Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
www.elsevier.com/locate/jfoodeng
a
Agricultural and Food Engineering Department, Indian Institute of Technology, Kharagpur, West Bengal 721 302, India
b
Faculty of Agricultural Engineering, Department of Food Technology, Bidhan Chandra Krishi Viswavidyalaya, Mohanpur, Nadia,
West Bengal 741 252, India
Received 28 July 2000; accepted 8 January 2001
Abstract
A laboratory model indirect type helical tube ultra-high temperature (UHT) milk sterilizer was designed and tested. The helical
triple tube heat exchanger modeling and simulation was carried out for heating milk from 90°C to 150°C in the annulus of the triple
tube using steam in the innermost and outermost tubes. The holding section was designed based on an Arrhenius model to reduce
B. stearothermophilus microorganisms in the milk by 8 log cycles at a sterilization temperature of 150°C with a residence time of 2.64
s. The helical double tube cooling section modeling and simulation was carried out to cool milk from 150°C to 90°C by using tap
water. The lengths of heating, holding and cooling sections were found to be 2.28, 1.0 and 2.08 m, respectively. The set-up was tested
for UHT processing of milk at the ¯ow rate of 135 l/h. The microbial eectiveness of the process was tested separately with in-
oculated samples of B. subtilis and B. stearothermophilus microorganisms and found to be eective during processing as well as
during storage of the UHT processed milk for two months. The quality attributes (i.e. color and viscosity) of the UHT processed
milk were also tested during the processing and storage periods. Overall there was no marked change in quality attributes for which
the processed milk can be considered unacceptable. Ó 2001 Elsevier Science Ltd. All rights reserved.
2. Design considerations
*
Corresponding author. Fax: +91-3222-82244. A UHT sterilizer consists of three sections, namely a
E-mail address: akd@agfe.iitkgp.ernet.in (A.K. Datta). heating section, a holding section and a cooling section.
0260-8774/01/$ - see front matter Ó 2001 Elsevier Science Ltd. All rights reserved.
PII: S 0 2 6 0 - 8 7 7 4 ( 0 1 ) 0 0 0 3 0 - 9
14 P.K. Sahoo et al. / Journal of Food Engineering 51 (2002) 13±19
Since milk ¯ows in the annular region of the tubes, the The length of the holding tube can be obtained from
above equation can be modi®ed, replacing
lb =lwall 0:14 L tp vmax :
18
by
Di2 =Do1 0:53 (Zuritz, 1990).
The heat transfer coecient for the helical tube heat
exchanger is obtained by multiplying
1 3:5Deq =Dc 2.3. Cooling section design
with heat transfer coecient of straight tube heat ex-
changer (Chopey & Hicks, 1984). So the ®nal expression The overall heat transfer coecient can be obtained
to evaluate the milk side heat transfer coecient for from
turbulent ¯ow in the annular region of the helical triple 0 1
Di1 D0i1 ln
D0o1 =D0i1 1
tube is given by Ui1 :
19
D0o1 hw 2kss hm
0:53
km 0:8 0:33 Di2 Deq Water and milk side heat transfer coecients are esti-
hm 0:027NRe NPr 1 3:5 :
Deq Do1 Dc mated by the following equations:
! 0 0:53 " #!
11 D0eq
kw 0:8 0:33 Di2
hw 0:027NRe NPr 1 3:5 ;
The equation is valid for NRe P NRec (Since the critical D0eq D0o1 D0c
Reynolds number indicates the onset of turbulence) and
20
0:7 < NPr < 700 (Geankoplis, 1997). 0:14 0
The heat exchanger length can be calculated from the km 0:8 0:33 lb D
hm 0:027NRe NPr 1 3:5 i10 :
heat balance equation, which is given below D0i1 lwall Dc
m_ m Cm DTm
Uo1 Ao1 Ui2 Ai2 DTlm
21
Uo1 pDo1 Ui2 pDi2 LDTlm :
12 By employing the heat ¯ux equivalence concept, the wall
temperature can be estimated as follows:
Tbm Ti1wall Ti1wall Tbw
2.2. Holding section design ;
22
0
1=hm Di1 ln
D0o1 =D0i1 1
2kss
hw D 0
o1
The average velocity of milk through the holding tube
Tbw To1wall To1wall Tbm
can be obtained from :
23
0
1=hw Do1 ln
D0o1 =D0i1 1
V_ 2kss hm D0
i1
v :
13
A The length of the cooling section can be obtained from
The maximum velocity of milk can be obtained from the heat balance equation as follows:
v m_ m Cm DTm m_ w Cw DTw ;
24
vmax
14
0:0336
log NRe 0:662 Ui1 Ai1 DTlm Ui1 pD0i1 LDTlm :
25
(For NRe > 2100) (Toledo, 1991).
