Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
Operator ’s Manual
Caution
・This analyzer system may cause unpredictable accident or injury, it if is improperly installed
or improperly operated. Therefore, please follow the instruction of operation manual strictly.
・Please read the【Safety caution】 certainly, in advance. You should not do the following action,
without our approval beforehand.
・If you conflict the above, you may loose our assurance and maintenance agreement does not
cover the trouble.
・For the safety operation of the analyzer, it is necessary to do daily maintenance by yourself,
but also the inspection of the whole analyzer by a qualified service engineer, is necessary at the
time of consumable replacement or at the time of proper interval to keep smooth operation.
・For the accident or the trouble, without above inspection, we are not responsible.
・After analyzer installation has finished, if position movement, transfer, resale or abolition is
necessary, please let us know. If the above action has done without presence of qualified
engineer, we will not responsible for the accident or trouble after that.
・The maintenance function parts, which are necessary to keep analyzer function, will be stored
for 7 years after the sales date. If you want to have these parts over 7 years, please consult to
us within storage period.
・The contents of this operation manual and software specification may change without notice.
・If you find any defects or mistakes in this manual, please let us know.
Registered trademark
Windows is a registered trademark in the United States of American Microsoft Corporation and
other countries.
CSA Standard
2 License
2.1 TMS permits to the customer a privilege to use the software, supplied by TMS, under
the condition of Software license agreement, antimonopoly and non-transferring basis.
Note A privilege to use the consented software, means that the privilege to install the
supplied software on a computer, having recommended operating system, and run the
software on the computer.
2.2 All the privileges which are not included in Software license agreement are reserved by
TMS.
2.3 Any customer cannot do the following action such as: let a third party use the consented
software, disclose, takeout to the other countries by means of setting the re-use rights
and other methods.
3 Copyright
3.1 Copyright, trademark right, and intellectual property right related with consented
software and operation manual belong to TMS.
3.2 Any customer cannot do the following action, such as transfer, sublease, disclosure,
related to the consented software and operation manual, without written permission of
TMS in beforehand. No customer can set security interest.
3.3 Any customer cannot do the following action, such as partial collection, modification,
reverse engineering, reverse compile, reverse assemble for the consented software. If a
customer did the above and any trouble occurred on the analyzer, based on the
modification, TMS has no responsibility.
4 Reproduction prohibition
Any customer cannot copy the consented software, even though partly.
5.2 TMS has no responsibility on the loss of customer or the third party, caused by the
V5.7
6 Forbidden items
Customer should not do the following action intentionally.
6.1 Transmitting harmful computer program, such as virus, to TMS or the third party.
6.3 The action inducing or promoting criminal acts and leading to criminal acts.
6.4 The action infringing copyright, trademark right, portrait right and other right and
profit of the third party.
6.5 The action of slander TMS or the third party and the action staining honor.
6.8 Other than the above items, any action which disturbs operation of TMS.
7 Effective period
7.1 This contract is active from the date of agreement, and active till failure of the contract,
by the procedure explained at the next article.
7.2 In spite of the above, the third article is active after the failure of the contract.
8.2 The customer can cancel this agreement by the abandonment of the consented software.
10 Conference
10.1 Regarding the subjects which are not covered by the agreement, customer and TMS
will discuss and reach to the conclusion by the basis of principle of good health.
10.2 If you have any question regarding the agreement, please contact to TMS by a
document.
Fast Detergent 2 5D Acid Washing Solution (CODE No. 50-24-0748), 2013/11/26 1/4
2. Hazards Identification
GHS classification and label elements of the products
GHS classification
HEALTH HAZARDS
Skin corrosion/irritation : Category 2
Eye damage /eye irritation : Category 2A
Label elements
PRECAUTIONARY STATEMENT
Prevention
P264 Wash contaminated parts thoroughly after handling.
P280 Wear protective gloves , eye protection/face protection.
Response
P302 + P352 IF ON SKIN: Wash with plenty of soap and water.
P332 + P313 If skin irritation occurs: Get medical advice/attention.
P362 + P364 Take off contaminated clothing and wash it before reuse.
P305 + P351 + P338 IF IN EYES: Rinse cautiously with water for several minutes. Remove
contact lenses, if present and easy to do. Continue rinsing.
P337 + P313 If eye irritation persists: Get medical advice/attention.
3. Composition/Information on Ingredients
Substance/Preparation: Preparation
General name : Nonionic surfactant、Organic acid (salt)、Metal corrosion inhibitor、
Antiseptic agent、Water
4. First-aid Measures
If inhaled
Move the victim to a place full of fresh air, rinse inside the mouth with water sufficiently and
keep at rest in a position comfortable for breathing. Get medical treatment, if necessary.
If on skin
Wash the affected area with a lot of cold water. Get medical treatment, if necessary.
If in eyes
If even a small amount of the product gets into eyes, immediately wash the affected eye by
flushing a lot of water for 15 minutes or more without rubbing (carefully wash eyeball and
eyelid completely). Get medical treatment by an oculist. In case you wear contact lens and can
take it off easily, you take if off and continue washing eyes.
Fast Detergent 2 5D Acid Washing Solution (CODE No. 50-24-0748), 2013/11/26 2/4
If swallowed
Rinse inside the mouth with water sufficiently. If the patient is conscious, let the patient drink
fresh water or milk. Never force the affected person to vomit. Immediately get medical
treatment.
5. Fire-Fighting Measures
Suitable extinguishing media
This product is non-flammable. In case of a fire around the container, as a fire - extinguishing
agent that uses the water, dry-chemical powder or carbon dioxide.
Specific fire-fighting measures
Evacuate non-essential personnel to a safe area. Move container from fire area if it can be done
without risk
Special protective equipment and precautions for firefighters
Firefighters should wear proper protective equipment. Be careful not to inhale smoke and
must work on the windward side of the fire.
2. H azar ds I dentification
G H S classification and label elements of the products
G H S classification
HE A L T H HA ZA R DS
S kin corrosion/irritation : C ategory 2
E ye damage /eye irritation : C ategory 2
E N V I R ON M E N T H A Z A R D S
H azardous to the aquatic env ironment-acute tox icity : C ategory 3
L abel elements
PR E C A U T IO N A R Y S T A T E M E N T
Prev ention
P273 A v oid release to the env ironment.
P264 W ash contaminated parts thoroughly after handling.
P280 W ear protectiv e glov es, eye protection/face protection.
R esponse
P302 + P352 I F ON S K I N : W ash w ith plenty of soap and w ater.
P332 + P313 If skin irritation occurs: G et medical adv ice/attention.
P362 + P364 T ake off contaminated clothing and w ash it before reuse.
P305 + P351 + P338 I F I N E Y E S : R inse cautiously w ith w ater for sev eral minutes. R emov e
contact lenses, if present and easy to do. C ontinue rinsing.
P337 + P313 If eye irritation persists: G et medical adv ice/attention.
D isposal
P501 D ispose of contents/container in accordance w ith local/national regulation.
4. F ir st-aid M easu r es
I f inhaled
M ov e the v ictim to a place full of fresh air, rinse inside the mouth w ith w ater sufficiently and
keep at rest in a position comfortable for breathing. G et medical treatment, if necessary.
I f on skin (or hair)
Wash the affected area w ith a lot of cold w ater. G et medical treatment, if necessary.
F ast D etergent 1 2.5D A lkaline W ashing S olution (C OD E No. 50-24-0747), 2013/11/25 2/4
I f in eyes
I f ev en a small amount of the product gets into eyes, immediately w ash the affected eye by
flushing a lot of w ater for 15 minutes or more w ithout rubbing (carefully w ash eyeball and
eyelid completely). G et medical treatment by an oculist. In case you w ear contact lens and can
take it off easily, you tak e if off and continue w ashing eyes.
I f sw allow ed
R inse inside the mouth w ith w ater sufficiently. If the patient is conscious, let the patient drink
fresh w ater or milk. Nev er force the affected person to v omit. Immediately get medical
treatment.
5. Fire-Fighting Measures
S uitable ex tinguishing media
T his product is non-flammable. I n case of a fire around the container, as a fire - ex tinguishing
agent that uses the w ater, dry-chemical pow der or carbon diox ide.
S pecific fire-fighting measures
E v acuate non-essential personnel to a safe area. M ov e container from fire area if it can be done
w ithout risk
S pecial protectiv e equipment and precautions for firefighters
F irefighters should w ear proper protectiv e equipment. B e careful not to inhale smoke and
must w ork on the w indw ard side of the fire.
I nternational rule
United Nation’ s classification /United Nation’ s number : Not applicable
IATA Dangerous Goods Regulations : Not restricted
Preface
Preface
Alarm code, guarantee, safety and caution for operation
Chapter1 [Outline]
Chapter5 [Maintenance]
Parts List
Periodic Check List
0-1
V5.7
0-2
Operator’s Manual Preface
In this manual, the items you have to be careful for safety are indicated by the following
levels.
Classification of display
Caution :If ignored, the accident may cause personal injury or heavy damage on the
instrument.
Warning :If ignored, the accident may lead to death or heavy injury.
:Biohazard.
There is the risk of your being infected.
In case of treating samples that may be hazardous or infectious, wear rubber
gloves not to touch directly.
Any spills on the skin should be washed off with a great volume of water and
sterilized.
And consult a physician if necessary.
Also any spills on the system should be wiped off and sterilized at once.
:This mark means laser light leakage. Do not see by naked eye.
0-3
V5.7
Connection, Biohazard
If you operate the dangerous parts, please wear
glove.
Location: Water supply and drainage panel
0-4
Operator’s Manual Preface
0-5
V5.7
Biohazard
If you operate the dangerous parts, please wear glove.
Location: Sample tray, Reagent tray
0-6
Operator’s Manual Preface
Biohazard
If you operate the dangerous parts, please wear glove.
Location: Reaction waste tank
0-7
V5.7
There are some warning labels. If you operate those parts, do not forget turn off
the main breaker.
Location: STM, RTM1, RTM2, MU1, MU2, CWS, DP, STM for ISE, STF, RGT,
CLB
0-8
Operator’s Manual Preface
Dangerous area
0-9
V5.7
0-10
Operator’s Manual Preface
Red frame shows dangerous moving parts. Never touch while running.
1. STF
2. STM
3. RGT
4. RTM-1
0-11
V5.7
5. RTM-2
6. MU-1
7. MU-2
8. CWS
0-12
Operator’s Manual Preface
1. Lamp house
2. Trans
0-13
V5.7
0-14
Operator’s Manual Preface
Caution
Warning
0-15
V5.7
(10) Grounding
・ Securely connect PE conductor of power supply cord to protective earth
ground of building installation. The resistance to ground should be less
than 10 .
・ This analyzer corresponds to Equipment Class 1.
・ Customer should not change power supply code by yourself.
If you find a scratch or cut, please let us know.
0-16
Operator’s Manual Preface
(14) Maintenance
・ When do maintenance follow this operation manual, and never touch other than
designated part.
・ When do maintenance operation, wear protective globe to avoid infection.
・ When you need to replace parts or units, please use parts and units supplied from us.
The maintenance operation done by a common operator are limited in the range of
operation manual.
・ Do not disassemble the analyzer.
・ For the electric shock protection, operator should not open front cover, back cover, left
and right covers. Authorized engineer, by TBMS, can open covers.
・ When analyzer is running, never open top cover. If an operator touch one of moving
parts, may cause trouble.
・ Before using an attachment, You have to read an operation manual of the attachment.
(17) Tube
・ When installation or liquid handling is done, watch tubes are not collapsed.
0-17
V5.7
(20) Emergency
・ For emergency stop, turn off main breaker at the back panel.
・ To do the above smoothly, keep the space at the back.
(21) Humidifier
・ Do not use a humidifier, an ultrasonic humidifier, or a hybrid type humidifier.
It might cause white turbidity outside of the light passing part on the reaction cell.
0-18
Operator’s Manual Preface
3. Warranty
0-19
V5.7
12. A defect or nonconformance that you found or could have found at the
time of receipt of the analyzer such as loss of or apparent damage to the
analyzer and that you failed to promptly notify to us in writing;
13. Minor defect or nonconformance that does not affect operation, function
and performance of the analyzer; and
14. Any other defect or nonconformance that is not attributable to us.
- The warrantees of consumables and spare parts of the analyzer are
to be accordance with its manufacturer's warranty.
- In case any defect or nonconformance is found in the analyzer, both
you and we will, at its own costs respectively, cooperate to
promptly determine the cause of the same, whereby you will, upon
our request, provide us with information in writing that is
necessary to verify such cause.
- If we perform any remedy for a defect or nonconformance of the
analyzer, such as repair and replacement of the analyzer and
payment of compensation money (including offer of price
discount),and subsequently it is found that any one of the events
contained in the foregoing (3) applies and that we are not
responsible for such defect or nonconformance, then you will,
within fourteen days after receipt of our invoice, reimburse us with
all the costs we have incurred associated with the remedy
(including costs associated with the investigation and dispatch of
our personnel) and refund the full amount of the compensation
money we have paid to you. Delay in reimbursement or refund is
subject to the five percent monthly compound overdue interest or
the maximum interest permitted by applicable laws from the due
date till paid in full.
0-20
Operator’s Manual Preface
5. Environment condition
0-21
V5.7
This instrument is very heavy. If you want to transport the instrument, you have to pack
by exclusive transportation box and transport by forklift. Please be sure to insert a fork to
the following red circle.
*Our company cannot take responsibility to occur because of packing and transporting by
you.
*If you want to pack and transport the instrument, please contact our service division.
Four wheels are attached to this instrument. If you push the instrument, you can move
easily. Please use in the case of movement of a short distance.
<Wheels>
And two stoppers are also attached to under the front panel of the instrument. If you use
stopper, you can fix the instrument. Stopper is fixed by turning.
<Stopper>
0-22
Operator’s Manual Preface
For the cleaning of analyzer, wear rubber globe and wipe by a cloth with neutral detergent
solution.
Do not use organic solvent. To remove biohazard contamination, paper waste by 70%
ethanol solution and paper waste by water thoroughly.
If you use ethanol for sterilization, be careful not to burn. Keep away fire source.
0-23
Operator’s Manual Table of Contents
Table of Contents
Chapter 1
Outline
Contents
T-1
V5.7
Chapter 2
Contents
T-2
Operator’s Manual Table of Contents
T-3
V5.7
Chapter 3
Analysis Operation
Contents
T-4
Operator’s Manual Table of Contents
T-5
V5.7
T-6
Operator’s Manual Table of Contents
Chapter 4
Contents
T-7
V5.7
T-8
Operator’s Manual Table of Contents
T-9
V5.7
Chapter 5
Maintenance
Contents
T-10
Operator’s Manual Table of Contents
T-11
V5.7
Chapter 6
Contents
Parts List
Periodic Check List
T-12
Operator’s Manual Chapter 1
Chapter 1 Outline
1-1
V5.7
1-2
Operator’s Manual Chapter 1
1.1 Specifications
Item Contents
Discrete, Single line random access, multi-tests
Method
analysis
480 tests/hr.
Throughput ・MAX 580tests/hr. (in case of loading ISE)
・160test/hr. (HbA1c Item Only)
End point method, Rate method
Analysis method
ISE
Linear, Factor, Non-linear( Logit-log, Spline,
Calibration curve
Exponential, Polynomial)
About 8.5 min.
Reaction time
(1st react. about 3.5 min+ 2nd react. about 5 min.)
Tray mode Software tray
Patient sample tray : Tray No. 1 - 900
Tray kind
Calibration tray : No.901,902
Serum, Plasma, Blood cell ,Urine, Dialysate, CSF
Contents of test
(ISE not available for CFS and Blood cell)
ISE washing :1
Dispensation
1.0 – 20.0 L (0.1 L steps)
volume
13 types, with pre-dilution
Dilution ratio
3,5,6,10,20,30,40,50,60,70,80,90,100 times
1-3
V5.7
Item Contents
Number of R1;36position
bottles on board R2;35position
Reagents are cooled at 8 – 12 Celsius degrees.
Reagent (If the room temperature is more than 25 ℃ , the
Storage
maximum cooling will be -13 ℃ from the room
temperature.)
R1;70mL
Bottle volume
R2;30mL
Remain check Liquid surface sensor or countdown
Mixing method By stirrers
Temperature
Direct block heating
control
Reaction
37 plus-minus 0.3Celsius degrees
temperature
Cell skip
Available
function
Reaction
Material:Plastics (Semi-disposable)
tray
Optical path length : 5 mm
Number of cuvette : 90 pieces (15 pieces x 6)
Reaction cuvette
Reaction volume : 100 – 300 micro-liter
Cell Wash: Dilution of stock detergent solutions
(alkaline, acidic) with water in the analyzer
Optical
Cuvette direct measurement
absorption
Fixed 12wavelengths
Wavelengths 340,380,405,450,505,546,570,600,660,700,750,
800nm
Spectropho
tometer Light detector Silicon photo-diode
Wavelength
Plus-minus 2nm
accuracy
Linearity range 0 – 3.0Abs.
Light source 12V 20W Halogen lamp
Operation Personal computer (available for touch panel)
User
Calculation Calculation based on user’s formula
interface
between item Correlation
1-4
Operator’s Manual Chapter 1
Item Contents
Reaction curve monitor
Monitor function
Reagent residual quantity, Run monitor
Option
Item Contents
1-5
V5.7
Item Contents
Calibration of
(Default condition)
serum
Automatic calibration is conducted once every 24 hours.
specimen
(Default condition)
Liquid feed
Solution (Calibrator 1) is automatically sent once every 2 hours.
Serum
Plasma 100 L
Sample volume
Dialysis
:Measurement
300 L
Urine
(Use sextuple diluted sample by urine diluent)
Serum
Plasma 45 seconds
Analysis time Dialysis
Urine 60 seconds
1-6
Operator’s Manual Chapter 1
Top view
1-7
V5.7
This figure shows a plan of the system when the top cover is removed.
