Sei sulla pagina 1di 165

System Installation Guide

for LABORIE UDS Systems with


Windows 7® OS

LABORIE
LABORIE OFFICES:
USA CANADA
400 Avenue D, Suite 10 2101 Boul. Lapinière
Williston, Vermont, USA 05495 Brossard, Québec, Canada J4W 1L7
Sales/Marketing: 802.857.1300 Montreal: 450.671.5901
Customer Service: 800.522.6743 Customer Service: 888.522.6743
Technical Support: 800.333.1039 Technical Support: 800.333.1039
Fax: 802.878.1122 Fax: 450.671.7182
service@laborie.com service@laborie.com

EUROPE INTERNATIONAL
Lumonics House, Valley Drive 6415 Northwest Dr., Unit 10
Rugby, Warwickshire CV21 1TQ UK Mississauga, Ontario, Canada L4V 1X1
Tel.: +44.01788.547888 Customer Service: 905.612.1170
Fax: +44.01788.536434 Fax: 905.612.9731
internationalsupport@laborie.com internationalsupport@laborie.com

www.laborie.com

UDS120, GOBY, Urostym, EVOX, AquaVu, and i-LIST are trademarks of LABORIE Medical Technologies Canada
ULC. Windows, Microsoft Office, the Start Button, and Internet Explorer are registered trademarks of Microsoft Corp.
Bluetooth is a registered trademark of Bluetooth SIG, Inc. HP is a registered trademark of the Hewlett Packard
Company. Adobe Acrobat and Adobe Reader are registered trademarks of Adobe Systems Incorporated in the U.S.
and other countries. Norton Antivirus is a trademark of Symantec Corporation or its affiliates in the U.S. and other
countries. PIXMA is a trademark of Canon Inc. SeeMore is a trademark of Interson Corp. TECRA is a trademark of
Toshiba Corporation. All other names are trademarks of their respective owners.

© Copyright 2015 by LABORIE.

All rights reserved. No part of this publication may be reproduced in any form whatsoever without the prior written
permission of LABORIE.

Document Number: UDS120-UM05


Version Number: 19
Release Date: January 30, 2015
Issued By: R.A.
REVISION HISTORY

Manual Date Note


Version
19 January 30, 2015  Updated Toshiba laptop requirements and removed DELL E6530 laptop from
minimum requirements (see page 2)
 Added HP ENVY 4500 all-in-one printer driver installation instructions (see
page 65)
 Added important notes to read before installation of GridConnect USB Serial
Adapter (see page 71)
 Updated configuration diagrams for Urostym installation (see page 110)
 Updated patch installation for i-LIST (see page 137)

18 October 23, 2014 Updated installation for recommended Nitro PDF Pro V10
17 May 5, 2014 Updated display resolution settings for PC configuration
16 March 14, 2014 Updated minimum requirements for installation; Updated PC configuration instructions;
Added configuration instructions for recommended Toshiba Tecra Z50 laptop; Added Grid
Connect USB serial adapter installation instructions; Updated EVOX installation section;
Updated i-LIST Office Reporter section to include patch installation instructions; Removed
COM Port Standardization from Appendix
15 December 5, 2013 Updated i-LIST Installation instructions
14 August 20, 2013 Updated StarTech USB to Serial installation instructions to add PER381; Updated laptop
connection diagram for Urostym installation; Added new COM port standardization
information
13 August 12, 2013 Removed previous USB to Serial adapter information and replaced with new StarTech USB
to Serial installation instructions; Added note for StarTech serial adapter use with
installation of UDS120 software with a wired connection; Updated Urostym installation
instructions to reference new StarTech USB to Serial adapter; Removed COM port
standardization information
12 May 17, 2013 Removed configuration for Toshiba Tecra A11 laptop and DELL T3500 PC models; Updated
minimum requirements; Updated configuration information for DELL E6530 laptops;
Updated configuration information for DELL Precision T3600 PC; Updated PDF installation
instructions; Updated Antivirus configuration information; Updated UDS120 software
installation instructions; Updated EVOX installation instructions; Updated iLIST Office
Reporter installation instructions
11 January 15, 2013 Updated minimum requirements; Added configuration information for DELL E6530 laptops;
Added configuration information for DELL Precision T3600 PC; Added Antivirus
configuration information for DELL E6530 laptop and DELL Precision T3600 PC; Added HP
OfficeJet 6100 Printer information; Updated COM port settings
10 June 1, 2012 Added installation information to enable use of Urostym system with Urocap IV standalone
(see page 150)
9.0 March 7, 2012 Added compatibility of UDS Goby software with HASP key driver installation; Added
reminder to plug the HASP key into all USB ports of PC once Goby software installed; added
HASP key pop-up message that appears with a new install; also updated Device Manager
screenshots; Added compatibility of UDS Goby software with current version of i-LIST Office
reporter software
8.0 January 13, 2012 Updated minimum requirements for installation; Updated service pack information and
Windows OS updates setup; Added step for increasing virtual memory (for desktop PCs
only); Added note at end of each type of UDS120 installation for UDS-94 hardware with a
V1.05 board set; Added software patch installation instructions for each type of
installation; Added note before installation of SeeMore™ ultrasound probe driver
7.0 December 6, 2011 Updated minimum requirements for installation; Added ATI FirePro Video Driver
installation instructions; Added HASP Key Driver installation instructions; Added installation
instructions for NuWav ultrasound probe with UDS120 software(optional feature); Updated
UDS120 installation instructions and removed patch installation instructions; Added TETRA
Manual Date Note
Version
installation note; Added Urocap IV standalone installation; Added EVOX installation
instructions; Updated installation instructions for iLIST Reporter V1.2.2.9; Added SeeMore™
ultrasound probe driver installation instructions; Added VISER installation instructions;
Added Turn off automatic updates step in System Finalization section; Added COM Port
Standardization
6.0 September 13, 2011 Added printer driver installation instructions for the Canon PIXMA™ iP100 series wireless
printer
5.0 August 16, 2011 Added software patch installation instructions for UDS120 V12.0.3 software
4.0 April 1, 2011 Added AquaVu software installation instructions
3.0 February 11, 2011 Updated Minimum Specifications; Added configuration information for laptop and PC
models recommended by LABORIE; Added step to Bluetooth configuration; Updated
Urostym connection diagrams
2.0 December 10, 2010 Updated Minimum Specifications; Added placement of Bluetooth Devices function;
Updated for UDS120 V12 Updated UDS120 installation instructions for version 12; Added troubleshooting
release information
1.0 October 1, 2010 Creation of first version of system configuration guide.
ABOUT THIS INSTALLATION GUIDE
This guide, divided in four parts, describes how to set up and install Laborie software and/or
accessories on your computer with the Windows® 7 OS.

IMPORTANT NOTES BEFORE START OF INSTALLATION


 NOTE 1: Plug only the power supply into the laptop or PC. Do not connect any
cables or devices that would enable an internet connection until instructed to do
so during installation.
 NOTE 2: It is recommended to have Windows 7 OS 64-bit version installed.

INSTALLATION PROCESS

System Setup
Part 1

Starting on page 1

Driver and Peripherals


Part 2 Installation

Starting on page 45

Laborie Software
Part 3 Installation

Starting on page 89

Part 4 System
Finalization

Starting on page 154

Follow the sections in this guide in the order as outlined for proper installation. If you
encounter any problems during installation, please contact LABORIE Technical Support at
1-800-333-1039. For troubleshooting information see page 157.
IMPORTANT NOTE! Installation and configuration must first
be performed under Administrator mode. The creation of
additional administrator and/or user groups will require a
repeat of the settings outlined in this guide.
LABORIE System Installation Guide for Windows 7 OS

Part 1: System Setup - required for all installations


This part outlines how to set up the computer before installing any drivers and/or Laborie software.
This part must be followed point-by-point. Once the sections in this part of the guide are
complete, you can then proceed to the installation of drivers and/or the Laborie software.

 Minimum Requirements for Software Installation (see page 2)

 Configure the Embedtek PC (see page 3)

 Configure the Toshiba TECRA™ Z50 Laptop (see page 14)

 Install any other software such as MS Office, Adobe Acrobat/Reader, or NitroPDF®


Professional (see page 32)

Page 1
LABORIE System Installation Guide for Windows 7 OS

Minimum Requirements for Software Installation

 Operating System: Microsoft® Windows® 7 64-bit Service Pack 1


 AntiVirus software (Microsoft® Security Essentials recommended)

EmbedTek Industrial PC (Base System):


 Intel Q67 Chipset-based platform
 Intel Core i3 3220S processor (55W) (embedded)
o 3.3 GHz, 2 cores, 4 threads, 3MB cache
 4 GB DDR-3 Dual Channel System Memory
o Upgradeable up to 32GB
 Dual GbE
 Wireless Networking (via USB)
 500GB 3.5” Hard Drive
 Dimensions: 3”H x 11.5”W x 11.5”D (black)
 Internal CD/DVD RW
 I/O
o 7.1 CH Audio, 1 x RS-232 serial port, 6 x USB 2.0, 1 x VGA port, 2 x HDMI 1.4 ports
 Windows 7

EmbedTek Industrial PC (High End):


Same Configuration as Base System plus:
 I7 3770T (45W) (Embedded)
o 2.5 GHz (3.7 GHz Max) , 4 cores, 8 threads, 8MB cache
 RAID 1 (2x 2.5” 500GB HD’s in place of 3.5”HD)
 8 GB RAM

Toshiba Tecra Z50 for all systems with laptop:


 Intel® Core™ i5-4210U Processor, (1.7GHz/2.7Ghz)
 Windows 7, 64-bit
 RAM: 4GB or higher
 Hard disk: 320GB or higher
 15.6" display
 1 built-in COM port
 USB 2.0 ports: Minimum 4
 Bluetooth 4.0 (867Mbps)

Page 2
LABORIE System Installation Guide for Windows 7 OS

Configure the Embedtek PC

Disable User Access Control (UAC)


 Click Start > Control Panel.
 In the Control Panel, select Large icons in the upper-right of the window.

 Scroll down the window and click User Accounts.


 Click the Change User Account Control settings option and set it to the lowest level possible of Never
notify.

 Click OK and then exit the Control Panel.

Set up MSMQ and .NET Framework


1. Click Start > Control Panel > Program and Features.
2. Click on ‘Turn Windows features on or off’ to open the Windows Features window.
3. Under this window, select all components under ‘Microsoft .NET Framework 3.x’. Under
‘Microsoft Message Queue (MSMQ) Server’ select only ‘MSMQ Triggers’ as shown in the
following image:

Page 3
LABORIE System Installation Guide for Windows 7 OS

4. Select “Internet Information Services”, under “Internet Information Services” click on “IIS 6
Management Console” and then click on “IIS Metabase and IIS 6 configuration compatibility” as
shown in the next screenshot.

5. Register ASP.NET on IIS by following steps.


a. Open command window, and type:

For 32-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727”

For 64-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727” for 64 bit OS,
and press Enter key.

Page 4
LABORIE System Installation Guide for Windows 7 OS

b. Type the command “aspnet_regiis -i” and press Enter key, and wait.

6. Enable 32-Bit Applications from IIS Manager for 64 bit Operating System.
a. Click Start icon and enter “iis” in ‘Search programs and files’, and select Internet Information
Services (IIS) Manger from the list as shown next screenshot.

Page 5
LABORIE System Installation Guide for Windows 7 OS

b. The Internet Information Services (IIS) Manger window will be opened as shown below.

c. Open Advanced Settings window by select Application Pools -> DefaultAppPool, and select
‘Advanced Settings…’ on the right pane. Select True on the Enable 32-Bit Applications option.

 Click OK and continue with the next step.

Page 6
LABORIE System Installation Guide for Windows 7 OS

Set up Driver Update Properties


 Click Start > Control Panel > System and click the Advanced system settings heading on the left
side of the window.

 In the resulting System Properties window click the Hardware tab and then click the Device
Installation Settings button.

Page 7
LABORIE System Installation Guide for Windows 7 OS

 Copy the selections as shown here:

 Click the Save Changes button and continue with the next step.

NOTE: Plug in the network cable to connect to the internet as it is now required for the
following steps.

Regional and Language Options Setup Procedure


See instructions on page 8.

Verify the Display Resolution


For single monitor setup:
 Click Start > Control Panel > Display.
 Verify the resolution of the monitor is set to 1440 x 900.
For dual monitor setup:
 Plug in the primary monitor first; then plug in the secondary monitor.
 Continue with the next step.

