Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
English
3
Elcometer 456
Basic Models
Operating Instructions
Op_456_3_English.book Page 0 Monday, February 14, 2011 3:32 PM
English
3
These instructions apply to the following Elcometer 456 models:
Ferrous (F), Non-Ferrous (NF) and Dual Ferrous/Non-Ferrous (FNF)
1
Op_456_3_English.book Page 2 Monday, February 14, 2011 3:32 PM
CONTENTS
Section Page
1 About your gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.3 What this box contains . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Conventions in these instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.5 Quick-start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2 Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Fitting the batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2 Battery condition. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Fitting probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 The controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.5 Switching the gauge on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.6 Switching the gauge off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.7 The screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.8 Selecting a language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3 Taking a reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
2
Op_456_3_English.book Page 3 Monday, February 14, 2011 3:32 PM
3
Op_456_3_English.book Page 4 Monday, February 14, 2011 3:32 PM
7 Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8 Personalised welcome screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
9 Storage and transit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
10 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
11 Statistics terminology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
12 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
13 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
14 Related equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
15 Fitting the wrist harness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
16 Probe measurement performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
17 Probe capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
18 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
19 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4
Op_456_3_English.book Page 5 Monday, February 14, 2011 3:32 PM
Thank you for your purchase of this Elcometer 4563 The gauge is available in
Coating Thickness Gauge. Welcome to Elcometer. three versions; Basic,
Elcometer are world leaders in the design, Standard and Top. This
manufacture and supply of coatings inspection manual describes the
equipment. Our products cover all aspects of operation of the
R
5
Op_456_3_English.book Page 6 Monday, February 14, 2011 3:32 PM
FERROUS (F)
ASTM B 499, ASTM D 1186-B, ASTM G 12,
BS 3900(C5), BS 5411 (11), DIN 50981,
IMO MSC.215 (82/84), ISO 1461, ISO 2063,
ISO 2808-7C, ISO 2808-12, ISO 19840,
NF T30-124, SSPC-PA2 (2004)
a. International Marine Organisation, Performance Standard for
Protective Coatings
6
Op_456_3_English.book Page 7 Monday, February 14, 2011 3:32 PM
Rechargeable batteries can be used but they will 16% to 33%, replacement required.
only have 25% to 30% of the life of alkaline
batteries. <16%, gauge beeps every 10 seconds
and symbol flashes - immediate
replacement required.
5 loud beeps, gauge switches off
automatically.
8
Op_456_3_English.book Page 9 Monday, February 14, 2011 3:32 PM
9
Op_456_3_English.book Page 10 Monday, February 14, 2011 3:32 PM
probes:
F1
456 F
Press key to
LED
switch on gauge, or R
On/Off key
Figure 3. Elcometer 4563 control keys Note: Before switching the gauge on for the first
time read “Selecting a language” on page 12.
10
Op_456_3_English.book Page 11 Monday, February 14, 2011 3:32 PM
2.6 SWITCHING THE GAUGE OFF When the gauge is switched on a welcome
To switch off all gauge information screen may be displayed briefly (4).
types, press and hold 3 seconds
11
Op_456_3_English.book Page 12 Monday, February 14, 2011 3:32 PM
12
Op_456_3_English.book Page 13 Monday, February 14, 2011 3:32 PM
13
Op_456_3_English.book Page 14 Monday, February 14, 2011 3:32 PM
456
09 / 10 / 2000
F1
F
Automatic-switching
CAL DATA STATS MENU
probe symbol
R
(FNF probes only)
R
Battery state
Substrate
Note: If CAL softkey symbol is flashing the gauge 4.2 MAIN MENU
should be recalibrated. This is due to the calibration Gauge configuration and measurement functions
adjustment method having been changed or a are controlled using menus (12). The structure of
probe change - see “Calibration adjustment” on the menus is shown on page 19.
page 23.
Menu title
Menu contents
Softkey functions
15
Op_456_3_English.book Page 16 Monday, February 14, 2011 3:32 PM
BACK softkey returns the gauge to a previous 4.3 MAIN MENU - EXTENDED MENU OFF
screen. Holding this softkey down will rapidly exit
from any menu and return to the reading screen.
16
Op_456_3_English.book Page 17 Monday, February 14, 2011 3:32 PM
17
Op_456_3_English.book Page 18 Monday, February 14, 2011 3:32 PM
D
USA GAUGE : Resets gauge to USA default
settings e.g. MM/DD/YY date format and imperial
units.
