and ground fault earth, or to some conducting body which
interruptors (GFIs). serves in place of the earth."
-8-
There are three reasons to ground devices have an isolated power
electrical systems: supply and grounding system. Plus,
1. Grounding of the system the manufacturers use plastic cases
minimizing the exposed metal.
protects against introduction
into the structure, via the
wiring, of high voltages (with SKIN RESISTANCE
Electrical
respect to ground) such as
might arise from lightning or
insulation failure in a high
voltage pole transformer.
Water affects skin resistance, and
for a given voltage, resistance
Safety
determines current. When dry, skin
2. Tying the circuit to ground has a resistance of upwards Made Easy
obviates the multiple problems 100,000 ohms. If there is an
that might ensue were the accidental application of 120 volts
domestic circuits "floating" at between the two hands, only
some undefined but high 1.2 milliamperes will flow. In a wet
voltage relative to ground. environment or on a hot and
3. To facilitate operation of over humid day, that same current path
current protective devices may come to have a resistance as
(fuses) which are located in the
"live" (ungrounded) side of
each circuit.
low as 1,000 ohms, resulting in a
current flow of 120 milliamperes. III.
If a patient is exposed to
electrical current, it takes about
CONTACT WITH 1 milliampere of 60 Hz AC for a
A GROUND threshold of sensation. The
sensation becomes uncomfortable
For purpose of illustration, we will if 5 milliamperes is attained. Strong
use an electrocardiograph (ECG) as muscle spasms appear at currents
our example of a medical device. of 10 to 20 milliamperes ("let-go"
Back in the 1950s, the standard current before sustained muscular
ECG was fitted with a two-prong contraction). A current flow of
plug. The patient was connected to 100 milliamperes or greater may
the chassis ground via the right-leg induce ventricular fibrillation and
electrode, and one wire of the death. These values are for currents
power cord was connected to the introduced at the body surface.
chassis through a 200,000-Ohm They are termed macroshocks and
resistor. This connected the chassis require two points of external body
to ground via the natural con- contact.
ductor in the power system. If the
plug was inserted with polarity There is a possibility of a direct
reversed, then the patient was electrical path to the heart via a
connected to the "hot" side of the needle or catheter in an artery or
power system. However, the 200K- vain. This directly reduces the
Ohm built-in resistor limited the resistance and current threshold.
current to 0.6 milliamperes if the Small amounts of current
patient or bystander provided a (100 microamperes) can be
pathway to ground. Added safety to potentially lethal. Electrical shock
the machine included a five- in these circumstances is termed
milliamp fuse in series with the "microshock".
right-leg electrode. As it turns out
this approach was not practical The frequency of the current is also
because the fuse kept blowing after important when considering
the patient was defibrillated. electrical shock. If the frequency is
raised above 1 KHz, these current
The codes changed this practice levels no longer produce such
and today’s ECG machines do not sensations or life-threatening
ground any of the electrodes to phenomenon. High frequencies in
earth ground. In fact, the "front the megahertz region will not cause
end" electronics on most medical shock at all. -9-
LEAKAGE CURRENT currents are equal. If the current is
unbalanced (i.e. current flowing
All electrically operated devices through ground) a net flux is
induced across the third coil. This

Electrical have some current that flows from


the energized electrical portions of
the device to the metal chassis.
current will trigger a switch
opening the "hot" line.
This current is referred to as leak-

Safety age current and has two com-


ponents, capacitive and resistive.
Capacitive leakage current results
The GFCI affords economical
protection against electrical shock.
However, the GFCI depends upon
an active system, and the integrity
Made Easy from distributive capacitance
between two wires or a wire and a of it’s mechanical operation is
metal chassis case. Components crucial.
that cause capacitive leakage
currents are RF filters, power
transformers, power wires, and any ISOLATION
device that has stray capacitance. TRANSFORMER
Resistive leakage current arises

III.
from the resistance of the The Isolation transformer offers
insulation surrounding the power electrical safety by converting
wires and transformer primary grounded power into ungrounded
windings. power. This is accomplished by
grounding the primary winding of
the transformer and not grounding
ELECTRICAL SAFETY the secondary winding. Isolation is
POWER SYSTEM not perfect for two reasons. First,
DEVICES the isolation transformer has some
stray capacitance to ground.
Several techniques are available to Second, every medical device that
protect clinicians and patients is attached to the transformer
from electrical shock. The most possesses stray capacitance which
common ones are the ground fault causes some degree of coupling
circuit interrupter (GFCI), the between its power-carrying wires
isolation transformer and and the grounded frame. To
equipotential grounding. monitor the system isolation, a line
isolation monitor (LIM) is em-
ployed. Its function is to analyze
the entire isolated circuit and
GROUND FAULT quantify its degree of isolation from
CIRCUIT INTERRUPTER ground. The LIM provides visual
and audible alarm signals when the
The GFCI acts like a circuit breaker predicted ground-seeking current
when it senses an inequality of as exceeds a specified magnitude. The
little as 6-ma between the "hot" LIM does not indicate an existing
and neutral wires of the circuit. It is current flow, rather it predicts the
mandated by the national electric current that would flow if a short
code where electrical outlets are so circuit were to develop between
situated that simultaneous contact isolated wire and ground.
with a grounded surface is
especially likely. "Wet locations" In the operating room, isolated
such as a whirlpool bath or power systems were first installed
bathroom are examples of GFCI as a measure directed against
usage. sources of ignition rather than
electrical shock. Flammable
In the GFCI, current in the "hot" anesthetics such as diethyl ether
wire passes through one and cyclo-propane were used.
transformer coil on the same Today, most anesthesiologists use
transformer core. From the design, non-flammable anesthetics and
-10- the net flux is zero, when the isolated power is not required.
EQUIPOTENTIAL and second, safety was blatantly
exploited for ego-serving,
GROUNDING bureaucratic, and commercial
gains.
Another technique that reduces
electrical shock is equipotential
grounding. This is accomplished by
Today, there are two standards that
specify electrical safety: NFPA 99
Electrical
adding another grounding wire
from each chassis to a central point
that is in parallel with the third wire
and the AAMI Standard for Safe
Current Limits. Safety
in the power cord. If the chassis of The 1996 edition of NFPA 99
all equipment is at the same includes numerous technical Made Easy
potential there will be no current changes that relate to leakage
leakage to the heart. This technique current and it now more closely
has its advantages and disadvan- correlates to the international
tages and is not typically used in standards. Originally, NFPA 76B
today’s health care environment. covered the electrical wiring system
and electrical appliances. The
summary of the standard as it
CODES AND
STANDARDS
relates to electrical safety is in the
Table on the following page. III.
One further comment about
In the 1960s, all aspects of hospital standards. On may 30, 1972,
activities involving fire and Underwriter’s laboratories (UL)
explosion hazards (including released the first edition of UL-544,
electrical shock and emergency Standard for Safety for Medical and
electrical power) were seen to be in Dental Equipment. UL-544 deals
need of some guidelines and/or with details regarding enclosure
standards. As it turns out, in the safety, mechanical stability, and
standards arena of the 1970s, integrity of insulation. It’s leakage
patient safety was twice corrupted. current, isolation requirements,
First, standards were generated in and test methods are similar to but
ignorance (solutions were imposed not identical with those of AAMI
before the problems were defined) and NFPA.