Page 8
LABORIE System Installation Guide for Windows 7 OS

Configure the Universal Serial Bus Controller Settings


Click Start > Control Panel. From the Control Panel window click Device Manager.
From the list locate and double-click on the entry for ‘Universal Serial Bus Controllers’. Double-click the
first listing of ‘USB Root Hub’ to open the USB Root Hub Properties window. In the Power Management
tab, confirm that no check marks are present beside the ‘Allow the computer to turn off this device to
save power’ setting. Repeat for all ‘USB Root Hub’ entries listed. Click OK.
Double-click the first listing of ‘Generic USB Hub’ (if visible) to open the Generic USB Hub Properties
window. In the Power Management tab, confirm that no check marks are present beside the ‘Allow the
computer to turn off this device to save power’ setting. Repeat for all ‘Generic USB Hub’ entries listed.
Click OK.
Click Apply and then click OK. Close all open windows and continue with the next step.

Configure Power Options


 Click on Start > Control Panel and click Power Options.
 In the Select a power plan window, click on Choose what the power button does located on the top-
left side of the window.
 Copy the settings illustrated in the following image:

 Click the Save changes button.


 In the Select a power plan window, select the Dell option if not already selected.
 Click the Change plan settings heading to open the Change Settings window.
 Copy the settings as shown in the following image:

Page 9
LABORIE System Installation Guide for Windows 7 OS

 Click the Change advanced power settings heading and make adjustments to each item under
the Advanced settings tab.
 Copy the settings for each item as shown in the following images:

Dell settings Hard disk settings

Page 10
LABORIE System Installation Guide for Windows 7 OS

Desktop background settings settings Wireless Adapter Settings settings

Sleep settings USB settings settings

Page 11
LABORIE System Installation Guide for Windows 7 OS

Power buttons and lid settings PCI Express settings

Processor power management settings Display settings

Page 12
LABORIE System Installation Guide for Windows 7 OS

Multimedia settings settings

 Click Apply, then click OK, and then click the Save changes button. Close all open windows.
 Continue with the installation of any other software (see page 32).

Remove SQL Server 2005 Application


Go to Control Panel > Program and Features and verify that the Microsoft SQL Server 2005
Compact Edition [ENU] application is not installed. If it is, un-install the program and then
restart the computer.

Continue by installing any other third-party software such as Adobe Reader, PDF creators.
etc… then go to Part 2.

Page 13
LABORIE System Installation Guide for Windows 7 OS

Configure the Toshiba TECRA™ Z50 Laptop

Disable User Access Control


 Click Start > Control Panel.
 In the Control Panel, select Large icons in the upper-right of the window.

 Scroll down the window and click User Accounts.


 Click the Change User Account Control settings option and set it to the lowest level possible of Never
notify.

 Click OK and then exit the Control Panel.

Page 14
LABORIE System Installation Guide for Windows 7 OS

Set up MSMQ and .NET Framework


1. Click Start > Control Panel > Program and Features.
2. Click on ‘Turn Windows features on or off’ to open the Windows Features window.
3. Under this window, select all components under ‘Microsoft .NET Framework 3.x’. Under
‘Microsoft Message Queue (MSMQ) Server’ select only ‘MSMQ Triggers’ as shown in the
following image:

4. Scroll up the list and then select and expand “Internet Information Services”. Under “Web
Management Tools” add a check mark to the “IIS 6 Management Console” option. Then select and
expand ‘IIS 6 Management Compatibility’ and add a check mark to the “IIS Metabase and IIS 6
configuration compatibility” option as shown in the next screenshot.

Page 15
LABORIE System Installation Guide for Windows 7 OS

5. Register ASP.NET on IIS with the following steps:


a. Open command window, and type:
For 32-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727”

For 64-BIT OS
“cd C:\Windows\Microsoft.NET\Framework64\v2.0.50727” for 64 bit OS,
and press Enter key.

b. Type the command “aspnet_regiis -i” and press Enter key, and wait.

6. Enable 32-Bit Applications from IIS Manager for 64 bit Operating System.
a. Click Start icon and enter “iis” in ‘Search programs and files’, and select Internet Information
Services (IIS) Manger from the list as shown next screenshot.

Page 16
LABORIE System Installation Guide for Windows 7 OS

b. The Internet Information Services (IIS) Manger window will be opened as shown below.

c. Open Advanced Settings window by select Application Pools -> DefaultAppPool, and select
‘Advanced Settings…’ on the right pane. Select True on the Enable 32-Bit Applications option.

Page 17
LABORIE System Installation Guide for Windows 7 OS

 Click OK and continue with the next step.

Set Up Driver Update Properties


 Click Start > Control Panel > System and click the Advanced system settings heading on the left
side of the window.

 In the resulting System Properties window click the Hardware tab and then click the Device
Installation Settings button.

Page 18
LABORIE System Installation Guide for Windows 7 OS

 Copy the selections as shown below:

 Click the Save Changes button and continue with the next step.

 Go to Start > Control Panel > Windows Update and click on Change settings.

Page 19
LABORIE System Installation Guide for Windows 7 OS

 Select Never check for updates under Important updates, and select Give me recommended
updates the same way I receive important updates under Recommended updates.

NOTE: Plug in the network cable to connect to the internet as it is now required for the following
steps.

Set up Regional and Language Options


NOTE: If you are not sure of the settings to apply for the Regional and Language Options, select the default settings
for the United States.
 Click Start > Control Panel and double-click the Regional and Language Options selection.
 In the Regional and Language Options window, click the Formats tab.
 Select the most appropriate Format setting. For example: If you are in the United States, select
the English (United States) format.
 Click Apply.
 In the Regional and Language Options window, click the Location tab.
 Select the most appropriate Location setting. For example: If you are in the United States, select
the United States location.
 Click Apply.
 In the Regional and Language Options window, click the Keyboards and Languages tab.
 Click the Change Keyboards button to open the Text Service and Input Languages window.
 In the Text Service and Input Languages window, click the Add button to open the Add Input
Language window.
 Double-click the most appropriate Language setting from the list.
 Double-click Keyboard.

Page 20
LABORIE System Installation Guide for Windows 7 OS

 Use the checkboxes to select the preferred keyboard settings.


 Click OK.
 Under the General tab in the Text Services and Input Languages window, select the Default
input language. If more than two selections appear in the drop-down box, select the preferred
choice; if only one entry is visible in the drop-down box then no change will be needed.
 Click Apply and then click OK to close the Text Service and Input Languages window.
For example: If you are in the United States, select the English (United States) language
in the Add Input Language window. Double-click English (United States) to view
Keyboard. Double -click Keyboard to view all the associated keyboard options. Use the
checkbox to select US. Click OK in the Add Input Language window. Make sure that
English (United States) – US appears in the Default input language section. Click Apply
and then OK to close the Text Service and Input Languages window.
 Click Apply and then OK to close the Regional and Language Options window.
 Continue with the next section.

Set up Mobility Options


 Click Windows Mobility Center in the control panel window.
 Under Battery Status, select High performance as shown.
 Under Wireless Network, select to Turn wireless off, to disable the wireless function.
 Keep all other settings as shown.

 Close this window and continue with the next step.

Install Bluetooth Monitor


 Click Start > Control Panel > Programs and Features and scroll down to locate in the
resulting list the Bluetooth Monitor 4 entry.
 If ‘Bluetooth Stack for Windows by Toshiba’ is listed instead and not ‘Bluetooth Monitor
4’, ONLY then you must un-install ‘Bluetooth Stack for Windows by Toshiba’ from the
Programs and Features list, and install the TC00091500E_Bluetooth_Monitor_v4.04.exe
application found in the Drivers & Utilities > Toshiba Tecra Z50 > Drivers > Windows 7
folder on the Laborie Released Software Disc.
 Continue with the next step.

Page 21
LABORIE System Installation Guide for Windows 7 OS

Place Bluetooth Devices Shortcut on Desktop


 Click Start > Control Panel.
 In the Control Panel window, type the word Bluetooth into the Search Control Panel field at the
top-right corner.

 As Bluetooth is typed into the search field, the available settings with automatically refresh. Click
the Change Bluetooth settings heading under Devices and Printers.

 In the resulting Bluetooth Settings window, make sure the Show the Bluetooth icon in the
notification area option is selected. This will place the Bluetooth icon shortcut in the task tray.

 Click OK and continue with the next step.

Page 22
LABORIE System Installation Guide for Windows 7 OS

Configure the Network, Universal Serial Bus Controller Power, and Bluetooth
Settings
Click Start > Control Panel. From the Control Panel window click Device Manager.

Network Settings:
Scroll down the list and double-click the ‘Network Adapters’ entry. Double-click the first hardware item
listed under Network Adapters [example: Intel® 82577LM Gigabit Network Connection] to open the
Properties window. If the Properties window contains Power Management tab then select it and
confirm that no check marks are present beside the settings ‘Allow this device to bring the computer out
of standby’, ‘Allow the computer to turn off this device to save power, ‘Allow this device to wake the
computer’, or ‘Only allow a magic packet to wake the computer’. Close the Properties window.

 Click OK.
 Repeat for every hardware item listed under the Network Adapters heading.

Page 23
LABORIE System Installation Guide for Windows 7 OS

Page 24
LABORIE System Installation Guide for Windows 7 OS

Page 25
LABORIE System Installation Guide for Windows 7 OS

Universal Serial Bus (USB) Controllers Settings:


 From the list, locate and double-click on the entry for ‘Universal Serial Bus Controllers’. Double-
click the first listing of ‘USB Root Hub’ to open the USB Root Hub Properties window. In the Power
Management tab, confirm that no check marks are present beside the ‘Allow the computer to turn
off this device to save power’ setting. Repeat for all ‘USB Root Hub’ entries listed. Click OK.
 Double-click the first listing of ‘Generic USB Hub’ to open the Generic USB Hub Properties window.
In the Power Management tab, confirm that no check marks are present beside the ‘Allow the
computer to turn off this device to save power’ setting. Repeat for all ‘Generic USB Hub’ entries
listed. Click OK.

Bluetooth Settings:
 Under the Bluetooth Radios, double-click each item to open its Properties window.
 If the Properties window contains Power Management tab then select it and clear the check
mark beside the ‘Allow the computer to turn off this device to save power’ or the ‘Allow this
device to save power’ options.

 Continue with next step.

Page 26
LABORIE System Installation Guide for Windows 7 OS

Configure Power Options


 Click on Start > Control Panel.
 Inside the Control Panel window click Power Options.
 Under Select a power plan, select the High Performance option.

 Click the Change plan settings option on the right to access the Edit Plan Settings window.
Copy the settings as illustrated in the following image.

 Click the Change advanced power settings heading.


 Double-click on the Power buttons and lid heading to open the list.
 Confirm the setting for Power button action and Start menu power button are both set to
Shut down and for Lid close action set and Sleep button action to Do Nothing. As
illustrated in the following images.

Page 27
LABORIE System Installation Guide for Windows 7 OS

 Double-click on Sleep and confirm that the settings for Sleep are set to Sleep after: Never;
Allow hybrid sleep: Off; Hibernate after: Never; Allow wake timers: Disable.

 Double-click on Hard Disk and


confirm that the settings under
Turn off hard disk after are set
to Never. If not then please
type in ‘Never’ into the
respective field.

Page 28
LABORIE System Installation Guide for Windows 7 OS

 Double click on Wireless Adapter Settings, and for both On battery & Plugged in, select
‘Maximum Performance’

 Double click on Display, and select the options as chosen in the following screens:

 Click Apply and then click OK.


 Click the Save changes button.
 After clicking the Save changes button, you will be brought back to the Power Options
window. In the Power Options window, click on Choose what the power buttons does
located on the top-left side of the window.
 Copy the settings illustrated in the following image.

Page 29
LABORIE System Installation Guide for Windows 7 OS

 Select the Don’t require a password option.


 Click the Save changes button and exit the Power Options window.
 Continue with the next step.

Configure Final Items


 Click Start > All Programs > TOSHIBA > Utilities > USB Sleep and Charge.
 Under the TOSHIBA USB Sleep and Charge Information window, place a check mark on the ‘In
the future, do not display this window’ option and then click ‘X’ in the upper-right corner.

 Unpin the media player and My TOSHIBA from the taskbar by right-clicking on each icon and
selecting the Unpin option.

Page 30
LABORIE System Installation Guide for Windows 7 OS

 Choose to ‘Close gadget’ for Yellow Pages, Norton, or any other gadgets that appear on the
desktop:

 Physically, and carefully, move the laptop along the counter or surface it is placed on. In the
resulting pop-up TOSHIBA HDD Protection window, place a check mark next to the ‘Do not
display this message in the future.’ option and click OK.

