AE-4020
SERVICE MANUAL
precision instrument, if one part is deteriorated, satisfactory result may not be obtained. Moreover, because it is
an analyzer using high technology, any operational trouble also affects measurement result significantly. In case
a trouble is occurred due to such a reason, refer to this service manual and operating manual for an appropriate
remedy.
CAUTION
This service manual is made for specially trained service representative only. This may be used as a
reference material for other persons, however, KDK Corporation bears no liability for any accidents or no
warranty of the product if undesignated person performs the service described on this manual.
Contents
Introduction
2. Composition
1. Diagram of assembly 2-2
2. Drawings 2-3
3. Service Mode
1. Service Mode Specifications 3-1
5. Standard Inspection
1. Test Items 5-2
2. Test Methods and Standard 5-3
6. Adjustment Method
1. Adjusting Items and parts 6-1
Appendix
1. External Output Specifications
2. Printing Formats Specifications
3. Error & Trouble Specifications
4. Work Standard
5. Ethernet settings
6.Maintenance parts
AE-4020 SERVICE MANUAL
1-1
Chapter 1 Outline of the Analyzer AE-4020 SERVICE MANUAL
1) Characteristics
Auto start,and non-directional test strip placement
Compact:210x328x164
Lightweight:3.6kg
Simple structure& Easy maintenance
Auto absorbing surplus sample urine
Color tone determination
Temperature correction
Abnormal color detection
No dipping detection
Test strip type auto-detection(assuming they have auto-classification marks)
2) Specifications
Name : Semi-automatic urine analyzer, the AUTION ELEVEN AE-4020
Apparatus make-up : Instrument , AC adapter (AC100V∼AC240V)
Sample : Urine
Test Strips : AUTION Sticks
AUTION Screen (micro albumin/creatinine)
Meas. Item : GLU, KET, BIL, NIT, PRO, URO, pH, BLD, LEU, Specific Gravity,
Color-tone, New items (ALB, CRE, A/C, P/C)
Meas. Method : Test strip; Single (BLD only ) or Dual Wavelength Reflectance Method
Meas. Wavelengths : 4-wavelength LED (450, 565, 635, 760 nm)
Dipping : The user dips test strips into samples.
Processing speed : 514 samples per hour (maximum processing mode: 7-second interval)
Test strip reaction time : 60 seconds
Warming up Time : 60 seconds
Caliblation : Calibrated automatically by a blank pad
Measurement types :Normal measurement, STAT measurement, Control measurement, Check
measurement
Display : Custom LC display (icons are used)
Operation part : (START,STOP,FEED,MENU,ENTER,0-9,−,←,→,ID)
Printer : Thermal printer (paper width: 58mm)
Data memory amount : 520 samples
Input Analytical curve : entering information on analytical curve through PC
Method of removing surplus urine
: Capilarity
Precision control : Measurement of The Aution Check (Control urine)
Auxiliary function : User mode (DATE, ID numbers, MEMORY , Printing of parameters, Test
strip type, Measurement result format, Printer settings, External output
ON/OFF)
Service mode (Test operation of system, Analytical curve setting, System
parameter setting, Initialize EEPROM)
Other function : Color tone determination, Temperature correction, Abnormal color
detection, No dipping detection )
Measurement environment : Temperature:10 to 30°C, Humidity: 30 to 60%R.H. (non-condensing)
*Temperature correction function used
Dimensions : 210 (width) x 328 (depth) x 164 (height) mm
Weight : Approx. 3.6 kg
Supply voltage : 100 to 240 VAC, 50/60 Hz (worldwide support, using plug adapters)
Pharmaceutical Affairs law : Conformity
Applicable laws, : Domestic EMC and CE marking (IVD Directive) Conformity
regulations and (UL, FCC, FDA) Compliance
standards
3) Principles of measurement
1-2
Chapter 1 Outline of the Analyzer AE-4020 SERVICE MANUAL
In the Optical Block, two different ranges of wavelength from the LED are irradiated on the reacted
parts of the Test Strip and the color-tone-compensated section. At that time, the combination of the
wavelengths is different for each measurement item, as shown in the table below.
Light irradiated to the color-tone-compensation section is adjusted for interfering factors such as
amount of reflected light and color of sample.
From the amount of reflected light measured in the Optical Block, the reflectance is calculated
using the following formula:
R = Tm ・ Cs / Ts ・ Cm
R: Reflectance
Tm: Reflective light amount with the meas. wavelength
Ts: Reflective light amount with the ref. wavelength
Cm:Reflective light amount at the color-tone-compensated section with the meas. wavelength
Cs: Reflective light amount at the color-tone-compensated section with the ref. wavelength
1-3
Chapter 1 Outline of the Analyzer AE-4020 SERVICE MANUAL
Measurement of Color-tone
● Color-tone
Y: Reflectance of 430nm
M: Reflectance of 565nm
C: Reflectance of 635nm
r: Reflectance of 760nm
1-4
AE-4020 SERVICE MANUAL
Chapter 2 Composition
1. Diagram of assembly
1) Diagram of assembly············································································································· 2-2
2. Drawings
1) Main Unit································································································································ 2-3
2)Assembly of Maintenance cover ····························································································· 2-5
3)Assembly of Under cover ········································································································ 2-7
4)Assembly of Chassis··············································································································· 2-8
5) Assembly of Optical Unit ······································································································ 2-10
6) Assembly of Table-base······································································································ 2-11
7) Assembly of drive part for Introduction················································································· 2-12
8) Assembly of drive part for conveyance ················································································ 2-13
2-1
Chapter 2 Composition AE-4020 SERVICE MANUAL
1. Diagram of Assembly
22-03211
Chassis 22-02929
Assembly of reinforcement metallic base 23-02926
Assembly of table 23-02943
Assembly of lever for introduction 24-02944
Assembly of drive part for introduction 23-02930
Assembly of motor for introduction 24-02931
Assembly of DC-motor 24-02988
PI-Assembly of drive part for introduction 24-02916
Assembly of table base 23-02932
Assembly(L) of mechanism for alignment 24-02933
Assembly(R) of mechanism for alignment 24-02934
Assembly of lever for conveyance 24-02935
Assembly of micro-switch 24-02920
Assembly of drive part for conveyance 24-02987
Assembly of lever-frame 24-02936
Assembly of pulley(F) 24-02945
Assembly of pulley(R) 24-02946
Assembly of motor for conveyance 24-02937
Assembly of DC-motor 24-02921
PI-Assembly of drive part for conveyance 24-02917
Assembly of table(R) for conveyance 24-02938
Assembly(adhesion) of table(R) for conveyance 24-03022
Assembly of photo-sensor 24-02919
Assembly of table(C) for conveyance 24-02939
Assembly of motor for optical part 24-02940
Assembly of optical unit 23-02941
Assembly of light measurement unit 24-02942
Assembly of light measurement 24-03013
Head-amp print circuit board 01-02620
PI-Assembly of optical part 24-02918
Assembly of cover for maintenance 23-02928
Main print circuit board 01-02574
Assembly of thermistor 24-03050
Assembly of under cover 23-02924
Driver print circuit board 01-02602
Assembly of external output 24-02925
Assembly of front cover 24-02927
2-2
Chapter 2 Composition AE-4020 SERVICE MANUAL
2. Drawings
1) Main Unit
5 6
2-3
AE-4020 SERVICE MANUAL
Note1:When removing the assembly of under cover, with great care not to damage connecting-wire.
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Assembly of lever for conveyance Lever for conveyance 96433 1
2 1 Assembly of table(C) for conveyance Tray for conveyance 96434 1
3 1 Assembly of table Tray 96430 1
4 1 Assembly of lever for introduction Arm 96432 1
5 1 Assembly of front cover - -
6 1 VR cover - -
7 1 Assembly of under cover - -
8 1 Waste Box 96431 1
9 1 Assembly of cover for maintenance - -
2-4
AE-4020 SERVICE MANUAL
1 2 3 4 5
Removal:
(M3:Screw)
6 7 8 9 10
11
Fig:Removing cover(SUS)
2-5
AE-4020 SERVICE MANUAL
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Key switch Set(①+②) 96469 1
2 1 Plate - -
3 1 Paper cutter 96480 1
4 1 Printer 96470 1
5 1 Paper cover 96478 1
6 2 Spring - -
7 1 Button(R) - -
8 1 Button(L) - -
9 1 Cover(SUS) - -
10 1 Main print circuit board 96451 1
11 1 Assembly of thermistor - -
2-6
AE-4020 SERVICE MANUAL
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Driver print circuit board 96452 1
2 1 Assembly of external output − −
2-7
Chapter 2 Composition AE-4020 SERVICE MANUAL
4) Assembly of Chassis
3
2 4
13
5
12
11
8
10
14
2-8
AE-4020 SERVICE MANUAL
2-9
AE-4020 SERVICE MANUAL
4
1 2 3
5 6
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Nut 96481 1
2 1 Drive screw(M6P1.5) - -
3 1 PI-Assembly of optical unit 96455 1
4 1 Belt 174-2GT-3 96477 1
5 1 Assembly of light measurement 96468 1
6 1 CU-joiner FCUJ(1.0)-10F-650-10S5 96472 1
2-10
AE-4020 SERVICE MANUAL
3 4
5
2
1 7
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Plate - -
2 1 Photo-IC TPS816 - -
3 1 AUTO-START print circuit board - -
4 1 LED LSF880N2 - -
5 1 LED block - -
6 1 Optical-block(A) - -
7 1 Black-seal - -
2-11
Chapter 2 Composition AE-4020 SERVICE MANUAL
2 1
Conclusion(×3)
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Assembly of DC-motor (For introduction) 96459 1
2 2 Bush(#80) 80B-0503 − −
3 1 Pinion − −
2-12
Chapter 2 Composition AE-4020 SERVICE MANUAL
2 3 4
1
6
x3
Note:When this unit is decomposed, adjustment with a Exclusive use jig is needed.
Item No. Qty. Parts Model No. Specification Item Code Cont.
1 1 Assembly of DC-motor (For conveyance) 96458 1
2 1 Coupling GJ5(4-3) 96475 1
3 1 Polyurethane circle belt MBN3-180 96476 1
4 1 Assembly of lever frame - -
5 1 PI-Assembly of drive part for 96454 1
conveyance
6 3 Bush(#80) 80F-0406 96474 1
2-13
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
1 General ............................................................................................................................................... 3
2 Analytical curve................................................................................................................................. 4
2-1 AE-4020 (Operation display [W AE4020 mg], [M AE4020 mg]) ......................................................... 5
2-2 AE-4020 (Operation display [W AE4020 SI], [M AE4020 SI]) ............................................................ 6
2-3 AE-4020 (Operation display [W AJ4270 mg], [M AJ4270 mg]) .......................................................... 7
2-4 AE-4020 (Operation display [W AJ4270 SI], [M AJ4270 SI]) ............................................................. 8
2-5 Change in analytical curve ................................................................................................................. 9
3-1
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-32 Item No. 400: Test operation of system (single unit operation)........................................................ 39
5-33 Item No. 410: Test operation of system (total, continuous).............................................................. 40
5-34 Item No. 420: Adjust system (photometric section light intensity).................................................... 41
5-35 Item No. 421: Check levelness ........................................................................................................ 42
5-36 Item No. 430: Adjust waiting position of introduction lever............................................................... 43
5-37 Item No. 440: Adjust surplus urine suction position of introduction lever......................................... 44
5-38 Item No. 450: Adjust transport lever waiting position ....................................................................... 45
5-39 Item No. 460: Adjust photometric section left side waiting position ................................................. 46
5-40 Item No. 470: Adjust black mark detection pulse (mark at right) ..................................................... 47
5-41 Item No. 471: Adjust black mark detection pulse (mark at left)........................................................ 48
5-42 Item No. 480: Adjust color tone (calculate C/M, Y/M, C/r correction value) ..................................... 49
5-43 Item No. 490: Set 430 nm wavelength correction (not used)........................................................... 50
5-44 Item No. 491: Set 500 nm wavelength correction (not used)........................................................... 50
5-45 Item No. 492: Set 565 nm wavelength correction (not used)........................................................... 51
5-46 Item No. 493: Set 635 nm wavelength correction (not used)........................................................... 51
5-47 Item No. 494: Set 760 nm wavelength correction (not used)........................................................... 52
5-48 Item No. 500: Display incoming strip detection level ....................................................................... 52
5-49 Item No. 510: Display temperature .................................................................................................. 52
5-50 Item No. 520: Input introduction lever waiting position (time: ms) ................................................... 53
5-51 Item No. 521: Input surplus urine suction position of introduction lever (time: ms) ......................... 53
5-52 Item No. 522: Input transport lever waiting position (time: ms)........................................................ 54
5-53 Item No. 523: Input photometric section left waiting position pulse ................................................. 54
5-54 Item No. 524: Set drive pulse for right drive photometric section 1 (mark at right).......................... 55
5-55 Item No. 525: Set drive pulse for left drive photometric section 1 (mark at right) ............................ 55
5-56 Item No. 526: Set drive pulse for right drive photometric section 2 (mark at left) ............................ 56
5-57 Item No. 527: Set drive pulse for left drive photometric section 2 (mark at left) .............................. 56
5-58 Item No. 528: Set color tone (C/M correction) ................................................................................. 57
5-59 Item No. 529: Set color tone (Y/M correction).................................................................................. 57
5-60 Item No. 530: Set color tone (C/r correction) ................................................................................... 58
5-61 Item No. 531: Set DIP detect threshold............................................................................................ 58
5-62 Item No. 550: Change analytical curve information (threshold, abnormal mark rank)..................... 60
5-63 Item No. 560: Set P/C ratio, DILUTE reference, PRO reference..................................................... 62
5-64 Item No. 561: Set P/C ratio, DILUTE reference, CRE reference ..................................................... 62
5-65 Item No. 600: Aution Net (entering information on analytical curve through PC) ............................ 64
5-66 Item No. 610: Set test strip to be selected ....................................................................................... 65
5-67 Item No. 700: Set device serial number ........................................................................................... 67
5-68 Item No. 701: Settings specific to place of destination .................................................................... 67
5-69 Item No. 710: Set upper limit of reflectance ..................................................................................... 69
5-70 Item No. 711: Print sample error mark ............................................................................................. 69
5-71 Item No. 712: Set abnormal color detection..................................................................................... 70
5-72 Item No. 900: Clear memory ............................................................................................................ 71
5-73 Item No. 990: Initialize EEPROM ..................................................................................................... 71
6 Item No. 142: The operation method at the time of AJ system selection................................... 73
6-1 Select the measurement mode.............................................................................................................. 73
6-2 Enter a Patient ID No............................................................................................................................. 73
6-3 After entering all the Patient ID Nos. ..................................................................................................... 73
3-2
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
1 General
Service mode, the following functions can be used.
As device settings, 3-digit item number and corresponding numeric figures are entered.
An item is given a number within No. 100 group or No. 200 group, depending on its function.
Configuration of screen and key assignment are sama as for normal mode. The following are functional
descriptions.
3-3
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
2 Analytical curve
The AE-4020 urine analyzer is used not only in Japan but in many countries. It must be capable of processing various analytical curves by offering simple and easier changeover
feature. These functions are all programmed and contained in ROM as default settings. Default analytical curve is 1. [W AE4020 mg] present unit (mg/dL). Selection of analytical
curve is made in Service mode.
Screen display Sales area Rank display Value Applicable test strips Remarks
1. [W AE4020 mg] Domestic (Japan), AE-4020 Present unit 10EA, 9EA, 8EA, 7EA, 6EA, 5EA, 4EA, 10PA,
(mg/dL) 2AC, 9UB, 8UA, 8UB, 7UA, 7UB, 5UA, 5UB,
4UA, 5EB, 6EB, 9EB
2. [W AE4020 SI] Domestic (Japan), AE-4020 SI (mmol/L) 10EA, 9EA, 8EA, 7EA, 6EA, 5EA, 4EA, 10PA,
2AC, 9UB, 8UA, 8UB, 7UA, 7UB, 5UA, 5UB,
4UA, 5EB, 6EB, 9EB
3. [W AJ4270 mg] Domestic (Japan), AJ-4270 Present unit 10EA, 9EA, 8EA, 7EA, 6EA, 5EA, 4EA, 10PA, Negative notation is output in the old
(mg/dL) 2AC, 9UB, 8UA, 8UB, 7UA, 7UB, 5UA, 5UB, format.
4UA, 5EB, 6EB, 9EB GLU: NORMAL, LEU: NEG.
