Notes: 1. This Table details recalibration and check intervals, but each item of equipment requires appropriate calibration and /or
Uncertainty
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
General
Chemical Use AR grade [grades other than AR grade 4 Consult UKAS In-house, N/A
reagents chemical reagent may be used where permitted guidance eg, labelling
by the test method] from reputable publications QC checks
manufacturer (if possible ISO 9000 approved). on chemical or charts
Establish date of expiry reagents
Force measuring (a) Compression Calibrate against BS 1610: Part 1 1 Must not exceed UKAS Yes
devices testing machines 1 year
for concrete & Calibrate machine performance using strain 1 Must not exceed UKAS N/A
mortars cylinder 1 year
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Force measuring Check rate of load pacer and that platens, 3 Yearly In-house No
devices auxiliary platens and spacing blocks meet the
— continued specifications given in the relevant standard
Hardness and surface texture of platens 2 Initial and after UKAS test No
and auxillary platens maintenance or
repair
(b) Loading Calibrate in accordance with the relevant 2 At least yearly In-house Yes
frames for standard, eg, BS 812, BS 598, BS 1377
soils, aggregates
Hydrometer Density Shall comply with BS 718 1 Initial only UKAS Yes
Soils.
Scale calibrate in accordance with 3 Yearly In-house No
BS 1377 : Part 2
2. Micrometers
(a) Reference Calibrate externally 1 Yearly UKAS Yes
(b) Working Calibrate against calibrated gauge blocks 2 Yearly In-house Yes
Page 11 of 29
Uncertainty
Length 3. Callipers
measuring (a) Reference Calibrate externally 1 Yearly UKAS Yes
devices (b) Working Calibrate against calibrated gauge blocks 2 Yearly In-house Yes
— continued meeting the appropriate grade requirements
of BS EN ISO 3650
4. Steel rules
(a) Reference Calibrate externally 1 5 yearly UKAS Yes
(b) Working Rules complying with BS 4372;
Engineers' rule not confirmed as BS 4372 check
against reference using suitable visual aid 2 Initial In house Yes
All rules – visual check for readability and wear 4 Yearly In-house No
All rules – recalibration 2 5 yearly In-house Yes
6. Straight edge/
Engineers' square
(a) Reference Calibrate externally 1 5 yearly UKAS Yes
(b) Working Carry out appropriate checks to satisfy the 3 Yearly In-house No
relevant standards
Muffle furnace Check temperature profile in working space 3 Initial and after In-house No
maintenance or
repair
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
For soils testing to BS 1377 : Part 9, carry out 3 As specified by In-house N/A
the following: method
(a) bulk density: verify by comparison with
in situ measurement or preferably the bulk
container method
(b) moisture content: verify by comparison to
moisture contents obtained by BS : 1377 : Part 2
method or by the method given in (a) above In-house
Ovens (a) Required Check temperature profile in working 3 Initial and after In-house No
(conventional) temperature space with a calibrated reference thermometer maintenance
tolerance or calibrated reference thermocouple. Clause or repair
< ± 2 °C 5.6.1 of prEN 932-5 may be suitable
Page 13 of 29
Uncertainty
Ovens (b) Required Check temperature profile in working space 3 Initial and after In-house No
(conventional) temperature with a BS 593 thermometer or calibrated maintenance
— continued tolerance thermocouple (with a unique serial number) or repair
³ ± 2 °C
Check temperature at the midpoint 3 At least yearly In-house No
of the working space with BS 593 depending on use
thermometer or calibrated thermocouple
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
(a) Reference: Calibrate for precision and range of 1 5 yearly UKAS Yes
liquid-in-glass measurement required
and
Check at ice point (or exceptionally, some 3 Yearly In-house No
other reference point).