1-8
Operator’s Manual Chapter 1
(Stop button)
If you push this button, all the functions stop.
1-9
V5.7
(Printer port)
To Printer (*option) D-SUB 9-pin male
connector for the printer
(Drainage connector)
1-10
Operator’s Manual Chapter 1
1.2.6 Tanks
Cable
Float Switch
Reaction Waste
Liquid Reservoir
Signal cable
Float switch
1-11
V5.7
Front view
1-12
Operator’s Manual Chapter 1
The analyzer and the peripheral devices (PC, printer, deionized water container and
waste line) are connected as shown in the following figure. Please ensure the correct
connection.
1. Connection diagram
Purified
water line
PC
Water
purification
Connection diagram
system
2. Interface (PC)
This is the connection diagram of Interface (Personal computers,
printer)
When you connect the instrument with PC or PC with printer, you have
to use LAN cable.
Personal computer specification, Please refer to Chapter 1 1.2.10
Personal computer.
PC Printer
LAN cable RS232C Straight
LAN port (Option)
3. Interface (Tubes)
Installation of water purification system in the analyzer differs from the
one out of the analyzer. Please refer to Setup manual about the detailed
connection method.
1-13
V5.7
A B C ② ③ ④ ⑤
PCB box
① TCD PCB
② SCU PCB
③ MCU PCB
④ UCN1 PCB
⑤ UCN2 PCB
B, C. Temperature controller
This unit controls temperature in the reagent tray at 8 – 12℃ .
1-14
Operator’s Manual Chapter 1
1-15
Operator’s Manual Chapter 2
Chapter 2
Functions and principles
2-1
V5.7
4. Exponential.............................................................................................................. 2-38
5. Polynomial ............................................................................................................... 2-39
2.6 ISE module .................................................................................................................. 2-40
2.6.1 Calibration ........................................................................................................... 2-40
2.6.2 Concentration calculation .................................................................................... 2-40
2.7 Linearity check function ............................................................................................ 2-41
2.8 Prozone check function................................................................................................ 2-42
2.9 Serum information (L-H-I) .......................................................................................... 2-43
2.10 Carryover avoidance.................................................................................................... 2-44
2.10.1 Kinds of carryover avoidances .............................................................................. 2-44
2.11 Automatic start-up, shut down.................................................................................... 2-45
2.12 Automatic maintenance .............................................................................................. 2-45
2-2
Operator’s Manual Chapter 2
2-3
V5.7
START
CUVETTE CLEANING
EVACUATION
R1 DISPENSATION
7.5sec
SAMPLE DISPENSATION
MIXING
3.5min
1ST Reaction
MIXING
5 min
2ND Reaction
END OF MEASUREMENT
CUVETTE CLEANING
2.75 min
AUTO-STOP
2-4
Operator’s Manual Chapter 2
1. Analysis process
<Explanation>
S : Sample dispensation
R1 : R1 dispensation
R2 : R2 dispensation
M : Mixing
2-5
V5.7
1. Sampler (STF)
1) Function
A sample tray holds cups and tubes.
Sampler rotates a sample tray and transfers samples to the sampling
position.
2) Operation
(Initialization) Rotates clockwise. For patient sample, No. 1 sample
is brought to the sampling position.
(In operation) Rotates clockwise. Transfers samples to the sampling
position one by one.
2-6
Operator’s Manual Chapter 2
1) Function
Sample probe aspirates the designated volume of a sample from
position and the sample is dispensed into a reaction cuvette.
The sample probe has a level sensor.
2) Operation
(Initialization) : Sample probe once moves to a reaction cuvette,
then back to the probe washing pot position.
(In operation) : Sample probe moves up and down following (a), (b),
(c) steps. At the washing station, the inside and
outside of probe walls are washed and water
droplets are removed.
3) Sample dispensation
Sample probe is filled with washing water till tip.
When aspirating sample, the probe aspirates a small amount
of air, and then aspirates a little excess amount of sample,
and dispenses designated volume.
2-7
V5.7
Reagent bottle
70ml (R1)
20ml (R2)
1) Function
Reagent tray holds reagent bottles and transfers them to the aspiration
position.
When main power is ON, these bottles are kept cooled.
2) Operation
(Initialization) Rotates clockwise. No.1 reagent bottle is
brought to the reagent aspiration position “A”.
(In operation) Transfers selected reagent bottle to the aspiration
position “A” (1st reagent) or “B” (2nd reagent).
2-8
Operator’s Manual Chapter 2
The bottles for additional washing program are set between the washing
trough and the sample tray.
The position of washing solutions bottles for the sample probe
The bottles for washing solutions and dilution bottles of R1 and R2 are set in
the outside of the reagent tray.
Bottle position
No.1:Acid solutions
No.2:Alkaline solutions
No.3:Dilutions (Only for R1)
2-9
V5.7
R1 probe R2 probe
R1 pump R2 pump
1) Function
Reagent probe aspirates the designated volume of reagent from a reagent
bottle and RTM carries the reagent probe to the cuvette position and the
reagent is dispensed into a reaction cuvette.
Reagent probe has a level sensor.
There are two reagent probes. R1 probe (for the 1st reagent) and R2
probe (for the 2nd reagent) are at the wash positions when
initialized.
R1 and R2 reagent pumps are commonly used for R1 and R2.
When pre-dilution is ordered, the diluent is aspirated from the diluent
bottle and the diluent is dispensed into a cuvette by R1 probe.
2) Operation
(Initialization) Reagent probe stops at the home position of
washing probe side.
(In operation) Reagent probe moves up and down following (a),
(b),(c) steps.
At the washing pot, the inside and outside of probe
walls are washed and water droplets are removed.
2-10
Operator’s Manual Chapter 2
3) Reagent dispensation
Reagent probe is filled with washing water till tip.
When aspirating reagent, the probe aspirates a small amount of
air, and then aspirates a little excess amount of reagent and
dispenses designated volume.
2-11
V5.7
1) Function
Holds reaction cuvettes at 37 Celsius degrees, and lets sample and reagent
react.
Reaction cuvettes are optical cells to measure light absorption.
(Reaction cuvette direct optical measurement)
Reaction cell has become a semi-disposable.
Refer to” Chapter 4, 4.11.6 Analyzer maintenance screen / 3.Cell check”.
The exchange of the reaction cell, see Chapter 5, 5.2.4 Reaction cuvette.
2) Operation
(Initialization): Rotates and stops at the home position.
(In operation) :Rotates to measuring all cuvettes in every cycle.
NOTE
When you remove a reaction cuvette, be careful not to make scratch on
neither inside nor outside wall.
Do not touch the optical face with fingers.
2-12
Operator’s Manual Chapter 2
7. Mixing unit(MU-1, 2)
Mixing unit-1
Mixing unit-2
1) Function
Mixing unit-1 : This unit is used for mixing of 1st reagent and the
sample.
Mixing unit-2 : This unit is used for mixing after adding 2nd
reagent.
2) Operation
Mixing unit-1
(Initialization) : If it is in the down position, it goes up.
(In operation) : Repeats rotation.
Mixing unit-2
(Initialization) : If it is in the down position, it goes up.
(In operation) : Repeats rotation.
2-13
V5.7
1) Function
The reaction drainage after the test is removed, and the cuvette is washed
with cleaning solution and rinsed with pure water.
And the pure water is dispensed for water blank measurement.
Then the pure water is removed and the cuvette is dried up by a Teflon
sponge.
Washing water is heated to 37 Celsius degrees to get good cleaning efficiency.
2) Operation
(Initialization) If it is in the down position, it goes up.
(In operation) Repeats going up and down.
2-14
Operator’s Manual Chapter 2
Position Markers
1 - CWS Reaction liquid Aspiration
Upper ASP Washing liquid aspiration
2
Lower W1 Washing liquid dispense
Upper ALK Alkaline washing liquid aspiration
3
Lower CWSALK Alkaline washing liquid dispense
Upper ACID Acid washing liquid aspiration
4
Lower CWSACID Acid washing liquid dispense
Upper W2 Washing liquid aspiration
5
Lower W2 Washing liquid dispense
Upper W3 Washing liquid aspiration
6
Lower W3 Washing liquid dispense
Upper W4 Washing liquid aspiration
7
Lower W4 Washing liquid dispense
Upper W5 Washing liquid aspiration
8
Lower W5 Washing liquid dispense
9 - CWS Washing liquid aspiration
10 - DR2 Residual water droplet removal
11 - DR1 Residual water droplet removal
2-15
V5.7
9. Spectrophotometer
3) Function
Measures optical absorption of 12 wavelength of pure water and reaction liquid
in cuvettes while the reaction tray rotating.
2-16
Operator’s Manual Chapter 2
Printer
Paper lamp Feed switch
Control panel
2-17
V5.7
ISE module is used for measurement of ion concentration (Na+, K+, Cl+) in the
sample. ISE uses a dedicated probe.
2-18
Operator’s Manual Chapter 2
2. Measurement cycle
(2)Calibration of urine
①Cal- 1 in the electrode unit is discharged.
②Cal-2 for urine in the electrode unit is moved.
③The electro-motive force of Cal-2 for urine in the electrode unit is read.
④Cal-2 for urine in the electrode unit is discharged.
⑤The sample line and the electrode unit are washed with Cal-1.
⑥At the end of washing, the electro-motive force of Cal-1 in the electrode unit is
read.
⑦Cal- 1 in the electrode unit is discharged.
⑧Cal-1 for urine in the electrode unit is moved and discharged.
⑨Cal-1 for urine in the electrode unit is moved and the electrode unit is read.
⑩Cal- 1 for urine in the electrode unit is discharged.
⑪ The sample line and the electrode unit are washed with Cal-1.
⑫ At the end of washing, the electro-motive force of Cal-1 in the electrode
unit is read.
2-19
V5.7
(5)Measurement of urine
Use urine diluted with the urine diluent by six times.
2-20
Operator’s Manual Chapter 2
① ② ③ ④
Sample prove Mixing Sample prove
← →
Blood cell
Hemolysis
Centrifugal Aspirates the red Hemolysis the red Aspirates the hemolysis
separate sample blood cell sample blood cell by mixing. sample and dispenses to
and dispenses to cell 2.
hemolysis cell 1.
2-21
V5.7
When the sample probe detects the clot by fibrin, perform the washing
automatically. The error flag “C” is attached to related result.
(Sample probe)
Sample aspiration
Clot detection
(Automatic-cleaning )
Sample probe moves to the
trough and wash 2 times.
【Clot is heavy】
【Dissolve the clot】
Can’t continue the measurement
Restart the measurement from the
next Item. 【Sampling stop】
2-22
Operator’s Manual Chapter 2
END RATE
One point end assay One point rate assay
Two points end assay 1 Two points rate assay 1
Two points end assay 2 Two points rate assay 2
Calculate the absorbance difference of the Calculate the absorbance rate of two
two different intervals, sub- and main- different intervals, sub-interval and main-
intervals. interval, and calculate concentration or
activity from the difference between these
two rates.
Two points end assay 1 Two points rate assay 1
Measure the optical absorbance at the fixed time after mixing the reagent
2-23
V5.7
MAIN
ABS
S : Sample dispensation
R1 : 1st reagent dispensation
R2 : 2nd reagent dispensation
ABS : Optical absorbance
MAIN :Main-measurement interval
R1+S R2
One point end assay
3) Concentration calculation
Cx : Concentration of unknown sample
(Ax−Ab)
Ax : Absorbance of unknown sample
Cx=―――――― × Cs
Ab : Absorbance of reagent blank
(As− Ab)
As : Absorbance of standard sample
Cs :Concentration of standard Sample
2-24
Operator’s Manual Chapter 2
S : Sample dispensation
R1 : 1st reagent dispensation
R2 : 2nd reagent dispensation
ABS : Optical absorbance
MAIN : Main-measurement interval
SUB : Sub-measurement interval
3) Concentration calculation
(Ax − AsL)
Cx= ――――――― × (CsH# - CsL#) + CsL#
(AsH − AsL)
2-25
V5.7
3) Concentration calculation
(Ax − AsL)
Cx= ――――――― × (CsH# - CsL#) + CsL#
(AsH − AsL)
2-26
Operator’s Manual Chapter 2
R1+S R2
One point rate assay
3) Concentration calculation
Cx= (△Ax−△Ab) × F
2-27
V5.7
R1+S R2
Two points rate assay 1
3) Concentration calculation
Cx= (△Ax−△Ab) × F
2-28
Operator’s Manual Chapter 2
S : Sample dispensation
SUB MAIN
ABS R1 : 1st reagent dispensation
R2 : 2nd reagent dispensation
ABS : Optical absorbance
MAIN : Main-measurement interval
t
SUB : Sub-measurement interval
R1+S R2
Two points rate assay 2
3) Concentration calculation
Cx= (△Ax−△Ab) × F + Cb
2-29
V5.7
3. Factor 4-4.Exponential
2-30
Operator’s Manual Chapter 2
One standard sample and a blank sample make a linear calibration line.
2-31
V5.7
ABS : Absorbance
ABS Conc. : Concentration
Sn
Asn B : Reagent blank
S3 Ab : Absorbance of reagent blank
As3
S2 Cb : Concentration of reagent blank
As2
Ax Ax : Absorbance of unknown sample
As1 S1
Cx : Concentration of unknown
B
Ab sample
Cb Cs1 Cs2 Cs3 Csn S1-Sn : Standard sample
Cx Conc.
As1-Asn: Standard sample absorbance
Linear 2
Cs1-Csn : Standard sample concentration
(1) Ax ≦ As1
(Cs1−0)
Cx= (Ax− Ab)×―――――――
(As1− Ab)
(Cs2−Cs1)
Cx= (Ax− As1)×――――――― + Cs1
(As2− As1)
2-32
Operator’s Manual Chapter 2
(Cs3−Cs2)
Cx= (Ax− As2) ×――――――― + Cs2
(As3− As2)
(Cs4−Cs3)
Cx= (Ax− As3) ×――――――― + Cs3
(As4− As3)
(Cs5−Cs4)
Cx= (Ax− As4) ×――――――― + Cs4
(As5− As4)
(Cs6−Cs5)
Cx= (Ax− As5) ×――――――― + Cs5
(As6− As5)
2-33
V5.7
2.5.3 Factor
1) Factor input
Input the factor into the input frame of Factor dialog box.
4) Factor calculation
Tv 1 1
F = ――― × ――― × ―――――― × 106
Sv L (ε 1−ε 2)
2-34
Operator’s Manual Chapter 2
2.5.4 Non-linear
This logic makes the approximate curve using 2-6 standard samples
including a blank sample, as a calibration curve. There are the following
five types.
The results that exceed the absorbance of the highest concentration
standard with multi-point curve (not including Spline) are considered to
be not reliable.
1. Logit.log 1
This logic is used for the calibration curve in which the absorbance
converges to the increase of concentration.
ABS
Asn Sn ABS :Absorbance
S3
As3 Conc. : Concentration
As2 B : Reagent blank
S2
Ax
Ab : Absorbance of reagent blank
As1 S1 Cb :Concentration of reagent blank
Ab Ax : Absorbance of unknown sample
B
Cx : Concentration of unknown
Cb Cs1 Cs2 Cs3 Csn sample
Cx Conc. S1-Sn : Standard sample
Logit.log1 A1-An : Absorbance of standard sample
2-35
V5.7
2. Logit.log 2
This logic is used for the calibration curve in which the absorbance converges
to the increase of concentration.
As this logic uses more parameters than logit.log 1, better approximation
may be expected.
ABS : Absorbance
Conc. : Concentration
ABS
Asn S3 Sn B : Reagent blank
As3 Ab : Absorbance of reagent blank
As2 Cb : Concentration of reagent blank
S2
Ax
Ax : Absorbance of unknown sample
As1 S1 Cx : Concentration of unknown
Ab sample
B
S1-Sn : Standard sample
Cb Cs1 Cs2 Cs3 Csn A1-An : Absorbance of standard sample
Cx Conc. C1-Cn : Concentration of standard
Logit.log2 Sample
2) Concentration calculation
Cx : Concentration of unknown sample
Cx= (C+ Cb) Ax : Absorbance or absorbance rate of
reagent blank
Cb : Concentration of reagent blank
Ab : Absorbance or absorbance rate of
reagent blank
K, a, b, c : Constants
( Calculated automatically )
K
Ax= Ab+―――――――――――――
(1+ exp(-a-b×lnC-c×C))
(Ax−Ab)
0 = a+ b ×lnC+ c× C− ln ―――――――――
(K− (Ax− Ab))
2-36
Operator’s Manual Chapter 2
3. Spline
ABS : Absorbance
Conc. : Concentration
ABS
B : Reagent blank
Asn Ab : Absorbance of reagent blank
As3 Sn
As2 S3 Cb : Concentration of reagent
Ax S2 blank
As1 S1 Ax : Absorbance of unknown
sample
Ab B Cx : Concentration of unknown
Cb Cs1 Cs2 Cs3 Csn S1-Sn : Standard sample
Cx Conc. A1-An : Absorbance of standard
sample
Spline sample
C1-Cn : Concentration of standard
sample
2) Concentration calculation
C00x : Concentration of unknown sample
Ax : Absorbance or absorbance rate of
Cx= (C+ Cb)
unknown sample
Cb : Concentration of reagent blank
Ab : Absorbance or absorbance rate of
reagent blank
Ax= a(I)+b(I)・(C-C(J))+c(I)・(Cx-C(J))2+d(I)・(C-C(J))3
f(C-C(J))=a(I)+b(I)・(C-C(J))+c(I)・(C-C(J))2+d(I)・(C-C(J))3-Ax
2-37
V5.7
4. Exponential
For being different from 4-1, 4-2, 4-3 above, this logic is used for the
calibration curve in which the absorbance diverges to the increase of
concentration.