-11-
NFPA - 99 1996 AAMI/AAMI ESI
1993

Electrical Chassis Source Current,


Cord Connected (Portable)
Ground
Open
Ground
Intact
Ground Ground
Open Intact
With Isolated Patient Connection 300µa 300µa 300µa 100µa

Safety With Nonisolated patient Connection


Likely to Contact Patient
300µa
300µa
300µa
300µa
300µa
300µa
100µa
100µa

Made Easy No patient Contact 300µa 300µa 300µa 100µa

Chassis Source Current, Ground Ground Ground Ground


Permanently Connected Open Intact Open Intact
With Isolated Patient Connection 5000µa 5000µa 5000µa 100µa
With Nonisolated patient Connection 5000µa 5000µa 5000µa 100µa
Likely to Contact Patient 5000µa 5000µa 5000µa 100µa

III. No patient Contact 5000µa

Ground
5000µa

Ground
5000µa 100µa

Ground Ground
Lead to Ground Current Open Intact Open Intact
With Isolated Patient Connection 50µa 10µa 50µa 10µa
With Nonisolated patient Connection 100µa 100µa 100µa 50µa

Ground Ground Ground Ground


Lead to Lead Current Open Intact Open Intact
With Isolated Patient Connection 50µa 10µa n/a n/a
With Nonisolated patient Connection 50µa 50µa n/a n/a

Ground Ground Ground Ground


Sink Current (Isolated Test) Open Intact Open Intact
With Isolated Patient Connection n/a 20µa n/a n/a
With Nonisolated patient Connection 50µa 50µa n/a n/a

Ground Ground Ground Ground


Open Intact Open Intact
Ground Impedance 0.5 Ohms
Existing System 0.2 Ohms
New Construction 0.1 Ohms

Ground Ground Ground Ground


Receptacle Indicators Open Intact Open Intact
Hospital Grade Listed
Emergency System Red
Isolated Ground Orange

Retention Force, Ground Ground Ground Ground


Ground Blade Open Intact Open Intact
4 Oz

Ground Ground Ground Ground


Open Intact Open Intact
-12- GFCI Trip Current 6mA
T he AAMI Electrical Safety
Committee began
deliberating the issue of
fault conditions and provides risk
current levels for each of these
states. Also included in this
safe risk current limits for standard are test methods for
electromedical apparatus in 1967
and first published recommended
limits and test methods in 1971.
equipment with nonconductive
enclosures and for double-
insulated equipment as well as
Electrical
The first edition of this American
National Standard was approved by
the American National Standards
limits for earth leakage or earth
risk currents. Safety
Institute (ANSI) in 1978, and a This standard is intended primarily
second edition was approved and for the testing of electromedical Made Easy
published in 1985. Work on this apparatus intended for use in or
third edition involved the near the patient care area.
dedicated efforts of concerned
health care professionals, industrial This standard should be considered
scientists, and government flexible and dynamic. As tech-
representatives. AAMI expresses its nology advances and new data are
gratitude for the service of all brought forward, the standard will
persons involved in the develop-
ment of this standard.
be reviewed and, if necessary,
revised. IV.
In this third edition of the standard, Suggestions for improving this
the risk current limits have been standard are invited. Comments
raised to be compatible with, and suggested revisions should be
although not identical to, the limits sent to Technical Programs, AAMI,
set forth by the International 3330 Washington Boulevard, Suite
Electrotechnical Commission 400, Arlington, VA 22201-4598.
(IEC) in its standard, Medical
electrical equipment—Part 1:
General requirements for safety
NOTE — This foreword does not
contain provisions of the American
The AAMI
(IEC 601-1-1988). As in the IEC
standard, this edition of Safe
National Standard, Safe current
limits for electromedical apparatus Standard For
current limits for electromedical (ANSI/AAMI ES1—1993), but does
apparatus (ANSI/AAMI ES1—1993)
introduces the concepts of normal
provide important information
about its development and
Safe Current
operating conditions and single intended use.
Limits For
Electromedical
Apparatus

-13-
SAFE CURRENT LIMITS In addition, this standard does not
apply to apparatus designed
FOR ELECTROMEDICAL primarily for nonmedical
APPARATUS applications and used in

Electrical 1 SCOPE
conjunction with electromedical
apparatus, but located outside of
the patient care vicinity.

Safety 1.1 Inclusions


This standard sets risk current
limits and referee test methods for
NOTE — As indicated above, all
devices cannot be readily covered
by this standard. Equipment such
Made Easy electromedical apparatus intended
for use in the patient care vicinity as video cassette recorders or
and also sets limits for non- computing devices are now being
patient-contact electromedical used in health care facilities. Such
apparatus. equipment is designed to other
standards. If such equipment is
The standard applies to line- and not located in the patient care
battery-powered apparatus and to vicinity, such devices are not
considered to pose a risk as they

IV. apparatus used singly or with


properly connected accessory
equipment. When more than one
are not likely to contact the patient.

2 DEFINITIONS
electromedical apparatus is
powered by a single power cord,
the equipment assembly acts like a For the purposes of this standard,
single apparatus in terms of risk the following definitions apply.
current limit requirements, and
shall be considered as such for the 2.1 accessory
purposes of this standard. The Device produced or recommended
safety and performance criteria by the manufacturer of an electro-
defined in this standard are medical apparatus, and intended to
intended for use in design quali- be electrically connected to that
fication by the device apparatus in order to make the
manufacturer. apparatus useful or to improve its
efficacy or versatility, and not a
NOTE — The referee test methods modular part of that apparatus.
of Section 5 are intended to provide
means by which conformance with 2.2 auxiliary apparatus
the standard can be established. Electromedical apparatus used in
These tests are not intended for use conjunction with other
in verifying the performance of electromedical apparatus to
individual devices in routine achieve a common purpose.
quality assurance inspections. Also,
referee tests allow for the use of NOTE — Auxiliary apparatus
alternative methods for design includes both interconnected
qualification, provided that devices apparatus and noninterconnected
so qualified will also meet the apparatus.
requirements of this standard when
tested in accordance with the 2.3 composite risk current
referee methods. Total risk current derived from the
risk currents of all the apparatus
1.2 Exclusions associated with the patient that can
This standard does not set limits flow through the patient, medical
for the composite risk current staff, or bystander.
when several devices are per-
forming different functions for the NOTE — This definition is included
same patient and are independ- for reference only. A method of
ently connected to the utility power derivation and limits for composite
system. This standard does not risk current are not covered in this
apply to therapeutic currents. standard.
-14-
2.4 electromedical apparatus removed or replaced without
Instrument, equipment, system, or affecting the operation of other
device that directly or indirectly assemblies in the apparatus.
uses electricity for any medical
purpose.