18
Op_456_3_English.book Page 19 Monday, February 14, 2011 3:32 PM
19
Op_456_3_English.book Page 20 Monday, February 14, 2011 3:32 PM
The following features are added to the MENU If there is no reading to delete, the gauge displays:
when EXTENDED MENU is active:
PRINT/OUTPUT
Outputs individual readings or a statistical summary
via the RS232 interface.
Figure 17. Delete last reading not available
To use this function, first setup using:
screen
MENU/SETUP/OUTPUT - see “OUTPUT:” on
page 21. Press OK to return to Delete menu.
If no printers have been setup, PRINT/OUTPUT will CAL METHOD
display a NOT AVAILABLE message.
Allows selection of calibration method from list of
DELETE options - see “Calibration method” on page 23 for
Deletes last reading only. Gauge displays ARE more details.
YOU SURE? (16). SETUP
Used to select, change or activate gauge features:
STATISTICS: Activates simple statistics feature.
Display - see “Display” on page 33.
Set NDFT - see “Set NDFT” on page 32
Figure 16. Delete last reading confirmation
screen Select stats - see “Select stats” on page 32.
Softkey enabled - switches STATS softkey on/off.
Press NO softkey to include reading in statistical
summary or YES softkey to delete reading.
20
Op_456_3_English.book Page 21 Monday, February 14, 2011 3:32 PM
PROBE: Only for dual function probes (FNF and F1 OUTPUT: Selects baud rate (18) and activates
2). Changes probe mode. data output via the RS232 interface - see
FNF probes - select from Automatic, F or N. “Interfaces” on page 13.
F1 2 probes - select from F1 or F2.
UNITS: Units are automatically set by the probe
type, however the user can manually override the
automatic setting. Select from µm, mm, mil, thou or
inch. Figure 18. OUTPUT screen
AUTO SWITCH OFF: Changes delay before BAUD RATE can be set at values from 1200 to
gauge switches off when displaying Reading 19200. The default value is 9600 baud.
Screen.
RS232 BIT IMAGE Toggle tick box to
• Minimum = 1 minute activate/deactivate. When activated, readings are
• Maximum = 10 minutes sent to the RS232 interface as they are taken. All
• Default = 1 minute images and characters are output as bit-maps. This
Auto switch off may be disabled by selecting ‘off’ (In allows printing on the Elcometer Miniprinter (see
this case, switch off using On/Off key .) “Miniprinter” on page 43).
OUTPUT: Activates data output. Toggle tick box to RS232 PLAIN TEXT Toggle tick box to
activate/deactivate. When activated, readings are activate/deactivate. When activated, readings are
sent to the mini portable printer as they are taken. sent to the RS232 interface as they are taken. The
See “Miniprinter” on page 43 for sales part number gauge sends standard ASCII characters from the
of this optional accessory. Courier New font sete. This allows printing on
devices other than the Elcometer Miniprinter, e.g.
21
Op_456_3_English.book Page 22 Monday, February 14, 2011 3:32 PM
RS 232 printers or PC via Elcometer software or via SOFTKEYS LOCKED: When ticked, the softkey
HyperTerminal. functions will always be displayed.
BEEP VOLUME: Changes volume.
• 0 = off
• 5 = loudest
• Default = 3
Figure 20. Softkeys locked/unlocked
LANGUAGES: Allows selection of language.
LARGE FONTS: When ticked, displays all menus When un-ticked, the softkey functions disappear
and screens using double height, single width fonts from the reading screen 5 seconds after the reading
(for improved legibility, if needed). screen has been displayed, or 5 seconds after
pressing any key. To view the functions again,
press any one of the four softkeys.
Softkey functions are always visible in menus.
OPENING SCREEN: Disables the opening
Figure 19. Large fonts enabled (welcome) screens so that the gauge switches on to
display the reading screen.
If a personalised welcome screen has been
downloaded into the gauge (see page 36), Opening
Screen must be activated to display this screen.
22
Op_456_3_English.book Page 23 Monday, February 14, 2011 3:32 PM
23
Op_456_3_English.book Page 24 Monday, February 14, 2011 3:32 PM
ZERO OFFSET: This is the method described in Note: When the calibration method is changed, e.g.
