Abstract
This Handbook is a three-volume compilation of approximately
150 selected papers previously published by the staff of the National
Bureau of Standards on precision measurement, calibration, and
related subjects. It was prepared to meet the urgent need of newly
established standards laboratories for a “textbook” and reference
source in these fields. Volume I contains papers in electricity and
electronics; Volume II, heat and mechanics; and Volume III, optics,
metrology, and radiation. Each volume contains a complete index
of the entire Handbook by author, subject, and title.
[i
Foreword
The National Bureau of Standards is charged with the responsibility of
establishing and maintaining the national standards of physical measurement,
and of providing means for their effective utilization. This responsibility
carries with it the mission of providing the central basis for a complete, con¬
sistent system of physical measurement, adequate for national growth in
research and technology.
The recent tremendous increase in industrial activity, particularly in the
missile and satellite fields, has led to an unprecedented demand for precision
measurement, which, in turn, is bringing about the establishment of hundreds
of new standards laboratories. Many of these new laboratories must cover
the entire field of measurement, and must do so with a staff not previously
trained in work on standards of precision measurement.
To aid these laboratories in transmitting the accuracies of the national
standards to the shops of industry, the Bureau has prepared this three-volume
Handbook. It is a compilation of publications by the Bureau staff that have
been found of value to those who are establishing and operating new standards
laboratories. Omitted are some extended works, as well as a few shorter
papers that are otherwise readily available.
It is hoped that this compilation will serve both as a “textbook” and
a reference source for the many scientists and engineers who must be trained
in the shortest possible time to fill responsible positions in this critical area.
A. V. Astin, Director.
hi
Preface
Because of the urgent need for this Handbook, it has been reproduced by
a photoduplication process. As a result, the individual publications that make
up the compilation will be found to vary in such details as style, size of type,
and method of pagination.
Each paper reproduced for the compilation is essentially complete as
originally published and retains its original page numbering. All pages have
also been numbered in regular sequence throughout the three volumes. Thus,
the volume page number and the original page number are combined, for
example 100/10.
Because of the short time available, a complete review of each paper
included was not possible. However, some efforts were made to bring the
older papers up to date, and only those papers have been included that are of
current value. Nevertheless, users should be cautioned that the state of the
art may have advanced beyond that represented in some of the older papers.
These three volumes, extensive as they are, include only a fraction of the
published work of the National Bureau of Standards relating to standards.
However, many of the reprinted papers contain extensive bibliographies that
will enable the user who is confronted with a special problem to locate additional
information.
The papers that appear or are cited in this three-volume Handbook were
originally published over a period of several years as circulars, research papers,
chapters of books, and as articles in scientific and technical periodicals. Thus
individual copies of many papers are no longer readily available. More recent
Bureau publications, and in some cases the older papers for which prices are
given, may still be obtainable by purchase from the Superintendent of Docu¬
ments, U.S. Government Printing Office, Washington 25, D.C. Other papers
may often be obtained directly from the authors or from the publishers of the
Journals in which the papers appeared. The papers referred to in the various
lists, if not generally available as stated above, are usually available for reference
in technical, university, Government depository, and public libraries.
IV
Contents
Electricity
Page
Electronics
v
Page
Accurate radio-frequency micro voltages. _ - 634
VI
Selected Papers on
Electricity
(Contents on page V)
VII
UNITED STATES DEPARTMENT OF COMMERCE • Sinclair Weeks, Secretary
NATIONAL BUREAU OF STANDARDS • A. V. Astin, Director
Francis B. Silsbee
1/1
Preface
Since World War II there has been a very marked increase in the number
and scope of private standardizing laboratories throughout the United States.
The function of each such laboratory is to maintain the accuracy and uniformity
of the measuring instruments and apparatus used by the organization or
organizations that the laboratory serves. A basic phase in this process is
necessarily to correlate the reference standards of the laboratory with those of
the National Bureau of Standards. This Circular has been prepared to sug¬
gest techniques and principles that experience has shown to be useful in such
operations. Although this Circular covers explicitly only the field of electrical
measurements, many of the principles involved are equally applicable in other
kinds of measurement.
A. V. Astin, Director.
Contents
Page
Preface_ n
1. Introduction_ 1
2. General principles_ 2
2.1. Competence of personnel_ 2
2.2. Position in organization_ 2
2.3. Appropriate scope and accuracy_ 2
2.4. Laboratory atmosphere_ 2
2.5. Care of standards_ 3
2.6. Timing of calibrations_ 3
3. Reference standards_ 4
3.1. Standard cells_ 4
3.2. Resistors_ 4
3.3. Inductors_-_ 4
3.4. Capacitors_ 4
3.5. Volt boxes_ 5
3.6. Time standards_ 5
4. Working standards_ 5
4.1. Standard cells_ 5
4.2. Resistors_ 5
4.3. Capacitors and inductors_ 6
4.4. Volt, boxes_ 6
4.5. Indicating instruments__._ 6
4.6. Watthour meters_ 6
4.7. Instrument transformers_^_ 6
5. Comparison apparatus_ 6
5.1. Consoles_ 7
5.2. Direct-current bridges, direct-reading ratio sets, universal ratio sets, direct-
current potentiometers_ 7
5.3. Lindeck potentiometer_ 7
5.4. Voltage-calibrating transformers_ 7
5.5. Instrument-transformer test sets_ 7
6. Interlaboratory standards_ 7
6.1. Shipment_ 8
6.2. Standard cells_ 8
6.3. Resistors, capacitors, and inductors_ 8
6.4. Instrument transformers_ 9
6.5. Indicating instruments_ 9
7. Summary_ 9
8. References_ _ 9
2/11
Suggested Practices for Electrical Standardizing Laboratories
Francis B. Silsbee
A number of basic principles are given that experience has shown to be important in
the operation of private standardizing laboratories. Types of standard apparatus are
classified and schedules appropriate for their intercomparison and for their checking at the
National Bureau of Standards are suggested.
3/1
2. General Principles is not done, it becomes difficult to fix responsibility
for errors that may be discovered subsequent!}7,
The following general principles have been and the laboratory workers lose the basis for
found by experience to be applicable to many much of their pride in workmanship.
of the problems involved in the maintenance of
accuracy in laboratory work and any program 2.3. Appropriate Scope and Accuracy
of calibration service testing should conform to
The nature and accuracy of the work done in a
them so far as possible.
standardizing laboratory may vary considerably
2.1. Competence of Personnel with the size of the plant it serves and the nature
of the output of the plant. However, there are
The person in direct charge of any standardiz¬ very few modern activities in which some phases
ing laboratory must be thoroughly familiar with do not require measurements of rather high
the fundamentals of electricity and magnetism accuracy. A standardizing laboratory must, of
and with the principles of operation of all types course, be equipped for calibration measurements
of electrical measuring apparatus [8, 9]. To be of a higher accuracy than any used elsewhere in
competent to carry on within the laboratory the the plant, because the accuracy of measurement
various sequences of calibration testing outlined can never rise above that of its source.
below, he must have taken a college-level course Many such local standardizing laboratories
in electrical measurements or have had laboratory provide service in many scientific fields, e. g.,
experience equivalent thereto. It is very desir¬ mechanics, heat, optics, and chemistry, in addi¬
able that he be a graduate in physics or electrical tion to their standardizing work in electricity
engineering from an accredited college. He must and magnetism, which alone is the subject of
lie mentally alert to detect discrepancies and this Circular. The direction and extent of such
inconsistencies in internal check tests and to rec¬ activities, as well as the degree of specialization
ognize abnormal behavior of any part of the in particular branches of electrical science, may
apparatus or circuits. Unless persons of this greatly affect the types and ranges of the electrical
type are available in a standardizing laboratory, standards required.
both its calibration service and the measurement
work that depends upon it throughout the plant 2.4. Laboratory Atmosphere
is certain to become unsatisfactory, regardless of
how often particular pieces of apparatus may be Atmospheric conditions are of considerable
tested by other persons in some higher echelon importance in an electrical standardizing labora¬
or at NBS. tory. If the relative humidity exceeds 65 per¬
cent, a conducting film of moisture is adsorbed
2.2. Position in Organization
on the surface of many types of insulating mate¬
Experience through the years has shown that rial. Leakage over such surfaces, particularly to
the most satisfactory results are obtained by the circuit of a sensitive galvanometer, may intro¬
maintaining a definite separation of the standard¬ duce serious errors.
izing laboratory from the rest of the organization. Certain types of apparatus, for example some
Responsibilities for accuracy are thereby made wire-wound resistors of high value, show changes
definite, and team spirit can be developed in the of several parts per Iff1 with prolonged changes in
staff of the standardizing laboratory that keeps humidity. Such variations can be detected by
them interested in then- work, which might other¬ comparing calibrations made near the middle of
wise seem routine and unimportant. This sepa¬ the winter heating season, during which indoor
ration protects them from pressures that demand humidities are normally very low, with those made
improper haste and sloppy workmanship, and near the middle of the summer when the ambient
gives them authority directly delegated from a humidity is high.
high level in management. Most electrical apparatus is compensated for
Closely related to the foregoing is the valuable temperature to some extent. However, if the
principle that team spirit can be developed and ambient temperature is subject to changes, the
maintained by recognizing that the quality of response of the various components may not be
the reference standards of a laboratory are to a equally prompt. In such circumstances the com¬
considerable extent a measure of the competence pensation may be temporarily ineffective. It is
of its staff (“a worker is known by the sharpness therefore important that the laboratory tempera¬
of his tools”). Hence, in cases where a given ture be held fairly constant (±2 deg F) at all
large organization maintains a plurality of stand¬ times. Such room control also makes easier the
ardizing laboratories in its plants in different task of the thermostats that control the baths in
cities, it is very desirable that each reference and which more temperature-sensitive apparatus, such
interlaboratory standard should be assigned per¬ as standard cells or standard conductivity samples,
manently to an individual standardizing labora¬ are kept. The particular temperature chosen is
tory, and returned to it following periodic checks of less importance. The conditions specified by
at any laboratory of a higher echelon. If this the American Society for Testing Materials for its
4/2
“Standard Laboratory Atmosphere” are a relative of equipment to be tested, the delicacy of the ap¬
humidity of 50 ±2 percent and a temperature of paratus, and particularly the type of personnel
23° ±1.1 deg C (73.4° ±2 deg F). This combi¬ by which it is to be handled. (For measurements
nation is widely used in tests of dielectrics and made by inexperienced students or by military
other materials, and, in addition, is very close to recruits, the likelihood of damage is much greater
the maximum of comfort for the workers. Hence than when a laboratory is staffed by experts.)
it is a good target to aim at in designing a new In every laboratory a real possibility is always
laboratory. present that an abrupt change may occur in some
Freedom from vibration and dust are also im¬ standard as a result of an unrecognized accident
portant. A shop in which electrical instruments or abuse. Fairly frequent checks made locally
may be opened for repair should be scrupulously are therefore an indispensable supplement to any
clean and dust free. Magnetic particles are program of calibrations to a higher echelon.
attracted to the air gaps of permanent magnet Cooperative arrangements with a similar lab-
instruments and must be constantly guarded oratoiy in the same city may allow frequent checks
against. Effective air filtering or electrostatic to be made against an independently maintained
precipitation are very desirable. group of standards. This may permit a much less
frequent checking schedule between either labora¬
tory and one of higher echelon.
2.5. Care of Standards
In any laboratory, the accuracy needed in meas¬
urement may differ widely on different projects.
In many categories, a good standard piece of
Hence the frequency with which the working in¬
apparatus improves with age but deteriorates
struments should be checked may vary consider¬
with handling and shipping. The shipping of
ably from project to project, even though instru¬
primary and working standards and of comparison
ments of the same accuracy class are used [2, 6, 7],
equipment can be minimized and the transfer of
The time intervals suggested elsewhere in this
the units be achieved best by the use of interlabo¬ Circular for check tests represent in most cases
ratory standards of particular types that are as
the considered opinion of persons who have had
rugged as practicable.
long experience with standard apparatus of the
The adjustment of a piece of measuring equip¬ types now commonly used. Similar test intervals
ment to be close to its nominal value is very often are listed in the American Standards Association
desirable in the shop instruments used in the Code for Electricity Meters [2], When a new
plant, because of the great saving in time and re¬ type comes into use, it must be appraised by being-
duction in possible erroneous applications of cor¬ checked at frequent intervals during its first few
rections that result. On the other hand, the exact years of service. After its good stability has thus
opposite is usually true of working and reference been demonstrated, the interval between tests
standards. In general, such standards should not may safely be made longer. Several types of
be readjusted, even if their values have gradually standard capacitors and inductors are currently
drifted materially away from their nominal values. in this probational stage.
Any standardizing laboratory naturally maintains
A procedure that is often of value, particularly
records of the values of its reference standards [10].
for the checking of consoles and complex equip¬
These are preferably kept on an individual card
ment such as a-c bridges, is to use an interlabora¬
or record sheet for each standard, so that its his¬
tory standard that is measured in one laboratory
tory is apparent at a glance. Each such record
and shipped to a laboratory of lower echelon for a
provides a valuable indication of the quality of the
repeat measurement, the lower laboratory not
standard, and gives warning of deterioration and
initially being aware of the value obtained at the
the need for replacement toward the end of its
higher. This procedure can detect significant de¬
useful life. Frequent adjustments tend to inter¬ fects in the wiring or insulation of the test console,
fere with such records and increase the proba¬ improper procedure and careless operations at the
bility of improper corrections, as the proper cor¬ lower-level laboratory, as well as changes or de¬
rections before and after a slight readjustment terioration in the reference or working standards
may not be conspicuously different. Even more of the lower laboratory.
important is the risk that the adjustment may When systematic local intercomparisons indicate
initiate a progressive drift in value of the standard changes or abnormal performance in primary or
as a result of introducing mechanical strain, or working standards, checks with a higher echelon
local heating. are in order regardless of whether any formal
scheduled time for such a check has been reached.
2.6. Timing of Calibrations If the local periodic checks uncover unsteadiness
of operation, large sudden changes in value, or
The most desirable sequence of steps and the other indication of a constructional defect in the
frequency with which comparisons of measuring reference or working standard, this standard
apparatus should be made depend, of course, upon should be sent to its manufacturer for repair
a great number of circumstances, such as the volume before it is submitted to NBS for calibration.
5/3
3. Reference Standards For laboratories requiring an accuracy of not
over 0.01 percent in their standard of voltage, the
The function of the reference standards of a reference standard often consists of a group of
laboratory is to maintain locally a continuity of three cadmium standard cells of the unsaturated
value in the units of measurement that they type. The individual cells are intercompared
embody. Successive comparisons with the higher- weekly by connecting them bv pairs in series oppo¬
echelon standards of NBS by means of interlabora¬ sition and measuring the differential emf. These
tory standards shipped back and forth will give cells have a very small temperature coefficient and
an indication of any slow drift of the reference can be shipped safely by parcel post (if carefully
standard. When this cumulative drift exceeds packed). However, they are less constant than
the confidence interval of the values derived by saturated cells and their emf usually decreases at a
the comparisons with the interlaboratory stand¬ rate ranging from 40 to 120 microvolts per year.
ards, the value assigned to the reference standard Therefore, their values should be reappraised at
is corrected accordingly. The reference standards least annually on the basis of a periodic check from
are used primarily to calibrate periodically the a higher echelon. A second group of 2 or 3
working standards of a lower echelon, such as unsaturated cells can be used as an interlaboratory
indicating instruments. Also, on occasion they standard for such checks. When the emf of an
may be used as working standards themselves. unsaturated standard cell has fallen below 1.0183
It is evident that both reference standards and volts, it is approaching the end of its useful life
working standards should be as permanent and and can no longer be used as a reference or an
reliable in construction as possible. Definiteness interlaboratory standard. Following a moderate
and repeatability are of major importance, whereas change in temperature, many cells tend to show a
sensitivity, low losses, and freedom from extrane¬ very considerable change in emf, which may per¬
sist for several days. This thermal “hysteresis”
ous influences are relatively much less important.
Closeness of adjustment to nominal value is of must be guarded against by allowing cells to stand
decidedly minor importance. An undetected for some time after shipment before taking read¬
ings. Temperature troubles with unsaturated
change in a reference standard may easily initiate
standard cells can be minimized by keeping them in
a chain of error that will propagate throughout
a thermally lagged copper-lined box. This re¬
the plant and cause losses in time and material
exceeding manyfold the original cost of the stand¬ duces temperature fluctuations and differences in
temperature between the two electrodes. It is
ard. Hence, these reference standards should be
important that the leads be brought out through
purchased only on very strict specifications, and
very high grade insulation to an external terminal
from manufacturers of high reputation and of long
board. [13,14],
experience in producing shop instruments and
3.2. Resistors
apparatus of demonstrated permanence.
If the laboratory possesses at le&»c two standard
3.1. Standard Cells resistors of each decimal value covering the range
A large and important laboratory, which is over which it expects to make accurate measure¬
expected to need a standard of voltage with an ments, one of these can be submitted to NBS
accuracy of 0.002 percent and to certify the emf of every 2 years as an interlaboratory standard, while
unsaturated standard cells used in its plant or by the other resistor of each pair remains undisturbed
its subcontractors, normally maintains as a refer¬ as a reference standard in its laboratory.
ence standard a group of 5 or 6 cadmium standard 3.3. Inductors
cells of the saturated type. These cells are rela¬
tively permanent but have a temperature coeffi¬ Many laboratories find it useful to possess two
cient of about 0.005 percent per degree Celsius fixed standard inductors of each decimal value
(centigrade) at room temperature. They must over the range it expects to cover with accurate
therefore be kept in a bath thermostatted to measurements. In recent years the quality of
±0.01 deg C [11, 12], A highly refined, water- standard inductors has been greatly improved,
white, acid-free mineral oil, having a viscosity of and this Bureau does not yet have sufficient data
about 0.25 poise at 25° C and a flash point of i70° on the stability of the newer types to make a
C, has been found suitable for baths for standard definite estimate of their expected stability. It
cells and for resistors. Such a group of saturated is suggested therefore that one inductance standard
cells should preferably be checked initially at NBS of each denomination be submitted annually to
and its value reassigned annually for the first 3 the NBS, until a sufficient history is obtained to
years and biennially thereafter, by using a second predict its performance.
group of 2 or 3 saturated cells as an interlaboratory 3.4. Capacitors
standard. The individual cells of the reference
Reference standards of capacitance include small
group can be intercompared monthly. The refer¬ fixed air-dielectric units, precision-variable air
ence group is used to check the laboratory’s work¬ capacitors, and solid-dielectric capacitors using
ing-standard cells of the unsaturated type and, mica or an equivalent dielectric. Fixed standards
on occasion, shop-standard cells also. 1,000 picofarads (micromicrofarads) or less in
6/4
value must be of three-terminal construction in to errors approaching 1 percent for short periods
order to avoid uncertainties due to stray capaci¬ of time, even though the average frequency over
tances. Precision-variable air capacitors are pre¬ a longer time as shown by a clock is very high.
ferably hand carried, although shipment of the These short-time fluctuations in frequency are
capacitor in its wooden container within a padded materially less if the supply is tied in synchronism
carton is usually satisfactory. Variable capacitors with a large power system.
with noticeable backlash must be adjusted or
repaired before submission to NBS for calibration. 4. Working Standards
Capacitors with worm reduction gear are cali¬ The working standards constitute the principal
brated at the cardinal points corresponding to each tools of the standardizing laboratory. They are
whole revolution of the worm wheel. If the worm calibrated at intervals by comparison with the
is eccentric, capacitance increments between reference standards, and used in the daily work of
cardinal points will depart from linearity, but it is checking shop instruments. The number needed
usually not necessary to calibrate the vernier dial of any one kind and range will depend upon the
for every turn. volume of testing service demanded in that range.
Each laboratory should possess two capacitance If the volume is very small, and also in special
standards of each decimal value and type needed cases where extreme accuracy is needed, a refer¬
to cover the range of concern. After its initial ence standard can be used as a working standard
check at NBS, one standard of each value can be also.
kept as a reference standard while the other is 4.1. Standard Cells
submitted to NBS at regular intervals as an
Most laboratories will need a number of un-
interlaboratory standard. This interval should
initially be 1 year for any one pair of standards. saturated cadmium standard cells to serve as
After the first 3 or 4 calibrations, an examination working standards and relieve their reference cells
of th.e record will indicate the appropriate future of excessive use and of the hazard of accidental
abuse. Such working-standard cells should be
frequency of calibration, taking into account the
actual accuracy demands made on particular checked against a reference standard at intervals
standards. On .variable air capacitors the eccen¬ of 1 or 2 weeks.
4.2. Resistors
tricity correction ordinarily will be determined
only during the initial calibration; thereafter, Depending upon the nature of the work in the
calibration at a few cardinal points is usually laboratory, there will probably be required, in
sufficient. addition to sets of fixed standard resistors, a num¬
3.5. Volt Boxes ber of dial-type resistance boxes, perhaps including
resistors in the megohm and multimegohm ranges,
Each laboratory should have one volt box with a and also resistors or shunts capable of carrying
plurality of ratios, which can serve as a reference larger currents than are appropriate to the refer¬
standard. This can be checked initially at NBS. ence standards. These working standards may
Unless the laboratory is rather large, this same volt initially be tested at NBS to determine the effects,
box may be rechecked at NBS at intervals of 2 if significant, of current and temperature on their
years, and thus serve as an interlaboratory resistance. Thereafter these working standards
standard also. need to be checked, using a moderate current, at
3.6. Time Standards intervals of about 6 months by comparison with
the reference standards of resistance, using a
Although time is not an electrical quantity, direct reading ratio set, Wheatstone bridge, or
many electrical laboratories require reference double ratio set [15], or the potentiometer method.
standards of time or of frequency. A high-grade If it is believed that a resistor has been overloaded,
seconds pendulum clock with photoelectric pickup it should be checked against the appropriate refer¬
or a standard crystal-controlled oscillator may ence standard without delay, and if a significant
form the reference standard. "Either can be change in value has occurred since the last regular
calibrated by reference to the standard-frequency check, the overloaded resistor should be checked
radio signals emitted by NBS stations WWV or at frequent intervals until its resistance again
WWVH.2 * By the use of a multivibrator in com¬ becomes steady.
bination with the reference-standard oscillator, the
Most standard resistors are made of manganin.
frequency of working-standard oscillators can be
This alloy has the valuable property of showing
calibrated over a wide range. Such an oscillator
low thermal electromotive force to copper and of
can be used to control a standard frequency changing relatively little in resistance with change
circuit to which, in turn, synchronous timers can in temperature. The resistance, Rt, at a tempera¬
be connected to serve as working standards for ture t° C is related to that, i?25, at 25° C by the
measuring time intervals. The use of the fre¬ formula
quency of local electric-power circuits as a time
Rt=Bu{l + a(jt-25)+p(t-25)*}.
standard, while very convenient, may be subject
Here the coefficient a is usually less than 10X10-6
2For information on this service, consult Radio Standards Division,
National Bureau of Standards, Boulder Laboratories, Boulder, Colo. and /3 usually lies between — 3 X10-7 and — 6 X 10~7.
7/5
4.3. Capacitors and Inductors 4.6. Watthour Meters
Each laboratory should compare all of its work¬ Laboratories that have occasion to test large
ing standards with its reference standards once a numbers of watthour meters, as do those of power
year, and also immediately after an interlaboratory companies, are best guided by the Electrical
group has been calibrated at NBS. Working Metermen’s Handbook [1]. This book was pre¬
standards that are used frequently, or upon pared by a committee of experienced meter
which great dependence is placed, may be com¬ engineers for the instruction of meter-laboratory
pared with the laboratory reference standards personnel.
whenever an important series of plant calibrations If the laboratory has occasion to test watthour
is undertaken. meters only rarely, it is sufficient to have available
an electrodynamic wattmeter of suitable range
4.4. Volt Boxes
and a standard of time accurate to the degree
A small laboratory may find it sufficient to use needed in the energy measurement. The watt¬
its reference volt box as a working standard also. meter can be calibrated by using direct current
This can be checked initially and at intervals of and a potentiometer, and then used to hold a
2 years at NBS. It is well to measure and record known constant power while the revolutions of
the resistance of each section of each volt box the watthour meter are timed.
initially, and monthly thereafter, as a means for In intermediate cases it may be desirable to
detecting possible changes such as might be caused install a group of three working-standard watthour
by inadvertently overloading one of the lower- meters. These can be calibrated as indicated in
voltage ranges. More often other working-stand¬ the preceding paragraph, and each can be used to
ard volt boxes will be used as auxiliaries to poten¬ check a number of other meters. To minimize
tiometers for the calibration of both working- friction they should be used without register
standard indicating instruments and shop instru¬ mechanisms but with photoelectric pickups. The
ments. These can be compared monthly with the complete but time-consuming calibration against
reference-standard volt box by a null method [16]. the wattmeter need be made only at intervals of a
month or so, provided quick intercomparisons
4.5. Indicating Instruments among the three working-standard meters shows
no relative change in their rates. Provision can
Direct-current and alternating-current amme¬ be made by means of suitable precision current
ters, voltmeters, and wattmeters of either the transformers so that the standard meter can be
0.1-percent or the 0.25-percent class [17] will be operated always at about its full-load speed, even
needed for the range of current, voltage, and when the meter under test is at light load.
power over which the shop instruments used in
the plant are to be checked. The a-c instruments 4.7. Instrument Transformers
must be of the electrodynamic, electrostatic, or
electrothermal (thermocouple) types, which can Many laboratories will need a set of multirange
be used on direct as well as on alternating current current and voltage transformers covering the
[6, 8, 18], The usual a-c instrument of the range of current and voltage over which other
moving-iron type is not suitable as a transfer shop transformers are to be calibrated. An initial
standard. The working-standard indicating in¬ calibration at NBS, using the burden of the stand¬
struments can be submitted to NBS for an initial ard circuit of the transformer testing set (plus
test, and the a-c instruments for the additional ammeter or voltmeter), can be made at 60 cycles
determination, on appropriate ranges, of the per second. The errors of current transformers in
ac-dc difference by comparison with a transfer general are smaller at higher frequencies, but the
instrument. This ac-dc difference test should initial tests should include tests at. 400 or 800
cover the full range of frequency over which the cycles per second if the transformer is to be used
instrument is likely to be used. After this at such frequencies. Subsequent tests at NBS
initial test, they can be retained in the laboratory need be made only at intervals of 5, or even 10,
and checked in terms of a standard cell and years.
standard resistor, using a potentiomenter. Pref¬
5. Comparison Apparatus
erably, the frequency of these checks ranges from
2 weeks to 2 months, depending upon the fre¬ The term “comparison apparatus” includes
quency of use of the working standards and their equipment by means of which the calibration of a
reliability as indicated by earlier check tests. shop instrument or standard is checked by com¬
Direct-current working-standard indicating in¬ paring it with an appropriate working standard
struments, when built into a console, usually have of the laboratory. In many cases the working
terminals so arranged that the instrument and its standard is substituted for the device under test
range-extending resistors can be checked by using in the same circuit of the comparison equipment,
a potentiometer and a bridge. Alternating- and the change in its indication is taken as the
current instruments may have to be checked by measure of their difference. Such substitution
comparison with interlaboratory standard instru¬ methods are in general capable of very high ac¬
ments of a multirange type. curacy, and should be used wherever practicable.
In other cases a working standard is in effect built the bridge or ratio set to measure or compare
in as part of the comparison equipment. Ex¬ interlaboratory-standard resistors. A potentiom¬
amples of the latter are (1) the working standard eter can be given a rough check at least annually
indicating instrument built into an instrument at one point by using it to measure a standard
testing console and (2) the rheostat arm of a Tell of known emf or, alternatively, by using a
Wheatstone bridge when used directly (as con¬ single cell first as the standard by which to adjust
trasted with its use by substitution). the potentiometer current and then as the
unknown to be measured.
5.1. Consoles
5.3. Lindeck Potentiometer
These devices, containing appropriate sources,
adjusting transformers and rheostats, and panel Any laboratory will find very useful a combi¬
instruments for approximate adjustment, will pro¬ nation of a standard resistor and milliammeter
vide the circuits for the comparison of shop indi¬ (of the 0.25% class) to use as a low-range Lindeck
cating and recording instruments with working- potentiometer for the purpose of intercomparing
standard instruments [19, 20], In some types of the various standard cells in the laboratory by
console the working-standard instruments and measuring the differences between them in pairs.
tlieir auxiliary range-extending apparatus also are The resistor and milliammeter can then be
permanently built into the equipment [20]. If checked on the.same schedule as the working-
suitable special terminals are available, these standard resistors and the working-standard in¬
working standards can be calibrated like portable struments, respectively [12, 21, 22]. Care must
standards. If not, some secondary procedure must be taken to minimize thermal emf in such a circuit.
be set up by which the console in effect tests other
working standards that have previously been 5.4. Voltage-Calibrating Transformers
checked over the full range. The console can be The ratios of the voltages of the various tapped
shipped direct from the manufacturer to the lab¬ sections of the secondary winding to the voltage
oratory. The person in charge of calibration of the tertiary winding to which the standard
work at the laboratory can then satisfy himself, voltmeter is connected can be determined initially
by appropriate measurements of insulation re¬ at each operating frequency. This should be
sistance and of circuit resistance, that the connec¬ done with no load and with rated load on the
tions are correct and that the leakage and lead re¬ secondary while the standard voltmeter, or an
sistance are not such as to introduce errors. In impedance duplicating it, is connected to the ter¬
general, these particular hazards are less if the tiary. Subsequent tests need be made only at
leads can be run directly in the open between the intervals of 5 or even 10 years.
working standard and the instrument under test.
The working-standard indicating instruments 5.5. Instrument-Transformer Test Sets
are sometimes mounted at an angle of 45° for
The instrument-transformer test sets can easily
greater convenience in reading. This arrange¬ be checked annually at the 100-percent point and
ment, however, tends to introduce additional
also at one other point.3 These checks can be
pivot friction. 1 Care must also be taken to insure supplemented by using the test set and working-
that no ferromagnetic material is located near standard instrument transformer to measure once
enough to either the instrument under test or a year the ratio and phase angle of an inter¬
the standard instrument to affect its calibration. laboratory-standard instrument transformer of
Even nonmagnetic metal supports can cause the same range, which is checked at NBS every
trouble by providing eddy-current circuits which 5 years. These test sets are currently available
can affect unshielded a-c instruments. Stray for use at 60 and 25 cycles per second, only.
magnetic fields from supply transformers and Radical modifications in procedure or in compo¬
rheostats (particularly those wound on enameled nent values are required if they are to be used
steel tubes) must be eliminated. The familiarity at other frequencies.
that the supervisor will obtain by carrying on
this acceptance test forms an essential part of his 6. Interlaboratory Standards
training for the job. His success in it can be
verified adequately by the use of a group of indi¬ Interlaboratory standards in general are similar
cating instruments as interlaboratory standards in nature and inherent accuracy to reference stand¬
in annual tests. The local checks of insulation ards and working standards. The smaller labor¬
and of lead resistance should be repeated every atories may well use some of their reference stand¬
6 months. ards to serve as interlaboratory standards also.
These should, so far as possible, be of rugged
5.2. Direct-Current Bridges, Direct-Reading Ratio Sets, construction to minimize damage and change of
Universal Ratio Sets, Direct-Current Potentiometers
value in shipment. On this account they usually
These can be tested initially at NBS and in are standards of fixed value rather than contin¬
most cases need only to be resubmitted at inter¬ uously adjustable devices.
vals of 3 years. During the interim, local checks
3Simple methods for such one-point checks on instrument-transformer
can be made at intervals of 6 months by using test sets are now being developed at the National Bureau of Standards.
572772 0-61-2
9/7
The function of an interlaboratory standard is instrument. Portable standard watthour meters
to transmit some one of the electrical units of (rotating standards) should also be individually
measurement from a laboratory of higher echelon, packed.
such as the NBS, to the local laboratory. It is Certain heavy accessories used with instru¬
sent systematically to NBS for a calibration and ments, such as ammeter shunts, current trans¬
is compared before and after this operation with formers, and voltage (potential) transformers,
the appropriate reference standard of the local should be packed in separate boxes to avoid
laboratory. In those cases in which a large organ¬ possible damage to the instruments. Heavy
ization maintains a plurality of separate stand¬ pieces should always be shipped in wooden boxes
ardizing laboratories under its authority, a single and held in place, if necessary, by checks or cleats.
set of interlaboratory standards may profitably Large transformers, especially those having oil-
be circulated to give a “round-robin test” by filled iron cases, should be crated singly, and
being sent to 3 or 4 of the company’s laboratories arranged whenever possible so that the terminals
in succession between trips to NBS. Such a pro¬ can be made accessible for tests without removing
gram provides a check both on the comparison the entire crate.
apparatus and on the operating competence of The tops of boxes and crates must be marked
the personnel at the various laboratories. It also “This Side Up.” Boxes containing delicate instru¬
performs the function of maintaining the assigned ments should be marked “Fragile, Handle With
values of the reference standards of each laboratory Care.” Those containing any glass parts should
in concordance with NBS standards. be marked “Glass.” Failure to use such markings
6.1. Shipment precludes recourse in the event of loss or damage
Electrical measuring instruments such as am¬ in shipping.
6.2. Standard Cells
meters, voltmeters, wattmeters, and watthonr
meters contain extremely delicate jewels and A laboratory having lelerence-standard cells of
pivots, upon which the operation of each instru¬ the saturated type would logically provide itself
ment depends. These delicate parts must be with a group of about three saturated standard
carefully protected from mechanical shocks and cells- which can be sent to the NBS annually,
jars during shipment. Sensitive instruments will while the reference cells are new, but biennially
not arrive in satisfactory operating condition un¬ thereafter. Cells of this type must be kept
less great care is taken in packing. Every effort upright at all times and protected as far as possible
is made to handle and to repack these instruments from shock and temperature changes. This
carefully at the Bureau, and whenever possible the necessitates hand carrying and arrangements at
return shipment is made in the original container. each end for installing them in a thermostatted
Before each instrument is packed, all binding hath [11, 12]. Smaller laboratories having un¬
posts should be tightened, and any externally saturated reference standards may use 2 or 3
operated clamping device for the moving system unsaturated cadmium standard cells as inter¬
should be switched to the “clamp” or “transit” laboratory standards. These can be shipped by
position. Plugs and other small accessories should parcel post. Shipments in extremely cold or hot
be enclosed in a small separate container tied to weather should be avoided. If each interlabora¬
the instrument. Glass windows of instruments tory cell is compared with the cells of the labora¬
lacking protective cases should be protected by tory reference group before and after their trans¬
pieces of thin wood or heavy cardboard before port to NBS, a very desirable check is obtained
wrapping. Each instrument should then be on any changes that may have occurred during
wrapped in heavy manila paper or similar cover¬ transport. If any one cell shows the same value
ing and sealed with gummed tape to exclude dust relative to the local group, before and giter its
and excelsior. travels, it is highly probable that its emf did not
Boxes in which instruments are packed should be change and recover by an equal amount. If all
strong, preferably of wood, with screwed-on tops cells are unchanged, the probability of the com¬
to avoid damage to pivots or jewels, which may parison being valid is greatly increased.
be caused by a hammer or nail puller.
6.3. Resistors, Capacitors, and Inductors
Clean, fresh excelsior or its equivalent in special
packaging material should be used as the shock- Fixed standard resistors, capacitors, and in¬
absorbing material. A layer of excelsior at least 3 ductors, whether of the fixed-unit type or groups
to 4 inches deep, pressed down firmly, should of these combined in dial-type boxes, are satis¬
surround each wrapped instrument. Instruments factory as interlaboratory standards and can be
having pivoted components should be packed used at intervals of 1 to 2 years for comparison
upside down. with NBS, depending on the stability of the
High-grade pivoted instruments of the labora- reference standards that they serve. Intercom¬
tory-standard type, which have comparatively parison between the laboratories of a single-
heavy moving systems without clamping devices, organization at intervals of 1 year may prove
should be packed with special care and should useful in cases where the volume of testing at the
always be individually shipped in wooden boxes individual laboratories is large and the working
with 4 to 6 inches of excelsior around the wrapped standards at these laboratories are therefore uu-
10/8
usually liable to deterioration or accidental (4) The laboratory must accept responsibility
damage. The values of the interlaboratory stand¬ for the internal consistency of its measurements,
ards of this group should be such as to cover the and should look to a higher echelon (such as
range of measurements with which the laboratory NBS) only for its initial calibration and for
is concerned. periodic checks to detect drifts in the values of its
reference standards.
6.4. Instrument Transformers
11/9
UNITED STATES DEPARTMENT OF COMMERCE • Charles Sawyer, Secretary
NATIONAL BUREAU OF STANDARDS • E. U. Condon, Director
of the
Electrical Units
by F. B. Silsbee
12/1
Preface
January 1, 1948, marked the end of an era in the evolution of the science of
electrical measurements. On that date the system of so-called “international”
electrical units was abandoned, and a new system of measurement, using what
are commonly called “absolute” units, was officially instituted. It is the
purpose of this Circular to put on record an account of the working of the
international system of electrical units, with particular reference to the main¬
tenance of the international units in the 37 years (January 1, 1911, to January 1,
1948) during which the system was in its final form, to point out the trends of
development that made them obsolete, to record the official steps by which
they were superseded, and to describe briefly the methods used in the measure¬
ments that now form the basis for the new absolute units.
In this record many of the illustrative data will be quoted from the records
of the National Bureau of Standards, merely because such data were the most
readily accessible to the author. It is to be presumed that a similar examina¬
tion of the records in any of the other national laboratories would show a
similar picture.
E. U. Condon, Director.
Contents
Page
Preface_ n
I. Historical _ 1
If. Maintenance of Units_5
1. General_ 5
2. Maintenance of the ohm at the National Bureau of Stand¬
ards_ 6
3. Maintenance of the volt at the National Bureau of Stand¬
ards_ 9
4. Maintenance of units at the International Bureau of
Weights and Measures_■_ 11
III. International Comparisons 1911 to 1948_ 12
1. Comparisons of the ohm_ 12
2. Comparisons of the volt_ 15
IV. Developments 1911 to 1948_ 17
1. Enlargement of scope of International Committee on
Weights and Measures_ 17
2. Trend toward absolute units_ 18
3. Official actions leading to the adoption of the absolute
units_ 19
V. Absolute Measurements_ 22
1. General_ 22
2. Absolute measurement of resistance_ 23
3. Absolute measurement of current_ 27
VI. Summary_ 30
VII. References_ 30
VIII. Appendix 1__T_._ 31
IX. Appendix 2_ 32
X. Appendix 3_ 35
XI. Appendix 4___ 35
XII. Appendix 5_ 35
XIII. Appendix 6_^_ 36
XfV. Appendix 7_ 37
14/11I
Establishment and Maintenance of the Electrical
Units
By F. B. Silsbee
Abstract
I. Historical
The early workers in electrical measurements 1 absolute volt = 108 cgs electromagnetic units
set up arbitrary standards of resistance by speci¬ of electromotive force.
fying the use of pieces of wire of a definite material, 1 absolute ohm = 109 cgs electromagnetic units
length, and cross section (or weight), and used the of resistance.
Daniell (zinc-copper) cell as a standard of electro¬ 1 absolute joule = 107 ergs.
motive force. The work of Gauss (1833) and 1 absolute watt = 107 ergs per second.
Weber (1851) showed, however, the possibility These factors were given international status
of measuring electric and magnetic quantities in by the Paris Electrical Congress of 1881.
terms of mechanical units. In 1861 the British Experiments made under the auspices of the
Association for the Advancement of Science British Association Committee, and by others,
established a committee on electrical units and led to the assigning of numerical values to the
standards [1] 1 * with Maxwell as chairman, which resistance of wire standards and to the electro¬
for over 50 years played a leading part in the motive force of certain types of standard cells,
development and coordination of electrical stand¬ which at the time were adequately close approxi¬
ards. This committee recognized the fundamen¬ mations to the absolute values. Numerous copies
tal value of correlating the electrical and mechani¬ of this “B. A. Ohm” were distributed to labora¬
cal units and, in 1863, recommended the particular tories in several countries by the committee as a
set of decimal multiples of the centimeter-gram- means of securing international consistency.
second electromagnetic units that form the “ab¬ In the early use of electricity for communica¬
solute practical” units. In this designation the tion, the principal need for quantitative results
word “absolute” signifies that the unit is defined was in the field of resistance measurements, but
directly in terms of the mechanical units by a in the 1880’s, with the beginning of the applica¬
numerically simple relation. The work“practical’ tion of electric energy to light and power, the need
signifies that the unit is (except for the unit of for measurements of other electrical quantities
capacitance) of a size that is convenient in prac¬ became acute. Numerous power stations, fac¬
tical engineering work. The particular factors tories for building electric machinery, and elec¬
chosen are given in the following tabulation. trical laboratories were rapidly being established
1 absolute coulomb = 0.1 cgs electromagnetic unit in many countries. At that time there were no
of charge. national standardizing laboratories, and the pro¬
1 absolute ampere =0.1 cgs electromagnetic unit cedures for making accurate absolute measure¬
of current. ments were (as now) both difficult and time-
consuming. There was, therefore, a very definite
1Figures in brackets indicate the literature references at the end of this
need for more convenient and universally accepted
paper.
15/1
procedures by which the electrical units could be cell has electrodes of zinc amalgam and of mercury
reproduced in an isolated laboratory. To satisfy with an electrolyte of mercurous and zinc sulfates.
these needs, much experimental work was carried Its emf is 1.434 volts at 15° C. Somewhat later
on to develop appropriate reproducible standards, Weston (1893) [42] introduced the use of cadmium
and a number of international electrical congresses in place of zinc. This change eliminated a num¬
were held to promote their acceptance. ber of the difficulties experienced with the Clark
As a reproducible standard of resistance, the cell, and the Weston cell lias come into universal
most promising seemed to be the resistance of a use. The positive electrode is mercury and the
column of pure mercury of specified dimensions at negative an amalgam containing about 10 percent
a specified temperature. A pair of such “mercury of cadmium. A paste of mixed mercurous-
ohms” is shown in figure 1. Mercury is easily sulfate and cadmium sulfate crystals is placed over
purified and, at room temperature, has a much the mercury electrode, and the electrolyte is a
higher resistivity and lower temperature coeffi¬ solution of cadmium sulfate. Cadmium standard
cient than any other pure metal. One such cells are made in two types. In one of these,
standard—the old Siemens Einheit—had been the
resistance at 0° C of a column of mercury 1 m
long and 1 sq mm in cross section. An interna¬
tional commission, appointed in accordance with
a resolution of the International Congress of
Electricians, Paris, 1881, adopted as the “legal
ohm” a mercury column having a length of 106
cm and a cross section of 1 sq mm. This unit was
never officially “legalized” by any government
action, and by the time of the Chicago Congress
of 1893 a length of 106.3 cm was recognized as -SATURATED
---- SOLUTION
much closer to the correct value.
Asa reproducible standard of current, the work
of Lord Rayleigh, Kohlrausch, and others gave
promise that high accuracy might be obtained
with the silver voltameter, in which the current
is measured by the rate at which silver is deposited
electrolytically from a silver-nitrate solution on
the interior surface of a platinum cup. At a
meeting of the British Association in Edinburgh
AMALGAM
in 1892, a rate of deposition of 0.001118 g per
second under specified conditions was adopted as Figure 2. Cross section of a saturated cadmium standard
that corresponding to 1 ampere. cell.
A reproducible standard of voltage had been The primary reaction in such a cell is
o/Q jn
developed by Latimer Clark (1873) [41] in the Cd(2-phase amalgam)+Hg2S0j(s)+^^gCdS04 mH.20(l) = ^^j^CdSOi
type of standard cell that bears his name. This 8/3H20(s)+2Hg«).
16/2
shown in figure 2, the Cd S04 solution is saturated, trical measure in the United States.” See
and surplus crystals of CdS04.8/3H20 are present Appendix 1.
at any ordinary temperature. Such cells show a In the years that followed, national standardiz¬
very remarkable constancy of emf with time but ing laboratories (Physikalisch-Technische Reichs-
have a temperature coefficient of about —0.004 anstalt 1887, National Physical Laboratory 1899,
percent per degree C. It is this type that is used National Bureau of Standards 1901), were created
for all work of high accuracy, such as maintaining in the more highly industrialized nations. The
nr comparing the national units of emf. Often results of work at these laboratories made it clear
cells of the saturated type are made using 0.05- that, to the higher accuracy then desired, the
normal or 0.1-normal sulfuric acid. Such cells units of -the two sets recommended at Chicago
are also very permanent but have an emf that is were not equal in magnitude but differed by
lower by 30 nv and 60 /av, respectively. The amounts that were large enough to be measured
second, or unsaturated type, has such a concen¬ definitely, and which would, if ignored, cause
tration of CdS04 as to make the temperature confusion in future measurements.
coefficient of emf so small as to be negligible for Another unfortunate feature of the Chicago
all engineering purposes. This particular con¬ recommendations was the specification on inde¬
centration corresponds to saturation at 4° C. pendent bases of all three units (the ohm, the
This unsaturated cell is the type generally pre¬ ampere, and the volt). In the case of the absolute
ferred in ordinary laboratory work in the United practical units this is permissible, because an
States because of its independence of temperature. inherent feature of the cgs electromagnetic system
It is, however, less constant in emf. of units leads as a theoretical consequence to the
The term “Weston normal cell” is often used relation that an absolute volt is the difference of
to designate any individual cadmium standard potential produced between the terminals of an
cell of the saturated t}^pe described above. Such absolute ohm by the presence of an absolute
cells may differ from one another in emf by 10 or ampere. However, with independent definitions
20 mv. When written “Weston Normal Cell,’’ for all three units, Ohm’s Law becomes a relation
the term takes on a slightly different and almost to be experimentally determined, and its expres¬
mystical significance as a cell that embodies the sion requires a numerical coefficient, which, in
essential features, namely, emf and temperature general, will not turn out to be exactly unity.
coefficient, shown by the average of an infinite The obvious procedure to avoid such an awkward
population of Weston normal cells. In the light coefficient is to define only two units independ¬
of measurements reported below, an alternative ently in terms of separate standards. There was
definition for “The Weston Normal Cell” would general agreement that one of the two primary
be a standard cell the emf E, of which at a tem¬ units should be the ohm, but opinion was divided
perature of f°C is given in international volts by as to whether the other should be the ampere or
the equation the volt. The discussion [3] waxed violent, one
group emphasizing the (apparent) simplicity of
U=l.018300-0.0000406(t-20) the voltameter, while their opponents, pointing
— 0.00000095(4 —20)2+ 0.00000001 (f — 20)3. out that the unit had to be ultimately maintained
by a standard cell, urged the simple process of
In the literature many instances will be found in specifying the construction of a cell and thus
which the author has chosen to express the emf saving one experimental step.
of a particular cell or group of cells as differing After preliminary discussions at international
by a specified amount from the Weston Normal gatherings in St. Louis, 1904, and Berlin, 1905,
Cell as defined in this way. an Internationa] Conference on Electrical Units
The International Electrical Congress held in and Standards was held in London in October
Chicago in 1893 recognized the fundamental value 1908. The resolutions drawn up at this con¬
of the absolute units based on the cgs system of ference were in such form (appendix 2) as to dis¬
mechanical units and confirmed the decisions of tinguish clearly between (a) the “fundamental
earlier congresses as to the particular integral units,” ohm, ampere, etc., which are exact decimal
powers of 10 by which the “practical” units multiples of the cgs electromagnetic units, and
should be related to the cgs units. However, the (b) a “system of units representing the above and
Chicago congress also recognized the utility of the sufficiently near to them to be adopted for the
three reproducible standards and formally recom¬ purpose of electrical measurements and as a
mended the legal adoption of units of both systems basis for legislation . . . the International
as being mutually equivalent. Ohm, the International Ampere, and the Inter¬
In accordance with these recommendations the national Volt . . . .” The “international”
United States Congress passed, on July 12, 1894, ohm was defined in terms of a mercury column,
Public Bill No. 105, making the practical units, which was specified in full detail, and the “inter¬
based upon the cgs system and embodied in the national” ampere was defined in terms of the silver
reproducible standards, the “legal units of elec¬ voltameter, the specifications for which, however,
17/3
were incomplete. The “international” volt was differed by only 10 ppm (parts per million), the
then defined as “the electrical pressure which, German unit being the larger. Preliminary re¬
when steadily applied to a conductor whose sults of work then in progress at the National
resistance is one International Ohm, will produce a Bureau of Standards [13] were also in agreement,
current of one International Ampere.” The and the committee therefore “recommended to
volt thus became a derived unit and the standard all countries for general use [5] as the inter¬
cell became a secondary standard, the emf of national ohm the mean of the values of the units
which was to be determined in terms of the two realized at the PTR (Phvsikalisch-Technische
primary international units. Reichsanstalt) and the NPL (National Physical
The previous practice of setting up cells as Laboratory).”
standards was still strong in the minds of the The basis for the ampere was found to be much
committee so that as a sort of byproduct it stated less satisfactory. In the operation of the silver
in its Schedule C, “The Weston Normal Cell may voltameter one complication arises from the forma¬
be conveniently employed as a standard of electric tion at the silver anode of “anode slime” that
pressure . . . and gave fairly definite must be prevented from getting to the cathode
specifications for its construction. A provisional where it might cause an error in the weight of the
value of 1.0184 volts was assigned, but it was deposit. To avoid such an error a number of
expected that this value would be revised from arrangements have been used, two of which are
time to time as a result of future experiments. shown in figure 3. In the Rayleigh form of
As a sequel to the London Conference, a special voltameter the anode is wrapped in filter paper.
International Technical Committee met in Wash¬ In the Richards (or porous pot) form a porous,
ington at the Bureau of Standards in April and unglazed porcelain pot surrounds the anode and
May 191 (Mo carry out work that the Conference catches the slime while permitting the current to
had recommended. The three foreign members 2 flow in the electrolyte in the pores. In the early
of the Committee brought from their respective Kohlrausch form, a shallow glass cup is placed
British, French, and German national laboratories below the anode to catch the slime, and in the
standard resistors and cells that had been eval¬ later Smith form, a glass cylinder can be lowered
uated in terms of their national units.- They made at the completion of the run so as to fit tightly
careful intercomparisons of these coils and cells on the upper edge of the cup and prevent any spill¬
and those in use at the National Bureau of ing of material when the voltameter is dis¬
Standards and also carried out an elaborate series assembled. A siphon form, in which the two
of experiments with silver voltameters of various electrodes are in separate vessels, has also been
types. As a basis for the value of the ohm, it used.
happened that both the Reichsanstalt and the The work of the committee showed clearly
National Physical Laboratory had each recently that voltameters of various types might give
completed an elaborate program of comparisons of consistently different results. For instance, the
the resistance of mercury columns with wire deposits from solutions in which filter paper
standard resistors [11, 12]. The comparisons of was used were heavier by about 150 ppm. The
their standards at Washington showed that they committee was therefore “of the opinion that
the specifications for the silver voltameter should
2 The International Technical Committee consisted of Dr. E. B Rosa not be completed until further experiments shall
(NBS), chairman: Dr. W. Jaeger (PTR); Prof. F. Laporte (LCE); Mr.
F. E. Smith (NPL); and Dr. F. A. Wolff (NBS). be made . . . ”. A great deal of work was
18/4
done later, particularly at the National Bureau that had originated in the four different countries,
of Standards, [21] in response to this recommend¬ and which had had time to settle down to constant
ation. A number of sources of error were tracked values. A total of 104 cells were included, the
down and definite operating procedures for the number in a group ranging from 15 to 40 cells.
silver voltameter were formulated, but the speci¬ The average deviations for the emf of an individual
fications were never officially adopted. cell from the mean of its group was ±12 /xv, while
In spite of the incompleteness of the specifica¬ the groups deviated by ±5 nv from the mean of
tions for the voltameter, however, the committee the four. The voltameter experiments, excluding
used the results of the voltameter experiments, the most doubtful results such as those with
together with the values in international ohms filter paper present, gave for the final mean the
assigned to various coils, to derive a value for value of 1.0183 international volts.
the international volt and to assign values in On the completion of the work of the Inter¬
terms of this for the emf of the various standard national Technical Committee [5], each delegate
cells used in the work.
carried back to his respective laboratory its
A further result of the work of the International
standard coils and standard cells each with its
Technical Committee was a revision of the value
tentatively assigned in 1908 for the emf of the newly assigned value in terms of a common
Weston Normal Cell in terms of the International system of units. The system of “international”
Volt. The committee took as a value represent¬ electrical units was thus successfully launched
ing the Weston Normal Cell the mean of the with universal agreement on the magnitude of
separate group means of the four groups of cells the units as of January 1, 1911.
19/5
significant. The use of a fairly large number In May 1909, values of resistance were assigned
(n= 10 or 25) of standards in the primary group to a group of 1-ohm standards of this new con¬
contributes very materially to reducing the effect struction, on the basis of the unit that had been
of a drift in any one of them. carried until then by the seventeen open-type
coils. From May 1909 until October 1932 the
2. Maintenance of the Ohm at the National ohm was maintained at the NBS by a "primary”
Bureau of Standards group of 10 resistors selected from these 1-ohm
standard resistors of the Rosa sealed type. The
Following the legalization of the electrical units unit thus maintained is that which later came to
by Act of Congress in 1894, the Office of Stand¬ be designated as the "Ohm—certified NBS”.
ards, Weights and Measures, then part of the Comparisons made in 1910, in connection with
Treasury Department, was authorized in 1897 to the work of the International Technical Com¬
appoint a verifier. As fast as the very limited mittee on electrical units, showed that at that
appropriations permitted, standard resistors of time this unit was larger by 2 ppm than the unit
various denominations were purchased in Ger¬ then used by the PTR, and larger by 7 ppm than
many and standardized at the Pliysikalisch- the mean value of the British and the German
Teclmische Reichsanstalt. When the National units. This mean value was adopted by the
Bureau of Standards was set up, following the Committee for purposes of international uniformity
passage of its organic act in 1901, it took over and is commonly designated as the "Washington
this equipment and the personnel that bad been unit”. (It is this Washington unit that is repre¬
operating it. The unit of resistance during this sented in figure 8.) The slight difference of 7
period was therefore that of the German institu¬ ppm was entirely negligible in the practical con¬
tion. At that time resistance apparatus made in duct of ordinary electrical measurements at that
Berlin by the firm of Otto Wolff was recognized time, and therefore the "book values” assigned to
as being of the highest quality available, and the the several resistance standards on the earlier
Reichsanstalt, the oldest of the national labora¬ basis were not changed.
tories, was currently active in setting up mercury Between 1909 and 1932, careful comparisons of
columns. Resistance standards and other resist¬ the relative resistances of the entire group of
ance apparatus of Wolff’s manufacture were pur¬ 1-ohm coils were made at somewhat irregular
chased by the Bureau in considerable quantity, intervals, averaging G months. In computing the
submitted to the PTR for standardization in results of each intercomparison, it was assumed
terms of the unit of that institution, and then that the average resistance of 10 selected coils of
shipped to Washington to equip the new lab¬ the highest grade had remained unchanged since
oratory. the prior intercomparison. On this basis a new
In 1907 a group of four recently certified 1-ohm value was assigned to each of the coils of the entire
resistors of the Reichsanstalt type were received group. An examination of the data shows that
from Berlin. The values of these resistances the average difference (without regard to sign)
formed the basis for assigning values to a larger between the two values thus successively assigned
group consisting of ten 1-olim and seven 0.1-ohm to a resistor was only ± 1.3 microhms, and that in
standard resistors. The mean of this group was 95 percent of the cases the difference in the suc¬
assumed to remain constant from that time until cessive values was less than ±3 microhms.
May 1909, and thus carried the unit forward. In general, the same resistors were kept in the
In the meantime, experimental work at Wash¬ standard group from year to year unless there was
ington had shown that open standards, such as some reason for a change. Also, five of the stand¬
those of the Reichsanstalt type [31], underwent ards that at one time had been removed from the
a marked seasonal fluctuation in resistance, which, primary group were later used in it again, so that
in the case of some 1,000-ohm resistors, amounted during the whole period of 23 years twenty stand¬
to as much as 200 parts per million in 3 months. ards were involved. During the period, standards
This effect, reported by Rosa and Babcock [32], were removed from the primary group on 1 5 occa¬
was traced to seasonal variations in atmospheric sions. On six of these occasions there was no
humidity. The effect is much less with 1-olim suspicion of a change or defect in the standard, but
resistors but, to avoid these fluctuations, Rosa the removal was occasioned by the use of the
[33] designed a new type of standard resistor. In standard for some other purpose, frequently for
this the insulated manganin coil is enclosed in a shipment abroad for use in an international inter-
hermetically sealed metal container. The con¬ comparison. Of the remaining nine removals, one
tainer is nearly filled with refined mineral oil, and was occasioned by the development of a loose ter¬
the terminals are brought out through a hard minal, which obviously caused the hermetic seal
rubber top. A considerable number of standards to be broken; five were occasioned by an abnor¬
of this construction were manufactured both in mally large increase in the resistance of the coil
the instrument shop at the NBS and also by over its value at the preceding intercomparison.
Otto Wolff. These increases ranged from a minimum of 5 to a
20/6
maximum of 12 microhms. Three were occa¬ out through a shellac seal. A group of such
sioned by decreases in resistance, the maximum of resistors showed such excellent behavior that two
which was less than nine microhms. Two of the of them were used in the primary grdup in Oc¬
standard resistors remained in the primary group tober 1930 and a third in April 1932. Beginning
for the entire period, and six resistors were in the in October 1932, the use of the Rosa sealed stand¬
group for more than 15 years. ards in the primary group was discontinued, and
Examination of the data shows that the resist¬ 10 resistors of the Thomas type made with a coil
ances of five of the resistors drifted at a slow and diameter of 6 cm formed the primary group until
fairly steady rate over the period while they were March 1939.
in the primary group. The rates of drift of their During this period the average difference ob¬
resistances ranged from a decrease of 1 ppm per served between successive measurements made at
year to an increase of 4 ppm per year. One intervals- averaging 6 months was only 0.3
standard that was used for a short period showed a microhm, and 80 percent of these differences
decrease at a rate of 6 ppm per year. The coil were less than 0.8 microhm. A least square line
that approximated most closely to the mean adjusted to fit the data over this period of 6
showed a decrease of 0.3 ppm per year. Several years, for the worst of these resistors, showed an
coils showed a drift that changed from an increase average slope of less than 0.5 microhm per year
of 0.6 ppm per year when first included to a de¬ and a standard deviation from a straight line of
crease of 2 ppm per year after several years had only ±0.4 microhm. It is thus evident that this
elapsed. type of construction produces a standard resistor
A comparison of the individual readings made the relative stability of which is better by an
on one of the standards, which showed a nearly order of magnitude than that of the Rosa type.
linear decrease of 1 ppm per year, with a In 1933, work was completed on the construc¬
least square straight line adjusted through these tion of an additional group of Thomas-type
points shows a standard (root mean square) devia¬ standards of somewhat larger dimensions in which
tion of ±2.8 ppm. It is probable that the lack the coil had a diameter of 8 cm, with a correspond¬
of precision in the electrical measurements in¬ ingly greater cooling surface. These standards
volved in each intercomparison could have con¬ were compared with the smaller double-walled
tributed only a very small part of this variation. standards between 1932 and 1939, and from March
Most of it probably results from minor secular 1939 to the present time a group of 10 of the large
variations, and perhaps some from uncertainties Thomas-type standards has been used to carry
in the temperature of the standard itself. the unit forward. The average difference ob¬
Beginning in 1928, Thomas [34] had constructed served between successive measurements in in¬
a number of standard resistors of a new type dividual coils is ±0.2 microhm, and 95 percent of
sometimes referred to as “double-walled.” One the differences do not exceed ±0.5 microhm.
of them is shown in figure 4. In these the bare The slope of a least-square line passed through
manganin wire is annealed in a vacuum at a tem¬ the points observed, in the case of the worst of
perature of 500° C. or above, while wound on a these standards, is 0.1 microhm per year, and the
mandrel. It is then slipped on an insulated brass standard deviation of an individual point is 0.1
form, spaced with linen thread, tied down, and microhm.
fastened with shellac, and then baked for about One method attempted at the NBS as a means
an hour at 80° C. The form is tubular and con¬ for independently corroborating the maintenance
stitutes the inner wall of a container that is sealed of the ohm was the use of standards formed by
by soldering its end flanges to an outer, slightly wires of pure metals. It would not be expected
larger, tube. Dry air at about atmospheric pres¬ that such standards would drift at the same rate
sure fills the container, and the leads are brought and in the same direction as those of the metallur-
21/7
gically more complex manganin alloy. In 1933, a creased in resistance. The tendency to increase
platinum and a gold standard were made by was presumably the result of oxidation or corro¬
winding wires of these metals on insulated brass sion of the resistance alloy, while the tendency to
tubes and enclosing each in a sealed metal con¬ decrease was perhaps the result of a slow release
tainer. When in use the container was immersed of strains, or of more obscure metallurgical pro¬
in an ice bath. Because of the high temperature cesses. Of course, the use of a considerable num¬
coefficient of resistance of pure metals, a change of ber of standards in the primary group decidedly
0.00025 deg C will change the resistance by 1 ppm. increased the probability that the group should
This accuracy in temperature seems to be about contain standards drifting in both directions, but
at the limit to which the ice-point can be re¬ even with as many as ten standards in the group
produced. Measurements at intervals of about a the drift in the mean value might well have been
year, between 1933 and 1939, in which these pure much greater than it apparently was. The basis
metal standards were compared with the small for these self-congratulatory remarks is found, of
Thomas-type primary group, showed for the gold course, not in the data resulting from intercompar¬
coil an apparent decrease in its resistance of only isons within the group but from the international
0.13 ppm per year, but the platinum coil showed intercomparisons shown in section III, 1, figure 8.
an apparent increase of nearly 0.7 ppm per year. Both the German and the British laboratories
The standard deviation of either of these rates of used a smaller number (four) of standard resistors
drift is ±0.2 ppm per year. When used in 1947 and apparently were unfortunate in choosing coils
it was found that the insulation on both of these that tended to increase in resistance, but their
standards had deteriorated to such an extent that painstaking work with mercury columns enabled
precise measurements could not be made. It is them to detect these drifts before they became
doubtful, therefore, whether much weight should significant in the industrial field and to make cor¬
be assigned to the results from the coils of pure responding adjustments in their units. The vari¬
metals. ous determinations of the ohm in absolute units
Perhaps the best indication of the constancy of made prior to 1925 are probably not accurate
the ohm as maintained in Washington is found in enough to be of much significance in insuring its
the results of comparisons with the standards of maintenance but give no indication of any con¬
other national laboratories and with the mercury tinuous drift. As indicated below, it is probable
columns set up occasionally by these other insti¬ that from 1939 onward absolute measurements of
tutions. The situation is discussed in section III higher precision may prove to be very valuable
and summarized by figure 8. for this purpose as well as for the primary estab¬
It seems evident from the foregoing data that lishment of the unit.
the apparent constancy of the unit of resistance The comparisons of the 1-ohm standard resistors
as maintained between 1909 and 1932, while Rosa- have been made since 1918 b}^ use of the high-
type standards were used, was in considerable precision bridge shown in figure 5 [36]. It is
measure the result of good luck in the use of stand¬ used as a Kelvin double bridge and the main ratio
ards, some of which increased while others de¬ arms are of the Waidner-Wolff type [37]. The
In use, the whole structure is immersed in a temperature-controlled oil bath to a level above that of the intermediate insulating slab. All resistors and
contacts are therefore at a uniform constant temperature. It may be connected as a Wheatstone or as a Kelvin bridge. The nominal ratio may be 1:1 or 10:1.
The standards under test are connected through amalgamated contacts.
22/8
contacts are only in adjustable shunt circuits of and nearly equal increases on the new basis, and
relatively high resistance in parallel with the main four of which had been in the reference group in
resistors, which are connected permanently. The 1908, was therefore used instead to carry the unit
lowest dial corresponds to 1 ppm per step and the forward from 1910 and continued to serve (with
galvanometer is sufficiently sensitive to permit of two eliminations) as the primary group until 1914.
extrapolation to the next significant figure. Several changes were then made, and a group of
eighteen cells carried on until July .1919 when
3. Maintenance of the Volt at the National three were dropped and five added. These
Bureau of Standards 20 cells carried on, with the loss of four, until
April 1937 when nine cells of the 0.05-normal acid
The Chicago Electrical Congress of 1893 and type, which had been made in December 1932,
the resulting law passed in 1894 had specified were added. These 25 cells were in use through
1.434 volts at 15° C as the emf of the Clark cell. 1947.
A number of Clark cells were made up prior to 1906 As an indication of the mutual consistency of
at this Bureau using specially purified materials. the neutral, saturated cadmium cells, data taken
After correcting for temperature and subtracting between 1919 and 1933 on the primary reference
0.0003 volt to allow for the higher purity of the group showed that 18 of these cells had changed
mercurous sulfate used, a value of 1.42110 volts relative to their mean by less than 15 mv (micro¬
was decided upon as the mean value at 25° C to volts). Nine cells showed a net increase, while
be assigned to the new Clark cells. Some satu¬ the other nine had decreased, the average net
rated Weston cadmium cells, made in May 1906, change during the 14 years, taken without regard
were then compared with the Clark cells, and as a to sign, being ±5.0 nv. This corresponds to an
result of the measurements, a mean value of average drift rate of ±0.25 /xv per year, the
1.01890 volts at 25° C was derived as the mean of greatest mean drift'rate being 1 nv per year.
a group of 12 of the Weston cells. This group, Superposed on these net drifts the cells showed
with certain losses and replacements, served to slight fluctuations in emf. Successive values for
carry the unit until the close of the meeting of the the same cell measured after the lapse of about a
International Technical Committee in May 1910. year differed on the average by ±2 mv. The
The experimental work of this committee with precision of each comparison as indicated by
silver voltameters of various types fixed the Inter¬ repeat measurements made a few hours or a few
national Ampere. Earlier measurements in Eng¬ days apart was about ±0.1 nv, so that these
land and Germany on mercury columns had secular fluctuations must be attributed to random
served to fix the values of wire standard resistors changes in the cells. None of the differences
in terms of the International Ohm. The com¬ between the annual values exceeded 10 iuv, and
bination of these units by the committee in its 83 percent of them were less than 4 nv.
voltameter experiments, which involved the NBS In contrast to this behavior of “good” cells, cer¬
reference cells, led to the conclusion that the volt, tain cells after showing similar good behavior for
as then maintained at the NBS, was smaller by many years suddenly begin to drop in emf at a rate
852 microvolts than the “International Volt” as of from 3 to IQ mv per year. Since 1919 four such
derived from the International Ohm and the cases have developed and after the “delinquency”
International Ampere. Accordingly, effective of each cell was definitely established by the con¬
January 1, 1911, new values, smaller by this tinuance of its drift, the cell was removed from
amount, were assigned to all the cells, and the the primary group. The unit was then “recap¬
volt and the other electrical units derived from it tured” by going back to a time before the abnormal
were changed abruptly on that date by corres¬ drift began and computing forward again on the
ponding amounts. This change was announced basis of the remaining cells only.- This process
in Bureau of Standards Circular C29 [4], which resulted in introducing three abrupt increases of
contained a brief history of the subject and a 4.6, 0.8, and 1.1, mv, respectively, in the volt on
summary of the reasons for making the change. September 19, 1933, April 10, 1937, and Septem¬
As a byproduct of the work of the International ber 7, 1944. The changes in the group between
Technical Committee, the Bureau came into January 1912 and July 1919 resulted from shorter-
possession of a group of 36 cells, which had been period troubles, which involved three broken cells.
included in the large group used to assign a value These did not involve a shift in the unit.
to the Weston Normal Cell, and which included Of the 16 neutral cells in the group in 1947,
cells of British and German as well as American 13 had been in the primary group since 1914 and
origin. Until January 1912, this group was used five since 1908. The nine acid cells had been in
to maintain the unit of emf, and the mean of these the group since 1937.
36 cells was assumed to remain at 1.018300 volts Theoretically, the procedure of maintaining a
at 20° C. By that time it became evident that unit of measurement solely by the use of a large
some of the cells had drifted rather badly. A group of standards is subject to the fundamental
smaller group of 19 cells, which had shown steady limitation that a simultaneous progressive drift
23/9
of all the standards cannot be detected by the ±7 /jv from a mean difference of 24 p.v, the British
intercomparisons and will result, in an equally volt being the larger. The reason for the earlier
rapid drift of the unit. It is therefore important drift of 18 mv between 1911 and 1925 (that is 1.3
to seek external corroborative evidence of the MV/year) is not known.
constancy of the unit. Another line of evidence lies in the comparison
One such line of evidence derives from the con¬ of acid and neutral cells. The emf of a 0.05-
struction of new groups of cells from freshly pre¬ normal acid cell is about 30 mv lower than that of
pared batches of pure materials. One example of a neutral cell, but the tendency for the mercurous
this is a batch of six cells constructed in 1925 the sulfate to hydrolyze and for constituents of the
mean value of which was found to be greater in glass container to be dissolved by the electrolyte
terms of the 1925 unit by only 12 mv than the are definitely less. On the other hand, acid cells
mean of the large number of cells set up 15 years sometimes develop objectionable gaseous films on
earlier by the International Technical Committee. the amalgam electrode. Because of these chem¬
Very similar evidence results from the comparisons ical differences, it seems reasonable to expect that
shown below (see figure 9 section III, 2) between if either type of cell showed a characteristic con¬
the British and American volts over a long time. sistent drift in emf its rate would be somewhat
At the British National Physical Laboratory fresh different than that for cells of the other type.
groups of cells are made up every few years and Over the 7 years from 1937 to 1944, the average of
the older groups are discarded at intervals so that the nine acid cells (0.05 N), which had been made
their unit lias been carried forward by cells that in 1932, increased relative to the 16 old neutral
range in age from 1 to 20 years averaging about 6 cells, which were at least 24 years old, at a rate of
years. Since 1912 most of the American neutral
only 0.9 mv per year.
cells ranged in age from 6 to 41 years, averaging
A better check on drift can be obtained by the
over 20 years. Another difference arises because
the British laboratory uses 0.1-normal acid cells, use of the silver voltameter. Even yet there are
while until 1937 the American cells were all no authoritative specifications that enable the
neutral. Yet since 1925 the difference between voltameter to be used unambiguously as a means
the two volts has oscillated only over a range of for reproducing the international ampere. How-
The stirred bath is thermostatically controlled at 28±0.01° C. Connections are made through mercury cups under the oil to the platinum leads sealed
through the glass H-tubes.
24/10
ever, experiments repeated with a given type tional emf developed at its terminals. By an
under given conditions after the lapse of a number arrangement of slots in the covers of the dials, the
of years are adequate to reproduce the same value instrument serves also as a computing device, one
of current to a high accuracy. In 1931 a series of part of which is set to read the known emf of the
measurements using the Smith form of voltam¬ chosen reference cell, while the other part, when
eter were carried out at the Reichsanstalt by a balance is obtained, indicates directly the emf
Vinal [52], Vigoureux [53], and Von Steinwehr of the unknown.
[54]. These indicated that the volt as maintained The greatest source of error in such measure¬
at the National Bureau of Standards was the ments is the presence in the circuit of spurious
same in 1931 within 10 ppm as it had been when emfs, such as those from thermoelectric action.
similar experiments with the Smith form of The principal features in the design and operation
voltameter were made in 1912 at Washington. of the comparator are devoted to minimizing such
Hence the average drift could not have exceeded emfs. Each division on the milliammeter corre¬
0.6 ppm per year. sponds to 1 mv, and the galvanometer is suffi¬
Figure 6 shows the oil bath that contains the ciently sensitive to make readings to 0.1 division
primary group of cells. Its temperature is con¬ significant.
trolled automatically to 0.01° C. As a precaution
against loss of the unit by fire or other catastrophe, 4. Maintenance of Units at the International
two supplementary reference groups of 10 cells Bureau of Weights and Measures
each are housed in other buildings. Although
these groups are not held at a constant tempera¬ Additional information on the accuracy with
ture, they have shown little tendency to drift as which electrical units can be maintained at a
compared with the primary group. properly equipped laboratory is afforded by the
The intercomparisons of the cells in the primary performance of the reference groups of resistors
group are made by opposing the emf of the cell and standard cells kept at the International
being measured to that of a selected reference Bureau. The report by M. Romanowski [58]
cell and measuring the small difference in emf. lists the results of intercomparisons between 1939
Because the difference in emf is small, only a. and 1946.
moderate percentage accuracy in its determination Data are given on the four resistors that form
is required to give very high accuracy in the final the primary group, G03, and on five other stand¬
result. Since 1933 the instrument used in this ard resistors. Romanowski in his report has
work at the NBS has been the Brooks Standard made an extrapolation on the assumption that the
Cell Comparator [43] shown in figure 7. It com- the resistances of these standards were increasing
o
The cover and the left-hand milliammeter are adjusted to read the known emf of the reference cell. The main galvanometer (not shown) is balanced by
turning the central handle and adjusting the current in the right-hand milliammeter. When a balance is reached, the emf of the unknown cell is read from the
digits in the right-hand slot and the right-hand milliammeter.
prises a specialized form of potentiometer in which in ^alue by 0.95 ppm per year. As will be shown
the bulk of the emf to be measured is balanced below, such a drift was probably not real but was
against the IR drop in an adjustable precision a result of the assumption that the mean of the
resistor but the last two digits of the result are ohms of the six national laboratories had not
obtained by reading a milliammeter, which meas¬ drifted between 1936 and 1939. If his data are
ures the current in a small additional resistor in reduced on the alternative assumption that the
series in the circuit and thus indicates the addi- mean value of G03 remained constant over the
572772 0-61-3
25/11
7-year period, all the standards except the one Table 1 . Performance of standard cells at International
open-type coil (designated R 2836) show a very Bureau 1989 to 1946
26/12
that laboratory and are watched (preferably for [13]. The results were found to agree with the
some months) to be sure that no changes have Washington unit within the experimental error
been stimulated by the transportation. They are estimated for the work with the mercury columns,
then carried back to the first laboratory and again and therefore no adjustment was made in the
measured. If any differences are obtained be¬ American unit.
tween the first and last measurements it is assumed At the National Physical Laboratory, Smith in
that the change is a result of a gradual drift, and 1912 [14] and Hartshorn [15] in 1924 made care¬
an intermediate value is assigned by linear interpo¬ ful determinations with mercury columns and
lation to represent the value in terms of the units found their Washington unit to be greater than
of the first laboratory at the time when the meas¬ that derived from the mercury column by 16 ppm
urements in the second laboratory were made. and by 41 ppm, respectively, on the two occasions.
The difference between this value and the value In view of this, in the international comparisons
measured in the second laboratory then consti¬ of 1931, the National Physical Laboratory de¬
tutes a measure of the difference in the units of cided to abandon the Washington unit which it had
the two countries involved. The amount of the used until then and to use a unit smaller by 30
drift is, of course, an indication of the reliability ppm.
of the intercomparison and has ranged from Between 1923 and 1925 Von Steinwehr and
changes of less than 1 ppm, in some instances, Scliultze [16] at the Reichanstalt carried out ex¬
to values as high as 80 ppm in cases where the tensive determinations with mercury columns.
coils have been carried to many intermediate As a result, in the comparisons with the Japanese
laboratories and ultimately half around the world. standards brought to them in 1927 by Professor
Although, according to the decision of the Jimbo, and in the comparisons with British and
London Conference in 1908, the mercury column American standards in 1931 and thereafter, they
was by definition the ultimate standard of resist¬ used a new unit smaller by 33 ppm than the
ance, in actual practice relatively little use has Washington unit that they had used previously.
been made of it. At the National Bureau of The results of the work with mercury columns in
Standards mercury ohm tubes were set up and the three laboratories are indicated by the isolated
careful measurements made in 1911 and 1912 points marked “Hg” in figure 8.
Figure 8. Differences from the adjusted mean of the values of the International Ohm, as maintained by various national
laboratories, 1911 to 19j8.
Points marked “Hg” indicate the results of measurements on mercury columns.
27/13
Between 1932 and 1939 frequent and syste¬ per year. The NBS has seen no reason to change
matic intercomparisons were made by the staff its unit, and it is assumed that its rate of drift
of the International Bureau of Weights and was 0. The mean of these three drift rates is the
Measures to determine the relative values of the 1.1 X10~6 assumed above.
units maintained in a number of countries. Between 1932 and 1939 the work of MM.
Standards from France, Germany, Great Britain, Perard and Romanowski at the International
Japan, Russia, and the United States of America Bureau provides a much better basis to give
were used in almost all of these intercomparisons. continuity between successive intercomparisons.
Following World War II comparisons at the In 1935 the Advisory Committee on Electricity
International Bureau were resumed. Although recommended that there be taken as the mean
the standards from the various countries could not international ohm the mean obtained that year
be assembled in Paris simultaneously, by the of the units of Germany, Great Britain, Japan,
middle of 1948 data had been 'obtained on the Russia, and the United States of America. The
units of all six countries. French unit, which had been used in this and in
The results of these intercomparisons are shown some earlier intercomparisons, was recognized as
in figure 8. These data are, for the most part, being out of line with the others. Therefore it
taken from the proceedings of the International was not included, and the Laboratoire Central
Committee on Weights and Measures [51]. In d’Electricite (LCE) decided to adjust its unit by
this figure, for various epochs, the relative values decreasing it by 69.5 ppm to coincide with the
of the ohm as recognized by the various countries new mean. The Russian laboratory also read¬
have been plotted to a very open scale. Many of justed its unit by reducing it by 10.6 ppm. This
the apparent small changes in value of the national latter readjustment, however, was made after the
units may well be the result of errors introduced international mean value for 1935 had been fixed.
by the hazards of transportation. It is probable In plotting figure 8, this 1935 mean value has been
that the actual uncertainty of the electrical used as a reference corresponding to 0 on the scale
measurements by which any two standard re¬ of ordinates. In other words, it has been assumed
sistors are intercompared is usually less than 1 to be the same as the mean value chosen in 1910
ppm, and therefore not appreciable in the figure. as a result of the intercomparisons at Washing¬
It should be kept in mind that, in the very ton. In projecting the 1935 mean value forward
nature of the case, the location of the point indi¬ or backward in time it has been assumed that the
cating the value of the unit of a given country at mean of the units of all six countries remained
a later time can be related to the position of the unchanged. In plotting the points for December
corresponding point at an earlier date only by 1932 and July 1933, when data on the Russian
making some arbitrary assumption as to the con¬ unit were not available, it was assumed that the
stancy of some unit between the two dates. ratio of the Russian unit to the mean of the other
Therefore, too much significance should not be five nations was the same as in November 1933.
given to the slope of the lines, which, in the dia¬ It will be noted that this procedure of equating
gram, have been arbitrarily drawn to connect the mean international ohm of 1935 with the mean
successive points. In the plotting of figure 8, Washington unit of 1910, when combined with the
for convenience, three different conventional pro¬ rate of drift postulated during the earlier years,
cedures have been used to relate the data at suc¬ gives a very satisfactory correlation between the
cessive epochs, as follows. 1931 data on the older basis and the 1932 data on
From 1910 to 1932 only data derived from meas¬ the newer basis. This correlation would seem to
urements at the NPL, PTR, and NBS were used justify the apparently arbitrary assumption of
to carry forward a common base from epoch to drift used in plotting the earlier data.
epoch. This convention was chosen because data When the international comparisons of the ohm
from other laboratories were very fragmentary. were resumed, following World War II, a question
The mean of the three values of the Washington arose as to the proper basis for carrying forward
unit as maintained during this period at the three the mean international ohm from 1933 to 1947.
laboratories has been assumed to have drifted The problem was unfortunately complicated by a
upwards at the rate of 1.1 X10-6 per year. The strange coincidence that illustrates the importance
justification for this somewhat arbitrary assump¬ of having some absolute standard as a fundamental
tion is as follows. In 1931 the NPL, as a result basis for preserving units of measurement. The
of earlier measurements on mercury columns, staff of the International Bureau had originally
decided that its Washington unit had increased assumed that the Mean International Ohm was
by 30 ppm in the course of the 21 years since the the same in 1939 as it had been in 1936. On this
unit was established at Washington. This cor¬ assumption it appeared that the resistances of the
responds to a drift of 1.4X10-6 per year. Simi¬ coils in their reference group G03 were increasing
larly, the PTR, in 1927, decided that its Wash¬ at the rate of 0.95 ppm per year. An extrapolation
ington unit had drifted upwards by 33 ppm in of this rate over the war period would predict an
the preceding 17 years, or at a rate of 1.9 X10-6 accumulated increase of 6.6 ppm by 1946. The
28/14
comparisons in 1946 between G03 and some which is several times as large as that of a manga-
American standards showed only a very slight nin resistor. Also, it is the net difference of the
increase (0.9 ppm) in the American unit relative separate temperature coefficients of the potentials
to G03. If this were combined with the predicted developed at each of the electrodes. Each of these
drift in G03 it would imply an increase of half-cells has a temperature coefficient of over 300
6.6+0.9 = 7.5 ppm in the American unit. Strangely ppm per deg C but the two are of opposite sign.
enough, comparisons in 1946 between British and Consequently, temperature gradients in the en¬
American standards indicated that the provisional closing oil bath may introduce serious error. Most
value of the British ohm was then 19.3 ppm smaller national laboratories maintain their standard cells
than the American ohm, although in 1939 it had at about 20° C, while some foreign laboratories
been only 3.3 ppm smaller. A splitting of the make measurements at temperatures as low as
difference in this shift would imply an increase of 16° C, but the climate at Washington requires
8 ppm in the American unit and an equal decrease that the bath be held at 28° C if condensation of
in the British. In spite of the apparent agreement moisture is to be avoided. The situation is fur¬
between these two figures 7.5 and 8 ppm, such an ther complicated because the temperature coeffi¬
increase in the American ohm seemed improbable cient of a standard cell varies somewhat with its
in view of the high quality of the standards in the age and perhaps with its acidity.
American group. The National Physical Labora¬ As one indication of the probable precision of
tory therefore made further comparisons of its such international comparisons, it may be noted
reference group with other British standards of that in 1929 six comparisons were reported, each
high quality and found a number of divergencies based on a different group of cells. The standard
which led them to “recapture” their unit by using deviation from the mean value of the six observed
as a basis for its maintenance other coils of prob¬ differences between the units at the NPL and the
ably better stability. On this final basis the NBS was 4 ppm. Among these comparisons were
British ohm in 1946 was larger than the provisional those involving Russian and Japanese cells, which
value by 18.2 ppm. If as an alternative basis the had been transported long distances, and which
international reference group G03 is assumed to had been handled many times in different labora¬
have remained constant from 1939 to 1946, it tories. In seven different years, comparisons
follows that the British and the American ohm between the same two laboratories were made by
have increased by only 4.8 and 0.9 ppm, respec¬ using more than one group of cells. A second
tively [57]. This latter basis has therefore been measure of the accuracy of these comparisons is
accepted by all concerned and was used in plotting obtained by noting for each of these seven years
the 1947 value in figure 8. the highest and the lowest value of the observed
ratio of the units of the two countries. The differ¬
2. Comparisons of the Volt ence between each pair of extreme values indicates
the uncertainty for that year. The average differ¬
In general the procedures used in the mainte¬ ence for the seven years was 9 ppm. In contrast
nance and in the intercomparisons of the volt at with such differences that appear when transporta¬
the various national laboratories have been similar tion of cells is involved are the results of work done
to those used for the ohm. For many years the in 1931 [53] at the Reichsanstalt, when a group of
comparisons were rather sporadic, and they have eight British cells was compared with a group of
been interrupted by the two World Wars. With 10 American cells on eight different days. The
the entry of the International Bureau of Weights observations were scattered over an interval of 5
and Measures into the field of electrical measure¬ weeks, but the cells were not moved between
ments, the intercomparisons became much more measurements. On only two of the eight days
frequent and systematic. The records of many of did the difference in the units reported by the two
the earlier intercomparisons, as well as of the groups differ by 1 ppm from the mean value of
later ones, will be found in the Proces Verbaux of 30 ppm; on one day the difference was 29, and on
the International Committee for Weights & another, it was 32.
Measures and in their various Annexes. The results of the various comparisons of the
The standard cell is a much more delicate device volt through the period from 1911 to 1948 are
than a standard resistor, and therefore more sus¬ shown in figure 9. In this figure, the plotted
ceptible to changes during shipment. Conse¬ points show at a given epoch the relative values
quently, in general, groups of from 4 to 10 such of the volt as maintained by the various national
cells have been used for important intercompar¬ laboratories. The correlations between the values
isons. For these precise comparisons, as well as at one epoch, and those at another, have been
for maintaining the value of the unit in the national made on the basis of the following somewhat
laboratories, cells of the “saturated” type are used. arbitrary assumptions. From 1911 to 1932,
This practice introduces some difficult problems inclusive, it has been assumed that the mean value
of temperature control. The temperature coeffi¬ of the units at the NPL and NBS was constant,
cient of such cells is about 40 ppm per deg C, and this mean is made the axis of abscissas. The
29/15
INTERNATIONAL VOLT
Figtjke 9. Differences from the mean of the values of the International Volt, as maintained by various national laboratories,
1911 to 1948.
reason that data from the PTR were not included resistors, the values of which had been carefully
is, first, that during the early years relatively few determined at their respective laboratories. In
comparisons were made with the German labo¬ Berlin, these standards were compared with those
ratory, and, second, that hi 1931, the PTR recog¬ of the PTR and an extensive series of silver de¬
nized that its standard had drifted by 82 ppm. It posits were made using several types of vol¬
would be difficult to decide when the major part tameters. The results have been reported by all
of this drift occurred and to make a reasonable three laboratories [52, 53, 54.] Following this
correction for it in the graph. After 1934, the work the PTR increased its unit by 82 ppm.
data in figure 9 have been plotted on the assump¬ The large change in the Russian unit between
tion that the “Mean International Volt”, Vm, as 1929 and 1934 was presumably the result of ex¬
defined in 1935, by the Consultative Committee tensive work with the silver voltameter reported
on Electricity [56], has remained constant. Dur¬ by Kolossov [22],
ing World War II, this is equivalent to assuming Comparisons at the International Bureau began
that the large group of standard cells maintained in 1932 [55] with measurements on standard cells
at the International Bureau and originating in from four countries, and in December 1934 data
various laboratories has also remained constant were obtained for cells from all six national labor¬
in emf. atories. The LCE, finding its unit to be out of
In 1931 it became evident tnat there was an line with the others, increased its volt by 76 ppm
increasing discrepancy between the unit as main¬ in 1935. The Consultative Committee on Elec¬
tained at the PTR, and that at the British and tricity suggested [56], in view of the very small
American laboratories. Consequently, at the differences between the units of the other five
invitation of the Reichsanstalt, arrangements were laboratories and the mean of the units, that each
made to have* an international committee carry should, if possible, adopt this mean value as its
out experiments with the silver voltameter, and, unit and thus obtain an immediate international
thus, to reestablish the international volt. P. unification. However, only the Russian labora¬
Vigoureux of the NPL and G. W. Vinal of the tory followed this suggestion and reduced its volt
NBS took to Berlin standard cells and standard by 13 ppm. In the intercomparisons of 1937 and
30/16
1939, the International Bureau has carried for¬ Between July 1945 and May 1948, standard
ward the mean value of VOT as defined prior to the cells were received at the International Bureau
from five of the six national laboratories (The
change in the Russian unit by assuming the mean
Reichsanstalt did not submit any). The results
of the six national units to be the same (that is, are plotted in figure 9 on the basis that the mean
1— 2.1X1CT6) as it was in 1935, after the Russian emf of the reference .group GV2 of 47 cells at the
adjustment. International Bureau has remained constant.
31/17
these objects it recommended the formation of resolution on the mercury ohm, he could not
an Advisory Committee on Electricity that refrain from saying, “Before putting the resolu¬
should advise the International Committee on tion, I should like to say a word or two. I have
questions relating to electrical standards and no doubt this proposition will commend itself to
systems of measurement. This recommendation a large majority of the Conference, but in my own
was approved by the Seventh General Confer¬ mind I feel some doubt as to whether the intro¬
ence held in 1927. The Advisory Committee on duction of the mercury column is not what we
Electricity is limited to 10 members, including a call a fifth wheel to the coach ... to define it
representative appointed by each of the national (the ohm) as the resistance of a column of mercury
laboratories designated by the International seems to me—I will not say illogical, but hardly
Committee, and also specialists named by the in accord with the precision that absolute measure¬
Committee. The chairman of the Advisory ments made now obtain ..." I only wish to
Committee is elected by the International Com¬ liberate my conscience by these few remarks . . .
mittee from among the membership of the latter. I will put the resolution to the Conference, that
The national laboratories designated were the The International Ohm shall be defined as the
National Physical Laboratory of Great Britain, resistance of a specified column of mercury. . . .”
the Laboratoire Central d’electricite at Paris, the A number of factors cooperated in the ensuing
Physikalisch-Technische Reichsanstalt of Ger¬ years to demonstrate the wisdom of Lord Ray¬
many, the Central Chamber of Weights and leigh. One of these was the fact that the need
Measures of the USSR, the Electrotechnical for standards that were conveniently reproducible
Laboratory of the Department of Communica¬ vanished with the growth of testing service by
tions of Japan, and the National Bureau of the national laboratories. A standard cell or
Standards of the United States. Two individual resistor could be shipped from San Francisco to
specialists, Prof. L. Lombardi, Director of the Washington, tested, and returned with a tiny
Electrotechnical Laboratory of the Royal School fraction of the time and expense involved in
for Engineers, Rome, and M. Ch.-Ed. Guillaume, setting up a voltameter or mercury column ex¬
Director of the International Bureau, were ap¬ periment. Laboratories in the smaller countries
pointed in 1928. could obtain similar service from the large na¬
At its meeting in 1929, the International Com¬ tional laboratories or, more recently, from the
mittee asked the Advisory Committee to consider International Bureau.
also problems related to photometric measure¬ A second factor in the diminishing utility of the
ments and standards, and specialists in pho¬ international units was the discovery of isotopes.
tometry were invited to attend the meetings. At Mercury has seven isotopes, of which five occur
this time the committee was often referred to as in considerable amounts, and which have masses
the -“Advisory Committee on Electricity and covering a range of 2 percent. Silver has two
Photometry.” However, in view of the differ¬ isotopes of almost equal abundance, which differ
ences between the two fields of work, a separate in mass by about 2 percent. The volume re¬
Advisory Committee on Photometry organized sistivity of mercury is presumably nearly the
along parallel lines was authorized by the Inter¬ same for all isotopes, but the Chicago Congress,
national Committee at its 1935 meeting. in an endeavor to make the definition more pre¬
The increase in the scope of work of the Inter¬ cise, changed the earlier specification for a column
national Bureau required increased laboratory 1 sq mm in cross section by specifying a column
space, and in November 1928 the International having a mass of 14.4521 g. With this definition
Education Board contributed 900,000 francs and a change in isotopic composition would be expected
the French government released an additional to make a change in the resistance of the column
small strip of land that made it possible to add proportional to the change in density. This
sufficient laboratory space to house the electrical difficulty is probably more academic than real
apparatus. Another assistant was added to the because specimens of mercury from widely dif¬
staff in 1929 to carry the additional work. ferent sources have been found to have very closely
the same density.
2. Trend Toward Absolute Units The difficulty from the silver isotopes is less
academic, because the process of electrolysis
In the decades following the London Conference may well introduce an appreciable change in
of 1908, the developments of science and of the isotopic composition. In fact, electrolysis is
electrical art materially changed the situation that the most common process for concentrating the
had led to the establishment of the system of inter¬ heavy isotope of hydrogen. If, in the deposit, the
national units. The genius of Lord Rayleigh, ratio of the light to the heavy isotope content is
who served as President of the Conference, may greater by 1 percent than it is in the solution, the
have enabled him to foresee these trends, for even deposit will be 100 ppm lighter than if no separa¬
when he was in the act of establishing the inter¬ tion had occurred. Moreover the extensive work
national system by formally putting to a vote the from 1911 to 1914 with the silver voltameter
32/18
revealed still other complications such as effects 3. Official Actions Leading to the Adoption
from colloidal silver, spurious inclusions in the of the Absolute Units
deposit, etc., which could cause appreciable
errors in work of the highest accuracy. The possibility of shifting back to the absolute
system of units had doubless been in the minds of
At first sight it might appear that the con¬
many scientists for a long time, and indications of
ventional choice of a fixed number, namely
it are evident in some of the discussions in 1921
0.00111800 grams of deposit per international
regarding the extension of the authority of the
coulomb, to define the unit of electric charge
International Committee. Perhaps the first offi¬
would have a theoretical advantage by eliminating
cial action in this direction was a set of resolutions
one experimental step hi the determination of the
formulated by the Committee on Instruments and
Faraday. This physical constant would then be
Measurements of the American Institute of Elec¬
given by F=Al/0.o6l 11800, where A is the atomic
trical Engineers and formally adopted June 27,
weight of the silver deposited. For the standard
1928, by its Board of Directors. This (see ap¬
atomic weight 107.880, F is 96,494 international
pendix 4) pointed out the unsatisfactory aspects
coulombs per gram equivalent, and the only
of the situation then existing and urged the na¬
experimental determination involved is that of the
tional laboratories to undertake the additional re¬
atomic weight. However, in the numerous theo¬
searches necessary to make it feasible to legalize
retical uses of the Faraday, as in the determination
the absolute units.
of Avogadro’s number, it is combined with other
At the same time Dr. G. K. Burgess, then
physical constants at least some of which involve
Director of the National Bureau of Standards and
the electromagnetic forces on elementary charges,
its representative on the international Advisory
which in turn are expressible primarily in absolute
Committee, called together an American Advisory
units. The “international Faraday” is therefore
Committee on Electrical Units and Standards.
useful only when the ratio of the international
The membership included representatives of the
to the absolute coulomb is also known. More¬
National Academy of Sciences, American Institute
over, the complications from inclusions, colloidal
of Electrical -Engineers, the American Physical
silver and possible isotopic separation constitute
Society, the National Electric Light Association,
serious sources of error when the silver voltameter
the Association of Edison Illuminating Companies,
is used to determine the Faraday. On the other
the National Electrical Manufacturers Associa¬
hand, they drop out when a single particular type
tion, and the American Telephone and Telegraph
of voltameter operating under closely controlled
Company. In 1930, two members, representing
conditions, is used to reproduce the same average
the Illuminating Engineers Society and the Optical
current in a repeat experiment even after the
Society of America, were added when photometric
lapse of many years.
questions were under consideration. This ad¬
The third and perhaps most important factor visory committee met first in Washington on June
was the tremendous increase in the scope of appli¬ 16, 1928, and discussed the functions that it was
cation of electrical methods of measurement in desirable to have the International Bureau under¬
many fields of science and engineering and the take in connection with the electrical units, and
simultaneous increase in the accuracy required. the consensus was as follows:
Electrical methods are used to measure tempera¬
(1) A central secretariat to arrange for systematic
tures, pressures, strains, acidity, concentrations, exchange of standards and compilation of results of
and a myriad of other physical and chemical intercomparisons thus made among the national labor¬
quantities. In some of these, particularly in the atories.
measurement of energy and in the calibration of (2) A laboratory to which concrete standards repre¬
senting the results obtained in the different countries
calorimeters, the desired precision has long been may be brought for precise comparisons.
so high that corrections had to be be applied for (3) A repository for international reference and
the difference between the international and the working standards with the necessary equipment so
absolute units. Electrical effects are fundamental that other standards may be compared with these
standards on request.
in the experiments by which a large number of
basic physical constants such as the Faraday, The committee then voted the following reso¬
electronic charge, mechanical equivalent of heat, lution:
Bohr magneton, and Zeeman displacement con¬
That in the opinion of this Committee, in view of
stant are measured. It soon became evident that improvements which are being made in absolute meas¬
the bother and confusion already experienced in urements, electrical standards should in future be
the field of thermochemistry would soon become based upon the absolute system of units.
serious in many other fields if the independently At almost the same time Dr. D. W. Dye, who
fined international units continued in use. represented the National Physical Laboratory on
33/19
the international Advisory Committee, laid the G. A. Campbell, in a paper entitled, “A definitive
same questions before the Electrical Units and system of units,” which he presented at a meeting
Standards Committee of the NPL.3 * The opinions in Chicago on June 24, 1933, of the Committee for
of this committee as to the desirable functions of Symbols, Units, and Nomenclature (SUN Com¬
the International Bureau were very closely con¬ mittee) of the International Union of Pure and
cordant with those of the American Advisory Com¬ Applied Physics, suggested that the ohm be fixed
mittee. They also regarded it “as of the utmost by a material standard resistor and that all other
importance that the use of the absolute units electrical units be derived from this and from the
should be introduced at the earliest possible mo¬ meter, the kilogram, and the second. On this
ment. * * * It is felt that this change can basis, the “permeability of space,” /jl„, would be¬
be made now (1928), but that, the difficulties in¬ come a derived quantity to be determined by
volved will increase with each year of delay.” experiment and would have a-value of about
The German position at this time, as set forth 0.9995X10-7 henry per meter. According to this
in a memorandum by Dr. von Steinwehr, was plan the prototype ohm was to be preserved at the
much less favdrable to the proposed change. International Bureau in the same manner as the
While admitting the theoretical advantages of the meter and the kilogram. It would follow that if
absolute system, they emphasized the inconven¬ this standard resistor should drift in time as did
iences that they feared would develop in practice those whose values are shown in figure 8, the ex¬
during the transition. They proposed that much perimentally determined value of p,, would also
additional work should be carried out first, to de¬ be found experimentally to drift and to have dif¬
termine the ratios of the international to the ferent values at different epochs. This proposal
absolute units. After these ratios were definitely was formally rejected by .the SUN Committee.
established, they would then urge that the defi¬ A similar suggestion was proposed by. Prof.-Ing
nitions in absolute terms be used as the basis for G. Giorgi and endorsed by F. Emde and by A.
defining the units, but that the established ratios Sommerfeld. A great deal of confusion has arisen
should then be applied so as to keep the magni¬ because the name Giorgi thus became attached to
tudes of the units actually used at the same values a second system of units, which was, in effect,
as the current international ones. almost a special case of an earlier system of units
On the other hand, a similar committee of that he had proposed. As early as 1901 Giorgi [2]
Italian scientists under the chairmanship of Prof. had made the constructive suggestion that the
L. Lombardi favored the ultimate transition to incomplete “practical absolute” system of eight
absolute units, although they felt that considera¬ electrical units (coulomb, ampere, volt, ohm,
ble further experimental work in absolute measure¬ henry, farad, joule, watt) could be extended into a
ments should be done first,. Professor D. Konova¬ complete system that would include mechanical
lov of the Central Chamber of Weights and Meas¬ and magnetic units as well. By the artifice of
ures, USSR, sent in proposals that had been ap¬ assigning to n;, the value 10-7 in place of the value 1,
proved by a group representing several central which is chosen in the centimeter-gram-second
scientific and technical institutions in Russia. electromagnetic system, Giorgi found that the
These proposals were closely parallel to those of meter, the kilogram, and the second became auto¬
the American and British members. Similar re¬ matically the mechanical units in his new com¬
sponses came from analogous committees of the plete system, and it is therefore usually called the
Societe Fran^aise des Electriciens and from the “MKS System”. The unit of force in this system
Japanese Electrotechnical Laboratory. has a magnitude of 105 dynes and the name
In view of this preponderance of opinion, the “newton” lias been proposed for it. This valuable
Advisory Committee, on June 8, 1929, adopted a suggestion has met with growing but rather belated
resolution translated as follows: approval.
In view of the great importance of unifying the In 1935 the International Electrotechnical Com¬
systems of electrical measurement upon a basis free mission meeting at Scheveningen gave its formal
from arbitrary characteristics, the absolute system, approval to this MKS system, recommended its
derived from the centimeter-gram-second system,
should replace the international system of units for general use for engineering purposes, and sug¬
all measurements in science and industry. gested that it be called the “MKS-Giorgi” system.
This was officially approved by the International In engineering uses, the distinction between the
Committee at its 1929 meeting and widely international units and the absolute units is neg¬
publicized. ligible and was probably not given any consider¬
As might have been expected, a few individuals ation by the IEC. Theoretically, if the absolute
disagreed with the proposed change in units, and units are used everything will be inherently
felt that the change in the ohm would involve consistent . On the other hand, with the suggestion
excessive inconveniences in laboratory work. of Campbell or the later suggestion of Giorgi,
which was not approved by the IEC, the joule and
3 The members of this Committee were Sir Richard Glazebrook, Chairman, watt would have retained their absolute values,
Sir William Bragg, Dr. D. W. Dye, Dr. Alexander Russell, Sir Arthur
Schuster, Dr. F. E. Smith, and Sir Joseph Petavel. the ohm would by definition have kept its inter-
34/20
national value, and the ampere and the volt would fixing the values. It published provisional values
have had neither their absolute nor their interna¬ for the ratios expressed to parts in 10,000 based on
tional values, but would have been respectively
1 international ohm= 1.0005 absolute ohms.
less and greater than their absolute values by the
1 international volt= 1.0004 absolute volts.
square root of the factor by which the international
ohm is greater than the absolute ohm. Later, at The Advisory Committee met in June 1939, and
its meeting in Torquay in 1938, the IEC explicitly the several members reported the status of the
showed its preference for the absolute system by work in their respective laboratories on determin¬
recommending “. . . as the connecting link be¬ ing the ratios of the international to the absolute
tween the electrical and mechanical units the units. The German, Japanese, and Russian labora¬
permeability of free space with the value At„=10-7 tories had not yet completed their programs.
in the unrationalized system or hv=4tt.10~7 in the However, the Committee felt that the good
rationalized system.” agreement between the British and American
The alternative suggestions of Campbell and of data, and the increasing inconvenience that the
Giorgi were duly considered by the International transition would entail if too long postponed
Committee, but at its meeting in 1935 the decision justified adhering to the proposed date of January
to shift to the absolute units was reaffirmed. 1, 1940. Resolutions were drawn up transmitting
It should be emphasized that the action of the this opinion to the International Committee,
International Committee for Weights and Meas¬ fixing provisionally the ratios to be announced,
ures in authorizing the change from the interna¬ and offering suggested wordings for exact theo¬
tional to the absolute electrical units has no bearing retical definitions of the new absolute units, which
whatever on the question also now under discus¬ could if desired be incorporated in the legislation
sion as to the relative merits of the MKS and cgs that might be enacted by the several adherent
systems. The absolute ohm, ampere, and volt nations to put the new units into full effect.
may be regarded equally well either (1) as units of The meeting of the International Committee
the absolute AIKS system (either rationalized or has been scheduled for September 1939, but the
unrationalized), or (2) as convenient decimal outbreak of World War II brought all activity in
multiples of the corresponding units of the cgs this field to a standstill.
system. Moreover, the action of the IEC in Following a brief preliminary session in Novem¬
specifically assigning a value to nT as the link ber 1945, the International Committee met in
between electrical and mechanical units should not plenary session in October 1946.4 A poll of the
be interpreted as giving any particular funda¬ laboratories represented on the Advisory Com¬
mental significance to permeability. The IEC mittee on Electricity showed that, with the
Committee on Electric and Magnetic Magnitudes exception of the Reichsanstalt, all still concurred
and Unit's (EMMU) explicitly stated “The in the opinion expressed by the Committee in 1939
Committee recognizes that any one of the following that the transition to the new units should be
practical units, ohm, ampere, volt, henry, farad, made promptly. The Germans still would have
coulomb, weber already in use may equally serve jmeferred to postpone the change until they could
as the fourth fundamental unit, because it is verify by their own experiments the correctness
possible to derive each unit and its dimensions of the new units. It was evident however that
from any four others mutually independent.” their flight from Berlin to Weida to escape the
At the Eighth General Conference on Weights bombings must have deranged the laboratory to
and Measures, meeting in 1933, the change to the such an extent that they would not be in position
absolute units was definitely confirmed (see ap¬ to make an.effective contribution for a very long
pendix 5). The Conference delegated to the time. Accordingly the International Committee
International Committee full authority to fix the formally voted resolutions (see appendix 6) based
ratios of the new to the old units and to set the on the recommendations made in 1939 by the
date for their adoption. It was hoped that the Advisory Committee. These (1) set the date
change might be made in 1935. January 1, 1948, for the change; (2) fixed the
However, absolute measurements of high ac¬ ratios
curacy are very time-consuming, and although
work was pushed vigorously at several national 1 mean international olim= 1.00049 absolute ohms.
laboratories, it became evident that reliable values 1 mean international volt= 1.00034 absolute volts.
of adequate accuracy for the ratios of the absolute as those governing the change on that date; and
to the international units could not be obtained (3) gave official formulations of definitions for the
so quickly. At the 1935 meeting, the International new units.