ABS : Absorbance
Conc. : Concentration
ABS B : Reagent blank
Asn Ab : Absorbance of reagent
As3 Sn
blank
As2 S3
Ax S2 Cb : Concentration of reagent
As1 blank
S1
Ax : Absorbance of unknown
Ab B sample
Cx : Concentration of unknown
Cb Cs1 Cs2 Cs3 Csn sample
Cx Conc.
S1-Sn : Standard sample
Exponential A1-An: Absorbance of standard
sample
C1-Cn: Concentration of standard
sample
2) Concentration calculation
Cx : Concentration of unknown
sample
Ax : Absorbance or absorbance
Cx = (C+ Cb)
rate of unknown sample
Cb : Concentration of reagent
blank
Ab : Absorbance or absorbance
rate of reagent blank
K, a, b, c : Constants
(Calculated automatically)
Ax-Ab
a ・ (lnC)+b・ (lnC)2+c・ (lnC)3 - ln ―――― =0
K
2-38
Operator’s Manual Chapter 2
5. Polynomial
This calibration curve is used for the case which optical absorbance
decreases as concentration increases.
ABS : Absorbance
ABS
B Conc. : Concentration
Ab S1 B : Reagent blank
As1
S2 Ab : Absorbance of reagent blank
As2
Ax Cb : Concentration of reagent blank
S3 Ax : Absorbance of unknown sample
As3 Sn
Asn Cx :Concentration of unknown sample
2) Concentration calculation
Cx : Concentration of unknown sample
Cx = (C+ Cb)
Ax : Absorbance or absorbance rate of
unknown sample
Cb : Concentration of reagent blank
Ab : Absorbance or absorbance rate of
reagent blank
K, a, b, c : Constants (Calculated automatically)
2-39
V5.7
ISE (ion selective electrode) measures electro-motive force (mV), induced by ion
concentration in the sample.
2.6.1 Calibration
2-40
Operator’s Manual Chapter 2
ABS
L1−L2
L1 : Main wavelength
L2 : Sub wavelength
△A
△A1 : The slope of measuring
△A1 range 1
△A2 : The slope of measuring
△A2
range 2
△A : The slope of the total curve
Time
Linearity check
1) Parameter setting
Input the numbers in the box of the linearity reference (%)
on Item Parameter screen, and set the check value.
※ The setting of Data Process Main should be more than 3
points.
3) Calculation Formula
2-41
V5.7
S : Sample dispensation
R1 : 1st reagent dispensation
R2 : 2nd reagent dispensation
ABS : Optical absorbance
FIRST : Interval (1)
SECOND: Interval (2)
* Use main wavelength
Prozone check
1) Prozone check value setting.
Input the Prozone check conditions in Item parameter setting column.
(1) First interval- Start, End : Starting and ending test points of basic
test interval.
Second interval-Start, End: Starting and Ending test points of second
interval.
*Input range is 9 – 76 point.
(2) Limit value: Input the comparison limit value(%).
(3) High, Low: Select the check high or low.
3) Calculation formula
Check value (%) = OD2/ OD1*100
OD1 : The gradient in the first interval
OD2 : The gradient in the second interval
(using a least-squares method.)
2-42
Operator’s Manual Chapter 2
The Lipemia, Hemoglobin and bilirubin which included in serum sample is influenced
the result.
The serum information measures the Lipemia, Hemoglobin, Bilirubin (L-H-I), and
judges the level of it.
The serum information result makes the judgement based on the information results.
Affect the measurement result. Lipemia(L)・Hemolysis(H)・Bilirubin(I).
2-43
V5.7
As sample probe, reagent probes and reaction cells contact with different kinds
liquid and sometimes built-in cleaning process is not sufficient.
It may cause carryover. To avoid these kinds of carryover, it is possible to do
special cleaning by carryover protection function.
2-44
Operator’s Manual Chapter 2
By the combination of automatic start-up, shut down and maintenance, shorten the
performance time when start-up and shut down the analyzer.
The settings are available for each day of the week.
Both ON and OFF time should be set. Only ON or OFF time cannot be set.
This is a function to automatically conduct washing, calibration and etc. when the
analyzer starts up.
Setting both automatic start up and maintenances shortens the time period before
opening for users.
The settings are available for each day of the week.
The time of automatic start up should be set after the system is turned ON.
2-45
Operator’s Manual Chapter 3
3-1
V5.7
3-2
Operator’s Manual Chapter 3
3-3
V5.7
3-4
Operator’s Manual Chapter 3
3.1.1 PC specification
There are two CDs (Disk1 & Disk2). Please be sure to install from
Disk1.
Disk 1 : Microsoft SQL Server 2012(2005)Express
Disk 2 : BIOLIS50i program
3.1.3 Installation
1. Disk1 installation
1). Insert Disk1 to the PC and installation starts automatically.
2). Select language.
3). Installation starts.
4). Click “Next” button.
5). Click “install” button.
6). Click “Finish” button.
*Disk1 installation is finished. Then install Disk2.
2. Disk2 installation
1). Insert Disk2 to the PC and installation starts automatically.
2). Select language.
3). This screen is shown. Click “Next” button.
4). Please read “Software license agreement” and click “I accept the terms
of the license agreement”.
5). Please input serial number, “User Name” and “Company Name”.
*If the number is “5000290212”, please input like the following photo.
3-5
V5.7
Note: Please input the correct serial number. The analyzer will not work
normally if the wrong number is inputted.
(Icon)
BIOLIS 50i
8).Please click the icon and confirm whether the installation is completed
correctly.
Default setting of use authority is User Level 1 (user name [Service], can
use all operations).
3. CAUTION
Disk1
Disk1 Ver. 3.03 is for Win8 and 10 and Disk1 Ver1.00 is for Win7.
Other combinations are not allowed.
Disk2
Disk2 Ver. 1.33 now supports Windows 8 and Windows 10, both 32bit and 64bit
versions OS.
Please refer to following matrix for your better understanding of the
combination of the Disk 1 & 2
32bit 64bit SQL Server
DISK1 V1.00
Windows7 2005 Express
DISK2 V1.33
DISK1 V3.03 DISK1 V3.03
Windows8 2012 Express
DISK2 V1.33 DISK2 V1.33
Windows1 DISK1 V3.03 DISK1 V3.03
2012 Express
0 DISK2 V1.33 DISK2 V1.33
3-6
Operator’s Manual Chapter 3
3.2.16 Shutdown
3-7
V5.7
Confirmation of the following points before starting the analyzer. Perform the
reagent refill or other requirement.
When analyzer is attached the ISE unit, add the following inspection.
3-8
Operator’s Manual Chapter 3
Login screen
4. Input [User name] and [Password] and click Login button. For User
Name and Password, please refer to the “Chapter4.4.12Logoff, Login s
creen”.
5. After a while, Routine Menu screen appears.
3-9
V5.7
IDLE WARM-UP
When ISE automatic calibration was failed, the ISE error screen will display
after parameter down loading. Confirm the error details by clicking the
ISE button at down part of monitor screen.
The error message shows by six digit of error code. Refer “Chapter6, 6.5
ISE Error code list”.
Note: The stand-by state means the temperature monitor shows [TEMP-OK].
It will take around 25 minutes after turning ON the system switch.
CAUTION
When analyzer system switch or PC power switch is turned OFF
do not turn them ON within three seconds.
If ISE is built in, do not turn analyzer system switch ON again
within one minute.
3-10
Operator’s Manual Chapter 3
Click
1) Each analyzer unit moves to the home position, (the starting position)
and stops there.
2) During movement of initialization, the status monitor shows [INIT].
3-11
V5.7
Click
Click
3-12
Operator’s Manual Chapter 3
Click
3-13
V5.7
Item list
Back color
Reagent 1: Yellow
Reagent 2: Blue
② Move the cursor to the [Item list] area and click the item you want
to register.
The back color of the item turns from white to blue.
③ Double-click the item.
Item number, item name, reagent structure (R1 or R2) will be
written into the designated position automatically.
④ Click save button.
3-14
Operator’s Manual Chapter 3
3-15
V5.7
Click
Click
3-16
Operator’s Manual Chapter 3
Click
Reagent screen
3-17
V5.7
1. Calibration registration
Click
Calibration screen
Tray position- Repetition No.
3-18
Operator’s Manual Chapter 3
Note: When the item is not on the screen, look for it by clicking or
in the scroll bar on the right side.
② To save the data, click Yes button. If you do not want to save,
click No button. The dialog box disappears.
3-20
Operator’s Manual Chapter 3
2. Calibration Order
Calibration screen
③ For Blank sample order entry, click [Blk Odr.] check box.
When order is accepted, レ mark appears in the box.
3-21
V5.7
① Click Item and select Item parameter which changing the concentration.
② Click calibration conc. box of calibration in Item parameter page 1.
Item screen
3-22
Operator’s Manual Chapter 3
Calibration screen
2. Select Round No. in Monitor menu screen. Then click Calib Start
button.
RUN
Click
3-23
V5.7
3. When the sampling of the last sample has finished, “Sampling has finished
is announced.
The status monitor display turns from [RUN] to [S. STOP].
A letter message: [Sampling has stopped.] flows at the bottom of the
screen.
3-24
Operator’s Manual Chapter 3
Calibration screen
3-25
V5.7
1. Operation flow
Click
3-26
Operator’s Manual Chapter 3
3-27
V5.7
⑭ Click Yes button to save the data. Click No button not to save the
data.
3-28
Operator’s Manual Chapter 3
Click Click
Select item
⑤ Click Yes button to store the data. Click No button not to store the data.
Dialog box disappears.
⑥ Repeat the procedure ② to ⑤ for the next test item.
3-29
V5.7
3-30
Operator’s Manual Chapter 3
⑥ Select test items by clicking each test item, in Order entry screen.
Name Dil
レ LDH |
レ ALB |
3-31
V5.7
Monitor screen
Click QC Start
button
3-32
Operator’s Manual Chapter 3
1. Current QC screen
① Click QC button.
② Clicking Current QC button at the sub menu screen, the following screen
appears.
Current QC screen
3-33
V5.7
3. Cumulative screen
① Click QC button.
② Clicking Cumulative button at the sub menu screen, the following
screen appears.
⑤ Clicking List button in the sub menu, QC test results are shown in list style
3-34
Operator’s Manual Chapter 3
3-35
V5.7
Click
R.Round
The reagent round number and round name are shown.
File Date
The date of built-in clock is automatically shown.
Date selection is possible by clicking button, to refer former order
entry.
Round- S.No.
There are 2 boxes, left box is used for sample round number input and
right box is used for sample number (position number) input.
Input method: number key and Enter key for both boxes.
Acceptable numbers as Round No.: 1 – 900
Acceptable numbers as sample No.: Patient sample 1 – 72
QC sample C1 – C30
STAT sample E1 – E30
Barcode ID
Character string of sample barcode is shown automatically or input if
sample barcode is applied.
Patient ID
Patient ID is shown automatically or input.
If the input patient ID is already registered one, stored information of
the patient are shown automatically.
Name
Input patient name. Insert a space between family name and given name.
Sex
Select one of male, female or unknown from pull down menu.
Age
When date of birth is input, age is shown automatically.
If age is less than one, the age is expressed as month and day.
It is possible to input age directly without date of birth input.
Draw Date
Input date and time. Insert a space between date and time.
Doctor
Input the name of doctor in charge.
The doctor name is registered by hitting Enter key.
3-37
V5.7
Section
Input the section in the hospital.
The Section name is registered by hitting Enter key.
Cup
レ Check when the sample cup using with tube as a adopter.
Comment 1
Select serum information out of Hemolysis, Lipemia and Bilirubin.
Judge by eyesight and input comment by Clicking list box.
#of Asp
Input the test repetition number and hit Enter key.
Default setting is 1(one), but you can set up to 10.
Specimen
Select one from the list box, Serum, Urine, Plasma, CSF, Dialysis and
Other. Default is serum.
Sampling stats
Character string of sample barcode is shown automatically or input if
sample barcode is applied.
Control
Select one from the list box of registered QC sample name.
By the selection, position number registration and item name registration
are automatically done.
For the QC sample registration, please refer to “Chapter 4, 4.4 QC”.
Sample Status
Automatically shown. There are 6 statuses: entry, under analysis, test
finished, cleared.
Sample Type
Select one from the list box, Patient, Control and STAT. (Standard sample
and STAT sample are shown automatically.) Default is Patient sample.
3-38
Operator’s Manual Chapter 3
① Select a Profile by double clicking the Profile name in Profile list box
at the right side of Order entry screen.
Profile List
② The grid lines of selected profile will be blue color from white color.
When selected, レ mark appears at the left of the Profile name.
If you want to cancel the selection, click Profile name or check mark itself.
③ Input the end sample number of the group into the box of End Sample No.,
and hit Enter key in keyboard.
Clicking OK button, the group order is registered.
To cancel group order entry, click Cancel button.
3-40
Operator’s Manual Chapter 3
③ Input the end sample number of the group into the input box of
End Sample No. , then hit Enter key.
Clicking OK button, the Group order is deleted.
If you do not want to delete, click Cancel button.
3-41
V5.7
1. Set the samples on the patient sample positions, following to the sample
numbers in the Order entry screen.
Click
Routine Menu screen
4. When the sampling of the last sample has finished, “Sampling has finished”
is announced, and following dialog box appears on the screen.
The status monitor display turns from [RUN] to [S. STOP].
5. After measurement of all the results and finishing pre-determined operation, the
an analyzer stops automatically.
Then “Measurement has finished” is announced and [END OF RUN] flows
at the bottom of the screen.
The status monitor display turns from [S. STOP] to [END].
3-42
Operator’s Manual Chapter 3
Test result screen and Reaction process monitor are used to confirm test
results. As for Test result screen, please refer to “Chapter 4, 4.3 Test re
sult screen”. As for Reaction process monitor, please refer to “Chapter 4,
4.5 Reaction monitor screen”.
3-43
V5.7
2. ISE
Please discard ISE waste liquid, according to the laws and
local rules. Also please
Clean the reagent pot. Please refer to “Chapter 5, 5.2.10-7 Cle
aning of sample line, and 5.2.10-8 Cleaning of reagent pot”.
4. Sample tray
Please clean the sample tray and wash off remaining samples.
When you clean the analyzer, please put rubber glove
8. Water line
Please close a water cock
3-44
Operator’s Manual Chapter 3
Click
Click
3-45
V5.7
Click
Click
3-46
Operator’s Manual Chapter 3
3.2.16 Shutdown
Finishing tests of all samples and test items, follow the procedure below to
exit analyzer.
Click
Program exit confirmation dialog box appears in the center of the screen.
3-47
V5.7
3-48
Operator’s Manual Chapter 3
Click
(2) After ordering of additional samples, please wait for the sampling
stop status.
When the sampling finishes, there is a voice message “Sampling has
completed.” and then “Sampling Stop” message appears in the lower
screen.
3-49
V5.7
Click
Note: No STAT sample order entry while STAT sample is under tests.
No double click Start STAT button.
3-50
Operator’s Manual Chapter 3
(3) Input starting sample number into [Start No.] box and stopping
sample number into [Stop No.] box.
If only one sample is registered, input the same number in both input
boxes.
(4) Click Start STAT button in main menu screen. Analyzing starts stat
sample measurement. When the last STAT sample aspiration has finished,
analyzer goes back to normal operation.
(4) Click STAT button on Routine Menu screen. The following STAT
sample number input screen is shown.
(5) Input starting sample number into [Start No.] box and stopping
sample number into [Stop No.] box.
If only one sample is registered, input the same number in both input
boxes.
(6) Click Start STAT button in main menu screen. Analyzing starts stat
sample measurement. When the last STAT sample aspiration has finished,
analyzer goes back to normal operation.
3-51
V5.7
If you want to stop analyzer immediately after analyzer starting up, follow the
procedure below.
Note: Please note, emergency stop deletes all the information under
measurement, except already measured sample.
If you want to stop sampling, click Sampling Stop button.
Click
(2) Emergency stop confirmation dialog box appears in the center of the screen.
3-52
Operator’s Manual Chapter 3
There are two kinds of re-run. One is manual rerun, which is performed
by manual operation after measurement results are given, and the other is
automatic rerun which is performed automatically by setting rerun
condition for each test item before measurement.
3-53
V5.7
Complete
Upper part of Order entry screen
3-54
Operator’s Manual Chapter 3
R&E screen
3-55
V5.7
Former result
R&E screen
3-56
Operator’s Manual Chapter 3
Click
Click
Click
3-57
V5.7
As sample probe, reagent probes and reaction cells contact with different
kinds of liquid and sometimes built-in cleaning process is not sufficient.
It may cause carryover.
To avoid these kinds of carryover, it is possible to do special cleaning by
carryover protection function.
To use this function, it is necessary to know when carryover may happen
by Exp.
3-58
Operator’s Manual Chapter 3
3-59
V5.7
A reaction cell used for registered reagent is not used for test at
the next cycle. Reagent probe dispenses cleaning solution into the
cell and washed.
Register test items which may cause carryover (Item) and cleaning
solution to use.
3-60
Operator’s Manual Chapter 3
There are 3 kinds of sample probe cleaning: Between items, between samples
and between sample kinds.
One kind of sample probe washing registration is possible.
① Click Order button. Sample Probe Wash button is shown in Order entry
screen.
3-62
Operator’s Manual Chapter 3
3-63
V5.7
Result printout
3-64
Operator’s Manual Chapter 3
2. Reagent setting
Lysing
Reagent 1
Reagent 2
Reagent screen
3. Calibration Order
Calibration screen
3-65
V5.7
Calibration screen
3-66
Operator’s Manual Chapter 3
6. Order entry
HbA1c
Order screen
HbA1c Items
Order screen
3-67
V5.7
7. HbA1c measurement
(1) Dispense lysing solution to the reaction cell from lysing bottle.
