Sei sulla pagina 1di 7

INSTALLATION GUIDE CPR Metronome Setting Program

First Edition: 11 Aug 2009


0614-903753A Second Edition: 30 Jan 2012 for AED-2100K/AED-2150 series
General You can download the Toshiba Bluetooth Stack from the following
web site. (As of October 2011)
This CPR metronome setting program enables the CPR http://aps2.toshiba-tro.de/bluetooth/?page=download
metronome function (100 times/minute) of the AED-2100K,
AED-2150C/K/L, AED-2151C/K/L and AED-2152C/K/L To use the Toshiba Bluetooth Stack, you need an end user license.
When you do not have the user license, you can download the test
automated external defibrillator. (Hereafter in this guide, referred
version of Toshiba Bluetooth Stack and test it for 30 days.
to as AED.)
If you use a Bluetooth driver and software other than the Toshiba
Hereafter in this guide, the AED-2150C/K/L, AED-2151C/K/L Bluetooth Stack for Windows, connection may be different. Refer to
the Bluetooth driver and software’s manual.
and AED-2152C/K/L automated external defibrillator is referred
to as AED-2150 series automated external defibrillator.

This installation guide provides the setting procedure.


Setting Overview
Following is the general setting procedure. For details, refer to
NOTE the following sections.
• Follow the procedures in this installation guide.
Otherwise, the AED might not operate. 1. Check the model and serial number of your AED.
• Before changing the setting, confirm that the AED battery 2. Add the AED to the Bluetooth Device List.
is fully charged. If power is lost while changing the 3. Change the AED Status to “Connected” on the Bluetooth
setting, the AED might not operate. Software.
• While changing the setting, check all the “Check Points”
4. Change the CPR metronome setting with this program.
in the procedure and write the results in the check sheet
5. Confirm that CPR metronome sound properly.
on the last page of this installation guide.
• Do not turn the AED power off while changing the setting.
This may damage the AED.
Procedure
Checking the Model and Serial Number
PC Requirements 1. Make sure that the model of your AED is one of the
To enable the CPR metronome setting of the AED-2100K/AED- following.
2150 series automated external defibrillator, a PC which meets • AED-2100K
the following requirements is necessary. • AED-2150C/K/L
• Operating system: Windows XP, Windows 7 • AED-2151C/K/L
• CPU: Celeron 1.2 GHz • AED-2152C/K/L
• Memory: 256 MB (512 MB is recommended) → Check point 1
• Hard disk free space: 500 MB
• Display resolution: 1024 × 768 dot 2. Check the AED serial number on the rear panel.
• Bluetooth module* → Check point 2

* If you use a PC which does not have a Bluetooth module, you need
a Bluetooth USB adapter to communicate with the AED. To set the
Bluetooth connection settings and install this program, you have to
install a Bluetooth driver and software before installing this program.

We recommend a Bluetooth USB adapter that can be used with the


Toshiba Bluetooth Stack for Windows. The Toshiba Bluetooth Stack
includes both the Bluetooth driver and Bluetooth Manager software.

The following versions of Toshiba Bluetooth Stack for Windows are


compatible with this program:
5.10.14, 6.10.10, 6.30.01, 6.40.00, 7.00.11, 7.10.04, 7.10.12, 8.00.12

1
Adding the AED to the Bluetooth Device List 3. Click the Custom Mode radio button and click the Next
To communicate with the AED, you have to add the AED as a button.
Bluetooth device in the Bluetooth Manager.

NOTE
• If your PC does not have a Bluetooth module, connect a
Bluetooth USB adapter to the PC.
• The following procedure is for when you use the Toshiba
Bluetooth Stack. If you use a different Bluetooth driver
and software, refer to the manual or online help for the
Bluetooth driver and software.
• When you do the following procedure, consult the facility
manager or administrator.

NOTE
1. Click the Windows Start menu and select Programs →
Make sure that Custom Mode is selected. If Express
TOSHIBA → Bluetooth → Bluetooth Settings.
Mode is selected, you cannot set the connection settings
correctly.

4. Select the AED from the list in the Bluetooth device area and
click the Next button.
Or, right-click the Bluetooth icon on the task bar and select
Bluetooth Settings from the pop-up menu.

2. Click the New Connection button on the Bluetooth Settings


window.

NOTE
• For the AED-2100K, the AED is displayed in the list
as “AED-2100K UR-0385 XXXXXX” (XXXXXX: serial
number of the AED main board).
• For the AED-2150 series, the AED is displayed in the
list as “AED-XXXX MAIN YYYYYY”. (AED-XXXX: AED
model, YYYYYY: serial number of the AED main board).
• Devices other than AED might also appear in the list.

The Bluetooth Manager – Bluetooth Security window


The Add New Connection Wizard window appears. appears.