Committee [8] postponed the date of adoption to
January 1, 1940, asked the Advisory Committee * The membership of the International Committee at this time was: L. de
Broglie (France); G. Cassinis (Italy); M. Chatelain (USSR); E. C. Crittenden
to meet early in 1939 to fix the ratios, and an¬ (USA); M. Dehalu (Belgium); W. J. de Haas (Holland); E. S. Johansen
nounced that only experimental work reported (Denmark); C. Kargatchin (Yugoslavia); W. Kosters (Germany); H. Na-
gaoka (Japan); Z. Rauszer (Poland); M. Ros (Switzerland); J. E. Sears
before the end of 1938 would be considered in (Great Britain); M. Siegbahn (Sweden); C. Statescu (Rumania); A. Perard.
35/21
This action was given ex post facto approval by prior to January 1,1948 to the absolute units used
the Ninth General Conference when it met in after that date are as given in the tabulation
October 1948. below. (It should be noted that the U. S.
Conformance with this formal action of the “Washington ohm” was 7 ppm smaller than the
International Committee was of course desirable ohm as certified for use in the USA).
on the part of the United States as a signatory
1 international ohm =1.000495 absolute ohms.
nation. Accordingly the National Bureau of 1 international volt =1.000330 absolute volts.
Standards proceeded to prepare for the impending 1 international ampere =0.999835 absolute ampere.
change. Bureau Circular C459, “Announcement 1 international coulomb = 0.999835 absolute coulomb.
of change in electrical and photometric units”, 1 international henry = 1.000495 absolute henries.
1 international farad =0.999505 absolute farad.
was published in May 1947 and given wide 1 international watt =1.000165 absolute watts.
circulation. Excerpts appeared in most of the 1 international joule =1.000165 absolute joules.
technical magazines concerned, and oral announce¬
ments were made at various scientific meetings. The legal status of the new units in the United
During 1947 all certificates issued by the Bureau States is, of course, exactly the same as that of the
giving precise electrical values contained a supple¬ older ones because the law of 1894 (appendix 1)
mentary statement of what the values were going mentions both sets of units on an equivalent basis.
to be in the new units. Similarly, during 1948 However, in order to remove the ambiguities of
such certificates contained not only the values in the old act, and also to provide a statutory basis
the new units but also a supplementary statement for photometric units that have hitherto been
of what the values would have been in terms of the based merely on the common consent of those
old units. interested, new legislation has been proposed.
Because of the differences, shown in figures 8 Substantially identical bills, HR4113 and S1850
and 9, between the mean international units as were introduced in the United States Congress in
defined by the intercomparisons at the Inter¬ July 1947 and December 1947, respectively, but
national Bureau and the Washington units as died with the expiration of the 80th Congress in
maintained at the National Bureau of Standards, December 1948. Bill S441 (see appendix 7) was
the ratios of the units as certified by the NBS introduced in the 81st Congress in January 1949.
V. Absolute Measurements
1. General As a second step, the inductance can be used in
one of several forms of measuring circuit at a
The many interactions between electrical and known frequency to determine the value in ab¬
mechanical phenomena have offered a considerable solute ohms of the resistance of a resistor. Be¬
variety of possible methods for deriving electrical cause of the fact that resistors rather than induc¬
quantities from mechanical measurements [61, 62, tors are commonly used for the maintenance of the
63]. However, experience has shown that rela¬ unit, most experimenters have combined the two
tively few of these give prospects of yielding an steps so that the final result of their experiment
accuracy as high as a few parts in 100,000. is the determination of the value of a resistance in
One such relation is that between the induct¬ absolute units. Hence such experiments are
ance of a circuit and its geometrical size and shape. referred to as absolute-ohm determinations, al¬
Many formulas [69] have been worked out for though the absolute henry is really the first
such relations. By constructing an inductor of product.
simple shape in such a way that all its significant It might appear that this single electrical
dimensions can be measured with sufficient determination of the ohm should suffice because
accuracy, and by ensuring that all materials used the ampere can theoretically be defined as the
in its construction or located in its neighborhood current that dissipates energy at the rate of 1
are nonmagnetic, a standard is obtained, the watt when flowing in a resistance of 1 ohm, and
inductance of which can be computed in absolute the watt is purely a mechanical unit. In practice,
henries. For results of high accuracy, extreme however, the mechanical measurement of energy
care and attention to detail are required in the or power with high accuracy is difficult. It has
construction of the inductor. The mechanical been found preferable to make use of a second
measurements involve a tedious multiplicity of typo of absolute measurement in which the current
measurements of diameter and length of various in a pair of coils develops a mechanical force of
parts of the structure, all of which must be of the attraction or repulsion between them. The force,
highest accuracy. An example of the effort to Fx, in the x direction developed in such a circuit
avoid systematic errors is the use by one group at by a current, /, is given by
the NBS of end-length standards only, while the
workers using another method used line standards
only.
36/22
where dM/dX[ is the rate of increase of the mutual apparatus. D and D' are disks mounted on the
inductance between the fixed and the moving coils rotating shaft, S'. Brushes b and b' make contact
as the dimension is increased. By arranging the with the periphery of the disks. Each pair of fixed
coils so that Fx is vertical, it may be measured by coils, for example, C\ and C2, or C3 and C4 may be
balancing it against the gravitational force, Mg, considered the primary windings of a mutual in¬
on a mass, M. The force is then readily obtained ductor, M, of the Campbell type, the secondary
and the major problem is the computation of of which is formed by the periphery of a slip-ring
dM/dX from the known proportions of the circuit. carried on a disk. When there is a current I in
Such an apparatus is commonly called a “current the coils, the magnetic flux linking the periphery
balance.” of one ring is <f>=MI. When the shaft, S, is ro¬
Because of the transitory nature of an electric tated, each conducting path (b d d' b') from a
current, the customary procedure is to connect in point on the periphery of one ring through the
series with the coils of the balance a resistor o-f disk and shaft to a corresponding point on the
resistance, B. The voltage drop, IB, in this periphery of the other ring will cut the total flux
resistor is opposed to the emf, E, of a standard embraced by the peripheries of the two rings, and
cell. When the current, /, is held at such a value this flux is given by (Mf-M') I. The emf
that IB balances E and at the same time the generated between the brushes is therefore propor¬
mass, M, of the weight has been adjusted to tional to the current and to the spexl of rotation.
balance the electrodynamic force, the relation
37/23
This.emf is balanced against the IR drop of the
same current through the resistor, B, which is
being measured. When a balance is obtained,
R—n Here B is in ohms if n is the
speed in revolutions per second and M and M' are
in henries.
One method used at the NBS by Curtis, Moon,
and Sparks [65] was to construct a large self in¬
ductor of copper wire wound in a carefully lapped
groove on the outer cylindrical surface of a form
of pyrex glass or of fused quartz. The inductor
on the pyrex form is shown in figure 11. The
computation of the inductance of such a coil is
made by the formulas developed by Snow.
Figure 12 shows the circuits used in the electrical
38/24
M, Mz A second method used at the NBS is the Wenner
method [68], which combines certain desirable
features of several of the preceding methods. The
mutual inductor, the primary of which is shown
in figure 15, is a modification of the Campbell type
with two gaps in the primary winding so placed as
to produce two annuli of zero field. By proper
design, the resulting annular space within which
the field is very small and in which the secondary
winding is to be located can be made relatively
large. The circuit used is shown in figure 16.
A direct current flowing through the four-terminal
Figure 14. Campbell method for the absolute measurement resistance, R, which is being measured (here
of resistance. shown with an adjustable shunt, Sn, across its
potential terminals) also flows in the primary
The values of the two mutual inductors are derived by comparison with a
computable standard inductor. The secondary emf of Mi is nearly equal winding, P, of the mutual inductor M, but is
in magnitude and opposite in phase to the drop of the primary current in r.
reversed in direction cyclically by the commutator,
Cp. The emf induced in the secondary of the
The first of these equations, together with the mutual inductor is rectified by the commutator,
measured ratio of R to r, gives a value for R (or r) Cs, and opposed through the galvanometer, G, to
in terms of M and co. the IR drop in the resistor, R. The two commu¬
tators are mechanically alined with such relative
phases that the secondary connections are re¬
versed after the primary current has had time to
settle to a steady value. The current flowing in
the secondary at the time of secondary commuta¬
tion is negligibly small. When a balance is
obtained the resistance is given by the equation
R—ft nM U/Ia,
39/25
Figure 16. Schematic circuit diagram for the Wenner method.
Resistor R is measured in terms of mutual inductor M, and frequency of reversal of primary and secondary commutators Cp and Cs. Electronic amplifier
-T'i and lr2 hold the current in R constant, while that in P reverses cyclically.
Two electronic amplifiers, Vx and V2, of the phase- Table 2. Absolute measurements of resistance
inverter type that have a “gain” of 1:1, give
further regulation and reduce any cyclic and Date Authors Method Value 1
irregular fluctuations in the potential across a and
c by a factor of 400. The current in the resistor,
1937 Vigoureux Lorenz 1. 000 499
B, is thus kept constant to better than 1 ppm
1937 Hartshorn and Campbell_ 1. 000 505
throughout the cycle. At the time of each Astbury.
secondary commutation, the resistor is directly in 1936-38 Curtis, Moon, Self-induc¬ 1. 000 472
series with the primary winding of the mutual and Sparks. tor.
1948 Thomas, Peter¬ Wenner_ 1. 000 484
inductor. Therefore Ir equals Ia to a high
son, Coo ter,
accuracy, and and Kotter.
R = 4 nM.
Mean 1. 000 490
This method has been found so convenient to
operate that it is planned to keep the apparatus
1 Number of absolute ohms in 1 mean international ohm.
in commission continuously and to measure the
resistance of the 1-ohm standards in the primary Many sources of error are common to all of the
group at intervals of not longer than 1 year. A methods, although they would be expected to ap¬
continued agreement with earlier values will be a pear to a very different extent in different methods.
very strong indication that neither the resistance One of these is the uncertainty in the dimensions
standards nor the mutual inductor has drifted in and the uniformity of the windings. Perhaps the
value. The chance that two such different struc¬ best inductors are those of Curtis, Moon, and
tures should show simultaneous and equal drift in Sparks, but even in these the outside diameter of
the same direction is very small. If discrepancies the helix has a maximum variation of 4 microns,
develop, the laborious process of measuring the me¬ although the variations in pitch correspond to axial
chanical dimensions of the inductor will have to be irregularities certainly less than a few tenths of a
repeated. It is hoped that this will not be neces¬ micron. Errors may arise from the change in
sary more often than once every 10 or 20 years. dimensions resulting from the temperature rise
Table 2 summarizes results obtained by these caused by the heating effect of the current if the
four methods for the absolute measurement of re¬ current used in the measurement is increased to
sistance and shows that the agreement is highly obtain higher sensitivity. Still another source of
satisfactory in spite of the very considerable dif¬ error is the uncertainty in the permeability of the
ferences between the methods. It therefore seems form on which the inductor is wound and the ef¬
probable that the systematic errors have been fects from magnetic ma terial in the neighborhood.
largely eliminated from all of them. For instance, to avoid such effects at the NBS, the
40/26
end of the building assigned for absolute measure¬ the cross section of wire is made large to permit
ments was constructed with bronze plumbing, re¬ the use of large currents.
inforcing rods of special nonmagnetic steel, lead Other sources of error applicable to particular
window weights, etc. methods that may be mentioned are: Imperfect
The frequency used in the main balance in the integration of the current by the galvanometer
second method, or in the Maxwell bridge in the used in the Maxwell bridge; changes in the resist¬
third method, and the speed of rotation of the ance of the intermediate circuit of the Campbell
generator or commutator shafts in the other two method as a result of heating of the copper coils
methods must be measured to the full accuracy that form part of this circuit; the low voltage de¬
desired in the final result. This, however, has not veloped in the Lorenz apparatus, which makes
constituted a serious limitation because the aver¬ thermoelectric and contact trouble at the brushes
age speed over a considerable time can readily rather significant, and the possibility of residual
be determined to high accuracy by comparison emf'in the secondary of the inductor at the time
with circuits timed by crystal oscillators. A more of commutation of the secondary circuit in the
subtle source of error may be the uncertainty in Wenner method.
the location of the current within the cross section In spite of all these possible errors, the great
of the primary winding. If the material were effort expended on these methods lias led to results
homogeneous one would expect the center of which, as shown in table 2, differ by only a very
gravity of the current distribution to lie somewhat few parts in 100,000.
inside the center of the wire because of the shorter
circumference of the inner edge of each turn. The 3. Absolute Measurement of Current
problem is complicated, however, because in the
process of winding, parts of the material are The only methods [70, 71, 72] used in recent
strained even beyond their elastic limits, and these times for the absolute measurement of current
strains may cause small changes in the resistivity, with high precision involve a current balance
which in turn will affect the distribution of cur¬ somewhat like that shown schematically in figure
rent. Such errors, are, of course, greater when 17. The current to be measured is passed in series
With the weight, lifted, the direction of currents is such that the moving coil is attracted by the lower and repelled by the upper fixed coil. When the switch
is reversed, the weight is lowered on to the balance pan and the electrodynamic forces are in the opposite direction.
572772 0-61-4
41/27
through the two fixed cods and the moving coil. precision than can be obtained by purely mechani¬
The direction of the windings is such that the cal measurements. This is done with the circuit
magnetomotive forces of the two fixed coils are in shown in figure 18. The two coils concerned are
opposition so that they conspire to force the mov¬ mounted concentrically with their planes in the
ing coil in the same direction. The current bal¬ magnetic meridian and are energized with cur¬
ance at the NBS uses coils the cross section of rents of such magnitude that the net magnetic
which is small relative to their radius and is com¬ field at the common center of the coils is zero.
monly designated as of the Rayleigh type. At This condition is indicated by a sensitive magne¬
the NPL all three coils are wound with single-layer tometer at the center. When the magnetometer
windings that have considerable axial length, and shows no deflection as a result of a reversal of
the balance is designated as the Ayrton-Jones both currents, the ratio of the radii is equal to
type. The NPL balance has two sets of coils, one the ratio of the currents. The ratio of the cur¬
moving coil being supported from either end of rents is then equal to the ratio of the resistances,
the balance. This has the advantage of doubling Rx and R2, when the galvanometer, G, also shows
the force and of giving some compensation between a balance.
the disturbing forces due to convection currents The principal limitation in the absolute measure¬
arising from the heating developed by the current ment of current seems to be in the uncertainty
in the moving coils. In operation, the rest point in the location of the coils, and it lias seemed
of the balance is observed before and after re¬ better to sacrifice the greater mechanical forces
versal of current in the fixed coils. A compensat¬ that can be obtained by using multilayer coils of
ing mass of known weight, nearly equal to the round insulated wire and to use instead single
change in electromagnetic force, is also added layer solenoids of bare wire or coils wound of
during one observation and removed during the metallic ribbon to form a coil of many layers but
other. of one turn per layer. Aluminum ribbon insu¬
The formulas for computing the force in terms lated. by an anodizing process has been found
of current are complicated, but in the Rayleigh convenient for such coils.
form involve principally the ratio of the effective Figure 19 shows the combination of a helical
radii of the fixed and moving coils. It has been moving coil with a pair of spiral fixed coils as
found possible to measure this ratio of the effec¬ mounted at the NBS. A comparison of the re¬
tive radii by an electrical method with a higher sults obtained when different coil combinations are
in use is shown in table 3. The values listed in
this table are the amounts' in ppm by which the
absolute ampere, as derived by the use of various
combinations of fixed and moving coils in the
current balance at the NBS, exceeds the NBS
international ampere as certified in the United
States. It will be seen that when the larger
(25-cm diam) moving coil was used with the
smaller (40-em diam) of the fixed coils, the differ¬
ence, 26 ppm, was 145 ppm smaller than the
value, 171 ppm, observed when the ribbon-wound
(25-cm diam) moving coil was used with the
fixed single-layer solenoids (46-cm diam). These
systematic difference are definitely larger than the
random errors of observation and are still unex¬
plained.
Fixed coils
Moving coil
Wire, Wire,
Spiral Helical
small large
42/28
Figure 19. Coils of current balance.
Here the two spiral fixed coils are wound of anodized aluminum strip on the outside of the water jacket and enclosed in a protective case. The moving coil
is a single-layer helix wound in a lapped groove in the glass form, and suspended by three rods from the perforated spider, which in turn hangs from one end of
the balance.
mass. The values reported from earlier work and On the basis of these results, the International
given in table 4 are based on the 1906 Potsdam Committee chose the reduced value 0.99985 for
determination. In the light of more recent the value of the mean international ampere in
measurements [751, it appears that the best value absolute amperes. Hence the international am¬
of g is probably less by 15 ppm than that derived pere as certified by the National Bureau of Stand¬
from Potsdam. ards is 0.999835 absolute ampere.
The results of the best recent absolute measure¬ It is evident from table 3, however, that the
ments of current are summarized in table 4. certainty of the final result is not as high as might
162]. be concluded from the (perhaps fortuitous) agree¬
Table 4. Absolute measurement of current
ment of the independent values listed in table 4.
There is an evident need for the development of
Date Authors Method Value a some radically different method for measuring cur¬
rent in absolute amperes that will not be subject
1936 Vigoureux Ayrton- 0. 999 863 to the same types of systematic error. The
Jones. Pellat balance [73] in which the current in two
1939 Curtis, Curtis, Rayleigh_ 0. 999. 860 coils placed with their axes at right angles produces
and Critchfield.
a mechanical torque, seems to offer such a possibili¬
1942 Curtis, Driscoll, Combina- 0. 999 856
and Critchfield. tion ty. The construction of such a balance was
begun at the NBS some time ago, and the mathe¬
Mean 0. 999 860 matical formulas needed in its use have been
worked out by Snow [74]. It is hoped that this
» Number of absolute amperes in 1 mean international ampere, using
values of gravity derived from the Potsdam determination. project can be resumed in the near future.
43/29
VI. Summary
The preceding record has shown that for the ment by which, at fairly long intervals, checks will
37 years from 1911 to 1947, inclusive, the national be made by one or another of the national labora¬
standardizing laboratories, with valuable coor¬ tories to detect any incipient progressive drifts in
dinating service from the International Bureau its standards. The older procedures for main¬
of Weights and Measures; succeeded in main¬ taining the ohm and the volt by groups of standard
taining throughout the civilized world a system of resistors and cells of the highest quality will con¬
electrical units that did not vary in time among tinue unchanged. The biennial intercomparisons
any of the six participating nations by more than at the International Bureau are being resumed and
0.01 of 1 percent. The units in the United States will serve both to detect relative drifts between
apparently never departed from the desired ideal the standards of different countries and to main¬
by more than 20 ppm. tain accurate values for the standards deposited at
The ideal pursued during these years was the Sevres, which in turn serve as primary standards
system of international electrical units, formu¬ for the laboratories of countries that do not have
lated by the London Conference of 1908, and national laboratories of their own.
experimentally realized by the Technical Com¬ It is to be expected that the International
mittee in Washington in 1910. The basic defini¬ Advisory Committee on Electricity will meet only
tions in terms of mercury columns and silver at fairly long intervals as additional experimental
deposits were brought into play apparently on realizations of the absolute units become available,
only four occasions: To readjust the British ohm or if an undue drift in the value of one or more
in 1927, the German ohm and the volt in 1932, nation’s unit relative to the mean indicates a need
and the Russian volt in 1930, when each of these for consideration by the International Committee
nations changed its unit to offset the effect of on Weights and Measures. Any readjustment of
cumulative drift in the standards it had been values resulting from such situations will presum¬
using to maintain the unit. The fact that such ably be very small.
drifts did accumulate makes it clear that in the However, the adoption of the new ideal, which
future as in the past some fixed definitions must is inherently consistent with the mechanical units,
be set up by which the standards can be checked lias automatically eliminated all future need of
periodically. The results of absolute measure¬ correcting precise scientific data to allow for the
ments by a variety of methods have shown that artificial distinction between the international
such absolute measurements can be used to serve and the absolute units.' Future absolute measure¬
as a check against cumulative drift with an accura¬ ments can be depended upon to keep the units
cy at least as great as that attainable in the dis¬ as maintained amply close to the ideal, because
carded international standards. with adequate attention they may be expected to
The change of January 1, 1948, involved both a increase in accuracy in parallel with any possible
change in the size of each unit by the factor shown future increases in the accuracy of the scientific
on page 22 and a change in the type of experi¬ measurements that they are to correlate.
VII. References
[11 Historical Reports of the Committee on Electrical [8] Proces Verb. Comite International des Poids et
Standards appointed by the British Association for Mesures 17, 74 (1935).
the Advancement of Science (reprinted 1913 by [9] Announcement of changes in electrical and photo¬
Cambridge Univ. Press). metric units, NBS Circular C459 (1947).
[2] G. Giorgi, Proposal concerning electrical and physical
units, Trans. Int. Electrical Congress 1, 136 (St.
Louis, Mo., 1904). Mercury Resistance Standards
[3] F. A. Wolff, The principles involved in the selection
and definition of the fundamental electrical units to [11] W. Jaeger, Iv. Ivahle, and H. Diesselhorst, Mercury
be proposed for international adoption, Bui. BS 5, standards of the Reichsanstalt, Wiss. Abhandl.
243 (1908-9) SI02. physik. tech. Reichsanstalt 4, 117 (1905).
[4] Announcement of a change in the value of the inter¬ [12] F. E. Smith, Mercury standards and resistance
national volt, NBS Circular C29 (1910). temperature coefficient of mercury, Phil. Trans.
[5] Report to the International Committee on Electrical 204, 57 (1904).
Units and Standards of a special technical commit¬ [13] F. A. Wolff, M. P. Shoemaker, and C. A. Briggs,
tee appointed to investigate and report on the Construction of primary mercurial resistance
concrete standards of the international electrical standards, Bui. BS 12, 375 (1915-16) S256.
units and to recommend a value for the Weston [14] F. E. Smith, Nat. Phvs. Lab. Report for the Year
Normal Cell, NBS Misc. Pub. M16 (1912). 1912, p. 40.
[6] Comptes Rendusdes Seance de la Sixieme Conference [15] L. Hartshorn, Nat. Phys. Lab. Report for the year
Generale des Poids et Mesures, pp 31 to 42 (Gau- 1925, p. 93.
thier-Villars, Paris, 1921). [16] von Steinwehr and Schulze, Die Quecksilbernormale
[7] Proces Verbaux, Comite International des Poids et der PTR fur das Ohm, Wiss. Abhandl. physik.
Mesures 13, 58 (1929). tech. Reichsanstalt 11, 75 (1927).
44/30
Silver Voltameter [55] M. Romanowski and M. Roux, Premiers Essais de
Comparison des Etalons Nationaux de Force Elec-
[2t] E. B. Rosa and G. W. Vinal, Summary of experi¬ tromotrice, Proces Verbaux, Comite International
ments on the silver voltameter at the Bureau of des Poids et Mesures, 16, 141 (1933).
Standards and proposed specifications, Bui. BS 13 , [56] Proems Verbaux, Comite International des Poids et
479 (1916-17) S285. Mesures 17, 94 (Resolution 4). (1935).
[22] A. C. Ivolossov, Determination by the method of the [57] Proems Verbaux, Comite International des Poids et
silver voltameter of the electromotive force of the Mesures 21, (1948).
International Weston Normal Cell, Proems Ver- [58] M. Romanowski, Conservation of the electrical units
baux, Comite International des Poids et Mesures at the International Bureau during the years 1939
,
16 150 (1933) Annexe No. 19. to 1946, Proces Verbaux, Comite International des
Poids et Mesures 20, 179 (1946).
Standard Resistors
Absolute Measurements
[31] W. Jaeger and S. Lindeck, Reichsanstalt type re¬
sistance standards, Ann. phys. 65 , 572 (1898); [61] H. L. Curtis, Electrical measurements (McGraw Hill
,
Electrician 59 626 (London, 1907). Book Co., Inc., New York, N. Y., 1937).
[62] H. L. Curtis, Review of recent absolute determinations
[32] E. B. Rosa and H. D. Babcock, On the variation of
resistances with atmospheric humidity, Bui. BS 4 , of the ohm and the ampere, J. Research NBS 33,
235 (1944) RP 1606.
121 (1907-08) S73.
[33] E. B. Rosa, A new form of standard resistance, Bui. [63] H. L. Curtis, A review of the methods for the absolute
BS 5, 413 (1908-09) S107. determination of the ohm, J. Wash. Acad. Sci., 32,
[34] J. L. Thomas, A new design of precision resistance 40 (1942).
standard, BS J. Research 5, 295 (1930) RP201. [64] P. Vigoureux, Determination of the ohm by the
[35] J. L. Thomas, Stability of double-walled manganin method of Lorenz, Nat. Phys. Lab. Collected
resistors, J. Research NBS 36, 107 (1946) RP1692. Researches 24, 277 (1938); 11,'209 (1914).
[36] F. Wenner, methods, apparatus and procedures for [65] II. L. Curtis, C. Moon, and C. M. Sparks, A deter¬
the comparison of precision standard resistors, J. mination of the absolute ohm using an improved
,
Research NBS 25 229 (1940) RP1323. self-inductor, J. Research, NBS 21, 375 (1938)
RP1137.
[37] F. Wenner and E. F. Mueller, The Waidner-Wolff
and other adjustable electrical resistance elements, [66] L. Hartshorn and N. F. Astbury, The absolute meas¬
,
J. Research NBS 15 477 (1935) RP842. urement of resistance by the method of Albert
Campbell, Phil. Trans. Rov. Soc. (London) [A]
236, 423 (1937).
Standard Cells [67] A. Campbell, On a standard of mutual inductance,
Proc. Roy. Soc. (London) [A] 79, 428 (1907).
[41] L. Clark, On a standard voltaic battery, Phil. Trans. [68] J. L. Thomas, C. Peterson, I. L. Cooter, and F. R
,
Roy. Soc. 164 1 (1874). Hotter, An absolute measurement of resistance by
[42] The Weston Standard Cell, Electrician 30, 742 the Wenner method, unpublished.
(London, 1893). [69] E. B. Rosa and F. W. Grover, Formulas and tables
[43] H. B. Brooks, The standard-cell comparator, a spe¬ for the calculation of mutual and self-inductance
cialized potentiometer, BS J. Research 11 211 , (3d ed. revised), Bui. BS 8, 1 (1912) S169.
(1933) RP586. [70] P. Vigoureux, An absolute determination of the
ampere, Phil. Trans. Rov. Soc. (London) [A] 236,
International Comparisons of Units 133 (1937).
[71] H. L. Curtis, R. W. Curtis, and C. L. Critchfield, An
[51] Proces Verbaux, Comite International des Poids et absolute determination of the ampere, using im¬
,
Mesures: D. W. Dye, 13 161 to 166 (1929); K. proved coils, J. Research NBS 22, 485 (1939)
,
Takatsu & S. Jimbo, 14 192 (1931); M. K. Malikov RP1200.
,
and A. C. Ivolossov 14 199 to 207 (1931); G. K. [72] R. W. Curtis, R. I;. Driscoll, and C. L. Critchfield,
,
Burgess, 14 312 (1931); A. Perard and M. Roman- An absolute determination of the ampere, using
,
owski, 17 288 (1935). helical and spiral coils, J. Research NBS 28, 133
[52] G. W. Vinal, International comparisons of electric (1942) RP1449.
standards, BS J. Research 8, 729 (1932) RP448. T73] H. Pellat, An absolute electrodvnamometer, .T. phvs.
[53] P. Vigoureux, International determination of the (2d series) 6, 172 (1887).
electromotive force of the Normal Weston Cell in [74] C. Snow, Mutual inductance and torque between two
International Volts, Coll. Res. Nat. Phys. Lab. 24 , concentric solenoids, BS J. Research 1, 686 (1928)
79 (1932). RP24.
[54] H. von Steinwehr and A. Schulze, Neubestimmung der [75] H. L. Dryden, A reexamination of the Potsdam ab¬
EMK des Internationales Westonelements, Z. solute determination of gravity, J. Research NBS
Instrumenk. 52, 249 (1932). 29, 303 (1942) RP1502.
VIII. Appendix 1
U. S. Law of 1894, 53d Congress, 28 Stat., Ch. 131, of electrical measure in the United States shall be as follows:
p. 102 First. The unit of resistance shall be what is known as
the international ohm, which is substantially equal to one
(Public—No. 105) thousand million units of resistance of the centimeter-
gram-second system of electro-magnetic units, and is
An Act To define and establish the units of represented by the resistance offered to an unvarying
electrical measure electric current by a column of mercury at the temperature
of melting ice fourteen and four thousand five hundred
Be it enacted by the Senate and House of Representatives and twenty-one ten-thousandths grams in mass, of a con¬
of the United States of American in Congress assembled, stant cross-sectional area, and of the length of one hundred
That from and after the passage of this Act the legal units and six and three tenths centimeters.
45/31
Second. The unit of current shall be what is known as charged to a potential of one international volt by one
the international ampere, which is one-tenth of the unit international coulomb of electricity.
of current of the centimeter-gram-second system of electro¬ Sixth. The unit of work shall be the Joule, which is
magnetic units, and is the practical equivalent of the equal to ten million units of work in the centimeter-gram-
unvarying current, which, when passed through a solution second system, and which is practically equivalent to the
of nitrate of silver in water in accordance with standard energy expended in one second by an international ampere
specifications, deposits silver at the rate of one thousand in an international ohm.
one hundred and eighteen millionths of a gram per second.
Seventh. The unit of power shall be the Watt, which
Third. The unit of electro-motive force shall be what
is equal to ten million units of power in the centimeter-
is known as the international volt, which is the electro¬
gram-second system, and which is practically equivalent
motive force, that steadily applied to a conductor whose
to the work done at the rate of one Joule per second.
resistance is one international ohm, will produce a current
of an international ampere, and is practically equivalent Eighth. The unit of induction shall be the Flenry,
to one thousand fourteen hundred and thirty-fourths of which is the induction in a circuit when the electro-motive
the electro-motive force between the poles or electrodes force induced in this circuit is one international volt while
of the voltaic cell known as Clark’s cell, at a temperature the inducing current varies at the rate of one Ampere per
of fifteen degrees centigrade, and prepared in the manner, second.
described in the standard specifications. Sec. 2. That it shall be the duty of the National
Fourth. The unit of quantity shall be what is known as Academy of Sciences to prescribe and publish, as soon as
the international coulomb, which is the quantity of elec¬ possible after the passage of this Act, such specifications
tricity transferred by a current of one international ampere of details as shall be necessary for the practical application
in one second. of the definitions of the ampere and volt hereinbefore given,
Fifth. The unit of capacity shall be what is known as and such specifications shall be the standard specifications
the international farad, which is the capacity of a condenser herein mentioned.
IX. Appendix 2
International Conference on Electrical electromotive force and current when set up under the
Units and Standards, 1908 conditions specified in Schedule C.
In cases in which it is not desired to set up the Standards
REPORT provided in the resolutions Schedule B, the Conference
recommends the following as working methods for the
The Conference on Electrical Units and Standards for realisation of the International Ohm, the Ampere, and the
which invitations were issued by the British Government, Volt.
was opened by President of the Board of Trade, the Right 1. For the International Ohm.
Hon. Winston S. Churchill, M. P., on Monday, 12th The use of copies, constructed of suitable material
October 1908, at Burlington House, London, S. W. and of suitable form and verified from time to time,
Delegates were present from twenty-two countries, and of the International Ohm, its multiples and sub¬
also from the following British Dependencies, namely, multiples.
Australia, Canada, India and the Crown Colonies. 2. For the International Ampere.
It was decided by the Conference that a vote each should
(a) The measurement of current by the aid of a cur¬
be allowed to Australia, Canada and India, but a vote was
rent balance standardized by comparison with a
not claimed or allowed for the Crown Colonies.
silver voltameter;
The total number of delegates to the Conference was
forty-six, and their names are set out in Schedule A to or (b) The use of a Weston Normal Cell whose electro¬
this Report. motive force has been determined in terms of
The officers of the Conference were; the International Ohm and International Ampere,
and of a resistance of known value in Interna¬
President tional Ohms.
3. For the International Volt.
The Right Hon. Lord Rayleigh, O. M., President of the
Royal Society. (a) A comparison with the difference of electric
Vice-Presidents potential between the ends of a coil of resistance
of known value in International Ohms, when
Professor S. A. Arrhenius. M. Lippmann carrying a current of known value in Interna¬
Dr. N. Egoroff. Dr. S. W. Stratton. tional Amperes;
Dr. Viktor Edler von Lang. Dr. E. Warburg. or (b) The use of a Weston Normal Cell whose electro¬
motive force has been determined in terms of the
Secretaries
International Ohm and the International Am¬
Mr. M. J. Collins. Mr. C. W. S.
pere.
Mr. W. Duddell, F. R. S. Crawley.
Mr. F. E. Smith. The duties of specifying more particularly the condi¬
tions under which these methods are to be applied has
The Conference elected a Technical Committee to draft been assigned to the Permanent Commission, and pend¬
specifications and to consider any matter which might be ing its appointment to the Scientific Committee to be
referred to the Committee and to report to the Conference. nominated by the President (see Schedule D), who will
The Conference and its Technical Committee each held issue a series of Notes as Appendix to this Report.
five sittings. The Conference has considered the methods that
As a result of its deliberation the Conference adopted should be recommended to the Governments for securing
the resolutions and specifications attached to this report uniform administration in relation to Electrical Units and
and set out in Schedule B, and requested the Delegates to Standards, and expresses the opinion that the best method
lay them before their respective Governments with a view of securing uniformity for the future would be by the
to obtaining uniformity in the legislation with regard to establishment of an international Electrical Laboratory
Electrical Units and Standards. with the duties of keeping and maintaining International
The Conference recommend the use of the Weston Electrical Standards. This Laboratory to be equipped
Normal Cell as a convenient means of measuring both entirely independently of any National Laboratory.
46/32
The Conference further recommends that action be taken II. As a system of units representing the above and
in accordance with the scheme set out in Schedule D. sufficiently near to them to be adopted for the purpose of
Signed at London on 21st October, 1908. electrical measurements and as a basis for legislation, the
By the Delegates of the Countries above written. Conference recommends the adoption of the International
For the United States of Ohm, the International Ampere, and the International
America: S. W. Stratton.v Volt defined according to the following definitions:
Henry S. Carhart. III. The Ohm is the first Prinary Unit.
Edward B. Rosa. IV. The International Ohm is defined as the resistance
For Austria: Victor Von Lang. of a specified column of mercury.
Ludwig Kusminsky. V. The International Ohm is the resistance offered to
For Belgium: P. Clement. an unvarying electric current by a column of mercury at
For Brazil: Leopold J. Weiss. the temperature of melting ice, 14.4521 grammes in mass,
For Chile: Victor Eastman. of a constant cross sectional area and of a length of 106.300
For Colombia: Jorge Roa. centimetres.
For Denmark and Sweden: Svante Arrhenius. To determine the resistance of a column of mercury in
For Ecuador: C. Nevares. terms of the International Ohm, the procedure to be
For France: G. Lippmann. followed shall be that set out in Specification I attached
J. Rene Benoit. to these Resolutions.
T. De Nerville. VI. The Ampere is the second Primary Unit.
For Germany: E. Warburg. VII. The International Ampere is the unvarying
W. Jaeger. electric current which, when passed through a solution of
St. Lindeck. nitrate of silver in water, in accordance with Specification
For Great Britain: Rayleigh. II attached to these Resolutions, deposits silver at the rate
J. Gavey. of 0.00111800 of a gramme per second.
R. T. Glazebrook. VIII. The International Volt is the electrical pressure
W. A. J. O’Meara. which, when steadily applied to a conductor whose resist¬
A. P. Trotter. ance is one International Ohm, will produce a current of
J. J. Thomson. one International Ampere.
For Guatemala: Francisco de Arce. IX. The International Watt is the energy expended per
For Hungary: Harsdnyi Desire. second by an unvarying electric current of one Interna¬
Vater Joisef tional Ampere under an electric pressure of one Inter¬
For Italy: Antonio Roiti. national Volt.
For Japan: Osuke Asano.
Shigeru Kondo. Specification I.
For Mexico: Alfonso Castello.
For Netherlands: Dr. H. Haga. Specification Relating to Mercury Standards of Resistance
For Paraguay: Max. F. Croskev.
For Russia: N. Egorolf. The glass tubes used for mercury standards of resistance
L. Swentorzetzky. must be made of a glass such that the dimensions may
For Spain: Jose Ma. de Madariaga. remain as constant as possible. The tubes must be well
A. Montenegro. annealed and straight. The bore must be as nearly as
For Switzerland: Dr. H. F. Weber. possible uniform and circular, and the area of cross-section
P. Chappuis. of the bore must be approximately one square millimetre.
Jean Landry. The mercury must have a resistance of approximately
For Australia: C. W. Darley. one ohm.
Threlfall. Each of the tubes must be accurately calibrated. The
For Canada: Ormond Higman. correction to be applied to allow for the area of the cross-
For Crown Colonies: P. Cardew. section of the bore not being exactly the same at all parts
For India: M. G. Simpson. of the tube must not exceed 5 parts in 10,000.
M. J. Collins.
The mercury filling the tube must be considered as
In the presence of: W. Duddell.
bounded by plane surfaces placed in contact with the
C. W. S. Crawley.
ends of the tulae.
F. E. Smith.
Secretaries. The length of the axis of the tube, the mass of mercury
the tube contains, and the electrical resistance of the mer¬
cury are to be determined at a temperature as near to 0° C,
Schedule B as possible. The measurements are to be corrected to 0° C.
For the purpose of the electrical measurements, end
RESOLUTIONS vessels carrying connections for the current and potential
terminals are to be fitted to the tube. These end vessels
1. The Conference agrees that as heretofore the magni¬ are to be spherical in shape (of a diameter of approxi¬
tudes of the fundamental electric units shall be deter¬ mately four centimetres) and should have cylindrical
mined on the electro-magnetic system of measurement with pieces attached to make connections with the tubes. The
reference to the centimetre as the unit of length, the outside edge of each end of the tube is to be coincident
gramme as the unit of mass and the second as the unit of with the inner surface of the corresponding spherical end
time. vessel. The leads which make contact with the mercury
These fundamental units are (1) the Ohm, the unit of are to be of thin platinum wire fused into glass. The point
electric resistance which has the value of 1,000,000,000 in of entry of the current lead and the end of the tube are to
terms of the centimetre and second; (2) the Ampere, the be at opposite ends of a diameter of the bulb; the potential
unit of electric current which has the value of one-tenth lead is to be midway between these two points. All the
(0.1) in terms of the centimetre, gramme, and second; (3) leads must be so thin that no error in the resistance is intro¬
the Volt, the unit of electromotive force which has the value duced through conduction of heat to the mercury. The
100,000,000 in terms of the centimetre, the gramme, and filling of the tube with mercury for the purpose of the
the second; (4) the Watt, the Unit of Power which has the resistance measurements must be carried out under the
value 10,000,000 in terms of the centimetre, the gramme, same conditions as the filling for the determination of the
and the second. mass.
47/33
The resistance which has to be added to the resistance For setting up the cell, the H form is the most suitable.
of the tube to allow for the effect of the end vessels is to The leads passing through the glass to the electrodes must
be calculated by the formula, be of platinum wire, which must not be allowed to come
into contact with the electrolyte. The amalgam is placed
0.80 in one limb, the mercury in the other.
ohm,
10637T The depolariser is place'd above the mercury and a
layer of cadmium sulphate crystals is introduced into each
where rx and r2 are the radii in millimetres of the end sec¬ limb. The entire cell is filled with a saturated solution
tions of the bore of the tube. of cadmium sulphate and then hermetically sealed.
The mean of the calculated resistances of at least five The following formula is recommended for the E. M. F.
tubes shall be taken to determine the value of Ahe unit of the cell in terms of the temperature between the limits
of resistance. 0° C. & 40° C.
For the purpose of the comparison of resistances with
a mercury tube the measurements shall be made with at
least three separate fillings of the tube. E<=E20—0.0000406(f—20°)
- 0.00000095 (t- 20°)2 + 0.00000001« - 20°)3
Specification II
48/34
X. Appendix 3.
Metric Convention. Article 7, as amended 1921 The bureau is also charged with the duty of making the
determinations relative to physical constants, a more ac¬
Art. 7. After the committee shall have proceeded with curate knowledge of which may be useful in increasing
the work of coordinating the measures relative to electric precision and further insuring uniformity in the provinces
units and when the general conference shall have so decided to which the above-mentioned units belong (article 6 and
by a unanimous, vote, the bureau will have charge of the first paragraph of article 7).
establishment and keeping of the standards of the electric It is finally charged with the duty of coordinating similar
units and their test copies and also of comparing with determinations effected in other institutions.
those standards,the national or other standards of precision.
XI. Appendix 4.
Resolutions Adopted by the Board of Directors, space permeability that, the accuracy and reproducibility
American Institute of Electrical Engineers, at of a system of electrical, units realized by such absolute
measurements would seem to be adequate for commercial,
its Meeting in Denver, Colorado, June 27, 1928 industrial and scientific purposes; and
Whereas, there is conclusive evidence that there are Whereas, the legalization of the absolute ohm and ampere
discrepancies between the statutorily established inter¬ and the units derived from them (these units to be realized
national electrical units (ohm, ampere, and volt) and the by the national standardizing laboratories) would avert the
fundamental ohm, ampere and volt which the international recurring proposals for revision of the values of the legalized
units were intended to represent, these discrepancies in the units, and would establish the electrical units on a per¬
case of the ohm and the volt amounting to approximately manent legal basis; therefor be it
one-twentieth of one per cent; and
Resolved, that the American Institute of Electrical
Whereas, differences of this magnitude are objectionably
Engineers hereby urges the Bureau of Standards and
large in comparison with the precision required and now
foreign national standardizing laboratories to under¬
being attained in the construction and use of standards
take as soon as possible, the additional researches nec¬
fundamental to all electrical measurements: therefore be it
essary in order that the absolute ohm and absolute
ampere based on the centimeter-gram-second electro¬
Resolved, that the American Institute of Electrical
magnetic system, with the absolute volt, watt and
Engineers hereby urges the Bureau of Standards and
other units derived from them, may be legalized in
foreign national standardizing laboratories to under¬
place of the international ohm and ampere and their
take, as soon as possible, the additional researches
derived units.
necessary in order that legislation to reduce these dis¬
crepancies to within acceptable limits may be enacted Resolved, further, that in order to avoid the confu¬
in the near future. sion which would result from an interim use of new
empirical units based on corrected values of the inter¬
And whereas, the present electrical units are defined by national units, the international electrical units should
statute in terms of material standards, namely, the mercury be continued in effect without any readjustment of
ohm and the silver voltameter, which it is now known only values until such time as the practicability of legalizing
approximately represent the absolute ohm and ampere the above-mentioned absolute units shall have been
and which experience has shown to have serious limitations, determined.
and Resolved, further, that copies of these resolutions be
Whereas, such progress has been made in recent years in communicated to the various national standardizing
the art of making absolute electrical measurements in laboratories and other interested bodies, by the
terms of the fundamental units of length, mass, time and Standards Committee.
XII. Appendix 5.
Translation of Resolution No. 10 adopted October national system;
Considering, however, that a number of national
6, 1933, by the 8th General Conference on
laboratories have not yet completed the measurements
Weights and Measures necessary to relate the international units to the absolute
units;
Resolution No. 10, Substitution of the absolute electrical
units for the units called “international”. It decides to postpone to the year 1935 the provisional
In accord with the first resolution relative to the elec¬ fixing of the ratio between each international unit and the
trical units proposed by the Advisory Committee and corresponding absolute unit;
approved by the International Committee on Weights For this purpose it gives to the International Committee
and Measures; the necessary authority to fix at that time and without
The Conference approves the principle of the substitu¬ waiting for another Conference, these ratios as well as
tion of the absolute system of electrical units for the inter¬ the date of adoption of the new units.
49/35
XIII. Appendix 6.
Translation of Resolutions adopted October 29,1946 ampere, volt and watt, thus defined, are identical with
by the International Committee on Weights those which were adopted at the London Conference of
1908.
and Measures The procedure to be followed for the establishment and
the conservation of the standards of reference required
RESOLUTIONS for certain chosen units is indicated in paragraphs 6 and
8, which are likewise equally intended to serve as a
Concerning the Change in Electrical Units guide for legislation.
4) Theoretical Magnitudes of the Units.—A. Definition
Resolution 1
of Mechanical Units Used in the Following Text:
The International Committee on Weights and Measures I. Unit of Force—The unit of force (in the system M. K.
meeting officially for the first time since 1937, adopts in S. (meter, kilogram, second)) is the force which gives to ?
principle the resolutions which were submitted to it by mass of 1 kilogram an acceleration of 1 meter per second
the Advisory Committee on Electricity in June, 1939. per second.0)
To adapt these resolutions to the current situation, which II. The Joule (unit of energy or work).—The joule is the
has resulted from the developments and scientific progress work done when the point of application of the M. K. S.
accomplished since 1939, it decides that: unit of force is displaced a distance of 1 meter in the
1) The date for putting in effect the absolute units direction of the force.
shall be January 1, 1948. III. The Watt (unit of power).—The watt is the power
2) The • relations for passing between the mean inter¬ which gives rise to the production of energy at the rate of
national units and the absolute units are: 1 joule per second.
B. Definition of the Electrical Units. The
1 mean international ohm = 1.00049 absolute ohm.
Committee offers the following proposals defining the
1 mean international volt =1.00034 absolute volt.
theoretical magnitudes of the electrical units.
The precision of the two relations given above will permit IV. The Ampere (unit of electric; current).—The
laboratories and industries to express all electrical quanti¬ ampere is the constant current which, if maintained in two
ties in terms of the new units, without introducing in this straight parallel conductors of infinite length, of negligible
conversion an error exceeding two units in the last place. circular sections, and placed 1 meter apart in a vacuum,
This error is hardly larger than that estimated as being will produce between these conductors a force equal to
the accuracy obtained in the national laboratories for 2X 10-7 M. K. S. unit of force per meter of length.
their absolute measurements. V. The Volt (unit of difference of potential and of
electromotive force).—The volt is the difference of electric
Resolution 2 potential between two points of a conducting wire carrying
a constant current of 1 ampere, when the power dissipated
1) Definitive Substitution of Absolute Electrical Units
between these points is equal to 1 watt.
for the International System.—In virtue of authority which
VI. The Ohm (unit of electric resistance).—The ohm is
was conferred on it by the General Conference on Weights
the electric resistance between two points of a conductor
and Measures in 1933, the International Committee on
when a constant difference of potential of 1 volt, applied
Weights and Measures announces at this time its decision
between these two points, produces in this conductor a
that the substitution of the system of practical absolute
current of 1 ampere, this conductor not being the seat of
electrical units for the international system shall be put
into effect beginning January 1, 1948. any electromotive force.
VII. The Coulomb (unit of quantity of electricity).—
The present resolutions constitute the “new announce¬
The coulomb is the quantity of electricity transported
ment’’ which the Circular Letter of January 1, 1940,
signed by the President and Secretary of the International in 1 second by a current of 1 ampere.
VIII. The Farad (unit of electric capacitance).—The
Committee on Weights and Measures, requested the
farad is the capacitance of a capacitor between the plates
various countries to await, before proceeding with any
change in units. of which there appears a difference of potential of 1 volt
2) Historic Continuity of the System.—The first definition when it is charged by a quantity of electricity equal to 1
of the practical absolute system of electrical units coulomb.
IX. The Henry (unit of electric inductance).-—The
adopted by the Committee was enunciated by the London
Conference of 1908 in the following manner: henry is the inductance of a closed circuit in which an
electromotive force of 1 volt is produced when the electric
(Here follows Schedule B of the London Conference, see Appendix 2.) current in the circuit varies uniformly at a rate of 1 ampere
per second.
31 General Considerations.—The definitions of the abso¬
lute electrical and magnetic units rest on the generally X. The Weber (unit of magnetic flux).—The weber is
recognized electromagnetic laws, which lead to a system the magnetic flux which, linking a circuit of 1 turn, pro¬
of inter-dependent relations among the various quantities duces in it an electromotive force of 1 volt as it is reduced
which are to be measured. Consequently the units can to zero at a uniform rate in one second.
be defined in various ways, according to the starting 5) Object of These Definitions.—The definitions given in
point chosen. paragraph 4) are intended solely to fix the magnitude of
In order to formulate legal decisions which are concerned the units and not the methods to be followed for their
solely with the size of the units and not with the processes practical realization. This realization is effected in accord
experimentally employed for their realization in accord¬ with the well-known laws of electro-magnetism. For
ance with the theory on which they are based, it is con¬ example, the definition of the ampere represents only a
venient to have a set of definitions, adequate for this pur¬ particular case of the general formula expressing the forces
pose, expressed in as simple and easily understood language which are developed between conductors carrying electric
as is possible. currents, chosen for the simplicity of its verbal expression.
In response to questions which have been addressed to It serves to fix the constants in the general formula which
it concerning a text, intended to serve as a guide for the has to be used for the realization of the unit.
wording of legislation, the Committee recommends con¬
sequently the adoption of the sequence of definitions given (i) It has been proposed to give the name “newton” to the M. K. S. unit of
in paragraph 4. The Magnitudes of the units: ohm, force.
50/36
The foregoing text refers to the M. K. S. system. It is The National Bureau of Standards transmitted to the
naturally possible to translate it into another system International Committee a formal recommendation re¬
(C. G. S., M. T. S., etc.) by appropriate modification by questing that the date of introduction of the absolute units
powers of ten. in practical use should be fixed as January 1, 1948.
6) Material Standards.—For practical comparisons the Moreover, this laboratory has made a very exhaustive
electrical units are represented by material standards for study including an objective and impartial discussion of
the. ohm and the volt to which are assigned appropriate all the experimental work carried out in the majority of
values expressed in absolute units. The standards for the the large scientific institutions of the world. The result
ohm at the present time take the form of resistance coils of this study together with the results of later work done
and those for the volt of voltaic cells (Weston cells for at the National Bureau of Standards is such that it seems
example.) proper to consider that the 5th decimal place has been
7) International Standards of Reference.—The values obtained in the measurement of the ratios between the
which are to be assigned to the standards of reference international and the absolute units. The ratios proposed
maintained at the International Bureau of Weights and by the National Bureau of Standards are therefore:
Measures will be fixed from time to time by the Interna¬
tional Committee, on the recommendation of the Advisory 1 mean international ohm= 1.00049 absolute ohms.
Committee on Electricity, in accord with the results of 1 mean international ampere = 0.99985 absolute ampere.
comparisons made between these standards and the na¬ 1 mean international volt = 1.00034 absolute volts.
tional standards the values of which will have been deter¬
mined directly by absolute measurements. the 6th decimal being zero.
8) National Standards of Reference.—The values to be The National Physical Laboratory declared itself in
assigned to the national standards of reference will be deter¬ accord with the proposals of the National Bureau of
mined in accordance with the results of comparisons made Standards and merely remarked that it would be preferable
with the standards of reference of the International Bureau. not to give the results of the standardization to more than
9) Ratio Between the Absolute Units and the Units of the five decimal places and to reserve strictly the use of the
International Systems.—These ratios are indicated in Reso¬ 6th decimal to the work of coordination between the
lution 1 above. national laboratories and their relations with the Inter¬
For the standards of each nation, or of particular labora¬ national Bureau of Weights and Measures.
tories, it will be necessary to take account, not only of the It is on the basis of'the above documents and on the
values indicated above for the ratios between the “mean consensus of the replies received from members of the
Advisory Committee on electricity, that the International
international units” £Lm,Am, and VM (which have been
Committee has adopted Resolution 1.
accepted by the International Committee) and the abso¬
The difference in opinion shown by the Physikalisch-
lute units, but also of the difference between the units
of the international system maintained by each laboratory Technische Reichsanstalt has not appeared such as to
justify a further delay in the introduction of the absolute
and the corresponding “mean international units.”
units which are so much desired by the electrical workers
in the majority of the nations. By force of circumstances,
RECOMMENDATION the Physikalisch-Technische Reichsanstalt has not been
For the purpose of avoiding confusion as far as possible able to participate in the recent discussions and its objec¬
tion appears to be the result of a certain doubt on the part
the Committee advises that, during the period of transi¬
of its physicists as to the accuracy which one should
tion, the adjective “international” (abbreviation: “int.”)
attribute to the ratios that appear in Resolution 1. The
be applied to the names of the electrical units for the units
reply of the Physikalisch-Technische Reichsanstalt ex¬
formerly in use and the adjective “absolute” (abbreviation:
presses the fear that the results stated by the Committee
“abs.”) for the units which the Committee has decided
to adopt. may have too provisional a character and may have to be
corrected in the future.
Annex to Resolution 1 The International Committee points out that in any
event the values for the ratios appearing in its Resolution 1
In anticipation of the meeting of the International Com¬ will be applied only on the present occasion. Any work
mittee in October, 1946, the International Bureau of on the absolute units will lead to slight future adjustment
Weights and Measures sent a circular letter to a number of the values assigned to the national standards and to
of national laboratories requesting their opinion on the those of the International Bureau. The concept of a
question of the absolute electrical units. The replies re¬ ratio between the absolute units and the international
ceived showed that on the whole the opinion was clearly units will itself lose all interest in the future with the
favorable to the immediate adoption of these units. disappearance of the latter.
XIV. Appendix 7
81st Congress, 1st Session. of the United States of America in Congress assembled, That
from and after the date this Act is approved, the legal
units of electrical and photometric measurement in the
S. 441 United States of America shall be those defined and estab¬
lished as provided in the following sections.
In the Senate of the United States January 13, Sec. 2. The unit of electrical resistance shall be the
1949 ohm, which is equal to one thousand million units of
resistance of the centimeter-gram-second system of electro¬
Mr. Johnson of Colorado (by request) introduced the magnetic units.
following bill; which wras read twice and referred to the Sec. 3. The unit of electric current shall be the am¬
Committee on Interstate and Foreign Commerce. pere, which is one-tenth of the unit of current of the cen¬
timeter-gram-second system of electromagnetic units.
A BILL
Sec. 4. The unit of electromotive force and of electric
To redefine the units and establish the standards of potential shall be the volt, which is the electromotive force
electrical and photometric measurements. that, steadily applied to a conductor whose resistance is
Be it enacted by the Senate and House of Representatives one ohm, will produce a current of one ampere.
51/37
Sec. 5. The unit of electric quantity shall be the is equivalent to the energy supplied by a power of one
coulomb, which is the quantity of electricity transferred thousand watts operating for one hour.
by a current of one ampere in one second. Sec. 10. The unit of intensity of light shall be the
Sec. 6. The unit of electrical capacitance shall be the candle, which is one-sixtieth of the intensity of one square
farad, which is the capacitance of a capacitor that is centimeter of a perfect radiator, known as a “black body”,
charged to a potential of one volt by one coulomb of when operated at the temperature of freezing platinum.
electricity. Sec. 11. The unit of flux of light shall be the lumen,
which is the flux in a unit of solid angle from a source of
Sec. 7. The unit of electrical inductance shall be the
which the intensity is one candle.
henry, which is the inductance in a circuit such that an
Sec. 12. It shall be the duty of the National Bureau
electromotive force of one volt is induced in the circuit by
of Standards to establish the values of the primary electric
variation of an inducing current at the rate of one ampere
and photometric units in absolute measure, and the legal
per second.
values for these units shall be those represented by, or
Sec. 8. The unit of power shall be the watt, which is derived from, national reference standards maintained by
equal to ten million units of power in the centimeter- the National Bureau of Standards.
gram-second system, and which is the power required to Sec. 13. The Act of July 12, 1894 (Public Law Num¬
cause an unvarying current of one ampere to flow between bered 105, Fifty-third Congress), entitled “An Act to
points differing in potential by one volt. define and establish the units of electrical measure”, is
Sec. 9. The units of energy shall be (a) the joule, which hereby repealed.
is equivalent to the energy supplied by a power of one watt
operating for one second, and (b) the kilowatt-hour, which Washington, March 1, 1949
52/38
UNITED STATES DEPARTMENT OF COMMERCE . Charles Sawyer, Secretary
NATIONAL BUREAU OF STANDARDS . A. V. Astin, Director
Francis B. Silsbee
53/1
Preface
One of the primary functions of the National Bureau of Standards is the
establishment, maintenance, and dissemination of units of measurement that
shall be uniform throughout the Nation and constant through the years. To
meet the needs of industry in the electrical field, which is developing so rapidly
in extent and in complexity, this duty has involved a continuous growth of
staff and equipment and the improving of old methods of measurement and
the inventing of new ones.
A report of the stewardship of the National Bureau of Standards with
respect to the establishment and maintenance of the fundamental electrical
units of electromotive force and resistance is given in the Bureau’s Circular
475, which describes both the older international system of electrical units
and the present absolute system, which was put into effect to replace the older
system on January 1, 1948. The present Circular carries the report further
by showing how the other electrical and magnetic units are derived from these
two, how the range of measurement is extended, and how the Bureau serves
the electrical industry by determining the corrections for apparatus submitted
for test. It is hoped that these brief descriptions of the methods that have
been found effective in the laboratories of the Bureau will be a useful guide
in the development of other laboratories that may have the duty of extending
the measurement process. This general picture of the procedures by which
accurate measurements are actually carried out should be helpful also to
educators who wish to give their students a definite and correct understanding
of the practical application of their teachings in the theory of electrical
measurements.
A. V. Astin, Director.
54/III
Contents
Page
Preface_ hi
1. Introduction_ 1
2. Fundamentals_ 1
3. Derivation of other units_ 3
3.1. Capacitance_ 4
3.2. Inductance_ 6
3.3. Current, power, and energy_ 7
3.4. Magnetic units_ 3
4. Transfer from d-c to a-e measurements_ 10
5. Extension of measurement ranges__ 13
5.1. The resistance seal*'__ . .. . 13
5.2. The scale of direct voltage_ 15
5.3. Idle scale of direct current_ 17
5.4. The scale of alternating current_ 18
5.5. The scale of alternating voltage_ 20
5.6. Measurement of crest anil surge voltage_ .. 24
6. Dissemination of units_ - 26
7. International relations_ 27
8. Bibliography_ 28
8.1. General_ 28
8.2. Absolute measurements_ 29
8.3. Standard cells_ 29
8.4. Standard resistors and resistance measurements_ 29
8.5. Potentiometers, bridges, etc_._ 30
8.6. Capacitance_ 30
8.7. Inductance (construction and measurement)__ 30
8.8. Inductance (computations)_ 31
8.9. Galvanometers and detectors_ 31
8.10. Instruments and meters_ 31
8.11. Instrument transformers_c._ 32
8.12. Magnetic measurements_ 32
5 5/IV
Extension and Dissemination of the Electrical and
Magnetic Units by the National Bureau of Standards
By Francis B. Silsbee
Starting with the ohm and the volt as maintained by groups of standard resistors and
cells, this paper describes the experimental processes by which the other electric and magnetic
units, e. g., farad, henry, ampere, watt, joule, gauss, and oersted are derived. It also de¬
scribes the series of steps by which the scales of measurement of resistance, direct and alter¬
nating current, and voltage are derived experimentally. Brief mention is made of the pro¬
cedures for the dissemination of these standards of measurement throughout the world by
the calibration of standard electrical measuring apparatus. An extensive bibliography lists
papers describing the measurement procedures in greater detail and serves as a historical
report of the work of the National Bureau of Standards in the field of electrical measurements
during its first 50 years.
1. Introduction
It is the purpose of this Circular to give an over¬ trial operations will consistently yield a product
all picture of the sequence of measuring processes of constant quality.
by which a self-consistent system of electrical Although many links in this network of measure¬
units is built up in the laboratories of the National ment are welded in remote laboratories, the Na¬
Bureau of Standards and thence disseminated tional Bureau of Standards has exerted for the
throughout the Nation. The branching chains past 50 years a very significant influence on the
of measurement thus initiated may be considered development of the system, both directly by its
as extending outward without a break from the testing service and by the invention of new meth¬
groups of primary standard cells and standard ods of measurement, and indirectly by example
resistors by which the units of voltage and of and precept. The particular experimental meth¬
resistance are maintained from year to year, and ods described herein are primarily those used in
as ending finally in the myriad measuring opera¬ the Bureau laboratories, but to a large extent the
tions throughout the Nation, on the basis of which same or closely similar methods are used elsewhere.
scientific research is conducted, manufactured The following paragraphs show the points of con¬
articles are adjusted and inspected, electric energy tact of the Bureau with the industrial world in
is bought and sold, and industrial processes are more detail. The extensive bibliography gives
precisely controlled. It is only as a result of this references in the field of electrical measurements,
self-consistency that scientists in different labora¬ not only to technical papers originating with the
Bureau staff, but also, on occasion, to other note¬
tories can talk in the same terms, that apparatus
worthy publications in this field.
formed by components manufactured in different
The extension of measurements into the radio
places at different times can be expected to func¬ field has not been included in this Circular because
tion properly, that competing sources of energy the methods used at these higher frequencies con¬
can be fairly compared, and that complex indus¬ stitute a quite distinct discipline.
2. Fundamentals
By international agreement [4, 6]1 the official consistent meter-kilogram-second-ampere, or Gi-
basis for electrical measurements throughout the orgi, system of units.
world is the system of practical absolute electrical The experimental processes by which these units
units. These units developed historically as con¬ are established in terms of length, mass, time, and
venient decimal multiples of the units of the- the conventionally assumed constant n„, usually
designated “permeability of space” are called
centimeter-gram-second electromagnetic system,
“absolute measurements.”
but they also form the electrical part of the self-
Because of the importance in such measure¬
1 Figures in brackets indicate the literature references at the end of this
ments of ready access to highly accurate stand¬
paper. ards of length, mass, and time, and because of the
56/1
long, tedious, and therefore expensive, observa¬ size, of the coil system. After such a current
tional programs required if significant accuracy balance has been constructed and its shape deter¬
is to be attained, the burden of periodic determi¬ mined by suitable measurements, it is connected
nations of this nature rests almost entirely on the in series with a resistor, the resistance of which
few larger national standardizing laboratories of is known in absolute units by reference to the
the world. Of the many 'theoretically possible former type of experiment.
forms of absolute electrical measurement, two The current is adjusted until the change in force
types of experiment have so far shown the great¬ that results when the current in one coil is reversed
est possibilities for high accuracy and, except in is balanced by the change in loading when a known
certain very special circumstances that will be weight is added to one pan of the balance. At
noted later, are universally used [17, 36]. the same time the potential drop produced by the
The first of these experiments involves the con¬ current in the known resistor is compared with the
struction of a standard of self- or of mutual in¬ electromotive force of a standard cell. By this
ductance, the geometrical configuration of which process, a value in absolute volts is obtained for
is such that (1) its significant dimensions can be the electromotive force of the cell. The cell then
directly and accurately measured [45, 46, 47] and serves to preserve the result of the experiment for
(2) its inductance can be calculated precisely future use. Such an experiment is usually called
from these measured values [142, 148, 149, 47]. an “absolute measurement of current,” although
As a preliminary for such procedures, Dr. E. B. the final result is the assignment of a value in
Rosa and his colleagues, in the early years of the absolute volts to a standard cell. Coils of a con¬
Bureau, carried on an extensive theoretical exam¬ siderable variety of shapes and construction have
ination of the various available formulas for in¬ been used in the current balance at the National
ductance and published a number of papers [132 Bureau of Standards, and, as a further check
to 141] in this field, the major results of which against systematic errors, it is planned to use a
are summarized in [135]. A large volume of radically different form, the Pellat [44, 145]
later work was done by Dr. Chester Snow and balance, in the near future.
others [142 to 159], the results of which are, of
course, applicable to the computation of the in¬
ductances of various circuit configurations for
many other purposes, as well as for absolute
measurements. After the value of the inductance
of a standard inductor in absolute henries lias
been computed, an experiment is performed by
which the value in absolute ohms of the resistance
of a standard resistor is obtained in terms of the
reactance of the standard inductor at the known
frequency used in the experiment. The net
result of the whole process is usually called an
“absolute measurement of resistance.” At the
National Bureau of Standards two completely
independent procedures have been used for this
purpose, one [38,39] starting with a self inductor
and the other [47] with a mutual inductor. The
two final results differed by only 12 ppm (parts
per million). The second of these procedures has
proved so convenient that it is planned to make
fairly frequent (perhaps annual) electrical com¬
parisons between the standard resistors and the
standard mutual inductor. It is hoped that
any secular changes in the dimensions of the in¬
ductor will be so slow that the tedious mechanical
measurements of its dimensions need be repeated
only at intervals of ten years or more.
The second type of experiment is the absolute
measurement of current by the use of a current
balance [40, 41, 42], This is an instrument bv
which it is possible to weigh accurately the me¬ Figure 1. Experimental establishment at the National
chanical force that is exerted between two coils Bureau of Standards of (he primary electrical units by
as a result of the electric currents in them. The absolvie measurements.
relation between the force and the current can The rectangles represent physical standards embodying the corresponding
units, the circles represent experimental procedures with their associated
be calculated by fundamental electromagnetic apparatus used in the measurements. The lines indicate the passage of
theory and is expressible by formulas that involve units and numerical values. The transfers to Washington of the foreign
values for it were made by the U. S. Coast and Geodetic Survey, using
only the geometrical shape, but not the absolute ‘•invariable pendulums."
58/3
3.1 Capacitance
59/4
in the range of capacitance, in spite of the inde¬
pendent bases on which the two ends rest.
The unit of dielectric constant for solid dielec¬
trics is derived from the standards of capacitance
by combining the electrically measured capacitance
of a specimen with its mechanical dimensions.
The capacitance is measured at frequencies from
60 c/s to 200 kc/s in a shielded Sobering bridge
(see fig. 6) that has been calibrated in terms of
the standard air capacitors. A major complica¬
tion arises from the fringing of the electric field
at the edges of the specimen, and a number of
approximate working formulas have been devel¬
oped for correcting for such effects [114], With
these edge-correction formulas, the accuracy is
about 1 percent. For higher accuracy, the guard¬
ring technique is used. The area of the specimen
is readily computed from its scaled dimensions,
and the accuracy is usually limited by the measure¬
ment of thickness. When feasible, this is deduced
from measurements of the mass, density, and area
of the samples. Under the best conditions,
results are repeatable to 0.1 percent.
Another method for minimizing edge effects
[117] is that of placing a 2-inch-diameter disk of
the dielectric in a micrometer electrode holder.
A capacitance measurement is made with the
specimen in place and another with it removed.
Figure 5. Cross section of standard capacitor of the guard- Either the electrodes may be moved closer together
well type. by a measured amount to give the same value
F. Guarded electrode, supported by an insulator of low expansion glass, of capacitance for the second measurement, or
J; E, high-voltage electrode; O, II, parts of guard ring. If the depth, d, of they may be left at the same spacing and the new
the well is greater than its radius, a, the direct capacitance between E and F
is very much smaller than if F is flush with the lower surface of H. value of the capacitance measured.
Standards for the measurement of loss factor
Auxiliary standards used in this range include in capacitors and dielectrics are usually air
(1) an adjustable capacitor of the guard-well type, capacitors. Studies have shown that such appa¬
in which the island electrode can be adjusted by ratus may have a loss factor as large as 1 X10_\
a micrometer screw to any desired depth in the and that if accuracies of 1X10-6 are needed, a
well, thus securing a working range 0.1 ppi down more precise standard is required. By combining
to 0.001 ppi. (2) an adjustable capacitor of a
modified Zichner type with a movable septum
between the electrodes by which the direct capaci¬
tance can be varied from 0.3 ppi down to zero
when the aperture in the septum is closed; (3)
a decade group of 0.1 ppi units, the direct capaci¬
tance of any unit of which can be reduced to
zero by inserting a grounded septum between its
electrodes; and (4) a similar group of 0.01 ppf units.
By the substitution of these standards in the
TY-arm of a commercial direct-capacitance test set
having inductive ratio arms with a fixed 1:1 ratio,
and a fixed fourth arm, they may be intercompared
at 465 kc/s, and other standards of direct capaci¬
tance up to 5 ppi can be measured. It is also
possible to step down from standards of higher
value with a shielded Schcring bridge to 5 ppi,
and at this level the two procedures have been
found to agree within 0.05 percent. The value
derived by extending this step-down process to a
0.1 -ppf capacitor was found to differ from the
value derived independently from the computable
standards by less than 0.1 percent. Because of Figure 6. Modified Sobering bridge and accessories for
measuring dielectric constant and loss factor tip to 200 kc/s.
these close agreements, in the range of overlap,
no significant inconsistencies can arise anywhere Oscillators arc below, detectors above, micrometer electrode holder at left.
60/5
(lata obtained with (a) a capacitor of constant of measurement of inductances in the Maxwell-
spacing and adjustable area and (b) a capacitor Wien bridge.
of constant area and adjustable spacing, a valid The circuit regularly used to measure induct¬
determination of the true zero of power factor can ance is that of the Maxwell-Wien bridge [17, p. 113]
be obtained [110, lit]. shown in figure 7. The a-c source is connected
Capacitors have recently come into use for across bd and the detector (a vibration galvanom¬
measuring the very small direct currents encoun¬ eter for 60 c/s, a telephone receiver with suitable
tered in ionization measurements. Here very low amplifier for 1,000 c/s, or an electronic null
absorption and leakage are essential qualities. detector) is connected across ca. The adjustable
The usual calibration procedure is to allow a capacitor, f\ may be either the group of air
known current to charge the capacitor for a capacitors shown in figure 4, which has been
measured time and to note the resulting voltage calibrated directly by the Maxwell commutator
with an electrometer. The current is chosen so bridge (H, figure 2) or a mica capacitor (B, figure
as to make the total charging time and the final 2) calibrated by the type-12 bridge at the particular
voltage approximately the same as they will be in frequency at which the inductor is to be tested,
the future use of the capacitor. usually 60 c/s or 1,000 c/s. This procedure is
convenient for inductors from 1 /zh to 10 h, and
3.2 Inductance gives results good to 0.01 percent over the middle
of this range.
In the practical work of inductance measure¬ Mutual inductances for use with alternating
ment at the National Bureau of Standards it current and of ranges up to about 50 mb are
usually has been found more expedient to derive regularly measured by the Curtis method [17,
the henry from the ohm and the farad than to p. 117] with the circuit as shown, including the
use directly the values computed from the dimen¬ dotted lines in figure 7. With switch S closed to
sions of the large computable inductors constructed the right, capacitor C is adjusted to a value CL,
for absolute measurements. These inductors are which is a measure of Lx. Switch S is then thrown
very bulky and hence have relatively large to the left and C readjusted to a new balance at
capacitance to ground and considerable electro¬ CM- The mutual inductance is then computed
magnetic coupling to other circuits. They are from the relation
wound with thick wire (0.7 mm) and would show
appreciable skin effect if used at high frequencies. M=(Cl-Cm)
However, they are very valuable for use occasion¬
ally to give an independent check on the accuracy with further correction terms involving the
residual inductances of the resistance arms.
For inductors of larger value, or at higher
c frequencies where the effects of distributed capac¬
itance are larger, this method becomes impracti¬
cal. In such cases, the primary and secondary
windings are connected in series, first aiding and
then opposing, and the self-inductance of each
combination is measured with a Maxwell-W ien
bridge. If the observed values are LA and L0l
respectively, and if LP and^Lg are the inductances
of the primary and secondary windings measured
separately,
4 2 2
61/6
Mutual inductors of the types used to calibrate be computed and the resistance of which is nomi¬
the ballistic galvanometers employed in measuring nally equal to that of the impedance under test,
the d-c properties of ferromagnetic materials (see is substituted for the unknown in the A-arm of
sect. 3.4) are tested ballistically. For this pur¬ the bridge. Each of the computable standards
pose such inductors are compared with the large for 10,000 and 1,000 ohms consists o'f a fine wire
computable mutual inductor that was constructed stretched down the axis of a brass tube about 5
primarily for the absolute measurement of resist¬ cm in diameter. For 100- and 10-ohm standards,
ance by the Wenner method [47], To do this, parallel wires are used. The time constants
the primary windings of the standard and test (L/R) of these standards can be computed to
inductor are connected in series. Their secondary 1 x 10-8 sec. For still lower resistances four-
windings are connected in opposition and with two terminal standards are used, either of the parallel-
resistance boxes also in the series circuit. A wire, reflexed-strip, or coaxial-tube type [128, 129].
galvanometer is bridged across between the Some of the coaxial tube type have their potential
junction of the resistance boxes and the junction leads inside the inner tube, whereas, others have
of the two secondary windings. The resistances them outside the outer tube. These four-terminal
are then adjusted until the galvanometer remains standards of inductance range in resistance from
balanced as the primary current is reversed. 0.1 to 0.0002 ohm and in current rating from 10
For this condition, the mutual inductances are to 2,500 amp.
proportional to the total resistance of the two
parts of the secondary circuit. This method 3.3 Current, Power, and Energy
avoids any complications that might enter an
a-c measurement because of capacitance in the
windings. The block diagram of figure 8 shows the deriva¬
tion of the units of current, power, and energy
At low values of inductance, errors from the
residual phase defects of the bridge arms may from those of resistance, voltage, and time. An
become appreciable. Hence, for the measurement electric current (if one excepts the persistent cur¬
of the residual inductances of nominally "non- rent in a superconducting circuit) is in the nature
inductive” two-terminal resistors [125, 126] it of things, a transitory affair, and a "standard
becomes essential to use a substitution method. ampere” cannot be preserved on a laboratory
In this a standard, the inductance of which can shelf as can a standard ohm. For the precise
measurement of direct current the normal pro¬
cedure is to insert a known resistance, R, in the
circuit and to compare the potential drop pro¬
duced in this resistance by the unknown current
with the emf of a standard cell. This comparison
is usually made with a potentiometer, but in rare
cases, where extreme accuracy is needed, special
resistors are used of such value that the IR drop
at the desired current is very closely equal to the
emf of the standard cell. Thus the circuit be¬
comes very simple, and errors from contacts,
thermal electromotive forces, and drift in an
auxiliary potentiometer current are minimized or
eliminated. With a potentiometer and suitable
standard resistor, it is easily possible to measure
a direct current much more accurately than with
an ammeter. Thus indicating ammeters of the
/jo-, }{-, and b-ppreent-accuracy classes are readily
tested, and the true values of current correspond¬
ing to a number of scale points are determined.
Such ammeters can in turn be used to check amme¬
ters of lower accuracy and also to check recording
instruments.
The measurement of power in a d-c circuit, as
for instance, in calibrating a calorimeter, can lx*
made most accurately by using a potentiometer
with an accessory shunt and volt box to measure
the current and the voltage of the circuit sep¬
arately and by then taking their product. The
checking of a wattmeter is done in the same way,
except that in this case separate storage batteries
are used to supply the. current and the voltage
Figure 8. Development of units and standards for current, circuits, as shown in figure 9. With switch Sre„
power, ami energy on direct and on alternating current. closed in either direction and with switch SAD
62/7
black spot on the meter disk and opens the circuit
of a synchronous timer after a predetermined
integral number of revolutions. The timer, or in
the case of the portable meter, the stepping relay,
derives its time signals from a multivibrator
operating in synchronism with the 100-kc standard
cpiartz oscillators.
The test of an a-c watthour meter is similar
except that not only the magnitudes of the cur¬
rent and the voltage, but also the phase angle
between them must be held constant. This is
accomplished by using the special astatic electro-
dynamic transfer wattmeter described in section
4. This wattmeter is first standardized for a
particular value of power using direct current
and voltage, which are measured with a poten¬
tiometer. It is then connected with its current
coil in series with the current circuit of the a-c
Figure 9. Circuits for transfer testing a wattmeter. watthour meter under test and with its voltage
With switch Sad closed to the left, both the test wattmeter and the stand¬ circuit in parallel with that of the watthour meter.
ard transfer wattmeter are energized with direct currents from batteries- The two circuits are then energized from separate
B\ and B2. With switch Sad closed to the right, both wattmeters are ener¬
gized with alternating currents from the two coupled generators, G1 and G2. transformers that are supplied from two coupled
Ammeters and voltmeters Aa, An, Va, Vd indicate approximate values of
current and voltage for control and for setting power factor. Potentiometer generators driven by a common motor. The cur¬
gives more exact measures of d-c voltage and current for d-c calibrations. rent and voltage measured by an auxiliary am¬
Switch Sre, by reversing direction of currents in wattmeter circuits permits
elimination of errors from the earth’s magnetic field. Voltmeter Vserves as meter and voltmeter are adjusted separately to
electrostatic tie and as indicator of stray leakage between batteries and
ground. the desired values by control of the generator
fields, and the stator of one generator is then
closed to the left, the voltage circuits of both the shifted in angular position until the desired rela¬
test wattmeter and the standard transfer watt¬ tive phase of current and voltage is secured. This
meter are energized by battery Bx, and the voltage condition is indicated by the reading of the trans¬
is measured both by voltmeter VD and more fer wattmeter, which then gives the same reading
accurately by the combination of the volt box, as on the earlier d-c calibration. One observer
VB, and potentiometer (with switch Sp closed up¬ then holds the transfer wattmeter at this constant
ward). The current circuit of the test watt¬ reading by occasional slight readjustments of the
meter and of the standard transfer wattmeter are field rheostats while a second observer counts and
energized by battery B2, and the current is meas¬ times the revolutions of the watthour meter as
ured by ammeter AD and more accurately by the described above for d-c watthour meters. At the
combination of the standard resistor, Bs. and the end of each “run” the d-c standardization of the
potentiometer. By this use of separate sources, wattmeter is repeated, and a small correction is
it becomes possible to test a wattmeter of high applied if any drift in its calibration has occurred.
range (many kilowatts) under its normal condi¬ This procedure is much more elaborate and time-
tions of use, while consuming only a few watts in consuming than those used by power companies
each of the two measuring circuits. A worth¬ in their routine testing of large numbers of meters.
while precaution is to connect the two (otherwise Its use is justified (a) because the NBS normally
insulated) circuits at one point, near the low- receives for test only master standards of high
voltage end of the voltage circuit, to eliminate quality, by means of which many other meters
any possibility of electrostatic forces affecting the are standardized, and (b) because the relatively
position of the movable system. The use of a small number but considerable variety of meters
voltmeter (V, fig. 9) as such an “electrostatic submitted would not justify a large investment
tie” gives warning of any serious leakage from one in special automatic test equipment.
battery to the other via the ground. In a project, currently active, it is planned to set
Electric energy is, of course, the integral of up a group of standard a-c watthour meters that
power with respect to time, and the normal way are to be standardized at long intervals by the
to test a d-c watthour meter is to maintain, by method just described and used as working stand¬
using a potentiometer, a known and constant cur¬ ards for the more convenient testing of other
rent and voltage at its terminals for a measured meters. No definite data are yet available on the
time. In the testing of portable watthour meters practicability of this procedure.
(“rotating standards”) the voltage and current
are held constant, and the voltage circuit is closed 3.4 Magnetic Units
by a relay and opened automatically by a stepping
relay after the lapse of a predetermined number of The unit of magnetic flux, the weber, is derived
seconds (usually 100). For a “house-type” meter from the henry and the ampere, the latter in turn
a photocell pickup counts the passages of the being derived from the volt and the ohm. The
63/8
procedure is to change, by a known amount, the outside. Hence, the value thus computed applies
current in the primary winding of a mutual in¬ to the interior of the specimen also. To approxi¬
ductor and to note the resulting deflection of a mate the desired distribution of magnetizing
ballistic galvanometer, which, together with the forces, the currents in auxiliary coils near the
test coil that is to be used in later measurements, joints and yokes as well as the current in the main
is connected to the secondary winding of the magnetizing coils are adjusted until the flux in the
mutual inductor. The value of the mutual in¬ central part of the specimen is equal to the
ductance in henries is obtained by comparing the average of the fluxes at two sections at the
working inductor with a large standard computable quarter points of its length. For high accuracy it
mutual inductor. This computable standard has is essential that, the test specimen be very uniform
an inductance of about 11 mb, and the working throughout its length [237].
standard inductors range from about ^ to 50 mil. The second method for realizing the unit of
The value in amperes of the change in current is magnetizing force, H, is to measure the tangential
determined by a potentiometer, which compares component of magnetic induction, B, in the air
the emf of a standard cell with the change in IR closely adjacent to a test specimen. This is
drop in a known resistor. The change, ArA4>, in numerically equal, in the centimeter-gram-seeond-
flux linkages with the N turns of the secondary electromagnetic system, to H in the air, and by a
winding of the working standard inductor, which fundamental theorem the tangential component
results from the change, A/, in the primary current of II is continuous across the boundary and into
is given simply by the specimen. The measurement of B in the air
NA<t>=MM. is preferably made by quickly rotating a pair of
“flip coils” of known area and number of turns
For convenience, the primary current is merely through 180°. The axes of these coils are parallel
reversed so that Al=2l. If / is in amperes and to the axis of the specimen at the beginning and
M is in henries, ATA</> is in webers. Normally end of the rotation. This method is used in the
the resistance of the complete secondary circuit High-// [241, 242] and the MH permeameters
(including all test coils) is adjusted so as to make [243], The latter is shown in figure 10.
the galvanometer scale direct reading [244], Because the value of II may vary with distance
The weber as thus established is used to measure out from the surface of the specimen, by reason of
the flux in test specimens in various types of per- leakage flux from other parts of the circuit, it is
meameter. The test specimen in a permeameter necessary to apply a suitable correction to the
is surrounded by a “B-coil,” of a known number value observed by the “flip coil” nearest the
of turns, which is connected in series with the surface. The difference between the flux linked
secondary winding of the calibrating mutual in¬ by this flip coil and the second located at a known
ductor in the galvanometer circuit. A measured greater distance from the surface of the specimen
change in the magnetizing current in the perme¬ gives a basis for extrapolating to the value at the
ameter produces a change in the flux linking the surface. Such multiple coils are provided in the
galvanometer circuit, and the resulting deflection High-// and MH permeameters. By applying
measures the flux change. such corrections and by careful calibration of the
A measurement of the cross-sectional dimen¬ galvanometer, accuracies of 0.5 percent in B or II
sions of the test specimen then permits the com¬ can be obtained. In ordinary testing practice, the
putation of the change in magnetic induction, accuracy is about 1 percent for either B or II.
AB=A(f)/A. If A is in square meters, AB is in Alternatively a fixed H-coil, adjacent to the
webers/meter2 (myriagausses). This result multi¬ specimen may be used and the magnetizing force
plied by 104 gives the induction in gausses.
Normally the area of the specimen and number of
turns in the A-coil are lumped with the other
factors in the calibration of the galvanometer, so
that the induction is read directly.
Two distinct methods are used for establishing
the unit of magnetizing force. In the older
method, exemplified by the Burrows Permeameter
[236, 249], the magnetomotive f rce is distributed
around the magnetic circuit in proportion to the
reluctance of the various portions so that all
portions carry the same flux and are at substan¬
tially the same magnetic potential. If this condi¬
tion is attained, the magnetizing force, //, at any
section, such as the central portion of the test
specimen, is directly computable from the current- Figure 10. The MH permeameter.
turns per unit length at that section. At the The bar specimen lies along the axis of the central coil. The four auxiliary
surface of the specimen the tangential component coils supply magnetomotive force for the joints and yokes. The plunger of
the iron-clad solenoid at the left “flips” the H-coils, which are located just
of H is, of course, the same inside the metal as below the specimen near its center.
64/9
deduced from the change in flux linkage in this the induced voltage. The derivation of the
coil when the magnetizing current of the perme- corresponding units is given elsewhere.
ameter is reversed. This is the normal procedure Recent work at the National Bureau of Stand¬
with the widely used Faliv Simplex Permeameter ards has made possible an alternative method for
[248, 250, 252].' measuring with extremely high accuracy the mag¬
Here the H-coil extends between two blocks of netic induction in fields above 1,000 gausses.
soft iron clamped to the ends of the specimen and The studies of Bloch [253] and of Purcell and
responds to the average value of the magnetic his colleagues [254] have shown that if a substance
potential gradient between the blocks. For is placed in a strong fixed magnetic field of intensity
accurate work a correction factor must be applied B0, on which is superposed a weak radio-frequency
to relate this to the true magnetizing force acting field, the material will show a resonance absorption
on the specimen. This factor is determined by at a particular frequency, /0. This resonant fre¬
the use of standard test bars that have been first quency is given by /0=yP0/27r, where y is the
calibrated in an absolute permeameter. gyromagnetic ratio of the atomic nuclei of the
With permeameters of these types, measure¬ substance used. At the Bureau, Thomas, Dris¬
ments of the normal induction curve, hysteresis coll, and Hippie [255] have performed this experi¬
loops, and of residual induction and coercive force ment with great care, using a material (water)
[244] are regularly made on sample bars sub¬ rich in protons and have determined the gyro-
mitted for test by the laboratories of steel com¬ magnetic ratio of the proton to be
panies and other organizations. The circulation
y= (2.67528 ± 0.00006) X 104 sec-1 gauss-1.
of such bars insures uniformity and accuracy in
this branch of industry. Also by such comparisons In this work the gauss was established by
the performances of permeameters of other types observing the mechanical force exerted by the
have been studied in detail [246 to 252], A field on the conductors that formed one side of a
closely related service is the calibration of search rectangular “force coil’’ hanging in the field and
coils to determine magnetically their effective area- carrying a measured current. The current was
turns. This is done by placing the coil near the measured in terms of a standard cell and standard
center of a long, uniformly wound, single-layer resistor. The force was measured by balancing
solenoid with the axis of the search coil parallel it against the weight of a known mass that could
to that of the solenoid. The mutual inductance, be added to or removed from the pan of the preci¬
M, between the coil and the solenoid winding is sion balance from which the force coil was sus¬
measured ballistic-ally. The area-turns, AN, of pended. The effective length of the conductors
the coil are then computed by the equation in the bottom side of the coil was measured by
a special micrometer in comparison with a known
AN=MI/B. end standard.
The ratio of the central flux density, B, td the By using this value of y and a resonance probe
current, I, is a constant that is computed from similar to the one developed for this work, other
the measured dimensions and pitch of the solenoid experimenters can measure the induction in an
winding. unknown magnetic field by a simple determination
Measurements of magnetic properties with of the resonant frequency in terms of the standard
alternating currents involve either power measure¬ radio-frequency emissions from NBS radio station
ments with wattmeters or the use of a-c bridges WWV [25]. Such a resonance detector can also
or potentiometers. The maximum cyclic change be used (1) to study the distribution of flux
in flux is measured by using a voltmeter of high density in a magnetic field and (2) with suitable
resistance (400 olnns/v) or more with a copper- additional circuits as a monitor to hold constant
oxide rectifier to indicate the average value of the induction at a given place in the field.
65/10
from an ideal electro dynamic instrument. The
first was a spurious torque due to eddy currents in
the aluminum damping vane. This effect is less
than 2 ppm at 60 c/s but increases as the square of
the frequency and has to be corrected for at
frequencies above 1,000 c/s. The second is an
effective phase defect that increases as the first
power of the frequency and at 60 c/s amounts to
only 0.38 minute, i. e., to an error of 0.011 percent
of the volt-amperes, even at zero power factor.
This transfer instrument is used principally in
checking the accuracy of electrodynamic watt¬
meters of the /10-percent or the h-percent class,
which in turn are used as standard transfer in¬
struments in other laboratories for the checking
of other wattmeters or watthour meters using
alternating current. In the usual test, the un¬
known and the standard transfer wattmeter are
connected with their current circuits in series and
their voltage circuits in parallel, as shown in
figure 9. With switch SAD closed to the right,
Figure 11. Electrodynamic ac-dc transfer instrument. alternating current and voltage are applied from
The links beside tlio instrument proper permit connecting the fixed coils separate transformers, T2 and 7j. and their rel¬
in parallel or series. The dials below the scale adjust the resistance is series ative phase is adjusted by shifting the stator of
with the moving coil.
G to give the desired power factor (usually 0.5).
The basic electrodynamic transfer instrument This can be calculated accurately enough from
for current, and power at power frequencies [177] is the readings of the wattmeter under test and of
shown in figure 11. It is an astatic electrodynamic the auxiliary ammeter, Aa, and voltmeter, VA.
wattmeter, the moving coils of which are carried The series resistor of the standard instrument is
by a taut suspension of phosphor-bronze strip. reduced until the spot of light comes to a con¬
Tbe fixed coils are rated at 2.5, 5, or 10 amp, venient position near the center of the scale, where
according as they are connected in series, series- the axes of the moving coils are at right angles to
parallel, or parallel. They are supported in slabs those of the fixed coils and the mutual inductance
of plate glass to which adjustable supports and a between the two circuits is zero. One observer
torsion head for the suspension are firmly attached. then adjusts the current to bring the pointer of
The angular position of the pair of movable coils the instrument under test as exactly as possible
is read by a mirror and scale at a radius of 2 m. on some scale mark, usually using either a pro¬
The movable coils are used in the position where jection equipment or a magnifying glass. The
the mutual inductance between them and the other observer reads as Hi the position of the spot
fixed coils is substantially zero. The deflection is of light on the instrument scale. The four-pole
usually about 60°. A capacitor shunts part of the double-throw switch, SAD, is then closed to the
series resistor and thus compensates for the self¬ left to substitute previously adjusted direct cur¬
inductance of the movable coils. A change in de¬ rent and voltage in place of the a-c quantities.
flection of 1 mm corresponds to 0.025 percent in Current is again adjusted until the instrument
power and (with the fixed coils in series) to a under test reads the same, and the standard in¬
change of 2X10-6 (amp)2 in the product of the strument reading, Dx, is recorded for this setting.
currents in the fixed and moving coils. Both current and voltage are then quickly re¬
Even though the electrodynamic transfer instru¬ versed by switch Sree and a second d-c reading,
ment was carefully designed to be as free as practi¬ D2, is obtained. The original connections are
cable from such sources of a-c error as eddy cur¬ then restored and a check a-c reading, A>, is re¬
rents, capacitance, and inductance, a great, many corded. The average of D[ and D2 is free from
checks were made on its performance before it was any effect of the earth’s magnetic field. The
put into commission as a transfer instrument. average of A\ and A2 corresponds to substantially
These included (1) a comparison up to 3,000 c/s the same time of observation as does the mean of
when connected as an ammeter, with a Hartman Dx and D2. Hence any linear drifts like t he effects
and Braun expansion-type hot-wire ammeter, (2) of spring fatigue or of heating in either instrument
a comparison at 60 c/s and full volt-amperes but cancel out, and their difference, when divided by
zero power factor with a sensitive quadrant elec¬ the mean equivalent deflection (3), B,
trometer, and (3) the measurement at 500, 1,000, s The standard transfer instrument is used with a uniform millimeter scale,
and 2,000 c/s of the loss in a 10-^if mica capacitor, the divisions of which do not exactly correspond with equal increments in
the measured power over the whole range. Hence to obtain relative changes
the phase defect of which had been previously in power from the observed changes in deflection, it is convenient to divide
measured in an a-c bridge. These check tests the latter by an “effective deflection.” P, defined as P= P(dDtdP), where
dDldP is the slope of the graph of deflection, O, versus power, P, at the part
showed the existence of two small departures of the scale used.
66/11
Aj -\-Ai—D\ —D2 of energy dissipation in a noninductive standard
w resistor that is in series with the ammeter under
test. Readings are made with alternating and
is a measure of the real difference in the perform¬ direct current alternately, as described above.
ance of the instrument under test on alternating It lias been found more convenient, however, in
current as compared with direct current. In the normal testing of a-c ammeters to use, as an
practice three such sets of four readings each are intermediate standard, an ammeter of the com¬
taken in rapid succession, and with a wattmeter posite-coil type [180], The a-c performance of
of good quality the mean difference thus obtained this instrument has been carefully compared with
and divided by the effective deflection correspond¬ that of the fundamentally simpler astatic watt¬
ing to the d-c reading will be found repeatable to meter, and it is regularly used on alternating
±0.02 percent. current, often with a standardized current trans¬
The performance of many wattmeters on alter¬ former, to check ammeters on alternating current.
nating-current can often be represented approxi¬ The basic instrument for the transfer of voltage
mately by an equation of the form from direct current to alternating current at
power frequencies is a suppressed-zero electro-
Pa=Pd{\ +a) -\-bEI sin 0, (1) dynamic voltmeter [179] in which the movable
system is carried by a taut suspension and its
where Pa is the reading on alternating current, position is read with a telescope and scale at a
and Pa is the reading on direct current for the scale distance, as now installed, of 3 m. A short
same true power, E and / are voltage and current, boom attached to the moving system is free to
and cos 0 is the power factor. A transfer test first play between two stops, only a few degrees apart
made at unity power factor (sin 0 = 0) gives a and insures that the suspension is held at all times
measure of a, which is usually very small. This with substantially the same twist as that present
combined with a second test at cos 0=0.5 (sin when readings are made. As normally used in the
0=0.866) gives a measure of b. transfer testing of voltmeters, the double optical
If eq (1) represented completely the performance lever and temperature compensating coils de¬
of a wattmeter, it would be more convenient to scribed in [179] have been found unnecessary. An
make the second test on alternating current at adjustable noninductive resistance in series with
zero power factor only. This can be done by ad¬ the instrument is set at such a value as to bring
justing the phase difference until the wattmeter the cross hair somewhere near the center of the
under test indicates zero with rated current and scale when the operating voltage is applied. Sets
voltage and then noting the reading, Pa', of the of alternately d-c and a-c readings are then ob¬
standard instrument, after which the latter is tained as described above for wattmeter testing.
calibrated using direct current. This procedure A measurement of the instrument sensitivity is
gives also made by applying two successive values of
direct voltage and noting the change in deflection
which results.
67/12
An alternative procedure for transferring from reactive power corresponding to the maintenance
d-c to a-c standards is the use of electro thermic of 1 amp in quadrature with a potential difference
instruments. Such apparatus of high precision of 1 v. The test of a single-phase varmeter is
has recently been installed [182, 207] and is regu¬ normally made by supplying the instrument with
larly used for testing voltmeters from 0.2 to 750 sinusoidal current and voltage from two gener¬
v and ammeters from 1 ma to 50 amp at frequen¬ ators coupled to a common driving motor. The
cies up to 20,000 c/s. Figure 12 shows some of power factor is adjusted to zero by mechanically
these instruments. In spite of the fundamental shifting the stator of one generator relative to the
simplicity of this method, a number of possible other until a wattmeter indicates zero The reac¬
sources of error exist such as Peltier temperature tive power in vars is then equal to the product of
shifts in the thermal converter, losses in the insu¬ the voltage and current, as measured by a volt¬
lating bead and resistance in common between meter and ammeter. A further check at unity
heater and couple circuits. With some individual power factor by connecting the varmeter at the
thermal converters, these effects may not be neg¬ terminals of a load of substantially nonreactive
ligible when an accuracy of 0.01 percent is sought. resistors, gives an indication of any phase defect
As finally worked out, however, the electrothermic in the varmeter circuit.
transfer for current and for voltage has been The calibration of polyphase varmeters and the
found to agree with the electrodynamic transfer use of varmeters with nonsinusoidal currents intro¬
to an accuracy of 0.01 percent at frequencies up to duces theoretical complications that are, in general,
500 c/s, and it is believed to be superior to the more serious than the purely instrumental errors.
latter and to be accurate to 0.02 percent for The Bureau has endeavored to bring some order
higher frequencies up to 20 kc/s. out of the earlier confused practices by cooperating
Still another unit, definable only in a-c circuits, in tbe formulation of standard definitions for the
is the var, the unit of reactive power. It is the numerous concepts involved [205].
68/13
These two groups, having a ratio close to 10:1
are connected to form two arms of the high-
precision bridge, the remaining two arms being the
adjustable ratios referred to above. The bridge
is then balanced by manipulating the four dials
and their setting, iV,, is noted. The paralleling
links are then shifted so that the 150-ohm coils of
the first set are put in parallel and form a 50-olnn
group, and those of the second set. are in series,
and, with the central 50-ohm coil, form a 500-ohm
group. These groups are turned end for end and
again connected in the bridge, and a second dial
setting, AT2, at balance is noted. The mean of
the two observed ratio settings, A\ and Ah, is then
adjusted by applying a small correction derived
from the observed differences between the 50-
ohm coil and the means of the groups of three in
parallel. A further correction for the resistance
of the paralleling links is applied, and after these
several corrections are taken into account, the
setting of the bridge dials for which the ratio is
exactly 10:1 is known within about 1' part in 3
million.
An alternative, but more laborious, procedure
sometimes used for obtaining an accurate 10:1
ratio, is to use a group of six 100-ohm standard
resistors. Each in turn is used as one arm of a
bridge; the other five in series constitute an
adjacent arm. The other two arms are the ad¬
justable 100-ohm arm of the precision bridge the dial A' to be tested, (2) an auxiliary box that
and a fixed 20-olim arm. The average of the six has a dial A of the same nominal value as A", and
successive settings on the adjustable arm then (3) a pair of mercury cups, C, into which may be
corresponds to an exact 100: 20 ratio to an order placed either (a) a short-circuiting link, L, or (b)
of accuracy given by the sum of the squares of a known standard resistor, S, of the same nominal
the departures of the individual 100-olim stand¬ value as one step of dial A' or A. The series
ards from their mean, provided proper corrections combination is connected to form one arm of a
are applied for the resistances of the contacts and Wheatstone bridge, the other arms of which are
connectors. This ratio is then used to compare formed by a direct-reading ratio set A-B and a
each of the two 50-ohm sections of a divided 100- fixed resistor, It. The dials are so set at each
ohm standard resistor with a 10-ohm standard, and stage in the process that the total resistance of the
from these two results the ratio of the divided arm is nominally 10 times the resistance of one
100-ohm to the 10-ohm is computed. The results dial step. Successive pairs of bridge balances are
obtained by this procedure are usually found to then made. In the first balance of each pair tin*
agree with those by the first-mentioned procedure standard resistor is in the circuit. In the second
to a few parts in 101. balance it is replaced by the link, and the A’ dial
By using the bridge, thus calibrated, it is easy is at a setting one step higher than during the
by successive steps to measure the values of wire- first balance. This constitutes in effect a sub¬
wound standard resistors of 10, 100, 1,000, or stitution of the dial step for the standard resistor,
10,000 ohms, and similarly to step down to 0.1-, and their difference is read from the small change
0.01-, 0.001-, and 0.0001-ohm standards. Re¬ required in the balancing arm of the bridge. The
sistances of intermediate value are occasionally auxiliary dial is then shifted one step, and a
measured by using analogous series or parallel second pair of similar balances gives a calibration
combinations of nominally equal resistors to for the next step of A’. Incidentally, the data
evaluate other bridge ratios that involve small thus obtained suffice for the calibration of the aux¬
integers. iliary dial also.
In contrast to the foregoing methods for testing When making precise measurements with re¬
individual standard resistors while immersed in a sistors of 1 ohm or less, it is essential to use re¬
thermostated oil bath, the procedure normally sistors of the four-terminal type to eliminate the
used for calibrating a decade dial of a bridge arm effects of contact resistance. The Kelvin bridge
or resistance box is to substitute a known standard or some equivalent circuit lias proved to be in
resistor successively for each of the steps of the general the most suitable for the comparison of
dial [80]. This is done, as shown in figure 14, by such resistors [7b]. Of the various procedures
connecting in series (1) a resistance box that has [77] that can be used to balance out the effects of
69/14
the resistance in the potential leads, that which source of error in this type of measurement. It is
involves 'applying a short-circuit across the ends used regularly in testing the unsaturated standard
of the main ratio arms is commonly used. cells that are sent to the Bureau periodically for
The Kelvin double bridge is usually employed check from a large number of scientific and indus¬
at the NBS in all measurements of low resistance, trial laboratories, as well as for the comparison
but remarkably good results can often be obtained of saturated cells.
with simpler apparatus, provided appropriate When an unsaturated standard cell is received
methods and precautions are used [78]. at the Bureau for test it is placed with others
At the other extreme the measurement of high in a large box that is thermally insulated, but not
resistance merges into the measurement of the thermostatically controlled, and its electromotive
resistivity of insulating materials [81]. High- force is measured daily for at least 10 days. If,
grade wire-wound resistance elements are available as is the usual case, the emf soon settles down
up to 1,000,000 ohms. Groups of such 1-megohm and remains within ±50 /zv of a steady value, a
resistors may be measured in a high-grade Wheat¬ certificate is issued giving this mean value to the
stone bridge while connected in parallel and can be nearest 10 yuv. However, if the fluctuations
used singly or in series to form a 10-megohm continue to exceed 0.01 percent or if the cell has
standard. Units of higher individual value con¬ an abnormally low value, below 1.0183 v, or if it
sisting of heterogenous compositions or of very shows an abnormally high internal resistance, no
thin film are likely to show changes in resistance formal “certificate” is issued, but instead the
with changes in temperature, voltage, and, unless data observed are sent to the customer in the
very carefully protected, in ambient humidity. form of a “report.” If properly used, unsaturated
Great precision in their measurement is therefore cadmium standard cells can constitute very accu¬
not required, and a simple direct deflection rate and reliable standards of voltage. They are,
method using a shunted galvanometer of high however, easily damaged by excessive current
current sensitivity (105 mm//za) is often satis¬ drain, and their electromotive force may be
factory. temporarily affected by sudden changes in tem¬
If the voltage is limited to a few volts, this pro¬ perature and also by any temperature difference
cedure fails at resistances exceeding 109 ohms. that may exist between the two electrodes [52].
For higher resistances the charge carried by the Modern- cell containers are made with copper
test resistor is accumulated on one electrode of an liners to minimize this latter effect.
adjustable air capacitor. The rate of ac¬ The extension of the d-c voltage scale.to lower
cumulation is measured by either changing the values is usually made by the use of a potentiom¬
capacitance or the potential applied to the other eter. For the normal precision testing of low-
electrode while using an electrometer as a null range voltmeters and millivoltmeters at the Bureau
indicator. When higher voltages can be used, a well-seasoned null potentiometer of the Feussner
correspondingly higher resistances can be measured type is used. The lowest dial has steps of 10 yuv.
[82, 86]. An accuracy of 0.2 percent can be ob¬ For moie rapid and less exacting calibrations a
tained up to 1013 ohms at 1.5 v. Brooks deflection potentiometer [96 to 99],
developed at the Bureau primarily for the rapid
5.2 The Scale of Direct Voltage calibration testing of electric indicating instru¬
ments, and for incandescent lamp testing, is used.
The starting point of the voltage scale is of It is, however, not suited for measurement of emf
course the primary group of saturated cadmium in circuits of high internal resistance. In this
standard cells [61], which maintain the volt apparatus the bulk of the. voltage to be Pleasured
during the long intervals between absolute measure¬ is balanced by the IR drop in the resistance of the
ments of current. This primary group [6] is main dial. Any unbalanced residue causes cur¬
picked from a much larger stock of saturated rent to flow in the calibrated galvanometer and
cadmium cells, batches of which are made up at produce a corresponding deflection. The value of
the Bureau from time to time, using very care¬ the total voltage is then indicated directly by the
fully purified materials [48 to 58]. one (or two) digits on the dial followed by three
The- members of the group are intercompared digits read from the galvanometer scale, the last
every few months by means of the standard cell digit corresponding to 0.1 scale division.
comparator [91]. This is a highly specialized A number of other types of potentiometer have
form of potentiometer that measures the difference been developed at the Bureau for special purposes.
in electromotive force between the cell under test The Wenner type [94, 95] is quite generally used
and a standard reference cell. A simple auto¬ for precise temperature measurement with thermo¬
matic computing device incorporated in the couples. On its lower range, one step on the
apparatus enables the observer to read the electro¬
lowest dial corresponds to 0.1 yuv. A special
motive force of the unknown cell directly. One
multirange potentiometer [90] has proved con¬
division on the instrument corresponds to 1 yuv.
