Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
H
tW
u
lo
a
c
d
M
s
n
m
ie
S
Common CT Systems
CT
Quality Assurance
System
10017857
08874427
08872017
08377520
07734713
07732444
07732436
07732428
07732410
07543106
07543015
07541985
07407039
07407013
07406502
07393114
07114023
07005502
04806118
03815490
03813933
03812950
03810160
03810152
03810137
03810079
03804890
© Siemens AG 2005
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Area of application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Documents required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Tools, text and auxiliary means . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Location of identification labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Informing the expert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
When is an acceptance test required? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
When is a partial acceptance test required? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Determining reference values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Protection against electrical voltages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Radiation Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Area of application 0
Documents required 0
You need the following tools, test and auxiliary means for the acceptance test (the mate-
rial numbers are located in SPC) (Tab. 1 / p. 4):
Tab. 1 Tools, text and auxiliary means
Name
Weight: approx. 135 kg for weighing down the table
- Tape measure, 100 cm
Angle support for set of phantoms
Phantom support1
20 cm water phantom
30 cm water phantom
Wire phantom
Wire phantom_100
Slice thickness phantom
Support for low-contrast phantom
Low-contrast phantom
CCR combi phantom
Dosimeter Dali CT, WK92 or DIADOS
CT measurement chamber with cable and holder
Inclinometer
X-ray film AGFA STRUCTURIX D7DW 13 x 18 cm
1. The phantom support is part of the 'Sliding Gantry' option. It is not required
for standard configurations.
CAUTION [ hz_serdoc_F13G10U01M01 ]
Poor image quality through wrong adjustment values may be cre-
ated during tune up.
Risk of misalignment!
¹ Only phantoms delivered with the CT system may be
used for system adjustments.
¹ The water phantom has to have the same temperature as
the examination room.
We recommend that you inform the expert in a timely manner about the date planned for
the acceptance test.
Advantages:
• Timely coordination with the acceptance testing of constructional radiation protection,
possible repeat tests are eliminated.
• The customer can work with the CT system earlier than planned.
Test positions 1
High-contrast resolution
Low-contrast resolution
Light localizer position
Dose measurements
X-ray tube voltage
Slice thickness
Event
Homogeneity
Dose profile
Water value
Pixel noise
Initial installation X X X X X X X X X X X X
Change in operator X X X X X X X X X X X X
1. X = test is mandatory
Test positions 1
High-contrast resolution
Low-contrast resolution
Light localizer position
Dose measurements
Repair
X-ray tube voltage
Replacement
Slice thickness
Homogeneity
Dose profile
Water value
New adjustment
Pixel noise
Test positions 1
High-contrast resolution
Low-contrast resolution
Light localizer position
Dose measurements
Repair
Slice thickness
Homogeneity
Dose profile
Water value
New adjustment
Pixel noise
X-ray Tube Unit X X X
DMS module X X X
DMS X X X X X X
Changing the operating
X X X X X X X X X
parameters/characteristics
1. X = test is mandatory
When putting a new CT system into operation or a component of the CT system, accesso-
ries or the test equipment that could change the test results, an initial constancy test has
to be performed directly after the acceptance test or status test that showed satisfactory
characteristics of the standard functions. The purpose of the initial constancy test is to
establish a new reference value (or values) for the parameter(s) tested.
Safety notes 0
WARNING [ hz_serdoc_F13G01U12M01 ]
Failure to observe the guidelines specified in this document may
lead to serious injury or loss of life.
Risk of accident and/or injury!
¹ The guidelines contained in this document have to be
strictly followed.
Carefully read through the following instructions as well as the newest version of the
• General Safety Notes (TD00-000.860.01 / General Safety Notes)
• Read through Product- specific safety notes (CT02-023.860.01 / Product-specific
safety notes)
carefully prior to working with the CT system.
WARNING [ hz_serdoc_F13G01U13M02 ]
High voltages and/or mechanical movements (e.g. gantry rotation)
may lead to accidents and injuries when opening system covers.
Risk of accident and injury!
¹ Only authorized service personnel is allowed to open
system covers.
Radiation Safety 0
CAUTION [ hz_serdoc_F13G09U01M01 ]
Not following X-rays protective regulations may lead to radiation
exposure to you and/or other persons.
Risk of radiation exposure!
Observe radiation protection regulations when X-rays are
switched on, e.g. :
¹ Never work inside the gantry room.
¹ Do not leave the system unattended.
¹ Make sure there is no one inside the gantry room.
¹ Make sure that no person can enter the gantry room
unnoticed (e.g. lock doors if necessary).
• The test phantoms are attached to the angle bracket at the head end of the CT patient
table.
NOTE In case the 'Sliding Gantry' option is installed, the test phantoms
are attached to phantom holder 478022K1130.
The phantom holder is located on the tabletop of the patient table.
As counterbalance weight, insert the CCR body phantom opposite
the holder.
You can adjust the height of the phantom holder.
• The water, slice thickness and wire phantom can be combined into a phantom set.
In addition, the low-contrast phantom can be mounted to the wire phantom via an
adapter ring.
• For all measurements, the phantom set has to be centered in the measurement field
and the gantry has to be fully upright.
• If the test phantoms do not include a serial number, identify them retroactively.
The identification allows for identifying the test means for subsequent constancy tests.
P • Enter the results of the acceptance test into the respective test protocol:
• The documentation of the acceptance test also includes pictures of the test bodies, the
test film of the measurements of the dose profile as well as the acceptance report on
film.
To generate the film documentation, use the camera function in the QUALITY platform.
• The respective Siemens office has to be keep the entire acceptance test documenta-
tion on file for 10 years.
A copy of the filled out test protocol has to be given to the operator of the X-ray
equipment and filed in the Owner Manual.