 Change the icon size by right-clicking on the desktop and select View > Small icons

Page 31
LABORIE System Installation Guide for Windows 7 OS

Install any other software such as MS Office, Adobe Acrobat/Reader, or


NitroPDF® Professional
If additional software is required for installation such as Adobe Acrobat™, Adobe Reader™, and/or
Microsoft Office® then read the installation guide provided with the product for instructions on how to
install the software.

To install NitroPDF® Professional, see NOTE 2.

NOTE 1: To finalize Microsoft Office installation, follow these steps:

Before starting installation make sure the computer is connected to the internet.
1. If the Microsoft Office Home and Business 2013 application is on a Disc, then once the disc is
inserted the following message must appear.
2. Click Run setup.EXE.

3. In the following window, enter the supplied product key.

4. Once the product key has been entered and verified to be correct, a tick mark will appear as
shown in the following screenshot. Click Continue.

Page 32
LABORIE System Installation Guide for Windows 7 OS

5. Select Redeem Online.

6. Under the following window, enter the account name and password. Once the password is
entered, click the Sign in button.

Note:
 If the following message window appears at the bottom of the screen, choose ‘No’ to disable
the AutoComplete.

 If the following message window appears at the bottom of the screen, choose ‘Ask me later’.

 If the following message window appears at the bottom of the screen, choose ‘Enable’

7. Under the following window enter the Country and the preferable Language. Click Continue
after making the required selections.

Page 33
LABORIE System Installation Guide for Windows 7 OS

8. Minimize the window that appears.

9. Click Finish Activation.

Page 34
LABORIE System Installation Guide for Windows 7 OS

10. Under the following window, enter a valid and active e-mail address and click Sign In.

11. Under the following window, enter the password used to access the e-mail address just entered.

Page 35
LABORIE System Installation Guide for Windows 7 OS

12. Under the following window, select the option Microsoft Office Home and Business 2013 (2).
Click Continue.

13. Click Next.

14. Select No thanks and click Accept.

Page 36
LABORIE System Installation Guide for Windows 7 OS

15. Click the ‘X’ in the top-right corner to close this window or wait for the video to end in about 1
minute.

16. Click Next.

Page 37
LABORIE System Installation Guide for Windows 7 OS

17. Select No Background from the drop down box and click Next.

18. Click No thanks.

19. Click All done.

Page 38
LABORIE System Installation Guide for Windows 7 OS

20. In the following window, under the ‘Account’ tab, click ‘Sign out’. If the following window does
not appear, click Start >All Programs > Microsoft Office 2013 > Word 2013 to start the Word
application. Then select ‘Blank Document’. Then click FILE and go into the ‘Account’ tab as
shown below and click ‘Sign out’.

21. Click ‘Yes’.

Page 39
LABORIE System Installation Guide for Windows 7 OS

22. Click Start > All Programs > Microsoft Office 2013 > Word 2013 to start the Word application.
When the Word application opens, confirm ‘Sign in to get the most out of office’ appears on the
top right corner of the Word application window, and not the credentials of the e-mail address
entered earlier in the installation process.

23. Select ‘Blank Document’.


24. From the Word application main menu, click FILE and then go into the ‘Account’ tab. Confirm
‘Sign in’ appears on the top right corner.
25. Click ‘Update Options’, and from the drop down choices, choose ‘Disable Updates’

MS Office installation complete.

Page 40
LABORIE System Installation Guide for Windows 7 OS

NOTE 2: If you are installing NitroPDF® Professional, follow these instructions:


For troubleshooting during installation see page 158.

Part 1: Uninstall Previous Version of Nitro PDF


If an older version of Nitro PDF (earlier than version 10) is installed, then you will need to
uninstall the previous version of NitroPDF since a new installation will not overwrite any
existing ones.

1. Click Start > Control Panel and click the Programs and Features heading.
2. Scroll down the list of installed programs to locate the currently installed version of Nitro PDF
(earlier than V10).

3. Click on the line with the name of the Nitro version installed and then click Uninstall on the top
toolbar.

Page 41
LABORIE System Installation Guide for Windows 7 OS

4. When asked, click Yes to confirm the removal.

5. Wait for the un-install to occur.

6. Click OK when alert message appears.

Page 42
LABORIE System Installation Guide for Windows 7 OS

7. Wait for the un-install to continue.


8. Continue with the installation of Nitro PDF Version 10.

Part 2: Install Nitro PDF


1. Double-click the nitro_pro10_x64_NLS.msi installation file [found under the Nitro PDF Pro 10 folder on the
Laborie Released Software Disc].

2. Click Run in the resulting Security Warning window and wait for the installation to occur.

IMPORTANT! Do not install any other software applications if prompted to do so.


3. Select the I accept
the terms in the
License Agreement
option and click
Install.

4. Wait for the


installation to occur
and then click Finish
when installation is
complete.

Page 43
LABORIE System Installation Guide for Windows 7 OS

5. Double-click the
NitroPDF
Professional
shortcut icon on
the desktop to start
the software.

6. Click the Activate button and type in the 18-digit license number to register the full program.

NitroPDF Professional installation complete.

Part 1 complete. Continue with parts 2 to 4.

Page 44
LABORIE System Installation Guide for Windows 7 OS

Part 2: Driver and Peripherals Installation


This section tells you how to install drivers. If you have one or all of the following products,
install them first before installing the Laborie software.

 Install the HP OfficeJet 6100 Printer (Optional) (see page 46)

 Install the HP Officejet Mobile 100 Printer (optional) (see page 53)

 Install the HP ENVY 4500e All-in-One Printer Driver (optional) (see page 65)

 Install the Grid Connect ATC 810 (PER410) USB to Serial Adapter Software (see page 71)

 Configure the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURiO](see page 83)

 Install HASP Driver for UDS120, UDS-Goby, NuWav Advanced, Urostym, Urocap IV
Standalone, and/or AquaVu (see page 88)

Page 45
LABORIE System Installation Guide for Windows 7 OS

Install the HP OfficeJet 6100 Printer (Optional)

Verify Printer Driver and Software are already installed


1. Click Start > Control Panel > Programs and Features.
2. In the resulting list scroll down and locate the HP OfficeJet 6100 Basic Device Software and
the HP Officejet 6100 Help software.
i) If it is visible in the list then go directly to Printer Driver Setup on page 51.
ii) If it is not listed then continue with the instructions in Install HP OfficeJet 6100
Driver and Software on page 47.

NOTE: Do not power-on and connect the printer to the computer until requested to do so
during the installation process.

Page 46
LABORIE System Installation Guide for Windows 7 OS

Install HP OfficeJet 6100 Driver and Software


1. Insert the installation disc for the HP OfficeJet 6100 printer into the DVD drive.
2. Click the Run Setup.exe option.

3. Click the Customize Software Selections button.

Page 47
LABORIE System Installation Guide for Windows 7 OS

4. Deselect all options except for the HP Officejet 6100 Help option and then click Next.

5. Select the No, do not enable the Web Services option and then click OK.

Page 48
LABORIE System Installation Guide for Windows 7 OS

6. Deselect all options in the Internet Connection Usage Settings window, if not already done. Then
click OK.

7. Select the I have reviewed and accept the installation agreements and settings option
and then click Next.

Page 49
LABORIE System Installation Guide for Windows 7 OS

8. Select the Connect later option and then click Next.

9. Click Finish.

10. Continue with the instructions in the Printer Driver Setup section on page 51.

Page 50
LABORIE System Installation Guide for Windows 7 OS

Printer Driver Setup


1. Connect one end of the USB cable to the printer and then power on the Printer.

2. Once the printer has initialized, plug the other end of the USB cable to any available USB
port on the computer. The computer will automatically start to install the necessary
drivers. Please do not disturb the computer during this process.

3. Once the driver installation is complete, go into the Device Manager – Universal Serial
Bus controllers, and confirm the presence of ‘Officejet 6100 (REST)’.

4. Click Start > Control Panel > Devices and Printers to open the Printers window.

Page 51
LABORIE System Installation Guide for Windows 7 OS

5. Check if a tick mark is next to the ‘HP Officejet6100’ printer. If not, then right-click
over this printer and select ‘Set as Default Printer’.

6. Confirm if the tick mark transfers onto ‘HP Officejet6100’.

7. Insert the printer USB cable into each USB port on the laptop or PC to initialize the
driver.
8. Print a test page and attach this test page to the Computer QC forms. Plug the cable into
any port to print the page.

The HP Officejet6100 is now available for all printing purposes.

Page 52
LABORIE System Installation Guide for Windows 7 OS

Install the HP Officejet Mobile 100 Printer (optional)

Option 1: Install Printer as a USB-Connected Device


NOTE: For a Bluetooth connection, go to Option 2 on page 56.

1. In the LABORIE installation disc, open the HP Officejet 100 installation program and
click Install USB-Connected Device.

2. Wait for the system check to occur.

Page 53
LABORIE System Installation Guide for Windows 7 OS

3. Click Next.

4. Select the checkbox “I have reviewed and accept the installation agreements and
settings. Then click Next and wait for installation.

Page 54
LABORIE System Installation Guide for Windows 7 OS

5. Connect the USB cable between the computer and the printer.

6. Click Finish.

7. Click Cancel.

Page 55
LABORIE System Installation Guide for Windows 7 OS

8. Click “Don’t remind me again”. The click OK.

9. Click Exit to finalize setup. If necessary print a test page to ensure printer is properly installed.

Option 2: Install Printer as Bluetooth-Connected Device


1. Click Install Bluetooth-Connected Device.

Page 56
LABORIE System Installation Guide for Windows 7 OS

2. Turn on Bluetooth radio on the printer. Then select the “The Bluetooth light is on”
checkbox and click Next.

3. Wait for system check then click Next.

Page 57
LABORIE System Installation Guide for Windows 7 OS

4. Select the checkbox “I have reviewed and accept the installation agreements and
settings. Then click Next and wait for installation to occur.

NOTE: If the printer software was previously installed the printer software, some of the files may need
to be deleted and computer restarted before proceeding. Follow the onscreen instructions if the restart
screen appears. Then continue with the steps below.

Page 58
LABORIE System Installation Guide for Windows 7 OS

5. Make sure Bluetooth radios in the computer and printer are turned on. Click Next.

6. Click Start > Devices and Printers and click Add a printer.

7. Click the Add a network, wireless or Bluetooth printer option.

Page 59
LABORIE System Installation Guide for Windows 7 OS

8. Locate and click on the printer name that starts with OJL411 followed by the S/N of
the printer (e.g. if the S/N of the printer is MY26M821TB, the printer name would
be OJL411MY26M821TB). Then click Next.

9. Click Next.

Page 60
LABORIE System Installation Guide for Windows 7 OS

10. Select “Do not share this printer” and click Next.

11. Make sure “Set as default printer” is selected. Then click “Print a test page”

12. Click Close.

Page 61
LABORIE System Installation Guide for Windows 7 OS

13. Click Finish.

14. Select “I have created a Bluetooth connection between the printer and the computer”. Then click
Next.

Page 62
LABORIE System Installation Guide for Windows 7 OS

15. Click Finish.

16. Click Cancel.

17. Select “Don’t remind me again” and click OK.

Page 63
LABORIE System Installation Guide for Windows 7 OS

18. Click Exit.

HP Officejet 100 is ready for use.

Page 64
LABORIE System Installation Guide for Windows 7 OS

Install the HP ENVY 4500e All-in-One Printer Driver (optional)

Part 1: Install Driver


1. In the LABORIE software disc, double-click the installation package in the folder named HP ENVY
4500e All-in-One Printer series.
2. Click Continue when the following window appears.

