Sei sulla pagina 1di 3

Proceedings of the 2018 Design of Medical Devices Conference

DMD2018
April 9-12, 2018, Minneapolis, MN, USA

DMD2018-6905

Downloaded from http://asmedigitalcollection.asme.org/BIOMED/proceedings-pdf/DMD2018/40789/V001T01A011/2787835/v001t01a011-dmd2018-6905.pdf by guest on 23 September 2020


A smartphone stethoscope and application for automated identification of
innocent Still’s murmur
Titus John Robin W. Doroshow Raj Shekhar
Sheikh Zayed Institute for Division of Cardiology Sheikh Zayed Institute for
Pediatric Surgical Innovation Children’s National Pediatric Surgical Innovation
Children's National Health Health System Children's National Health
System Washington, DC, USA System
Washington, DC, USA AusculTech Dx Washington, DC, USA
AusculTech Dx Dayton, MD, USA AusculTech Dx
Dayton, MD, USA Dayton, MD, USA

BACKGROUND

Primary care physicians (PCPs) often lack the skills to


distinguish the common innocent Still’s murmur from far
less frequent but potentially serious pathological heart
murmurs. This leads to approximately 800,000 children
being referred to pediatric cardiologists each year for
evaluation of heart murmurs in the United States [1-2].
The murmur is ultimately diagnosed as an innocent Still’s
murmur in approximately 78% of these children Figure 1. The current StethAid prototype that includes a
(Children’s National Health System data). These novel digital stethoscope and a custom mobile
unnecessary referrals and associated tests cost the application.
healthcare system over half a billion annually, and are a
source of avoidable anxiety for children and families while METHODS
waiting to see a pediatric cardiologist.
We have created a novel digital stethoscope (see For our first goal, we evaluated the developed digital
Figure 1), which connects to a smartphone, and stethoscope. Sounds (sinusoidal waves and normal heart
developed a custom mobile application that records heart sounds) for the given comparison test were played
sounds, using the developed hardware, and analyzes through speakers into the diaphragms of the StethAid and
them for the presence of Still’s murmur [3]. We call this Littmann stethoscopes. Sounds were played at the same
application and digital stethoscope StethAid. volume for the two devices.
Our first goal was to compare our device with a Next, we evaluated the frequency response of the two
commercially available electronic stethoscope (3M devices by recording the sinusoidal signals from 40 Hz to
Littmann model 4100) in terms of recording quality and 400 Hz in 20-Hz steps, corresponding to 19 recordings for
demonstrate that the two devices produced comparable each of the stethoscopes. Discrete 1024-point Fourier
sound representations in both the time and frequency transforms (FFTs) were performed on each of the
domains. recorded sine waves to produce one-sided power spectra
Our second goal was to demonstrate that our corresponding to each of the 19 recordings. These
automated Still’s murmur identification algorithm respective power spectra were summed in the frequency
produced comparable results, in terms of specificity and domain and a third order polynomial was fitted to the
sensitivity in identifying a Still’s murmur, for each of the summed response as shown in Figure 2.
two stethoscopes. Then we compared the reception of prerecorded
normal heart sounds on the devices. The sounds were
played through speakers into the diaphragms of the

1 Copyright © 2018 ASME


stethoscopes. Raw signals were band passed filtered
from 40 Hz to 400 Hz. Five cardiac cycles from each of
the recordings were manually segmented. An example
cardiac cycle recorded on each device is presented in
Figure 3. A 1024-point FFT was then performed on each
of the five cycles of each recording, and the resulting
spectra were averaged in the frequency domain to

Downloaded from http://asmedigitalcollection.asme.org/BIOMED/proceedings-pdf/DMD2018/40789/V001T01A011/2787835/v001t01a011-dmd2018-6905.pdf by guest on 23 September 2020


produce the respective curves seen in Figure 4. The
recordings from both devices were also compared to the
prerecorded normal heart sounds, in both time and
frequency domains.
For our second goal, we compared our automated
Still’s murmur identification algorithm [3] on Still’s and
pathological murmurs recorded using StethAid
stethoscope. Utilizing the Littmann stethoscope, we built
a database of different heart murmur types and their
associated diagnoses as provided by an expert pediatric
cardiologist. The identification algorithm was trained and
tested on the database of the Littmann sounds. 70% of
the murmur library was used to train the algorithm, while
30% comprised the test set.

RESULTS

Hardware Analysis - We show a comparable


frequency response between our device and the Littmann
stethoscope as shown in Figure 2. This response is
characterized as the bell-shaped response with increased
power in the 100 Hz - 350 Hz range.

