Sitrans F M

Mechanical Installation

Restricted © Siemens AG 2013 All rights reserved. siemens.com/flow


SANEPAR –Digitrol
Training

Agenda

 Finding a suitable location

 Electrical Installation

 Most common errors

 Menu structure

Restricted © Siemens AG 2013 All rights reserved.


Page 2
SANEPAR –Digitrol
Training

Agenda

 Finding a suitable location

 Electrical Installation

 Most common errors

 Menu structure

Restricted © Siemens AG 2013 All rights reserved.


Page 3
Selecting a suitable location

Restricted © Siemens AG 2013 All rights reserved.


Selecting a suitable location

Choose a location that provides at least 5 x diameters before and 3 x diameter after
the sensor for the best accuracy.

Restricted © Siemens AG 2013 All rights reserved.


Installation

Horizontal pipes

Risk of air

Risk of particles

Partly filled pipe


BAD LOCATION!
Restricted © Siemens AG 2013 All rights reserved.
Installation

Vertical pipes

Flow in upward direction

Restricted © Siemens AG 2013 All rights reserved.


Grounding MAG 3100 and MAG 5100 W

Sensors with build in earthing electrodes


Grounding via grounding electrodes

Sensors with no earthing electrodes( MAG 3100 PTFE/PFA)


Conductive pipe
Use grounding strap

Sensors with no earthing electrodes( MAG 3100 PTFE/PFA)


Non Conductive pipe Use grounding ring

Restricted © Siemens AG 2013 All rights reserved.


Installing MAG 3100 MAG 5100 W

Use proper gaskets Use torque tables for correct tightening

Restricted © Siemens AG 2013 All rights reserved.


Installing MAG1100, MAG 3100 and MAG 5100 W

Update to IP68 Direct burial

Pea Gravel

IP68 Potting kit

Restricted © Siemens AG 2013 All rights reserved.


Grounding the MAG 1100

Restricted © Siemens AG 2013 All rights reserved.


Installing MAG 1100

Use proper gaskets Use torque tables for correct tightening

Restricted © Siemens AG 2013 All rights reserved.


Grounding MAG 1100 FOOD

Grounding via adapters

Restricted © Siemens AG 2013 All rights reserved.


Mounting MAG 1100 Food

Do not turn the sensor after clamps are fasten

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- Immediately after a slide valve

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s

Acceptable measuring with


DN150

the valve 100 % open DN150

SLIDE VALVE
Vanne type Opercule
Add. error <1 % on the
condition that the valve is
100 % open Valve open [%] Ref.
Flow 100 % 50 % 25 % 10 % error
3
m /h Error % %

250 -0,5 6,1 -18 -0,04

188 -0,5 6,3 -0,11

125 -0,5 6,5 -0,14

63 -0,5 6,6 -0,16

38 -0,8 6,9 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- 4 x diameter after a slide valve

MAG 6000/3100, DN 150


Qmax = 700 M3/h, 10 m/s

Add. error < 1 % irrespective


of the valve position 0.6m DN150

DN150
Valve type Opercule

Vanne type Opercule

V alve open [% ] R ef.


F low 100 % 50 % 25 % 10 % error
m 3 /h E rror % %

250 -0,3 0,6 -0,1 -0,04

188 -0,1 0,7 -0,6 -0,11

125 -0,5 0,7 -0,6 0,6 -0,14

63 -0,4 0,7 -0,5 0,6 -0,16

38 -0,3 0,9 -0,04 0,8 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Unathorized Iinstallations
- Immediately before a slide valve

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s

Add. error < 0.5 % irrespective


DN150
the valve position.
Valve type Opercule

A valve or any other distortions


should always be mounted after
the flow meter.
V alve open [% ] R ef.
F low 100 % 50 % 25 % 10 % error
m 3 /h E rror % %

250 -0,3 -0,3 -0,3 -0,04

188 -0,3 -0,3 -0,3 -0,11

125 -0,3 -0,4 -0,3 -0,4 -0,14

63 -0,4 -0,4 -0,5 -0,4 -0,16

38 -0,3 -0,4 -0,4 -0,5 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- Immediately after a butterfly valve

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s

Add. error ~ 5% with 100%


DN15
open valve 0

DN150
BUTTERFLY VALVE

Not acceptable Vanne type Opercule

V alve open [% ] R ef.


F low 100 % 50 % 25 % 10 % error
m 3 /h E rror % %

250 3,2 20 -0,04

188 3,4 20 -0,11

125 3,4 20 -0,14

63 3,9 19 -0,16

38 3,9 18 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- 4 x diameter behind a butterfly valve

MAG 6000/3100, DN 150


Qmax = 700 M3/h, 10 m/s

Add. error ~ 3% irrespective


DN150
the valve position. 0.6m

DN150
BUTTERFLY VALVE

Butterfly valves should be


placed more than 5 x dia. in
front of an electromagnetic
V alve open [% ] R ef.
flow meter. F low 100 % 50 % 25 % 10 % error
m 3 /h E rror % %

250 -2,3 1,2 -0,04

188 -2,0 1,1 -0,11

125 -2,4 1,1 0,6 -0,14

63 -2,7 1,1 0,3 -0,16

38 -2,7 1,0 0,4 0,4 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- Installed after a bending