3. Click the Web Services Enablement heading.

Page 65
LABORIE System Installation Guide for Windows 7 OS

4. Select the No, do not enable the Web Services option and click OK.

5. Click the Internet Connection Usage Settings heading.

Page 66
LABORIE System Installation Guide for Windows 7 OS

6. Remove all check marks from the settings and click OK.

7. Select the I have reviewed and accept the installation agreement and settings option and click
Next.

Page 67
LABORIE System Installation Guide for Windows 7 OS

8. Select the Connect later - Do not connect to the printer now. I will connect it later. option and
click Next.

9. Click Finish to complete the printer installation over the computer

Page 68
LABORIE System Installation Guide for Windows 7 OS

Part 2: Set Printer as Default Printer and Print a Test Page


1. Turn on the printer and connect it to the computer through the USB cable.
2. Under the notification area on the desktop of the PC a bubble message should appear. Click on
this message. Under the message there must be a check mark next to all the installed drivers as
follows:

3. Go into the Device Manager of the computer and check that the following entries are found
under each section:
 Imaging devices - HP ENVY 4500 series (USB)
 Universal Serial Bus controllers - HP ENVY 4500 series (REST), USB Composite Device,
USB Printing Support.
4. Click Start > Control Panel > Devices and Printers.
5. Confirm that HP ENVY 4500 series appears with a check mark next to it as shown in the
following screenshot:

6. If a check mark does not appear as shown above, right-click over the HP ENVY 4500 series icon
and choose to Set as default printer.
7. Double-click the HP ENVY 4500 series icon.
8. Double-click the HP Printer Assistant option to open the HP Printer Assistant window.

Page 69
LABORIE System Installation Guide for Windows 7 OS

9. Double-click the Maintain Your Printer option to open the HP ENVY 4500 series Toolbox
window.

10. Under the Device Services tab, click the icon next to the Print Quality Diagnostics Page option.

11. Click Close when page is printed. Setup complete.

Page 70
LABORIE System Installation Guide for Windows 7 OS

Install the Grid Connect ATC 810 (PER410) USB to Serial Adapter Software
IMPORTANT NOTES! READ BEFORE INSTALLATION

 When installing a second/additional unit of the GridConnect USB Serial Convertor, do not plug
the second unit into the same USB port of the computer used when installing the first unit.
Instead, choose a different USB port to plug and install the second unit.

 Never directly unplug the GridConnect USB Serial Convertor from the USB port of the computer.
Always go first into the notification area and right-click the Safely Remove Hardware and Eject
Media option, choose Eject USB High Speed |RS-232 Converter, and then unplug the unit.

1. Open the Device Manager and under Other devices right-click on USB High Speed RS-232 Converter.

Page 71
LABORIE System Installation Guide for Windows 7 OS

2. Click Update Driver Software.

3. Select the Browse my computer for driver software option.

Page 72
LABORIE System Installation Guide for Windows 7 OS

4. Select the Let me pick from a list of device drivers on my computer option.

5. Select Show All Devices and click Next.

Page 73
LABORIE System Installation Guide for Windows 7 OS

6. Click Have Disk.

7. Select ftdibus.inf and click Open.

Page 74
LABORIE System Installation Guide for Windows 7 OS

8. Ensure the correct address is displayed for driver location and click Open.

9. Select USB Serial Converter and click Next.

Page 75
LABORIE System Installation Guide for Windows 7 OS

10. Click Close.

11. Open the Device Manager once again and right-click USB Serial Port under Other devices.

Page 76
LABORIE System Installation Guide for Windows 7 OS

12. Click Update Driver Software.

Page 77
LABORIE System Installation Guide for Windows 7 OS

13. Click Browse my computer for driver software.

14. Select Let me pick from a list of device drivers on my computer.

Page 78
LABORIE System Installation Guide for Windows 7 OS

15. Select Show All Devices and click Next.

16. Click Have Disk.

Page 79
LABORIE System Installation Guide for Windows 7 OS

17. Select ftdiport.inf and click Open.

18. Ensure the correct address is displayed for driver location and click Open.

Page 80
LABORIE System Installation Guide for Windows 7 OS

19. Select USB Serial Port and click Next.

20. Click Close.

Page 81
LABORIE System Installation Guide for Windows 7 OS

21. Open Device Manager:


 USB Serial Port should now be active under Ports and USB Serial Converter should
be visible under Universal Serial Bus Controllers.

22. Plug the Grid Connect USB Serial Convertor into every port on the computer and check the
notifications/Device Manager for any COM Port assignment failures or exclamation marks.

NOTES:
 When installing a second/additional unit of the GridConnect USB Serial Convertor,
do not plug the second unit into the same USB port of the computer used when
installing the first unit. Instead, choose a different USB port to plug and install the
second unit.
 Never directly unplug the GridConnect USB Serial Convertor from the USB port of
the computer. Always go first into the notification area and right-click the Safely
Remove Hardware and Eject Media option, choose Eject USB High Speed |RS-232
Converter, and then unplug the unit.
23. Shutdown then power ON the computer. Unplug and then plug the Grid Connect adapter back
in to the same UDS port to which it was previously connected. Open the Device Manager and
confirm that the COM port assignment has not changed.

Grid Connect setup complete.

Page 82
LABORIE System Installation Guide for Windows 7 OS

Configure the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURiO]


NOTE: This step does not need to be performed for the DELL E6530 laptop or the for the DELL
Precision T3600 desktop PC.

! IMPORTANT!
COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] is a Plug-and-Play device;
therefore, after the COM098 is plugged into the USB Port of the computer and the Device
Manager lists TDK Bluetooth USB Adapter and Microsoft Bluetooth Enumerator under Bluetooth
Radio, there will be no need to follow this procedure. The following procedure must be
followed to its completion only if the Device Manager, under Bluetooth Radio, does not list TDK
Bluetooth USB Adapter and Microsoft Bluetooth Enumerator.

1. Plug the power cord of the UDS system into an electrical outlet.
2. Insert the Laborie Released Software disc BuildXX into the CD-ROM drive of the
computer.
3. Copy and paste the folder called Windows 7 to the computer desktop. The folder can be
found inside the CD in the following location:
D:\[UDS Hardware Model]\Software\Bluetooth Drivers\COM098
[Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO]\
OR
D:\COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO]\
4. Insert the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] Bluetooth
dongle into any USB port of the computer.
Result: The Found New Hardware Wizard window will automatically appear.

5. Select No, not this time and click Next.


6. In the resulting window select Install from a list or specific location [Advanced] and click Next.
7. Select the Don’t search. I will choose the driver to install option and click Next.

Page 83
LABORIE System Installation Guide for Windows 7 OS

8. Choose Bluetooth Radios from the Common hardware types list and click Next.

9. Click the Have Disk button.

Page 84
LABORIE System Installation Guide for Windows 7 OS

10. Click the Browse button.

11. Select the bth.ini file in the Windows 7 folder located on the desktop and click Open.
12. Confirm that the path location appears correctly and then click OK.
Result: The Found New Hardware Wizard window appears with the COM098 driver.
13. Click Next.
14. Select Yes if the following message screen appears.

15. Wait for 2-3 minutes while the wizard searches for the COM098 driver.
Result: The screen with the ‘TDK Bluetooth USB Adapter’ will appear confirming that the driver
has been installed onto the computer.
16. Click Finish. The computer will automatically initialize the driver installation of the Microsoft
Bluetooth Enumerator.

If you see the Found New Hardware Wizard window again, then follow steps below.
a. Choose ‘Install the software automatically [Recommended]’ and click Next.

Page 85
LABORIE System Installation Guide for Windows 7 OS

b. Wait 2-3 minutes while the wizard searches for the Microsoft Bluetooth Enumerator.
c. The following screen with the ‘Microsoft Bluetooth Enumerator’ will appear confirming that the
driver has been installed onto the computer.

d. Click Finish.
e. Plug the COM098 [Bluetooth, USB Adapter, EDRv2.0, MS Stack, EZURIO] into each USB port of
the computer, doing so, you may be asked by the computer to repeat ALL the steps explained
above (i.e. steps 1 to 16).
f. Continue with the following steps.

17. Delete the Windows 7 folder that was copied on the desktop in step 3.
18. Click Start > Control Panel and type the word Bluetooth into the search field at the top-right corner.

 As Bluetooth is typed into the search field, the available settings will automatically refresh. Click
the Change Bluetooth settings heading under Devices and Printers.

Page 86
LABORIE System Installation Guide for Windows 7 OS

 In the resulting Bluetooth Settings window, make sure the Show the Bluetooth icon in the
notification area option is selected. This will place the Bluetooth icon shortcut in the task tray.

 Click OK.

19. Click the Start button , click Control Panel, click Hardware and Sound, and then click
Bluetooth Devices.

20. Click Add and follow the on-screen instructions.

Page 87
LABORIE System Installation Guide for Windows 7 OS

Install HASP Driver for UDS120, UDS-Goby, NuWav Advanced, Urostym, Urocap
IV Standalone, and/or AquaVu Software
NOTE: This step does not need to be performed for the DELL E6530 laptop. However, you
must plug the Aladdin Hasp Key into each USB port on the laptop or PC before proceeding
with the other installations. At times it may take a minute for the HASP key to activate when
plugged in to the USB port. Wait for the hasp key light to illuminate before continuing to plug
into another USB port.
Do not connect probes to the computer until instructed to do so.
1. Insert the LABORIE software disc
into the CD or DVD drive and
double-click the
HASPUserSetup.exe application.

2. Click the I accept the license


agreement option and click Next.

3. Click Next in the following screen and wait for installation to occur.
4. Click Finish to complete installation.

5. Continue with the installation of the


UDS120 software, and/or the
NuWav Advanced software, and/or
the Urostym software, or the UDS-
Goby software, and/or the Urocap
IV standalone software, and/or the
AquaVu software.

Driver and peripherals installation complete. Continue with parts 3 and 4.

Page 88
LABORIE System Installation Guide for Windows 7 OS

Part 3: L ABORIE Softw are Installation

 Install NuWav Advanced Ultrasound Software with UDS120 Software (see page 90)

 Install UDS120 Software (for all Laborie UDS systems except Goby) (see page 92)

 Install UDS120 Software for a Single UDS Device Connection (see page 92)

 Install UDS120 Software for a Multiple UDS Device Connection (see page 93)Install the GOBY
Software (see page 98)

 Connect GOBY Components (see page 100)

 Install Urocap IV - Standalone Model (see page 104)

 Install Urostym Software (see page 110)

 Install the UROSTYM Hardware (see page 110)


 Install the Standalone UROSTYM V6 Software for a UROSTYM unit (PRO180) (see
page 113)
 Install the UROSTYM V6 Software for a UROSTYM Unit (PRO180) with a UDS Unit
(see page 118)
 Install the UROSTYM V6 Software Application for use with a UDS Unit Only (see
page 119)
 Configure the UROSTYM Software (see page 121)
 Configuring the UROSTYM V6 Software Application after UDS Software
Installation for Multiple BT Connection or for use with Urocap-III (see page 122)
 Transferring Patient Data (see page 125)
 Install EVOX Electrodiagnostic Software (see page 126)

 Install i-LIST Office Reporter Software (see page 131)

 Install AquaVu Software (see page 140)

 Install SeeMore™ Ultrasound Probe Driver and Software(see page 142)

 Install VISER Software (see page 147)

 Enable Urostym Reports with Urocap IV (see page 150)

 Upgrading UDS120 Software (see page 152)

Page 89
LABORIE System Installation Guide for Windows 7 OS

Install NuWav Advanced Ultrasound Software with UDS120 Software


1. Before installation make sure to configure the HASP key. If not already configured, see
page 88.
2. Insert the Laborie Disc Buildxxx software disc into the disc drive of your computer.
Result: The Welcome to UDS 120Vxx.x-ServerX Vx.xx Setup welcome screen will appear.

NOTE 1: If the CD does not automatically start, select the UDS120-


V1X.X.X_ServerXx.x.xx.exe program on the software disc to start installation.
NOTE 2: Screens in this guide may slightly differ from actual installation screens.

3. Click Next.
Result: The End User License Agreement will appear.

4. Click I Accept.
Result: The Connection Type window appears.

5. Select the hardware and type of connection and click Next. For Goby KT select UDS94 for
Delphis, Delphis IP, Delphis KT and the Bluetooth (MS Stack) options.

IMPORTANT! If you have selected a wireless option a screen will appear to remind you
to install Bluetooth software. Follow the appropriate Bluetooth Manuals for installing
the Bluetooth support on the computer or laptop.

Page 90
LABORIE System Installation Guide for Windows 7 OS

6. Select NuWav – Advanced plus any other Options and click Next.

7. Select the Language and click Next.

8. Select the default Folders to store software and patient data.

9. Click Next.
Result: The program is ready for installation.

10. Click Finish to complete installation.

Page 91
LABORIE System Installation Guide for Windows 7 OS

Install UDS120 Software (for all Laborie UDS systems except Goby)

Install UDS120 Software for a Single UDS Device Connection


NOTE: Before start of installation, make sure that the HASP Key driver is installed. If not already
installed, see page 88 for installation instructions.

