Rev. AB
Model 3095FT
Flow Transmitter
Model 3095FT
Pressure-159
Model 3095FT Flow Transmitter
3095-001AB
Greater meter flexibility with 100:1 rangeability
• Low power design that allows remote
installation with optional solar power system
FIGURE 1. Model 3095FT Flow Transmitter with
RTD Assembly.
The Model 3095FT Flow Transmitter is the world’s • Chart recorders have high maintenance costs,
Model 3095FT
most compact electronic flow measurement (EFM) including recommended monthly calibration,
device. Developed in conjunction with the Gas and replacement pens and charts.
Research Institute, the multivariable Model 3095FT • Charts must be integrated to calculate
is a high-performance extension of the Rosemount® flow volumes.
pressure family. In one package, this transmitter
measures differential pressure, absolute pressure, • Chart recorders are prone to losing data
and process temperature, as well as calculates flow. or recording inaccurate data—ink runs dry,
The three process variables along with the painted charts, vibration effects, and errors
calculated flow are available at all times in response due to moisture, temperature, drift, and
to a single request. mechanical wear.
• Integration is slow and cumbersome.
Model 3095FT communications occur on a two-wire
Billing, accounting, and auditing may take
system using the digital HART® (Highway
weeks, and inaccurate or late data may result
Addressable Remote Transducer) protocol, which is
in prohibitive penalties.
based on the Bell 202 Frequency Shift Keying (FSK)
standard. The Model 3095FT performs flow
With the Model 3095FT, chart recorders can now be
calculations and relevant data storage in nonvolatile
replaced with state-of-the-art measurement
memory per applicable American Gas Association
technology with minimal disruption to your business
(A.G.A.), American Petroleum Institute (API), and
methods. The Model 3095FT can be installed in
Gas Processors Association (GPA) orifice meter and
many types of flow monitoring and custody transfer
electronic flow measurement standards.
applications, ranging from monitoring flow at
remote locations, to a comprehensive system that
provides instantaneous flow, DP, AP, and PT via the
NEW SOLUTIONS FOR digital HART protocol signal.
GAS FLOW MEASUREMENT Advantages with the Model 3095FT include superior
Traditionally, mechanical chart recorders have been accuracy, thereby eliminating penalties and
installed in low-volume flow metering applications. reducing unaccounted-for natural gas, high
However, chart recorders do not provide the reliability, low maintenance, and no integration fees,
timeliness or accuracy needed to compete in today’s all resulting in lower long-term operating costs.
gas industry. Following are some of the problems
associated with chart recorders:
© 1995, 1996, 1997, 1998 Rosemount Inc.
May be protected by one or more of the following U.S. Pat. Nos. 4,370,890;
4,612,812; 4,791,352; 4,798,089; 4,798,089; 4,818,994; 4,833,922; 4,866,435;
4,926,340; 5,028,746; Des. 358,782. MEXICO PATENTADO NO. 154961.
Other U.S. and Foreign Patents Issued and Pending.
Pressure-160
Pressure-160
Rosemount Inc.
Introduced in 1994, the Model 3095FT Flow Transmitter New in 1996, the Model 3095FB Multivariable Transmitter
provides a field-proven solution for natural gas flow with Modbus Protocol provides a low cost measurement
monitoring and custody transfer applications. solution for monitoring natural gas flow.
• Accurately measures differential pressure, absolute • Accurately measures differential pressure, absolute
pressure, and process temperature (or gage) pressure, and process temperature
• Digital HART protocol communications • Provides RTUs (Remote Terminal Units) or flow
computers with real-time process variable information
• Performs flow calculations per American Gas
Association (A.G.A.), American Petroleum Institute • Transmission speed is configurable from 1200 to
(API), and Gas Processors Association (GPA) orifice 9600 baud
meter and electronic flow measurement standards • Easily integrates into existing SCADA applications
• User-configurable data logging that exceeds API • Capable of multidroping up to 32 transmitters on
MPMS Chapter 21.1 requirements the same RS-485 bus
• Most cost-effective EFM replacement for mechanical • For additional information, see Pressure-145
chart recorders
Model 3095FT
SYSTEM INTEGRATION
The Model 3095FT communicates via HART
protocol, which uses the Bell 202 Frequency Shift
Keying (FSK) technique. Communication is
accomplished by superimposing a high-frequency
signal on top of the fixed output signal. Because all
communication uses HART, the Model 3095FT does
not provide a 4–20 mA output signal.