(2) Sampling from layer of red blood cell when set the primary
tube.
(3) Wash outside the sample probe using washing water by up and
down movement in trough position.
(4) Dispense red blood cell after aspiration of washing water, lyse
with mixing.
(5) HbA1c reagent 1 dispenses next reaction cell of lysing cell.
(6) Wash tip of sample probe, aspirate the lysing sample and
dispense reagent 1 reaction cell.
(7) Dispense reagent 2 to continue reaction, measure fixing time
under Item parameter setting.
3-68
Operator’s Manual Chapter 3
ISE
ISE module automatically do calibration for serum sample every 24 hours by using
Cal 1 and Cal 2.
Manual calibration, please refer to the normal calibration of “Chapter4, 4.11.5.
Analyzer maintenance screen-2 ISE-4.Calibration”.
1. How to do ISE calibration order entry for urine and dialysis sample
Check
Click
Calibration screen
Note: If you select urine and dialysis, position numbers of calibration tray are
assigned as followings: CE29 is dialysis standard sample, CE27 is urine
standard sample 1 and CE28 is urine standard sample 2
ISE module automatically do calibration for serum sample every 24 hours by using
Cal 1 and Cal 2.
Manual calibration, please refer to the normal calibration of “Chapter4, 4.11.5
Analyzer maintenance screen-2 ISE-4.Calibration”.
ISE calibration for urine and dialysis is done using sample tray.
① To initialize analyzer, click Ready button.
② Set sample cups containing standard samples for necessary ISE sample kinds.
Serum ;Use CAL-1 and CAL-2 packs
Urine ;Set a cup with Urine CAL-1 500 micro-litter at C27 position, and set a
cup with Urine CAL-2 500 micro-litter at C28 position.
Dialysis ;Set a cup with CAL-D 400 micro-litterC29 position.
3-70
Operator’s Manual Chapter 3
Click
3-71
V5.7
Click
Click
Calb.result screen
Note: If ISE unit was in operation, the error will pop up display.
If an error has occurred during the calibration measurement, Error
code is displayed in the normal calibration results confirmation
screen.
Refer to "Chapter 6, 6.3 Error Number and message list, 6.5 ISE
Error code list".
3-72
Operator’s Manual Chapter 3
2. Calibration screen
(1) Clicking ISECalresult, serum ISE calibration results are shown.
Click
Calibration screen
Serum
Note: If ISE unit was in operation, the error will pop up display.
If an error has occurred during the calibration measurement, Error
code is displayed in the normal calibration results confirmation
screen.
Refer to "Chapter 6, 6.3 Error Number and message list, 6.5 ISE
Error code list".
3-73
V5.7
ISE order entry is performed in Order entry screen like other order entries.
ISE Na,K,Cl
If any error occurred, popup display shown and an error code of 6 digits is
printed in data message area.
Refer to “Chapter 6, 6.5 ISE error code list”. In some cases, the error code
is out multiple at the same time. Numerical value is placed in the 1-6-digit
multiple digits if that occurred multiple errors at the same time.
3-74
Operator’s Manual Chapter 3
3-75
V5.7
Upper limit
Lower limit
Note: Label bottom should be more than 20 mm above the tube bottom.
If you use 12.4 x 100 mm (7 mL), label bottom should be less than 35 mm
above the tube bottom.
Note: Bar code should be set to always face the outside of the tray.
Blood collecting tube setting are 72 positions (outside and middle side) of
the sample tray.
Use rubber gloves when handling blood collection tube.
3-76
Operator’s Manual Chapter 3
Click
⑤ Select [by Sample ID] by clicking input box of [Sample ID mode] area.
⑥ Clicking >> button in [Sample ID mode] area, register End barcode.
Click OK button. If you do not use End barcode, skip this step.
3-77
V5.7
STAT sample
STAT sample tests is acceptable only under operation.
If you input common patient sample position numbers of sample tray into
[Start No.] and [Stop No.] in STAT sample order entry screen, these position
numbers are used as STAT sample position number.
When you use STAT positions (E-) for STAT sample test, as an item
registration, [E-] for [S No.] in Order entry screen. (No group order entry)
If the last on the measured specimen of emergency specimen for position (E
~). Read the bar code of the youngest position No. of sample tray.
If the last on the measured specimen is of general sample for position. Read
the bar code of the next position of interrupted by urgent sample
measurement position. Reading 5position bar code. Then to perform the
measurement if there is a specimen that has not been measured in the
meantime.
General sample for position of the measured in the same round, cannot be
used for emergency specimen.
To perform an emergency measure in the general sample for position, in the
case of the same position No. and the last specimen was measured to be
entered in the emergency interrupt [Stop No.], the bar code of the next
position of the [Stop No.] read.
Control sample
3-78
Operator’s Manual Chapter 3
Calibration
3-79
V5.7
Click
Stat
ID Sample No.
Sequence number
ID List screen
(2) Clicking Comm. settings button in the sub menu, Communication settings
screen is shown.
(3) Click input box ON of [Communication] row, in [LAN (Result send)] section.
(4) Transfer condition registration.
TCP Receive port :Accepting LAN port
Default is “3010”
Send port :Default is “3011”
Delimit character Default is “,” (Comma).
IP address acquisition is automatic.
(5) Clicking Order button, following message box is shown.
3-81
V5.7
Message box
(7) Connect analyzer LAN port and PC LAN port with LAN cable.
Refer “Bi-directional Communication specifications LAN TCP/IP, in detail.
3-82
Operator’s Manual Chapter 3
Analyzer has Simple operation mode. Using this mode, operator can run analyzer
easily.
Here explain how to use Simple operation mode.
Maintenance screen
(2) Click Profile for Simple operation button of [File maintenance] section.
Simple operation profile screen is shown.
(3) Select Profile number from pull down menu, by clicking button.
Profile Name and Round are shown automatically.
3-83
V5.7
3-84
Operator’s Manual Chapter 3
The measurement of Simple operation Mode is necessary after the setting by Login.
Adoption of simple operation mode is as follows.
Login screen
3-85
V5.7
(1) By the above operation, analyzer starts under simple operation mode.
Refer to “Chapter 3.6.2 Adoption of simple operation mode”.
The next Simple Operation Mode screen is shown.
Sample round
Sample position
Profile
Item Name
Item Name
Sample number.
Test results of the above sample are
shown below
Result
(2) As each Profile has own sample positions, set the sample at the sample position,
which belongs to the selected Profile.
(4) Click the sample position of the selected [Profile] area, where the sample is set,
in Simple operation mode screen.
※Only one test is done under the same Sample round and sample tray position.
If you want to use the same sample position again, please change Sample
round.
3-86
Operator’s Manual Chapter 3
(6) When test has finished, test results shown in Simple operation mode screen.
If you want to see test results of another sample, click < or > button in sample
number box, to seek sample number of interest.
3-87
V5.7
Note: If you do to fit the automatic start-up operation setting and the system power
supply automatic ON by the timer setting.
Set before the automatic start-up operation start time the system power ON
time.
If the system power ON time was later than the automatic start-up operation
start time, it does not perform the operation that has been set.
Note: If you do to fit the automatic start-up operation setting and the system power
supply automatic ON by the timer setting.
Set before the automatic start-up operation start time the system power ON
time.
If the system power ON time was later than the automatic start-up operation
start time, it does not perform the operation that has been set.
3-88
Operator’s Manual Chapter 3
Note: Click "Cancel" if you do not perform start-up automatic operation of the day.
(3) When it comes to standing up during automatic operation start time changes
to "Automatic start-up run screen", behavior that was set start.
※ If you do not turn off the system power, Automatic startup standby screen
even when the operation start time is not displayed. Start the operation
that was set up to display the execution screen.
② ③ ④ ⑤
⑥
⑦
Automatic start-up run screen
3-89
V5.7
○; Operation is completed
×; Operation failure
[Blank]; Not yet working
⑥ This button is stop button. Click if you want to stop the operation.
Will be displayed next stop confirmation screen.
⑦ Click if you close the screen. When all of the operation is finished button
is enabled.
3-90
Operator’s Manual Chapter 3
Manually system switch ON. Only setting “automatic operation start-up run”
① Manually system switch ON.
② Automatic start-up waiting screen will be displayed.
※If the automatic start-up waiting screen is displayed, you cannot be
operated in the other screen. If you do not turn off the system power, it
does not display.
Note: If you ON the system power after the start-up operation start time set.
"Automatic start-up waiting screen" is not displayed. And it does not also set
worked.
If you do not turn off the system power.
① When it comes to standing up operation start time, display is "Automatic
start-up run screen". Will be setting works.
② If another operation was running in start-up operation start time set, does
not stand up when the automatic operation of the day.
Click "Cancel" if you do not perform start-up automatic operation of the day.
After the other operation prior to start-up operating time that is set, once in the
system power is OFF, if again you have to turn ON the system power, to display
again "Automatic start-up waiting screen".
② ③ ④ ⑤
⑥
⑦
Automatic start-up run screen
3-91
V5.7
○; Operation is completed
×; Operation failure
[Blank]; Not yet working
⑥ This button is stop button. Click if you want to stop the operation.
Will be displayed next stop confirmation screen.
⑦ Click if you close the screen. When all of the operation is finished button is
enabled.
3-92
Operator’s Manual Chapter4
Chapter 4
Screens and programming
4-1
V5.7
4-2
Operator’s Manual Chapter4
4-3
V5.7
4-4
Operator’s Manual Chapter4
4.1.1 Function
This screen is the main screen, which is used for selection operation such as
analyzer operation control, routine operation program, circumstance setting
program and analysis condition setting program selection. And also displays
run monitor and analyzer status monitor Input and setting of sample number
test range is possible in this screen. Each switch and button has the following
function.
Routine operation
Condition setting
Environmental setting
4-5
V5.7
Run monitor
Status message
Routine Menu screen
4-6
Operator’s Manual Chapter4
QC start When this button is ON, clicking the button starts tests for
QC sample which test order has already entered
This button is ON, while under Ready state and sampling stop
state.
4-7
V5.7
Exit This switch is used to shut down analyzer when test has
finished and Priority level switching.
[Temperature monitor]
Displays analyzer temperature control state.
Shows [---], [WARM-UP] or [TEMP-OK].
Search method
① Select the specimen to be displayed.
② Click Update button.
4-8
Operator’s Manual Chapter4
Printing method
① Select the result row and click the Add line
button.
② The selected row is displayed in .
※ If you intend to print lines of 1-10 and 21-30, enter
"1-10, 21-30" in
③ Click Print button to print.
Reagent info. Reagent name, bottle position and residual volume are
shown by table format.
4-9
V5.7
4.2.1 Function
This screen is used for new order entry, modification or deletion of test
items for the samples, except standard samples, and above function is
possible even test is running. You can designate a sample round and
position in order to measure sample.
4-10
Operator’s Manual Chapter4
Round – S.No. * ** There are 2 input boxes for sample Round No.(*)
and sample No.(**) * means Round No. input
and ** means sample No. for new order entry,
reference, modification and deletion.
Click the box, and do the following input operation.
4-11
V5.7
Trace sample Check the when you use to put the sample cup in a
test tube or adapter.
Sample information
Test condition
4-12
Operator’s Manual Chapter4
レ UA Order box
When item name is clicked, a check mark
appears at the left box.
If click the item name again, order is cancelled.
In case of rerun --- At the end of the test, already
tested item name is shown in blue color. Input a
check mark to do rerun order entry.
Dil
UA * | Dilution ratio box
Sample dilution ratio setting (13 kinds)
Clicking desired dilution ratio in dilution ratio list, the
ratio is written into the box.
Select the desired ratio or select the blank to cancel.
4-13
V5.7
Profile screen
4-15
V5.7
In case of “All”:
Clicking YES button, Host is requested to send the
information.
In case of “Appoint”:
Input Start sample ID and End sample ID. The inputted
ID No. information is requested to the HOST by clicking
this button.
4-16
Operator’s Manual Chapter4
File Date
Clicking button, select File
2011/01/19 date to display.
2011/01/18
2011/01/17
Setting of sample ID
4-17
V5.7
ID list screen
4-18
Operator’s Manual Chapter4
4.3.1 Function
4-19
V5.7
Age * Patient age display (For the child less than one year, show
months and days).
Trace samples * This means whether you use test-tube or sample cup.
4-20
Operator’s Manual Chapter4
(In case of “by Sample No”) (In case of “by Sample ID”)
Host Send screen
4-22
Operator’s Manual Chapter4
Database
Results Test results list display and print out is
possible.
Patient History
Sequential test results display for one
patient and print out and graphic
display is possible.
Abnormal Value
Abnormal test results list display and
print out is possible.
Report
Single Report Patient single report format.
(Please refer to the input of patient ID)
Cumulative Report
Patient cumulative report format.
(Please refer to the input of patient ID)
Error Expression
In case that data error occurred, you
can select whether to print out the
result in value or in asterisks“******”.
Header Settings
Inputting header of report is possible.
4-24
Operator’s Manual Chapter4
1. Database
(1). Results
Clicking Database and Results buttons in R&E screen, the following
screen is shown.
4-25
V5.7
Print The data can be printed out, using the external printer.
Print operation is the same as “1. Test results”.
Exit Clicking the button, the screen goes back to Data list
menu screen.
4-27
V5.7
Print The data can be printed out, using the external printer.
Exit Clicking the button, the screen goes back to Data list
menu screen.
Save Select a data cell to save by clicking, and click this button.
4-29
V5.7
2. Report
You can show the patient single report screen or the cumulative report screen by
clicking the button.
Delete Selecting the cell to delete and clicking this button, the item will be
deleted.
Exit Clicking the button, the screen goes back to Database menu
screen.
Save Clicking this button, the inputted Error Expression setting are
saved.
Exit Clicking the button, the screen goes back to Database menu
screen.
(6). Exit Clicking the button, the screen goes back to R&E screen.
4-31
V5.7
4.4.1 Function
Current QC screen
4-32
Operator’s Manual Chapter4
Name
CTRL-1 * QC sample name
| After clicking the box, Name + Enter key
CTRL-30 *
4-33
V5.7
4-34
Operator’s Manual Chapter4
Item Name * **
These boxes are used for Item parameter setting of each
item.
* means this test item number
**means item name
Firstly click in the * box first, then click item name.
4-35
V5.7
Clicking Item Name in QC range menu and select item, following QC range
setting screen appears.
This screen is used to input the following target values for each registered QC
sample. Target value of data range, mean and SD.
・Range
・Mean
・SD
Range Mean SD After clicking each box, you can input a target value of
QC and click Enter key.
4-36
Operator’s Manual Chapter4
1. Current QC screen
4-37
V5.7
P Displays P flag.
Refer to “Chapter 6, 6.1.1 Data message flag”.
4-38
Operator’s Manual Chapter4
Clicking QC button and Daily button, the following Daily QC screen appears.
In this screen, the test results graph within a day of QC samples are shown.
Daily screen
4-39
V5.7
Cumulative screen(Graph)
4-40
Operator’s Manual Chapter4
Westgard notation
Mean graph
Number of test results
4-41
V5.7
If the sample is tested once a day, variation width of the sample is shown as 0, in the
graph.
Central part / between the graphs shows the data of statistic calculation.
Statistics calculation
4-42
Operator’s Manual Chapter4
4.5.1 Function
This is used for optical absorption data of all reaction process and graphic
display.
This screen can display test results of main-wavelength, sub-wavelength and
the difference between two wavelengths.
4-44
Operator’s Manual Chapter4
Reaction Graph Graphic display with X-axis is test point and Y-axis is
optical absorbance value By the selection of condition
setting, controlled and displayed.
4-45
V5.7
(1) Press Ctrl + D key simultaneously, when Reaction monitor screen is shown.
(2) The following screen is shown.
(3) Input [Start No. - Stop No.] of RC# to be stored in [Register number].
Output file name can be changed.
(4) Click New button. Reaction monitor file is stored as CSV file.
Click Add button if you want to add data to the file until now.
Clicking Cancel button, Output Reaction monitor file screen is closed.
4-46
Operator’s Manual Chapter4
4.6.1 Function
This screen is used for Item parameter analysis condition, depending on each
item, registration and modification. The following selection buttons are appear.
Clicking Optical button (Item No.1–77), in Item number pull down menu in Item
parameter screen, Common test item parameter screen first page is shown.
Common test item parameter screen is composed of five pages.
Clicking Item and click of Item No., following Item selection menu screen
appears.
Click
Note: The test item names, which are already registered, are shown on the
screen.
This screen is used for the designation of test items, to be register or
modify.
Item No.| This input box is used to select test item number to
register or to modify. Clicking button, test item list is
shown on the screen.
Clicking the number, the number appears in the box and
back color turns to blue.
4-47
V5.7
Item Name Test item name appears in the box if already registered.
For registration of new test item, input tem name into
the box.
Save Clicking this button, the contents of the screen are stored in
the Item Parameter file.
Data information
4-48
Operator’s Manual Chapter4
Analysis
Calibration
If the other than Factor is selected as calibration type, input standard sample
concentration.
4-49
V5.7
Correlation
Mixing Speed
Select mixing speed of MU1 or MU2.
If you want to use Single reagent, mixing of M2 is not
done.
4-50
Operator’s Manual Chapter4
Aspiration
Reaction monitor
0 Level Point * Basic test point, which is used for reaction monitor or
graphic display. Default is “0”.
After clicking the box, Number + Enter
4-51
V5.7
Data process
Read
Start End Stating test point and ending test point of main
Main * * test interval, which is used for test result value
calculation. (Input; 9 to 76, Double reagent; 36~)
1 step is 7.5 seconds.