D150 is estimated from the Arrhenius equation.
3. Materials and methods
Ea 1 1
D150 D121 exp ;
15
R T T0 An isometric view of the sterilizer set-up is shown in
where T
150 273 K, T0
121 273 K Fig. 5. The sterilizer is made of 316 grade stainless steel.
It consists of a heating section, a holding section and a
Ea ln
10 T T0
:
16 cooling section. The dimensions of tubes of all sections
R z are given in Fig. 3. The porting and the instrument ar-
Ea =R is calculated within the temperature range 373±398 rangements at inlet and outlet of each section are shown
K (Burton, 1988). in Fig. 4. The heating and holding section of the steril-
D121 181 s; izer was well insulated with high-class insulating mate-
rial to avoid heat loss during the experiment. A steam-
z 9:43°C: jacketed pan was attached to a centrifugal pump to
The holding time of process ¯uid (milk) in holding tube pump milk from the pan to the heating section of the
can be estimated as follows. sterilizer. Five thermocouples were connected at the inlet
and outlet of the heating section, the outlet of the
tp
SV
D150 :
17
holding section, the outlet of the cooling section and in
For the target microorganism, B. stearothermophilus the steam condensate to record observations during the
(most heat resistive spore forming bacteria), the steril- experiment. A digital temperature indicator was used to
izing value (SV) is 8 (Burton, 1988). indicate temperature. Two Bourdon tube pressure
16 P.K. Sahoo et al. / Journal of Food Engineering 51 (2002) 13±19
gauges were ®xed at the inlet and outlet of the heating ¯ow control valve was used for a ®nal experiment where
section and one at the end of the cooling section to re- the desired sterilization temperature of milk was at-
cord pressures at dierent time intervals during the ex- tained. Glass bottles and caps were sterilized in an au-
periment. A VASPA electrode steam boiler was used to toclave at 15 psi pressure for 15 min for packaging the
supply steam as the heating medium to the heating milk. Prior to sterilization, the milk was preheated in a
section of the sterilizer. The working pressure range of steam-jacketed pan to 93°C to minimize the fouling
such a boiler is 0±10 bar and the electric supply is 3- problem in the heat exchanger and the developed cul-
phase, 440 V and 50 Hz. A nozzle was ®tted with a tures of B. stearothermophilus and B. subtilis were in-
discharge tube for easy packaging (Figs. 3±5). oculated separately at two batches and at the same time
The testing was carried out in the constant wall a boiled milk sample was taken for milk quality study.
temperature helical triple tube UHT sterilizer with raw Preheated milk was pumped to the heating section of
clari®ed whole milk collected from the local market. the sterilizer for ®nal heating to the sterilization tem-
Prior to the experiment, the sterilizer as well as the perature using steam at 160°C. Heating was followed by
steam-jacketed pan was cleaned using cold and hot holding in the holding section at the sterilization tem-
water with a surface active agent. All connections were perature to attain perfect sterility. Then the sterilized
checked for leakage. The equipment was sterilized by milk was cooled to 90°C in the cooling section using tap
using hot water at 150°C. Then the trial was initiated water and aseptically packed into glass bottles using two
using milk by regulating the ¯ow control valve. A ball- spirit lamps at the nozzle tip and then capped using a
type ¯ow control valve is located at the end of cooling capping machine. Packed milk was cooled to room
section to regulate the pressure in the sterilizer so that
the process ¯uid does not boil at the sterilization tem-
perature. The pressure, which is maintained in the
sterilizer due to regulation of ¯ow control valve, is called
back pressure. Dierent observations of milk ¯ow rate,
temperature and back pressure were recorded at dier-
ent settings of ¯ow control valve. That setting of the
Table 1
Experimental observations in helical triple tube heat exchanger
Back pressure (bar) Flow rate (l/h) Milk inlet temperature (°C) Milk outlet temperature (°C) Reynolds number
Table 2
Bacterial counts (cell/ml) for various milk samples
Acknowledgements
References