Sei sulla pagina 1di 12

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles

for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

Designation: D7181 − 20

Standard Test Method for


Consolidated Drained Triaxial Compression Test for Soils1
This standard is issued under the fixed designation D7181; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

1. Scope of this test standard to consider significant digits used in


1.1 This test method covers the determination of strength analysis methods for engineering design.
and stress-strain relationships of a cylindrical specimen of 1.7 Units—The values stated in SI units are to be regarded
either intact or reconstituted soil. Specimens are consolidated as standard. The inch-pound units given in parentheses are
and sheared in compression with drainage at a constant rate of mathematical conversions, which are provided for information
axial deformation (strain controlled). purposes only and are not considered standard. Reporting of
1.2 This test method provides for the calculation of princi- test results in units other than SI shall not be regarded as
pal stresses and axial compression by measurement of axial non-conformance with this test method.
load, axial deformation, and volumetric changes. 1.7.1 The gravitational system of inch-pound units is used
when dealing with inch-pound units. In this system, the pound
1.3 This test method provides data useful in determining (lbf) represents a unit of force (weight), while the unit for mass
strength and deformation properties such as Mohr strength is slugs. The slug unit is not given, unless dynamic (F = ma)
envelopes. Generally, three specimens are tested at different calculations are involved.
effective consolidation stresses to define a strength envelope. 1.7.2 It is common practice in the engineering/construction
The stresses should be specified by the engineer requesting the profession to concurrently use pounds to represent both a unit
test. A test on a new specimen is required for each consolida- of mass (lbm) and of force (lbf). This implicitly combines two
tion stress. separate systems of units: that is, the absolute system and the
1.4 If this test method is used on cohesive soil, a test may gravitational system. It is scientifically undesirable to combine
take weeks to complete. the use of two separate sets of inch-pound units within a single
1.5 The determination of strength envelopes and the devel- standard. As stated, this standard includes the gravitational
opment of relationships to aid in interpreting and evaluating system of inch-pound units and does not use/present the slug
test results are beyond the scope of this test method and must unit for mass. However, the use of balances or scales recording
be performed by a qualified, experienced professional. pounds of mass (lbm) or recording density in lbm/ft3 shall not
be regarded as non-conformance with this standard.
1.6 All observed and calculated values shall conform to the 1.7.3 The terms density and unit weight are often used
guidelines for significant digits and rounding established in interchangeably. Density is mass per unit volume whereas unit
Practice D6026. weight is force per unit volume. In this standard density is
1.6.1 The procedures used to specify how data are collected, given only in SI units. After the density has been determined,
calculated, or recorded in this standard are regarded as the the unit weight is calculated in SI or inch-pound units, or both.
industry standard. In addition, they are representative of the
1.8 This standard may involve hazardous materials,
significant digits that generally should be retained. The proce-
operations, and equipment. This standard does not purport to
dures used do not consider material variations, purpose for
address all of the safety concerns, if any, associated with its
obtaining the data, special purpose studies or any consideration
use. It is the responsibility of the user of this standard to
for the user’s objectives; and it is common practice to increase
establish appropriate safety, health, and environmental prac-
or reduce the significant digits of the reported data to be
tices and determine the applicability of regulatory limitations
commensurate with these considerations. It is beyond the scope
prior to use.
1.9 This international standard was developed in accor-
1
This test method is under the jurisdiction of ASTM Committee D18 on Soil and dance with internationally recognized principles on standard-
Rock and is the direct responsibility of Subcommittee D18.05 on Strength and ization established in the Decision on Principles for the
Compressibility of Soils.
Current edition approved Jan. 1, 2020. Published February 2020. Originally
Development of International Standards, Guides and Recom-
approved in 2011. Last previous edition approved in 2011 as D7181 - 11. DOI: mendations issued by the World Trade Organization Technical
10.1520/D7181-20. Barriers to Trade (TBT) Committee.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
1
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
2. Referenced Documents 3.2.4 failure, n—a maximum-stress condition or stress at a
2.1 ASTM Standards: 2 defined strain for a test specimen.
D653 Terminology Relating to Soil, Rock, and Contained 3.2.4.1 Discussion—Failure is often taken to correspond to
Fluids the maximum principal stress difference (maximum deviator
D854 Test Methods for Specific Gravity of Soil Solids by stress) attained or the principal stress difference (deviator
Water Pycnometer stress) at 15 % axial strain, whichever is obtained first during
D1587 Practice for Thin-Walled Tube Sampling of Fine- the performance of a test. Depending on soil behavior and field
Grained Soils for Geotechnical Purposes application, other suitable failure criteria may be defined, such
D2216 Test Methods for Laboratory Determination of Water as maximum effective stress obliquity, σ1’/σ3max’, or the
(Moisture) Content of Soil and Rock by Mass principal stress difference (deviator stress) at a selected axial
D2435 Test Methods for One-Dimensional Consolidation strain other than 15 %.
Properties of Soils Using Incremental Loading 4. Summary of Test Method
D2487 Practice for Classification of Soils for Engineering
Purposes (Unified Soil Classification System) 4.1 The test specimen, either intact or reconstituted, is
D3740 Practice for Minimum Requirements for Agencies mounted in the testing apparatus using either a dry or wet
Engaged in Testing and/or Inspection of Soil and Rock as mounting procedure. The test specimen is cylindrical in shape
Used in Engineering Design and Construction and dimensions are measured prior to mounting. The test
D4220 Practices for Preserving and Transporting Soil specimen is then back pressure saturated. After saturation, the
Samples specimen is isotropically or anisotropically consolidated. The
D4318 Test Methods for Liquid Limit, Plastic Limit, and test specimen is then axially loading at a constant rate and with
Plasticity Index of Soils the drainage lines open to allow the sample to drain.
D4753 Guide for Evaluating, Selecting, and Specifying Bal- 5. Significance and Use
ances and Standard Masses for Use in Soil, Rock, and
5.1 The shear strength of a saturated soil in triaxial com-
Construction Materials Testing
pression depends on the stresses applied, time of consolidation,
D4767 Test Method for Consolidated Undrained Triaxial
strain rate, and the stress history experienced by the soil.
Compression Test for Cohesive Soils
D6026 Practice for Using Significant Digits in Geotechnical 5.2 In this test method, the shear characteristics are mea-
Data sured under drained conditions and are applicable to field
D6913 Test Methods for Particle-Size Distribution (Grada- conditions where soils have been fully consolidated under the
tion) of Soils Using Sieve Analysis existing normal stresses and the normal stress changes under
D7263 Test Methods for Laboratory Determination of Den- drained conditions similar to those in the test method.
sity (Unit Weight) of Soil Specimens 5.3 The shear strength determined from this test method can
D7928 Test Method for Particle-Size Distribution (Grada- be expressed in terms of effective stress because a strain rate or
tion) of Fine-Grained Soils Using the Sedimentation load application rate slow enough to allow pore pressure
(Hydrometer) Analysis dissipation during shear is used to result in negligible excess
3. Terminology pore pressure conditions. The shear strength may be applied to
field conditions where full drainage can occur (drained
3.1 Definitions: conditions), and the field stress conditions are similar to those
3.1.1 For definitions of common technical terms, refer to in the test method.
Terminology D653.
5.4 The shear strength determined from the test can be used
3.2 Definitions of Terms Specific to This Standard: in embankment stability analyses, earth pressure calculations,
3.2.1 back pressure, n—a pressure applied to the specimen and foundation design.
pore-water to cause air in the pore space to compress and to NOTE 1—The quality of the result produced by this standard is
pass into solution in the pore-water thereby increasing the dependent on the competence of the personnel performing it and the
percent saturation of the specimen. suitability of the equipment and facilities used. Agencies that meet the
criteria of Practice D3740 are generally considered capable of competent
3.2.2 effective consolidation stress, n—the difference be- and objective testing/sampling/inspection/etc. Users of this standard are
tween the cell pressure and the pore-water pressure prior to cautioned that compliance with Practice D3740 does not assure reliable
shearing the specimen. results. Reliable results depend on many factors; Practice D3740 provides
a means of evaluating some of those factors.
3.2.3 effective consolidation stresses, n—for anisotropic
(9.4), the vertical and lateral stress applied magnitudes are not 6. Apparatus
equal by design, with lateral stress equal to the cell pressure 6.1 The requirements for equipment needed to perform
minus pore-water pressure, and the vertical stress equal to the satisfactory tests are given in the following sections. See Fig.
desired total vertical stress (9.4.2) minus the pore pressure. 1