1. Double‐click the UDS 120V12.x.x-ServerX Vx.xx.exe program in the Laborie software disc.
Result: The Welcome to UDS 120V12.x.x-ServerX Vx.xx Setup welcome screen will appear.
2. Click Next.
Result: The End User License Agreement will appear.
3. Click I Accept the license agreement and click Next.
Result: The Connection Type window appears.
4. Select the hardware and type of connection and click Next. For Goby KT select UDS94 for
Delphis, Delphis IP, Delphis KT and the Bluetooth (MS Stack) options.
5. Click Ok in the resulting warning screen.
6. Select any Options and click Next. This is where you will select options like video,
Ultrasound (for NuWav), infusion transducer, and/or infrared remote. If any of these are
part of your system you must select them here to ensure proper software configuration.
7. Click Ok in the resulting warning screen.
8. Select the Language and click Next.
9. Confirm the Folders and click Next.
10. Click Next.
Result: The program will be installed on the computer.
11. Click Finish to complete installation.
12. (optional) If the UDS120 software will be used with UDS-94 hardware having a v1.05 board
set, then open the uds120.ini file (located in C:\Windows\uds120.ini\) and change
the EMGDeltaAD value under the EMG section from 837 to 755.

If no other programs are required at this point then go to Part 4 System Finalization to
complete set up.

IMPORTANT NOTE for UDS120 Wired Connection Installation with the Adapter:
 Once the UDS installation process for Wired Connection is complete, change the uds120.ini
settings under the [UDS-120] section from CommPortLoopCount=300000 to
CommPortLoopCount=600000

Page 92
LABORIE System Installation Guide for Windows 7 OS

Install UDS120 Software for a Multiple UDS Device Connection


This installation process pertains to the installation of a Triton/Aquarius, Delphis/KT, TETRA, Urocap,
and/or Mini-D device combination.
The multiple device connection installation follows this procedure:
i. Install the UDS120 software (page 93)
ii. Configure channels in the UDS120 software (page 95)

NOTE 1. The ONLY Multiple Device connections supported with this software release are the
following:
 1 UDS device (Triton/Aquarius, Delphis/KT) with wireless urocap
 1 UDS device (Triton/Aquarius, Delphis/KT) with wireless urocap and a Mini-D device
 1 UDS device (Triton/Aquarius, Delphis/KT) with 2 Mini-D devices
 Goby-KT with Wireless Uroflow
 Tetra NIRS (wired) with Triton
 Tetra NIRS (wired) with AquariusTT
 Tetra NIRS with Urocap / Wireless Uroflow
 Urocap with Mini-D

LIMITATIONS: The pair of Multiple Devices NOT supported in this present release is the
following:
 Tetra NIRS with Mini-D

Note 2. Functions ONLY with Bluetooth (MS Stack) and NOT Bluetooth (Widcomm). Therefore,
Bluetooth (Widcomm) should not be using this present release of UDS120 for Multiple Devices.

Note 3. A maximum of three wireless connections are recommended for connection through Bluetooth.

Note 4: To install a multiple device connection with Urostym V6, please contact the Service
department.

Note 5: Make sure COM Port settings are configured.

IMPORTANT NOTE for UDS120 Wired Connection Installation with the Adapter:
 Once the UDS installation process for Wired Connection is complete, change the uds120.ini
settings under the [UDS-120] section from CommPortLoopCount=300000 to
CommPortLoopCount=600000

Page 93
LABORIE System Installation Guide for Windows 7 OS

To install the UDS120 software for a multiple UDS device connection:


NOTE: Before start of installation, make sure that the HASP Key driver is installed. If not already
installed, see page 88 for installation instructions.

1. Double-click the UDS 120V12.x.x-ServerX Vx.xx.exe application on the LABORIE


software disc.
Result: The Welcome to UDS 120V12.x.x-ServerX Vx.xx.exe Setup welcome screen will
appear.
2. Click Next.
3. Click I Accept the license agreement and click Next.
4. In the Connection Type window select Multiple Devices (Via Connection Manager) and
click Next.
5. Select the UDS device combination for connection. If you have more than two devices
then select one of the combinations in this window; the third device will need to be
configured in the Connection Manager of the UDS120 software, which occurs once the
software is completely installed.

IMPORTANT! If you have more than two UDS devices to connect and you are not sure of
which combination to select in this screen then follow the list below. Selections are listed in
order of importance.
1. Triton
2. Delphis
3. Tetra
4. Urocap
5. Mini-D
For example:
 If you have a Triton device, a Tetra device, and a Urocap device, then you would
select the Tetra NIRS with Triton option and configure the third device (the Urocap)
in the Connection Manager screen in the UDS120 software.
 If you have a Delphis device, a Mini-D device, and a Urocap device then you would
select the Delphis with Urocap option and configure the third device (the Mini-D) in
the Connection Manager screen in the UDS120 software.

6. Click Next.
7. Select any Options and click Next.
8. Select the Language and click Next.
9. Click Next in the Destination Folder screen.
10. Click Next to start the installation. Installation may take a few minutes to complete.
11. Click Finish to complete the software installation.
12. (optional) If the UDS120 software will be used with UDS-94 hardware having a v1.05
board set, then open the uds120.ini file (located in C:\Windows\uds120.ini\)
and change the EMGDeltaAD value under the EMG section from 837 to 755.
13. Continue with the configuration of the Multiple Device Configuration as described in
the following section.

Page 94
LABORIE System Installation Guide for Windows 7 OS

UDS120 Client Software Multiple Device Configuration


 For the Multiple Devices (Via Connection Manager) option with Bluetooth support, only
Bluetooth (MS Stack) is supported; Bluetooth (Widcomm) is not supported.

1. Start the UDS120 software by double-clicking the UDS 120 shortcut on the desktop.
2. Click Connection Manager to open the Connection Manager window.

3. You will see two UDS devices already listed in the Device column. These were
chosen during the installation process. If you have more than two UDS devices
then select any additional devices in this column.

IMPORTANT!
Do not click the
Play button
until ALL
devices have
first been set up
in the Device,
Connection
Type, and Wired
COM
Step 7 Step 3 Step 4 Step 5
Port/Wireless
Settings
columns.

4. Select Wired or Wireless as the connection type.

5. Select one of the Settings:


 If Wired, select the Wired COM Port to which the UDS device is connected.
 If Wireless, click the Search BT button to connect the UDS device to the
software. The name and serial number of the UDS device will appear in the
Bluetooth Name and Address fields.

6. Repeat these selections for each UDS device.

7. Once all connections are set, close the UDS120 software and then restart the
UDS120 software.

8. Once ALL the devices have been added click the button next to each device to
start the UDS device:

Page 95
LABORIE System Installation Guide for Windows 7 OS

Play button = turn UDS device power ON.


Stop button = turn UDS device power OFF.
9. Click the Calibration button next to a UDS device to open the Calibration
window.

10. Complete the calibration procedure as required.

11. When the Calibration button for another UDS device is clicked a message box
appears asking for the selection of the Flowmetry device to use for calibrating
the Pump transducer. For example: If Urocap1 is to be chosen as the Flowmetry
device to use with the pump, then calibrate the Urocap1 device before starting
the calibration of the Pump transducer inside the Delphis1 device.

12. After selecting the Flowmetry device to use with the pump, a Transducer
Calibration window appears. Continue with the calibration of the other UDS
device and then continue with the following steps.

13. Click the Channel Mappings tab. This will display the channels that are needed to
configure the UDS graph for testing with the UDS devices. The channels from the
devices selected during installation will already be listed.
* NOTE: If any changes are made to the Channel Mappings tab after initial configuration, the UDS120
software must be restarted.

* NOTE: If you have more than two UDS devices to connect, you will need to map the channels in this
list for any additional device(s). Scroll down the list to the first empty line and begin by selecting the
Device Name, then select the Device HW Channel and then select the UDS Screen Channel Name.
Refer to the following charts for additional channel setup:
Additional Mini-D1 Device
UDS Screen Channel Name Device Name Device HW Channel
Ch9 Mini-D1 Ch1
Ch10 Mini-D1 Ch2
Ch11 Mini-D1 Ch3
EMGR Mini-D1 EMGR
Additional Mini-D2 Device

Page 96
LABORIE System Installation Guide for Windows 7 OS

UDS Screen Channel Name Device Name Device HW Channel


Ch12 Mini-D2 Ch1
Ch13 Mini-D2 Ch2
Ch14 Mini-D2 Ch3
EMGR2 Mini-D2 EMGR2

Additional Urocap II/III Device


UDS Screen Channel Name Device Name Device HW Channel
Flow_2 Urocap Flow
Volume_2 Urocap Volume

Additional Tetra NIRS Device


UDS Screen Channel Name Device Name Device HW Channel
Ch15 Tetra NIRS Ch1
Ch16 Tetra NIRS Ch2

14. Once all selections are set click OK.


NOTE: If any changes are made to the Channel Mappings tab after initial configuration, the UDS120
software must be restarted.

15. In the UDS120 main window click Config > Set up/Modify.
16. Create a configuration file that contains all the channels listed in the UDS Screen
Channel Name column from the Channel Mapping tab. For example, following the
channel listing as seen in the screen from step 13 the configuration file will look
like this:

17. Type a name in the Test Name box. This will be the basic configuration file that
you will use for testing with all the connected devices.
18. Click OK.

If no other programs are required at this point then go to Part 4 System Finalization to complete
set up.

Page 97
LABORIE System Installation Guide for Windows 7 OS

Install the GOBY Software


1. Insert the LABORIE software disc into the disc drive of the laptop.
2. Double-click the UDS-GOBY installation application on the LABORIE software DVD.
3. Click Next in the welcome
screen.

4. Click the I Accept the license


agreement option and then
click Next.

Page 98
LABORIE System Installation Guide for Windows 7 OS

5. Confirm the Destination


Folder location and click
Next.

6. Click Next to start the installation.


7. Wait for the system to
configure the software. This
will take a few minutes.

8. Click Finish to complete


installation.

Page 99
LABORIE System Installation Guide for Windows 7 OS

If the following message appears


when you click Finish, plug the
HASP key into any USB port and
click OK to continue. This message
appears when there is a
completely new install of the HASP
driver and Goby software.

9. Once installed two software


shortcuts are placed on the
desktop and an additional
GOBY Device Manager icon is
placed in the system tray in
the lower-right.

10. (optional) Insert the Bluetooth dongle (COM093) into a USB port on the computer and
wait for the configuration of the Bluetooth.

11. Insert the HASP key into one of the USB ports on the computer and wait until Windows
configures the HASP key driver. Plug the HASP key into every USB port of the computer.

12. Continue with the installation of the i-LIST Office Reporter Software (see page 131). Once
i-LIST Office Reporter is installed return to this page and continue with the connection of
the GOBY components (as outlined in the following section).

Connect GOBY Components


1. Place the UDS Roam cartridge into the
GOBY Hub and make sure the connector
UDS Roam
at the bottom of the UDS Roam cartridge
clicks securely into place into the Hub.

2. Plug the Hub into the charger and plug GOBY


Hub
the charger into an outlet. If using a GO
Cart, plug the power cord of the cart into
an outlet. Make sure the cart is on a level
surface and the wheels are locked.

3. Switch the GOBY Hub to ON with the power switch at the bottom of the Hub.

Page 100
LABORIE System Installation Guide for Windows 7 OS

4. Plug the Urocap IV into the charger and plug the charger into an outlet.
 If using a GO Cart, place the Urocap
into the charger at the bottom of the
GO Cart and then plug the power
cord of the cart into an outlet. Make
sure the cart is on a level surface and
the wheels are locked.
5. Double-click the GOBY Device Manager
icon on the computer desktop.

6. The GOBY Device Manager window will


display the status of the GOBY
components.

7. Start by connecting the Urocap IV Uroflowmeter. Make sure it is near the computer.

8. Click Setup to connect the


equipment.

9. In the Urocap Setup window, click


the Device Search button to start
locating the Uroflowmeter that
needs to be connected for the
system.

10. Click Cancel in the Bluetooth Device


Server window.

11. Click Device Search again.

Page 101
LABORIE System Installation Guide for Windows 7 OS

12. Wait for the Urocap to be located.


When the Urocap is located in the
Bluetooth Device Server window,
highlight the name and then click OK.

13. When a connection is established,


the Bluetooth address and serial
number fields will display the
component’s information.

14. Click OK to go back to the GOBY Device Manager window.

15. Repeat steps 7 to 14 to connect the GOBY Hub and the UDS Roam to the system.

16. Once the components are set in the


setup window, click once on each
component button (Urocap IV, GOBY
Hub, and UDS Roam) in the Device
Manager window and wait for the
connection to establish and finalize.