NOTE — Also included are all parts


2.11 nonoperational
environmental conditions
Temperature, humidity, altitude, or
Electrical
that are connected to such
equipment and are required for the
normal use of the equipment,
acceleration limits specified by the
manufacturer for storage or
shipment.
Safety
including associated patient wiring
or cables. 2.12 normal condition (NC) Made Easy
Condition in which all means
2.5 enclosure provided against safety hazards are
Exterior surface of the electro- intact and the device is operating
medical apparatus, including all as desired.
accessible parts, knobs, grips,
and shafts. 2.13 output part:
Part of the electromedical
2.6 exposed electrically
conductive surface
External metal or otherwise
apparatus, other than a patient-
applied part, that is intended to
deliver output signal voltages or
IV.
electrically conductive surface that currents to other equipment.
is connected to the internal
circuits, mechanisms, or enclosure 2.14 patient-applied part
of an electromedical apparatus. Entirety of any part of the
equipment that comes
2.7 input part intentionally into contact with the
Part of the electromedical patient via a patient connection.
apparatus, other than a patient-
applied part, that is intended to 2.15 patient-applied risk current
receive input signal voltages or Current flowing from the
currents from other equipment. electromedical apparatus through
the patient to power ground (earth)
2.8 isolated patient connection or between patient-applied parts.
Connection between the patient
and the electromedical apparatus 2.16 patient care vicinity
that is isolated from power ground Space, within a location intended
(earth)1), the utility power system, for the examination or treatment of
and other supporting circuitry to patients, extending 6 feet (ft)
such a degree that the risk current (1.8 meters [m]) beyond normal
flowing through the connection location of the bed, chair, table,
does not exceed the limits given in treadmill or other device that
Table 1—Summary of risk current supports the patient during
requirements in rms microamperes examination and treatment. The
(mA), provided in Section 4.2. patient care vicinity extends
vertically to 7 ft, 6 inches (in)
2.9 modular apparatus (2.3 m) above the floor.
Electromedical apparatus that
includes modules in its 2.17 patient connection
construction. Deliberate connection that can
carry current between an electro-
2.10 module medical apparatus and a patient.
Self-contained assembly that This can be a surface contact (e.g.,
performs a function or class of an ECG electrode), an invasive
functions in support of the major connection (e.g., an implanted wire
function of an electromedical or catheter), or an incidental long-
apparatus. term connection (e.g., connective
tubing).
NOTE — Modules can generally be -15-
NOTE — As used in this standard, 3.1 Classification of
"patient connection" is not electromedical apparatus
intended to include adventitious or For purposes of this standard, four
casual contacts, such as push categories of electromedical

Electrical buttons, bed surfaces, lamps, and


hand-held appliances.
apparatus have been defined. For
each category, risk currents are
established. These four categories

Safety 2.18 patient isolation risk current


Current flowing from the patient to
power ground (earth) through a
are listed below :

a) Electromedical apparatus with


part applied to the patient due to isolated patient connection:
Made Easy the unintended introduction of a Electromedical apparatus
voltage from an external source on intended to be connected to
the patient. the patient with the patient
circuit isolated from power
2.19 risk current: ground (earth), utility power
Nontherapeutic current that can systems, and other circuitry.
flow through the patient, medical
staff, or bystander as a result of the b) Electromedical apparatus

IV. use of electromedical apparatus.

2.20 single fault condition (SFC)


with nonisolated patient
connection: Electromedical
apparatus intended to be
Condition in which a single means connected to the patient.
of protection against a safety
hazard in equipment is defective, a c) Electromedical apparatus
component failure could increase likely to contact the patient:
the risk current, or a single external Electromedical apparatus that
abnormal condition exists. does not have a patient-
applied part, but that is
2.21 sink current: intended for use in the patient
Current that flows into a device or care vicinity.
any part thereof, when an external
voltage is applied to it. NOTE — See Section 2 for
definition of patient care vicinity.
2.22 source current
Electrical current that flows from d) Electromedical apparatus with
any part of an electromedical no patient contact:
apparatus to any other part or to Electromedical apparatus that
power ground (earth), when no is intended for use outside the
external voltages are applied. patient care vicinity and that
has no patient connections.
2.23 therapeutic current:
Current that is intentionally 3.2 Classification of
applied to the patient for treatment measurement conditions
of disease or disorder. The following Sections define
normal and fault conditions.

3 CLASSIFICATION OF 3.2.1 Normal operating conditions


ELECTROMEDICAL Under normal operating
conditions, a device is operating as
APPARATUS AND designed with all means provided
MEASUREMENT for protection against safety
CONDITIONS hazards intact, connected properly
and securely to an approved power
The following Sections define source and, if the device includes
specific classes of electromedical patient-applied parts, with such
apparatus and measurement parts applied according to the
conditions and detail how these manufacturer's instructions. The
classifications are applied in this following are considered normal
-16- standard. operating conditions:
a) power switch on/power switch NOTE — Labeling of isolated
off; patient connections shall comply
with symbol number 3 of table DII,
b) power polarity normal/power page 329, in the International
polarity reversed (cord-
connected apparatus only);
Electrotechnical Commission (IEC)
standard 601-1, second edition
(IEC, 1988).
Electrical
c) patient grounded; patient not
grounded. 4.1.2 Information manuals
The manufacturer shall supply the
Safety
3.2.2 Single fault condition user with operating and
A single failure of a device's maintenance instructions Made Easy
protection mechanism against a specifying how the electromedical
safety hazard or the failure of a apparatus should be operated and
single device component can maintained to prevent the device's
introduce a hazard condition or risk current from increasing
lead to the existence of an external beyond the limits set by this
hazardous condition. The standard for its particular category
following are considered to be (refer to Section 3.1). In addition,
single fault conditions :

a) power ground (earth)


the manufacturer shall disclose the
risk current category for which the
apparatus is designed and shall
IV.
conductor open; identify the specific limits defined
by this standard for that category.
b) short circuit of either barrier of
double insulation; 4.2 Risk current requirements
(general)
c) failure of a single component Electromedical apparatus shall
that can produce a hazardous meet the applicable risk current
current; limits of this standard under
normal conditions and under the
d) (for equipment that is not single fault conditions specified in
intended to be grounded) the the test methods of Section 5.
application of line voltage to Table 1 (see next page) summarizes
an input or output part or to these requirements.
accessible conductive
hardware of the enclosure; 4.2.1 Apparatus interconnection
Electromedical apparatus shall
e) (for electromedical apparatus meet the risk current limits of this
with isolated patient standard when manufacturer-
connections) the application of designated auxiliary apparatus,
line voltage on a patient- modular apparatus, or accessories
applied part. are attached in the quantity and
combinations stipulated by the
manufacturer. The manufacturer
shall supply the user (and shall
4 REQUIREMENTS label the apparatus) with
limitations and with directions for
the interconnection of modular
4.1 Labeling and documentation apparatus, accessories, and
requirements auxiliary apparatus, and with
directions for the use of
convenience receptacles.
4.1.1 Isolated patient
connections (labeling) 4.2.2 Cleaning and sterilization
Patient connections that meet the Electromedical apparatus shall
requirements of this standard for meet the risk current limits of this
isolated patient connections shall standard after exposure to any
be identified as being isolated at disinfection or sterilization process
the connector of the apparatus. specified by the manufacturer. -17-
4.2.3 Environmental conditions NOTES :
Electromedical apparatus shall 1) The frequency-weighted
meet the risk current limits of this network compensates for the
standard after exposure to the allowable increase in risk

Electrical nonoperational environmental


conditions (e.g., storage, trans-
portation, etc.) and under the
current limits with increasing
frequency. For measurement
purposes with a voltmeter as

Safety worst-case environmental


operating conditions specified
by the manufacturer.
shown, the limit remains
constant at 1 mA/mV,
independent of frequency.
With the meter connected, the
Made Easy 4.3 Enclosure risk current entire circuit is called the "risk
current tester."
4.3.1 General
Enclosure risk current, when 2) Refer to Section 5.7.2 for
measured with the AAMI standard component requirements and
test load, is that current that flows tolerances.
between power ground (earth) and
4.3.2 Risk current limits

IV. a) exposed chassis conductive


surfaces or hardware; or
Limits for enclosure risk current for
all categories of electromedical
apparatus, whether battery-
b) a 200 cm2 (centimeters powered, cord-connected, or
squared) foil in contact with a permanently connected, and under
nonconducting enclosure. both normal and single fault
conditions, are shown in Table 1 on
next page.

10,000 Ω

0.015 µF
MILLI-
INPUT 1,000 Ω VOLTMETER
100 Ω

Figure 1
AAMI standard test load .