2
6.2 Axial Loading Device—The axial loading device may be
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
a screw jack driven by an electric motor through a geared
Standards volume information, refer to the standard’s Document Summary page on transmission, a hydraulic loading device, or any other com-
the ASTM website. pression device with sufficient capacity and control to provide

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
2
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20

FIG. 1 Schematic Diagram of a Typical Consolidated Drained Triaxial Apparatus

the rate of axial strain (loading) prescribed in 9.5.2. The rate of supplied to the chamber and inlets leading to the specimen base
advance of the loading device should not deviate by more than and provide for connection to the cap to allow saturation and
61 % from the selected value. Vibration due to the operation drainage of the specimen when needed.
of the loading device shall be sufficiently small to not cause 6.5 Axial Load Piston—The piston passing through the top
dimensional changes in the specimen. of the chamber and its seal must be designed so the axial load
NOTE 2—A loading device may be judged to produce sufficiently small
vibrations if there are no visible ripples in a glass of water placed on the
due to friction does not exceed 0.5 % of the load on piston at
loading platform when the device is operating at the speed at which the failure and so there is negligible lateral bending of the piston
test is performed. during loading. For triaxial cell with internal load, cell piston
6.3 Axial Load-Measuring Device—The axial load- friction is not as important.
measuring device shall be an electronic load cell, hydraulic NOTE 3—The use of two linear ball bushings to guide the piston is
recommended to reduce friction and maintain alignment.
load cell, or any other load-measuring device capable of the NOTE 4—A minimum piston diameter of 1⁄6 the specimen diameter has
accuracy prescribed in this paragraph and may be a part of the been used successfully in many laboratories to reduce lateral bending.
axial loading device. The axial load-measuring device shall be 6.6 Pressure and Vacuum-Control Devices—The chamber
capable of measuring the axial load to an accuracy of within pressure and back pressure control devices shall be (a) capable
1 % of the axial load at failure. If the load-measuring device is of applying and controlling pressures to within 62 kPa (0.25
located inside the triaxial compression chamber, it shall be lbf/in.2) for effective consolidation pressures less than 200 kPa
insensitive to horizontal forces and to the magnitude of the (28 lbf/in.2) and to within 61 % for effective consolidation
chamber pressure. pressures greater than 200 kPa, and (b) able to maintain the
6.4 Triaxial Compression Chamber—The triaxial chamber effective consolidation stress within 2 % of the desired value
shall have a working chamber pressure capable of sustaining (Note 5). The vacuum-control device shall be capable of
the sum of the effective consolidation stress and the back applying and controlling partial vacuums to within 62 kPa.
pressure. It shall consist of a top plate and a base plate The devices may consist of pneumatic-pressure regulators,
separated by a cylinder. The cylinder may be constructed of combination pneumatic pressure and vacuum regulators, or any
any material capable of withstanding the applied pressures. It is other device capable of applying and controlling pressures or
desirable to use a transparent material or have a cylinder partial vacuums to the necessary tolerances. These tests can
provided with viewing ports so the behavior of the specimen have a duration of several days, therefore, an external air/water
may be observed. The top plate shall have a vent valve such interface is recommended for both the chamber-pressure or
that air can be forced out of the chamber as it is filled. The base back-pressure systems.
plate shall have an inlet through which the pressure liquid is NOTE 5—Many laboratories use differential pressure regulators and