ISO 19840 for coatings on steel surfaces from Smooth to Rough, the gauge will display a
roughened by blast cleaning. The calibration uses message (22).
the smooth surface technique, and a correction
value (zero offset) is applied to each reading to
account for the effect of the roughened surface; the
value depends on the surface profile - see Table 1.
SET OFFSET: This screen sets and changes the Figure 22. Recalibration required screen
offset for different surface roughness This value is
used only with the Zero Offset calibration method. If the NO softkey is pressed the CAL softkey
Table 1: Correction values from ISO 19840 symbol on the Readings Screen will flash to warn
that calibration adjustment is still required.
Profile according Correction Value
If the YES softkey is pressed the calibration
to ISO 8503-1 (µm) (Zero Offset)
adjustment procedure is activated - see “Calibration
Fine 10 adjustment procedure” on page 25.
Medium 25
Coarse 40 5.2 CALIBRATION FOILS AND STANDARDS
Calibration adjustment should be carried out with
the appropriate probe on the same type of metal,
the same curvature and similar finish to the item to
be measured. It is best to use an uncoated sample
of the items to be tested.
Calibration can be carried out using measured foils
or coated standards.
24
Op_456_3_English.book Page 25 Monday, February 14, 2011 3:32 PM
FOILS (SHIMS): These are coating thickness errors due to placing the foil labels between the
standards which have been measured using foils.
techniques independent of the gauge. They are
ideal for calibration because they provide a known Serial no. TK1009
Inspected by
value of thickness on the actual substrate to be micron mils/thou
502 19.75
measured. Calibration certificates for foils are
available upon request.
mils/thou
6.93
Inspected by
When using foils care must be taken to keep the
176.1
micron
foils clean and free from dust and to avoid damage
by creasing particularly the thinner foils. Always
remove a foil from its storage wallet before use. Figure 23. Stacking foils to increase thickness
When calibrating a High Temperature PINIP™
Probe use the special thickness standards supplied COATED STANDARDS: Thickness standards using
with the probe - see “Calibrating High Temperature typical substrate materials coated with hardwearing
PINIP™ Probes” on page 30. materials and measured using techniques
independent of the gauge.
Note: To calibrate 5 mm (200 mils) and 13 mm
(500 mils) range gauges it will be necessary to Coated standards are most often used to confirm
stack the foils (23). Care must be taken to avoid that the gauge meets its specifications if it is not
possible to use foils (shims).
25
Op_456_3_English.book Page 26 Monday, February 14, 2011 3:32 PM
26
Op_456_3_English.book Page 27 Monday, February 14, 2011 3:32 PM
these readings and the last reading. Repeat Note: - - - indicates over-range (26). Taking a
this action until a stable reading is obtained. reading within range clears this screen.
Average
27
Op_456_3_English.book Page 28 Monday, February 14, 2011 3:32 PM
reading. Repeat this action until a stable The gauge will display the option to test the
reading is obtained. calibration of the gauge.
28
Op_456_3_English.book Page 29 Monday, February 14, 2011 3:32 PM
29
Op_456_3_English.book Page 30 Monday, February 14, 2011 3:32 PM
30
Op_456_3_English.book Page 31 Monday, February 14, 2011 3:32 PM
6.1 ENLARGE STATS Figure 35. Reading screen with all statistics
Displays the chosen statistical values as double-
height characters. The example screens (34) 6.3 CLEAR STATS
appear when all the statistical values are selected. Resets to zero all statistical values selected in
The Up/Down softkeys can be used to move STATS MENU/DISPLAY.
through the list. OK softkey returns to the Reading
Screen.