-18-
Patient- Patient
Category Cord- Applied Risk Isolation General Other

Electrical
Connected/ Perma- Current Risk Current Cord- Cord- Perma-
Battery- nent (source (sink connected connected)1 nent2)
current) current)
Normal
Condition

ISOLATED
Single Fault
100 mA 100 mA 10 mA N/A 500 mA 2,500 mA 5,000 mA
Safety
Condition 300 mA 5,000 mA 50 mA 50 mA 1,000 mA 5,000 mA 10,000 mA Made Easy
Normal
Condition 100 mA 100 mA 10 mA N/A 500 mA 2,500 mA 5,000 mA

NON-ISOLATED
Single Fault
Condition 300 mA 5,000 mA 100 mA N/A 1,000 mA 5,000 mA 10,000 mA

Normal
Condition 100 mA 100 mA N/A N/A 500 mA 2,500 mA 5,000 mA IV.
LIKELY TO CONTACT PATIENT
Single Fault
Condition 300 mA 5,000 mA N/A N/A 1,000 mA 5,000 mA 10,000 mA

Normal
Condition 100 mA 100 mA N/A N/A 500 mA 2,500 mA 5,000 mA

NO PATIENT CONTACT
Single Fault
Condition 500 mA 5,000 mA N/A N/A 1,000 mA 5,000 mA 10,000 mA

Table 1
Summary of risk current requirements in rms microamperes (dc to 1 kHz)

1) Equipment that has no 2) Equipment specified to be


protectively grounded permanently installed with a
(earthed) accessible parts and protective power ground
no means for protective (earth) that is electrically
grounding (earthing) of other connected and secured at a
medical equipment and which specific location so that the
complies with the applicable connection can only be
requirements for enclosure loosened or moved with the
leakage current and patient aid of a tool.
leakage current; also mobile
x-ray equipment and mobile
equipment with mineral
insulation.

-19-
4.4 Patient-applied risk current 4.5.2 Risk current limits
(source current)2) Limits for patient isolation risk
current for electromedical
4.4.1 General apparatus with isolated patient-

Electrical Patient-applied risk current, when


measured with the AAMI standard
applied part(s), whether cord-
connected or permanently
connected, are shown in Table 1.

Safety test load, is that current that flows


between any patient-applied part
and :
NOTE — The current shall be
measured at the patient end of the
cable when connected to the
Made Easy a) power ground (earth); device.
b) exposed chassis conductive
surfaces or hardware; 4.6 Earth risk current
c) a 200 cm2 foil in contact with
a nonconducting enclosure; or 4.6.1 General
d) any other patient-applied Earth risk current, when measured
parts. with the AAMI standard test load, is
that current that flows in the

IV. Patient-applied risk current is also


that current that flows between all
patient connections tied together
protective earth conductor (ground
conductor).

and (a), (b), (c), and (d) listed above NOTE — Not applicable to double
when measured with the AAMI insulated devices using a two-wire
standard test load. power cord.

4.4.2 Risk current limits 4.6.2 Risk current limits


Limits for patient-applied risk Limits for earth risk current for all
current for all categories of categories of electromedical
electromedical apparatus, whether apparatus, whether cord-
battery-powered, cord-connected, connected or permanently
or permanently connected, under connected, and under both normal
both normal and single fault and single fault conditions, are
conditions, are shown in Table 1. shown in Table 1.

NOTES : 4.7 Risk current limits


versus frequency
1) These limits are not applicable The risk current limits specified in
to electromedical apparatus Table 1 are for frequencies from dc
without direct patient-applied to 1 kilohertz (kHz). Above 1 kHz,
connections. the limit is increased
2) The current shall be measured proportionally to a maximum value
at the patient end of the cable 100 times the limit at 1 kHz. Above
when it is attached to the 100 kHz, the limit is that which is
device. The cable is specified determined for 100 kHz (see Figure
by the manufacturer. 2). The use of the AAMI test load
automatically compensates for
4.5 Patient isolation risk current frequency.
(sink current)
5 Tests
4.5.1 General This Section contains referee tests
Patient isolation risk current, when and procedures by which
measured with the AAMI standard compliance with the requirements
test load, is that current that would of Section 4 and Table 1 can be
flow into a patient-applied part if determined.
the patient came into direct
contact with a potential of 120 volts WARNING — These tests can
(V), 60 hertz (Hz) with respect to expose personnel to hazardous
power ground (earth). electric shock and must be carried
-20- out with caution.
100 mA
Electrical
NORMALIZED CURRENT

Safety
Made Easy
1 mA

500 Hz 1000 Hz 10 KHZ 100 KHZ 1 MHZ 10 MHZ


FREQUENCY (HZ)

Figure 2 — Normalized current limits versus frequency

5.1 Compliance with the Instruments that indicate true rms


IV.
labeling requirements microamperes and have internal
Compliance with the labeling and frequency compensation identical
documentation requirements of to that shown in Figure 1 meet the
Section 4.1 shall be verified by requirements of this Section if the
inspection. measurement indicated does not
display an error of greater than 5
5.2 Compliance with the percent of reading and resolves a
risk current requirements signal as small as 1 mA in the band
(general test procedures) from dc to 100 kHz.
The risk currents of electromedical
apparatus shall be measured by the 5.2.1.2 Power source
methods described in this Section.
5.2.1.2.1
5.2.1 Test equipment For line voltage powered
and power system equipment, the tests shall be
performed on a grounded power
5.2.1.1 Measuring instruments system at the rated line voltage
The risk current tester consists of plus 10 percent. In the grounded
the AAMI standard test load and a system, the potential between the
millivoltmeter as shown in Figure 1. neutral and grounding conductors
The millivoltmeter shall measure at the receptacle selected for the
true rms volts; however, it may be test shall not exceed 3 V.
calibrated to true rms micro-
amperes (mA) by employing a 5.2.1.2.2
conversion factor of one micro- The power ground (earth) terminal
ampere per millivolt (mV). The used in these tests shall be the
millivoltmeter shall have an input grounding terminal of the specific
impedance of at least 1 megohm receptacle powering the
and have a bandwidth of dc to at instrument under test.
least 1 megahertz (MHz) (–3
decibels). In the band from dc to 5.2.1.2.3
100 kilohertz (kHz), the indicated Battery-powered apparatus shall be
measurement shall not display an tested while powered by the type of
error of greater than 5 percent of battery recommended by the
reading, and shall resolve a signal manufacturer and, if applicable,
as small as 1 mV. while connected to line power.
-21-
5.2.2 Test conditions delivers therapeutic energy to the
patient (e.g., a pacemaker), the
5.2.2.1 General therapeutic energy shall be zero
First, the apparatus shall be during the test. Otherwise, the

Electrical disconnected from all other


apparatus except auxiliary
apparatus, modular apparatus, or
instrument shall be in the active or
operable mode; i.e., with output
switches closed, with electrodes

Safety accessories, as defined in the


normative definitions (see Section
2). A single- or multi-function
properly connected to dummy
loads, and with final circuit stages
properly functioning but without a
apparatus in a cabinet or in physiological drive signal.
Made Easy multiple cabinets with a single
power cord connection is tested as 5.2.2.4 Operation
a single apparatus. Each individual During the test, the apparatus shall
apparatus shall also be tested run through a normal cycle and
independently if described by the activate all accessories and/or
manufacturer as a stand-alone auxiliary apparatus.
apparatus. Tests shall be con-
ducted at the rated line voltage