The Model 3095FT meets the needs of a stand-alone
gas flow measurement instrument. The integration
of multivariable measurement, advanced flow
calculations, and audit trail capabilities into one
cost-effective, compact package make the Model
3095FT the ideal choice for orifice plate gas flow
measurement installations.
3095-DATAA22B
Remote communication is possible using cellular
phone, RF radio, satellite, and land-line
communications. In addition, Rosemount Inc. can
offer or recommend a variety of protocol converters
to convert HART into industry standard, or
proprietary protocols for integration into an existing FIGURE 2. Model 3095FT Single Site Metering Installation.
SCADA system. For more information on how the
Model 3095FT can be used to meet your telemetry
system requirements, contact your local Rosemount
Sales Engineer.
Pressure-161
161
Model 3095FT Flow Transmitter
MODEL 3095FT (1) The Model 3095FT sensor operating limits may limit the pressure
SOFTWARE FUNCTIONALITY and temperature range.
3095-30950005
installations, and automatically performs the
corresponding AGA calculations.
The Model 3095FT fast update and computation
frequencies contribute to its high overall flow FIGURE 3. Logged Variables Screen.
accuracy. Sensor updates occur nine times per
second and flow calculations update once per second.
Gross versus Detail Characterization Detail characterization covers all pressure,
The Model 3095FT calculates the gas temperature, and gas composition ranges for
compressibility factor using either gross or detail which A.G.A. computes compressibility factors.
characterization methods. Gross characterization is Table 1 identifies the acceptable ranges for both
a simplified method that is acceptable for a narrow of these characterization methods.
Model 3095FT
Analysis of the log file is a critical part of gas flow The Model 3095FT system includes two types of
measurement. Model 3095FT log files can be saved transmitter security to prevent unauthorized or
Model 3095FT
as an ASCII file or as a comma separated value file. accidental changes to the configuration. Mounted on
This information can then be printed as a text file or the electronics board is a transmitter security
retrieved using a spread sheet program. switch, which when enabled, prevents changes to the
transmitter configuration.
Model 3095FT logged data can also be imported into
Flow-Cal,™ an editing and graphing software In addition, the Model 3095FT User Interface
package developed by Coastal Flow Measurement Software provides three levels of password security:
Inc. This software provides edit, recalculation, and System Administrator (one password), Maintenance
sophisticated graphing capabilities. For further (three passwords), and Operation (six passwords).
information concerning this software package,
contact Coastal Flow.
USER INTERFACE
Configuration and Security
The Model 3095FT has extensive configuration SOFTWARE PACKAGES
capabilities, including gas properties, gas The Model 3095FT User Interface Software is a
composition, calculation methods, flow parameters, separate software package that allows easy
LCD display, and multiple audit trails. Figure 5 configuration of the Model 3095FT Flow
illustrates the Model 3095FT flow parameters Transmitter and retrieval of audit trail data. This
configuration screen. tool serves as a communications interface to the
Model 3095FT.
For best performance of the User Interface Software,
the following computer hardware and software is
recommended:
• DOS-based 386 computer or above
• 4 MB RAM minimum
3095-30950001
Pressure-163
163
Model 3095FT Flow Transmitter
Differential Pressure
In the differential pressure sensor, process pressure
is transmitted through the isolating diaphragm and
Meter Assembly fill fluid to the sensing diaphragm in the center of
the capacitance cell. Capacitor plates on both sides
Meter of the sensing diaphragm detect its position. The
Cover
differential capacitance between the sensing
diaphragm and the capacitor plates is directly
proportional to process pressure.
3051-3031C05C
Meter
Connection
Pressure-164
Pressure-164
Rosemount Inc.