When all components are connected and functioning, the GOBY Device Manager window will
look similar to this:

Page 102
LABORIE System Installation Guide for Windows 7 OS

GOBY installation and setup complete. Continue to Part 4.

Page 103
LABORIE System Installation Guide for Windows 7 OS

Install Urocap IV - Standalone Model


Before installation, make sure that:
 The laptop is configured as outlined in Part 1 and Part 2 of the UDS System / Computer
Configuration and Setup Guide for LABORIE UDS Systems with Windows 7® OS.

Order of Installation
When installing Urocap IV software, the order for proper installation is as follows:

1. Install the Sentinel HASP key driver (if not already installed). See page 88.

2. Install the Urocap IV software. i-LIST Office Reporter software is part of the Urocap IV
standalone install so there is no need for separate install for i-LIST Office Reporter software.

Step 1: Install the GOBY Urocap IV Software


1. Insert the LABORIE software disc into the disc drive of the laptop.
2. Double-click the UDS-GOBY installation application on the LABORIE software DVD.
3. Click the Accept button.

Page 104
LABORIE System Installation Guide for Windows 7 OS

4. Click Next.

5. Confirm the Installation Folder


location and click Next.

Page 105
LABORIE System Installation Guide for Windows 7 OS

6. Click Next to start the


installation.

7. Wait for the system to


configure the software. This will
take a few minutes.

8. Click Close to complete


installation.

Page 106
LABORIE System Installation Guide for Windows 7 OS

9. Once installed two software


shortcuts are placed on the
desktop and an additonal GOBY
Device Manager icon is placed
in the system tray in the lower-
right.

10. (optional) Insert the Bluetooth dongle into a USB port on the computer and wait for the
configuration of the Bluetooth.

11. Insert the HASP key into a USB port on the computer.

12. Continue with Step 2.

Step 2: Connect Urocap IV to the UDS120 Goby Software


The Goby Device Manager window allows you to connect the Urocap IV to the software. Once
the connection is established in this window, the equipment will automatically connect each
time the computer is powered on.

1. Double-click the Goby Device Manager


icon on the computer desktop.

2. The Goby Device Manager window will


display the status of the Urocap IV
equipment.

Note: The Goby Hub and UDS Roam are not


required for this connection.

3. Make sure the Urocap IV is near the computer.

4. Click Setup to connect the


equipment.

Page 107
LABORIE System Installation Guide for Windows 7 OS

5. In the Urocap Setup window, click


the Device Search button to start
locating the Uroflowmeter that
needs to be connected for the
system.

6. Click Cancel in the Bluetooth Device


Server window.

7. Click Device Search again.

8. Wait for the Urocap to be located.


When the Urocap is located in the
Bluetooth Device Server window,
highlight the name and then click OK.

9. When a connection is established,


the Bluetooth address and serial
number fields will display the device
information.

10. Click OK to go back to the Goby Device Manager window.

Page 108
LABORIE System Installation Guide for Windows 7 OS

11. In the Device Manager window click


the Urocap IV device button and wait
for the connection to establish and
finalize.

When the Urocap IV is connected and functioning, the Goby Device Manager window will look
similar to this:

Setup complete. Continue with the final setup process in Part 4: System Finalization.

Page 109
LABORIE System Installation Guide for Windows 7 OS

Install Urostym Software


This section outlines the installation process for the UROSTYM unit and software setup.

Important notes: If the UDS120 software application is to be installed on the same computer
with UROSTYM software, then the UDS software must be installed first; do not register the
modules again through the UROSTYM Utility program, but instead keep using the modules
registered through the UDS120 software.

The HASP key is required for this version of UROSTYM. Do not plug the key in before installation
of the software is completed. If it is already plugged into the computer, remove it and re-insert
it when asked to do so during installation.

Install the UROSTYM Hardware

1. If the Line Isolation Transformer (LIT) has a power switch, make sure it is in the OFF
position.
2. Connect one end of the power cord (LIT080) to the LIT and connect the other side to the
UROSTYM.
3. Plug the AC adaptor of the laptop computer into an outlet. If using power adaptor
POW320 with a laptop, plug it into the LIT. If using a desktop computer, plug the
computer and monitor power cords into the LIT. NOTE: LIT is recommended to meet
Hospital Grade power protection requirements in the United States.
4. Plug the printer power cord into the LIT.
5. Connect the medical grade power cable between the (LIT) and an electrical outlet. If the
LIT has a power switch, turn on the LIT with the power switch.

Refer to the following setup diagrams for more information.

Page 110
LABORIE System Installation Guide for Windows 7 OS

Page 111
LABORIE System Installation Guide for Windows 7 OS

Page 112
LABORIE System Installation Guide for Windows 7 OS

6. Turn on the UROSTYM unit.

 The USB-to-Serial Adapter driver must be installed before continuing. If it is not already
installed go to page 71 for installation instructions.
 No more than one adapter should be configured and connected to the computer.
Configuration of a second adapter unit to work on the same computer will result in
communication failure and may require re-installing the operating system on the
computer. If for any reason a new adapter is ordered, the aforementioned failures will
result.
 If using Urostym with other Laborie UDS equipment, then the wired connection to the
Urostym system must be configured before the Bluetooth connection to UDS or Urocap
systems is configured.
 Never change the COM Port number of the ‘Unknown Device’ (i.e. the PER380 or
PER381) located under the Device Manager – Ports (COM & LPT).

Install the Standalone UROSTYM V6 Software for a UROSTYM unit (PRO180)


1. Back up the C:\Windows\uds120.ini file.
2. Insert the UROSTYM software disc into the CD or DVD drive.
3. Double-click the HASPUserSetup.exe application.

4. Click the I accept the license agreement option and click Next.

Page 113
LABORIE System Installation Guide for Windows 7 OS

5. Click Next in the following screen and wait for installation to occur.

6. Click Finish to complete installation.

7. In the Laborie software disc double-click


Urostym_v6.3.0_Bxx_Win7_SvX3.0.75_UG141_CDB150.exe.
8. Click Next.
9. Specify the user’s Full Name, the name of the Organization and click Next.
10. Select the Destination Folder for the UROSTYM software and click Next.
11. Click Next in the ready to install screen.
12. During installation the window may appear “frozen” for a few minutes. Please wait for
the installation to continue.
13. Click Finish to complete the installation.

During the installation process, files are copied onto the PC and a UROSTYM shortcut

is placed on the desktop.

Page 114
LABORIE System Installation Guide for Windows 7 OS

14. Plug the HASP key into a USB port and wait until Windows configures the HASP key
driver.
15. Start the UROSTYM software by double-clicking the UROSTYM shortcut on the desktop.
16. Click OK in the resulting pop-up messages.

17. Set up the UROSTYM components in the resulting CheckDB window section by section.

Section 1. Database 6.3.0 (in the upper-left of the screen)


 Click the Update DB to Version 6.3.0 button.

 Click the Check Protocols button.


 Click the Check EMG/Uroflowmetry Tests button.

Section 2. HASP Key (on the upper-right side of the screen)


 Click the HASP Info button to check the available features (available features have a
check mark next to them).

Page 115
LABORIE System Installation Guide for Windows 7 OS

Section 3. UROSTYM Connection (in the middle of the screen)


 Set the proper COM port number for the UROSTYM Unit. To set the COM port
number click the Change Com Port button and select Port #1 if using a laptop or
Port #2 if using a desktop PC.
 Click the Connect To Urostym button to check the connection.
Result: The connection indicator changes to a green color, and the UROSTYM
Firmware Version text box is populated by the Firmware version number.

Section 4. UROSTYM UI Language (on the right side of the screen)


 Select the language for the User Interface.

Section 5. Components (on the bottom-left side of the screen)


 If the Registered Version line (indicated in the circled area below) of the ServerX
component is empty,

Page 116
LABORIE System Installation Guide for Windows 7 OS

Then, click the Register Components button and select the select the ServerXMSB.ocx for a
PC with Bluetooth (MS Stack) component to be registered:

Result: The RegSvr32 message will appear. Click OK to close the pop-up message

 If the installed version line of the ServerXShell component is empty (indicated in


the circled area below) and the PC has Bluetooth installed, then click the
Register Components button and select the ServerXShellMS component to be
registered.

18. Click the Exit button to close the Application. In the resulting message click Yes to start
the UROSTYM application or click No to exit.

The software installation is complete. Go to the configuring the software section on page
121.

Page 117
LABORIE System Installation Guide for Windows 7 OS

Install the UROSTYM V6 Software for a UROSTYM Unit (PRO180) with a UDS Unit
1. Install the UDS120 software if not already installed.
2. Install the UROSTYM V6 Software Application (see page 114, start from step 7 and go to
step 18 (but do not exit the checkDB utility window).
3. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in
the bottom-left of the screen).
Result: The UDS120.ini File Configurator window appears.

4. Set the path of the UDS120.ini file by clicking the path selection button in the
upper-left of the screen and then selecting the UDS120.INI file located inside the
C:\Windows\ folder.

5. Click the Connect to UDS/Urocap button.


Result: The UDS unit will connect in a few seconds (connection time depends on
connection type and UDS hardware model). In the ServerX section the green indicator
will appear, and the section will be populated by the ServerX version. If not visible, it
means that the connection was not established. Check the connection type and click the
Connect to UDS/Urocap button again.

Page 118
LABORIE System Installation Guide for Windows 7 OS

6. Tap the Uroflowmeter transducer to verify that the device is responding.


Result: The UDS indicator is green and the progress bar and text box (located below the
Connect to UDS button) change displays. For the UDS 94 Triton two progress bars will be
displayed for Uroflow1 and Uroflow2 transducers.
7. Close the CheckDB application.
The software installation is complete. Go to the configuring software section on page 121.

Install the UROSTYM V6 Software Application for use with a UDS Unit Only
1. Install the UDS120 software if not already installed.
2. Install the UROSTYM V6 Software Application (see page 114, start from step 7 and go to
step 18 (but do not exit the checkDB utility window).
3. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in
the bottom-left of the screen).
Result: The UDS120.ini File Configurator window appears.

Page 119
LABORIE System Installation Guide for Windows 7 OS

 If the UDS120.ini file is configured for a Multiple Bluetooth connection, then the
following message will be displayed:

 Click OK.

4. Click the Connect to UDS button.


Result: The UDS unit will connect in a few seconds (connection time depends on
connection type and UDS hardware model). In the ServerX section the green indicator
will appear, and the section will be populated by the ServerX version. If not visible, it
means that the connection was not established. Check the connection type and click the
Connect To UDS button again.

5. Tap the Uroflowmeter transducer to verify that the device is responding.


Result: The UDS indicator is green and the progress bar and text box (located below the
Connect to UDS button) change displays. For the UDS 94 Triton two progress bars will be
displayed for Uroflow1 and Uroflow2 transducers.
6. Close the CheckDB window.

The software installation is complete. Go to the configuring software section on page 121.

Page 120
LABORIE System Installation Guide for Windows 7 OS

Configure the UROSTYM Software


Once the software is installed it will need to be configured for use.
1. Start the UROSTYM application.
2. Click the Set COM button (if the Set COM window was not displayed
automatically).

3. Depending on the equipment used for the system make the selections as
follows:

UROSTYM unit and UDS UDS unit only


UROSTYM unit only unit (Urocap-III, Dorado -KT, Delphis,
(Urocap-III, Dorado -KT, Delphis, Triton, Aquarius TT)
Triton, Aquarius TT)
Check off the UROSTYM Check off the UROSTYM Unit Check off the UDS Unit
Unit check box only and and the UDS Unit check check box only.
confirm the COM port boxes and confirm the COM Select the desired
number. port numbers. channel(s) and/or
Select all three channels Select the channels which Uroflowmetry radio
for will be used for button.
Biofeedback/Stimulation. Biofeedback/Stimulation
studies with each of the
units.

4. If a UDS unit is used alone or with a UROSTYM unit for


Uroflowmetry/Biofeedback studies:
o Select desired transducers (for example: EMG1, EMG2, P1, P2, P3, etc...) for
biofeedback channels (if the UDS unit is not the Urocap-III).
o For Uroflowmetry channel select desired transducer if the UDS unit has
more than one channel (Dorado -KT, Triton and Aquarius TT units have two
Uroflowmetry transducers)
o Click the Uroflowmeter Adjustment button.
o Tap the Uroflowmeter to make sure that the device is responding.