Great pains were taken in the design of this com¬ venient for measuring small temperature differ¬
parator to minimize spurious thermoelectric ences. For the measurement of very small inten¬
effects, which usually constitute the most serious sities of radiation by a physical photometer,
70/15
another special potentiometer [93] reading to 0.001 current, and (2) the possibility that leakage cur¬
fxv is used. rents in the insulating supports may bypass part
Potentiometers, in effect, measure merely the or all of Ra but flow through Rb, thus invalidating
ratio of the unknown electromotive force to that the assumption implicit in eq (3) that the same
of the standard cell. Their calibration therefore current exists throughout the entire series circuit.
needs involve no direct measurement of resistance, To minimize such errors, the extension of the
but merely the determination of ratios of resist¬ voltage scale at the Bureau is made by using an
ances. Hence, for the testing of potentiometers especially constructed standard volt box rated for
[87] and also for other applications, it has been 750 v, but satisfactory for use as high as 1,500 v
found convenient to use a universal ratio set, [100], In the design of this apparatus (1) the
which is equivalent to a slide wire having a total resistance has been chosen so high (333/3 olims/v)
resistance of 2111.1 ohms, with a tap adjustable in that heating effects are negligible, and (2) guard
steps of 0.01 ohm. plates have been provided to intercept any leak¬
In the measurements of voltages of extremely age paths over the insulating supports and bush¬
low values, as well as in the highly precise measure¬ ings. The guard plates are connected to taps on
ment of resistance, one limiting factor may be the an auxiliary guard circuit so that the potential dif¬
sensitivity of the galvanometer. This was rec¬ ference across any insulator is never more than
ognized early, and considerable effort has been 75 v. and at most bushings is nominally zero.
devoted to theoretical studies and practical de¬ The measuring circuit is composed of groups of
signs of galvanometers for both direct [168, 169, five coils of nominally equal resistances alternating
92] and alternating [163, 164, 165, 166] current. with single coils, each having a resistance nomi¬
The galvanometers used at the Bureau for preci¬ nally equal to the total resistance between it and
sion resistance measurements are of a special the grounded end of the circuit. To determine
Bureau design, with taut suspensions and with the ratio of the divider with high accuracy, it is
the center of gravity slightly out of line with necessary merely to substitute its successive equal
the suspensions. With this construction, it is coils or groups of coils in one arm of a sensitive
possible to adjust the effective restoring torque by and stable, but uncalibrated, Wheatstone bridge.
properly tilting the instrument. With a 10-second The relative values of the coils are noted, and the
period, these have a sensitivity of 40 mm/fiv. value of the total resistance to any tap point is
In most of the other electrical measurements, computed from these relative values, and an as¬
commercial galvanometers of appropriate design sumed nominal value for the lowest step, on the
are used. The more complex amplifying combina¬ assumption that the resistances are truly additive.
tions, which can push the sensitivity to the limit In taking the ratio of any total resistance to the
set by Brownian motion (0.001 juv for a 10-ohm resistance up to a tap point, the assumed nominal
circuit and 2-second period), are seldom peeded. value cancels out leaving only the ratio. With
The Coblentz [167] moving-magnet galvanometer reasonable care, an accuracy of 0.005 percent can
lias been superseded by moving-coil galvanometers, be obtained.
ns they are not similarly affected by external In use the volt box under test is connected in
magnetic fields. parallel with the standard volt box, using on I lie
The extension of the scale of direct voltage to latter a tap giving the same nominal ratio. The
higher values while still using the potentiometer is difference in voltage, if any (exists, between the tap
made by using instrument voltage dividers (more points of the two dividers is measured with a small
commonly called “volt boxes”) [100]. Such a potentiometer of the Lindeck type. The ratio of
divider consists of two resistors, Ra and Rbl this difference to the total applied voltage, meas¬
which are connected in series, across the voltage ured with a voltmeter, gives the ratio error of the
to be measured. A potentiometer is connected to volt box under test relative to the standard volt,
measure the voltage, Eb, at the terminals of Rb, box. Corrections for lead resistance, etc., are de¬
and the total voltage, Ea+b, is then calculated by rived by auxiliary measurements with the. poten¬
the relation tiometer. 'The equipment used for such tests is
rp T,' (Ra+Rh) /.)\
shown in figure 15.
T^a+b — n ' O') In the measurement of the higher direct voltages
txb
used in X-ray generators a new source of error
In principle the ratio of resistances might be com¬ enters the picture. 'Phis is (lie corona discharge
puted from measurements of Ra and of Rb sep¬ from points and edges that are at a considerable
arately, such measurements being based on the potential above their surroundings. This can
10:1 resistance ratio described in section 5.1. This result in a significant current leakage to ground
method is not desirable, however, because of the directly through the air and can also cause trouble
two principal sources of error that arise in the use from the chemical action of the ozone and oxides
of volt boxes: (1) the possibility that at the high of nitrogen formed and from the electrostatic pre¬
values of voltage applied and hence larger values cipitation of dust. To obviate these troubles, a
of current in Ra, it may be so heated that its value construction has been developed at the Bureau
becomes significantly different from that observed [102] in which the resistance, Rn, is subdivided into
in a resistance-bridge measurement at smaller units of 5X10® ohms. Each is surrounded with a
71/16
I
72/17
The, function has the usual parabolic form char¬ the ratio of which is deduced from the resistance
acteristic of manganin, with a maximum resistance of its parts. With such networks very low cur¬
at 32° C. At higher currents, even though the rents can be produced and used to calibrate a-c
temperature is held at the same value by vigorous amplifiers. The amplifiers in turn can be used to
cooling, the resistor shows an increase in resistance. measure unknown currents of low value. The
This effect amounts to 100 ppm at 3,000 amp, and principal limitations in such work arise from
if proportional to the square of the current, would superposed noise and instability in amplifier gain
be 0.1 percent at 10,000 amp. It seems probable rather than from any difficulty in correlating the
that the effect is the result of mechanical strain results with the ampere.
set up in the resistance alloy sheets by their In the early years of the Bureau the scale of
differential expansion relative to the cooler rigid alternating current was extended upward by us¬
copper terminal blocks. Pending the construction ing a group of astatic electrodynamic instruments
of an improved high-current resistor free from [175] having suspended moving coils and fixed
such effects, the upper end of the precise scale of coils of cable formed of insulated strands of fine
direct currents must be considered as tapering wire (litzendraht). The highest rated (5,000 amp)
off from an accuracy of 0.01 percent at 3,000 amp instrument was a “tubular” electrodynamometer
to 0.1 percent at 10,000 amp. It appears that [176] in which the fixed “coil” consisted of two
there would be no serious difficulty in constructing coaxial copper t ubes. With the coming into general
a high-current resistor that could extend the high- use of current transformers [227], the number of high-
accuracy limit well above 10,000 amp, if a need range, self-contained ammeters, wattmeters, and
for such measurements should arise. The capacity watthour-meters submitted for test has fallen to
of the storage batteries presently available at the almost zero. The old high-range standard instru¬
National Bureau of Standards permits a current ments have been discarded, and when occasion¬
to be held at this latter value for about % hour. ally a high-range instrument is tested, the meas¬
Figure 17 shows this high-current circuit. urements are made with the 5-amp standard trans¬
fer instrument and a carefully calibrated current
5.4 The Scale of Alternating Current transformer.
The ratio and phase angle of a current trans¬
The electro thermic transfer ammeters offer a former depend to a very appreciable extent on the
convenient means for measuring with high accu¬ conditions of burden and frequency at which it is
racy alternating currents down to 1 ma at fre¬ operated and, to a much less extent, on the ar¬
quencies up to 20 kc/s. Lower values are read¬ rangement of the primary leads. The extension
ily derived by using a resistive shunting network, of accurate measurements to large alternating cur-
572772 0-61 - 6
73/18
rents has therefore required the careful study of For primary currents exceeding 2,500 amp use
the performance of current transformers [211, is made of a general principle in instrument trans¬
225, 226] and the development of methods for former design, namely, if the primary winding of
determining their ratio and phase angle with high an instrument transformer consists of two or
accuracy [210, 213, 217]. It happens that it is more sections, all having the same number of
much easier to compare the relative performances turns, then at the same frequency, secondary
of two transformers of the same nominal ratio burden, and secondary current (or voltage) the
than it is to determine the ratio and phase angle ratio correction factor (i. e., the quotient of true
of a single transformer [215, 216]. Hence many ratio divided by nominal ratio) and the phase
public utility companies and manufacturers of angle will be the same, to a high degree of pre¬
electric machinery submit to the National Bureau cision, whether the primary sections are con¬
of Standards for test groups of transformers of nected in series or in parallel. In other words,
assorted current ratings which they expect to use the core and secondary winding cannot distin¬
as standards for the testing of other transformers guish between the additional magnetomotive
of the corresponding ratings which they regularly force caused by the same current linking the core
use for measurement purposes. in an additional series turn and that caused by
The circuit used in testing current transformers additional current linking the core in a parallel
[224] up to primary ratings of 2,500 amp is shown turn. Deviations from this principle will arise,
for a current transformer, only in proportion to
the product of two factors each of which is
unlikely to be large: (a) the relative difference in
the distribution of current among the several
primary sections when they are in parallel, and
(b) the relative difference in the coupling (mutual
inductance) between each of the several primary
sections and the secondary winding. If their
relative differences do not exceed 1 percent each,
the principle will be valid to 0.01 percent.
To apply this principle, a special high-range
standard current transformer having eight pri¬
mary ranges has been constructed. This has 24
sections of primary winding, and with these in
series, can be calibrated by the usual resistance
method at ratings from 500 to 2,500 amp. It is
Figure 18. Circuit for testing a current transformer by the
resistance method. then used with its primary sections in parallel
as a standard up to 12,000 amp. A careful
in figure 18. Here Rr is a resistor of high current- study [224] of its performance has shown that
carrying capacity and of low and known induc¬ the equality of the coupling of the various sec¬
tance; R2 is a four-terminal resistor adjustable, tions of the secondary and the equality of the
from 0.01 to 0.1 ohm of 5-amp rating and known division of current among the primary sections
low inductance; M is an adjustable mutual induc¬ are both satisfactorily close. When the trans¬
tor continuously adjustable to 50 yh. that serves former is measured with only 500 secondary turns
to measure the phase-angle between the primary (i. e., 2,500 amp turns at full current) and with
and secondary currents. When the detector, D, each of the eight possible connections of its
shows a balance the ratio of the transformer is primary turns, its ratio correction factor at any
equal to R2/Ri, and its phase angle can be com¬ secondary current was constant to 0.02 percent
puted from the setting of M. A limiting factor in and its phase angle to 1 minute, and when operat¬
this method has been the determination of the ing at its normal 12,000 amp turns, the perform¬
effective four-terminal self-inductances of the re¬ ance of the transformer is presumably even better.
sistors used as Rx [128, 129]. The oil-cooled resis¬ Even if the flux distribution is greatly distorted
tors rated at 2,500 amp and at 1,000 amp are of by omitting one of the primary turns, the ratio
tubular construction, and their inductance can be correction factor is affected by less than 0.01
computed from their dimensions. The air-cooled percent.
resistors for currents of 500 amp and less are It seems probable that serious difficulty will be
formed of flat strips and have their potential encountered in an attempt to construct non-
leads so placed that their mutual coupling to the. inductive primary resistors for sustained currents
strip largely compensates for the self-inductance exceeding 2,500 amp because the lack of chemical
of the latter. The small residual inductances are uniformity in the resistance alloy will affect the
measured by a substitution process that compares distribution of current and thus the inductance.
them with those of a group of special resistors However, there is every indication that the
constructed so as to attain accurately computable measurement range at power frequencies can be
inductances, although at the sacrifice of current extended well above 12,000 amp by the use of
capacity and stability of resistance. multirange standard transformers.
74/19
Figure 19. Equipment for testing high-range current transformers.
Current is supplied by the step-down transformer, K, and measured either by the standard multirange transformer S or by using tubular standard
resistors in the oil tank M. The transformer under test is at 0 in the center of the “cage” at the right. The heavy copper tubes comprise the outer
conductors of the coaxial secondary leads.
Figure 19 shows the circuit used in measure¬ the secondary currents of the two transformers
ments from 500 to 12,000 amp. The transformer being compared. Therefore, a high accuracy can
under test is placed at the right in the “cage” be obtained in their relative ratios, although only
formed by the four 10- by %-inch copper bars, moderate accuracy is required in the adjustment
which, in parallel, constitute the “return” lead of the circuit constants. Nevertheless, as a
of the circuit. The central “outgoing” lead either check on possible incorrect adjustment at the
passes through the window of the transformer or factory or on major drifts in calibration, such
is bolted to the transformer primary terminals if transformer test sets are frequently submitted to
it is of the bar type. This circuit geometry [187] the National Bureau of Standards for test. The
offers two advantages. The first is that the test consists in the measurement of the resistance
components of magnetic field at the transformer of the slide wire at 13 points on the ratio scale
in the center, produced by the action of the four and of the mutual inductance of the inductor at
return bars cancel one another so that the result¬ 15 points on the phase-angle scale. The test set
ant effect is closely the same as if the return is also experimentally checked in actual operation
circuit were at infinity. This situation constitutes at midscale and one other point. The true values
a logical basis with which the effects of other of transformer ratio and phase angle are then
configurations can be compared [225, 226]. The computed and tabulated against the correspond¬
second advantage is that the resultant magnetic ing settings used in the test.
field of the entire circuit drops off very rapidly
with distance from the axis (roughly as r‘5) and 5.5. The Scale of Alternating Voltage
is negligible in those parts of the laboratory
where the measuring circuits are located. Between 1 v and 300 v, two alternative bases
Most of the transformers submitted for test are are available for the measurement of alternating
intended to be used as standards with which to voltages. The first, applicable at power fre¬
compare the performance of other current' trans¬ quencies, is to use the electrodynamic transfer
formers. Such comparisons are usually made by voltmeter, described in section 4, to transfer from
means of the Silsbee current transformer testing direct to alternating voltage at the 100- or 200-v
set, the circuit of which was developed at the level, and then to use a three-winding step-down
Bureau in 1918 [216], In the use of this set the transformer. The ratios of the voltages of the
measuring elements are required to measure various sections of the two secondary windings to
merely the difference in magnitude and phase of each other are checked by comparison with a
75/20
special resistive voltage divider. Normally, one Phase Shifting Trf.
winding supplies the standard transfer voltmeter,
already calibrated with direct voltage, and the
other tapped secondary supplies the a-c voltmeter
under test. Such a transformer has a very con¬
stant ratio over a wide range of primary voltages.
The second procedure, applicable also at audio
as well as at power frequencies, is to use the
electrothermic transfer voltmeters directly to fill
in the scale from 1 to 750 v.
Below 1 v, known voltages can be obtained by
using resistive voltage dividers or attenuators
that have been checked at somewhat higher
voltages against the standard volt box or by
measuring the resistances of their component
parts. "The low voltages thus derived, down to
about 20 (j.v, can be used to check low-range
Figure 20. Circuit for testing voltage-transformer test sets.
electronic voltmeters and as standard signals for
Voltage AC is supplied to both the comparator and the “Standard” cir¬
testing amplifier gain. To measure an unknown cuit of the test set. Voltage BC is applied to the “X” circuit of the test .set.
voltage in this low range, the known low voltage By adjustment of the magnitude and phase of the auxiliary supply, voltage
.-ICmay be set with any desired phase and magnitude relative to voltage PC.
with properly adjusted phase, can be bucked
against the unknown. The difference, after the accuracy of their adjustment. This is done
amplification, is applied to a detector, and the by supplying both the set under test and a special
known low voltage is adjusted to balance the voltage transformer comparator from a pair of
unknown. 60-cycle voltages (a-c and d-c) which differ in
From about 300 up to 250,000 v almost all magnitude and phase by adjustable amounts
precise a-c voltage measurements make use of (see fig. 20). The set under test is set on a
step-down voltage (potential) transformers [227]. succession of 11 points on its ratio scale and 13
The Bureau has contributed to the establishment points on its phase-angle scale, and also as a
of this alternating-voltage scale by developing check at the four “corner” points corresponding to
methods and equipment for measuring the ratio combinations of the extreme settings of the two
and phase angle of voltage transformers, by scales. The true voltage relations are read on
suggesting methods for the comparison of the the comparator for each point, and the resulting
performance of a voltage transformer with a corrections are reported in tabular form.
standard transformer of the same nominal rating The comparator is also used to compare trans¬
[214, 215], and by its testing service for calibrat¬ formers submitted for test with one of the standard
ing such standard transformers and comparison transformers owned by the Bureau. These seven
test sets [229]. standard transformers have the ratings shown in
The use of standard voltage transformers as a table 1. Their ratios and phase angles are deter¬
basis for the checking of others of the same mined with great care at intervals of about a year
nominal ratio is even more desirable than the by the methods outlined in the following para¬
analogous use of standard current transformers graphs, and have been found to be very stable.
both (1) because very simple apparatus suffices By using them as intermediate working standards,
for the comparison [214,228] and (2) Because of the need for operations involving direct metallic
the greater safety when the operator is isolated connections between the observer and the high-
from the high-voltage circuit by the transformer voltage circuit is limited to the annual performance
insulation and need handle only circuits con¬ checks.
nected to the low-voltage secondary windings.
It is almost universal practice to use voltage Table 1. Ratings of standard voltage transformers
76/21
would have the same phase angle on all ranges,
and its ratio would be strictly proportional to the
number of turns in series on each range. Depar¬
tures from this ideal might result either: (1) from
inequalities in the magnetic coupling from the
several primary sections to the secondary or (2)
from the presence of capacitance currents flowing
between portions of the primary circuit in a dif¬
ferent manner when different connections are used.
The applicability of this principle has been demon¬
strated in a large number of multirange transform¬
ers in which both the higher and the lower ranges
could be tested by the use of the 30-kv shielded
resistor [193, p. 327]. Auxiliary measurements of
equality of sections and of capacitance between
sections indicate that the departures from the
Figure 21. Circuit for testing a voltage transformer by the
ideal are negligible in the two higher-range trans¬
use of a shielded resistor.
formers used in the extension of the Bureau’s scale
secondary voltage of the transformer is then bal¬ of alternating voltage. The first of these trans¬
anced by the potential drop in an adjustable por¬ formers is calibrated with its four sections in par¬
tion, Re, of the resistor circuit at its grounded end. allel up to 30 kv and list'd with them in series up
A mutual inductor, M, which has its primary in to 120 kv. The second is compared with the first
series with the resistor and its secondary in series up to 120 kv and used with its two sections in
with the detector, serves to balance out any phase series up to 240 kv. Figure 22 shows some of the
displacement between primary and secondary standard transformers and tin1 shielded resistor.
voltages. The principal source of error in such a The chain of measurement of voltage from the
method is the capacitance to ground of the various electromotive force of a standard cell to an alter¬
sections of the resistor. To minimize this error, nating voltage of 240,000 is a long one of many
the working circuit is divided into sections of links, and it was felt desirable to check the result
20,000 ohms each, and each section is enclosed in by some independent method. For ibis purpose
a metal shield. Each shield is connected to an
appropriate tap point on an auxiliary high-resist¬
ance guard circuit, Rs. This guard circuit, also
of about 500,000 ohms, is in parallel with the
working circuit, and its taps are adjusted to be
at the same potentials as the midpoints of the
sections of the working circuit they shield. The
capacitance currents from the shields to ground
therefore flow in the guard circuit only and pro¬
duce only a second-order error in the working
circuit. With such an arrangement the second-
order error increases roughly as the fourth power
of the primary voltage, with the result that the
method becomes rather impractical above 30 kv.4
Above 30 kv several procedures are possible.
One is the use of a capacitance voltage divider as
used by Sharp and Crawford [222] and in more
convenient form, by Bousman and Ten Broeck
[223]. Such a circuit is now under development
at the National Bureau of Standards as a further
check on the performance of the standard multi-
range voltage transformers that are currently in
use to carry the testing service to 250 kv.
These multirange voltage transformers consti¬
tute another application of the general principle
enunciated earlier, which forms the basis for ex¬
tending the range of alternating current. Each
Figure 22. Voltage-transformer testing bay in National
is made with its high-voltage winding in several
Bureau of Standards High-Voltage Laboratory.
sections that can be connected in parallel for cali¬
The 30-kv shielded resistor is in the center at the back; standard trans¬
bration and in series for use as a high-voltage formers for 25 kv are at the left and for 100 kv at the right (on insulating sup¬
standard transformer. Ideally, such a transformer ports). The observers on the mezzanine are setting the burden and bal¬
ancing the secondary voltage against the IR drop in the grounded section of
the resistor. The large 350-kv transformer in the pit in the left foreground
4 By using an autotransformer of high rating (75 kva) to supply the shields can be used as a supply. One bushing of the 125/250-k v standard transformer
Weller [2211 found it possible to push this method to 132 kv. shows at the right edge.
77/22
an absolute electrometer of the attracted-clisk type involved different transformer connections, elec¬
suitable for voltages up to 275 kv was designed trometer spacings and disk diameters, frequencies
and constructed [192], In this instrument a light (60 c/s and 25 c/s), wave forms and magnitudes
Duralumin disk (16 cm in diameter) hangs from (10,000 to 100,000 v) of the voltage measured, and
one arm of a sensitive balance and is centered with changes in the potentiometer circuit used to
small clearance in an opening in the center of a calibrate the voltmeter. The two methods were
large guard ring. A flat circular plate is supported found to agree with an average discrepancy with¬
below the disk and guard ring, parallel to them out regard to sign of only 0.01 percent. The
and at a distance that can be adjusted over a range correctness of the extension of the a-c voltage
from 2 cm for low voltages (10 kv) to a maximum scale by the use of voltage transformers is thus
of 110 cm for 275 kv. The rms value of the voltage confirmed by an entirely independent method.
applied between the disk and the lower plate can Because of the much greater convenience of the
be computed in electrostatic units from the meas¬ transformer method, it will be used almost ex¬
ured diameter of the disk, the spacing between it clusively in the future. It is planned, however, to
and the plate, and the force of attraction, which reassemble the absolute electrometer again and to
is weighed by the balance. Figure 23 shows the make a similar cross check between the two
general arrangement of the electrometer, and figure methods at the 275-kv level.
24 shows some details at the balance. The interrelations of the two methods in terms
A long series of experiments [193] was performed of the present absolute systems of units is shown
in 1936 in which an alternating voltage (usually of in figure 25. It will be seen that bbtli methods
60 c/s) was measured simultaneously with the use a mass subject to local gravity as the measure
electrometer and with a sensitive electrodynamic of force (at F) and that the electrometer and the
voltmeter fed by one of the calibrated standard current balance occupy corresponding positions,
voltage transformers. The conditions were varied each involving only the ratios of its significant
from one experiment to another by changes that dimensions.
In the electrostatic method, the conventional
constant, e„, usually designated as “the permit¬
tivity of space,” is arbitrarily chosen as 1 (in the
centimeter-gram-second-electrostatic system). On
the other hand, in the electromagnetic method, it
is the conventional constant, usually designated
as “the permeability of space,” that is chosen
arbitrarily either as 1 (in the cgs-emu system) or as
47T-10-7 (in the rationalized mksa system).
The absolute ohm is evaluated by a separate
process, from the units of length and time and the
chosen value of fj.„. Combining the absolute ohm
with the absolute ampere, evaluated by the current
balance, fixes the absolute volt, in terms of which
values are assigned to NBS standard cells. The
long chain of step-up and d-c to a-c transfer
procedures then yields the electromagnetic value
for a high alternating voltage shown at A. The
electrometer yields directly at B the value of the
same voltage in statvolts. According to electro¬
magnetic theory ev and are not independent, but
are connected (as indicated by the dot-dasli lines
in figure 25) by the relation
78/23
Figure 24. Balance and attracted disk of the Brooks absolute electrometer.
Disk is shown with draft cover removed; above it are the Poynting clamp and the stirrup for the measuring weights; the compensated
counterweight and the “chainomatic” chain hang from the other end of the beam; the optical path is enclosed in the black tubing and prism boxes.
accuracy, but the slight observed residual differ¬ the United States is that of the American Institute
ence in the voltage measurements might be re¬ of Electrical Engineers [202], which is based on the
garded as an independent determination of c as pioneer work of the large electric manufacturing
being 299.783. companies. The values in it differ somewhat
from those in the corresponding standard of the
5.6 Measurement of Crest and Surge International Electrotechnical Commission, which
Voltage is generally used in Europe. One of the projects
still ahead of the National Bureau of Standards is
a study of the spark-over voltage of standard
In the higher voltage range, engineers are usually
more concerned with crest values than with rms sphere gaps with a view to reconciling the differ¬
values of alternating voltages. This is because ences and perhaps improving the accuracy of meas¬
the breakdown of insulation is to a large extent urement. Preliminary experiments have indicated
the feasibility of determining the crest factor of an
dependent upon the crest value of the voltage
applied to it, and the testing of insulation is the alternating voltage by using a capacitance poten¬
principal application of voltages above the normal tial divider and a point-by-point measurement of
operating levels. The commonly used measure¬ wave form with a quadrant electrometer as a
ment technique in this field is to note the spark- detector. This method is slow but capable of an
over voltage between metal spheres of known accuracy better than 0.1 percent; it will form the
diameter and spacing in air under standard test basis for calibrating other techniques.
conditions. The standard currently recognized in An alternative method for measuring cres't volt-
79/24
Figure 27. Surge current generator.
This hank of 40 capacitor units mounted in five tiers (only two of which are
visible in the photograph) is shown delivering a surge having a crest value of
about 100.000 amp. The outer conductor of the tubular shunt used in meas¬
uring such currents is visible below the table. The 100-kv charging line and
the triggering circuit are visible at the top.
80/25
age is the corona voltmeter of Whitehead. This erator now in use at the National Bureau of Stand¬
was studied by the Bureau in 1928 and its possi¬ ards for studying methods of surge-voltage
bilities and limitations examined [194], measurement, and figure 27 shows the 50,000-joule
A. closely related field is the measurement of surge-current generator. This has a rating of 10
surge voltages and currents. In recent years there nf at 100,000 v and has delivered currents c.f
has been a decided trend toward the use of tran¬ 200,000 amp.
sient voltage surges ("artificial lightning”) for the The complications which arise from inductance
proof-testing of electric apparatus. The reasons and skin effect in measuring transient currents are
for this are twofold. First, the duration of such closely related to those encountered on the precise
a test and the distribution of electric stress within measurement of the phase angle of current trans¬
the apparatus during the test can be made to formers. By applying similar procedures, stand¬
approximate the extreme conditions to which the ard forms of shunt and of inductor have been
equipment will be exposed in actual service, and developed [130] for measuring surge currents and
the results are therefore more significant than
the rates of change of such currents, respectively.
those of a test at an abnormally high value of sus¬
The design of a voltage divider that will be accu¬
tained alternating voltage. Second, in spite of
its complicated construction, the cost of such test rate for rapidly changing voltage, especially within
equipment is usually less than that of an equiva¬ the first microsecond of the surge, constitutes a
lent conventional transformer. Figure 26 shows difficult problem, and the Bureau’s work on it has
the 2,000,000-volt 33,000-joule surge-voltage gen¬ just begun.
6. Dissemination of Units
The unit of time as represented by standard but also by larger organizations as a check on the
radio carrier and modulation frequencies can be steps in the measurement sequence as performed
disseminated conveniently and with extreme ac¬ in their own laboratories. As a result a stream of
curacy by broadcasting. National Bfireau of about 2,000 high-grade electric instruments and
Standards radio stations WWV and WWVH per¬ pieces of related apparatus flows through the
form this service [23 to 26], and anyone interested electrical laboratories of the Bureau annually.
can just tune in. The electrical units cannot be The Bureau is required by law [1, 2] to charge
transmitted so easily, and hence the values have appropriate fees for such testing service unless it
to be disseminated by the more laborious physical is for a branch ol the Federal or State Government.
transportation of electric instruments, meters, and Regular fee schedules [5] list these charges for the
other .measuring apparatus. Electric measuring more usual types of test, and arrangements can
apparatus to be standardized is sent to the Bureau often be made by correspondence for special
from manufacturers of electric instruments and unlisted tests at appropriate fees.
machinery, public utility companies, State public A considerable fraction of the apparatus sub¬
utility commissions, university laboratories, in¬ mitted to the Bureau for test is newly manufac¬
dustrial research laboratories, as well as from tured, and in some cases (e. g., standard resistors)
Federal agencies and private individuals. it is desirable to have it held in the laboratory for
Some organizations send in groups of saturated several months to enable successive measurements
standard cells and 1-ohm standard resistors and to insure that secular drifts in value, originating
carry on the rest of the measurement chain in in the process of manufacture, have steadied down
their own laboratories. A number of foreign to a negligible rate. Other standard apparatus,
governments that are currently in the process of however, is usually submitted at regularly sched¬
founding or expanding their national standardizing uled intervals to insure the sustained accuracy of
agencies come in this class. More often, un¬ measurements based on it. The desirable fre¬
saturated standard cells, sets of standard resistors, quency of sffeh periodic checks depends on a
capacitors, inductors, and instrument transformers great number of factors, including (1) the accuracy
covering a considerable range of values are sub¬ required, (2) the extent to which intercomparisons
mitted. Many power companies submit their are made among the standards in the local labora¬
standard transfer wattmeters for comparison with tory, (3) the ruggedness and stability of the
the NBS instrument to verify the relation of its particular type of apparatus, and (4) the care¬
a-r performance to that on d-c, although they are fulness and skill of those handling: the apparatus.
well equipped to calibrate it on direct current A standard cell is relatively delicate ami can be
themselves. Manufacturers of electric measuring ruined in a few seconds by an excessive drain of
devices submit for test their own "laboratory current, yet if a group of at least three cells is
tools,” by the use of which they adjust their available and if its members are intercompared
product, and thus disseminate the units to their frequently, a change of any one relative to the
ultimate customers, who therefore may not need rest is quickly evident, and the volt as fixed by the
any direct contact with the Bureau. Indicating mutually consistent remaining cells may be trusted
instruments are submitted not only by small to an accuracy of 0.01 percent for 6 months or a
laboratories that use them as primary standards, year. Most standard cells of the unsaturated
81/26
type show a gradual decrease in emf, which on the This d-c performance, however, can be ade¬
average is about 80 nv/yenr. quately checked by measurements made at the
By contrast, a standard resistor, intended for local laboratory with a potentiometer and standard
the measurement of currents of several thousand cell.
amperes, is a very rugged structure.and is not at Recognizing the desirability of having precise
all likely to be damaged by carelessness in the apparatus checked initially at the Bureau, several
laboratory. Barring corrosion from acids in its manufacturers of electric measuring apparatus
cooling oil, if this is allowed to become rancid, or make it a practice to submit for test groups of
very slow metallurgical changes, such a resistor usually 10 or 20 similar standards. After the test
may be trusted to an accuracy of 0.01 percent for these are returned to the maker’s stock and later
5 or even 10 years. Resistors made with fine wires sold “with a National Bureau of Standards Certif¬
are decidedly more prone to change as a result of icate” at an increase in price to cover the Bureau’s
corrosion and mechanical strains, and should be fees and the handling costs. By this process, the
rechecked every 2 or 3 years. This periodicity necessary seasoning time does not introduce a
also applies to resistance boxes, measurement delay between the sale of the standard and its
voltage dividers, bridges, and potentiometers. ultimate use.
In many cases an initial test at the Bureau The results of calibration tests of measuring in¬
may be very desirable, and further periodic checks struments and apparatus are issued in one of two
may be unnecessary. An example is the transfer forms: “Certificates” and “Reports.” A Certif¬
test of a high-grade wattmeter, which determines icate is usually issued if the results have rela¬
the difference, if any, between its a-c and its d-c tively permanent validity, and if the apparatus
performance. This difference depends upon such tested meets certain standards of precision and
factors as eddy currents, inductance, and capac¬ tolerance.
itance in its windings and is very unlikely to Reports are issued when unusually long and
change appreciably with time or use. The d-c complex measurement programs are involved,
calibration as a wattmeter, which depends also when the apparatus is such that it can be too
on its springs and the dimensions and relative readily thrown out of adjustment, or when it
positions of its coils, may show secular changes. shows symptoms of instability.
7. International Relations
By the Convention of the Meter, as amended in values of its electrical units [6]. Also, the results
1921 [6], the International Bureau of Weights and of any new and improved absolute measurements
Measures at Sevres, France, is authorized to can be disseminated with high precision to the
coordinate the work of the various national stand¬ rest of the world by these biennial intercomparisons.
ardizing laboratories in the electrical field. The Most countries derive their electrical units by
National Bureau of Standards cooperates ac¬ sending standards to the International Bureau for
tively with the International Bureau and sends certification. As a matter of practical convenience,
groups of standard cells and of standard resistors however, a number of national laboratories have
to Sevres every 2 years for comparison with the preferred to have their standards tested at the
groups of standards kept at the International National Bureau of Standards. This is particu¬
Bureau. Other nations do the same, and in this larly expedient when the standard in question is
way the relative values of the units as maintained manufactured in the United States and can readily
in the different countries are derived. Graphs of be standardized in Washington before shipment
these values are given in [6, p. 13, 16]. In ab¬ abroad.
stract theory, all electrical units are supposed to Occasionally, when a standard capacitor or in¬
be derived by an unbroken chain from those main¬ dicating instrument of extreme accuracy happens
tained at the International Bureau. In the final to be shipped, say, from England to the United
analysis, however, because of errors in comparison, States, it is found feasible to arrange to have it
changes during transportation, and drifts with tested at the National Physical Laboratory before
time, the unit actually used in practice in any in¬ shipment and again at the National Bureau of
dividual laboratory is bound to differ somewhat Standards on its arrival in this country [109].
from the ideal. The relative differences between The agreements found in such comparisons are
countries seldom exceed ±20 parts per million. usualty not as precise as those based on the ship¬
The international comparisons of high precision ment of standard resistors and cells, but give a
serve to give early warning to any laboratory in very valuable assurance that the methods used in
case its units should show a drift relative to those the two laboratories for deriving other units from
of the other laboratories. The appearance of the more fundamental ones lead to results that
such relative drifts in the German standards led are consistent to an accuracy well within the needs
in 1931 to cooperative experiments in Berlin in of commercial operations.
which representatives of the British National The dissemination of scientific units of measure¬
Physical Laboratory and the National Bureau of ment throughout the world offers an almost unique
Standards took part [12], As a result of these example of harmonious international relations
experiments the German laboratory changed the which have persisted for many years. The gain
82/27
to society from world-wide uniformity in the field national organization which deals with weights and
of measurement is very great, and the sacrifices measures, established in 1875, has filled such a
required to maintain uniformity, at least in the definite need and was devised with such farsighted
electrical fields, are small. The workers in the wisdom that it has survived two world wars and
national standardizing laboratories, confronted has grown in prestige, scope, and influence. The
as they are with similar problems, have learned a precise unification of electrical measurements is
mutual respect for their confreres. The inter¬ one of its outstanding accomplishments.
8. Bibliography
The following list gives references to a large Bureau of Standards Journal of Research; J.
number of technical papers, standards, and books Journal of Research of the National Bureau of
originating for the most part at the National Standards; C, Circular (of the National Bureau of
Bureau of Standards and dealing more or less Standards).
directly with the extension and development of
methods for electrical and magnetic measurements. 8.1 General
A number of papers on absolute measurements
have also been included because the references [1] 31 Stat., ch 872, p 1449 11901) Establishment of the
listed in Bureau Circular 475 included only a few National Bureau of Standards.
[2] Public Law 619—81st Congress, (ch 486—2d Session)
selected papers in this field. Many NBS papers
(S2201). Amendment of section 2 of the Act of
that give the results of applying these methods of March 3, 1901 (31 Stat. 1449).
measurement in various fields, such as to the [3] Public Law 617—81st Congress (ch 484—2d session)
properties of insulating and of magnetic materials An act to redefine the units and establish the
standards of electrical and photometric measure¬
or to the performance of engineering devices,
ments.
have not been included. A few of the references {4] Announcement of changes in electrical and photo¬
are included for their historic value only, but in metric units, C459 (1947) 50.
those cases where the content of an older paper [5] -Testing by the National Bureau of Standards (policy,
general information, fee schedules) C483, 250.
has been included as part of a later publication,
[6] F. B. Silsbee, Establishment and maintenance of the
the older reference has been omitted. electrical units, C475 (1949) 250.
The following letter-symbol designations are [7] Frank A. Wolff, The so-called international electrical
used in combination with the Bureau publication units, B 1, 39 (1904-05) S3.
numbers, and these letters should be included [8] Frank A. Wolff, The principles involved in the selec¬
tion and definition of the fundamental electrical
with the numbers in all references to Bureau units to be proposed for international adoption.
publications: S, Scientific Papers; T, Technologic B 5, 243 (1908-09) S102.
Papers; RP, Research Papers; C, Circulars; [9] E. C. Crittenden, Present status of the international
H, Handbooks; ,M, Miscellaneous Publications; electrical units, Trans. AIEE 40, 990 (1927).
[10] G. W. Vinal, Units of electrical measurement, Trans.
LC, Letter Circulars (mimeographed). Am. Electrochem. Soc. 55, 43 (1929).
When a price is stated, the publication may be [11] H. B. Brooks, The unit of electrical resistance, P'ast
purchased from the Superintendent of Documents, historv and impending change, Trans. AIEE 50,
Government Printing Office, Washington 25, 1318 (1931).
[12] G. W. Vinal, International comparison of electrical
D. C. Many of the older papers are now out of standards, BJ 8, 729 (1932) RP 448.
print, but can be consulted in large public or [13] E. C. Crittenden, The Work of the Bureau of Stand¬
university libraries which maintain sets of Bureau ards in electricity and radio, Sci. Monthly 37, 405
publications. (1933).
[14] E. C. Crittenden, Revision of Electrical Units, Elec.
A complete catalog of Bureau publications Engin. (April 1940).
covering the period from 1901 to June 30, 1947, [15] J. H. Dellinger, International system of electric and
with author and subject indexes, NBS Circular magnetic units, B 13, 599 (1916-17) S292.
C460, is available from the Superintendent of [16] Electric Units and Standards, C60, 2d ed. (1920).
[17] H. L. Curtis, Electrical measurements, McGraw-
Documents for $1. A mimeographed supplement Hill Book Co., New York (1937).
to this is included. Copies of articles published [18] F. B. Silsbee, Standards for electrical measurement,
in other technical journals are not obtainable Elec. Engin. 70, 202 (March 1951).
from the Government. The standards mentioned [19] Radio instruments and measurements, C74, 2d ed.
(1918). $1.50.
may be purchased from the American Standards [20] P. G. Agnew, W. H. Stannard, J. L. Fearing, A
Association, 70 East 45th Street, New York. 17, system of remote control for an electric testing
N. Y., or from the American Institute of Electrical laboratory, B 13, 581 (1916-17). S291.
Engineers, 30 West 39th Street, New York 18, [21] A. B. Lewis, Calculations of electrical surge-generator
circuits, J 17, 585 (1936) RP929.
N. Y. [22] H. L. Curtis, Shielding and guarding electrical
In the following references, to save space, the apparatus, Trans. AIEE 48, 1263 (1929).
code letters listed below have been used preceding [23] G. H. Lester, Experimental standard-frequency trans¬
the volume number as abbreviations to indicate mitting station, WWVH, Communications 29, 20
(19491.
the corresponding series of Bureau publications:
[24] John M. Shaull, Adjustment of high-precision fre¬
B, Bulletin of the Bureau of Standards; S, Scien¬ quency and time standards, Proc. IRE 28, 6-15
tific Papers of the Bureau of Standards; BJ, (1950).
83/28
[25] Standard frequencies and time signals, WWV and [49] F. A. Wolff and C. E. Waters, Clark and Weston
WWVH, NBS Letter Circular LC974. standard cells, B 4, 1 (1907-08) S70.
[26] J. M. Shaull and C. M. Kortman, Comparing outputs [50] F. A. Wolff and C. E. Waters, The electrode equi¬
from precision time standards, Electronics 24, 102 librium of the standard cell, B4, 81 (1907-08) S71.
(1951). [51] F. A. Wolff, The temperature formula of the Weston
[27] E. B. Rosa and G. W. Vinal, Summary of experi¬ standard cell, B5, 309 (1908-09) S104.
ments on the silver voltameter at the Bureau of [52] J. H. Park, Effect of service temperature conditions
Standards and proposed specifications B 13, 479 on the electromotive foree of unsaturated portable
(1916-17) S285. (This includes a summary of the standard cells, BJ 10, 89 (1933) RP518.
results of 10 earlier papers on the voltameter.) [53] E. F. Mueller and H. F. Stimson, A temperature-
control box for saturated standard cells, J 13, 699
(1934) RP739.
8.2 Absolute Measurements 154] G. W. Vinal and M. Langhorne Howard, Effect of
glass containers on the electromotive force of
[28] E. B. Rosa and N. E. Dorsey, A new determination of Weston normal cells, BJ 11, 255 (1933) RP588.
the ratio of the electromagnetic to the electrostatic [55] D. N. Craig, G. W. Vinal and Francis E. Vinal,
unit of electricity, B 3, 433 (1907) S65. Solubility of mercurous sulphate in sulphuric-acid
[29] E. B. Rosa and N. E. Dorsey, A comparison of the solutions, J 17, 709 (1936) RP939.
various methods of determining the ratio of the [56] G. W. Vinal and L. H. Brickwedde, Metastability of
electromagnetic to the electrostatic unit of elec¬ cadmium sulfate and its effect on electromotive
tricity, B 3, 605 (1907) S66. force of saturated standard cells, J 26, 455 (1941)
[30] H. L. Curtis, Note on the ratio of the electromagnetic RP1389, 50
to the electrostatic unit of electricity as compared [57] L. H. Brickwedde and G. W. Vinal, Electromotive
to the velocity of light, BJ 3, 63 (1929) RP83. force of saturated Weston standard cells containing
[31] Paul R. Heyl and Guy S. Cook, The value of gravity deuterium oxide, J 20, 599 (1938) RP1094, 50
at Washington, J 17, 805 (1936) RP946. [58] L. H. Brickwedde and G. W. Vinal, Relation of elec¬
[32] H. L. Drvden, A reexamination of the Potsdam ab¬ tromotive force to the concentration of deuterium
solute determination of gravity, J 29, 303 (1942) oxide in saturated standard cells, J 27, 479 (1941)
RP1502, 50
RP1435, 50
[33] H. L. Curtis, The determination of the electrical [59] G. W. Vinal, D. N. Craig, and Jb. H. Brickwedde,
units bv mechanical measurements, Jour. Wash. Standards of electromotive force, Trans. Electro-
Acad. Sci. 22, 193 (1932).
chem. Soc. 68, 139 (1935).
[34] H. L. Curtis, Principles involved in determining the [60] G. W. Vinal, Maintenance of the volt, Trans. Am.
absolute values of the electrical units, Am. Physics Electrochem. Soc. 54, 247 (1928).
Teacher 5, 7 (1937).' [61] G. W. Vinal, Primary batteries (John Wiley & Sons,.
[35] H. L. Curtis, A review of methods for the absolute ^ 1950, ch 6).
determination of the ohm, Jour. Wash. Acad. Sci. [62] G. W. Vinal, Standard cells and the change from inter¬
32, 40 (1942). national to absolute electrical units, Trans. Electro¬
[36] H. L. Curtis, Review of recent absolute determinations chem. Soc. 93, 95 (1948).
of the ohm and the ampere, J 33, 235 (1944) [63] G. W. Vinal, L. H. Brickwedde, and W. J. Hamer,
RP1606, 10*. New quartz containers for standard cells at the
[37] H. L. Curtis, Absolute electrical measurements, Sci. National Bureau of Standards, C. R. 15th Confer¬
Monthly 69, 9 (1949). ence, Inti. Union of Pure and Applied Chemistry,
[38] H. L. Curtis, Charles Moon, and C. M. Sparks, An p 92 (1949).
absolute determination of the ohm, J 16, 1 (1936)
RP857.
[39] II. L. Curtis, Charles Moon, and C. M. Sparks. A 8.4 Standard Resistors and Resistance
determination of the absolute ohm using an im¬
proved self-inductor, J 21, 375 (1938) RP1137, 150. Measurements
[40] E. B. Rosa, N. E. Dorsey, and J. M. Miller, A de¬
termination of the international ampere in absolute [64] E. B. Rosa and H. D. Babcock, On the variation of
measure, B 8, 269 (1912) S171. resistances with atmospheric humidity, B 4, 121
[41] H. L. Curtis and Roger W. Curtis, An absolute de¬ (1907-08) S73.
termination of the ampere, BJ 12, 665 (1934) [65] E. B. Rosa, A new form of standard resistance, B 5,
RP685. 413 (1908-09), SI07.
[42] H. L. Curtis, Roger W. Curtis, and C. L. Critchfield, [66] J. L. Thomas, A new design of precision resistance
An absolute determination of the ampere using im¬ standard, BJ 5, 295 (1930) RP201.
proved coils, J 22, 485 (1939) RP1200, 100 [67] J. L. Thomas, Stability of double-walled manganin
[43] R. W. Curtis, R. L. Driscoll, and C. L. Critchfield, resistors, J 36, 107 (1946) RPi692, 50
An absolute determination of the ampere using [68] J. L. Thomas, Some experiments with pure-metal
helical and spiral coils, J 28, 133 (1942) RP1449 100 resistance standards, BJ 12, 313 (1934) RP657.
[69] J. L. Thomas, Reproducibility of the ice point, BJ 12,
[44] H. Pellat, An absolute electrodynamometer, Jour, de
323 (1934) RP658.
Physique, 2d Series, 6, 172 (1887).
[70] J. L. Thomas, Gold-chromium resistance allovs, J 13,
[45] Charles Moon, Apparatus for comparison of length of 681 (1934) RP737.
gages, BJ 10, 249 (1933) RP528. [71] T. B. Godfrey, Further data on gold-chromium re¬
[46] Charles Moon, A precision cathetometer, J 14, 363 sistance wire, J 22, 565 (1939) RP1206, 50
(1935) RP774. [72] J. L. Thomas, Gold-cobalt resistance alloys, J 14,
[47] J. L. Thomas, Chester Peterson, I. L. Cooter, and 589 (1935] RP789.
Ralph Kot.ter, An absolute measurement of re¬ [73] J. L. Thomas, Electrical-resistance alloys of copper,
sistance bv the Wenner method, J 43, 291 (1949) manganese and aluminum, J 16, 149 (1936) RP863.
IIP2029, 300
[74] Frank Wenner and J. L. Thomas, A method of
adjusting the temperature coefficient and resistance
8.3 Standard Cells of low-valued resistance standards, BJ 12, 147
(1934) RP639.
[48] F. A. Wolff and C. E. Waters, Preliminary specifica¬ [75] E. F. Mueller and Frank Wenner, The Waidner-Wolff
tions for Clark and Weston standard cells, B 3, 623 and other adjustable electrical-resistance elements,
(1907) S67. J 15, 477 (1935) RP842.
84/29
[76] Frank Wenner, The four-terminal conductor and the [105] F. W. Grover, The simultaneous measurment of the
Thomson bridge, B 8, 559 (1912) S181. capacity and power factor of condensers, B 3, 371
[77] Frank Wenner and Ernest Weibel, Adjustments of the (1907) S64.
Thomson bridge in the measurement of very low [106] H. L. Curtis, Mica condensers as standards of
resistances, B 11, 65 (1915) S225. ,
capacity, B 6 431 (1909-10) S137.
[78] Frank Wenner and Alva Smith, Measurement of low [107] F. W. Grover, The capacity and phase difference of
resistance bv means of the wheatstone bridge, S 19, paraffined paper condensers as functions of tem¬
297 (1923-24) S481. perature and frequency, B 7, 495 (1911) Si66.
[79] Frank Wenner, Methods, apparatus, and procedures [108] H. L. Curtis and Charles Moon, Absolute measure¬
for the comparison of precision standard resistors, ment of capacitance by Maxwell’s method, S 22,
J 25, 229 (1940) RP1323, 15*. 487 (1927-28) S564.
[80] J. L. Thomas, Precision resistors and their measure¬ [109] H. L. Curtis, C. M. Sparks, L. Hartshorn, and N. F.
ment, C470 (1948) 20*. Astbury, Capacitance and power factor of a mica
[81] H. L. Curtis, Insulating properties of solid dielectrics, capacitor as measured at the National Bureau of
,
B 11 359 (1915) S234. Standards and the National Physical Laboratory,
[82] L. S. Taylor, Accurate measurement of small electric BJ 8, 507 (1932) RP431.
,
charges by a null method, BJ 6 807 (1931) RP306. [110] A. V. Astin, Measurement of relative and true power
[83] H. B. Brooks, Accuracy tests for meggers, Elec. factors of air capacitors, J 21, 425 (1938) RP1138,
World, 85, 973 (1925). 10*.
[84] Master test code for resistance measurements, AIEE [111] A. V. Astin, Nature of energy losses in air capacitors
Std. No. 550 (1947). at low frequencies, J 22, 673 (1939) RP1212.
[85] Copper wire tables, C31, 3d ed. (1914) 30*. [112] C. Moon and C. M. Sparks, Standards for low values
[86] P. J. Higgs, A method for measuring high insulation of direct capacitance, J 41, 497 (1948) RP1935,
resistance, Jour. Sci. Instr. 10, 169 (1933). 10*.
[113] J. H. Dellinger and J. L. Preston, Methods of meas¬
urement of properties of electric insulating ma¬
8.5 Potentiometers, Bridges, etc. terials, S 19, 39 (1923-24) S471.
[114] A. H. Scott and H. L. Curtis, Edge correction in the
[87] Frank Wenner and Ernest Weibel, The testing of determination of dielectric constant, J 22, 774
potentiometers, B 11, 1 (1915) S223. (1939) RP1217, 5*.
[88] C. W. Waidner, H. C. Dickinson, E. F. Mueller, and [115] Chester Snow, A standard of small capacitance,
D. R. Harper, 3d, A Wheatstone bridge for resist¬
,
ance thermometry, B 11 571 (1915) S241.
J 42, 287 (1949) RP1970, 15*.
[116] W. D. Voelker, An improved capacitance bridge for
[89] E. F. Mueller, Wheatstone bridges and some accessory precision measurements, Bell Lab. Record 20, 133
apparatus for resistance thermometry, B 13, 547 (1942).
(1916-17) S288. [117] L. Hartshorn and W. H. Ward, Measurement of the
[90] H. B. Brooks and A. W. Spinks, A multirange poten¬ permittivity and power factor of dielectrics at
tiometer and its application to the measurement of frequencies ’from 104 to 10s cycles per second,
small temperature differences, BJ 9, 781 (1932) J. Inst. Elec. Engrs. (London) 79, 597 (1936).
R P506.
[91] H. B. Brooks, The standard-cell comparator, a
specialized potentiometer, BJ 11, 211 (1933) RP586. 8.7 Inductance (Construction and
[92] Frank Wenner, A method of reducing the effect of
disturbances in the galvanometer branch of a poten¬
Measurement)
tiometer circuit, J 22, 425 (1939) RP1194, 5*. [118] E. B. Rosa and F. W. Grover, Measurement of
[93] R. P. Teele and Shuford Schul imann, A potentiometer inductance by Anderson’s method, using alter¬
for measuring voltages of 10 microvolts to an accu- nating currents and a vibration galvanometer,
^ racy of 0.01 microvolt, J 22, 431 (1939) RP1195, 10*. B 1, 291 (1904-05) S14.
[94] F. Wenner, Note on potentiometer design, Phvs. Rev. [119] E. B. Rosa and F. W. Grover, Use of serpentine in
31, 94 (1910). standards of inductance, B 1, 337 (1904-05) S15.
[95] Leo Behr, The Wenner potentiometer, Rev. Sci. [120] J. G. Coffin, Construction and calculations of abso¬
Instr., 3, 109 (1932). lute standards of inductance, B 2, 87 (1906) S29.
[96] H. B. Brooks, A new potentiometer for the measure¬ [121] H. B. Brooks and F. C. Weaver, A variable self and
ment of electromotive force and current, B 2, 225, mutual inductor, B 13, 569 (1916-17) S290.
(1906) S33. [122] H. B. Brooks and A. B. Lewis, Improved continu¬
[97] H. B. Brooks, A deflection potentiometer for volt¬ ously variable self and mutual inductor, J 19, 493
meter testing, B 4, 275 (1907-08) S79. (1937) RP1040, 10*.
[98] H. B. Brooks, Deflect ion potentiometers for current [123] H. B. Brooks, Design of standards of inductance,
and voltage measurements, B 8, 395, (1912) S172. and the proposed use of models in the design of
[99] H. B. Brooks, Outline of design of deflection potenti¬ air-core and iron-core reactors, BJ 7, 289 (1931)
ometers with notes on the design of moving-coil RP342.
galvanometers, B 8, 419 (1912) Si73. [124] H. L. Curtis and Leon W. Hartman, Dual bridge for
[100] F. B. Silsbee and F. J. Gross, Testing and perform¬ the measurement of self-inductance in terms of
ance of volt boxes, J 27, 269 (1941) RP1419, 10*. resistance and time, J 25, 1 (1940) RP1310.
[101] Melville F. Peters, George F. BlacKburn, and Paul T. [125] F. W. Grover and H. L. Curtis, The measurement
Hannen, Theory of voltage dividers and their use of the inductances of resistance coils, B 8, 455
with cathode rav oscillographs, BJ 9, 81 (1932) (1912) Si75.
RP460. [126] H. L. Curtis and F. W. Grover, Resistance coils for
[102] L. S. Taylor, Apparatus for the measurement of alternating current work, B 8, 495, (1912) S177.
high constant or rippled voltages, BJ 5, 609 (1930) [127] Frank Wenner, Ernest Weibel and F. B. Silsbee,
RP217. Methods of measuring the inductances of low-
resistance standards, B 12, 11 (1915-16) S246.
8.6 Capacitance [128] F. B. Silsbee, A study of inductance of four-terminal
resistance standards, B 13, 375 (1916-17) S281.
[103] E. B. Rosa and F. W. Grover, The absolute measure¬ [129] F. B. Silsbee, Notes on the design of four-terminal
ment of capacity, B 1, 153 (1904-05) S10. resistance standards for alternating currents,
[104] E. B. Rosa, Wattmeter methods of measuring power BJ 4, 73 (1930) RP133.
expended upon condensers and circuits of low power [130] J. H. Park, Shunts and inductors for surge-current
factor, B 1, 383 (1904-05) S18. measurements, J 39, 191 (1947) RP1823. 25p.
85/30
[131] B. Hague, Alternating current bridge methods (Sir [157] H. L. Curtis, An integration method of deriving the
Isaac Pitman & Sons, Ltd., 5th ed., 1946). alternating-current resistance and inductance of
conductors, S 16, 93 (1920) S374.
[158] C. N. Hickman, Alternating-current resistance and
8.8 Inductance (Computations) inductance of single-layer coils, S 19, 73 (1923-24)
S472.
[132] E. B. Rosa, The self-inductance of a coil of any [159] Chester Snow, Alternating current distribution in
length and any number of layers of wire, B 4, 369 cylindrical conductors, S 20, 277 (1924-26) S509.
(1907-08) S83. [160] Walter Bartky, Numerical calculation of generalized
[133] E. B. Rosa and Louis Cohen, The mutual inductance complete elliptic integrals, Rev. Modern Physics
of two circular coaxial coils of rectangular section, 10, 264 (1938).
B 2, 359 (1906) S42.
[134] Louis Cohen, The self-inductance of a solenoid of
any number of layers, B 4, 383 (1907-08) S84. 8.9 Galvanometers and Detectors
[135] E. B. Rosa and F. W. Grover, Formulas and tables
for the calculations of mutual and self-inductance,
[161] L. W. Austin, Detector for small alternating cur¬
B 8, 1 (1912) S169, 55<t.
rents and electrical waves, B 1, 435 (1904-04) S22.
[136] F. W. Grover, Additions to the formulas for the
[162] M. G. Lloyd, Function of a periodic variable given
calculations of mutual and self-inductance, B 14,
by the steady reading of an instrument; with a note on
537 (1918-19) S320.
[137] F. W. Grover, Calculation of the maximum force the use of the capillary electrometer with alternating
between two coaxial circular currents, B 12, 317 voltages, B 4, 525. (1907-08) S90.
(1915-16) S255. [163] F. Wenner, A theoretical and experimental study
[138] F. W. Grover, Tables for the calculation of the induct¬ of the vibration galvanometer, B 6, 347 (1909-10)
ance of circular coils of rectangular cross sections, Si 34.
S 18, 451 (1922-23) S455. [164] P. G. Agnew, A new form of vibration galvanometer,
[139] F. W. Grover, Formulas and tables for the calcula¬ S 16, 37 (1920) S370.
tion of the inductance of coils of polygonal form, [165] H. L. Curtis, A vibration electrometer, B 11, 535
S 18, 737 (1922-23) S468. (1915) S239.
[140] H. L. Curtis and C. M. Sparks, Formulas, tables, and [166] Ernest Weibel, A study of electromagnetic moving
graphs for computing the mutual inductance of coil galvanometers for use in alternating-current meas¬
two coaxial circles, S 19, 541 (1923-24) S492. urements, B 14, 23 (1918-19) S297.
[141] F. W. Grover, Tables for the calculation of the [167] W. W. Coblentz, Sensitivity and magnetic shielding
mutual inductance of circuits with circular sym¬ tests of a thomson galvanometer for use in radiometry,
metry about a common axis, S 20, 1 (1924-26) B 13, 423 (1916-17) S282.
S498. [168] F. Wenner, General design of critically damped
[142] Chester Snow', Formula for the inductance of a helix galvanometers, B 13, 211 (1916-17) S273.
made with wire of any section, S 21, 431 (1926-27) [169] H. B. Brooks, Sensitivity of a galvanometer as a
S537. function of its resistance, BJ 4, 297 (1930) RP150.
[143] F. W. Grover, Methods for derivation and expansion [170] L. S. Taylor, Accurate measurement of small elec¬
of formulas for the mutual inductance of coaxial tric charges by a null method, BJ 6, 807 (1931) RP306.
circles and for the inductance of single-layer [171] L. F. Curtiss, A resistance-coupled amplifier for
solenoids, BJ 1, 487 (1928) RP16. measuring ionization currents, BJ 10, 583 (1933)
[144] Chester Snow; Mutual inductance of any two circles, RP550.
BJ 1, 531 (1928) RP18. [172] W. W. Nicholas, An electrostatic voltmeter, BJ 8,
[145] Chester Snow, Mutual inductance and torque 111 (1932) RP404.
between two concentric solenoids, BJ 1, 685 (1928) [173] P. G. Agnew, A device for measuring the torque of
RP24. electrical instruments, B 7, 45 (1911) S145.
[146] F. W. Grover, A comparison of tne formulas for the [174] H. B. Brooks, Temperature compensation of milli-
calculation of inductance of coils and spirals voltmeters, J 17, 497 (1936) RP926.
wound with wire of large cross section, BJ 3, 163 [175] E. B. Rosa, The compensated two-circuit electro¬
(1929) RP90. dynamometer, B 3, 43 (1907) S48.
[147] Chester Snow, The mutual inductance of two parallel [176] P. G. Agnew, A tubular electrodynamometer for
circles, BJ 3, 255 (1929) RP94. heavy currents, B 8, 651 (1912) S184.
[148] Chester Snow, The effect of small variations in pitch [177] J. H. Park and A. B. Lewis, Standard electrodv-
upon the inductance of a standard solenoid, B 6, namic -wattmeter and ac-dc transfer instrument, J 25,
777 (1931) RP304. 545 (1940) RP1344, 10fh
[149] Chester Snow, A simplified precision formula for the [178] Paterson, Rayner, and Kinnes, The use of the elec¬
inductance of a helix, with corrections for the trostatic method for the measurement of power, J. Inst.
lead-in-wires, BJ 9, 419 (1932) RP479. Elec. Engrs. (London) 51, 294 (1913).
[150] Chester Snow, Mutual inductance and force between [179] F. K. Harris, A suppressed-zero electrodynamic
two coaxial helical wires, J 22, 239 (1939) RP1178, voltmeter, BJ 3, 445 (1929) RP105.
15fb [180] F. B. Silsbee, Composite-coil electrodynamic instru¬
[151] Chester Snow, Theory of a single-layer, bifilar ments, BJ 8, 217 (1932) RP411.
absolute standard of mutual inductance, J 24, [181] J. H. Dellinger, High-frequency ammeters, B 10,
597 (1940) RP1302, 10<h 91 (1914) S206.
[152] Chester Snow, Mutual inductance of two helices [182] F. L. Hermach, A precision electrothermic volt¬
whose axes are parallel, J 25, 619 (1940) RP1346, meter for measurements between 20 and 20,000 c/s,
10fh Trans. AIDE 67, 1224 (1948).
[153] F. W. Grover, Inductance calculations (D. Van
Nostrand Co., Inc., New York, N. Y., 1946).
[154] J. G. Coffin, Influence of frequency upon the self- 8.10 Instruments and Meters
inductance of coils, B 2, 275 (1906) S37.
[155] Louis Cohen, The influence of frequency on the [183] E. B. Rosa, M. G. Lloyd and C. E. Reid, Influence of
resistance and inductance of solenoidal coils, B 4, wave form on the rate of integrating induction watt¬
161 (1907-08) S76. meters, 1, 421 (1904—05) S21.
[156] John M. Miller, Effective resistance and inductance [184] T. T. Fitch and C. J. Huber, A comparative study
of iron and bimetallic wires, B 12, 207 (1915-16) of American direct-current watthour meters, B 10,
S252. 161 (1914) S207.
86/31
[185] " T. T. Fitch and C. J. Huber, A comparison of [212] P. G. Agnew and F. B. Silsbee, Accuracy of the fof-
American direct-current switchboard voltmeters and mulas for the ratio, regulation and phase angle or
ammeters, B 7, 407 (1911) S163. ,
transformers, B 10 279 (1914) S211.
[186] A. E. Peterson, Performance of portable electrical [213] P. G. Agnew and F. B. Silsbee, The testing of instru¬
instruments in magnetic fields, Trans. AIEE 67, 1228 ment transformers, Trans. AIEE 31, 1635 (1912).
(1948). [214] H. B. Brooks, Testing potential transformers, B 10,
[187] F. B. Silsbee, Precautions against stray magnetic 419 (1914) S217.
fields in measurements with large alternating currents, [215] P. G. Agnew, A watthour meter method of testing
Trans. AIEE 48, 1301 (1929). ,
instrument transformers, B 11 347 (1915) S233.
[188] P. G. .Agnew, An approximate experimental method [216] F. B. Silsbee, A method for testing current trans¬
for the analyses of EMF waves, B 6, 95 (1909-10) Si 19. formers, B 14, 317 (1918-19) S309.
[189] F. W. Grover, Analyses of alternating-current waves [217] F. B. Silsbee, Methods for listing current trans¬
by the method of Fourier, with special reference to formers, Trans. AIEE 43, 282 (1924).
methods of facilitating the computations, B 9, 567 [218] F. B. Silsbee, Lead resistance for current transform¬
(1913) S203. ers, Elec. World 81, 1082 (1923).
[190] F. K. Harris, A new cathode-ray oscillograph and [219] H. B. Brooks and F. C. Holtz, The two-stage current
its application to the study of power loss in dielectric transformer, Trans. AIEE 41, 382 (1922).
materials, BJ 12, 87 (1934) RP636. [220] F. B. Silsbee, A shielded resistor for voltage trans¬
[191] Chester Snow, Effect of clearance and displacement former testing, S 20, 489 (1924-26) S516.
of attracted disk and also of a certain arrangement [221] C. T. Weller, A 132-kv shielded potentiometer for
of conducting hoops, upon the constant of an elec¬ determining the accuracy of potential trans¬
trometer, BJ 1, 513 (1928) RP17. formers, Trans. AIEE 48, 790 (1929).
[192] H. B. Brooks, F. M. Defandorf, and F. B. Silsbee, [222] C. H. Sharp and W. W. Crawford, Some recent de¬
An absolute electrometer for the measurement of velopments in exact alternating-current measure¬
high -alternating voltages, J 20, 253 (1938) ments, Trans. AIEE 29, 1517 (1910).
RP1078, 15£. [223] H. W. Bousman and R. L. Ten Broeck, A capaci¬
[193] F. B. Silsbee and F. M. Defandorf, A transformer tance bridge for determining the ratio and phase
method for measuring high alternating voltages angle of potential transformers, Trans. AIEE 62,
and its comparison with an absolute electrometer, 541 (1943).
J 20, 317 (1938) RP1079, 10*. [224] F. B. Silsbee, Ray L. Smith, Nyna L. Forman and
[194] H. B. Brooks and F. M. Defandorf, An experimental J. H. Park, Equipment for testing current trans¬
study of the corona voltmeter, BJ 1, 589 (1928) formers, BJ 11, 93 (1933) RP580.
RP21. [225] J. H. Park, Accuracy of high-range current trans¬
[195] F. M. Defandorf, The measurement of high voltage, formers, J 14, 367 (1935) RP775.
Jour. Wash. Acad, Sci. 38, 33 (1948). [226] J. H. Park, Effect of wave form upon performance
[196] F. D. Weaver, An easily assembled console for rapid of current transformers, J 19, 517 (1937) RP1041.
testing of electrical indicating instruments, Instru¬ [227] B. Hague, Instrument transformers (Sir Isaac
ments 22, 396 (1949). Pitman & Sons, Ltd. 1936).
[197] F. D. Weaver, Notes on the care and use of electrical [228] Leeds & Northrup Co., Potential transformer test
instruments, Instruments 23, 1236 (1950). set, Cat, E-50-501 (2) 1942.
[198] Electrical measuring instruments, C20, 2ded. (1915). [229] F. B. Silsbee, New high-voltage laboratory at the
[199] H. B. Brooks, Accuracy of commercial electrical National Bureau of Standards, Elec. Engin. 59
measurements, Trans. AIEE 39, 495 (1920). (June 1940).
[200] H. B. Brooks, Standardization of electrical measur¬
ing instruments, Trans, AIEE 42, 894 (1923).
[201] Am. standard for electrical indicating instruments , 8.12 Magnetic Measurements
(ASA Std. C39.1-1951).
[202] Am. standards for measurement of test voltage in [230] Charles W. Burrows, The best method of demag¬
dielectric tests (ASA Std. C68.1-1942, AIEE netizing iron in magnetic testing, B 4, 205 (1907-
No. 4). OS) S78.
[203] Code for electricity meters, ASA Std. No. C12 (4th [231] M. G. Lloyd and J. V. S. Fisher, An apparatus for
ed. 1941). determining the wave form of magnetic flux, B 4,
[204] Electrical metermen’s 'handbook, (Edison Electric 467, (1907-08) S87.
Institute, 420 Lexington Ave., New Arork, 6th ed., [232] M. G. Lloyd, Dependence of magnetic hysteresis
1950). upon the wave form, B 5, 381 (1908-09) S106.
[205] H. L. Curtis and F. B. Silsbee, Definition of power [233] M. G. Lloyd. Errors in magnetic testing with ring
and related quantities, Trans. AIEE 54, 394 specimens, B 5, 435 (1908-09) S108.
(1935). [234] M. G. Lloyd and J. V. S. Fisher, The testing of
[206] F. B. Silsbee, Power factor of polyphase- svstems, transformer Steel, B 5, 453 (1908-09) S109.
Proc. AIEE 39, 59 (1920b [235] C. W. Burrows, Uniformity in magnetic testing and
[207] F. L. Hermach, Thermal converters as ac-dc transfer in the specification of magnetic properties, Proc.
standards for current and voltage measurements ASTM, 8, 658 (1908).
at audio frequencies, J 48, 121, (1952) RP2296. [236] Charles W. Burrows, The determination of the mag¬
netic induction in straight bars, B 6, 31 (1909-10)
Si 17.
8.11. Instrument Transformers [237] R. L. Sanford, Determination of the degree of uni¬
formity of bars for magnetic standards, B 14, 1
[208] E. B. Rosa and M. G. Lloyd, The determination of (1918-19) S295.
the ratio of transformation and of the phase rela¬ [238] W. L. Cheney, Magnetic testing of straight rods in
tions in transformers, B 6, 1 (1909-10) Si 16. intense fields, S 15, 625 (1919-20) S361.
[209] M. G. Lloyd and P. G. Agnew, The regulation of [239] W. L. Cheney, Measurement of hysteresis values
potential transformers and the magnetizing cur¬ from high magnetizing forces, S 16, 281 (1920)
rent, B 6, 273 (1909-10) S129. S383.
[210] P. G. Agnew and T. T. Fitch, The determination of [240] M. F. Fischer, Apparatus for the determination of
the constants of instrument transformers, B 6, the magnetic properties of short bars, S 18, 513
281, (1909-10) S130. (1922-23) S458.
[211] P. G. Agnew, A study of the current transformer [241] R. L. Sanford and Evert G. Bennett, An apparatus
with particular reference to iron loss, B 7, 423 for magnetic testing at high magnetizing forces,
(1911) S164. BJ 10, 567, (1933) RP548.
87/32
[242] R. L. Sanford, and E. G. Bennett, An apparatus for [249] R. L. Sanford, Standards for testing magnetic per-
magnetic testing at magnetizing forces up to 5,000 meameters, BJ 4, 177 (1930) 11P140.
oersteds, J 23, 415 (19391 RP1242. [250] R. L. Sanford, Performance of the Fahy simplex
[243] R. L. Sanford and P. H. Winter, A permeameter for permeameter, BJ 4, 703 (1930) RP174.
magnetic testing at magnetizing forces up to 300 [251] R. L. Sanford, A method for the standardization of
oersteds, BJ 45, 17, (1950) RP2109, 10^. permeameters at high magnetizing forces, BJ 6 ,
[244] R. L. Sanford, Magnetic testing, C456 (November 355 (1931) RP279.
1946) 10«b ' [252] R. L. Sanford and Evert G. Bennett, Determination
[245] R. L. Sanford, Permanent magnets, C448 (August of magnetic hysteresis with the Fahy simplex
1944) lOtf. permeameter, J 15, 517 (1935) RP845.
[246] Charles W. Burrows, An experimental study of the [253] F. Bloch, Nuclear induction, Phys. Rev. 70, 460
Koepsel permeameter, B II, 101 (1915) S228. (1946).
[247] Charles W. Burrows and R. L. Sanford, An experi¬ [254] E. M. Purcell, H. C. Torrev, and R. V. Pound,
mental studv of the Fahy permeameter, B 14, 267 Resonance absorption by nuclear magnetic
(1918-19) S306. moments in a solid, Phys. Rev. 69, 37 (1946).
[248] R. L. Sanford and James M. Barry, Determination [255] H. A. Thomas, R. L. Driscoll, and J. A. Hippie,
of the magnetic induction in sheet steel, S 21, 727 Measurement of the proton moment in absolute
(1926-27) S545, 10£. units, J 44, 5.69 (1950) RP2104.
88/33
Journal of Research of the National Bureau of Standards Vol. 60, No. 4, April 1958 Research Paper 2845
The value of an electric current has been determined in absolute measure by means of
an electrodynamometer, and simultaneously by standard cells and standard resistors as
currently maintained. The electrodynamometer used was of the Pellat type, and featured
a fused silica balance beam and single layer helical coils.
The relation of the NBS ampere to the absolute ampere, from this determination, may
be expressed as
1 NBS ampere= 1.000013 absolute amperes.
The uncertainty in this result from all known sources is estimated to be eight parts per
million.
572772 0-61-7
89/287
Figure 1. Side view of the clectrodijnamometer. Figure 2. Rotatable coil and balance mechanism.
90/288
coil has already been mentioned. Although the
diameter of the stationary coil was not needed with
the highest accuracy, considerable pains were taken
on accoimt of its potential use as a standard of self
inductance, and the values obtained will be recorded
here. The m.ethod used in measuring both coils was
similar to that described by Curtis, Moon, and
Sparks [6]. The experimental arrangement of appa¬
ratus used in the measurement of the rotatable coil
is shown in figure 3 and that used in the case of the
stationary coil is given in figure 4. The results are
given in tables 2 and 3 and in figures 5 and 6. The
correction given in table 2 to the diameter of the
Diameter of wire
Origin of sample
Sample
No.
Beginning End of
of winding winding
vim mm
1 0. 5613 0. 5620
2 .5619 . 5619
3 . 5617 . 5620
Average 0.5618
Diameter, over-all
Tempera¬
Position ture coeffi¬
Top half Bottom Bottom Bottom cient
half half half
Diameter of wire
AXIAL POSITION ALONG COIL,Cm
Origin of sample j
Sample Figure 6. Variation in the diameter of the stationary coil.
No.
Beginning End of
of winding winding
i 6. Length of the Balance Arm
mm mm
1 0. 6995 0. 6990
2 .6992 . 6989 The distance between the outer knife edges of the
3 .6992 . 6990
Average . 6993 . 6990 | beam, figure 7, was evaluated by comparison with a
calibrated interval on a line standard. The balance
Average over-all diameter at 25.0° C__ 280.3239 mm beam was mounted on a comparator intended origi¬
Wire diameter_ _ 0.0991 mm
Mean diameter of winding, 2d\. at 25° C_ 279.6248 mm nally for the purpose of comparing line standards.
Force on wire during winding_ 15 1b In this instrument the unknown and standard are
Initial strain in wire, Ki__ 16.8X10-4
placed side by side on separately adjustable mount¬
Correction lor current distribution,—V (see section 7)_ —2.1 ppm
ings. These mountings were attached to a carriage
Effective radius, ai = ai+A«i_ 139.8121 mm
Thickness of fused silica form_ 2.5 cm that could be moved to bring either the standard or
Magnetic susceptibility of form_ —0.8X10-6
Resistance of winding at 25° C_ 40.40 ohms unknown into the field of two rigidly supported
Change in winding resistance with temperature_ 0.15 ohm/° C microscopes. It was found possible to set the cross
hairs of the microscopes over the apexes of the knife
edges on the beam in the same manner as on a gradu¬
ation of a line standard. The difference between the
o POS 1 45°
spacing of the outer knife edges and a known interval
h□
o
o
O
on the line standard was obtained from the readings
u <i Q
o ) o ( ° ’ o <
of the filar micrometer eyepieces of the microscopes.
Through the cooperation of L. V. Judson and
B. L. Page of the Metrology Section, facilities w-ere
provided for making this measurement.
The ratio of the arm lengths was determined by
weighing known masses. As indicated in table 4 no
significant difference was found in this ratio over a
period of 3 years whether 20-gram or 3-gram masses
were used. The masses were transposed as far as
possible but since the arm lengths differed by 1.5
percent a number of small masses had to be left on
one pan. Dividing the unloaded length (fi + /2) by
(1 -\-l2/li), where U/U is the ratio of the arms when
symmetrically loaded, gives a first approximation
to the length, fi, of the arm chosen to receive the
balancing mass. As pointed out by R. D. Outkosky
a further correction must lie applied to allow for the
increase in the length of the arm wdien the pan and
balancing mass are applied. To measure this dis¬
tortion, the beam was mounted on a surface plate,
the four bosses under the central part of the beam
Figure 5. 1 an a lion in the diameter of tin rotatable coil. I (fig. 7) being secured to the plate with beeswax.
92/290
Table 4. Length of balance arm arises when the standard mass is added to the scale
pan and the current in the stationary coil is reversed,
Distance between end knife edges, (/1 + /2)
that is, there is then an additional increase in the
length of the balance arm. Both distortions affect
Temperature (/1+/2)
Date, 1951 Corr. to the position of all the load on the end knife edge
25° C including, though to a lesser extent, the distributed
Beam Standard
mass of the balance arm.
°C °C cm In order to discuss these effects, let the following
28 June 29. 0 29.0 50. 59226 notation be used:
29 June 27. 0 27. 1 . 59238
29 June 27.4 27. 2 . 59234
29 J une 28.0 28. 1 . 59236
2 July 26. 2 26.2 . 59248 P=Mass of scale pan,
m = standard balancing mass,
(i July 26. 0 26.0 . 59251
g= acceleration of gravity,
1 Average_ 50. 59239 lx — unloaded length of balance arm,
8lP = change in /1 on adding load P,
b.fPp=change in torque to allow for distortion of
Above values based upon calibrated interval 390 to
the distributed mass of beam 011 adding
896 mm on meter bar 153-R.
load P,
Ratio of balance arms, l2/Ii torque on rotatable coil with mass m re¬
moved and stationary coil current in
Condition of weighing
forward direction,
Date h/b
./T= torque on rotatable coil with mass m on
19-20 Aug. 1952 0.984546 20-gram masses scale pan and stationary coil current
15 Dec. 1953 . 984548 3-gram masses transposed reversed.
22 July 1955 . 984550 3-g masses transposed
23 July 1955 . 984547 3-g masses transposed
(It was on these bosses that the beam rested during where
the measurement of the distance between the end JT= P(j (11 blp) -j- 6 ?p
knife edges.) By means of a microscope equipped
with a filar micrometer eyepiece the endwise move¬ and
ment of the knife edge at E figure 7 was measured .^2— (P~\~m)g(lx-\- dlp+ri) -i-&-Sr+m.
when a load equal to the scale pan was applied to
that end of the beam. A further complication Subtracting the above expressions
93/291
The subscripts 1 and 2 refer to the stationary coil
,%-^mgh [n+ jf-(2) and the rotatable coil respectively, and
5^m=[MT+3/8Mt\8lm,
where Mt is the mass of the arm between B and the where
cross rod at E. With this value of b^~m eq. (2) 1- 3-5 • • • (2n—l)
A,
becomes 2- 4-6 • • • 2n
Numerical values obtained for the quantities in Pniv) is Legendre’s polynomial of degree n, and
the brackets of eq (2') are as follows:
P'M is (d/dn) P,(m).
ra=1.481 grams Mt=31.5 g
P= 18.3 grams 5/w,//;= 1.0X 10-6
Also
Mr=4.4 grams blP+Jlx = 13.8X10-6
When these numerical values are applied in eq (2') (1 Mi) (1' "M2
Sv Z) (-1)”+|
the corrected torque is 6 (P n= 1
^-^=^[1+37.1 X10"6].
Pl) P
{nP^A"\?§r(n+^)T$
In effect the unloaded length lx of the balance arm
is increased by 37.1 ppm as a result of the distortion The radius of the rotatable coil, a2, in eq (3) is the
under load. The correction for this is applied in effective radius rather than the mean radius, a2, of the
table 4. winding, determined by mechanical measurement.
The difference between the effective and mean radii
7. Geometrical Factor of the Electro¬ depends on the variation of current density over the
dynamometer cross sect ion of the wire. The “natural” distribution
of current density (inversely proportional to distance
The maximum torque, T, produced by unit cur¬ from the axis of the coil) would certainly apply if the
rents in concentric helical wires is given by Snow [3] resistivity were constant everywhere in the wire.
as follows: Some measurements by Wells [7] indicate that for a
wire in simple tension, the resistivity-strain coeffi¬
T ran cient can be represented empirically by
La* J 9. ir/2
Ay jAl
L2t 2 =0+7
—4v2 -j- azNiHi X1 + S'+aS'i+ (4)
Ua(x2)dx2 Vo I lo
L,l 2
+Pl P/2 v (x)_ where, <r0, is the resistivity and l0 is the length of the
2MiJ_£i/2L {x\+a\y'*LaW unstrained specimen. The constants 0 and 7 were
determined by Wells on oxygen-free copper wire to
3aiX! have the values 0=1.13 and 7=—2.5X104. To the
2^ 5/2
Ux(x i)J (3)
(zi+a?) extent that the strain at any point in the wire can be
94/292
determined from the initial tension and bending, the
resistivity-strain relation can be incorporated in an
expression for the current density as a function of the
distance from the axis of the coil. A section in the
wire cut by a plane through the axis of the coil is
shown in figure 8. The coordinates of a point,
P(x, y'), are shown as functions of r and 0 in the wire.
The current density, i(y'), is assumed to be inversely
proportional to (a2 ++) and the resistivity, a(y'),
and is independent of x. The initial strain, K2, is
known from the winding tension and is taken to be
independent of x and y'. The strain due to bending
around the form is assumed to be equal to y'/a2 and
independent of x. On the basis of these assumptions,
the current in a filament of area r dd dr is given by
Crdddr
i(y')rdddr- (5)
2ir(a2-\-y')a(y')
a (y')= °o (1+0AT2+
(6)
where
Figure 8. Section through one turn of rotatable coil cut by an
axial 'plane.
3+37i+2
l+pK2+yK/ makes unit current in the wire is
7> 37/iV
2 1+^+7K/
)+
On combining eqs (5) and (6) and neglecting terms of
higher order than the third in (y')d2), the current
density is given by:
«{(0-KS)'}-«(OT (8)
and its value, by the integral I 2 j i(y') rdddr that A«i=-^ I1 (1+2,8+67^?).
Jo jo o CL\
95/293
Numerical values of Aa2 and &ax are given in tables The magnetic charge distributed over each end of
2 and 3. the stationary coil with one ampere flowing was thus
A summary of the computations on the constant estimated to be 1.5X10-3 pole. The contribution
T is given in table 5. The following expression, of the field of these charges to the field at the center
obtained by differentiating the main term of eq (3), of the stationary coil is found to be positive and of
was used to take account of small changes in the the order of one part in 107 even if the charges are
linear dimensions of the coils: regarded as point charges on the axis of the coil.
The magnetic charges on the ends of the rotatable
<$T_0 da2 4a\ bax_ L\ bLx coil induced by its own magnetic field are estimated
T a2 (L?+4a?) ax (Z,?+4a?) Lx to be 0.56 pole when one ampere is flowing. These
charges contribute to the magnetic moment of the
The constant T can be set equal to Flx where F is coil in the amount of —5X10-3 dyne cm/gauss.
the force constant referring to the scale pan on bal¬ The magnetic moment of the rotatable coil in the
ance arm lx. From the results of tables 4 and 5 the absence of the form is 1,400 dyne cm/gauss. The
value of F is correction to the torque on account of the rotatable
coil form is thus estimated to be —3.0 ppm. An
F=T/lx = 70029.91 dynes/cgsu2. attempt was made to ascertain experimentally the
effect of the aluminum alloy used in the balance
The acceleration of gravity g at the scale pan, on the support by placing more of this material in the
basis of the Dryden reduction [8] and a gravity opposite end of the stationary coil. No significant
survey made at this Bureau in 1948 by the Geological change in the measured torque was observed when
Survey, is 980.081 cm/sec2. The required balancing approximately twice the original amount of alumi¬
mass per unit current squared is num alloy was added. It must be admitted that it
would be difficult by this kind of test to guarantee
Flg=7l.45318 grams/cgsu2. that the effect was not as large as 1 ppm.
On reversing the current in the stationary coil the 8. Measurement of the Torque
change in torque on the rotatable coil is doubled.
After applying a correction to the torque of —3 ppm The electrical connections of the balance are shown
on account of the magnetic susceptibility of the coil in figure 9. During the measurement of the torque,
forms and changing to the practical unit of current , the current was held constant at about 1.02 amp as
the calculated balancing mass per ampere squared is determined by the standard cell and standard re¬
sistor. These standards were maintained at a con¬
2F(1 —3.0X 10-6)/100<7= 1.429059 grams/amp2. stant temperature and compared from time to time
with this Bureau’s primary standards. When the
The correction applied above to the electrodyna¬ current in the stationary coil was reversed, the
mometer constant on account of the magnetic sus¬ change in torque was compensated by placing a
ceptibility of the coil forms was estimated in the mass on the balance arm. The mass (a platinum-
following manner. From the known susceptibility, iridium rod) was adjusted by trial and later evaluated
k, of each coil form, the area A of its cross section, by comparison with known standards. Small correc¬
and the mean axial component of field h existing in tions to the balancing mass were made from readings
the form, the quantity, q, of free magnetism existing on the calibrated scale of the balance. The sensi¬
on its ends was calculated by the expression q=kAH. tivity of the balance was 1.8 cm/mg and the corre¬
sponding half period was about 10 sec. The contri¬
Table 5. Computation of balance constant bution of the current in the leads to the torque was
[Value used for the Independent variables:
determined separately and subtracted from the total
torque. A series of runs, table 6, shows the fluctua¬
a, = 13.98121 cm. Li = 100.0087 cm. Wi = 1000.
tions in the experimental result after all known
02= 5.79192 cm. L2= 9.3333 cm. iV2= 140.]
corrections have been applied. The result of each
run is derived from ten resting points of the balance
Quantity Numerical value each of which is calculated from nine observed
turning points. Resting points were obtained at
0.963064 intervals of about 4 min with the current in the
.6274
1.785461X106
stationary coil alternately in the forward and re¬
Li 2 versed direction.
CiUi)_ 1.0
0.873111
-.0612
C2<J12)_ — .212
s.' . _ -63i0X10-6
9. Results
Si__ 1.0X10-8
cm mq 0 C mg ° C mq ° C mq mq
23 Apr. 0. 002 0. 001 .36.9 -0. 166 30.8 0. 002- 29. 0 0. 002 -0.161
24 Apr_ .012 .005 36. 9 -.166 30.8 .002 29. 0 .002 -. 156
25 Apr_ .011 .006 .36. 8 -. 164 30.7 .002 29.0 .002 -. 154
26 Apr_ .016 .009 36.7 -. 163 .30.6 . 002 29.0 .002 -. 150
27 Apr_ .013 .007 36. 9 -. 166 30.7 .002 29.0 .002 -. 155
28 Apr_ .002 .001 37.0 -. 167 .30.9 .002 29.0 .002 -. 162
28 Apr_ .002 .C01 37. 1 -. 168 31.0 .002 29.0 .002 -. 16.3
29 Apr_ .007 .004 36.7 -. 163 30.6 . 002 29.0 .002 -. 155
30 Apr.-... .010 . 005 36.9 -. 166 30.8 .002 29. 0 .002 -. 157
1 May. . .006 .003 37.0 -. 107 30.9 .002 29.0 .002 -. 160
2 May - .004 .002 37.3 -. 171 31.1 .002 29.0 .002 -. 165
10 May. . .002 .001 36. 6 -. 161 30. 5 .002 29.0 .002 -. 156
26 May_.. .002 .001 36.8 -. 164 .30. 7 .002 29.0 .002 -. 159
97/295
This undertaking extending over a number of 10. References
years has accumulated the contributions of many
other staff members. In addition to those already [1] R. W. Curtis, R. L. Driscoll, and C. L. Critchfield, J.
named, the author wishes to thank the following: Research NBS, 28, 133 (1942) RP1449.
[2] J. C. Maxwell, Treatise on Electricity and Magnetism,
R. D. Cutkosky for valuable support in verifying
Clarendon Press, Oxford, Vol. II, Third Ed., Section 583.
the results by careful experimental checks; L. P. [3] Chester Snow, Torque between concentric single layer
Slivka for assistance in the design and construction coils, NBS J. Research 22, 607 (1939) RP1208.
of the balance arrestment mechanism; Catherine Law ,
[4] M. H. Pellat, J. Phys. 6 175 (1887); Bull. Soc. Intern.
des Electriciens 2d Series 8, 573 (1908).
for the calibration of the standard cell; W. H. Wood
for the calibration of the standard resistor; Eleanor [5] H. L. Curtis, Charles Moon, and C. Matilda Sparks,
J. Research NBS 21, 375 (1938) RP1137.
Clinton for the calibration of the standard mass;
[6] H. L. Curtis, Charles Moon, and C. Matilda Sparks,
and A. G. Strang for the calibration of the length J. Research NBS 16, 1 (1936) RP875.
standards. Helpful suggestions were received from [7] T. E. Wells, Proc. Instr. Soc. of Am., Vol. II, (1956).
J. L. Thomas and F. B. Silsbee. [8] H. L. Dryden, J. Research NBS 29, 303 (1942) RP1502.
98/296
Journal of Research of the National Bureau of Standards Vol. 60, No. 4, April 1958 Research Paper 2846
Prior to the adjustment of the electrical units in 1948, the value of a current had been
determined in absolute units by means of a current balance and simultaneously measured
in NBS amperes by comparison with standard resistors and standard cells. This work
was reported in RP1449. Similar measurements made recently with an electrodynamom¬
eter indicate a possible change in the values of the standards. The present paper reports
a repetition of the work described in RP1449. The purpose of this remeasurement was
to determine whether or not the standards had changed. Only minor changes were made
in the equipment in order that factors which might have introduced small systematic errors
in the results would remain unchanged.
According to the work described in this paper, 1 NBS ampere = 1.000008 absolute
amperes. Recent work with the Pellat electrodynamometer gave the result 1 NBS ampere
= 1.000013 absolute amperes. The weighted mean of these two values is
The results given above for the current balance differ by 6 ppm from those obtained in
1942. This indicates, in view of the uncertainties of measurement, that any change in
the ampere as maintained by standard resistors and standard cells does not exceed a few
parts in a million.
1. Introduction
99/297
Figure 2. Side view of current balance showing reversing
switches and operating rod.
100/298
rigid. These can be seen in the photographs. Also, o ‘
_ CL
iam 4 6.19115 cm
>
m
z
1 O .> O <
o
a sliding joint was put into the switch rod. The ’ o o
Hi Pi
Average outside diameter of coil . _ _cm.. 46.26090 46.26111 24.51360 24. 51342
Diameter of wire... .. - . _ _ .cm.. 0.06996 0. 06996 0.05123 0. 05123
Mean diameter of coil_ _ . __ _cm.. 46. 19094 46. 19115 24.46237 24. 40219
Current distribution correction..- _ern. -0. 00003 -0. 00003 -0. 00003 -0. 60003
Weighting correction . . _ _ _cm.. +. 00002 +. 00002 .0 .0
Effective mean diameter_... _cm.. 46. J9093 46.19114 24 46234 24. 46216
Axial length of coil_ __ . .. _cm.. 27. 51654 27. 51642 2. 6050 2. 6649
Number of turns. 344 41
Mean pitch . _ _ _ ... __ .cm.. 0.0799899 0.0799896 0.065000 0.064998
f 11.8016 11.7806
Resistance of winding j[ - _ohms.. } 2.8220 2. 8169
[ 11.7974 11. 780S
Resistance-temperature fUpperl / 0.0443 0. 0436
relationship . _[Lower/ " _ohms/°C . | 0.01079 0. 01034
l .0443 .0442
Temperature coefficient of expansion. _ .--Ppm/°C-_ 3.6 3. 5 3.9 3. 7
Winding tension__ _. ... ks? 10 3
101/299
the old measurements. The temperature coefficients where (8=1.13 and 7=—2.5X104;
of resistance do not agree, but this is because they
were assumed, not measured, in the earlier work.
At that time the temperature coefficients were taken
from tables of copper-wire characteristics. Since the
wires are under considerable strain, it is not surpris¬
ing that the measured temperature coefficients differ
:<70
[ 1+X-+7
l0 m
at a point in wire 1 is related to the initial strain
from the values assumed in RP1449. In
A new measurement of the diameter of the wire in the wire, Ku and the position of the point, 37, by
on the movable coil was made, and the result agreed A//Z0= (3/187) +.K1, where K1=2.5X 10~3) (and K2=
with the previous measurement. In view of the 1.4X.10-3). We then have to first order in 37/77
excellent agreement it was felt unnecessary to re¬
measure the fixed coil wire diameter. (8+37^) “I
The current distribution corrections contained in a = <r0(l+/9^1+7X?)^l+^
(l + 8#l+7#?)J,
table 1 correct for the variation of current density
over the cross section of each wire; no corresponding and
corrections were made in the work reported in
RP1449 because the net effect based on Snow’s I=_1_f1 , yi (8+37K\) ~|
assumption of the “natural” distribution [6] was a «To(l+8#i+7#?)L n (1+81£i+72£?)J
small. Recently Wells [7] has measured the resist¬
ance-strain relation in copper wire, making it now leading to the result
possible to give further expression to the variation
of current density over the cross section of the wires. g n
Using Snow’s formula for the helix equivalent to a “,(ri',%,<r0(l+,3.K1+7i??)l.1~
helical wire, we have
2/i (l+8iU+7#?+8+37gQ
ri (1+8/U+7#?)
-+2.4,
We then have
where r^the effective coil radius, , _ l + 8+37i^+8#i+7#?
l + 8#i+7PT?
17=the mean coil radius,
and similarly for A2. For the fixed coil, we have
P! = the wire radius, Ai = —0.07, At7 = — 1.7X 10-5 cm; and for the mova¬
ble coil, A2= —0.13, Ar2= —1.5X10-5 cm. A/7
and A77 are doubled and applied in table 1 as diame¬
ter corrections.
An attempt has been made to determine higher
order corrections to the effective diameter based on
and 1/1(77) is the volume density of current in the Wells’ resistivity determinations, but for the coils
wire as a function of the distance 77 from the x axis. used here such corrections are negligible.
A relationship similar to the above holds for the
movable coil, whose coefficients will be denoted in 3.2. Pitch
what follows by the subscript “2.” Snow has shown
that the radius corrections do not depend upon the The pitch of IIx was measured as described in
second derivative of u{r). This means that a first RP1449, and found to be insignificantly different
from the earlier value. The pitch of P, was last
order expansion of u(r) will lead to a radius correc¬
tion which is correct to second order. measured in 1934, and was not remeasured for the
The current density at a given point in the fixed 1942 work. A new determination was felt to be in
order for the completeness of this determination,
coil is given by ul{r\) = Glar[={G/arl)[l — (7/1/77)] to
even though the force constant is not strongly de¬
first order in 2/1/77, where a is the resistivity of the
pendent, upon the pitch of the movable coil.
copper, 37=77—77,. and G is a constant. The anal¬
ysis to follow is carried out for the fixed coil only, For the measurement of the movable-coil pitch
a meter bar was set up vertically, parallel to the coil
wire 1, but it is to be understood that the equations
axis. A telescope was clamped to a vertical bar
are valid for the movable coil, wire 2, also.
The resistivity of the copper has been measured which was free to pivot in such a way as to swing the
in terms of the strain by Wells [7b He finds that telescope from the meter bar to horizontal gradua¬
tions ruled on the wires of the coil. The telescope
Act was equipped with a filar micrometer eyepiece. A
reading was made of the distance between a gradua¬
tion ruled on a wire and a graduation on the meter
bar. Then, the telescope was raised to measure the
position of another wire. The distance between the
102/300
two wires is given by the distance between the two tX
)'a(X,a)
meter bar graduations plus the difference between
the readings of the filar micrometer eyepiece.
4
<XW2+(ri+r2)2L1-p {i *-2k]
The accuracy of the measurement depended upon
how well the coil and meter bar remained fixed with
irX / a\ ._, /, a\ fc(2 — k2)
(cos |)sin_1 (ioosi)-!(i=F|vi-*w!
respect to each other, and upon the repeatability of 4V^T2
the pivot of the vertical bar. The measurement is
not as good as that used to measure the fixed-coil
pitch, which used two telescopes, both of which were k sin
in 1+^1- k2 COS2 ;
^Zj{x)r{x)
-v-_hFh
2~'' r,„= ”S/(x)
>X2+(r1 + r2)2 _
K-n
Y'>
jj
culated with the variation coefficients dF/d>\, dF/dr2,
etc.
103/301
Table 2.—Summary of computations on force due to unit currents
The following values of the independent variables were used in the computation:
Azimuthal correction term: [2<o„(.Xj, <*)+“ <,(X3, a)— Ua(X3, a)] = +0.1600. The following adjustments wee made for differences between the values of
dimensions used in the above computation and the dimensions given in table 1:
Axial correction term: [2alr(.XiJot)+<3x(X2,a)— wI(A,3,a)] =+0.0045.
Fixed coil diameter An = +0.1 micron; AF= — 0.0016 dynes.
7=33,709.163 dynes. Movable coil diameter Ar2= — 0.1 micron; AF= — 0.0030 dynes.
Fhh = 1348.3665 dynes. Total force adjustment = —0.0046 dynes.
Hence for Hi and Pi at the dimensions of the coil corresponding to 30° C,
The computation of the coefficients in the variation formula for these two helices
gave the equation: Fhh = 1348.3619 dynes.
5. Experimental Determination of the Force Readings were made of nine turning points of the
Between the Coils balance with the current in one direction, then the
current in the fixed coil was reversed and the measure¬
The experimental determination of the force ment of turning points repeated. A set of ten
between the coils was made as described in RPi449. measurements involving nine reversals of current
The wiring diagram is shown in figure 6 for more was averaged and entered in table 3 as one determi¬
convenient reference. nation.
The force between the fixed-coil leads and the
movable coil was measured by removing the fixed
coil from the circuit without changing the lead-wire
configuration. The movable-coil lead effect was
measured in a similar way. These forces must be
subtracted from the total force between the coils,
and appear in table 4.
It was found that the mechanical dimensions of
the coil supports were not as stable as had been
hoped. Even with the straps around the case as
described earlier, the vertical position of the movable
coil with respect to the fixed coil changed about 0.2
mm in one month. The change was ascribed to
dimensional changes in the wooden case due to a
change in humidity, and the observed balancing mass
was corrected for this change under the assumption
that the shift was proportional to time. The num¬
bers given in table 3 are corrected for the effect,
which was never more than 4 ppm in the current.
The temperatures listed in table 3 are the temper¬
atures of the wires, computed from their resistances
and measured temperature coefficients. Because of
temperature gradients in the coil forms, the mean
104/302
Table 3. Results of measuremenits of force have been referred to the same electrical standards
Observed difference in rest points equals average difference in scale reading of and to the same value of the acceleration of gravity.
rest points of the balance, S, corresponding to “on” and “off” positions of This makes it possible to interpret the results directly
weight, multiplied by the sensitivity of the balance (2.33 mg/cm). Mass
of weight (a platinum cylinder): 1.425 569 g. as an apparent change in the electrical standards.
The measured values of the currents are expressed in
Temperature ° C “NBS amperes,” which is taken in this paper to
Observed Difference
Date (Mav difference in rest mean the current with reference to the present NBS
1956) n, in rest points
points corrected
standards of resistance and electromotive force,
P,
to 30° C which went into effect in 1948 [1]. The value of the
Upper Lower
acceleration of gravity is based on the Dryden reduc¬
mg 771(1
tion [8] and a gravity survey made at the National
18_ 24. 79 29. 84 28. 60 -0.041 -0. 049 Bureau of Standards by the Geological Survey. To
21_ 29. 45 29. 45 28. 26 -.001 -.009
21. 29. 55 29. 56 28. 37 -.002 -.010 make the comparison complete, the new diameter
22_ 29. 54 29. 56 28. 37 +.004 -.004 weighting, current distributions, and temperature
22_ 29. 59 29. 64 28.42 -.029 -.037
gradient corrections are applied in this paper to
23_ 29. 88 29. 93 28.71 -.049 -.057
23_ 29. 92 29. 95 28. 73 +.014 +. 006 both the 1942 and 1956 work. It may be pointed
23_ 29. 96 30. 00 28. 78 -.063 -.071 out here that these last three corrrections tend to
23_ 29. 99 30. 05 28. 80 -.063 -.071
24_ 29. 95 30.05 28. 82 -.046 -.053 cancel, and do not change the 1942 result by more
24_ 29. 95 30. 05 28. 82 -.019 -.026 than 1 ppm.
24_ 29. 96 30. 06 28. 82 -.002 -. 009
24_ 29. 97 30. 07 28. 82 -.007 -. 014
6. Permeability of the Forms
Averaee 6 times sensitivity of the balance_ -0. 031 It was assumed in the earlier work on the current
balance that the permeability of the coil forms had a
Table 4. Calculation of final results negligible effect on the force constant. Inasmuch as
[Value of acceleration of gravity 980.081 cm/sec 2]
the susceptibility of each form was only — IX 10~6,
the correction would certainly be small; but an order
1942 1956
of magnitude calculation was felt desirable.
In the following computation, the permeabilities
Standard mass (a platinum cylinder) of the movable coil and of the fixed coil are treated
grams. 1. 427 655 1. 425 569 separately. Unit current (1 amp) is assumed flow¬
Buovance correction_grams. -0. 000 077 -0. 000 079
-.000 068 -. 000 044 ing in each of the coils. It is necessary with the
-.000 018 -.000 022
5 times sensivity of balance_grams. +. 000 222 -.000 031 method used to compute the magnetic fields of the
Correction for temperature gradients solenoids at various points. This can be done in all
in coil forms_grams. -.000 014 — .000 014
Net compensating mass=A/...grams. 1.427 700 1. 425 379 cases by means of formulas given in a paper by
Measured force (M times gravity) =
Fm_dynes.. 1399.262 1396. 987 Snow [9].
Calculated force for unit current at 30° The field of the movable coil serves to induce
C = Fc_ dynes- 1348. 394 1348.361 9
magnetic poles on the ends of the movable coil form,
Equivalentcurrent, /abs=-«/A5t
V Fc whose magnitude can be computed through the
absolute amperes. 1. 018 688 1. 017 871 4
Emf. of standard cell_volts. 1.018 702 1. 017 870 3 relation m=(HxV)/l, where H= 3.5 oersteds is the
Resistance of standard resistor, Ri
ohms. 1. 000 015 1. 000 007 6 mean field intensity in t he form, x=__l X10~6 is the
Resistance of standard resistor, R? susceptibility, U is the volume of the coil form, and
ohms. 1. Q00 019 1. 000 005 6
Equivalent current_XBS amperes 1. 018 686 1.017 863 0 / is its length, m is from this approximately
/ ab.