4. [W AJ4270 SI] Domestic (Japan), AJ-4270 SI (mmol/L) 10EA, 9EA, 8EA, 7EA, 6EA, 5EA, 4EA, 10PA, Negative notation is output in the old
2AC, 9UB, 8UA, 8UB, 7UA, 7UB, 5UA, 5UB, format.
4UA, 5EB, 6EB, 9EB GLU: NORMAL, LEU: NEG.
5. [M AE4020 mg] Overseas in general AE-4020 Present unit 10EA,10PA,2AC Only for application examination paper,
(mg/dL) the difference from 1 is.
6. [M AE4020 SI] Overseas in general AE-4020 SI (mmol/L) 10EA,10PA,2AC Only for application examination paper,
the difference from 2 is.
7. [M AJ4270 mg] Overseas in general AJ-4270 Present unit 10EA,10PA,2AC Only for application examination paper,
(mg/dL) the difference from 3 is.
8. [M AJ4270 SI] Overseas in general AJ-4270 SI (mmol/L) 10EA,10PA,2AC Only for application examination paper,
the difference from 4 is.
3-4
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
3-5
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
3-6
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
* Negative notation of qualitative value is different from "W AE4020 mg" in the following items. (old format)
GLU: NORMAL, LEU: NEG.
3-7
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
* Negative notation of qualitative value is different from "W AE4020 SI" in the following items. (old format)
GLU: NORMAL, LEU: NEG.
3-8
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Item name, semiquantitative symbol and concentration are reflected upon printing and external output. The unit
can use the following external output formats: AE-4020, AX-4280 (SA-4250 compatible), ticket printer
(AJ-4270), AM-4290 and MA-4240.
Note that AJ-4270 is not included in format selection options because it may be considered equivalent of AX-4280
format when used with S.G. external output position setting.
As an independent device, the analyzer accepts manual change of rank threshold and abnormal mark ranks except
for relative density (threshold in ROM for China cannot be changed). However, printing order of items,
transmission order of items, semiquantitative symbol of item, concentration, and detailed information on
color tone cannot be changed by the analyzer itself.
When it is necessary to change these items, connect external output terminal of the analyzer to RS-232C terminal
of the PC through a cross cable and use "Analytical curve input program" on the PC. (This feature cannot be
used when ROM for China is installed.)
3-9
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) The screen as shown below will pop up. To enter through the service mode, type "-4020-".
The number of settable items is refer to "Functions and item No."
Upon pressing ENTER key, the password is compared with stored one. As both match, the screen changes
to step (3) Self-diagnosis screen. If a wrong password, it is cleared, requiring the correct password input.
(3) The self-diagnosis screen will appear. Wait for approx. 10 seconds.
AE-4020 01.00
(4) The item No. input screen (reverse mode) will appear.
No. 000
If the following screen appears in place of the screen described in (4) above, it indicates a warning.
EEPROM failure
T001 Ent
This massage implies that the problem is associated with main ROM, for example, version up
related problem. Press ENTER key and the item No. input screen will appear. Perform "Initialize
EEPROM". Never use functions other than "Initialize EEPROM" in this case.
Should "T001 Ent" appear while the ROM version is kept unchanged, the board may be defective.
Power failure
E001 Ent
This warning will appear when power failure occurs during measurement.
3-10
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Version change
T102 Ent
This warning will appear when version of main ROM is upgraded.
Pressing ENTER clears all backup data.
E002 Ent
This warning will appear when the voltage of battery used to retain data, e.g. measurements, drops
or the data is lost.
When the voltage is low, leave the power supply on to charge the battery.
Pressing ENTER clears all backup data.
As described before, if a wrong password is typed, the screen clears it and waits for correct password
without entering service mode.
To prevent from changing to service mode while typing the password, press STOP key. The standby
screen will pop up after completion of system initialization.
If reverse mode menu screen will not appear after typing "-4020-", press a numeric key and then
type "-4020-" again.
(2) When "-4020-"is accepted, service mode item No. input screen appears.
No. 000
No. 000
3-11
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) System initialization process starts in the same way as in the case of power up.
12:34
CAUTION: During service mode entering process, measurement start No. and measurement type will be
changed upon typing the password. Note the changes when returning from the service mode
during measurement.
3-12
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
4 Screen flowchart
Power ON
Full screen
System initialization
20
Standby screen
MEAS
No. 001 10EA -4020- to service mode
Print parameter
Set No. of surplus urine Select with/without temperature Select with/without specific gravity
Set No. of test strip dispositions Set date printing Select temperature output
suctions correction correction
No. 130 100 No. 131 200 No. 132 0 No. 133 1 No. 134 1 No. 135 1
No. 136 0 No. 137 1 No. 138 1 No. 139 1 No. 140 0 No. 141 0
No. 300 0 No. 301 0 No. 302 5 No. 303 005 No. 304 0 EEP 0-1
EEP 0-1
Set device ID
No. 320 0
Set IP address
No. 321-1 192 No. 321-2 168 No. 321-3 225 No. 321-4 001
No. 322-1 255 No. 322-2 255 No. 322-3 255 No. 322-4 255
No. 323-1 192 No. 323-2 168 No. 323-3 225 No. 323-4 001
3-13
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Test operation of system
(total, continuous)
Adjust system
(photometric section light intensity)
No. 420 -430- 2500 -565- 2500 -635- 2500 -760- 2500 -000- 150
Adjust levelness
No. 470 No. 470 ------ No. 470 STA No. 470 1748 707 EEP 0-1
No. 471 No. 471 ------ No. 471 STA No. 471 698 1748 EEP 0-1
No. 480 1 No. 480 ------ No. 480 STA No. 470 0 1 EEP 0-1
No. 500 68
Display temperature
Enter introduction surplus urine suction Enter photometric section left waiting Set photometric section Set photometric section
Enter introduction waiting position Input transport lever waiting position
position position right drive pulse 1 right drive pulse 2
No. 520 0 No. 521 0 No. 522 10 No. 523 67 No. 524 1654 No. 525 713
Set photometric section left drive pulse 1 Set photometric section left drive pulse 2 Set color tone (C/M correction) Set color tone (Y/M correction) Set color tone (C/r correction) Set DIP detect threshold
No. 526 704 No. 527 1663 No. 528 1000 No. 529 1000 No. 530 1000 No. 531 923
No. 550 01 No. 550 01 No. 01 80.0 No. 02 30 72.0 EEP 0-1
Set P/C ratio, DILUTE reference, PRO Set P/C ratio, DILUTE reference, CRE Check EEP writing
No. 610 No. 610 10EA 1 No. 610 9EA 1 EEP 0-1
Set upper limit of reflectance Print sample error mark Set abnormal color detection Check EEP writing
Clear memory
Initialize EEPROM
3-14
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Functions and item No. (550-610 are not available on Chinese version)
Default Mode
Function Item No. Unit Input range
value Service Mode
Print parameter 100 1 - 0 to 6
Print parameter No. list 110 - - -
Measurement related process settings
Select analytical curve 120 1 - 1 to
Set No. of test strip dispositions times 0,100 to 200
Set No. of surplus urine suctions 131 200 times 0,200 to 400
Set date printing 132 1 - 0 to 2
Select with/without temperature 133 1 OFF/ON 0 or 1
correction
Select temperature output 134 1 OFF/ON 0 or 1
Select with/without specific gravity 135 1 OFF/ON 0 or 1
correction
Select check measurement print 136 0 - 0 to 3
Select measurement auto stop 137 1 OFF/ON 0 or 1
Eliminate effect of creatinine on albumin 138 1 OFF/ON 0 or 1
With/without DIP detection 139 1 OFF/ON 0 or 1
Set qualitative value indication format 140 0 - 0 to 2
Select JCCLS print 141 0 OFF/ON 0 or 1
Select ID# setting 142 0 - 0 or 1
Unused ID# 143 0 - 0 or 1
External output settings
Set communication port 300 0 - 0 to 1
Set communication format 301 0 - 0 to 4
Set communication speed 302 5 - 0 to 6
Set time gap between communication 303 005 0.1 sec 001 to 999
blocks
Set S.G. external output destination 304 0 - 0 to 2
Set communication timeout 1 310 15 sec 01 to 99
Set communication timeout 2 311 30 sec 01 to 99
Set communication timeout 3 312 20 sec 01 to 99
Set No. of retransmissions 313 06 times 01 to 99
Set communication standby time 1 314 10 sec 01 to 99
Set communication standby time 2 315 01 sec 01 to 99
Ethernet settings
Set device ID 320 000 - 000 to 255
Set IP address 321 000 - 000 to 255
Set sub-net masking 322 000 - 000 to 255
Set default gateway 323 000 - 000 to 255
Set DHCP 324 0 OFF/ON 0 or 1
System test and adjustment
Test operation of system (single unit 400 - - 0 to 8
operation)
Test operation of system (total, 410 - - 0 to 1
continuous)
Adjust system (photometric section light 420 - - 0, 1, 4, 7, 3, 6, 9,
intensity) START
Check levelness 421 - - -
Adjust waiting position of introduction 430 0 msec 0, 1, 4, 7, 3, 6, 9,
lever ENTER, START
Adjust surplus urine suction position of 440 0 msec 0, 1, 4, 7, 3, 6, 9,
introduction lever ENTER, START
3-15
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Default Mode
Function Item No. Unit Input range
value Service Mode
Adjust transport lever waiting position 450 50 msec 1, 3, ENTER, START
3-16
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5 Operation of function
5-1 Item No. 100: Print parameter
This function is related to parameter setting and enables printing of various settings including service settings,
system settings, user settings, test strip information, test strip item information and color tone information.
These printing items can be individually or totally specified.
(1) In the item No. input screen, press keys "1", "0" and "0" in this order.
(3) From keys "0" through "6", select the No. of table to be printed.
For test strip information, 01 to 32 can be input (32: print all tables).
For test strip item information, 00 to 17 can be input (17: print all tables).
(4) Printing
No. 100
During printing, a print mark (printer
Print mark
mark) is lit.
At the end of printing, the item No. input screen will appear.
3-17
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Print confirmation screen will appear. To start printing, press "1" key.
(3) Printing
No. 110
During printing, a print mark (printer
Print mark
mark) is lit.
At the end of printing, the item No. input screen will appear.
3-18
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using keys "0" through "9", select the analytical curve information.
(3) Overwrite confirmation screen will appear. Press "1" key when overwriting.
(4) Overwrite initialization reconfirmation screen will appear. Press "2" key when initializing and
overwriting.
3-19
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) From numeric keys 0-9, type the number of test strip dispositions.
Upon pressing ENTER key, the unit acknowledges the input and goes to the input screen (2) of the item
No. 131: Set No. of surplus urine suctions.
Upon pressing STOP key, the unit ignores the input value and returns back to the item No. input screen.
(2) From numeric keys 0-9, type the number of surplus urine suctions.
Upon pressing ENTER key, the screen changes to the input screen (2) of the item No. 132.
If STOP key is pressed instead, the unit ignores the input value and returns back to the item No. input
screen. See Note 1.)
(No figure is displayed if 000 is set. If a large number of test strips are measured, measurement data
will be affected.)
3-20
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Use "1" key when using correcting function. Otherwise, use "0".
0: "OFF (disable)"
1: "ON (enable)"
Upon pressing ENTER key, the unit accepts the input and goes to the item No. 134 Input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-21
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Use "1" key when outputting temperature. Otherwise, press "0" key.
(2) Press "1" key when using this feature. Otherwise, press "0" key.
0: "OFF (disable)"
1: "ON (enable)"
Upon pressing ENTER key, the unit accepts the input and goes to the item No. 136 Input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-22
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Use keys "0", "1", or "2" to select ON/OFF of abnormal mark output.
(2) Press "1" key when using this feature. Otherwise, press "0" key.
0: "OFF (disable)"
1: "ON (enable)"
Upon pressing ENTER key, the unit accepts the input and goes to the item No. 138 Input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-23
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Press "1" key when using this feature. Otherwise, press "0" key.
0: "OFF (disable)"
1: "ON (enable)"
Upon pressing ENTER key, the unit accepts the input and goes to the item No. 139 Input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
(2) Press "1" key when using this feature. Otherwise, press "0" key.
0: "OFF (disable)"
1: "ON (enable)"
Upon pressing ENTER key, the unit accepts the input and goes to the item No. 140 Input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-24
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Select desired qualitative value output format by using "0", "1" or "2" key.
0: "4+"
1: "+4"
2: "++++"
Upon pressing ENTER key, the unit accepts the input and returns back to the item No. input screen.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
*Note that if this output format change is ignored, the qualitative value may have been changed by
Aution net AE. To enable the new setting in Item No. 140, first change the format to "4+" through
Aution net AE and then repeat the above steps.
3-25
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
0: Use
1: Delete
3-26
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-27
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
0: "RS-232C"
1: "Ethernet"
Upon pressing ENTER key, the unit accepts the input and goes to the communication format setting input
screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
(2) Using one of keys "0" through "5", select desired external output format.
When outputting in AJ-4270 format, select AX-4280 and then select [output to item information area] in
Item No. 304, S.G. external output destination.
3-28
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using one of keys "0" through "6", select the desired baud rate.
5-21 Item No. 303: Set time gap between communication blocks
Set the time gap between external output data blocks in units of 0.1 sec, in the range of 001 to 999. Factory setting
is 005.
(1) In the item No. input screen, press keys "3", "0" and "3" in this order.
(2) Using keys "0" through "9", type the desired time gap.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set external
output destination of S.G. option.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-29
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using key "0", "1" or "2", select the destination of S.G. external output.
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-30
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-23 Item No. 310: Set communication timeout 1 (required in two-way communication)
External output used in RS-232C two-way communication system. Set the ACK waiting time following
transmission of ENQ, in units of second, in the range of 01 to 99. Factory setting is 15.
(1) In the item No. input screen, press keys "3", "1" and "0" in this order.
(2) Using keys "0" through "9", set ACK waiting time.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set
communication timeout 2.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
5-24 Item No. 313: Set No. of retransmissions (required in two-way communication)
External output used in RS-232C two-way communication system. Set the No. of retransmissions upon receiving
NAK from external HOST, in the range of 01 to 99. Factory setting is 06.
(1) In the item No. input screen, press keys "3", "1" and "3" in this order.
(2) Using keys "0" through "9", set No. of retries upon receiving NAK.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set
communication standby time 1.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
3-31
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-25 Item No. 314: Set communication standby time 1 (required in two-way communication)
External output used in RS-232C two-way communication system. Set the standby time to be maintained upon
receiving NAK in response to transmitted ENQ, in units of second, in the range of 01 to 99. Factory setting is 10.
(1) In the item No. input screen, press keys "3", "1" and "4" in this order.
(2) Using keys "0" through "9", enter communication standby time 1.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set
communication standby time 2.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-32
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-26 Item No. 315: Set communication standby time 2 (required in two-way communication)
External output used in RS-232C two-way communication system. Set the standby time to be maintained upon
receiving ENQ in response to transmitted ENQ, or upon receiving NAK in response to transmitted frame, in units
of second, in the range of 01 to 99. Factory setting is 01.
(1) In the item No. input screen, press keys "3", "1" and "5" in this order.
(2) Using keys "0" through "9", enter communication standby time 2.
Upon pressing ENTER key, the unit accepts the input and returns back to the item No. input screen.
Note 1)
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-33
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using keys "0" through "9", enter the device ID.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set IP address.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
3-34
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set sub-net
masking.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-35
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-29 Item No. 322: Set sub-net masking (required in Ethernet communication)
Set sub-net masking used over Ethernet. A sub-net masking is composed of 4 sets of integer (0-255).
e.g. 255. 255. 255. 0
As a rule, input a whole number 4 times.
(1) In the item No. input screen, press keys "3", "2" and "2" in this order.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set default
gateway.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-36
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-30 Item No. 323: Set default gateway (required in Ethernet communication)
Set default gateway used over Ethernet. A default gateway is composed of 4 sets of integer (0-255).
e.g. 192. 168. 100. 300
As a rule, input a whole number 4 times.
(1) In the item No. input screen, press keys "3", "2" and "3" in this order.
(2) Using keys "0" through "9", type address 1 of default gateway.
(3) Using keys "0" through "9", type address 2 of default gateway.
(4) Using keys "0" through "9", type address 3 of default gateway.
(5) Using keys "0" through "9", type address 4 of default gateway.