IV. plus 10 percent.

5.2.2.2 Nonconducting enclosure


5.3 Enclosure risk current

5.3.1 Application
The risk current shall be measured The enclosure risk current tests
from an electrically conductive foil shall apply to cord-connected, line-
the size of the enclosure, but not to powered apparatus, to battery-
exceed 200 cm2, in immediate powered apparatus with the
contact with the enclosure. The foil charger connected, and to
shall be placed at a location— permanently connected apparatus.
determined by experimentation—
such that the current measured to 5.3.2 Cord-connected, normally
power ground (earth) is a grounded apparatus
maximum. If exposed chassis
hardware is likely to be touched by 5.3.2.1
personnel, then the hardware shall Using the test circuit of Figure 3,
be treated as an exposed the enclosure risk current shall be
electrically conductive surface. measured:
Testing of nonconductive exposed
surfaces of patient wiring and a) between enclosure and power
cables is not required. ground (earth);
b) between electrically
5.2.2.3 Controls conductive surfaces and
During risk current tests, all power ground (earth);
operator-accessible controls shall c) between a 200 cm2 foil in
be adjusted to yield the largest risk contact with the non-
current found by experiment. If the conducting enclosure and
electromedical apparatus normally power ground (earth).
Reversing
Switch
200 cm2 fail
S1
Rated
Line Apparatus Select
Voltage Under Per
+10% Test 5.3.2.1

Ground M
Ground Switch
S2 Ground Wire Exposed
Conductive Surface Risk Current
Tester
-22-
Figure 3 — Enclosure risk current test circuit (normally grounded)
NOTE — The 200 cm2 foil line 5.3.3
leading from the "select" box to the Cord-connected, normally
apparatus under test refers to the ungrounded apparatus
connective mode with insulated
apparatus. The line with an arrow
leading from the "select" box to the
apparatus under test refers to
5.3.3.1
Using the test circuit of Figure 4
(see next page), the enclosure risk
Electrical
connections made with conductive
enclosure.
current shall be measured:

a) between enclosure and power


Safety
5.3.2.2 ground (earth);
Each measurement is performed b) between electrically conduc- Made Easy
when: tive surfaces and power
ground (earth);
a) the utility electricity supply c) between a 200 cm2 foil in
polarity is normal and when contact with the non-
the utility electricity supply conducting enclosure and
polarity is reversed (by power ground (earth).
reversing S1). These are
normal conditions;
b) the apparatus power switch is
on; the apparatus power
5.3.3.2
Each measurement is performed
when:
IV.
switch is off. These are
normal conditions; a) the utility electricity supply
c) the ground switch (S2) is open; polarity is normal; the utility
the ground switch is closed. electricity supply polarity is
The first is a single fault reversed (by reversing S1).
condition; the second is a These are normal conditions;
normal condition; b) the apparatus power switch is
d) each barrier of double insul- on; the apparatus power
ation is short circuited. These switch is off. These are
are single fault conditions. normal conditions;
c) each barrier of double
NOTE — The test methods for all insulation is short-circuited.
measurement conditions are not These are single fault
supplied in this standard because conditions.
they are device- and circuit-
specific. NOTE — The test methods for all
measurement conditions are not
5.3.2.3 The power on/power off supplied in this standard because
test also applies to apparatus with they are device- and circuit-
nonrechargeable batteries. specific.

Reversing 200 cm2


Switch fail
S1
Rated Line Apparatus Select
Voltage Under Per Risk Current
+10% Test 5.3.2.1 Tester

Exposed
Conductive Surface

Ground

Figure 4 — Enclosure risk current test circuit (normally ungrounded) -23-


5.3.3.3 The power on/power off e) all patient connections
test also applies to apparatus with shorted together and power
nonrechargeable batteries. ground (earth);
f) all patient connections

Electrical 5.3.4 Permanently


connected apparatus
Before line-powered apparatus is
shorted together and any
exposed, electrically
conductive surface;

Safety permanently installed, the


enclosure risk current shall be
measured according to the
g) all patient connections
shorted together and a
200 cm2 foil in contact with
procedures described in 5.3.2. the nonconducting enclosure;
Made Easy h) any patient connection and all
5.4 Patient-applied risk current other patient connections
(source current) connected together.

5.4.1 Application NOTE — The test methods for all


The patient-applied risk current measurement conditions are not
tests of this Section shall apply to supplied in this standard because
line-powered and battery-powered they are device- and circuit-

IV. electromedical apparatus that has a


patient connection(s).
specific.

5.4.2.2
5.4.2 Apparatus with isolated Each measurement is performed
patient connection when:

5.4.2.1 a) the utility switch (S1) is


Using the test circuit of Figure 5 normal/the utility switch is
(see next page), the patient-applied reversed. These are normal
risk current (source current) shall conditions;
be measured between: b) the apparatus power switch is
on/the apparatus power
a) any patient connection and switch is off. These are normal
power ground (earth); conditions;
b) any patient connection and c) the ground switch (S2) is
any exposed, electrically open/the ground switch is
conductive surface; closed. The first is a single
c) any patient connection and a fault condition; the second is a
200 cm2 foil in contact with normal condition;
the nonconducting enclosure;
d) any patient connection and
any other patient connection;

120K
Select Per
6.4.2.2.d 120V / 60Hz

Reversing
Patient
Switch
Connection
S1
Rated Line Select Per
Apparatus
Voltage 5.3.2.1
Under
+10%
Test

200 cm2 Select Per


Ground Switch
fail 5.3.2.1
S2

Exposed Risk Current


Conductive Surface Tester

-24-
Figure 5 — Patient-applied risk current test circuit
d) line voltage is applied to an other external electrical connection
input or output part or to on the apparatus grounded. This
accessible conductive hard- test shall be performed with the
ware of the enclosure, if not apparatus both on and off and
grounded under normal
conditions. These are single
fault conditions;
properly connected to its electrical
supply. The patient cable shall be
placed 20 cm away from a
Electrical
e) each barrier of double
insulation is short circuited.
These are single fault con-
grounded surface.

NOTE — The 120 K resistance is


Safety
ditions. intended to protect the test
operator. Made Easy
NOTE — The 120 K resistance is
intended to protect the test 5.6 Earth risk current
operator.
5.6.1 Application
5.4.3 Apparatus with nonisolated The earth risk current test shall
patient connection apply to cord-connected apparatus,
The patient-applied risk current to battery-powered apparatus with
shall be measured by the
procedures described in 5.4.2.
the charger connected, and to
permanently connected apparatus. IV.
5.5 Patient isolation risk current 5.6.2
(sink current) Cord-connected, normally
The patient isolation risk current grounded apparatus
shall be measured in each
individual patient connection that 5.6.2.1
is labeled "isolated" when a Using the test circuit of Figure 7
potential of 120 V rms, 60 Hz, is (see next page), the earth risk
applied through a series 120 kilohm current shall be measured in the
resistance to the labeled patient protective power ground (earth).
connection, as shown in Figure 6.
The patient isolation risk current is 5.6.2.2
measured with respect to power Measurement shall be performed
ground (earth). For solely battery- when:
powered apparatus, the patient
isolation risk current is measured a) the utility electrical supply is
with respect to an electrically normal/when the utility
conductive surface on which the electrical supply is reversed (by
apparatus is positioned, and with reversing S1). These are normal
an exposed conductive surface or conditions;

1:1

Rated Apparatus 120V


Select Per
Line Under 60Hz
5.5
+10% Test
120K

Ground

Risk Current
Tester

Figure 6 — Patient isolation risk current test circuit -25-


b) the apparatus power switch is shall be the largest current
on/the apparatus power switch measured during any of the
is off. These are normal required tests and conditions. The
conditions; apparatus must meet all applicable

Electrical c) each supply conductor is


interrupted, one at a time
(opening S2 and S3 in turn).
limits of Table 1.