1. 000 002 1. 000 008 — 3.0X10-4 pole. The axial component of the fixed
/nbs
coil field intensity, Hx, at the end of the movable
coil form is computed to be Hx= 0.65 oersted. The
temperatures of tlie coils are slightly different from force on each end of the coil form is then F'—
the wire temperatures. A measure of this effect — 0.65X3.OX 10-4 dyne. Since the total force be¬
was made and applied to the work reported in tween the coils is F—1348/2 = 674 dynes without
RP1449. The 1942 temperature gradient measure¬ reversal of the current, the permeability of the
ments were corrected by the better resistance measure¬ movable coil has an effect of —[(2X2.OX 10-4)/674] =
ments made recently; and it was found that under — 0.6 ppm in the force, considering both ends of the
equilibrium conditions with one ampere through the form.
coils, the mean fixed coil form temperature was 1.1° Because of the complicated field distribution
C below the fixed coil wire temperature, and the mean inside the fixed coil form due to current in the fixed
movable coil form temperature was 0.1° C below coil, a rather elaborate calculation was made of the
the movable coil wire temperature. Application of fixed coil permeability effect. A rough estimate
the computed force-diameter variation coefficient indicated that the form, although diamagnetic,
from table 2 and the temperature coefficients of would cause an increase in the radial component of
expansion from table 1 leads to —0.014 mg as the field at the movable coil, which is not what one would
required correction. This will be found applied in at first expect.
table 4. The magnetic charge distributed over the surface
For the comparison of the present work with the of the form was calculated using the normal com¬
work of 1942 in table 4, both of these determinations ponent of field given by Snow’s formulas, and the
105/303
572772 0-61-8
form susceptibility. The distribution was broken Table 5. Sources of error causing an uncertainty, e, in
up into a series of rings of charge one centimeter ratio Iabs!Inbs as large as 1 ppm
wide extending around the form, and the total
charge per ring was determined. This charge was e(50%
error)
then assumed concentrated on a circle located at the
center of the ring. A formula for the potential of a ppm
circle of charge has been given by Smvthe [10], in Measurement of the force_ ... _ 2
Lead-corrections _ .. .__ _ . 2
terms of Legendre polynomials, but this did not Radius of fixed coil_ ... . ... _ 2
converge satisfactorily for our purposes. A solution Radius of movable coil_ ... . _ 3
Current distribution over wires... . _ . 1
was found in terms of elliptic integrals, which leads Calibration of length standards . _ . 1
Calibration of standard mass. . .... 1
to an easier numerical calculation. Calibration of electrical standards.. .. . 1
It can be shown that the potential of a circle of Adjustment of coils.. ..... .. _ . . .. 1
Temperature of coils.. . ... ... _.. 1
charge at a point a distance r from the axis of the Permeabilitv of forms. . . . _ _ .. <1
Acceleration of gravity .. ... . _ _ .. 3
circle and a distance d from the plane of the circle is
given by V=[2QK(k)]/(Tr&), ks=(4rR)/<%2, ,^2= 50-percent error in final result, -y'Ze-.. 6
107/305
UNITED STATES DEPARTMENT OF COMMERCE • Charles Sawyer, Secretary
NATIONAL BUREAU OF STANDARDS • E. U. Condon, Director
Precision Resistors
and
Their Measurement
by James L. Thomas
108/1
Preface
109/11
Contents
Page
Preface_ n
I. Introduction_ 1
1. Definition of resistance_ 1
2. Importance of resistance measurements_ 1
3. Types of resistors_ 2
II. Resistance Materials and Construction Methods_ 2
1. Resistance alloys_ 2
2. Spools, winding and adjustment_ 4
3. Sheet-metal resistors_ 5
4. Accelerated aging_ 6
5. Annealed resistors_ 6
6. Effects of humidity_ 7
7. Load coefficients_ 9
8. Stability of resistors with time_ 9
ill. Methods of Comparison of Resistors_ 10
1. Ammeter-voltmeter methods_ 10
2. Ohmmeters_ 11
3. Potentiometer methcrd_ 11
4. Differential-galvanometer method_ 12
5. Bridge methods_ 13
IV. Special Apparatus for Precision Measurements_ 19
1. Direct-reading ratio set_ 19
2. Universal ratio set_ 26
V. Calibration of Precision Bridges_ 28
1. Calibration of Wheatstone bridges_ 28
2. Calibration of thermometer bridges_ 29
VI. Resistivity of Solid Conductors_ 31
1. Resistivity, definition and units_,_ 31
2. Measurement of resistivity_ 31
VII. References_ 32
110/III
Precision Resistors and Their Measurement
By James L. Thomas
Abstract
I. Introduction
112/2
same time a small temperature coefficient of same. From the ingot, manganin is usually
resistance at ordinary room temperatures. On worked cold with an occasional softening by
the other hand, if the thermoelectric power is of heating to a “red heat.” If this annealing is done
no importance the resistance of an alloy of about in air there is a selective oxidation of the surface
20 percent of nickel and 10 percent of manganese that breaks down the alloy and leaves a coating
would be most constant with temperature in the with a relatively high conductivity. This surface
ordinary range of laboratory temperatures. The layer may be removed by “pickling” in an acid,
accepted composition of manganin, 84 percent of but again the action of the acid is somewhat
copper, 12 percent of manganese and 4 percent of selective, and the surface layer is left with a
nickel is reasonably close to the optimum for a slightly different composition from that of the
general purpose resistance alloy. Fortunately interior. These complications add to the diffi¬
the proportions need not be exact, as the melting culty of the control of the quality of manganin
losses make the composition somewhat difficult wire. In addition, the completed resistance coils
to control. are usually baked at about 150° C for 24 to 48
If the resistance of a sample of manganin is hours in order to stabilize the resistance. This
plotted against temperature, the curve will be found baking also affects the temperature coefficient of
to be of the general shape of that shown in figure 1. the wire by amounts depending upon the size
of the wire and the length of time of the baking.
All of these factors make it difficult to produce
resistors with the maximum at the best tempera¬
ture, usually between 20° and 25° C. However,
precision resistors of good quality usually can
be obtained with temperature coefficients in that
interval of not more than 10 ppm/0 C.
Temperature coefficients of resistance change
little with time, and if changes do occur in their
values there is very probably an accompanying
large change in resistance. Standards that are
stable in resistance need not have redetermina¬
The maximum, M, of this curve is in the neighbor¬ tions of their temperature coefficients. Although
hood of room temperature, being generally be¬ the data are meager and only the order of magni¬
tween 20° and 50° C, while the minimum, N, is tude is known, it appears that for manganin
at about 350° C. Somerville [3] found the differ¬ resistors a change in resistance of one part in a
ence in resistance between the maximum and the thousand will be required to change the slope
minimum to be about V/2 percent of the value at of the temperature-resistance curve at room
the maximum. temperature by one or two parts in a million
For an interval at- least 15° to 20° C on each side per degree centigrade.
of the maximum, the resistance-temperature Constantan.—-A series of alloys of nickel and
curve for manganin is symmetrical with respect copper containing 40 to 60 percent nickel, with
to a vertical axis. Its equation can be accurately a small amount of manganese to improve their
represented by mechanical properties, all have practically the
same electrical properties. These alloys are
Rt=R25[l + a(t-25) + (3(t-25)2], (4) sold as “constantan,” or under various trade
names, for use as thermocouple materials, and
where Rt is the resistance at t°C and R05 is the have thermoelectric powers against copper of
resistance at 25° C. The coefficient a is the slope about 40 mv/0 C. However, except for their
of the resistance-temperature curve at 25° C, and large thermoelectric powers the electrical prop¬
for manganin of good quality a has a value of erties of these alloys are remarkably similar to
10X10-6 or less. The value of <3, which deter¬ those of manganin.
mines the curvature at any point in the interval, The resistance-temperature curve for constantan
is usually between — 3X10-7 and — 6X10-7. is similar to that for manganin shown in figure 1.
This means that 10° C either side of the maximum Its maximum is at or near room temperature with
the resistance is less than at the maximum by the minimum around 500° C. The difference in
from 30 to 60 parts per million (ppm). resistance between the maximum and minimum is
The temperature at which the maximum of somewhat less than for manganin, and the curva¬
resistance occurs is a function of the thermal and ture in the neighborhood of the maximum is also
mechanical treatment of the manganin as well less. As a consequence, constantan changes some¬
as of its composition. Different size wires drawn what less in resistance over the ordinary range of
from the same melt will have maxima at different atmospheric temperatures than does manganin.
temperatures, that is, at any given temperature Its stability with time is about the same as that of
their coefficients of resistance will not be the manganin. At room temperature the resistivity
113/3
of constantan is 45 to 50 microhm-cm as it is for small. Resistors of this material have been
manganin. produced such that the total change in resistance
Resistance coils of constantan are sometimes in the interval 20° to 30° C did not exceed a few
used instead of manganin in values of 1,000 ohms parts in ten million. The thermoelectric power of
and above. They may also be used in smaller this gold-chromium alloy against copper is several
denominations in cases where no difficulty will times that of manganin, being 7 or 8 /jlv/° C at
arise from the large thermal emf’s, as for example 25° C. The stability of this alloy with time has
in alternating current circuits. not been thoroughly tested but preliminaiy results
Therlo.—When manganin was first developed a were promising.
small amount of nickel was added to the copper For many applications the extremely small
and manganese in an attempt to reduce the ther¬ temperature coefficient of gold-chromium alloy
moelectric power of the alloy against copper. makes its use desirable. However, the tempera¬
With the proportion of nickel now used the ther¬ ture coefficient must be adjusted for each coil by
moelectric power against copper at room tempera¬ baking, and the cost of this adjustment limits the
ture is almost the same, although of opposite sign, use of the material. Although the temperature
as when the nickel is omitted. However, the coefficient may be made small at room tempera¬
nickel improves the mechanical properties of the ture, the interval over which the coefficient is
alloy and probably reduces the surface action small is not more than 20° or 30° C. The tem¬
during forging and annealing. perature-resistance curve is similar to that for
The development of therlo was another attempt manganin, as shown in figure 1, but with M and N
to reduce the thermoelectric power of copper- much closer together in temperature and in re¬
manganese alloys against copper. Instead of sistance.
nickel, an equal percentage by weight of aluminum Other Alloys.—It is probable that all resistance
was added to the copper and manganese. As its alloys that have small temperature coefficients at
name implies, the resulting alloy had a very small room temperature have temperature-resistance
thermoelectric power against copper at room tem¬ curves that are cubics, similar to that of manganin
peratures, less than l/xv/°C. Its other electrical shown in figure 1. These curves are nearly
properties are almost identical with those of man¬ straight in the neighborhood of the inflection
ganin. However, as the thermoelectric power ol points between the maxima and minima. The
manganin against copper is only of the order of 2 ideal resistance alloy for use in instruments would
or 3yuv/°C, the improvement was of little signifi¬ have this inflection point at room temperature
cance. with a zero slope. Moreover, the maximum and
At the National Bureau of Standards [4] an minimum should be at widely separated tempera¬
investigation has been made of copper manganese- tures so that the zero slope would be obtained over
aluminum alloys of the therlo type. There it was the usual range of atmospheric temperatures. Of
found that for a resistance alloy the best composi¬ the alloys already discussed, only gold-chromium
tion is 85 percent of copper, 9.5 percent of man¬ has the inflection point in the neighborhood of
ganese, and 5.5 percent of aluminum. This alloy ordinary room temperatures, the others having
has nearly the same resistivity as manganin, its small slopes because of use near the point of
temperature coefficient at 25° C can be brought to maximum resistance. None of these have small
zero by a suitable heat treatment, and the change coefficients over a very large temperature interval.
in the temperature coefficient with temperature is Alloys of nickel and chromium are commercially
about half that of manganin, at least in the ordi¬ available that have practically linear temperature-
nary range of room temperatures. Its thermo¬ resistance curves over an interval of several
electric power against copper at 25° C is only hundred degrees centigrade, which interval in¬
about 10 percent of that of manganin, and this cludes ordinary atmospheric temperatures. Al¬
thermoelectric power may be further reduced by though the temperature coefficient is constant, it
the addition of a very small percentage of iron, is too large for use in apparatus where the highest
without materially affecting the other properties accuracy is required. Recent attempts to reduce
of the alloy. The stability of such alloys with the coefficients of these alloys by the addition of
time was found to be equal to that of manganin. comparatively small amounts of other materials,
Except for unusual applications, the difference such as copper and aluminum, appear very promis¬
between manganin and therlo is of little importance ing. It is quite possible that an alloy and heat
and the alloys may be used interchangeably. treatment will be developed such that no correction
Gold-chromium.—An alloy of recent introduc¬ for temperature will need be made, at least
tion, which appears to be veryT promising for some throughout the range of laboratory temperatures.
applications, is gold with slightly over 2 percent of
chromium [5]. This alloy has a resistivity at 2. Spools, Winding and Adjustment
room temperatures of about 20 times that of cop¬
per. By baking at fairly low temperatures the In the beginning of the electrical instrument
temperature coefficient can be made extremely industry, wire coils were wound on wooden spools
114/4
like those that are still used for thread. How¬ layer of wire. Although this requires the use of
ever, because of the demand for increased accu¬ smaller wire than for multilayer coils, there are
racy, these wooden spools have been entirely re¬ several advantages. In the first place, the heat
placed by metal spools for resistors of high quality. dissipation is more satisfactory for a single-layer
The reason for the change to metal has been two¬ coil, since a considerable temperature rise may be
fold. In the first place, the wooden spools absorb obtained in the center of a multilayer coil as a
moisture in amounts dependent upon the humidity result of the passage of the current through the
in the air, and expand or contract therefrom. This coil, and the load coefficients are usually large.
results in varying stresses applied to the wire, with Moreover, multilayer coils are more subject to
accompanying changes in resistance. change in resistance because of changes in atmos¬
A more important reason for the use of metal pheric humidity (see section II, 6), and are
spools is the fact that they more readily dissipate usually found to be less stable in resistance with
the heat produced in a coil by the passage of the time.
current. The wire is wound in rather intimate After being wound, the coils are artificially aged
thermal contact with the metal spools, and the by baking in air at about 150° C for 48 hours,
heat is readily transferred to the spools. The en¬ after which they are kept for a considerable period
tire surface of the spools, both inside and outside, of time before final adjustment. Some manu¬
is effective in dissipating heat to the surrounding facturers impregnate coils before baking with a
air. However, for wooden spools the area that is shellac varnish, while others impregnate them
effective in dissipating heat is largely the exposed after baking with special waxes. The final adjust¬
outer surface of the resistance wires. When metal ment is usually accomplished in two steps. The
spools are used, the temperature rise for a given excess wire is cut off in order to make the resist¬
heat dissipation depends primarily upon the size ance just slightly less than the nominal value.
of the spool and only to a minor extent upon the Copper lead-wires, which are usually somewhat
size of the wire. However, the wire size should larger in diameter than the resistance wire, are
be selected so as to cover the spool as completely then silver-soldered to the ends of the coil.
as possible, and if necessary the turns should be Final adjustment is made by filing or scraping the
spaced to prevent bunching of the coil at one end resistance wire near the end, care being taken to
of the spool. see that the metal cuttings are not forced into the
For many alternating-current applications, the insulation. Any filed part is finally painted with
use of metal spools is undesirable or even out of shellac varnish, which is allowed to air dry. For
the question. For such applications, when wire- coils of the highest precision the interval between
wound coils are required, wooden spools may be baking and final adjustment should be as long as
used, although for these purposes ceramic spools practicable, an entire year being desirable.
have come into rather general use. The objections
to ceramic spools for resistors of high precision
are their poor heat conductivity and the fact that 3. Sheet-Metal Resistors
their temperature coefficients of linear expansion
are very much smaller than for the resistance wire. Precision resistors having values of 0.1 ohm or
Metal spools are ordinarily of brass, which has below are usually made of sheet manganin brazed
a coefficient of thermal expansion nearly the same or silver-soldered to heavy copper terminal posts.
as that of the resistance alloys. This avoids large Potential leads are attached as shown in figure
changes in stress in the coils because of tempera¬ 2. Here C and C are copper rods, with binding
ture changes. The spools are ordinarily mounted posts at the top, attached to the sheet of resistance
with their axes vertical, and both ends should be material, S. The resistance material may be a
left at least partially open in order to allow a single sheet of manganin or several sheets in
ready flow of convection currents of air through parallel. The sheets are often not straight but
the spools. Before being wound, the spools are are bent in an S-shape in order that greater
enameled or covered with a single layer of silk, lengths can be used. Additional binding posts
which is impregnated with shellac varnish and P, P are used as potential terminals and are
allowed to air dry. connected by copper wires soldered at some point
The resistance wire is generally double-silk or on the copper terminal bars. For a resistor of
silk and cotton covered, and often the wire is this type the resistance is measured between the
enameled before these coatings are applied. The two branch points. This is to say, the resistance
correct length of resistance wire is cut., doubled at is equal to the ratio of the potential difference
its center, and the center is attached by means of between the potential terminals P, P to the current
a thread near one end of the insulated metal spool. flowing in and out the current terminals C, C.
The two halves are then wound side-by-side (bi- For standards, sheet-metal resistors are mounted
filarly) after which the free ends are tied down in perforated containers with hard rubber or
with silk thread. bakelite tops. These are often used in oil baths
High-quality resistors are wound with only one in order to facilitate the dissipation of the heat
115/5
PC CP are lowered on the current posts [6], This effect
is large when the resistance of the copper terminal
posts is high in proportion to that of the resistance
element. For standards of very low resistance,
large temperature coefficients may result unless
the potential branch points are carefully located.
4. Accelerated Aging
After being wound, resistance coils are impreg¬
nated with a shellac varnish and given an accel¬
erated aging by baking. The temperature for
baking is limited by the silk insulation, which
should not be heated above about 150° C, and the
coils are usually baked at this temperature for 48
hours. As a result of the baking the resistance of
a coil may decrease, sometimes as much as 1 or
2 percent, and a sufficient length of wire must be
used to compensate for this change. After being
baked, the resistance of the coils is much more
stable with time than is that of unbaked coils.
This aging process is often called “annealing”,
but it is doubtful that the improvement in stability
results from the relieving of internal stresses in
the wire, as happens during true annealing. In
the two or three months immediately following
their baking, resistance coils will ordinarily in¬
crease in resistance by an amount that is usually
of the order of 0.01 percent. They then are ready
for final adjustment.
caused by the flow of current through the resistance Sheet-metal resistors are usually painted with
clement. a lacquer or shellac varnish as a protection of
An accelerated aging of these resistors is ob¬ the surface. This coating limits the temperature
tained by heating in air at 150° C for 48 hours, as to which these resistors can be raised during aging,
is done for wire-wound resistors. Final adjust¬ and they are usually treated in the same way as
ment of the resistance may be accomplished by insulated wire-wound coils, being baked at 150° C
changing the points of attachment of the potential for 48 hours. Sheet resistors may be heated at
leads, the resistance decreasing as the points of high temperatures before being lacquered, but
attachment are lowered. Increases in resistance such treatment is apparently no improvement
may be obtained by scraping or by drilling holes over baking at 150° C, as far as subsequent
in the resistance material. Small changes can stability is concerned.
be obtained by filing the copper current posts,
but the direction of the change is dependent upon 5. Annealed Resistors
the location of the potential leads.
Adjustment of sheet metal resistors is often .As has already been stated, the usual baking of
accomplished by making U-shaped saw cuts in wire-wound and sheet-metal resistors is not done
the sheet, with the open ends of the U’s pointing at a sufficiently high temperature to anneal the
in the direction of flow of the current. The U- resistance material. As baking at 150° C im¬
sliaped tabs so formed are bent outward and to proves the stability with time, it would be logical
them are attached the potential leads. Adjust¬ to expect greater stability if the heating were
ment is made by increasing the length of the U carried on at a sufficiently high temperature to
cut. Lengthening one cut will increase the four- obtain actual annealing. In the case of manganin
terminal resistance, whereas lengthening the other this takes place between 500° and 600° C.
cut will decrease the resistance. The annealing of metal is a complicated process,
Not only may the resistance of a four-terminal but the first step is probably a reforming of the
resistor of the type shown be altered by changes metallic crystals to the shape they had before
in the points of attachment of the potential leads, being cold-worked. This results in the relieving
but it is possible to change the temperature co¬ of many of the internal stresses that resulted from
efficient of resistance in the same way. In fact, the distortions of the crystals. The amount of
the temperature coefficient is lowered at the same this restoration of the metal to its preworked
time as the resistance if the points of attachment condition depends upon both the annealing tem-
116/6
perature and the time. In general, the higher instability would be avoided or reduced if a coil
the temperature the shorter the time required were annealed and mounted in a vacuum.
for a given annealing. If the annealing is con¬ Annealed resistors mounted in vacuo have been
tinued after crystals have been restored to their tested at the National Bureau of Standards and
original condition, there may result an actual found to be very stable. It is difficult, however,
uniting of adjacent crystals, with an accompany¬ to seal the coils in suitable containers for high
ing decrease in the mechanical strength of the evacuation. Equally good results have been
material. obtained with annealed coils mounted in sealed
For base-metal alloys, annealing should take metal containers filled' with dry air. Whatever
place in a vacuum or in an inert atmosphere to the effect of the air, an equilibrium condition is
avoid a reaction between the metal and the sur¬ soon obtained when the supply of air is limited.
rounding air. Such reaction might take place Probably the most stable resistors that have been
inside the metal at the intercrystalline boundaries made are a group of annealed 1-ohm manganin
as well as at the surface of the metal. These resistors mounted in double-walled air-filled con¬
reactions do not necessarily decrease the stability tainers, now being used at the National Bureau of
of resistance with time if the products are stable. Standards [8J for maintenance of the unit of re¬
There may, however, be a selective reaction that sistance.
makes the material inhomogeneous, and this might Although good annealing improves the stability
have a considerable effect on the resistivity and of sealed resistors, it is apparently of no special
temperature coefficients. value for unsealed coils. When mounted in open
When a resistor is made by winding wire on a containers, annealed resistors cannot be expected
spool, the wire usually will not straighten if it is to be any better than, if as good as, those haked
removed from the spool. This means that parts at 150° C. This is true even if the resistors are
of the wire have been stressed past their elastic varnished or enameled after the annealing. None
limit and a permanent deformation has taken of these coatings is impervious to the atmos¬
place. For such a bent wire the portions farthest phere, and they merely retard any reaction be¬
from the center of the spool have been elongated tween the air and the resistance material.
past their elastic limit, while the filaments nearest
the center of the spool have been compressed 6. Effects of Humidity
beyond the elastic limit. Intermediate filaments
are subjected to stresses that depend upon the It lias long been known that wire-wound re¬
changes in their length, which resulted from the sistors undergo seasonal variations in resistance,
bending. The stress distribution is from a maxi¬ being higher in resistance in summer and lower in
mum in tension to zero and then to a maximum winter. This effect is most noticeable in high
in compression. These stresses are superimposed resistance coils of small wire, and even in high-
upon the stresses that were produced in the wire grade resistors may amount to several hundredths
as it was being fabricated. of a percent of the resistance. The effect is to a
These internal stresses in a wire result in a large extent a result of changes in average humid¬
slight change in the shape of its cross section. In ity, and is greatest in climates where there is a
addition to this change there is a change in re¬ large difference in humidity between winter and
sistivity, which results from the presence of the summer. This seasonal change was first observed
stress [7]. The resistivity is therefore not uni¬ in the case of manganin resistors made with silk-
form across the wire, and the difference between covered wire. The accepted explanation was
parts of the wire may amount to nearly 1 percent that the resistance changes resulted from dimen¬
in the case of manganin, and perhaps more for sional changes of the shellac with which the coils
other alloys. Although the resistivity changes were impregnated, as the shellac absorbed ir gave
considerably when a wire is bent, there is not off water vapor.
necessarily much change in resistance, as the The effect of moisture on resistors has been
change in resistance of the parts under compression thoroughly investigated by Dike [9], who came
may compensate for the change in the parts under to the conclusion that the effect of changes in
tension. humidity is to change the tension in the silk with
The effect of annealing of a coil of wire is to which the wire is customarily insulated. This
reduce the internal stresses, thus making the change in tension changes the pressure trans¬
resistivity more nearly uniform in the wire. It is mitted to the wire by the insulation and hence
probable that a slow annealing takes place at changes the resistance. He also found that the
room temperatures, and the accompanying reduc¬ effect of humidity on cotton insulation is opposite
tion in the internal stresses may be one reason to that on silk, and by using a mixture of cotton
for the change with time of the resistance of a coil and silk fibres for insulating the wire he was able
of wire. Another cause for change might well to eliminate most of the seasonal changes in re¬
be some reaction between the wire and the sur¬ sistance that result from changes in humidity.
rounding atmosphere. Both of these sources of For standard resistors the effect of humidity
117/7
heating. The objection to the oil is the fact that
it may in time become somewhat acid, and the
acid may corrode the resistance wire or injure
the insulation.
The advantage of ready dissipation of heat
combined with the advantage of hermetic sealing
is found in the double-walled type of standard
resistor [8] developed at the Bureau. In this type
the container is made of coaxial cylinders only
slightly different in diameter with the space
between the cylinders sealed. The resistance
element is mounted in this sealed space in good
thermal contact with the smaller cylinder, which
serves as the inside wall of the container. One
of these resistors is shown in figure 4. The outside
diameter of the container is 9 cm and its length
13 cm. The series of holes near the top are just
above the double-walled part and are intended to
increase the facilities for cooling, and the con¬
tainers are left open at the bottom for the same
purpose. These double-walled resistors readily
give up heat to an oil bath yet are not affected
by humidity changes. The sealed space in which
the coil is mounted is filled with dry air, and no oil
comes in contact with the resistance material.
The seasonal changes in resistance of standard
resistors that arise from changes in humidity
are readily eliminated, as has just been discussed,
by sealing in metal containers. This arrangement
is not satisfactory for large instruments and meas¬
uring apparatus, which are not readily sealed.
If the cases of such equipment are reasonably
tight a drier, such as calcium chloride, may be
kept inside the case. This procedure is somewhat
may bo also eliminated by mounting the coils hazardous, since if not replaced with a sufficient
in sealed containers. This procedure was first frequency the drier may become dissolved in
advocated by Rosa [10] who designed the resistor
shown in figure 3, known as the NBS type of
standard resistor. In this figure, R represents
the manganin coil mounted on a silk-insulated
brass spool and baked as described in section
II, 4. This coil is supported from the liard-
rubber top, T, by means of the thermometer
tube, W, which is so arranged that a thermometer
can be inserted from the outside. The copper
lead wires, which are silver-soldered to the ends
of the resistance coil, are in turn soft-soldered to
the copper binding posts, B. The hard-rubber
top is screwed into the metal container, C, which
is filled with a good quality light mineral oil.
The binding posts, thermometer well and the
threads by which the hard-rubber top is connected
to the container are all sealed with shellac, which
is not soluble in oil.
The purpose of the oil is to give good thermal
contact between the resistor and the case and to
facilitate the dissipation of the heat developed
in the resistor by the current through it. In
addition, the oil in effect increases the heat
capacity of the resistor, thus increasing the cur¬
rent that it can carry temporarily without over¬ Figure 4. Double-walled standard resistor.
118/8
adsorbed water and be spilled on the coils. Tlie changes of resistivity. For such resistors the
use of silica gel avoids this danger. A more effect of the heating may usually be taken into
satisfactory method is to mount a heater in the account by measuring the temperature change and
apparatus and, by means of a thermostat, main¬ calculating the change in resistance from the
tain the temperature of a metal box containing temperature coefficients of the resistor. For
important resistors at a constant value, above that sheet-metal resistors the effect of stress changes
of the laboratory. In this way the relative may be important and the calculated change cannot
humidity is kept always low and seasonal varia¬ be relied upon.
tions may be reduced. Additional advantages It should be evident from the above discussion
are that no temperature corrections need be that the load coefficient of a resistor is a rather
applied when calibration is made under conditions indefinite quantity, varying with time of flow of
of use, and that resistors will usually be more the current and with the environment of the resis¬
stable in resistance if kept at a constant tempera¬ tor. To be of value it should be measured under
ture instead of being allowed to follow variations conditions of use. Measurement is usually made
in laboratory temperatures. by passing the desired current through the resistor
and a second resistor connected in series, measuring
7. Load Coefficients the ratio of resistance both with negligible and
with the required heating. This ratio may be
The load coefficient of a resistor is defined as measured by means of a bridge or a potentiometer,
the proportional change in resistance caused by and the comparison resistor should be one that
the production of heat in the resistor at the rate is not appreciably affected by the test current.
of one watt. This change, however, is a function A resistor of one-tenth or less the resistance of that
of time. When a current starts flowing in a under test should be used, and its‘load will be
wire-wound resistor the wire very quickly takes one-tenth or less than for the resistor under test.
up a temperature above that of the spool on which For the reference lower-valued resistor one that is
the wire is mounted. The amount of this change known to have a small load coefficient should be
depends upon the thermal contact between the chosen. If such a resistor is not available one
wire and the spool. If the current continues to with a low temperature coefficient should be se¬
flow, this initial difference in temperature is lected, on the assumption that its load coefficient
maintained, but the coil and spool continue to is correspondingly low, and its load coefficient
rise in temperature together. The maximum should bo roughly determined to make sure that
temperature that will then be attained depends it is small. This can be done by balancing the
upon the facilities for dissipation of heat by the resistor in any bridge using a small test current.
spool, and often upon the facilities for cooling From an external source a large current is then
the material to which heat from the spool flows. sent through the resistor under test for several
Suppose we have a standard resistor of the NBS minutes, after which the extra circuit is discon¬
type resting on a table and we send through it a nected and the bridge circuit is again balanced.
sufficient current to liberate one watt in the coil. This last balance must be made quickly before the
Within about 30 seconds the temperature of the coil heat from the large current is dissipated.
will rise to a steady value about 1° C above that If a resistor is being used under conditions where
of its spool. Spool and coil will then continue to the load changes its resistance, the amount of the
rise in temperature at the rate of about 10° or 15° change may often be determined experimentally.
C per hour, and if allowed to continue will in about To do this the heating may be doubled, by in¬
2 hours reach a steady temperature of some 10° creasing the current by 40 percent, and the result¬
to 20° C above that of the room. This may cause ing change noted. To a first approximation this
a permanent change in resistance, which, however, doubling of the heating doubles the change in re¬
ordinarily amounts to only a few parts in a million. sistance, and twice the change should be subtracted
If, instead of being mounted in an oil-filled con¬ algebraically from the final value. Such a pro¬
tainer, the same resistor had been left open in the cedure is especially satisfactory in the case of
air of the room, its final rise in temperature would Wheatstone bridges, as the procedure will correct
probably have been less by some 50 percent and back to zero test current, whichever branch or
would have been reached in 15 to 20 minutes. branches are being changed by the current.
The rise in temperature of the resistance coil will
be accompanied by a change in resistance, whose 8. Stability of Resistors With Time
amount depends upon the temperature coefficient
of resistance of the wire and also upon changes in In applications where stability with time is of
stress in the wire because of dimensional changes importance, as for instance for standard resistors or
of the coil and its support. For single-layer precision measuring apparatus, manganin is used
manganin coils mounted on brass spools the relative almost exclusively. In such applications low tem¬
changes in dimensions are small, and the change in perature coefficients of resistance and small ther¬
resistance results primarily from temperature mal emf’s against copper are usually also required,
119/9
and few alloys other than manganin are suitable. changes in the unit as maintained by the manganin
Consequently, a discussion of the stability of re¬ resistors.
sistors is largely a discussion of the stability of The average user of standard resistors is interest¬
manganin. ed in the stability that may be expected from
Most national standardizing laboratories keep a standards available commercially. In this con¬
selected group of manganin resistors, which are nection, an analysis made in 1941 at this Bureau is
regularly intercompared and used for maintenance relevant. Of nearly 600 standard resistors that
of the unit of resistance. The relative values of had been submitted more than once to this Bu¬
such standards are remarkably constant, the in¬ reau for test, the average yearly change in resist¬
dividual resistances not changing by more than ance, without regard to sign, was found to be
one or two parts in a million per year with refer¬ 8 ppm. Of the total only 2 percent averaged
ence to one another and for some groups very much greater than 60 ppm per year, and for nearly 90
less. It is supposed that the average of a group re¬ percent of all standards tested the annual change
mains constant to a high degree, but such stability was 10 ppm or less. If signs were neglected there
can be only assumed. No method of measurement was no significant difference between the average
has been used that would detect with certainty yearly change of sealed and unsealed resistors.
This would appear to mean that sealing merely
changes of less than 10 or 20 parts in a million in
reduces seasonal variations without improving the
the group as a whole. The international ohm was
long-time stability. However, if regard is taken
defined as the resistance of a mercury column of of sign, the performance of sealed and unsealed
specified dimensions at the temperature of melting resistors was quite different. In this case the
ice. However, such resistors have not been con¬ average yearly change was about —.3 ppm for
structed with sufficient accuracy to demonstrate sealed and about +5 ppm for unsealed standards.
the performance of manganin resistors used to That is to say, sealed standards about as often
maintain the unit. Likewise, absolute ohm deter¬ decrease as increase in resistance with time,
minations have not been sufficiently reproducible whereas the change in unsealed standard resistors
to give such information. If the unit as main¬ is predominantly upward. In connection with
the sealed resistors it is interesting to note that
tained by means of manganin resistors were tested
practically the same result would have been obtained
every ten years by comparing against mercury
if the unit of resistance had been maintained by
ohm. or absolute ohm determinations, the apparent supposing the average value of the 400 sealed
change would probably not exceed 10 or 20 parts resistors had remained constant, as was obtained
per million. This could be just as well attributed by assuming the average of a group of 10 selected
to errors in realizing the unit experimentally as to resistors of special construction to be constant.
120/10
may be neglected and R may be calculated as the requiring high precision such as the determination
ratio of the instrument readings. If the resist¬ of temperature rise or of the conductivity of line
ance of the voltmeter is not large as compared conductors. When used with circuits that are
with R, the latter may be calculated from the highly inductive or capacitive, the precautions
equation appropriate for such resistance measurements
should be observed.
Ohmmeters may be classified according to
“~I(l-E/IRt)
either their principles of operation, their source
where Rv is the resistance of the voltmeter, its of energy, or their range.
reading being E volts, and the ammeter reading / The principles commonly used are: Simple
amperes. If the resistance of the ammeter is ohmmeter, ratio ohmmeter, Wheatstone bridge.
known or if it is negligible as compared with the In the simple ohmmeter a source, the voltage
of which is assumed to be definite and to corre¬
unknown resistance, the voltmeter may be con¬
spond to the calibration of the instrument, is
nected across both the resistance and the ammeter,
as shown in figure 6. In this case the ratio of applied to the unknown resistor and the resulting
current causes an indicating instrument to deflect
over a scale. This scale is so graduated that the
pointer indicates directly the resistance in ohms
(or megohms). In many cases provision is made
by a magnetic or electric shunt to adjust the in¬
strument at one point, usually at zero resistance,
to fit the existing value of the voltage. In some
of these instruments the final indication is by a
vacuum-tube voltmeter, which measures the drop
produced in a very high resistance by the current
through the specimen.
Figure 6. Alternate connection for ammeter-voltmeter
In the ratio meter or “crossed-coil” type of
measurement of resistance.
ohmmeter, the current through the unknown
resistor flows in one of the coils, while the other
voltmeter to ammeter readings gives the total carries a current that is proportional to the voltage.
resistance between the points of attachment of The current is led to the coils by ligaments, which
the voltmeter, that is, the resistance of the am¬ exert a negligible torque so that the moving system
meter and of the connecting leads is included. takes up a position that depends on the relative
That is magnitude of the currents in the two coils. The
scale can therefore be laid off to indicate resistance
/* -f-tf.!, (6) directly, and the indication will be independent of
the voltage used, provided that the resistor under
where i?4 is the resistance of the ammeter and of test obeys Ohm’s Law.
all lead wires between the points of attachment of The designation “ohmmeter” is also applied
the voltmeter leads. (though perhaps incorrectly) to certain forms of
With suitably calibrated, instruments it is possi¬ the Wheatstone bridge in which the dial that
ble to measure resistance to 0.1 or 0.2 percent pro¬ adjusts the balance is calibrated to read directly
vided the instruments are of such ranges that the value of the unknown resistance.
large deflections are obtained. For this accuracy
it is usually necessary to calibrate the voltmeter 3. Potentiometer Method
with the same leads as those that are to be used
in the resistance measurements. It was pointed out in section III, 1 that the
ammeter-voltmeter method for measuring resist¬
2. Ohmmeters ance is complicated by the current drawn by the
voltmeter. Such complications may be avoided
Ohmmeters are instruments for indicating by using a voltmeter of some type that requires
directly on a scale, with a minimum of manipula¬ no current from the circuit being measured, i. e.,
tion or computation, the resistance of the circuit electrostatic or vacuum tube voltmeters.
connected across their terminals. They are avail¬ Potentiometers are also suitable for the measure¬
able in a wide range from milliohmmeters reading ment of potential difference when it is desired to
to 0.001 ohm to megolimmeters reading to 50,000 avoid drawing a current from the source of
megohms. Their accuracy is limited both by the potential difference. They are especially good in
calibration and reading of the indicating instru¬ cases where an accuracy of 0.1 percent or better is
ment and in the simple ohmmeter by their depend¬ required, as such accuracy is difficult to attain with
ence upon a fixed value of voltage. They are, deflecting instruments. Where a potentiometer
therefore, in general, not suited for applications is available, the ammeter-voltmeter method may
122/12
bridge is more suitable. The choice between the
simple and the double bridge is usually made on
the basis of the required accuracy.
It is difficult to attach a copper lead wire to a
resistor by means of binding posts or other
clamped connections without introducing an un¬
known contact resistance of the order of 0.0001
ohm. For resistors above 1 ohm such an uncer¬
tainty is usually negligible, whereas for a resistor
of say 0.01 ohm the uncertainty is 1 percent.
For standard resistors the contact resistance is
often reduced by amalgamating the contact sur¬
faces. For clean well-fitted contacts the resistance
then amounts to only a few microhms, but the re¬
sistance of such contacts will rise with time as
the copper combines with the mercury to form a
granular material. This material should be re¬
meet, and it is necessary to devise methods of use moved every few' rqonths by scraping and wiping
for which the conditions need be only approxi¬ the surfaces.
mately met, if high accuracy is to be attained. Where accuracy is required, low-valued resistors
Such is that due to Kohlrausell [ 11J, and known are usually of the four-terminal type. For these,
as his method of overlapping shunts. two leads are soldered or brazed to each end of the
For Kolilrausch’s method the circuit is as shown resistance material, as shown in figure 9. The
in figure 8. The two galvanometer windings, G, p p
are connected respectively across A and S, and in f f
addition each .shunt bridges the resistance, L,
which is used to connect A" and S in series. The
resistance of L is usually small as compared with
the other resistances. In series with the galvanom¬ •-l-V\A/VWWWJ-•
eter windings are connected the resistors, Rx and Fihure 9. Four-terminal resistor.
R2, one or both of which are adjustable. In addi¬
tion, a special switch not shown must be used, resistance in question is that between the branch
which will interchange the battery B and the lead points at the two ends. That is to say, the resist¬
resistance L. This must be done without materi¬ ance is the ratio of the potential difference between
ally changing the current furnished by the battery. the terminals P, P to the current flowing in and
The effect of the interchange is to reverse the out the current terminals, C, C. Methods of
current through both galvanometer windings. measurement are used such that any effect from
The values of the adjustable standard, S, and the the lead resistances is avoided or reduced to a
rheostat, or R2, are adjusted until there is no negligible amount. For such a purpose the
deflection of the galvanometer for either position potentiometer method is suitable as no appreciable
of the batfery, or until the deflection is the same current is drawn through the potential leads, and
in magnitude and direction for both battery posi¬ the potential drop between branch points is inde¬
tions. Under either condition of adjustment, pendent of the magnitude of the lead resistances in
and X are equal. the current circuit. Double-bridge methods bal¬
In actual practice, where (S' and A" are standard ance out the lead resistances or connect them in
resistors under comparison, the larger is made ad¬ high-resistance branches where they are negligible.
justable by a precision rheostat connected in Simple Wheatstone bridge.—The simple Wheat¬
parallel, and the amount of the shunt required to stone bridge is primarily a group of four resistors
make A and S equal allowTs an easy calculation connected in series-parallel as shown in figure 10.
of the difference in resistance between the two. A current, /, is passed through the two parallel
The balance is exact even if the two circuits of branches, and G is a detector connected to the
the galvanometer are not exactly alike electrically junctions of the resistors as shown. It may be
or magnetically. readily shown that if there is no potential differ¬
ence across the detector, G, the relation between
5. Bridge Methods the resistances of the four arms is as follows:
123/13
tion for a shorter time. Other things being equal,
ratio coils of from 10 to 100 ohms are therefore
about the best compromise.
The choice of resistance for the adjustable arm
of a Wheatstone bridge is influenced by the same
factors as the choice of resistances for the ratio
arms. That is to sav, the lower the value the
greater the stability in resistance but the more
troublesome the contact resistances become, and
in the adjustable arm several contact resistances
are required. Steps smaller than 0.1 ohm, which
allow readings of the adjustable arm to about 0.01
ohm by interpolation from galvanometer deflec¬
tions, are seldom used as it is not wise to rely upon
the combined effect of the several contacts of the
arm to be definite to much better than 0.01 ohm.
Wheatstone bridges for the measurement of re¬
sistance to 0.1 percent are available commercially
at moderate prices. To this accuracy these can
usually be relied upon without the application of
corrections to the readings. However, for meas¬
urements to 0.01 percent, corrections to the read¬
ings of the ratio arms and of the rheostat arm must
The detector, G, is usually a galvanometer, and usually be applied, and the bridge must be main¬
the lack of a potential difference across it is evi¬ tained at the temperature of calibration within a
denced by a lack of motion of the galvanometer few degrees. Primarily because of the effects of
coil if a switch or key in series with G is opened or changes in humidity, calibrations of a Wheatstone
closed, or better, if the battery circuit is opened bridge must be made rather frequently if an accu¬
or closed. Opening or closing of the battery, cir¬ racy of 0.01 percent is to be attained. This is
cuit should be avoided, however, if there is especially true where there is a marked change in
inductance or capacitance in the circuit of the ambient conditions. For example, between winter
unknown resistance or in any arm of the "Wheat¬ and summer the ratio coils and the rheostat arm
stone bridge. In measuring the resistance of a may each change by 0.01 percent or more, and the
field winding of an electric motor, for example, errors may be additive rather than compensating.
the galvanometer may be damaged if its circuit is The use of air-conditioned laboratories improves
left closed and the battery circuit of the bridge is the performance markedly, but even then it is
opened. In this case the current should be left advisable to make occasional spot checks by
on until a steady state is reached before testing the measuring standard resistors. Whenever possible,
bridge balance by opening and closing the gal¬ ratio arms below 10 ohms or above 1,000 ohms
vanometer circuit. should be avoided, as should rheostat readings in
From the above equation it is evident that the excess of 1,000 ohms. This means that the Wheat¬
resistance of any one of the four arms, say .V, can stone bridge is best suited for the measurement of
be calculated if the resistances of the other three resistance in the range 10 to 10,000 ohms.
are known, or if one of the three and the ratio of When an accuracy greater than 0.01 percent is
the other two are known. The usual general pur¬ required, special bridges are required or special
pose Wheatstone bridge is made in such a way that techniques are used. One of the best of the special
any one of several known resistors may be selected methods is that of substitution, in which the un¬
for use as either A or B and hence their ratio may known resistor is replaced with a standard resistor
have any one of a number of values. & is then a or resistors having the same nominal resistance as
known resistor whose value may be adjusted in that of the unknown. The bridge that is being
small steps over a wide range of resistance. used is then relied upon to. determine only the
In commercial instruments the ratio arms A and difference between the standard and the unknown,
B are usually coils that are connected into the and this difference need not be accurately meas¬
bridge by inserting suitable plugs. As there are ured. If the unknown and standard differ by 0.1
resistances in these plug contacts, which may be percent , the difference need be determined to only
rather variable, it is desirable to use relatively high 1 percent to give the unknown to 10 parts in a
values of resistance for A and B in order to reduce million in terms of the standard. Nearly any good
the uncertainty in the resistance of the ratio arms. bridge can be so used without calibration, provided
On the other hand, high-resistance coils are more contact resistances are sufficiently constant that
affected by humidity and are less stable in resist¬ readings can be repeated to the desired precision.
ance with time and therefore retain their calibra¬ It is evident that the calibration of the measuring
124/14
bridge is of least importance when the standard
and unknown resistances are nearest equal.
Standard resistors are usually available only in
decimal multiples or submultiples of an ohm and
for many resistance measurements the lack of a
suitable standard prevents the use of the substitu¬
tion method.
Mueller bridge.-—For the accurate measurement
of odd-sized resistances, one of the best bridges
is that designed by E. F. Mueller [12] of the
National Bureau of Standards for use in resistance
thermometry. This is a special bridge intended
for the accurate measurement of resistances up to
about 110 ohms. The effects of humidity changes
are greatly reduced by mounting the resistors in a
compartment that is electrically heated and whose
temperature is maintained at 35° C by means of
a thermostat. As this is a temperature to which
the laboratory, temperature seldom rises, the
relative humidity in the compartment housing
the resistors is kept low and variations in the low
humidities have little effect on the resistances of
the coils.
For the Mueller bridge the ratio arms are equal, Figure 12. Connection of bridge to 4-lead thermometer.
and the arms can be interchanged to test the
equality. A small slide wire is connected between the resistor A' and the lead resistor are connected
the ratio coils, and the sliding contact is used as into the right-hand arm, and the lead resistor Lx is
the branch point for the bridge as shown in figure in the adjustable arm, R. With equal ratio arms,
11. The operator can set the ratio arms to equality R and X would be equal for a balance of the
by properly proportioning the resistance of the bridge only if Lx and Z4 are equal. Instead of
slide wire between the two ratio arms by changing adjusting Lx and i4 to exact equality, their con¬
nections are interchanged and a second bridge
balance is obtained, the average of the two balances
being that which would be obtained with Lx and
Li equal. Alien Lx and Z4 are interchanged, it is
necessary to shift the branch point from lead 2 to
lead 3 in order to keep X in the right-hand arm.
In actual use, instead of the lead wires being inter¬
changed, the internal connections of the bridge
arms to the binding posts 1 and 4 are interchanged
to obtain the same result. This interchange is
effected by means of an amalgamated switch for
which the uncertainty in contact resistance is only
Figure 11. Ratio arms of Mueller bridge. a few microhms.
The rheostat arm of the Mueller bridge is ad¬
justable in steps as small as 0.0001 ohm. In order
the setting of the slide contact. This setting is
that such steps should not be masked by changes
correct if no change in the bridge results when the
in contact resistance, special types [13] of decades
arms A and B are interchanged. The remaining
are used. The 0.0001-ohm-per-step decade is
two arms are the adjustable rheostat and the un¬
approximately as shown in figure 13. With the
known resistor, which must be of the four-terminal
type. Resistance thermometers are usually of the 10
four-lead type, and other resistors should be pro¬
vided with four leads when they are to be measured
on a bridge of this type. The method of connec¬
tion of the unknown resistance is as shown in figure
12. R is the adjustable rheostat arm that will be
discussed in detail later. X is a four-terminal
resistor with lead wires Lu Z2, L3, and Z4 connected
to the binding posts 1, 2, 3, and 4, respectively. If
the galvanometer is connected to binding post 2, Figure 13. Shunt-type decade.
125/15
switch contact on the 0 stud, the resistances of 12
ohms and 0.110 ohm are connected in parallel, and
the resistance between the terminals £ and T is
about 0.1090 ohm. When the switch contact is
moved to stud 1, the 12-ohm branch is increased
sufficiently to increase the parallel resistance from
0.1090 to 0.1091 ohm, an increase of 0.0001 ohm.
Likewise when the switch is moved to 2 the parallel
resistance increases to 0.1092 ohm, etc. Thus the
resistance between the terminals S and T may be
increased in steps of 0.0001 ohm, although the
value is not zero when the switch is set to read zero.
The 0.001- and 0.01-ohm-per-step decades are
made in the same way as the 0.0001-ohm decade.
When these three decades are connected in series
and all set to zero, there is a series resistance of
about 1.6 ohms. However, since the bridge has
equal ratio arms an equal resistance of 1.6 ohms
may be included in the A'-arm, and the value of
X is measured by observing the increase in the
resistance of the rheostat arm when X is inserted 0.1-ohm resistors are in series in the A'-arm.
in its arm. A short-circuiting plug is provided When the dial reading is increased, the resistance
with the bridge that connects together the ter¬ is removed from the A'-arm, the effect being the
minals 1 and 4 of figure 12. The bridge is then same as if equal resistance were added to the
balanced with and without this shorting plug, rheostat arm, R, since the bridge has equal arms.
and the difference in the readings of the rheostat This may be seen from the following consideration.
arm gives the resistance of A', and leads. Assume that the bridge is balanced with the un¬
The purpose of the shunted type of decade that known resistor and the 10 steps of the 0.1-ohm
was described above is to reduce the effect of dial all connected in the A'-arm. If now A' is
variations in contact resistances and transient increased by 0.1 ohm, the A'- and i?-arms may be
emf’s in the switch. Referring again to figure 13, brought again to equality by either increasing R
it is seen that the switch contact is in the high by 0.1 ohm or by decreasing the resistance in
resistance branch. The ratio of resistance for the series with A' by 0.1 ohm. The readings of the
two branches is about 100 to 1, and it may be dial in the A'-arm are such that increases in
readily shown by differentiation that the effect readings of the dial correspond to decreases in the
of a variation of switch contact resistance is resistance in the A'-arm.
reduced by the square of this ratio. Thus a con¬ The 10-ohm-per-step decade of the rheostat,
tact variation as high as 0.01 ohm would vary the not shown, is the only one that has contact resist¬
resistance from S' to T by only 1 microhm, which ances directly in series with the arm. A special
is negligible in the rheostat arm where the mini¬ type of dial that has amalgamated contacts is used
mum steps are 100 microhms. For the 0.001 and for this purpose. Such contacts are uncertain by
0.01 decades, the ratio of currents in the branches only a few microhms, which may be tolerated in
is less than 100 to 1, and less variation in the the rheostat arm where the minimum steps are
switch-contact resistance can be tolerated. 100 microhms. For Mueller bridges available
The 1 -ohm and 0.1 -ohm steps of the rheostat arm commercially, the galvanometer and battery are
are not made of the shunt type, as the resistance interchanged from the positions assumed in the
with the dial set on zero would be rather large. above discussion. This in no way affects the
The contact resistances for the 1-ohm dial are validity of the conclusions.
thrown into the ratio arm A, as shown in figure 14. Adjustable-ratio bridges. As has already been
The bridge current is introduced through the stated, the Wheatstone bridge shown in figure 10
10 1-ohm coils of this decade, and the switch is balanced if
merely changed the point of connection of the
ratio arm, without opening the circuit of the 1-ohm X=S^, (10)
decade. In this case a variation o: the switch
contact resistance changes the resistance of ratio and this balance may be realized by keeping A
arm A. This effect is reduced by using high and B fixed and adjusting S, or by keeping S fixed
resistance ratio arms, A and B each being 1,000 and adjusting the ratio A/B. In most commercial
ohms. The 0.1-olim-per-step dial is arranged in Wheatstone bridges the balance is obtained with
the same way as the 1-ohm-per-step dial, with the S. However, for comparing nominally equal
switch contact in series with the B ratio arm. resistances, bridges are sometimes constructed in
With the 0.1-ohm dial set at zero, all 10 of the which the ratio is adjustable.
126/16
For measurement, a four-terminal resistor is
usually connected in series with a four-terminal
standard. The ratio of the potential differences
across the two resistors is then determined when a,
current flows through the two. This ratio may be
determined by means of a potentiometer or in
terms of a resistance ratio by means of the Kelvin
double-bridge.
The circuit of the Kelvin double bridge is
shown in figure 16, where A" and S denote the
unknown and standard resistors, each with both
current and potential terminals. The two are
connected by means of a conductor L, preferably
of low resistance as compared with A" or S. The
resistors A and B and also a and b provide resist¬
ance ratios that must be known, as must also be
the value of S. This bridge is balanced if there
is no change in the deflection of the galvanometer
when the current circuit is opened or closed.
When balanced the following relation holds
between the values of the resistance:
An adjustable-ratio bridge is very convenient
for use in sorting large numbers of resistors that
must be equal within given tolerances. Such a A
bridge may be constructed as shown in figure 15, AAfS Is (11)
B aA b± L
where A and B are the two arms of a resistance
ratio that is adjusted by changing A. A resistor This equation would be exactly the same as that
is connected to the binding posts S, and the resist¬ for a simple Wheatstone bridge if the last term
ors that are to be compared with it are connected on the right side were zero or negligibly small as
in turn to the X terminals. If the resistors are compared with the term S A/B. That last term
to be the same to, say, 0.1 percent, the arm A is can be made zero by making a/b —A/B, irrespective
made to have either of two values, which give of the value of the link resistance, L. However,
ratios, A/B, of 0.999 or 1.001. This could be the smaller the value of L, the less important is
realized by making B 100 ohms and A either 99.9 any lack of equality between the ratios A/B and
ohms or 100.1 ohms, the latter accomplished, for
a/b.
instance, by making A equal to 100.1 ohms and In actual use two methods are employed for
having a plug to short-circuit 0.2 ohm of this total. balancing the double-bridge. The resistors A
The procedure is then to observe the direction of and B, and also a and b may be fixed ratio coils
deflection of the galvanometer with either ratio so chosen that AIB=a/b and the bridge balanced
setting and then change to the other ratio. If by varying S, which is an appropriate adjustable
this causes a reversal of the direction of deflection
standard; or S may be a fixed standard resistor
of the galvanometer, the correct balance of the and the bridge balanced by adjusting the ratios
bridge is between the two ratio settings. In other A/B and a/b, keeping these ratios at all times
words the ratio of X/S is between the two ratios equal.
0.999 and 1.001, i. e., the two are equal within When a double-bridge is to be balanced by means
±0.1 percent.
Besides this use in limit bridges, adjustable ratio
bridges may be used for precise measurements of
resistors in terms of nominally equal standards.
Such bridges when used in conjunction with the
substitution method permit very accurate meas¬
urements of resistance. Such an adjustable ratio
is the “ direct-reading ratio set” described in section
IV, 1, where the appropriate techniques are also
discussed.
Kelvin double-bridge.—As has been already
stated, four-terminal resistors are used in order
to avoid uncertainties arising from variations in
contact resistances. For precision work, resistors
of 1 ohm or below are usually of the four-terminal
type.
127/17
of an adjustable standard, care should be taken
in connecting up such a standard in order to keep
the resistance small between it and the unknown.
Adjustable standards are usually made such that
the position of one or both of the potential
terminals is adjustable. As a result, varying
amounts of resistance are left at the ends in series
with the used part of the resistor. The adjustable
standard should be connected in the circuit in
the way that connects the larger part of unused
resistance in the external circuit rather than
between resistors so as to form part of the link
resistance, L.
When a fixed standard is used for S, and the
bridge is balanced by adjusting the ratios A/B
and a/b, the most convenient arrangement is to
have the ratios adjustable together. This is ratio A and B are such that their ratio is known
accomplished by having the same dial handle only for the resistance between the two binding
operate two dials together so that both ratios are posts s and s'. The lead resistors LA and LB and
changed simultaneously. This arrangement re¬ the internal leads in A” and S' add to the resistance
quires special apparatus, but such double-ratio of A and B. This ordinarily prevents the ratio
sets are commercially available. of the bridge arms from being the same as the
If not “ganged” together, the two ratios are known ratio of A to B. In other words
changed separately but always by the same
amount. Actually, however, it is not necessary to A A La A
(12)
be able to adjust the auxiliary ratio, a/b, in known B-TLb B
steps, and any convenient adjustable ratio, such
except under the condition that LA and LB are
as a slide wire, may be used. In this case the
respectively negligible as compared with A and B,
balance of the double bridge is by successive ap¬
or under the condition that the ratio of the leads
proximation. It is first balanced by adjusting the
is the same as the ratio of A to B. That is
main ratio, A/B, after which the circuit is opened
at the link, L, and a balance now obtained by A-\-La_A
varying the auxiliary ratio a/b. The link circuit (13)
B-\-Lb b
is again closed and a new balance obtained with
the main ratio. The procedure is repeated until La=A
if (14)
the same balance is obtained with or without the Lb B
link circuit’s being open. Ordinarily the final
If the lead resistances are small as compared with
balance is obtained after only a few sets of ad¬
A and B, the second equation need be only ap¬
justments, and accurate results may be so at¬
proximately satisfied.
tained. Since only one ratio need be known any
The same considerations apply to the lead re¬
apparatus having such a ratio, as for example a
sistance La and Lb in the auxiliary ratio arms.
simple Wheatstone bridge, may be used in con¬
Their effect may be made negligible by making
junction with a slide wire and a fixed standard
their ratio the same as the ratio a/b.
resistor to make good measurements by the
The following procedure for balancing the bridge
double-bridge method.
and eliminating the effects of the leads by adjust¬
In work of the highest accuracy, account must
ing their ratio may be followed. Arrangement is
be taken of the resistance of the leads that are
made for readily opening and closing the circuit
used to connect the ratio arms to the unknown
at L and for short-circuiting the main ratio arms
and standard resistors. There is also lead re¬
by connecting together s and .s'. Shortings to s'
sistance in these resistors, between the branch
is the equivalent of reducing A and B both to zero,
points and the terminal binding posts, which is
leaving only the lead resistances in the arms. If
not necessarily negligible. One way of taking
the ratios are ganged and nominally equal, the
these lead resistances into account is To use such
bridge is balanced in the following steps:
high resistances in the ratio arms that the con¬
1. With the circuit as shown, the ratios A/B
necting leads have a negligible effect. Usually,
and a/b are adjusted until the galvanometer
however, this reduces the sensitivity of balance of
indicates a balance. This gives
the bridge. A better arrangement is to make the
lead resistances adjustable and select their ratio A+La X . , ,
in such a way as to balance out their effects. A A-, , T = o approximately. (15)
JL> -f- O
convenient method of adjusting the lead re¬
sistances is as follows: Referring to figure 17, the 2. The main ratio A/B is now short-circuited
128/18
and the bridge is again balanced, by changing the
ratio of the lead resistances. This may be done
by adjusting the length of one lead wire, or a small
rheostat may be used in series with one of the
leads. This step makes the ratio of the lead
resistances the same as X/S and hence to a first
approximation also to A/B.
3. The short is removed from the main ratio
and the link opened. The bridge then becomes
the simple Wheatstone bridge shown in figure 18.
To a first approximation
X/S=A/B=a/b=LA/LB, (16)
and the simple bridge is unbalanced only if the
resistances of the leads La and Lb are not in this
same ratio. They are made in this ratio by ad¬
justing La or Lb until the simple bridge shows a
balance.
4. The link, L, is restored, giving again the
double-bridge, and a second balance is obtained
by varying the ratios A/B and afb. This balance
is more nearly correct than the first, as the leads
have been approximately adjusted to their proper
ratio.
5. The entire procedure is repeated until no In cas'e a/b is obtained from a slide wire, or a
further change is required in the settings of the ratio that is adjustable but not known, the same
procedure is followed as that outlined above. In
main ratio, under which condition
this case, however, the balance for step 3 is obtained
X/S=A/B=a/b, (17) by varying the ratio a/b, if necessary making a
fine adjustment of the ratio by varying the lead
X — SA/B—Sa/b. (18) resistances La and Lb.
129/19
ten steps each of 0.01 and 0.001 ohm respectively. between studs 1 and 2 there is a resistance of
When at their center positions, the resistance of 3.145 ohms so that when the dial is at 2, the shunt
the three dials is 0.555 ohm, and with the fixed arm totals 139.811 ohms and the parallel combina¬
coil of 99.445 ohms the total of the arm is 100 tion is 24.7 ohms. The successive steps of the
ohms. Starting from these center positions, a dial are not of equal magnitude, but the parallel
step on Dx will either raise or lower the M-arm by resistance may- be changed by 0.1-ohm steps
0.1 ohm. The ratio of A to B is then changed from 24.5 to 25.5 ohms. For this the steps on
from 100/100 to 100.1/100 or to 99.9/100, that the dial average about 3.6 ohms each.
is to say, it is raised or lowered by 0.1 percent. Since a change in the high-resistance arm of
Likewise, steps on I)2 and D3 change the ratio about 3.6 ohms is required to change the parallel
by steps of 0.01 and 0.001 percent respectively. resistance by 0.1 ohm, it is obvious that switch
By interpolation of the steps on the lowest dial, contact variations in the high resistance arm will
changes of 0.0001 percent may be determined. have their effect reduced by a factor of about 36 to
Although the ratio set described above is 1. Their effect could be still further reduced by
correct in theory, variations in the resistances increasing the resistance of the shunt, the effect
of the contacts of the dial switches would make being reduced as the square of the ratio of the
the readings uncertain b}r several steps on the current in the high resistance branch to the total
lowest dial when ordinary dial switches are current.
used. The use of mercury switches will reduce In the same way, it is possible to make a decade
the variations to a few microhms, but such switches for changing in 0.01-olim or 0.001-olim steps by
are somewhat difficult to operate and keep in means of a 150-ohm shunt in parallel with a
condition. 30-ohm resistor, the 150-ohm shunt to be changed
To avoid difficulties from variations in switch- in steps that average about 0.36 or 0.036 ohm re¬
contact resistances the design is usually modified spectively. Three decades with appropriate shunts
so as to reduce their effect. This is done by plac¬ and a 25-olim series resistor could be used to ob¬
ing the switches in high-resistance shunt circuits tain the equivalent of the A-arm of figure 19, but
that require comparatively large changes in resist¬ with the ratio not appreciably affected by normal
ance to obtain small changes in the parallel re¬ variations in switch contact variations. An
sistance, as was done in the rheostat of the Mueller equivalent instrument is available commercially
bridge (see section III, 5). The shunt circuits and, together with a group of standard resistors,
may be made sufficiently high that switch contact is one of the most useful pieces of apparatus for
variations are negligible even when switches of the measurement of electrical resistance, especially
moderate quality are used. An example of a for the calibration of resistance apparatus. Some
shunted dial for obtaining steps of 0.1 ohm is of the procedures will now be described.
shown in figure 20. With the dial set at 0 the Comparison of two-terminal resistance standards,
resistance of the shunt arm totals 133.636 ohms, substitution methods.—In nearly all measurements
and this in parallel with the 30-ohm branch gives where the highest possible accuracy is desired, a
a total resistance of 24.5 ohms. When the dial is substitution method is used. That is, the change
moved to stud 1, the resistance of the parallel required to restore balance after replacing a stand¬
combination increases by 0.1 ohm to 24.6 ohms, ard with an unknown is measured. In comparing
which requires that the shunt arm be 136.666 two-terminal standard resistors by substitution,
ohms. That is to say, between studs 0 and 1 the direct-reading ratio set is very rapid and con¬
there is a resistance of 3.030 ohms. Likewise venient, and accurate results may be obtained.
For this comparison a Wheatstone bridge is set¬
up as shown in figure 21. A and B are the two
arms of the ratio set, and Y is an auxiliary resis¬
tance of the same nominal value as the standard
resistors under comparison. The two resistors are
in turn placed in the mercury cups Q, and the
bridge balanced by varying the ratio A\B. The
difference in the ratio for the two balances gives
the percentage difference between the two stand¬
ard resistors. Thus, if the difference in the ratio
for the two balances is one step on the 0.001 dial,
one resistor is 0.001 percent higher than the other.
If we are comparing 1,000-olim coils, for example,
the difference is 0.001 percent of 1,000 ohms, i. c.,
0.01 ohm. Which coil is the larger is determined
by observing whether the ratio is increased or
decreased when the standard resistor is replaced
by the unknown resistance. Actually the dif-
130/20
change in temperature of 1 deg C should not
change the ratio more than 2 or 3 parts in a million.
In spite of these sources of error, it is probably
possible to compare standard resistors ranging
from about 10 ohms to 1,000 ohms to within
1 or 2 parts in a million by the use of this ratio
set. This, of course, requires that the set be well
constructed, that a fairly sensitive galvanometer
be used and that the resistances differ by not more
than about a tenth of a percent. W' ith resistances
less than 10 ohms, the resistance of the mercury
cup contacts may cause trouble. For this reason
standard resistors smaller than 10 ohms are now
almost always four-terminal resistors. Methods
of comparing such resistors will be discussed later.
Another method for comparing nominally equal
two-terminal standard resistors, which is practi¬
cally equivalent to the preceding substitution
method, is what might be called a double substi¬
tution method. In this latter arrangement, the
two resistors under comparison are used to form
two arms of a Wheatstone bridge, the ratio set
forming the other two arms. The resistors are
ference is determined in percentage of the arm Y, both mounted in mercury cups so that they may
but for small differences any departure of Y from be interchanged without affecting lead resistances
its nominal value is of no consequence. in series with them. After a balance is obtained
Although the dials of a direct-reading ratio set by adjusting the ratio set, the resistors are inter¬
change the ratios in decimal parts of a percent, it changed, and a new balance is obtained. The
is desirable to mark the dials in parts per million. percentage difference between the two resistors is
Thus the steps on the 0.001-percent dial would be half the difference between the two readings of
marked, 0, 10, 20, etc., and the 0.01-percent dial the ratio set. This method is used for the same
would be 0, 100, 200, etc. Readings of the instru¬ range of resistances as those measured by the
ment and data taken from it are then whole num¬ simple substitution method, and about the same
bers of moderate sizes rather than decimals. In* accuracy may be attained.
the discussion that follows, it is assumed that the Comparison of four-terminal resistances with two-
ratio set is marked in parts per million, and that terminal resistance.—The direct-reading ratio set is
corrections to resistors are also expressed in parts convenient for the comparison of two-terminal with
per million. Bridge readings and corrections to four-terminal resistances. Although the occasion
standards are then readily added and subtracted seldom arises for the comparison of two-terminal
and calculations are greatly simplified. with four-terminal standard resistors, it is often
In comparing nearly equal resistances by sub¬ necessary to compare a standard resistor of one
stitution, the lack of correct adjustment of the type with a resistance coil of the other type. Thus,
coils of a well-made ratio set will cause errors of in measuring the coils of many pieces of electrical
much less than 1 part in a million. The variation apparatus it is impossible to make connection with
in the contact resistance of the dial switches may the coils except through comparatively large con¬
be appreciable unless they are kept clean. The necting resistances. However, it is generally pos¬
switches should be frequently cleaned and lubri¬
sible to make potential connections to the two
cated with a little good-quality light oil. When
ends of the coils and measure them as four-terminal
kept in good condition, the resistance of the three
conductors. In doing so, it is often convenient to
switch contacts should not vary as much as a
compare them with two-terminal standard re¬
thousandth of an ohm. From the values of the
sistors.
coils on the 0.001 percent dial, we see that a vari¬
In figure 22, is a four-terminal resisfor to he
ation of a thousandth of an ohm will cause vari¬
measured, having the current terminals Tx and T->,
ations in the ratio of 1/3.6 of a part in a million.
and potential terminals Pi and P>. The potential
In comparing resistances by the substitution
leads may contain considerable resistance in addi¬
method, the two balances of the ratio set should
tion to that of the leads. A and B are the two
be made quickly so that temperature changes will arms of the direct-reading ratio set, and .1/, and
not cause variations in the ratio set between read¬ M2 are mercury cups into which either a two-
ings. It is best to have all coils of the ratio set terminal standard resistor or a short circuiting
made from the same lot of resistance material. link may be placed.
If this is done and manganin coils are used, a Suppose we start with the galvanometer con-
131/21
nected at Pi, a standard resistor nominally equal
to X inserted in Mx, and with M2 shorted. We
then have a simple Wheatstone bridge, which is
balanced by varying the setting of the ratio set.
After this balance is obtained,'the galvanometer
connection is shifted to P2, and the standard re¬
sistor is placed in M2, Mx being shorted with the
link. We again have a Wheatstone bridge but
with the standard resistor and unknown inter¬
changed. This interchange has been obtained
without making any change in the resistance of the
leads of the measuring circuit, except for possible
variations in the resistances of the mercury cup
contacts, which will be small if the mercury con¬
tacts are clean. The bridge is now again balanced
by means of the ratio set. The percentage differ¬
ence between the unknown and the standard re¬
sistor is half the difference in reading of the ratio
set for the two balances. Unless the resistances Figure 22. Measurement of 4-terminal resistor.
under comparison are fairly large, it will he neces¬
sary to take into account the resistance of the
short-circuiting link. This is done by subtracting
the link resistance from the resistance of the
standard resistor and considering that the resistor
has this new value and is being interchanged with
a link of zero resistance.
The resistance of the link can be measured as
follows: Connect the link between two 1-ohm
resistors to form two arms of a Wheatstone bridge,
using the direct-reading ratio set for the other
two arms, as shown in figure 23. L is the link,
and A and B are the arms of the ratio set. Two
balance readings are taken, first with the gal¬
vanometer connected at one end of L and then
at the other. Half the difference in the readings
is the value of the link resistance in percentage
of the 1-ohm arms. If the link resistance is large,
it may be necessary to use larger resistances in
place of the 1-ohm coils. This method is very
convenient for the measurement of small resist¬
ances such as links, connecting wires, switch
contact resistances, etc. It is not a precision
method but usually is sufficiently accurate for the
measurement of resistances such as those just
mentioned, which are to be used in series with
larger resistances.
Substitution method for decades.—In the calibra¬
tion of precision rheostats the occasion often
arises for the measurement of a series of resistors
of the same nominal value. This is readily done
by the substitution method, using a standard
resistor of the same value as the steps of the
rheostat, reading differences on a direct-reading
ratio set. The procedure is illustrated in figure
24. In this figure PR is the precision rheostat
to be calibrated, and let us assume that the 10-
ohm-per-step dial is to be checked. PB is then
a plug box or any decade with 10-ohm steps, and
M is a pair of mercury cups in which is placed a
standard 10-ohm resistor. The arm ]’ is a 100-
ohm resistor whose value need not be accurately
132/22
known, and A and B are the two arms of a direct
reading ratio set. The reading of the rheostat
under calibration is set at zero, the plug box is
set at 90 and the 10-ohm resistor is placed in
M, making the total nominal resistance of the
arm 100 ohms.
After the bridge has been balanced by changing
the setting of A, the standard resistor is removed
and the mercury cups shorted with an amal¬
gamated copper link, and also the reading of
PR is changed from 0 to 10 ohms. This has in
effect substituted a 10-olim step of PR for the
10-ohm standard resistor. The change in the
reading of A required to again obtain a balance
of the bridge is a measure of the difference between
the step on the rheostat and the standard resistor,
such difference being a percentage of 100 ohms.
The second 10-ohm step on the rheostat is obtain¬
Figure 26. Lead resistances in Wheatstone bridge.
ed by leaving Pi? at its 10-olim reading, again
placing the standard resistor in M and reducing
the PB resistance to 80 ohms. These three case when the individual resistors are measured
resistors still total 100 ohms, nominally. The directly. This may be seen by reference to figure
standard is now again replaced by a short-circuit¬ 25, which represents the connections to one dial
ing link, and.Pi? is set to read 20 ohms, the result¬ of the rheostat. The switch contact may be set
ing change being read from A. The change was on the contact studs marked 0, 1,2, 3, etc., thus
produced by the substitution of the second 10-olim connecting 0, 1, 2 or more resistors between the
step of the precision rheostat for the 10-ohm dial terminals S and T. It should be noted that
standard. the resistors are usually connected in series and
The procedure is continued, the steps of PR lead wires are connected from the junctions of the
being successively substituted for the standard resistors to the switch contact studs. If the resis¬
resistor. It should be noted that the steps of tors are calibrated by measuring between studs 0
the auxiliary plug box, PB, are also being replaced and 1 ,1 and 2, 2 and 3, etc., two of these leads are in¬
by the standard resistor as PR is being increased cluded with each resistor. In actual use all the
and PB decreased in reading, the standard being leads to the coils in use are not in the circuit, but
cut in and out of the circuit. Data are therefore only one lead is used for any setting of the dial.
obtained for calibration of both PR and PB in The above method of calibration determines the
terms of the standard resistor. Hence by this step as in actual use, the only additional data
method, two precision rheostats may be calibrated required being the resistance between terminals
simultaneously. with the dial set at zero. This is readily measured
When 10-olim steps are calibrated in a 100-olim by the method just described for the determination
arm, the accuracy is reduced by one order. That of link resistances. For any setting, this “zero
is to say, differences in readings of the ratio set resistance” must be added to the sum of the steps
must be obtained to 0.0001 percent if the 10-olim as determined by substitutions.
steps are to be determined to 0.001 percent. The Comparison of four-terminal resistance stand¬
method has the advantage that changes in the ards.—Before taking up the question of the
over-all resistance of the rheostat are determined comparison of four-terminal standard resistors,
under the conditions of use, which is often not the let us consider briefly the effect of the lead wires
of a simple Wheatstone bridge. In figure 26, Ar
and Y are nominally equal resistances, and A and
B are the two arms of a direct-reading ratio set.
The conductors x, y, x', and y' are used to connect
up the bridge, and we will also denote their resist¬
ances by x, y, x', and y', respectively. The ratio
of the resistances of the two arms containing A'
and Y is not in general the same as the ratio of
A" to Y because of the resistance of these con¬
necting leads. If we could select leads such that
the ratios xjy and x'ly' were the same as X/Y, the
ratio of the two arms would be independent of the
actual values of the lead resistances. That is, the
balance would be the same as if the resistances of
133/23
the leads were negligibly small. It is possible to bridge balanced is 5497 millionths, and with the
make the leads adjustable and make tlie ratios galvanometer connection changed to b, the reading
x/y and x'\y' the same as X/Y. In fact, sueli an is 5613. Let us further assume that when we
arrangement is used at the National Bureau of change the battery leads to a' b', leaving the
Standards to reduce the effect of the leads when galvanometer connected at b, the balance reading
using the Kelvin double bridge. Instead of adjust¬ changes to 5835.. Then with leads of the proper
ing the leads, it is possible to balance the bridge ratio, or leads with negligibly small resistances,
with fixed leads, and then find what this balance the balance would have been at (5,613 +5.497)/2 +
would have been with the proper ratio of lead re¬ 10/11(5,835 — 5,613); i. e., at 5,757.
sistances, or with negligibly small lead resistances. As a matter of fact, this scheme for taking into
A' and Y were assumed to have practically equal account the connecting leads is practically never
resistances. Instead of connecting the galvanom¬ used in comparing two-terminal resistors. As the
eter as shown, suppose we balance the bridge with substitution method requires no consideration of
the upper galvanometer connection first at a and the lead resistances, except to see that they are
then at b. The average of the two readings is reasonably small, it is generally used. However,
the value that would have been obtained had the in comparing four-terminal resistances with a
conductors x and y been equal or negligibly small simple Wheatstone bridge we follow exactly the
in resistance. Thus, by taking two readings we steps outlined above. Figure 27 shows a bridge
can take into account the effect of these two set up for this purpose. A and B are the two
connecting resistances. This average balance arms of the direct-reading ratio set. A" and Y are
reading is not the correct reading, however, unless the four-terminal resistors under comparison,
also x. =y', which is probably not true. We must with current terminals Tu To, -TV, and TV, and
now find what this balance would have been with potential terminals Pu P2, P/, and P2'. The
x' = y'. To do this suppose we shift the current bridge is balanced when connected as shown, and
connections from the points shown to the points a second balance is obtained after shifting the
a' and b'. In doing so we remove the lead resis¬ galvanometer connection to P2'. The third bal¬
tances x' and y' from the X and Y arms and ance is obtained after now changing the current
connect them in series with the ratio arms A and B. connections from Tj and T/ to Pi and P/, and
If x' and y' are not equal, we will change the calculations are made as above. The two arms
balance of the bridge by removing them from the X and Y are now interchanged and the three
arms Ar and Y. We will have still further changed readings again obtained. From these two sets of
this balance by adding them, to the ratio arms A readings, we get two balance points on the ratio
and B. Which of these changes is the greater set. Half the difference between these two balance
depends upon the relative sizes of the two pairs readings is the percentage difference between the
of arms. If A and B are equal to X and Y, the two four-terminal resistors.
changes will be equal. That is, with all arms Although the direct-reading ratio set was
nominally equal the balance would be changed a developed for use in the comparison of nominally
certain amount if the battery connections were equal resistances, it can be readily adapted for the
shifted to a' and b'. This change is twice that comparison of resistances of any ratio provided
which would have been obtained had we only some independent means is available for accurately
removed x' and y' from the X and Y arms. The
average of the readings before and after changing
the battery connections is then the reading that
would have been obtained with x' — y' or both
negligibly small.
Suppose, however, that A and B are ten times
as large as X and I". Then connecting x' and y'
in series with A and B produces only a tenth as
large a change as is caused by their removal from
X and Y. Then ten-elevenths of the change in
balance when the battery connections are shifted
is due to the removal of x' and y' from the X and I"
arms, and the remaining one-eleventh is due to
the connection of the leads in the ratio arms.
This enables us to calculate what the balance
would be with x' and y' equal or negligibly small.
As an example, suppose X and Y are each 10
ohms, and the arms of the direct-reading ratio set
are 100 ohms. Assume that with the battery
connected as shown in figure 26 and the galva¬
nometer at a, the reading of the ratio set with the Figure 27. Comparison of J+-terminal resistors.
134/24
realizing the same resistance ratio. As an example
of such a use let us consider the calibration of a
25-ohm resistor by comparison with a 100-ohm
standard.
As has been described above, the direct-reading
ratio set consists of the arm A, which is adjustable
in small steps from values slightly below to slightly
above 100 ohms. The 5-arm is ordinarily a fixed
100-ohm arm. Suppose we change the 5-arm
by connecting an additional 300 ohms in series.
The ratio A/B then becomes 100:400, which is
adjustable in the same percentage steps as was
the 100:100 ratio. Using two sets of mercury
cups, let us set up the bridge shown in figure 28.
The 25-ohm resistor is placed at X and the 100-
ohm standard at S, and the bridge is balanced
by adjusting A, and let us call this reading A.
Also let A0 be the reading A would have had if the
X and S arms had been exactly in the ratio 25:100.
If cx and cs designate the proportional corrections
of X and S, the actual balance A, of the bridge,
will be
A—A0A-Cx—cs- (19)
resistors do. not need to be equal to a very high
Since A is obtained experimentally and cs is precision. However, they must not differ by more
known, this equation could be solved for cx if A0 than 0.1 percent if the average of the five ratios is
were also known. To obtain A0, the arms X and to be correct to one part in a million.
S must be replaced by a resistance ratio of exactly A procedure analogous to this method of obtain¬
1:4. ing a ratio of 1:4 may be followed to determine any
To realize a 1:4 ratio it is necessary merely to ratio 1 :n, where n is an integer, by taking the
have five resistors that are reasonably nearly equal. average of n+1 ratios. It may also be used to
If one of these five is connected in place of X and determine any ratio r:n, where both r and n are
the other four in series in place of S, the ratio will integers. Starting with r+n equal resistors, r in
be only approximately 1:4. However, if the five one arm and n in the other, the resistors are
resistors are placed one after the other in X, the rotated in a cyclic order until each resistor occupies
remaining four each time being connected in .S, each position once. The average of the r-\-n
the average of the five ratios will be 1:4 to a very readings is the reading that would be obtained if
high accuracy. In other words, the average of the all resistors were equal. Moreover, the resistors
five readings of A with the five resistors in turn in either arm may be in parallel rather than in
at X, the remaining four in series in the 5-arm, series.
will be Aq, the reading of A for an exact 1:4 ratio. In the calibration of high resistances, use may
This value of A0 can be substituted in eq 19 to be made of the fact that the proportional correc¬
allow the calculation of cx, the proportional cor¬ tion to a group of nominally equal resistors is the
rection to the unknown 25-ohm resistor. same when they are connected in parallel as when
Another method of obtaining A0 would be to connected in series. For example the ten 100,000-
balance the ratio set with one resistor at A" and ohm sections of a megohm box may be connected
the other four at S, thus obtaining A of eq 19. in parallel and measured against a 10,000-olun
The five resistors could then be substituted in any standard resistor. If the parallel group is high
bridge and their differences determined. The in resistance by 0.01 percent, the series resistance
terms cx and cs are then the amounts in propor¬ of 1 megohm will also be high by 0.01 percent.
tional parts that the resistor in X and those in S Here it is assumed that the ten sections are suffi¬
differ respectively from the average value of the ciently near to equality that in the expansion
five.
To get an accurate value of the 1:4 ratio, the
five resistors should have large enough resistances yy— = 1 — c+c2— c3, etc. (20)
1 Ac
that the lead resistances of the X and S arms are
negligible, or the lead resistances should be in the
ratio 1:4. They may usually be so set with a the second and higher powers of c are negligible,
sufficient accuracy by shorting Xand S and adjust¬ where c is the proportional amount by which the
ing the bridge circuit so obtained to a balance by resistance of any section differs from the average
changing the length of the lead wires. The five of all. If this condition is satisfied, the ratio of
135/25
the resistance of n resistors in parallel to their
resistance in series is exactly 1 :n2.
The methods that have just been presented
assume the use of a direct-reading ratio set. These
methods, however, are entirely satisfactory when
use is made of any ratio set that has small and
definite steps. Such a set may be assembled from
ordinary laboratory apparatus. For example, the
adjustable 100-ohm arm might be made of a
105-ohm coil with a parallel decade box reading
about 2,500 ohms. A change in reading of the
decade box by 0.1 ohm would change the 100-ohm
arm by about 1 ppm. The changes in the parallel
resistance are not directly proportional to the
change in the high resistance arm, but they may
be readily calculated. This adjustable 100-ohm
arm, together with a fixed 100-ohm resistor,
constitutes an adjustable ratio set.
iflt2/i?3,4 where B12 and i?3,4 are the resistances of
2. Universal Ratio Set the slide wire between the points 1 and 2, and
3 and 4, respectively.
Precision standard resistors are usually made It is very difficult to make a wire of sufficient
only in integral multiples or submultiples of an
length and uniformity that the ratio i?i,2/i?3,4 can
ohm. Consequently odd-valued resistors usually be accurately determined. Instead of a slide
cannot be measured by a substitution method, wire, the universal ratio set makes use of a group
except in a few cases where standard resistors can of wire wound resistors so that the resistances
be combined to give a resistance nearly that of the
between the balance points 1, 2, 3, and 4 can be
unknown. The comparison of odd-valued resis¬
accurately known. The arrangement of its dials
tors with standards is then not possible with a
is as follows. The highest dial consists of twenty
direct-reading ratio set, but a ratio set is required
100-ohm resistors in series, with the dial contact
that is accurately adjustable over a wide range,
acting as the potential connection to the instru¬
at least from a 1:1 to a 5:1, or preferably to a
ment, as shown in figure 30. As the dial is
10:1 ratio.
rotated in a clockwise direction, the 100-ohm
A very convenient wide-range ratio set is one
resistors are successively changed from the right
used at the National Bureau of Standards and
to the left side of the contact. To change resist¬
called a “universal ratio set” [ 14]. This instru¬
ance from the right to left side in 10-ohm steps,
ment is one having a constant resistance, between
two more dials are used, each having ten 10-ohm
two external terminals, of about 2,111 ohms. An
steps as seen in figure 31. These two dials are
arrangement of dials is such that in effect a poten¬
operated by the same handle but with one dial
tial connection may be made at any point of the
increasing as the other decreases in resistance.
2,111 ohms to the nearest 0.01 ohm. The ratio
The total resistance between S and T remains
of the resistance between the potential point and
constant for any setting of the 10-olim decades,
one terminal of the set to the resistance between
the potential point and the other terminal is
therefore adjustable in small steps over a very
large range. The device is the equivalent of a
long slide wire with a movable contact, and its
uses are analagous.
Suppose it were desired to measure a resistance
of say 6.8 ohms by comparing it with a 10-ohm
standard resistor. The two resistors could be
connected in series and the combination connected
across a slide wire as shown in figure 29, X being
the unknown, S the standard, and IF a slide wire.
If a galvanometer, G, is connected to terminal a
of the unknown and to the slide wire, a Wheat¬
stone bridge is obtained that will balance with
the slide wire at some point, 1, near the end. If
now the galvanometer connection is changed suc¬
cessively to b, c, and d, successive balances will be
obtained with the slide wire at 2, 3, and 4, respec¬
tively. The ratio X/S is then the same as
136/26
IOOASTEPS
and of course it is unaffected by the position of some resistance is common to both the standard¬
the 100-ohm dial switch that is merely a potential cell and main-dial resistance. The arrangement
contact. for such a test is shown in figure 32. The ratio
Steps of 1, 0.1, and 0.01 ohm-per-step are set, URS, which is shown as a slide wire, is con¬
obtained in the same way as the 10-ohm steps, nected in parallel with the potentiometer, the
that is with two decades of each denomination connection-to the latter being made to the battery
operating together, one on each side of the 100-ohm binding posts, BA-\- and BA—. Readings on
dial. This scheme for the change of the potential URS are made with the galvanometer connected
point along a fixed resistance is essentially the successively to the SC and emf binding posts, for
same as that of the Feussner type potentiometer. all settings of the emf dials. It should be noted
The large number of contacts in series limits its that changes in the emf dials are merely changes
use for precision work to ratio sets of fairly in potential points and do not affect the readings
high resistance. obtained for the SC dial.
The universal ratio set is used only for the Potentiometers are provided with rheostats in
determination of resistance ratios and hence may the battery circuit for adjusting the potentiometer
be calibrated in terms of any unit. It is most current. A change in this battery rheostat will
conveniently calibrated in terms of an average change the differences in readings obtained on the
step on the highest or lowest dial. This is done universal ratio set but not the ratio of the differ¬
by comparing each step of a decade with the ten ences. It is possible to make the ratio set direct-
steps of the preceding dial starting with the reading by adjusting the battery rheostat until the
smallest dials. Only the steps on one side of the difference on the ratio set for the standard-cell
100-ohm dial need be tested, as the function of posts is a decimal multiple, preferably 1,000, of
the other group is to keep the total resistance the reading of the standard-cell dial. Corrections
constant. A check of the constancy of the over-all for the emf dials may then be read directly from
resistance needs to be made for all readings of the differences across the emf terminals.
double dials.
The use of a universal ratio set is the same as
that of a slide wire as described in conjunction
with figure 29, for the measurement of odd-sized
resistances. It may be used in the measurement
of four-terminal resistors and is especially con¬
venient for the tests of potentiometers. The test
of a potentiometer consists in the measurement of
the ratio of the emf-dial resistance to the stand¬
ard-cell resistance for all settings of the emf and
standard-cell dials. These resistors are of the
four-terminal type with potential connections
brought out to emf and standard-cell binding
posts. The method is satisfactory even when Figure 32. Check of potentiometer with universal ratio set.
138/28
post to the other. That is to say, the current and order to read to 0.001° C, it is necessary to read
potential connections at point 2 may be inter¬ these thermometers respectively to 0.00001 or
changed. This will place the variable galvanometer 0.0001 ohm. For such measurements, bridges of
switch resistance in series with the galvanometer special design are usually used.
where it will not affect the bridge balance. In this country thermometer bridges for preci¬
The rheostat arm, R, may be calibrated like sion work are usually made with equal ratio arms.
any other precision rheostat as described in These arms are interchangeable, so that the use
section IV, 1. Connection should be made of an average value eliminates errors from lack of
through the appropriate X and GA terminals. equality of the ratio arms, or the ratio arms may
The resistance of the rheostat arm with all be interchangeable and adjustable so that they
dials set at zero may be measured by connecting may be made equal at any time. With such
the arm between two equal resistors, using a interchangeable ratio arms, actual calibrations of
direct-reading ratio set to connect up a Wheat¬ them are unnecessary. The calibration of a ther¬
stone bridge. The change in the reading of the mometer bridge requires only a calibration of the
ratio set is determined when the galvanometer rheostat arm.
connection is changed from one branch point of In order to measure platinum resistance ther¬
the rheostat arm to the other. This is the same mometers, which as stated above are usually not
as the method given in section IV, 1 for the very high in resistance, with a bridge having equal
measurement of lead resistances. In a similar ratio arms, it is necessary to have the rheostat arm
way, the lead resistances in the X arm may be adjustable in small steps. In fact, steps as low as
determined, taking readings with the galvanom¬ 0.0001 or 0.00001 ohm are needed for work of the
eter connection to branch points 1 and 2 and to highest precision. In order to obtain such small
the external X binding posts, with the latter steps, contact resistances cannot be used directly
connected by a shorting wire. The calculation in the rheostat arm. Hence recourse is had to
must be made in two parts in order to exclude the decades of the Waidner-Wolff type described above
resistance of the shorting wire. in section III, 5, in which the changes in resistance
For calculating X from the calibration data, it result from changes in the values of high-resistance
is convenient to express the corrections to the shunts on comparatively small resistances.
readings of the rheostat arm in ohms and to ex¬ Switches are placed in the high-resistance shunts,
press the corrections to the ratio arms in pro¬ where variations in their resistance will have a
portional parts. The calculation in proportional negligible effect. Decades of the Waidner-Wolff
parts involves merely a division of the correction type cannot be set to have zero resistance, but
in ohms by the nominal value in ohms. The the small changes start from an appreciable mini¬
value of X for a given balance is calculated from mum value. However, for equal-arm bridges,
the equation compensating resistance may be placed in the
A-arm. The value of the unknown is then meas¬
W=-g(l-(-a—b)(R-f-r-j-r0)—X0- (21) ured by the increase in reading of the rheostat arm
when the unknown is connected into the circuit.
The use of a resistance thermometer to measure
In this equation a and b are the proportional cor¬ temperature involves merely the determination of
rections to the ratio arms A and B respectively, resistance ratios. Hence in calibrating thermom¬
7?+r is the sum of the dial readings and correc¬ eter bridges it is necessary to determine only rela¬
tions, and r0 is the resistance of the rheostat arm tive values of resistance, which does not require
with all dials at zero. The term X0 is the resist¬ the use of standard resistors. This calibration is
ance in ohms of the lead wires in the A-arm of readily made by the user, especially for bridges
the bridge. The factor (l+a— b) is an approxi¬ having equal ratio arms.
mation for (l+a)/(l + &) and is accurate if a and b The calibration of such thermometer bridges
are small as compared with unity. If neither a requires as auxiliary equipment only an adjustable
nor b exceeds 0.001, the error from neglecting the resistor that has the same range as that of the
second-order terms does not exceed two parts in bridge rheostat. This resistor needs to be accu¬
a million. rately adjustable, although the resistance need
not be known for any position. Such an adjust¬
2. Calibration of Thermometer Bridges able resistor may be assembled, for example, from
a decade box with minimum steps of 0.1 ohm in
Because of the space limitations and structural series with an 0.1- or 1-ohm resistor, which is in
difficulties, platinum resistance thermometers usu¬ turn shunted by a slide wire or a rheostat of fairly
ally have resistances of less than 100 ohms. The high resistance. This adjustable resistor is con¬
common values are about 2.5 or 25 ohms, the nected across the A-terminals of the bridge and is
actual values being so chosen that the change in used to balance the bridge after the rheostat arm
resistance is very close to 0.01 or 0.1 ohm per is set to certain required readings. The shunted
degree centigrade change in temperature. In 0.1- or 1-ohm should be attached to the A-termi-
139/29
nals so as to be adjacent to the rheostat arm. If setting, and balance the bridge by adjusting X.
a slide wire is used for the shunt, the sliding con¬ The galvanometer deflection is now read when the
nection should be used as the galvanometer branch lowest dial reading is changed from 0 to 1. Leav¬
point so that its variable resistance will be in the ing this dial on 1, the bridge is again balanced
galvanometer branch where it will not affect the with the X-rheostat, and the galvanometer deflec¬
balance. This, in effect, throws the shunted re¬ tion then again read for a shift of the lowest rheo¬
sistance partly in the X and partly in the rheostat stat dial from 1 to 2. These alternate balances
arm. We then have the equivalent of the bridge with rh and X are continued until the value of
shown in figure 35, the ratio arms A and B being all the steps on the smallest dial of rh are deter¬
mined in terms of galvanometer deflections. Un¬
less one of the steps is defective, these should be
the same to the nearest 0.1 step.
The steps on the second dial of rh should each
equal the 10 steps of the lowest dial. They are
measured in terms of the lowest dial by setting
the lowest dial at 10 and the second dial at 0, the
other dials being at any convenient value, usually
small in order to obtain good sensitivity. With
the lowest dial at 10 and the next larger dial at 0,
the bridge is balanced by adjusting the value of X,
after which the smallest dial is turned to 0 and
the other dial shifted from 0 to 1. If there is now
any galvanometer deflection, it is because the first
step on the second dial is not the same as 10 steps
on the smallest. The amount that they differ, in
terms of steps of the lowest dial, is determined by
reading the galvanometer deflection and evaluat¬
ing this deflection by reading the additional change
in deflection resulting from a change of the smallest
dial setting from 0 to 1.
To measure the second step on the second dial
its reading is left at 1 and the lowest dial now set
at 10, a balance being obtained with X. The
lowest dial is now set back to 0 and the other
increased to 2, thus substituting the second step
of the higher dial for 10 steps on the lower. The
lack of balance is again translated into fractions
of a step on the lowest dial by interpolation, using
galvanometer deflections. This procedure is con¬
Figure 35. Calibration of rheostat arm of thermometer tinued, each step of the second dial being compared
bridge. in turn with the 10 steps of the lowest dial. A
table is now made showing the value, in terms of
accurately adjusted to equality. For a balance steps of the lowest dial, of the first step of the
the rheostat arm, rh, and the X-arm must be also second dial, the sum of the first two steps, sum of
equal. If then X is changed by any amount, as the first three steps, etc. This table will give the
by the insertion of a resistance thermometer in resistance corresponding to any reading of this dial
series, the rheostat would have to be changed by in terms of steps on the smallest dial, which we
an equal amount to again balance the bridge. might call “bridge units”.
Thus, actual values of the rheostat arm need not The values of the steps on the next higher dial
be known if changes in the arm are accurately are now determined in the same way but in terms
determined. of the 10 steps on the second dial, the unit again
The procedure for checking the rheostat arm is being a step on the smallest dial. The continua¬
usually as follows. The resistances of all steps in tion of this process gives the resistance of each step
the arm are determined in terms of an average of each dial in terms of the 10 steps on the preced¬
step on the smallest dial. The steps of the small¬ ing dial, from which finally is calculated the re¬
est dial are first intercompared to see that they sistance for each setting of each dial in terms of
are equal within 0.1 step, which is as accurately bridge units.
as the steps may be readily read by interpolation The resistances as determined may be used
from galvanometer deflections. An easy way to with any resistance thermometer without con¬
intercompare the steps on the lowest dial is to set verting their values to ohms. It is sufficient to
it at 0, the other dials being at any convenient standardize the resistance thermometer on this
140/30
same bridge, by measuring its resistance at known “zero” resistance of the rheostat need not be
temperatures. However, it is often desirable to determined, if two bridge balances are obtained,
convert the values to ohms in order that the bridge the first with the unknown connected to the X-
may be used for the measurement of resistance terminals and the second with the unknown re¬
other than that of thermometers. This is done by sistance short-circuited.
measuring the resistance of a standard by balanc¬ Many precision thermometer bridges provide a
ing the bridge with the standard resistor connected shorting plug for short-circuiting the resistor con¬
to the AT-terminals. All readings of the rheostat nected to the W-terminals. The galvanometer
should then be multiplied by the ratio of the connects to the center of the shorting connector
resistance of the standard in ohms (rather, the so that equal amounts of its resistance are inserted
difference between its resistance and that of the in the X and rheostat arms thus giving the same
shorting connector) to its resistance in bridge units
balance as if the shorting connector had a negligible
in order to convert the rheostat calibration to ohms.
The above procedure may be used for the cali¬ resistance. With this method it is unnecessary to
bration of the rheostat arm of any Wheatstone know the resistance of the shorting connector,
bridge, in terms of steps of the lowest decade and although such resistance may usually be esti¬
then in ohms, by comparison with a standard mated with sufficient accuracy from its length and
resistor. When used with equal ratio arms, the gauge size.
141/31
density of the conductor is determined by the liquid of known resistivity. In this case measure¬
equation ments of dimensions may be avoided, as the known
and unknown are given the same dimensions by
placing them in turn in the same container or
“conductivity cell.” The ratio of their resistivi¬
ties is then the same as the ratio of their resis¬
where wa and ww are the weights of the specimens tances. Mercury is often used as the liquid of
in air and water, respectively, and Dw is the known resistivity, since its value is accurately
density of the water. At a temperature of 21° C, known.
the density of water is 0.998 gm/cm3 and this value Although the resistivity of mercury is known
decreases uniformly to 0.997 gm/cm3 at about to a high accuracy, its use as a liquid of known
25° C. resistivity may lead to errors of considerable
Instead of calculating A separately, the value of magnitude. In comparing the resistivity of
A from eq 26 may be substituted into eq 23, giving liquids by placing them in turn in the same cell
and measuring their resistance, it is tacitly as¬
sumed that the current distribution through the
P=jXfx/,- (28)
cell is the same for both liquids. This may be
incorrect unless the liquids have very nearly the
From this equation it is seen that the resistivity, p, same resistivity, as the distribution of current in
equals the product of resistance per unit length the cell depends to some extent upon the resistance
and mass per unit length, divided by density. of the metal electrodes. For high-resistivity
The length need not be the same for the resistance liquids the resistance of the electrodes may play
and mass measurements if the material is uniform. a negligible part in determining the current dis¬
For two conductors having the same density, the tribution, but in the case of mercury, which is a
ratio of their resistivities is the same as the ratio relatively good conductor, the current distribu¬
of their values for the product E/lXm/l. This tion may depend to a considerable degree upon the
product is called “mass resistivity”, and is a dimensions of the electrodes and upon the resis¬
constant that is characteristic of the material of tivity of the metal of which they are made.
a conductor, being D times as large as the or¬ In determining the resistance of a liquid, it is
dinary, or volume, resistivity. The mass resis¬ necessary to use an alternating-current bridge.
tivity is often specified in the purchase of con¬ When a direct current is used, the ions in the
ductors for electrical uses, and it is sometimes more liquid will drift towards the electrodes thus making
readily measured than is volume resistivity. the density nonuniform. Moreover, polarization
However, it is doubtful that this advantage is will often be produced by the liberation of gases
efficient to compensate for the confusion that at the electrodes. When measured with alter¬
rises from the use of two types of resistivity. nating current a conductivity cell is found to be
It is usual commercial practice to specify electrically the equivalent of a resistor and a
percentage conductivity rather than resistivity, capacitor in parallel. This requires a balance
especially in the purchase of copper conductors. of reactance as well as of resistance when the cell
Conductivity is the reciprocal of resistivity, and is measured in an alternating current Wheatstone
percentage conductivity is obtained by dividing bridge.
the resistivity of the given sample into that of the VII. References
standard and multiplying by 100.
By international agreement, the resistivity of [1] N. B. Pilling, Trans. Am. Electrochem. Soc. 48, 171
annealed copper is taken as 1.7241 microhm-em (1925).
[2] R. S. Dean and C. T. Anderson, Trans. Ain. Soc.
at 20° C [15]. This value was selected from Metals 29, 899 (1941).
measurements of the resistivity of a large number [3] A. A. Somerville, Phvs. Rev. 31, 261 (1910).
of samples of high-purity commercial copper wire [4] J. L. Thomas, J. Research NBS 16, 149 (1936) RP863.
from both American and European refiners. The [5] J. L. Thomas, J. Research NBS 13, 681 (1934) RP737.
[6] F. Wenner and .T. L. Thomas, BS J. Research 12, 147
value was agreed upon as a standard for reference (1934) RP639.
and was not intended to be the value for abso¬ [7] P. H. Bridgman, Proc. Am. Acad. Arts Sci. 52, 573
lutely pure material. Copper has been produced (1917).
with a conductivity of several percent greater [8] J. L. Thomas, J. Research NBS 36, 107 (1946) RP1692.
[9] P. H. Dike, Rev. Sci. Inst. 7, 278 (1936).
than 100, which probably indicates a material [10] E. B. Rosa, BS Bui. 5, 413 (1908-09) S107.
of higher purity than that of the standard. [11] F. Kohlrausch, Wied. Ann. 20, 76 (1883).
Some effort has been made to secure interna¬ [12] E. F. Mueller, BS Bui. 13, 547 (1916) S288.
tional agreement for a standard for the resistivity [13] E. F. Mueller and F. Wenner, J. Research NBS 15,
485 (1935) RP842.
of aluminum, but so far copper is the only metal [14] F. Wenner and E. Weibel, BS Bui. 11,27 (1914) S223.
for which such a value has been adopted. [15] BS Circular C31 (1914).
The measurement of the resistivity of a liquid
may often be made by comparison with another Washington, March 30, 1948.
142/32
U. S. Department of Commerce National Bureau of Standards
abstract
I. Introduction_-_ 230
II. Review of earlier work_ 231
III. Sensitivity of bridges_ 232
1. Damping of galvanometers_232
2. Wheatstone bridge_ 233
3. Thomson bridge_ 239
IV. Load coefficients_■_ 241
V. Galvanometer used with NBS precision bridge_ 244
VI. NBS precision bridge_ 244
VII. Bridge equations_248
1. Substitution method_ 248
2. Interchange method_251
3. Known ratio method_251
VIII. Establishment of known ratios_ 252
IX. Adjustments of Thomson bridge_ 255
X. Precision and accuracy_ 256
XI. Appendices_258
Appendix 1. Terminals and contacts_ 258
(a) Standard resistors_ 258
(b) Terminal blocks of bridge_ 260
(c) Terminal blocks of auxiliary apparatus_ 261
(d) Thomson bridge_ 265
Appendix 2. Thermoelectromotive forces_ 266
(a) Thermoelectric balances_267
(b) Thermal shunts_ 267
143/229
Journal of Research of the National Bureau of Standards [Voi.ss
XI. Appendices—Continued.
Appendix 3. Insulation_ 268
(a) Theory in case of defective insulation_ 269
(b) Procedure in case of defective insulation_ 270
(c) Test of insulation_ 271
(d) Wagner branches__ 272
Appendix 4. Optical systems of galvanometers_ 274
(a) Description of optical system_ 274
(b) Quality of optical systems and sensitivity of bridges_275
Appendix 5. Methods of analyses (three expedients)_277
(a) Classical form of solution of unbalanced Wheatstone
bridge_277
(b) First expedient (hypothetical battery of zero
resistance_277
(c) Second expedient (hypothetical battery of high electro¬
motive force and high resistance)_-_278
(d) Third expedient (separation and synthesis of the
effects of the six independent resistances of a four-
terminal conductor)_279
Appendix 6. Ohm’s law_285
Appendix 7. Units of resistance_■_ 288
Appendix 8. References_ 290
(a) Authors_ 290
(b) Subjects_ 293
XII. Index_ 294
1. INTRODUCTION
144/230
H-enner] Resistance Comparisons
145/231
Journal oj Research oj the National Bureau of Standards [Voi.
146/232
Wenner] Resistance Comparisons
2. WHEATSTONE BRIDGE
147/233
Journal oj Research oj the National Bureau oj Standards [Voi. S6
Obviously then
z, E(X+ AX)
1 X+AX+Y’
(2)
EA
En > and (3)
~A+B
AX) , EA
(4)
+X+AX+Y± A+B
Since the direction of Eg taken as positive is immaterial, to avoid
complications in what follows the T sign will be taken as positive,
and the ± sign will be taken as negative. Then, since from eq 1
it follows that
A _ X
AAB x+y’ w
eq 4 takes the form
ex r i+ax/x
‘] or
E„= (6)
X+Y[l+AXJ(X+Y)'
148/234
Wenner] Resistance Comparisons
Q.-Q^DEXYdXKX+Y)2, (11)
and
Q.2-Q=-DEXYdX/(X+Y)2. (12)
Subtracting eq 12 from eq 11 eliminates Q and gives
dQ=2DEXYdX/(X+Y)2> (13)
S=dQ/dX. (14)
Then it follows from eq 13 and 14 that
S=2DEXY/(X+Y)2. (15)
Since from eq 1 it follows that
XY/(X+ Y)2=AB/(A+B)2, (16)
another expression for the sensitivity is
S=2 DEAB/(A+B)2. (17)
However, if the resistance W of the bridge between its galvanometer
terminals is greater than the resistance V, which gives the specified
damping of the galvanometer, the resistance U is placed in parallel
with the galvanometer .and adjusted so that
UW/(U+W) = V (18)
With the galvanometer branch only open, the current in the U branch
is EguKJJ+W), while the potential drop across the break in the
galvanometer branch, Eg, is U times this current, or EguUI(U-\-W).
Therefore, since Uj(U-\-W) = V/W.
That is, the effect of the resistance in parallel with the galvanometer
(in parallel with the bridge, if considered from the standpoint of the
electromotive force developed in the galvanometer coil as a result of
its motion), when of such value as to give the specified damping of
the galvanometer, is a reduction of the electromotive force in the
galvanometer circuit, and consequently of the combined sensitivity
149/235
Journal of Research of the National Bureau of Standards ivoJ.es
150/236
Wenner] Resistance Comparisons
the potential drop in each bridge arm, and here the potential drop
will be used. How these factors, especially the precision desired in
the measurement and the load coefficients, limit the permissible
sensitivity may be seen by considering the following example.
Assume that
D=10 millimeters per microvolt,
V—15 ohms,
X—10 ohms,
Y=5 ohms,
A= 50 ohms, and
B—25 ohms.
Assume that the bridge is to be balanced to 1 part in a million
and that investigation of the load coefficients has shown that a change
of 1 part in a million occurs
in X when Ex, the potential drop in X, is 0.75 volt,
in Y when Ev, the potential drop in Y, is 0.75 volt,
i*n A when Ea, the potential drop in A, is 2 volts, and
in B when Eb, the potential drop in B, is 1.5 volts.