Upon pressing ENTER key, the unit accepts the input and goes to the input screen (2) of Set DHCP.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-37
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(1) In the item No. input screen, press keys "3", "2" and "4" in this order.
Note 1) After completing the setting sequence and before returning to the item No. input screen, the
EEPROM write confirmation screen will appear if updating is required. To save the change on
EEPROM and make it valid on the next power-up, press "1" key. If the current modification is for
temporary use and will not be used again, press "0" key (not save to EEPROM). Upon pressing
STOP key, all current modifications will be undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-38
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-32 Item No. 400: Test operation of system (single unit operation)
Test individual units independently.
(This test can be run once the target units are connected. Since these target units can freely run this test
program, they may hit each other, causing damage.)
(1) In the item No. input screen, press keys "4", "0" and "0" in this order.
3-39
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(1) In the item No. input screen, press keys "4", "1" and "0" in this order.
(3) After entering operation mode, the unit starts continuous operation of system and displays the number of
operations.
3-40
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-34 Item No. 420: Adjust system (photometric section light intensity)
Use this function to adjust light intensity of photometric section. The unit continuously samples the photometric
section on white board and displays A/D counts of every wavelengths and A/D counts of dark in sequence. Pulse
drive the photometric section with key.
(Since the photometric section can operate freely, it may hit associated parts, causing damage to them.)
(1) In the item No. input screen, press keys "4", "2" and "0" in this order.
Press ENTER key to accept the input
No. 420 and move photometric section to the
white board.
(2) A/D counts are displayed. These counts can be toggled in the order shown below by pressing "-" key.
If the photometric section fails to operate, the following trouble screen will appear.
T131
Description of the trouble code.
T131: Photometric section is malfunctioning.
3-41
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) After initialization process, install the levelness checking tool on the photometric section and press
START key.
No. 421
– – – – – – – – – – –
Center point
Photometric section
[1] [6] [11]
25 mm Checking tool
70 mm
115 mm
Photometric table
3-42
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) The introduction lever moves to the waiting position and the current control value is displayed.
Adjust the control value (0-100) with keys "1" through "9" and move the lever to controlled position with
START key.
3-43
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-37 Item No. 440: Adjust surplus urine suction position of introduction lever
This option is to adjust surplus urine suction position of the introduction lever. Input the control value with keys
"1" through "9". Determine appropriate surplus urine suction position by driving the introduction lever with
START key and "0" key, and then press ENTER key to accept the current value.
(CAUTION: This option allows the introduction lever to move without control, causing to hit surrounding
parts and damage them.)
(1) In the item No. input screen, press keys "4", "4" and "0" in this order.
(2) The introduction lever moves to the surplus urine suction position and the current control value is
displayed.
Adjust the control value (0-100) with keys "1" through "9" and move the lever to controlled position with
START key.
Enter the number of item to be operated.
3-44
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) The transport lever moves to the waiting position and the current control value is displayed.
Adjust the control value (0-100) with keys 1, 4, 7, 3, 6 and 9, and move the lever to controlled position
with START key.
Enter the number of item to be operated.
No. 450 0 10
[2] [1]
[1]: Time delay (ms) from detection of transport PI to stop.
[2]: Status of transport PI
Input keys and transport lever movement
7: Changes control value by 10 ms in normal direction (test strip transport direction).
4: Changes control value by 5 ms in normal direction (test strip transport direction).
1: Changes control value by 1 ms in normal direction (test strip transport direction).
9: Changes control value by 10 ms in reverse direction (opposite to test strip transport direction).
6: Changes control value by 5 ms in reverse direction (opposite to test strip transport direction).
3: Changes control value by 1 ms in reverse direction (opposite to test strip transport direction).
START key: Moves the lever to the current control position.
ENTER key: Accepts the current control value and returns to the item No. input screen.
STOP key: Ends the adjustment operation and the item No. input screen.
3-45
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-39 Item No. 460: Adjust photometric section left side waiting position
Adjust the left side waiting position of the photometric section. While pulse-driving the photometric section with
key, determine appropriate waiting position and press ENTER key to set the number of control pulses.
(Since the photometric section can operate freely, it may hit associated parts, causing damage to them.)
(1) In the item No. input screen, press keys "4", "6" and "0" in this order.
3-46
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-40 Item No. 470: Adjust black mark detection pulse (mark at right)
The unit measures Aution Sticks and displays number of control pulses of photometric section.
* When item No. 004: Test strip placement direction is "1 (mark at left)" or "2 (mark at right)", change
this setting to "0" and then start adjustment.
* Be sure to use test strip having 4 mm wide black mark or white check sample. Set MEAS mode test strip
to 4 mm wide mark (10 EA).
If these adjustments are not done, pulse settings will be indicated as "----".
(1) In the item No. input screen, press keys "4", "7" and "0" in this order.
(2) Set Aution Sticks on the introduction section and press START key.
No. 470
The unit goes to EEPROM writing screen upon pressing ENTER key.
The unit returns back to the item No. input screen upon pressing STOP key.
The unit starts photometry again upon pressing START key.
T137 Ent
The trouble code identifies the problem as shown below:
T130: photometric section fails during initialization
T131: photometric section fails to operate
T132: A/D value overflows (quantity of reflected light is abnormal)
T133: range over (quantity of reflected light is too large)
T134: range under (quantity of reflected light is too small)
T135: dark over (quantity of stray light is large)
T137: test strip without black mark
T122, T138: no test strip
For other codes, refer to "Error and trouble specification".
3-47
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-41 Item No. 471: Adjust black mark detection pulse (mark at left)
The unit measures Aution Sticks and displays number of control pulses of photometric section.
* When item No. 004: Test strip placement direction is "1 (mark at left)" or "2 (mark at right)", change
this setting to "0" and then start adjustment.
* Be sure to use test strip having 4 mm wide black mark or white check sample. Set MEAS mode test strip
to 4 mm wide mark (10 EA).
If these adjustments are not done, pulse settings will be indicated as "----".
(1) In the item No. input screen, press keys "4", "7" and "1" in this order.
(2) Set Aution Sticks on the introduction section and press START key.
No. 471
The unit goes to EEPROM writing screen upon pressing ENTER key.
The unit returns back to the item No. input screen upon pressing STOP key.
The unit starts photometry again upon pressing START key.
T137 Ent
The trouble code identifies the problem as shown below:
T130: photometric section fails during initialization
T131: photometric section fails to operate
T132: A/D value overflows (quantity of reflected light is abnormal)
T133: range over (quantity of reflected light is too large)
T134: range under (quantity of reflected light is too small)
T135: dark over (quantity of stray light is large)
T137: test strip without black mark
T122, T138: no test strip
For other codes, refer to "Error and trouble specification".
3-48
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-42 Item No. 480: Adjust color tone (calculate C/M, Y/M, C/r correction value)
Measure 5 Aution Sticks dipped in water. Based on the results, C/M, Y/M, C/r correction value (zero point
correction) will be calculated.
* Be sure to use test strip set in MEAS mode.
(1) In the item No. input screen, press keys "4", "8" and "0" in this order.
(2) Press START key and set Aution Sticks to the introduction section in timing with buzzer.
T137 Ent
The trouble code identifies the problem as shown below:
T130: photometric section fails during initialization
T131: photometric section fails to operate
T132: A/D value overflows (quantity of reflected light is abnormal)
T133: range over (quantity of reflected light is too large)
T134: range under (quantity of reflected light is too small)
T135: dark over (quantity of stray light is large)
T137: test strip without black mark
T138: no test strip
For other codes, refer to "Error and trouble specification".
(4) At the end of photometric process (5 strips), EEPROM write screen will appear.
If the modified settings will be used at the next power-up, press 1 key to save on EEPROM. If the current
modification is for a temporary use, press 0 key. The modification will be reflected on the printed
parameters but not be saved on EEPROM once power is shut down. Pressing STOP key cancels all
modifications.
EEP 0-1
3-49
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-43 Item No. 490: Set 430 nm wavelength correction (not used)
Set 430 nm wavelength correction value. Practical setting method is not developed.
(1) In the item No. input screen, press keys "4", "9" and "0" in this order.
(2) Using keys "0" through "9", enter 430 nm wavelength correction value.
Upon pressing ENTER key, the unit acknowledges the input value and goes to input screen (2) of 500 nm
wavelength correction.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
5-44 Item No. 491: Set 500 nm wavelength correction (not used)
Set 500 nm wavelength correction value. Practical setting method is not developed.
(1) In the item No. input screen, press keys "4", "9" and "1" in this order.
(2) Using keys "0" through "9", enter 500 nm wavelength correction value.
Upon pressing ENTER key, the unit acknowledges the input value and goes to input screen (2) of 565 nm
wavelength correction.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
3-50
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-45 Item No. 492: Set 565 nm wavelength correction (not used)
Set 565 nm wavelength correction value. Practical setting method is not developed.
(1) In the item No. input screen, press keys "4", "9" and "2" in this order.
(2) Using keys "0" through "9", enter 565 nm wavelength correction value.
Upon pressing ENTER key, the unit acknowledges the input value and goes to input screen (2) of 635 nm
wavelength correction.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
5-46 Item No. 493: Set 635 nm wavelength correction (not used)
Set 635 nm wavelength correction value. Practical setting method is not developed.
(1) In the item No. input screen, press keys "4", "9" and "3" in this order.
(2) Using keys "0" through "9", enter 635 nm wavelength correction value.
Upon pressing ENTER key, the unit acknowledges the input value and goes to input screen (2) of 760 nm
wavelength correction.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
3-51
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-47 Item No. 494: Set 760 nm wavelength correction (not used)
Set 760 nm wavelength correction value. Practical setting method is not developed.
(1) In the item No. input screen, press keys "4", "9" and "4" in this order.
(2) Using keys "0" through "9", enter photometric section initial position correction value.
Upon pressing ENTER key, the unit accepts the input and returns back to the item No. input screen.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
(2) The current incoming test strip detection voltage (A/D count) is displayed.
3-52
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-50 Item No. 520: Input introduction lever waiting position (time: ms)
Input the time for adjusting position where the introduction lever detects PI and stops. Using keys "1" through "9",
input control value and press ENTER key.
(1) In the item No. input screen, press keys "5", "2" and "0" in this order.
(2) Using keys "0" through "9", input introduction section waiting position control value and press ENTER
key.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
521.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
5-51 Item No. 521: Input surplus urine suction position of introduction lever (time: ms)
Input the time for adjusting position where the introduction lever detects PI and stops. Using keys "1" through "9",
input control value and press ENTER key.
(1) In the item No. input screen, press keys "5", "2" and "1" in this order.
(2) Using keys "0" through "9", input introduction section waiting position control value and press ENTER
key.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
522.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-53
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-52 Item No. 522: Input transport lever waiting position (time: ms)
Input the time for adjusting position where the transport lever detects PI and stops. Using keys "1" through "9",
input control value and press ENTER key.
(1) In the item No. input screen, press keys "5", "2" and "2" in this order.
(2) Using keys "0" through "9", input introduction section waiting position control value.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
523.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
5-53 Item No. 523: Input photometric section left waiting position pulse
Set the number of pulses for photometric section left waiting position in the 001 to 999.
(1) In the item No. input screen, press keys "5", "2" and "3" in this order.
(2) Using keys "0" through "9", input the No. of drive pulses for photometric section left waiting position.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
524.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-54
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-54 Item No. 524: Set drive pulse for right drive photometric section 1 (mark at right)
Set the number of drive pulses for photometric section. Input range is 0001 to 9999.
(1) In the item No. input screen, press keys "5", "2" and "4" in this order.
(2) Using keys "0" through "9", input the No. of drive pulses for photometric section.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
525.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
5-55 Item No. 525: Set drive pulse for left drive photometric section 1 (mark at right)
Set the number of drive pulses for photometric section. Input range is 0001 to 9999.
(1) In the item No. input screen, press keys "5", "2" and "5" in this order.
(2) Using keys "0" through "9", input the No. of drive pulses for photometric section.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
526.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-55
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-56 Item No. 526: Set drive pulse for right drive photometric section 2 (mark at left)
Set the number of drive pulses for photometric section. Input range is 0001 to 9999.
(1) In the item No. input screen, press keys "5", "2" and "6" in this order.
(2) Using keys "0" through "9", input the No. of drive pulses for photometric section.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
527.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
5-57 Item No. 527: Set drive pulse for left drive photometric section 2 (mark at left)
Set the number of drive pulses for photometric section. Input range is 0001 to 9999.
(1) In the item No. input screen, press keys "5", "2" and "7" in this order.
(2) Using keys "0" through "9", input the No. of drive pulses for photometric section.
Upon pressing ENTER key, the unit accepts the input value and moves to input screen (2) of item No.
528.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-56
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Upon pressing ENTER key, the unit accepts the input value and moves to the input screen (2) of the item
No. 529.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
Upon pressing ENTER key, the unit accepts the input value and moves to the input screen (2) of the item
No. 530.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-57
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Upon pressing ENTER key, the unit accepts the input value and moves to the input screen (2) of the item
No. 531. If STOP key is pressed, the unit ignores the input value and returns back to the item No. input
screen. Note 1)
(1) In the item No. input screen, press keys "5", "3" and "1" in this order.
Upon pressing ENTER key, the unit accepts the input value and moves to the EEPROM write
confirmation screen.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-58
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-59
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-62 Item No. 550: Change analytical curve information (threshold, abnormal mark rank)
Among information on test strip, threshold of reflectance and abnormal mark rank can be set. Change in threshold
may be left unsaved to EEPROM. (Threshold contained in Chinese version ROM cannot be changed.)
(1) In the item No. input screen, press keys "5", "5" and "0" in this order.
(2) Using keys "0" through "9", input the items of test strip to be changed.
Input values and corresponding test strip items are as shown below.
Item number vs item name
Item number Item name Item number Item name Item number Item name
01 GLU 07 PH 13 A/C
02 PRO 08 BLD 14 LPR
03 KET 09 LEU 15 LP/C
04 BIL 10 S.G. 16 P/C
05 URO 11 ALB
06 NIT 12 CRE
Upon pressing STOP key, the unit returns back to the item No. input screen.
If an item not registered as test item is selected, the unit will not go to the next value change input screen.
(3) Using "-" key, display the rank to be changed and set the threshold of that rank in terms of reflectance (in
unit of 0.1%). (Concentration is expressed as upper 4-digit value.)
Upon pressing START key, a symbol [ ] is displayed. Ranks below this rank will be given abnormal
mark. (Specific gravity cannot be set.)
Upon pressing ENTER key, the unit sets the input value and returns back to (2) test strip item input.
Upon pressing STOP key, the unit returns back to the item No. input screen. *See Note 1.
3-60
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
To perform measurement by using modified standard analytical curve, print the changed test strip
items by using "Print parameter". (To restore initialized analytical curve)
During reprinting and retransmission, backed up data will result in changed test strip item
information when output.
When "Change analytical curve" is selected again in the item No. input screen, the test strip item
information will be read from EEPROM. If the last process ended with "Not save in EEPROM",
the information is not the desired one.
3-61
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-63 Item No. 560: Set P/C ratio, DILUTE reference, PRO reference
Determine PRO reference used to judge DILUTE by P/C ratio. Input range is 000 to 999.
(1) In the item No. input screen, press keys "5", "6" and "0" in this order.
Upon pressing ENTER key, the unit accepts the input value and moves to the input screen of the item No.
561.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
5-64 Item No. 561: Set P/C ratio, DILUTE reference, CRE reference
Determine CRE reference used to judge DILUTE by P/C ratio. Input range is 000 to 999.
(1) In the item No. input screen, press keys "5", "6" and "1" in this order.
Upon pressing ENTER key, the unit accepts the input and returns back to the item No. input screen.
Note 1)
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
3-62
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-63
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
5-65 Item No. 600: Aution Net (entering information on analytical curve through PC)
Feed information on test strip, test strip item and color tone to the PC that will send the information to the unit.
The information may not be stored in EEPROM.
(This function is not contained in Chinese version ROM.)
(1) In the item No. input screen, press keys "6", "0" and "0" in this order.
(2) A screen will appear to indicate that communication is being processed. Send the settings from the PC.
Pressing STOP key cancels the process.
-------------
(3) When the PC ends the transmission of the information, EEPROM write confirmation screen will appear.
If the modified settings are to be used at the next power-up, press "1" key to save the settings on
EEPROM. If the modified settings are for temporary use and not required to be saved on EEPROM, press
"0" key. Pressing STOP key cancels all inputs.
EEP 0-1
(4) While writing to EEPROM, write screen appears that will be replaced by the item No. input screen as
writing process finishes.