HART
Communications
Piezoresistive Capacitive
AP Sensor DP Sensor
H L
RTD
Input Pressure Pressure Personal
3095-0181A
Computer
Model 3095FT
Absolute Pressure
The absolute pressure sensor is fabricated utilizing a The Model 3095FT can accept a signal from any
processing method called chemical vapor deposition 100-ohm platinum RTD that conforms to IEC-751
(CVD). This technique, which is superior to other Class B. The Model 3095FT Flow Transmitter can be
technologies that are vulnerable to drift over time, supplied with an optional Rosemount Series 68 or 78
isolates the sensing element from the silicon RTD temperature sensor. For further information on
substrate to achieve high accuracy and repeatability. Rosemount temperature sensors and accessory
hardware, contact your Fisher-Rosemount Sales
The absolute sensor consists of a Wheatstone bridge
Engineer.
circuit made from polysilicon resistors deposited on
a silicon substrate. The absolute pressure sensor is
hydraulically connected to the high pressure side of
Digital Compensation
the transmitter. Process pressure is transmitted The Model 3095FT uses a dedicated micro-processor,
through the fill fluid to the sensing element, creating located inside the sensor module, to linearize and
a very small deflection of the silicon substrate. The correct the raw sensor outputs. To ensure premium
resulting strain on the substrate changes the bridge performance, this sensor microprocessor uses the
resistance in proportion to the pressure applied. absolute (or gage) pressure measurement to
compensate for zero line pressure effects and an
Process Temperature internal temperature measurement to compensate
for thermal effects.
Process temperature is measured using an input
connection on the sensor module for a standard
resistance temperature device (RTD). Rosemount
Inc. offers a special shielded cable with connector
for connecting the RTD input to the Model 3095FT
(see ordering information for details).
Pressure-165
165
Model 3095FT Flow Transmitter
Certification Label
Housing
Terminal Block
Cover O-ring
Electronics Board
Cover
Nameplate
Module O-ring
Model 3095FT
Sensor Module
Housing Locking Screw
RTD Connector
Drain/Vent Valve
Coplanar Flange
Optional
Flange Adapters
3095-3095B08A.EPS
Bolts
Pressure-166
166
Rosemount Inc.
5.0 4.20
(127) (107)
½–14 NPT Conduit 4.09 Certification
4.3 Label
(110) Connection (104)
Meter Cover (Two Places)
(Optional) 2.15
(55)
0.75 (19)
0.75 (19) Clearance for
Clearance for Cover Removal
Cover Removal
Transmitter Transmitter
7.07
Circuitry Connections
(180)
This Side This Side
8.17
Nameplate (208)
Drain/Vent
Valve Housing
Rotation
Set Screw
6.4
(163)
3095-3095G05B, H05A
Model 3095FT
½–14 NPT on optional mounting adapters.
Adapters can be rotated to give connection ¼–18 NPT on Coplanar flange
centers of 2.00 (51), 2.125 (54), or 2.25 (57). for pressure connection without
the use of mounting adapters
NOTE
Dimensions are in inches (millimeters).
7.07
(180)
1.10 (28)
6.15
(156)
2.81
3095-3095J04B , K04A, L04B
(71)
6.25
3.54 (159)
(90)
4.73
(120)
NOTE
Dimensions are in inches (millimeters).
Pressure-167
167
Model 3095FT Flow Transmitter
pressure sensor.
Temperature Limits
Process:
Operating Region –40 to 185 °F (–40 to 85 °C).
250
0 Ambient:
7.5 11.0 35 –40 to 185 °F (–40 to 85 °C).
Power Supply Voltage
Storage:
Communication requires a minimum loop resistance of 250 ohms.
–50 to 212 °F (–46 to 100 °C).
Zero Suppression Humidity Limits
Can be set anywhere within the sensor limits as long 0–100% relative humidity.
as the span is greater than or equal to the minimum LCD Meter
span, the lower range value does not exceed the Optional dual-row, 11-digit, alphanumeric, scrolling
lower range limit, and the upper range value does liquid crystal display.
not exceed the upper range limit.
LCD Meter Temperature Limits
NOTE Operating: –13 to 185 °F (–25 to 85 °C).