Obviously, if the resistance of X is to be calculated from values
assigned to A, B, and Y, the potential drop in none of the four re¬
sistances can be permitted to exceed the value just stated, and pref¬
erably it should not exceed two-thirds this value. In the absence
of known compensating effects, it will be assumed that the maximum
permissible sensitivity is obtained when the potential drop in the
bridge is as high as possible, without that in any arm of the bridge
being higher than two-thirds that which results in a proportional
change in the resistance equal to the precision sought in the
measurement.
With the battery and galvanometer in their normal positions, it is
readily seen that IF=20 ohms, also that as E is increased ET is the first
to reach the maximum permissible value and that Ex= EX/(X-\-Y).
Therefore, eq 20a may be written
YV
S=2DEX (22)
{X+Y)Wt
and taking Ex =0.5 volt gives
o_2X10X106X.5X5X15
= 2.5X106. (22a)
* 15X20
That is, the change in deflection of the galvanometer following a
reversal of the connections to the battery is 2.5 mm per part per million
lack of balance of the bridge.
With the positions of the battery and galvanometer interchanged,
it is readily seen that W is less than V) that, as e is increased, Eh is the
first to reach the maximum permissible value; and that Eb=eB/(B-\-Y).
Therefore, eq 21b may be written
YV
S=2DEb (23)
(B+Y)W.
Taking Eb— 1 volt and V/W—l gives
iSr=2X10X106XlX5_Xl__3 3X1Q6
(23a)
151/237
Journal of Research of the National Bureau of Standards [Voi. es
(27 a)
S~2DE\A.+X)W’
and
S=2DE>(B+Y)W- (27 b)
Now, if it is understood that Ex, Ev, Ea, and Eb each represents the
maximum permissible potential drop in X, in Y, in A, and in B, each
of eq 26 and each of eq 27, in general, gives a different sensitivity.
However, that one of eq 26 which gives the lowest sensitivity gives
the maximum permissible sensitivity with the battery and galva¬
nometer in their normal positions; while that one of eq 27 which gives
152/238
Weriner ] Resistance Comparisons
3. THOMSON BRIDGE
572772 O - 61 - 11 153/239
Journal of Research of the National Bureau of Standards [voi.
(F+ Y')V
(28a;)
\X+X'+Y+ Y')W’
(X+X')V
(281/)
\X+X'+Y+ Y') W’
BV
(28a)
\A+B)W’
and
AV
(286)
\A+B)W’
if the positions of the battery and galvanometers are as shown in
figure 2, and
AV
S=2DEX (29a;)
(A+X+X')W’
BV
S—2DEy (29 y)
{B+Y+ Y')W
(XYX')V
S=2DEa (H+W-l-W/)IF, (29 a)
and
(Y+Y')V
S=2DEb (296)
(B+Y+Y')W’
if the relative positions of the battery and galvanometer are inter¬
changed.
With reference to the differences between eq 26 and 28 and eq 27
and 29, it should be pointed out that for the F-connected conductors
to be equivalent to the A-connected conductors it is necessary that
154/240
Wenner] Resistance Comparisons
155/241
Journal of Research of the National Bureau of Standards \voi. gs
(1.) From among the available 0.1- and 0.01-ohm standards select
from each denomination one of the better from the standpoint of
facilities for dissipating heat and low temperature coefficient.
(2.) Inspect these two standards and make an estimate of their
relative facilities for dissipating heat.
(3.) From the estimate of their relative facilities for dissipating heat
and their known temperature coefficients make an estimate of the
ratio of the load coefficient of the 0.01-ohm standard to the load co¬
efficient of the 0.1-ohm standard, and designate this ratio k.
(4) Place the 0.1-ohm standard in the X arm and the 0.01-ohm
standard in the Y arm of a bridge, and balance the bridge by ad¬
justments of the A arm, first with 0.1-watt and then with 0.5-watt
power dissipation in the 0.1 ohm-standard. Assuming that the power
dissipated in the A and B arms of the bridge has no appreciable
effect on their resistances, the load coefficient of the 0.1-ohm standard
(that is, the change in the resistance of the 0.1-ohm standard result¬
ing from the dissipation of 1 watt in it) is taken as 2){ times the pro¬
portional increase in the resistance of the A arm divided by (1 —
0.1 k).
(5) Place this 0.1-ohm standard in the Y arm of the bridge and a
1-ohm standard in the X arm, and balance -the bridge first with one
and then another potential drop across the bridge. The load co¬
efficient of the 1-ohm standard is taken as equal to the proportional
increase in the resistance of the A arm of the bridge divided by the
increase in power dissipation in the 1-ohm standard plus 0.1, the load
coefficient of the 0.1-ohm standard as determined in (4). It will be
noted that the result obtained involves the initial estimate of the
load coefficients of the 0.01-ohm standard relative to that of the
0.1-ohm standard to the extent of only 1 percent. The value thus
obtained for the load coefficient of the 1-ohm standards may, there¬
fore, be presumed to be somewhat more precise than the value obtained
for the load coefficient of the 0.1-ohm standard.
Load coefficients of standards of higher nominal values may be
determined as outlined above, but for standards of the lowest nominal
value used a different procedure is required. If among these there
are two standards of such construction that it may be assumed that
the temperature rises for equal power dissipations are equal and that
the temperature inequalities cause little or no mechanical strain, and
if these two standards have markedly different resistance-temperature
coefficients, their load coefficients may be determined from their
resistance-temperature coefficients and the difference of their load
coefficients obtained by direct comparison.
Having determined the load coefficients of one or more standard
resistors, the load coefficients of others of the same nominal value
are readily determined by direct comparisons, whether or not their
performances are normal.
Investigation of wire standard resistors of the Physikalisch-Tech-
nische Reichsanstalt design having winding areas of approximately
40 cm2 has shown that the proportional change in their resistance
resulting from the dissipation of 1 watt in their windings is about the
same as the proportional change in their resistance resulting from
increasing the temperature of the oil bath 1° C. It may be concluded
therefore that the temperature rise of the resistance material above
the temperature of the oil is about 1° C per watt power dissipation.
156/242
Warner) Resistance Comparisons
On the same basis it has been concluded that for the sealed standard
resistors developed in this Bureau about 1907 and the double-walled
type constructed prior to 1930 [89] the temperature rise is also about
1° C per watt power dissipation. For standard resistors of lower
denominations in which the resistance material is in the form of sheets,
the temperature rise per watt dissipation is less. For those of the
Physikalisch-Technische Reichsanstalt design, of the smaller size, the
temperature rise is about 0.2° C for the 0.01 ohm, 0.4° C for the 0.001
ohm, and 0.3° C for the 0.0001 ohm per watt dissipation.
Since most of the temperature coefficients are less than 20 parts per
million, errors resulting from heating will, in general, be less than
1 part in 2 million if the potential drops are limited to the following
values:
0.15 volt for 1-ohm,
.5 volt for 10-ohm,
1.5 volts for 100-ohm,
5. volts for 1,000-ohm,
15. volts for 10,000-ohm,
0.05 volt for 0.1-ohm,
.035 volt for 0.01-ohm,
.0075 volt for 0.001-ohm, and
.0025 volt for 0.0001-ohm standard resistors.
The temperature rises, and consequently the load coefficients,
depend on the viscosity and other factors affecting the circulation of
the oil, the design of the standards, and their temperature coefficients,
so there must of necessity be large variations. However, the potential
drops stated above are about the maxima permissible in measurements
to 1 part per million, unless the load coefficients are known to be
abnormally low, or a procedure is followed for eliminating the error
which otherwise would result from the heating by the test current.
In routine testing the schedule of potential drops given above is
followed approximately in the comparisons of resistors having nominal
values of 1 ohm and less. In the comparisons of resistors having
nominal values of 10 ohms and more the potential drops used are
much less than those given in the schedule. In special cases the
criterion used for the maximum permissible potential drops, without
an application of a correction for the heating, is a very small but
definitely noticeable change in the balance of the bridge on increasing
the potential drop across the bridge by a factor of 2.
If there is occasion to use larger potential drops, if there is reason to
suspect one or more abnormally large load coefficients, or if the load
coefficient of one or more of the bridge arms is not known to the
accuracy necessary for obtaining the desired precision, a valid correc¬
tion for the effect of the heating may be obtained by balancing the
bridge first with what is presumed to be a suitable potential drop
across the bridge and second with a somewhat larger potential drop
across the bridge. If the second potential drop is twice the first, the
effects of the temperature rise in each arm of the bridge will be 4 times
as large as with the first potential drop. Consequently, if the balances
are established by adjustments of the A arm (or A and a arms) of
the bridge and the value taken for A is four-thirds the first minus
one-third the second, a correction is applied for the effects of heating
by the test current in all arms of the bridge.
157/243
Journal of Research of the National Bureau of Standards [voi. ts
The galvanometer which has been and still is being used in most of
the precision resistance measurements made in the National Bureau of
Standards was designed and constructed especially for the purpose
about 1914 [116]. It has an adjustable shunt on the magnet, all¬
copper circuit, and taut suspensions with the center of gravity of the
moving system slightly off the axis of rotation. This latter feature
makes it possible by tilting to adjust the period, T, over a range from
about 5 to about 15 seconds. In cases in which really high sensitivity
is desired the performance seems to be most satisfactory with a period
of about 10 seconds'. The shunt of the magnet is so adjusted that with
a period of 10 seconds the external resistance giving critical damping
is 35 ohms. This adjustment once made is seldom changed as the
shunt is not readily accessible. The galvanometer is used with a
scale distance of 1.5 m. With critical damping the operating con¬
stants then are
T=10 seconds,
D=30 mm/^v,
I7=35 ohms,
158/244
Wenner] Resistance Comparisons
1 The lack of an inductive balance manifests itself as ballistic deflection of the galvonometer, following a
reversal of the test current. If this ballistic deflection is large, it limits, somewhat, the precision of the
resistive balance. However, there is a more important reason why the ballistic deflections should be kept
small. Most sensitive galvanometers when deflected alternately in one and then in the other direction,
by equal amounts, have their rest points shifted slightly in the direction of the last deflection. For the
galvanometer used with the NBS precision bridge, if the rest point is observed after a deflection in one
direction and again after an equal deflection in the opposite direction, the difference between the two
observed rest points is from 1 to 2 percent of the amplitude of the deflections. If, therefore, systematic
errors from this source are to be insignificant, inductive balances must be such that the ballistic
deflections are less than 50 times the change in deflection resulting from lack of resistive balances correspond¬
ing to the precision sought in the comparisons.
In the comparisons of standard resistors of the usual construction, having resistances in the range from
0.1 to 100 ohms, usually this condition is realized without a special device for making inductive balances
and without special precautions on the part of the observer. However, if the resistors in the X and Y arms
of the bridge have low resistances of different nominal values (such, for example, as 0.001 and 0.0001 ohm), and
if a high precision is desired, a means for compensating the effect of the difference between their time con¬
stants and of mutual inductances between different parts of the bridge circuit is necessary. Likewise, if the
resistors in the X and Y arms of the bridge are of the usual bifilar construction and have high resistances of
different nominal values (such, for example, as 10,000 and 1,000 ohms), and if a high precision is desired,
provision should be made for compensating the effect of the distributed capacitances. The effects of self
and mutual inductances and of distributed capacitances may be compensated by a mutual inductor
having one of its windings connected in a galvanometer lead and the other winding connected in a battery
lead between the current-reversing switch and a current terminal of the bridge, provided the inductance
is adjustable over a suitable range of positive and negative values.
159/245
Journal of Research of the National Bureau of Standard [ voi. sb
Iu general, the notation here is the same as in figures 2, 9, and 16. Here 1 and 2 represent main terminal
blocks (shown in detail in fig. 11); 3 and 4, current-terminal blocks; 5 and 6, terminal blocks for the less fre¬
quently used ratios of A to B', 7 and 8, hinged terminal blocks to accommodate standard resistors with different
spacing of terminals; 2 50, the three adjustable resistance sections shown in figure 4; Rt>,^compression carbon-
plate rheostat for adjusting large test currents; P, potential divider for adjusting small test currents; AM,
ammeter shunt; VM, voltmeter leads; ki, ki, and S, keys and switch for closing battery branch and reversing
connections to battery leads, k\ and k2 for small test currents, S for large test currents; Qi and Q$, copper
links and terminal blocks with amalgamated contacts for connecting two sections of the A arm ana of tnea
arm, either in series or In parallel. For the Wheatstone bridge method usually terminals 1 and 2 are used
instead of 3 and 4, xz and yz are left disconnected, and the galvanometer circuit is closed by a connection
between M and q.
160/246
Wenner] Resistance Comparisons
161/247
Journal oj Research oj the National Bureau of Standards ivoi.25
162/248
Journal of Research of the National Bureau of Standards Research Paper 1323
163
Figure (i. -General view of the XUS precision bridge (removed from oil bath).
Journal of Research of the National Bureau of Standards Research[Paper 1323
Wenner] Resistance Comparisons
Y
X2 ~jjAn (1 -f- (Z2 ®n)
r <33>
X3 ~jffXn (1 T Ch ®n)
Y
Xi~ "g-A„(l ~\~a4 <z„)
etc.
Letting N represent the nominal value of X and of YAJB, and
x, y, and b represent the amounts in proportional parts by which
X, Y, and B exceed their nominal values, eq 33 may be written in
the form
N( 1 -fa:) =N(l -f y) (1 -f a—an)/(l -f b). (34)
Expanding eq 34 and neglecting 3d and higher-order terms gives
x=y+ (a—an)— b+y(a—an)~ b(a—an)+ b2—by. (35)
Letting 2 represent that part of eq 35 which, for the series of measure¬
ments, is a constant, and, for the present, leaving out of consideration
the second-order term (y—b)(a—an) gives
Zi = oq-f 2
x2=02-f 2
■ (36)
X3 — 03 + 2
z4=a4 + 2 ,
etc.
Since X2 and X± are known, ?2 and x4 are known. Furthermore, all
a’s will be read from the bridge. Therefore, 2 may be determined from
the second or fourth of eq 36, or what is better, from both of these
equations and the mean value used. Substituting the mean value of
2 in each of eq 36 gives a value for each of the x’s and consequently
for Xu X2, X3, Xi} etc. in terms of the unit in which values of X2 and
Xi are expressed.
To see to what extent the values thus found are in error as a result
of the approximations involved in eq 36, it is necessary to consider
the second-order term (y—b){a—an) of eq 35, the magnitudes of
X\, x2, x3, x^ etc. and what constitutes nominal values of each of the
bridge arms. It was intended that resistances of the sections of the
A arm be so proportioned that differences of readings of the dials
equal differences in resistance in parts per million of the resistance the
A arm has when the reading of the dials is 5555. This resistance
(instead of 25, 100, or 1,000 ohms, depending on what sections are
used and the way they are connected) should be taken as the nominal
value of A. The appropriate multiple of the unit used is taken as the
165/249
Journal oj Research of the National Bureau of Standards [Voi. ns
166/250
Wenner] Resistance Comparisons
2. INTERCHANGE METHOD
Xl=Xrg(l+al—an). (38)
If, therefore, conditions are such that the second-order terms under
the radical and terms of the order of (a2—ax)2 may be neglected,
x2=xx+(a2—ax)/2, (42)
167/251
Journal of Research of the National Bureau of Standards ivoi.zs
of the dials of the A arm, for which the ratio of A to B has the nominal
ratio of Xx to X2. From eq 43 it follows that
xx=x2-\-ar—ax. (45)
S=Mx-\-M2AM^sA^tiC. (46)
1_1_
P~MX +K+ M3 + etc. (47)
Now, let Mbe their mean resistance and mx, m2, m3, etc. be the depar¬
tures of each in proportional parts from the mean of all. Then
and
S=nM, (51)
168/252
Wenner] Resistance Comparisons
and
p=iKi+Sm!) (52)
572772 0-61-12
169/253
Journal of Research of the National Bureau of Standards iva.ts
170/254
Wenner] Resistance Comparisons
When it is desired to find the reading of the A arm for which the
ratio of A to B is 2 to 1, the procedure is as follows:
1. The A arm is connected so that its nominal resistance is 100
ohms, the B arm is connected so that its nominal resistance is 50
ohms, the X arm is arranged to receive two standard resistors con¬
nected in series, and the Y arm is arranged to receive one standard
resistor.
2. Three standard resistors having the same nominal value (pref¬
erably 100 ohms) are used.
3. The bridge is balanced by an adjustment of the A arm with
each of the three standard resistors placed one after the other in the
Y arm, the other two being in the X arm.
4. The mean of the three readings of the A arm, after a correction
is applied to account for the resistance of connectors, is taken as the
reading of the A arm for which the ratio of the resistance of the A
arm to the resistance of the B arm is 2 to 1.
A similar procedure is followed in determining the reading of the
A arm for which the ratio of A to B is 3 to 2, 3 to 1, 2 to 3, 1 to 3,
1 to 2, or involves other small integers.
Figure 9.-—Circuit diagram of the Thomson bridge arranged for explanation of the
procedure followed in establishing balances of the bridge in such a way as to make
A/B=X/Y.
171/255
Journal of Research of the National Bureau of Standards [Voi. ts
(2) With the connectors L and 0 both closed, the bridge is balanced
by an adjustment of xx. These two adjustments make
x X
~=Y approximately. (55)
(3) With the connectors L and 0 both open, the bridge is balanced
by an adjustment of x2 [74]. This makes
X+x2+a_A+Xi (55a)
YJr'y2Jr ft B-\-yi
(4) With L closed and 0 open, the bridge is again balanced by an
adjustment of the direct-reading arms which, as stated above, are
mechanically connected so always read the same.
These four adjustments give
X_A
(56)
Y~B’
usually to the precision desired or attainable. The process is really
one of successive approximations, but the observer soon learns to
judge from the magnitude of the change in the reading of the dials in
making the fourth adjustment and the known precision required in
the various adjustments, if the adjustments need be repeated. Equa¬
tion 56 is the same as the Wheatstone bridge equation. Therefore,
in making a series of measurements, the data recorded and all calcu¬
lations are the same as though the Wheatstone bridge method had
been used. That is, only the readings of the dials for each of the
final balances of the bridge are recorded, and these are substituted in
the appropriate equation 36, 42, 44, or 45 considered above.
Since xfA and yfB seldom exceed 0.001, while in the final adjust¬
ment X\ is in series with A and y1 is in series with B, the precision
required in the second adjustment may be two or three orders lower
than that sought in the measurement. Furthermore, the precision
required in the third adjustment need not be high, and in many cases
this adjustment may be omitted. An important feature of the pro¬
cedure is that the adjustments result in making X/Y=A/B, whether
or not the four-terminal conductors, X and Y, are linear.
The simplicity of the adjustments results more from having suitably
designed and well-constructed apparatus than from the procedure
followed, which differs only in a minor detail from that described by
Jaeger and Diesselhorst [45]. Procedures for determining the cor¬
rection terms [P] of the Thomson bridge equation or for making these
terras negligibly small have been discussed by several of those who
have had occasion to use the Thomson bridge method.
172/256
Wmner\ Resistance Comparisons
the job well, be neither easily fatigued nor perturbed, recognize slight
disturbances promptly and be able to locate then* source and correct
the difficulty, and, above all else, be able to differentiate between that
which is essential and that which is not essential in each of eight
orders of magnitudes.
At present the resistance of no standard resistor can be presumed
to be known (either in international ohms 3 or in absolute ohms) to an
accuracy within one order of trhe precision to which two or more
nominally equal resistances can be compared (if within the range con¬
sidered here) or quite to the precision to which an 0.0001-ohm standard
may be compared with a 10,000-ohm standard (provided a sufficient
number of intermediate steps is used). It might seem, therefore,
that the precision considered here is unnecessarily high. However,
the precision of the resistance measurements is a factor contributing
somewhat to the accurac}^ to which units of resistance may be realized
and is an important factor in selecting the particular standard resistors
used in maintaining a unit of resistance. For judging the relative
quality of standard resistors, a precision of 1 part in 1 million is really
needed, and it is a great convenience that for the most part this is
readily obtained. As an indication of a limited need for a somewhat
higher precision, it may be pointed out that the resistance of each of a
fairly large group of 1-ohm standard resistors (of the double-walled
type constructed by Thomas in 1933) is remaining so constant relative
to their mean resistance that only by making the comparisons to a
precision of about 1 part in 10 million can the changes occurring during
a few months be detected.
In the above discussions a stated precision represents twice the
probable error of a single measurement. The precision is estimated
from the extent of the agreement of results on repetitions of the
measurement under different conditions, or the consistency of results
obtained in case more than the minimum required number of measure¬
ments is made.
For example, if five 1-ohm standards are to be intercompared,
they may be substituted one after another in the same arm of the
bridge. The five balances of the bridge thus obtained give data from
which the resistance of each standard may be calculated in terms of
the resistance of any one or the mean resistance of any two or more.
Repetitions of these measurements under conditions giving different
distributions of the systematic errors furnish data for determining the
probable error of a single measurement by the substitution method,
that is, the probable error of each of the five results obtained from five
balances of the bridge.
The same five standards may be compared by interchange between
the X and Y arms of the bridge in all possible combinations. The
data obtained from the 20 balances of the bridge then serve in calcu¬
lating the resistance of each standard in terms of the resistance of any
one or the mean resistance of any two or more, the probable error in
the results and the probable error in a single measurement by the
interchange method. If all balances of the bridge were equally
reliable, the probable error in the results obtained from the 20 balances
of the bridge would be smaller than that obtained from five balances
3 Here by “international ohm” is meant a unit of resistance realized by the use of a column of mercury
under specified conditions, as distinguished from the unit of resistance of this country or the mean of the
units of resistance of several countries.
173/257
Journal of Research of the National Bureau of Standards [voi.25
174/258
Wenner] Resistance Comparisons
sary in order that the resistance be definite to 1 part in 1 million, considering only
the effects of contacts and current and potential distributions in terminals. But
first it will be necessary to state more precisely what is to be understood by the
resistance, or rather the resistances, of a conductor.
Since the more important resistances of a four-terminal conductor are more
definite than is the resistance of a two-terminal conductor, four-terminal con¬
ductors will be considered first. Referring to figure 10, the heavy lines desig¬
nated 1, 2, 3, and 4 will be considered as representing surfaces on which con¬
nections to other conductors may be made, that is, they will represent terminals.
In case 1 and 4 are normally used as current terminals and 2 and 3 are normally
used as potential terminals, the resistance may be defined as the ratio of the
drop in potential from 2 to 3 to the current entering on 1 and leaving on 4.
This will be designated the resistance (1234). In case 2 and 4 are used as cur¬
rent terminals and 1 and 3 as potential terminals, the resistance may be defined
as the ratio of a drop in potential from 1 to 3 to the current entering on 2 and
leaving on 4. This will be designated the resistance (2134).
It will thus be seen that a four-terminal conductor has as many four-terminal
resistances as there are permutations of four numbers. However, only two of
these resistances are independent of each other,
and the relations between all are rather simple.
Since there are so many resistances, it is a con¬
venience to divide them into groups and to attach
distinctive names to these groups. The four
equal resistances (1234), (2143), (4321), and
(3412) will be called the direct resistance. The
four equal resistances (1432), (4123), (2341), and
(3214) will be called the cross resistance, and the
four equal resistances (1243), (2134), (3421), and
(4312) will be called the diagonal resistance.
When defined in this way, the direct resistance
four-terminal conductor. minus the cross resistance equals the diagonal
resistance, as was pointed out in a previous paper
[101]. Furthermore, any one of the 12 remaining four-terminal resistances is
equal in magnitude but opposite in sign to either the direct, the cross, or the
diagonal resistance.
It is also a convenience to divide four-terminal conductors into groups in ac¬
cordance with the relative magnitudes of the resistances and attach names to
these groups. Conductors for which the cross resistance is less than a millionth
part of the direct resistance will be called linear,4 and all others will be called
nonlinear.
If variations in the current and potential distribution on terminals are to affect
the direct resistance by certainly less than 1 part in a million, the terminals must
be so designed that the sum of the following potential ratios will be less than
0.000 001. With 1 and 4 used as current terminals, (1) the potential difference
between any two points on 2 to the potential difference between 2 and 3; (2)
the potential difference between any two points on 3 to the potential difference
between 2 and 3, and with 2 and 3, used as current terminals; (3) the potential
difference between any two points on 1 to the potential difference between 1 and
4; and (4) the potential difference between any two points on 4 to the potential
difference between 1 and 4. This condition is generally realized in standard
resistors designed for precise measurements, but is seldom realized in standard
resistors designed for use with large currents. In many cases the resistance
depends upon the manner of connecting the current leads to the current terminals
to the extent of 0.01 percent, and in exceptional cases to the extent of 1 percent.
The resistance of a two-terminal resistor may be defined as the ratio of the po¬
tential drop to the current, taken between two equipotential surfaces lying par¬
tially in the terminals of the resistor and partially in the conductors through
which the current enters and leaves the resistor. From an experimental stand¬
point, it is better to define the resistance of a two-terminal resistor as the differ¬
ence in resistance of two four-terminal conductors which are the same in all re¬
spects, except that one does and the other does not include the two-terminal
resistor.
* It should be noted that the distinction made here between a linear and nonlinear conductor does not
involve a proportionality and lack of proportionality of the potential drop to the current.
175/259
Journal oj Research oj the National Bureau of Standards [voi. ss
To illustrate what is meant by this statement and to show that the resistance
so defined is not strictly a constant, let it be assumed that there is at hand a short
copper rod 1 cm in diameter and having plane end surfaces which are amalga¬
mated and that the resistance of this rod is to be measured. Also let it be assumed
that there are at hand two sets of cylindrical copper terminal blocks having plane
amalgamated end surfaces, the diameter of one set of blocks being 1 cm and the
diameter of the other set of blocks being 2 cm. Each terminal block has a flexible
current lead and a flexible potential lead. If (1) the copper rod is placed between
the amalgamated surfaces of the terminal blocks having the smaller diameters
and the resistance of the four-terminal conductor thus constituted is measured;
if (2) the copper rod is removed, the amalgamated surfaces of terminal blocks are
placed in contact, and the resistance of the four-terminal conductor thus consti¬
tuted is measured; and if (3) the latter measured value is subtracted from the
former measured value, a value is obtained for the resistance, as defined, of the
copper rod. This resistance will in general be slightly higher than that which
may be calculated from the known resistivity and dimensions of the copper rod.
Now, if this procedure is repeated, using the terminal blocks having the larger
diameter, a still higher value will be obtained for the resistance of the copper rod,
provided the contacts between the rod and terminals are equally good in the two
cases, and provided also that when terminal blocks are placed together, the con¬
tact is such as to give a substantially uniform current density over the contact
area. In the first case, one amalgamated contact is included in the measured
resistance. In the second case, there is in addition a nonuniformity of current
density in the vicinity of the contacts, which in effect increases the measured
resistance.
Obviously, therefore, the resistance of a two-terminal standard resistor depends
to a greater or less extent on the manner in which it is connected into a circuit.
For example, when a two-terminal standard resistor is placed in an arm of a Wheat¬
stone bridge, the bridge arm, which is a four-terminal conductor, consists of the
standard and the end connectors. If the terminals of the standard and of the end
connectors are amalgamated, the end connectors consist of two amalgamated sur¬
faces or mercury cups and two terminal blocks of the bridge. With end connectors
of the usual type, that is, having deep mercury cups, the resistance of the bridge
arm may depend on the amount of the mercury in the cup and the position of the
terminals of the standard in the cup to the extent of a few microhms. This diffi¬
culty is obviated to a considerable degree by the use of plane amalgamated
surfaces, instead of mercury cups.
What is of more importance than the form of the contact surfaces is the gradual
accumulation of copper amalgam in the solid phase on the terminals of standard
resistors and on end connectors or contact blocks. Cases have been observed in
which the removal of the amalgam in the solid phase has resulted in a lowering
of the resistance by more than 10 microhms (though a definite reason for so large
a change has not been found). Even when the amalgamated surfaces of the
terminals of the standard resistor and of the terminal blocks with which it may
be used are apparently in good condition, the resistance may be expected to depend
on the resistances of the amalgamated contacts and the distribution of the current
density in their vicinity to the extent of a few microhms.
The more important terminal blocks of the NBS precision bridge and all ter¬
minal blocks of the auxiliary apparatus used in establishing known resistance
ratios have four or more terminals.
A diagram of a main terminal block of the bridge which is used when the A
arm is 25 or 100 ohms is shown in figure 11 (a). The block is constructed of copper,
and terminals 1, 2, 4, and 6 are copper wires soldered in holes drilled in the under
side of the block. Terminal 3 is a plane amalgamated surface (not a mercury cup),
and terminal 5 is a small mercury cup. Terminals 1 and 5 are so located with
respect to the other terminals that they may be considered as equivalent, unless
both are used at the same time. Terminal 6 is connected to the 900-ohm section
of the A arm. The main terminal block normally used with the B arm of the
bridge, when this arm has a resistance of 25 ohms or 10 ohms, is similar to that
shown in figure 11 (a) except that it has no terminal 6. When the Wheatstone
bridge method is used, terminals 1" (or 5), 3 and 4 are used as shown in figure
11 (b). When the Thomson bridge method is used, terminals 1, 2, and 4 are used
as shown in figure 11 (c). Consequently, the current distributions in these blocks
are not the same in the two cases, and as a result the contributions of one terminal
176/260
Wenner] Resistance Comparisons
block to the resistance of the A arm and of the other terminal block to the re¬
sistance of the B arm are not the same. When the Thomson bridge method is
used the resistance contributed to the A arm or to the B arm by its terminal block
is higher than when the Wheatstone bridge method is used, by the four-terminal
resistance (1234).
Since the reading of the A arm of the bridge for which the ratio of A to B is
10 to 1 is determined by the Wheatstone bridge method and this reading is used
with both the Wheatstone and the Thomson bridge methods, the effect of these
resistances should be known. Measurements of these resistances give values rang¬
ing from —0.3 to +0.3 microhm, depending upon the point on the amalgamated
a
Figure 11.—A main terminal block of the NBS precision bridge.
a shows the six terminals of the terminal block, 6 shows the three terminals used with the Wheatstone
bridge method, and c shows the three terminals used with the Thomson bridge method.
177/261
Journal of Research of the National Bureau of Standards [Voi. ts
tacts and a portion of each of two terminal blocks. Terminal 2 is also used as a
current or potential terminal, both when the three 150-ohm resistance sections
adjacent to it are connected in series and when they are connected in parallel.
Terminal 1 serves in connecting the apparatus into the bridge. With the parallel
connection two-thirds of the current is through terminal 3, one-third of the cur¬
rent is through terminal 4, and all of the current is through terminal 2 or terminal 1.
In measuring the resistance of the paralleling connection, all of the current is
through terminal 3 and all through terminal 1 or terminal 2. With the series
connection, all the current is through terminal 1 or terminal 2, and all through
terminal '4.
One of the terminal blocks to which two 150-ohm sections are connected is
shown in elevation in figure 13 and in plan in figure 8. Here terminal 3 is an
amalgamated surface on which a copper link rests when the resistance sections
terminating in the block are connected in parallel, and 2 is a small mercury cup
used as a terminal in measuring the resistance of the paralleling connection.
178/262
Wenner] Resistance Comparisons
3 3
ance of the 50-ohm section resulting from the change of the connection of three
150-ohm sections from series to parallel. In eq 57, Lj represents the four-terminal
resistance of the paralleling connection between terminal blocks a and h, using
1 and 2 of terminal block a, and 1 and 2 of terminal block h, as the terminals (the
measurement being made with the other paralleling connection open). A similar
statement applies to L2, and (1234) „ represents the direct resistance of terminal
block a, (1432) a the cross resistance of terminal block a, etc. A similar solution
applies to the other group of three 150-ohm
sections. To this point, three different cur¬ ^ 2 6
rent distributions in each of the eight termi¬
nal blocks of the auxiliary apparatus have
been considered.
However, the relative values of resistance
sections must be determined, and measure¬
ments of these require additional current
distributions in at least some of the terminal
blocks. Of these additional current distri¬
butions, the two in each of the central termi¬
nal blocks, d and d', which are of the type
shown in figure 15, are of most importance.
The additional current distributions in the
central terminal blocks are involved in a
determination of the ratio of the resistance 50 150
of the 50-ohm section to the resistance of Figure 15 .—A central terminal
the six 150-ohm sections connected in series. block of the auxiliary apparatus.
The measurements used in this determination
consist in comparisons of the resistance of the 50-ohm section with the resistance
of one and then of the other of the groups of three 150-ohm sections connected in
parallel.
In the measurement of the resistance of one group of three 150-ohm sections
connected in parallel, 1 and 2 of terminal block a and 5 and 2 of terminal block d
are used as the four terminals. In the measurement of the resistance of the other
group of three 150-ohm sections connected in parallel, 1 and 2 of terminal block
o' and 5 and 2 of terminal block d' are used as the four terminals. These resist¬
ances will be referred to as the resistances (la2a2d5d)j> and (la-2a'2d-5d0 v. The
resistance of the 50-ohm section is measured, using 5 and 2 of terminal block d
and 5 and 2 of terminal block d' as the four terminals. This will be referred to
as the resistance (5d2d2d'5d0- The addition of the resistance (2a5dld2d)P to the
resistance (la2a2d5d)z> gives the resistance (la2a2dld)p, which, when multiplied by
the ratio given by eq 57, gives the resistance (la202dld)«, that is, the four-terminal
resistance of the three 150-ohm sections in series and with the normal distribution
179/263
Journal oj Research oj the National Bureau of Standards ivoi.gs
180/264
Wenner] Resistance Comparisons
When the Thomson bridge method is used with the battery and galvanometer
in their normal positions there is current through each of the four terminals of
each of the two four-terminal conductors designated X and Y in figures 9 and 16.
There may therefore be some question as to which if any of the four-terminal
resistances of X and Y are compared.
If the bridge is balanced according to the procedure outlined above or any other
giving the same result, the connection through L may be opened or closed without
changing the balance of the bridge; also the connection through 0 may be opened
or closed without changing the balance of the bridge. That is, the balance of the
bridge is independent of the resistance of the branch L and independent of the
resistance of the branch O. As will be shown in the appendix on methods of
analysis, a conductor having a resistance, R, and in which there is a current, I, may
181/265
Journal oj Research of the National Bureau of Standards [voi.bb
If, for example, terminal 1 of X were connected to the A arm of the bridge,
terminal 2 were used as a connection to the battery, and other connections were
as shown in figure 16, the adjustment would establish the relation
A_the diagonal resistance of X
B the direct resistance of Y ' '
The conductors, X and Y, might be connected into the bridge in such a way that
the adjustment would make the ratio of A to B equal to any one of several ratios
of a four-terminal resistance of one, to a four-terminal resistance of the other.
Finally, it should be noted that whether the Thomson bridge method, the poten¬
tiometer method, the Matthiessen and Hockin bridge method, or the Kohlrausch
differential galvanometer method is used the same value will be found for the
ratio of a four-terminal resistance of X to a four-terminal resistance of Y, provided
proper adjustments are made in each case and equal^precisions are attained in
these adjustments.
182/266
Wenner] Resistance Comparisons
13. The dial switches serving to adjust balances of the bridge operate in the oil
bath, so are well lubricated, and this reduces the heat developed in the contacts
by the operation of the switches.
14. The resistances in series with the dial switches are high relative to the
resistances of the sections shunted. Consequently, variations of the thermo-
electromotive force, introduced into the A arm of the bridge by the dial switches,
are fully one order smaller than the variations of the thermoelectromotive forces
in the dial switches.
Since thermoelectric balances and thermal shunts are not generally used else¬
where, an explanation of each may be in order.
(a) THERMOELECTRIC BALANCES
If a complete electric circuit is made up in part from one kind of metal and in
part from another kind of metal, there must be at least two and in any case an
even number of junctions between dissimilar metals. That is, the junctions occur
in pairs. When these junctions are at different temperatures usually there is a
thermoelectromotive force in the circuit. However, it is not necessary that all
junctions be at the same temperature to avoid a thermoelectromotive force. The
thermoelectromotive force would be zero if the two junctions constituting one pair
were at one temperature, the two junctions constituting another pair were at
another temperature, etc. This is the end sought in making thermoelectric
balances. In the simplest case a thermoelectric balance consists in making the
two junctions of a pair of similar construction and of placing these two junctions
in such relative positions that both will be exposed to very nearly the same
ambient temperature. More generally, a thermoelectric balance consists of con¬
structions and arrangements of parts such that some type of symmetry both
from the thermal and from the thermoelectric points of view is realized for two or
more of several junctions in an electric circuit. Tor a circuit to be thermoelectri-
cally balanced, each junction between dissimilar metals must be taken into
consideration.
For the purpose of illustrating a method of realizing a thermoelectric balance,
consider a soldered connection between two copper wires of the same size which
are exposed to air currents of varying temperature. The connection is made
similar in the two directions parallel to the circuit. The conductor is supported
from opposite walls of the laboratory, and the connection is placed at some dis¬
tance from either support. Consequently, air currents heat and cool the wires
on opposite sides of the connection at very nearly the same rate, and keep them at
very nearly the same temperature. Therefore, the two junctions, solder to copper
and copper to solder, always assume very nearly the same temperature, so both
the magnitude and the rate of change of the thermoelectromotive force are very
small. In this case, one of the junctions is balanced against the other.
For the purpose of illustrating another method of realizing a thermoelectric
balance, consider two sections of a circuit in each of which there is a soldered
connection between a copper wire 1 mm in diameter, and a copper wire 0.5 mm
in diameter. The two soldered connections are made similar and so located
with respect to each other that in case of a changing temperature, the rate of
heat transfer from a larger to a smaller wire is the same through one as through
the other of the soldered connections. As a consequence, the difference in tem¬
perature of two junctions, copper to solder and solder to copper, of one connec¬
tion, is very nearly the same as that of the other connection. Therefore, the two
soldered connections, each constituting a thermocouple, develop very nearly the
same electromotive force. Further more the arrangement is made such that in
case of a current in the circuit, this current is from the larger to the smaller wire
through one of the soldered connections, and from the smaller to the larger wire
through the other connection. Therefore, the electromotive force developed in
one of the soldered connections is in opposition to that developed in the other.
In this case, one of the thermocouples is balanced against the other.
(b) THERMAL SHUNTS
183/267
Journal of Research of the National Bureau of Standards ivoi. ss
the larger wire, and this portion thoroughly impregnated with shellac. Then,
although air currents may heat and cool the smaller wire more rapidly than the
larger, the heat transfer between the two is mainly through the electrical insula¬
tion. The result is a material reduction in the temperature difference and rate
of change of the temperature difference between the two junctions, copper to
solder and solder to copper, and consequently, a material reduction in the thermo¬
electromotive force and its rate of change.
Thermoelectric balances and thermal shunts constitute two of the simpler
and more generally applicable of the effective means of reducing troubles from
thermoelectromotive forces. Furthermore, each may be made to supplement
the other and sundry means employed for this purpose, such as a thermal shield
the circulation of air in an enclosed space, the blocking of air currents, etc.
With the arrangement used and procedure followed in establishing balances
of the bridge, thermoelectromotive forces seldom constitute a limiting factor on
the precision of measurement, even when there is forced circulation of the air
from the general heating system at a temperature 10° C above the temperature
of the bridge and 15°C above the general temperature of the laboratory or from
the refrigerating system at a temperature 15°C below the temperature of the
bridge and laboratory. Only in exceptional cases is it necessary to place tem¬
porarily an obstruction to the normal circulation of the air in the vicinity of the
galvanometer, the terminals of standard resistors, or the connectors extending
above the surface of the oil.
APPENDIX 3. INSULATION
184/268
Winner] Resistance Comparisons
the temperature of the oil below the dew point of the air of the laboratory, as this
would result in the condensation of water in the oil. (3) The oil is selected on
the basis of original freedom from acids and tendency to develop acids, is protected
from strong light, is kept covered when the apparatus is not in use, and is re¬
newed at intervals of about 2 years. (4) Most of the insulating material is pro¬
tected from strong light, especially when not in use. (5) Care is taken not to
permit the accumulation of excessive amounts of dust or hygroscopic sulfur com¬
pounds on the surfaces of the insulators of the hard-rubber type. (6) The bridge
proper and the rheostat used in regulating the damping of the galvanometer are
kept under oil. (7) The galvanometer leads have but few supports and these are
of amber. (8) The galvanometer case is supported on amber posts, and the
galvanometer circuit is insulated from the case. (9) The battery is supported on
blocks of paraffin, and the battery leads are well insulated.
However, the resistance of the insulation of the battery and its leads, the
rheostats for regulating the test current, the ammeter-voltmeter, keys, etc., gen¬
erally is from two to four orders lower than the resistance of the insulation of the
galvanometer and galvanometer leads. It should be noted that it is not neces¬
sary that the insulation of both the battery branch and the galvanometer branch
be exceptionally good. Nevertheless, when the humidity is very low, the insula¬
tion of the entire bridge circuit is so good that there may be trouble from electro¬
static actions. To avoid electrostatic effects arising from the stirring of the oil,
the circulating propeller is electrically connected to the guard, and the belt driv¬
ing the circulating propeller is located beneath the metal tank containing the oil
bath. The more serious of the remaining sources of electrostatic effects arise
from movements of the observer. In case these movements cause trouble, the
guard is temporarily extended under the observer so he becomes a part of it,
the observer is thoroughly insulated from the floor, or a galvanometer lead is
connected to the galvanometer case. Should slight defects in the insulation be
suspected, in any case, their presence or absence may be confirmed easily. If
present, but not very serious, usually it is a simple matter to carry out the measure¬
ment in such manner as to avoid an error from this cause.
(a) THEORY IN CASE OF DEFECTIVE INSULATION
The correction for the effect of slightly defective insulation will be designated
as /. Then, following the notation used in the paper, the Wheatstone bridge
equation may be written
In cases in which the resistances of the battery leads and galvanometer leads are
negligibly small in comparison with the resistances of the bridge arms, as is usually
assumed in discussions of this subject, figure 17 represents the bridge circuit with
sufficient exactness for the purpose at hand. Here a, b, c, and d represent branch
points of the bridge; e represents the guard extending under all mechanical
supports of the apparatus; and m, n, o, and p represent the resistances of the
insulation between leads, or branch points of the bridge and the guard. For this
somewhat ideal arrangement, if the bridge is balanced, the relation between
A, B, X, Y, m, n, o, and p is
X(l-\-Ymp/F) AA + Brnn/F) „„
Y{l + Xop/F)~ Bil + Ano/F)’
where
F=mno-\-nop-\-opm-\-pmn. (64)
572772 O - 61 - 13 185/269
Journal oj Research oj the National Bureau oj Standards [voi. ss
Equation 66 shows the effect of slight defects in the insulation and the relation
between these defects necessary for them to compensate each other. However,
it is of little use in the solution of the particular problem at hand, since, in general,
m, n, o, and p will not be known. Furthermore, it is not sufficiently general to be
applicable in all cases.
Frequently a resistance is placed in one or both battery leads to reduce the
potential drop across the bridge to a value less than the electromotive force of the
battery and in one or both galvanometer leads or in parallel with the galvanometer
Figure 17.—Wheatstone bridge circuit arranged to show the effect of slightly defective
insulation of the battery leads and of the galvanometer leads.
to adjust the damping of the galvanometer. Also major parts of the leakage
through or over insulating material may be distributed near the battery and
galvanometer. When this is the case, resistances in the leads make slight defects
in the insulation more serious.
(b). PROCEDURE IN CASE OF DEFECTIVE INSULATION
For such a bridge circuit as shown in fig. 18 or the simpler one (shown in fig. 17), to
a first-order approximation, the correction term, /, of eq 62 may be considered to
be the product of two factors.5 One of these, which will be designated J, com¬
prises two or more resistance ratios involving only the resistances of bridge arms,
galvanometer leads, and leakage paths between galvanometer leads and the guard.
The other, which will be designated K, is the potential drop from the guard to the
branch point b (or branch point c) divided by the potential drop from a to d.
Following the notation used above and replacing / by the product of J and K,
* The basis of the discussion given here will be explained in the appendix on methods of analysis, first
expedient.
186/270
Wenner] Resistance Comparisons
J=ad—aa. (72)
From eq 69 and 72, or 71 and 72, it follows that
In like manner it may be shown that if balances of the bridge are made with
branch points b and c alternately connected to the guard
187/271
Journal of Research of the National Bureau of Standards [Voi.ss
It was mentioned that originally the bridge was equipped with a Wagner
branch. Usually when a Wagner branch is used with a Wheatstone or other
four-arm bridge, it is attached as shown in figure 19. Here Wa and Wb represent
the arms of the Wagner branch. The bridge and the Wagner branch are adjusted
by successive approximations. First, the bridge is adjusted with the key kw open,
so that there is no current in the galvanometer. Second, with the galvanometer
connected between branch point b (or c) and the guard or with the galvanometer
in its normal position, and the key, lcw, closed, the Wagner branch is so adjusted
that there is no current through the galvanometer. This second adjustment, if
it involves any marked changes in the ratio of Wa to Wh, disturbs the first adjust¬
ment to some extent. Therefore, the first adjustment is repeated and this dis¬
turbs the second adjustment slightly, etc. However, successive adjustments re¬
quire smaller and smaller changes, and after the changes become insignificant, the
galvanometer branch is at the potential of the guard. Then there is no leakage
current through the leakage path, q, or any arm of the bridge. Therefore, since
there is no current through the galvanometer, the Wheatstone bridge equation
gives the relation between X, Y, A, and B.
If this adjustment were valid with the key k\ against the + battery terminal
block and the key, k2, against the — battery terminal block, the relation between
the resistances would be as follows:
Wa(l + Wb/q)__X_A
^ Y~B'
However, with this adjustment the galvanometer branch will not be at the poten¬
tial of the guard with both keys k\ and k2 against + battery terminal blocks, as
shown in figure 18, or with Aq against a —, and k2 against a + battery terminal
188/272
Warner] Resistance Comparisons
block. More specifically, when a switch, such as is shown in figure 19, is used in
reversing the current through the bridge and this switch is located between the
bridge and a leakage path between a battery lead and the guard, no adjustment
can be made such that the galvanometer branch will be at the potential of the
guard for all permutations of the settings of the keys kly k2, and kw. If then there
is a leakage path from a galvanometer lead to the guard, no adjustment can be
made such that there is no current through the galvanometer with the current
through the bridge alternately in one and then in the other direction, when the
key kw is open and also when the key kw is closed.
Notwithstanding this difficulty it is possible, in some cases only, to make an
adjustment such that the relation between X, Y, A, and B is that given by the
Figure 19.—A Wagner branch connected in parallel with the battery branch of a
Wheatstone bridge.
Wa and Wb represent the arms of the Wagner branch, and p and q represent leakage paths between
the bridge circuit and the guard.
189/273
Journal of Research of the National Bureau of Standards [voi.ss
Figure 20.—A Wagner branch connected in parallel with the galvanometer branch
of a Wheatstone bridge
Here the arrows indicate the direction of the leakage current in each branch of the network when the
potential of all bridge arms is higher than the potential of the guard.
While balances of the bridge are being established, the deflection of the gal¬
vanometer is ever changing, so it requires the concentrated attention of the
observer. Watching the deflection and separating those components of the
deflection resulting from a lack of balance of the bridge from those resulting from
other causes constitute the most fatiguing part of a series of measurements. As
fatigue is a factor affecting the precision of a series of measurements the optical
system should be as suitable for the purpose as can be obtained without excessive
refinements. What constitutes the best obtainable optical system depends on
the skill of the available optician, personal preference, and physiological factors.
190/274
Wenner] Resistance Comparisons
both eyes and does not require that the observer keep his head in a very definitely
fixed position. The graduations are fine, distinct, black lines with 1-mm
spacing.
The lamp and scale are supported from a side wall and the ceiling, both of
which are masonry. Also the galvanometer is supported from a side wall. Con¬
sequently, mechanical vibrations of the scale and image are seldom troublesome.
With the lamp operated at normal intensity, the image appears rather wide
and excessively bright. Therefore, the power supplied to the lamp is reduced
somewhat below the point at which, on dull days and with the window shades
open, the diffraction fringes do not contribute to the apparent width of the image
and a suitable contrast is obtained. As the outside illumination changes the
intensity of the light source is not changed but occasionally one or more of the
window shades are closed or the lights of the laboratory turned on. The general
illumination is so adjusted that the contrast between the graduations and ground
glass is approximately the same as the contrast between the ground glass and
image.
In cases in which the combined sensitivity of the bridge and galvanometer are
barely sufficient for obtaining the desired precision, repetitions of series of meas¬
urements show, that an exceptionally skillful observer can so nearly balance the
bridge that that component of the electromotive force resulting from the lack of
balance of the bridge does not exceed 0.001 microvolt. Since the sensitivity of
the galvanometer then would be 30 mm/microvolt, it follows that 0.001 micro¬
volt corresponds to changes in the deflection of 0.06 mm, following reversals of
the direction of the test current. To associate a change in the deflection that is
only slightly larger than this with a change in the direction of the test current
(while the deflection is changing gradually as a result of one or more of several
causes, abruptly on reversals of the test current because of a lack of a perfect
inductive balance, and erratically as a result of one or more of several possible
causes); requires that the deflection be observed continuously while the test
current is reversed several times, at intervals corresponding approximately to
the period of the galvanometer. However, a balance of the bridge to 0.001
microvolt can be made only when the thermoelectromotive forces are remaining
reasonably constant, the inductances of the bridge are reasonably well balanced,
local and microseismic vibrations are not troublesome, electromotive forces
induced by changes of the magnetic field in the laboratory are not troublesome,
and conditions otherwise are favorable.
Since the optical system of the galvanometer is the last of the factors affecting
the sensitivity of bridges to be discussed at this time, it is in order to give an
illustration of how a consideration of most of these factors leads to a fairly definite
conclusion. For this purpose let the reader assume that he has the problem of
comparing a number of 10-ohm standard resistors, that he has at hand a bridge
suitable for the purpose, and that he has prepared a specification for a galvanom¬
eter, but before purchasing a galvanometer wishes to know if the specification is
adequate and satisfactory for the realization of a precision of 1 part in 1 million
in the comparisons.
Let it be assumed that:
1. The period of the galvanometer is to be 7 seconds.
2. The galvanometer is to be critically damped with an external resistance of
60 ohms.
3. The sensitivity of the galvanometer is to be such that with 1 microvolt, in
the circuit giving critical damping, the deflection is five scale divisions, with the
distance between the mirror and scale that which is to be used.
4. The optical system of the galvanometer is to be of such quality that a change
of the deflection of 0.1 scale division is readily detectable.
5. It is estimated that local disturbances will cause random change in the
apparent equilibrium position of the galvanometer of not more than 0.1 scale
division during time intervals of 7 seconds.
6. The resistance between the galvanometer terminals of the bridge is 60 ohms.
7. The comparisons are to be made by substituting the 10-ohm standard
resistors successively in the X arm of the bridge.
8. A 10-ohm standard resistor is to be placed in the Y arm of the bridge and is
not to be disturbed during the series of measurements.
9. Balances of the bridge are to be checked by a single observation of the change
in the deflection of the galvanometer following a reversal of the test current, with
no provision for establishing inductive balances.
191/275
Journal of Research of the National Bureau of Standards [voi. ts
10. The divergence of.the load coefficients of the series of 10-ohm standard
resistors, to be compared, is 10 parts in 1 million per watt dissipation.
As previously pointed out (eq 26a:), the change in the deflection of the galvanom¬
eter following a reversal of the test current is
2 DEX Y/(X+Y),
lack of balance of the bridge may be taken as 1 in 25 million Ex. Plotting the
minimum discernible lack of balance of the bridge as ordinates and Ex as abscissas
gives curve A of figure 21.
From 7 and 10, since the resistance in the X arms is 10 ohms, it follows that for
the series of measurements the divergence of the effects of heating is Hx2X10'6.
Plotting Ex2 as ordinates and Ex as abscissas gives curve B. Adding the ordinates
of curves A and B gives curve C, the sum of the various factors limiting the
sensitivity. It will be observed that the maximum of what may be called the
effective sensitivity is 2.2 in 10 million, or about 1 in 4)4 million, and that this
sensitivity is obtained with a potential drop of approximately 0.27 volt in the
X arm of the bridge. Obviously this sensitivity is somewhat more than that
required for a precision of 1 part in 1 million.
In this analysis it has been assumed that the combined effect of the various
factors limiting the effective sensitivity is obtained by the addition of their
individual effects, that is, the analysis is based on the worst, not the probable,
of the assumed conditions. It may be concluded, therefore, that a galvanometer
having the operating characteristics specified would be suitable for obtaining
the desired precision in the series of measurements, and inasmuch as there would
be no difficulty in obtaining a galvanometer having these operating constants,
it may be concluded that the specification is satisfactory.
192/276
Wenner] Resistance Comparisons
Before considering any one of these three expedients a few of the many possible
expressions for the electromotive force or current in the galvanometer circuit of
an unbalanced Wheatstone bridge will be stated, for the purpose of comparison
with corresponding expressions, one of which will be derived by the use of the
first and'another by the use of the second of these expedients.
The expression given by Maxwell and others, and which will be referred to as
the “classical solution,” is
EhBAX
Ig~RG(A + B+X+ AX + Y)+R(A + X + AX){B+ Y) + (76)
+ G(A + B)(X+AX+Y) + XY(A + B) + AB(X + AX+Y)+AXY(A + B).
Here Ig is the current in the galvanometer branch and G is the resistance of the
galvanometer branch. The significance of the other characters are as explained
in connection with figure 1.
It has been shown in this paper (eq 6) that
EX T l + AX/X
(77)
~ x+ y|_ i+ax/(x+ Y)
from which it follows [30, 71, 88, 108] that
ex r i+ax/x
(78)
1 *~ x+ FL1+AX/(X+ Y)
where R9 is the resistance to an electromotive force in the galvanometer branch,
that is, the" resistance of the galvanometer branch plus the resistance of the
bridge between its galvanometer terminals with the galvanometer branch open
and the battery branch closed.
193/277
Journal oj Research of the National Bureau of Standards [Voi. n
AB (X+AX)Y
Cr_t' A + B^X + aX+Y'
Consequently,
194/278
Wenner] Resistance Comparisons
electromotive force Eb sufficiently to give this current. If this were done the
galvanometer branch could be opened without changing the current I in the
battery branch. Furthermore, from the standpoint of an electromotive force
in the galvanometer branch, the resistance of the battery branch would be
infinite. Consequently, an expression for the resistance to an electromotive
force in the galvanometer branch could be written from an inspection of figure
1. With the galvanometer branch open it is readily seen that the current in
the X arm is I (A A- B)/(X-\- AX+ Y+ A-\-B) and in the A arm is I(XAX A-Y)/
(X+AX-f- F+-A + B). These currents multiplied by the resistances (X-f-AX)
and A give the potential drop from a to b and from a to c, while the difference be¬
tween these two potential drops is the electromotive force appearing in the galva¬
nometer circuit when the galvanometer branch is closed. With the galvanometer
branch closed it is readily seen that the resistance to an electromotive force in the
galvanometer branch is G-\- (A”+AX + A) (F+ B)/(X-\- AX + F+ A + B). Conse¬
quently, the current in the galvanometer branch
which reduces to
Before considering the separation and synthesis of the effects of the independent
resistances of a nonlinear four-terminal conductor, it is important to know the
number and nature of these resistances and the purpose of the separation and
synthesis. In the appendix on terminals and contacts it was pointed out that a
four-terminal conductor has but two independent four-terminal resistances.
These are the direct resistance and the cross resistance, which here will be desig¬
nated D and C. The remaining independent resistances are the terminal resist¬
ances. which are of the three-terminal type. These are
Bi = (1124) = (1142) = (4211) = (2411)
R2= (2231) = (2213) = (1322) = (3122)
tf3= (3342) = (3324) = (2433) = (4233) and
I?4= (4431) = (4413) = (1344) = (3144).
(See appendix on terminals and contacts for notation)
The purpose of the separation and synthesis of the effects of these six resistances
is to obtain analytical solutions of complicated networks composed of linear con¬
ductors and nonlinear four-terminal conductors. Normally, little or nothing
would be gained by separating the effect either individually or collectively of
the terminal resistances of the four-terminal conductors from the effect of the
resistances of the linear conductors connected to the terminals. Therefore the
only separations which will be considered are (a) of the effect of the direct resist¬
ances from the combined effect of the cross resistances, the terminal resistances,
and the resistances of the linear conductors; and (b) of the effect of the cross
resistances from the combined effect of the direct resistances, the terminal resist¬
ances, and the resistances of the linear conductors. Since the solution must
contain the combined effect of the terminal resistances and the resistances of
linear conductors, in general it is necessary to determine three effects. These
three effects may be determined by considering (1) that each nonlinear four-
terminal conductor, constituting a part of a network, has four linear resistances
arranged as shown in figure 23 (a), (2) that each nonlinear four-terminal conductor
195/279
Journal of Research of the National Bureau of Standards [voi.ss
has five linear resistances arranged as shown in figure 23 (b), and (3) that each
nonlinear four-terminal conductor has five linear resistances arranged as shown
in figure 23 (c).
When in the analysis of a network problem any of the well-known procedures
is followed in the treatment of the linear conductors, the first consideration gives
a partial solution which is complete, except that it contains neither the effect of
the direct resistances nor the effect of the cross resistances of the nonlinear four-
terminal conductors. The second consideration gives a partial solution which is
complete, except that it does not contain the effect of the cross resistances of the
nonlinear four-terminal conductors. The third consideration gives a partial
solution which is complete, except that it does not contain the effect of the direct
resistances of the nonlinear four-terminal conductors.
The second partial solution minus the first partial solution gives the effect of
the direct resistances, that is, the separation (a); while the third partial solution
minus the first partial solution gives the effect of the cross resistances, that is, the
separation (b). The addition of the effect of the direct resistances and the effect
of the cross resistances to the combined effect of the terminal resistances and
resistances of linear conductors constitutes a synthesis of these three effects, and
consequently gives a complete solution of the network problem. Obviously, the
complete solution may be obtained by the addition of the second and third, and
subtraction of the first of the partial solutions. If the second or third of these
partial solutions contains the effect of the direct resistances or of the cross resist¬
ances as additive terms only, that is, if the separation (a) or the separation (b)
occurs more or less automatically, the transfer of the additive terms occurring
in one of these partial solutions to the other of these partial solutions gives the
complete solution, so the first partial solution is not required. In cases in which
the combined resistances of terminals of the nonlinear four-terminal conductors
and linear conductors connected to the terminals are sufficiently high relative to
the four-terminal resistances of the nonlinear four-terminal conductors that first
order approximations of the effects of the four-terminal resistances are sufficient,
the effect of the direct resistances can always be obtained as additive terms in
the second partial solution, and the effect of the cross resistances can always be
obtained as additive terms in the third partial solution.
To show how this method of analysis works out in practice, consider that three
linear conductors having resistances of 0.43, 1, and 1.37 ohms respectively, are
connected to a nonlinear four-terminal conductor, Q, as shown in figure 24 (a),
that the six independent resistances of Q are
R1 = .07 ohm,
R2 = .13 ohm,
R3—.O5 ohm,
i?4 = .03 ohm,
D =.08 ohm, and
C = .03 ohm,
and that the problem is to find first the potential drop from M to 3 with 1 ampere
in the circuit and second the resistance of that part of the circuit included between
M and N. To more definitely visualize these problems, it may be assumed that
Q consists of a circular piece of sheet metal to which terminal posts are soldered.
196/280
Wenner] Resistance Comparisons
The first partial solution is obtained from the arrangement of linear conductors
in figure 24 (b). For this arrangement the resistance between M and the junction
R\, Ri, Rz, and Rt (see fig. 23, a) is (,5X1.5)/2, or .3750 ohm. Therefore,
since the current is 1 ampere, the potential drop from M to 3, E, = .3750 volt.
The second partial solution is obtained from the arrangement of linear conduc¬
tors shown in figure 24 (c). If E2 represents the potential drop from M to 3 for
a, a complete circuit in which Q represents the nonlinear four-terminal conductor; 6, a circuit showing the
terminal resistances of 0 as linear resistances; c, a circuit showing the terminal resistances and the direct
resistance of 0 as linear resistances; and d, a circuit showing the terminal resistances and the cross resist¬
ance of Q as linear resistances.
197/281
Journal of Research oj the National Bureau oj Standards [voi.ss
it is readily seen that the current in the upper branch is (.43 + .07)/(.43 + .07+.03 +
.13+1.37) = .2463 ampere. This current multiplied by 1.5, the resistance of
the upper branch between M, and junction of R2, C, and R3 (see fig. 23, c) gives
E2 = .3695 volt. Therefore, the potential drop from M to 3 of the actual circuit,
E= E2-\- E3—Ei = .4495 volt.
Since in the second partial solution the effect of the direct resistance of Q ap¬
peared as an additive term, namely 0.08 volt, the first partial solution might have
been omitted and the complete solution obtained by transferring this additive
term to the third partial solution. This would have given E=E3 + .08=.4a95
volt.
Considering now the second problem and letting rh r2 and r3 be the resistances
of that part of the circuit between M and A7'for the arrangements shown in figures
23 b, c, and d, it will be seen that the effect of the direct resistance, which is 0.08
ohm, will appear in the solution for r2 as an additive term with a coefficient of
unity. Therefore, it is not necessary to solve for rx or r2.
Referring to figure 23 (d), it is readily seen that
The values stated for the six independent resistances of the nonlinear four-
terminal conductor, Q, may be realized by the use of eight linear conductors
having the resistances and arrangement
.4 XL shown in figure 25. If therefore these eight
linear conductors are substituted for the
nonlinear four-terminal conductor, the
problems just considered become problems
in linear networks, and the solutions given
may be checked by well-known methods.
A somewhat more complicated problem
is a determination of the effects of the three
different current distributions in each of
the four terminal blocks involved in chang¬
ing three resistance coils from a series to a
parallel connection and in measuring the
resistances of the paralleling connections.
A solution of this problem would serve to
give a better idea as to how the method of
analysis under consideration works out in
Figure 25.—Eight linear conductors practice, and at the same time give an
so arranged, and having such resist- answer to a question that frequently arises
ances as to be equivalent to the non- in connection with precise resistance com-
linear four-terminal conductor, Q, of parisons.
figure 24 (a). Referring to figure 14, the problem is
to find the ratio of the four-terminal
resistance of the three 150-ohm resistance sections with the paralleling connectors
Li and L2 removed to their four-terminal resistance with these connectors as
shown, using in each case terminals 1 and 2 of terminal block a and terminals 1
and 2 of terminal block d as the four terminals. The resistances of the paralleling
connections L\ and L2 are to be measured as four-terminal conductors. In
the measurement of Lu L2 is removed and terminals 1 and 2 of terminal block
a and terminals 1 and 2 of terminal block b are used as the four terminals. In the
measurement of the resistance of L2, L\ is removed and terminals 1 and 2 of
terminal block c and terminals 1 and 2 of terminal block d are used as the four
terminals.
It will be assumed that all conductors are linear except the terminal blocks
a, b, c, and d; and that each of these is a nonlinear four-terminal conductor. Aside
from what is obvious, the problem resolves itself into a determination of the
effects of the direct and cross resistances of the terminal blocks a, b, c, and d.
With reference to the terminal resistances of the terminal blocks, it will be
seen by reference to figure 14 that Rs of each terminal block is included in the
measurement of Lx or L2, R\ of terminal blocks a and d are outside the resistances
198/282
Warner] Resistance Comparisons
0 b
Figure 26.—Two arrangements of linear resistances for determining the the effect
of three different current distributions in each of the four nonlinear four-terminal
blocks shown in figure IS.
consideration of the arrangement shown in figure 26(a), and the third partial
solution may be obtained from a consideration of the arrangement shown in
figure 26 (b). Here Da, Di, Dc, and Dd represent the direct resistances and
Ca, Cb, Cc, and Cd represent the cross resistances of the terminal blocks, a, b, c,
and d.
Let
[i2s]2 = the four-terminal resistance of the entire combination with the series
connection, excepting that part contributed by the cross resistances
Ca, Cb, Cc, and Cd of the terminal blocks.
[ijp]2=the four-terminal resistance of the entire combination with the parallel
connection, excepting that part contributed by the cross resistances.
[Xi]2 and [X2]2=the resistances of the paralleling connections, excepting that
part contributed by the cross resistances.
[f2s]3=the four-terminal resistance of the entire combination with the series con¬
nection, excepting that part contributed by the direct resistances,
Da, Db, De, and Dd.
[-KP]3 = the four-terminal resistance of the entire combination with the parallel
connection, excepting that part contributed by the direct resistances.
[Xi]3 and [X2J3=the resistances of the paralleling connections, excepting that
part contributed by the direct resistances.
199/283
Journal of Research of the National Bureau of Standards [voi. ?s
[X2]2=L2-Dc-Dd. (86)
[72s]3==450, (89)
It is readily seen that the resistance 72j, of terminal block a is higher by 2C0/3
and the resistance 72b of terminal block d is higher by 2C<j/3 with the parallel
connection than with the series connection. In fact, this is the reason why Ca
and Cd appear in eq 90.
Obviously the effects of the different current distributions in each of the four-
terminal terminal blocks depend on their four-terminal resistances. Further¬
more, by design or adjustment, these four-terminal resistances may be made
exceedingly small. As an illustration of a design having exceedingly small four-
terminal resistances consider an equilateral tetrahedron, using the apices as the
terminals. Also, by design, it is possible to make the four-terminal resistances
definite to almost any extent desired. In cases in which the four-terminal resist¬
ances are sufficiently small to be neglected, and are equally definite, a nonlinear
four-terminal conductor may be considered as equivalent to four linear con¬
ductors connected in star.
In the solution of network problems, it is frequently possible to select for the
independent variable any one of two or more quantities which are dependent on
each other. Making different selections leads to different forms. Then, by
making approximations, still other forms are obtained of the solution. As an
illustration of exact expressions for the current in the galvanometer branch of an
unbalanced bridge consider eq 76, 78, 79, and 81. The second of these involves
the drop in potential across the bridge with the galvanometer branch open, the
third the drop in potential across the bridge with the galvanometer branch
200/284
Wenner] Resistance Comparisons
closed, the fourth the current through the bridge, while the first, that is the
usual classical solution, involves the electromotive force in the battery branch.
The second of these involves the resistance of the battery branch in a rather com¬
plicated way, as does also the. classical solution. Therefore, if only these four
were available it is conceivable that cases might arise in which it would be more
convenient to use either the third or the fourth rather than the second or the
classical form of solution. In many cases approximate solutions may be obtained
in forms much more convenient to use than any of the forms of exact solutions,
yet giving all the accuracy required.
572772 0-61-14
201/285
Journal oj Research of the National Bureau of Standards [voi.ts
202/286
Warner] Resistance Comparisons
the total current) following the establishment or reversal of the test current. In
still other cases it appears that there is a gradual mechanical yielding of the
insulating material under the stresses caused by the temperature difference be¬
tween the resistance material and its mechanical support. In such cases the
effect of the test current upon the resistance may be larger at the end of 15 seconds
than at the end of a longer time.
Changes in resistance resulting from the test current, whether or not the way
in which they are brought about is known, and provided they are not proportional
to the square of the current, will be referred to as the secondary effects of the
current upon the resistance. There are many possible causes of secondary effects,
so it is not possible in all cases to obtain as complete information concerning them
as might be desired.
However, correlation of the secondary effects with details of construction of
many resistors makes it possible to predict with some degree of certainty their
magnitudes in others in which they are too small to be observed. For relatively
few resistance coils and standard resistors under normal conditions of use is it
probable that the secondary effects of the current upon the resistance amount to
less than 1 in 100 million. On the other hand, for relatively few standard resistors
having resistances in the range from 0.0001 to 10,000 ohms, if of the precision
ty-pe and of good quality, are the secondary effects of the normally used test
currents in excess of 1 in 100,000.
Question 2.—It appears that the first precise measurements having a direct
bearing on the relation between a change in current from one steady value to
another steady value and the resulting
change in potential drop were made by
F. A. Wolff of this Bureau. At a meet¬
ing of the American Association for the
Advancement of Science in the summer
of 1899 he presented a paper entitled
“Experimental Test of the Accuracy of
Ohm’s Law.” Unfortunately, no further
record of this investigation has been
found either in the form of a published
paper or unpublished notes or data.
The circuit used was essentially that
shown in figure 27. Here xu x2, x3, and
x4 represent the four arms of a balanced
Wheatstone bridge. This bridge consti¬
tutes the X arm of a second Wheatstone
bridge, the other arms of which are Figure 27.— Wolff’s circuit for testing
designated Y, A, and B. This bridge is the accuracy of Ohm’s law.
also balanced. Dr. Wolff refers to the
bridge whose arms are designated xu x2, x3, and x4 as the auxiliary bridge, the other
bridge as the main bridge, and the circuit as a bridge within a bridge.
This method was tried about 1920 [105], when it gave promise of reasonably
satisfactory results, though some difficulties were encountered. These difficulties
were, for the most part, obviated by replacing the battery of the auxiliary bridge
by a source of alternating current and omitting a provision for balancing the
auxiliary bridge. Standard resistors having nominal values of 10 ohms were used
as the xu x2, x3, and x4 arms of the auxiliary bridge and the Y arm of the main
bridge. The direct current in the Y arm was adjusted to a value somewhat larger
than that regularly used in testing 10-ohm standard resistors and the alternating
current in the supply branch of the auxiliary bridge was adjusted to a root mean
square value equal to approximately twice the value of the direct current in the
Y arm of the main bridge.
Since xu x2, x3, and x4 were approximately equal in resistance very little of the
alternating current passes through the A, B, and Y arms, or through the battery
or galvanometer branch of the main bridge. Presumably, therefore, the alternat¬
ing current had no appreciable effect except in xu x2t x3, and x4, the arms of the
auxiliary bridge. Following the usual procedure in balancing the main bridge,
the precision of the balances was about 1 part in 10 million. On opening the
connection to the source of the alternating current, the balance was changed by
an amount not perceptibly different from that corresponding to the change in
resistance of the series parallel combination of x4, x2, x3, and x4 resulting from the
reduced power dissipation.
203/287
Journal of Research of the National Bureau of Standards [Voi. 25
204/288
Wenner] Resistance Comparisons
Standard resistors from Germany, France, and England were compared with
standard resistors of the United States from time to time during a period of about
6 weeks. Values were then assigned to all the standard resistors used in the inter-
comparisons in terms of the new unit. In assigning the new values, it was as¬
sumed that the mean value of one of the standard resistors from the National
Physical Laboratory during the time of the intercomparisons was the same as the
value previously assigned in the National Physical Laboratory, and that the
mean of the values of the resistances of the two standard resistors from the
Physikalisch-Technische Reichsanstalt during the time of the intercomparion was
the same as the mean of the values previously assigned in the Physikalisch-
Technische Reichsanstalt. Equal weights were given to the previously assigned
value in the National Physical Laboratory and the mean of the previously as¬
signed values in the Physikalisch-Technische Reichsanstalt.
Following the assignment of new values to the standard resistors of this Bureau
used in the intercomparison, they were intercompared either directly or indi¬
rectly with many other standard resistors of this Bureau. From time to time
since then, selected groups (usually of 20 or more) 1-ohm standard resistors have
been intercompared, and on the basis of relative changes and previous good per¬
formances a group of 10 standards has been selected from the larger group. In
assigning new values to all the standard resistors of the larger group and to others,
it has been assumed that the mean resistance of the group of 10 had not changed
since the previous similar intercomparison. This in general is the procedure by
which the unit established in 1910 has been maintained in this Bureau.
From 1910 to 1929 only standard resistors of the sealed type developed by
Rosa [76] were included in the groups of 10. A total of 20 different standards
have been included in the groups of 10, while 3 of these 20 have been included
in every group of 10. From 1932 to 1938 only standards of the smaller double-
walled type [89] were included in the groups of 10. More recently, only standards
of the larger double-walled type constructed by Thomas in 1933 have been in¬
cluded in the groups of 10.
The unit established in 1910 and maintained in this Bureau in the way described
above has been referred to by different names. The name now being used is the
NBS International Ohm.
Obviously, a unit of resistance maintained in this way changes with time to
the extent of the proportional decrease (or increase) of the resistances of the
standard resistors used in its maintenance. Therefore, it is not to be presumed
that the NBS International Ohm has remained constant during the time since
its establishment. To obtain information on this point various investigations
have been made in this Bureau. Most of those since 1927 have been made by
Thomas. In addition, groups of standard resistors have been sent abroad for
measurements of their resistances in other national laboratories, and standards of
other national laboratories have been measured in this Bureau. More recently
these international intercomparisons have been made at the International Bureau
of Weights and Measures.* Comparisons of the values found for the resistances
of the same standard resistors in different national laboratories have given the
relations between the units of resistance of the different countries. From infor¬
mation as to the manner in which the units of the different countries were estab¬
lished and have been maintained, the results of the international intercomparisons
and the investigations made in this Bureau, it may be concluded that the probable
change in the NBS International Ohm during the 30 years since its establishment
has been less than 30 parts in 1 million. Since the standard resistors now being
used in this Bureau in the maintenance of the units are fully one order better
than those formerly used for this purpose, it is reasonable to presume that the
present rate of change of the NBS International Ohm is less than 1 part in 1
million per year.
The relation between the NBS International Ohm, the mean international ohm,
and the units of resistance of other countries is given in table 1, which contains
data taken from the 1933, 1935, and 1937 reports of the International Committee
of Weights and Measures (Comit4 International des Poids et Mesures, Proc4s-
Verbaux des Stances) and Comptes Rendus, page 24, volume 209, 1939.
205/289
Journal oj Research oj the National Bureau oj Standards [voi.es
1 In this table a plus sign signifies that the unit of the country was larger than the Mean International
Ohm.
The Mean International Ohm was defined as the mean of the units of electrical
resistance, in 1935, of Germany, United States, Great Britian, Japan, and Russia.
After this mean was obtained, it was taken by France as the new French unit
and by Russia as the new Russian unit. The data given in the table are based
on the assumption that units of resistance are being maintained with equal
fidelity in each of the six countries.
The unit of the United States is the unit of resistance referred to above as the
NBS International Ohm. Therefore, in the early part of 1939,
1 NBS International Ohm = 0.999 997 Mean International Ohm.
From measurements made in this Bureau by Wenner, Thomas, Cooter, and
Kotter, the results of which were reported to the International Bureau of Weights
and Measures in December 1938,
1 NBS International Ohm= 1.000 485 absolute ohm.
From measurements made in this Bureau by Curtis, Moon, and Sparks, the
results of which were reported to the International Bureau of Weights and Meas¬
ures in December 1938,
1 NBS International Ohm= 1.000 484 absolute ohm. In each case the un¬
certainties in the measurements were estimated to be 2 parts in 100,000.
In 1937 the International Committee on Weights and Measures adopted pro¬
visionally the relation
1 Mean International Ohm =1.000 48 absolute ohm.
From this relation and the data given in table 1, it follows that provisionally
1 NBS International Ohm= 1.000 477 absolute ohm
as of February 1939.
APPENDIX 8. REFERENCES
(a) AUTHORS
The letters included in brackets at the end of each reference indicate the sub¬
jects discussed. The key to these letters is given under the heading “Subjects”
p. 293.
[1] R. Appleyard, Proc. Phys. Soc. (London) 14, 243 (1895-96) [P].
[2] W. E. Ayrton and T. Mather, Electrician 32, 627 (1893) [G].
[3] A. C. Becquerel, Ann. Chim. 32, 425 (1826) [D].
[4] A. Blondel, Ann. phys. [9] 10, 219 (1918) [G].
[5] P. W. Bridgeman, Proc. Am. Acad. Arts Sci. 57, 129 (1922) [I].
[6] H. B. Brooks, BS J. Research 4, 297 (1930) RP150 [F].
[7] H. B. Brooks, BS J. Research 11, 211 (1933) RP586 [O].
[8] H. B. Brooks and A. W Spinks, BS J. Research 9, 781 (1932) RP506 [O].
[9] H. L. Callendar, Proc. Phys. Soc. (London) 22, 220 (1909-10) [K].
[10] G. Carey-Foster, Teleg. Eng. J. 1, 196 (1872-13) [S],
[11] C. D. Child and O. M. Stewart, Phys. Rev. 4, 502 (1897) [K].
206/290
Wcnner] Resistance Comparisons
207/291
Journal of Research of the National Bureau of Standards [voi. s&
[68] A. Paalzow and H. Rubens, Ann. phys. [3] 37, 529 (1889) [K],
[69] H. A. Perkins, Phvs. Rev. 18, 131 (1921) [I],
f70] H. A. Perkins, J. phys. [6] 4, 246 (1923) [I],
[71] Lord Rayleigh, Proc. Roy. Soc. (London) 49, 203 (1891) [C, K].
[72] Lord Rayleigh, Trans. Roy. Soc. (London) 173, 697 (1882) [L],
[73] E. H. Rayner, Proc. Phys. Soc. (London) 27, 384 (1914-15) [M],
[74] J. H. Reeves, Proc. Phys. Soc. (London) 14, 166 (1895) [B, P],
[75] H. Rohmann, Phvsik. Z. 14, 203 (1913) [F],
[76] E. B. Rosa, Bui. BS 5, 413 (1908) S107 [R].
[77] A. Rosen, J. Inst. Elec. Eng. 62, 916 (1924) [N],
[78] A. Schuster, Phil. Mag. [51 39, 175 (1895) [KR
[79] L. Schwendler, Phil. Mag. [4] 31. 364 (1866) [K],
[80] G. F. C. Searle, Electrician 66, 999 (1911) [E].
[81] G. F. C. Searle, Electrician, 67, 12 (1911) [M, P],
[821 F. E. Smith, Electrician, 57, 976 (1906) [B, D, H, J, K, L, S, U].
[83] F.. E. Smith, Electrician, 57, 1009 (1906) [K, L, P, S].
[84] F. E. Smith, Dictionary of Applied Physics, R. Glazebrook edition (Mac¬
millan, London, 1922) 2, 718 (1922) [K, L, M, P, S],
[85] S. G. Starling, Electricity and Magnetism (Longmans, London, 1912)
[86] H. von Steinwehr, Handbuch der Physik (Springer, Berlin, 1927) 16, 445
[D, F, J, K, U],
[87] F. Streintz, Ann. Physik [4] 3, 1 (1900) [S].
[88] L. Thevenin, Comptes rendus 97, 159 (1883) [C].
[89] J. L. Thomas, BS, J. Research 5, 295 (1930) RP201 [R].
[90] J. L. Thomas, BS., J. Research 12, 313 (1934) RP657 [R].
[911 J. L. Thomas, J. Research NBS 13, 681 (1934) RP737 [R].
[921 J- L. Thomas, J. Research NBS 14, 589 (1935) RP789 [R].
[93] J. L. Thomas, J. Research NBS 16, 149 (1936) RP863 [R].
[94] W. Thomson, Phil. Mag. [4] 24, 156 (1862) [K, P],
[95] W. Volkman, Ann. Physik [4] 10, 217 (1903) [G],
[96] K. W. Wagner, Elektrotech. Z. 32, 1001 (1911) [T].
[97] K. W. Wagner, Elektrotech. Z. 33, 635 (1912) [T].
[98] C. W. Waidner, H. C. Dickinson, E. F. Mueller, and D. R. Harper, 3d.
Bui. BS 11, 575 (1915) RP241 [RU].
[99] F. Wenner, Phys. Rev. 17, 384 (1903) [S],
[100] F. Wenner, Bui. BS 6, 347 (1909) RP134 [C, F, K],
[101] F. Wenner, Bui. BS 8, 559 (1912) RP181 [B, E, F, K, P, S].
[1021 F. Wenner, J. Wash. Acad. Sci. 2, 114 (1912) [R],
[103] F. Wenner, BS Sci. Pap. 12, 469 (1915) S258 [S].
[104] F. Wenner, Bui. BS 13, 211 (1916) S273 [F],
[105] F. Wenner, Phys. Rev. [2] 15, 531 (1920) [T].
[106] F. Wenner, Phys. Rev. [2] 23, 780 (1924) [I],
[107] F. Wenner, J. Opt. Soc. Am. 11, 495 (1925) [G].
[108] F. Wenner, BS Sci. Pap. 21, 191 (1926) S531; Proc. Phys. Soc. 39, 124
(1927) [C],
[109] F. Wenner, N. L. Forman, and A. R. Lindberg, Phys. Rev. [2] 20, 589
(1922) [II.
[110] F. Wenner and J. L. Thomas, BS J. Research 12, 147 (1934) RP639 [R].
[111] F. Wenner and A. Smith, BS Sci. Pap. 19, 297 (1924) S481 [H],
[112] F. Wenner, E. H. Smith and F. M. Soule, BS J. Research 5, 711 (1930)
RP223 [U].
[113] F. Wenner and F. M. Soule, Phys, Rev. [2] 23, 780 (1924) [I].
[114] F. Wenner and E. Weibel. Bui. BS 11, 1 (1914) S 223 [S],
[115] F. Wenner and E. Weibel, Bui. BS 11, 65 (1914) S225 [B, S].
[116] F. Wenner, E. Weibel, and F. C. Weaver, Phvs. Rev. [2] 3, 497 (1914) [F],
[117] C. Wheatstone, Trans, Roy. Soc. (London) 133, 303 (1843) [U].
[118] C. Wheatstone, Pogg. Ann. 62, 535 (1844) [U].
[119] W. P. White, Phvs. Rev. 19, 305 (1904) [F].
[120] W. P. White, Z. Tnstrumentenk. 34, 71 (1914) [A, G, O].
[121] F. A. Wolff, M. P. Shoemaker, and C. A. Briggs, Bui. BS 12, 449 (1915-16)
S256 [B, H, P],
[122] A. Zeleny, Phys. Rev. 32, 297 (1911) [F],
[123] F. Zernike, Koninkl. Akad. Wetenschappen Amsterdam 24, 239 (1922) [F].
[124] L. A. Zurcher and A. C. Seletzky, Trans. Am. Inst. Elec. Engrs., 1939 [Kj.
208/292
Winner] Resistance Comparisons
(b) SUBJECTS
The numbers included in brackets at the end of each subject refer to publica¬
tions in which the subject is discussed. The key to these numbers is given under
the heading “Authors.”
[A] Adjustable resistance elements. [63, 120].
[B] Balancing of resistances of connectors. [45, 74, 82, 101, 115, 121].
[C] Current distribution in a linear network. 129, 30, 49, 100, 108].
[D] Differential galvanometer method. [3, 31, 33, 34, 39, 40, 50, 51, 52, 54, 65,
82, 86].
[E] Four-terminal conductors. [25, 35, 58, 80, 101].
[F] Galvanometers. [6, 14, 15, 28, 37, 38, 40, 41, 43, 44, 54, 56, 59, 61, 62, 65,
75, 86, 100, 101, 104, 116, 119, 122, 123],
[G] Galvanometer shunts. [2, 4, 17, 44, 65, 95, 107, 120].
[H] Measurement of resistances of connectors. [44, 46, 55, 82, 111, 121].
[I] Ohm’s law. [5, 13, 57, 66, 67, 69, 70, 105, 106, 109, 113].
[jj Potentiometer method. [40, 60, 82, 86],
[K] Sensitivity of bridges. [9, 11, 19, 20, 21. 23, 25, 26, 28, 29, 32, 39, 40, 42, 51,
58, 65, 68, 71, 78, 79, 82, 83, 84, 85, 86, 94, 100, 101, 125].
[L] Series-parallel build up. [24, 72, 82, 83, 84].
[M] Shunting of standard resistors. [55, 73, 81, 84].
[N] Star-polygon transformations. [48, 53, 77].
[01 Thermoelectromotive forces. [7, 8, 22, 27, 29, 120],
[P] Thomson bridge. [1, 19, 21, 25, 36, 45, 46, 54, 55, 65, 74, 81, 83, 84, 94, 101,
121],
[Q] Shielding. [16, 47],
[R] Standard resistors. [76, 89, 90, 91, 92, 93, 98, 102, 110].
[Sj Various methods and their modifications. [10, 18, 19, 25, 54, 58, 64, 65, 82,
83, 84, 85, 87, 99, 101, 103, 114, 115].
[T] Wagner branch. [16, 29, 47, 96, 971.
[U] Wheatstone bridge. 112, 29, 47, 82, 86, 98, 112, 117, 118],
209/293
Journal oj Research of the National Bureau of Standards [voi. 26
XII. INDEX
Page Ohm’s law—Continued. Page
Adjustable resistance elements_ 245 separation of potential drop caused by cur¬
Adjustments of Thomson bridge__255 rent from total potential drop_288
Amalgamated contacts Wolff’s method of testing . _ 287
procedure in maintenance of low resistance_264 Optical system of galvanometer
Bridge equations. 248 description . . . _ _ 274
approximations invoiced .. _ 249 a factor in the effective sensitivity__276
limitations with Thomson bridge method_250 Precision and accuracy.. _ 256
Comparisons of standards Resistance measured by Thomson bridge
of different nominal values_ 251 method .265
Four-terminal conductors Resistanceof two terminal conductor: definit ion . 257
cross resistance. . —- 259 Resistances of terminal blocks; analysis of effect. 283
definition of four terminal resistances..259 Resistance ratios 252
design for exceedingly low four-terminal apparatus for establishing 10 to 1 ratio_253
resistances . 284 procedure for establishing 2 to 1 ratio..255
diagonal resistance_259 ratio of resistance of n standards connected in
direct resistance ......_ 259 series to the resistance of the same standards
equivalent to eight linear conductors_282 connected in parallel_ 252
six independent resistances. 279 Sensitivity of bridges:
requirement for definite direct resistance_259 dependent on electromotive force developed
requirement to be classed as linear.. . 259 by relative motion of coil and magnet of
Four-terminal resistances of four-terminal con¬ galvanometer ..._ ..... ._ 232
ductor dependent on external damping resistance of
conditions for one or more to be definite_259 galvanometer ... ... . 235
definition. 259 dependent on procedure followed in use of
measurement by Thomson bridge method . .. 265 bridge ..._ _ .. _ __239
measurement by Wheatstone bridge method.. 254 dependent on quality of optical system of gal¬
number ____- 259 vanometer . . . _ ... ... _276
number which are independent... 259 dependent on sensitivity of galvanometer_234
relation between_ 259 doubled by reversing test current _ _239
Galvanometer limited by heating effect of the test current.. 236
description_ 244 effective sensitivity .. 272
operating characteristics_ 244 maximum permissible sensitivity_ 238
Ouard of Thomson bridge _ 277
description __ 268 of Wheatstone bridge_233
how used in correcting for effect of slight de¬ reduced by shunting galvanometer_ 235
fects in th-insulation ... __ 270 Standard resistors: quality used in maintenance
how used in testing the insulation_271 of unit of resistance_ 257
Inductance Substitution method_ 248
method of compensating_ 245 Terminals and contacts of standard resistors_258
Interchange method..___ 251
Insulation Terminal blocks
means employed for maintaining--- 268 of auxiliary apparatus__ 261
procedure in case of slight defects-270 of bridge. _260
quality required. _ . _ 268 Thermal shunts_ 267
theoretical correction for slight defects..269 Thermoelectric balances... 267
test for slight defects. 271 Thermoelectromotive forces
Load coefficients effect principally eliminated by reversal of
definition. _ 241 test current . 239
determination_ 242 means of mitigating effect_ 266
limits sensitivity of bridges . 2?6 Thomson bridge
Matthiessen and Hockin bridge method_265 adjustments _ 255
Methods of analyses resistances compared by. 265
hypothetical battery of high resistance and sensitivity _239
high electromotive force .. --- 278 Units of resistance_ 288
hypothetical battery of zero resistance_ 277 mean International ohm
hypothetical opening of a branch of a network 277 definition .. 290
separation and synthesis of the effects of the tentative relation to absolute ohm..290
independent resistances of a four-terminal NBS Intel-national Ohm
conductor. _.. _ 279 how established. 288
NBS precision bridge; description__ 244 how maintained_ 289
Ohm’s law_ 285 NBS International Ohm and Absolute Ohm
accuracy of formulae expressing resistances relation as determined in this Bureau ... 290
of combinations of conductors in terms of relative values of units of different countries.. 290
their individual resistances __ 288
conclusions based on tests made by Chrystal 285 Wagner branches
more important parts .... 285 connected in parallel with battery branch_273
primary effects of test current_ 286 connected in parallel with galvanometer
proportionality of potential drop and current. 287 branch_ 274
questions concerning_ _286 Wheatstone bridge: different forms of the solu¬
secondary effects of test current_ 286 tion of the network.... 277
210/294
U. S. Department of Commerce National Bureau of Standards
abstract
The apparatus and method used in the precise measurement of the ratio of volt
boxes at the National Bureau of Standards are described in detail. The method
is applicable with rated voltage applied to the volt box and has an accuracy ap¬
proaching one part in 100,000. The results of such measurements on a number of
volt boxes under various conditions of humidity and temperature are given as a
guide to their use and an index of their accuracy.
CONTENTS pag8
I. Introduction_269
II. Method_ 270
III. The standard volt box_272
IV. The Lindeck potentiometer_276
V. Auxiliary features_277
VI. Test procedure_.-278
VII. Typical performance characteristics of volt boxes_278
1. Accuracy of manufacturer’s adjustment of ratio_279
2. Secular change_ 280
3. Seasonal variation of ratio_280
4. Effect of surface resistance of insulating panel on the ratio of
volt boxes_282
5. Effect of self-heating_ 282
VIII. Appendix I. Alternative test method_283
IX. Appendix II. Derivation of test schedule for calibrating the standard
volt box_285
I. INTRODUCTION
211/269
Journal oj Research of the National Bureau oj Standards [Voi. 27
212/270
SiZs&ee,]
QT0S3 J Testing and Performance of Volt Boxes
Figure 1.—Diagram of circuit used to intercompare volt boxes by the first method.
213/271
Journal oj Research of the National Bureau of Standards [Voi. 27
1,000 microvolts) unless the volt box under test is seriously out of
adjustment. Consequently in measuring the small potential differ¬
ences, v, and the applied voltage, V, only a moderate relative accuracy
is sufficient to give a high precision in Fx. The applied voltage,
V, is conveniently measured with a voltmeter, but the small potential
differences, v, should be measured with a low-range potentiometer.
The three sections which follow describe in some detail the main
components of the equipment used in this method at the National
Bureau of Standards.
214/272
Journal ol Rcseari li ol the National Bureau ol Standards Research Paper I4ll>
215
Gross*'] Testing and Performance of Volt Boxes
A, Metal guard ring connected to guard circuit. 7, A resistance coil of the guard circuit.
B, A terminal of the standard volt box. J, Air-tight metal box.
C, Bakelite top panel. K, Sweated air-tight joint.
7), Insulator bushing. L, Rubber gasket.
E, Brass support for panel F. E is connected to the AT,Metal angle.
guard circuit. N, Brass rod, which extends to bottom of 7. A tap
F, Hard rubber panel. lead of guard circuit connects to bottom of N.
0, Two resistance coils of the working circuit. 0, Brass rod extending to bottom of O. A tap from
77, A guard-circuit terminal. the working circuit connects to bottom of 0.
216/273
DEPARTMENT OF COMMERCE
NATIONAL BUREAU OF STANDARDS
Washington 25, D. C.
Errata to Accompany
216a/274
oloss*’] Testing and Performance of Volt Boxes
using such an improvised arm have shown that the ratio of the 50,000-
ohm sections to the 500-ohm section is the same at 300 volts and at 150
volts as at 30 volts, within 3 parts per million. Routine calibrations
are therefore made at only 30 volts. In the case of the higli-voltage
sections, the measurement should be made with the corresponding
section of the guard circuit energized by connecting it to the battery
which supplies the bridge current, as shown in figure 4 (A).
Table 1 (see p. 287) is the schedule used for computing the final
volt-box corrections from the bridge readings. In the table each value
of p is the observed resistance of the precision rheostat in the bridge
circuit of figure 4 (A and B). The computed values of s in column 7
are the corrections in parts per million which, when applied to the
nominal ratio of the resistance of the range in column 8 to the resist¬
ance of the 0 to 0.15 range, will give the actual ratio. To obtain the
actual ratio for any range in column 8 to any other range listed in that
column, subtract the value of s for the second range from that of the
first and apply this difference as a correction in parts per million to the
nominal ratio of the two ranges. The mathematical derivation of the
equations of table 1 is given in appendix II.
The actual carrying out of a test by this procedure is very rapid, a
complete test and computation for all ranges being obtainable in less
than 3 hours. The sole theoretical error in the procedure as described
arises from the resistance of the rods which lead from the junction
points of the sections to the binding posts. The resistance of each
rod is only 0.0006 ohm. If the rods are equal in resistance, the error
will be equal to the resistance of two rods divided by the total resist¬
ance of the volt-box range used. Differences in rod resistance will
cause errors equal to these differences divided by the resistance of the
section measured. To eliminate this latter error in the measurements
of the 50-ohm sections, the galvanometer and battery connections of
the bridge circuit are made through the sections next adjacent to the
one under test, as shown in figure 4 (B), and thus the rods are put into
arms of higher resistance. The effect of the rod resistance on the volt-
box ratio is less than 3 parts per million.
As a measure of the precision readily attainable in this process of
calibration, data obtained over a period of 12 years were examined
and 27 successive independent determinations of the ratio of two
sections of identical construction, namely that between the 450- and
600-volt taps and that between the 600- and 750-volt taps, were
tabulated. The standard deviation of these values from their mean
was found to be 7 parts per million, the greatest individual deviation
being 19 parts per million. This deviation, of course, includes any
drift in the relative values of the coils involved as well as the errors
in the setting up of their ratio.
The individual coils are wound with a single layer of silk-covered
manganin wire, all coils being wound with wire from the same spool
so as to minimize differences in the temperature coefficient of resistance
of the various parts. Nevertheless, early data showed a tendency to a
cyclical annual fluctuation in the ratio of the 750-volt to the 1.5-volt
section amounting to about 30 parts per million. This is apparently
the result of changing atmospheric humidity. To minimize such
effects, a small air pump is arranged to maintain the average pressure
within the standard volt box at 15 mm of water above atmospheric.
Air from the pump passes through a drying vessel containing CaCl2
j
and is thus thoroughly dried before entering the box. A rubber gasket
between the metal box and the Bakclite top seals the joint sufficiently
so that an air flow of about 0.1 m3 per 24 hours is enough to maintain
the pressure differential. Since the installation of this dry-air supply,
the cyclic fluctuation has been very small.
218/276
arofse'] Testing and Performance oj Volt Boxes
V. AUXILIARY FEATURES
All of the wires in the measuring circuits are run in shielded cable,
and each shield is maintained at the same potential as the wire inside
it by means of the auxiliary “guard” circuit of the standard volt
box mentioned in section III above. The galvanometer and the
Lindeck potentiometer are mounted on metal guard plates which are
so located as to intercept any possible leakage currents. When volt¬
age Vi, v2, or v3 is being measured, these guard plates are connected
to the low-voltage terminal of the auxiliary guard circuit of the
standard volt box; but when v4 is measured, these plates are con¬
nected to the high-voltage terminal of the guard circuit. A special
plug-connector panel is used to connect the Lindeck potentiometer
circuit to measure i\, v2, v3, and v4, and this panel is designed so that
guard plates intercept leakage currents between the different measur¬
ing circuits.
For test voltages from 400 to 750 volts the batteries shown in
figure 1 are replaced by a motor-driven d-c generator. The driving
motor and the field of the generator are energized by means of
batteries.
With this equipment, 750 volts is the nominal limit at which volt
boxes are tested. However, experiment has shown that the calibra¬
tion of the standard volt-box can be relied upon up to 1,500 volts,
and by special arrangements the generator can be connected in series
with a group of batteries to obtain this voltage.
Since this method of calibrating a volt box involves the ratio of the
small voltages measured by the potentiometer to the larger voltage
measured by the voltmeter, it is necessary that the readings of these
two instruments be mutually consistent, though neither need be
accurate. Means were therefore provided for readily intercomparing
them. This is done by applying the same voltage to the voltmeter
and the standard volt box and measuring the voltage drop across the
0 to 1.5-volt range of the standard volt box with the Lindeck poten¬
tiometer. The “volts” and “common” terminals (see fig. 5) of the
potentiometer are used. Knowing the volt-box ratio, the agreement
of the two instruments can be readily checked. Since the volt-
219/277
Journal of Research of the National Bureau of Standards [voi. 27
220/278
Silsbee, ]
Gross J Testing and Performance of Volt Boxes
factors which affect the voltage ratio so that those who use volt
boxes may know what factors may affect measurements of a given
precision.
At the time of the construction of a volt box the departure of its
actual voltage ratio from the nominal value will depend upon the care
with which the manufacturer adjusts the resistances of the various
coils. Thereafter the departure of the voltage ratio from the intial
value may depend upon (a) secular change of the resistance of the
coils, (b) ambient temperature, (c) temperature changes caused by
self-heating, (d) slow seasonal variations in ratio caused by seasonal
changes in relative humidity, and (e) changes in insulation resistance
between terminals and between the coil studs.
Standards between the years 1927 and 1940. For each plotted point
in this figure the ordinate represents the percentage of volt-box ranges
tested the ratio of which was within ±0.005 percent of the ratio
indicated by the abscissa. If the volt box was tested on more than
one occasion, only the results of the initial test were used in locating
this curve. Curves of this type for individual ranges had the same
221/279
Journal of Research of the National Bureau of Standards [Voi. 27
2. SECULAR CHANGE
One would expect that the secular change of ratio would be less than
the secular change of resistance of the individual coils, provided that
all of the coils are of similar construction. Data are available on seven
volt boxes which have been sent in to the National Bureau of Stand¬
ards for annual test over periods of from 6 to 10 years. Of these seven
volt boxes, the voltage ratio of the 150-volt to the 1.5-volt range has
varied 0.01 percent, or less, in six of them, and the ratio of the seventh
has increased uniformly by a total of 0.04 percent in a 10-year period.
The other ranges of each volt box in this group varied in the same
manner as the 150-volt to the 1.5-volt range. It may be supposed
that this performance is typical-of high-grade volt boxes.
used to insulate the manganin wire had a greater length at high relative
humidity than at low, and he concluded that this property of the silk
was responsible for the fact that the resistance of silk-insulated man¬
ganin wire coils is greater at the higher humidities. He also found that
if such coils were coated with certain waxes and suddenly changed
from an atmosphere of one relative humidity to that of another, the
resistance change would require approximately 3 weeks to complete,
but that coils not coated with wax required less than 24 hours although
the magnitude of the change in resistance was tbe same. The per¬
centage change of resistance is greater the finer the wire.
Figure 7 shows the seasonal variation in ratio for a volt box in which
the wire size is much larger on the low-voltage coil than on the coils
nearer the high-voltage end. Dike’s theory explains this curve very
well, because from that theory we would expect the humid summer
3 P. H. Dike, Rev. Sci. Instr. 7, 278 (1936).
222/280
Si7,<6ee,1
Gross J Testing and Performance of Volt Boxes
01 23456789 10
DAYS
Figure 8.—Drift of voltage ratio after a sudden increase in relative humidity.
Each volt box stood for several weeks in a room having a relative humidity between 25 and 40 percent, and
was then placed in a closed cabinet in which a saturated solution of sodium chloride maintained a relative
humidity of 75 percent. Each curve represents measurements made on the 150-volt to the I.5-volt range of
a volt box, with the exception of the curve for volt box A on which the 150-volt to the 0.15-volt range was
measured.
Volt box A has a resistance of 40 ohms per volt, its wood spools are wound with silk-insulated wire, and
the diamoter of the wire on the low-voltage coils is approximately 10 times that of the wire on the nigh-
voltage coils.
Volt box B has a resistance of 200 ohms per volt, its metal spools are wound with silk-covered wire, and
the ratio of the diameter of the wire in the low-voltage coil to that of the wire in the high-voltage coil is inter¬
mediate between that of volt boxes A and C.
Volt box C has a resistance of 26% ohms per volt, its wood spools are wound with silk-covered wire, and
the diameter of the wire in the low-voltage coil is twice that of the wire in the high-voltage coils.
Volt box D has a resistance of 333 ohms per volt; its metal spools are wound with enameled wire which is
insulated with a patented mixture of silk and cotton, and the ratio of the diameter of the wire in the low-
voltage coils to that of the wire in the high-voltage coils is approximately the same as that of C.
The coiis of all four volt boxes are wax-coated.
223/281
Journal oj Research oj the National Bureau oj Standards [Voi. 27
the volt box’ from the cabinet. These ratio measurements were
made on the 150-volt range of each volt box with only 30 volts applied
so as to make any self-heating effects negligible.
Volt boxes A, B, and C have coils wound with silk-covered manganin
wire. The manganin wire of volt box D, however, is insulated with
a patented 4 combination of silk and cotton, which is claimed to cause
much less seasonal variation of resistance with humidity. The
improvement is obvious from the curves.
These curves for volt boxes in which silk-covered manganin is used
bear out two implications from Dike’s paper. First, for a given volt¬
age ratio the effect of a change in humidity will be least for the volt
box having the fewest ohms per volt, since such a box will usually
have thicker wire; this would explain the small effect on the ratio of
C in figure 8. Second, the greater the difference between the wire
diameters of the low-voltage and high-voltage coils, the greater will
be the effect of humidity. This is shown by the curves for A and B,
where A has one-fifth of the ohms per volt of B but the greater differ¬
ence in wire size.
After the data for figure 8 were obtained, volt box A was kept in
the cabinet for 4 weeks, and its ratio was found to have increased 0.048
percent in that time. Upon removing it from the cabinet to the dry
air of the room, the ratio decreased exponentially, with a change of
0.029 percent the first week. Although the experiments reported
above were made on a limited number of volt boxes, the results
probably indicate the magnitude and character of the effects that
would be experienced in the actual use of volt boxes.
Some of the coil studs and binding post terminals of volt box C,
which had the customary hard-rubber top, were disconnected from
the coils and the insulation resistance between them was measured.
Computations based on these measurements showed that surface
leakage between coil studs and between terminals of a volt box
would cause variations in the ratio of less than 5 parts per million
for a relative humidity of 75 percent or less, provided the resistance
was less than 400 ohms per volt, terminal spacings were 1% inch or
more for 150 volts, and the coil stud spacings were 1 inch or more.
However, it is probable that at some relative humidity higher
than 75 percent the surface resistivity of the panel of a volt box would
become low enough to have a serious effect on the ratio.
5. EFFECT OF SELF-HEATING
For most of the volt boxes tested at the National Bureau of Stand¬
ards during the past 13 years, it has been found that the ratio decreases
during the first hour or two that rated voltage is applied, and remains
nearly constant thereafter until the voltage is removed. The de¬
crease is an exponential one, and the magnitude of the change for the
300-volt range in volt boxes having a resistance of 200 ohms per volt
has been found to range from 0.005 percent to 0.03 percent, the average
value being about 0.02 percent. One-half of this change occurs in
about 20 minutes for most volt boxes.
4 P. H. Dike. U. S. Patent 2,026,616, issued Jan. 7, 1936.
224/282
S’iZsftee,']
Gross J Testing and Performance of Volt Boxes
Two volt boxes having a resistance of 333 ohms per volt and a
maximum resistance per coil of 5,000 ohms were found to show no
measurable change in ratio even after several hours at rated voltage.
The 750-volt range of a volt box having 100 ohms per volt showed
a change in ratio of 0.04 percent in 2 hours at rated voltage.
Two factors are responsible for this self-heating effect. First, the
resistance wire has an appreciable temperature coefficient of resist¬
ance; and second, for reasons of economy in construction the resistance
coils in the higher voltage sections are usually on spools of approxi¬
mately the same physical dimensions as those between the poten¬
tiometer taps, but have a much greater resistance per spool and hence
a much greater temperature rise. The grade of manganin usually
used in volt boxes has its maximum resistance near room temperature;
and hence, as the higher resistance coils heat up, their resistance
becomes a smaller percentage of the cold value than does that of the
coils between the potentiometer taps. Hence, the voltage ratio of
the volt box decreases.