Abs. limit
Low * Input box for the lower and upper limit of optical
∼ absorbance. (-3.0 – 3.0 Abs)
High * After clicking the box, Number (Abs) + Enter
LOW ≦ HIGH
If calculated variation is out of the designated low
range, “B” flag appears in the result printout.
If out of the high range, “G” flag appears.
Correction value
4-52
Operator’s Manual Chapter4
End Point Limit * Limit value to check optical absorbance upper limit,
for END point assay. (-3.0~3.0 Abs)
After clicking the box, Number (Abs) + Enter
If optical absorbance exceeds limit value, within the test
points designated by main interval, “E” mark appears in
result.
Linearity check (%) * Linearity standard value for RATE assay (0~99%)
After clicking the box, Number (%) + Enter
As for calculation method of the number, refer to “Chapter 2,
2.7 Linearity check, Calculation method”.
If calculated linearity exceeds standard value, “L”
mark appears in result.
Prozone check
Start End
First * ** Starting (*) and ending (**) test points of basic test interval.
(9 – 76) After clicking the box, Number + Enter
Second * ** Starting (*) and ending (**) test points of reference test
interval. (9 – 76) After clicking the box, Number + Enter
Limit (%)
*** Comparison limit value (-99999 - 99999) %
After clicking the box, Number + Enter
4-53
V5.7
Minimum Abs.
Average *** Enter the minimum absorbance average value of main
wavelength in the first interval. (0~3.0) (Default 3.0)
No Prozone check is done if the average value exceeds the
input value.
Variation *** Enter the absorbance variation of main wavelength in the first
interval per cycle. ( ⊿ OD/cycle) ( 0 ∼ 3.0 ) 1 cycle = 7.5sec
(Default 0)
No Prozone check is done if the slope value exceeds the input
value.
Note: If the set value of the minimum Abs. average and the
variation are “0” or are not set, Prozone check is not
conducted.
Normal range
Input boxes of lower and upper limit of male or female
normal range for various sample kind.
(Serum, urine, plasma, CSF and dialysis)
If the normal range is out of the range, “L” or “H” flag
appears.
Panic range
Input boxes of lower and upper limit of male or female
normal range for various sample kind.
(Serum, urine, plasma, CSF and dialysis)
If panic range is out of the range, “PL” or “PH” flag appears.
Reaction check
Reaction check ON-OFF switch will be established to select
whether to check a reaction absorbance or not. A checkpoint
can be set.
The upper limit and the lower limit of the check range can
4-54
Operator’s Manual Chapter4
be inputted.
The average of the main wavelength absorbance of three
points centered on the read point you enter will display the
"AB" in the case beyond the setting range.
[Condition of the setting value range]
Check Point : 1 90
LOW: -3.0 - +3.0
HIGH: -3.0 - +3.0
Auto rerun SW
ON ◎OFF Auto-rerun selection by clicking On or OFF
If select ON here, you have to do the following settings.
Set Auto-rerun ON in System parameter screen.
Register the diluent to be used in System screen.
If selected OFF, the following settings are ignored and no
rerun.
4-55
V5.7
The amount of dispend: Set the sample volume at the time of automatic
rerun.
Qualitative value
4-56
Operator’s Manual Chapter4
Parameter is fixed.
Item No.| Item number is 96. Is fixed.
Item Name * Test item name (Max input 8 characters)
Data Information
Wavelength
Item Main Wave- Sub Wave-
Lipemia 660nm 700nm Parameter is fixed.
Hemolysis 570nm 600nm
Bilirubin 450nm 505nm
4-57
V5.7
Correlation
Parameter is fixed.
Aspiration
Parameter is fixed.
Kind ◎ Single Single is fixed.
Vol
Sample * Dispense volume is 16 L.
Reagent 1 * Dispense volume is 144 L.
Reag. vol Bottle vol.
Blank value
Parameter is fixed.
Reaction Monitor
Parameter is fixed.
Data Process
Read Parameter is fixed.
Abs.Limit
Mixing Speed
Parameter is fixed.
4-58
Operator’s Manual Chapter4
Qualitative value
4-59
V5.7
Clicking of ISE button, in ISE item parameter screen first page is shown.
Data information
4-60
Operator’s Manual Chapter4
Analysis
Correlation
Normal range
Low High
Male * ** Input boxes of lower and upper limit of male or
Auto-rerun setting
4-61
V5.7
Auto-rerun setting(conc.)
4-62
Operator’s Manual Chapter4
Clicking of HbA1c, the Item number 101,102,106, in the Item number pull
down in Item parameter screen, HbA1c parameter screen first page is shown.
Item No.101, 102 The No. 101 and 102 are used for one parameter
methods same as common Items.
(Example : Latex reagent)
Item No.106 The No. 106 is used for two parameters method.
Parameter set Hb and A1c individually.
(Example: Enzymatic method)
4-63
V5.7
Item Name * Test item name. (Max input 8 characters)Clicking the box,
input item name + Enter.
Data information
Analysis
4-64
Operator’s Manual Chapter4
Calibration
In case of Factor
Factor * Input Factor value.
(In case of decrease reaction, minus – is necessary to
input.)
Factor value input box
If the other than Factor is selected, the following is
shown. (-99999 - 99999)
If the other than Factor is selected as calibration type, input standard sample
concentration.
Blank * “0” concentration, such as physiological saline, input
is necessary.
#1 * After clicking this box, Concentration value + Enter
∼ Usable up to 6 samples.
#6 * Registration or modification starts from the younger
number.
Concentration setting order also starts from younger
to larger. In case of multiple calibration, you need to
input more than 3 concentration.
Correlation
Reaction monitor
0 Level Point * Basic test point, which is used for reaction monitor
graphic display. Default is “0”.
After clicking the box, Number + Enter
Data process
Read
Start End Stating test point and ending test point of main
Main * * test interval, which is used for test result value
calculation. (Input; 9 to 76, Double reagent; 36~)
1 step is 7.5 seconds.
Abs. limit
Low * Input box for the lower and upper limit of optical
∼ absorbance. (-3.0 – 3.0 Abs)
4-66
Operator’s Manual Chapter4
Correction value
(K value)
K value calculation method
1. In case of two reagents system and sub test
interval is before R2 dispense, main test interval is
after R2 dispense.
End Point Limit * Limit value to check optical absorbance upper limit,
for END point assay. (-3.0~3.0 Abs)
After clicking the box, Number (Abs) + Enter
If optical absorbance exceeds limit value, within the
test points designated by main interval, “E” mark
appears in result.
4-67
V5.7
Prozone check
Start End
First * ** Starting (*) and ending (**) test points of basic test
interval. (9 – 76)
After clicking the box, Number + Enter
Second * ** Starting (*) and ending (**) test points of reference test
interval. (9 – 76)
After clicking the box, Number + Enter
Limit (%)
*** Comparison limit value (-99999 - 99999) %
After clicking the box, Number + Enter
Minimum Abs.
Average *** Enter the minimum absorbance average value of main
wavelength in the first interval. (0~3.0) (Default 3.0)
No Prozone check is done if the average value exceeds the
input value.
Variation *** Enter the absorbance variation of main wavelength in the first
interval per cycle. ( ⊿ OD/cycle) ( 0 ∼ 3.0 ) 1 cycle = 7.5sec
(Default 0)
No Prozone check is done if the slope value exceeds the input
value.
Note: If the set value of the minimum Abs. average and the
variation are “0” or are not set, Prozone check is not
conducted.
4-68
Operator’s Manual Chapter4
Normal Range
Panic Range
Aspiration
Lysing sample preparation
Lysing Aspi.level * Input aspiration position (mm) from the bottom of
HbA1c primary tube. (1-7mm)
After clicking the box, Number (micro-liter) +Enter
HbA1c
Single ◎ Double Single reagent or double reagent selection Select one
out of two by clicking one of them.
Black circle appears if selected.
4-69
V5.7
Mixing speed
Item No.| When 106 is selected as Item number, the first page of
item parameter screen is shown.
Item Name * Test Item name Clicking the box, input Item name + Enter.
Data information
4-70
Operator’s Manual Chapter4
Analysis
Calibration
In case of Factor
Factor * Input Factor value.
(In case of decrease reaction, minus – is necessary to
input.)
Factor value input box
If the other than Factor is selected, the following is
shown. (-99999 - 99999)
If the other than Factor is selected as calibration type, input standard sample
concentration.
4-71
V5.7
Correlation
Reaction monitor
0 Level Point * Basic test point, which is used for reaction monitor
graphic display. Default is “0”.
After clicking the box, Number + Enter
Input warrant range of low and high limit for male and
female.
The B flag shows with result if the result exceeds the
low limit.
The G flag shows with result if the result exceeds the
high limit.
4-72
Operator’s Manual Chapter4
Data process
Read
Start End Stating test point and ending test point of main
Main * * test interval, which is used for test result value
calculation. (Input; 9 to 76, Double reagent; 36~)
1 step is 7.5 seconds.
Abs. limit
Low * Input box for the lower and upper limit of optical
∼ absorbance. (-3.0 – 3.0 Abs)
High * After clicking the box, Number (Abs) + Enter
LOW ≦ HIGH
If calculated variation is out of the designated low
range, “B” flag appears in the result printout. If out of
the high range, “G” flag appears.
4-73
V5.7
Correction value
Prozone check
Prozone phenomenon check parameter input area, for immuno-reaction.
For theoretical explanation of the principle, refer “Chapter2 2.2.2 Prozone
check function”. If the prozone check is out of the value, “P” flag appears.
Start End
First * ** Starting (*) and ending (**) test points of basic test
interval. (9 – 76)
After clicking the box, Number + Enter
4-74
Operator’s Manual Chapter4
Second * ** Starting (*) and ending (**) test points of reference test
interval. (9 – 76)
After clicking the box, Number + Enter
Limit (%)
*** Comparison limit value (-99999 - 99999) %
After clicking the box, Number + Enter
Minimum Abs.
Average *** Enter the minimum absorbance average value of main
wavelength in the first interval. (0~3.0)
(Default 3.0)
No Prozone check is done if the average value exceeds the
input value.
Variation *** Enter the absorbance variation of main wavelength in the first
interval per cycle. (⊿OD/cycle)(0∼3.0)1 cycle=7.5se
(Default 0)
No Prozone check is done if the slope value exceeds the input
value.
Note: If the set value of the minimum Abs. average and the
variation are “0” or are not set, Prozone check is not
conducted.
4-75
V5.7
Computing Method
Item1 * Hb display.
Item2 ** A1c display.
Method *** Input the formula for calculated the HbA1c.
Data Information
Units * Unit name for test result value. (Up to 8 letters)
After clicking the box, Unit to be registered+Enter
Decimals * Number of digits below decimal point for test
result value (0 – 3)
After clicking the box, Number of character+Enter
Computing Method List
New Formula Edit the formula, create a formula list.
Normal Range
4-76
Operator’s Manual Chapter4
Panic Range
Aspiration
Lysing sample preparation
Lysing Aspi.level * Input aspiration position (mm) from the bottom of
HbA1c primary tube. (1-7mm)
After clicking the box, Number (micro-liter)+Enter
HbA1c
Single ◎ Double Single reagent or double reagent selection Select one
out of two by clicking one of them.
Black circle appears if selected.
4-77
V5.7
First、Second
Input read starting point and ending point for the first
and second interval.
Check Value
Input check value against lysing condition optionally.
Mixing speed
4-78
Operator’s Manual Chapter4
4-79
V5.7
Computing Method
After clicking each box, select an item in the Item list and double click it.
Data Information
Item List
4-80
Operator’s Manual Chapter4
Normal Range
Input boxes of lower and upper limit of male or
female normal range for various sample kind.
(Serum, urine, plasma, CSF and dialysis)
If out of range value, flag “L” or “H” mark appears.
Panic Range
Input boxes of lower and upper limit of male or
female normal range for various sample kind.
(Serum, urine, plasma, CSF and dialysis)
If out of range value, flag “PL” or “PH” mark
appears.
4-81
V5.7
1. Set the detergent capacity. Capacity that was set here, probe washing,
the reaction cell washing It is used in the purification.
Settings for each cleaning is done with of "system screen-wash1".
2. A bottle capacity setting of diluent-hemolytic agent.
3. [Bottle set position]
Position1: Acid detergent Position2: Alkaline detergent
(Position3: Diluent R1 only)
Detergent screen
4-82
Operator’s Manual Chapter4
4-83
V5.7
4.7.1 Function
This is used for the registration of reagent bottle position on the inside and outside,
from No.1 to 71.
Depending on the System parameter setting, residual test display method on the
screen is decided to be test count or residual %.
(If reagent barcode is used, input automatically.) Each item can hold 2 bottles.
Click button
4-84
Operator’s Manual Chapter4
Round Name After selecting above the Kind of Round, name for the
kind of round in Round Name.
Set Current Round Clicking this button, the shown round name is
Registered as the current round.
For the current round, the back color turns to white,
and the back color is yellow for the not current round.
Item The display of the test item name of the set reagent
Test Item registration or modification procedure
① Clicking bottle position which you register or change.
The color of the box turns from white to blue.
② After the selection of test item to be registered or
modified from the item list by clicking, double click it.
If the item to be registered or modified is not in the list
search the item, clicking or button on the right.
③ The Test item is set in the position.
4-85
V5.7
In case of “Count”
Reagent residual volume renewal by volume input
1. Click a position box to be renewal.
2. Press Ctrl key. Test volume input box is shown.
3. Input Volume (ml) + Enter. T. Counts renew
Dual No. When two reagent bottles of the same reagent are set on
the same round, bottle position number of the second
bottle is shown automatically. Maximum 2 bottles are
acceptable.
4-86
Operator’s Manual Chapter4
Alarm * T.count.(%)
This box shows alarm level of short reagent. When
residual test number or residual % reaches this [residual
test volume alarm], the back color of [Alarm T. count]
box becomes to yellow.
FILL ALL This button is shown (in case of test count) screen.
Clicking this button, residual test number of all items
are input automatically.
4-88
Operator’s Manual Chapter4
4.8.1 Function
1. Calibration screen
Calibration screen
4-89
V5.7
Graph
Item Name Test Item Name display (upper part), corresponding to the
UAL test item number.
*Lot 001 *Reagent Lot No. is displayed when using reagent barcode.
(lower part)
How to input
1. Input “Sample position” into above (**) and press Enter key
2. Input “The number of sampling” into above (*) and press Enter key
3. Input “Absorbance value” into above (***) and press Enter key
4-90
Operator’s Manual Chapter4
Clicking this button, enter orders for all the test items
on this screen except ISE, by one action.
Clicking the button on the left of each item, a check
mark appears in the box. To cancel the order, click the
button again. The mark disappears.
4-91
V5.7
Stability screen pops up. The valid date and time of the
calibration curve for each item is displayed.
CalData screen
4-92
Operator’s Manual Chapter4
Recalculate as follows.
1. Click check box of recalculating order.
2. Click current displayed absorbance value for
recalculating test item.
3. Calibration history is shown at right bottom of
the calibration screen. By double clicking, select
one absorbance value which is needed
re-calculation from there.
4. Current absorbance changes to above 3 Abs.
5. Clicking Re CAL button, the following dialog box is
shown.
4-96
Operator’s Manual Chapter4
4.9.1 Function
Clicking each “System program select button”, the next setting screen is
displayed.
4-97
V5.7
Clicking System button and Sys.Param button, the following screen is shown.
Unknown Tray For Unknown Tray, the measurement order follows the
◎by sample order of sample number.
Calobration Tray For Calibration Tray, you can select By Sample or By Item.
by sample
◎by Item Default setting is By Item.
by Item : Measure the blank sample for each item and
standard sample in pairs.
by Sample: Measure the blank sample for all the items
before measuring each standard sample.
Control Lock-out
Control Yes ◎No If set as Yes, when a test result of control sample exceeds
2SD, no order entry of the test item is acceptable.
Default is No.
4-98
Operator’s Manual Chapter4
Reagent remain
◎Count Reagent residual volume display selection.
Level sensor Count : If Count is selected, residual volume is
shown by test count number.
Level sensor : If Level sensor is selected, residual
volume is shown by %.
Warming…
If unable is Measurement is selected, no test run during warm-up.
Default is Measurement enable.
◎Measurement enable; In case “Measurement is unable”, measurement is not.
Measurement is unable; started during warming up.
Auto Rerun
On ◎ off Select whether to execute automatic rerun or not.
This selection has the highest priority. If OFF is
selected here, even if ON is item parameter, no
auto-rerun.
System
ISE If you perform ISE measurement, click this box.
Reagent Barcode If you perform Reagent Barcode, click this box.
Clicking the button, Barcode reading button appears in
Reagent screen.
Clot Sensor If you perform check the clogging of the sample probe,
clicking this button.
Mix Speed
During automatic dilution, or make the mixing speed
setting of automatic hemolysis (at the time of HbA1c
measurement).
Set the mixing speed. (1∼5)
The number, the faster the more mixing speed large. to
select by clicking on the button.
After speed set, clicking Save button.
4-99
V5.7
4-100
Operator’s Manual Chapter4
Single Report
* Auto print Make the check in the case of performing automatically
patient report printing.
Date Settings
A calendar of the month is shown.
System date is marled in the calendar.
Click a date to change the system date.
Date Format
4-101
V5.7
4-102
Operator’s Manual Chapter4
Open System screen and click Wash 1 button, the following screen is shown.
Registration of washing condition for reagent probes and reaction cell are input
here.
Line Delete Clicking this button, data of the selected line are
deleted.
All Delete Clicking this button, all the input data are deleted.
Save Clicking this button, all the data on the screen are
registered.
4-103
V5.7
Cell Wash
To avoid reagent carryover through reaction cell, register effecting test item and cell
cleaning solutions to be used.
※The upper limit of setting registration is 60.
Detergent
* Input box of cleaning solution to avoid carryover.
Clicking button and select a cleaning solution from the
pull down menu.
