Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
www.lumexinstruments.com
Application Examples
DETECTION
MATRIX SAMPLING ATTACHMENTS
LIMIT
Ambient air 0.5 ng/m³ 10 L/min
Natural and other RA-915M
2–500 ng/m³ 1–10 L/min RP-91NG
gases
Water 0.5 ng/L 1–20 mL RP-92
Urine 1 ng/L 1–5 mL (Cold vapor)
RA-915M
Solid samples (soils, with PYRO-915+
0.5 µg/kg 10–400 mg
rocks, sediments)
Water and solutions 1 µg/L 10–200 µL
Coal 1 µg/kg 10–300 mg
Sorbent traps
0.5 ng PYRO-915+
for stack gases*
RA-915M with RP-92
Biological samples
(tissues, liver, blood, 10–300 mg
hair, etc.) 1–5 µg/kg
Specifications
POWER ANALYSIS
DIMENSIONS (L × W × H) WEIGHT POWER SUPPLY
CONSUMPT ION TIME
RA-915M 90–240 VAC, Direct
470 × 110 × 210 mm 40 W
Air and gas analysis 7 kg 50/60 Hz, 12 VDC on-line
RP-92 attachment 360 × 225 × 125 mm 20 W 40–80 sec
430 × 340 × 135 mm (pyrolyzer) 100–240 VAC,
PYRO-915+
400 × 280 × 135 mm 17.5 kg 50/60 Hz 700 W 60–180 sec
attachment
(power supply unit)
RP-91NG attachment 320 × 290 × 160 mm 6.5 kg 35 W
Services
Optional Upon request, installation and commissioning
Analyzer can be used in combination with solid, liquid and gas of RA-915M analyzer and attachments can be
matrix analysis attachments: carried out at Сustomer’s site by our service
engineers.
ÿÿ PYRO-915+ attachment for liquid and solid samples
analysis (Thermal decomposition method without Support and training of operators, meeting
sample pre-treatment and thermoscanning option) Customer’s specific needs, are provided
by highly qualified application engineers.
ÿÿ RP-92 attachment for liquid sample analysis (Cold vapor method)
ÿÿ RP-91NG attachment for natural gas analysis
The information and specifications in this publication are subject to change without notice. 19LEN08.02.01-1