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
3
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
transducers to achieve the requirements for small differences between surface of the specimen base and cap that contacts the
chamber and back pressure. membrane to form a seal shall be smooth and free of scratches.
6.7 Pressure- and Vacuum-Measurement Devices—The 6.11 Porous Disks—A rigid porous disk shall be used to
chamber pressure-, back pressure-, and vacuum-measuring provide drainage at each end of the specimen. The coefficient
devices shall be capable of measuring the ranges of pressures of permeability of the disks shall be greater than that of fine
or partial vacuums to the tolerances given in 6.6. They may sand (1 × 10-4 cm/s (4 × 10–5 in./s)). The disks shall be
consist of electronic pressure transducers, or any other device regularly cleaned by ultrasonic or boiling and brushing and
capable of measuring pressures, or partial vacuums to the checked to determine whether they have become clogged.
stated tolerances. If separate devices are used to measure the
6.12 Filter-Paper Strips and Disk—Filter-paper strips are
chamber pressure and back pressure, the devices must be
used by many laboratories to decrease the time needed for
normalized simultaneously and against the same pressure
testing. Filter-paper disks of a diameter equal to that of the
source. Since the chamber and back pressure are the pressures
specimen may be placed between the porous disks and speci-
taken at the midheight of the specimen, it may be necessary to
men to avoid clogging of the porous disks. If filter strips or
adjust the zero-offset of the devices to reflect the hydraulic
disks are used, they shall be of a type that does not dissolve in
head of fluids in the chamber and back pressure control
water. The coefficient of permeability of the filter paper shall
systems.
not be less than 1 × 10-5 cm/s (4 × 10-6 in./s) for a normal
6.8 Volume Change Measurement Device—The volume of pressure of 550 kPa (80 lbf/in.2). To avoid hoop tension, filter
water entering or leaving the specimen shall be measured with strips should cover no more than 50 % of the specimen
an accuracy of within 60.05 % of the total volume of the periphery. Many laboratories have successfully used filter strip
specimen. The volume-measuring device is usually a burette cages. An equation for correcting the principal stress difference
connected to the back pressure but may be any other device (deviator stress) for the effect of the strength of vertical filter
meeting the accuracy requirement. The device must be able to strips is given in 11.3.3.1.
withstand the maximum back pressure and of sufficient capac- NOTE 6—Grade No. 54 Filter Paper has been found to meet the
ity for the performance of the test. Volume changes during permeability and durability requirements.
shear are often on the order of 620 % or more of the specimen 6.13 Rubber Membrane—The rubber membrane used to
volume. Either allowing for resetting of the system during encase the specimen shall provide reliable protection against
shear or having a total capacity capable of measuring the entire leakage. Membranes shall be carefully inspected prior to use
change may meet the needed capacity. and if any flaws or pinholes are evident, the membrane shall be
discarded. To offer minimum restraint to the specimen, the
6.9 Deformation Indicator—The vertical deformation of the
unstretched membrane diameter shall be between 90 and 95 %
specimen is usually determined from the travel of the piston
of that of the specimen. The membrane thickness shall not
acting on the top of the specimen. The piston travel shall be
exceed 1 % of the diameter of the specimen. The membrane
measured with an accuracy of at least 0.25 % of the initial
shall be sealed to the specimen cap and base with rubber
specimen height. The deformation indicator shall have a range
O-rings for which the unstressed inside diameter is between 75
of at least 20 % of the initial height of the specimen and may
and 85 % of the diameter of the cap and base, or by other
be a dial indicator, linear variable differential transformer
means that will provide a positive seal. An equation for
(LVDT), extensometer, or other measuring device meeting the
correcting the principal stress difference (deviator stress) for
requirements for accuracy and range.
the effect of the stiffness of the membrane is given in 11.3.3.2.
6.10 Specimen Cap and Base—The specimen cap and base 6.14 Valves—Changes in volume due to opening and closing
shall be designed to provide drainage from both ends of the valves may result in inaccurate volume change and pore-water
specimen. They shall be constructed of a rigid, noncorrosive, pressure measurements. For this reason, valves in the specimen
impermeable material, and each shall, except for the drainage drainage system shall be of the type that produces minimum
provision, have a circular plane surface of contact with the volume changes due to their operation. A valve may be
porous disks and a circular cross section. It is desirable for the assumed to produce minimum volume change if opening or
mass of the specimen cap and top porous disk to be as minimal closing the valve in a closed, saturated pore-water pressure
as possible. However, the mass may be as much as 10 % of the system does not induce a pressure change of greater than 0.7
axial load at failure. If the mass is greater than 0.5 % of the kPa (60.1 lbf/in.2). All valves must be capable of withstanding
applied axial load at failure and greater than 50 g (0.1 lb), the applied pressures without leakage.
axial load must be corrected for the mass of the specimen cap NOTE 7—Ball valves have been found to provide minimum volume-
and top porous disk. The diameter of the cap and base shall be change characteristics; however, any other type of valve having suitable
equal or only nominally greater than the initial diameter of the volume-change characteristics may be used.
specimen. The specimen base shall be connected to the triaxial 6.15 Specimen-Size Measurement Devices—Devices used to
compression chamber to prevent lateral motion or tilting, and determine the height and diameter of the specimen shall
the specimen cap shall be designed such that eccentricity of the measure the respective dimensions to four significant digits and
piston-to-cap contact relative to the vertical axis of the speci- shall be constructed such that their use will not disturb/deform
men does not exceed 1.3 mm (0.05 in.). The end of the piston the specimen.
and specimen cap contact area shall be designed so that tilting NOTE 8—Circumferential measuring tapes are recommended over
of the specimen cap during the test is minimal. The cylindrical calipers for measuring the diameter.

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
4
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
6.16 Data Acquisition—Specimen behavior may be re- change is kept to a minimum. Where removal of pebbles or
corded manually or by electronic digital or analog recorders. If crumbling resulting from trimming causes voids on the surface
electronic data acquisition is used, it shall be necessary to of the specimen, carefully fill the voids with remolded soil
calibrate the measuring devices through the recording device obtained from the trimmings. If the sample can be trimmed
using known input standards. with minimal disturbance, a vertical trimming lathe may be
6.17 Timer—A timing device indicating the elapsed testing used to reduce the specimen to the necessary diameter. After
time to the nearest 1 s readability, shall be used to obtain obtaining the necessary diameter, place the specimen in a miter
consolidation data (9.3.3). box, and cut the specimen to the final height with a wire saw
or other suitable device. Trim the surfaces with the steel
6.18 Balance—A balance or scale conforming to the re- straightedge. Perform one or more water content determina-
quirements of Specification D4753 readable to four significant tions on material trimmed from the specimen in accordance
digits. with Test Method D2216. Determine and record the mass and
6.19 Water Deaeration Device—The amount of dissolved dimensions of the specimen using the devices described in 6.16
gas (air) in the water used to saturate the specimen shall be and 6.20. A minimum of three height measurements (120°
decreased by boiling, by heating and spraying into a vacuum, apart) and at least three diameter measurements at the quarter
or by any other method that will satisfy the requirement for points of the height shall be made to determine the average
saturating the specimen within the limits imposed by the height and diameter of the specimen.
available maximum back pressure and time to perform the test. 7.3 Reconstituted Specimens by Compaction—
6.20 Testing Environment—The consolidation and shear Reconstituted specimens shall be prepared at the conditions
portion of the test shall be performed in an environment where specified for the test. Soil used for a reconstituted specimen
temperature fluctuations are less than 64 °C (67.2 °F) and shall be thoroughly mixed with enough water to produce the
there is no direct exposure with sunlight. desired water content. If water is added to the soil, store the
6.21 Miscellaneous Apparatus—Specimen trimming and material in a covered container for at least 16 h prior to
carving tools including a wire saw, steel straightedge, miter compaction. Reconstituted specimen may be prepared by
box, vertical trimming lathe, apparatus for preparing reconsti- compacting material in at least six layers using a split mold of
tuted specimens, membrane and O-ring expander, water con- circular cross section having dimensions meeting the require-
tent cans, and data sheets shall be provided as necessary. ments enumerated in 7.1. Specimens may be compacted to the
desired density by either: (1) kneading or tamping each layer
7. Test Specimen Preparation until the accumulative mass of the soil placed in the mold is
compacted to a known volume; or (2) by adjusting the number
7.1 Specimen Size—Specimens shall be cylindrical and have of layers, the number of tamps per layer, and the force per
a minimum diameter of 33 mm (1.3 in.). The average-height- tamp. The top of each layer shall be scarified prior to the
to-average-diameter ratio shall be between 2 and 2.5. An addition of material for the next layer. The tamper used to
individual measurement of height or diameter shall not vary compact the material shall have a diameter equal to or less than
from average by more than 2 %. The largest particle size shall 1⁄2 the diameter of the mold. After a specimen is formed, with
be smaller than 1⁄6 the specimen diameter. If, after completion the ends perpendicular to the longitudinal axis, remove the
of a test, it is found based on visual observation that oversize mold and determine and record the mass and dimensions of the
particles are present, indicate this information in the report of specimen using the devices described in 6.14 and 6.17. A
test data (12.2.5). minimum of three height measurements (120° apart) and at
NOTE 9—If oversize particles are found in the specimen after testing, a
particle-size analysis may be performed on the tested specimen in least three diameter measurements at the quarter points of the
accordance with Test Method D6913 to confirm the visual observation. height shall be made to determine the average height and
7.2 Intact Specimens—Prepare intact specimens from large diameter of the specimen. Perform one or more water content
intact samples or from samples secured in accordance with determinations on excess material used to prepare the specimen
Practice D1587 or other acceptable intact tube sampling in accordance with Test Method D2216.
procedures. Samples shall be preserved and transported in NOTE 10—It is common for the density or unit weight of the specimen
accordance with the practice for Group C samples in Practices after removal from the mold to be less than the value based on the volume
D4220. Specimens obtained by tube sampling may be tested of the mold. This change occurs as a result of the specimen swelling after
removal of the lateral confinement provided by the mold.
without trimming except for cutting the end surfaces plane and
perpendicular to the longitudinal axis of the specimen, pro- 7.4 Reconstituted Specimens by Other Methods—Prepare
vided soil characteristics are such that no significant distur- reconstituted specimens in the manner specified by the request-
bance results from sampling. Handle specimens carefully to ing agency. These methods will usually require a forming
have negligible disturbance, change in cross section, or change jacket used to form the specimen directly on the base pedestal.
in water content. If compression or any type of noticeable The forming jacket will allow the specimen to be prepare in the
disturbance would be caused by the extrusion device, split the membrane. There are steps in Section 8 that may not be
sample tube lengthwise or cut the tube in suitable sections to required by some of the methods. Common methods include:
facilitate removal of the specimen with minimum disturbance. 7.4.1 Pluviation Through Water Method—For this specimen
Prepare trimmed specimen, in an environment such as a preparation method, a granular soil is saturated initially in a
controlled high-humidity room where soil water content container, poured through water into a water-filled membrane