No. 600
Stored
(CAUTION): When using this feature to modify the standard analytical curve before starting
measurement, print the changed items (test strip information, test strip item information)
by using "Print parameter" option. Save the printed document. This is to restore the
modified analytical curve in case the test strip information and test strip item information
are initialized.
During reprinting and retransmission, backed up data will result in changed test strip
information and test strip item information when output. (Color tone results are not
irrelevant.)
When process finished with "Not save to EEPROM", do not use other setting functions.
If "Aution Net" is selected later on the item No. input screen, test strip information, test
strip item information, color tone information and system parameters will be read from
EEPROM. If "Not save to EEPROM" was chosen at the end of the preceding process,
undesired information will be supplied.
3-64
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) The name of test strip and current setting (1: display as an option, 0: not display as an option) will be
displayed. When changing the current setting, press "0" or "1" key as appropriate.
(3) At the end of settings, EEPROM write confirmation screen will appear.
To acknowledge these settings, press "1" key. Remember that the backup data will be erased.
To cancel the inputs, press "0" key.
EEP 0-1
(4) Following "1" or "0" key operation, the screen will change to the item No. input screen.
3-65
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(5) Set test strip for each measurement mode. On the item No. input screen, press "2" key.
(6) Set test strip for each measurement mode. On the input screen, press "2" key.
(7) Select test strip for normal measurement mode with "-" key.
MEAS
(8) Select test strip for STAT mode with "-" key.
STAT
(9) Select test strip for control mode with "-" key.
CONT
3-66
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) With keys "0" to "9", input the serial number of device.
Acknowledge the input value by pressing ENTER key. The screen will change to the item No. 701 input
screen.
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
Note 1)
3-67
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-68
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using keys "0" through "9", input the reflectance higher threshold.
Upon pressing ENTER key, the unit accepts the input value and goes to the item No. 711: Print sample
error mark input screen (2).
If STOP key is pressed, the unit ignores the input value and returns back to the item No. input screen.
(2) Using key "1" or "0", turn ON or OFF error mark printing function.
3-69
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Using key "1" or "0", turn ON or OFF abnormal color detection feature.
Note 1) After each setting and before returning to the item No. input screen, the EEPROM write
confirmation screen appears. To save the change on EEPROM and make it valid on the next
power-up, press "1" key. If the current modification is for temporary use and will not be used again,
press "0" key (not save to EEPROM). Upon pressing STOP key, all current modifications will be
undone.
EEP 0-1
Saving on EEPROM
No. XXX
Stored
3-70
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(2) Erasure confirmation screen will appear. To erase contents, press "1" key. If not, press "0" key.
(2) Using keys "0" through "8", select the item(s) to be printed.
3-71
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
(3) Initialization confirmation screen will appear. To start initialization, press "1" key.
EEP 0-1
If "0" key is pressed, the unit will return to the item No. input screen without performing initialization.
(4) Initialization confirmation screen will appear. To start initialization, press "2" key.
EEP 0-2
At the end of initialization process, the unit returns to the item No. input screen.
If "0" key is pressed, the unit will return to the item No. input screen without performing initialization.
(CAUTION): When "0: All", "3: Test strip information", "4: Test strip item information" or "6:
Factory setting" is initialized, the backup data is cleared.
Initialization process on "Test strip information" or "Test strip item information" deletes
all the test strip information or test strip item information. If threshold or printing
sequence has been changed, make sure that contents to be erased by initialization process
can be restored (for example, print the test strip item information).
Also note that initialization of "Test strip information" or "Test strip item information"
makes measurement impossible. Be sure to perform "Select analytical curve".
Never turn off power during initialization.
3-72
Chapter 3 Servise Mode AE-4020 SERVICE MANUAL
6 Item No. 142: The operation method at the time of AJ system selection.
6-1 Select the measurement mode
1) On the Ready screen, press ID
The Meas Type screen is displayed.
The Meas icon flashes.
MEAS. STAT.
ID#
MEAS. MEAS.
Enter a Patient ID by using the numeric keys. A maximum of 13 digits including hyphen can be entered.
When you are not entering Patient ID for every sample, press the ENTER key to skip the unnecessary
sample.
2) Press ENTER.
The number is stored in memory.
Repeat the same procedure for the following Patient IDs.
MEAS. STAT.
ID# 1-0
Press 1.
The screen returns to the Standby.
To cancel the entering, press 0. The screen returns to the Standby.
3-73
AE-4020 SERVICE MANUAL
1 Circuit Diagrams
4-1
AE-4020 SERVICE MANUAL
5-1
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
1. Test Items
6) Waste Box
7) Detaching and reattaching the suction
ports
8) Surplus urine suction position of the
introduction arm
9) Check test strips feeding
10) Detaching and reattaching the
introduction arm, test strip tray, test
strip feed tray and feed lever
3 Functional 11) Check at Power ON
Tests
12) Operator Keys
13) Display
5-2
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
2. Mechanical Check
4) Power switch
Check the attachment condition of Power Switch, and ON/OFF.
[Standard]
The Power Switch should be attached firmly without looseness or stop. The switch correctly
correspond to ON/OFF.
5) Maintenance cover
Remove and attach the Maintenance Cover for several times and check the condition.
[Standard]
The Maintenance Cover should be removed and attached easily.
6) Waste Box
Remove the Waste Box several times and check the condition.
[Standard]
The Waste Box should smoothly be removed and inserted. No abnormal sound should be
heard.
7) Detaching and reattaching the suction ports
Remove and attach the suction ports for several times and check the condition.
[Standard]
The suction ports should be removed and attached easily.
5-3
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
10) Detaching and reattaching the introduction arm, test strip tray, test strip feed tray and feed
lever.
Remove and attach the introduction arm, test strip tray, test strip feed tray and feed
lever for several times and check the condition.
[Standard]
The introduction arm, test strip tray, test strip feed tray and feed lever should be removed
and attached easily.
3. Functional Tests
11) Check at Power ON
Check the operation and contents of the display.
[Standard]
The display and operation should be as follows:
At power ON→Full screen lighting
●Name of the analyzer, version No. are displayed for 2 seconds, then Self Check starts.
At Self Check→ AE-4020 01.00
●The analyzer checks the ROM and back up RAM, then initializing starts.
Initializing(Subtraction display)→ 21→20→……
●After initializing, the display goes to the Ready screen.
Ready screen(MEAS. Icon lighting, Measurement number/Test strips
classification display)→No,0001 10EA
(Initializing all functions - disposal of Test Strips)
5-4
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
13) Display
Check each letter of the display.
※It checks at the time of the all-points light display at the time of a power supply ON.
[Standard]
The letter should be clear without no missing part of dotted letter.
5-5
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
18) Bar-code
Connect the Hand-held Bar-code Reader (BL-N60VN : KEYENCE) and check the reading
condition.
[Standard]
Letter and number should be correctly read and displayed.
4. Performance Test
19) Check Strip
Equipment is made into check sample mode.
A check sample is measured each N=1 every right and every left.
(Choose by the hyphen key.)
[Standard]
All results of wavelengths should fall within the standard range.
5-6
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
「-4020-」→No,133→0(Invalid) →ENTER)
(Specific gravity compensation release:It sets up out of measurement standby as follow.
「-4020-」→No,135→0(Invalid) →ENTER)
[Standard]
Pro and PH: Fiducial %→ +2.5% and -4.5% of within the limits as a result of measurement.
The other item is ±3% of fiducial %.
Moreover, standard deviation (S. D) is less than 0.5% of thing.
5-7
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
5. Others
21) Exterior
Check any flaw of deteriorating the product value such as scratch, dirt or deformation.
[Standard]
No flaw that may deteriorate product value should be found.
23) Accessories
Check any flaw of deteriorating the product value such as scratch, dirt or deformation on all
accessories. Check the specifications and quantity of accessories are the same as
described on the drawings.
[Standard]
No flaw that may deteriorate product value should be found. The specifications and quantity
of accessories should be the same as described on the drawings.
24) Labeling
Check all the labelings adhered to the analyzer.
[Standard]
The following items should be permanently printed.
Approval Code
Manufacturer
Serial No.
Product name
Electrical standard
5-8
Chapter 5 Standard Inspection AE-4020 SERVICE MANUAL
5-9
AE-4020 SERVICE MANUAL
6-1
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
[Standard]
Adjust the time so that the end of auto start detect window and end of the test strip are
located at 0 - 0.5 mm with the test strip placed along the introduction arm.
Standard: 0 - 0.5 mm
Test strip
Introduction arm
• Key operation
7 key: Change adjusted value 50 ms frontward (standby position)
4 key: Change adjusted value 10 ms frontward
1 key: Change adjusted value 1 ms frontward
9 key: Change adjusted value 50 ms backward
6 key: Change adjusted value 10 ms backward
3 key: Change adjusted value 1 ms backward
0 key: Move back to the previous standby position
START key: Move to the currently adjusted position
ENTER key: Accept the currently adjusted value and return back to the item number input
screen.
STOP key: Terminate adjustment and return back to the item number input screen without
storing the value.
6-2
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
Suction ports
(suction plate)
Introduction arm
Not go checking
Introductionarm section
positioning fixture Go checking section
(74-12158)
Place the introduction lever alignment fixture on the suction ports as show below.
6-3
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
Introduction arm
and pusher
Adjust the distance from the pusher tip to the depth to 4.5 - 4.85 mm.
A pair of pushers have 0.5 mm max. longitudinal deviation at tip. Correct the deviation at
the pusher having deviation at front.
6-4
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
OK NG
Adjusting alignment arm standby position<3)-3> Readjusting test strip alignment position<3)-4>
End of adjustment
< 3)-1> Test strip alignment center position.
Prior to this adjustment, position of the introduction arm for surplus urine suction (Reverse mode 440)
shall be completed.In the single unit operation mode 400 (Reverse mode), press No.2 to move the
introduction arm to the surplus urine suction position.
Place the alignment lever adjustment fixture (74-11794) on the 1st port.
Adjust the lateral position of alignment mechanism with setscrew so that the clearance between the tip
of alignment arm rib and the fixture is 0 mm. Tighten the setscrew.
Specification : Clearance between tip of alignment arm rib and the fixture:0 ± 0.25 mm.
(At both left and right side)
Specification : Space between left and right alignment arm ribs: 111 ± 0.5 mm
<3)-2> Verifying test strip transportation status. Test strip 10EA contacts with arm
With alignment arm positioned as specified in ib
3)-1, test strip 10EA shall be placed to touch
the rib of right-hand arm.
6-5
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
Press No.1 key. While leaving the installed test strip as it is, return the introduction arm and
alignment arm to the standby position. Make sure that the test strip is undisturbed. If disturbed, use
"2" key to move the alignment arm to alignment position to bring the test strip touch the arm.
Under the condition shown in the figure above, press "5" key to move the test strip to a port under
the photometric section.
(Standby position on the transport member is adjusted (Reverse mode 450) and the alignment
spring under the photometric section touches Ctr of transport table.)
Alignment spring
Transport table
Ctr
Clearance between
test strip and guide
Refer to the figure above. Pass a 0.5 mm block gage through the clearance between test strip and
guide. (If the block gage fails to pass, jump to 3)-4.)
Next, reverse the direction of the test strip as shown in the figure below (holding portion of the strip
at left), touching it to left-hand arm.
6-6
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
In reverse mode 400, use "1" key to move the introduction arm and alignment arm to the standby
position.
Specification: Clearance between tip of alignment arm rib and the fixture:
0 ± 0.25 mm. (At both left and right side)
Specification: Distance between ribs of left and right alignment arm at standby position is
121 ± 0.5 mm.
6-7
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
Specification:
Clearance between tip of alignment arm rib and the fixture : 0 ± 0.25 mm.
Specification:
Space between left and right alignment arm ribs: 111 ± 0.5 mm
Distance from the guide of transport member to the top of alignment arm as the test strip is moved
rightward:
Distance from the guide of transport member to the top of alignment arm as the test strip is moved
leftward:
6-8
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
Specification:
The clearance between L/R guide and the end of coming test strip measured under
If the clearance is outside this specification, the bonded pulley and associated parts around it
may have play or looseness, or transport motor is not positively installed or secured.
6-9
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
[Standard]
• The position equivalent of +7 (ms) from the position where test strip alignment springs (at
left and right) contact the test strip tray.
[Functional requirement]
[1] Test strip alignment springs shall positively contact with the test strip tray.
[2] No test strip shall be caught by the transport lever, bent backward, at test strip waste
stage.
No gap or clearance
Test strip
Transport lever
• Key operation
7 key: Change adjusted value 10 ms frontward (test strip feed direction)
4 key: Change adjusted value 5 ms frontward (test strip feed direction)
1 key: Change adjusted value 1 ms frontward (test strip feed direction)
9 key: Change adjusted value 10 ms backward (opposite to test strip feed direction)
6 key: Change adjusted value 5 ms backward (opposite to test strip feed direction)
3 key: Change adjusted value 1 ms backward (opposite to test strip feed direction)
START key: Move to the currently adjusted position
ENTER key: Accept the currently adjusted value and return back to the item number input
screen.
STOP key: Terminate adjustment and return back to the item number input screen without
storing the value.
6-10
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
[Standard] At all 3 locations in front of the photometric position, 0 - 0.5 mm gap as shown in
the figure below shall be maintained.
Photometric
section
Transport table
0 - 0.5 mm
• Key operation
7 key: Shift adjusted value leftward by 100 pulses
4 key: Shift adjusted value leftward by 10 pulses
1 key: Shift adjusted value leftward by 1 pulse
9 key: Shift adjusted value rightward by 100 pulses
6 key: Shift adjusted value rightward by 10 pulses
3 key: Shift adjusted value rightward by 1 pulse
START key: Move to left standby position according to the entered value.
ENTER key: Accept the currently adjusted value and return back to the item number input
screen.
STOP key: Terminate adjustment and return back to the item number input screen without
storing the value.
6-11
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
• 430 nm:
VR01 (Driver print circuit board): 2700 ± 50 counts
• 565 nm:
VR03 (Driver print circuit board): 2700 ± 50 counts
• 635 nm:
VR04 (Driver print circuit board): 2700 ± 50 counts
• 760 nm:
VR05 (Driver print circuit board): 2700 ± 50 counts
* Reading of 2700 ± 200 counts at a wavelength after adjustment is allowable when taking into
account initial fluctuation and temperature variation. Reading outside this range requires
readjustment.
6-12
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
7) Adjusting black mark detecting pulse (black mark placed on right side)
• Prepare a test strip having a black mark at a position within the dimensional tolerance specified
below.
• Perform adjustment with upper cover closed.
4 ± 0.3
83.5 ± 0.3
110 ± 0.5
6-13
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
8) Adjusting black mark detecting pulse (black mark placed on left side)
• Place the test strip used in paragraph (4) on the photometric measurement position with the
black mark on its left side.
• Start adjustment with the upper cover closed.
4 ± 0.3
83.5 ± 0.3
110 ± 0.5
6-14
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
* CAUTION [1]: Stop turning VR07 CW once the LED turns on. Higher sensitivity causes sensor
error: LED is kept lit.
* CAUTION [2]: Do not touch test strips N8, N5 and N3 with bare hand. ND surface must be free
from dust and flaw: if contaminated or degraded, replace with new one. Or
replace periodically, at least every 6 months. Store test strips at room
temperature in lightproof condition.
Detection window
6-15
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
9)-2. Verification 1
[1] Hold the introduction lever at introduction position during the step below.
[2] Wile sliding the test strip N5 for adjustment (with dull N5 attached) from the front of detection
window to the depth of the window, verify that the automatic start sensor displays "0: not
detected" on the screen.
* As noted before, ND paper attached surface is facing detection surface (detection window
side).
NO. 400 1. 0. 0. 1. 0. 0
9)-3. Verification 2
[2] Leave the screen display as it is. Position the test strip N3 for adjustment (ND paper: dull N3
attached) 2-3 mm from the front of detection window.
* As noted before, ND paper attached surface is facing detection window.
[3] Put your hand over the detection window until "1: detected" is displayed on the screen.
[4] Put back the hand. Display "1: detected" will change to "0: not detected"
[5] Move the test strip N3 for adjustment (ND paper: dull N3 attached) to a position 2-3 mm from
the rear of detection window; and repeat steps [2] and [3] above. Verify "0: not detected"
message.
