Some possible explanations why the error has occurred:

- No connection between transducer board and FM I/O board


- No connection between FM I/O board and protective driver board
- Transducer board mounted in incorrect position on FM I/O board
- Incorrect transducer ID. Please update transducer ID (I2C scan).
- Transducer board is broken
- Serial bus error on protective CPU board
- Uncalibrated transducer. Please calibrate transducer
- Broken FM I/O board

P PTRA 000 104 ERROR READING PARAMETERS FROM VENOUS PRESSURE


TRANSDUCER
Unconditional error, blood pump not available

General description of conditions for occurring:


- An error occurred while reading transducer parameters during init
(start-up).

Some possible explanations why the error has occurred:


- No connection between transducer board and BM I/O board
- No connection between BM I/O board and protective driver board
- Transducer board mounted in incorrect position on BM I/O board
- Incorrect transducer ID. Please update transducer ID (I2C scan).
- Transducer board is broken
- Serial bus error on protective CPU board
- Uncalibrated transducer. Please calibrate transducer
- Broken BM I/O board

P PTRA 000 105 ERROR READING PARAMETERS FROM PH TRANSDUCER


Unconditional error, blood pump not available

General description of conditions for occurring:


- An error occurred while reading transducer parameters during init
(start-up).

Some possible explanations why the error has occurred:


- No connection between MM I/O board and protective CPU board
- Incorrect transducer ID. Please update transducer ID (I2C scan).
- MM I/O board is broken
- Serial bus error on protective CPU board
- Uncalibrated transducer. Please calibrate transducer

P PTRA 000 106 ERROR READING PARAMETERS FROM ARTERIAL PRESSURE


TRANSDUCER
Unconditional error, blood pump not available

General description of conditions for occurring:


- An error occurred while reading transducer parameters during init
(start-up).

Some possible explanations why the error has occurred:


- No connection between transducer board and BM I/O board
- No connection between BM I/O board and protective driver board
- Transducer board mounted in incorrect position on BM I/O board
- Incorrect transducer ID. Please update transducer ID (I2C scan).
- Transducer board is broken
- Serial bus error on protective CPU board
- Uncalibrated transducer. Please calibrate transducer
- Broken BM I/O board

HCEN12371 Rev. 10.2011 AK 96® - Error Codes 107 (115)


Program version 3.50
P PTRA 000 107 ERROR READING PARAMETERS FROM DIASCAN
TRANSDUCER
Unconditional error, blood pump not available

General description of conditions for occurring:


- An error occurred while reading transducer parameters during init
(start-up).

Some possible explanations why the error has occurred:


- No connection between transducer board and FM I/O board
- No connection between FM I/O board and protective driver board
- Transducer board mounted in incorrect position on FM I/O board
- Incorrect transducer ID. Please update transducer ID (I2C scan).
- Transducer board is broken
- Serial bus error on protective CPU board
- Uncalibrated transducer. Please calibrate transducer
- Broken FM I/O board

P SDT, CPU P system


P SDT 003 001 CHECK SUM ERROR
Unconditional error, blood pump not available

General description of conditions for occurring:


- The check sum of the protective system presets is not correct.
Enter service/ preset and preset the machine (make a default
preset).

Some possible explanations why the error has occurred:


- Broken/ new CPU P safe-RAM

P SDT 004 001 CPU C DOES NOT SEND SHUT DOWN ACKNOWLEDGE

Some possible explanations why the above error has occurred:


- Internal software error

P SDT 004 002 CPU W DOES NOT SEND SHUT DOWN ACKNOWLEDGE

Some possible explanations why the above error has occurred:


- Internal software error

P SDT 004 003 SHUT DOWN FAILURE

General description of conditions for occurring:


- The AK 96 was not switched off within 5 seconds after shut down
request.

Some possible explanations why the above error has occurred:


- Protective driver board broken (IC16 and components)

P SDT 004 004 CPU C DOES NOT SEND RESET ACKNOWLEDGE

Some possible explanations why the above error has occurred:


- Internal software error

P SDT 004 005 CPU W DOES NOT SEND RESET ACKNOWLEDGE

Some possible explanations why the above error has occurred:


- Internal software error

108 (115) AK 96® - Error Codes HCEN12371 Rev. 10.2011