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
5
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
placed on a forming mold, and then densified to the required 8.2 Depending on whether the saturation portion of the test
density by vibration; refer to reference by Chaney and Mullis.3 will be initiated with either a wet or dry drainage system,
mount the specimen using the appropriate method, as follows
NOTE 11—A specimen may be vibrated either on the side of the mold
or the base of the cell using a variety of apparatus. These include the in either 8.2.1 or 8.2.2. The dry mounting method is strongly
following: tapping with an implement of some type such as a spoon or recommended for specimens with initial saturation less than
metal rod, pneumatic vibrator, or electric engraving tool. 90 %. The dry mounting method removes air prior to adding
7.4.2 Dry Screening Method—For this method a tube with a backpressure and lowers the backpressure needed to attain an
screen attached to one end is placed inside a membrane adequate percent saturation.
stretched over a forming mold. A dry uniform sand is then NOTE 12—It is recommended that the dry mounting method be used for
poured into the tube. The tube is then slowly withdrawn from specimens of soils that swell appreciably when in contact with water. If
this membrane/mold allowing the sand to pass through the the wet mounting method is used for such soils, it will be necessary to
screen forming a specimen. If a greater density of the sand is obtain the specimen dimensions after the specimen has been mounted. In
such cases, it will be necessary to determine the double thickness of the
desired the mold may be vibrated. membrane, the double thickness of the wet filter paper strips (if used), and
7.4.3 Dry or Moist Vibration Method—In this procedure the combined height of the cap, base, and porous disks (including the
compact oven-dried, or moist granular material in layers thickness of filter disks if they are used) so that the appropriate values may
(typically six to seven layers) in a membrane-lined split mold be subtracted from the measurements.
attached to the bottom platen of the triaxial cell. Compact the 8.2.1 Wet Mounting Method:
material for each lift by vibration to the dry unit weight 8.2.1.1 Fill the specimen drainage lines and the pore-water
necessary to obtain the prescribed density. Scarify the soil pressure measurement device with deaired water.
surface between lifts. It should be noted that to obtain uniform 8.2.1.2 Saturate the porous disks by boiling them in water
density, the bottom layers have to be slightly under compacted, for at least 10 min and allow to cool to room temperature.
since compaction of each succeeding layer increases the 8.2.1.3 Place a saturated porous disk on the specimen base
density of sand in layers below it. After the final layer is and after wiping away all free water on the disk, place the
partially compacted, put the top cap in place and continue specimen on the disk. Next, place another porous disk and the
vibration until the desired dry unit weight is obtained. specimen cap on top of the specimen. Check that the specimen
7.4.4 Tamping Method—For this procedure tamp air dry or cap, specimen, and porous disks are centered on the specimen
moist granular or cohesive soil in layers into a mold. The only base.
difference between the tamping method and the vibration
NOTE 13—If filter-paper disks are to be placed between the porous disks
method is that each layer is compacted by hand tamping with
and specimen, they should be dipped in water prior to placement.
a compaction foot instead of with a vibrator, refer to reference
by Ladd, R.S.4 8.2.1.4 If filter-paper strips or a filter-paper cage are to be
7.4.5 After the specimen has been formed, place the speci- used, saturate the paper with water prior to placing it on the
men cap in place and seal the specimen with O-rings or rubber specimen. To avoid hoop tension, do not cover more than 50 %
bands after placing the membrane ends over the cap and base. of the specimen periphery with vertical strips of filter paper.
Then apply a partial vacuum of 35 kPa (5 lbf/in.2) to the The filter paper should extend to porous disks on top and
specimen and remove the forming jacket. If the test confining- bottom of sample.
pressure is greater than 103 kPa (14.7 lbf/in.2), a full vacuum 8.2.1.5 Proceed with 8.3.
may be applied to the specimen in stages prior to removing the 8.2.2 Dry Mounting Method:
jacket. 8.2.2.1 Dry the specimen drainage system. This mounting
method may be accomplished by allowing dry air to flow
8. Mounting Specimen through the system prior to mounting the specimen.
8.1 Preparations—Before mounting the specimen in the 8.2.2.2 Dry the porous disks in an oven and then place the
triaxial chamber, make the following preparations: disks in a desiccator to cool to room temperature prior to
8.1.1 Inspect the rubber membrane for flaws, pinholes, and mounting the specimen.
leaks. 8.2.2.3 Place a dry porous disk on the specimen base and
8.1.2 Place the membrane on the membrane expander or, if place the specimen on the disk. Next, place a dry porous disk
it is to be rolled onto the specimen, roll the membrane on the and the specimen cap on the specimen. Check that the
cap or base. specimen cap, porous disks, and specimen are centered on the
8.1.3 Check the porous disks and specimen drainage tubes specimen base.
are not obstructed by passing air or water through the appro- NOTE 14—If desired, dry filter-paper disks may be placed between the
priate lines. porous disks and specimen.
8.1.4 Attach the pressure-control and volume-measurement 8.2.2.4 If filter-paper strips or a filter paper cage are to be
system and a pore-pressure measurement device to the cham- used, the cage or strips may be held in place by small pieces of
ber base. tape at the top and bottom.