5.7.2 AAMI standard test load

Safety This is a single fault condition;


d) each barrier of double
insulation is short-circuited.
As shown in Figure 1, the test load
shall be constructed using metal-
film resistors with a tolerance of
This is a single fault condition. 1 percent or better, and a mica- or
Made Easy plastic-dielectric (extended foil)
5.6.3 Permanently connected capacitor with a tolerance of 5
apparatus percent or better. The AAMI
Before the line-powered apparatus standard test load has an
is permanently installed, the earth impedance frequency charac-
risk current shall be measured teristic (Figure 8) which is the
according to the procedures approximate inverse of the curve of
described in 5.6.2. Figure 2, which shows risk current

IV. 5.7 Risk current limits


versus frequency
versus frequency.

5.7.3 Risk current calculation


Using the AAMI standard test load
5.7.1 General of Figure 1 and a voltmeter
When multiple risk currents of calibrated to indicate rms
various frequency and phase millivolts, the weighted risk current
relationships are present during a is read directly from the meter,
single test, the resultant risk because:
current is related to the voltage
across the AAMI standard test load. V(mV rms)
The risk current of an apparatus I(mA rms) =
Z(k ohms)
Line Interruption Switches
S2•
Reversing
Switch S1
Rated Line
Voltage Apparatus
+10% Under
S3• Test

Protective
Earth Risk
(Ground Current
Conductor) Tester

Figure 7 — Earth risk current test circuit


Figure 8 —
Relative Magnitude (dB)

Relative
0 frequency
characteristics of
millivoltmeter
-20 reading in AAMI
standard test
load of Figure 1
-40
-26- 10 101 102 103 104 105
A.1 General the electromedical apparatus itself.
The rationale discusses the need The standard does not require
for the standard and describes the nonisolated patient connections to
basic underlying principles, carry any special labeling.
empirical data, assumptions, and
sources that support the
requirements and test methods
Electrical
adopted in the standard.

A.2 Need for the standard


A.4.1.2 Information manuals
By identifying the risk current
classification and risk current
Safety
This standard seeks to reduce the limits, the manufacturer is
risk of inadvertent electric shock informing the user of the device's Made Easy
from medical devices. In intended purpose as that purpose
particular, it concerns itself with relates to the risk of electric shock.
the risk of injury from the small Any special user actions required to
currents that inevitably flow from ensure that the risk current limits
or to electromedical apparatus. are maintained throughout the life
The intent of the AAMI Electrical of the equipment should be
Safety Committee was to develop a described in the operating
general baseline standard. The
extent to which the standard
should be applied is to be
instructions or maintenance
manuals. IV.
determined by individual A.4.2 Risk current requirements
institutions, standards groups, and (general)
other authorities. The committee felt that grounding
should not be the primary
A.3 Classification of approach to limiting the risk of
electromedical apparatus and electric shock because it is possible
measurement conditions to have a single fault failure in the
Changes from the second edition of grounding system. Redundant
the standard (AAMI, 1985) have means of grounding are possible
been made in keeping with but are controlled by the user and
changes to the requirements given not the manufacturer of the
in Section 4. electromedical apparatus. If the
grounding is lost or if other safety
A.4 Rationale for the specific means fail, the risk current
provisions of the standard available from the enclosure should
not represent a substantial hazard
A.4.1 Labeling and to the patient.
documentation requirements
The risk current limits were
A.4.1.1 Isolated patient changed for certain categories as
connection (labeling) compared to the previous version
Fault conditions can contribute to of this standard (ANSI/AAMI ES1—
patient risk due to source and sink 1985). These changes have been
currents. The greatest risk is with made in order to bring this
direct cardiac applications; such document into closer harmon-
applications should utilize isolated ization with risk current limits
patient connections. In order to specified in the International
better manage patients requiring Electrotechnical Commission
direct cardiac connection, the user standard, Medical electrical
should be able to readily identify equipment—Part 1: General
electromedical apparatus with requirements for safety (IEC 601-1,
isolated patient connections. 1988).
Therefore, labels should appear on

-27-
In its review of the allowable risk disturbance in rhythm in
current levels, the committee humans was 80 mA. In a
considered the following: second study, Watson
a) Likelihood of stimulation of (1973) found that the

Electrical excitable tissue. The likelihood


of stimulation of excitable
tissue depends upon:
smallest current that pro-
duced ventricular fibril-
lation in humans was

Safety 1) the location of the sites at


which current enters and
leaves the body;
15 mA.
4) Mechanically induced
ventricular fibrillation has
2) the area of contact; been observed during
Made Easy 3) the amount of current cardiac catheterization at
flowing; zero current.
4) the susceptibility to 5) The human data obtained
mechanical stimulation; by Starmer (1973) and
5) the presence of a fault Watson (1973) follow,
condition; reasonably well, a normal
6) the probability of the distribution for currents to
current having a given 300 mA. All patients are

IV. value. Medical devices have


been classified into the risk
categories described in
not equally susceptible to
current-induced ventricular
fibrillation. According to
Section 3 because of the Figure A.1, there is approx-
different magnitudes of risk imately a 1 percent prob-
associated with these ability of fibrillation at
categories. 30 mA.
b) Survey of published data. 6) Current perception is a
During the several years since function of contact
the publication of the previous location, contact pressure,
version of this standard skin condition, moisture,
(ANSI/AAMI ES1—1985), and contact area. Experi-
experience has been gained ments report a wide range
with respect to the incidence of current perception.
of problems related to risk Dalziel (1968) reports that
current and the probability of only approximately 1 per-
occurrence of the potential cent of the population can
hazard. The available pub- perceive 500 mA passing
lished data on currents from the fingers of one
causing ventricular fibrillation hand to the fingers of the
in humans have also been other hand. Tan and
reexamined. The following Johnson (1990) report that
have been noted: 300 mA produces a strong
1) The combination of an sensation for electrodes
open power ground (earth) placed 10 cm apart on the
wire and a person touching upper arm. Levin (1991)
both a conductive part of reports that nearly all the
the enclosure and the population will perceive
patient is a low probability 500 mA without any reac-
event. tion for current passing
2) The combination of an from the finger on one
open power ground (earth) hand to the underside of
wire and a person touching the wrist on the other hand.
both a conductive part of Levin further reports that,
the chassis and the distal on the underside of the
end of an invasive cardiac wrist, the stratum corneum
connection is a low (layer of dead material on
probability event. the skin surface) is not as
3) In one study, Raftery (1975) thick as that on the fore-
found that the smallest finger and, therefore, the
-28- current that produced a
sensitivity to current Auxiliary power outlets may be
perception might be higher. provided for powering additional
Startle current is that level devices. The labeling requirements
of perception current that, provide some assurance that the
when first perceived, might
result in a nurse or other
clinician's involuntary
user has appropriate guidance
about the limitations applying to
equipment or accessories con-
Electrical
reaction to the sudden
sensation of perception
current. This uncontrolled
nected to an auxiliary power
receptacle. Safety
reaction is of great concern. A.4.2.2 Cleaning and sterilization
7) Worldwide, since the The long-term effects of repeated Made Easy
advent of risk current disinfection or sterilization of a
standards, concern about device on risk currents must be
grounding, and use of considered because of possible
better practices in handling degradation of insulating materials.
catheters and invasive
cardiac connections, there A.4.2.3 Environmental conditions
have been no reports of Temperature, humidity, atmos-
incidents involving risk
current passing through
the patient.
pheric pressure, mechanical shock,
and similar environmental con-
straints can have an effect upon
IV.
the risk currents. To protect the
NOTE — A minority of the AAMI patient, the risk current limits must
Electrical Safety Committee were also be met in the intended
opposed to increasing risk current environment.
limits unless scientific studies
supported higher limits. A.4.3 Enclosure risk current
The 100 mA NC values were
Also, in order to harmonize with selected for cord-connected and
IEC 601-1 and to allow for permanently connected apparatus
additional fault conditions as on theoretical grounds. For a
compared to the current standard, typical ground resistance of
the test measurement classifi- 0.2 ohms, 100 mA of enclosure risk
cations of "normal condition" (NC) current, measured as in Figure 3,
and "single fault condition" (SFC) requires that 500 mA flow in the
were introduced. power ground (earth) wire. This
would be a major fault condition.
Measuring risk current under SFC
is important because: The 300 mA SFC values for isolated,
— components can fail; nonisolated, or likely-to-contact-
— single faults exist that are not patient, cord-connected apparatus
now considered in the were selected on the basis of
standard; reaction current measurements by
— faults can occur in accessory Levin and the low probability of
equipment. risk current reaching the distal end
of an invasive heart connection via
A.4.2.1 Apparatus interconnection another person. Levin reported
The total risk current associated that currents of 300 mA will not
with a device can be a function of produce sensations leading to a
the modules, accessories, and "startle" reaction (Levin, 1991).
interconnections used with the
device. Voltage differences can The 500 mA limit for SFC for cord-
occur between different parts of a connected, no-patient-contact
device, particularly if current flows equipment was selected because
in the grounding circuit. Thus, a there is no concern about this
remote accessory powered from the current reaching the patient.
device or from a separate source is, The 5,000 mA limit for SFC for
for purposes of the standard, permanently connected equipment
considered part of the device. is basically a power ground (earth) -29-
wire current, because the that produced ventricular fibril-
apparatus is, by definition, lation in humans are considered,
permanently grounded. This is the the probability is approximately
current allowed from the enclosure zero.

