1. Combined low pressure chamber.


The combined chamber of reduced pressure with a working area of at least 12.0 cubic meters. m. complete with a multi-
channel precision temperature meter (16-channel, measurement accuracy of at least 0.3 mK, simultaneous operation with
different types of sensors (TS, TP, pressure, humidity, communication with a computer via RS232C and USB)) and platinum
resistance thermometers 16 vibration-proof on a flexible heat-resistant cable in the amount of 16 pcs. for testing large-sized
products for high, low temperature, high humidity, low atmospheric pressure.

The size of the working area of the climate chamber of cold, heat and moisture is not less than 2000x2800x2000 mm
(WxHxD).
Overall dimensions with engine room no more than 2350x3350x3600 mm (WxHxD).
Temperature Range:
a) lower temperature value:
- minus 50 ° C for testing products without heat;
- minus 35 ° C for testing products with heat up to 15 kW;б)
upper temperature value plus 85˚С.

Maintenance Accuracy:
- temperature not less than ± 1,0˚С;
- relative humidity not less than ± 3.0%.
Uneven maintenance:
- temperature in volume not more than ± 2.0˚С;
- humidity not more than ± 3.0%.
Temperature change rate:
- heating not less than 2.0 ° C / min;
- cooling not less than 1.0 ˚С / min.

Relative humidity in steady state, in an unloaded chamber in the range from 20% to 95% at a temperature of from 20 ° C to 55 ° C;
The maximum permissible working load on the bottom of the chamber is at least 1000 kg.
The equipment should be equipped with accessories to ensure a tight input into the chamber of supply and signal circuits with the
following parameters:
- the number of contacts designed for a rated current of 100A - at least 10 pcs .;
- the number of contacts designed for a rated current of 3 A - at least 4 pcs .;
- the number of contacts for signal circuits (connecting thermocouples dTPL-HK) - at least 12 pcs.
The door is single-leaf, the doorway is at least 1750x3000 mm.

Heated viewing window on the wall of at least Ø500 mm.


2 technological openings of at least Ø100 mm must be provided.
Power supply: 380V, three-phase, 50 Hz.
Used refrigerants R404A, R23.
The total mass of the camera is not more than 4500 kg.
The internal working volume of the chamber must be made of stainless steel, the material of the chamber body is powder-
coated steel, the thickness of the sandwich panels is at least 160 mm.
Low Pressure Range 1 mmHg

2. The combined chamber of cold, heat, moisture and solar radiation


The supply of a chamber of cold, heat, moisture and solar radiation with a working area of at least 3.5 m3 complete with a
hygrometer with a set of 3 temperature and humidity probes is designed to test products for elevated, low temperature, high
humidity and solar radiation.
Characteristics of a hygrometer with a display: design for simultaneous connection of 3-7 probes of relative humidity and
temperature. The hygrometer should record the date and time, the measurement results from all the probes at the same time,
key events at a removable flash card at a specified time interval. The hygrometer must have protection against power failure,
protection of the measured data from unauthorized access or modification (provided by recording in a protected format with
the insertion of a digital signature, setting passwords for access to data).
It must be possible to connect control relays and open / close sensors to the hygrometer. Pre-installed software for the
hygrometer should provide convenient and intuitive access to settings and measurement results, provides the construction and
analysis of graphs and tabular data, the formation of measurement protocols, and data transfer to the user database.

1
Characteristics of the temperature and humidity probe:
Accuracy: ± 0.8%, ± 0.1 kat10 - 30 ° C
Humidity measurement range: 0 - 100%
Operating temperature range: -100 to +200 ° C *
Diameter: 15mm
Length (probe head): 100 mm
Cable length 5 m
Case Material: PEEK
Weight no more than 470 g.