3
8.3 Place the rubber membrane around the specimen and
Chaney, R., and Mulilis, J., “Wet Sample Preparation Techniques,” Geotech-
seal it at the cap and base with two rubber O-rings or other
nical Testing Journal, ASTM, 1978, pp. 107-108.
4
Ladd, R.S., “Preparing Test Specimens Using Under-Compaction,” Geotech- positive seal at each end. A thin coating of silicon grease on the
nical Testing Journal, ASTM, Vol. 1, No. 1, March, 1978, pp. 16-23. vertical surfaces of the cap and base will aid in sealing the

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
6
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
membrane. If filter-paper strips or a filter-paper cage are used, specimen. After approximately 10 min, allow deaired water to
do not apply grease to surfaces in contact with the filter paper. slowly percolate from the bottom to the top of the specimen
8.4 Attach the top drainage line and check the alignment of (Note 15).
the specimen and the specimen cap. If the dry mounting 9.2.1.1 There should always be a positive effective stress of
method has been used, apply a partial vacuum of approxi- at least 13 kPa (2 lbf/in.2) at the bottom of the specimen during
mately 35 kPa (5 lbf/in.2) (not to exceed the consolidation this part of the procedure. When water appears in the burette
stress) to the specimen through the top drainage line prior to connected to the top of the specimen, close the valve to the
checking the alignment. If there is any eccentricity, release the bottom of the specimen and fill the burette with deaired water.
partial vacuum, realign the specimen and cap, and then reapply Next, reduce the vacuum acting on top of the specimen through
the partial vacuum. If the wet mounting method has been used, the burette to atmospheric pressure while simultaneously
the alignment of the specimen and the specimen cap may be increasing the chamber pressure by an equal amount. This
checked and adjusted without the use of a partial vacuum. process should be performed slowly such that the difference
between the pore pressure measured at the bottom of the
9. Procedure specimen and the pressure at the top of the specimen should be
9.1 Prior to Saturation—After assembling the triaxial allowed to equalize. When the pore pressure at the bottom of
chamber, perform the following operations: the specimen stabilizes, proceed with back pressuring of the
9.1.1 When possible, bring the axial load piston into contact specimen pore-water as described in 9.2.3. To check for
with the specimen cap several times to permit proper seating equalization, close the drainage valves to the specimen and
and alignment of the piston with the cap. During this measure the pore pressure change until stable for at least 2 min.
procedure, take care not to apply an axial load to the specimen If the change is less than 5 % of the effective stress, the pore
exceeding 0.5 % of the estimated axial load at failure. When pressure can be assumed to be stabilized.
the piston is brought into contact, record the reading of the NOTE 15—For saturated clays, percolation may not be necessary, and
deformation indicator. water can be added simultaneously at both top and bottom.
9.1.2 Fill the chamber with the chamber liquid, being 9.2.2 Starting with Initially Saturated Drainage System—
careful to avoid trapping air or leaving an air space in the After filling the burette connected to the top of the specimen
chamber. with deaired water, apply a chamber pressure of 35 kPa (5
9.2 Saturation—The objective of the saturation phase of the lbf/in.2) or less and open the specimen drainage valves. When
test is to fill all voids in the specimen with water without the pore pressure at the bottom of the specimen stabilizes,
undesirable prestressing of the specimen, allowing the speci- according to the method described in 9.2.1.1, or when the
men to swell, or causing migration of fines. Saturation is burette reading stabilizes, back pressuring of the specimen
usually accomplished by applying back pressure to the speci- pore-water may be initiated.
men pore water to drive air into solution after saturating the 9.2.3 Applying Back Pressure—Simultaneously increase the
system by either: (1) applying vacuum to the specimen and dry chamber and back pressure in steps with specimen drainage
drainage system (lines, porous disks, pore-pressure device, valves opened so that deaired water from the burette connected
filter-strips or cage, and disks) and then allowing deaired water to the top and bottom of the specimen may flow into the
to flow through the system and specimen while maintaining the specimen. To avoid undesirable prestressing of the specimen
vacuum; or (2) saturating the drainage system by boiling the while applying back pressure, the pressures must be applied
porous disks in water and allowing water to flow through the incrementally with adequate time between increments to per-
system prior to mounting the specimen. It should be noted that mit equalization of pore-water pressure throughout the speci-
placing the air into solution is a function of both time and men. The size of each increment may range from 35 kPa (5
pressure. Accordingly, removing as much air as possible prior lbf/in.2), up to 140 kPa (20 lbf/in.2), depending on the
to applying back pressure will decrease the amount of air that magnitude of the desired effective consolidation stress, and the
will have to be placed into solution and will also decrease the percent saturation of the specimen just prior to the addition of
back pressure required for saturation. In addition, air remaining the increment. The difference between the chamber pressure
in the specimen and drainage system just prior to applying back and the backpressure during back pressuring should not exceed
pressure will go into solution much more readily if deaired 35 kPa (5 lbf/in.2) unless it is deemed necessary to control
water is used for saturation. The use of deaired water will also swelling of the specimen during the procedure. The difference
decrease the time and backpressure required for saturation. between the chamber and back pressure must also remain
Many procedures have been developed to accomplish satura- within 65 % when the pressures are raised and within 62 %
tion. The following are suggested procedures: when the pressures are constant. To check for equalization after
9.2.1 Starting with Initially Dry Drainage System—Increase application of a backpressure increment or after the full value
from partial vacuum acting on top of the specimen to the of backpressure has been applied, close the specimen drainage
maximum available vacuum. If the final effective consolidation valves and measure the change in pore-pressure over a 1-min
stress is less than the maximum partial vacuum, apply a lower interval. If the change in pore pressure is less than 5 % of the
vacuum to the chamber. The difference between the partial difference between the chamber pressure and the back
vacuum applied to the specimen and the chamber should never pressure, another back pressure increment may be added or a
exceed the effective consolidation stress for the test and should measurement may be taken of the pore pressure Parameter B
not be less than 35 kPa (5 lbf/in.2) to allow for flow through the (see 9.2.4) to determine if saturation is completed. Specimens