Line Clear Clicking this button, data of the selected line are
deleted.
All Clear Clicking this button, all the input data are deleted.
Save Clicking this button, all the data on the screen are
registered.
4-104
Operator’s Manual Chapter4
Open System screen and click Wash 2 button, the following Sample Probe
Wash registration screen is shown.
Sample probe wash has 3 kinds of washing “by Item”, “by Sample” and “by
Specimen”. Only one washing kind is selectable.
In case of by Item and by Specimen, set washing condition from Wash 2 screen.
*When by Sample is selected in Sys.Param screen of System screen,
Sample probe washing button is displayed in order screen.
4-105
V5.7
4-106
Operator’s Manual Chapter4
*Only by Sample,
set washing
condition from
order screen.
Click
Sample probe wash
Wash bottle position Select cleaning solution cup position ACID or ALK
◎ ACID on a sample tray by clicking one.
ALK
4-107
V5.7
User Interface; System program version number and date are shown.
Sequence Control; Control program version number and date are shown.
4-108
Operator’s Manual Chapter4
Host
Communuication
◎On Off Using the bi-directional communication of ASTM,
click ON. Default is Off.
Request
◎ All Appoint Select All or Appoint by clicking input box of
[Request] area.
Default is Appoint.
4-109
V5.7
Reaction Info.
Sample ID mode
◎ by Sample No
lf this button is selected, a sample is identified by the position
number on a sample tray.
by Sample ID >>
lf this button is selected, a sample is identified by barcode.
Clicking >> button in Sample ID mode area, register End
barcode within 16 characters and click OK button.
If you do not use End barcode, skip this step.
4-111
V5.7
Re-run inputted ID
ON ◎ OFF
When "by sample ID" setting, perform the operation setting
of the time bar code reading failure.
Communication
ON ◎OFF Communication is connected when ON is selected.
4-112
Operator’s Manual Chapter4
Click the "Configuration" button in the System screen, initial setting screen is
displayed.
The Configuration screen is initial setting. This screen is protected by password.
Sample barcode
Lamp
4-113
V5.7
Language select
Reagent remain
Calibration
Internal printer
AUTO START
4-114
Operator’s Manual Chapter4
4-115
V5.7
4.10.1 Function
Analyzer has Simple operation mode. Using this mode, operator can run analyzer
easily.
Refer to “Chapter3, 3.2 Routine operation and “Chapter3, 3.6 Simple operation
mode”
Sample round
Item No.
Profile Name
Item Name
Sample No.
Result
4-116
Operator’s Manual Chapter4
Sample position Therefore sample number is decided after sample cup is set on a
tray.
*Only one test is done under the same Sample round and sample
position.
*If you want to use the same sample position again, please
change Sample round. Sample round of Simple operation mode
are 991 to 999.
Color of registered position changes depending on the test stage.
4-117
V5.7
4.11.1 Function
This function is used for daily maintenance, such as cell blank check,
lamp life check, ISE information and error list confirmation.
Maintenance screen
Maintenance list
Maintenance list Daily maintenance list for user is shown.
AUTO START This is the screen to set the automatic operation contents at the
start-up.
Monitor This is the screen that monitors display the device status.
File Maintenance
Samp. File This button is clicked when test results are stored in an
external memory.
4-118
Operator’s Manual Chapter4
Item File This button is clicked when item parameters are stored
in external memory.
Maintenance
Cell Check The latest cell blank value list is displayed. Check
Service maintenance
4-119
V5.7
1. Maintenance list
Click Maintenance list button, the following maintenance list screen is shown.
The maintenance list for Daily, Weekly, Monthly and every 6 months are made
in the screen. The conditions of the maintenance are checked by check box and
updated date.
Note; If the date is not updated within each period, the background color of related
button is change to yellow.
2. Auto stert
The auto-start procedures (washing, calibration etc.) are possible to set after
auto-start up.
※Refer to "Chapter 4, 4.13 Timer setting for automatic start-up".
Note;
Before setting of the ISE calibration,
two kinds of settings are essential.
4-120
Operator’s Manual Chapter4
4-121
V5.7
3. Monitor
Clicking Monitor button in Maintenance screen, the following screen is
shown.
This is the screen that monitors display the device status.
Monitor screen
4-122
Operator’s Manual Chapter4
1. ProFile
Clicking ProFile button in Maintenance screen, the following screen is
shown. Register set items in this screen to display in Order entry screen.
Message guide
Profile Name *
Profile name input box. (Up to 10 characters)
To change the profile name, click this box and delete
the old name by BS or Delete key, then input
a new name.
4-123
V5.7
2. Sample File
Save ->> This button is used to store the test results from
Hard disk to USB or other external memories.
Select the date by clicking in Current Sample File,
and click this button.
<<- Read This button is used to store the test results from
USB or other memories to Hard disk.
Select the date by clicking in External Sample File,
and click this button.
4-124
Operator’s Manual Chapter4
3. Item File
Current Item A date list, which Hard Disk has Item file.
file
External Item
A date List, which USB has Item file.
file
Save ->> This button is used to store the QC file from Hard
Disk to USB or other external memories.
Select the date by clicking in Current QC File, and
click this button.
<<- Read This button is used to store the Item file from
USB or other external memories to Hard Disk.
Select the date by clicking in External Item, File and
click this button.
4. QC File
QC file screen
Extrnal
A date List, which USB has QC file.
Sample File
Save ->> This button is used to store the QC file from Hard
Disk to USB or other external memories.
Select the date by clicking in Current QC File, and
click this button.
4-126
Operator’s Manual Chapter4
Clicking Maintenance button and ProFile for Simple operation button, the following
screen is shown.
Message guide
Profile Name *
Profile name input box.(Up to 10 characters)
To change the profile name, click this box and delete
the old name by BS or Delete key, then input
a new name.
Test Item list Test Item selection area Clicking the box, check mark
appears in this box.
Click all the boxes of Test Items to register for the
Profile.
To cancel the Test Item, click the box again.
4-127
V5.7
1. User Maintenance
Clicking Maintenance button and User Maint button, the following user
maintenance screen is shown.
Message guide
Cell Drain Clicking this button, all the reaction cuvettes are
drained.
PRIM Clicking this button, all the filling lines are filled with
water.
It is convenient to use this function at the system
set up or the piping tubes exchange.
4-128
Operator’s Manual Chapter4
Special wash 2;
Set the detergent bottles to the position 2 (Alkaline 15%)
before start the washing.
4-129
V5.7
2. ISE
4-130
Operator’s Manual Chapter4
Electrode area
Clicking Maintenance button, ISE button and Electrode button, the following
screen is shown.
[Reference]
days is Left Input box for the residual days to be alarmed.
Alarm D
_/365 days is Left
[Pump Tube]
days is Left Input box for the residual days to be alarmed.
Alarm D
_/90 days is Left
Reagent area
Open Maintenance button and click ISE button and Reagent button, the
following screen is shown for Cal 1 and Cal 2.
[Reagent 1]
Alarm T Input box for the residual test numbers to be alarmed.
_/ 1321 Residual test numbers display.
This residual test count can be over-written by
inputting a number.
[Reagent 2]
Alarm T Input box for the residual test numbers to be alarmed.
_/ 40 Residual test numbers display.
This residual test count can be over-written by
inputting a number.
4-132
Operator’s Manual Chapter4
Bottle
ml Input box of Cal-A bottle volume.
Maintenance
Clicking ISE button.
Clicking Maintenance button, the following screen appears.
This screen is used to discharge the liquid in the tubes and electrode,
or to fill them with liquid when you exchange electrodes or Cal-1,
Cal 2 or pump tubes.
ISEWash This button is used for ISE wash to clean up the ISE
unit. Set a sample cup contain with 500 uL
ISE wash solution. Clicking this button, ISE wash will
start.
Maint mode This button is used for maintenance mode.
Position Setting This button is used to set the ISE probe position.
4-134
Operator’s Manual Chapter4
1. ISE wash
Sample tray of C, E30 is done by setting the ISE cleaning solution.
3.Purge
4.Calibration
4-135
V5.7
6. U-Calib.(Urine calibration)
8. Rom version
9. Flow sensor
Note ; After the liquid surface check, please be sure to return the nozzle position.
(Please initialization)
① ② ③ ④ ⑤
⑦ ⑧
⑥ ⑨ ⑩ ⑪
1. No. display.
2. Measured value display.
4-137
V5.7
4-138
Operator’s Manual Chapter4
3. Cell Check
This screen shows current reaction cell blank data, of various wavelengths.
Cell blank data are taken during the test.
When the cell blank value exceeded the check value, boxes will be colored.
Please follow the indication on the bottom.
Refer to Chapter 5.5.2.4 Maintenance for Cuvette replacement.
4-139
V5.7
Stop Clicking this button, will stop the cell check operation.
Save Clicking this button, will save the cell check value.
Start Clicking this button, will start a cell check operation.
You cannot click if you do not in the ready state.
Close Clicking this button, the display goes back to the
Maintenance screen.
Note : When any box lights up in red, it is about the time for all cuvettes to
be exchanged.
Please exchange all of 90 cuvettes at once.
4-140
Operator’s Manual Chapter4
4. LAMP
Lamp screen
4-141
V5.7
[Date Time] Shows date and time, which the error occurred.
[Error] Shows error number of Help list.
[Unit No.] Shows the unit number of the error.
[Error No.] Shows error number.
[Function No.] Shows the code number of unit movement.
[Sample No.] Shows the sample number.
[Item Name] Shows Test item name which the error occurred.
[Detail No.] Shows detail number.
[Message] Shows the message.
[Comment] Shows the comment.
4-142
Operator’s Manual Chapter4
6. Result Log
4-143
V5.7
Add line * Print line selection button and input box of line
range to be printed.
CSV Outputs the result log that was displayed as a CSV file.
4-144
Operator’s Manual Chapter4
7. Clot Sensor
The sample probe is checked the clot. The clot sensor error will display when the
result value is above
The check level is fixed in the program.
The checking is done on the analyzer status is [Ready] and [Run].
The sampling is to stopping automatically after automatic washing when the clot
sensor value is above.
Clicking Maintenance button and Clot Sensor button, the following screen appears.
Close Clicking the button, the screen goes back to Maintenance menu
screen.
4-145
V5.7
8. Auto-gain
Auto-gain screen
4-146
Operator’s Manual Chapter4
9. Cell Skip
When the value exceeded the cell skip value, Abs. box will be colored.
Abs.box is yellow, when the value exceeded one time.
Abs.box is red, when the value exceeded two times. And then the cell is not used for
the measurement.
Skip the cell, the cell check value is not used to measure until the normal
value.
Cell No. Display cell number (No.1-90)
Abs. The result value of cell skip.
Yellow: The value exceeded one time.
Red : The value exceeded continues two times.
The red color keeps Until the value change
to normal after operation of cell check.
Note ; When the cell skip is greater than 20 cell, it will display a warning.
Refer to "Chapter 6 Check flag・Alarm function".
4-147
V5.7
4-148
Operator’s Manual Chapter4
Priority level
1. Operator level: No result edit, item parameter edit and service
button operation.
2. Manager level: All modification, operation and edition are
possible except service button operation.
3. Service engineer level: All modification, operation and edition are
possible.
4. Simple operation level: Simple operation mode only.
Log-in screen
Input [User name] and [password] and click Login button.
Log in Clicking this button, start the system program at the specified level.
Cancel Clicking this button, exit the system program.
Default user name and password are followings.
Priority level User Name Password
Operator level User1
Manager level Manager
Service engineer level Service
Simple operation level Easy
4-149
V5.7
4-150
Operator’s Manual Chapter4
Example 2: From Monday to Friday, ON at AM 8:30 and OFF at PM 6:00 and on Saturday,
ON at AM 8:30 and OFF at PM 1:30.
① Push MODE button under [time setting]. Timer mode is [action time setting
mode].
② 1 is shown at right bottom of the screen and P flickers.
③ Push SELECT button. S appears right top of the screen.
④ Input ON time by Pushing h button and m/PWD button till AM 8:30.
⑤ Push WRITE button.
⑥ Input OFF time by Pushing h button and m/PWD button till PM 1:30.
⑦ Push WRITE button.
⑧ S disappears.
⑨ Input ON time by Pushing h button and m/PWD button till AM 8:30.
⑩ Push WRITEbutton.
⑪ Input OFF time by Pushing h button and m/PWD button till PM 6:00.
⑫ Push WRITEbutton.
⑬ Push Mode button.
⑭ Mode change to [Day of the week setting] mode.
Note: Timer does action (power ON) on the day of the week, which [ ] mark
under the day of the week sign. [ ] mark flickers for the SELECT.
⑮ Repeat pushing d button till mark comes to SA.
⑯ Push WRITE to flicker [ ] mark under SA.
Note1: [ ] mark changes circularly ON - OFF - flicker -, by pushing WRITE
button.
Note2: In the same way, OFF [ ] mark under SU.
⑰ Push MODE button.
⑱Screen change to current time display.
4-151
V5.7
When the date is changed during the system program is on, the following
date change confirmation screen is shown.
※ The screen is shown after end of run when the analyzer is running.
4-152
Operator’s Manual Chapter 5
Chapter 5 Maintenance
5-1
V5.7
5-2
Operator’s Manual Chapter 5
Caution
When do maintenance operation, wear protective globes, mask and glasses
to avoid infection. (For infection prevention)
5-3
V5.7
5.2.10
Glove
Washing of Probe, Cuvette, Stirrer 1
ALK or ACID Washing
1.Special wash
Solution
5.2.11
1
Exchange of washing solutions Glove
ALK or ACID Washing
1
Solution
5-4
Operator’s Manual Chapter 5
Caution
When do maintenance operation, wear protective globes, mask and glasses
to avoid infection. (For infection prevention)
① Turn off PC and analyzer power switch before starting the maintenance operation.
② Remove the sampler cover.
③ Pull up STM arm by hand and move and rotate to above the sampler.
※Move it to sample aspirating position or detergent bottle position etc. where to be easily
cleaned.
④ Wipe out the dirt at the outside of probe tip with paper wiper or other material soaked in
pure water.
5-5
V5.7
Driver
Remove cover
5-6
Operator’s Manual Chapter 5
⑦ Insert the probe into the arm and fix it by probe fixing screw.
Washing trough
Vacuum hole
5-7
V5.7
① Turn off PC and the analyzer system power switch before starting the
maintenance operation.
② Remove reagent tray cover.
③ Remove a reagent bottle of position.
④ Move up RTM arm by hand and to the reagent bottle position.
⑤ Wipe out the dirt at the outside of probe tip by paper wiper or other material
soaked in pure water.
R1
R2
5-8
Operator’s Manual Chapter 5
Reagent probe
Driver
Connector
Reagent probe
Probe fixing screw
Remove of the probe
5-9
V5.7
Probe
⑨ Connect the connector by using tweezers and fix the probe fixing screw.
⑩ Joint part And replace the new probe and fix the probe fixing screw.
Tweezers
Connector
5-10
Operator’s Manual Chapter 5
Washing trough
Vacuum hole
5-11
V5.7
5.2.3 Stirrer
1. Stirrer cleaning
① Turn off PC and analyzer system power switch before starring the
maintenance operation.
② Turn the arm to the cuvette position by hand.
③ Wipe out the dirt at the outside of stirrer by paper wiper or other material
soaked in pure water.
④ Clean up both reagent 1 and reagent 2 stirrer.
⑤ Return the units by hand.
5-12
Operator’s Manual Chapter 5
Note ; The surface of the cell will be tend to scar easily. Please be careful.
1. Cuvette exchange
① Turn off the PC and the analyzer in advance.
② Remove the cell cover.
Stopper
Cell 5-13
V5.7
⑧ After set of all cells, rotate the cell slowly and check if the cell set
correctly or not.
⑨ Turn of the PC and the analyzer and do initialization.
⑩ Set the cell cover.
*Match the concavity and convexity of the cell cover and the RCT.
5-14
Operator’s Manual Chapter 5
Float switch
① Loosing Cap of the reservoir and pull up the float switch unit.
② Wipe out water around the filter and other part by Paper wiper for better
working condition.
③ Confirm if the float switch moves up and down smoothly.
④ If you find dart around the float switch, wash out dart by tap water with
brush.
⑤ Return the float switch unit into the drainage reservoir.
⑥ Tighten the cap.
5-15
V5.7
1. Lamp Exchange
After the lamp replacement, please Auto-gain adjustment.
Caution
Be sure to wait for ten minutes to cool the lamp and lamp house for
maintenance after turning off the system.
The temperature of the lamp rises during operation, and it might cause
burn on your fingers.
Lamp house
① Turn off the PC and the system power switch and wait for ten minutes to
cool the lamp and lamp house before starting the maintenance operation.
② Open the top cover.
③ Pull up STM arm and turn it to STF side a little.
5-16
Operator’s Manual Chapter 5
Lamp house
Lamp holding plate
⑦ Remove the lamp fixing screws (2) at lamp holding plate by (+) driver
and remove the lamp.
Slit
Pin
5-17
V5.7
⑩ Set the lamp leads to the block and fix them by (+) driver.
⑪ Set the terminal cover of the terminal block.
Terminal block
Lamp leads
Terminal block
Note : The lamp leads are made a little longer than necessary length.
Push these leads into a slit below, so that these leads are not
damaged.
5-18
Operator’s Manual Chapter 5
2. Gain Adjustment
When finishing the lamp exchange, turn on the system to adjust the gain
(adjustment of the intensity of light).
*Measure the water blank after 15 min from the lamp ON.
Auto-gain screen
③ Clicking Start button, the gain for all the wavelengths is automatically
executed.
※If you do not in the ready state, you cannot click the Start button.
④ After auto gain is completed, clicking Save button.
⑤ Clicking Close button, return to the Maintenance Menu Screen.
5-19
V5.7
3. Cell check
When Auto-gain is finished, perform cell check measurement.