6-16
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
9)-4. Verification 3
[1] Place a white base film (test strip for adjustment N8, N5 or N3 with ND attached surface
facing upward) at a position approx. 5 mm from the front. Move the white base film slowly
backward while observing the screen for the message generated by automatic start sensor.
When the message changes from "0: not detected" to "1: detected", verify that the distance
is within 1.5 mm from the end of detection window.
[2] In the same condition as for the above step, move the base film from a position 5 mm behind
the rear slowly toward front. When the message changes from "0: not detected" to "1:
detected", verify that distance is within 1.5 mm from the end of detection window.
z Test strip for adjustment (N8, N5, N3) ND8 (dull ND8 attached)
* CAUTION: Excessively increased sensitivity of automatic start detection will keep "ON" state
during measurement, and may generate E008 automatic start sensor error. After
increasing the sensitivity of automatic start sensor, be sure to perform verification 1
(ND5) and verification 2 (ND3).
[Reference] Reflectance (760 nm) of test strips for adjustment measured on AE-4020
[1] ND8: approx. 70%
[2] ND5: approx. 22%
[3] ND3: approx. 7%
[Remarks] Do not touch test strips for adjustment (N8, N5, N3) with bare hand. Keep ND surface
free from dust and flaw. Save strips at room temperature under lightproof condition.
Replace strips as they become dirty or degraded, or periodically replace (approx.
every half a year).
6-17
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
10)-2. Adjustment
[1] Open the incoming strip sensing PI level check screen. (Reverse mode 500 → ENTER)
[2] Without using test strip, perform the following settings.
* Main board cover installed and upper cover closed
• Control: VR06 (driver PCB)
• Detect level: 100 ± 5 counts
• LED: DO3 (driver PCB) unlit
10)-3. Checking
[1] Place a test strip on the incoming strip sensing PI section and close the upper cover.
[2] Verify the following:
• Detect level: 900 counts or more
• LED: DO3 (driver PCB) is lighting
6-18
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
[Specification] Reflectance at 565 nm, 635 nm and 760 nm shall be equal to or lower than
the % value shown below at all positions for PAD1 through PAD10.
• 100 ± 2.0 %
6-19
Capter 6 Adjustment Method AE-4020 SERVICE MANUAL
12)-2. Adjustment
• Reverse mode: 480 → ENTER
• Press START key.
• Start measurement of the ion exchanged water on required items, 5 times in synchronous with
buzzer sound.
• Measured color tone corrections and DIP detection thresholds and adjustments will be printed.
• Press "1" key to save the corrected values.
12)-3. Verification
• Measure the ion exchanged water 5 times by repeating procedure described in 5.3.
• Verify that the following specifications are met.
• When OK, press 0 key.
[Specification]
• Averaged (Avr) C/M, Y/M, C/r: 0±5
• Variation (R) in C/M, Y/M, C/r: 20 or less
6-20
AE-4020 SERVICE MANUAL
Appendix
No. Description Appendix._Rev._
1 External Output Specification Appendix.1_Rev.F
2 Print Format Specification Appendix.2_Rev.F
3 Error & Trouble Specification Appendix.3_Rev.E
4 Program initialization Appendix.4_Rev._
5 Rewriting main program (only for a factory.) Appendix.5_Rev._
6 Ethernet setup Appendix.6_Rev._
7 Packing Appendix.7_Rev._
8 Work standard (only for a factory.) Appendix.8_Rev._
9 Maintenance parts Appendix.9_Rev._
Appendix AE-4020 SERVICE MANUAL
App 1
App 2
App 3
App 4
Program initialization
Appendix AE-4020 SERVICE MANUAL
App 5
App 6
Ethernet setup
Appendix AE-4020 SERVICE MANUAL
App 7
Packing
App 8
Appendix AE-4020 SERVICE MANUAL
Work standard
Appendix AE-4020 SERVICE MANUAL
App 9
Maintenance parts
Appendix.1 AE-4020 SERVICE MANUAL
Revision History
Revision Date of
Revised Item Description/Date of Execution
Number Revision
A 2004/02/24 AS-4020 format Measurement error has been
added in the field of test order
record.
B 2004/05/28 Entire configuration. Temperature has been added in
AS-4020 format the field of test order record.
Test strip name has been added in
the test order record and result
SA-4230 format record.
Device name changed. SA-4230 format has been deleted.
AS-4020 → AE-4020
C 2004/09/07 MA-4240 format MA-4240 format has been added.
AE-4020 format Ethernet communication.
Changed into ACK/NAK at every
message.
Changed the packet detection
system into end data.
SOH has been added to the
beginning of sending message.
RS-232C communication.
ETB has been added to a frame
other than the final record.
Test order record.
Sample type has been added.
Message header record.
The device version number has
been fixed to a 10-digit number.
D 2005/01/10 General Errors corrected.
E 2005/04/18 4-1. Communication Ethernet port number added.
Setting
F 2005/07/12 8.AM-4290 format Errors corrected.
10.SA-4230 format SA-4230 format has been added.
App.1-2
Appendix.1 AE-4020 SERVICE MANUAL
CONTENTS
1. OVERVIEW .......................................................................................................................................... 5
2. RS-232C ................................................................................................................................................. 5
2-1. External Output System................................................................................................................... 5
2-2. Applicable Connector...................................................................................................................... 5
2-3. Communication Specification ......................................................................................................... 5
2-4. Transfer Rate ................................................................................................................................... 5
2-5. Data Format..................................................................................................................................... 6
2-6. Flow Control.................................................................................................................................... 6
2-7. Output Timing ................................................................................................................................. 6
2-8. Time Gap (Interval between texts) .................................................................................................. 6
4. ETHERNET ........................................................................................................................................ 13
4-1. Communication Setting ................................................................................................................. 13
4-2. Ethernet Communication Frame Structure .................................................................................... 13
App.1-3
Appendix.1 AE-4020 SERVICE MANUAL
9. SA-4230-COMPATIBLE FORMAT.................................................................................................. 38
9-1. Block Configuration...................................................................................................................... 38
9-2. Details of First Block (Sample Information) ................................................................................. 38
9-3. Details of Second Block (Measurement Result 1) ........................................................................ 40
9-4. Details of Second Block (Measurement Result 1) ........................................................................ 40
7-7. Details of Fourth Block (Measurement result of specific gravity) ................................................ 42
App.1-4
Appendix.1 AE-4020 SERVICE MANUAL
1. Overview
This section describes the details of the specifications about external output of AE-4020, especially
data format and RS-232C two-way communication protocol and Ethernet
AE-4020 external output can make an output in a format in compliance with ASTM-1394. This
output in the AE-4020 standard format is made in the two-way communication between the device
and PC.
Moreover, AE-4020 is compatible with a conventional device and it can output a data in
AX-4280-compatible format, AM-4290-compatible format (supports AUTION Sticks only) and
MA-4240-compatible format. The output in these compatible format is made in a conventional
unidirectional communication.
(Note) The details of this specification are subject to change or revision according to the version upgrade of ROM
(built-in software) of AE-4020.
2. RS-232C
2-1. External Output System
RS-232C-compliant bit serial output.
App.1-5
Appendix.1 AE-4020 SERVICE MANUAL
App.1-6
Appendix.1 AE-4020 SERVICE MANUAL
Device PC
<ENQ>
<ACK>
To transfer phase
(Frame transmission)
App.1-7
Appendix.1 AE-4020 SERVICE MANUAL
If the device receives <NAK> after transmitting <ENQ>, the device transmits <ENQ> again after waiting
10 seconds.
If the device does not receive <ACK> although trying to transmit <ENQ> six times (including the first
time), the device regards the status as trouble after transmitting <EOT>. (Note)
Device PC
<ENQ>
<ENQ>
Wait 10 seconds <NAK>
・ ・
・ ・
・ ・
<ENQ>
<NAK>
<EOT>
If the device receives <ENQ> after transmitting <ENQ>, the device transmits <ENQ> again after waiting
1 second. (Note)
If the PC receives <ENQ> after transmitting <ENQ>, the PC interrupts the transmission and waits
<ENQ>. If the PC does not receive <ENQ> after waiting 20 seconds, the PC transmits <ENQ> again.
Device PC
<ENQ>
<ENQ>
Wait 1 second
<ENQ> Interrupts
transmission and
waits <ENQ>.
<ACK>
To transfer phase
(Frame transmission)
(Note) If a collision of <ENQ> occurs, a priority is given to the device and the PC waits the next <ENQ>.
App.1-8
Appendix.1 AE-4020 SERVICE MANUAL
If no response is made for 15 seconds after transmitting <ENQ>, the device regards the status as trouble
after transmitting <EOT>. (Note)
Device PC
<ENQ>
Waits a response
for 15 seconds
<EOT>
If no response is made for
15 seconds, the status is
regarded as trouble.
(Note) The status is regarded as trouble to detect the line breakage (cable disconnection, etc.).
Device PC
<ACK>
<ACK>
End frame (ETX)
<ACK>
App.1-9
Appendix.1 AE-4020 SERVICE MANUAL
If the device receives anything other than <ACK> after transmitting a frame, the device performs the
same process as receiving <NAK>. (Note)
(Note) The device treats the <EOT> response after transmitting the frame as <NAK>, not regarding it as
transmission interruption.
If the device has no more frame to be transmitted when it has received <ACK> after transmitting the
frame, the device enters the transmission end phase. If the device has any frame to be transmitted, the
device transmits the next frame.
When the device has transmitted all the frames, the device sends <EOT>.
Device PC
Transmitting a record
The device has the
<ACK>
frames to be transmitted.
If the device receives <NAK> after transmitting a frame, the device transmits the frame again after
waiting 1 second.
If the device does not receive <ACK> although trying to transmit the same frame six times (including the
first time), the device regards the status as trouble after transmitting <EOT>. (Note)
Device PC
Frame
<NAK>
Wait 1 second
Frame
<NAK>
Wait 1 second
・ ・
・ ・
・ ・
Frame
<NAK>
<EOT>
App.1-10
Appendix.1 AE-4020 SERVICE MANUAL
If no response is made for 15 seconds after transmitting a frame, the device regards the status as trouble
after transmitting <EOT>. (Note)
Device PC
Frame
Waits a response
for 15 seconds
<EOT>
If no response is made for
15 seconds, the status is
regarded as trouble.
(Note) The status is regarded as trouble to detect the line breakage (cable disconnection, etc.).
App.1-11
Appendix.1 AE-4020 SERVICE MANUAL
[Example]: Communication made when the device sends a measurement result to PC.
App.1-12
Appendix.1 AE-4020 SERVICE MANUAL
4. Ethernet
The Ethernet communication in our device is made using the Ethernet microcomputer.
PC Device
A
Application Application
Winsock Driver
Driver
Ethernet Ethernet
Controller Microcomputer
App.1-13
Appendix.1 AE-4020 SERVICE MANUAL
5-2. Transfer
(1) Record Transmission by Device
If the device receives anything other than <ACK> after transmitting a record, the device performs the
same process as receiving <NAK>.
The device terminates the transmission when it has received <ACK> after transmitting an end record.
Otherwise, the device transmits the next record.
Device PC
Transmitting a record
First packet
Packet #2
Packet #3
Last packet
<ACK>
App.1-14
Appendix.1 AE-4020 SERVICE MANUAL
If the device receives <NAK> after transmitting a record, the device stops transmission and regards the
status as trouble. (Note)
Device PC
Transmitting a record
<NAK>
(Note) A priority is given to the device in the communication and the device does not receive <NAK> usually.
If the device receives any <NAK>, the status is regarded as trouble.
If no response is made for 15 seconds after transmitting a record, the device regards the status as trouble .
(Note)
Device PC
Transmitting a record
Waiting 15 seconds
(Note) The status is regarded as trouble to detect the line breakage (cable disconnection, etc.).
App.1-15
Appendix.1 AE-4020 SERVICE MANUAL
[Example]: Communication made when the device sends a measurement result to PC.
Device: <SOH>
Device: H|$^&|||AE-4020^10405001^V01.00 |||||||||200203111010<CR>
Device: P|1<CR>
Device: O|1|0203110000001|1|^^^10EA|R||||||||||U||||26^<CR>
Device: R|1|^^^10EA^GLU|+2^150|^mg/dL||!||F|||200203111010<CR>
Device: R|2|^^^10EA^PRO|+4^OVER|^mg/dL||*||F|||200203111010<CR>
Device: R|3|^^^10EA^BIL|-^|^mg/dL||||F|||200203111010<CR>
Device: R|4|^^^10EA^PH|^6.5|||||F|||200203111010<CR>
Device: R|5|^^^10EA^S.G.|^>1.030|||||F|||200203111010<CR>
Device: L|1|N<CR>
Device: <EOT>
PC: <ACK><EOT>
App.1-16
Appendix.1 AE-4020 SERVICE MANUAL
Communication
ETB
(transmission) data
STX FN or CSH CSL CR LF
(A maximum of 24
ETX
bytes)
App.1-17
Appendix.1 AE-4020 SERVICE MANUAL
6-2. Ethernet
Unlike RS-232C, the frame does not contain STX, FN, ETX and checksum. For the details of data structure,
refer to Section 6-3. (The data structure is the same as that of RS-232C.)
Communication (transmission)
SOH EOT
data
Each record consists of several fields. The content of the record is defined by a record type. The number of
fields in a record is fixed at every record type. Any data in a field is variable (the smallest size is 0 by default).
Every field is delimited by a field delimiter. A blank field is also delimited by a field delimiter, so we can see
the field is blank. (The last field may not be delimited by a field delimiter.)
(Note 1) A result record is appended per measurement item, therefore, several result records are appended to
one measurement result. The number of result records varies depending on test strip.
App.1-18
Appendix.1 AE-4020 SERVICE MANUAL
<Example> P|1<CR>
App.1-19
Appendix.1 AE-4020 SERVICE MANUAL
<Example> O|1|0203110000001|1|^^^10EA|R||||||||||U||||25^DRIFT<CR>
(Note 1) Measurement error:
The details of error are shown below.
App.1-20
Appendix.1 AE-4020 SERVICE MANUAL
<Example> R|1|^^^10EA^GLU|+2^150|^mg/dL||!||F|||200203111010<CR>
R|2|^^^10EA^BIL|-^|^mg/dL||||F|||200203111010<CR>
R|3|^^^10EA^PH|^5.0|||||F|||200203111010<CR>
R|4|^^^10EA^COLOR|RED|||||F|||200203111010<CR>
App.1-21
Appendix.1 AE-4020 SERVICE MANUAL
Device: <SOH>
Device: H|$^&|||AE-4020^10405001^V01.00 |||||||||200203111010<CR>
Device: P|1<CR>
Device: O|1|0203110000001|1|^^^10EA|R||||||||||U||||26^STRIP DEVIATE<CR>
Device: R|1|^^^10EA^GLU|^|^mg/dL||||F|||200203111010<CR>
Device: R|2|^^^10EA^PRO|^|^mg/dL||||F|||200203111010<CR>
Device: R|3|^^^10EA^BIL|^|^mg/dL||||F|||200203111010<CR>
Device: R|4|^^^10EA^PH|^|||||F|||200203111010<CR>
Device: R|5|^^^10EA^S.G.|^|||||F|||200203111010<CR>
Device: L|1|N<CR>
Device: <EOT>
PC: <ACK><EOT>
App.1-22
Appendix.1 AE-4020 SERVICE MANUAL
7. AX-4280-Compatible Format
7-1. Compatibility with SA-4250, AX-4280 and AJ-4270 Standard Format
AX-4280-compatible format has some compatibility with the SA-4250 and AX-4280 standard formats, but it is
slightly different from them. For example, because there is no rack number nor port number in AJ-4270 and
AE-4020 and these are not equipped with an independent specific gravity/turbidity measurement unit, AJ-4270
and AE-4020 do not output the concrete value for those items (a blank value or “ –”is output). In measurement
with the test strip AUTION Sticks, the item “ S. G.”can be measured when a specific test strip (10EA, 9EA) is
used. For the SA-4250 and AX-4280 formats, the measurement result of specific gravity meter is output in the
specific gravity area of the fourth block. On the other hand, for the AX-4280-compatible format, “ S. G.”is
output as the measurement result of the item (in either the second or third block), otherwise the measurement
result of specific gravity meter is output in the specific gravity area of the fourth block. You can select any of
these two types of output.
The table below shows the major differences between the SA-4250 format and the AX-4280-compatible
format.
Difference AX-4280 Format AX-4280-Compatible Format
Device name “AX-4280€35 “A E-4020€35
Port number Uses a rack number and port The value is blank.
number.
