Precision measurements: Pass/Fail: Automated testing:
The 1865+ provides resistance Measured results are automatically For automated system applications, the
measurement capability over a range of 1 compared to an operator-programmed 1865+ includes an I/O interface connection
kΩ to 100 TΩ (test voltage dependent), with limit for pass/fail testing. Pass/Fail indicator with remote start and pass/fail outputs. An
a basic accuracy of 0.5%. To meet the test lights or a pass/fail output from the optional IEEE-488 interface is also available
requirements for a wide range of devices, instrument's I/O interface give a clear (or can be added later), which enables
the test voltage is fully programmable up to indication of the measurement results. the 1865 to be remotely controlled by a
1000 Vdc. Sensing the proper resistance computer.
Program data storage:
measurement range is done automatically,
Test setup conditions and measured Display mode (Resistance or Current):
eliminating setup errors. The operator can
results can be stored in CSV format for The operator may select display mode
initiate an automatic test leads zeroing
easy use with Microsoft Excel. Storage to read either measured resistance or
routine in order to eliminate lead or fixture
options include: measured current.
errors.
• 512 MB of internal instrument
Safety features
Easy to use: memory,
For protection of the operator, the 1865+
The 1865’s menu programming system, • A computer (via the IEEE interface)
provides safety features such as current
simple controls and indicators combine • A flash drive (via the USB port)
limiting, a safety interlock, and a warning
for efficiency of test and productivity. Its New programs can also be uploaded via
indicator when high voltage is active.
multi-function keypad provides the operator the IEEE interface or the USB port.
with an easy way to program and make
measurements.
OPTIONAL ACCESSORIES
IEEE Interface (7000-23): Rack Mount Kit (1865-50): shielded case reduces electrical noise and
IEEE to RS-232 Adapter This kit allows the 1865 unit to be installed interference and includes a cover interlock
in a standard 19" rack mount fixture. switch and remote start for maximum
Rear Panel Input (1865-03): operator safety.
Rear input is available as an alternative Shielded Lead Set (1865-51):
to front input. This option places This set is available to reduce external noise
the banana jacks on the rear panel. during the calibration.
Component Test Fixture (1865-52):
An accessory fixture which accommodates
a variety of component types, including
radial, axial, and chip components. Its
SPECIFICATIONS
Resistance range: Output voltage impedance:
1 kΩ (103 Ω) to >100 TΩ (1014 Ω) 1 kΩ, ±5%
No of ranges: Test cycle:
7 manually settable plus auto-ranging Manual: Charge, Measure, Discharge
Automatic: Charge time: 0 - 300 seconds
Resistance range for set voltage
Dwell time: 0 - 300 seconds
Voltage
Rmin Rmax 1 Measure time: 0 - 999 seconds
Setting Discharge time: 0 - 300 seconds
1 Vdc 1 kΩ (103 Ω) 100 GΩ (1011 Ω)
Measurement limits:
10 Vdc 10 kΩ (104 Ω) 1 TΩ (1012 Ω) Pass/Fail (1 limit)
100 Vdc 100 kΩ (10 Ω)
5
10 TΩ (10 Ω)
13
Interfaces:
1000 Vdc 1 MΩ (106 Ω) >100 TΩ (1014 Ω) Standard: RS-232, I/O port with safety interlock
1
USB Host Port for Data/Program storage
Rmax is stated for > 25% accuracy and resistive load only.
Optional: IEEE-488
Capacitance and/or noise will reduce useful resistance range.
Input terminals:
Voltage range: Four sheathed banana jacks, front-mounted (with optional
1 to 1,000 volts, programmable in two ranges
rear-mounting)
Voltage accuracy at front panel connectors: + unknown (red)
1 - 100 V: ±[(1% of setting + 1 V) - unknown (black)
Resolution: 25 mV guard (blue)
100 - 1,000 V: ±[(1% of setting + 2 V) ground (green)
Resolution: 250 mV
Additional features:
Resistance accuracy: • Fully programmable via on-screen menu
±[0.45% +{(RX/VX)(0.0005FS + 2pA) + 30Ω/RX}100%] • Internal zeroing
where: • 512 MB of internal memory for storing test conditions and results
RX = Measured resistance in ohms • Measurement averaging (1-400 readings)
VX = Programmed voltage in volts • Stop on pass
FS = Full scale current range in amperes • Safety interlock
• Keypad lockout
Measuring current:
0.1 pA (10-13 A) to 1 mA (10-3 A) Power:
• 90 - 250 V
Current accuracy at front panel connectors: • 47 - 63 Hz
1 nA to 1 mA ±[0.5% + (0.0005 FS + 2 pA)]
• 40 W max
100 pA to 1 nA ±[1% + (0.0005 FS + 2 pA)]
1 pA to 100 pA ±[10% + (0.0005 FS + 2 pA)] Display:
LCD graphic display
Short-circuit current: High-Voltage warning indicator
<2 mA
Pass/Fail indicator
Input impedance: Environmental conditions:
5 kΩ, ±5%
Operating temperature: 0ºC to 50ºC, <45% RH
ORDERING INFORMATION