Customer Bulletin

Chemistry Systems: Miura ONE and all its OEMs

Date: November 22, 2012


Issue: Washing Station new settings
Author: A.Ballini

1. INTRODUCTION
Extensive testing on cuvette washing station process led us to modify pumps setting parameters in
order to improve the dispensation. The new setting adopted makes pumps running slower than
previously in order to prevent splashes and further contamination of the cuvette.
The washing station peristaltic pump setting for dispensing pumps must be changed with new
parameters:
• In case of systems using Gardner-Denver pumps, the “percentage of time” during
dispensing must be changed to 40% and the “First Dispension Time” must be changed to 7
[10/sec].
• In case of systems using new Welco pumps, the “percentage of time” during dispensing
must be changed to 30% and the “First Dispension Time” must be changed to 7 [10/sec].
In the following paragraph detailed instructions have been given.

2. WASHING STATION PARAMETERS SETTING.


After instrument Warming-Up is completed enter Service Software.
In the Washing Station panel parameters related to the section “Exempt Pumps” must be changed
as follows depending on two different cases:

Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-60242-00-A

Via delle Driadi, 45 – 00133 Roma Rev. A del 22/11/12

pag. 2 / 3
Customer Bulletin

a) Systems mounting older Gardner-Denver peristaltic pumps, see picture below for references,
considering only washing station pumps below indicated:

57ML SR10_30 PUMP ASSY 24Vdc - p/n 10060-01-A


This assembly is based on the Gardner/Denver pump p/n 20300348
38ML SR10_30 PUMP ASSY 24Vdc - p/n 10638-00-A
This assembly is based on the Gardner/Denver pump p/n 20300346

PERISTALTIC PERISTALTIC PERISTALTIC PERISTALTIC PERISTALTIC


PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE
“DIS5” “DIS3” “DIS4” “DIS1” “DIS2”

In this case following settings have to be used:

• %Time= 40
• First Dispension Time: 7 [Sec./10]

See picture below for further reference:

Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.
I.S.E. S.r.l.
CBL-60242-00-A

Via delle Driadi, 45 – 00133 Roma Rev. A del 22/11/12

pag. 3 / 3
Customer Bulletin

b) Systems mounting new Welco peristaltic pumps, see picture below for references,
considering only washing station pumps below indicated:

WPM1 57ML 2.5 24V Pump - p/n 11053-00-A


This assembly is based on the Welco pump p/n WPM1-P2.5(155)AA-WP
WPM1 57ML 2.0 24V Pump - p/n 11052-00-A
This assembly is based on the Welco pump p/n WPM1-P2(155)AA-BP

PERISTALTIC PERISTALTIC PERISTALTIC PERISTALTIC PERISTALTIC


PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE PUMP NEEDLE
“DIS5” “DIS3” “DIS4” “DIS1” “DIS2”

In this case following settings have to be used:

• %Time= 30
• First Dispension Time: 7 [Sec./10]

See picture below for further reference:

Content of this document is property of I.S.E. S.r.l. and neither the document nor part of it
can be reproduced, copied or disclosed without the authorization of I.S.E. S.r.l.
Unauthorized copies, total or partial, could cause legal actions in order to preserve owner’s interests.

Potrebbero piacerti anche