The number of 10 characters 7 characters and 3 spaces
characters for test strip
name
Temperature Value (00 to 99) Value (0 to 55)
Output of the item “ S. Disabled “Enable” or “ Disable” can be
G.” in the second and selected in the setting.
third blocks
Output of the “ S. G.”Outputs the measurement “ Enable” or “ Disable” can be
(Specific Gravity) area in result of specific gravity selected in the setting. A blank is
the fourth block meter. output if “Disable”is selected.
Turbidity The measurement result of Nil
turbidity meter is output to
the measurement results 1 to
12.
(Note) To maintain the compatibility with SA-4250, the measurement number of control measurement is not
output. (A blank is output for a sample number.)
App.1-23
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
V A E - 4 0 2 0 V .
19 20 21 22 23 24 25 26 27 28 29 30 31 32
B
33 34 35 36 37 38 39 40 41 42 43
D / /
44 45 46 47 48 49
T :
50 51 52 53 54 55 56 57
N .
58 59 60 61 62 63 64
P -
65 66 67 68 69 70 71 72 73 74 75
S
76
ETB
App.1-24
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11 12 13 14
15 16 17 18 19 20 21 22 23 24 25 26
27 28 29 30 31 32 33 34 35 36 37 38 39
40 41 42 43 44 45 46 47 48 49 50 51
52 53 54 55 56 57 58 59 60 61 62 63 64
65 66 67 68 69 70 71 72 73 74 75 76
77 78 79 80 81 82 83 84 85 86 87 88 89
90 91 92 93 94 95 96 97 98 99 100 101
102 103 104 105 106 107 108 109 110 111 112 113 114
▲
115 116 117 118 119 120 121 122 123 124 125 126
127 128 129 130 131 132 133 134 135 136 137 138 139
140 141 142 143 144 145 146 147 148 149 150 151
152
ETB
For the item name, semi-quantitative symbol, concentration and unit symbol, refer to the
calibration curve specifications.
App.1-25
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11 12 13 14
15 16 17 18 19 20 21 22 23 24 25 26
27 28 29 30 31 32 33 34 35 36 37 38 39
40 41 42 43 44 45 46 47 48 49 50 51
52 53 54 55 56 57 58 59 60 61 62 63 64
65 66 67 68 69 70 71 72 73 74 75 76
77 78 79 80 81 82 83 84 85 86 87 88 89
90 91 92 93 94 95 96 97 98 99 100 101
102 103 104 105 106 107 108 109 110 111 112 113 114
115 116 117 118 119 120 121 122 123 124 125 126
127 128 129 130 131 132 133 134 135 136 137 138 139
140 141 142 143 144 145 146 147 148 149 150 151
152
ETB
App.1-26
Appendix.1 AE-4020 SERVICE MANUAL
7-7. Details of Fourth Block (Measurement result of specific gravity and color tone)
(1) Format
1
STX
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
40 41 42 43 44 45 46
47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65
C O M .
66 67 68
69
ETX
App.1-27
Appendix.1 AE-4020 SERVICE MANUAL
a) Left justification
The table below shows the details of error.
Details of Measurement Error Meaning Data
Normal Available
DRIFT Change in DARK light quantity Available
REFLEX OVER Considerably high reflectance Available
STRIP DEVIATE Black mark misaligned Unavailable
DIP MISS No dip Available
ILLEGAL PAPER Test strip type error Unavailable
App.1-28
Appendix.1 AE-4020 SERVICE MANUAL
8. AM-4290-Compatible Format
8-1. Details of Format (Concentration Value Output)
(1) Format
1
STX
2 3 4 5 6 7 8 9 10 11 12 13
D - - ,
14 15 16 17 18 19 20 21 22 23 24 25 26 27
T : , N ,
28 29 30 31 32 33 34 35 36 37 38 39 40
T e m p E A CR LF
41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57
G L U , ,
58 59 60 61 62 63 64 65 66 67 68
, , CR LF
69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85
P R O , ,
86 87 88 89 90 91 92 93 94 95 96
, , CR LF
97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113
B I L , ,
114 115 116 117 118 119 120 121 122 123 124
, , CR LF
125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141
U R O , ,
142 143 144 145 146 147 148 149 150 151 152
, , CR LF
153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169
P H , ,
170 171 172 173 174 175 176 177 178 179 180
, , CR LF
181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197
S G , ,
198 199 200 201 202 203 204 205 206 207 208
, , CR LF
209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225
B L D , ,
226 227 228 229 230 231 232 233 234 235 236
, , CR LF
237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253
K E T , ,
254 255 256 257 258 259 260 261 262 263 264
, , CR LF
265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281
N I T , ,
282 283 284 285 286 287 288 289 290 291 292
, , CR LF
293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309
L E U , ,
310 311 312 313 314 315 316 317 318 319 320
, , CR LF
321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337 338
, ,
339 340 341 342 343 344 345 346 347 348
, , CR LF
349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365 366
C O L O R , ,
367 368 369 370 371 372 373 374 375 376
, , CR LF
377
App.1-29
Appendix.1 AE-4020 SERVICE MANUAL
ETX
App.1-30
Appendix.1 AE-4020 SERVICE MANUAL
*Note The AM-4290 format supports the conventional test strip (10EA, 9EA, 8EA, 7EA, 6EA,
5EA and 4EA) only.
* If 10PA is measured in this format, a data of CRE is not output because the output item is
fixed.
* If 2AC is measured, the measurement result becomes blank because there is no applicable
item.
The table below shows whether a data is enabled or disabled when a measurement error occurs.
Details of Measurement Error Meaning Data
Normal Enabled
DRIFT Change in DARK light quantity Enabled
REFLEX OVER Considerably high reflectance Enabled
STRIP DEVIATE Black mark misaligned Disabled
DIP MISS No dip Enabled
ILLEGAL PAPER Test strip type error Disabled
App.1-31
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11 12 13
D - - ,
14 15 16 17 18 19 20 21 22 23 24 25 26 27
T : , N ,
28 29 30 31 32 33 34 35 36 37 38 39 40
T e m p E A CR LF
41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57
G L U , ,
58 59 60 61 62 63 64 65 66 67 68
, , CR LF
69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85
P R , ,
86 87 88 89 90 91 92 93 94 95 96
, , CR LF
97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113
B I L , ,
114 115 116 117 118 119 120 121 122 123 124
, , CR LF
125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141
U R , ,
142 143 144 145 146 147 148 149 150 151 152
, , CR LF
153 154 155 156 157 158 159 160 161 162 163 164 165 166 167 168 169
P H , ,
170 171 172 173 174 175 176 177 178 179 180
, , CR LF
181 182 183 184 185 186 187 188 189 190 191 192 193 194 195 196 197
S G , ,
198 199 200 201 202 203 204 205 206 207 208
, , CR LF
209 210 211 212 213 214 215 216 217 218 219 220 221 222 223 224 225
B L D , ,
226 227 228 229 230 231 232 233 234 235 236
, , CR LF
237 238 239 240 241 242 243 244 245 246 247 248 249 250 251 252 253
K E T , ,
254 255 256 257 258 259 260 261 262 263 264
, , CR LF
265 266 267 268 269 270 271 272 273 274 275 276 277 278 279 280 281
N I T , ,
282 283 284 285 286 287 288 289 290 291 292
, , CR LF
293 294 295 296 297 298 299 300 301 302 303 304 305 306 307 308 309
L E U , ,
310 311 312 313 314 315 316 317 318 319 320
, , CR LF
321 322 323 324 325 326 327 328 329 330 331 332 333 334 335 336 337
, ,
338 339 340 341 342 343 344 345 346 347 348
, , CR LF
App.1-32
Appendix.1 AE-4020 SERVICE MANUAL
349 350 351 352 353 354 355 356 357 358 359 360 361 362 363 364 365
, ,
366 367 368 369 370 371 372 373 374 375 376
, , CR LF
377
ETX
App.1-33
Appendix.1 AE-4020 SERVICE MANUAL
App.1-34
Appendix.1 AE-4020 SERVICE MANUAL
9. MA-4240-Compatible Format
9-1. Details of Format (Measurement Result)
(1) Format
001
STX
002 003 004 005 006 007 008 009 010 011
D - - ,
012 013 014 015 016 017 018 019 020 021 022 023 024 025
T : , N ,
026 027 028 029 030 031 032 033 034 035 036 037 038
S CR LF
039 040 041 042 043 044 045 046 047 048 049 050 051 052 053 054 055
, ,
056 057 058 059 060 061 062 063 064 065 066
, , CR LF
067 068 069 070 071 072 073 074 075 076 077 078 079 080 081 082 083
, ,
084 085 086 087 088 089 090 091 092 093 094
, , CR LF
095 096 097 098 099 100 101 102 103 104 105 106 107 108 109 110 111
, ,
112 113 114 115 116 117 118 119 120 121 122
, , CR LF
123 124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139
, ,
140 141 142 143 144 145 146 147 148 149 150
, , CR LF
151 152 153 154 155 156 157 158 159 160 161 162 163 164 165 166 167
, ,
168 169 170 171 172 173 174 175 176 177 178
, , CR LF
179 180 181 182 183 184 185 186 187 188 189 190 191 192 193 194 195
, ,
196 197 198 199 200 201 202 203 204 205 206
, , CR LF
207 208 209 210 211 212 213 214 215 216 217 218 219 220 221 222 223
, ,
224 225 226 227 228 229 230 231 232 233 234
, , CR LF
235 236 237 238 239 240 241 242 243 244 245 246 247 248 249 250 251
, ,
252 253 254 255 256 257 258 259 260 261 262
, , CR LF
263 264 265 266 267 268 269 270 271 272 273 274 275 276 277 278 279
, ,
280 281 282 283 284 285 286 287 288 289 290
, , CR LF
291 292 293 294 295 296 297 298 299 300 301 302 303 304 305 306 307
, ,
308 309 310 311 312 313 314 315 316 317 318
, , CR LF
319
ETX
App.1-35
Appendix.1 AE-4020 SERVICE MANUAL
a) Classification number:
Fixed to [0] for normal measurement.
[C] for the control measurement.
[S] for the STAT measurement.
b) Test strip name:
8EA, 8UA [URIFLET 8A]
8UB [URIFLET 8B]
7EA, 7UA [URIFLET 7A]
App.1-36
Appendix.1 AE-4020 SERVICE MANUAL
Any other test strips are left-justified when those are output.
Example: 10EA [10EA ]
* For 10PA, the number of output item is 10 or more. So, any items after the tenth are
not output.
* The number of characters for semi-quantitative symbol output is 5 and any character
(the first character) may be omitted.
* Troubleshooting in the MA-4240-compatible format (no measurement result)
W001: Drift [3]: Test strip setting error
W002: Considerably high reflectance [3]: Test strip setting error
W003: Test strip misaligned [3]: Test strip setting error
W004: No dip [3]: Test strip setting error
W005: Test strip type error [7]: Test strip type judgment disabled
App.1-37
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4
M ■ ■
5 6 7 8 9 10 11 12 13 14 15 16 17 18
B ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
19 20 21 22 23 24 25 26 27 28 29
D ■ ■ ■ ■ / ■ ■ / ■ ■
30 31 32 33 34 35
T ■ ■ : ■ ■
36 37 38 39 40 41 42 43
N ■ ■ . ■ ■ ■ ■
44 45 46 47 48 49 50
P □ □ □ − □ □
51 52 53 54 55 56 57 58 59 60 61
S ■ ■ ■ ■ ■ ■ ■ □ □ □
62
ETB
App.1-38
Appendix.1 AE-4020 SERVICE MANUAL
App.1-39
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11
■ ■ ■ ■ □ □ ■ ■ ■ ■
12 13 14 15 16 17 18 19 20 21 22 23
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
24 25 26 27 28 29 30 31 32 33
■ ■ ■ ■ □ □ ■ ■ ■ ■
34 35 36 37 38 39 40 41 42 43 44 45
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
46 47 48 49 50 51 52 53 54 55
■ ■ ■ ■ □ □ ■ ■ ■ ■
56 57 58 59 60 61 62 63 64 65 66 67
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
68 69 70 71 72 73 74 75 76 77
■ ■ ■ ■ □ □ ■ ■ ■ ■
78 79 80 81 82 83 84 85 86 87 88 89
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
90 91 92 93 94 95 96 97 98 99
■ ■ ■ ■ □ □ ■ ■ ■ ■
100 101 102 103 104 105 106 107 108 109 110 111
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
112 113 114 115 116 117 118 119 120 121
■ ■ ■ ■ □ □ ■ ■ ■ ■
122 123 124 125 126 127 128 129 130 131 132 133
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
134
ETB
App.1-40
Appendix.1 AE-4020 SERVICE MANUAL
1
STX
2 3 4 5 6 7 8 9 10 11
■ ■ ■ ■ □ □ ■ ■ ■ ■
12 13 14 15 16 17 18 19 20 21 22 23
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
24 25 26 27 28 29 30 31 32 33
■ ■ ■ ■ □ □ ■ ■ ■ ■
34 35 36 37 38 39 40 41 42 43 44 45
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
46 47 48 49 50 51 52 53 54 55
■ ■ ■ ■ □ □ ■ ■ ■ ■
56 57 58 59 60 61 62 63 64 65 66 67
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
68 69 70 71 72 73 74 75 76 77
■ ■ ■ ■ □ □ ■ ■ ■ ■
78 79 80 81 82 83 84 85 86 87 88 89
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
90 91 92 93 94 95 96 97 98 99
■ ■ ■ ■ □ □ ■ ■ ■ ■
100 101 102 103 104 105 106 107 108 109 110 111
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
112 113 114 115 116 117 118 119 120 121
■ ■ ■ ■ □ □ ■ ■ ■ ■
122 123 124 125 126 127 128 129 130 131 132 133
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
134
ETB
App.1-41
Appendix.1 AE-4020 SERVICE MANUAL
STX
2 3 4 5 6 7 8 9 10 11 12
□ □ □ □ □ □ □ □ □ □ □
13 14 15 16 17 18 19 20 21 22 23
C O M . ■ ■ ■ ■ ■ ■ ■
24 25 26 27 28 29 30 31 32 33 34
■ ■ ■ ■ ■ ■ ■ ■ ■ ■ ■
35
ETX
a) “Enable”or “
Disable”can be selected in the setting. A blank is output if “
Disable”is selected.
b) REFERENCE : 7-7
App.1-42
Appendix.2 AE-4020 SERVICE MANUAL
App 2
App.2-1
Appendix.2 AE-4020 SERVICE MANUAL
Revision History
Revision Date of Revision Revised Item Description / Date of Execution
Number
A 2004/02/03 Error/Trouble Number only. Comment deleted.
B 2004/03/19 Print format Measurement result format changed.
Serial number changed into 8-digit
format.
C 2004/07/16 System setting format Setting for elimination of influence of
creatinine upon albumin added.
Optical system right-standby positition
adjustment setting deleted
EEPROM cleared. Default setting
User setting added.
Start with START key deleted from
inlet operation mode.
D 2004/09/21 System setting format DIP detection threshold input changed
Print format into service setting.
JCCLS-compliant format added.
Chinese character print format
modified.
500 nm deleted.
E 2004/12/17 Fine print of check Color paper measurement result added.
sample
App.2-2
Appendix.2 AE-4020 SERVICE MANUAL
CONTENTS
1. Overview ..........................................................................................................................................4
4. Print Format.....................................................................................................................................5
App.2-3
Appendix.2 AE-4020 SERVICE MANUAL
1. Overview
This document describes the print specifications for AE-4020.
2. Printer Specifications
The table below shows the overview of the printer used in this device. For details, refer to attached documents.
Item Specifications
[1] Manufacturer Seiko Instruments Inc.
[2] Product name (Product number) Thermal Line Printer (LTPH8235)
[3] Paper width (Effective print width) 58 mm (48 mm)
[4] Total number of dots 288
[5] Number of print characters Standard font: 32 characters
[6] Maximum printing speed 200 dots/s, 40 mm/s
[7] I/O system Parallel I/O
3. Details of Print
3.1.Type of Print
There are 9 types of print as shown below. For every type, the print layout and print item are described and print
timing is shown.
y User Mode
yMeasurement result print
yTrouble message print
yCheck sample print
yList print
yUser setting information print (Parameter print)
y Service Mode
ySystem setting information print (Parameter print)
yTest strip information print (Parameter print)
yReagent item information print (Parameter print)
yColor tone setting information print (Parameter print)
App.2-4
Appendix.2 AE-4020 SERVICE MANUAL
4. Print Format
4.1.Measurement Result
The measurement results contains a normal sample, STAT sample and control sample.