Page 121
LABORIE System Installation Guide for Windows 7 OS

Result: The UDS indicator on the main window status bar is green, the
Samples value on the Adjustment form goes up, and the curves in the Flow
and Volume graph areas show a curve deflection.
o Click OK.

5. Click OK.
6. Close the UROSTYM application.

UROSTYM software is configured and ready for use.

Configuring the UROSTYM V6 Software Application after UDS Software Installation


for Multiple BT Connection or for use with Urocap-III
If in the existing system the UROSTYM software was installed for using the UDS unit, referenced
on the C:\Windows\UDS120.ini file, and the UDS software is added and configured for Multiple
BT connection, or the UROSTYM software should be configured for using the Urocap-III as an
Uroflowmeter, then UROSTYM software will not be able to establish connection with UDS or it
can stop working with the UDS or Urocap-III unit. To secure that it works properly:

1. Double-click the UrostymCheckDB icon on the Desktop:

2. Click the Configure the UDS120.ini file button (in the 6.Path to UDS120.ini file section in the
bottom-left of the screen).

Page 122
LABORIE System Installation Guide for Windows 7 OS

Result: The UDS120.ini File Configurator window appears and the warning message may
display. Click OK to continue.

3. Click the path selection button in the upper-left of the window.

4. In the resulting window, set the path to the UDS120.ini file to C:\Program Files\Laborie
Medical Technologies\Urostym\UDS120.INI.UC-III.

5. In the Connection/Hardware Type section select the Hardware type and connection type.
ii. For a wired connection click the Change COM Port button and select the COM port
that you wish to use:
COM1 – for Triton/Delphis family

Page 123
LABORIE System Installation Guide for Windows 7 OS

COM4 – for Urocap-III wired unit


iii. For BT connections click the Search BT Device button and select the BT device in the
pop-up window.
Note: The Search BT Device button is enabled if the ServerXShell
component has registered and the Connection radio button is set to
Bluetooth (MS Stack) or Bluetooth (Widcomm).

6. Click the Connect to UDS/Urocap button.


Result: The UDS unit will connect in a few seconds (connection time depends on connection
type and UDS hardware model). In the ServerX section the green indicator will appear, the
section will be populated by the ServerX version. If not visible, it means that the connection
was not established. Check the connection type and click the Connect to UDS/Urocap
button again.

7. Tap the Uroflowmeter transducer to verify that the device is responding.


Result: The UDS indicator is green and the progress bar and text box (located below the
Connect to UDS button) change displays. For the UDS 94 Triton two progress bars will be
displayed for Uroflow1 and Uroflow2 transducers.

Page 124
LABORIE System Installation Guide for Windows 7 OS

8. Click the Transducers Calibration button and calibrate all necessary transducers according
to UDS Manual or follow the on-screen instructions.

9. Click Exit and close the CheckDB window.

UROSTYM software for multiple Bluetooth connection or for Urocap III use is complete.

Transferring Patient Data


To transfer patient data from one computer to another, follow these steps:

1. Install the UROSTYM 6.3 software on a new computer.

2. Back up the Urostym.mdb file found on the old computer with the patient data.
Note: By default the Urostym.mdb file is located in the folder:
For version 6.0.3 or older – “C:\Program Files\Laborie\Urostym\"
For version 6.1.0 - "C:\Program Files\Laborie Medical Technologies\Urostym\"

3. Insert a USB memory key into the old computer and copy the Urostym.mdb file to the key.
Remove the memory key.

4. Insert the memory key into the new computer and paste the Urostym.mdb file to the new
computer to the C:\Program Files\Laborie Medical Technologies\Urostym\ folder.
You will need to overwrite the existing Urostym.mdb file.

5. Click the UrostymCheckDB icon on the desktop.

6. Click the Update DB to version 6.3.0 button in the CheckDB window located in section 1
Database 6.3.0.

7. Close the CheckDB window.

UROSTYM software is ready for use.

Page 125
LABORIE System Installation Guide for Windows 7 OS

Install EVOX Electrodiagnostic Software

IMPORTANT NOTES: PLEASE READ BEFORE INSTALLATION

 If both the UDS120 software application and the Evox software application are to be installed on
one computer, first install the UDS120 software application and then continue by installing the
Evox software application named EvoxV1.1.x_ServerX3.0.x_Setup.exe [located inside the folder
named :\\Laborie Software Disc\Evox\Software\].

 If you currently have, or previously had, any EP (Evoked Potential) or Evox software applications
installed on your computer, please back up all configuration (.CFP, .CFT), settings (.INI), and UDS
data (.EDA) files on the computer before EVOX Software installation.

 Each file required for installation is located in the LABORIE Software Disc:
o UDS120 v12.X.X-ServerX3.0.XX.exe: \\Laborie Software Disc\[UDS Device
Name]\Software\V12.x.x_ServerX3.0.x\UDS120-V12.x.x_serverX3.0.x.exe.
o EvoxV1.X.X_ServerX3.0.XX_Setup.exe:\\Laborie Software Disc\Evox\Software\
EvoxV1.1.x_ServerX3.0.x_Setup.exe.

 When UDS120 V12.1.xx (ServerX3.0.xx) and Evox V1.1.xx (ServerX3.0.xx) are installed on the
same computer, do the following before launching UDS120 software.
i. Plug-in the Aladdin HASP key programmed with the required features.
ii. Install UDS120 software
iii. Install Urostym software (optional)
iv. Install Evox software
v. Install iLIST Office Reporter software
vi. Launch Evox software and then exit the Evox software (to initiate the ServerX
registry)

Step 1: Install EVOX Software


1. Once the UDS120 software has
been installed, double-click the
EvoxV1.1.x_ServerX3.0.x_Setu
p.exe file in the Laborie
software CD.
Result: The Welcome to the
Evox Version 1.x.x Installation
Wizard screen appears.

2. Click Next.

Page 126
LABORIE System Installation Guide for Windows 7 OS

3. Click the I accept the license


agreement option and click
Next.

4. Select the Hardware type and


the Connection type.

5. Select the COM Port where the


infrared remote will be
connected.

6. Click Next.

7. Click Next.

Page 127
LABORIE System Installation Guide for Windows 7 OS

8. Wait for the software to install.


This may take a few minutes.

9. Click Finish to complete


installation.

10. Plug the HASP key into a USB


port on the computer and
Double-click the EVOX shortcut
icon on the desktop to start
the software.

NOTES:
i. After EVOX Installation, the following directories are created on the PC:
 C:\Data\Evox (the path for EVOX data files)
 C:\CFG\Evox (the path for EVOX configuration files)
 C:\UDS120\ (the path for EVOX program files)

ii. When UDS120 and EVOX Software are installed on the same computer, the EVOX software and the
UDS software CANNOT RUN AT THE SAME TIME. One needs to be closed prior to opening the other
software.

11. Continue with the next step.

Page 128
LABORIE System Installation Guide for Windows 7 OS

Step 2: Connect the Equipment

Page 129
LABORIE System Installation Guide for Windows 7 OS

Step 3: Establish EVOX Connection

Establish a Bluetooth Connection


1. Start the EVOX software by clicking the EVOX shortcut icon on the computer
desktop.
Result: The Laborie Neurophysiology System welcome screen appears.
2. Click System > Choose Bluetooth.
3. Select the Bluetooth device that matches your hardware and click OK.
4. Wait one full minute and look for the icon in the upper-right corner to display the
connected status.

Connected Disconnected
NOTE: You may be required to restart the software and hardware.
If you power off the equipment, please wait at least one minute before powering it on again.

Step 4: Upgrade HASP Key (Optional)


In some instances the hasp key may need to be updated. A message will appear if this is
necessary. Please contact Laborie’s Service department for help with updating the HASP key.

Step 5: Copy Serial Numbers (Optional)


If you are installing the EVOX software application on a computer not supplied by Laborie
please note the following:
a. After having completed both the UDS120 and the Evox software application installation,
secure the uds120.ini and Uds120_Evox.ini from the C:\Windows folder to a safe
location on the computer (This will help in case any unexpected events may occur at the
end of the complete installation).
b. Once step a. is complete, transfer the UDSSerialNumber.ini file [for example: TSKE
09070012.ini] from the Laborie Product Disc [located inside the folder: \\[UDS
Hardware]\Software\V1X.X.X_ServerX3.0.XX] and paste into the C:\Windows folder.
c. After having performed the transfer to the new location, rename the UDSSerialNumber.ini
file to uds120.ini.
d. Transfer the UDSSerialNumber_Evox.ini file [example: TSKE 09070012_Evox.ini] from the
Laborie Product Disc [located inside the folder: \\Evox\Software] and paste into the
C:\Windows folder.
e. After having performed the transfer to the new location, rename the
UDSSerialNumber_Evox.ini file to Uds120_Evox.ini.

EVOX installation complete.

Page 130
LABORIE System Installation Guide for Windows 7 OS

Install i-LIST Office Reporter Software


NOTE: Remove Norton Antivirus before start of iLIST Office Reporter Installation if installed on
laptop or PC.

1. Double-click iLIST-Snapshot.exe in the Laborie Disc Buildxxx\iList Reporter\Software


folder on the software disc.

2. Depending on the system components present on the system, the installation process
will occur in one of the following ways:
i. If .Net Framework 3.5 SP1 and SQL Server 2008 Express Edition are
already installed then the i-LIST Office application will automatically
begin. Go to step 10 on page 133 .
ii. If .Net Framework 3.5 SP1 and SQL Server 2008 Express Edition are not
installed go to step 3 on page 131.

3. To begin .Net Framework 3.5 SP1 installation click Accept in the following three setup
windows as they appear on screen.

Page 131
LABORIE System Installation Guide for Windows 7 OS

4. Wait for the setup to continue. Ignore the “Not Responding” message and let the
installation run.

5. When the request for restart window appears, click Yes.

6. Click Yes.

Page 132
LABORIE System Installation Guide for Windows 7 OS

7. After restart, the i-LIST Prerequisites Setup will continue, and the SQL Server
2008 Express Edition installation will automatically start.

8. Click Yes to restart the computer.

9. After restart, double-click the iLIST-Snapshot.exe program again and continue


with the following steps.

10. The i-List Office Reporter 1.3.0.9 Setup Wizard will begin.

11. Click Next.

Page 133
LABORIE System Installation Guide for Windows 7 OS

12. Click Next.

13. Click Next.

14. Wait for installation to occur. The program will show a series of screens that
track the installation processes.

Page 134
LABORIE System Installation Guide for Windows 7 OS

15. Click Close to complete installation,

16. Click OK to close the message box.

17. Continue with Configure/Verify.

Page 135
LABORIE System Installation Guide for Windows 7 OS

Configure/Verify UDS configuration files


Configure each of the of the following UDS configuration files (.CFG files): Uroflow, CMG,
Micturition, UPP.

1. Start the UDS120 or UDS Goby software and open the Uroflow configuration file by
either clicking the Uroflow button on the UDS control panel or clicking Config > Open
and then double-clicking the uroflow.cfg file.
2. Click Config > Set up/Modify.
3. Ensure that the Test Type box has the correct test type (for example, Uroflow is set in
the following illustration) selected.
4. Click the ILIST Events Linking tab and then click the Full Test Graph Settings button.

5. Ensure all the settings on the i-LIST Event Graph are set as in the image below:

6. Repeat steps 1 to 5 for all remaining configuration files (Micturition, UPP, etc…). For each
test, enter the correct test type and select the correct channels in the Specify the Channels
section in the i-LIST Event Graph box.

Page 136
LABORIE System Installation Guide for Windows 7 OS

7. Continue with the patch installation as detailed below.

Install Patch
NOTE: Make sure all other applications are closed before installing the patch.

1. In the LABORIE software disc, double-click the RunMe.exe file in the Patch-
ArmReport2And3-for-v1-3-0-18 folder.

2. Click Run.

Page 137
LABORIE System Installation Guide for Windows 7 OS

3. Click OK in the warning message.

4. Wait for the installation to occur.


5. Click OK when installation is complete.

6. Continue with Verification of Installation in the following section.

Verification of Installation
1. Right-click on the Computer icon on the computer desktop and select Manage.
2. From the Computer Management window check if the snapshotqueue and udsadmin
entries appear in the Message Queuing group as illustrated in the following image:

Page 138
LABORIE System Installation Guide for Windows 7 OS

i-LIST Office Reporter software installation complete. Restart the computer and continue with the
installation of any remaining applications.

