Sei sulla pagina 1di 1

ILF Consulting Engineer

Al Majan House, SHEET NO.: 2 OF 7


CONDUCTIVITY ANALYSER
1st. floor, Al Khuwair SPEC NO.: A4-C925-ID-02-13
OMAN JOB NO.: 925
CLIENT: SEPTECH MUSCAT LLC SL.NO BY CHKD APPRD DATE REV
CONSULTANT : ILF Consulting Engineer - OMAN 1 RS KRR CS 24.11.09 0
PROJECT: 5 MIGD MOBILE SEA WATER DESALINATION PL 2 SS KRR CS 21.12.09 A

1 Tag No 925-AIT-45 925-AIT-46


2 Service Filtered Sea Water A Filtered Sea Water A
GENERAL

3 Line No. 300-PVCU-FW-10 300-PVCU-FW-15


4 Pressure Bar G (Operating) 1.5 1.5
5 Viscosity Cp 1 1
6 Temperature OC (Operating) 35 35
7 Type Electrolytic Conductivity Electrolytic Conductivity
8 Mounting Type In-line In-line
9 Process Connection / Size 25 mm Style Fitting 25 mm Style Fitting
10 Insertion Length 100 mm 100 mm
11 Material SS316 SS316
SENSOR

12 Response Time 90% in <45 Sec 90% in <45 Sec


13 Sensitivity * *
14 Range 0 TO 100000 µS/cm A 0 TO 100000 µS/cm A
15 Sensor Cable length 6 Meters 6 Meters
16 Power Supply From the Transmitter From the Transmitter
17 Temperature Compensation Pt1000, Built-in Pt1000, Built-in
18 Type Electronic Electronic
19 Accuracy ±1% OF FS ±1% OF FS
Anodized Aluminium, Powder Coat Anodized Aluminium, Powder Coat
20 Housing
Polyurethane Polyurethane
TRANSMITTER

21 Mounting 2" Pipe 2" Pipe


22 Power Supply 24 VDC, 2 Wire 24 VDC, 2 Wire
23 Transmitter Output Isolated 4 - 20mA DC Isolated 4 - 20mA DC
24 Local Indication Dual line LCD Dual line LCD
25 Enclosure Class WP To IP-65 WP To IP-65
26 Area Classification Safe Safe

27 Cable Entries 1/2" NPT (F ) Sensor & Output 1/2" NPT (F ) Sensor & Output

28 Mfr. * *
29 Sensor model no. * *
30 Transmitter Model no. * *
* Vendor to Specify
Notes :

1 Automatic Calibration and Cleaning system required.


2 The Analyzer shall have tag number plates(SS plate) firmly attached at a visible place.