2
5. Enter the 5 digit AED serial number (printed on the label on 9. Enter a connection name in the Connection Name text box
the back of the AED) in the Bluetooth Passkey (PIN) text and click the Next button.
box and click the OK button.

6. Select “Serial Port” in the Service selection list and click the
Next button.

We recommend a connection name that includes the COM


port number. This makes it easy to check the COM port
number afterwards.
Example: Device name displayed in the step 4 + COM port
number

10. Click the Finish button.

7. Check the “Use default COM port” check box and click the
Next button.

Changing the AED Status to “Connected” on the


Bluetooth Software
1. Open the AED lid to turn the power on.

8. Confirm the COM port and click the Next button.

2. For the AED-2100K, check that pads placement indicators of


the AED are lit.
AED-2100K

3
For the AED-2150 series, check that instruction is displayed 5. The icon for the selected AED changes and shows that the
on the LCD display. AED is connected to the PC.
AED-2150 series

This icon is
displayed when the
AED is connected to
the PC.

3. Click the Windows Start menu and select Programs →


TOSHIBA → Bluetooth → Bluetooth Settings.

Or, right-click the Bluetooth icon on the task bar and select
Bluetooth Settings from the pop-up menu. If you turn the AED power off after connecting the AED to
the PC, the Bluetooth Manager – Bluetooth Security window
appears. To connect the PC to the AED again, enter the 5
digit serial number printed on the back of the AED in the
Bluetooth Passkey (PIN) text box.

Applying the CPR Metronome Setting


1. Double-click the setup file.
For AED-2100K: AED-2100K CPRMetronome.exe
4. Double-click the AED in the list on the Bluetooth Settings For AED-2150 series: AED-2150 CPRMetronome.exe
window.
The setting program is started and the CPR Metronome
Setting Program window opens.

This icon is displayed


when the AED is not
connected to the PC.

Connection name that


you specified when
you added the AED.

Click the Next button.

2. Click the Connect button.

4
3. Select the COM port in the Port list box and click the OK • For the AED-2150 series, the battery status indicators on
button. the AED slowly blink and the following illustration is
displayed on the LCD display.

NOTE 5. Click the Next button on the CPR Metronome Setting


When you select the COM port, select the port that Program window.
you selected when you added the AED as a Bluetooth
device. Refer to “Adding the AED to the Bluetooth
Device List”. You do not have to change any settings
other than the COM port.

6. Check that the following information is displayed on the


CPR Metronome Settings Program window.
• Model: AED-2100K (for AED-2100K) or AED-2150 (for
AED-2150 series)
• CPR Metronome: 100 times/minute
• COM port:
COM port selected in step 3 in this procedure.
The Communication in Progress window appears and shows
→ Check point 3
the connection progress bar. When the AED is connected,
there is a voice message “Communication mode” from the
AED and the AED enters the communication mode.

4. Check that the AED is connected to the PC.

When the AED is connected to the PC:


• The Next button on the CPR Metronome Settings Program
window becomes available.
• For the AED-2100K, the battery status indicators on the
AED slowly blink and the pads placement indicators are
not lit.

Battery status indicators Pads placement indicators

5
7. Click the OK button to start installation.

While applying the setting, a progress bar appears on the


CPR Metronome Settings Program window.

NOTE
• The installation consists of two phases: uploading and
applying the setting to the AED. When applying (the
second phase of installation) starts, there is a beep
sound; this is not an error indication.
• Do not turn the AED power off during installation.

When installation finishes, the AED power automatically


turns off and the following screen appears.
→ Check point 4

8. Click the OK button.

9. Close the AED lid.

Confirming the CPR Metronome Sound


1. Open the AED lid to turn the AED power on.

2. Check that the CPR metronome sounds in both Adult and


Child modes at the rate of 100 times/minute.
→ Check point 5

Checking that the Check Sheet is Complete


When you finish installation, confirm that every item in the check
sheet on the last page is checked. Make a copy of the check sheet
and keep it for 6 years.

6
Check Sheet for CPR Metronome Setting Program
Customer Information

Customer name:
Install department:
Facility:
Address:

Installation Check Points

1. The AED model is one of the following. AED-2100K


AED-2150C AED-2150K AED-2150L
AED-2151C AED-2151K AED-2151L
AED-2152C AED-2152K AED-2152L
2. AED serial number: ( )
3. CPR Metronome setting is 100 times/minute. Confirmed
4. A “Setting applied” message box appears after installation. Confirmed
5. CPR metronome sounds in both Adult and Child modes at Adult mode
the rate of 100 times/minute. Child mode

Other Information

Installation comment:

Installed by:
Service company:

Copyright Notice
The entire contents of this manual are copyrighted by Nihon
Kohden. All rights are reserved.

Manufacturer European Representative


NIHON KOHDEN CORPORATION NIHON KOHDEN AMERICA, INC. NIHON KOHDEN EUROPE GmbH NIHON KOHDEN TRADING (SHANGHAI) CO., LTD