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
7
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
shall be considered to be saturated if the value of B is equal to 9.3.1 When the saturation phase of the test is completed,
or greater than 0.95, or if B remains unchanged with addition bring the axial load piston into contact with the specimen cap,
of backpressure increments. The B Parameter could also be and record the reading on the deformation indicator. During
checked following consolidation stage. this procedure, take care not to apply an axial load to the
specimen exceeding 0.5 % of the estimated axial load at
NOTE 16—Although the pore pressure Parameter B is used to determine
adequate saturation, the B-value is also a function of soil stiffness. If the
failure. If continuous deformation monitoring is not being
saturation of the sample is 100 %, the B-value measurement will decrease used, after recording the reading, raise the piston a small
with increasing soil stiffness. Therefore, when testing stiff soil samples, a distance above the specimen cap, and lock the piston in place.
B-value of 95 % or even below may indicate a saturation approaching 9.3.2 With the specimen drainage valves closed, hold the
100 %.
maximum back pressure constant and increase the chamber
NOTE 17—The back pressure needed to saturate a specimen may be
higher for the wet mounting method than for the dry mounting method pressure until the difference between the chamber pressure and
because of the added difficulty of flushing out the air before back-pressure the back pressure equals the desired effective consolidation
saturation and may be as high as 1400 kPa (200 lbf/in.2). pressure. Consolidation to the final stress conditions may be
9.2.4 Measurement of the Pore Pressure Parameter performed. If continuous deformation monitoring is being
B—Determine the value of the pore pressure Parameter B in used, loads must be applied to the piston to keep it in contact
accordance with 9.2.4.1 – 9.2.4.4. The pore pressure Parameter with the specimen cap.
B is defined by the following equation: 9.3.3 Obtain an initial reading on the volume change device,
and then open appropriate drainage valves so that the specimen
∆u
B5 (1) may drain from both ends into the volume change device. At
∆σ cp
increasing intervals of elapsed time (0.1, 0.2, 0.5, 1, 2, 4, 8, 15,
where: and 30 min and at 1, 2, 4, and 8 h, and so forth) observe and
∆u = change in the specimen pore pressure that occurs as record the volume change readings, and, if not already doing
a result of a change in the chamber pressure when the so, after the 15-min reading, record the accompanying defor-
specimen drainage valves are closed, and mation indicator readings obtained by carefully bringing the
∆σcp = isotropic pressure change in the chamber pressure. piston in contact with the specimen cap. If volume change and
deformation indicator readings are to be plotted against the
9.2.4.1 Close the specimen drainage valves, record the pore square root of time, the time intervals at which readings are
pressure, and increase the chamber pressure. Commonly, an taken may be adjusted to those that have easily obtained square
increase of 70 kPa (10 lbf/in.2) is used. roots, for example, 0.09, 0.25, 0.49, 1, 4, and 9 min, and so
9.2.4.2 After approximately 2 min, determine and record the forth. Depending on soil type, time intervals may be changed
maximum value of the induced pore pressure. For many to convenient time intervals that allow for adequate definition
specimens, the pore pressure may decrease after the immediate of volume change versus time.
response and then increase slightly with time. If this occurs,
values of ∆u should be plotted with time and the asymptotic NOTE 18—In cases where significant amounts of fines may be washed
from the specimen because of high initial hydraulic gradients, it is
pore pressure used as the change in pore pressure. A large permissible to gradually increase the chamber pressure to the total desired
increase in ∆u with time or values of ∆u greater than ∆σ3 pressure over a period with the drainage valves open. If this is done,
indicate a leak of chamber fluid into the specimen. Decreasing recording of data should begin immediately after the total pressure is
values of ∆u with time may indicate a leak in that part of the reached.
pore pressure measurement system located outside of the 9.3.4 Plot the volume change and deformation indicator
chamber. readings versus either the logarithm or square root of elapsed
9.2.4.3 Calculate and record the B-value using Eq 1. time. Allow consolidation to continue for at least one log cycle
9.2.4.4 Reapply the same effective consolidation stress as of time or one overnight period after 100 % primary consoli-
existed prior to the B-value measurement by reducing the dation has been achieved as determined in accordance with one
chamber pressure or by, alternatively, increasing the back of the procedures outlined in Test Method D2435. A marked
pressure by the amount of the chamber pressure increase. If B deviation between the slopes of the volume change and
is continuing to increase with increasing back pressure, con- deformation indicator curves toward the end of consolidation
tinue with back pressure saturation. If B is equal to or greater based on deformation indicator readings indicates leakage of
than 0.95 or if a plot of B versus back pressure indicates no fluid from the chamber into the specimen, and the test should
further increase in B with increasing back pressure, initiate be terminated. The plot can be used to also determine t50 or t90.
consolidation.
9.4 Anisotropic Consolidation (CAD)—If required, the
9.3 Consolidation Isotropic (CID)—The objective of the specimen can be loaded with the axial loading device, applying
consolidation phase of the test is to allow the specimen to reach load at rate determined in 9.5.2. This stage follows the isotropic
equilibrium in a drained state at the effective consolidation consolidation and requires both the chamber and specimen
stress for which a strength determination is required. During drainage lines are open during the anisotropic consolidation.
consolidation, data is obtained for use in determining when Some equipment may require active operator interaction to
consolidation is complete and for computing a rate of strain to maintain the load constant prior to axial loading stage. When
be used for the shear portion of the test. The consolidation the specified vertical stress has been achieved, the axial load is
procedure is as follows: held constant.

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
8
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
9.4.1 Allow consolidation to continue for at least one log device or for piston friction. However, if an internal load-
cycle of time after 100 % primary consolidation has been measuring device of significant flexibility is used in combina-
achieved or overnight. tion with an external deformation indicator, correction of the
9.4.2 The load, P, to be applied to the piston to reach σvc’ deformation readings may be necessary. In both cases, record
(vertical effective stress) may be calculated from Eq 2. the initial reading on the pore-water pressure measurement
P 5 ~ σ 'vc 2 σ 'hc ! *A c 2 K1 ~ σ 'hc 1 u b ! *a (2)
device immediately prior to when the piston contacts the
specimen cap and the reading on the deformation indicator
K 5 W 2 @ ~ A c 2 a ! *h c *γ # (3) when the piston contacts the specimen cap.
where: 9.5.1.3 Check for pore pressure stabilization. Record the
σvc’ = vertical effective stress, defined at the center of the pore pressure. Close the drainage valves to the specimen and
specimen, measure the pore pressure change until stable. If the change is
σhc’ = lateral effective stress, defined at the center of the less than 5 % of the effective stress, the pore pressure is
specimen, assumed to be stabilized. Reopen the drainage lines.
Ac = area of specimen, after isotropic consolidation com- 9.5.2 Axial Loading—Open the drainage valves before ap-
puted using Eq 7, plying axial load to dissipate excess pore pressures throughout
a = area of loading piston, at point it enters cell, the specimen at failure. Determine loading rate which will
ub = back pressure, allow pore pressure to dissipate. Assuming failure will occur at
= unit weight of cell fluid, about 4 % axial strain, a suitable rate of strain may be
hc = distance from top of loading cap to mid-height of determined using Eq 4 or Eq 5:
specimen, after isotropic consolidation, and with side drain:
W = weight of piston + top cap + top half of specimen.
4%
9.4.3 After anisotropic consolidation the sample height and ε̇ 5 (4)
16t 90
area should be recomputed before shear accounting for the
additional change in height and volume. Eq 6 and 7 should be without side drain:
used for computations. 4%
ε̇ 5 (5)
10t 90
9.5 Shear—During shear, the chamber pressure shall be kept
constant while advancing the axial load piston downward where:
against the specimen cap using controlled axial deformation as t90 = time value obtained in 9.3.4.
the loading criterion. Specimen drainage is permitted during
If, however, it is estimated that failure will occur at a strain
shear, and volume changes will be read from the burette.
value other than about 4 %, a suitable strain rate may be
Failure is reached slowly so that excess pore pressure is
determined using Eq 5 by replacing 4 % with the estimated
dissipated under drained conditions.
failure strain. This rate of strain will provide for the sample to
9.5.1 Prior to Axial Loading—Before initiating shear, per-
build up minimal pore pressure during shear.
form the following:
9.5.2.1 At a minimum, record load, deformation, and vol-
9.5.1.1 Place the chamber in position in the axial loading
ume change values at increments of 0.1 % strain up to 1 %
device. Be careful to align the axial loading device, the axial
strain and, thereafter, at every 1 %. Take sufficient readings to
load measuring device, and the triaxial chamber to prevent the
define the stress-strain curve; hence, more frequent readings
application of a lateral force to the piston during shear.
may be required in the early stages of the test and as failure is
9.5.1.2 Bring the axial load piston into contact with the
approached. Continue the loading to 15 % strain, except
specimen cap to permit proper seating and realignment of the
loading may be stopped when the principal stress difference
piston with the cap. During this procedure, care should be
(deviator stress) has dropped 20 % or when 5 % additional
taken not to apply an axial load to the specimen exceeding
axial strain occurs after a peak in principal stress difference
0.5 % of the estimated axial load at failure. If the axial
(deviator stress).
load-measuring device is located outside of the triaxial
chamber, the chamber pressure will produce an upward force
10. Removing the Specimen
on the piston that will react against the axial loading device. In
this case, start shear with the piston slightly above the 10.1 When shear is completed, perform the following:
specimen cap, and before the piston comes into contact with 10.1.1 Close the specimen drainage valves.
the specimen cap, either (1) measure and record the initial 10.1.2 Remove the axial load and reduce the chamber and
piston friction and upward thrust of the piston produced by the back pressures to zero.
chamber pressure and later correct the measured axial load, or 10.1.3 With the specimen drainage valves remaining closed,
(2) adjust the axial load-measuring device to compensate for quickly remove the specimen from the apparatus so the
the friction and thrust. The value of the axial-load measuring specimen will not have time to absorb water from the porous
device reading should not exceed 0.1 % of the estimated failure disks.
load when the piston is moving downward prior to contacting 10.1.4 Remove the rubber membrane (and the filter-paper
the specimen cap. If the axial load-measuring device is located strips or cage from the specimen if they were used), and
inside the chamber, it will not be necessary to correct or determine the water content of the total specimen in accor-
compensate for the uplift force acting on the axial loading dance with the procedure in Test Method D2216. Free water