Fig. 2: Enter the tester's name and the phantom serial numbers
Pos. 1 Holder
Pos. 2 Water phantom
Pos. 3 Slice phantom
Pos. 4 Wire phantom
Pos. 5 Low-contast-phantom or Allignment-phantom
CAUTION In addition you will need the 30 cm water phantom for the IEC ac-
ceptance test.
¹ Otherwise, the water value, homogeneity, and noise will
not be determined correctly.
1. Combine the water, slice thickness, wire and low-contrast phantom into a test phantom
set.
2. Set the gantry tilt to 0 degree (gantry is fully upright)
3. Place a weight of approx. 135 kg onto the tabletop of the patient table.
NOTE Item number 3 is omitted when the 'Sliding Gantry' option is in-
stalled.
NOTE The test results with the acceptable tolerance margins are shown
in the result window. After you completed the test run, the image
of a condensed test protocol is shown. The test protocol also in-
cludes the tolerances, as well as the markers (< / >) for falling be-
low or exceeding the tolerance margins.
The correlation of the axial PATIENT positioning light and scan plane is tested by posi-
tioning and scanning a thin absorber.
1 2
The test phantom set is centered in the measurement field. It cannot be tilted downward.
The slice thickness phantom is positioned in the measurement field so that the light band
of the inner light markers is exactly centered above the intersection points of the reference
markers on the phantom. A measurement is made and the position of the test phantom in
the slice plane is assessed.
The position of the phantom is correct when the long and short bar "1/Evaluation of the
phantom positon after the measurement" are positioned one above the other. If the long
bar can be seen to the left of the short bar "2/Evaluation of the phantom position after the
measurement", the CT table (refer to note) has to be shifted by 1 mm with FEED IN prior
to each new measurement. If the long bar is shown on the right of the short bar, then
move the table with FEED OUT by one millimeter before each new measurement.
The deviation of the light marker from the slice plane results from the sum of the displace-
ment steps of the table (refer to note) that were required to position the phantom correctly
in the slice plane.
NOTE Is the 'Sliding Gantry' option installed, the gantry is shifted accord-
ingly.
Position the quality phantom so that the light marker cuts across the front of the phantom.
The respective mode is loaded and the topogram is started.
A (green line) is displayed in the topogram that should be pulled to the center of the slice
thickness phantom.
1. The phantom has to be positioned exactly in the Z direction of the slice plane. The Z
position is correct if the long and the short bars lie above one another in the tomogram.
Refer to "1/Evaluation of the phantom position after the measurement" for testing the
position of the light marker.
2. - The two aluminum strips must appear lying horizontally in the tomogram image.
The program tomographs the aluminum strips for each selectable slice thickness and cal-
culates the current slice thickness from the tomograms.
Homogeneity 0
NOISE 0
NOTE This measurement is not required for the acceptance test in Ger-
many!
The contrast scale CS states the change of the effective attenuation coefficient of a
water-equivalent material that is required to change the mean CT number of this material
by 1.
Two density values are required for determining the contast scale. Only an air scan is per-
formed for this. As a second value, the water value is taken over from previous measure-
ments.
Table Pos. 0
Positional accuracy of the PATIENT SUPPORT includes both longitudinal positioning and
backlash evaluation.
The accuracy of longitudinal PATIENT SUPPORT positioning is evaluated by moving the
PATIENT SUPPORT a defined distance in one direction and confirming the distance trav-
elled.
The accuracy of moving the PATIENT SUPPORT in one direction and moving it back to
the starting position is known as backlash.
The actual distance moved is measured in each case at the table with a tape measure.
The exact measurement value (e.g. 300.5) has to be entered into the program platform.
The program takes this value and computes the difference between the actually driven
and electronically displayed travel path.
Dose Profile 0
NOTE The base width closely aproximates the width of the film blacken-
ing, if the film is not too over or underexposed.
The film is positioned in the system axis perpendicular to the central ray. As the film car-
rier, a CCR combination phantom is used. The upper edge of this phantom is moved into
the center of rotation of the measurement phantom at a tolerance margin of 5 mm. The
film is shifted in Z direction before each exposure with changed slice thickness.
After the film has been developed, the base width (width of filme blackening) each individ-
ually exposed strip is measured in the center and entered manually into the input field.
For that purpose, call the DOSE PRO/EVALUATE function in the platform.
Dose 0
NOTE For 80 kV, the measurement is performed in the head mode. The
body mode is used for 120/130/140 kV.
The program computes the dose in the system axis according to the following formula:
Test Procedure
CTDIW for the typical axial head and body conditions of operation.
Position CTDI Head (16 cm Plexi) / Body (32 cm Plexi) phantom in the center of the scan
field.
Position the phantom such that the inner light marker hits the engraved marking.
Click GO when ready.
Gantry tilt: 0
NOTE The tilt shown at the exposure device cannot deviate by more than
+/- 2° from the measured value.
Using the configuration under technical, the language for the protocols is selected in the
service menu.
During the acceptance test the protocols are automatically generated together with the
results of the measurement and stored as pdf files.
The path for the protocols is:
C:\Somaris\service\pdf\report\pdfreport
If the customer has a network printer with connection to the CT system, the protocols may
be printed out with this printer.
If not, the protocols can be printed with the help of the Explorer from C:\Somaris\ser-
vice\pdf\report\pdfrepor and copied to one or several floppies.
Take the floppies to a PC with a printer and print out the files.
hz_serdoc_F13G01U12M01 . . . . . . . . . 8
hz_serdoc_F13G01U13M02 . . . . . . . . . 8
hz_serdoc_F13G09U01M01 . . . . . . . . . 8
hz_serdoc_F13G10U01M01 . . . . . . . . . 5