② Clicking Cell Check in the maintenance menu screen, the Cell check screen
is shown.
③ Clicking Start button, the cell check for all the wavelengths is
automatically executed.
※If you do not in the ready state, you cannot click the Start button.
※Click the Stop button, aborts the measurement operation.
5-20
Operator’s Manual Chapter 5
Lamp screen
Note: “100%” is 1200 hours, but it is not the indication of its life.
Please get a rough idea to use.
5-21
V5.7
1. Daily inspection
Tube joint
Pump head
Seal screw
Tube joint
Reagent 1 Reagent 2
Reagent washing
pump unit
⑤ Check if there is not a leakage at the plunger seal nor at the tube joints.
・If the plunger seal leaks, tighten the
seal screw a little by fingers.
・If the tube joint leaks, tighten the joint screw by fingers.
Note If the above operation fails to stop plunger seal leakage, U-packing
has to be replaced.
5-22
Operator’s Manual Chapter 5
2. U-packing replacement
① Turn off PC and the analyzer system power switch before starting the
maintenance operation.
② Remove the front panel.
③ Remove the tube joints of both side.
④ Loosen the pump fixing screws, using a Phillip type screwdriver and
remove pump unit.
At the same time, disconnect all connectors, connecting pump unit and
analyzer.
⑤ Loosen head fixing screws(2 screws) and remove a pump head from
the body. When removing the pump head, pull straight up.
U-packing
Spacer
The constitution of
pump head seal
5-23
V5.7
⑩ Insert a spacer, and set and turn plunger seal screw slightly.
⑪ Assemble pump head to the body.
・If pump head is hard to insert, rotate the head right or left a little while
insertion.
⑫ Tighten seal screw by fingers.
⑬ Set the reagent pump unit to the analyzer.
⑭ Power on PC and analyzer, then initialize.
5-24
Operator’s Manual Chapter 5
1. Daily inspection
Note; If the above operation fails to stop plunger seal leakage, U-packing has to be
replaced.
Pump head
Seal screw
5-25
V5.7
2. U-packing replacement
① Turn off PC and the analyzer system power switch before starting the
maintenance operation.
② Remove the front panel.
③ Remove the tube joints of both side.
④ Loosen the pump fixing screws, using a Phillip type screwdriver and remove
pump unit.
・At the same time, disconnect all connectors, connecting pump unit and
analyzer.
⑤ Loosen head fixing screws(2 screws) and remove a pump head from the body.
When removing the pump head, pull straight up.
U-packing Spacer
5-26
Operator’s Manual Chapter 5
⑩ Insert a spacer, and set and turn plunger seal screw slightly.
⑪ Assemble pump head to the body.
・If pump head is hard to insert, rotate the head right or left a little while
insertion.
⑫ Tighten seal screw by fingers.
⑬ Set the reagent pump unit to the analyzer.
⑭ Power on PC and analyzer, then initialize.
5-27
V5.7
1. Preparation
① Air removal tool for clot sensor [parts code :50-26-0744]
This tool consists of socket, packing, joint and Poaron tube (1 , 35 cm)
Note : Do not use syringe other than 5 mL size. If the size is larger
than 5 mL clot sensor can be damaged.
Connect ① Air removal tool for clot sensor and ② syringe and tip to
complete the air removal jig
2. Procedure
① Main power supply turns ON, and start software program.
② Turns ON the system power supply.
③ Run [READY] to initialize the analyzer.
④ Remove STM arm cover, and remove Teflon tube from the sample probe.
5-28
Operator’s Manual Chapter 5
Teflon tube
Teflon tube
Note : Even if repeating above procedures does not remove air from
clot sensor, or continues to entrap air, air may be entering from
the membrane or the joint area.
In such case, replacement of the clot sensor is recommended.
5-29
V5.7
5.2.10 Air trap liquid level check and detergent aspiration method
Acid Alkaline
Air trap
Air trap
5-30
Operator’s Manual Chapter 5
SUS pipe
⑤ Remove the syringe, and replace the tube with stop pin.
5-31
V5.7
For using ISE, you need to care about the following maintenance regularly.
1. Exchange of calibrator 1 and calibrator 2
2. Exchange of electrode
3. Exchange of Pump tube
4. Exchange of PV tube
ISE screen
5-32
Operator’s Manual Chapter 5
ISE Maintenance
5-33
V5.7
Connecting Cap
③ Hooking the new Cal-1 pack, screw the cap into the pack, holding it upside
up.
④ Click PURG in ISE Maintenance screen.
⑤ Repeat ordering Purge till no bubble comes out from the tube and no error
occurs.
⑥ Then click Cancel button in ISE Maintenance Screen
⑦ Click Reagent in ISE maintenance screen.
Check Change This boxes on Reagent1 and then click Update Set button.
Residual amount of Cal-1 is updated.
5-34
Operator’s Manual Chapter 5
Connecting Cap
Hook
Cal-2 pack
Cal-1 pack
③ Hooking the new Cal 2 pack, screw the cap into the pack, holding it upside
up.
④ Click Calibration in ISE Maintenance screen.
⑤ Repeat ordering Calibration till no error occurs. Click Clib.result and check
the Calibration result. Then click Cancel button in ISE Maintenance screen.
⑥ Click Reagent in ISE maintenance screen. Check Change This boxes on
Reagent2 and then click Update Set button. Residual amount of Cal-2 is
updated.
5-35
V5.7
4. Exchange of Electrode
Holding plate
Connecting plate
④ Remove all the old electrodes (Na, K, Cl, Reference) from the connecting
plate.
⑤ Put all the new electrodes (Na, K, Cl, Reference) on a connecting plate.
Keep the holding plate down, and insert the new electrode to the backs.
⑥ Click Mainte Cancel and calibrator 1 is filled in ISE unit.
⑦ Executing Mainte Cancel , confirm there is no leakage from the electrodes.
If leakage is found, repeat from 1 to 5.
5-36
Operator’s Manual Chapter 5
① Remove CAL-1 pack and CAL-2 pack from the hook to prevent leaking from
the tube.
② Open ISE Maintenance screen ( Maintenance ISE Maintenance). Click
Mainte mode button and the liquid in ISE unit drains away to take off the
liquid in the tube.
Lever presser
③ Turn the lever presser to outside to loosen the tubes. Perform the 4 lever.
④ Remove the tube joint from hook and remove the old pump tube.
⑤ Exchange the new pump tube.
⑥ The new tube place the center of roller and hook the joints. Then put back the
lever presser.
⑦ Hang Cal-1 pack and Cal-2 pack on the Click Mainte Cancel button and
calibrator 1 is filled in ISE unit.
⑧ Click Close button.
⑨ Check Change These boxes of Pump Tube on Electrode screen.
⑩ Then click Update Set button. Residual test number is updated.
5-37
V5.7
6. Exchange of PV tube
※Please exchange once a year.
① Remove CAL-1 pack and CAL-2 pack from the hook to prevent leaking from
the tube.
② Open ISE Maintenance screen ( Maintenance ISE Maintenance). Click
Mainte mode button and the liquid in ISE unit drains away to take off the
liquid in the tube.
PV valve, PV tube
③ Pushing head of PV valve, remove PV tube in the groove of the pinch valve.
④ Exchange the new PV tube.
⑤ Set the new PV tube, inserting in the groove.
Pull
Push
PV tube
5-38
Operator’s Manual Chapter 5
Note: After electrode cleaning, calibration is necessary before starting ISE test.
5-39
V5.7
Maintenance screen
ISE screen
⑥ Cleaning solution is dispensed in ISE to clean the inside of ISE line Cleaning
can start at Ready state.
5-40
Operator’s Manual Chapter 5
Reagent pot
ISE cover
Cotton swab
Reagent pot
Note: Be careful not to leave cotton thread in the pot. The flow of the liquid
might be stuffed.
⑦ Click Mainte cancel button after cleaning. ISE unit is filled with Calibrator
1.
⑧ Put on the ISE cover after confirming that any error message is not shown.
5-41
V5.7
In case there is any bubble in the ion passage of Ref electrode, remove the
bubble as follows.
① Peel off the seal on the air hole of the electrode, and leave it open.
Passage joint
Ion passage
② Fill the micropipette with Cal-2, and insert the tip of the micropipette from
the end of the passage joint to the ion passage.
③ Make a rush of Cal-2 from the micropipette.
The bubble is removed, and Cal-2 spills over from the end of the passage joint.
④ Pull out the micropipette.
⑤ Make sure that no bubble is in the ion passage of the electrode.
If there is still any bubble, repeat the procedure from 1.
⑥ Wipe off the spilt liquid all over the electrode, if any.
5-42
Operator’s Manual Chapter 5
In case of turning off the power of ISE unit more than four days, keep the
electrodes as follows.
The explanation is given for Ref electrode, but keep other electrodes likewise.
Micropipette
Passage joint
② Fix the attached seals to the both ends of the passage not to let the liquid leak
out. Wipe off around the exits before fixing the seals. Also, wipe out the
spilt liquid.
Seal
5-43
V5.7
Maintenance
1. Special wash
Wash the analyzer same procedure as normal measurement. (about 26 minutes)
① Set the detergent bottles to the position 1 (Acid 30%) and position 2 (Alkaline
15%) before start the washing.
② Start the washing by clicking of “Maintenance”, “User maintenance” and select
“Special wash 1 (use Acid detergent)” or “Special wash 2 (use Alkaline
detergent)”.
2. All washing
Sequent washing is done automatically in this order, “Probe wash” “Cell
wash” “Mixing stirrer” (about 15 minutes.)
① Set “Alkaline detergent” in each detergent position (No. 2) for sample, R1,
and R2 before washing.
② Click Maintenance User Maint All washing and the washing
starts.
③ The reagent probe aspirate 240 uL of the detergent and the sample probe
aspirates 25uL of the detergent.
5-44
Operator’s Manual Chapter 5
Detergent bottle
5-45
V5.7
Caution Biohazard
The content of waste bottle contains sample, which may cause infectious disease.
When do maintenance operation, wear protective globe to avoid infection and be careful
not to spill the waste liquid.
Caution
The content of waste bottle contains sample, which may cause infectious
disease.
When you discard these waste, follow the instruction of qualified operator
and related law.
Some reagent, standard sample and control sample may contain regulated
substance.
When discard, consult to the maker about regulation for these substance.
Strictly follow laws and regulations of your country when discarding waste
water, reagents and samples.
5-46
Operator’s Manual Chapter 5
Caution
2. Replacement procedure
Fuse
Main power
5-47
V5.7
Note: Make sure that [SQL Server Management Studio] and [Analyzer software]
are closed before [Database Manager] is used.
⑥ Click Backup
⑦ If “Success!” is displayed, saving Backup Data is completed.
⑧ Click OK
⑨ Go back to the Backup Tool screen.
⑩ Click × resulting in closing the Backup Tool screen.
5-48
Operator’s Manual Chapter 5
Note: Save the backup data in a flash memory such as a USB memory.
※It is recommended that you use a USB memory only for backup data which
virus is removed from in terms of security.
※1 Use the backup data file which is saved in the same PC for recovery.
If a backup data file saved in another PC is used for recovery, it is not
guaranteed that the analyzer properly works.
5-49
V5.7
Paper cover
Paper
(Left side)
Power switch
(Front)
Power lamp
Printer body Control panel
5-50
Operator’s Manual Chapter 5
Power switch
Caution When you check レ , certainly turn on printer power for operation.
If you operate the analyzer while printer power is OFF (error
indicated in start-up), some of malfunctions probably occur.
5-51
Operator’s Manual Chapter 6
6-1
V5.7
6-2
Operator’s Manual Chapter 6
There are three kinds of check flags, out of range error, data error and liquid
shortage error.
Each message flag has its priority order. If multiple messages would
occur, the message with the highest priority appears.
The following messages are shown in order of priority.
Priority for display; C >T > E >B, G >P > L >AB > t
※ In case of “V”, priority for display;
C >E >B, G >P >L >AB >V >T, t
※ ”Q”only displays in case of no other flag.
C The cell blank value is lower than Cell check Level 1 or higher
than Level 3.
T Reaction tray temperature is out of the set temperature plus-minus 0.1
degree Celsius, while measurement.
E In case of END point assay:---Cell blank subtracted Abs of main
wavelength, is higher than [Endpoint Limit].
B (1) or (2) and (3): (1) The Abs at main wavelength at start test
point is lower than Abs limit [Low] in item parameter screen. (2)
The Abs at end test point is lower than the limit [Low]. (3) The
Abs of any test point of consecutive 3 points from start test point
is lower than the limit [Low].
Or the result density is lower than the result warrant limit
[Low] when HbA1c (Item number 106) is measured.
G (1) or (2) and (3): (1) The Abs at main wavelength at start test
point is higher than Abs limit [High] in item parameter screen.
(2) The Abs at end test point is higher than the limit [High]. (3)
The Abs of any test point of consecutive 3 points from start point
is higher than the limit [High].
Or the result density is higher than the result warrant limit
[High] when HbA1c (Item number 106) is measured.
P Prozone is detected. The prozone check result exceeds the upper or
lower limit.
L In case of RATE assay:---Linearity check result exceeds [LINEARITY
CHECK] value.
AB The Abs average of 3 points, which is centered upon set test points
in item parameter screen, is out of range.
Or the result is higher than the Lysing absorbance check when
HbA1c (Item number 106) is measured.
6-3
V5.7
V Data is too high or low. Rerun with dilution or confirm the reaction
Abs..
Q Measurement results of quality control sample was out of the SD
management range.
6-4
Operator’s Manual Chapter 6
For the short liquid flag, check sample, reagent and hemolyzing volume
in container.
6-5
V5.7
6-6
Operator’s Manual Chapter 6
① ② ③
① [Error]number
② [Unit No]number
③ [Error No]number
After searching the above numbers from "Error number and message" pages,
please see the "Error causes and solutions" pages.
6-7
V5.7
6-8
Operator’s Manual Chapter 6
6-9
V5.7
6-10
Operator’s Manual Chapter 6
Ref
No.
Cause Solution
UNIT1 or UNIT2 controller may be broken.
1 When [Mixing1 rotation] or [Mixing2 rotation]
errors occur, UNIT1 or MCPU PCB may be broken.
UNIT1 or UNIT2 controller may be broken.
2 When [Mixing1 rotation] or [Mixing2 rotation]
errors occur, UNIT1 or MCPU PCB may be broken.
UNIT1 or UNIT2 controller may be broken.
3 When [Mixing1 rotation] or [Mixing2 rotation]
errors occur, UNIT1 or MCPU PCB may be broken.
①Encoder may be broken.
4
②UNIT1 or NIT2 controllers may be broken.
・This message is shown if an unit doesn't return
home position within a certain time.
・While doing measurement, if the software try Please call our service division.
to let the unit do other movement, this message
5 is shown.
Causes are as following.
① Sensor may be broken.
② Motor may be broken.
③ PCBs may be broken.
①Home sensor may be broken.
② There are some differences between actual
value and theoretical value of Rotary encoder
6 (except for vertical motion units and pumps).
* In an unit without encoder, when the unit
return to 0 position, if the unit is not in the
sensor position, this message is shown.
① Please do "Ready" mode.
Other unit is in the position disturbing
7 ②If the symptom is not improved,
movement of the unit which the message shows.
please call our service division.
There are some error except Limit abnormality,
① Please do "Ready" mode.
alarm, emergency stop, EA/EB input error. The
8 ②If the symptom is not improved,
causes are as following.
please call our service division.
UNIT1 or UNIT2 PCB may be broken.
①Please change the cable
The connection between the machine and PC may be ②If there are some causes around
impossible. the machine and PC, please away from
9
①There are some noises in communication line. it.
②There is a problem in the communication line. ③If the symptom is not improved,
please call our service division.
①A motor of RCT may be broken.
②Connection between RCT motor and PCB may have
been disconnected.
10 Please call our service division.
③There is a problem in the communication with
SCPU PCB.
(One of PCBs may have been damaged.)
6-11
V5.7
Ref
No.
Cause Solution
Please change acid detergent
11 There is no acid detergent.
bottle.
Please change alkaline detergent
12 There is no alkaline detergent.
bottle.
①Please check whether the pure
water machine operate normally.②
Please check whether there are no
obstacles in hose (inside/outside)
13 There is no pure water in pure water tank.
for connection between pure water
machine and pure water tank.
③If the problem doesn't solve,
please call our service division.
①Please discard waste liquid.
14 Waste tank is full. ②If there is no waste liquid,
please call our service division.
①Please discard waste liquid.
15 Separate Collection Reservoir is full. ②If there is no waste liquid,
please call our service division.
16 There is no sample. Please add or change the sample.
Please add or change reagent 1
17 There is no reagent 1. depending on [Item Name] of error
list.
Please add or change reagent 2
18 There is no reagent 2. depending on [Item Name] of error
list.
①Please wipe off overflowed water.
②If there are some obstacles in
waste liquid line, please remove
the obstacles. If waste liquid
①There are some obstacles in waste liquid line
line is bent, please mend the line.
or waste liquid line is bent.
19 If the problem does not solve,
②DP has some problems or is broken.
please call our service division.
③CWP has some problems or is broken.
③If there is no problem in waste
liquid line, please call our
service division with keeping the
machine OFF.
The machine wash the probe
automatically. After washing, if
The inside of sample probe might be clogged up
20 jams are solved, "Recovery" message
with fibrin。
is shown. If jams are not solved,
"Big problem" message is shown.
Jams were solved as result of
21 Jams of sample probe were solved.
washing.
①Please clean the inside of the
The machine washed the probe automatically. probe by nylon wire.
22
However jams were not solved. ②If jams are not solved, please
call our service division.
6-12
Operator’s Manual Chapter 6
Ref
No.
Cause Solution
①Please check room temperature.