Figure 3. Prerecorded source normal heart sounds


(black, top). Recorded sounds with StethAid (red, middle)
and with Littmann (blue, bottom) stethoscopes.

Figure 2. Comparison of the frequency response of


StethAid (red) and Littmann (blue) stethoscopes.

The comparisons shown in Figure 3 are of normal


heart sounds in the time domain. Generally, the
phonocardiogram of normal heart sounds has a Figure 4. Frequency response of normal heart sounds for
frequency content varying from approximately 40 Hz up to the two hardware devices.
300 Hz [4].
Murmur Analysis - The analysis algorithm has been
developed and tested using an in-house murmur library
with clinically documented diagnoses. The library

2 Copyright © 2018 ASME


consisted of 119 Still’s murmurs, 84 other innocent improve its accuracy as the murmur library grows and
murmurs, and 502 pathological murmurs [3]. introduce the improved algorithm to all users. The
Our improved algorithm offers a true positive algorithm can also be trained on StethAid recordings only
identification rate (Still’s murmur identified correctly) of or on mixed recordings from different hardware types.
89% with a false positive identification rate (pathological Once the algorithm is fully developed with clinically
murmur misidentified as Still’s) of 3% on Littmann acceptable sensitivity and specificity, it will be made
recordings. We have also evaluated our algorithm on the available for use by PCPs.

Downloaded from http://asmedigitalcollection.asme.org/BIOMED/proceedings-pdf/DMD2018/40789/V001T01A011/2787835/v001t01a011-dmd2018-6905.pdf by guest on 23 September 2020


heart sounds of 64 children (33 Still’s murmurs and 31 By equipping PCPs with automated murmur
pathological murmurs) recorded utilizing StethAid classification at the point of care, StethAid offers the
hardware. For this analysis we observed a true positive possibility of avoiding unnecessary referrals of children
identification rate of 85% with a false positive rate of with innocent heart murmurs to pediatric cardiologists or
3.2%. These preliminary results indicate that the for echocardiography. This would save costs to the
diagnostic accuracy of our algorithm is similar whether healthcare system and avoid needless anxiety and
the recordings are made using the StethAid hardware or inconvenience to affected families.
the Littmann stethoscope.
Acknowledgments
INTERPRETATION This work has been supported by National Institutes of
Health grant 1R41HL131081 and a grant (RE159-G5)
We compared our StethAid digital stethoscope with a from the Atlantic Pediatric Device Consortium (funded
commercial electronic stethoscope. We analyzed through Food and Drug Administration grant
frequency response between the two stethoscopes and 5P50FD004193).
found that the StethAid stethoscope’s frequency response
was comparable to that of a commercial electronic REFERENCES
stethoscope, and that our hardware reproduced heart 1. Fogel, D. H. (1960). The innocent systolic murmur in
sounds with the same fidelity as that of the Littman children: a clinical study of its incidence and
hardware. characteristics. American heart journal, 59(6), 844-
The bell-shaped frequency response seen in Figure 3 855.
was as expected because diaphragms of stethoscopes
are tuned for recording frequencies present in normal 2. Vukanovic-Criley, J. M., Criley, S., Warde, C. M.,
heart sounds and pathologies while attenuating other Boker, J. R., Guevara-Matheus, L., Churchill, W. H.,
frequencies. The frequency response seen with normal ... & Criley, J. M. (2006). Competency in cardiac
heart sounds was also shown to follow the same shape examination skills in medical students, trainees,
and follows previously characterized examples [4]. physicians and faculty. Arch Intern Med, 166(6), 610-
Although we obtained respectable true positive and 6.
false positive identification rates, these numbers need
refinement per the clinical needs. From a clinical 3. Kang, S., Doroshow, R., McConnaughey, J., &
standpoint, false positives, which represent pathological Shekhar, R. (2017). Automated Identification of
murmurs incorrectly labeled as Still’s murmurs, pose a Innocent Still's Murmur in Children. IEEE
risk. However, if the current false positive identification Transactions on Biomedical Engineering, 64(6),
rate is deemed high, adjusting the threshold of the 1326-1334.
classifier can reduce this rate. This would minimize the
possibility of not referring pathological murmur cases, but 4. Debbal, S. M., & Bereksi-Reguig, F. (2008).
would come at the expense of a slightly reduced true Computerized heart sounds analysis. Computers in
positive identification rate and a smaller savings in the biology and medicine, 38(2), 263-280.
number of Still’s murmur referrals. Given the current high
rate of over-referrals, the potential for decreasing them
would still exist.
The Still’s murmur identification algorithm is cloud-
based, which provides a convenient paradigm to further

3 Copyright © 2018 ASME

Potrebbero piacerti anche