Thermocouples were attached to the coils of a volt box of the usual
construction having 200 ohms per volt. It was found that for the
9,000- and 10,000-ohm coils the temperature rise was 10° C as compared
with 4° C for the 300-ohm coil. The ratio decreased with time on a
smooth exponential curve for both the 150-volt range and the 300-volt
range, the total change being 0.005 percent.
For volt boxes of normal construction the temperature rise of the
coils is negligible at 20 percent of rated voltage, so that if the ratio
of a volt box at rated voltage differs from that at 20-percent voltage
by an amount that is negligible for a given measurement then the
heating effect will also be negligible for that measurement.
VIII. APPENDIX I
The galvanometer is first connected between the terminals 2 and 2' and the slider
between 1 and 1' is adjusted to give a balance. This assures that the resistances of
the two sides of the rheostat, together with the leads to a and to d, are in the ratio
Lx_P x -{- II x ~\~R x_is
Ls Ps~\~IIs~\~Ra IK
The galvanometer is then shifted to join 3 and 3'’, ana ci-e oi the auxiliary resistors
Rs or Rx as needed is increased until a balance is again reached. This gives the
relation
LxpP x_II x +R x_is'
L spPs Hs-\-Rs is’
225/283
Journal of Research of the National Bureau of Standards [Voi. 27
Whence F x =F t
(1+j?«+V.)
_0+h!+p,).
In actual practice Rx or R$ is left zero according as Fx is greater or less than Fb, so
that the formula becomes very simple. This procedure does not fully eliminate
the effect of the connections between points 1 and a and between 1' and d, but such
lead resistances are usually quite negligible. It also requires the measurement,
though with only very moderate precision, of Hs + Ps (or Hx + Px)- For boxes of
different resistances a wide range of values of Rx (or Rs) is required.
226/284
olosa*'] Testing and Perjorraan.ee of Volt Boxes
IX. APFENDIX II
where pm, k is the resistance of the rheostat arm when measuring section (m, k).
Hence
(2)
Am ‘ Am
where nm, h is defined as pm, & — pm, i. The total resistance of the first five sections
of the group is given by
(3)
The bridge ratio is now changed to and the bridge is balanced first
across the combination of the first five sections in series giving
(4)
(5)
Hence
(7)
18)
227/285
Journal of Research of the National Bureau of Standards [Voi. 27
Inserting these values and noting that this total resistance of the mth group
is identical with the resistance of the first section of the (m+l)th group we get
ff(m+i,n = 10 Rm,i +2 X10m-3gm|5 + 10m~2nm,t (10)
as the equation connecting resistance values in successive groups.
If this relation is applied successively, beginning with the first group (m = 1)
we get for the resistance of the first section of the mth group in terms of the
resistance of the first section of the first group and the observed differences
k'=k
Defining qm,h as 'fdnm,k' and inserting the values of ratio from eq 9, we get
k'=i
-mjh —
106 / su—
.
— k-10™-
A 2,000/
1,5)
/ ~~ ~Rjd (91>5+?2,5 * * * ?»
k 10' =•0 R 1,1
10,000, Q rn 1 k \
+ Rui \
^m—1)1
k y
(15)
228/286
Gross6'] Testing and Performance of Volt Boxes
Table 1.—Schedule for computing corrections for the standard volt box
Ad-
Section in Bridge just-
n’s ?’ S r’s Correction Range
bridge arm ratio 1 able
arm
1 o 3 4 5 6 7 8
l
0-0. 75 P l,t
0. 75-1. 5 Put nut —Put—Pits ffl,l=7?l,I n, 1=20051,1 S2,l = Sl, 5+71,1 +1.5
0-1.6 P2,l _ _
4,000/10,000
O
O
0-3. 0
u-5
1. 5-3.0 |P2,2 W2,2 = P2,2—P2,l 52,2=712,2 S2,2 = 52,1+72,2
3. 0-4. 5 l>2,3 «2,3 = p2,3 — P2,l 02,3 = 02,2+712,3 J'2,3 = 6752,3 S2)3 = S2*l"h^2,3 0-4.5
4. 5-6. 0 P2,l «2,4 = P2,4—P2,l 52,4 = 52,3-1-712.4 r2,4=5052,4 S2,4 = S2,ri-r2,4 0-6.0
6. 0-7. 5 Ip 2,5 tl2,5 = P2,5—P2,l 52,5 = 52,4+712,5 72, 5=4052,5 S2,5=S2,H-72,5 0-7.5
0-7.5 P2,s
7. 5-15 P2,t «2,I = P2,1“P2,« 52,1=742,1 r2,1=20052,1 S3,1=«2,5+72,1 0-15
0-15 P3,l
15-30 1,000/1,000 (P3,2 B3,2 = P3,2—P3,l 53,2 = 713,2 73.2=10053,2 6'3,2 = S3,l+73,2 0-30
30-45 Phi 1‘3,S = P3,3 — 7)3,1 53.3 = 53,2+713,3 73.3 = 6753.3 S3,3 = 53,1+73,3 0-45
45-60 P3-4 W3,4=P3,4“P3,l 53.4 = 53,3+713.1 73.4=5053.4 S3,4=S3,l+7s,l O-CO
60-75 \P3,S W3.5=P3.5—P3.1 53.2 = 53.4+)l3.5 73.5=4053.5 S3,6 = 53,1+73,5 0-75
0-75 'P3,.
S’
s.
s,
75-150
II
73,1=20053,1 +150
L
0-150 PM
40,000/1,000
150-300 ]P4,2 7!4,2 = Pl,2“p4,l 51.2 = 714,2 74,2 = 10054,2 S4,2 = Sl,l+74,2 0-300
300-450 Pi ,3 Wl,3=P4,3“pl,l 54,3 = 54,2+714,3 74,3 = 6751,3 54,3 = Si, 1 + 74,3 0-150
450-600 Pl,4 72i,4 = p4,4“P4,l 54,4 = 54,3+714,4 74,4 = 5054,4 S4,4=54,1+74,4 +000
600-750 'P4,S 7‘4»5 = P-!>5 p4»1 54,5 = 54,4+714,6 74,5 = 4054,5 54,5=54,1 + 74,6 0-750
229/287
journal of Research < >f the National Bureau of Standards V- L No. 3, March i'^Sd Resear ;n Paper 2402
The method by which multimegohm resistors art1 measured at the National Bureau of
Standards is a null method using an electrometer as the null detector. The charge flowing
through tlie resistor during the time of measurement is obtained from a variable* air capacitor
maintained at a fixed potential. The potentials across the variable air capacitor and thus
across the specimen are maintained constant as indicated by the null reading of the electrom¬
eter by decreasing the capacitance of t he air capacitor at just l he right rate. The capacitance
of the capacitor is varied by a small direct-current motor geared to tilt1 shaft of the capacitor
and whose speed can lx* controlled. Several multimegohm resistors-of two different makes
have bet'ii studied over a period of about thr< ■c years. Although these are tlx* most stable
mullimegohm resistors available, it was found that they had erratic (luctuatiotis of 0.5 to l
percent and were generally voltage sensitive, With the impressed voltage varied from 1.5
to JSO volts, various resistors showed resistant* [• changes ranging from 0.4 to 20.9 percent.
230/147
Figure 2. Equipment used to measure multimegohm resist¬
ances when the vibrating-reed electrometer is the null indicating
instrument.
magnitude dependent upon the setting of me contact rant electrometer and a vibrating-reed electrometer.5
P. The procedure is to decrease the value of Cs The equipment used in the latter case is shown in
continuously and smoothly in such a manner that figure 2.
the potential difference across it remains constant When the Compton quadrant electrometer is used,
at its initial value as its charge diminishes. The the point A is directly connected to ground. The
rate at which Cs is decreased is such that the poten¬ quadrant electrometer has a sensitivity such that a
tial difference across open switch s remains zero as potential of 1 mv produces a deflection of about 1.5
indicated by the detector. It will be obvious, there¬ mm at a scale distance of 2 m. Its period is about
fore, that current through X is being maintained 11 sec, and this relatively long interval is a serious
constant at its initial value i, with the difference disadvantage when trying to observe potential
that the energy dissipated in X is now furnished by changes and to compensate for them.
the capacitor rather than the battery. Under these Of the two, the vibrating-reed electrometer is both
conditions the pertinent equation is: more sensitive and more rapid in its response. Its
sensitivity is such that a potential of 1 mv produces
Vx At a meter deflection of 50 scale divisions. The response
(1) is almost instantaneous. A disadvantage is that
E Vs AC’ neither of the measuring terminals may be con¬
nected to ground. This requires that all equipment
where R is the resistance of the specimen, AC is the be insulated from ground. Another disadvantage is
change in capacitance during time interval At and that on the most sensitive range it has a zero drift
Vx and Vs are the potentials across the specimen that must be determined and for which a correction
and variable air capacitor, respectively. Since the must be applied. The vibrating-reed electrometer
current through the specimen is negligibly small is, however, generally more satisfactory than the
compared to the current in the potential divider, the quadrant electrometer.
ratio Vx/Vs may be taken equal to rx/rs and eq (1) The capacitor, Cs, must have negligible leakage
becomes: between its plates. A three-terminal variable air
capacitor is used. Each set of plates is insulated
At
/£=- (2) from the frame and case. There is no dielectric
AC direetly between the plates. Conductance to ground
on the high-potential side causes no error because
3. Description of Equipment the leakage current returns to the battery without
affecting the electrometer. There is no difficulty in
Any good electrometer having the required sen¬ keeping the leakage to ground on the low side
sitivity may be used as the null instrument to indicate sufficiently low because this part of the circuit is
when points A and B of figure 1 arc at the same maintained at, or near, zero potential at all times,
potential. The insulation between the terminals of and the leakage current is then'fore negligible.
the electrometer must be of such quality that the 6 This electrometer converts a (1-c potential into an a-c potential by means
of a reed vibrating near an anvil on which the d-c potential appears. The a-c
leakage current will be negligible compared to the potential is then amplified and rectified. A negative feedback stabilizes the
cuiTcnt flowing through the specimen. Two types circuit. A full description of this type of electrometer is given by H. Palevsky.
R. K. Swank, and R. Grenchik, i)esign of dynamic condenser electrometers,
of electrometers have been used, a Compton quad¬ Rev. Sci. Instr. 18, 298-314 (May 1947).
231/148
It lias been found that when the relative humidity
rises above about 40 percent appreciable conduction
does occur directly between the plates of the capaci¬
tor. This is probably due to very fine fibers that
are generally floating in the air and which settle onto
the plates, occasionally bridging them. When a
drying agent is placed in the capacitor, this con¬
ductance disappears.
Any one of three capacitors having ranges of 9.5 to
20.5, 15 to 125, and 1O0 to 1,100 picofarads (micro¬
microfarads) are available for use in the measure¬
ments- of resistance. With these capacitors, and
with proper potentials applied to them, measure¬
ments can be made involving currents from 10-9 to
10“13 amp. For instance, if a 109-ohm resistor is
being measured at 1.5 v, it will be seen from eq (1)
that a potential of 500 v will be required on the vari¬
able air capacitator for a capacitance change of 900
picofarads in 300 sec. If a 1013-ohm resistor is being
measured at 1.5 v, a potential of 15 v will be re¬
quired on the variable air capacitor for a capacitance
change of 9 picofarads in 900 sec.
It is necessary to decrease the capacitance of the Figure 3. Motor drive for variable air capacitor
capacitor, Cs, at a specific uniform rate. For this
purpose, a small d-c motor is geared to the shaft of S and starts the motor driving the capacitor Cs so
the capacitor, as shown in figure 3. When using the that all three events occur simultaneously.
vibrating-reed electrometer, the motor must be insu¬ For attainment of the desired accuracy (limit of
lated from the shaft and frame of the capacitor, so a error of 0.1 percent in the present case), the following
fiber gear and insulating supports are used. The requirements must, be met. (1) The capacitance
speed of the motor is controlled by a series resistance. change, AC, during measurement must be large
This resistance is adjusted so that the speed of the enough so that it can be determined with the required
motor is just below the proper speed, and small accuracy.. This depends upon the range of the ca¬
changes in the speed are made by shorting part of pacitor being used. (2) The time of measurement
the resistance by means of a key. By closing this must be long enough so that At can be determined
key momentarily with varying regularity, it is with the required accuracy. The timer used in this
possible to keep the po-tential between points A and work can be read to 0.1 sec, so a time of measurement
B within plus or minus 1 mv as indicated by the of about 300 sec is generally used. (3) The sensi¬
electrometer. tivity of the electrometer to voltage must be such
that in combination with proper regulation, the
The potential divider is a plug box having a total
deviations of the potential between points A and B
resistance of 1 megohm. This resistance is low com¬
from zero are negligible in comparison to the voltage
pared to the resistances being measured, and yet is
applied to the specimen.
high enough to prevent undue drain on the battery
used for potential source. The potential source
4. Measurement Procedure
must be very steady. The voltage on the variable
air capacitor must not change sufficiently during the The measurement procedure can be best described
period of measurement to cause a change of charge on by an example such as the following. Suppose that
the .capacitor, which is appreciable compared with a resistor having a nominal value of 1010 ohms is to
the total charge that flows through the specimen. be measured at 1.5 v-. A time of measurement of
The charge Qx that flows through the specimen during 300 sec is chosen. If the capacitor having a range
measurement is given by QX=VS.AC, aud the charge of 1,000 picofarads is used, a AC of 900 picofarads is
change due to voltage change is given by QS—C-AV. chosen. As seen from eq (1), choice of the above
If the ratio Qs/Qx is to be less than the limit of error L, three values will require that the voltage on the
then AV must be less than L(AC/C)VS. If the limit capacitor be 50 v. As this is also the value of the
of desired error is 0.1 percent, AC is 9 picofarads, and voltage across the voltage divider, the battery con¬
C is 20 picofarads, then AV must not be more than nections are adjusted to give that value as nearly as
0.045 percent of Vs. It has been found that heavy- possible. The battery pack is so constructed that
duty radio B batteries will satisfy this requirement its voltage output can be adjusted in steps of 1.5 v.
As 0.5 ma is the current chosen for the potential
when connected to a potential divider that causes a
divider, the resistance of the potential divider is
current drain not exceeding 0.5 ma.
adjusted to 100,000 ohms. The tap for the specimen
The time is measured by means of an electric timer is placed at 3,000 ohms to give 1.5 v. The value of
that indicates to tenths of a second. This timer is the voltage across the potential divider is watched
started by a switching mechanism that opens switch by means of a potentiometer connected across 300
232/149
t
ohms. To keep the sensitivity of the circuit within across switch, s, would change the voltage on the*
a practical working range, the value of Cg is adjusted specimen and hence the current by about 0.03
until the potential difference across the open switch, percent. The potential that appears across the
s, changes at the rate of about 20 mv/sec when the switch, s, is random in nature, being both positive
capacitance, C, remains fixed. An appreciably and negative, so the average effect is less than
higher rate than this makes it difficult to adjust the 0.03 percent.
speed of the driving motor on the variable air capac¬ The appearance of a potential across switch s
itor quickly enough to keep the potential difference also changes the potentials across the capacitors,
across the open switch, S, from fluctuating more than Cs and (\. The appearance and disappearance of
1 mv. The equipment is now adjusted for measure¬ these random potentials do not affect the measure¬
ment . ment except as they produce a leakage of charge
To make a measurement, switch s is closed, the over the insulation from the low-potential side of
capacitor is set at its highest value, and the batterv the capacitor, Cs, to the shield or through the insu¬
is connected to the potential divider. At time / = 0, lation of the capacitor, Cg. If the insulation is good
switch s is opened at the same time that a switch is (10 10 ohms or greater), this leakage will be neglibible.
closed, which starts both the motor drive on the However, if the measurement is terminated at a
capacitor and the timer. The potential difference time when the potential across the switch s is a
across the open switch, s, is kept less than 1 mv maximum of 0.5 mv, then the resistance of the
(with practice it can be kept less than 0.4 mv) by specimen as computed by eq (2) will be in error
slightly changing the speed of the driving motor. because part of the charge that flowed through the
At the end of about 300 sec the switch is opened, specimen came from a change in potential of the
which stops the motor and timer. Switch s is then capacitors, Cs and Cg, rather than from a change
closed. The capacitance difference is determined in capacitance of capacitor Cs. The error due to
from its original and final readings, and the time this can be estimated by recalling that CQ is ad¬
interval is read from the timer. The resistance of justed so that the rate of change of potential across
the specimen is computed by eq (2). The capacitor switch s is about 20 mv/sec when the capacitance,
is then reset to its highest value, the timer reset to Cs, remains fixed. This is equivalent to about 6,000
zero, the battery to the potential divider is reversed, mv in the 300 sec during which the measurement is
and another measurement is made, as described made, so an error of 0.5 mv at the end of the measure¬
above. An average of this pair of measurements is ment would cause an error of about 0.01 percent.
considered a single determination. Two determina¬ In those cases where the current flow is less
tions are made on each specimen. than 1(U12 amp (resistance greater than 1012 ohms
at- 1.5 v), the capacitance desired for C\ becomes
5. Accuracy and Precautions less than the minimum capacitance of the system
A limit- of error of 0.1 percent can be obtained and cannot be realized. Therefore, the measure¬
ment time must be increased correspondingly if
with this equipment if the following precautions are
the same acurracy is to be obtained.
observed. The capacitance difference, At?, should be
large enough so that it can be determined with the Because of the small currents involved, it was
required accuracy. The variable air capacitors used found necessary to take special precautions in
are calibrated to 0.01 percent of their maximum shielding to eliminate spurious effects. It was
range. Because only about nine-tenths of the maxi¬ found that contact-potential differences of different
mum range is used in a measurement and inaccura¬ metals in the shielding and conductor can cause a
cies due to stopping the motor are involved, AC drift of the electrometer even though no potential
cannot be determined with an accuracy greater than is applied. It was not practical to make all parts
about 0.04 percent. of the equipment from the same metal. However,
The time interval, At, must be of such duration the effect was virtually eliminated by the proper
that it can be determined with sufficient accuracy. selection by a cut-ancl-try process of metals for the
Because the timer reads to 0.1 sec, 300 sec is chosen. various parts. Correction for the residual can be
Care must be taken to assure that the timer is started made by making measurements with the battery
at the same instant switch s is opened. Under connected direct and reversed. An average of these
present procedure, the error due to lack of the two measurements gives a value from which the
synchronization is not greater than 0.1 sec. Thus error is eliminated, provided the effect does not
At can be determined with an accuracy of 0.03 change during the time of measurement. The
percent. terminals on some of the resistors are especially
The resistances in the potential divider can be bad in this respect. The trouble from specimen
readily determined to 0.01 percent, so these present terminals may be eliminated by surrounding the
no problem regarding limit of error. terminal with a sheath of metal that does not produce
It is not possible to keep the potential difference a drift of the electrometer. A bare copper wire
across the open switch, s, always at zero. The wound into a tight helix makes an excellent sheath.
appearance of this potential causes an error in two There is a drift in the vibrating-reed electrom¬
ways. It changes the potential across the specimen eter that is equivalent to a current of about 10~16
and across the capacitor. Because a voltage of 1.5 amp. This is not significant when measuring resist¬
v is applied to the specimen, a voltage of 0.5 mv ances below 1012 ohms, but becomes appreciable for
572772 0 - 61 - 15 233,
resistances above this value. This drift is sufficiently 6. Aging and Voltage Coefficient of Certain
constant over the period of measurement, so that an Commercial Resistors
average of two measurements with reversed battery
practically eliminates this effect. A number of multimegohm resistors have been
To test the equipment for spurious currents, a studied over the past 3 years, during which time they
measurement is made with the resistor removed from have been kept under standard laboratory conditions
the circuit. When the circuit is set for measuring (23° C and less then 50 percent relative humidity).
resistances in the range 109 up to 1011 ohms, and when Most of these resistors have been measured by the
it is operating properly, no detectable change in method described in this paper, but a few of these
potential (equivalent to less than 0.02 percent of the resistors had resistances too low to be measured in
resistance for which the circuit is set to measure) is this manner. These latter resistors were measured
observed across open switch s in 300 sec. When the either by the “comparison method, using a galva¬
circuit is set for measuring resistances of the order of nometer” described in ASTM Tentative methods of
1013 ohms, the readings are appreciable and somewhat test for electrical resistance of insulating materials
erratic. The potential appearing across open switch D257-52T, or more lately by means of the Wheat¬
s always has the same polarity, regardless of the stone bridge. The study included two types of
polarity of the battery across the potential divider resistors. One type (1.2—1 through 1.2-9) Consisted
and is great enough to produce an error of from 0.1 of a small coated rod sealed in a glass envelope. The
to 0.5 percent Averaging a pair of measurements glass envelope had been treated to reduce surface
with the battery reversed tends to cancel this effect, conductance. The other type (1.2-10 through
but because this effect is somewhat erratic, the limit 1.2-14) consisted of molded composition material.
of error for measurements of 1013 ohms is 0.2 to 0.3 The values of resistance fluctuated with time, as
percent. shown in figure 4 and were a function of the voltage
L- H -\ \A o +|
cr.
UJ
CL
r pL^
1.2-8,
_!_Cl.__
TYPE A, 10® OHMS
.o L O
UJ
o L
2
< +1
X
o 1.2-9, TYPE A, 10® OHMS
UJ
o 0
2
<
f—
—O'
L
if)
if)
Ul
K 0
-I
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
MONTHS
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
MONTHS
k\ iva. -4
*2
T 1.2-4, TYPE A, 10'° OHMS
z
UJ
+1 f/
o /
<E
£ o 1/
1.2-5, TYPE A, I09 OHMS
+i
o—— —o ■q r
*•< P
+i
1.2-6, TYPE A,' I09 0HMS
« 0
0 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 38 40
MONTHS
Fir. i re 4. Change of resistanee of m ultimegohm resistors with time.
The crosses arc values obtained after I lie resistors had been subject to a voltage of ISO volts.
234/151
at which they were measured, as shown in table 1. Table 1. Change of resistance of multimegohm resistors with
The values of the molded composition resistors (1.2-10 voltages above 1.5 volts
to 1.2-14) fluctuate more with time than do the glass-
sealed resistors (1.2-1 to 1.2-9), which may be due Volts
Nominal
to humidity effects. One of the molded composition Resistor
value
resistors (1.2-12) now has over twice the resistance 1.5 to 45 1.5 to 180
235/152
Reprinted from NBS Circular 524,
Electrochemical Constants.
Issued August 14, 1953.
12. Standard Cells and the Unit of Electromotive Force
By Walter J. Hamer, Langhorne H. Brickwedde, and Phyllis R. Robb 1
Introduction
236/103
History of the United States Unit of Electromotive
Force
In 1893 the Chicago International Electrical Congress 4 chose the
Clark cell5 as the standard of emf, and assigned to it a value of 1.434 v
at 15° C, based on units of current and resistance then accepted
and in conformity with the report of November 1892 of the Com¬
mittee of the Board of Trade of Great Britain. By an act of the
United States Congress, July 12, 1894, definitions were adopted for
electrical units substantially equivalent to those adopted at Chicago.
The National Academy of Sciences, acting on authority of the United
States Congress, prepared and published on February 9, 1895, de¬
tailed specifications for realizing the international ampere and the
international volt, by the silver coulometer and the Clark standard
cell. Thus, in 1895 the Clark cell was the legal standard of emf in
the United States. It was assigned a value of 1.434 v at 15° C.
In the years immediately following 1893, most countries adopted
1.434 v for the Clark cell at 15° C. Later work showed that cells
made with specially purified Hg2S04 had an emf of 1.4337 v at 15°
C, or an emf 0.0003 v lower than the cells made prior to the meeting
of the Chicago International Electrical Congress. Germany, how¬
ever, recommended and adopted 1.4328 v at 15° C for the Clark
cell in 1898. Their lower value was based on different values of
the ohm and ampere. Although the German value was not uni¬
versally accepted, it was a more nearly correct value as later experi¬
ments showed. Also, during the years 1893 to 1905, the Weston
cell (see footnote 2) was found to have many advantages over the
Clark cell and at an informal international conference6 at Cliar-
lottenburg in October 1905, the Weston normal cell was first pro¬
posed for adoption' in maintaining the volt. This informal rec¬
ommendation, however, was not officially adopted until 1908 (see
below).
In 1906 the United States standard of emf was defined in terms
of the Clark cell made with specially purified Hg2S04 and of the
Weston normal cell, whose emf was expressed in terms of its
measured emf difference from the Clark cell. The Clark cells were
assigned a value of 1.42110 v at 25° C. This was based on the
value 1.434 v at 15° C adopted by the Chicago International Elec¬
trical Congress and legalized by the U. S. Congress, less the correc¬
tion of 0.0003 v for the use of specially purified Hg2S04, and less the
temperature correction calculated from the emf-temperature co¬
efficient determined by Callendar and Barnes [10], i. e., 1.434 —
0.0003 —0.0126= 1.421 i v. The mean of the Weston normal cells
made at NBS, and then on hand, was 1.01890 v at 25° C in terms
of the Clark cells at 25° C. Thus in 1906 the standard of emf in
the United States was a group of Weston normal standard cells,
whose mean emf was taken to be 1.01890 v at 25° C.
In 1908 the International Conference on Electrical Units and
Standards [11] met in London and officially accepted the Weston
4 This Congress was composed of delegates from Austria, Canada, France, Germany, Great Britain,
Italy, Mexico, Sweden, Switzerland, and the United States.
6 The Clark cell was devised by Latimer Clark in 1872 [9], It consists of a Zn-amalgam negative, a Hg-
Hg2S04 positive in a saturated solution of Z11SO4.
6 The international Electrical Congress in St. Louis in 1904 recommended that the various governments
concerned appoint members to an International Commission to consider inequalities known to exist in
the electrical units as maintained in the various countries. In accordance with this recommendation,
the Physikalisch-Technische Reichsanstalt invited representatives of the various standardizing labora¬
tories to a conference at Charlottenburg.
237/104
/
Normal Cell on the recommendation of the Charlottcnburg Con¬
ference of 1905. The London Conference further adopted pro¬
visionally 1.0184 v as the emf of the Weston Normal Cell at 20°
C, and recommended for the emf-temperature coefficient of the
Weston Normal Cell the formula based on the measurements of
Wolff [12] at NBS. By this formula, the Weston Normal Cell
at 25° C had an emf of 1.018174 v, which is 0.000726 v lower than
the value then accepted in the United States. This discrepancy
was caused largely by the fact that the United States had accepted
1.4337 v for the Clark cell at 15° C, whereas the German value
of 1.4328 v at 15° G had proved to be a more nearly correct value.
The London Conference still felt, however, that further work was
needed and' recommended that additional experiments be made.
Meon of the Group Meons 1910
26
852 pv
336 346
PV pv
NBS certified interactional
New basis adopted Jan. 1.-1911
volue 1947 flOpv
- 1.01830 v Mean International value 1947
Figure 12.1. Effect of changes in the volt for the United States from 1906 to 1951,
inclusive.
(Not drawn to scale.)
238/105
other cells made in 1906, which had shown exceptionally good char¬
acteristics, were made the Bureau’s reference cells in February 1908.
The mean of this group of seven cells was 14 /xv below the group of
1906. The group of seven cells served officially as the standard
of emf from 1908 to 1911 and had a mean value of 1.019112 v. In
1910 the mean of the group means of the cells submitted by England,
France, Germany, and the United States was 40 nv higher than the
mean of the NBS older cells, or 26 ,uv higher than the value assumed
by NBS in 1906 for the primary group. This difference of 40 /xv
(0.04 fxv) was later found to be largely caused by an aging effect
of these cells, the duration of which was about 2 years. In other
words, the difference between the Bureau’s older group of cells, and
the new cells made by the four countries in 1910 decreased by 87 /xv,
and hence became insignificant within about 2 years. As the Bureau’s
older group was then 6 years old (4 years at start of comparison),
the new cells made by the four countries were judged to decrease
in emf with time; an increase in emf of the older cells was deemed
extremely unlikely.
Those who met in Washington in 1910 still realized, however, that
the Hg ohm and the Ag coulometer were not fundamental standards
for resistance and current, and believed that absolute units based
on the cgs system were more desirable. Accordingly, the standard¬
izing laboratories of several nations continued their absolute measure¬
ments and determinations of the relation between absolute units
and the international units then in use. 7 It was not until 1948,
however, that the absolute scale based on mechanical units become a
reality. Two World Wars had caused delays in the necessary experi¬
mental work. As a result of determinations of the relation between
absolute units and the international units in various countries, it was
found that the mean ratio was 1.000490 for the ohm and 0.99985
for the ampere [13]. Therefore, through Ohm’s law the ratio for the
volt is 1.00034. As international comparisons of the ampere arc not
feasible, the conversion factors in transportable units were chosen,
and the International Committee on Weights and Measures in October
1946 announced [14]:
The International Committee set the date January 1, 1948, for the
7 This does not imply that methods for absolute measurements were not known in 1910, but that sufficient
accuracy of the measurements had not been attained. Actually the work of Gauss in 1833 and of Weber in
1851 showed the possibility of measuring electric and magnetic quantities in terms of mechanical units.
[13].
239/106
change to absolute units.8 Likewise, other countries had their own
conversion factors. In essence, therefore, on January 1, 1948, the
absolute volt and absolute ohm were identical in all countries.
Between 1906 and 1951, the progress in the science of electri¬
cal measurements, as formally recognized by international agreements,
had led to a number of changes in the unit of emf in the United
States. The magnitude and significance of these changes are sum¬
marized in table 1. The emf of a typical saturated cadmium cell
and the standard electrode potential of the Ag-AgCl electrode at
25° C in terms of the national standard at various periods in the history
of the United States are listed. The effect of the conversion from
international to absolute units is shown in terms of a quantity well
known to chemists. Although the changes are not large, care should
be exercised in making comparisons of literature data prior to 1948
with data reported since January 1, 1948.
Table 1.—Effect of changes in the unit of the United States standard of electro¬
motive force on the electromotive force of saturated cadmium standard cells at 20°
C and on the standard electrode 'potential of the silver-silver chloride electrode at
25° C as certified by NBS.
Silver-
Standard silver
Year
cell chloride
electrode
1 rolts 1rolts
1906 1.019152 0.22258
1910 1.019152 .22258
1911 1. 018300 .22239
1912 a 1.018337 .22240
1913 1. 018300 . 22239
1947 1.018310 b. 22239
1948 1. 018646 . 22246
1951 1. 018646 . 22246
a Effect of aging of reference group. Correction for this was made and incorporated by 1913.
b Value of 0.22239 v is that given by H. S. Harned and R. W. Ehlers [15].
(E—E°) volts
+ln mu rn ci — hi 7h 7ci>
(RTJF) volts
and the volt cancels on the left side of this equation. Therefore, the
value of the activity coefficient of HC1 is the same whether the emf
of the cell is measured in international or in absolute volts. Of
course, E, E°, and PT/Y must be measured or expressed in the same
units. The same holds for pH values, ionization constants, and like
quantities. Conversion from international to absolute units does
8 It had been suggested that absolute volts be used to designate the new unit to distinguish it from the older
unit. The National Bureau of Standards, however, took the view that it was preferable to drop the adjec¬
tive “absolute” and use only “volt.” It was deemed advisable to speak of only one kind of volt and to do
so from the time of the change. When the National Bureau of Standards uses the term “volt,” “absolute
volt” is therefore implied.
240/107
affect the values of the faraday, standard electrode potentials, redox
potentials, electron charge, electron volt, free-energy changes, and
ionic mobilities. Conversion factors for these and other quantities
are given elsewhere [16].
Figure 12.2. Ratios and differences in acid and neutral saturated cadmium cells.
241/108
mium amalgam. These cells were studied for a number of years and
then incorporated into the primary group in 1937.
In figure 12.2, the ratio of the emf of acid and neutral cells in the
NBS primary group is shown as a function of time for the past 15
years (in the figure (Ratio—1)X106 is plotted for convenience). ■ This
ratio has varied with the neutral cells decreasing and the acid cells
increasing in emf in terms of the group mean. In the lower part of
the figure, the differences in emf between acid and neutral cells are
shown as a function of time. Over a 15-year period this difference
has shifted by 9.2 juv, or 9.2 ppm, or 6 parts in 10 million (0.0006 mv)
per year. This shift is very much beyond the precision and accuracy
of most scientific investigations, but is a real effect.
Of course, we cannot decide from, these measurements which, type
of cell is changing in emf with time, or which type is more stable.
Recourse, therefore, has been made to measurements of the tempera¬
ture coefficient and effect of age on the stability of the two types. In
table 2, the differences of the emf at 20° C. from those at 25°, 28°, and
30° O. are given for a group of 46 cells of different ages and acidities
but of similar construction to the cells in the NBS primary group.
Number Aa Aa Aa
Age (20-25)° C (20-28)° C (20-30)° C
of cells
a Mean deviation.
It will be noted that the average deviation from the mean is much
larger for neutral cells than for acid cells, being approximately 5 juv for
the former and 1 nv for the latter. Therefore, acid cells exhibit less
variation in behavior than neutral cells and are probably more reliable.
It will also be noted that the emf-temperature coefficients of acid and
of neutral cells differ and that the emf-temperature coefficient for acid
cells is more constant with age. This observation also suggests that
acid cells are more reliable. Our observations on acid cells are not
very different from those given by Vigoureux and Watts [17].
The second modification that was effected was the substitution of
deuterium oxide for ordinary water as the electrolyte solvent. An
account of these cells is given in papers by Brickwedde and Vinal
[18, 19]. In figure 12.3, the relation between the emf of saturated
standard cells and the mole percent of deuterium oxide is shown.
The total change in going from pure-water solution to pure deuterium-
242/109
oxide solution is only 395/zv, which is well within the range of 2,100^v
of the specialized potentiometer [20] used in the intercomparison
of primary-group cells. The cause of the change in emf resulting
from substitution of deuterium oxide for water has been given in the
papers of Brickwredde and Vinal [18, 19].
2.5
1.5
O 020 ( 20 CELLS )
DjO GROUP 1 ( 10 CELLS )
1.0. I
0.4 0.5 0.6 0.7
Figure 12.4. Statistics on the changes in the electromotive forces at 28° C from
1949 to 1951 of standard cells made with water or deuterium oxide.
243/T10
of experimental technique rather than differences inherent in the
chemistry of the cells. Definite conclusions will be possible only
after more extensive observations.
The third modification that was effected was the substitution of a
fused-quartz cell container [21] for the Kimble standard flint glass
container that has been used at the National Bureau of Standards for
many years. It has been realized for a long time that the alkali of
glass may neutralize the acid in acid standard cells (neutral cells
should not be affected as Hg2S04 should hydrolyze to maintain the
equilibrium concentration of H2S04), thereby affecting the stability
of the cell [22, 23, 24], It was felt that fused quartz may, therefore,
be a better container. Details of their construction have been given
elsewhere [21] and will not be repeated here.
Seven satisfactory saturated standard cells were made in these
containers. Eight control cells in Kimble glass were made at the
same time. Five acid cells (0.02 N in H2S04) were made in both
types of containers and two neutral cells in quartz and three neutral
cells in glass. Readings of the emf were taken at intervals during
1949, 1950, and 1951. If we take the cell average for 1951 and sub¬
tract the average for the same cell in 1949, we obtain an emf difference
that is a measure of the change in emf of that cell over a 2-year period.
These emf differences at 28°C are given in table 3.
Table 3. Differences in the electromotive force of standard cells in quartz and glass
containers between 1951 and 1949 (1951 value minus 1949 value)
Quartz Glass
nv tin M*'
1.9 1. 7 3.8 3.9
3.4 3.5 3.6 9.4
l.G 2.2 11.9
1.6 3. H
j. i 1.9
The average change in emf for glass containers is higher than for
quartz containers. In addition, the changes in emf vary less in the
quartz group than in the glass. Greater average change and greater
variability in emf seem to be the case for glass in both neutral and
acid cells. The sample size, however, was too small to draw definite
conclusions between the behaviors of acid and neutral cells in the two
types of containers. Additional cells will be obtained and further
studies made.
International Comparisons
Annually or biennially, the National Bureau of Standards transports
by messenger saturated standard cells that have been compared with
the NBS primary standard cells to the Bureau International des
Poids et Mesures9 for comparisons against the international group
maintained there. The National Bureau of Standards measures the
emf of its cells immediately prior to transport to the International
# For a brief history of the International Bureau, see the discussion of F. B. Silsbee under standard cells
at the end of this paper.
244/111
Bureau and immediately after their return. Except on very rare
occasions, the two measurements have agreed within a microvolt,
showing that transportation by messenger had hot affected the cells.
Comparisons with the International Bureau (BIPM) since World
War II are shown in figure 12.5. Differences between the NBS and
BIPM values reported for the NBS cells are given. Reported differ¬
ences are represented by the shaded circles. In 1946, 1948, and 1950
the values for the NBS cells differed by +10.0, +9.0, and —0.79/xv
from the BIPM values for the same cells. However, BIPM reported
their results in international volts in 1946 and 1948 and in absolute
volts in 1950. If their international values of 1946 and 1948 are
converted to absolute volts, using their conversion factor of 1.00034,
the differences between the BIPM and NBS values for the same cells
are —0.2, —1.2, and — 0.79 for 1946, 1948, and 1950, respectively.
The difference between the two groups of cells, those at NBS and those
at BIPM, each maintained differently and by different staffs, has
not varied on the average by as much as 1 pv since World War II.
Of course, in arriving at the respective conversion factors, cognizance
had been taken of the 10-ppm difference between the BIPM and
NBS units that existed prior to January 1, 1948.
In table 4, comparisons made in 1950 with BIPM and other coun¬
tries are listed. In column 2 are shown the difference in the volt from
BIPM, of the National Physical Laboratory (England), Laboratoire
Central des Industries Electriques (France), the Electrotechnical
Laboratory (Japan), the Institut de Metrologie d’U. R. S. S. (Russia),
and the Deutsches Amt fur Mass and Gewicht (Germany), who sub¬
mitted cells to BIPM in 1950. The mean of the differences from the
BIPM unit for the countries contributing to international comparisons
is —2.02 ppm. If we omit the Russian and the German values, the
mean difference is only +0.16 ppm. The numbers represent the
relative changes in the unit since the comparison of 1947, on which
were based the various conversion factors that equalized the absolute
units adopted in the several countries on January 1, 1948.
245/112
In column 3 of table 4, the differences of the units of other countries
from the U. S. unit are listed. The mean difference for the partici¬
pating countries is only —1.25 ppm. Comparisons have been carried
out at the National Bureau of Standards for the national laboratories
of Sweden, Israel, and the Union of South Africa.
ppm ppm
United States. -0. 77
England. _ -2. 2o -1.43
F ranee _ +0. 09 +0.KS
Japan_ +3. Si +1.3i
Russia... -23.0 -22.2
Germany_ + 10.2 + 11.0
Mean.. -2.02
References
[1] Edward Weston, German Patent 75,194 (Jan. 5, 1892); The Electrician 30 ,
741 (1893).
[2] J. L. Thomas, C. Peterson, I. L. Cooter, and F. It. Kotter, J. Research NBS
,
43 291 (1949) RP2029.
[3] H. R. Curtis, C. Moon and C. M. Sparks, J. Research NBS 21, 375 (1938)
RP1137.
[4] ,
P. Vigoureux, Nat. Phys. Lab. Collected Researches 24 277 (1938); F. E.
,
Smith, 11 209 (1914).
[5] L. Hartshorn and N. F. Astbury, Phil. Trans. Roy. Soc. (London) [A] 236,
423 (1937).
[G] II. L. Curtis, R. W. Curtis, and C. L. Critchfield, J. Research NBS 22, 485
(1939) RP1200.
[7] R. W. Curtis, It. L. Driscoll, and C. L. Critchfield, J. Research NBS 28, 133
(1942) RP1449.
[8] P. Vigoureux, Phil. Trans. Roy. Soc. (London) [A] 236, 133 (1937).
[9] ,
Latimer Clark, Proc. Roy. Soc. (London) 20 444 (1872).
[10] H. L. Callendar and H. T. Barnes, Proc. Roy. Soc. (London) 62, 117 (1897).
[11] Report of International Conference on Electrical Units and Standards, 1908
(refer to appendix 2 of reference [13] for reprint of this report).
[12] ,
F. A. Wolff, Trans. Am. Electrochern. Soc. 13 187 (1908).
[13] F. B. Silsbee, NBS Circular 475 (June 30, 11)49).
[14] NBS Circular C459 (1947;; see also appendix 6 of reference [13].
246/113
[15] II. S. Harned and R. W. Elders, J. Am. Chem. Soc. 55, 2179 (1933).
[16] ,
G. W. Vinal, J. Wash. Acad. Sei. 38 265 (1948).
[17] P. Vigoureux and S. Watts, Proc. Plivs. Soc. 45, Part. 2, 172 (1933).
[18] ,
L. H. Rrickwedde and G. W. Vinal, J. Researcli NPS 20 599 (1938) RPE094.
[19] E. II. Rrickwedde and G. W. Vinal, J. Research N RS 27, 479 (1941) RP1435.
[20] II. R. Rrooks, .1. Research NRS 11, 223 (1933) RP586.
[21] G. W. Vinal, E. II. Rrickwedde, and W. J. Hamer, Compt. Rend, de la
Quinzieme Conference de l’Union Internationale de Chemie Pure et
Appliquee, Amsterdam, 92 (1949).
[22] II. von Stoinwehr, Proc. Congres Int. d’Electricite, Paris, 2d section, Report
No. 7, 9 (1932).
[23] A. N. Shaw, H. E. Reillev, and R. J. Clark, Phil. Trans. Roval Soc. Loudon
229, 134 (1930).
[24] G. W. Vinal and M. E. Howard, RS J. Research 11, 255 (1933) R P588.
Discussion
247/114
change—H2S04 may come in. Would it not be advisable to add the
acid in the form of a buffer solution?
Dr. W. J. Ha mbr: I think that is a very good suggestion. I believe
some work with buffer solutions has been done by Vosburgh. When
you add a buffer solution you are, of course, adding more materials
that could cause barm. Nevertheless, I think your suggestion is a
good one.
Dr. H. H. Uhlig, Massachusetts Institute of Technology, Cam¬
bridge, Mass.: I should like to ask why heavy water changes the
potential of a standard cell.
Dr. H amer: In a standard cell, water takes part in the cell reaction.
The cell reaction is
=^(CdS04H‘°>)+2Hs(1)’
where m is the number of moles of water associated with 1 mole of
CdS04 in the saturated solution. Mrs. Brickwedde, in making up the
cells, did not only change the solvent but used cadmium sulfate with
heavy water of hydration. The difference in emf between a cell with
ordinary water and one with heavy water is related merely to the
differences in the free-energy changes of the cell reactions and involves
both the solvent and the solid phases.
Dr. Uhlig: I asked the question because I wondered whether, in
the standard cell, one might not be measuring the polarized Hg
potential. This means that if Cd is added to Hg, we measure essen¬
tially the potential of an alloy, with Hg polarized by small corrosion
currents to the same potential as the alloyed Cd. The potential,
therefore, is of Hg on which hydrogen atoms are adsorbed, and would
change the potential to an.active value almost hut not exactly equal
to the reversible potential of Cd in Hg. In heavy water, deuterium is
adsorbed instead of hydrogen, which polarizes Hg to a slightly dif¬
ferent potential, because the overvoltage value differs from that of
hydrogen. I wonder if this might he a possible explanation.
Dr. Hamer: 1 would not think so. I think it is mostly just the
difference in the cell reactions. In other words, we could change the
solvent to ethyl alcohol and get a difference in emf corresponding to a
difference in free-energy changes, or a change hi the solid phases.
Dr. W. F. K. W YNNE-f To nes, University of Durham, Newcastle,
England: I agree with Dr. Hamer. Surely in any cell of this sort the
thermodynamics should show quite clearly that one is dealing simply
with the different phases present. The only possible effect is the one
Dr. Hamer mentioned, and that is the change in the solid phase.
Dr. Uhlig’s idea is interesting, hut it just can’t he right. We certainly
haven’t the wrong chemical process hi the cell.
Dr. U hlig: Wouldn’t it be worthwhile to check this by arranging a
cell where water of hydration does not enter? It seems to me there
would be a smaller effect of heavy water if no water of hydration were
involved.
Dr. H amer: It could he checked in an unsaturated cell, hut CdS04
is less soluble in heavy water than in ordinary water, and a comparison
based on equal concentrations (or solubility) would show that the emf
difference would he larger rather tharj smaller, as the emf increases on
dilution of electrolyte.
248/115
Dr. V. K. LaMer, Columbia University, New York, X. Y.: This
is one of the most reversible cells we ever had, isn’t it?
Dr. M. Eppley, The Eppley Laboratory, Inc., Newport, R. I.: Yes,
but a reversible cell owes its reversibility to the maintenance of the
original conditions. If you come to other conditions, the cell is no
longer reversible. This cell comes as nearly to being reversible as a
cell can be.
Dr. Uhlig: This is what I am driving at. The separate open-
circuit potentials of Hg, and Cd in mercury, are represented as shown
in figure A. Upon alloying Hg with Cd and immersing in an electrolyte
a small corrosion current is set up between Hg and Cd atoms, which
cathodically polarizes Hg to the potential Eobs (corrosion potential).
Similarly, Cd is anodieallv polarized from its reversible value also lo
T _ 2.303 RT i
log
" * 6.52 F (VXK)^2’
250/117
Hence for the Weston Cell, the reversible potential E (therm.) can
be expressed in terms of the polarized or observed potential E (obs.) as
follows: E (therm.) = E (obs.) +/(i), where J(i) is the additive term to
the potential resulting from polarization of the Cd electrode in accord
with the local-action current, i. Hence, the thermodynamic potential
E (therm.) is always greater than the observed value, the difference in
general being small, although presumably measurable. From Wagner
and Traud’s data on anodic polarization of Zn amalgam,/(f) is in the
order of (13 mv/ma)/cm2; hence, the corresponding potential shift for
2.1 X10-9 amp/cm2 is 3X10-8 v.
In this regard, the value 2.1 X10-9 amp/cm2 should be looked upon
as a minimum value, with the actual value probably being higher.
An increase, in some instances, by a factor of 100 is not inconceivable.
For one thing, metallic impurities in the amalgam would tend in
general to decrease hydrogen overvoltage, thereby increasing corrosion
and local-action currents and producing hydrogen far in excess of the
calculated (0.008ml/cm2)/vr, as is sometimes observed. Even if the
amalgam were pure, the overvoltage on either solid or liquid Cd
amalgam may be lower than that for pure Hg. Dissolved oxygen and
depolarizers from cork, possibly contaminating the electrolyte, would
tend to increase the local-action currents above the calculated value
without accompanying accumulation of hydrogen gas, but with the
expected effect on the polarized potential of Cd and of the cell.
All these effects point to local-action currents on the surface of Cd
amalgam as an important if not primary source of potential drift in
Weston standard cells.
Dr. Hamer: I think your discussion refers to the stability of the
cell rather than to the question of the difference in the emf (open-
circuit voltage) of standard cells made with water or deuterium oxide.
Dr. J. O. M. Bockris, Imperial College of Science and Technology,
London, England: I don’t fully understand Prof. LThlig’s idea. The
cell must be working reversibly if no actual change in the concen¬
tration of reactants is detectable over a period of years. It would
not suffice to test the explanations for the effect of heavy water on
the emf of the cell merely by removing the water of hydration from the
reaction. Change of medium changes the solvation energy of the ions
and hence the emf of the cell. It would be relatively easy to calculate
the change in emf of the cell to be expected upon change from water to
deuterium oxide throughout, assuming the usual cell reaction stated
and complete reversibility. Then we could see if this new figure was
in agreement with the experimental one.
Dr. Eppley: I tried to measure the potential of Hg in a CdS04
solution. There is no definite potential that I have ever been able
to find. Of course, you would expect there would be practically no
Hg ion there because the Cd would be forcing it the other way. That
is the explanation I have given. Is that a correct explanation? Xow
the depolarizer, so-called, is not a depolarizer but is a tie vice whereby
you maintain a concentration of Hg ions over the Hg. Of course, if
we use it up, everything goes wrong.
Dr. G. W. Vinal, Princeton, N. J.: I agree.
Dr. Hamer: Your explanation for the potential of Hg in solutions
of CdS04 appears plausible. We have made no studies of this type
at the Bureau. The Hg-Hg2S04 electrode is one of the second kind,
but ceases to be of this type when Hg2S04 becomes depleted.
251/118
RP518
ABSTRACT
CONTENTS
Pafie
I. Introduction and purpose_ 89
II. Effect of temperature inequality of the limbs_ 90
III. Hysteresis effects_ 96
IV. Summary_ 98
1 Eppley, Trans, of the A. I. E. E., vol. 50, No. 4, p. 1293, December, 1931.
252/89
Bureau oj Standards Journal oj Research [ voi. 10
2 Weston, The Electrician, voi. 30, p. 741, 1892. Vosburgh and Eppley, J. Opt. Soc. of Am., vol. 9, p.
72 1924.
3F. E. Smith, Phil. Mag., vol. 19, p. 272, 1910.
253/90
Park] Temperature Effects on Standard Cell Emj
and cells of model B were in metal cases. Several cells of each model
were tested, but since all cells of the same model gave practically
the same performance, data are given here for only one cell of each
model.
The results of the experiments to determine the effect of tempera¬
ture inequalities are shown by the curves in Figures 2 to 6. Curves
in Figure 2 show the results on a standard cell of model A, when it
was placed near a 16-inch slide rheostat dissipating 288 watts. The
inserted diagram shows the position of the cell with respect to the
rheostat. Curves A and B show the effect on emf when the cell was
placed 10 inches from the heated rheostat. When the positive
limb is nearer to the rheostat (curve A), its temperature is higher
SHEET BRASS
WOOD
than that of the negative limb, and the cell emf increases. When the
negative limb is nearer to the rheostat (curve B), its temperature is
higher than that of the positive limb, and the cell emf decreases.
These curves illustrate the fact that the temperature coefficients of the
two limbs of the cell taken separately have the opposite effect on the
cell emf, also that the effects are about the same per degree difference
in temperature. In curve A the emf of the cell does not return to its
normal value until about 36 hours after removing the source of heat;
this and other slight irregularities which may occur in the cooling
part of the curves are due to the hysteresis effect which will be taken
up later. Curve E shows the temperature difference between the
limbs of the cell under the same heating conditions as for curves
A and B. As seen from curves A, B, and E, a temperature difference
254/91
Bureau of Standards Journal of Research [ voi. 10
Curve C (fig. 2) shows the effect on the emf when the cell was
placed 20 inches from the heated rheostat. By comparing this curve
with curve A, it is seen that the change in emf due to unequal heating
happens to vary approximately inversely as the square of the distance
from the rheostat.
255/92
Park\ Temperature Effects on Standard Cell Emf
Curves D and F (fig. 2) show the results when the cell was placed
in a protecting container (shown in fig. 1) and submitted to the same
heating conditions as for curves A, B, and E. As seen from curve F,
a container of this sort practically eliminates temperature inequality
in the cells. The change in the emf of the cell, shown by curve D, is
partially due to the fact that the temperatures of both limbs of the
cell were increased about 5° C. due to the radiation of heat from the
rheostat.
The curves in Figure 3 show the results obtained when the experi¬
ments just described were repeated, using a standard cell of model B.
o2
256/93
Bureau of Standards Journal of Research [ Vol. 10
due to the heat of the sun’s rays, shining continuously on one side of
the cell case for over two hours. For both of the above conditions the
cell and the source of heat were arranged to give the maximum
opportunity for temperature inequality of the two limbs of the cell.
Curves are shown for the three models of cells in their usual cases and
the differences between the behavior of these cells under the same
conditions are largely due to the differences of their cases.
257/94
Park] Temperature Effects on Standard Cell Emj
Figure 6.—Change in standard cell emf due to exposure of one side of cell
case to sunlight
Curve A, standard cell model A. Curve B, standard cell model B. Curve C, standard cell
model C
ture or condition of the cell may cause cells of the same model to have
different changes in emf for the same external temperature disturb¬
ance. However, the results shown may be used as a guide for the user
to determine whether protection against temperature inequality is
258/95
Bureau of Standards Journal of Research [ voi. 10
needed for the accuracy required. The methods mentioned here for
protection against temperature inequality are probably sufficient for
all practical work. However, where high precision (0.01 of 1 per cent
or better) is required, more effective methods of protection against
temperature disturbances should be used.
259/96
Park] Temperature Effects on Standard Cell Emj
much greater hysteresis than one of the oldest cells with a low emf.
Therefore, it is very difficult, if not impossible, at the present time to
predict just what the hysteresis error of any given cell will be. How¬
ever, from the curves in Figure 7, an approximation may be had as
to within what limits the hysteresis error of any unsaturated portable
cell in good condition 6 might fall. Eppley 7 found that when an
unsaturated cell was suddenly cooled from 30° to 25° C. the average
hysteresis error was from 0.02 to 0.027 per cent. This average
error lies within the limit shown here.
A method of protecting cells from hysteresis errors was applied to
Cell IV, which was chosen because of its large hysteresis. It was
placed inside a container (such as shown in fig. 1) and this container
Curve A, effect on emf of Cell II, when temperature was changed'from 23° to 27° C. Curve B,
effect on emf. of Cell II, when temperature was changed from 27° to 23° C. Curve C, effect on
emf of Cell IV, when temperature was changed from 23° to 27° C. Curve D, effect on emf of Cell
IV. when temperature was changed from 27° to 23° C. Curve E, same as curve D with cell in
insulated bos
was placed in a wooden box which was sufficiently large to leave about
3 inches of space between the container and the box on all sides, mineral
wool being loosely packed in this space. The cell, thus thermally
insulated, was placed in the constant-temperature cabinet, set for
27° C. and maintained at this temperature until the emf of the cell
became steady. Then the air temperature was lowered to 23° C.
and held there, readings of the emf being taken regularly until it
again became constant. The results of this experiment are shown by
curve E in Figure 7. By comparing this curve with curve D it may be
seen that thermally insulating the ceil decreases the hysteresis effect
by about 50 per cent in the case of a sustained temperature change.
However, under ordinary laboratory conditions, where the average
temperature over a period of 24 hours does not change much from day
6 Cells which are not in good condition may show much greater hysteresis errors.
* See reference 1, p. 1299.
260/97
Bureau of Standards Journal of Research [ voi. 10, Park]
IV. SUMMARY
Temperature inequality between the two limbs of a standard cell,
which may occur (1) when the cell is placed near a heat source, such
as a hot radiator or a rheostat carrying current; (2) when the cell is
placed where the sun’s rays will strike it; (3) when the cell is placed
where cold air will strike it; or (4) when the cell is held in the hand,
caused an error in the emf of the cell of about 300 juv per °C. difference
in temperature. The temperature inequality between limbs can be
practically eliminated by placing the cell in a container such as shown
in Figure 1, or, where possible, by lining the cell case with sheet
copper one-sixteenth inch thick.
The hysteresis error, which occurs in portable unsaturated cells
when their temperature is changed, varies widely among different
cells, and the only way to determine this error for a given cell is to
check its emf immediately after its temperature has been changed.
Errors due to hysteresis may be reduced by sufficiently insulating the
cell against changes in temperature; but the only way to eliminate these
errors is to keep the cell in a box whose temperature is thermostatically
controlled. Under normal laboratory conditions precautions against
hysteresis need be taken only if accuracies of 0.03 per cent or better
are required.
Washington, October 12, 1932.
261/98
Reprinted from Journal of The Washington Academy of Sciences
Vol. 38, No. 2, February 15, 1948
262/33
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES vol. 38, no. 2
For the measurement of voltage it has factory temperature range. The use of a
been found convenient to adopt as a pri¬ potentiometer or of a tap resistor (volt box)
mary standard, the voltage of the standard is the simplest method for extending the
cell. The unsaturated standard cell, the range of measurements of direct voltages by
voltage of which is approximately 1.0180 ± reference to the standard cell. If r is the
volts, is almost universally used as a lab¬ resistance from one end to the tap point and
oratory reference standard for voltage R is the total resistance including r, then
measurements. When kept in a thermo- the voltage e across r can be measured with
stated enclosure and tised under favorable reference to a standard cell by using a
conditions, it exhibits a remarkable con¬ potentiometer and the voltage U across R
stancy of voltage. The small differences that is then V = e- R/r.
may arise in a thermostated group of cells ' Although this simple resistance method
from year to year are of the order of micro¬ is adequate for small steady direct voltages
volts. Unfortunately, similar electrochem¬ it requires modification for the precise
ical devices having 100, 1,000, or 1,000,000 measurement of the high varying direct-,
times the voltage of a standard cell do not alternating-, surge-, and pulse-voltages that
exist. Therefore, in the usual measurement one encounters in practice. The manner in
of higher voltages, it has been necessary to which the voltage at any instant varies
develop devices which ultimately refer back with the time is a fundamental factor in the
to the standard cell. problem of measurement. It is necessary to
Resistance methods of extending the express alternating voltage in terms of di¬
range of voltage upward from the value of rect voltage which can be evaluated by
the voltage of the standard cell make use of reference to the standard cell. Instruments
the fact that by selecting suitable alloys and designed to read the same on the effective
heat treating them properly resistors can be value of alternating voltage as on direct
constructed to have a resistance, or a ratio voltage are called “transfer instruments.”
of resistance, that remains constant to As transfer instruments, electrostatic in¬
within a few parts in 100,000 Over a satis- struments (1) are favored in Great Britain,
V,. ... _
DC VOLTAGE
o
Time
^mox
263/34
Feb. 15, 1948 defandorf: the measurement of high voltage
while electrodynamic instruments (2) find 7. Force arising from the attraction or repul¬
sion of electrostatic charges on electrodes.
favor in this country. An electrostatic volt¬
S. The deflection of a stream of charged par¬
meter, for instance, should give the same ticles by means of a known field, either electro¬
indication for a direct voltage V as it does static or magnetic, after their acceleration in
in the case of an alternating voltage of pure vacuo by the voltage to be measured.
sine-wave form for which the crest or maxi¬
The devices used in measuring high volt¬
mum voltage Vm — \2V.
age will be considered in accordance with
If rectified alternating voltage is applied
the above classification and in the order
to a direct current voltmeter the instru¬
listed.
ment will read the average value of voltage
applied to it, i.e., Vave = (1/0/,! vdt- SERIES IMPEDANCE METHOD
If, however, completely rectified alter¬
The simplest method of measuring high
nating voltage is applied to a good electro¬
voltage would appear to be to connect a
static voltmeter the voltmeter will read the
high impedance in series with a sufficiently
crest voltage. Crest voltage can also he de¬ sensitive current measuring instrument hav¬
termined from the length of a spark gap ing a negligible impedance compared to the
across which it will just cause a discharge,
value of the high series impedance, Z. Val¬
or it can be readily evaluated from the wave
ues of the indicated current, i, would then
form determined by use of a calibrated os¬ give the high voltage, Y — iZ. The imped¬
cillograph.
ance Z may, of course, be primarily resis¬
A more detailed knowledge of the varia¬
tive, inductive, or capacitive or combina¬
tion of the voltage with time such as is
tions of these elements.
given by an oscillograph becomes of special
Series resistors.—This basically simple
interest in the case of pulse and surge volt¬
method has been widely used in nuclear dis¬
ages. Thus if would seem appropriate to ap¬
integration work for direct voltage measure¬
praise the value of the several devices used
ments, wherein the impedance Z is built of
in the delineation of wave form, and to men¬
many high resistance units in series (dia¬
tion each type under the particular voltage
gram at A in Fig. 2), care being taken to
divider or device with which it is generally
insure that the current entering at the high
associated.
voltage end of the resistor is the same as
A consideration of high-voltage measur¬
that leaving through the deflecting instru¬
ing devices appears to lend itself better for
ment at the low voltage end. For reliable
discussion under a classification of methods
measurements it is necessary to be sure that
of measurement or types of devices rather
the electrical leakage across insulating sup¬
than under the classification of character¬
ports of the resistor and from section to sec¬
istics of voltage previously outlined. The
tion is negligible and that there is negligible
devices which are useful in measuring high
corona current from the units. Changes
voltage may be conveniently classified as
arising from self-heating must be made
to type by considering whether the method
negligibly small or must be allowed for. The
of measurement employs:
design of such a series resistor should be
1. A high series impedance with a low-imped¬ substantially the same as for the potential
ance instiument to indicate current through the
divider type to bo discussed later in detail.
impedance.
2. A potential divider in which a fraction of the
Scries reactors.—For alternating voltage
total voltage is measured across tap points of measurements, series reactors have been
the impedance. used as the series impedance (3). They have
3. A voltage transformer that permits measure¬ the drawback, however, of requiring iron
ment of a low voltage having a direct ratio to the
cores at lower frequencies if the inductive
high voltage.
4. A generating voltmeter in which a voltage reactance is to be made large in comparison
proportional to the field intensity in the region of with the resistance of the windings. Stray
the instrument is indicated. and distributed capacitance effects raise
d. A spark discharge in which the length of the
additional objections so that series reactors
spark gives a measure of the voltage.
6. The cooling effect of an “electric wind” as in have been little used in high voltage meas¬
the ionic wind voltmeter. urements.
264/35
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES vol. 38, no. 2
value of voltage measured between taps, sembly is sectionalized, each section con¬
when multiplied by the ratio of the value sisting of uniformly spaced wire-wound re¬
of the total impedance to that of the tapped sistors located between toroidally shaped
section then gives the value of voltage corona shields, which are electrically con¬
across the total impedance. Potential di¬ nected to the end of each group of resistors.
viders are commonly constructed of sections Isolant.ite insulation is used for supporting
of resistance, of resistance shunted by the resistors and the uniformly spaced co¬
capacitance, or of capacitance. Types of rona shields. The shields were constructed
each of these will be described. of smooth copper-tubing of circular section
Corona shielded resistor.—For a resistive in order to provide satisfactory field grading
divider one may use n resistors, each of re¬ along the resistor and thus to insure free¬
sistance r, all exactly alike (or nearly so), dom from corona discharge currents at the
connected in series, or one resistor (a—1) highest intended operating voltage. The de¬
times as large as the smaller one connected sign provides for adequate insulation, me¬
in series with it, and across which the volt¬ chanical support, and centering of this
age drop is measured. For use at higher
voltages, both the electrical and mechanical
designs of the physical device require care.
Improvements in his earlier designs (10)
have been incorporated by L. S. Taylor in a
resistor, not yet described, which is in
use in the measurement of the high rec¬
tified voltage supply for a 1.4-million-
volt X-ray tube in the X-ray Labora¬
tory of the National Bureau of Standards.
As seen in Fig. 3, the column at the left
side contains the corona-shielded resistor,
which is installed immediately adjacent
to the column of 10 large cascaded-rectifier
tanks, each of which at maximum volt¬
age contributes 140 kv to the total voltage
of 1.4 million. This arrangement provides
double shielding. The external system
seen in Fig. 3 has tubular connections
from the spun-metal corona shields to
each section of the cascaded supply. The
internal resistor assembly, also separately
corona shielded, is thoroughly insulated
from this outer system. Thus the currents
required by the outer shields are supplied
as a sectionalized direct load on the supply
independently of the measuring circuit.
This outer shield therefore takes care of
current flow arising from external ioniza¬
tion, migratory dust particles, and minor
surface discharges as well as surface leakage
across the insulating columnar supports,
and relieves the internal resistor of the
resulting irregularly varying effects. In
addition, the separate outer shields provide
mutual field grading and the large top
shield is effective in shaping the general Fig. 3.—1.4-million-volt rectifier column with
electrostatic field. The internal resistor as¬ shielded resistor column on left side.
266/37
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
column inside the outer corona shield. Each decreasing it to values much below 0.05
section was adjusted and thoroughly tested milliampere because of the instability of the
at rated voltage before assembly and subse¬ resistance insulation, and of other troubles.
quent measurements of individual sections For much of the pioneering nuclear disinte¬
show their resistance to be practically free gration work (11) carbon resistors available
from drift. These precautions were taken to in units of much higher resistance per unit
make sure that the current throughout the length than the wire-wound type have
resistor will be the same as at its low voltage proved both economical and useful.
end where measurement of the current is It should be observed that the relatively
made; i.e., in order that the high voltage, high shunting capacitance of the corona
V, being measured is equal to nri where nr shields of this type of resistor limits its
is the total resistance of the stack of n usefulness as a resistor for alternating cur¬
shielded resistors. Actually the current, i, rent measurements.
is often determined by adding a small re¬ Capacitance shielded and guarded resistors
sistor, r0, and measuring v — ir0 by means of for alternating current.-—The resistor di¬
a potentiometer. vider, in high favor for the measurement
Manganin wire-wound resistors and a of direct voltage, has certain specific defects
relatively large current of 1 milliampere in addition to its relatively large consump¬
were chosen as design values in order to tion of power. It necessarily has residual
minimize parasitic effects of leakage, ab¬ inductance and stray capacitance associ¬
sorption, and corona currents, and thus to ated with its resistance. As shown at A in
insure a resistor that would permit measure¬ Fig. 4, each section ?t • • • rn may be con¬
ments to 0.1 percent or better. A current of sidered as having at its terminals a lumped
1 milliampere corresponds to a load of 1.4 shunting capacitance of value Ci • • ■ Cn,
kilowatts, so that the resistor was designed and between its upper terminal and ground
to dissipate a corresponding amount of a capacitance Ca • ■ • Cm. If the value of
heat. Carbon or the so-called “metallized” capacitive reactance \/2irfC\ is »ri and
radio-type resistors have a much higher l/27r/Ca»>UY and if the impedance used in
temperature coefficient and are in general the measuring circuit shunting rn is »r„,
less stable than wire-wound resistors. Con¬ the voltage division by resistance will be
siderable care must therefore be taken in trustworthy. As a result of the increase in
their use to insure freedom from thermal shunting effect of the stray capacitances
and voltage effects. Although reduction in with frequency, these inequalities grow
self-heating and in energy loss in such resis¬ less until the division of the applied voltage
tors may be secured by reducing the current is no longer proportional to resistance.
say to 0.1 milliampere, the effects of leakage The shielded a-c resistor (12) minimizes
currents become relatively more important the effects of the unavoidable ground ca¬
and may contribute to some extent in re¬ pacitances at the cost of additional energy
ducing the attainable accuracy to the order dissipation, by employing a second or guard
of 0.5 percent. resistor in parallel with the first or “work¬
Still another consideration that should ing” resistor. This guard resistor is so con¬
not be overlooked is the shunting effect on nected to the shields that it supplies current
the resistance of the capacitance of the to the ground capacitances that otherwise
shields. In the case of a resistor of the high would have to be charged through and
resistance type, sudden surges are easily along portions of the working resistor. In
transmitted through the shunting capaci¬ an a-c shielded resistor the working, or
tances so that the measuring instrument, shielded resistor, is composed of sections
which must be correspondingly more sen¬ of value r each contained within a metal
sitive, requires increased surge-protection. shield or box as indicated at B in Fig. 4.
Thus there are drawbacks to either in¬ Each shield is maintained at a potential
creasing the current in the measuring resis¬ corresponding to the midpoint of its en¬
tor to values greatly exceeding one milli¬ closed resistor by connecting it to a tap
ampere because of increased energy loss, or point on the proper section Rn of the guard
267/38
Feb. 15, 1948 defandokf: the measurement of high voltage
resistor. Thus this arrangement makes the are filled with oil in order to improve ther¬
value of each shunting capacitance definite mal characteristics. This resistor is ar¬
and supplies the ground capacitance cur¬ ranged in tiers, each tier consisting of four
rent from the guard circuit. This is ac¬ working resistor boxes with corresponding
complished at the cost of a slight increase sections of the guard resistor supported on
in the shunting capacitance of the indi¬ a mahogany framework. Tiers are as¬
vidual resistors because of the capacitance sembled one above the other by means of
to the shield in which the}'- are located, porcelain bus-type insulators, which serve
since one end of the enclosed resistor is to provide additional insulation and to
above and the other below the potential of permit bolting together of the resistor-
the shield. However, because the potential supporting frameworks to form a mechan¬
difference across the capacitance of either ically stable assembly. Each box contains
end of the resistor element to its shield is 20 flat mica cards wound unifilarly with
only one-half the voltage drop in this sec¬ manganin wire. It is not feasible to extend
tion of resistance r, the shunting effect t he range of a-c shielded resistors and retain
need not be excessive at low frequencies, the same order of accuracy (0.01 percent)
say 25 or GO cycles. Such a resistor, com¬ by this procedure much above 30 kv because
posed of about 25 shielded sections, is used of the effects of the capacitances which
in a voltage transformer testing setup for shunt the guard resistances. The resulting
phase-angle measurements and ratio meas¬ error increases as the fourth power of the
urements to 0.01 percent at the National voltage.
Bureau of Standards. This resistor is rated An a-c shielded resistor of this type is
at 0.05 amperes in each of the two (working also satisfactory for use as a series resistor
and guard) 500,000-ohm circuits. In this in conjunction with an indicating volt¬
device, which at 25 kv absorbs 2.5 kw, the meter or sensitive oscillograph. It may be
individual sections of the working resistor used equally well on direct current but
are enclosed in metal shield boxes, which would ordinarily not be used because its
268/39
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
A
Fig. 5.—Surge resistor.
load on the source is greater than that for inductance. The cable connection to the
the simpler corona shielded resistor. cathode ray oscillograph has a surge im¬
Weller (13) devised a shielded a-c resistor pedance usually of the order of 50 ohms,
for use in transformer testing up to 132 kv whereas the full-scale sensitivity of the
in which the shield current was supplied by cathode ray oscillograph is of the order of
auto-transformers rated at 75 kva, instead 2,000 volts for a cold cathode-type and 200
of by a tapped resistor. volts for a hot cathode-type cathode-ray
Surge resistors.—In surge voltage testing oscillograph. Thus for full-scale deflection
equipment used for studies of surge effects the currents to a cable considered as a
on electric power and transmission equip¬ resistance would be of the order of 40 or
ment, the fundamental component of volt¬ 4 amperes, respectively.
age in a 1^X40 microsecond wave4 is of Resistance dividers for high-voltage
the order of 105 cycles per second. Hence surge measurements are not provided with
components as high in frequency as 107 corona shields because they would add
cycles per second are of significance in excessive stray capacitance but are often
fixing the wave form. The cathode ray so located adjacent to the surge generator
oscillograph used to delineate such surge itself that some degree of shielding and
wave forms in conjunction with a potential- field grading along the divider is secured.
dividing resistor is generally located at some Since a resistor for a 2,000,000-volt surge
distance from the resistor and main dis¬ measurement must be at least 700 cm. long
charge circuit in order to avoid induction to provide adequate longitudinal insulations
effects from the large surge currents. Fig. 5 and avoid flashover troubles in air, its
indicates the common method used for distributed and stray capacitances intro¬
connecting the resistor tap point through a duce disturbing effects unless the resistance
coaxial cable to the deflecting plates of the is held proportionately low. Thus a divider
oscillograph. The resistor is composed of a resistor for a 2,000,000-volt surge may have
series of resistance cards. Each card has two a resistance as low as 5,000 ohms and, if
similar windings wound in opposite direc¬ wire-wound, the wire must be of sufficient
tions and connected in parallel to reduce diameter to carry high momentary currents
without damage. As a rule, accuracies of
41.e., one that rises to crest in \\ micro¬
seconds and falls again to one-half crest value in the order of one percent are all that are
40 microseconds. required. If it is desired to study surge com-
269/40
Feb. 15, 1948 defandorf: the measurement of high voltage
ponents of the order of 108 or 109 c}rcles per low frequency range where a higher order
second the residual inductance of each ele¬ of accuracy is required.
ment and the distributed and particularly Capacitor dividers.—Because residual in¬
the non-uniform stray capacitance effects, ductance and residual capacitance effects
assume especial significance in surge resis¬ in a resistance divider are unavoidable, not
tor dividers. For those frequencies, sup¬ only in the resistance elements themselves
plementary capacitance elements (14) nray but also in their electrical connections, it is
be added to the resistance elements of the natural to turn to the use of capacitance
divider in such a manner as to make uniform elements for alternating voltage division.
the capacitance shunting effect on each The capacitance divider generally consists
element and thus insure proper division of of a single high-voltage guarded-electrode
the surge voltage and its correct delineation capacitor in series with a low-voltage capac¬
by the oscillograph. The present interest in itor of very much higher capacitance. The
better voltage measurements of surge wave-
fronts of duration less than one microsecond
should lead to improved designs of surge-
voltage resistor dividers.
The cathode ray oscillograph plays such
an important role in the field of surge-
voltage (and surge-current) measurements
that its contribution should be mentioned
at this point. Although its value for use at
lower frequencies should not be underesti¬
mated, it stands alone in versatility for use
in high-frequency, surge, and pulse meas¬
urements. It owes this versatility to the
small inertia of its electron beam, which
serves at the same time as the moving ele¬
ment and pointer. The cathode-ray oscillo¬
graph, with a suitable sweep circuit, am¬
plifiers, shunts, and potential dividers, has
been developed into one of the most useful
pieces of electrical laboratory equipment
and covers a range from a few cycles per
second to frequencies of millions of cycles
per second. Developments in electron optics
and of new phosphors are constantly en¬
larging the place of the cathode-ray oscillo¬
graph as a useful precision device in spite
of the complication of its accessory equip¬
ment. In the future an order of accuracy
better than 1 percent may be expected.
For most measurements, and especially
those of high voltage surges, its high effec¬
tive impedance results from the low capaci¬
tance between its deflection plates. The
energy loss, arising largely from stray ions
and electrons within the tube, is so small
as to be of little concern. Thus in the visual
or photographic delineation of wave, surge,
and pulse shape the starting point now Fig. 6.—Compressed-gas capacitor. Diagram of
Bousman and Ten Broeck capacitor altered to
seems to be a calibrated cathode-ray oscillo¬ include complete shielding of the working elec¬
graph except in very special instances in the trode.
270/41
JOURNAL- OF THE WASHINGTON ACADEMY OF SCIENCES vol. 38, no. 2
residual inductance effects in capacitors are A rather recent and desirable circuit ar¬
minimized by properly shaping their sizable rangement using a compressed-gas shielded
conducting parts. The residual resistance capacitor for testing potential transformers
effects are minimized (1) by using high is described by Bousman and Ten Broeck
grade solid insulation for the electrode sup¬ {16). Their circuit arrangement is similar to
ports in order to keep electrical leakage a Schering bridge, i.e., a bridge in which
through and across the surface of the insu¬ the A and X arms are capacitances instead
lation low; (2) by using free-air or com¬ of resistances. The difference lies primarily
pressed gases {15, 16) as the dielectric in the mode of supplying the voltage to the
because of their low dielectric loss; and bridge, the high-voltage arm (A) being
(3) by using electrodes with well-rounded supplied by connection to the ungrounded
contours and smooth surfaces to insure end of the high-voltnge winding of the
freedom from corona discharge. It is cus¬ transformer, and the low-voltage arm (X)
tomary to support the guard electrode on being connected to the ungrounded end of
its own insulation opposite the high-voltage the low-voltage winding. Thus if this bridge
electrode and in turn to support the guarded is balanced when supplying power to one
low-voltage electrode on the guard. The vinding of the transformer, one may de¬
insulation between the guard and the termine from the settings of the bridge arms
guarded electrode is arranged to lie outside both the voltage ratio and phase angle of
the high electrostatic field so that dielectric the transformer. A simple reconnection of
losses as null as surface and volume leak¬ the bridge arms to a suitable supply per¬
age to the working capacitance are mini¬ mits quickly checking the constancy of the
mized. bridge arm components before and after
The range of various types in free air is ratio and phase angle measurements. Thus
limited by the breakdown voltage of air to in this bridge reliance for the ratio measure¬
a practical value of about 15,000 v/cm ment is placed primarily not on capacitanc¬
gradient. Churcher {17) has described a es but on the constancy of resistance
capacitor for use at 300 kv, cylindrical in coils which are more suitable as reference
form, with an over-all diameter of 2 meters standards because of their stability, while
and a height of 7 meters. This unit is to the phase angles are based on the air ca¬
some extent an absolute standard as it pacitors. An extension in range to higher
permits computation of its capacitance voltages might logically folloMr the arrange¬
from dimensions measured under normal ment of Bousman and Ten Broeck without
working conditions. To that extent it is serious reduction in the accuracy of 0.1
preferable to a smaller compressed gas percent claimed for their equipment.
type {16) shown in Fig. 65 for use up to 300 The subject of high-voltage nrave form
kv, having an outside diameter of less than should also be considered in connection
1 meter and height over the bushing of with capacitance dividers. Offhand, capaci¬
3 meters, and a shielded capacitance to the tance dividers would appear to be ideal for
high-voltage electrode of 50 micromicro¬ use with the cathode-ray oscillograph, as
farads. However, when the space available its impedance is essentially capacitive reac¬
is limited, the compressed-gas equipment tance. In surge measurements, honrever,
may be built to occupy about one-fourth it is usually both desirable and convenient
the space of a free air unit of the same rating. to locate the cathode-ray oscillograph
The one advantage, which may be in part at some distance from the surge circuit
psychological, of constructing a capacitor in order to minimize inductive effects.
whose capacitance can be calculated from This involves the use of a fairly long
its dimensions lies in the care and precision high-quality cable (preferably coaxial) con¬
demanded in its design and manufacture. necting the divider to the oscillograph so
that the surge impedance of the cable,
6 In this figure the original design of Bousman primarily resistive, rather than the capaci¬
and Ten Broeck has been modified to show com¬
plete shielding of the outer guarded cylindrical tive impedance of the oscillograph plates
section by the guard. assumes the major role in the measurement
271/42
Feb. 15, 1948 defaxdorf: the measurement of high voltage
circuit. Thus, as suggested earlier, a surge conditions have been found to remain re¬
resistor is to be preferred although surge markably constant over long periods of
capacitor dividers (18) have been used. For time, variations in ratio requiring for their
lower-frequency measurements this objec¬ detection measurements having an accuracy
tion does not exist to the same extent and of the order of 0.01 percent. Departures
capacitor dividers with cathode-ray oscillo¬ from nominal ratio have been determined
graphs as well as with amplifiers supplying by resistance bridge methods (20) up to 30
electromagnetic oscillographs (string or kv with a precision of 0.01 percent and by
loop in use up to several thousand cycles capacitance bridge methods up to 132 kv
per second) have been found useful in with a precision of 0.1 percent or better
delineating wave form with an accuracy of (16).
a few percent. Where symmetrical multiple high-voltage
A capacitance-divider method yielding a windings are provided in a voltage trans¬
high order of precision in delineating low- former the following “series-parallel prin¬
frequency wave form is due to Silsbee (19) ciple” has been found valid for extrapolat¬
who, in line with the early work of Rosa, ing measurements of ratio factor at low
used a point-by-point method. In his device voltage to permit their use at higher volt¬
a potentiometric balance by means of a age. Within the voltage limits of a high-
quadrant electrometer is obtained across voltage shielded resistor or capacitor, and
the low voltage portion of a capacitance po¬ with the high-voltage windings in parallel,
tential divider for as many points in a measurements of ratio factor are made to
repeated voltage wave as may be desired, cover the voltage per coil range of the low-
thus permitting the evaluation of crest-, voltage winding. The nominal ratio for
average-, and effective-values of the wave series connection is then multiplied by the
form. This method requires a synchro¬ measured ratio factor at the same volts per
nously driven contactor which may be set coil for the parallel connection. Reliable
accurately for each balance point selected measurements of effective alternating volt¬
for delineating the wave. The accuracy of age by this method can be made to better
measurement is very high and appears to be than 0.1 percent (20) and are in use up to
limited primarily by the steadiness of the 250 kv. The cost of transformers with such
alternating voltage source, say to a few symmetrical multiple windings increases
parts in ten thousand. rapidly with voltage. Other less expensive
and less accurate (1 percent to 0.5 percent)
transformer methods
transforming devices such as (1) a number
Voltage transformer.—The method of of small chain-connected or cascaded trans¬
measuring high alternating voltage in com¬ formers (21, 22) and (2) a high-voltage
mon use in the United States employs a resistor (23), reactor (2Jf), or capacitor
step-down transformer termed a voltage or (25) in series with a small transformer, have
“potential” transformer by the manufac¬ been introduced in Europe for measuring
turer in order to designate its intended use high voltage but have not met with general
in voltage and power measurement. The favor in the United States.
high-voltage winding is connected across Supply transformer with high-voltage wind¬
the terminals of the voltage source to be ing tap or with voltmeter coil.—Although the
measured, a voltmeter is connected across use of a voltage transformer with only an
the low-voltage winding and its reading is instrument connected as the burden on the
multiplied by the ratio of transformation to low-voltage winding represents the ideal
obtain the value of the high voltage. The arrangement, it should be mentioned that
ratio of voltage of the high-voltage winding adequately precise values of high voltage
to that of the low-voltage winding of a may often be deduced from voltage meas¬
well-designed transformer remains nearly urements made on the low-voltage input
constant over a wide range of voltage. The windings. This is particularly true if the
measured values of ratio of well-built resistance and leakage reactance of the
transformers kept under normal laboratory high-voltage winding are low and the cur-
272/43
i
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
rent drawn by the load on the high-voltage connection giving an accuracy of voltage
winding is small. Better still, an instrument measurement of about 2 percent at lull
connected between ground and a tap point load and correspondingly better accuracy
on the high-voltage winding (26) near its at lighter loads. Still another arrangement
grounded end may serve for measuring consists in providing a third winding or
a relatively low voltage which will be pro¬ voltmeter coil {27) so located relative to the
portional to the high voltage except in so low- and high-voltage windings that the
far as the voltage across the tapped section magnetic flux linked by it automatically
of winding is affected by distributed and takes into account any voltage drop in the
stray capacitance currents which flow high voltage winding arising from the load
through it from the rest of the high-voltage connected to it. An accuracy of one-half
winding. Fig. 7 shows three such 350 kv of 1 percent at full-load leading current and
60 c/s 1,000 kva transformers in the High better accuracy at smaller loads is claimed
Voltage Laboratory of the Nat ional Bureau for a good design. These latter arrange¬
of Standards. These units are shown con¬ ments are not considered to be as trust¬
nected in cascade to give 1,000,000 volts. worthy as the use of a separate voltage
Each unit has a high-voltage winding tap transformer.
273/44
Feb. 15, 1948 defaxdorf: thf, measurement of high voltage
± =F
274/45
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
measuring device associated with high- maker has done his best to produce them.
voltage generators of the Van de Graaff With reasonable care and cleanliness in
type. assembly, dust has not been found to be a
When the generating voltmeter is used problem in generating voltmeters operating
as a voltage measuring component (i.e., as within a pressure chamber at higher gradi¬
an instrument for determining a value of ents (31).
gradient to be multiplied by a constant When used with suitable precautions, the
factor in order to obtain the total voltage generating voltmeter offers a means of ob¬
difference) it is not sufficient that the fields taining information not readily available in
in the immediate neighborhood of the in¬ other ways. For instance when its com¬
strument shall be below corona or dis¬ mutating device is provided with a phase-
charge-forming values, but it is equally shifting mechanism and the rotor is syn¬
necessary that the gradients at surfaces in chronously driven it provides a means of
the neighborhood of the sample field as determining wave form (32) at high voltage
well as those at the high-voltage electrode with practically no load on the source being
be below discharge-forming values and that investigated. It is effective on a rippled
there be no nearby relatively large sources direct voltage as well as on alternating
of air ionization or dust. The effect of dust voltage. When used for determining wave
and charged particles may indirectly pro¬ form it may be located in an undisturbed
duce serious distortions in high fields by and readily accessible portion of the high
being deposited and aligned on otherwise voltage field. The generating voltmeter
smooth polished metal surfaces so as to method of determining wave form should
form the necessary points for initiating be good to a few tenths of 1 percent.
discharges and consequent space charges
that would otherwise not exist. If the AIR SPARK-GAP BREAKDOWN
charge on the surface of the segments al¬ Sphere and rod gaps.—If an accuracy of
ternately exposed to the electric field is to the order of 3 percent in determining crest
lie in definite ratio to the total field, then voltage is sufficient, as in the case of insu¬
at no place in the field may the gradient lator testing, the sphere spark gap in air is
exceed the approximate breakdown gradi¬ useful as a crest voltage measuring device.
ent, 30 kv crest/cm, in air at NPT. The For voltages above 17 kv spheres of 6.25
practical limit is actually much lower. For cm diameter and larger (up to 2 meter
instance, the value of average gradient at diameter for approximately 2,000,000 volts)
which self-propagating discharges occur operated under controlled conditions serve
in the case of thunderstorms appears to be as voltage standards for electrical break¬
about 10 kv/cm. In the case of the Brooks down measurements of dielectrics (33).
electrometer (30) discharges have been Earlier theories of breakdown of sphere
known to occur between the parallel faces gaps assumed that pure air, as well as other
of the electrometer plates when the voltage gases, has a definite breakdown strength or
applied divided by the spacing had a value breakdown gradient at normal pressure and
of 5 kv eff/cm. Thus in air at I\TPT it is temperature. Paschen’s law relating the
doubtful that for a generating voltmeter an length of the breakdown gap with air den¬
average value of gradient exceeding 7 kv sity permits correction for usual tempera¬
max/cm with an upper limit of 20 kv/cm at ture and pressure variations. Russel, Dean,
the electrode surfaces can be employed con¬ Peek, and others have given empirical re¬
tinuously without some error in measure¬ lationships based in part on electrostatic
ment arising from the above causes. Pre¬ field theory for both sphere and cylindrical
sumably this practical limit arises from gaps, which, although they fit well in a
the almost ever-present small particles of limited domain of pressure and tempera¬
dust (insulating or semi-conducting) which ture, are not so satisfying as relations (3d)
when deposited on the surface upset the developed at a later date on the basis of the
ideal condition of pointless smooth surfaces newer atom-physical background. The de¬
one wishes to assume after the instrument tailed work on discharges in gases by Loeb
275/46
Feb. 15, 1948 defandorf: the measurement of high voltage
(35) and his collaborators as well as much ness of the surface, and cleanliness of the
other valuable work in this field has pro¬ air, appear to be limited to a rather low and
vided a fairly satisfactory explanation of questionable accuracy, say from 1 to 3
some of the scattering of breakdown values percent. When extreme precautions are
of spark gaps. Meek (36) suggests that in used by following a ritual of cleansing, the
the case of the shorter gaps the electron use of ultraviolet irradiation of the spheres,
avalanche initiated at the cathode con¬ limitation by resistance of surface pitting
stitutes the usual initial process whereas in by the spark current, and bv insuring
the case of larger gaps the mechanism is “cleanup” of the sparking surfaces of the
more akin to the positive streamer dis¬ sphere gap through preliminary sparking, a
charge of lightning and originates within series of 10 or more sparkover values often
the gap at some distance from the electrode. may be observed to agree to within ±0.1
For gaps of intermediate length there lies a percent. However, this apparent high pre¬
domain in which the initiation may be of cision, equal to that of the high-grade indi¬
either variety and such gaps show a larger cating voltmeter employed as part of the
scattering of breakdown values. The prob¬ equipment in making such observations, is
ability of a free electron existing in the right deceptive. Painstaking observations made
location to initiate a discharge is, of course, the following day under seemingly identical
an important factor in the scattering of the conditions may agree among themselves to
initial breakdown voltage when the time the same precision but their average value
of application of voltage is extremely short. will almost invariably differ by several
Irradiation (37) of small electrodes of the tenths of one percent, and sometimes by
gap with radium or with ultraviolet light to mote than 1 percent from those made on the
produce photoelectrons is helpful in reduc¬ previous day. In spite of its low order of
ing this type of scattering but appears not accuracy the sphere gap serves as a com¬
to be necessary in the case of large spheres mercial standard (33) for high alternating-
because of the much higher probability, in and surge-voltage measurements appar¬
the larger volume of air between the spheres ently because of its basic simplicity. In
of the presence of an initiating ion or elec¬ larger sizes it gives a useful measurement of
tron. the maximum value of surge voltage and
An idea of the effectiveness of ultraviolet essentially serves as a voltage limiter when
irradiation in reducing scatt ring may be it is used in parallel with a device under¬
cited in the case of 12.5 cm diameter brass going voltage-withstand tests.
spheres. Irradiation of the spheres by an Simple needle gaps, as a matter of his¬
open carbon arc reduced the scattering of torical interest, were once accepted as al¬
individual GO-cycle sparkover values by a ternative standard voltage measuring gaps,
factor of 5 as compared with the results but as a result of the inherently high scat¬
obtained without irradiation. On the other tering of values of sparkover voltage, varia¬
hand the average values of sparkover volt¬ tion in sparkover values with changes in
age were lower by from two to five percent humidity, and large scattering in sparkover
in the irradiated case, the amount of lower¬ voltage when used for measuring surge
ing being dependent on the intensity of voltage, they have been discarded as stand¬
radiation. ards. Rod gaps (39) in which the elec¬
Basing his work on the detailed informa¬ trodes are used repeatedly as contrasted
tion now available on the mechanism of with the formerly prescribed use of new
spark formation Ver Planck (38) appears to #00 needles for each sparkover of the needle
have successfully correlated the enormous gap, have sparkover values which are also
amount of data on sphere spark gaps. affected by humidity to about the same
From the very nature of the spark be¬ extent as the flashover of porcelain insula¬
tween spheres and of corona on cylinders, tion and appear to have replaced needle
voltage measurements based on sphere (or point) gaps for those uses where gaps
gaps and corona cylinders (34), because of are desired for correlating the flashover
their dependence on surface shape, cleanli¬ voltage of insulators.
276/47
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES vol. 38, no. 2
277/48
Feb. 15, 1948 defaxdorf: the measurement of high voltage
change in range by adjusting the spacing nearly spherical. The moving system is
between the high-voltage electrode and rather massive with a separate damping
grounded electrode of which the moving chamber within the guard. The Imhof (44)
element usually forms a part. It is interest¬ (1926) design, Fig. 12, provides a flattened
ing to consider the developments in these guard ring and fits the light-weight at¬
instruments over a few years as shown in tracted electrode into its own damping
Figs. 11, 12, and 13. In the early design of chamber. Starke and Schroeder (43) (1928)
Abraham and Villard (41) (1911), Fig. 11, in one model, Fig. 13, employed a relatively
the curvature of the disk and guard is . larger flat-guarded electrode and a flat-
strip suspension for the movable rectangu¬
lar flat electrode, P, providing it with a
mirror instead of a mechanical pointer, as
well as with a balanced damping chamber
arrangement. Nearby objects would be
increasingly less effective in producing de¬
flection errors at equivalent spacings in the
later voltmeters. These voltmeters may be
read to closer than 1 percent but unless
calibrated in place, especially when used at
large spacings at maximum rated voltage,
are likely to be affected by nearby objects.
Ellipsoidal voltmeter.—The ellipsoidal
voltmeter of Thornton and Thompson
(58), illustrated in Fig. 14, is of the nature
of an electrometer and satisfactory theoret¬
ical equations have been developed for it.
It, like the electrometer, depends on a
relatively undistorted axial field if it is to
be used as an absolute instrument.
Fig. 11.—Abraham and Villard electrostatic
voltmeter. The moving element consists of a metallic
ellipsoid of revolution carried on a bifilar
silk suspension and is provided at its lower
end with a reflecting mirror and damping
trostatic voltmeter.
278/49
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NT. 2
vane. The ellipsoid is suspended with its The instrument described was designed
long axis horizontal and is centrally located for measurements up to 200 kv. As shown
between two vertical circular plates facing schematically in Fig. 14 the plates were
each other and at an adjustable distance 140 cm in diameter and were spaced as
apart. Polarization of the ellipsoid by the much as 100 cm apart (i.e., average gradi¬
electric field established between the plates ent of 2 kv per cm). An exploration of the
by the source of voltage to be measured, field indicated that it was quite uniform at
gives rise to a mechanical couple tending to the midpoint between the plates within a
align the ellipsoid, which is initially set. at radial distance of 25 cm; however, no study
an angle to the horizontal axis perpendicu¬ of its longitudinal variation appears to have
lar to both plates. Although the change in been made. Great care was taken to insure
the angular deflection of the ellipsoid may that the ellipsoids of revolution Avere ac¬
be used as a measure of the voltage applied, curately shaped so that the theoretical
a more sensitive and rather interesting- relationships would apply. One ellipsoid
method of voltage indication has been con¬ of duralumin Avas 3.9692 cm long and 0.5970
trived which involves measuring the in¬ cm in diameter and Aveighed 2.0496 grams.
crease in frequency of its swings when volt¬ The electric gradients at the tips of the
age is applied. Thus: ellipsoids are considerably higher than the
average gradient betAveen the plates. These
E — k(n2 — n02)112 (1)
gradients must be kept Avell beloAv corona¬
where E is the electric field strength, k is a forming values if disturbing effects from
constant found from dimensions, n is the electric Avind are to be avoided. In its prac¬
number of swings per second with the volt¬ tical form Avhere it is to be used as a de¬
age on, and n0 is the number of swings per flection instrument for laboratory or shop
second with the voltage off. measurements an insulating enclosure is
Sufficiently large plates are used and at provided for the suspension, ellipsoid, mir¬
such a separation that the electric gradient ror, and damping mechanism to shield
E throughout a considerable volume along them against Avind and dust. At the bottom
the axis would be substantially uniform if end of the moving-system assembly the
the ellipsoid were not present. Since the damping vane is suspended in a damping-
disturbance effected by the ellipsoid is chamber attached to a tube Avhose upper
small, the voltage applied to the plates is end is cemented to the bottom of a IioIIoav
glass sphere in which the ellipsoid is cen¬
V = Ed, (2)
tered. The upper end of the sphere is ce¬
where d is the plate separation. mented to a second tube housing the bifilar
279/50
Feb. 15, 1948 defandorf: the measurement of high voltage
suspension and an adjustable suspension used. An oil-cup damper made the sphere
control for changing the period of swing. motion nearly aperiodic. An accuracy of 1
The control mechanism and the torsion percent was claimed.
head are mounted at the supported upper Sorensen (47, J+8) employed a horizontal
end of the tube. A spherical shape was used arrangement of 1-meter spheres with rather
for the hollow glass sphere enclosure for the long slender shanks presumably in order
ellipsoid so as to permit a theoretical eval¬ to reduce effects of attraction arising from
uation of the effect of the dielectric of the the shanks. The electrostatic attraction of
sphere on the electric field E at the ellip¬ the grounded sphere could be readily meas¬
soid. The theoretical correction derived ured as it was supported by suspending its
for this spherical glass enclosure agreed shank by four ropes tied to the shank at
well with experimental results when the the apices of the two thus-formed F-sus-
relative humidity was not high enough to pensions. The upper ends of the ropes were
cause electrical surface leakage. attached to ceiling members. This laterally
The ellipsoidal voltmeter is said to be stable suspension possessed only a small
accurate to 0.1 percent and to be only longitudinal stability so that differences in
slightly affected by humidity. Because of longitudinal electrostatic forces of several
low average gradient, 2 kv eff/cm, it ap¬ hundred grams could be measured to better
pears to be more bulky than other electro¬ than one gram. A small wire in line with and
static voltmeters and is of interest mainly attached to the end of the grounded sphere
because it is a unique arrangement per¬ shaft ran over the rim of a bicycle wheel
mitting absolute measurements. thus insuring low friction. A small weight
Sparkless sphere-gap voltmeter.—Large pan attached to the end of the wire per¬
spheres ordinarily used as sphere spark-gap mitted weighing the force of attraction.
voltmeters in measuring high crest-voltage These two sphere-electrometer devices
have been modified to permit their use as represent useful laboratory tools for they
electrostatic voltmeters for measuring the can be calibrated and used as voltmeters
effective value of voltage at spacings for measuring effective voltage. They can
slightly in excess of sparking distances.
This arrangement has been called a spark¬
less sphere-gap voltmeter (47).
Hueter (4d) employed a vertical ar¬
rangement of 1-meter spheres. The upper
high-voltage sphere was supported on a
spring whose additional extension as a
result of the electrostatic force was magni¬
fied by a lamp-mirror-scale arrangement.
The spring and the mirror optical-lever
arrangement were mounted within the
sphere shank which was provided with a
small window. An external arc-lamp and
scale were mounted on an adjacent wall
in the laboratory and gave satisfactory
readings in daylight. The vertically adjust¬
able lower sphere was grounded and its
driving screw mechanism was arranged to
indicate the gap length. The weight of the
upper one-meter sphere was 60 kg and for a
75 cm gap the electrostatic attraction was
approximately 800 grams at 1,000,000
volts. In order to minimize effects of
changing gap length, only small displace¬
ments (less than 0.5 percent of the gap
length) of the spring-suspended sphere were Fig. 14,—Ellipsoidal voltmeter,
280/51
JOURNAL OF'THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
also be employed in t lie usual manner as from coplanarity, Snow (54) has developed
sphere spark-gaps for the measurement of a mathematical solution which takes into
crest voltage. If one is satisfied with the account the change in force arising from
accuracy of the sphere spark-gap as a change in position of the disk. This change
high-voltage standard, measurement of the in force might at first appear only to vary
voltage wave form permits calibration of inversely with the separation d (shown by
these devices as electrometers for the meas¬ a simple differentiation of equation (3)).
urement of effective voltage without re¬ However, Snow has evaluated the addi¬
course to other methods. To a very limited tional change in force arising‘from the
extent they may be considered to be abso¬ redistribution of the charges, which in the
lute electrometers, particularly if adequate case of a protruding disk results in an
relative clearances to the floor, walls, leads increase in concentration of charge at the
and other conductors and insulators are edge of the disk at the expense of the charge
maintained so that approximate corrections on the adjacent edge of the guard plate
can be made for the presence of these ob¬ and vice versa in the case of a retracted
jects. Simple theoretical calculations of the disk. Troublesome instability in disk elec¬
force of attraction can be made by the trometers arising from this latter com¬
method of image charges in the case of two ponent of force has for a long time been
insulated or one grounded and one insulated recognized as a weakness in electrometers
sphere. For a 25-cm spacing of his 1-meter designed to cover a wide range of voltage
spheres Sorensen considered that no cor¬ measurement by adjusting the spacing d.
rection was necessary, and at 35 cm the One may either elect to provide a linear
effect of shanks and other nearby objects restoring force adequate for all spacings d
was said not to exceed 1.5 percent. at a considerable sacrifice in sensitivity at
One may hope for an accuracy approach¬ large spacings or provide for some adjust¬
ing 0.5 percent with the design of Ilueter ment of the restoring force with change in
in making relative measurements up to one spacings at nearly constant maximum al¬
million volts. This would involve ample lowable gradient. The restoring force re¬
clearance for the spheres and first calibrat¬ quired to balance the electrostatic attrac¬
ing by an accurate lower voltage method tion in high voltage electrostatic volt¬
using a separation of the spheres adequate meters and electrometers has been pro¬
for one million volts and making certain vided by the following devices:
that corona-free electric field conditions (1) Suspension of the moving electrode on one
exist in the neighborhood. arm of a gravity balance, i.e., change of restoring
Disk electrometers.—About 1880 Lord force secured by adjustment of c.g. of balance
relative to central knife edges (52, 30).
Kelvin made an outstanding contribution
(2) Suspension of moving electrode on a coiled
by incorporating a guard ring for the disk spring (4-6, 61).
of the attracted disk electrometer. The (3) Pendulous suspension of electrode (49).
guard ring not only validified the use of the (4) Suspension of disk electrode assembly on a
simplifying mathematical assumptions in flat-strip (torsion) suspension (43, 55).
(5) Suspension of the disk from a metallic
computing the electrostatic force of attrac¬
membrane (50, 51).
tion of the disk but it established the basis (6) By combining the torque provided by a
for a design whose readings were less ef¬ flat-strip supporting suspension with that pro¬
fected by neaidrg objects. When coplanar duced in a current-balance arrangement of coils
which permits adjustment of the restoring force
with the guard ring the force
(55) by change in current.
572772 0-61-19
281/52
Feb. 15, 1948 defandorf: the measurement of high voltage
nection that permits adjustment of sta¬ calibration with known weights against
bility as well as of sensitivity. The restoring current is made in order to permit subse¬
force arises in part from the vertical strip quent checking of the calibration of the
suspension which carries the disk with its electrometer by means of the current in¬
plane vertical as well as the two similarly stead of weights after it has been filled with
mounted astatically connected current bal¬ compressed gas.
ance coils. One pair of field coils with cur¬ Brooks absolute electrometer.—The ab¬
rent if is connected in series aiding one solute high voltage electrometer of Brooks
moving coil and provides a torque propor¬ (30) (Figs. 16 and 17) is illustrative of what
tional to if for balancing the electrostatic can be accomplished in the way of precision
force on the disk; the other pair of coils when the attracted disk is supported on
with current im is connected in series op¬ one arm of a gravity-type balance. Brooks’s
posing so that by reversing im and adjusting modification of the Kelvin electrostatic
its value relative to the current in the mov¬ attracted disk electrometer with guard ring
ing coil the restoring force of the strip was designed for use in free air up to 275
suspension toward the null or coplanar kv on alternating voltage and was arranged
position of the disk may be either opposed to allow a step by step experimental evalua¬
or aided thus altering the stability of this tion of errors not readily calculable. In ad¬
system by the simple adjustment of the dition to a guard ring, it employed guard
current in this pair of coils. Thus this ar¬
rangement permits a desirable adjustment
toward higher sensitivity at large spacings
of the electrodes where the deflecting force
decreases.
The fifth of the above arrangements is
exemplified in the devices of Rogowski and
Bocker (50, 51) (illustrated in Fig. 15) in
which an elastic diaphragm D provides the
restoring force for the attracted disk. The
diaphragm supports an iron-cored coil
(above) which is actually part of a current
transformer (“Messdose”) constructed
with two air gaps in its iron core. The fixed
coil with its core form the other part of the
current transformer which is supplied from
a voltage-regulated alternating-current sup¬
ply. Very small changes in the air gap
(moving coil position) suffice to produce
full scale deflections of an ammeter con¬
nected across the moving coil and this am¬
meter is calibrated to read the high voltage.
As part of this particular moving system,
arranged for use in a compressed gas en¬
closure, a force coil (Druckspiile) K is sus¬
pended below in an iron-clad solenoid es¬
pecially constructed to permit a measure¬
ment of force in terms of solenoid current.
This arrangement permits, prior to assem¬
bly of the unit in the compressed gas cham¬
ber, a direct calibration with known weights
on the disk for current in the solenoid
against deflections of the diaphragm as in¬ Fig. 15.—Rogowski and Bocker compressed-
dicated by the ammeter. This preliminary gas electrostatic voltmeter.