App.2-5
Appendix.2 AE-4020 SERVICE MANUAL
[3]Date (a), time (b), test strip type (c) and device temperature (d)
a, b) Date (10 one-byte characters), time (5 one-byte characters)
Date and time of measurement. Three print formats are available, i.e., Year-Month-Day, Month-Day-Year and
Day-Month-Year.
Example 1996-05-21 09:53
21-05-1996 09:53
05-21-1996 09:53
c) Test strip type (7 one-byte characters)
Example 10EA
d) Device temperature (3 characters)
The device temperature is shown by a maximum of three figures and the unit is “°C”.
Example 38°C
App.2-6
Appendix.2 AE-4020 SERVICE MANUAL
[4]Item error mark (a), measurement item name (b), rank (C), concentration value or reflectance (D) and measurement
unit (e)
a) Item error mark (1 one-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
b) Measurement item name (4 one-byte characters)
GLU, PRO, BIL, URO, BLD, KET, NIT, LEU
C) Rank (Semi-quantitative symbol) (6 double-width characters, left-justified)
D) Concentration value or Reflectance (4 double-width characters, right-justified)
If a measured value is OVER, “OVER” is printed in double-width character.
The reflectance is indicated at the same place as the measurement unit.
e) Measurement unit (6 one-byte characters, right-justified)
For the reflectance, “%” is printed as the measurement unit.
[5]pH error mark (a), pH measurement result (D), reflectance (D) and measurement unit (e)
If pH is not indicated in the reagent item, nothing is printed.
a) pH error mark (1 two-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
D) pH measurement result (3 double-width characters) or reflectance and measurement unit (right-justified 4
two-byte characters and 1 one-byte character for unit)
For the concentration value, the pH measurement result (4 double-width characters) is printed.
For the reflectance, the reflectance and measurement unit (%) are printed.
[6]Specific gravity error mark (a), specific gravity measurement result, reflectance and measurement unit (D)
If specific gravity is not indicated in the reagent item, nothing is printed.
a) Specific gravity error mark (1 two-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
D) Specific gravity measurement result (6 double-width characters) or reflectance and measurement unit
(right-justified 3 double-width characters and 1 one-byte character for unit)
For the concentration value, the specific gravity measurement result (6 double-width characters) is printed.
For the reflectance, the reflectance and measurement unit (%) are printed.
App.2-7
Appendix.2 AE-4020 SERVICE MANUAL
yPrint timing
Printed when the measurement of one sample has been completed or when reprint has been executed
in the User menu.
yPrint speed
It takes about seven seconds to print the measurement result.
App.2-8
Appendix.2 AE-4020 SERVICE MANUAL
App.2-9
Appendix.2 AE-4020 SERVICE MANUAL
[4]-1.Item error mark (a), measurement item name (b), rank (C), measurement result (D) and measurement unit (e)
a) Item error mark (1 one-byte character)
Normal: Blank
Error value : “*”
b) Measurement item name (4 one-byte characters, left-justified)
GLU and PRO
C) Rank (4 double-width characters, left-justified)
A rank (semi-quantitative symbol) is printed.
However, if the measured value is NORMAL, “NORMAL” is printed in double-width character. A format of
[4]-2.
D) Measurement result (4 double-width characters, right-justified)
If the measurement is OVER, “OVER” is printed in double-width character.
e) Measurement unit (6 one-byte characters, left-justified).
App.2-10
Appendix.2 AE-4020 SERVICE MANUAL
App.2-11
Appendix.2 AE-4020 SERVICE MANUAL
[3]Date (a), time (b), test strip type (c) and device temperature (d)
a) Date (10 one-byte characters), b) time (5 one-byte characters)
Date and time of measurement. Three print formats are available, i.e., Year-Month-Day, Month-Day-Year and
Day-Month-Year.
Example 1996-05-21 09:53
21-05-1996 09:53
05-21-1996 09:53
c) Test strip type (4 one-byte characters, left-justified)
Test strip type measured.
Example 10EA
d) Device temperature (2 one-byte characters, right-justified)
The device temperature is shown by a maximum of two figures and the unit is “°C”.
Example 38°C
[4]Item error mark (a), measurement item name (b), rank (C), measurement result (D) and measurement unit (e)
a) Item error mark (1 one-byte character)
Normal: Blank
Error value : “*”
b) Measurement item name (6 one-byte characters, left-justified: Chinese image data)
C) Rank (4 double-width characters, left-justified)
A rank (semi-quantitative symbol) is printed. However, if the measurement is NORMAL, “NORMAL” is
printed in double-width character.
D) Measurement result (4 double-width characters, right-justified)
If a measured value is OVER, “OVER” is printed in double-width character.
e) Measurement unit (6 one-byte characters, right-justified)
App.2-12
Appendix.2 AE-4020 SERVICE MANUAL
[6]Specific gravity error mark (a), specific gravity measurement result (D)
If specific gravity is not indicated in the reagent item, nothing is printed.
a) Specific gravity error mark (1 one-byte character)
Normal: Blank
Error value: “*”
D) Specific gravity measurement result (6 double-width characters, right-justified)
Printed for the concentration value.
App.2-13
Appendix.2 AE-4020 SERVICE MANUAL
4.2.Trouble Message
y Print format and print item
It is notified that an error has occurred in a device to stop the measurement operation and the following is printed.
[1] Date/Time (the same as measurement result)
yPrint timing
Printed after the measurement result is printed when a trouble has been detected.
App.2-14
Appendix.2 AE-4020 SERVICE MANUAL
4.3.Check Sample
4.3.1.Normal Printing of Check Sample
yPrint format and print item
yPrint timing
Printed when the check measurement has been completed.
App.2-15
Appendix.2 AE-4020 SERVICE MANUAL
App.2-16
Appendix.2 AE-4020 SERVICE MANUAL
App.2-17
Appendix.2 AE-4020 SERVICE MANUAL
[3]Item error mark (a), measurement item name (b), rank (C), concentration value or reflectance (D) and measurement
unit (e)
a) Item error mark (1 one-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
b) Measurement item name (4 one-byte characters)
GLU, PRO, BIL, URO, BLD, KET, NIT and LEU
C) Rank (semi-quantitative value) (4 double-width characters, left-justified)
However, “NORMAL” is printed in 6 two-byte characters using a field of concentration value.
D) Reflectance (6 double-width characters, right-justified)
e) Measurement unit (6 one-byte characters, right-justified)
“%” is printed.
If the reflectance is 100.0% or more, the first decimal place is rounded off to show 3 two-byte characters + % (1
one-byte character)
[4]pH error mark (a), pH measurement result (D), reflectance (D) and measurement unit (e)
If pH is not indicated in the reagent item, nothing is printed.
a) pH error mark (1 two-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
D) pH measurement result (3 double-width characters) or reflectance and measurement unit (right-justified 4
two-byte characters and 1 one-byte character for unit)
Reflectance and measurement unit (%) is printed.
[5]Specific gravity mark (a), specific gravity measurement result, reflectance and measurement unit (D)
If specific gravity is not indicated in the reagent item, nothing is printed.
a) Specific gravity mark (1 two-byte character)
Normal: Blank
Abnormal: “*”
Abnormal color: “!”
(“!” is shown first and then “*”.)
D) Specific gravity measurement result (6 double-width characters) or reflectance and measurement result
(right-justified 3 double-width characters and 1 one-byte character for unit)
Reflectance and measurement unit (%) is printed.
App.2-18
Appendix.2 AE-4020 SERVICE MANUAL
y Print timing
Print is executed when the check measurement has been completed. This print is performed only if the values of
Service Mode – Service Setting Information – Printer Setting – Check Sample Print are supported.
App.2-19
Appendix.2 AE-4020 SERVICE MANUAL
4.4.List Print
yPrint format and print item
[2]List print header (Time: Time of measurement result, Meas: Measurement type, No.: Measurement result number,
Mark: Error mark)
[4]Measurement result (a: hour, b: minute, c: measurement type, D: measurement number, E: error mark)
y Print timing
Printed when the List Print has been executed in the User menu.
App.2-20
Appendix.2 AE-4020 SERVICE MANUAL
[1]Date of print
[2]Date/Time of trouble
y Print timing
Printed when the Trouble List Print has been executed in the User menu.
App.2-21
Appendix.2 AE-4020 SERVICE MANUAL
[1]Version
[2]Date
App.2-22
Appendix.2 AE-4020 SERVICE MANUAL
[7]Printer setting
-1. Setting of printer ON/OFF
-2. Setting of number of printed sheets
-3. Number of line breaks setting
-4. Additional data setting “No. Only”: measurement number only, “+Condition”: measurement number + status
(date + test strip type + temperature),
”+ID#”: measurement number + status + ID number
-5. Selection of language of print item English: English, Chinese: Chinese * Note: Printed for the Chinese
version only.
[9]Barcode setting
-1. Setting of first reading characters
-2. Setting the number of effective reading characters
y Print timing
Printed when Parameter Print (Item No. 001) has been executed in “SET UP” (User setting).
App.2-23
Appendix.2 AE-4020 SERVICE MANUAL
Example of print
¦----+----1----+----2----+----3--¦
¦ AE-4020 V01.00 ¦
¦ 2003-10-01 13:45 ¦
¦ User Information Item No. ¦
¦ ¦
¦No.001 Information Print ¦
¦No.002 Strip Type ¦
¦No.003 Data Type ¦
¦ [0-1] 0:Concentration ¦
¦ 1:Reflection ¦
¦No.004 Strip Direction ¦
¦ [0-2] 0:Auto 1:Right ¦
¦ 2:Left ¦
¦No.005 Introduce Mode ¦
¦ [0-1] 0:Auto Start ¦
¦ 1:Cycle mode ¦
¦No.006 Buzzer ¦
¦ [0-1] 0:OFF 1:ON ¦
¦No.007 Abnormal Marking ¦
¦ [0-1] 0:OFF 1:ON ¦
¦No.008 Meas No. Reset ¦
¦ [0-1] 0:OFF 1:ON ¦
¦No.009 Printer ¦
¦ [0-1] 0:OFF 1:ON ¦
¦No.010 Copies ¦
¦ [1-3] (sheet) ¦
¦No.011 Line Feed ¦
¦ [0-9] (Line) ¦
¦No.012 Header ¦
¦ [0-2] 0:No. ¦
¦ 1:No.+Condition ¦
¦ 2:No.+Condition+ID#¦
¦No.013 On Line ¦
¦ [0-1] 0:OFF 1:ON ¦
¦No.014 Barcode ¦
¦ [1-32] Starting Digit ¦
¦ [1-13] Reading Digit ¦
*¦No.015 Language ¦
¦ [0-1] 0:English ¦
¦ 1:Chinese ¦
¦No.090 Trouble List Print ¦
¦No.099 Information Initialize ¦
¦- - - - - - - - - - - - - - - - ¦
y Print timing
Printed when the parameter item number print (Item No. 000) has been executed in “SETUP” (User setting).
App.2-24
Appendix.2 AE-4020 SERVICE MANUAL
App.2-25
Appendix.2 AE-4020 SERVICE MANUAL
1.Version
2.Date of printing
3.System setting information title
4.Serial number of device
5.Calibration curve setting
6.Setting of the number of measurement operations to be performed until the request to discard test strip is made.
7.Setting of the number of measurement operations to be performed until the request to discard surplus urine (inlet
table cleaning).
8.Date print format
9.Temperature correction function ON/OFF
10.Temperature print ON/OFF
11.Specific gravity correction function ON/OFF
12.Check measurement result format (normal [Normal], detail [Detail], A/D count [A/D ])
13.Automatic measurement stop function ON/OFF
14.ALB correction function ON/OFF
15.DIP detection function ON/OFF
16.Qualitative value display format setting
17.JCCLS print setting
18.Select ID# setting
19.Unused ID# setting
20.External output setting sub-title
-1. Setting of communication port
-2. Setting of communication format
-3. Setting of communication speed
-4. Setting of time gap between blocks
-5. Setting of specific gravity measurement result output position
-6. Setting of timeout 1
-7. Setting the number of retransmissions
-8. Setting of standby time 1
-9. Setting of standby time 2
21.Ethernet-related setting sub-title
-1. Setting of device ID
-2. Setting of IP address
-3. Setting of subnet mask
-4. Setting of default gateway
-5. Setting of ON/OFF of DHCP usage
22.Drive-related setting sub-title
-1. Setting of inlet standby position
-2. Setting of inlet suction position
-3. Setting of carrier standby position
-4. Setting of left standby position of photometric section
-5. Setting of right drive position of black mark set at the right
-6. Setting of left drive position of black mark set at the right
-7. Setting of right drive position of black mark set at the left
App.2-26
Appendix.2 AE-4020 SERVICE MANUAL
-8. Setting of left drive position of black mark set at the left
23.DIP detection sub-title
-1. DIP judgment threshold
24.ROM version
-1. Version of BOOT ROM
-2. Version of MAIN ROM
-3. Version of ETHERNET microcomputer
y Print timing
Enter the reverse mode in the service mode. Print is executed when [System Setting] or [All] has been selected in
Parameter Print (Item No. 100).
App.2-27
Appendix.2 AE-4020 SERVICE MANUAL
App.2-28
Appendix.2 AE-4020 SERVICE MANUAL
1.Device version
2.Date of print
3.System setting information title
4.Serial number
5.Setting of calibration curve
6.Setting of the number of measurement operations to be performed until a comment of request to discard test strip is
shown.
7.Setting of the number of measurement operations to be performed until a request to discard surplus urine (inlet table
cleaning) is made.
8.Date print format
9.Temperature correction function ON/OFF
10.Temperature print ON/OFF
11.Specific gravity correction function ON/OFF
12.Check measurement result format (normal, detail, A/D count)
13.Automatic measurement stop function ON/OFF
14.ALB correction function ON/OFF
15.DIP detection function ON/OFF
16.Qualitative value display format setting
17.JCCLS print setting
App.2-29
Appendix.2 AE-4020 SERVICE MANUAL
App.2-30
Appendix.2 AE-4020 SERVICE MANUAL
-7. Suction standby position for check measurement (difference from normal measurement)
27.Debug setting
yPrint timing
Enter the reverse mode in the service mode. Print is executed when [System Setting] or [All] has been selected in
Parameter Print (Item No. 100).
App.2-31
Appendix.2 AE-4020 SERVICE MANUAL
App.2-32
Appendix.2 AE-4020 SERVICE MANUAL
y Print timing
Enter the reverse mode. Print is executed when [Test Strip Information] or [All] has been selected in Parameter Print
(Item No. 100).
App.2-33
Appendix.2 AE-4020 SERVICE MANUAL
App.2-34
Appendix.2 AE-4020 SERVICE MANUAL
[1]Version name
[2]Date
[3]Item information title
[4]Item No. aa (2 one-byte characters)
[5]Item name aaaaa (5 one-byte characters, right-justified)
[6]Item unit aaaaaaa (7 one-byte characters, right-justified)
[7]Measurement wavelength aaa [nm] (3 one-byte characters)
[8]Reference wavelength aaa [nm] (3 one-byte characters)
[9]Error mark display rank aa (2 one-byte characters, right-justified)
[10]Semi-quantitative value, concentration value output position a (1 one-byte character)
[11]Temperature correction parameter sub-title
-1. Parameter A
-2. Parameter B
-3. Parameter C
-4. Parameter D
-5. Parameter E (mantissa part)
-6. Parameter E (exponent part)
-7. Parameter F (mantissa part)
-8. Parameter F (exponent part)
-9. Correction lower limit reflection
-10. Correction upper limit reflection
-11. Correction lower limit temperature
-12. Correction upper limit temperature
[12]Abnormal color detection parameter sub-title (a maximum of two conditional expressions can be set.)
-1. Abnormal color reflectance judgment expression [none, <, >, =<, =>, !=, =]
-2. Abnormal color detection wavelength 1
-3. Abnormal color detection wavelength 2
-4. Abnormal color threshold
-5. Abnormal color positive judgment threshold (reflectance)
App.2-35
Appendix.2 AE-4020 SERVICE MANUAL
y Print timing
Enter the reverse mode. Print is executed when [Test Strip Item Information] or [All] has been selected in Parameter
Print (Item No. 100).