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
9
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
remaining on the specimen after removal of the membrane where:
should be blotted away before obtaining the water content. In ∆Hsat = change in height of the specimen during saturation.
cases where there is insufficient material from trimmings for
11.2.2.2 Method B:
index property tests, that is, where specimens have the same
diameter as the sampling tube, determine and record the V wf1V s
Ac 5 (8)
specimen mass prior to removing material for index property Hc
tests and obtain a representative portion of the specimen used where:
to determine its final water content. Prior to placing the
Vwf = final volume of water (based on final water content),
specimen (or portion thereof) in the oven to dry, sketch or
and
photograph the specimen showing the mode of failure (shear
Vs = volume of solids as follows: ws/(Gsρw)
plane, bulging, and so forth).
where:
11. Calculation
ws = specimen dry mass,
11.1 Initial Specimen Properties—Using the dry mass of the Gs = specific gravity of solids, and
total specimen, calculate and record the initial water content, ρw = density of water.
volume of solids, initial void ratio, initial percent saturation, 11.2.3 Using the calculated dimensions of the specimen
and initial dry unit weight. Calculate the specimen volume after consolidation and either an assumed or measured specific
from values measured in 7.2 or 7.3. Calculate the volume of gravity of solids, calculate the consolidated void ratio and
solids by dividing the dry mass of the specimen by the specific percent saturation.
gravity of the solids (Note 19) and dividing by the density of
water. Calculate the void ratio by dividing the volume of voids NOTE 20—The specimen will absorb water from the porous disks and
drainage lines during the time it is being removed from the apparatus.
by the volume of solids where the volume of voids is assumed When this effect is significant, Method A will yield more reasonable
to be the difference between the specimen volume and the values.
volume of the solids. Calculate dry density by dividing the dry NOTE 21—In this test method, the equations are written such that
mass of the specimen by the specimen volume. compression and consolidation are considered positive.

NOTE 19—The specific gravity of solids can be determined in accor- 11.3 Shear Data:
dance with Test Method D854 or it may be assumed based on previous test 11.3.1 Calculate the axial strain, ε1, for a given applied axial
results. load as follows:
11.2 Specimen Properties After Consolidation—Calculate ∆H
the specimen height and area after consolidation as follows: ε1 5 (9)
Hc
11.2.1 Height of specimen after consolidation, Hc, is deter-
mined from the following equation: where:
∆H = change in height of specimen during loading as deter-
H c 5 H o 2 ∆H o (6)
mined from deformation indicator readings.
where: 11.3.2 Calculate the cross-sectional area, A, for a given
Ho = initial height of specimen, applied axial load as follows:
∆Hc = change in height of specimen at the end of
V c 2 ∆V ε
consolidation, and A5 (10)
H c 2 ∆H ε
Hc = height of specimen after consolidation.
11.2.2 The cross-sectional area of the specimen after where:
consolidation, Ac, shall be computed using one of the following Vc = specimen volume after consolidation, V0 – ∆Vc,
methods. The choice of the method to be used depends on ∆Vε = change in volume from beginning of shear to any
whether shear data are to be computed as the test is performed strain, and
(in which case Method A would be used) or on which of the ∆Hε = change in height from beginning of shear to any
two methods, in the opinion of a qualified person, yield strain.
specimen conditions considered to be most representative of NOTE 22—The cross-sectional area computed in this manner is based on
the assumption that the specimen deforms as a right circular cylinder
those after consolidation. Alternatively, the average of the two during shear. In cases where there is localized bulging, it may be possible
calculated areas may be appropriate. to determine more accurate values for the area based on specimen
11.2.2.1 Method A: dimension measurements obtained after shear.
NOTE 23—For the following equations σv is replaced by σ1 and σh is
V o 2 ∆V sat 2 ∆V c replaced by σ3. This can be done only if σh is less than σv.
Ac 5 (7)
Hc
11.3.3 Calculate the principal stress difference (deviator
where: stress), σ1 – σ3, for a given applied axial load as follows:
Vo = initial volume of specimen, P1K1σ 3 ~ A 2 a !
∆Vc = change in volume of specimen during consolidation ~ σ 1 2 σ 3! 5 A
2 σ3 (11)
as indicated by burette readings, and
∆Vsat = change in volume of specimen during saturation as where:
follows: 3Vo [∆Hsat/Ho]. A = corresponding cross-sectional area.