②If the message is displayed
23 The temperature of RCT is abnormal.
easily, please call our service
division.
①Please check room temperature.
②If the message is displayed
24 The temperature of WSH is abnormal.
easily, please call our service
division.
①Please check whether pure water
machine works normally.
②Please check whether pure water
is full in pure water tank.
③Please check whether there is no
25 There may be no water in WSH.
air in tube connected to CWS. If
there is air in tube, please
reconnect the tube.
④If the problem is not solved,
please call our service division.
The machine tried to check reagent remaining in After "ready", please operate
26
the state that was not "ready". reagent remaining check.
There is position problem about reagent 1
27 Please check round information.
position of round setting.
There is position problem about reagent 2
28 Please check round information.
position of round setting.
There is position problem about hemolysis agent
29 Please check round information.
of round setting.
①Please restart PC and the
PC cannot transmit information to the machine
machine.
30 from a PC and information may overflow in the
②If it seems to occur frequently,
machine side.
please call our service division.
①If there are causes between PC and
the machine, please remove the
A communication between PC and the machine is causes.
31
impossible by noises. ②Please change the cable.
③If it seems to occur frequently,
please call our service division.
Measurement is impossible. Causes are
①Please check parameter input.
following.
32 ②If the problem does not solve,
①PCB may be broken.
please call our service division.
②There is a problem in parameter input.
A kind of calibration is
Calibration data and a kind of calibration of mismatched. As this is a system
33
item parameter are different. problem, please call our service
division.
Please check the round setting.
There is no data in round information which PC
34 Please check whether reagents are
tried to calculate a result of calibration.
registered correctly.
Number of characters is abnormal. (too long or Number of characters is mismatched.
35
too short) Please recreate the bar-code.
6-13
V5.7
Ref
No.
Cause Solution
It is impossible to read a bar-code because of
36 Please recreate the bar-code.
a problem of bar-code.
Please check whether the bar-code
37 A bar-code is not affixed.
is affixed in primary tube.
Please put a cup or a primary tube
38 A cup is not put.
in Sampler.
There might be a request in position which the
39 machine is analyzing. (Only at the time of online Please check a request from host PC.
communication).
①Please do Auto-gain again.
②If the problem does not solve,
40 Absorbance is too low. (C mark) 。
please replace the lamp with new
one.
Check reaction monitor. Judging
from absorbance value, select a
Absorbance of measurement is out of calibration solution from the following and run
41
range of standard sample (V flag). the process.
①Dilute and rerun.
②Rerun without dilute.
①Check position of calibrator.
②Check "Calibration CV (%)" in
Serious difference between previous result and
42 "Configuration" screen. If the
current result.
value is small, input 20-30.
③Draw calibration again.
①Do Auto-gain again.
43 Absorbance is high (too dark). Lamp may be out. ②If the problem is not solved,
change the lamp to new one.
44 Abnormality of calculation item.
This is calculation error. Check
45 Abnormality of calculation item. the following points.
① Result value of each item.
46 Abnormality of calculation item. (Check reaction procedure whether
reaction is correct seeing reaction
There is no item for calculation in calibration
47 monitor.)
file.
② Parameter of calculation item.
48 Abnormality of calculation item. (Whether value is input
correctly.)
49 There is no item information. After checking above two points,
rerun.
50 There is no specimen kind.
There is no item in item parameter against
Cancel request from host. Check
51 request from host. Which host requested in item
item parameter.
parameter.
52 There is no item which is able to measure. Check remain of reagent.
After checking of reagent
There is no item in Calib and Round against
53 information and round setting, draw
request from host.
calibration again.
6-14
Operator’s Manual Chapter 6
Ref
No.
Cause Solution
It is impossible to measure because of short of Remaining amount of reagent is
54
test count. nothing. Add reagent.
55 Reagent expired. Change reagent.
56 Power has been turned OFF. Check power of analyzer.
Abnormality of communication data from
57
analyzer.
Abnormality of communication data from
58
analyzer.
Check cable between analyzer and
59 No response from analyzer.
PC. MCPU PCB may be broken. If the
Abnormality of communication data from problem is not solved, please call
60
analyzer. our service division.
Abnormality of communication data from
61
analyzer.
Communication error occurs at measurement
61’
result messages while measurement.
Other software uses the port.
Check the following two points.
①Restart PC and start up only
62 Serial port has not been opened. analyzer software.
②Check port number on device
manager in Windows and correspond
the number with analyzer.
63 Printer has been turned OFF. Turn ON printer.
64 In case of the skip cells are above 20. Please change cells.
Conversion from external representation to
This is system error. Please call
65 internal representation is impossible. ( LAN
our service division.
transfer)
Please prepare new Lamp. Please
66 The exchange time of Lamp approaches.
change it as soon as a lamp is OFF.
Please check the error code ins ISE
67 There are some problem during measurement. error code list.
(Please see ISE error code list)
You may have measured other sample except urine, Please check sample type in sample
68
serum, dialysis. order screen.
The urine calibration parameter is an error
69 which was sent to the ISE unit after the system Please calibrate urine.
start.
Please change Na electrode and
70 The lifetime of Na electrode is expired. change the date in ISE maintenance
screen.
Please change K electrode and
71 The lifetime of K electrode is expired. change the date in ISE maintenance
screen.
Please change Cl electrode and
72 The lifetime of Cl electrode is expired. change the date in ISE maintenance
screen.
Please change Ref electrode and
73 The lifetime of Ref electrode is expired. change the date in ISE maintenance
screen.
6-15
V5.7
Ref
No.
Cause Solution
Please change ISE pump and change
74 The lifetime of ISE pump is expired.
the date in ISE maintenance screen.
This message is displayed when the data was not After finishing of ISE movement,
75
able to send to ISE unit. please do "Ready" again.
The exchange time of the Na electrode
76 Please prepare new Na electrode.
approaches.
77 The exchange time of the K electrode approaches. Please prepare new K electrode.
The exchange time of the Cl electrode
78 Please prepare new Cl electrode.
approaches.
The exchange time of the Ref electrode
79 Please prepare new Ref electrode.
approaches.
80 The exchange time of the Calibrator1 approaches. Please prepare new Calibrator1.
81 The exchange time of the Calibrator2 approaches. Please prepare new Calibrator2.
82 ASTM communication open error
83 ASTM communication breakdown
84 ASTM - Send message unable
85 ASTM - Timeout
86 ASTM - NAK received
87 ASTM - Over record length
88 ASTM - Measurement request registration error
ASTM - Requested sample is during measurement or
89
completed the measurement
90 ASTM - Position error
91 ASTM - No request specified Contact your host software company.
92 ASTM - No Patient information
93 ASTM - Item error
94 ASTM - Abnormal message error
95 ASTM - Position check error
96 ASTM - Short field
97 ASTM - Delimiter split error in message
98 ASTM - Existence check error
99 ASTM - Lens check error
100 ASTM - Value check error
101 ASTM - Retry over
You may have washed the ISE despite not the
102 Please do "Ready" and wash the ISE.
"Ready".
You may have calibrate ISE despite not the Please do "Ready" and calibrate the
103
"Ready". ISE.
①Please do "Ready".
②Once turn OFF the PC and the
104 ISE probe is not able to go back to home position. analyzer and then turn ON again.
③If the problem is not solved,
please call our service division.
After finish in purge or
105 ISE automatically purges or maintains.
maintenance, please retry.
Check cup set position. Set cup in
106 There is no cup in designated position.
correct position.
6-16
Operator’s Manual Chapter 6
Ref
No.
Cause Solution
107 You may have operate a motor which is undefined.
Designated position error of motor movement
108 Please call our service division.
table.
109 Index error of motor movement table.
Please wait until ISE movement
finish.
110 The communication with ISE unit is impossible. After finishing the movement,
please do "Ready" and check whether
ISE unit move correctly.
Default setting of sample barcode reader has not There are some problems in a cable
111
been succeeded. connected with sample barcode
Default setting of reagent barcode reader has reader.
112
not been succeeded. Please call our service division.
The causes are as following.
①PCB may be broken.
113
②The line between level sensor and PCB has been
broken.
The causes are as following.
①PCB may be broken.
114
②The line between level sensor and PCB has been
broken. MCPU PCB may be broken.
The causes are as following. Please call our service division.
①PCB may be broken.
115
②The line between level sensor and PCB has been
broken.
The causes are as following.
①PCB may be broken.
116
②The line between level sensor and PCB has been
broken.
117 System error of MCPU
118 System error of MCPU MCPU PCB may be broken.
119 System error of MCPU Please call our service division.
120 System error of MCPU
6-17
V5.7
Error Ref.
No. Code
Contents No.
In measurement of sample or 1
1 100000 Noise error of Na electrode :
calibrator2
In measurement of sample or
2 200000 Noise error of K electrode :
calibrator2
In measurement of sample or
3 300000 Noise error of Na, K electrodes :
calibrator2
In measurement of sample or
4 400000 Noise error of Cl electrode :
calibrator2
In measurement of sample or
5 500000 Noise error of Na, Cl electrodes :
calibrator2
In measurement of sample or
6 600000 Noise error of K, Cl electrodes :
calibrator2
In measurement of sample or
7 700000 Noise error of Na, K, Cl electrodes :
calibrator2
8 S00000 Air detection at flow sensor : In measurement of sample 2
In measurement of
9 B00000 Air detection at flow sensor :
Calibrator2
In measurement of sample or 3
10 E00000 Poor drainage detection at flow sensor :
Calibrator2
In measurement of 4
11 010000 Noise error of Na electrode :
Calibrator1
In measurement of
12 020000 Noise error of K electrode :
Calibrator1
In measurement of
13 030000 Noise error of Na, K electrodes :
Calibrator1
In measurement of
14 040000 Noise error of Cl electrode :
Calibrator1
In measurement of
15 050000 Noise error of Na, Cl electrodes :
Calibrator1
In measurement of
16 060000 Noise error of K, Cl electrodes :
Calibrator1
In measurement of
17 070000 Noise error of Na, K, Cl electrodes :
Calibrator1
In measurement of 5
18 0A0000 Air detection at flow sensor :
Calibrator1
In measurement of 6
19 0E0000 Poor drainage detection at flow sensor :
Calibrator1
In measurement of 7
20 001000 Potential drift error at Na electrode :
Calibrator1
In measurement of
21 002000 Potential drift error at K electrode :
Calibrator1
Potential drift error at Na, K In measurement of
22 003000 :
electrodes Calibrator1
In measurement of
23 004000 Potential drift error at Cl electrode :
Calibrator1
6-18
Operator’s Manual Chapter 6
Error Ref.
No. Code
Contents No.
Potential drift error at Na, Cl In measurement of 7
24 005000 :
electrodes Calibrator1
In measurement of
25 006000 Potential drift error at K, Cl electrodes :
Calibrator1
Potential drift error at Na, K, Cl In measurement of
26 007000 :
electrodes Calibrator1
27 000100 Potential overshooting of Na electrode : (OUT OF RANGE) 8
28 000200 Potential overshooting of K electrode : (OUT OF RANGE)
Potential overshooting of Na, K
29 000300 : (OUT OF RANGE)
electrodes
30 000400 Potential overshooting of Cl electrode : (OUT OF RANGE)
Potential overshooting of Na, Cl
31 000500 : (OUT OF RANGE)
electrodes
Potential overshooting of K, Cl
32 000600 : (OUT OF RANGE)
electrodes
Potential overshooting of Na, K, Cl 9
33 000700 : (OUT OF RANGE)
electrodes
(Out of slope range, slope 10
34 000010 Calibration slope error of Na electrode :
drift)
(Out of slope range, slope
35 000020 Calibration slope error of K electrode :
drift)
Calibration slop error of Na, K (Out of slope range, slope
36 000030 :
electrodes drift)
(Out of slope range, slope
37 000040 Calibration slope error of Cl electrode :
drift)
Calibration slope error of Na, Cl (Out of slope range, slope
38 000050 :
electrodes drift)
Calibration slope error of K, Cl (Out of slope range, slope
39 000060 :
electrodes drift)
Calibration slope error of Na, K, Cl (Out of slope range, slope
40 000070 :
electrodes drift)
Pulse over error of return movement to 11
41 000001
home position (nozzle X axis)
Pulse over error of return movement to 12
42 000002
home position (nozzle Z axis)
Timeout error of movement (nozzle X 11
43 000003
axis)
Timeout error of movement (nozzle Z 12
44 000004
axis)
45 000007 Not movement error of nozzle X axis 13
46 000008 IC error of pulse motor controller 14
47 000009 IC error of pulse motor driver
48 00000N Samp_OK Timeout error 15
6-19
V5.7
Ref
No.
Error code Causes Solutions
1 100000 Air bubble in sample line Make sure there is enough sample in cup.
200000 Confirm positioning of sampling probe and
300000 sample cup.
400000 Poor L-GND function Wipe off inside of L-GND pipe by thread.
500000 If the above action is not effective, change
600000 L-GND pipe.
700000 Dirty sample tube Clean sample tube by ISE cleaning solution,
(Propagation of noise) supplied from us. If above cleaning is not
effective, replace sample tube with new one.
Dirty drainage tube Exchange drainage tubes with a new tube.
If drainage trough is dirty, clean up the trough
thoroughly.
Electrode is damaged. If error occurs for particular items such as
100000, 200000, 400000, replace with the new
corresponding electrode.
If error occurs for all items (700000), replace
with new Reference electrode.
Drying of the electrode If the electrodes are dry, it takes time to
recover.
If recovery is impossible during 1-2 hours,
please change the electrodes.
2 S00000 No sample or short of Make sure if there is enough sample in cup.
B00000 sample Make sure if sampling nozzle is clogging.
Check Cal-2 residual volume in the bag. If
the volume is too small, replace it with new
bag.
Pump tube degradation cause Replace pump tube with new one.
poor liquid transfer Change pump tube.
Pinch valve tube degradation Replace pinch valve tube with new one.
cause poor liquid transfer. Change pinch valve tube.
Liquid flow line clogging If sampling nozzle or electrode inlet is clogged,
clean using fishing line or mandolin string.
Make sure if piping tube is crushed.
If piping tube is clogging, change tube.
6-20
Operator’s Manual Chapter 6
Ref
No.
Error code Causes Solutions
6-21
V5.7
Ref
No.
Error code Causes Solutions
4 010000 Defective or degraded electrode If error occurs for particular items like 01000,
020000 02000, 04000, change the electrode.
030000 If error occurs for all items (07000), change
040000 Reference electrode.
050000
Dried electrode Once electrode is dried, it takes a couple of
060000
hours to recover. Wait 1 to 2 hours, and try it
070000
again.
If it is not successful, replace electrode.
5 0A0000 Check if Cal-1 bag is connected. If not
No Cal-1 bag or short
connect it. Check Cal-1 volume in the bag.
Cal-1 volume in the bag.
If the volume is too small, replace it.
Short sample, caused by pump
Trial pump tube replacement.
tube degradation
Short sample, caused by pinch
Trial pinch valve tube replacement
valve tube degradation
Liquid flow line clogging
If sampling nozzle or electrode inlet is clogged,
clean using fishing line or mandolin string.
Make sure if piping tube is crushed.
If piping tube is clogging, change tube.
6-22
Operator’s Manual Chapter 6
Ref
No.
Error code Causes Solutions
6 0E000 Flow sensor is not Working To recover flow sensor function, do the
properly. following action. Exceed command <PURG>,
and do air level setting.
Clean inlet and outlet holes of flow sensor
using thread.
Flow sensor is damaged Change a flow sensor or sensors.
Or change tube of flow sensor and reagent pot.
Poor connector electric Pull out connectors of pump and pinch valves,
conductivity. and insert them again.
Peristaltic pump has trouble. Change roller pump.
If roller bearing fixing screws are loosen,
tighten them. If the above action is not
effective replace pump
Pinch valve is not working. Replace pinch valve itself.
7 001000 Bubble in Cal-1 If there is not enough Cal-1, change Cal-1.
002000 If there is bubble in Cal-1 tube, perform
003000 <PURG>.
004000 Poor electrode function Clean electrode using proper cleaning
005000 solution, supplied from us.
006000
007000 Defective or degraded electrode If error occurs for particular items like
001000, 002000, 004000, change the electrode.
If error occurs for all items (007000), change
Re electrode.
Dried electrode Once electrode is dried, it takes a couple of
hours to recover.
Wait 1 to 2 hours, and try it again. If it is not
successful, replace electrode.
Poor L-GND function Clean inside of L-GND pipe by a thread.
If it is not effective, replace L-GND pipe.
Unstable temperature Do not use Cal-1, which is just taken out from
refrigerator. Leave it to be room
temperature.
Do not operate analyzer when the room
temperature or analyzer temperature are
changing rapidly.
8 000100 Air bubble in sample line Make sure there is enough sample in cup.
000200 Confirm positioning of sampling probe and
000300 sample cup.
000400 Check sample level in the sample cup. Check
000500 if sample probe tip goes down into sample
000600 surly.
000700
Air bubble in Cal-1 line Check Cal-1 volume in the bag. If the
residual volume is too small, replace the bag.
If you find an air bubble in Cal-1 line, perform
<PURG>.
6-23
V5.7
Ref
No.
Error code Causes Solutions
6-24
Operator’s Manual Chapter 6
Power on LED 24V power is not supplied. Check a cable and connector.
(green) of pump (Poor connect of power supply Change ISE unit or power supply
part is not part) part.
working. Disconnection of fuse Change fuse on driver PCB.
Result become 0
Calibration is not finished. Do calibration.
without error.
6-25
V5.7
6-26
V5.7
Packing
Code Parts Name Remarks
Qty.
50-24-0747 Alkaline Washing Solution 2.5D (1000mL) 4
20-27-0474F Na Electrode F 1
20-27-0476F Cl Electrode F 1
Packing
Code Parts Name Remarks
Qty.