31
Op_456_3_English.book Page 32 Monday, February 14, 2011 3:32 PM
32
Op_456_3_English.book Page 33 Monday, February 14, 2011 3:32 PM
• none of the remaining measurements shall be When PSPC statistics are selected the standard
below 0.9 x NDFT. statistics displayed on the screen are replaced by
those for PSPC (38):
34
Op_456_3_English.book Page 35 Monday, February 14, 2011 3:32 PM
NON-FERROUS
F N FNF N1 right angle 8 9 9
F1 * 9 8 9 N1A anodiser’s 8 9 9
F2 * 9 8 9 N1 PINIP™ 8 9 9
F1 2 * 9 8 9 NM3 miniature straight 8 9 9
F3 * 9 8 9 NM3 miniature 45° 8 9 9
F1 right angle 9 8 9 NM3 miniature 90° 8 9 9
F2 right angle 9 8 9 FNF1 * 8 8 9
DUAL
F1 2 right angle 9 8 9 FNF1 right angle 8 8 9
FERROUS
7.6 F1 2 HIGH TEMPERATURE PINIP™ Care must be taken to ensure that the
PROBES calibration conditions are not affected by
These probes are capable of measuring changes in the zinc or aluminium coating
on surfaces up to 250°C (480°F). Wear thickness. This can be determined by checking
appropriate protective clothing and take the zero over an area of the galvanised or
care to avoid bodily contact with the hot metal-coated steel. Metal coatings on steel
surface during measurement. above 50 µm (2 mil/thou) should be consistent
enough to obtain a stable zero on the layer of
Do not exceed the maximum measurement speed
when using these probes - see page 40. metal.
3. Take readings.
Note: Refer to special calibration procedure - see
“Calibrating High Temperature PINIP™ Probes” on
page 30.
8 PERSONALISED WELCOME SCREEN
A personalised welcome screen can be created and
7.7 COATINGS ON GALVANISED OR downloaded into the gauge.
METALLISED STEEL
Screen dimensions are 128 pixels x 64 pixels. The
The FNF probe in fixed N1 mode may be used to welcome screen is typically used to personalise the
measure paint coatings on galvanised, aluminium gauge with a logo, serial number, user name, etc.
(Al) or zinc (Zn) sprayed steel substrates. This is the first screen displayed when the gauge is
1. Set the gauge to the N1 mode switched on.
(MENU/SETUP/PROBE).
2. Zero and calibrate the gauge on a sample of 8.1 CREATING THE SCREEN
the coated steel - see “Calibration adjustment” 1. Download Elcometer ‘Welcome Screen
on page 23. Wizard’ software. This software is available
36
Op_456_3_English.book Page 37 Monday, February 14, 2011 3:32 PM
37
Op_456_3_English.book Page 38 Monday, February 14, 2011 3:32 PM
38
Op_456_3_English.book Page 39 Monday, February 14, 2011 3:32 PM
11 STATISTICS TERMINOLOGY
Term Meaning
Coefficient of Variation. The standard deviation divided by the mean for a group
COEF OF VARIAT’N
of readings, expressed as a percentage.
HIGHEST READING The value of the maximum thickness in a group of readings.
LOWEST READING The value of the minimum thickness in a group of readings.
The average of a group of readings; the sum of the individual readings divided
MEAN
by the number of readings.
Number of Readings. The running value for the number of readings taken in a
NO. OF READINGS group. In the case of the mode, the Number of Readings is the number of
values recorded, not the total number of readings taken.
Standard Deviation. A statistical measure of the spread of values in a group of
STD DEVIATION
readings.
PERCENT GREATER The percentage (%) of readings greater than or equal to the value of NDFT (see
OR EQUAL TO NDFT “Set NDFT” on page 32)
PERCENT BETWEEN The percentage (%) of readings less than the value of NDFT but greater than
90% AND 100% NDFT 90% of the value of NDFT.
90/10 test results: = pass, = failed on 90% test, = fail due to a reading
90:10 PASS / FAIL below 0.9xNDFT (if a number between 1 and 10 is displayed, insufficient
measurements have been taken to be statistically valid - take more measurements)
39
Op_456_3_English.book Page 40 Monday, February 14, 2011 3:32 PM
40
Op_456_3_English.book Page 41 Monday, February 14, 2011 3:32 PM
41
Op_456_3_English.book Page 42 Monday, February 14, 2011 3:32 PM
42
Op_456_3_English.book Page 43 Monday, February 14, 2011 3:32 PM
43
Op_456_3_English.book Page 44 Monday, February 14, 2011 3:32 PM
44
Op_456_3_English.book Page 45 Monday, February 14, 2011 3:32 PM
a. Whichever is the greater. Lower value achieved when calibrated close to the thickness to be measured.