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
10
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
11.3.3.1 Correction for Filter-Paper Strips—For vertical F = force applied to stretch the membrane,
filter-paper strips that extend over the total length of the L = unstretched length of the membrane,
specimen, apply a filter-paper strip correction to the computed ∆L = change in length of the membrane due to the force, F,
values of the principal stress difference (deviator stress), if the and
error in principal stress difference (deviator stress) due to the Am = area of the membrane = 2tmWs
strength of the filter-paper strips exceeds 5 %.
where:
(1) For values of axial strain above 2 %, use the following
equation to compute the correction: tm = thickness of the membrane, and
Ws = width of circumferential strip, 15 mm (0.5 in.).
K fpP fp
∆ ~ σ 1 2 σ 3! 5 (12) NOTE 25—A typical value of Em for latex membranes is 1400 kPa (200
Ac lbf/in.2).
where: NOTE 26—The corrections for filter-paper strips and membranes are
based on simplified assumptions concerning their behavior during shear.
∆(σ1 – σ3) = correction to be subtracted from the measured Their actual behavior is complex, and there is not a consensus on more
principal stress difference (deviator stress), exact corrections.
Kfp = load carried by filter-paper strips per unit length
of perimeter covered by filter-paper, 11.4 Principal Stress Difference (Deviator Stress) and
Pfp = perimeter covered by filter-paper, and Change in Volume (∆V) Versus Strain Curves—Prepare graphs
Ac = cross-sectional area of specimen after showing relationships between principal stress difference (de-
consolidation. viator stress) and change in volume (∆V) with axial strain,
For values of axial strain of 2 % or less, use the following plotting deviator stress and ∆V as ordinates and axial strain as
equation to compute the correction: abscissa. Select the principal stress difference (deviator stress)
and axial strain at failure in accordance with 3.2.4.
50ε 1 K fpP fp
∆ ~ σ 1 2 σ 3! 5 (13) 11.5 Determine and record the major and minor effective
Ac
stresses at failure based on stresses, σ1f’ and σ3f’ respectively as
where: follows:
ε1 = axial strain (decimal form) and other terms are the same '
σ 3f 5 σ 3 2 u f □effective lateral confining stress at failure (16)
as those defined in subparagraph (1) of 11.3.3.1.
' '
NOTE 24—For filter-paper generally used in triaxial testing, Kfp is σ 1f 5 ~ σ 1 2 σ 3 ! at failure 2 ∆ ~ σ 1 2 σ 3 ! fpf 2 ∆ ~ σ 1 2 σ 3 ! mf 1σ 3f
approximately 0.19 kN/m (1.1 lbf/in.). (17)
11.3.3.2 Correction for Rubber Membrane—Use the follow- where:
ing equation to correct the principal stress difference (deviator
uf = pore pressure at failure,
stress) for the effect of the rubber membrane if the error in
∆(σ1–σ3)fpf = paper correction at failure, and
principal stress difference (deviator stress) due to the strength ∆(σ1–σ3)mf = rubber membrane correction at failure.
of the membrane exceeds 5 %:
4E m t m ε 1 11.6 Mohr Stress Circles—If desired, construct Mohr stress
∆ ~ σ 1 2 σ 3! 5 (14) circles at failure based on axial and radial stresses on an
Dc
arithmetic plot with shear stress as ordinate and normal stress
where: as abscissa using the same scales. The circle based on total
∆(σ1 – σ3) = correction to be subtracted from the measured stresses is drawn with a radius of one half the principal stress
principal stress difference (deviator stress), difference (deviator stress) at failure with its center at a value
Dc = =4A c /π diameter of specimen after equal to one half the sum of the major and minor principal
consolidation, stresses.
Em = Young’s modulus for the membrane material,
tm = thickness of the membrane, and 12. Report: Test Data Sheet(s)/Form(s)
ε1 = axial strain (decimal form).
12.1 The methodology used to specify how data are re-
(1) The Young’s modulus of the membrane material may be corded on the test data sheet(s)/form(s), as given below, is
determined by hanging a 15-mm (0.5-in.) circumferential strip covered in 1.6 and Practice D6026.
of membrane using a thin rod, placing another rod through the
bottom of the hanging membrane, and measuring the force per 12.2 Record as a minimum the following general informa-
unit strain obtained by stretching the membrane. The modulus tion (data):
value may be computed using the following equation: 12.2.1 Identification data, such as project number, boring
number, sample/specimen number and depth. Include other

Em 5
S D F
Am
(15)
pertinent data that helps identify the sample/specimen.

S D ∆L
L
12.2.2 Visual description of the specimen, including soil
classification and whether the specimen is intact, compacted, or
otherwise prepared.
where:
12.2.3 Values of plastic limit and liquid limit, if determined.
Em = Young’s modulus of the membrane material, These values can be determine by Test Method D4318.

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
11
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.
D7181 − 20
12.2.4 Value of specific gravity of solids including a nota- 12.3.11 Axial strain at failure, percent,
tion if the value was assumed or determined specifically for the 12.3.12 Rate of strain, percent per minute,
soil. This can be determined by Test Methods D854 or 12.3.13 Principal stress difference (deviator stress) and
assumed. change in volume versus axial strain curves as described in
12.2.5 Particle-size analysis, if determined. This can be 11.4,
determined by Test Methods D6913, D7928 as applicable. 12.3.14 Mohr stress circles based on axial and radial
12.2.6 Name of person(s) who prepared and tested the stresses, (optional),
samples/specimens, including the date(s) performed. 12.3.15 Slope of angle of the failure surface (optional),
12.3 Record as a minimum the following test specimen 12.3.16 Failure sketch or photograph of the specimen, and
data: 12.3.17 Remarks and notations regarding any unusual con-
ditions such as slickensides, stratification, shells, pebbles,
12.3.1 Initial specimen dry unit weight, void ratio, water
roots, and so forth, or other information necessary to properly
content, and percent saturation, (specify if the water content
interpret the results obtained, including any departures from
specimen was obtained from cuttings or the entire specimen),
the procedure outlined.
NOTE 27—The specific gravity determined in accordance with Test
Methods D854 is needed for calculation of the saturation. An assumed 13. Precision and Bias
specific gravity may be used provided it is noted in the test report that an 13.1 Precision—Test data on precision is not presented due
assumed value was used.
to the nature of the soil materials tested by this test method. It
12.3.2 Initial height and diameter of specimen, is either not feasible or too costly at this time to have ten or
12.3.3 Method followed for specimen saturation dry or wet more laboratories participate in a round-robin testing program.
method, Also, it is either not feasible or too costly to produce multiple
12.3.4 Total back pressure, specimens that have uniform physical properties. Any variation
12.3.5 Effective consolidation stress CID or effective con- observed in the data is just as likely to be due to specimen
solidation stresses, variation as to operator or laboratory testing variation.
12.3.6 Time to 100 % primary consolidation, 13.1.1 Subcommittee D18.05 is seeking any data from users
12.3.7 Specimen dry unit weight, void ratio, water content, of this test method that might be used to make a limited
and percent saturation after consolidation, statement on precision.
12.3.8 Specimen cross-sectional area after consolidation 13.2 Bias—There is no accepted reference value for this test
and method used for determination, method, therefore, bias cannot be determined.
12.3.9 Failure criterion used,
12.3.10 The value of the principal stress difference (deviator 14. Keywords
stress) at failure and the values of the minor and major 14.1 back pressure saturation; consolidated drained
principal stresses at failure, (indicate when values have been strength; effective stresses; non-cohesive soil; strain-controlled
corrected for effects due to membrane or filter strips, or both), loading; stress-strain relationships

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

Copyright by ASTM Int'l (all rights reserved); Tue Apr 14 10:47:42 EDT 2020
12
Downloaded/printed by
DALHOUSIE UNIVERSITY (DALHOUSIE UNIVERSITY) pursuant to License Agreement. No further reproductions authorized.

Potrebbero piacerti anche