Flow calculations will cease with negative DP readings. Storage: –40 to 185 °F (–40 to 85 °C).
Pressure-168
Pressure-168
Rosemount Inc.
Model 3095FT
±1.0 °F (0.56 °C) for 12 months.
(including Linearity, Hysteresis, Repeatability)
±0.075% of span for spans from 1:1 to 10:1 of URL.
Physical Specifications
For spans less than 10:1 rangedown, Electrical Connections
½–14 NPT, M20 3 1.5 (CM20), PG-13.5.
Accuracy = 0.025 + 0.005 ------------ % of Span
URL
Span Process Connections
Transmitter: ¼–18 NPT on 2 1/8 -in. centers.
Ambient Temperature Effect per 50 °F (28 °C) RTD: RTD dependent (see ordering information).
±(0.025% URL + 0.125% span) spans from 1:1 to 30:1.
±(0.035% URL – 0.175% span) spans from 30:1 to Process Wetted Parts
100:1. Isolating Diaphragms
Static Pressure Effects 316L SST or Hastelloy C-276.®
Zero error = ±0.10% of URL per 1,000 psi (6894 kPa). Drain/Vent Valves
Span error = ±0.20% of reading per 1,000 psi 316 SST or Hastelloy C.®
(6894 kPa). Flanges
Stability Plated carbon steel, 316 SST, or Hastelloy C.
±0.1% of URL for 12 months. Wetted O-rings
Absolute Pressure Glass-Filled TFE.
Range 3 (absolute) Non-Wetted Parts
0–8 to 0–800 psi (0–55.16 to 0–5515.8 kPa) Electronics Housing
(100:1 rangeability is allowed). Low copper aluminum.
Range 4 (absolute) Bolts
0–36.26 to 0–3,626 psi (0–250 to 0–25000 kPa) Plated carbon steel per ASTM A449, Grade 5;
(100:1 rangeability is allowed). or austenitic 316 SST.
Reference Accuracy Fill Fluid
(including Linearity, Hysteresis, Repeatability) Silicone oil.
±0.075% of span for spans from 1:1 to 6:1 of URL.
Paint
For spans less than 6:1 rangedown, Polyurethane.
Accuracy = 0.03 + 0.0075 ------------ % of span
URL O-rings
Span Buna-N.
Pressure-169
169
Model 3095FT Flow Transmitter
Weight ACCESSORIES
Component Weight in lb (kg) Model 3095FT User Interface Software Packages
Model 3095FT Transmitter 6.0 (2.7) The User Interface software package is available
LCD Meter 0.5 (0.2) with or without the HART modem and connecting
SST Mounting Bracket 1.0 (0.4) cables. All configurations are packaged separately.
12 ft (3.66 m) RTD Shielded Cable 0.5 (0.2)
12 ft (3.66 m) RTD Armored Cable 1.1 (0.5) Part No. 03095-5100-0001: Windows User Interface
24 ft (7.32 m) RTD Shielded Cable 1.0 (0.4) Software–Single PC License, HART Modem, and
24 ft (7.32 m) RTD Armored Cable 2.2 (1.0) Cables.
Part No. 03095-5110-0001: Windows User Interface
Software–Single PC License.
OPTIONS Part No. 03095-5110-0003: Windows User Interface
Standard Configuration Software– Site License.
Unless otherwise specified, the transmitter is Part No. 03095-5105-0001: HART Modem and
shipped as follows: Cables.
Engineering units: Remote Power Supply Options
Differential inH2O (Range 2) The Model 3095FT can be powered from a solar
Absolute/Gage psi (all ranges) panel/battery system. Although other systems can
Output: Digital HART protocol signal be used if they meet the power requirements of the
Flange type: Specified model code option 3095FT, Rosemount can supply the following
Flange material: Specified model code option systems:
O-ring material: Specified model code option
A/ ASP5127: Base System: powers one 3095FT
Drain/vent: Specified model code option
(5W solar panel, 7Ah battery).