(b) Reference: Calibrate for precision and range of 1 1 – 5 yearly UKAS Yes
platinum measurement required depending on
resistance requirement of
(c) Working: Calibrate against reference PRT or reference 2 Yearly In-house Yes
platinum thermometer
resistance
(e) Working: Use a BS 593 thermometer (with unique serial 2 5 yearly; In-house Yes
Page 15 of 29
Uncertainty
Time measuring (a) Required Calibrate appropriately according to 1 Yearly UKAS Yes
devices tolerance required accuracy
£ ± 0.5 sec
Volumetric Where Calibrate by calculating the volume or amount 2 Initial, then In-house N/A
glassware calibration is of distilled water contained or delivered at a consult UKAS
specified or measured temperature (referring to the guidance
Grade A required temperature correction tables in BS 1797) publications
by test method) or by reference to appropriate test method on volumetric
glassware
Volume-change (a) Transducers Calibrate by weighing the amount of distilled 2 Yearly In-house Yes
Edition 2 ✺ October 2000
(b) Burette Calibrate by weighing the amount of distilled 2 2 yearly In-house Yes
type indicators water delivered at a measured temperature
(referring to the correction tables in BS 1797)
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Water baths, (a) Required Check bath or tank temperature with an 3 During use In-house No
tanks etc, temperature appropriate calibrated thermometer or
constant tolerance calibrated continuous recording device
temperature < ± 2 °C
Check temperature profile with a calibrated 3 Initial and after In-house No
thermometer; refer to the relevant BS or maintenance
standard method for individual requirements or repair
(b) Required Check bath or tank temperature with a 3 During use In-house, No
temperature BS 593 thermometer or calibrated max/min eg,test
tolerance thermometer worksheet
(b) Spring Check with working masses at suitable points. 3 Before each use In-house No
balances
Class E1 & E2
Uncertainty
Aggregates
Abrasion machine Check:
dimensions of moulds, trays and plates; 3 2 yearly In-house No
speed of rotation of machine; 3 2 yearly In-house No
mass of weights; 3 Before each run In-house No
rate of flow from feed mechanism 3 At start of run In-house No
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Corn emery Natural corn Check the grading of a sample from each batch 3 Before In-house No
emery for PSV commissioning
determination each batch
Uncertainty
Impact testing Aggregate Check (refer to BS 812: Part 112: clause 5.1.1):
device impact value
apparatus dimensions of cylindrical steel cup 3 2 yearly In-house No
and metal hammer;
mass of circular metal base
and metal hammer; 3 2 yearly In-house No
height of drop and free fall of hammer. 3 Before each test In-house No
thickness of reaction block/floor 3 Initial and In-house No
if moved
Thickness gauges Flakiness Check with calibrated caliper or micrometer 3 Yearly In-house No
index/elongation
Bituminous materials
Compaction Automatic Check hammer mass, freefall distance, foot 3 Yearly In-house No
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Hardness number Mastic asphalt Calibrate penetration dial gauge 2 Yearly In-house No
apparatus measuring dial Calibrate force applied 2 Yearly In-house No
Check dimensions of indenter pin 2 Yearly In-house No
Check general condition and wear on 4 Regularly, No
indentor pin depending on use
Penetrometer Needle Calibrate needle against requirements of 1 100 uses or UKAS Yes
BS 2000: Part 49: Section 5.3 5 years,
whichever occurs first
Softening point Apparatus Check frame assembly, guides, balls and rings 3 Yearly In-house No
against requirements of BS 2000 : Part 58
Page 21 of 29
Page 22 of 29
Uncertainty
Concrete
Air content apparatus(a) Type A Calibrate apparatus. 3 Yearly, & when In-house No
Refer to BS 1881: Part 106: clause 5.3 site of operation
varies in elevation
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
(b) Check against test block (s) 3 Before each use In-house No
Curing tanks (a) Required Check tank temperature with a calibrated 3 During use In-house No
temperature max-min thermometer or calibrated continuous
tolerance temperature recording device
< ± 2 °C
Check temperature profile of tank with a 3 Initial and In-house No
a calibrated thermometer after maintenance,
adjustment or
repair
(b) Required Check tank temperature with BS 593 3 During use In-house, eg No
temperature thermometer, calibrated continuous test worksheet
Page 23 of 29
Uncertainty
Half cell (a) Millivolt Check millivoltmeter output against reference 3 Monthly In-house No
equipment meter calomel electrode of known millivolt output
(b) Copper & Check millivolt meter 1 Yearly UKAS Yes
silver cells
Slump rod/ Check dimensions and mass for tamping bars 3 Yearly In-house No
tamping bar
Soils
depending on use
Depth gauge Calibrate, eg using calibrated gauge blocks 2 Yearly In-house Yes
Hanger weights Consolidation & Check weight using appropriate calibrated 3 2 yearly In-house No
direct shear tests balance
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Refer to
BS 1377: Part 6: clauses 3.2.5 & 3.2.6
Uncertainty
Penetration Field Calibrate and check components of apparatus 2,3,4 As appropriate In-house Yes/No
resistance tests to ensure compliance with BS 1377: Part 9
Edition 2 ✺ October 2000
apparatus
pH meter Check with at least two standard buffer solutions 3 Before each use In-house No
of appropriate pH
Pore water Calibrate pressure transducer or gauge using a 2 6 monthly In-house Yes
pressure calibrated reference pressure gauge
measuring
device
Edition 2 ✺ October 2000
Uncertainty
Type of budget
Item Type Calibration or check Level Interval certificate required?
Ring shear For component items such as dial gauges 2, 3 Refer to In-house Yes/No
apparatus and load application, refer to individual
BS 1377 : Parts 1 and 6 listings
Sand In situ density Check a sample from the batch for grading 3 Before In-house No
replacement and purity requirements. Refer to commissioning a
sand BS 1377 : Part 9 batch