Install i-LIST Enterprise


Due to its complexity, the variety of options available, and the high degree of customization this
product is currently installed by the i-LIST Team only. Please contact the i-LIST Team for
support at 1-800.522.6743.

Page 139
LABORIE System Installation Guide for Windows 7 OS

Install AquaVu Software


Before installing AquaVu software for use with UDS systems, make sure that the UDS120 and i-
LIST Office Reporter software are already installed.

IMPORTANT! INSTALLATION MUST OCCUR IN ADMINISTRATOR MODE

Do not connect the probe to the computer until instructed to do so during installation.

NOTE: Before start of installation, make sure that the HASP Key driver is installed. If not already
installed, see page 88 for installation instructions.

1. Double-click the setup.exe file in the AquaVu product software folder on the LABORIE
software disc.

2. Click Next in the Welcome to the AquaVu Setup Wizard screen.

3. Click Next in the Confirm Installation screen.

4. Wait for installation to occur. It will take a few minutes to install.

5. If a Windows warning message


appears, click Install this driver
software anyway to continue
installation.

6. Click Close in the Installation Complete window.

7. Plug the probe cable into one of


the USB ports on the computer
and wait a few moments for probe
driver installation to occur.

8. Keep the probe plugged into the


computer and double-click the
AquaVu shortcut icon on the
desktop to start the software.

Page 140
LABORIE System Installation Guide for Windows 7 OS

9. Wait for probe detection to occur


and display in the software.

NOTE: If the following window


appears select the first option and
click Next.

AquaVu Installation complete.

Set up AquaVu for User Mode (optional)


If you have installed AquaVu software, additional important installation steps for AquaVu
support under user mode are listed below:

1. Once the Set up Security Permissions for User Mode Support process has
been performed on a computer with AquaVu installed, contact your IT department to request
and to enable user mode support.

2. Log in to the user account.

3. Make sure the probe is un-plugged.

4. Under user account, double-click the AquaVu desktop icon.


The installation process for the AquaVu software will appear to have restarted, however
this behavior is normal and expected. The installation process will take about 5 minutes
to complete.

5. Once the installation process is complete, the AquaVu software will open. At this point, close
the application, plug in the ultrasound probe into any USB port on the computer, and restart the
AquaVu software application.

AquaVu installation process is now complete and AquaVu enabled to be used under
user mode.

Page 141
LABORIE System Installation Guide for Windows 7 OS

Install SeeMore™ Ultrasound Probe Driver and Software


IMPORTANT NOTE - READ BEFORE START OF INSTALLATION!
Do not install the SeeMore application on any laptop or desktop PC if:
 it already has the NuWav ultrasound application installed
or
 it previously had the NuWav application installed and was then un-installed

1. Browse the Laborie Disc BuildXX\NuWav\Software\ and double-click the


SeeMoreSetup1.3.02Gall.exe file.

2. Click Run in the Open File – Security Warning screen.

3. Click OK in the self-extractor screen.

Page 142
LABORIE System Installation Guide for Windows 7 OS

4. Click Next to start installation.

5. Keep the destination folder as assigned by the installation program and click Next.

Page 143
LABORIE System Installation Guide for Windows 7 OS

6. Keep the start menu folder as assigned by the installation program and click Next.

7. Select the Create a desktop icon option and click Next.

Page 144
LABORIE System Installation Guide for Windows 7 OS

8. Click Next in the Ready to install screen.

9. Click Install this driver software anyway in the security message window.

10. Wait a few minutes for installation to occur.

Page 145
LABORIE System Installation Guide for Windows 7 OS

11. Click Finish in the Completing the SeeMore USB Application Setup Wizard screen.

Plug the probe into each USB port of the computer and restart the computer after each plug-in.
NOTE: If the following window
appears when probe is plugged in
or software starts then select the
first option and click Next.

Page 146
LABORIE System Installation Guide for Windows 7 OS

Install VISER Software

Important! Please read before starting installation


 VISER software cannot be installed with a multiple device and/or multiple
calibration connection.
 VISER software must only be installed after the installation of the UDS120
software.
Part One: Install the VISER Software
1. Insert the VISER software CD into the
CD or DVD drive of your computer and
start Setup.exe.
Result: The Welcome to VISER 1.1.5
Installation Wizard welcome screen
appears.

2. Click Next.
Result: The User Information screen
appears.

3. Add the user’s Full Name, the name of the Organization and click Next.
Result: The Destination Folder window appears.

Page 147
LABORIE System Installation Guide for Windows 7 OS

4. Select the Destination Folder for the


VISER software and click Next.
Result: The Ready to Install the
Application window appears.

5. Click Next when ready to install.

6. Click Next to start installation.

7. Click Finish to complete the


installation.

Software installation is complete. Continue with pressure transducer selection.

Part Two: Select the Pressure Transducer


1. Double-click the VISER shortcut icon on the desktop to start the software application.

Page 148
LABORIE System Installation Guide for Windows 7 OS

2. In the main menu bar click Patient and then click Preferences.
3. Select the Sensors Sets heading on the left side of the window.

add pressure
channel number
here

4. Apply pressure onto the pressure transducer and specify which channel is responding.
For example, if Channel 5 is responding to pressure placement then add the number 5
into the Pressure Channel text box in the Sensors Set 1 section.
5. Click OK.
6. Click OK in the resulting pop-up message.

The configuration is saved on the computer hard drive and will be used as the current setting
for all tests.

VISER Installation is complete.

Page 149
LABORIE System Installation Guide for Windows 7 OS

Enable Urostym Reports with Urocap IV (optional)

IMPORTANT NOTES TO READ BEFORE INSTALLATION:


 The Urocap-IV must use V12.0.10A or higher for Urocap-IV standalone with iList 1.2.2.9 or
higher.

 The Urostym must use Urostym software V6.3.0.10 with iList V1.2.2.9. Use of other versions of
Urostym software will not be compatible.

 Once the above conditions are satisfied, then you can use ONE hasp key programmed with
Urocap-IV and LAB596 Urostym software.

 When Urostym and Urocap IV are on the same system with the same HASP key, the
EMG/Uroflow feature in Urostym will not work with the Urocap IV.

1. Install the HASP Key driver. See page 88 for installation instructions.

2. Install Urostym software. See page 110. NOTE that only Urostym V6.0.3.0.10 is
supported for use with this reporting feature.

3. Install Urocap IV Standalone with iLIST V1.2.2.9 software. See page 104.

4. Run the installation utility as outlined below:

a. Double-click the EnableUrosytmReports 1.2.2.9 for Urocap-IV.exe application on the


LABORIE software disc to start the utility.

b. Click Run.

Page 150
LABORIE System Installation Guide for Windows 7 OS

c. Click OK and wait for the


confirmation message to
appear.

d. Click OK in the
confirmation message.

Installation complete.

Page 151
LABORIE System Installation Guide for Windows 7 OS

Upgrading UDS120 Software

Upgrade Software

1. Backup the following files: (this step is needed only to ensure no data is accidentally lost)
 C:\windows\uds120.ini file
 C:\windows\saturnDirectShow.ini file (if available)
 C:\CFG folder

2. Uninstall the following software through the Windows Control Panel :


 UDS
 i-LIST Office Reporter

3. If you are updating software on a "live" system and wish to preserve the current
software settings, then go to step 4; otherwise perform the following:
a. Please ensure the following files are removed from the C:\Windows folder
(Please note that some of the items might be already removed when uninstalling
the software in which case it is OK)
 uds120.ini
b. Please ensure the following folder is removed:
 C:\CFG
4. If you wish to remove the i-LIST Database, perform the following steps:
 Click Start and type "cmd" in the Run field.
 Type the following command (iLIST database will be removed and existing
iLIST data will not be recoverable after this step): OSQL -E -S .\SqlExpress
-Q "drop database [iList-Snapshot]
 Ensure no error messages are displayed. The cmd window should appear
as shown:

5. Run UDS installation.


6. Run iLIST Installation.

Page 152
LABORIE System Installation Guide for Windows 7 OS

NOTES REGARDING UPGRADING OF EXISTING V12.2.X INSTALLATION


i. Installation will not overwrite the CFG and DATA folder if path exists.
ii. Installation will not overwrite the uds120.ini file if path exists.
iii. Installation will create CFG, DATA, UDS120 and all Windows *.ini files and components
during clean install.

Files and Folders backup performed by Service: Note* After installation, the files and
folders, if applicable, should be restored if the Installation has overwritten the existing files
and folders mentioned below. See Upgrading of Existing Installation notes above.

Files and Folders remaining during Installation Removal 12.2.x:


 C:\CFG\*
 C:\DATA\*
 C:\Windows\uds120.ini
 C:\Windows\saturnDirectShow.ini
 C:\Windows\ilist.ini
 C:\Windows\saturn.ini
 C:\Windows\vbnwinloc.ini
 Camera file

Files and Folders removed during Installation Removal:


 C:\UDS120\*
 C:\Windows\ServersXMSB.ocx
 C:\Windows\ServersXB.ocx
 C:\Windows\ServersX.ocx
 C:\Windows\FornixMS.ocx (for Multiple Connections)
 C:\Windows\ServerO.exe (for Multiple Connections)
 C:\Windows\3dPlotAX.ocx
 C:\Windows\ServerXShell.dll (for Multiple Connections)
 C:\Windows\SaturnCOM.dll
 C:\Windows\UMSService.exe (for Goby)

Page 153
LABORIE System Installation Guide for Windows 7 OS

Part 4: System Finalization – Required for All


Systems
 Create a Restore Point (see page 155)

 Finalize AntiVirus Settings (see page 156)

 Turn off Automatic Updates (see page 156)

Page 154
LABORIE System Installation Guide for Windows 7 OS

Create a Restore Point


1. Click Start and right-click Computer, and then select Properties.
2. Click the System protection link on the left side of the window.

3. Click the System Protection tab and then click Create.

Page 155
LABORIE System Installation Guide for Windows 7 OS

4. You will be asked for a name. Give it a useful name that you will be able to easily
identify later.
5. Click the Create button. The system will start to create the restore point.

When finished, you will see a message saying “The restore point was created successfully”.

Finalize AntiVirus Settings


1. Run the Full Scan, but before that please check for any updates by clicking the Update button.

2. Run a full scan.


Tip: Since the full scan requires a few hours’ time, it’s best to have it run overnight.

Turn off Automatic Updates


 Click Start > Control Panel and click the Windows Update icon.
 Select the Change Settings heading in the upper-left side of the window.
 In the resulting window select the Turn off automatic updates option.
 Click OK and close the control panel window.

System installation and configuration complete. The system is now ready for use.

Page 156
LABORIE System Installation Guide for Windows 7 OS

Appendix
Troubleshoot Installation

If you see this…. Do this…..


“Error Code: 66a (Can't install KB2160841)”  Click Start > Control
Panel > Programs >
Programs and
Features
 In the resulting list,
right-click the
Microsoft .NET
Framework 4 Client
Profile file and select
Uninstall/Change.
 Select to Repair the
file.
 Wait for the repair to
complete before
continuing with
installation.
Copy and paste the
EncrytAES.dll file back into
C:\UDS120 folder on the PC
or laptop.

Page 157
LABORIE System Installation Guide for Windows 7 OS

Troubleshoot PDF Installation


If an ERROR message appears as illustrated below:

1. Open the Zip file (V6_License_Manager_x86.zip) from the Laborie Disc Build1XX\Driver and Utilities\
2. Extract LicenseManager.exe from the Zip file (V6_License_Manager_x86.zip).

IMPORTANT: Make sure you complete steps 3 to 5 otherwise you will receive an error when running
the alternative License Manager.

3. Right click LicenseManager.exe and select Copy from the context menu.
4. Open up My Computer on your PC and browse to the default location of your Nitro Pro installation:

Nitro Pro 8
* 32-bit: C:\Program Files\Nitro\Pro 8
* 64-bit: C:\Program Files (x86)\Nitro\Pro 8
Nitro Pro 6 or 7
* 32-bit: C:\Program Files\Nitro PDF\Professional
* 64-bit: C:\Program Files (x86)\Nitro PDF\Professional)

5. Once the above directory is open, right click and select Paste to move License Manager.exe into that
location.
6. Double click on the License Manager.exe file.
7. Enter your serial number into the License Number field (screenshot below).
8. Select Activate at the bottom of the screen (screenshot below)

Page 158
LABORIE System Installation Guide for Windows 7 OS

You should receive a Product successfully Activated! Message:

You should then be able to double click the Nitro Pro icon on your desktop.

Configuration fully complete.

Page 159