Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
1/3
5. Results display: After the test is completed, the instrument determine whether exist seroconversion in first infection or a
automatically calculates the SARS-CoV-2 antibody concentration elevation in specific IgM and IgG antibody titre.
value(AU/mL) of the current sample according to the standard 3. A high titre specific IgG Ab will fight with IgM Ab for antigen
curve, and simultaneously displays the patient information. Then binding site,which will decrease sensitivity of the test and cause
you can print the test results directly. Click again to quickly return false negative result.
to the main screen. 4. Patients with impared immunity or under immunosuppressive
6. Quality control: Select the quality control test process to therapy are not recommended to refer this interpretation of their
determine the control and the results are automatically classified test results.
into the quality control system of the instrument for judgment. The 5. Patients with blood transfusion or under blood products
control can be determined periodically according to the laboratory treatment in recent momths may have false positive results.
requirements. The results obtained should be within the set limits. 6. The hemolytic, lipemia, jaundice, and contaminated samples
If quality control is out of control, appropriate corrective actions may affect the test results. Such samples should be avoided.
must be taken in accordance with laboratory rules and regulations. 7. If the sample contains interfering substances such as rheumatoid
For detailed instructions on quality control, refer to the Quality factor,HAMA,etc, the test results will also be abnormal.
Control Application and Quality Control Inquiry sections of the 8. We do not test all types of collection tubes that may be used for
IFU of the semi-automatic chemiluminescence immunoassay this kit; therefore, for blood sample collection tubes from different
analyzer. manufacturers, different results may be obtained due to different
Calculations raw materials and additives.
The system software uses a standard curve to automatically
[Product Performance Index]
determine the test results of a patient. The results are given in
1. Negative conformity rate: testing positive reference material of
AU/mL.
the company ,there is no false negative result.
[Reference Interval] 2. Positive conformity rate:. testing negative reference material of
Cut-off value: 1.0AU/mL the company ,there is no false positive result.
Note: Due to the complexity of clinical conditions, differences 3. Repeatability: Coefficient of variation (CV) ≤ 10%.
between individuals and regional differences may affect the test
[Precautions]
results; therefore, each laboratory is recommended to establish its
1. This product is only used for in vitro diagnosis, not for other
own reference interval based on the actual situation of its own
purposes; do not use expired reagents.
laboratory. The above data is for reference only.
2. All reagent components, samples and various wastes should be
[Interpretation of Test Results] treated as infectious agents. At the same time, this product is a
1. When the test result of the sample is <1.0 AU/mL, it means that one-time use product, and it should be destroyed centrally in
the test result of the sample is negative. accordance with relevant laboratory requirements after use.
2. When the test result of the sample is ≥1.0 AU/mL, it means that 3. Please read the instructions carefully before operation, and
the test result of the sample is positive. follow the instructions. During use, all laboratory reagent handling
3. The test results of this kit are for clinical reference only, and precautions must be followed.
cannot be used as the sole basis for clinical diagnosis. 4. Wear appropriate gloves. If the test strip accidentally leaks, and
Comprehensive determination should be made in combination the skin accidentally comes into contact with the reagent, rinse
with clinical symptoms and epidemiological history. with plenty of water.
5. The test strip should be stored at 2 - 8 ℃, and should not be
[Limitations of Test Method]
stored frozen. Equilibrate to room temperature before use.
1. Due to limitations of methodologies or antibody specificity,
6. Due to methodologies or antibody specificity, different results
laboratory technicans should notice negative results and make
may be obtained on the same sample when different
decision in combination with other testing result.It’s recommended
manufacturers' reagents are used. The results obtained with
to retest debatable samples through virus isolation culture
different reagents should not be directly compared with each other,
identification.
so as not to cause erroneous medical interpretation; the laboratory
2. During early infection, when IgM Ab isn’t formed or the
is recommended to specify the characteristics of the reagents used
concertration is very low, it will cause a negative result. If there is
in the test report issued to the clinician.
a suspected infection,it’s recomended to retest in 7-14 days to
2/3
[Explanation of Labels]
[References]
1. Boscato Lm, et al. Heterophilic antibodies: a problem for all
immunoassays. Clin Chem, 1988, 34:27-33.
2. Kricka L. Interferences in immunoassays - still a threat. Clin
Chem, 2000, 46: 1037–1038.
3. Bjerner J, et al. Immunometric assay interference: incidence and
prevention. Clin Chem, 2002, 48: 613–621.
4.Feng Renfeng, The Quality Management Technique Bases in
Clinical laboratory, 2007, Apr
5.Peng Wenwei, Lemology (V5.0), PMPH,2001.
[Basic Information]
Registrant/Manufacturer Name: Shenzhen Watmind Medical Co.,
Ltd.
Address: 8th Floor, Building A, 16-1 Jinhui Road, Jinsha
Community, Kengzi Street, Pingshan District, Shenzhen, China
Phone: +86 755-86538356 Fax: +86 755-26658059
Postcode: 518055 Website: http://www.watmind.com/
Production address: 8th Floor, Building A, 16-1 Jinhui Road,
Jinsha Community, Kengzi Street, Pingshan District, Shenzhen,
China
3/3