App.2-36
Appendix.2 AE-4020 SERVICE MANUAL
[1]Version name
[2]Date
[3]Color control information title
[4]Color tone correction sub-title
-1. C/M value correction
-2. Y/M value correction
-3. C/r value correction
App.2-37
Appendix.2 AE-4020 SERVICE MANUAL
y Print timing
Enter the reverse mode. Print is executed when [Color Tone Information] or [All] has been selected in Parameter
Print (Item No. 100).
App.2-38
Appendix.2 AE-4020 SERVICE MANUAL
[1]Version name
[2]Date
[3]Serial number
[4]Drive-related setting sub-title
-1. Setting of inlet standby position
-2. Setting of inlet suction position
-3. Setting of carrier standby position
-4. Setting of left standby position of photometric section
-5. Setting of right drive position of black mark set at the right
-6. Setting of left drive position of black mark set at the right
-7. Setting of right drive position of black mark set at the left
-8. Setting of left drive position of black mark set at the left
[5]DIP detection sub-title
-1. DIP judgment threshold
[6]Color tone correction sub-title
-1. C/M value correction
-2. Y/M value correction
-3. C/r value correction
App.2-39
Appendix.2 AE-4020 SERVICE MANUAL
Example of print of service setting item Example of print of system setting item number
number list list
¦----+----1----+----2----+----3--¦ ¦----+----1----+----2----+----3--¦
¦ AE-4020 V01.00 ¦ ¦ AE-4020 V01.00 ¦
¦ 2003-10-01 13:45 ¦ ¦ 2003-10-01 13:45 ¦
¦ System Information No. ¦ ¦ System Information No. ¦
¦ for ¦ ¦ for ¦
¦ Service ¦ ¦ System ¦
¦ ¦ ¦ ¦
¦ Information Print <100> ¦ ¦ Information Print <100> ¦
¦ Information ItemNo. Print<110> ¦ ¦ Information ItemNo. Print<110> ¦
¦ Rank Select <120> ¦ ¦ Rank Select <120> ¦
¦ [01-XX] ¦ ¦ [01-XX] ¦
¦ Strip Dispose <130> ¦ ¦ Strip Dispose <130> ¦
¦ [0 or 100-200] ¦ ¦ [0 or 100-200] ¦
¦ Table Cleaning <131> ¦ ¦ Table Cleaning <131> ¦
¦ [0 or 200-400] ¦ ¦ [0 or 200-400] ¦
¦ Date Format Order <132> ¦ ¦ Date Format Order <132> ¦
¦ [0-2] ¦ ¦ [0-2] ¦
¦ Temp. Correct <133> ¦ ¦ Temp. Correct <133> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ Temp. Print <134> ¦ ¦ Temp. Print <134> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ S.G. Revise <135> ¦ ¦ S.G. Revise <135> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ CHECK Result <136> ¦ ¦ CHECK Result <136> ¦
¦ [0-2] ¦ ¦ [0-2] ¦
¦ Auto Meas Stop <137> ¦ ¦ Auto Meas Stop <137> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ ALB Revise <138> ¦ ¦ ALB Revise <138> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ Dip Miss Judge <139> ¦ ¦ Dip Miss Judge <139> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ Rank Mark <140> ¦ ¦ Rank Mark <140> ¦
¦ [0-2] 0:4+ 1:+4 2:++++ ¦ ¦ [0-2] 0:4+ 1:+4 2:++++ ¦
¦ JCCLS <141> ¦ ¦ JCCLS <141> ¦
¦ [0-1] 0:OFF 1:ON ¦ ¦ [0-1] 0:OFF 1:ON ¦
¦ ID# SELECT <142> ¦ ¦ ID# SELECT <142> ¦
¦ [0-1] 0:AE 1:AJ ¦ ¦ [0-1] 0:AE 1:AJ ¦
¦ Unused ID# <143> ¦ ¦ Unused ID# <143> ¦
¦ [0-1] 0:Use 1:Delete ¦ ¦ [0-1] 0:Use 1:Delete ¦
¦ Aux ¦ ¦ Aux ¦
¦ Port <300> ¦ ¦ Port <300> ¦
¦ [0-1] 0:RS232C 1:Ethernet ¦ ¦ [0-1] 0:RS232C 1:Ethernet ¦
¦ Format <301> ¦ ¦ Format <301> ¦
¦ [0-5] 0:AE-4020 1:AX-4280 ¦ ¦ [0-5] 0:AE-4020 1:AX-4280 ¦
¦ 2:Ticket 3:AM-4290 ¦ ¦ 2:TICKET 3:AM-4290 ¦
¦ 4:MA-4240 5:SA-4230 ¦ ¦ 4:MA-4240 5:SA-4230 ¦
App.2-40
Appendix.2 AE-4020 SERVICE MANUAL
App.2-41
Appendix.2 AE-4020 SERVICE MANUAL
y Print timing
Enter the reverse mode. Print is executed when the print of parameter item number list (Item No. 110) has been
performed.
App.2-42
Appendix.3 AE-4020 SERVICE MANUAL
App 3
App.3-1
Appendix.3 AE-4020 SERVICE MANUAL
Revision History
Revision Date of
Revised Item Description/Date of Execution
Number Revision
2003/11/27 New
A 2004/01/13 3. Trouble Change trouble numbers to
distinguish them from error
numbers
B 2004/08/20 General Modifying user corrective
procedure
C 2005/01/27 General Change notation
D 2005/03/12 T134 Low A/Ded value Change reference 1500 → 1650
T135 Excessive A/Ded Change reference 500 → 700
dark 2500 → 2700
White board A/D correction
E 2005/04/28 T137 Undetected black Add processing method
mark
App.3-2
Appendix.3 AE-4020 SERVICE MANUAL
1. Warning 4
1.1. Definition of warning 4
1.2. Upon occurrence of warning 4
1.3. Warning 5
1.3.1. W001 (Drift) 5
1.3.2. W002 (Reflectance is excessively high) 5
1.3.3. W003 (Incorrect positioning of test strip) 6
1.3.4. W004 (No DIP) 6
1.3.5. W005 (Wrong test strip type) 7
2. ERROR .................................................................................................................................................. 8
2.1. Definition of error 8
2.2. Upon occurrence of error 8
2.2.1. E001 (Power-down) 9
2.2.2. E002 (Abnormal backup voltage) 9
2.2.3. E004 (Printer paper out) 10
2.2.4. E005 (Waste box full) 10
2.2.5. E006 (Surplus urine full) 11
2.2.6. E007 (No applicable data) 11
2.2.7. E008 (Auto start sensor error) 12
2.2.8. E009 (No test strip on transport section) 12
3. TROUBLE ........................................................................................................................................... 13
3.1. Definition of trouble 13
3.2. Upon occurrence of trouble 13
3.2.1. T101 (EEPROM error) 14
3.2.2. T102 (Version change) 14
3.2.3. T110 (No analytical curve) 15
3.2.4. T120 (Inlet error) 16
3.2.5. T121 (Transport section error) 17
3.2.6. T123 (Incoming test strip detection error) 18
3.2.7. T130 (Photometric section initialization error) 19
3.2.8. T131 (Photometric section operation error) 20
3.2.9. T132 (A/D overflow) 21
3.2.10. T133 (A/D overrange) 22
3.2.11. T134 (Low A/Ded value) 23
3.2.12. T135 (Excessive A/Ded dark) 24
3.2.13. T137 (Undetected black mark) 25
3.2.14. T138 (Test strip feed error) 26
3.2.15. T160 (Initialization failure) 27
3.2.16. T170 (External output initialization failure) 28
3.2.17. T171 (Two-way communication error) 29
App.3-3
Appendix.3 AE-4020 SERVICE MANUAL
1. Warning
1.1. Definition of warning
Warning indicates that the result of measurement is not reliable although no operational problem occurred
during measurement.
App.3-4
Appendix.3 AE-4020 SERVICE MANUAL
1.3. Warning
1.3.1. W001 (Drift)
Description y Difference between Max. and Min. dark counts exceeds the threshold.
y Difference was determined at the end of photometry of a sample.
Possible cause y Disturbing light entered through front maintenance cover.
y Disturbing light entered through slit of waste box.
y Error or fault, e.g. excessive offset, in electrical circuit (detector and associated
circuitry).
Action taken by y Install waste box in place.
user y Change orientation of the device.
Action taken by y Check photometric section (optical system).
manufacturer y In the reverse mode “420”, check A/Ded dark value.
Process y Includes W001 on printout. (Does not print measurement result.)
y Continues measurement.
Display None
Printing
| ******************************** |
| COM: W001 |
| ******************************** |
Remarks y Threshold (dark max. − dark min.): 50 counts
App.3-5
Appendix.3 AE-4020 SERVICE MANUAL
App.3-6
Appendix.3 AE-4020 SERVICE MANUAL
App.3-7
Appendix.3 AE-4020 SERVICE MANUAL
2. Error
2.1. Definition of error
An error message indicates that the device must be stopped because the user misuses the device or operates it
in a wrong wary.
E001 Ent
ENTER key releases the error and returns to the standby screen.
App.3-8
Appendix.3 AE-4020 SERVICE MANUAL
Printing None
Remarks y Definition of “during measurement”: time span from the start to the end of measuring
operation.
y The previous measurement is made invalid (valid if measurement result has been
output).
y Error message remains displayed until [ENTER] key is pressed.
Printing None
Remarks y Error message remains displayed until [ENTER] key is pressed.
App.3-9
Appendix.3 AE-4020 SERVICE MANUAL
Printing None
Remarks y Error message remains displayed until [ENTER] key is pressed.
y Once the paper out occurs during measurement, the screen will not return to the
standby screen until the paper is supplied and remaining measurement records are
printed.
y Power supply pins:
J2 1–2 pins + 5 V
3–4 pins + 5 V
5–6 pins + 5 V
E005 [1]
Printing None
Remarks y Displayed error message will be replaced with [1] upon pressing [ENTER] key. Press
[1] key to return to standby screen.
y When waste box becomes full during measurement, FULL icon blinks, prompting
removal of test strips.
App.3-10
Appendix.3 AE-4020 SERVICE MANUAL
E005 OFF
Printing None
Remarks y Displayed error message will be replaced with [OFF] upon pressing [ENTER]. While
[OFF] is displayed on the screen, no subsequent operation is accepted.
y When surplus urine full condition is detected during measurement, FULL icon blinks,
prompting removal of surplus urine.
Printing None
Remarks y Error message remains displayed until [ENTER] key is pressed.
App.3-11
Appendix.3 AE-4020 SERVICE MANUAL
Printing None
Remarks y Error message remains displayed until [ENTER] key is pressed.
Printing None
Remarks y Error message remains displayed until [ENTER] key is pressed.
y PI detection level: 100±5 counts without test strip
900 counts or more with test strip placed
App.3-12
Appendix.3 AE-4020 SERVICE MANUAL
3. Trouble
3.1. Definition of trouble
Trouble is defined as a problem created in AE-4020, causing itself to halt.
T121 Ent
Pressing ENTER key will cancel the trouble and go to system initialization comfirmation screen.
T121 [1]
App.3-13
Appendix.3 AE-4020 SERVICE MANUAL
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T101 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Even if the problem is removed, EEPROM is not initialized. Initialization will be
performed in service mode.
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T102 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
y Pressing [ENTER] clears data contained in the RAM and starts system initialization.
App.3-14
Appendix.3 AE-4020 SERVICE MANUAL
T110 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T110 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-15
Appendix.3 AE-4020 SERVICE MANUAL
T120 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T120 ¦
¦ ------------------------------- ¦
Remarks y Type of trouble detected
1. Travel time of the introduction lever from waiting position PI to surplus urine
suction position PI becomes longer, 1 sec minimum.
2. Introduction lever has moved from surplus urine suction position to transport
waiting position, but the surplus urine suction position PI still “detects” the
introduction lever.
3. Travel time of the introduction lever from transport waiting position to surplus
urine suction position PI becomes longer, 360 ms minimum.
4. Travel time of the introduction lever from surplus urine suction position PI to
waiting position PI becomes longer, 1 sec minimum.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-16
Appendix.3 AE-4020 SERVICE MANUAL
T121 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T121 ¦
¦ ------------------------------- ¦
Remarks y Trouble detection
Time lag from start of transport operation to PI detection becomes longer: 2 sec or more.
Transport lever was not at waiting position at the beginning of photometry.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-17
Appendix.3 AE-4020 SERVICE MANUAL
T123 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T123 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
y PI detection level: 100±5 counts without test strip
900 counts or more with test strip placed
App.3-18
Appendix.3 AE-4020 SERVICE MANUAL
T130 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T130 ¦
¦ ------------------------------- ¦
Remarks y Trouble detection
Difference in the number of driving pulses for rightward movement and leftward
movement was 33 (approx. 1 mm) or more.
Movement toward the left waiting position by 144.8 mm did not cause PI detection.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-19
Appendix.3 AE-4020 SERVICE MANUAL
T131 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T131 ¦
¦ ------------------------------- ¦
Remarks y Trouble detection
Difference in the number of driving pulses for rightward movement and leftward
movement was 33 (approx. 1 mm) or more.
Movement toward the left waiting position by 144.8 mm did not cause PI detection.
Optical system was not at waiting position at the beginning of transporting sequence.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-20
Appendix.3 AE-4020 SERVICE MANUAL
T132 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T132 ¦
¦ ------------------------------- ¦
Remarks y Standard A/Ded value is 4,000 counts.
y White board A/D correction is 2,500 2,700±50 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-21
Appendix.3 AE-4020 SERVICE MANUAL
T133 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T133 ¦
¦ ------------------------------- ¦
Remarks y Standard A/Ded value is 4,000 counts.
y White board A/D correction is 2,500 2,700±50 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-22
Appendix.3 AE-4020 SERVICE MANUAL
T134 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T134 ¦
¦ ------------------------------- ¦
Remarks y Standard A/Ded value is 1,500 1,650 counts.
y White board A/D correction is 2,500 2,700±50 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-23
Appendix.3 AE-4020 SERVICE MANUAL
T135 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T135 ¦
¦ ------------------------------- ¦
Remarks y Standard A/Ded value is 500 700 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-24
Appendix.3 AE-4020 SERVICE MANUAL
T137 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T137 ¦
¦ ------------------------------- ¦
Remarks y White board A/D correction is 2,500 2,700±50 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-25
Appendix.3 AE-4020 SERVICE MANUAL
T138 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T138 ¦
¦ ------------------------------- ¦
Remarks y White board A/D correction is 2,500 2,700±50 counts.
y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-26
Appendix.3 AE-4020 SERVICE MANUAL
T160 [1]
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T160 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-27
Appendix.3 AE-4020 SERVICE MANUAL
OFF
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T170 ¦
¦ ------------------------------- ¦
Remarks
HOST connection has no relation to this problem.
y Trouble message is displayed. Press [ENTER]. Turn off power when “OFF” is
displayed.
App.3-28
Appendix.3 AE-4020 SERVICE MANUAL
Printing ¦ ¦
¦ 2005-01-20 16:17 ¦
¦ T171 ¦
¦ ------------------------------- ¦
Remarks y Trouble message will be displayed. Pressing [ENTER] will clear the problem.
Pressing [1] starts system initialization.
App.3-29
Appendix.4 AE-4020 SERVICE MANUAL
App 4
Program initialization
App.4-1
Appendix.4 AE-4020 SERVICE MANUAL
Initializing EEPROM
END
App.4-2
Appendix.4 AE-4020 SERVICE MANUAL
[LCD display]
AE-4020 01.00
Version No.
• If not the latest version, write the newest program by using the program writing software (from PC).
* For further information on operating method, refer to reverse mode operation manual.
* For further information on operating method, refer to reverse mode operation manual.
App.4-3
Appendix.6 AE-4020 SERVICE MANUAL
App 6
Ethernet setup
App.6-1
Appendix.6 AE-4020 SERVICE MANUAL
A seal is stuck.
Ethernet is set up in the reverse mode.The following setup is performed for a "reverse mode operation
procedure document" to reference.
① Communication port setup
② Equipment ID setup.
③ IP address setup
④ Subnet mask setup.
⑤ Default gateway setup.
⑥ DHCP setup
Completion
App.6-2
Appendix.7 AE-4020 SERVICE MANUAL
App 7
Packing
App.7-1
Appendix.7 AE-4020 SERVICE MANUAL
App.7 Packing
Clearing data
Fastening components
END
App.7-2
Appendix.7 AE-4020 SERVICE MANUAL
[2] [4]
[5]
[1] [3]
App.7-3
Appendix.8 AE-4020 SERVICE MANUAL
App.8-1