Temperature range:
- cooling -70˚С;
- heating (with the source of solar radiation turned off) + 120˚С;
- heating (when the source of solar radiation is on) from + 40 ° C to + 55 ° C; *
- relative humidity in steady state, in an unloaded chamber in the range from 20% to 95% at a temperature of from 20 ˚ С to 60
*С *
- integral heat flux density of not less than 1120 W / m2, including the flux density of the ultraviolet part of the spectrum
(wavelength 280 ÷ 400 μm) - 68 W / m2

Maintenance Accuracy:
- temperature not less than ± 1,0˚С;
- relative humidity not less than ± 3.0%;
- the integral heat flux density is not more than 10.0%, while the accuracy of maintaining the density of the ultraviolet part of
the spectrum is not less than 25.0%.
The rate of temperature change in the unloaded chamber volume:
- cooling not less than 1,0˚С / min;
- heating not less than 3.0 ° C / min.
The time to reach maximum relative humidity in an unloaded chamber is 40 minutes.
The uneven distribution of the working volume:
- temperature not more than ± 2.0 ° C;
- humidity not more than ± 3.0%.

Dimensions of the tested plane of solar radiation of at least 740 x 740 mm.
The distance from the simulated solar radiation source to the test plane is 900 mm.
The size of the working area of the camera (WxHxD) not less than 1300x2100x1250 mm
The door is single-wing, hinged with a left hanger on the front side of the camera, the doorway is 1300x2100 mm.
Heated multi-layer viewing window in the camera door 500h600mm.
Two process openings of at least Ø100 mm are provided.
The internal working volume of the chamber must be made of stainless steel, the material of the chamber body is made of
powder-painted steel.
Thermal insulation material - non-combustible polyurethane foam.
Through holes 2 pcs. not less than Ø100 mm.
The case and the heating elements of the steam generator are resistant to corrosion.
The camera cooling system is air.
Used refrigerants R404A, R23.
Power supply: 380V, three-phase,
Power not less than 45 kW.
The bottom of the chamber must withstand a normally distributed load of at least 1000 kg;
The total mass of the camera is not more than 1900 kg.
The design of the solar radiation simulator is located on the outside of the camera on the roof and has the ability to move
along the guides in the working (above the previously open hatch in the roof of the camera) and the inoperative position, away
from the hatch, allowing it to be closed.
The position of the manhole cover (closed or open) and the position of the solar radiation simulator must be mutually blocked
from accidental switching on.
Protective transparent glass should be installed in the hatch from the side of the plane of the ceiling of the camera.
Lighting the internal volume of the camera.
The camera has a running hours counter.
The creation of increased relative humidity should be carried out by means of a separate closed steam generator.

2
Camera control
Camera control should be carried out by means of a Russified programmable temperature and humidity controller with a color
7-inch touch screen.
The main parameters of the controller:
Screen resolution of at least 640x480;
Control modes: programmable and fixed values ("manual mode");
Memory capacity: at least 1000 programs with 120 steps in each and at least 99 cycles in each program.
Set and display ranges:
- temperatures from minus 100 ° C to 200 ° C;
- humidity from 0 to 100%;
- time 9999 hours 59 minutes.
Automatically save information when power is turned off.
Set resolution not less than:
- temperature, ± 0.01 ° C;
- humidity, ± 0.1%;
- time, 1 minute.

Temperature sensors and a psychometric humidity sensor based on 2 sensors of the Pt100 type;
Monitoring and control using the PID controller.
The controller has the following menu functions:
- Dialog mode input from the touch panel;
-Display of set and current values of temperature and humidity;
- Display of program number, step and working time;
- Editing, allowing you to enter multiple values of temperature and humidity on one page. Displays the status of the running
program.
The function of backlighting the screen, setting the on and off time, and the delayed start of the program should be provided.
Locking the controller display with a user-entered password.
The presence of a switching interface, RS-485, Ethernet.
The presence of a warning system (light and sound signals) and a power outage in the event of fire, excess of supply voltage,
overheating of the chamber, excess of freon pressure and compressor current.
The power level of acoustic noise when the camera is operating at a distance of 1 m from the camera, not more than 70 dB.
The estimated operating time of the camera for one failure is at least 8000 hours.