Flow Configuration
Model 3095FT
Tagging
Three customer tagging options are available:
Rosemount and the Rosemount logotype are registered trademarks of
1. Standard SST tag is wired to the transmitter. Rosemount Inc.
Tag character height is 0.125 in. (3,18 mm), Coplanar and Multivariable are trademarks of Rosemount Inc.
85 characters maximum. HART is a registered trademark of the HART Communication Foundation.
Modbus is a trademark of Modicon, Inc.
2. Tag may be permanently stamped on transmitter Hastelloy C and Hastelloy C-276 are registered trademarks of Cabot Corp.
nameplate upon request. Tag character height is Windows is a registered trademark of Microsoft Corp.
Flow-Cal is a trademark of Coastal Flow Measurement, Inc.
0.0625 in. (1,59 mm), 65 characters maximum.
3. Tag may be stored in transmitter memory. Cover photo: 3095-002AB
Pressure-170
Pressure-170
Rosemount Inc.
ORDERING INFORMATION
Model Product Description
3095F Flow Transmitter
Code Output
H Digital HART Protocol Signal
Code Differential Pressure Range
2 0–2.5 to 0–250 inH2O (0–0.62 to 0–62.2 kPa)
3 0–10 to 0–830 inH2O (0–2,48 to 0–206 kPa)
Code Absolute/Gage Pressure Ranges
3 0–8 to 0–800 psia (0–55.16 to 0–5515.8 kPa)
4 0–36.26 to 0–3,626 psia (0–250 to 0–25000 kPa)
Code Isolator Material Fill Fluid
A 316L SST Silicone
B(1) Hastelloy C-276 Silicone
Code Flange Style, Material
A Coplanar, CS
B Coplanar, SST
C Coplanar, Hastelloy C
0 None (Required for Option Code S5)
Code Drain/Vent Material
A SST
C Hastelloy C
0 None (Required for Option Code S5)
Code O-ring
1 Glass-filled TFE
Model 3095FT
Code Process Temperature Input (RTD ordered separately)
0 Fixed Process Temperature (no cable)
1 RTD Input with 12 ft (3.66 m) of Shielded Cable (Intended for use with conduit.)
2 RTD Input with 24 ft (7.32 m) of Shielded Cable (Intended for use with conduit.)
3 RTD Input with 12 ft (3.66 m) of Armored, Shielded Cable
4 RTD Input with 24 ft (7.32 m) of Armored, Shielded Cable
7 RTD Input with 75 ft (22.86 m) of Shielded Cable (Intended for use with conduit)
8 RTD Input with 75 ft (22.86 m) of Armored, Shielded Cable
Code Transmitter Housing Material Conduit Entry Size
A Polyurethane-covered Aluminum ½–14 NPT
B Polyurethane-covered Aluminum M20 3 1.5 (CM20)
C Polyurethane-covered Aluminum PG 13.5
D Polyurethane-covered Aluminum G½
J SST ½–14 NPT
K SST M20 3 1.5 (CM20)
L SST PG 13.5
Code Terminal Block
A Standard
B With Integral Transient Protection
Code Meter
0 None
1 LCD Meter for Flow Units, Aluminum Housing
Code Bracket
0 None (Required for Option Code S5)
1 Coplanar SST Flange Bracket for 2-in. Pipe or Panel Mount, SST Bolts
Code Bolts
0 CS bolts
1 Austenitic 316 SST Bolts
N None (Required for Option Code S5)
Code Approvals
0 None
A Factory Mutual (FM) Explosion-Proof Approval
C Canadian Standards Association (CSA) Explosion-Proof Approval
H ISSeP/CENELEC Flameproof Certification
Pressure-171
171
Model 3095FT Flow Transmitter
{
Plated CS
DF Flange Adapters — Adapter Type Determined by Selected Flange Material: SST
Hastelloy C
Typical Model Number 3095F H 2 3 A B A 1 1 A B 1 1 0 A A
30535C2A
IN
© 1998 Rosemount, Inc. U. S. A. Fax 44 (1243) 867 5541 Tlx RS 61117 FRSPL
http://www.rosemount.com
¢00813-0100-4015• ¤
00813-0100-4015 Rev. AB 11/98
Pressure-172