45
Op_456_3_English.book Page 46 Monday, February 14, 2011 3:32 PM
17 PROBE CAPABILITIES
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
46
Op_456_3_English.book Page 47 Monday, February 14, 2011 3:32 PM
47
Op_456_3_English.book Page 48 Monday, February 14, 2011 3:32 PM
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
48
Op_456_3_English.book Page 49 Monday, February 14, 2011 3:32 PM
49
Op_456_3_English.book Page 50 Monday, February 14, 2011 3:32 PM
a. This is the recommended maximum calibration foil value to achieve the specified accuracy under these measurement
conditions
50
Op_456_3_English.book Page 51 Monday, February 14, 2011 3:32 PM
51
Op_456_3_English.book Page 52 Monday, February 14, 2011 3:32 PM
52
Op_456_3_English.book Page 53 Monday, February 14, 2011 3:32 PM
18 ERROR MESSAGES
Under certain conditions the gauge will display error messages (39). These messages are normally cleared
by pressing any one of the softkeys. The cause of the error will be indicated by the message and should be
corrected before proceeding.
Error message Causes Action to take
#1 - PROBE Probe-to-gauge communication failure. Integral gauge - return to Elcometer*.
Separate gauge - remove probe and refit.
If error persists, return to Elcometer*.
#2 - PROBE Corrupt data output from probe. Integral gauge - return to Elcometer*.
Separate gauge - try new probe.
If error persists, return to Elcometer*.
#3 - PROBE Internal error. Return to Elcometer*.
PROBE a) Probe is dual FNF, but gauge is ferrous a) Gauge model F and gauge model N
UNSUPPORTED only, or non-ferrous only. See page 34. cannot be used with a dual FNF probe.
b) Old gauge software does not support b) Return to Elcometer* for software
new probe. upgrade.
PROBE Probe has been changed. Calibration adjustment required.
CHANGED
UNSTABLE a) External electro-magnetic interference. a) Isolate from electro-magnetic
READING b) Gauge faulty. interference.
b) Return to Elcometer*.
53
Op_456_3_English.book Page 54 Monday, February 14, 2011 3:32 PM
54
Op_456_3_English.book Page 55 Monday, February 14, 2011 3:32 PM
19 INDEX
Numerics BS 5411 (3) 6 Coatings
2 Point calibration method 23 BS 5599 6 On galvanised steel 36
90/10 rule 32 On metallised steel 36
C Coefficient of Variation 39
Computer
A Cable
PC to gauge 43 Connection cable 43
Accessories 41
AS 2331.1.4 6 CAL softkey
Flashing 24 D
AS 3894.3-B 6
AS/NZS 1580.108.1 6 Calibration Delete last reading 20
ASTM B 499 6 Adjustment 24 Dimensions 40
ASTM D 1186-B 6 Coated standards 24 DIN 50981 6
ASTM D 1400 6 Coated thickness standards 25 DIN 50984 6
ASTM D 7091 6 FNF probes 23 Display
ASTM E 376 6 Foils 24 Changing size of fonts 22
ASTM G 12 6 Locking 16 Character size 11
Methods 23 Locking softkeys 22
Shims 25
B
Calibration Adjustment 25 E
Batteries Procedure, 2Point 29
Fitting 8 ECCA T1 6
Procedure, Rough 29 Elcometer 456
Life of 41 Procedure, Smooth 26
Precautions 37 Features 6
Procedure, Special substrate 29 Overview 5
Rechargeable Calibration certificates 42
Charging 41 EN 13523-1 6
Calibration methods Error messages 53
Life of 8 2 POINT 23
Specification 40 ESC 26
ROUGH 23 Extended menu
Baud rate 21 SET OFFSET 24
Beep Contents 20
SPECIAL SUB 23
Changing volume 22 ZERO OFFSET 24
Switching off 22 F
Character size 11
BS 3900 (C5) 6 Coated thickness standards 25 Factory calibration 17
BS 3900(C5) 6 Spares 42 FNF 34
BS 5411 (11) 6 Foils 25
55
Op_456_3_English.book Page 56 Monday, February 14, 2011 3:32 PM
56
Op_456_3_English.book Page 57 Monday, February 14, 2011 3:32 PM
57
Op_456_3_English.book Page 58 Monday, February 14, 2011 3:32 PM
Probes, maximum 40
Test certificates 42
Test readings 28
Testing calibration of gauge 29
Tick box 15
U
Units
Automatic setting of 21
Manual setting of 21
Up/Down 15
US gauge settings 18
US NAVY NSI 009-32 6
US NAVY PPI 63101-000 6
W
Weight 40
Welcome screen
Creating 36
Disabling 22
Wrist harness, fitting of 44
Z
Zero offset calibration method 24
58