Electrical of permanently connected


equipment if the power ground
(earth) wire were opened. The
The 50 mA SFC limit for isolated
equipment was selected because

Safety permanently connected power


ground (earth) wire is not expected
to open.
the probability of causing ventri-
cular fibrillation is low if the data of
Figure A.1 are extrapolated and is
approximately zero if only
Made Easy A.4.4 Patient-applied risk current observed values of currents
(source current) producing ventricular fibrillation in
The 10 mA NC limit was selected as humans are considered.
the current that may flow directly
into the heart continuously. Figure The 100 mA SFC limit was selected
A.1 shows that the probability of for nonisolated equipment,
inducing ventricular fibrillation is because if this current enters and
very small if the data on which the exits the surface of the body, then

IV. chart is based are extrapolated. If


only observed values of currents
only a fraction will reach the heart.

Starmer 60 Hz D = 1.25mm

Watson 60 Hz D = 2.0mm

99.9

99

95 +
Cumulative Percent

+
90
+
+
+
50 +
+
+
20
+

5 +

0.1
0 50 100 150 200 250 300 350 400 450 500

Microamperes
Figure A.1 — Normal probability plot

-30-
A.4.5 Patient isolation risk current ground (earth). Contact with such
(sink current) elements and a second device
The 50 mA SFC value for isolated under test will result in current
equipment was allowed because of flow. Thus, the committee felt that
the low probability of line voltage
appearing on a patient, and
because of the low probability of
leaving the earth risk current
unmeasured and unlimited con-
stituted a potential hazard that
Electrical
50 mA inducing ventricular
fibrillation. For line voltage to
appear on a patient, a power
should be avoided.

The allowable values for earth risk


Safety
ground (earth) wire must be open current detailed in this standard
and there must also be a fault in are not thought to pose a direct Made Easy
basic insulation. If the data in hazard as the current is safely
Figure A.1 are extrapolated, then returned to earth. The values
50 mA has a low probability of selected were chosen to avoid any
inducing ventricular fibrillation. If significant increase in the current
only recorded human data are flowing through the protective
considered, the probability of grounding system of the instal-
inducing fibrillation at 50 mA is lation and to be consistent with the
approximately zero.

A.4.6 Earth risk


limits of IEC 601-1 for power
ground (earth) and nonconducting
enclosures. Further, the 2.5 mA
IV.
(ground risk) current limit in normal mode is within the
The original standard of 1978 and limit for isolation monitors set by
the revised edition of 1985 did not the National Fire Protection
include earth current as a potential Association (NFPA, 1993, Section
risk current. This was not 3.4.3.3), which specifies that
considered to be an issue as most isolation monitors should not
medical devices of the era utilized alarm at 3.75 mA.
conductive enclosures requiring
grounding. Thus, earth current was A.4.7 Risk current limits
effectively measured as enclosure versus frequency
(chassis) current under the open Figure 2 of the standard was
ground condition. derived from strength/frequency
data for perceptible and lethal
In the last 25 years, however, the currents (Geddes and Baker, 1971).
change to nonconductive The flat portion between 100 kHz
enclosures negated this equality. and 1 MHz does not reflect
The enclosure current is now physiological data obtained with
measured as the capacitive- purely sinusoidal currents.
coupled current to a 200 cm2 foil in Stimulation has been observed
contact with the enclosure. This with complex waveforms at high
current bears little resemblance to frequencies, but little data are
the earth risk current due to the available. In the absence of data, it
current-limiting characteristics of was deemed prudent not to
the capacitive coupling of the extrapolate beyond 100 kHz.
nonconductive enclosure.
A.5 Tests
The earth risk current does not The test procedures documented in
pose a direct risk to the patient or Section 5 of the standard provide
medical personnel. However, referee test methods for verifying
excessive earth risk current, either compliance with the requirements
by design or internal breakdown, of Section 4. These referee tests are
will raise the potential of the not necessarily intended for
device's ground with respect to purposes of manufacturing or
true power ground (earth) as quality control (although these
represented by structural elements, applications are not precluded), as
modular wall units, cold water equivalent measurements may be
pipes, and other installed piping, or obtainable by other means.
by an adjacent receptacle power -31-
ASSOCIATION FOR THE ADVANCEMENT OF MEDICAL
INSTRUMENTATION. Safe current limits for electromedical apparatus.
ANSI/AAMI ES1—1985. Arlington (Vir.): AAMI, 1985. ISBN 0-910275-50-5.

Electrical DALZIEL, CF. Reevaluation of lethal electric currents. IEEE Trans Indus
Gen Appl, GA-4, 1968, vol. 1, no. 5, p. 467-476.

Safety GEDDES, LA. and BAKER, LE. Response to the passage of electric currents
through the body. J Assn Adv Med Instrum, 1971, vol. 5, p. 13-18.

INTERNATIONAL ELECTROTECHNICAL COMMISSION. Medical electrical


Made Easy equipment—Part 1: General requirements for safety, 2nd ed. IEC 601-1.
Geneva: IEC, 1988.

LEVIN, M. Perception of chassis leakage current. Biomed Instrumentation


and Technology, 1991, vol. 25, no. 2, p.135-140.

NATIONAL FIRE PROTECTION ASSOCIATION. Standard for health care


facilities. NFPA 99-1993. Quincy (Mass.): NFPA, 1993.

IV. RAFTERY, EB., GREEN, HL., and YACOUB, MH. Disturbances of heart
rhythm produced by 50 Hz leakage currents in human subjects.
Cardiovascular Research, March 1975, vol. 9, no. 2, p. 263-265.

STARMER, CF. and WHALEN, RE. Current density and electrically induced
ventricular fibrillation. J Assn Adv Med Instrum, 1973, vol. 7, no. 1, p. 3-6.

TAN, KS. and JOHNSON, DL. Threshold of sensation for 60 Hz leakage


current: Results of a survey. Biomed Instrumentation and Technology, 1990,
vol. 24, no. 3, p. 207-211.