282/53
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
hoops equally spaced between the guarded equally spaced hoops are maintained at
and grounded electrodes. These guard equally spaced potentials corresponding to
hoops, spaced 2 cm apart, are connected to the voltage applied between the high-volt¬
tap points on a potential dividing capacitor age (upper) and grounded (lower) plates of
also across the voltage source. Thus the the electrometer. The hoops not only screen
283/54
Feb. 15, 1948 defandorf: the measurement of high voltage
the disk from external fields, but by their Compressed gas electrometer.—A consider¬
potential distribution tend to correct, ation of the compressed gas electrometers
within their diameter, electric field distor¬ of Tschernyscheff (56), Palm (57), Rogow-
tion that woidd otherwise arise from edge ski, Bocker (50, 51), and others who at¬
effects at the upper guard ring and lower tained accuracies approaching 0.1 percent
plate. A mathematical solution of the con¬ leads one to consider whether additional
tribution to force on the disk arising from design features might be incorporated in
these hoops has been carried through by the compressed gas type in order to give
Snow (54) and Silsbee (30) and has been an accuracy approaching that of the Brooks
demonstrated to be adequate by suitable electrometer in free air.
experimental tests.6 For instance by omit¬ Actually the matter of precise weighing
ting hoops, or electrically shorting various of the forces (less than 2.5 grams) on a
sets of hoops, their potential distribution 16-cm diameter disk, involved in the Brooks
was changed drastically. For such condi¬ electrometer offers no serious problem in
tions when the corresponding corrections free air aside from the necessity of providing
were applied based on the mathematical a carefully thermostated enclosure to
solution and experimentally measured po¬ avoid air currents. The possibility of having
tential distribution, very good agreement in correspondingly higher forces to measure
the measured values of voltages was ob¬ has its appeal. In the Rogowski and Becker
tained. voltmeter the maximum force of attraction
As a result of the work on the Brooks of the disk may reach 250 grams and is
Absolute Electrometer, which was com¬ measured to 0.2 grams. Thus the forces
pared with the transformer voltmeter employed in measurements are < ne hundred
method of measuring voltage, it was con¬ times as large as in the Brooks electrom¬
cluded that this instrument is reliable for eter, however, the relative accuracy of
absolute determinations to about 0.01 per¬ measurement of this larger force is less in
cent. This work was limited to 100,000 their device.
volts—approximately one-third its rated At the outset one would have to provide
voltage—because the clearances in the a force-measuring device approaching the
space in which the equipment was housed accuracy and repeatability of a high-grade
were inadequate. Results up to full rated gravity balance, i.e., something better than
value, 275 kv, are not yet available. (With that incorporated in the Rogowski and
the cessation of the war it is expected to Bocker electrometer. Smaller-scale length
extend the measurements to higher values measurement would have to be made with
of voltage in the modern High Voltage about the same relative accuracy and the
Laboratory of the National Bureau of flat metal surfaces of the plates would re¬
Standards.) quire a high quality optical finish. The
greatest loss involved in the use of com¬
6 The simple equation (3) may be rearranged pressed gas rather than free air, as in the
VB=2dv2F/A. This equation may be thought Brooks electrometer, lies in the relative
of as being satisfactory for measurements yield¬ difficulty of making and checking mechani¬
ing an accuracy of 1 per cent. IT an accuracy of
0.01 per cent is desired, correspondingly refined cal measurements under pressure before
experimental techniques must be employed and and after voltage observations.
additional physical measurements must he made
The immediate gain from the use of
as indicated by the larger number of terms in the
complete equation for the voltage in electro¬ compressed air (15) or other gases (51B)
magnetic units of the Brooks electrometer, is to increase sparkover voltages nearly in
proportion to pressure. For instance in the
4v(c+hd) \/2Mpg I" Av — A0 design for use to 400 kv alternating, Bocker
(rb+rd) L 2MFg used carbon dioxide at 15 atmospheres
pressure which permits a gradient of 100
(VI f Zq Z kv eff/cm as compared with 2.5 kv eff/cm
in the Brooks electrometer as limited by the
- £ C.G.-W
k=1 2 \ rd) \ rhJ J present potential dividing capacitor con-
284/55
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
285/56
Feb. 15, 1948 defandorf: the measurement of high voltage
device for everyday use in the measurement lished the High Voltage Scale. A short di¬
of high voltage by deflecting a stream of gression in explanation of how this was
free-moving positively charged particles accomplished seems in order prior to -a
has not been developed, the nuclear phys¬ presentation of the contribution of the
icist has used such a device in his work for free-moving charged-particle deflection-
a number of years. To help him on his way method in this work.
in measuring high voltages he has estab¬ The study of atom physical phenomena
286/57
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
led to the physicists’ need for a voltage From plotted curves of gamma radiation
reference standard of the order of 1,000,000 against accelerating voltage (proton en¬
times the voltage of the standard cell. ergy) the rather sharp maximum of gamma
Historically, one may go back to Planck’s ray effect fixes the value of voltage chosen
equation hv = Ve wherein h is Planck’s as the resonance voltage. By selecting
constant, v is the maximum frequency cor¬ elements for these reactions in which a
responding to the maximum accelerating- single sharp maximum occurs and by care¬
voltage V, and e is the value of the ele¬ fully determining the corresponding volt¬
mentary charge. Duane and Hunt (60) ages in terms of the standard cell, the volt¬
used a steady high-vcltage storage battery ages at which particular reactions occur
in order to evaluate h from a careful meas¬ were established as reference points on the
urement of the accelerating voltage V and High Voltage Scale. In a similar manner
the corresponding maximum frequency but using a boron fluoride ionization cham¬
limit of the continuous X-ray spectrum ber (paraffin surrounding the target tube
produced. Since those measurements were and nearby chamber) instead of the Geiger-
made the value of h has been well estab¬ Mueller counter, neutron counts may be
lished by other methods. Thus this relation made for similar reactions in which neu¬
affords a method of determining V by a trons are ejected from the bombarded tar¬
measurement of the maximum frequency of get by voltage accelerated protons. In this
emitted continuous X-radiation. little use case points for the scale are determined by
has been made of this method as a high- gradually reducing the accelerating voltage
voltage reference standard largely because V and choosing that voltage at which there
of the accuracy required in evaluating the is an abrupt decrease in ionization chamber
frequency. With the advent of work on current as the limiting voltage for ejection
nuclear disintegration a pressing need for of neutrons is reached.
high voltage reference standards became Early experimental values of voltage cor¬
apparent as the usual extrapolation meth¬ responding to the resonant radiation in the
ods of measuring V, the particle accelerat¬ bombardment of lithium by protons was
ing voltage, were both cumbersome and fixed at 440 kv for Li (py) with a probable
none too reliable. In early bombardment error of 2 percent and a relative accuracy
work it was found that resonance radiation of 1 percent by Tuve, Hafstad, and Hey-
of gamma rays occurred within a rather denburg (11) of the Carnegie Department
narrow range of the voltage used in accel¬ of Terrestrial Magnetism. A redetermina¬
erating the protons. The early careful tion of voltage values for some of the fixed
measurements by Tuve, Hafstad, and Hey- points was reported in 1944 by Hanson and
denburg (11) of the voltage at which these Benedict (59), of the University of Wiscon¬
radiations occurred provided the basis for sin. Their electrostatic analyzer (a device
their adoption of certain values of voltage for determining the value of the electro¬
corresponding to the resonance gamma ra¬ static field at right angles to a beam of
diation for selected elements as fixed points charged particles that will deflect the beam
on the High Voltage Scale. a given amount) was carefully constructed
The measurement process for establishing so that its deflection constant could be
values for the fixed points on the High Volt¬ computed from dimensions as a check on
age Scale consists in: values determined experimentally. Their
(1) Providing a source of protons which are ac¬ experimental method employed an elec¬
celerated by a carefully measured adjustable volt¬ tron beam in place of the proton beam
age V. used later in their evaluation of the fixed
(2) Directing this beam of protons on a target
of a selected element such as lithium or one of its points. The use of a low-voltage electron
salts. beam (accelerating voltages of 8 to 20
(3) Measuring the gamma-ray resonance radia¬ kv) and of small deflecting voltages (150
tion effects by means of Geiger-Mueller counters
to 300 volts) permitted precise voltage
or similar devices which permit a quantitative
measurement of this radiation as a function of the measurements and higher precision for de¬
accelerating voltage. termining the deflection constant of their
287/58
Feb. 15, 1948 DEFANDORF: THE MEASUREMENT OF HIGH VOLTAGE
analyzer than the absolute method consist¬ come to us from space or may be accelerated
ing of a computation of the deflection con¬ in the laboratory by resonance techniques
stant using carefully measured dimensions. to have electron velocity equivalents ap¬
Various refinements including the use of a proaching one hundred times that value.
highly stabilized source for the deflecting Presumably the future holds in store some
voltage and automatic regulation of the new insulating arrangement for the col¬
ion-accelerating voltage reduced fluctua¬ lector of the high energy particles that man
tions arid contributed to highly precise is able to produce—an arrangement that
measurements of the fixed points on the will permit the collector to build up to the
High Voltage Scale. Hanson and Benedict unbelievable potentials we think of when
consider the following values expressed in we hear of a new machine in the rumor or
terms of the Million Electron Volt scale blueprint stage that will produce 1,000
to be accurate to 0.3 percent: Li(py) million electron-volt particles. Just what
0.4405, F(py) 0.877, Li(pn) 1.883, Be(pn) such voltages will be used for and how they
2.058. Their apparatus gave relative values will be measured I wish to leave as part of
agreeing to better than 0.1 percent. Their your field of conjecture.
values are seen to agree within the toler¬
REFERENCES
ances given for the values of Tuve, Haf-
stad, and Heydenburg although they are (1) Rayneii. Journ. Inst. Electr. Eng. 59:
138. 1921.
approximately 1.5 percent higher.
(2) A. Rosa. Bull. Bur. Stand. 3: 43.
The establishment of the High Voltage 1906.
Scale based on a phenomenon unaffected by B. Harris. Bur. Stand. Journ. Res. 3:
temperature, pressure, and humidity except 445. 1929.
as their abnormalities plague the collateral (3) Keinath. Die Technik elektrischer Mcss-
gerate 2: 17. 192S.
work of t he investigator, thus represents a (4) Chubb. Proe. Inst. Electr. Eng. 35:
distinct step forward in the process of better 121. 1916.
measurements and standards for high volt¬ (5) Palm. Elektrotechn. Zeitschr. 47: S73,
ages. 904. 1926.
(6) Ivonig. Helv. Phys. Acta 1929: 357.
SUMMARY AND CONCLUSION (7) Ilaefelv et Cie, Basle. German Patent
No. 394014. 1923.
The principal methods in use for measur¬ (8) Van Cauwenberghe and Landsberg.
ing high voltages have been outlined Bull. Techn. Assoc. Ing. Bruxelles 4.
1929.
through the discussion of a number of
(9) Davis, Bowlder, and Standring.
devices. Mention was first made of the Journ. Inst. Electr. Eng. 68: 1222.
correlation of different methods of high- 1930.
voltage measurements through extrapola¬ (10) Taylor. Bur. Stand. Journ. Res. 5:
tion techniques in which the standard cell 609. 1930.
(11) A. Tuve, Hafstad, and Heydenburg.
was used as the primary standard of volt¬ Phys. Rev. 50: 504. 1936.
age. Absolute high-voltage electrometers B. Hafstad, Heydenburg, and Tuve.
were later discussed and the good agree¬ Phys. Rev. 53: 230. 1938.
ment between their independently deter¬ C. Heydenburg, Hafstad, and Tuve.
Phys. Rev. 56: 1078. 1939.
mined values when compared with a stand¬
(12) .4. Silsbee. Bur. Stand. Sci. Pap. 20:
ard-cell voltage-extrapolation method was 489. 1925.
noted. Concluding remarks cited the use of B. Davis. Journ. Inst. Electr. Eng.
both the standard-cell voltage extrapola¬ 69: 1028. 1931.
tion-technique and a less precise absolute C. de la Gorce. Rev. Gen. d’Electr.
29: 427. 1931.
method in establishing the voltage values (13) Weller. Trans. Amer. Inst. Electr.
for certain fixed points on the High Voltage Eng. 48: 790. 1929.
Scale. (14) A. Elsner. Arch, fiir Elektr. 36: 329.
At the moment it appears that high 1942.
voltages as we know them in the laboratory
B. Arch, fiir Elektr. 35: 663.
Hohl.
1941.
stop with values of the order of 10,000,000 (15) Palm. Zeitschr. fiir techn. Phvs. 14:
volts. On the other hand, charged particles 390. 1933.
288/59
JOURNAL OF THE WASHINGTON ACADEMY OF SCIENCES VOL. 38, NO. 2
(16) Bousman and Ten Broeck. Trans. (39) Bellaschi and Teague. Electr. Journ.
Amer. Inst. Electr. Eng. 62: 541. 32:56. 1935.
1943. (40) Thornton, Waters, and Thompson.
(17) Churcher and Dannatt. Journ. Inst. Journ. Inst. Electr. Eng. 69: 533.
Electr. Eng. 69: 1019. 1931. 1931.
(18) Bellaschi. Trans. Amer. Inst. Electr. (41) Abraham and Villard. Journ. de Phvs.
Eng. 52:544. 1933. 1911: 525.
(19) Referred to in Bur. Stand. Journ. Res. 1: (42) Schumann and Klemm. Arch, fiir Elektr.
610. 1928. 12: 553. 1923.
(20) Silsbee and Defandorf. Bur. Stand. (43) Starke and Schroeder. Arch, fur
Journ. Res. 20: 317. 1938. Elektr. 20: 115. 1928.
(21) Pfiffner. Elektrotechn. Zeitschr. 47: (44) Imhof. Arch, fiir Elektr. 23: 258. 1929.
44. 1926. (45) Th. Wulf. Phys. Zeitschr. 31: 315.
(22) A. Keinath. Siemens Zeitsclir. 8: 629. 1930.
1928. (46) Hueter. Elektrotechn. Zeitschr. 55:
B. Wirz. Arch, fiir Elektr. 21: 563. 833. 1934; 56: 1319. 1935.
1929. (47) Sorensen, Hobson, and Ramo. Electr.
C. Camilli. Gen. Electr. Rev. 39: 95. Eng. 54: 651. 1935.
1936. (48) Sorensen and Ramo. Electr. Eng. 55:
(23) Imhof. Schweiz. Electr. Ver. Bull. 19: 444. 1936.
741. 1928. (49) Rogowski. Arch, fiir Elektr. 25: 521.
(24) Kuchler. Elektrotechn. Zeitschr. 58: 1931.
203. 1937. (50) Rogowski and Bocker. Arch, fiir
(25) A. Wellings and Mortloch. Journ. Elektr. 32: 44. 193S. Elektrotechn.
Inst. Electr. Eng. 79: 577. 1936. Zeitschr. 59: 123. 1938.
B. Probst. Elektrotechn. Zeitschr. 28: (51) Bocker. Arch, fiir Elektr. 33: 801.
750. 1924. 1939. Elektrotechn. Zeitschr. 61: 729.
(26) Hendricks, Hubbard, and Vallin. 1940.
Gen. Electr. Rev. 42: 420. 1939. (52) Kelvin. Papers on electrostatics and
(27) A. Caiiall. U. S. Patent No. 1979096. electromagnetism, ed. 2, sect. 360: 287.
1934. London, 1884.
B. Hendricks. Gen. Electr. Rev. 46: (53) A. Thornton and Thompson. Journ.
477. 1943. Inst. Electr. Eng. 71: 1. 1932.
(28) Kirkpatrick. Electr. Eng. 51: 863. B. Bruce. Journ. Inst. Electr. Eng. 94
1932. (2): 129. 1947. (Published sub¬
(29) Workman and Holzer. Rev. Sci. Instr. sequent to delivery of this paper.)
10: 160. 1939. (54) Snow. Bur. Stand. Journ. Res. 1: 513.
(30) Brooks, Defandorf, and Silsbee. 1928.
Bur. Stand. Journ. Res. 20: 253. 1938. (55) Nacken. Arch, fiir Elektr. 33: 60.
(31) Herb, Parkinson, and Kerst. Rev. 1939; 36: 678. 1942.
Sci. Instr. 6: 261. 1935. Phvs. Rev. (56) Tschernyscheff. Phvs. Zeitschr. 11:
51: 75. 1937. 445. 1910.
(32) Kirkpatrick. Rev. Sci. Instr. 3: 1,
(57) Palm. Zeitschr. fiir techn. Phvsik 1:
430. 1932; 5: 33. 1934.
137. 1920.
(33) ASA American Standards for Measure¬
ment of Test Voltage in Dielectric (58) A. Charlton and Cooper. Gen. Electr.
Tests, C68.1-1942. Rev. 40: 438. 1937.
(34) VerPlanck. Amer. Inst. Electr. Eng. B. Trump, Safford, and Cloud.
Techn. Pap. 41-4. Trans. Amer. Inst. Electr. Eng. 60:
(35) Loeb. Fundamental processes of electrical 132. 1941.
discharge in gases. New York, 1939. C. Schumb. Phys. Rev. £9: 692. 1946.
(36) Meek. Journ. Franklin Inst. 230: 229. (59) Hanson and Benedict. Phys. Rev. 65:
1940. 33. 1944.
(37) Meek. Journ. Inst. Electr. Eng. 93 (60) Duane and Hunt. Phvs. Rev. 6: 166.
(11-32): 97. 1946. 1915.
(3S) VerPlanck. Trans. Amer. Inst. Electr. (61) Winkelbrandt. Arch, fiir Elektr. 31:
Eng. 57:45. 193S. 672. 1937.
289/60
ratio-correction factor derived from the
Measurement of Voltage Ratio at measured quantities. The phase angle y
is the difference in phase between the
primary and the secondary voltage of the
Audio Frequencies transformer or divider.
The circuit diagram for the derivation
of the following equations is shown in Fig.
WILBUR C SZE
MEMBER AIEE
1(A).
increasingly important in computer fields. are continued successively until the de¬ \-\-jtaCiRz
Accuracies of 0.01% or better are de¬ tector D indicates a null in both the test
When detector D indicates a null, then
sired- Equipment such as precision ratio and the guard circuit.
Is—0, V v—I Zz), and Vs=IvZ\.
transformers are used in these applica¬ In addition to these two circuits there
Let
tions. However, methods have been are two auxiliary circuits: 1. the Schering
lacking for verifying the ratios of this bridge, Fig. 1(B), used for the precise Fj> = |Fpl(cos y—j sin y)
equipment to that accuracy at the desired measurement of the phase angles and 2. Vt-\Vt\ = \V,\/NF
frequencies of 100 to 10,000 cps (cycles the Wheatstone bridge, Fig. 1(C), used
per second). for the precise determination of the volt¬ Then
The purposes of this paper are: 1. to age ratios. Vp Z1+Z2
describe the use of a resistance-ca¬ The transfer from one circuit to another NF(cos y—j sin 7)=
Vs Zi
pacitance voltage divider in the measure¬ is accomplished by the arrangement of
R1FR2 -\-juRiRi( C1 + C2)
ment of voltage ratio and phase angle at switches shown in Fig. 2, described later. (D
i?l( 1 —(—Ci )
audio frequencies, 2. to describe the special The guard circuit, for both the trans¬
equipment used for the construction of the former test circuit and the Schering From equation 1 equate the real terms
circuit, 3. to analyze various factors affect¬ bridge, is arranged to eliminate the effect to get
ing the accuracy with which precision of ground admittances by completely
measurements can be made, and 4. to show
the test results obtained from
000 cps for a ratio transformer.
100 to 10,-
shielding the detector, detector circuit,
and associated apparatus.
1(^+1
N\Ri X cos y-\-o}RiCz sin
290/1
Fig. 1. Circuit diagrams Substitute Ri = Ri(Ci/Ci) in equation 3
to obtain
A—Transformer calibration cir¬
cuit RiRi
B—Schering bridge circuit u(Ci'—Ci) radians (4)
RiirRi
C—Wheatstone bridge circuit
By the same token, C2 may be adjusted
to Ci rather than varying C\. The
following is then obtained
(A)
Ri
co(Ci — Ci) radians (5)
R1-FR2
291/2
Fig. 2. Wire dia¬
gram
Test Procedure AC, Test, and Sobering. This gives the field between the working electrodes.
circuit of Fig. 1(B). Obtain a null in¬ One of these supporting structures is
The procedure in testing a voltage dication on the detector by varying C2" shown in Fig. 4. The dielectric loss
transformer or .voltage divider with this and C2 or C\ and C\. Again the guard angle of these capacitors was measured
circuit arrangement consists of three circuit must be balanced as in step 1. and found to be about 1 second at 1,000
separate steps. Steps 1,2, and 3 should follow one an¬ cps.
Step 1. Connect the test transformer other with as little delay as possible be¬ Ci and C2 are two precision variable air
in the circuit and set the selector switches tween each step, so that the variation in capacitors specially developed and de¬
in Fig. 2 to marks AC, Test, and Trf, resistance values of R\ and R-> due to signed for this setup, 20 to 120 picofarads
essentially giving the circuit shown in Fig. temperature change is small. and 50 to 1,100 picofarads respectively,
1(A). Next select the proper values of With the values of settings recorded used as shielded 3-terminal capacitors.
settings for all the apparatus in the circuit, from the procedure, the voltage-ratio The dielectric loss angles were measured
so that correction factor and the phase angle of and found to be approximately 2 seconds
the test transformer can be computed by and 5 seconds respectively at 1,000 cps.
S2/Si ~ R2/Rl Rgl/Rgl « Ci/C2«
Cl' ICC ~ Cgl/Cgi~N- 1 substituting the proper values into equa¬ Such small values are relatively unim¬
tions 2 and 4 or 5, depending upon portant in balances of the circuit as long as
Vary either Ci and R\ or C2 and i?2 to
whether C2 or Cj was varied. the corresponding conductance values
obtain a null indication on the detector.
are very small in comparison with the
Now set the center switch to Guard and
Description of Apparatus resistors Ri and R2.
adjust Cgi and Rg\ or Cff2 and Rg2 to ob¬
tain a null indication on the detector. Resistors Ri and R2 are two precision
This procedure must be repeated until The two low-loss variable air capacitors a-c decade resistors, 11,111 ohms total in
no further adjustment is required for a Ci" and C2' are iV.B5-type, parallel-plate, steps of 0.1 ohm. Their resistance ele¬
balance indication of the detector with 100 to 5,000 picofarads (micromicro¬ ments have very small residual reactances.
the switch set to either the Test position farads), modified for use as shielded 3- However, in precise a-c measurements,
or the Guard position. terminal capacitors. Their general con¬ especially at higher audio frequencies,
struction is shown in Fig. 3. One set of the effect of these reactances should be
Step 2. Turn the top selector switch
the plates is mounted on the central shaft taken into account. A special bridge
in Fig. 2 to DC; this essentially gives
and may be rotated by the shaft relative described later was therefore set up to
the same circuit as shown in Fig. 1(C).
to the fixed set. The central shaft is evaluate the change of the effective resist¬
Close the galvanometer switch and vary
electrically connected to the case. The ance over the range from direct current to
Si or S2 for a zero-current indication of other set is stationary and supported on 10,000 cps.
the galvanometer G, with the battery
three guarded quartz-rod insulators. The The null detector is connected to the
connections reversed.
guards are arranged to shield completely several bridge arrangements through a
Step 3. Reset the selector switches to the quartz insulation from the active preamplifier and an impedance-matching
292/3
Fig. 3 (left).
Three - terminal
air capacitor,
case removed
Fig. 4 (right).
Shielded support
lor air capacitor
shielded transformer; see Fig. 2. The the results desired. Corrections for exact change in frequency at the same time the
battery-operated preamplifier is a single value, temperature coefficient, and lead phase angle is measured, without altering
pentode amplifier. It is completely resistances must be applied to all the re¬ the test setup.
shielded by its metal container connected sistors. Speical precautions must be Fig. 5 shows a diagram of the circuit
to the shields for the detector leads. The taken to ensure that the leakage currents used for measuring the phase angle and
impedance-matching shielded transformer across the ratio arms are insignificant. effective resistance of the resistors R\ and
had individually shielded windings. With this procedure, the uncertainty of R2. Capacitors Cw2 and Cw 1 and a selec¬
A commercial wave analyzer was used ratio S2/S1 should not exceed 20 ppm tor switch were added to the Wagner arms
in balancing the test and guard circuits. (parts per million) or 0.002%. in order to provide for balancing the
The over-all resolution of the complete Changes in value of these standard circuits at frequencies up to 10 kc. In
detecting system is better than 1 micro¬ resistors, since their initial measurements, this figure Rs' and L/ represent the resist¬
volt. have been trivial. As a guard against any ance and equivalent inductance of the
significant unpredicted changes in resist¬ parallel-wire standard. Rx' and Lz'
D-C Wheatstone Bridge ance, a schedule has been established represent the resistance and equivalent
whereby these resistors and the resistance inductance of the resistance boxes whose
Only those special features of the d-c box can be intercompared to 10 ppm after phase angles are to be measured.
Wheatstone bridge that are essential to a about 2 hours of systematic work. To determine the a-c/d-c difference of
high order of accuracy of measurement the resistance boxes the ohmic value of the
are described. A-C Tests of Resistors unknown resistance box is set equal to
Precision standard resistors were used that of the parallel-wire standard at a low
for the Si and S2 arms of the Wheatstone Since the voltage ratio is derived from base frequency, 400 cps or less, then with¬
bridge. The resistance elements are measurements with d-c resistance stand¬ out changing the original settings of the
hermetically sealed in oil to eliminate ards, it is necessary to know the change of box, the frequency is increased in succes¬
effects from the humidity variations in the effective resistance of the a-c resistors with sive steps and the bridge circuit is re¬
laboratory. frequency. In this connection Berberich balanced with arm C for both the standard
The arrangement of standard resistors developed an excellent substitution and the box at each step.
in the ratio arms S2/Si are determined by method for determining the phase angle of The proportional change of the effective
the ratio R2/R\. These standard resistors a resistor.1 By means of a bridge circuit resistance of the computable parallel-wire
are connected in series-parallel arrange¬ he compares a resistor with a parallel-wire standard a is computed from
ments so that reference standard of computable react¬
a = (2w2LC/3) - (2co2C2f?2/15)
ance which has approximately the same
,/SiesN-1
resistance value as that of the resistor. The change d of the effective resistances
An adjustable resistance box which is The substitution method he developed of the resistance box due to frequency is
not hermetically sealed is connected in makes it simple to minimize the effect of computed from the difference of the resist¬
one arm, preferably in the arm with large parasitic influences in the circuit, does not ance settings of Rx and Rs dials at suces-
resistance values, for the fine adjustments require special apparatus, is easy to use, sive increasing frequency, i.e.,
of the d-c bridge. The lowest dial incre¬ and gives highly accurate results. It is
0 {Rx R»)/Rs
ment on the resistance box must give readily adaptable for measuring the
sufficient adjustment for the accuracy of effective resistance of a resistor with However, the effective resistance and
293/4
Fig. 5 (left).
R= 1 200A
Modified Ber-
berich's bridge
*3
■y\
£ -60 0\
X *? ]
-80
Fig. 6 (right).
Frequency cor¬
rection for GR-
type 1432-N 0 1000 3000 5000 7000 9000
decade resistor C/S FREQUENCY
100 CPS 400 CPS 1,000 CPS 3,000 CPS 7,500 CPS 10 Kc
Nominal y* 7» y9 y, 7» 7»
Ratio F Minute F Minute F Minute F Minute F Minute F Minute
294/5
can also be inverted in relation to the sistance box in arm A, and Let
295/6
Reprinted from IRE TRANSACTIONS
ON INSTRUMENTATION
Volume 1-7, Numbers 3 & 4, December, 1958
HE use of tightly coupled inductive ratio arms Standards Laboratory of Australia. By combining the
rather than resistive ratio arms in a 4-arm bridge best techniques for constructing ratio transformers,
for the comparison of impedances was suggested completely shielded 3-terminal capacitors and detec¬
by Blumlein1 in 1928, and the use of a 3-winding trans¬ tors of high sensitivity, together with a cylindrical cross
former in such a bridge circuit was described by Starr2 capacitor as a calculable standard, there is now promise
in 1932. Other bridges using Blumlein’s principle have of being able to assign values to capacitance standards
been described by several workers3 in the past 30 years. comparable with, or perhaps even better than, the ac¬
Historically, it is of interest to note that conjugate curacy assigned to our present standards of electro¬
bridges making use of 3-winding transformers were de¬ motive force and resistance.
scribed by Elsas4 in 1888 for resistance comparison and The present paper describes a transformer bridge con¬
by Trowbridge5 in 1905 for capacitance and inductance structed at the National Bureau of Standards for meas¬
comparisons. uring the direct capacitance of 3-terminal capacitors
Thus the basic principle of operation and the general ranging in values up to 1 pti and having a least count of
arrangement of transformer bridges have been known 1 Mpf- Although the transformers and network com¬
for many years. However, the possibilities of such ponents described below were designed specifically for
bridges for the precise comparison of very low value operation at 1 kc, the operation is by no means limited
capacitors had never been fully exploited before the to this frequency. Voltage output of the ratio trans¬
formers constitutes the most serious limitation at lower
frequencies, but it is reasonable to suppose that, with
* Manuscript received by the PGI, August 14, 1958.
relatively minor modifications, satisfactory operation
t National Standards Lab., Chippendale, N.S.W., Australia.
t General Radio Co., Concord, Mass. should be possible over the audio-frequency range to at
§ National Bureau of Standards, Washington, D. C.
least 10 kc.
1 British Patent No. 323037.
2 A. T. Starr, “A note on impedance measurement,” W. Eng. and While some of the present bridge components differ
Exp. W., vol. 9, pp. 615-617; November, 1932.
substantially from their counterparts at NSL, it should
3 C. H. Young, “Measuring inter-electrode capacitances,” Bell
Labs. Rec., vol. 24, pp. 433-438; December, 1946. be understood that no more is involved generally than
H. A. M. Clark and P. B. Vanderlyn, “A. C. bridges with induc¬
modifications and in some cases improvements of de¬
tively coupled ratio arms,” Proc. IEE, vol. 96, pp. 365-378; May,
1949. signs already proven by Thompson and his group
C. W. Oatley and J. G. Yates, “Bridges with coupled inductive
in Sydney.
ratio arms for the comparison of standards of resistance or capaci¬
tance,” Proc. IEE, vol. 101, pp. 91-100; March, 1954. There has been little detailed information published
A. M. Thompson, “A bridge for the measurement of permit¬
up to now concerning these components, and the present
tivity,” Proc. IEE, vol. 103, pt. B, pp. 704-707; November, 1956.
4 A. Elsas, “Ueber Widerstandsmessungen mit dem Differential- paper must be considered primarily as a discussion of
inductor,” Ann. Phys., vol. 35, pp. 828-833; 1888.
the constructional details and performance of the NBS
s A. Trowbridge, “On the differential transformer,” Phys. Rev.
vol. 20, pp. 65-76; 1905. transformer-ratio bridge.
296/253
IRE TRANSACTIONS ON INSTRUMENTATION December
Balance Conditions
Vi_C2
V ~ ci.'
2
1 1
-j -;
oiCiy «Cq
e\ Ci
€2 Ci
572772 0 - 61 - 20 297/254
1958 McGregor, et al.: New Apparatus at the NBS for Absolute Capacitance Measurement
shielding the secondary from the primary and from ex¬ cumference. Connection to the outer shield is by a tube
ternal fields, and the effect of leakage flux on the ratio coaxial with the primary-lead tube but insulated from
has been minimized by a uniform distribution of each it. The outer shield is covered with an impregnated
winding completely around the toroid. The loading error glass-fiber insulation which can withstand the heat and
resulting from the resistance of the windings has been mechanical abrasion involved in assembling the second¬
kept low by using copper strip rather than wire for the ary windings. The use of two separate shields between
secondaries. primary and secondary makes it possible to separate the
Two bridge translormers of similar design have been grounds so that, even with considerable capacitance
built, differing in some details and in the ratios of their current from the primary to the inner shield, the outer
secondary voltages. The first of these was a 10/1 ratio shield, next to the secondary, can be kept at the ground
transformer and its construction will be considered in potential of the bridge. In addition the double copper
detail. Its core is a toroid, wound of continuous 1-mil shield provides some magnetic shielding as a result of
Supermalloy tape,6 and has an OD of 4.5 inches, an eddy current action and so assists in confining the
ID of 3 inches, and a height of 1.5 inches. This Superm¬ primary flux, with the result that all the secondary
alloy has an initial dc permeability of about 75 X 103, turns more nearly link all the primary flux than would
and the thinness of the tape permits operation at fre¬ be the case without this shielding.
quencies of the order of 1 kc without serious diminution The two secondaries which provide bridge voltages
of its permeability. The large cross-sectional area of the of precise and stable ratio are wound around the
core, 1| square inches, enables the required secondary shielded and electrostatically isolated primary. As noted
voltage to be obtained with few turns and, hence, with above, a winding of about 70 turns provides the desired
low winding resistance. When operated at 1 kc in its re¬ 100-volt output. For a 1:1 ratio the secondary, winding
gion of maximum permeability (flux density about 5 would consist of two sections of 70 turns each. The
kilogausses) it produces about 1.6 volts per turn. A coupling of these sections to the primary could be made
reasonable bridge voltage is of the order of 100 volts, nearly identical by the familiar use of a bifilar winding
which is obtained with a secondary winding of 70 turns. in which corresponding turns of the two sections are
This gives good sensitivity without excessive voltage on made to occupy, as nearly as possible, the same position
the capacitors. relative to the core. Any nonuniformity in flux distribu¬
The primary winding is of rectangular copper wire tion around the core would then influence both sections
0.036 inch X0.104 inch so that 73 turns just fill the inner alike, and the ratio of secondary voltages would not be
circumference of the toroid. This heavy wire was sel- affected. For ratios other than unity the winding cannot
lected, not from any concern for low primary resistance, be truly bifilar, but the turns of the two sections can be
but to insure a uniform and nearly continuous distribu¬ distributed to effect a similar sampling of the flux
tion of magnetomotive force along the toroid. Small around the core by both. The secondary windings in the
flexible leads are connected with minimum separation 10:1 ratio transformer are distributed in the following
to the ends of the primary winding, and one of the leads manner. The secondary is divided into twenty 7-turn
is returned around the circumference of the toroid in sections. The seven turns of each section are wound as
such a direction as to cancel the field produced by the a uniform spiral around the toroid so that each section
“single-turn” effect of the toroidal winding. samples the flux around the entire toroid and ends at a
It has already been pointed out that the ratio of volt¬ point adjacent to its beginning. The ends of all 20 sec¬
ages induced in the secondary windings, which is of in¬ tions are brought out at the outer circumference in such
terest in a bridge transformer, must be as completely a manner that 10 sections can be connected in series to
independent of the primary as possible. To insure this form a 70-turn winding, and the other 10 sections in
independence, the secondary is electrostatically shielded parallel to form a 7-turn winding. Further, the.connec-
from the primary by two copper shields. The inner tions are such that the turns of the 70-turn winding are
shield is a toroidal cup of 20-mil copper fitting snugly always separated by turns of the 7-turn winding as they
over the primary and closed by a copper lid soldered progress around the toroid, thus approximating the uni¬
to the cup around the outer circumference. An insulated form flux sampling of a bifilar winding.
gap along the inner edge prevents a shorted turn. One The requirement of low error in the secondary ratio,
of the primary leads is connected to this shield, and the when the transformer is heavily loaded, necessitates a
other is brought out through a small copper tube winding resistance of only a few milliohms. To achieve
soldered to the lid. This tube and the wire inside consti¬ this, the cross section of the copper must be large, so
tute a coaxial pair of leads to the primary Winding. that edge-wound strips rather than wire are used in the
After insulation with Teflon tape the inner shield is secondary to utilize best the limited space inside the
fitted into another toroidal cup which is closed with a toroid. The requirement that 140 turns of minimum re¬
similar lid having an insulated gap around its inner cir- sistance separated by 2-mil mica strips must be fitted
into the toroid leads to dimensions of 0.040 inch X0.255
6 The authors greatly appreciate the help of J. E. Mitch, chief
inch for the cross section of the strips inside the toroid.
engineer of the Arnold Eng. Co., Marengo, Ill., in selecting and mak¬
ing available the core material. It is convenient to use stock of 40-mil thickness for all
298/255
IRE TRANSACTIONS ON INSTRUMENTATION December
the winding strips, but the width of the other three sides
of the turn can be increased to \ inch to reduce the re¬
sistance. The turns are constructed from this copper
strip by using a C-shaped piece to enclose the top, in¬
side, and bottom of the toroid and curved strips fitting
the contour of the outer circumference to space the Cs
at intervals of y of the circumference (see Fig. 3). The
strips are connected by soldering with a special clamp-
type resistance heater having graphite jaws thin enough
to fit the small gaps between turns. After assembly the
strips inside the toroid are separated by insulating strips
of 2-mil mica and on the outer circumference by strips
of 6-mil impregnated glass-fiber insulation.
The transformer is enclosed in a Mu-metal can with
the ends of all 20 sections of the secondary, and the co¬
axial primary and shield leads, brought outside the can
through slots. The connections of the secondary sections
in series and in parallel are made with additional copper
strip outside the can, where the stray field of the trans¬
former should be negligible. An outer brass case covers
the Mu-metal shield and connections and is used to
mount the BPO7 connectors through which the trans¬
former leads are brought out to the bridge and driving
amplifier.
The characteristics of the transformer can be deter¬
mined in the capacitance bridge by the intercomparison
and summation of a group of closely matched capaci¬
,0287fl 0.75/ih
tors, as described by Thompson.8 The ratio error (for
light loads) was measured using 100-pf (picofarad) ca¬
pacitors. The ratio, defined as Fio/V\ = 10(1 -\-<xJrjfi),
has a magnitude correction a = 0 3X10-6 and a phase
correction /3 = 2.7 X 10-6. The effective leakage induct¬
ance and series resistance of the two secondary wind¬
ings have been determined by measuring the ratio
change when known loads are connected across one sec¬
tion. The equivalent circuit thus determined at the BPO
connectors to the transformer is shown in Fig. 4.
A second transformer, having a toroidal core of a Fig. 4—10/1 transformer characteristics.
somewhat smaller cross section (0.94 in2) and with an
80-turn primary, was built in a similar manner. This boxes covering a range of 12 orders of magnitude from
transformer has two equal 80-turn secondaries in 8-turn 10-6 pf to 1 fJLf.
sections around the toroid, which are intermingled as The range from 1 pf to 1 /if is covered in a 6-decade
described above. The ratio of this transformer at 1 kc box having a special switching system that allows the
has a magnitude correction at light loading of 0.1 X 10~6, selection of any individual capacitor separately for cali¬
and a phase angle of 0.8 yur. The leakage inductance of bration and the addition of the capacitors iirparallel to
each secondary winding is about 1 yuh and its resistance balance an arbitrary unknown. The switching system
is 0.030 ohm. is such that each capacitor is either connected directly
between line and detector, or is grounded at both
Design and Construction of Capacitors terminals. The switch decks controlling line and detector
An unknown capacitor supplied by one secondary of connections, respectively, are completely shielded from
the transformer (see Fig. 1) may be balanced in the each other, so that switch and lead capacitances are
bridge by a capacitance of appropriate magnitude sup¬ connected to the shield system rather than to the other
plied by the other secondary. This is accomplished active electrode of the capacitor [see Fig. 1(b)]. With
through the use of two 3-terminal decade capacitance this system and with the complete shielding incor¬
porated in the design, the residual direct capacitance
7 These coaxial connectors are of British manufacture and may with all switches on zero is not observable and is cer¬
be described as British Post Office Pattern 8.
tainly less than 10-6 pf. Each decade consists of ten
8 A. M. Thompson, “The precise measurement of small capaci¬
tances,” this issue, pp. 245-253. equal capacitors, permitting quick intercomparison of
299/256
1958 McGregor, et al.: New Apparatus at the NBS for Absolute Capacitance Measurement
300/257
IRE TRANSACTIONS ON INSTRUMENTATION December
301/258
1958 McGregor, et al.: New Apparatus at the NBS for Absolute Capacitance Measurement
MICROMHOS
OR
EQUIVALENT PICOFARADS
IN+
302/259
IRE TRANSACTIONS ON INSTRUMENTATION December
their increments. The desired flexibility is incorporated method was required that would eliminate this end
into the unit by providing frequency selection switching effect from the calculation.
which changes the value of the resistor if the reading is Cylindrical gauge bars (“reference-grade” end stand¬
desired in equivalent picofarads and the phase trimming ards) of high quality were available in a variety of
capacitor if the reading is desired in micromhos. The lengths. The added requirements of uniformity and
unit is calibrated in the bridge by comparison with re¬ equality of diameter among the various bars in the as¬
sistors having known or negligible residual reactances. sembly could be met within one or two ten-thousandths
of an inch.16 The cylindrical cross capacitor, designed as
Computable Standard of Capacitance a computable standard for the present work, is con¬
structed of such gauge bars. The end sections are 2-inch
The sensitivity and range of the bridge, the accuracy
gauge bars having an axial hole for bringing shielded
of its ratio transformers, and the stability of the air
connections through to the central guarded electrode. A
capacitors described above require a better low-value
2-inch gauge bar is permanently bolted to, but insulated
standard of capacitance than has been available hereto¬
from, each of the guard bars to form a portion of the
fore at NBS.
defined central electrode. The inner bars can either be
Parallel plate and guard-well capacitors11 have been
wrung together mechanically to form a 4-inch central
used as low-value calculable standards. Plans for ma¬
electrode, or each can be wrung to an end of a 10-inch
terially improving such capacitors by using optical in¬
bar to form a 14-inch central electrode. If the guard as¬
terference techniques in determining their mechanical
sembly is not disturbed, the difference in values of the
dimensions were abandoned when it became apparent
4-inch and 14-inch capacitors should equal the com¬
that a better computable capacitor could be constructed
puted value of a 10-inch capacitor without the need for
using the electrode configuration suggested by Thomp¬
end corrections. The attainable accuracy should, under
son and Lampard.12 Lampard13'14 had shown that an
favorable circumstances,17 very nearly equal the ac¬
electrode assembly made up of equal right circular
curacy to which the mechanical length of the 10-inch
cylinders, located at the corners of a square, fulfilled the
gauge bar can be determined.
necessary conditions of symmetry, and that for certain
types of small departures from symmetry, the mean
value of the cross capacitances was not significantly O0O
affected. Further, measurements by Thompson and one ©0©O©
of the authors15 on such an assembly had shown that (a)
the re-entrant nature of the gaps between the cylinders
002 -.002
made possible substantial separation' of the electrodes -*!*-■
303/260
1958 McGregor, el al.: New Apparatus at the NBS for Absolute Capacitance Measurement
can be made using 6 as the detector bar, 1 and 3 in paral¬ Sample Data on Computable Cross Capacitor
lel as the line bars, and the remaining bars, 2, 4, 5, 7, 8, Mean Cross CAPAciTANCE, Corrected to 20°C, Referred
to Mean of Temperature-Compensated Capacitors
together with the end sections of 6 grounded. In this in¬
stance the capacitances of 1-6 and 3-6 add in parallel. Date 14-Inch Assembly 4-Inch Assembly
The difference between the balance readings taken
7-11-58 1.3897864 Arbitrary Units
with 2 and 6 alternatively used as detector bars is a 7-11-58 1.3897860
7-11-58 1.3897855
measure of any lack of symmetry in the assembly. The
7-11-58 1 .3897855
mean of the values is very closely equal to the value 7-14-58 1.3897859
computed for the capacitance.18 A close approximation 7-14-58 1.3897866
7-14-58 1.3897862
to symmetry is accomplished by using bars of equal and
uniform diameter and by separating them with equal 1 .3897860 (Average)
insulating spacers near their ends. The spacers are ar¬ 7-15-58 0.3973392
7-15-58 0.3973395
ranged in the end sections of the assembly so that there
7-16-58 0.3973400
is no solid dielectric in the electric field of the calculable 7-16-58 0.3973400
7-16-58 0.3973391
central portion. The entire assembly rests on a granite
7-16-58 0.3973399
surface plate and is enclosed in a steel housing. This 7-16-58 0.3973402
container can be evacuated to a pressure below 0.1 mm
0.3973397 (Average)
Hg, eliminating the need for an air correction to the
computed value of capacitance. 7-17-58 1.3897864
7-17-58 1.3897863
A sample set of measurements on the computable 7-21-58 1.3897866
capacitor is given in Table I. The individual obser¬ 7-21-58 1.3897862
7-23-58 1.3897856
vations are stated in terms of an arbitrary unit, the 7-23-58 1.3897864
mean of the four temperature-compensated capacitors. 7-25-58 1.3897875
2C0 = ~~8 \ Co / ’
cially express their appreciation of the skills and pa¬
tience of Chidester, Guatney, Graef, Matwey, Pararas,
where Ci, C2 are the measured values and Co is the computed value of
the capacitor. and Stadler.
304/261
JOURNAL OF RESEARCH of the National Bureau of Standards—C. Engineering and Instrumentation
Vol. 64C, No. 1, January-March 1960
(October 8, 1959)
Step-calibration methods are used in many physical laboratories for the extension of
measurements to quantities far removed from the magnitude of greatest accuracy at which
absolute determinations are made. The excellent precision of repetitive substitution pro¬
cedures is exploited by step-up or step-down methods to extend measurements to higher or
lower magnitudes without serious degradation of accuracy. The application of step-up
techniques to the calibration of variable air capacitors is described in this paper as a practical
example of the method.
305/75
bridge that will be used, or if a capacitor can be con¬ where x is the correction to the reading (or setting)
nected or disconnected by a precise switching ar¬ Xn. The calibration will consist of the determina¬
rangement. The 100-pf capacitor should be ad¬ tion of the relatively small correction, x, to each
justed close to the nominal value, but it need not be 100-pf division mark, Xn, and since only capacitance
calibrated. It must, however, be free from signifi¬ differences are of interest, the observer is free to
cant drift over the period of a quarter-hour or so choose any one of the division marks as a reference
during which the test is being run. point. It is frequently convenient to choose as a
A calibrated 1,000-pf air capacitance standard, S', reference the first marked point on the dial. In this
is needed to relate the results of the step-up test to case the 100-pf mark is considered as a reference and
the national reference standard of capacitance. a correction of 0.00 is arbitrarily assigned to it.
The bridge used for this step calibration need not For this example it will be assumed that the small
have great accuracy but must be stable, for it is used variable capacitor, V, has corrections that are
with a sensitive detector for substitution measure¬ negligibly small.
ments. A small variable capacitor, V, is required, The capacitor under test is first carefully set to
having a least count (smallest readable increment) the 100-pf mark (ACi = 100 pf) avoiding backlash
one-tenth that of X or smaller. It is advantageous errors by approaching the mark in the direction of
to choose the smallest possible precision variable increasing dial readings. The small variable capaci¬
capacitor, V, consistent with other limitations so tor, V, is set to any convenient mark near the center
that the corrections to Fare negligibly small relative of its range. The 100-pf capacitor, S, is connected
to the corrections to X. The total range of V nlust in parallel with A" and V. Now the bridge must be
be at least a little larger than the range of errors in balanced using the controls on the bridge itself If
the capacitor to be calibrated. The readable a balance cannot otherwise be obtained, V may be
accuracy of this capacitor, if expressed in percent of used to attain balance. When the bridge is balanced
total range, need not be very great. the reading VA is recorded. The fixed capacitor,
The equipment described is assembled as shown in S, is then removed and X set to the 200-pf mark
figure 1. It is most important that the cables usecl (A„2 = 200 pf), again approaching the mark in the
to connect components be shielded and rigid, or if same direction. Without changing any other com¬
flexible cables are used, it should be ascertained that ponent the circuit is rebalanced by changing V
variations in cable capacitance are negligible. The alone, and the reading VB recorded as before. In
cables must be fixed in position and must not be the first balance the external bridge arm consisted
disturbed during the entire calibration. This pre¬ of N-hA:'+ VA, and for the second balance, with the
caution is intended to emphasize the importance of bridge unchanged, the external arm consisted of
particular care to one of those sources of systematic X2-\-VB- These can be equated to yield
error that could impair good calibration accuracy.
The operator must have a good technical apprecia¬ S+Ah + VA=AT2 + VB (1)
tion of the apparatus and quantities measured,
gained through study and experience. The cable and connector capacitance, as well as resid¬
uals within the bridge, contribute equally to both
BRIDGE balances and are therefore deliberately disregarded.
It is convenient to work with small numbers, and
eq (1) can be expressed as
and since
Sn=Xn2—Xnl and x’! = 0 (3)
306/76
VB, recorded. When the first balance is equated in parallel with X and V. With X set at 10 0 pf, the
to the second balance bridge is balanced with the bridge controls and V,
if necessary, and the reading, VA, recorded. S' is
S+X2+VA=X3+VB (5) removed, X is set to the 1,100-pf mark, the bridge
or rebalanced using F alone, and the reading, VB,
recorded. Then
Sn-^S-^-Xn 2 + ^2+Fa = A’„3 +£3+FB (6)
and since
S' -\-Xi-\-VA=Xn-\-VB (12)
or
Sn~Xnz— Xn2 (7) Sn+s' JcXnl-\-VA=XnnJrxn+VB (13)
and since
Xs = X2+(Va—Vb)» + 8. (8) Sn—Xn n Xni (14)
307/77
4. Dual Calibration BRIDGE
O
W
s X u V Va-Vb X{VA- 718 X a Vb- Va 718 u ft
1
1 b)
1
0 100 1100 5.00
0.34 2.05 -1.38 0.67
100 100 1000 5.34
0.31 0. 31 -0.14 0.17
0 200 1000 5. 03
.04 1.71 -1.24 ■ 47
100 200 900 5.07
-.03 .28 — 28 .00
0 300 900 5.10
.07 1.67 -1.10 .57
100 300 800 5.17
.06 ■34 -41 -.07
0 400 800 5.11
. 24. 1.60 -0.97 .63
100 400 700 5. 35
.37 .71 -.55 .16
0 500 700 4. 98
.36 1.36 -.83 .53
100 500 600 5. 34
.02 .73 -.69 .04
0 600 600 5.32
.31 1.00 -.69 .31
100 600 500 5. 63
.05 .78 -. 83 -.05
0 700 500 5.58
.26 0.69 -.55 ■ 14
100 700 400 5. 84
.26 I.04 -.97 .07
0 800 400 5. 58
.29 .43 -■41 .02
100 800 300 5.87
.31 1.35 -1.10 .25
0 900 300 5. 56
.01 .14 —. 28 -.14
100 900 200 5. 57
.03 1.38 -1.24 04
0 1000 200 5. 54
. .13 .13 -.14 -.01
100 1000 100 5.67 •
■H 1.92 -1.38 .54
0 1100 100 5.13
“ The columns x and u are the corrections to the dial readings of the variable capacitors X and U. Th£ good precision of the method is.noticeable by comparing-the
test of Xin table 2 with table 1, which represents a test of the same capacitor about an hour earlier.
308/78
Standards of grounded capacitance, often called two- magnitudes sufficiently large that negligible errors
terminal capacitors, are characterized by having one result from differences between the connectors used
of the capacitor electrodes connected to the case, in in the several laboratories involved. Accurate cali¬
contrast with standards of direct capacitance (three- bration of small fixed capacitors can only be accom¬
terminal capacitors) having both capacitive elec¬ plished if mating connectors are submitted.
trodes insulated from the case. The direct capaci¬ The accuracy of measurements on fixed standards
tance, CD, between the two active electrodes, as of grounded capacitance having electrodes termi¬
shown in figure 3, is definite to the extent that the nated in binding posts or unshielded plugs is limited
separate terminals and associated leads are shielded by the variation in the geometrical design of the
from each other. Adequate shielding that does not instrument panels, cables, and connectors to which
interfere with the direct capacitance is relatively the standard can be attached. Differences as large
easily obtained, and excellent accuracy in direct as several tenths of a picofarad are possible with
capacitance measurements is possible to a fraction present commercially available standards and appa¬
of a micropicofarad (10-18 f). ratus with which they may be used. It is under¬
standable that differences of several tenths of a
percent are to be expected if a fixed 100-pf standard
of this type is measured in several laboratories or on
different equipment, while if measurements are
performed on 1,000-pf standards the uncertainties
at the connectors would be only several hundredths
of a percent of the quantity measured. In the
step-up procedure described in this paper it is evident
that the precision of repeated measurements, the
freedom from the effects of residuals in test apparatus
obtainable by substitution methods, and the ac¬
curacy of measurements at magnitudes closer to
optimum, are combined in a manner favorable to
the accurate calibration of the capacitance differ¬
ences of variable capacitors. The method is quite
Figure 3. The direct (three-terminal) capacitance, Cd, is
applicable at any frequency although at higher
made definite by complete shielding. The grounded (two-
terminal) capacitance, Cq, is indefinite because of variations frequencies residuals in components can be trouble¬
in the stray capacity. some and may require special attention. For
example, it may be necessary to apply corrections
for errors introduced by residual inductance in the
. The grounded capacitance, CG, shown in figure 3,
cables connecting the apparatus.
is more difficult to define in a precise manner, be¬
cause the capacitance between the ungrounded
terminal and all grounded objects, Ct, is indistin¬ 6. Conclusion
guishable from C0 unless a separate “zero-balance”
of the measuring apparatus is made with the leads Step-calibration methods can be employed for the
attached but the capacitor disconnected. Even this calibration of variable capacitors. The few neces¬
procedure will not insure good precision unless care sary items of equipment are generally available in
is used to connect the capacitor to the measuring any electrical measurements laboratory. Reference
apparatus in identically the same way every time. to the national electrical standards is made through
The best precision in practical measurements of the use of a single fixed capacitance standard that
grounded capacitance is possible only if the method can be transported to other standardizing labora¬
of connection to the measurement circuit is well de¬ tories more easily, and calibrated less expensively
fined. An adequately shielded rigid adapter or than variable capacitors.
connector is necessary as an auxiliary part of the
capacitor and must be used with it for every meas¬ 7. References
urement. If the capacitor is to be used as a stand¬
ard for accurate capacitance measurements, the same [1] Lord Rayleigh, Phil. Trans. 173, 661 (1882).
connector must be used with the capacitor when it [2] F. Wenner, J. Research XBS 25, 252 (1940). RP 1323.
is calibrated, and the assembly becomes a standard of, [3] F. B. Silsbee and F. J. Gross, J. Research XBS 21, 269
(1941). RP1419.
capacitance added to a circuit, or in other words, [4] J. L. Thomas, XTBS Circ. 470 (1948).
capacitance difference. Thus, good precision of re¬ [5] L. H. Ford and N. F. Astbury, J. Sci. Instr. 15, 122 (1938).
peated measurements is simply obtained in any
laboratory if the successive measurements are accom¬
plished using rigid wiring and the same connectors
every time. The best accuracy in terms of the na¬
tional reference standards can be obtained only for Boulder, Colo. (Paper 64C1-27)
309/79
NATIONAL BUREAU OF STANDARDS
57
May, I960
by
Thomas L. Zapf
DISTRIBUTED BY
Price $ *50
310/ i
CONTENTS
Page
ABSTRACT ii
1. INTRODUCTION 1
2. EQUIPMENT 1
5. CONCLUSION 4
REFERENCES 5
LIST OF ILLUSTRATIONS
311/ii
VARIABLE CAPACITOR CALIBRATION WITH AN
ABSTRACT
The use of an inductive voltage divider bridge for the calibra¬
discussed.
1. INTRODUCTION
2. EQUIPMENT
accuracy of ratio of the arms thus formed often far exceeds that
tances to ground are then across either the detector or the ratio
312/1
more than reduce the sensitivity, and the effect on accuracy of
loading the closely coupled ratio arms, which have very small
The shields of these cables, the other terminal of the detector, and
bration point and the inductive voltage divider adjusted until the
If good quality air capacitors are used in the above bridge cir¬
572772 0-61-21
313/2
in series with one or the other (as needed) of the capacitors in the
sary to place a guard cap over, but not touching, the unused terminal.
314/3
capacitor as shown in Figure 4. Capacitance differences corre¬
5. CONCLUSION
315/4
REFERENCES
3 (1959)
Errara to Accompany
316a/6
Journal of Research of the National Bureau of Standards Vol. 48, No. 2, February 1952 Research Paper 2296
Thermal converters and associated equipment that are used as ac-dc transfer standards
at the National Bureau of Standards for the precise measurement of current and voltage at
power and audio frequencies are described. The standards and the equipment are primarily
used to standardize a-c ammeters and voltmeters submitted to the Bureau for certification.
The ac-dc transfer may be made with these thermal converters at currents from 1 milliampere
to 50 amperes, voltages of 0.2 to 750 volts, with an accuracy of 0.01 percent at frequencies
from 25 to 20,000 cycles per second.
The special tests to insure the required accuracy of the transfer standards are described,
and the results are presented. A number of factors that limit the transfer accuracy of ther¬
mal converters have been discovered, and the results of special tests and theoretical work to
evaluate these factors are discussed. The solutions, by an approximation method, of certain
pertinent nonlinear differential equations governing the heating of a conductor by an electric
current are given.
317/121
with a test on direct current gives more information transfer voltmeter. A careful experimental and
about the performance of the test instrument. theoretical study of their performance led to the
Because the ac-dc difference of an instrument de¬ purchase of additional commercial thermal converters
pends upon geometrical factors that are relatively of suitable ranges and the design and construction of
permanent, subsequent tests for checking the con¬ equipment for using them as transfer standards for
stancy of calibration need generally be made only voltage and current.
on direct current. Occasionally transfer tests arc
made by using a selected low frequency (such as 3.2. Description of Transfer Standards
60 c/s) as the reference in place of direct current.
An example of the accuracy with which such Each thermal converter purchased in the milli-
tests are made, laboratory standard instruments, ampere ranges is of the vacuum type, mounted in an
with scales 12 in. long, are regularly standardized evacuated glass bulb with supports of copper wire
and certified at the Bureau to 0.05 of a scale division. embedded in the glass. The conductor, heated by
For a 150-division instrument this requires an the current to be measured, is a short straight wire
accuracy of measurement of 0.03 percent or better. generally less than ){ cm long and often less than
0.001 in. in diameter. The heater alloy and dimen¬
3. Development and Description of Electro- sions are chosen to give a temperature rise of about
thermic Transfer Standards 200° C at the center where the hot junction of the
thermocouple is fastened by a ceramic bead. This
3.1. Choice of Standards bead provides electrical but not thermal insulation
between the heater and thermocouple circuits, with a
coupling capacitance less than 1 /x/xf and an insulation
Several types of instruments were considered in
choosing transfer standards to cover the full audio¬ resistance greater than 50 megohms at 25° C.
Because high current sensitivity is not required,
frequency range. Electrodynamic instruments have
been highly developed [1, 2, 3]2 at the Bureau for the thermal converters in the 1- to 50-amp ranges are
measurements at the commercially important power not evacuated. The hot junction of the thermo¬
frequencies and can be used with proper corrections, couple is welded directly to the midpoint of a thin-
up to about 2,000 e/s. However, unavoidable in¬ walled tubular heater. These thermal converters
ductance and stray capacitance errors limit their are temperature compensated [6].
useful frequency range and present little hope for the For current measurements, thermal converters in
desired extension. Electrostatic instruments have the series 1, 2, 5, 10, etc., with rated output electro¬
been carefully studied [4, 5] at the National Physical motive forces of 10 mv were purchased, with ranges
Laboratory. They are useful over wide frequency from 1 ma to 50 amp. Thermal converters of 7.5-
and 30-ma ratings, with appropriate series resistors,
ranges, but because of their low-torque—weight ratio
at low voltage are not readily adaptable for either are list'd for voltage measurements. Figures 1 and 2
the low voltage or the current ranges desired in this show some of the thermal converters and the model
application. Electrothermie instruments, which use A and model B voltmeter elements. It should be
an effect produced by the heating of a conductor emphasized that the caption NBS on the nameplate
carrying a current to be measured, seemed more shown in figure 2 signifies only that the equipment
promising. Of the several kinds, thermocouple in¬ was assembled for use at the National Bureau of
struments, in which the temperature rise of the con¬ Standards. All the thermal converters were pur¬
ductor (heater) is measured by a thermocouple, chased commercially, as were the components of tin1
seemed most feasible. Straight-wire heaters allow voltmeters. It should also be emphasized that the
a wide frequency range, and the thermocouple thermal converters shown in these figures are the
measurements permit high precision of reading. The transfer elements only. An indicator (to be dis¬
usual disadvantages of poor stability and large tem¬ cussed in the next section) is necessary for all meas¬
perature influence are not important in their use as urements, and a potentiometer and accessories are
transfer instruments, and their low overload capac¬ also necessary, except for transfer tests of other
ity is not a serious limitation in careful laboratory instruments.
The model A voltmeter was developed as a proto¬
work. Fortunately, good thermal converters are
commercially available. They have been used for type instrument to meet an immediate need for
voltage measurements. It consists of a 30-ma
measurements at frequencies up to about 65 Me, with
thermal converter connected with its heater in
a d-c millivoltmeter as the indicator, and are gener¬
series with a commercial decade resistance box.
ally considered in the L- to 2-percent accuracy
classes. So far as is known, they have not been modified to have two I .()()()-ohm-per-step decades,
previously studied for use as transfer instruments a t and a 1 ()()-, a 10- and a 1-ohm-per-step decade, all
the frequencies and the accuracies needed in this in series. As shown in figure 3, one end of the heater
application. of thi' thermal converter is connected to the low-side
! terminal of the box, with the shield of the box con¬
For these reasons, commercial thermal converters
nected to the other end of the healer and to the
were selected for study to determine their value as
j shield of thi' coaxial cable connecting the instrument
ac-dc standards in a wide-range audio-frequency
to the circuit. The center lead of this cable is con¬
• Figures in brackets indicate the literature references at the end of this paper. nected directly to the high-side terminal of the box.
318/122
mounted in a suitable shielded box with the shield
connected to the low-side terminal of the instrument,
as shown in figure 3. For both instruments the
effects of capacitance currents and of the self-
inductance of the resistors were carefully considered;
the limiting factors are discussed in another section
of this paper.
Wv^^\AA^vL^X\Jv^A^
L__ _I
B
Figure 3. Circuit diagrams of the model A and model B volt¬
meter elements.
A, Model A voltmeter element; B, model B voltmeter element.
3.3. Indicator
319/123
meter, which generally sets the limit on the accuracy 3.4. Use of Instruments
obtainable with this form of potentiometer.
Such a potentiometer has been incorporated in a These transfer standards are used almost solely for
panel constructed for the audio-frequency testing. It standardizing other a-c instruments. The necessary
was designed to have adequate stability, freedom from switching and control circuits for such tests have been
changing extraneous electromotive force and the re¬ incorporated in the panel shown in figure 5. In
quired ranges, with sufficient precision of reading. Its addition to these circuits and the Lindeck potenti¬
circuit is shown in the central portion of figure 4. ometer, the panel contains the impedance-matching
The resistor, R, is made of manganin, and a special transformers for the high-voltage circuits used in
tliermofree key [8] and copper binding posts are used testing voltmeters and the high-current circuits used
in the electrothermally sensitive emf circuit. The in testing ammeters. The present transformers pro¬
components are mounted in the central part of the vide ranges up to 50 amp and 800 v, with a nominal
panel, as shown in figure 5, and are enclosed by a input voltage of 150 v and a power level up to 100 w.
C signifies coaxial lead with shield grounded; S signifies shielded lead with shield grounded.
grounded metal shield. No other thermal pre¬ Motor-generator sets and voltage stabilizers are
cautions were found necessary. A reflecting wall- available for tests at 60 c/s and audio-frequency
mounted galvanometer is used, with a voltage sensi¬ oscillators and power amplifiers for tests at other
tivity (in the circuit) of 1 mm//iv. A Xo. 0 dry cell frequencies. Batteries, controlled by suitable ad¬
supplies the potentiometer current, which is adjust¬ justable resistance voltage dividers and series resis¬
able from 0.2 to 1.2 ma. At these current levels, tors, are used for the necessary direct-current sources.
repeated tests have shown that the steady drift in tin* The potentiometer, standard cells, volt boxes, and
voltage across the four terminal resistor is less than resistors are those normally used for instrument
0.02 percent per hour, and that superposed fluctua¬ testing and are periodically standardized to insure a
tions are much less than 0.01 percent. The four- continued accuracy of considerably better than 0.01
terminal manganin resistor has taps at 0.1, 0.5, 2,5, percent.
5, 10, and 25 ohms, giving milivolt ranges, at a A schematic diagram of the major circuits and
current of 1 ma, of these same numerical values. equipment on this panel is shown in figure 4. For an
The lowest range is used in transfer tests as a quick a-c test of an ammeter, the transfer thermoelement
check of the potentiometer. With the thermocouple is connected to the AC-DC binding posts and its
of the thermal converter connected to the “TC” thermocouple to the TC posts on the panel. The
posts of the potentiometer, but with no current test instrument is connected to the AC posts, and a
through its heater, a galvanometer deflection of 10.0 suitable four-terminal standard resistor is connected
cm to the left on the scale with the Lindeck set for 0.1 to the DC posts for use in measuring the d-c current
mv serves as a valuable partial check of the required by means of a potentiometer. For low currents at
accuracy and correctness of the Lindeck circuit. tlie higher audio frequencies, shielded leads with the
320/124
shields grounded are used to minimize errors due to readings. This is fortunately true for each of the
stray capacitance currents. The alternating current thermal converters used. After an initial warm-up
is adjusted for the desired deflection of the test instru¬ period of a minute or so, the maximum observed
ment, and the Lindeck potentiometer is adjusted for drift under those laboratory conditions of any of the
an “on-scale” deflection of the galvanometer. The thermal converters tested was less than 0.5 percent
standard instrument is then switched to direct cur¬ per hour, and the average less than 0.05 percent per
rent, which is adjusted to give the same galvanometer hour. If it were necessary, the effect of significant
deflection and is then measured with the external drift could be practically eliminated by using a deflec¬
potentiometer. The direct current through the tion method as outlined for transfer tests, calibrating
heater of the thermoelement is then reversed and the the standard instruftient on direct current at the
d-c measurement repeated.1 * 3 nominally correct current or voltage before and after
each a-c setting.
For ac-dc difference or transfer tests, which are
more frequent and usually more important than
straight a-c tests, the test and standard instruments
are connected in series to the AC-DC posts 4 in the
case of ammeters, or in parallel with suitable coaxial
leads to the AV-DV receptacles in the case of volt¬
meters. Suitable short leads are used to connect the
two AV receptacles together and to connect the two
DV receptacles together. The current or voltage is
adjusted for the desired deflection of the test instru¬
ment, and the Lindeck potentiometer is adjusted for
a deflection near the null position at the center of
the galvanometer scale. The two instruments are
then in quick succession connected to alternating,
direct, reversed direct, and then alternating current.
The current is adjusted to give the same deflection
of the test instrument in each case. The resulting
deflections of the galvanometer of the Lindeck poten¬
tiometer are observed with the potentiometer con¬
trols undisturbed. This procedure practically elim¬
inates any effect of drift in either instrument, and
errors in repeating the settings of the instrument
under test can be made very small by using a low-
power microscope and setting the image of the
instrument pointer in coincidence with a cross-hair
in the eyepiece. The control circuits have been
arranged to facilitate switching, and with two
observers the procedure is rapid even with moder¬
Figure 5. Instrument testing panel, including Lindeck
ately fluctuating voltages as only one of the two
element.
instruments is set to a given deflection.
From the average of the deflections on alternating
For an a-c test of a voltmeter, a similar procedure
and on direct current, the ac-dc difference of the test
is used. The transfer standard is connected to the
instrument is readily computed by a simple formula.5
AV-DV receptacle on the panel with a coaxial lead,
with due care to connect its thermal converter to the 4. Evaluation of Transfer Performance
grounded shield of this lead, and the thermocouple
is connected to -the TC posts on the panel. The The ac-dc difference of each transfer standard
test instrument is connected to the circuit with a must be known to the full accuracy desired in the
special twin lead consisting of two coaxial lines per¬ measurements to be made. As, in a sense, there are
manently paralleled at the instrument and connecting no absolute transfer standards, such an evaluation
to the two AV receptacles on the panel. Similar essentially consists of four steps.
leads of the same length are used to connect the high 1. Choice of a type of instrument that theory in¬
side of a volt box for use with the external poten¬ dicates is suitable over the desired ranges.
tiometer to measure the direct voltage, to the two 2. Study of all known effects that can cause such
D\ receptacles. The use of these leads eliminates an instrument to depart from this ideal.
the effect of any stray magnetic field or lead resistance 3. Construction and use of an instrument in such
on the measurements. a way that these effects, by computation or
This simple procedure is reliable only if the drift direct tests of the separate components, are
of the response of the standard instrument is negli¬ negligible or known.
gibly small in the time taken for a series of such 1 One terminal of the heater is connected to the post that is grounded on
alternating current.
3 Such reversed d-c readings are common practice for highest accuracy with all 3 The formulas used in computing the results of the a-c and transfer tests are
types of ac-dc instruments. developed in appendix 1.
321/125
4. Comparison of actual transfer performance of cation, and their direct-current and low-frequency
the standard with that of some other instru¬ responses were verified by special tests described in
ment, preferably of a quite different type, other sections. The effective impedance of each
whose performance has also been evaluated. voltmeter element was approximately computed
The evaluation of these electrothermic transfer from inductance and capacitance data supplied by
standards was considered of primary importance the manufacturer of the resistance cards, and was
because of the high order of accuracy demanded in verified by special tests to be described later. The
this application. The unique facilities of the Bureau capacitance and resistance of the Lindeck potenti¬
made this quite feasible, particularly for step 4, ometer to ground were measured and found to be
which is necessary to guard against errors from un¬ 1,600 ppi and 10,000 megohms at an ambient relative
suspected causes. Prototype instruments such as humidity of 40 percent. Thus the computed stray
these, in which previously available ranges are currents are sufficiently small so that no appreciable
greatly extended, cannot be directly compared with fraction of the currents measured will be diverted
other types of standards over their full range. The through the thermocouple at audio frequencies when
evaluation therefore consisted of the study, compu¬ one end of the heater is grounded. The use of
tations, and special tests of steps 2 and 3 during the shielded leads where necessary, the avoidance of a
design and construction of these instruments and ground between the test instrument and the standard
their panel, the intercomparisons of thermal con¬ in the current circuits, and the use of the special co¬
verters of adjacent current ranges in the series, the axial lines in the voltage circuits assure that the
comparison at low frequencies of selected thermal fundamental principle is observed even at the
converters with the electrodynamic transfer stand¬ smallest current and voltage in -the present ranges.
ards at the Bureau, and the comparison of these The comparison tests of step 4 were planned to
thermal converters with certain portable hot-wire give maximum possible assurance against errors due
and electrostatic instruments also available. to unsuspected causes. Each comparison consisted
A thermal converter can be used as a transfer of repeated careful transfer (ac-dc difference) tests,
standard for alternating-current measurements over as described in section 3.4. The 1-, 2- and 5-amp
the range of frequencies for which the response char¬ thermal converters were compared with the NBS
acteristic, E=j{I), is the same as with direct current , electrodynamic transfer ammeter at selected fre¬
where E is the output emf, and / the current through quencies from 25 to 200 c/s. The model A voltmeter
the heater of the thermal converter. For a volt¬ was compared with the NBS suppressed-zero electro-
meter element consisting of a resistor in series with a dynamic voltmeter at frequencies of 30 and 60 c/s and
thermal converter, the magnitude of the effective im¬ voltages from 20 to 240 v. Both of these electro-
pedance Z= V/I must also be equal to the effective dynamic instruments have been carefully studied
resistance on direct current, where V is the applied [1, 2], They can be read with a precision of better
voltage. In addition to these two requirements for than 0.005 percent. The results of these compari¬
instrument testing, the fundamental principle that the sons, corrected for the small known transfer errors of
standard and the instrument under test “see” exactly the electrodynamic instruments, are shown in table 1.
the same quantity must be carefully observed. They are estimated to be accurate to 0.005 percent or
The response characteristic of the types of thermal better. The results of these tests disclosed that the
converters used were believed to be suitable at fre¬ model A voltmeter and the electrodynamic voltmeter
quencies far higher than those used in this appli¬ were in excellent agreement at these frequencies.
amp amp % o? % %
% % /V % % %
1_ Electrodynamic ammeter.__ 0.4 -0. 001 +0. 001 -0. 001 -0. 002
i _ . _do 1 -.053 — -.008 -.001
i_ do _ _ _ 1. 2 -.067 —. 013 —. 0C8 -.005
2_ _do 0. 8 -.022 -.001 +.001 + .001 —
2_ _ _ do . 2 115 -.025 -.009 + .001
2_ _do 2.4 -. 154 — -.035 -.015 -.002
5_ _do__ _ . 4 -.004 -.006 -.005
5_ __ do... . _ 5 -.010 — -.005
5_ ..do . . 6 -.011 -.018 -.008
5_ 4 -0. 006
_ do _ 5 — -.007 -0.004 -.002
V
Model A voltmeter.. Electrodynamic voltmeter 20 -.004
Do_ _ do 50 — -.005 —
Do_ . do 120 -0. 002
Do_ __do.__ __ _ . 150 — -.003 —
Do_ _do. 240 -.002
Do_ Electrostatic voltmeter 150 +.004 -.002 +i004 "-.004 -.002
Do_ __ .do_ _ _ 300 +.004 -.004 -.010 -0. C08 +.009
322/126
They disclosed significant disagreements between tlie Table 2. Results of intercomparisons of thermal converters
ampere-range thermal converters and the electro-
dynamic ammeter, which were considered to be due Comparative ac-dc difference
Test
to a low-frequency error in the 1- and 2-amp thermal Thermal converter
cur¬
ranges
converters and a small error independent of frequency rent 25 100 400 1,000 10,000 20,000
c/s c/s c/s c/s c/s c/s
in the 5-amp thermal converter.
The model A voltmeter was also compared with ma or or Of c7
/c /V /c %
two new high-grade portable pivoted electrostatic 1 to 2 tna_ 1. 2 /ooo 0. 000 +0.006
2 to 5 ma.... 2. 4 +. 006 +.003 -.003
voltmeters, having scales over 5 in. long, at selected 5 to 10 ma... 6 +.004 . 000 .000
frequencies from 30 to 20,000 c/s. For increased 10 to 20 ma.. 12 .000 — +.004 -.006
20 to 30 ma* 24 +.001 -0.003 -.001 . 000
precision a low-power microscope was focused on
30a to 50 ma... 36 +.004 . 000 . 000
the pointer of the electrostatic instrument under 50 to 100 ma... 60 +.024 +0. 028 +.036 +. 027
test. This instrument was always set to the same 100 to 200 ma.. 100 -.046 — -.044 -0.050 -.049
200 to 500 ma__ 240 +.026 +. 039
deflection and was tapped lightly before each read¬ 500 to 1,000 ma 600 -.002 -. 002 +. 002 +.002
ing. Repeated sets of careful readings were taken amp
at each frequency, and all tests were at the full-scale 1,000 ma to 1 amp. 1.2 +.034 —. 065 -.056 -.065
1 to 2 amp_ 1 -.028 -.002 -.002 -. 002 +. 001
deflection of the electrostatic instrument. Because 2 to 5 amp_ 2 -. 123 -.011 _ +.003 +.004 +.009
5 to 10 amp_ 5 -.008 -.006 -.006 -.007 -.004
of these repeated readings and the favorable condi¬ 10 to 20 amp_ 10 -.010 -.010 -.008 -.012 -.009
tions, the averaged results are estimated to be accu¬ 20 to 50 amp_ 20 -.014 -.015 -.014 -. 018 -.035
323/127
response of the thermal converter on alternating and Table 5. Results of comparisons of model A and model B
on direct current the response of the shunted model thermal voltmeters
A voltmeter was greater on alternating than on
direct current. The results of tests of the 1-, 2- and ac-dc difference of model B voltmeter
5-amp thermal converters, duplicating those with
Voltage
the electrodynamic ammeter, are not shown in the model B
25 1,000 5,000 15,000 20,000
c/s c/s c/s c/s c/s
table. The average difference between eight such
duplicate results obtained with this ammeter and
Volts V % % % % %
with the shunted model A voltmeter was 0.004 per¬ 3 through 75.. Rated_ O) M 0\> (“) V)
cent, and the largest difference was 0.013 percent. 150__ _do_ _ +0. 013
300.. _do. +.02x
A few additional check tests of these thermal con¬ 600.. 400_ +0.0i4 +6. 044 +.05x
750_ 400_ +0.003 +o.oii +. 028 +.064 +. 082
verters with the shunted voltmeter were made at
other frequencies but are not tabulated.
• Less than 0.01 percent, 25 to 20,000 c/s each range,.
The results of all these tests were combined to
assign reliable values of ac-dc difference, 5, to each
As a result of these extensive series of tests, the
thermal converter in the series from 1 ma to 50 amp transfer performance of each of this first set of ther¬
and to the model A voltmeter. These values were mal converters and of each of the two voltmeters
less than 0.01 percent for the model A voltmeter and was uniquely determined. Small but significant ac-
for all the thermal converters of ranges less than 100 dc differences of both known and unsuspected causes
ma. The values for the other converters (rounded
were found. Corrections for these differences can
off to the nearest 0.005%) are shown in table 4. readily be applied when these thermal converters are
used as transfer standards. However, an investiga¬
Table 4. Observed 'percentage ac-dc differences of thermal
converters (to the nearest 0.005 percent) tion of such differences is an especially important
objective of a development such as this. As a result
of such an investigation, a number of factors that
(a) Differences independent of frequency
limit the transfer accuracy of an electrothermic in¬
strument were discovered and evaluated, with the
Percentage, of rated current
Converter results given in the following sections.
range
40 50 60 80 100 120
ma 5. D-C Response
1 through 50. (») (a) (B) (*) (•) (*>
100. -0.030 -0.060 -0.065
200. O) C) (“) (“) (■) In all measurements with a transfer standard the
500...- -0.0*5 -0.055 -.070
1000_ -.040 "-.060 -.060 average of the d-c response of the standard for the
amp two directions of current is taken as the reference upon
5.. .000 .000 -.005 .010 -.015
in. -0.005 .015 -.020
which the a-c measurements are based. For all
20. -.005 . 0x5 -.025 three types of such standards, the reversed d-c aver¬
50. -.010 -.010
age gives a better basis than does the response for one
direction alone as effects that lead to small differ¬
(b) Differences dependent upon frequency
ences in response for the two directions of direct cur¬
rent are reduced to the second order of smallness
Frequency (cycles per second)
when the average is taken.
range current
25 60 100 200
Reversal differences may be expected in a contact
thermal converter because of the flow of heater cur¬
* amp amp rent through the hot junction of the thermocouple.
(0.4 -0.0i0 0.000 0. 000 0. OCO
1__.. h.o -.055 -.010 .000
This can cause a voltage drop in the thermocouple
b-2 -.065 -.015 -.010 -.005 circuit, which may lead to second-order errors be¬
(0.8 -.020 .000 .000 .000 cause of the nonlinearities of the thermal and electric
2.. •,2. 0 -. 120 -.025 -.010 .000
[2.4 -. 155 -.035 -.015 .000
circuits. This cause of reversal difference is mini¬
mized by the special constructions used in most
» Less than 0.01 percent throughout. high-range thermal converters, and is eliminated by
the use of the small insulating bead between the
The model B voltmeter, which was developed to heater and the thermocouple in the low-range ther¬
give a wider voltage range and greater convenience mal converters. Surprisingly, however, early tests of
of use than the model A, was compared with thd the first thermal converters purchased for the model
model A voltmeter at rated voltage on each range A voltmeter showed appreciable reversal differences
up through the 300-v range and at 400 v on the higher (up to 0.2%) even in these insulated-heater thermal
ranges, at selected frequencies from 30 to 20,000 c/s. converters. It was soon realized that this was prob¬
The results are shown in table 5 and represent ably due to thermoelectric effects in the heater cir¬
directly the ac-dc difference, 5, of the model B volt¬ cuit. Peltier and Thomson heating of the heater can
meter. They disclosed significant but not unex¬ cause dissymmetry in the temperature rise of the
pected errors for the higher voltage ranges at the heater, and thus cause differences in the emf of the
higher frequencies. thermocouple unless the hot junction is exactly at the
324/128
midpoint of the heater.8 This supposition was veri¬ tribution is unchanged by these thermoelectric
fied by an analysis of the data, which showed that effects. Calculations show (see appendix 2) that
the differences were larger in those thermal converters with direct current the temperature rise at the
with heaters having larger thermoelectric effects. midpoint of the conductor, where the hot junction
Further verification was secured by tests of special of the thermocouple is attached, is unaffected by
thermal converters, constructed by one of the manu¬ the Peltier heating, for completely symmetrical
facturers 9 at the Bureau’s request, with the hot construction of a thermoelement. They show,
junction of each deliberately and considerably off- however, that it is changed by Thomson heating,
center. The reversal differences were much larger, and becomes, approximately,11
and the signs and magnitudes were found to be
roughly in agreement with calculations based on
thermoelectric data for the alloys used. <2)
Early intercomparisons of these few thermal
converters indicated that no significant ac-dc dif¬ where 6t is the midpoint temperature rise in the
ferences should be expected, even with rather large presence of Thomson heating, in °C; 9d, the rise
reversal differences from this cause, provided that without Thomson heating; <7, the Thomson voltage
the average for the two directions of current is coefficient in volts-degree _1, Vc, the voltage across
taken as the d-c response. However, intercom¬ the conductor in volts.
parisons and tests of the much larger number of This was derived by considering a a constant,
thermal converters later purchased for current whereas both experimental results and the electron
measurements showed that a good proportion had theory of metallic conduction indicate that a—BT,
measurable ac-dc differences, all of the same sign where B is a constant and T is the absolute tempera¬
and independent of frequency. This suggested ture. However, a solution by a perturbation
a d-c error, for which these thermoelectric efforts method,12 of the nonlinear equation that results when
seemed the most likely cause. This was inves¬ <j=BT, shows that the above results give a suffi¬
tigated theoretically (see appendix 2) by consid¬ ciently close approximation for the temperature rise
ering the temperature rise of a homogeneous con¬ of about 200 deg C encountered in typical thermal
ductor heated by an electric current and cooled converters. Thus eq 2 shows that an ac-dc difference,
solely by conduction to two relatively massive or transfer error, should be expected in a thermal
terminals; i. e. by neglecting, for simplification, converter, and enables the approximate calculation
of this error. Such calculations have been made for
radiation and convection losses and the heat ab¬
some common metals and some of the alloys used as
stracted by the attached thermocouple.
heaters in thermal converters, on the assumption of
With only the normal joule, or resistance, heating,
a 200 deg C rise at the midpoint, and a voltage drop
and with the terminals of the conductor at equal
of 0 .2 v at rated currents.13 The results, converted to
and fixed temperature, the temperature distri¬
ac-dc difference for the same temperature rise, rather
bution along the conductor is parabolic, and the
than temperature difference for the same current, are
temperature rise at the midpoint is, as Goodwin
shown in table 6. For the computation of these
showed,10 equal to
results, the Thomson coefficient <r was considered to
be equal to B(d0-\-2/36d-\-273), where B is the second
derivative of the characteristic curve of emf versus
temperature for a thermocouple composed of the
where Vc is the voltage drop across the conductor, indicated metal and lead.14 60 is the ambient temper¬
and p and k are the electrical resistivity and ther¬ ature in deg C. The results show approximately the
mal conductivity, respectively. In addition, how¬ errors to be expected for these materials.
ever, there is Peltier heating or cooling at the junc¬ Table 6. Calculated ac-dc difference at rated current due to
tions of dissimilar metals (the two junctions between Thomson effects in thermal converter heaters
the heater and its two supports) and Thomson
heating along each half of the wire. Unlike joule ac-dc
Metal B
heating, these are dependent upon the direction difference
325/129
The results may be compared with the observed When a constant direct current is applied, the
ac-dc differences shown in table 4. From information heater of a thermal converter reaches a constant
supplied by the manufacturers, the 1-, 2-, 5-, 10-, 20-, temperature. When an alternating current is ap¬
50-, and 200-ma thermal converters have heaters of plied (after the transient state is over) the tempera¬
either carbon, nickel-chromium, or platinum- ture of the heater varies cyclically about a mean
rhodium-ruthenium alloys. The observed ac-de dif¬ value, with an amplitude that decreases with in¬
ferences of these converters were less than 0.01 per¬ creasing frequency because of the thermal capacity of
cent. The 100-, 500- and 1,000-ma thermal convert¬ the element. At frequencies high enough so that the
ers had heaters of constantan (Advance). The cyclic variations are negligible, the temperature rise
observed differences for these converters (at rated of the heater is independent of frequency (provided
current) were close to the calculated value of —0.06 that at very high frequencies the skin effect and
percent. The 1- and 2-amp converters, of platinum- electromagnetic radiation are negligible and that
iridium, had observed differences (from table 4,b) standing waves do not alter the temperature distribu¬
less than 0.005 percent, except at low frequencies, tion). However, at lower frequencies the mean
also close to the calculated values. The 5-, 10-, 20- value, to which the d-c instrument or potentiometer
and 50-amp converters, of a special platinum-3- connected to the thermocouple responds, may not be
percent-copper alloy, had differences of —0.010 to the same because of nonlinearities in the thermal and
— 0.015 percent. Thus the agreement between the electric circuits. These nonlinearities are caused
computed and observed values serves as excellent principally by radiation losses, the dependence upon
verification of this source of error in transfer thermal temperature of the electrical resistivity and the ther¬
converters. Very fortunately, the error is signifi¬ mal conductivity of the heater, and the curvature of
cantly large only for materials having unusually the emf versus temperature characteristic of the
large thermoelectric effects, such as constantan. thermocouple. A theoretical evaluation of these ef¬
However, this does lead to the interesting conclusion fects for the conditions in a thermal converter, in
that for operation at a low temperature a thermal which most of the cooling is by conduction to the
converter should have large thermoelectric effects heater, is difficult, but does serve as a guide in deter¬
in the thermocouple, but that for the highest transfer mining the approximate limiting frequencies for an
accuracy it shoidd have very small thermoelectric arbitrarily chosen error and for determining which
effects in the heater. factors influence that error. Such an approximate
Additional experimental verification of this cause analysis has been carried out by a perturbation
of transfer error was secured by constructing15 at the method (see appendix 3). The results show that the
Bureau two 5-amp thermal converters differing only average (time average) temperature rise at the mid¬
in that one had a heater of manganin and the other of point of a conductor heated by a sinusoidal electric
constantan. These were 5-amp thermal converters current and cooled solely by conduction to massive
with a small Borax bead between the hot junction of terminals is approximately
a copper-constantan thermocouple and the center of
the heater wire, to eliminate contact effects. Careful
ac-dc difference tests with the standard electro¬
dynamic ammeter as the reference established agree¬ and that the percentage ac-dc difference of a thermal
ment within 0.005 percent for the manganin element converter having the hot junction of a thermocouple
and indicated an ac-dc difference of —0.06 percent fastened to the midpoint of such a heater is approxi¬
for the constantan element at an emf of about 5 mv.
mately
As a result of this work, a second set of thermal 5= —25Hddcq2 (4)
converters in the 1- to 1,000-ma range was purchased
where
under specifications prohibiting heaters of constantan 0ac=tbe midpoint temperature rise in de¬
or similar alloys. All of these converters were found grees centigrade on alternating cur¬
to have ac-dc differences less than 0.01 percent at rent of rms value / amperes
audio frequencies. 0dc=the midpoint temperature rise in de¬
grees centigrade for the same value
6. Low-frequency Response of direct current.
h = a—/3—2NTol2, where a is temperature
It lias long [10] been known that there is a low- coefficient of electrical resistivity.
frequency limit to the correct rms response of a ther¬ /3=the temperature coefficient of thermal
mal converter, for reasons analogous to those causing conductivity of the conductor,
a similar low-frequency limit in all other types of N=p£K/ak.
transfer instruments. Below this limit the response p and a=the perimeter and area of the cross
characteristic E=/(I) is not independent of frequency section of the conductor, respec¬
(except for symmetrical square waves), and may be tively.
expected to be in error by an amount that is inversely f=the emissivity.
proportional to the square of the frequency. As A = tlie Stefan-Boltzmann constant in
table 4, b, shows, such an error was observed for the watts-cm_2-(deg AT)-4
1- and 2-amp thermal converters. k = the thermal conductivity in watts-
These thermal converters were constructed by E. S. Williams. cm_1-deg_1
326/130
To=the ambient temperature in degrees It is evident that the most significant factor in
Kelvin the error term of eq 4 is the length of the heater
q=d/ul2 and is (approximately) the ratio as the error is dependent upon the fourth power
of the crest value of the cyclic of this quantity.17 It is of interest to note that
fluctuation of temperature at the the lengths, 21, of the heater of the 1- and 2-amp
midpoint to the average temperature thermal converters are only 0.28 and 0.19 cm,
rise at the midpoint respectively, and that the length of each of the
u=2irf, where j is the frequency of the higher-range thermal converters is about 0.5 cm.
applied current in cycles per second. The value of H can be determined from the depar¬
(f=the thermal diffusivity of the heater ture from “square law” of the characteristic equa¬
material in cm2-sec_1. tion £’=/(/) of a thermal converter. For these
1=one-half the length of the conductor in thermal converters it was considerably less than
centimeters. the factor given for platinum in table 7. The
H=h-\-B/2A, where A and B are the con¬ estimated frequencies for a 0.1-percent error, on
stants in the emf-temperature char¬ the assumption that the diffusivity of the plat¬
acteristic for the thermocouple it¬ inum-iridium alloy used in these thermal converters
self, that is, in the expression, is the same as that of platinum, were about one-half
E=Ad+(B/2)62 the observed values.
These equations indicate the factors upon which
the low-frequency error depends. They show that
7. High-Frequency Response
the temperature rise can correctly be used as a
measure of the rms alternating current, at fre¬
quencies high enough so that the second tertn in As these thermal converters are inherently suit¬
the bracket of eq 3 is negligible for the accuracy able at frequencies up to 100 Me or more, the upper-
desired; in other words, at frequencies for which frequency limit is set by the circuits in which they
the heater integrates reasonably well. This is true are used. The precautions indicated in section 4
even if the thermal and electrical conductivities are sufficient to control capacitance and leakage
of the heater are dependent upon temperature. currents in tests of ammeters at frequencies con¬
As a corollary, these equations show that a thermal siderably higher than the 20,000 c/s established
converter correctly indicates rms response, except as the present upper limit by the demand for this
at low frequencies, even though its response char¬ testing service. The actual upper-frequency limit
acteristic E=f(I) is not quadratic. has not been established.
The above equations are valid only at frequen¬ The magnitude of the effective impedance sets the
cies for which the second term in the brackets of upper frequency limit for the model A and the model
eq 3 is small compared to 1. B voltmeters. For an accuracy of 0.01 percent at
Equation 4 may be used to calculate approxi¬ 20,000 c/s, a time constant of less than 1X10-7 and a
mately the frequencies for which ac-dc differences skin effect less than 0.01 percent are necessary for
may be expected to cause an error of an arbitrarily the impedance of each step of the model A and each
chosen value. This has been done for some typical range of the model B instruments. The time con¬
materials for an error of 0.1 percent, a temperature stant of each decade of the model A voltmeter was
rise of 200° C, and a typical heater length of 0.4 cm. computed from the inductance and direct-capaci¬
The results are shown in table 7.16 tance values furnished by the manufacturer of the
resistance box used. This was less than 5X10-8 for
each decade. The skin effect was also negligible for
Table 7. Frequencies for which the computed ac-dc differ¬
ence is 0.1 percent each decade. The resistance box is connected as a
three-terminal resistor, with the shield connected to
[TC-A signifies a thermocouple of Chrome] P and Alumel, and TC-B a thermo¬ the grounded side of the line. For this connection
couple of copper and constantan, for which B/2A = — O.IXHH and + 0.9XHH,
respectively] the errors caused by the capacitance currents to the
shield are much less than those that would occur if
Approximate temperature Frequency for the shield were connected to a resistor terminal.
coefficient 6=0.1*
Thermal They are significant only for the two 1,000-ohms-
Heater material
diffusivity per-step decades. The effects of the capacitance of
Electrical Thermal
TC-A TC-B
resistivity conductivity the resistance cards and switch studs of these decades
to the shield were computed from the manufac¬
%I°C %I°C cm2-sec-1 c/s c/s turer’s capacitance figures of about 10 p.p.1 per decade
Copper- +3.9X10"3 -0. 2X10-3 1.1 00 TO
Platinum_ . +3.5 +.5 .25 j2 14 by considering the capacitance as distributed along a
Manganin_ < -1 -f-2. 7 .075 3 2
Constantan... . < -1 4-2.3 .004 3 2 uniform transmission line, composed of resistance
80 Ni-20 Cr_ .1 b+3.5 .038 2 2
Carbon- -. 5 .2 5 I
elements and short-circuited at the end. From the
+ .4
usual transmission line formulas, the magnitude of
» For a heater 0.4 cm long with a temperature rise of 200° C.
b Estimated from Lorenz’ Law.
17 For a given heater alloy, the low-frequency range can be extended by using
longer heaters, with proportionately greater cross-sectional area to give the same
>6 The effect of radiation losses has been neglected in preparing this table. temperature rise for the same current. If solid heaters are used, the improved
In general, this effect is significant only in the very low and very high range low-frequency performance will be accompanied by increased error due to skin
thermal converters. effect at high frequencies.
327/131
the current at the short-circuited end was computed and between the high-side binding post and the top
as approximately shield. This shield is brought as close as 2.5 cm to
the row of high-side posts.
|/!=^[1 -0.0056 {uCR)2], (5)
8. Other Effects
where
V=the applied voltage The intercomparisons of insulated-heater thermal
I — the current at the short-circuited end converters disclosed an unsuspected ac-dc difference
i?=the resistance in the circuit that was dependent upon the voltage between the
(7= the total capacitance to shield of the heater and ground and upon the heater current of
windings in use each of these elements. This was traced to a very
C0 = 27r/. small leakage current flowing through the ceramic
bead between the heater and thermocouple and then
At 20,000 c/s for a capacitance of 30 ju/zf this is
through the stray capacitance of the Lindeck poten¬
changed by less than 0.005 percent at 20,000 ohms,
tiometer to ground. This current caused additional
the maximum setting. The effect of the capacitance
heating of the bead and thus an error as on direct
(about 10 mA) of each rather large switch blade to
current the insulation resistance of the Lindeck was
the shield was separately computed, and found to be
much higher than the capacitive reactance, even at
less than 0.005 percent at 20,000 c/s. Thus it was
20 c/s. In addition, the ceramic bead of each of
concluded that the errors of this voltmeter should
these thermal converters exhibited marked dielec¬
be a maximmn of 0.01 percent at the highest settings
tric absorption, so that its a-c resistance was much
and frequencies used, and less at lower settings and
less than its d-c resistance. For a typical thermal
frequencies. The performance at the top settings
converter at rated current a leakage current of 0.7 ^a
was verified by the tests outlined in section 4.
resulted from a difference of 50 v between the heater
The impedance of the model B voltmeter was simi¬
and ground, resulting in a power dissipation of 35 /xw
larly computed, and only the capacitance to the shield
in the bead. This was sufficient to cause a 0.2-per-
was found to. be significant. As a rough but ade¬
cent error, which was proportional to the square of
quate approximation this may be considered the
the voltage difference and was very markedly de¬
same for each step, which leads to the approxima¬
pendent upon the heater current.19 The reactance
tion of considering the capacitance as distributed
of a capacitance as small as 100 /x^f is less than the
along a nonuniform transmission line composed of
computed bead resistance, even at a frequency as
resistance elements and short-circuited at the end.
low as 50 c/s, so that in almost any application this
The effective impedance was computed by a pertur¬
source of error makes it necessar}^ to maintain the
bation method,18 on the rough assumption that the
heater near ground potential.
resistance per unit length was proportional to the
square of the electrical distance from the output end.
The magnitude of the current at the short-circuited 9. Other Applications
end was found to be approximately
This equipment was designed for the testing of
other instruments. Portability and the ease of mak¬
|/|=^[1 —0.0022 (wCR)2]. (6) ing other a-c measurements were not considered to
be important. However, the thermal converters and
voltmeter elements are readily portable and have
From the measured capacitance of 50 mil and the
been used, with a small portable millivolt potenti¬
total resistance of 100,000 ohms for the 750-v range,
ometer as the indicator, for measurements at other
the change in current was computed as —0.09 per¬
locations. For measurements of current or voltage
cent at 20,000 c/s, which is in very good agreement
to 0.01 percent, it is necessary to use a transfer
with the observed results of table 5. Moderately
method, calibrating the thermal converter or volt¬
good agreement for the other ranges was also ob¬
meter on direct current at the time of use. For a
tained, but the change in current was not observed
more moderate accuracy of about 0.1 percent, it
to increase in proportion to the square of the fre¬
should be feasible to use a low-range thermal con¬
quency as indicated by eq 6. However, the agree¬
verter, shunting it for current measurements and
ment is good enough to show that the ac-dc difference
adding a series resistor for voltage measurements.
of this instrument is due to capacitance currents to
Only the thermal converter itself need be transferred
the shield. These currents cannot easily be reduced in
to direct current if the resistances of the shunts and
a shielded multirange instrument of reasonable size.
the series resistor are properly adjusted. Thus d-c
(A shield is desirable to control the currents.) They
sources of extended range, volt boxes, etc., should be
could be compensated by connecting appropriate
unnecessary. The whole apparatus could be built
capacitors across sections of the resistors, but this
in a convenient case, containing a small potenti¬
seems less desirable than the procedure of applying
ometer for the measurements of the current through,
small corrections and has not been done in the model
or the voltage across the thermal converter heater.
B voltmeter. Much of the capacitance is that
between the binding posts of each range (see fig. 2) 19 This was due to the marked decrease in the resistance of the ceramic bead
with an increase in temperature. With a fixed voltage across the head, the error
18 See appendix^. was in some cases proportional to the tenth power of the heater current.
328/132
A suitable low-range thermal converter should also accuracy not unreasonable. The cost of the required
make feasible an a-c potentiometer with an inherent number of standards and the equipment is not pro¬
accuracy of 0.01 percent. The basic circuit for a hibitive. Thus electro thermic instruments may be
polar form of such a potentiometer (to measure the ranked on a par with electrodynamic and electro¬
magnitude but not the components or phase angle of static instruments for a-c measurements of the high¬
an alternating voltage) is shown in figure 6. With est accuracy.
the switch thrown to direct-current the potenti¬ As a result of this work, thermal converters of
ometer is standardized in the usual manner, and the negligible or small known transfer error are now
deflection of the indicator of the thermal converter is available as transfer standards with which the trans¬
observed. The switch is then thrown to alternating fer performance of other thermal converters may be
current, which is adjusted for the same deflection evaluated at the NBS'to 0.01 percent at audio
of this indicator. The circuit is surely not new, but frequencies.
a proved thermal converter makes possible a con¬
venient transfer with accuracies comparable to those The author acknowledges the help of Murray Blitz,
of a d-c potentiometer. The other basic limitation who made most of the intercomparisons described in
remains that any a-c potentiometer measures only section 4 of this report and checked all the tables and
the fundamental component of the unknown voltage. equations, and of Earl Williams, who constructed all
However for an alternating quantity having a har¬ the equipment and assisted with the measurements.
monic content as large as 2 percent, the rms value
differs from the fundamental by only 0.02 percent.
Audio-frequency oscillators and amplifiers with am¬ 11. References
plitude distortion much less than this are now avail¬
[1] F. K. Harris, A suppressed-zero electrodynamic volt¬
able. Such an oscillator could be used to supply the
meter, BS J. Research 3, 445 (1929) RP105. -
potentiometer directly and to supply the test circuit [2] J. H. Park and A. B. Lewis, Standard electrodynamic
through an amplifier preceded by a suitable phase- wattmeter and ac-dc transfer instrument, J. Research
shift network. NBS 25, 545 (1940) RP1344.
[3] F. B. Silsbee, Composite-coil electrodynamic instruments,
BS J. Research 8, 217 (1932) RP411.
10. Summary [4] C. C. Patterson, E. H. Rayner, and A. Kinnes, The use
of the electrostatic method for the measurement of
power, J. Inst. Elec. Engrs. (London) 51, 294 (1913).
The electro thermic transfer standards used at the
[5] R. S. J. Spilsbury and A. Felton, The electrostatic volt¬
Bureau for the precise measurement of voltage and meter as a dc-ac transfer instrument, J. Inst. Elec.
current at audio frequencies, and designed primarily Engrs. (London) 89-11, 129 (1942).
for the standardization of electric instruments, have [6] W. N. Goodwin, Jr., The compensated thermocouple
ammeter, Trans. Am. Inst. Elec. Engrs. 55, 23-33
been described. The extensive tests to establish the
(1936).
transfer performance of the standards have been out¬ [7] F. L. Hermach, A precision electrothermic voltmeter for
lined, and the results have been given. The causes measurements between 20 and 20,000 cycles, Trans.
of the observed ac-dc differences have been outlined Am. Inst. Elec. Engrs. 67, 1224 (1948).
[8] H. B. Brooks and A. W. Spinks, A multi-range potentiom¬
and the errors evaluated. The equations governing
eter and its application to the measurement of small
the temperature rise of a conductor heated by an temperature differences, BS J. Research 9, 781 (1932)
electric current and cooled by conduction to relatively RP506.
massive terminals have been solved to establish the [9] G. Borelius, Bestimmung des Thomsoneffektes aus dem
thermischen Gleichgewicht in einem stromdurchflos-
steady-state midpoint temperature rise on direct
senen Drahte, Ann. Physik 63, 845 (1920).
current with Peltier and Thomson heating, as well as [10] M. G. Lloyd, Elec. Rev. & Western Elec. 58, 65 (Jan. 14,
the ordinary resistance heating, in order to obtain 1911).
the d-c error of a thermal converter. The equations
have been solved to establish the average midpoint
temperature rise on alternating current, in order to
Appendix 1. Development of Working
obtain the low-frequency error of a thermal converter Equations
in which the electrical and thermal conductivities of
By definition, the percentage ac-dc difference of an instru¬
the heater are temperature dependent, in which some
ment is the percentage difference in the quantity required
loss of heat occurs by radiation, and in which the to give the same response on alternating and direct current.
characteristic of the thermocouple is nonlinear. It is
Possible applications and modifications of the equip¬ g = 100 (7)
ment for measurements under less stringent con¬ V dc
572772 0-61-22
329/133
where Ode is the quantity measured with tjfie potentiometer, Therefore, for a transfer test of an ammeter with a thermal
8S is the ac-dc difference of the transfer standard, and Qac converter that has no transfer error, the required ac-dc
is the a-e value required for the observed deflection of the difference of the test instrument is by eq 7 and 10
test instrument.
AC
where the alternating and direct currents are adjusted for the
same response of the test instrument, and the resulting
galvanometer deflections are Dac and Ddc, respectively.
If the transfer thermal converter has a percentage ac-dc
difference, <5S, at the current and frequency at which the test is
made, then the ac-dc difference of the instrument under test
is, to the desired order of approximation,
Figure 6. Elementary form of Polar a-c potentiometer. in which 8S is the percentage ac-dc difference of the range (or
setting) of the standard voltmeter used.
In this deflection method of using the Lindeck potentiom¬
When a thermal converter is used for transfer tests the eter for rapid transfer tests, the resistance of the circuit seen
small percentage difference, 8t, between the values required by the galvanometer must be constant to better than 1 per¬
for the same indication of the test instrument on alternating cent. The resistance of the thermocouple of each thermal
and direct current is determined from the resulting deflection converter has been adjusted to the same value so that the
of the Lindeck galvanometer. Therefore, for each transfer simplified formulas may always be used. The temperature
thermal converter, the small percentage difference in output coefficient of the resistance of the Lindeck circuit has been
emf resulting from a small percentage difference in heater computed as less than 0.1 percent per deg C, and computations
current must be evaluated. This relationship may be show that the resistance seen by the galvanometer is changed
expressed as by less than 0.2 percent by the adjustable battery circuit, for
AE AI ail ranges of the potentiometer except the topmost.
(9)
E 7’
330/134
one terminal by a small amount A9 and to lower the other by that of eq 18. The method is powerful, however, and was
an equal amount. The differential equation then has as its used later in evaluating other perturbing effects, so that it
boundary conditions, 9=90 + A9 at x=+l and d—90—A9 at will be outlined here. The Thomson heating, which is much
x=—l. The solution is ' smaller than the Joule heating, can be considered as perturbing
the temperature distribution of the conductor. Its effect can
be approximated by solving the differential equation without
e-e0=^(l*-x*) + AelJC’ (16) the perturbing term, then substituting the result, and its
derivatives as necessary, in the perturbing term only, and
solving the resulting equation.
and is therefore unchanged at the midpoint.22 Of course,
complete thermal symmetry is never attained, so that Peltier Without the Thomson heat term, eq 19 becomes
heats may cause a change in the midpoint temperature.
However, for massive terminals the temperature changes, T"= — b
A6, are very small, so that Peltier heating need not be con¬
sidered significant.
‘Thomson heating occurs in homogeneous conductors having with the boundary condition given, the solution of which is
a temperature gradient in the direction of the current flow.