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s
DN150
Pos. A Pos. B

Reference measuring
• Error -0.04 to -0.18%

Pos A
• Error 2.3 to 2.7%

Pos B, 90° turned %


Pos.A Pos. B Sans perturbation
5
• -0.3 to 0.2% 4

0
0 50 100 150 200 250
-1 m^3/h

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- Installed after a T-point

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s
P os . A P os . B

Reference measurings DN150

• Error -0.04 to -0.18 %

25 cm
Pos A
• Error 1.9 to 2.3 %

Pos B, turned 90 ° %
5 Pos. A Pos. B Sans perturbation
• Error 0.9 to 1.3 % 4

0
0 50 100 150 200 250
-1
m^3/

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- DN 150 installed in DN 400 pipe

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s

Higher flow => higher inaccuracy DN 400 DN 400


DN150

The inaccuracy however,


is within +0.5 %, with a
diameter ratio of 2 : 1
%
1
0,8
Sans perturbation Avec retrecissements
0,6
0,4
0,2
0
-0,2 0 50 100 150 200 250 300
-0,4
-0,6
-0,8
-1

Restricted © Siemens AG 2013 All rights reserved.


Unathorized installations
- Installed 5 x dia. after a pump

MAG 6000/3100, DN 150


Qmax = 700 m3/h, 10 m/s Pompe centrifuge

Add. error ~ 0.5% due to the influence


from the pump. DN 400

5 x dia is not always enough after a 0.6m DN150 DN 150

pumping installation.
0.25m

M ea. R ef.
F low error error
m 3 /h % %

250 0,5 -0,04

188 0,3 -0,11

125 0,5 -0,14

63 -0,6 -0,16

38 -0,3 -0,18

Restricted © Siemens AG 2013 All rights reserved.


Agenda

 Finding a suitable location

 Electrical Installation

 Most common errors

 Menu structure

Restricted © Siemens AG 2013 All rights reserved.


Page 24
Electrical installation

Compact installation.

Remove plugs from Connect plugs and earth Connect power and output
connection blocs wire to connection board to connection board

Restricted © Siemens AG 2013 All rights reserved.


Electrical installation

Remote installation – At the sensor

Remove Sensorprom to be Connect coil and electrode Tightening cable glands to


mounted at Transmitter cables. fit to cables

Restricted © Siemens AG 2013 All rights reserved.


Electrical installation

Remote installation – At the transmitter


Locate the Sensorprom

MAG5000/6000 MAG 6000 I (Exd) 19” version


wall mounting

Restricted © Siemens AG 2013 All rights reserved.


Electrical installation

Remote installation – cable connections

Special Electrode cable


(81/82/0/83/84)
Long distance or
empty pipe detection

Electrode cable (82/0/83)

Coil cable (85/86)

Restricted © Siemens AG 2013 All rights reserved.


Agenda

 Finding a suitable location

 Electrical Installation

 Most common errors

 Menu structure

Restricted © Siemens AG 2013 All rights reserved.


Page 29
MAG remote installation

Forgot to relocate the Sensorprom

Restricted © Siemens AG 2013 All rights reserved.


Compact installation

Forgot to reconnect plugs to connection board

Restricted © Siemens AG 2013 All rights reserved.


MAG 5100 W (MAG 3100)

Overtigtning and/or no gasket used

Restricted © Siemens AG 2013 All rights reserved.


MAG 1100( MAG 3100)

Wrong selection of liner and/or electrode materail

Restricted © Siemens AG 2013 All rights reserved.


Common errors

 Air in the pipe line ( Too high flow, sometimes erratic unstable flow))

 Wrong cable connections ( between sensor/transmitter or on output)

 Use cable of bad quality ( local cable – ”cheaper solution”)

 Mixed up sensorproms ( wrong flow rate )

 Bad grounding (PTFE lined sensors with no grounding electrodes)


( erratic unstable reading with and without flow)

 Wrong sensor selection due to amount of solids.


( stable with no flow, erratic unstable flow with flow
due to electrode noise)

Restricted © Siemens AG 2013 All rights reserved.


Agenda

 Finding a suitable location

 Electrical Installation

 Most common errors

 Menu structure

Restricted © Siemens AG 2013 All rights reserved.


Page 35
Display

PRIMARY FIELD FOR UNIT FIELD


NUMETIC VALUE
ALARM FIELD TWO
FLASHING TRIANGLE

MODE FIELD

TITLE LINE LOCK FIELD

SUBTITLE FIELD FORWARD KEY

BACKWARD KEY
TOP UP KEY

LOCK / UNLOCK KEY


CHANGE KEY

SELECT KEY

Restricted © Siemens AG 2013 All rights reserved.


Menu - Overview

Restricted © Siemens AG 2013 All rights reserved.


Password – Sensor Characteristics

Sensorprom must be installed. Otherwise


all data must be programmed manually

If Sensorprom is NOT installed


Error code 40 can be suppressed

Sensor size – if no sensorprom sensor size


must be programmed in metre. Eg, DN 100 = 0.100

Calibration Factor from wet calibration – can be found


on sensor label and calibration report.

Factory password: 1000 Correction Factor - must be 1.0000


To be changed in special conditions only

Excitation frequency according to Sensor size.


See Handbook for correct selection.

Restricted © Siemens AG 2013 All rights reserved.