Bibliography WATSON, AB., WRIGHT, JS., and LAUGHMAN, J. Electrical thresholds for
ventricular fibrillation in man. Med J Australia, 1973, vol. 1, p.1179-1182.

You can contact BAPCO:


BAPCO
3200 Sencore Drive
Sioux Falls, SD 57107

1-800-419-4000
http://www.bapcoinfo.com
email: bapco@bapcoinfo.com

-32-
FREQUENCE: The number of oscillations, per unit of to a patient, but may be conveyed from exposed metal

Glossary time, of a particular current or voltage waveform. The


unit of frequency is the Hertz (Hz). (The unit of frequency
used to be "cycles per second", a term no longer
parts of an appliance to ground or to other accessible
parts of an appliance.

LINE ISOLATION MONITOR: An instrument that


Of Terms preferred). Note : The waveform may consist of
components having many different frequencies , in which
case it is called a complex or nonsinusoidal waveform.
continually checks the hazard current from an isolated
surface to ground.
AMPACITY: Current-carrying capacity of electrical
conductors expressed in amperes. GROUND-FAULT CIRCUIT INTERRUPTER: MACROSHOCK: The effect of large electric current
A device whose function is to interrupt the electric circuit (milliamperes or larger) on the body.
ANESTHETIZING LOCATION: Any area of the to the load when a fault current to ground exceeds some
facility that has been designated for the administration of predetermined value that is less than that required to MICROSHOCK: The effect of small electric currents
operate the overcurrent protective device of the supply (as low as 10 microamperes) on the body. To be
any flammable or nonflammable inhalation anesthetic
circuit. hazardous, such currents must be applied to a conductor
agents in the course of examination or treatment,
inside or very near the heart.
including the use of such agents for relative analgesia.
GROUNDING SYSTEM: a system of conductors
CONDUCTIVE: Materials, such as metals, that are that provides a low-impedance return path for leakage mV: Millivolt.
commonly considered electrically conductive, and and fault currents. It coordinates with, but may be locally
materials that, when tested, have a resistance not more extensive than, the grounding system described in mA: Milliampere.
exceeding 1,000,000 ohms. Such materials are required Article 250 of NFPA 70, National Electric Code.
where electrostatic interconnection is necessary.
PATIENT CARE AREA: Any portion of a health care
HAZARD CURRENT: for a given set of connections facility where patients are examined or treated.
in an isolated power system, the total current that would Note : Business offices, corridors, lounges, day rooms,
CRITICAL EQUIPMENT: Equipment that is
flow through a low-impedance if it were connected dining rooms, or similar areas are not classified as patient
essential to the safety of the occupants in the facility.
between either isolated conductor and ground. The care areas.
CRITICAL SYSTEM: A system of feeders and branch various hazard currents are:
circuits in nursing homes and custodial care facilities Fault Hazard Current: The hazard current of a given PATIENT-CARE-RELATED ELECTRICAL
arranged for connection to the alternate power source to isolated power system with all devices connected APPLIANCE: An electrical appliance that is intended
except the line isolation monitor. to be used for diagnostic, therapeutic, or monitoring
restore service to critical receptacles, task illumination,
Monitor Hazard Current: The hazard current of the purposes in a patient care area.
and equipment.
line isolation monitor alone.
DIRECT ELECTRICAL PATHWAY TO THE Total Hazard Current: The hazard current of a given PATIENT EQUIPMENT GROUNDING
HEART: An external conductive pathway, insulated isolated system with all devices, including the line POINT: A jack or terminal that serves as a collection
isolation monitor, connected. point for redundant grounding of electrical appliances
except at its ends, one end of which is in direct contact
serving a patient vicinity or for grounding other items in
with the heart muscle and the other outside the body,
that is accessible for inadvertent or intentional contact IMPEDANCE: Impedance is the ratio of the voltage order to eliminate electromagnetic problems.
with grounded objects or energized, ground-referenced drop across a circuit element to the current flowing
sources. Catheters filled with conductive fluids and through the same circuit element. The circuit element PATIENT LEAD: Any deliberate electrical connection
may consist of any combination of resistance, capaci- that may carry between an appliance and a patient. This
electrodes, such as may be used for pacing of the heart,
tance, or inductance. The unit of impedance is the Ohm. may be a surface contact (such as an ECG electrode), an
are examples of direct electrical pathways to the heart.
invasive connection (such as an implanted wire or
DOUBLE-INSULATED APPLIANCES: INTRINSICALLY SAFE: As applied to equipment catheter), or an incidental long-term connection (such as
and wiring, equipment and wiring that are incapable of conductive tubing). Adventitious or casual contacts such
Appliances having an insulation system comprising both
releasing sufficient electrical energy under normal or as a push button, bed surface, lamp, hand-held
basic insulation necessary for the functioning of the
abnormal conditions to cause ignition of a specific appliance, and so fourth, are not considered patient
appliance and for basic protection against electrical
hazardous atmospheric mixture. Abnormal conditions leads.
shock and supplementary insulation. The supplementary
insulation is independent insulation provided in addition may include accidental damage to any part of the
to the basic insulation to ensure protection against equipment or wiring, insulation or other failure of PATIENT VICINITY: In an area in which patients are
electrical components, application of overvoltage, normally cared for, the patient vicinity is the space with
electric shock in case of failure to the basic insulation.
adjustment and maintenance operations, and other surfaces likely to be touched by the patient or an atten-
similar conditions. dant who can touch the patient. Typically in a patient
ELECTRODE: A device intended to probe an
room, this is a space within the room 6 ft. (1.8m) beyond
electrically conductive connection through a cable to a
patient. There are several types: ISOLATED PATIENT LEAD: A patient lead whose the perimeter of the bed in its normal location and
impedance to ground or to a power line is sufficiently extending vertically within 7 ft. 6 in. (2.3m) of the floor.
Active Electrode: An electrode intended to generate a
surgical effect at its point of application to the patient. high that connecting the lead to ground or to either
Dispersive Electrode: An electrode intended to com- conductor of the power line results in a current flow REACTANCE: The component of impedance
below a hazardous limit in the lead. contributed by inductance or capacitance. The unit of
plete the electrical path between patient and appli-
reactance is the Ohm.
ance, and at which no surgical effect is intended. It is
often called the "indifferent electrode", the "return ISOLATED POWER SYSTEM: A system
electrode", the "patient plate", or the "neutral comprising an isolating transformer or its equivalent, REFERENCE GROUNDING POINT: A terminal
a line isolation monitor, and ungrounded circuit bus that is the equipment grounding bus, or an extension
electrode".
conductors. of the equipment grounding bus, and is a convenient
collection point for installed grounding wires or other
EXPOSED CONDUCTIVE SURFACES: Those
surfaces that are capable of carrying electric current and ISOLATION TRANSFORMER: A transformer of bonding wires where used.
that are unprotected, uninsulated, unenclosed, or the multiple-winding type, with the primary and
unguarded, permitting personal contact. secondary windings physically separated, that inductively WET LOCATIONS: Those patient care areas that are
couples its ungrounded secondary winding to the normally subject to wet conditions, including standing
grounded feeder system that energizes its primary water on the floor, or routine dousing or drenching of the
FAULT CURRENT: A current in an accidental
winding. work area. Routine housekeeping procedures and
connection between an energized and a grounded or
incidental spillage of liquids do not define a wet location.
other conductive element resulting from a failure of
insulation, spacing, or containment of conductors. LEAKAGE CURRENT: Any current, including
capacitively coupled current, not intended to be applied
1-800-419-4000 • www.bapcoinfo.com

Potrebbero piacerti anche