Sei sulla pagina 1di 1

INSTRUMENT CALIBRATION SHEET

Tag No Calibration Date 26-Jun-20


Facilities Instrumentation Labarotary, UTHM Pagoh Transmitter Type Pressure Transmitter
Range 4-20 mA output Transmitter Serial No. 91W225450
Allowable Span error ±0.2 % Transmitter Model No. EJA11OE

RISING FALLING
Standard Pressure/mbar Multimeter/mA Pressure/mbar Multimeter/mA
Bacaan Bacaan Bacaan
% Input mbar input Bacaan 1 Bacaan 2 Bacaan 3 Bacaan 1 Bacaan 2 Bacaan 3 Bacaan 1 Bacaan 2 Bacaan 3 Bacaan purata Bacaan 1 Bacaan 2 Bacaan 3
purata purata purata
0 0 0.00 0.00 0.00 0.00 4.00 4.00 4.00 4.00 0.00 1.00 0.00 0.33 4.01 4.00 4.00 4.00
25 250 255.00 251.00 248.00 251.33 6.08 6.06 6.01 6.05 256.00 255.00 254.00 255.00 6.06 6.05 6.04 6.05
50 500 497.00 500.00 501.00 499.33 7.97 8.01 8.02 8.00 495.00 501.00 507.00 501.00 7.95 7.96 8.04 7.98
75 750 747.00 748.00 751.00 748.67 9.99 10.00 10.01 10.00 757.00 756.00 745.00 752.67 10.08 10.06 10.03 10.06
100 1000 994.00 1004.00 1010.00 1002.67 11.99 12.03 12.10 12.04 996.00 1001.00 1005.00 1000.67 11.98 12.01 12.04 12.01

Remarks: All measurements were taken approximately because of the higher sensitivity of the equipment.

Test Equipment used Pressure Transmitter


Serial No. 91W225450
Calibration Expiry Date

Calibrated By Witnessed By Accepted By


Name Ts. MOHD FAIZAL BIN ESA Dr. ROHANI BINTI RAHMAD MUHAMMAD IRFAN BIN ROSLI
Sign
Date 26-Jun-20

Potrebbero piacerti anche