For the one-dimensional case here considered, the Thomson
heat in an element of length dx having a temperature differ¬ Tr-T0=| (l2-x2),
ence dy is <rIdd watts where a is the Thomson voltage coefficient.
Experiment and theory indicate that a is proportional to the
where the subscript 1 indicates the first approximation.
absolute temperature, but, as the Thomson heating is small
Placing this value of T\ and T\ in the perturbing term of 19,
compared to the joule heating, we might, as a first approxi¬
reduces this to a linear equation,
mation, consider <x constant. Equation 14 then becomes
By substituting the values for b and c and noting that Appendix 3. Low-Frequency Error of a
2Ipl/a=V, the voitage drop across the conductor, eq 1 and 2
of the text can be obtained. They show that Thomson heat¬ Thermoelement
ing causes a small (second order of smallness) decrease in the
temperature rise at the midpoint of the conductor, and can The problem is to find the steady-state average (over an
thus lead to a transfer error in a thermal converter. However, integral number of cycles) temperature rise at the center of a
the approximation of considering a a constant is a drastic one, uniform conductor of length, 21, and cross-sectional area, a,
and a more accurate approach will be to consider a = BT, carrying a sinusoidal current and cooled chiefly by conduction
where B is a constant,23 and T is the absolute temperature. to massive terminals that are at fixed and equal temperatures.
Equation 17 then becomes Assume that the current remains sinusoidal and that only
Joule heating of the conductor occurs. If all the cooling is by
,d2T . 72p . T dT conduction to the terminals, the differential equation govern¬
akd7>+a+B,Tlh=0’ ing the temperature of the conductor is, under these assump¬
tions,
which is of the form
, d29 , Bp d9
ak -=ams (21)
T" + wTT' + b=0, (19) ox2 a ot
with the boundary conditions T=T0 at x= ±1, where with the boundary condition 9( + l.,t) =0, where
w= Bl/ak 9 = temperature in deg C
This is a nonlinear equation and very difficult to solve. f? = the square of the instantaneous value of the current=
Chester Snow showed that it could be reduced to a Ricatti and /2( 1 — cos 2cot)
then to a Bessel equation, but suggested an approximation or /=rms value of the current
perturbation method to give the desired results more simply w=27r/, and / is the frequency of the current in cycles
and directly. The answer is not significantly different from per second
Z = time in seconds
22 Since Peltier heating depends upon the junction temperature, a second- m—density of the conductor in grams-cm-3
order change in midpoint temperature may occur.
23 B=d2EldT2, where E is the emf of a thermocouple composed of the metal in s=thermal capacity of the conductor in joules-grams-1
question and a metal such as lead, which has very small Thomson heating. -deg C-1.
331/135
The procedure will be to consider p and k constant and to Thus the average value over an integral number of cycles of
solve for the periodic steady-state temperature rise, then to the temperature rise is the same as that found in appendix 2
approximate p and k as p = po (1 + «0) and k=ko (1 + /30), to in¬ for the same numerical value of direct current, so that under
troduce radiation loss, and to solve the resulting equation by these conditions the temperature rise of the conductor can
the perturbation method just outlined. For the cases con¬ be used as a measure of the current without ac-dc error.
sidered ad and (39 are considerably less than 1, and radiation We shall now consider that p=p0(l + ad) and fc=fco(l + /30)
loss is much less than the Joule heating of the conductor. and that some small fraction of the total heat is lost by
radiation. If p is the perimeter of the heater in centimeters,
With a and /3=0, eq 21 is the form then the heat radiated per second from the surface of a
differential length, dx, of the heater is
d29 d9
dtf~gdt = -i(1~cos 2a)0 dHr=pt;K(Ti— Tftdx,
where g=ms/k0. and b = Pp0!a2k0. where f is the emissivity of the heater material, a numerical
Let 0=u(x,t) -\-\j/{x). Then constant, and K is the Stefan-Boltzmann constant in watts-
cm~2(deg A)-4. T is the absolute temperature of the ele¬
dn-d_d2u ay ment; 2’o is the absolute temperature of the surroundings.
dx2 dx2 dx2 Introducing these terms in eq 22 gives, as /30<1,
Let dy/dx3 = — b. Then <p= — (5x4/2+ CiX + c2. g-0(l-/^=-&( 1-cos 2ut)(l+vd)+'^(T*-Tl)
Applying boundary conditions gives
(28)
d(-\-l,t) 4-</'(/) =0
where n=a — (3. The perturbing terms are
d(-l,t) = u(-l,t)+*(-l)= 0.
dd
(1) dt
We may take ip(l) =\p( — l)=0.
Then \p = b/2(l2—x2) and eq 22 becomes
(2) 5( 1 —cos 2ut)t\9
d2u du
Q dt
= b cos 2cd, P?K (T4-T4)
dx2 (3)
ak
with u(±l,t) =0.
From complex number theory if u is harmonic in time, it As these perturbing terms are relatively small, the solution
can'be represented by the real part of Uei2u,t, where U is a will differ by only a small amount from that given by eq 27.
complex function of the real variable x, and j=V—1. Simi¬ Hence a sufficient expression for these small terms will be
larly, b cos 2cd can be represented by the real part of be’2"1, given by substituting the known values of the dependent
where b is a real number. Substituitng these quantities and variables y from eq 27, and its derivatives as necessary,
the appropriate derivatives in eq 23 and dividing by e’2o“ in these perturbing terms. When this is done, it is found
results in that the first of these terms reduces to functions of x, each
d2U multiplied by sin md or cos nut, where n=2, 4. The second
a-^-j2ugU=b, (24) reduces to similar harmonic terms plus the terms — b[t)P—
(vUa/2)], where P={l2—x2)b/2. The third term becomes
with U=0 when x= ± l. The solution of this is unnecessarily complex. It can be simplified by making the
further approximation that for this term
cosh pxl
(25) T— To=9r=P(l —g sin 2ut),
cosh filj
and
where p = ^j2ug. At the midpoint T*-Tt=(Ta+dry-Ti~49rTl+6d?T20> 9r<T0
b
U= — - — 1-—■
cosh plj J 2<x>g'
and also that k=k0. With these approximations, this term
[‘ reduces to harmonic terms plus the terms
0]=-^- (l2— x2) -pUa cos 2ut — Ub sin 2ut. (27) ^ — gvr = Z(Pn cos nut + Gn sin nut) (30)
332/136
For each frequency oBthe harmonic driving terms in the If the hot junction of the thermocouple is fastened to the
right-hand side of this equation, the solution of the equation midpoint of the heater, the emf of the couple may be repre¬
that vanishes at the end points will be of the form Ua cos sented by
nut — Ub sin nut, where Ua and Ub are functions of x. The
average value, over an integral number of cycles of the E=
lowest frequency, will therefore vanish. Since this is a
linear equation, the solution is the sum of such solutions where 0m is the temperature of the hot junction in deg C, and
and will therefore also vanish. Thus the average value A and B are constants to be evaluated at the temperature of
of 0, which we are interested in, is equal to (x). the cold junction. On direct current 0m = 0dc and
Evaluation \p (x) in the same manner as before gives for
the value of \j/ at the midpoint where x=0,
E = A0dc 1 + 2 ,j
and on the assumptions that Ua2{ — 1) = Uai(l) and f7O2(0) = 0. The average value of emf over an integral number of cycles of
alternating current is then (to the desired accuracy, and if
The required expression for Ua can be evaluated from eq 25 (B/2A) Ode is small compared to unity)
and 27 by noting that
where y = -\[ug, so that cosh ,ux=cosh yx cos yx-\-j sinh yx sin From these expressions
yx, and cosh ftl= cosh yl cosyZ+j sinh yl sin yl. At the fre¬
quencies we are interested in /d)>5, so that Ea-Ed H
9dcq 2
Ed 2
where H=h-\-B/2A.
cosh jut~-7^(coS yl-\-j sin yl). For table 7, it is desired to compute the frequency at which
the ac-dc difference of a thermal converter due to this effect
Since w = real part of is equal to 0.1 percent. From eq 7 and 9 and for n = 2 the
difference in current for the same emf is one-half the difference
in emf for the same current, and is opposite in sign. There¬
Ue^f=re [1- e>'2“ j = Ua cos 2o>t-Ub sin 2cot, fore the percentage ac-dc difference, 5, is approximately
-25 H0dcd2
it will be found that 5= — 25H0dcq2
(2trf)H* ’
Ua =—i [sinh yx sin yx cos yl — cosh yx cos yx sin yd. where in addition to the symbols already defined, d=l/g=
uge^1 k0/ms and is the thermal diffusivity in appropriate units.
Performing the desired integration results in
Appendix 4. Effect of Shield Capacitance
Uo2— — ,y 'gyOeyi (cos yl cosh yx cos yx + sin yl sinh yx sin yx). on the Model B Voltmeter
From which We may approximate the effects’ of capacitance to the
shield of this voltmeter by considering the voltmeter as
Ua2(l) = Ua2(-l)~-j4land.
dco Q Ua2(0)«Ua2(l). consisting of a transmission line of rx=rox2 ohms per unit
length, having a constant capacitance Co farads per unit
Therefore, eq 31 becomes length to the shield, witha sinusoidal voltage of rms value V
volts applied to the high-resistance end. The other end may
be considered short-circuited, since the thermal converter,
^(0)=eoc=^[i + ¥+(’-hivr^M3] = which is connected between the low-resistance end and the
ground post, is of low resistance. The shield is also connected
to this ground post. We wish to find the magnitude of
0d j^l + M-f 2^i22J’ (32)
current, io, at the short-circuited end. For an element of
length dx at a distance x from this end, we have
where M is equal to the remaining terms in eq 31 and h~t\ —
2NTIV. dV T dl.
By the same procedure, eq 28 can be solved for the d-c case
^ = Irx = Irtf’and—^roF, (35)
by considering dd/dt=0 and i—I, a constant of the same
numerical value as the rms value of the alternating current. where dV is the rms voltage drop across the element, and
The temperature rise at the midpoint will be found to be dl is the rms current from the element to the shield. Differ¬
entiating the first of these and substituting the values of I
0ac=0d[\ + M]. (33) and dl/dx in the resulting expression gives
Thus the temperature rise on direct current is the same as on d2V 2 dV
alternating current when the frequency is high enough so that \2x2V = 0, (36)
dx'2 x dx
the third term in the brackets of eq 32 is sufficiently small
compared to unity. Therefore the temperature rise at the subject to the boundary conditions "F = 0 at x=0 and V = Va
midpoint on alternating current is at x=l and where \2=juc0r0.
When A=0 this becomes, with the same boundary
0ac=edc [ 1 +2-^^)] **e4c [1 +\odccf^ (34) conditions,
drV_2 dV=
if M is small compared to unity. dx 2 x dx ’
333/137
the solution of which b From eq 35
Vax3 J_ dV
TV (38)
rox2 dx
\\ e now put this value of T’ in the perturbing term containing Performing the indicated differentiation of eq 37 and sub¬
A2 in eq 36, to get stituting in 38, and noting that the total resistance is
R = r0l3/3 leads to
d2V_ 2 dV \2Vax5
dx2 x dx l3 V
[l_g + 8,X10-.X,, + V(,-f)+^]
R
The solution of this, with the same boundary conditions, is
Solving for T0 at .r=0, and noting that the total capacitance
Vax3,
Y, AV\ , A'-dV' is C=cJ gives
V
l3 V 28/ 28P '
/o — 1—jO. 107a)C R — 7.9 X 10"3(o)C/? )2^j-
As \2=jcocuro, we will need to repeat this process to secure
terms in X4, for we are interested in the magnitudes of 1 ’ and I.
Putting the above value of F2 in the last term of eq 36 and Since the terms in the bracket are small compared with unity
solving as before gives
|/0| = ^1-0.0022 (t,-Ctf)2J-
334/138
THERMAL VOLTAGE CONVERTERS FOR ACCURATE VOLTAGE
MEASUREMENTS TO 30 MEGACYCLES PER SECOND
F. L. Hermach E. S. Williams
This is a publication of the American Institute of Electrical Engineers
SYNOPSIS
INTRODUCTION
The r-f thermal voltage converters make use of the transfer principle,-*- in which a
d-c voltage is substituted for the a-c voltage to be measured. The d-c voltage is adjusted
to give the same output emf of the thermoelement that was obtained with the a-c voltage
applied, and is measured with a potentiometer or other suitable means. Thus only good
short-time stability, high precision of reading and small known frequency influence are
required of the transfer standard.
The dimensions of the converters and additional data are given in the caption of
Fig. 1. Each converter is assembled by first soldering the resistor, R, to the thermo¬
element, TE, and to the coaxial input connector, A, which is mounted in disc, D^. This
disc is then fastened to the cylinder, F, and the two thermocouple leads are soldered
to the 2-pin output connector, B. Disc,D£, which has a small hole with a cross wire
at G, is then fastened to F and the free end of the heater of the thermoelement is
335/1
soldered to the cross wire. For clarity both thermocouple leads are shown in the figure.
However the plane of the thermocouple leads and the pins of the output connector is
actually at right angles to the axis of the cylinder so that minimum emf is induced in
the output circuit. In use F must be grounded through the input connector.
The 100- and 200-volt converters contain two resistors and two cylindrical inner
shields, Si and S2, as shown in Fig. 2, to reduce the effect of the distributed capaci¬
tance from the resistors to the outer cylinder. This considerably extends their fre¬
quency range. Shield, S£, is centered in F by a narrow ring, and is fastened to F by
3 equally spaced screws. Slots in F make axial adjustment of S2 possible. The end-
plate of S-l is soldered to the input lead and to the resistor in assembly and this
shield is fastened to by small insulators. The assembly is otherwise similar to
that described for the other converters. The 50-volt converter has a single resistor
and'a fixed inner shield like Si in a cylinder of the dimensions given in the caption
of Fig. 1.
Additional information on the voltage converters is given in the first three columns
of Table 1. The commercially available deposited carbon resistors are 1/8 inch in dia¬
meter and 2 inches long, and are vacuum sealed within a 3/16 inch diameter glass cylinder
(Some resistors of a different construction showed unaccountable ac-dc errors above 10 me
when used in earlier converters.) The 5- and 10-ma UHF thermoelements, each with coli-
near supports and heater wire, are also commercially available, with an output voltage
of about 7 mv at rated current. Each is in an acorn-shaped evacuated glass bulb, and
has a small bead between the heater and hot junction of the thermocouple to electrically
but not thermally insulate them.
THEORY
Because of the geometry, the resistor, R, in the cylinder of each low-voltage con¬
verter (without inner shields) may be represented as a transmission line of length, £,
with a uniformly distributed series impedance of z ohms per unit length and a uniformly
distributed shunt admittance of y mhos per unit length. The line is terminated by the
heater of the thermoelement. We are interested in the magnitude of the trans-impedance
Zc = Vi/IQ where is the input voltage being measured and IQ is the current through
the heater. If the heater resistance, R^, is much less than that of the resistor, R,
we may to a first approximation consider this a short circuited line. (It will be
apparent later that this is not a drastic approximation at low frequencies since we are
concerned only with the effect of the line reactance on Zc.) Then by ordinary steady-
state transmission-line theory,
336/2
We may define the parameters a = wCR and b = wL/R. If each of these is less than
unity the hyperbolic function may be approximated by the first 3 terms of its series
expansion, so that
z c = za ++ ^
6
+ izir
120
+ ...)
Then if all terms in anbm for which m+n>3 are discarded we find that
e2
Zc « R(l+jb)(l - ^ +
6 120 6
rac - Vdc
S = (3)
Vdc
where Vac and Vdc are the rms alternating and direct voltages required to obtain the same
response (output emf) of the converter. If the thermoelement has no ac-dc difference and
if we define k = b/a, this becomes
So (4)
where SQ signifies that other possible causes of ac-dc differences are neglected.
The relation makes it possible to estimate SQ (to this approximation) very simply.
For the cylindrical construction of figure 1 with a cylindrical unspiraled resistor,
L = 0.012iM microhenries and C = 0.61i/M picofarads, where M = log^Q(g/h) and g and h
are the diameters of the cylinder and resistor, respectively, and i is in inches. Thus
numerically, k = 1.9x10^(M/R)^ and a = 3.8xlO~^OR/M, where f is the frequency.
For example if R = 2000 ohms, g and h are 2 in. and 1/8 in., and i = 2.0 in.,
SQ = +0.001 (+0.1%) at 40 me.
For low voltage converters the effect of the heater resistance R^, cannot be
neglected. To the next approximation the heater can be considered as a lumped resis¬
tance termination for the transmission line. Then from steady-state transmission-line
formulas,
337/3
zc = cosh \Tzy + -j== sinh sfzY (5)
When k<0.038, distributed capacitance from the resistor to the cylinder predominates.
At a given frequency and geometry, the resultant error is proportional to the square of
the resistance. It can be minimized by the construction shown in Figure 2 in which two
inner cylindrical shields are used, with one connected to the input lead. Thus there are
now two transmission lines in series. If R^«R, application of transmission line formulas,
with careful attention to signs, gives for the desired trans-impedance,
where the subscripts 1 and 2 refer to the left and right hand lines (resistors and shields)
of figure 2, and the resistance of the thermoelement is neglected.
If Z\ = Z2 = Z and = Y2 = Y we have
zc = sinh
Thus the trans-impedance is twice that of either line, and the ac-dc difference, SQ, will
.be that of half the total line. The parameter, a, of eq. 4 will then be l/4th that for
a line of the same dimensions without inner shields. Thus at a given frequency for which
a<l the use of the double shield rather than a single cylinder of the same diameter should
reduce the frequency error (for resistors of the same value and dimensions) by a factor
of 16.
However more detailed analysis indicates that even further improvement may be possible.
For these converters k«0.038, so that the series inductance of each line may be neglected.
By expanding the hyperbolic functions of eq 7 and discarding higher order terms it can be
shown by lengthy and rather tedious algebraic manipulation that SQ ~ 0 when 1-032,
where and are the lengths of the left and right hand lines of fig. 2, and a<l.
For the convenient dimensions shown in figure 1, the parameter, a, of each low-
voltage converter now being used (up through 20 volts) is low enough that it has not been
necessary to proportion the converters for minimum error in accordance with eq 6. The
higher voltage elements are constructed with dual shields, with the grounded shield
axially adjustable. This shield is set to give minimum frequency error as explained in
the next section.
338/4
A useful theoretical study of coaxial lines, terminated by resistors, was published
by Crosby and Pennypacker.1^ They show conditions for minimizing the input reactance.
TESTS
The shield of the Lindeck potentiometer provides reasonable immunity from induced
fields. In some tests at 10 me a current of 2 ma from the shield to ground through the
lead from the potentiometer to the converter caused no significant error.
For transfer measurements the most important requirement is that the ac-dc difference
of each of these converters be known to.the full accuracy desired. The general principles
on which such determinations are based have been given.^ In the frequency range studied,
the major error of each of these converters is caused by the reactance. For a given
converter this error should be independent of voltage level. Thus it was feasible to
evaluate the relative errors by intercomparing converters of adjacent voltage ranges to
determine their differences in frequency response. A complete series of such comparisons
was made at 2 voltage levels for each pair of converters at frequencies up to 40 me. In
each comparison the two converters were connected in parallel to a coaxial lead through
a tee fitting (GR 874), and a shielded potentiometer was connected to each output. Each
potentiometer was adjusted for zero deflection at the test voltage. The converters were
then supplied in succession with alternating, direct, reversed-direct, and alternating
voltage. Each voltage was adjusted to produce the same emf of the higher rangd> converter
and the deflection of the galvanometer connected to the other converter was observed.
From the differences in emf (directly determined from these differences in deflection)
the difference in the frequency response of the two converters was determined.1 In these
tests the movable inner shields of the 100- and 200-volt converters were adjusted for
best performance over the desired frequency range, by comparison with the 20- and 50-volt
converters. The errors of these adjustable converters were found to be complicated func¬
tions of the frequency and shield position.
339/5
The results of these intercomparisons from 0.1 to 40 me are given in Table 2. They
show that for each pair of converters the relative ac-dc differences were independent of
the applied voltage to 0.02%, or less, and were less than 0.4% up to 30 me. For the low-
voltage converters they were less than 0.057, to 20 me, without exception and without
evidence of systematic errors.
The relative ac-dc differences of most of the thermoelements used in the voltage
converters were also determined (before the thermoelements were installed) by making
similar ac-dc intercomparisons (as current-measuring elements) at 40 me. For these
tests the two thermoelements were connected in series along the axis of a brass cylinder
having the same dimensions as in fig. 1 but with a coaxial input connector at each end
and with two 2-pin output connectors. (A split cylinder with one part readily removable
facilitated changing thermoelements.) The input was applied to one connector with the
other shorted. Rough calculations indicated that at 40 me the current should change by
about 0.5% along the transmission line formed by the heaters of the thermoelements in
this cylinder, but that the distribution should not be greatly dependent on the resistance
of the heaters (less than 100 ohms each). Therefore two determinations of relative ac-dc
differences were made with each pair of thermoelements, with first one end then the
other of the cylinder shorted. For each pair the two determinations differed by about
0.47o at 40 me, but in each case their algebraic average was less than 0.027,.
These calculations and tests also provide assurance that even at 40 me the current
along the heater of a single thermoelement terminating a voltage converter is well within
0.1 percent of the value at the mid-point of the heater.
The 1-volt converter was compared at rated voltage with a wire-wound thermal voltage
converter of known ac-dc difference at 3 cps and 20.kc, with observed differences of
+0.027, and 0.00% respectively.
The response of the 1-volt converter, with a shielded d-c millivoltmeter connected
to its output, was determined to better than 1% at frequencies from 1 to 400 me by the
bolometer bridge of Selby & Behrent5} with the results shown in Table 3. It is evident
that the frequency influence is very small to 100 me. It increases rapidly at higher
frequencies, becoming -5.27, at 300 me. The calculated ac-dc difference at 300 me by
equation 6 is only -0.9% (this converter has a 1/4 in. diameter unsealed resistor).
Some of the discrepancy may be accounted for by the voltage rise in the connector (a
type N-UG58/U input connector was substituted for the type 874 for these tests).
The intercomparisons of Table 2 show that the ac-dc differences of the five low-
voltage converters without inner shields (1- to 20-volt ranges) are all equal to better
than 0.057, to more than 20 me. At 40 me they agree to 0.1% with the values calculated by
eq 6. (Values cannot readily be computed for the higher voltage elements with the inner
shields.) This unanimity between converters having such a wide range of resistors and
different thermoelements gives good assurance against unknown sources of error. It is quite
unlikely (but not impossible) that each converter would have the same ac-dc error. However
the test of the 1-volt converter with the bolometer bridge provides most valuable additional
340/6
assurance. The large errors of this converter above 100 me decrease rapidly with decreas¬
ing frequency, well within the stated accuracy of the bridge measurements. For conver¬
ters without inner shields almost all known causes of such errors (such as the effect of
reactance, eq 6, skin effects in the resistor or thermoelement, etc.) should cause ac-dc
differences approximately proportional to the square of the frequency, over the range for
which the errors are small. Thus the authors believe that the large measured errors
above 100 me can be extrapolated downward to lower frequencies to indicate with consider¬
able confidence that the ac-dc difference of this converter is less than 0.27o at 40 me
and less than 0.1%, below 30 me.
Based on the above consideration the authors assigned a value of zero to the ac-dc
difference of the 1 volt converter to 40 me, and then determined the ac-dc differences
of all the other converters from the intercomparison data in Table 1. The results,
rounded to the nearest 0.05% to 10 me and 0.1% to 40 me, are given as observed values
in Table 4.
USE
These converters are most conveniently used to measure the ac-dc difference or
frequency influence of other rms instruments such as thermocouple voltmeters (which are
now available with accuracies of 1/27, to 10 me). A typical set-up for this is shown in
Fig. 5. The scale calibration of the voltmeter can then easily be checked on reversed
direct current. The ac-dc difference tests are similar to the intercomparisons already
described, and d-c calibration of the converter is not necessary once the scale factor
of the Lindeck potentiometer is determined for each converter. This is the percent
change in input voltage per centimeter change of galvanometer deflection.1 Either direct
or low-frequency alternating current may be used as the reference frequency and the test
may be made rapidly and accurately. The results are only slightly affected by drifts in
either instrument. At frequencies above about 20 me a small lead correction may be
necessary if the connectors between the junction plane and the two instruments are not
electrically equal, but this is readily determined to the required accuracy.
An r-f generator of at least 5 watts output and good voltage stability is required
but only moderate frequency accuracy and stability are needed. It is very difficult to
construct broad-band generators of good wave form at this power level. However, a simple
adjustable LC tuned circuit, with the instruments connected across the capacitor, can be
used to simultaneously improve the wave form, match the impedance of the instruments to
that of usual 50-ohm source, eliminate the capacitance loading of the instruments on the
source, and provide the higher voltages often required. For these combined purposes the
values of inductance and capacitance should be chosen so that u)L = 1/uC , where
and R£ are respectively the resistances of the source (including the inductor) and the
load (instruments). For low voltages a 50-ohm resistor can be connected in series with
the LC circuit, and the instruments and attenuators as needed can be connected in parallel
with this resistor for improved wave form.
341/7
CONCLUSIONS
ACKNOWLEDGEMENT S
It is a pleasure to acknowledge the help of R. Richardson and J. Scofield who checked
the equations and carried out the computations which made Fig. 4 and the last column of
Table 3 possible. J. E. Hill assisted with some of the measurements. Sincere thanks are
also extended to M. C. Selby and L. Behrent for their careful calibrations of the 1-volt
converter.
REFERENCES
1. Thermal converters as AC-DC Transfer Standards for Current and Voltage Measurements
at Audio Frequencies, F. L. Hermach, J. Research NBS, vol. 48, pp. 121-138, 1952.
2. Wide Range Volt-Ampere Converters for Current & Voltage Measurements, F. L. Hermach
and E. S. Williams, Communications & Electronics (AIEE) 1959 (paper 59-161).
3. Tables and Chart Atlas of Complex Hyperbolic & Circular Functions, A. E. Kennelly,
Harvard University Press, 1921.
1 5 90 0.2
3 5 90 0.5
5 5 90 1
10 5 90 2
20 10 25 2
50 10 25 5
100 10 25 10<2>
200 10 25 20<2)
(2) Two equal resistors in configurat ion of figi
342/8
TABLE 2. COMPARATIVE AC-DC DIFFERENCES OF VOLTAGE CONVERTERS
Voltage Range
Converter Converter Applied Comparative AC-DC Difference, D, (Percent)
A B Volts 0.1 me 1 me 5 me 10 me 20 me 30 me 40 me
1 0.997
10 0.998
30 0.998
100 0.999
200 0, 982
300 0.945
400 0.883
343/9
F
F
344/10
Fig. 4. Computed AC-DC Difference as a
function of the parameters a and k.
and audio frequencies depend, at the present time, upon these same laws of the interchange of energy. They are:
ac-dc transfer instruments which have nearly equal ac 1) electrodynamic instruments, which depend upon the
and dc response. Such instruments may be calibrated force between current-carrying conductors; 2) electro¬
on direct current and then used on alternating current, static instruments, which depend upon the force be¬
or alternatively, may be used directly to make ac-dc tween charged conductors; and 3) electrothermic instru¬
difference measurements, as in the certification of other ments, which depend upon some effect produced by the
heating of a current-carrying conductor. The first and
instruments.
In addition to the instantaneous values as deter¬ third respond essentially to current and the second to
mined by an oscilloscope or point-by-point methods, voltage, but series or shunt resistors make all three types
there are at least seven other quantities that may be suitable for both measurements, while other circuit ar¬
determined for a periodic ac wave in which all values rangements make power measurements almost equally
separated by the periodic time, T, are equal. These are: feasible.
dM~I
U6f = B\I2 where B\ =-
dd J e-$f
5) and 6) the positive and negative crest or peak values; The differential equation governing the angular deflec¬
and 7) the crest-to-crest or peak-to-peak value. tion is1
Instruments for measuring each of these values of an
ac voltage or current wave are commercially available P6" + Ad' + Ud = Bi2
(with widely varying accuracies). Ordinarily, however, where P and A are the inertial and damping constants,
only one of these values is really required in a measure¬ respectively. If the mechanical inertia is sufficiently
ment. Unfortunately, it is difficult to deduce accurately great that with alternating current applied the periodic
one value from the measurement of another, except for fluctuations of 6 are negligible compared with the aver¬
a few waveforms (which, of course, must be known). In age value, termwise integration over an integral number
most cases the rms value is the one really desired, since of cycles gives
the rate of transformation of electrical to other forms of
* Manuscript received by the PGI, August 14, 1958. 1 F. K. Harris, “Electrical Measurements,” John Wiley and Sons,
f National Bureau of Standards, Washington, D. C. Inc., New York, N. Y., pp. 47 and 413; 1952.
346/235
IRE TRANSACTIONS ON INSTRUMENTATION December
347/236
1958 Hermach: AC-DC Transfer Instruments for Current and Voltage Measurements
most national standardizing laboratories. In general, sonably rugged and portable pointer-and-scale instru¬
there are three rather distinct steps in a complete de¬ ments of the 0.1 per cent accuracy class, based on
termination. These are: 1) choice of two different types Weston’s designs, have been available for a number of
of instruments that are inherently suitable as ac-dc decades. For increased resolution and accuracy, recently
transfer instruments, and a theoretical study of all developed commercial instruments, following Silsbee’s
known effects that can cause ac-dc differences in each composite-coil design,6 now use a taut-suspension sys¬
type; 2) construction of instruments of each type in tem with the moving coil of a second instrument rigidly
such a way that each of these effects by computations fastened to it to provide an electromagnetic restoring
and tests can be evaluated over the required1 ranges; torque.7-9 This torque can be precisely determined by
and 3) comparison of the actual transfer performance measuring the current through the second instrument
of the two instruments to guard against unknown with a resistor and dc potentiometer.
sources of error. The tests of step 2) will often take three As the torque equations indicate, an electrodynamic
forms. These are: a) tests of components of the instru¬ instrument responds to the square of the current, the
ment for the effects of known factors (such as magnetic scale being marked to indicate current directly. As an
susceptibility, for example; b) intercomparisons of unshunted ammeter (limited to about 0.1 ampere be¬
instruments of the same type but of different ranges, in cause of the springs or ligaments which must carry the
which some of the effects (such as residual reactance) current to the moving coil), it will be useful as a transfer
may be expected to be very different; and c) repeated instrument over the range of frequencies for which the
intercomparisons, in which each known source of error average torque, T =/(/), is the same as with direct cur¬
in one instrument is accentuated by definite amounts rent. The chief factors which affect this in an actual
(such as by the deliberate introduction of known react¬ instrument are: 1) capacitances between turns and be¬
ance). Normally the comparisons of step 3) can be made tween coils, which alter the currents; 2) eddy currents
over only part of the useful ranges of frequency, voltage, in. neighboring metals, which alter the field that links
etc., and the performance must also be judged by the re¬ with the moving coil; and 3) electrostatic torque, caused
sults of steps 1) and 2). Normally also one of the two in¬ by differences in potential between the fixed and mov¬
struments will have better-known characteristics, ing coils. In general, these set an upper limit of about 1
wider or more suitable ranges, or better-behaved ac-dc kc for such instruments. The frequency ranges of am¬
differences than the other, and thereafter is used as the meters, which use shunts across the moving coils, and
principal transfer instrument for that laboratory, but voltmeters, which use series resistors to limit the cur¬
both are important. rent, are more sharply limited by the inductances of the
The accuracy of ac measurements is often limited by coils. The manufacturer must insure that the ratio of the
the stability of the source rather than the instrument. impedances of the two parts ol the divided circuit of the
For the calibration of instruments with a transfer instru¬ ammeter is equal to the ratio of resistances, to the full
ment, stabilized sources free from short-time fluctua¬ accuracy of the measurement over the frequency range
tions are required, but long-time stability and good load of interest. Similarly he must insure that the magnitude
regulation are of less importance.4’5 Repeated sets of of the impedance of the voltmeter (defined as the ratio
readings make it possible to determine the ac-dc differ¬ of the applied voltage to the current through the coils)
ence to a precision considerably better than the fluctua¬ is equal to its dc resistance. By connecting resistors in
tions in the source, and the procedure described tends series with the coils of the ammeter and capacitors
to eliminate drifts in both the test and standard instru¬ across portions of the voltmeter, the upper frequency
ments. If both instruments inherently respond to the limit for 0.1 per cent error in commercial instruments
same values (rms for example), low-order harmonics up can be raised from 100 cps to about 1000 cps, and for
to several per cent ordinarily cause the test to be in error 0.25 per cent error to about 2500 cps.10
by much less than 0.1 per cent. Because the mutual inductance between the coils
changes with scale position (the operating torque de¬
Electrodynamic Instruments
pends upon this) complete compensation cannot be at¬
The electromagnetic force between two conductors tained at all scale positions. Mutual inductance intro-
with reasonable geometry and currents is so small that
6 F. B. Silsbee, “Composite coil electrodynamic instruments,”
to obtain satisfactory torque the effect is magnified by J. Res. NBS, vol. 8, pp. 217-264; 1932.
coiling the conductors. The torque-weight ratios thus 7 G. F. Shotterand H. D. Hawkes, “A precision ac/dc comparator
for power and voltage measurements,” Proc. IEE, vol. 93, pp. 314—
obtained are still considerably less than those enjoyed 324; 1946.
by dc instruments with permanent magnets, but rea- 8 J. Sorge, “A new precision instrument for ac power measure¬
ments,” VDE Fachberichtc, vol. 17, pp. 27-30; 1953.
9 R. F. Estoppey, “Inductronic Electrodynamometer,” Confer¬
4 A. H. M. Arnold, “Alternating-current instrument testing ence on Electronic Standards and Measurements, paper 35; August,
equipment,” Proc. FEE, vol. 101, pp. 121-133; 1954. 1958.
5 F. L. Hermach, “Power supplies for 60-cycle tests of instru¬ 10 J. H. Miller, “Frequency compensation of ac instruments,”
ments and meters,” Proc. ISA, vol. 11, paper 56-21-3; 1956. Trans. AIEE, vol. 70, pp. 217-221; 1951.
348/237
IRE TRANSACTIONS ON INSTRUMENTATION December
duces other errors as well, so that the electrodynamic
ac-dc transfer instruments designed and constructed at
the U.S. National Bureau of Standards11 and the Na¬
tional Physical Laboratory of South Africa12 are nor¬
mally operated only over a narrow range of deflections
about the position of zero mutual inductance, by 'the
use of a continuously adjustable external multiplier for
the first instrument and a torsion-head for the second.
In the NBS instrument (see Fig. 1), two sets of coils are
astatically arranged to eliminate the effect of uniform
external fields, and a strip suspension and light-beam
pointer with a scale 2 meters from the instrument insure
definite readings and permit a scale factor at 0.01 per
cent/mm at currents from 0.1 to 20 amperes. The instru¬
ment is normally used as a wattmeter or ammeter, and
a similar but older companion instrument as a voltmeter
(10 to 600 volts).13 The NPLSA instrument is quite sim¬
ilar, but may be connected as a voltmeter as well (0.05
to 5 amperes and 25 to 500 volts). The ac-dc difference
of each of these instruments is believed to be known to
better than 0.01 per cent at power frequencies and to
perhaps 0.1 per cent up to 3 kc (above 50 volts), verified
by careful study and by comparison with electrothermic
and electrostatic instruments.
one-to-one basis, so that torque-weight ratio remains charges on insulation in the electric field, which can
low. Thus electrostatic instruments, as distinct from modify the field with direct voltage applied, and 3) im¬
perfectly conducting films on the plates which can cause
electrodynamic instruments where the fixed coil may be
the field strength between the plates to differ on direct
designed to have more ampere-turns than the moving
coil, seem rather delicate except at voltages above about and alternating voltage.14 None of these limit the fre¬
1 kv, and are rarely available commercially for precision quency range as sharply as does the inductance of
measurements. However, taut-suspension, light-beam- electrodynamic instruments, and electrostatic voltme¬
pointer constructions, using improved optical systems ters may be accurate to perhaps 1 me. However, when
an electrostatic instrument is used as an ammeter the
and modern materials, have begun to appear commer¬
impedance of the 4-terminal resistor must be equal to
cially in Europe and may presage a renaissance of this
type of instrument. its dc resistance. (II a transformer is used, it is no longer
an ac-dc transfer instrument.) This is difficult to achieve
Electrostatic instruments inherently respond to volt¬
age but may be adapted for current measurements by above about 20 kc.
Electrostatic voltmeters with ranges from 50 to 160
measuring the voltage across a resistor carrying the
volts have been the principal ac instruments for over
current. However, the voltage must be rather high (50
40 years at the National Physical Laboratory in Eng¬
to 100 volts) for adequate accuracy and the shunt must
land14 and somewhat modified instruments of the same
then dissipate considerable power at high currents, or
basic design are similarly used at the National Stand¬
an instrument transformer must be used (either a cur¬
ards Laboratory in Australia. The ac-dc differences of
rent transformer with a resistor in its secondary, or a
the NPLE instruments are known to 0.01 per cent at
voltage transformer between a resistor and the instru¬
ment). power frequencies and to better than 0.05 per cent up to
100 kc (verified by calculations of the known sources of
11 J. H. Park and A. B. Lewis, “Standard electrodynamic watt¬
error and by comparison with electrodynamic and elec¬
meter and ac-dc transfer instrument,” /. Res. NBS, vol. 25, pp. 545- trothermic instruments and with other electrostatic
579; 1940.
12 J. W. Whittaker, “A precision electrodynamometer standard and
ac/dc transfer instrument,” Proc. IEE, vol. 101, pp. 11-20; 1954. 14 R. S. J. Spilsbury and A. Felton, “The electrostatic voltmeter
13 F. K. Harris, “A suppressed zero electrodynamic voltmeter,” as a dc/ac transfer instrument,” J. IEE, vol. 89, pp. 129-137;
J. Res. NBS, vol. 3, pp. 445-457; 1929. 1942.
349/238
1958 Hermach: AC-DC Transfer Instruments for Current and Voltage Measurements
instruments having noble-metal conductors). Special 10 me.16 The calculated ac-dc differences of the convert¬
transformers and resistors have been constructed and ers and of the voltmeter elements were verified by inter¬
tested to permit the measurement of current and of comparisons of different ranges and by comparisons
voltage below 20 volts, and resistance voltage dividers (over limited ranges) with electrodynamic and electro¬
have been evaluated for voltage measurements to 1000 static instruments. A Lindeck deflection potentiometer
volts (at audio frequencies).4 is used at NBS with these converters and voltmeter ele¬
ments to provide high resolution (0.005 per cent/mm at
Electrothermic Instruments rated output emf) and rapid reading.
Easily measurable effects are produced by the heating Similarly it has been shown that a properly selected
of short straight metallic conductors or very small beads and mounted pair of indirectly heated thermistors may
of semiconductors, having very low electrical time- be used in a bridge as an ac-dc transfer instrument for
constants. These effects can thus be independent of fre¬ an audio-frequency potentiometer to better than 0.02
quency up to 100 me or more. In addition, the effects per cent at frequencies from 0.2 to 20,000 cps.17 Ther¬
can be measured electrically, rather than mechanically, mistors are also used for ac measurements at the Physi-
with high precision and freedom from mechanical kalisch-Technische Bundesanstalt in Germany.18 Their
resonances. These are the chief advantages of electro- rather long thermal time constants (about 10 seconds)
thermic instruments. provide good low-frequency response but make highly
Because their output is electrical, and because of their stable sources necessary.
low reactance, thermocouple instruments and bolometer The small ac-dc differences of thermal converters and
bridges are the only two forms now generally used. In thermistor bridges at audio frequencies are caused al¬
commercial thermocouple instruments the temperature most solely by: 1) thermoelectric effects in the heater,
rise of a conductor (heater) is measured with a thermo¬ which affect the temperature rise on direct current, but
couple and millivoltmeter. The combination of the because of thermal inertia do not affect it on alternating
heater and thermocouple is called a thermal converter. current; 2) integration errors at low frequencies, which
Many commercial bolometer bridges now contain bead are inversely proportional to the square of the fre¬
thermistors (temperature-sensitive resistors) and are quency; and 3) small reactance effects in the voltmeter
unbalanced by the temperature rise caused by the heat¬ elements, such as capacitance currents between parts of
ing of the applied current. The unbalance voltage is the multiplier, and from the multiplier to its surround¬
measured or balance is restored by additional currents ing shield. Fortunately, it is possible to obtain thermal
of a different frequency, which are measured. Thermo¬ converters with heaters of manganin or other alloys of
couple instruments of the \ to 2 per cent class have long low thermoelectric effects and of sufficient length to
been commercially available for current and voltage make the ac-dc differences 0.01 per cent or less from 20
measurements, while most commercial bolometers are cps to 200 kc.
used for power measurements (absorbing all of the A single 10-ma thermal converter is used with mul¬
measured power). In either case undesired thermal ef¬ tirange shunts and series resistors for ac measurements
fects, such as high ambient temperature coefficients, in the U.S.S.R., both in a central standardizing lab¬
accelerated aging of the heater at the elevated tempera¬ oratory and in convenient consoles at branch labora¬
ture, and transfer of heat to other parts of the measur¬ tories.19 Series and shunt resistors are also used in the
ing element have limited the sustained accuracy obtain¬ laboratories of the Germany Authority for Weights and
able. However, by using such instruments solely as ac-dc Measures, but with two converters in a differential
transfer instruments, calibrating them before and after circuit.20 As a further modification of the transfer princi¬
each reading, these effects can largely be eliminated. ple, the heater of the thermal converter alone (without
Studies at the National Bureau of Standards16 have the shunt or series resistors) can be switched to direct
shown that properly designed thermal converters may current for the dc calibration.21 Only a small dc source is
be used as ac-dc transfer instruments at currents from 1
ma to 50 amperes with an accuracy approaching 0.01
16 F. L. Hermach, “Electrothermic instruments for the measure¬
per cent at power and audio frequencies and very prob¬ ment of alternating current and voltage,” Proc. NPL Symposium on
ably up to 200 kc. Their low reactance also makes possi¬ Precision Electrical Measurements, HMS Stationary Office, London,
Eng., paper No. 15; 1955.
ble transfer voltmeters from 0.2 to 600 volts (at 133 17 F. C. Widdis, “The indirectly heated thermistor as a precise
ohms per volt) with wire-wound series resistors that ac-dc transfer device,” Proc. IEE, vol. 103, pt. B, pp. 693-703; 1956.
18 H. J. Schrader, “A-C potentiometer measurements with in¬
have ac-dc differences less than 0.03 per cent at audio directly heated NTC resistors,” Electrotech. Z., vol. 73-A, pp. 547-
frequencies. Voltmeter elements with film resistors in 549; 1952.
19 K. P. Shirokov, “An installation for calibrating ammeters and
coaxial lines terminated by thermal converters have voltmeters at high frequencies,” Proc. Russ. Inst. Meteorology, vol.
been constructed and intercompared to 0.05 per cent to 24, p. 24, 1954.
20 W. Rump, “On the exact absolute measurement of ac voltages
and a potentiometer for testing ac instruments,” Eleklrotechnik, vol.
5, pp. 64-67; 1951.
iof. L. Hermach, “Thermal converters as ac-dc transfer stand¬ 21 F. L. Hermach and E. S. Williams, “Multirange audio-fre¬
ards for current and voltage measurements at audio frequencies,” quency thermocouple instruments of high accuracy,” J. Res. NBS,
J. Res. NBS, vol. 48, pp. 121-138; February, 1952. vol. 52, pp. 227-234; 1954.
350/239
IRE TRANSACTIONS ON INSTRUMENTATION December
Other Instruments
351/240
circuit, the current through the heater is
A Wide-Range Volt-Ampere Converter h=IdRm'/{Rn + Rm') = VpRm'/RP(R» +
Rm'). If the a-e-d-c difference of the
converter is negligible, /<,=//,, and if
for Current and Voltage Measurements Rm'=Rm, Va=VvRn/Rv. The a-c volt¬
age to be measured is thus the product
of the measured potentiometer voltage
F. L. HERMACH E. S. WILLIAMS and a ratio of resistances, and is inde¬
MEMBER AIEE NONMEMBER AIEE
pendent of both the emf-versus-current
characteristic and the heater resistance of
T echnological developments
have brought increasing needs for
measurements to 0.02% over the same
frequency range.
the converter.
From Fig. 1(B), with the converter
connected to Rs, the current through the
convenient, portable instruments for
heater is Ih =IaRs/(Rh-{-Rs)- With the
measuring current and voltage with high Method
converter in the d-c circuit, Vp — Id(Rh-{-
accuracy over a wide range of frequencies.
Rs'). If Ih—h and RS'=RS, and if the
A “universal” volt-ampere (VA) con¬ For a-c measurements, a single thermal
a-c impedance of the heater is equal to its
verter has been developed for making such converter (often called a thermoelement
d-c resistance, Ia — Vp/Rs, again desirably
measurements to 0.05% with alternating or thermocouple) serves as an a-c-d-c
independent of the emf-versus-current
current at frequencies from 5 cps (cycles transfer standard. Its heater is con¬
characteristic and the resistance of the
per second) to 50 kc and with direct nected in series with an a-c resistor for
converter.
current. The new model is based voltage measurements and in parallel
on an earlier, much more limited proto¬ with an a-c shunt for current measure¬
type converter,1 and, like it, is used with ments, as shown in Fig. 1. The electro¬
a null-type d-c potentiometer. With this motive force (emf) of the converter is
single, self-contained, portable, 44-range manually balanced by the voltage from
instrument, the 1.5-volt range of a suitable a built-in d-c source, with a galvanom¬
potentiometer can be used to make a-c eter as the indicator (not shown in the
(rms) and d-c measurements from 7.5 figure). The heater of the converter is
milliamperes (ma) to 20 amperes and from then switched to a second d-c source,
0.5 volt to 600 volts. With a more sensi¬ which is adjusted to rebalance the
tive external galvanometer, the instru¬ galvanometer. The voltage across a
ment can also be used as an a-c-d-c portion of this circuit is then measured
transfer standard for a-c-d-c difference with the potentiometer and multiplied
by a simple number to obtain the alter¬
nating voltage or current. For d-c
Paper 59-161, recommended by the AIEE Indicat¬ measurements, the same multiplier is
ing and Integrating Instruments Committee and
approved by the AIEE Technical Operations connected as a volt box and the same
Department for presentation at the AIEE Winter shunts are used, without the thermal con¬
General Meeting, New York, N. Y,, February 1—6,
1959. Manuscript submitted November 1, 1957; verter.
made available for printing December 1, 1958.
From Fig. 1(A), with the thermal con¬
F. L. Hermach and E. S. Williams are with the verter connected to Rm, Va=Ia(Rm-\-RK).
National Bureau of Standards, U. S. Department
of Commerce, Washington, D. C. With the converter connected to the d-c
352/1
A-l
Fig. 2. Wiring diagram of VA converter. Switch-position markings from top to bottom on diagram: switch C, AV, AC, DV, DC;
switch D, AV No. 1, AV No. 2, AC No. 1, AC No. 2. Dotted lines signify potential leads except at A-2 and B-3 where they
indicate intermediate switch positions are not shown
For direct voltage measurements, the so that hc = Vp/Rs, and again the nominal
heater of the thermal converter is dis¬ multiplying factor, Ft = l/Rs, is the same a= ^L l + Rs/Rk J
connected, the multiplier, Rm, is connected as that for the corresponding a-c range. = VpF{(l+w) (2)
directly to the voltage to be measured, If the resistors Rp, Rs, Rm, Rm', and R/
and a 1.5-volt section of it is connected have small proportional corrections of where the capital letters now signify the
to the potentiometer. If this section is ap, a,, a^, dm', and a/ respectively, more nominal values. For the useful ranges,
designated as Ra, Vdc = VpRm/R^, and if detailed analysis shows that, to a sufficient Rn>Rh> Rs, so that the accuracy re¬
RC = RP the nominal multiplying factor, approximation, quired of the compensating resistors Rm'
Fv=Rm/Ra, for the volt box thus formed and R3' is roughly inversely proportional
VpRm I” _ am + dm'Rh/Rm'1
is the same as that for the corresponding Fa = to the current or voltage range, while
Rp L ap 1 + Rk/Rm J essentially full accuracy is required of
a-c voltage range. For d-c measurements
= VpFt( 1 + B) (1) Rm, Rs, and Rp. In practice the over-all
the potential terminals of the shunt Rs are
connected directly to the potentiometer, and correction factors, v and w, are directly
353/2
measured and switch D is the a-c-d-c
transfer switch for the thermal converter.
The switch markings are identified in the
note on Fig. 2. For a d-c measurement,
the position of switch D is immaterial,
and it is only necessary to set the range
and selector switches and connect the
potentiometer. For an a-c measurement,
after these are set switch D is rotated to
AV No. 1 or AC No. 1, connecting the
thermal converter to the desired a-c
source, and the Ei circuit is adjusted with
controls marked “No. 1” (coarse and fine)
for zero deflection of the built-in galva¬
nometer (or any other convenient deflec¬
tion). Switch D is then rotated to AV
No. 2 or AC No. 2 connecting the thermal
converter to the E2 circuit which is ad¬
justed with controls marked “No. 2” to
rebalance the galvanometer to the same
deflection. The external potentiometer
is then read in the usual way. In switch
positions marked “Rect. Inst.,” (rectifier
evaluated by d-c tests of each range of the ance upon temperature and therefore cur¬ instrument) two additional poles of
instruments. rent.) switch D (not shown in Fig. 2) connect a
panel-type rectifier instrument to the
An instrument which responds to both
Description a-c input in place of the thermal converter.
direct and alternating current is often
best tested by evaluating its scale correc¬ This is useful for preliminary settings and
A wiring diagram of the VA converter for checking to make sure that the thermal
tions on direct current and then deter¬
is shown in Fig. 2 and a photograph of the converter will not be overloaded in the
mining its a-c-d-c difference at the de¬
instrument in Fig. 3. The ranges and No. 1 position.
sired frequencies2 (defined as the difference
between the direct and alternating cur¬ the nominal shunt and series resistances A 7.5-ma thermal converter having a
rent or voltage required to produce the are listed in the first two columns of heater resistance of 662/3 ohms is used in
same deflection). This difference is Tables I and II, respectively. There are the instrument. Because of the com¬
usually small and relatively permanent. separate input binding posts for the cur¬ pensating resistors, the actual heater
Such a-c-d-c difference tests can readily be rent and voltage circuits, with separate resistance need not be closely adjusted.
made with the VA converter by connecting range switches, A and B in Fig. 2. The High-quality a-c resistance cards are
the two instruments together and to a instrument can thus be used to measure, used for the series and shunt resistors of 1
source of the desired voltage or current, in succession, the voltage and current in a ohm or more. Shunts of lower resistance,
making a preliminary setting of the test d-c'or a-c circuit without reconnections. for the higher current ranges, were spe¬
instrument, and then balancing the emf It may also be useful for wattmeter cially made in our laboratories of bifilar
of the thermal converter, E, to give zero testing on direct current or at unity power strip, ranging from 2 to 20 mils in thick¬
deflection of the galvanometer. The factor on alternating current (more ness, of nickel-chromium-aluminum-iron
instruments are then successively switched strictly at the power factor of maximum alloy, with insulation of 1-mil polyester
to a-c, d-c, reversed d-c, and a-c sources, deflection of a wattmeter) by the familiar film. The dimensions were chosen as a
which are adjusted to give the same deflec¬ “phantom loading method.”3 compromise between the requirements of
tion of the test instrument, and the re¬ Switch C selects the quantity to be large area for cooling and of thin strip for
sulting galvanometer deflections are ob¬
served, without rebalancing. The per¬ Table I. Results of D-C Tests of Current Ranges
centage a-c-d-c difference of the test in¬
strument, T, is computed as T = AD/KE Per-Cent Correction
where AD is the difference between the
D-C Ranges, at 9 Months* A-C Ranges, at Rated I
mean of the two a-c and the two d-c Current
galvanometer deflections, and A is a sensi¬ Range, Rs, 20% At 6 At 18
Amperes Ohms Rated I Rated I Months* Months *
tivity factor. This factor is readily
determined by observing the changes in
20. . 0.025. .+0 05. .-0 07. .+0.03. .+0 05
galvanometer deflection resulting from 10. . 0 05. .-0 01. .-0.09. .+0.01. .+0,02
5 . 0.1 . .-0.02. .-0 01. .+0 04. .4-0.03
small measured differences in applied .+0,02.
2.5. . 0.2 . .+0.03. .+0 03. .+0 05
direct current or voltage at several differ¬ 1 .05 . . 0 00. .+0 03. .+0 03 .... .4-0.05
. 1 . .-0 09. .-0.08. .-0 07. .— 0.021
ent levels of heater current, and plotting . 2 . .-0.04. .-0.03. .-0.02. . 0 00
K versus E. Tests show that only two ... 5 . .-0.01. .-0.01. .-0 01. . 0 00
.10 . .— 0 04. .-0.04. .-0.04. .—0.03
such curves are necessary, one of which .20 . - 0 03. .-0.03. .-0.02. .-0.02
applies to all current ranges and the other 7.5 ma. . None . .+0.01
to all voltage ranges. (They differ be¬ * Number of months after completion of the instrument,
cause of the dependence of heater resist¬ t Readjusted before this test; adjustment affects lower-current ranges.
354/3
Table II. Results of D-C Tests of Voltag e Ranges not exceed 0.04% per minute for any 7.5-
ma converter tested to date, so that, since
Per-Cent Correction
1 /2 minute is ordinarily sufficient time for
D-C Ranges, at 9 Months* A-C Ranges, at Rated V these balances, the warm-up drift error
Voltage
Range, Rm» 20% At 6 At 18 should not exceed 0.02% even for the
Volts Kilohms Rated V Rated V Months* Months* largest change in heater current. How¬
ever, for best results, a warm-up period
600. . 80. .+0.01_ .+0.01. .+0.02
300. . 40. .+0.00_ . 0 00. .+0.02. .+0.01 of at least 2 minutes should be allowed.
150. . 20. .+0 00. . . . . 0.00. . :_+0.02. .+0.01 After the initial change, the steady-state
75. . 10. .-0.01_ .-0.01. .+0 01. . 0 00
60. . 8. .-0 01_ .-0 01. . 0.00. . 0.00 drift of the galvanometer in either the
30. . 4. . 0 00_ . 0 00. .+0.01. .+0.02
15. . 2. . 0 00_ .4-0 02... . .4-0.02
No. 1 or No. 2 positions of switch D
7.5. . 1. .-0.02_ .-0 02. .-0.02 is less than 0.003% per minute, with No.
6. .0.8. .-0.01.... .-0.02. . 0 00
3. ...0 4. . 0 00_ .-0.03 . . — 0 03 6 dry cells as the internal d-c sources.
15... .0.2. . 0.00_ .-0.01. . -0 03 The a-c resistors permit some response
0 5t. .None. .+0.02
of the thermal converter at frequencies
* Number of months after completion of the instruments,
f Alternating-current range only.
up to 100 megacycles (me). In areas of
high television signal strengths, errors
minimum reactance, using inductance are peculiar to the use of the thermal con¬ can result, since at these frequencies lab¬
and skin-effect formulas developed by verter in this instrument and deserve oratory leads of ordinary lengths can
Silsbee.4 The compensating resistors, special mention. approximate quarter-wave antennas.
which carry only small direct current, were Thermal converters have been studied These difficulties were avoided in this
chosen to meet the accuracy require¬ and used as highly accurate a-c-d-c instrument by a low-pass L-C filter with a
ments dictated by equations 1 and 2. transfer standards.5 Normally the aver¬ cutoff frequency of 10 me per second and a
A built-in taut-suspension light-beam age for the two directions of current serves characteristic impedance of 67 ohms, con¬
galvanometer with 50-millimeter (mm) as the d-c reference in transfer measure¬ nected to the heater terminals of the
scale provides a deflection of 2 mm for ments. In this instrument, however, only thermal converter. The filter changed
0.1% change in input current or voltage one direction is used for convenience and the a-c-d-c difference of the converter by
at rated heater current. This is adequate rapidity of measurement. A converter less than 0.04% [0.003 db (decibels)] at
for the zero settings required in a-c tests, of small d-c reversal difference is thus 100 kc.
but for a-c-d-c transfer tests to 0.02%, desirable. Fortunately, converters of the
in which differences in deflection must be bead type, with an insulating bead be¬ Tests
read, a more sensitive external box-type tween the heater and thermocouple, can
galvanometer with a longer scale is con¬ be selected with reversal differences of The results of d-c tests of the d-c ranges
nected in place of the internal galvanom¬ 0.02% or less. They are then completely (shunts and voltage divider), and of re¬
eter. For these tests an external 1.5-ma interchangeable without requiring recali¬ versed d-c tests of the a-c ranges of the
d-c milliammeter is also connected at a bration of the instrument. completed instrument are given in Tables
pair of binding posts marked M in Fig. 2, Most low-range thermal converters I and II. The results of a-c-d-c differ¬
to measure the battery current through the have ambient temperature coefficients of ence tests from 20 cps to 50 kc are given in
10-ohm resistor, e, and therefore, at bal¬ 0.1 to 0.2% per degree centigrade (C), and Table III. The d-c shunts were tested
ance, the emf, E, of the thermocouple. should be thermally protected for precise with an accurate Kelvin double bridge,
The VA converter has over-all dimen¬ transfer measurements. A cotton-filled and the volt box tested in comparison
sions of 17x14x9 inches and weighs 30 Bakelite cylinder around the converter with the National Bureau of Standards
pounds. and the wooden case of the instrument
provide a thermal time constant of 40
Table III. Results of A-C-D-C Difference
Precautions minutes, which is ample for normal labora¬
Tests at Rated Current and Voltage
tory use of this instrument. A wooden
A number of factors must be evaluated divider separates the shunts and multiplier Per Cent A-C--D-C Difference
in the design and construction of an' from the converter and the d-c circuits
Range 20 Cps 20 Kc 30 Kc 50 Kc
instrument such as this. Some are com¬ to minimize self-heating changes.
monly encountered in the design of any After a sudden change in heater current,
600 volts. . . . ..0.00.. . 0.00. . 0 00. . . +0 03*
equipment of high accuracy such as a many bead-type thermal converters ex¬ 300 volts. . . . . -0.01. . . -0.01. . . -0.02
d-c potentiometer, but are accentuated hibit an exponential “warm-up” drift 150 volts. . . . ..0 00.. . . -0 01. . . -0 02
75 volts.. . . .. 0.00. . -0.02
by the high voltage used in some of the having a time constant of 1 to 2 minutes. 60 volts. . . . .. 0.00. . . -0.01
30 volts. . . 0 00. . 0 00
circuits. These are: 1. insulation, con¬ This drift, an increase of current with time . . +0.01.
15 volts. . . . 0 00
tact, and lead resistance of the wiring, for the same emf, can amount to almost 7.5 volts. . . . . .+0.01. . 0 00
6 volts. . . . . . +0.01. . . +0.01
switches, and other components; 2. self¬ 0.1% for some converters following a 3 volts.. . . . +0.01. . . +0.01
heating and ambient temperature effects change from zero to rated current. The 1.5 volts. . . . .0.00.. .. 0.00. +0.01
0.5 volt. ..0 00.. . . +0,02. . . +0.03
in the shunt and series resistors; and 3. percentage error caused by it is 20 amperes . +0.01. . . -0.01 . . -0.02
the stability of these resistors. In addi¬ 10 amperes. . -0.01. . . -0.01. . . -0.04
d = /;i'2 Sdt 5 amperes. ..0.00.. . -0.02. . . - 0 03. . . -0.04
tion, because of the unusual frequency 2.5 amperes. .. 000. . . +0.01
1 ampere. . 0 00. . . +0 04
range in the instrument, the effects of where 5 is the percentage rate of change 500 ma. .,0.00.. . . -0 01 . . -0 03
residual reactances, of the a-c resistors, of heater current at constant emf and 250 ma. . . -0 02. . . -0 01
100 ma. . 0 00 . . -0 01
associated circuits, and of the frequency and ti are the elapsed times at which the 50 ma. .. 0.00. . . +0 01
influence of the thermal converter must be No. 2 controls and the potentiometer are 25 ma. .0 00.. ..+0.01. . . +0 01
7.5 ma. . . -0.01. . .+0.01
carefully considered. Still other factors balanced, respectively. The drift rate did
355/4
(NBS) standard volt box. The reversed 0.5-volt and 7.5-ma ranges showed that urements of very high accuracy over
d-c tests were made with a potentiometer the a-c-d-c difference of the converter truly wide ranges of voltage, current,
and auxiliary shunts and volt boxes. The itself is not more than 0.01% at fre¬ and frequency. Its accuracy depends
a-c-d-c difference tests were made with quencies down to 5 cps. Both theory and primarily upon wire-wound resistors and
the NBS standard thermal converters,5 experiment indicate that the difference strip shunts of excellent stability and low-
which have recently been evaluated down should also be negligible on all other temperature coefficient. After initial a-c-
to 2 cps, and with a newly constructed ranges of the instrument. d-c difference tests of the instrument,
transfer voltmeter element, which has The small discrepancies in Tables I and routine periodic tests need ordinarily be
been evaluated to 100 kc. II between the corrections for the corre¬ made only with direct current, thus
The results of the d-c tests show gen¬ sponding d-c and a-c ranges are caused simplifying the standardization of the
erally satisfactory performance, adjust¬ by the d-c reversal difference of the ther¬ instrument.
ment, and stability of the instrument, ex¬ mal converter, the difference between Ra The VA converter should be partic¬
cept for the 10- and 20-ampere ranges in and Rp, and the influence of the com¬ ularly suitable for testing other instru¬
which much larger shunts should have pensating resistors Rm' and R/. While ments, since all three types of tests,
been used. Separate correction curves these are small (the cause of the relatively (d-c, a-c, and a-c-d-c transfer tests) may
have been prepared for these ranges to large discrepancy [0.06%] between the be made with it.
show the change in resistance caused by corrections for the d-c and the a-c 5-
the self-heating of the measured current. ampere ranges is unknown), they are References
The a-c-d-c differences, defined as difficult to evaluate separately, so that
1. Multirange, Audiofrequency Thermo¬
i^ac Idc)/I<ic or (kac 1^dc)/Vdc for the future d-c tests must probably be made couple Instruments of High Accuracy, F. L.
Hermach, E. S. Williams. Journal of Research
same thermocouple emf with alternating on both a-c and d-c ranges. In most
National Bureau of Standards, Washington, D. C.
and direct current, are very small. They laboratories a volt box and shunt box vol. 52, 1954, p. 227.
depend upon the ratios of small reactances could be reserved as standards for these 2. The Testing of Electrical Instrument:
F. L. Hermach. Proceedings, Instrument Societ
to resistances in the instrument and may and other tests and could be sent to a of America, Pittsburgh, Pa., vol. 8, 1953, Pap
be expected to remain constant to the suitably equipped standardizing labora¬ 53-2-3.
accuracy desired. Thus, unless the com¬ tory periodically, or the instrument 3. Electrical Measurements (book), F.
Harris. John Wiley & Sons, Inc., New Yoi
ponents are changed, future periodic tests itself could be sent to such a laboratory. N. Y. 1952, chap. 11.
of the instrument need normally be made 4. Notes on the Design op Four Tbrmd
with direct current only, with a-c-d-c Conclusions Resistance Standards for Alternating C
rent, F. B. Silsbee. Journal of Research, Natic
checks made only occasionally. This Bureau of Standards, vol. 4. 1930, p. 73.
greatly simplifies the standardization of This portable, VA converter can serve 5. Thermal Converters as Ac-Dc Tran:
the instrument. as an accessory to a null-type d-c potenti¬ Standards for Current and Voltage Meas
ments at Audio Frequencies, F. L. Hern
Additional low-frequency tests of the ometer to make possible d-c and a-c meas¬ Ibid., vol. 48,1952, p. 121.
356/5
JOURNAL OF RESEARCH of the National Bureau of Standards—C. Engineering and Instrumentation
Vol. 64C, No. 3, July-September 1960
A continuously variable, 0- to 180-degree, phase shift standard for 400 cycles per second
is described in detail. It consists of a -r-section line made up of twelve 14.6 degree and three
4.3 degree sections to provide for two sizes of coarse steps and an RC circuit at the input
to the line to provide for fine steps and a continuous fine control. A method for accurately
adjusting the characteristic impedance of all 7r-sections to the same value, which is used as
the termination, was devised. Under these conditions it is shown that the phase shift
introduced by each 7r-section can be accurately computed from a measured value of induct¬
ance. The phase shift of each 7r-section was also determined by an experimental procedure
dependent upon a 180-degree phase shift introduced by a toroidal transformer. The values
obtained by these two independent methods agree to within 0.01 degree.
357/229
Under these conditions the phase shift of each section 4. Theory and Method of Measurements on
can be computed from a measured value of induct¬ 7r-Sections
ance. Also the input impedance is independent of
the number of sections so that the input can be part
The circuit diagram in figure 2 shows a generalized
of an R-C circuit used to get fine control of phase
7r-section delay line which can be used as a phase
angle.
shifter. To eliminate reflections so that the shift
The final design consisted of twelve 14.6° 7r-sections
introduced by any one section is constant irrespective
and three 4.3° 7r-sections with switching arrange¬
of other sections in the line, the terminating im¬
ments so that any desired combination of sections
pedance, Z0, must be made equal to the input im¬
could be connected as a delay line. The input to
pedance, Zln. Considering one section only and
this line was used as part of an R-C network which
solving for Z0 when Z0=Zln gives
incorporated 10 capacitor steps giving 0.44° phase
shift each and a variable air capacitor for continuous
Z,z2
fine control. (1)
Several circuit connections and switching arrange¬ 1 A
+ 4Z2
ments for the complete phase shifter were considered
and tried. The one finally selected is shown sche¬
matically in figure 1. The “input” or voltage to be Also,
shifted is connected to a 4,000-ohm noninductive
resistor, shunted by a capacitance-remove type vari¬ e
able air capacitor (1,100 to 100 pf), in series with e0= (2)
Z, . Z)
the input to the delay line. A decade set of mica 1+
2Z2 ' Z0
capacitors (0.004 to 0.044 /A) and a zero adjusting
capacitance are connected in parallel with the delay
Equation (2) shows that for each section the output
line input. A set of 13 single-pole, two-position,
voltage lags the input voltage by the angle repre¬
mercury-contact switches are arranged to connect
sented by the vector, l-j-(Z1/2Z2) + (Z1/Z0).
any number from 0 to 12 of the 14.6° 7r-sectious into
For a properly terminated line made up of pure
this delay line. Also a set of 4 switches is arranged
inductors and capacitors the phase shift per section
to connect any number from 0 to 3 of the 4.3°
7r-sections into the delay line. The end of the delay
line is connected to a resistor of 1,000 ohms shunted
by a capacitor of 0.0021 pif which was chosen to be
equal to the characteristic impedance of the line.
The terminated end of the line is connected to the
“output” binding posts of the phase shifter through
the “test-check” switch when it is in the “test”
position. Figure 2. Generalized w-section line.
3-4.3° 7T-
P ig ure 1. Schematic wiring diagram for 0 to 180 degree reference standard phase shifter 400 cps.
358/230
could be readily computed from the known values of
inductance and capacitance. However, there are
several practical considerations which tend to reduce
the accuracy of the computed value. For Zx=jwL
and Z2—l/ja)C, the terminating impedance as deter¬
mined from eq (1) must be
'2 L 1
(3)
'° C qo2LC
C02L2 C0*1*
(4)
4Zq ' 8Zq
359/231
Because of the procedure used in matching charac¬ C']2 and C"/2, using fine trimmers m parallel with
teristic impedance for each 7r-section, as described the main capacitance. This last balance is somewhat
in the following paragraphs, this computation does “tricky” to obtain because the two unbalanced
not require that the exact value of capacitance for voltages put in by C'/2 and C"/2 respectively are
the 7r-section be known. It does require an exact not 90° apart in phase—but after a little practice
knowledge of the terminating impedance. it can be obtained quite rapidly. By using a highly
selective tuned amplifier as detector this method
4.2. Procedure Used in Matching the Characteristic gives a very sensitive wav to adjust the characteristic
Impedance of Each 7r-Section impedance of the 7r-section to be exactly equal to
7?o and C0 in parallel.
By arranging a potentiometer-bridge circuit with The derivation of the relations between the
the 7r-section termination (C0 and R0 in parallel) as 7r-section circuit constants and the terminating
one of the arms and using a sensitive detector to impedance, for which the input and terminating
obtain a balance, any small variation in this ter¬ impedance are equal, is given in appendix 1. The
minating impedance can readily be detected. If all procedure used was to derive an expression for the
circuit elements are kept the same after getting an input impedance of the 7r-section (terminated by
initial balance, except that one 7r-section is added Co and 7?0 in parallel) and equate this to C0 and
ahead of the termination, the detector would remain R0 in parallel, giving a complex equation in circuit
in balance if the characteristic impedance of the parameters. The two separate equations obtained
7r-section were exactly equal to C0 and R0 in parallel. from the complex equation can then be used to
If the characteristic impedance is off slightly the determine any two of the circuit parameters in
detector can be rebalanced by a fine adjustment on terms of the others. First, it is assumed that the
the two capacitor legs of the 7r-section—thus experi¬ capacitors in the two legs of the 7r-section are equal,
mentally making the impedance of the 7r-section — /2 = C" /2=C/2. The expressions derived for
exactly equal to C0 and R0 in parallel. C0 and C/2 are given by eqs (9) and (10), respectively,
The circuit arrangement actually used in making in appendix 1. These values of C0 and C/2 can now
these adjustments is shown schematically in figure 4. be put in the two original equations bv letting
Most of the circuit elements shown in this diagram C'72 = C/2 (1+X") and C'/2 = C/2 (1+X'). Equa¬
either represent or correspond directly to those used tions (11) and (12) obtained in this manner can be
to make up the 7r-section phase standard (as shown solved for actual values of X' and X" which are meas¬
in detail in fig. 1). Ri is the 4,000-ftresistor. Ci is ures of the adjustments required in the capacitance
the fine control capacitor. C2 represents the capaci¬ legs to maintain the same characteristic impedance
tor decade and the zero set capacitor. The single for slight changes in other 7r-section parameters.
7r-section is used to represent the entire delay line. To help clarify the above procedure an example
C0 and R0 are the 7r-section termination. The only using actual values for the 14.6° 7r-section is worked
additional circuit elements required are the ratio out in appendix 1.
transformer and detector (the detector transformer The values of C'/2, i.e., C/2 (1+X'), obtained in
only being shown). The procedure used was to first this manner, together with the known or assigned
balance the detector with the 7r-section disconnected values of the other circuit parameters can be used
(switch A up and B open) by adjusting the ratio to compute phase shift as explained in the following
transformer and capacitor C2. Then witli a 7r-section section.
connected ahead of C0 and R0 (switch A down and
B closed) the detector was balanced by adjusting
4.3. Computation of Phase Shift for a 7r-Secticn
360/232
i?'=107th For R0= 1,000 0 and C'/2=C"/2=C/2, Table 2. Phase shift for various settings of the capacitors in
the R-C network
U0=0.002099X 10-6 and <7/2=0.0508191 X1CT6 (as
computed in appendix 1). Putting these values in (See figure 8)
eq (19), the phase shift angle 0, as shown in appendix
2, can be expressed as tan 0=0.25968. For the Pi =4,000 ohms; i?o=l,000 ohms
maximum changes in r, R' and R" encountered in
the twelve 14.6° 7r-sections as determined from Phase AB for 500 A0 for 3,900
Ci Ci shift, 0 pf change pf change
actual measurements, the changes in tan 0, as com¬ in Ci in Ci
puted in appendix 2, are given in the following table.
Pf Pf den deg deg
1,200 4, 000 f0.092U
700 -. 13823 0. 2303s
200 -.36862 . 2303g
r R’ R" Tan 6
1,200 7,900 -. 3570s 0. 4491s
700 -.58742 . 23039 .44919
2.80 10^ 107 0. 25968 .44922
107 200 -.81784 .23042
3.10 107 .25968
2.80 10s 107 .25967
.25969 1,200 11,800 -.8O6O9 .44906
2.80 107 10s
. 2304, .4490s
700 — 1. 0365o
2.80 10s 105 . 25968 .44909
200 -1.2669s .2304s
x-sections can be derived. As shown in appendix 2 1,200 23, 500 -2. 1519s . 44834
700 -2. 38244 . 23051 .4483s
this expression is 200 -2. 6129s . 23054 .44843
By using the special measuring techniques and the x-section line. As shown in table 2 the average
theoretical relations described in the previous sec¬ value of this phase shift is 0.2305° for a change in
tion, a computed value of phase shift for each of the Ci of 500 pf. (0.0046° for 10 pf or 0.01° corresponds
x-sections can be derived, and such values are listed to 21.7 pf).
in table 1. The phase shift inserted by the R-C The last column in table 2 indicates that the phase
network, connected ahead of the x-sections as shown shift put in by a 3,900-pf change in C2, i.e., by one
in figures 1 and 4, may also be computed using step on the decade capacitor dial, depends to a
known values of R and C. Appendix 3 shows how slight extent upon the total values for Oi and C2.
this can be done, and actual phase shifts for various For any given setting of C2 the change in this phase
values of C\ and C\ are given in table 2. shift for settings of C\ from 1,200 to 200 pf is less
One requirement for a standard phase shifter is than 0.00015°. Thus, the effect of the fine control
that the phase shift introduced by any given change capacitor setting on ’ the decade capacitor steps is
in the continuous fine control be independent of the negligible. As C2 is increased the phase shift for a
setting of other phase shift dials. From the fourth 3,900-pf change in C2 decreases somewhat, the
column in table 2 it can be seen that the phase shift maximum change being about 0.003°. The cali¬
put in by a 500-pf change in the fine control capac¬ bration for each step on the capacitor decade can
itor is the same to within 0.0005° for all settings on take this change into account, but this does indicate
(\ (the 0.46° steps) up to 43,000 pf. Thus, if the that another capacitor decade could not be added.
x-sections are all matched so that the impedance The phase shift for each step on the decade capacitor
looking into the line is the same no matter how could be computed if the value of capacitance added
many sections are connected, the phase shift put in for each step were known to a high accuracy (to
by a given change in the setting of the fine control about 1 pf for 0.0001°), but it was decided that the
capacitor will be essentially the same irrespective of experimental method as explained in the following
phase shift put in by the “capacitor decade” and section would be easier to apply.
572772 0 - 61 - 24
361/233
6. Experimental Calibrations 0, i.e., maximum capacitance and the capacitor
decade switch on step one, the auxiliary phase shifter
An experimental calibration can be devised for was adjusted (together with the ratio transformer)
any phase shifter which fulfills the requirements until the detector showed a balance. Then, with the
given in section 2. In brief, the procedure is first auxiliary phase shifter setting held constant and the
to obtain the phase shift of each step of each dial in decade capacitor switch set back to zero the detector
terms of settings on the other dials (including the was balanced by adjusting the fine control capacitor
fine control). These relative values can then be and ratio transformer. Since the auxiliary phase
converted to absolute values either by using a shifter remains unchanged for these two balances
known reference phase shift such as 180° or by assum¬ the phase shift put in by the first step on the decade
ing that the calibration of the fine dial is correct. capacitor must equal that read on the fine control
In the procedure actually followed the computed dial after the second balance. The change in phase
values for the fine control dial were used to get a shift when the decade capacitor switch is changed
calibration on each of the decade capacitor steps. from 2 to 3, 3 to 4, etc., can be measured in the same
Using these, a preliminary calibration of first the manner. The actual values as measured are given in
small and then the large 7r-sections was obtained. table 3.
Then the 180° point was checked using a special
Table 3. Phase shift for each step of capacitor decade switch—
toroidal core transformer. Details of the method
as determined by comparison with fine control dial
used will now be described.
A potentiometer-bridge circuit arrangement plus
Phase shift
an auxiliary phase shifter (see fig. 5) were used to Decade capacitor step Phase shift 2 corrected 2
of each step for 180°
compare the phase shift settings on the various dials check
of the 7r-section standard. The auxiliary phase
shifter, consisting of an R-C circuit supplied through deg
a ratio transformer, need not be accurately cali¬ 1_ 0.4468 0. 4470
2_ . 4429 0. 8897 .4431 0.8901
brated, but it must be stable and continuously 3_ .4454 1. 3351 .4456 1.3357
4_ . 4340 1. 7691 .4342 1.7699
variable from 0° up to at least 15°. A sensitive 5_ .4436 2. 2127 .4438 2. 2137
amplifier with shielded input transformer was used 6_ .4440 2. 6567 .4442 2. 6579
to indicate when the phase shifts put in by standard 7_ .4344 3.0911 .4346 3. 0925
8_ .4447 3. 5358 .4449 3. 5374
and auxiliary phase shifters were equal. The ratio 9_ .4406 3. 9764 .4408 3. 9782
transformer was used to give a magnitude balance. 10_ .4428 4. 4192 .4430 4.4212
362/234
than 0.005°. In order to correct for these fairly large
errors in the 14.6° 7r-sections introduced by very small
errors in the calibration of the capacitor decade, an
independent check of the 180° point was required.
This was done by connecting the primary of a
uniformly wound toroidal transformer in parallel
with the input to the standard phase shifter. The
secondary of the toroidal transformer was used to
get an exact 180° phase shift by (1) reversing polarity
or (2) using its two coils as a center tap winding
with the center grounded. The phase shifter zero
was set with the secondary polarity giving zero
phase shift, and then with the opposite polarity
the 180° point was measured on the phase shifter.
The magnitude was balanced by connecting a ratio
transformer either at the primary or at the secondary Figure 6. Setup used for calibrating a phasemeter.
of the toroidal transformer. The 180° point was
measured using five different arrangements for the
windings of the toroidal transformer and with the
“input” to the 7r-section phase standard is supplied
two locations of ratio transformer for each of the
by a 2-w low distortion oscillator with a regulated
live arrangements. Since the results obtained on
output of 5 v. This value of voltage was chosen
all measurements were the same to within 0.01° it
because it puts 0.25 v across each 14.6° 7r-section,
was concluded that the phase shift being measured
at which value the change of inductance with voltage
was exactly 180°. Using the preliminary experi¬
is minimum for the iron core inductors. It was
mental calibration, based on computed values of
found that a 5-percent change in voltage at this level
phase shift for the fine control capacitor, the mea¬
was required to produce a detectable change in
sured value came out to be 179.92°. Thus a correc¬
inductance. Actually the voltage is maintained
tion factor of 180/179.92 or (1 +0.000417) was applied
constant to within 2 percent. The reference or
to the preliminary calibration values to get the
unshifted voltage input terminals of the meter under
correct values as listed in table 1 for the 14.6 and
test are connected in parallel with the standard
4.3 degree 7r-sections and in columns 4 and 5 of table
“input.” At this point the value of impedance of
3 for the capacitor decade steps. As indicated in
the instrument under test is not critical because it
table 1 these measured values agree quite well with
merely acts as an additional load on the oscillator,
the values computed from measured values of induct¬
and its only possible effect would be a slight change
ance for each 7r-section. Although the change in
in waveform of the supply voltage. This effect
the capacitor decade calibration introduced by the
would be entirely negligible for most instruments
180° check is quite small, as seen from table 3, it
used as reference standards. If a value of voltage
should be included for accurate measurements.
other than that required for the 7r-section standard
(5 v) is required by the meter under test, a trans¬
7. Requirements for Use of Phase Standards former would normally be used. In such a case the
phase shift introduced by the transformer must be
known so that a correction for it can be applied.
The master phase angle 7r-section standard
If the impedance looking into the phasemeter
described in this paper is a continuously variable
phase shifter whose absolute calibration has been through the shifted voltage binding posts were in¬
finitely high, the output of the 7r-section standard
accurately determined. It is primarily intended to
could be connected directly to these “shifted volt¬
be used to establish corrections for other continu¬
age” binding posts without affecting the terminat¬
ously variable phase shifters and phasemeters used
ing impedance of the 7r-section line. Under this
as standards. Herein, a phasemeter is defined as
condition the “auxiliary termination” would not be
an instrument capable of measuring the phase shift
needed. However, since the performance of the ir~
between two alternating voltages of the same fre¬
section line is sensitive to small changes in it# ter¬
quency. Throughout this kind of calibration testing
mination, Z0, some compensation will normally be
the two factors most likely to affect, accuracy are
(1) the impedance of the instrument under test and required for the impedance at- the “shifted voltage”
binding posts. This compensation is provided for
(2) waveform distortion in supply voltage. These
two points are of sufficient importance to be con¬ by two insulated binding posts at the output end of
sidered separately the 7r-section phase standard, (see figs. 6 and 1), one
being connected to the output end of the 7r-section
line and the other to i?0 and C0 (in parallel) which
7.1 Compensation for Impedance of Instrument form the normal line termination. When these two
Under Test binding posts are connected by a copper link the 7r-
section line is terminated inside the 7r-sec.tion stand¬
The connections used for calibrating a phasemeter ard case and the normal output terminals can be
are shown in the block diagram in figure 6. The used. By removing the link the internal line ter-
363/235
mination is disconnected and the “auxiliary termi¬ the phase shift put in by the phase shifter being
nation” may be connected as shown in figure 6. tested is equal to that of the standard. For a given
This auxiliary termination consists of a stable ad¬ setting of the phase shifter being tested the detector
justable R-C network which is connected in parallel would be balanced by adjusting the phase shift put
with the leads from the 7r-section standard and in by the x-section standard and the magnitude of
arranged so that with the “shifted volts” leads also the output voltage from the phase shifter being
connected in parallel, the total impedance thus tested by using a ratio transformer.
formed can be adjusted to be equal to Z0 (the x-
section impedance). 7.2. Effect of Waveform Distortion
In order to make this adjustment a ratio trans¬
At the very beginning of this paper phase angle
former and a sensitive detector with a shielded input was defined to be the quantity used as a measure of
transformer are connected as shown in figure 6.
the time phase relation between two voltages having
First the leads from the 7r-section standard to the
pure sine waveform. So far the effects of distortion
auxiliary termination are disconnected, the link re¬ have not been mentioned. However, since it is
placed on the 7r-section standard and the switch put
impossible to obtain pure sine waves in practice,
in “check” position (see fig. 1). Under these condi¬
such effects must be considered. For a distorted
tions the x-section line termination is the same as it
waveform, phase angle is herein defined as the time
was when the capacitor legs were adjusted, i.e., it phase relation between the fundamental sine-wave
consists of Co, Ro, and the compensating capacitor components of the two voltages. Thus, there is no
Cs in parallel. With the detector transformer lead change in definition but a definite basis is established
in position “1” and all 7r-section standard dials set
for a discussion of the effects of unwanted harmonics.
on “zero,” the detector is balanced by adjusting the If all the circuit elements were linear (i.e., no iron
ratio transformer and the “zero set capacitor.” cores or lossy dielectrics were used) and if measuring
Next the line termination Co, Ro, Cs is replaced by and detecting instruments were only sensitive to the
the “auxiliary termination.” This is done by re¬ fundamental, then the use of a distorted waveform
moving the link from the x-section standard binding for making the measurements would make no differ¬
posts, putting the switch in the “test” position, and ence. Unfortunately, these postulations cannot be
connecting the auxiliary termination to the x-section
met in practice even though very selective amplifier
standard and the phasemeter under test as shown in detectors are available.
figure 6. The detector is then balanced by adjusting Since the x-section standard as described in this
the impedance of the auxiliary termination. To paper is made up of iron core inductors the voltage
correct for any phase shift introduced by the “auxil¬ drop across each inductor is always kept low enough
iary termination” and its leads, the detector lead so that flux density in the core is well below satura¬
position is changed from “1” to “2” and the zero set tion. Yet it is possible that distortion could be
capacitor is adjusted to balance the detector. Any introduced by the inductors even though the supply
correction for the phase shift in the ratio transformer voltage has a pure sine wave form. To investigate
can be made during this balance by actually setting this possibility, first an oscillator of very low distor¬
this phase shift on the x-section standard fine control tion was used to supply the phase shifter “input,”
dial before making the final balance. The phase¬ and distortion was measured at the “output” posts
meter under test can now be calibrated for any phase for various phase angle settings, i.e., for different
shift angle by first disconnecting all leads from the numbers of x-sections in the “line.” The waveform
phase shifters to the detector transformer, then set¬ with phase shifter set on “0” i.e., no iron core
ting the required angle on the x-section phase stand¬ inductors connected contained 0.02 percent 2d
ard and reading the corresponding value on the harmonic, 0.01 percent 3d harmonic, and 0.01 percent
phasemeter. of higher harmonics. As inductors were connected
If a voltage higher than the 1 v available from the all harmonics remained at about the same value
x-section standard output must be applied to the except the 3d which gradually increased to about
phasemeter under test, an amplifier can be used 0.04 percent with all twelve 14.6° x-sections in
between the auxiliary termination and the “shifted circuit. Thus for a nearly pure sine-wave input the
volts”' binding posts. The only requirements are inductors in the x-sections slightly increase the 3d
(1) the amplifier must have low distortion and (2) harmonic distortion.
the three balances as described in the previous para¬ Similar tests were made using an oscillator with
graph must be made with the amplifier and its leads somewhat greater distortion, i.e., about 0.03 percent
in position and the amplifier gain set to the same 2d, 0.5 percent 3d, and 0.16 percent 4th. For this
voltage as to be used during the calibration. The input voltage the effect of adding x-section inductors
phase shift introduced by the amplifier maybe fairly on the distortion was too small to be significant (less
large, thus requiring considerable change in the zero than about 0.02 percent). In order to check for
set capacitor during the final detector balance before possible effects of this distortion on measurements
starting actual calibration. involving only the fundamental component, the 180°
For calibrating another continously variable phase point was checked as described in section 6 using each
shifter the procedure would be essentially the same of the oscillators whose distortions have just been
as that just described for a phasemeter except that given. The results using the oscillator with 0.5
an external detector must be used to indicate when percent 3d harmonic agreed with those using the
364/236
nearly pure sine wave pscillator to within about two capacitor legs are not assumed to be equal.
0.005°. Thus it was concluded that for accuracies R" represents the total losses in C"/2 expressed as
within 0.01° the total wave form distortion should a parallel resistance and likewise R' for C/2. The
be less than 0.5 percent. It should be borne in total losses in the inductor are represented by r, as
mind that in making these tests a highly selective, a series resistance. Five circuit equations may be
feed-back tuned, amplifier was used in the detector written by inspection of figure 7:
circuit.
Waveform may also have an effect on or may be ixa—i2a—V,
affected by the phase shifter or phasemeter being
calibrated. For testing another phase shifter of the — —1— — ^3 R =0,
same design as the 7r-section standard the restrictions
on distortion should be the same as those given above. —i2R"-{-i3d— iiR'=0,
For other types of phase shifters or phasemeters the
effects of wave-form distortion should be determined ~\~iif ^5£7=0,
by repeat tests using two oscillators of significantly
different waveform. From a limited number of such 'M7+ '^(■Zo+ g)=0,
tests it is believed that in general for all oscillators
with less than 0.5 percent total distortion the phase- where
angle test results would be the same to within
about 0.01°.
b — R" -)-cyi’ d—R" YrJr R' Yj^L,
". C
8. Summary and Conclusions JU-
A continuously variable, 0° to 180° phase shifter
for 400 cps has been described in detail. By using
a special procedure for matching the characteristic f=R'+—ir> and g
C’ . C'
impedance of the L-C 7r-sections used in this phase J lo-
shifter, a method for experimentally measuring the
phase shift of each section has been devised. This
phase shift has also been computed from measured Using these equations and putting 1/Z0= (l//?0) +
values of inductance for each 7r-section. The results juC0, the following expression for input admittance
obtained by these two methods agree to within 0.01°. Fin is obtained:
These results indicate that this phase shifter can be
used as a master standard in calibrating other phase
shift devices.
Y in=ju
C",
^l+ju>R0\
(c°+f)I][1+1P+^"F7]
Similar phase angle standards for higher audio¬
frequencies could be designed, but on the basis of
the experience gained it is thought that the upper
2 1 ^1+jco^c!
M)] (r+jcoA)
r rR' r w2LR' /
1 R'
1 R" R0 1 R0R" 1 R" R" \
365/237
By equating the imaginary parts of the complex Likewise eq (7) gives
equation the following is obtained:
y+y'-^(y-\")+u2Lc0
Kq
(v-x")
2r(702 Z2
-co2zf (A' + A" + X'X")
£0<7 +A+7
0-T)]
co2Z(7q2 w2LC n . 2n
(12)
(7) ~2T
where
Equations (11) and (12) can be used to compute
/ l 1 Ro \ actual changes in C /2 and C"\2 that are made experi¬
71 /Ro = uL
\R' R" R'R"J mentally when adjusting the characteristic imped¬
ance of a given 7r-section to exactly match the ter¬
minating impedance (R0 and (70 in parallel). First
-\-corR0 the values for (7/2 and (70 are computed using eqs (10)
R' R"
and (9) by assuming average values for L, r, R' and
For (772= (7+2= (7/2 these two equations can be R". A value for R0 must also be chosen. Then if
used to solve for (7/2 and C0 when L, R0, r R', R" are these values of R0 and (70 are kept the same the effect
known. Equation (6) is used to solve for (70. of small changes in r, R', and/or R" on values of
Putting (7/2= (L/2RI) (1+5) where 5+1, this equation C" 12 and C’/2 can be computed using eqs (11) and
becomes: (12). The actual value of C'\2 can then be used to
accurately compute the phase shift for each indi¬
^R„rCl + 2^LC„-r^~ (l+S)!~+m=0. (8) vidual 7r-section as indicated in appendix 2.
The above procedure may be best illustrated by
Solving it for (70 gives taking actual values for the 14.6° 7r-section:
C02L2 5A+
1+
4 Rl 8 64 + > Putting these values in eqs (9) and (10) gives
(70=0.002099X 10“6,
cc2L2
[ 1+
4 Rl
(1+i)2+ m]-
(7/2 = 0.050819, XIO"6.
An estimated value is used first for 5. The exact
value is determined later from eq (10) and a recom¬ When these values are put in eqs (11) and (12) they
putation made if necessary. Using this value for give X' = X" = 0 for the values of r, R', and R" given
C0 and (7'/2 =(7"/2 = (7/2, eq (7) becomes above. For small changes in r, R', and R" these
equations give the corresponding values for X' and
X" required to- keep the characteristic impedance
_2+ A • 3r2
" 2^ Rl [> 4 c+A2 ( the same.
First keep all values the same as above except
Solving for (7/2 gives change r to 3.1 ohms. Putting values in eq (11)
gives
C C02L2
(1+7)2+||(1+T)3+...]
2 2i?2[1+7+ 4 Rl 0.99849 X' —1.00164 X" = 0.00952. (13)
(10)
where Putting values in eq (12) gives
3r2
T_4cc2L2 0.96613 X'+O+ege? X" = 0.000065. (14)
Now going back to eq (6) and putting (7'/2=(7/2 Solving eqs (13) and (14) for X' and X" gives
(1 + X') and 0"'/2 = <7/2 (1 = X"), where (7/2 is the
value from eq (10), it follows that X'= 0.00480 and X" =-0.00472.
X'-X"+^T (X'-X")(X' + X" + X'X") Next keep all values the same as assumed above
except put i?' = 108. Then eq (11) becomes:
=jrRpC rR{p\2 2(702 r2_m2_
L 2 LC C ^JLRoC u2LC ( U 0.99864 X' —1.00148 X"=0.00275. (15)
366/238
Equation (12) becomes l r
Solvings eqs (15) and (16) for X' and X" gives:
tan 0=0.25969.
— ^\R' (R'~\ f3 (y=®’
\ 2 / Now for all values the same, except R'—R" = 10SQ,
giving (772 = 0.05095 pi (from X' = 0.00274) and
Co+(772=0.05305 pi
— V—J7> + ^3 / —^7 + Zo\=0-
tan 0=0.25968.
H V*2 )
The changes in r, R'fi-R" as assumed above are
For Z0=R0/l+ju>C0R0, these equations give the the maximum to be encountered among the twelve
relation 14.6° sections, as found from actual measurements.
Since tan 0 is very nearly the same for all of these
Vput _1_ assumed values, the average values based on the
ii JL_|_ 1 wLA I JrA I I actual measurements may be used, i.e., r=2.8 ohm,
^Ro^R' R0 Ro^ R' J Ro R'=R"—107 ohm. For R0= 1,000 ohms these
367/239
values give C/2 — 0.0508191 pf and (70=0.002099 pf. The following circuit equation may be written by
When these values are put in eq (19) it reduces to inspection of figure 9:
a constant times L and for/=400 cps this gives
h(h\ + 7?0) ^2-^1 ^sRo=E,
tan 0=2.59678 L.
Thus for the actual inductors used, if the termi¬
nating impedance is held constant (R0= 1,000 ohms
and C0— 0.002099 /uf) and if the trimmers on each
capacitance leg are adjusted so that the impedance h-Serb^3 ^-S0-bj^r ^=0.
looking into the line equals the terminating imped¬
ance (by the method explained in section 4.2), the
phase shift introduced by each section is equal to These equations yield the following relation:
a constant times its inductance.
_ So_1+jttCiS,_
Appendix 3. Computation of Phase Shift of E S1-f-S0 ,/co/b 1C ~ (20)
the R-C Network 1 ' Sj -)- RG (Cl+Cx)
Figure 9 is a simplified circuit diagram of the R-C By clearing the j term from denominator the fol¬
network. Rx is a precision resistor accurately ad¬ lowing expression for tan 6 is obtained where 6 is the
justed to 4,000 ohms. (7, is a precision air capacitor angle by which e leads E:
with a capacitance remove dial variable from 1,050
to 50 pf. The change in (7, can be read from the dial R\ _Eq \
setting to within 1 pf so that it can be used as a Rx-\-Ro “ R\-\-Ro)
fine control of phase shift. S0 is the resistance tan 6= (21)
co2R\Ro
component of the ir-line input impedance (1,000 C1(C1+C2)
ohms). C-2 includes the capacitance component C0 R: + Ro
of the 7r-line together with the decade capacitor
(ten steps of 0.44° each), the zero setting capacitor For
and any lead capacitance. The purpose of the follow¬
ing derivation is to get an expression for phase shift 7f1=4,000 ohms, R0
between voltages E and e in terms of the circuit f co=2513.27
constants. = 1,000 ohms, and ]
[w2=6.31653X 106,
the expression becomes
368/240
U. S. Department of Commerce National Bureau of Standards
abstract
A description is given of the design and construction of the standard instrument
used at the National Bureau of Standards in testing wattmeters and watthour
meters.
The investigation to determine the accuracy of this instrument for alternating-
current testing at frequencies up to 2,000 cycles per second is described and the
results are given.
CONTENTS
Page
I. Introduction_546
II. Design and construction_547
1. Criteria of design-547
2. General features of design_548
'3. Details of design and construction_548
(a) Supporting framework_548
(b) Fixed coils_ 550
(c) Moving system_552
(d) Preliminary tests_ 554
(e) Series resistor_555
(f) Shielding_557
(g) Wattmeter constants_558
(1) Fixed coils_ 558
(2) Moving system_558
III. Performance tests_^- 559
1. Direct-current accuracy_ 559
2. Analysis of ac-dc errors_ 562
3. Errors depending on E or I separately—“Aa”_563
(a) Interaction between parts of voltage circuit_563
(b) Magnetic impurities in the moving system_563
(c) Induced and capacitance currents in the moving system. 564
4. Errors depending upon El—test at unity power-factor_567
(a) Method_567
(b) Procedure_ 568
(c) Results_569
5. Errors depending upon phase angle 0—tests at zero power-
factor_569
(a) At 60 cycles per second, using quadrant electrometer._ 569
(1) Method_569
(2) Procedure_570
(3) Results_572
(b) At high frequencies, using capacitance load_573
(1) Method and procedure_573
(2) Derivation of equation_574
(3) Results_ 575
IV. Summary_ 575
V. Appendix I. Computation of the effect of variation in wattmeter sen¬
sitivity_ 577
VI. Appendix II. Phase difference between the fixed-coil current and the
quadrant voltage on the electrometer and the relation of this phase
difference to 0, the phase angle between Eand I_578
369/545
Journal oj Research oj the National Bureau of Standards iva. 25
I. INTRODUCTION
370/546
LaeTJi’3] Standard Electrodynamic Wattmeter
that it does not respond to direct current, and with all instruments the
range of which exceeds the range (10 amp and 300 v) of the standard
transfer instruments. In this latter case instrument transformers
must be used to extend' the range of the transfer instrument. It may
be noted that any self-heating or shift of zero which occurs in the
instrument under test during the course of this straight a-c test will be
included in the final result, and the accuracy of a test of this type is
therefore definitely less than that of the transfer test.
The transfer instruments usually used at the National Bureau of
Standards are of the electrodynamic type. The voltmeter in use at
the present time has been described in a previous publication.3 The
present paper describes the wattmeter now being used. Although
designed primarily as a wattmeter, this instrument may also be used
as an ammeter if it is connected to measure the power consumed in an
external shunt.
1. CRITERIA OF DESIGN
371/547
Journal oj Research oj the National Bureau oj Standards ivoi.ts
372/548
Lewii\ Standard Electrodynamic Wattmeter
These glass panels are hinged to each other, one being stationary and
the other opening and closing like a door. In the closed position the
corresponding halves of the fixed coils and the two halves of the damp¬
ing box fit snugly together to form two complete fixed coils and one
damping box.
The difficulty of attaching hinges directly to the glass forced the use
of supporting frames of mahogany for the two glass panels. The
hinges are attached to these frames. These frames also allow the lack
of squareness of the comers of the glass panels to be corrected for and
concealed. The use of any wood at all is an undoubted disadvantage.
However, half of each fixed coil, half of the damping box, and the
entire moving system are supported entirely by the stationary glass
panel. The other half of the fixed-coil system and the other half of the
damping box are supported by the second glass panel, which is at¬
tached to the first panel by the hinges and the mahogany frame. Any
warping of the wooden frames can result only in a displacement of the
entire second panel with respect to the first panel. The possibility
of warping was reduced to a minimum by carefully varnishing the
mahogany inside and out.
In order to provide an immediate and visible check on any gross
misalinement of the two glass panels with respect to each other, two
groups of fiducial marks were etched on the panels. Those groups were
placed at widely separated intervals on the panels. Each group con¬
sists of a large cross etched on the rear surface of the rear panel, a
smaller cross on the rear surface of the front panel, and a still smaller
cross on the front surface of the front panel. With the instrument as
originally assembled, the panels properly alined and the moving
system free from sticking, the three crosses constituting each group lie
accurately on a straight line perpendicular to the face of the panels.
They may therefore be lined up by eye so as to fall one upon the other.
Any subsequent displacement of one of the panels with respect to the
other will be evident as a displacement of these fiducial marks from a
straight line. It will be possible to bring the panels back into their
original alinement by inserting suitable shims between the glass panels
and the wooden frames.
The glass panels were made of plate glass of rectangular shape,
1 in. thich and 8 in. wide by 44 in. long. The only machine work
required was to cut the glass to size and drill holes through it for
mounting the wattmeter parts. Round holes are readily put through
glass by using a tubular tool in a drill press with carborundum and
water as a grinding compound. Holes of other shapes can be made by
drilling a series of small round holes around the outline desired and
then smoothing down the edges with a flat brass tool worked by hand
with carborundum. The only prerequisite in working this glass was
found to be patience. The location and size of the holes drilled in the
glass panels are shown on the assembly drawing of the wattmeter in
figure 1.
The fixed coils were mounted on the glass panels by first embedding
them in bakelite holders which could be inserted into round holes in
the glass panels. An attempt was made to mold the fixed coils into
the Bakelite holders under heat and pressure by constructing a die
which was used in a vulcanizing press. The heat and pressure of the
press destroyed the insulation of the coils and the attempt had to be
abandoned. The die was remade for molding cylmdrically shaped
373/549
Journal of Research of the National Bureau of Standards ivoi.ss
coil holders closed at one end. The fixed coils were cemented into
these holders with sealing wax. Coils thus made have been in service
5 years now and still appear to be very solid. The coils in their holders
were fixed in the glass panels with fiber shims pressed and cemented in
place with a cellulose-acetate cement. The Bakelite holders were
slightly smaller than the holes, and the holes in the glass were slightly
conical to permit this construction. In constructing another instru¬
ment it might be desirable to make the holes even more conical de¬
liberately for this purpose.
The damping box was made of molded Bakelite. It was molded
in two pieces, the same die being used for each piece. Each half of the
damping box fits into an opening in the bottom of each glass panel,
as is shown in figure 1, and was fixed firmly in position by small
wedges cemented in place. An opening was also made in the upper
half of the movable glass panel to allow the free passage of a beam
of light to and from the mirror attached to the moving system. The
trim surrounding this opening was also made of molded Bakelite.
A thin piece of cover glass, cemented on the rear surface of this piece
of trim, allows the passage of light but protects the interior of the
instrument from the stray air currents of the room.
(b) FIXED COILS
374/550
Journal of Research of the National Bureau of Standards Research Paper 1344
375
ja
o
a
C/5
6
o
376
Park, I
Lewis \ Standard Electro dynamic Wattmeter
to the flux of the lower coil, thus giving astatic construction. The
two ends of the fixed-coil cable were brought out to a terminal board
located on the wall near the wattmeter. The terminal board is ar¬
ranged so that the four coils formed by the four wires in the fixed
coil cable can be connected in series, series-parallel, or parallel.
572772 0 - 61 - 25 377/551
Journal oj Research of the National Bureau of Standards [ voi. n
5 If the entire moving system is not rigidly interconnected, it is possible, particularly at very low power
factors, for forced internal vibrations to take place between its component parts. The result is a broadening
of the reflected spot of light. Because of the varying scale law of the instrument, particularly if the moving
coils are not at the position of zero mutual, the true position of the spot of light will not correspond to the
midpoint of this broadened band of light. The apparent indication of the instrument will therefore be in
error.
378/552
Park, 1
Lewis] Standard Electro dynamic Wattmeter
Figure 3.—Details of the movable head to which the upper end of the suspension
is attached.
the larger moving coils were chosen for permanent mounting in the
wattmeter.
The details and dimensions of the brass forms used for winding
these coils are given in figure 4. Each coil contains 100.5 turns of
No. 32 silk-enamel copper wire. After each layer of a coil was wound
(about 19 turns), a coat of thin shellac was applied and allowed to
dry. This served to hold the completed coil together after removing
the form. The completed coil was then dipped in thin insulating
varnish and baked at about 120°C for 24 hours. This procedure
was repeated four times to form a tough protective coating over the
coil and to give it rigidity. The coils were mounted on the moving
system by inserting the suspension tube through the two holes in the
379/553
Journal oj Research oj the National Bureau of Standards [voi. ts
top and bottom of the coil. After the moving; coils were located on
the tube so as to fit into the hollow centers of the fixed coils and the
two moving coils were twisted into the same plane, they were lashed
to the tube with silk thread and shellac was applied to hold them
rigidly to the tube. The two moving coils were connected in series
in such a manner that the direction of winding on the upper coil is
reversed with respect to the direction of winding on the lower coil.
The two remaining free ends of the coils were then soldered to the
upper and lower suspensions.
The damping vane, iy2 in. high by 4 in. wide, is made of 10-mil
sheet aluminum with the edges rolled for stiffness. This damping
vane is riveted to the suspension tube at the proper distance below
the lower moving coil to fit into the damping box when the moving
system is in place.
Two mirrors are attached to the suspension tube above the upper
moving coil by tbe use of brass collars and spiders. One mirror (%
in. in diameter) is located so that with the moving coils in the position
of zero mutual inductance with the fixed coils it reflects a beam of
light from a straight filament lamp onto the center of a 50-cm ground-
glass scale located 2 m directly in. front of the wattmeter. The other
mirror (% in. in diameter) is located on the tube so that its plane makes
an angle of 60 degrees with the plane of the first mirror. The mechan¬
ical zero of the moving system is normally set so that the lightbeam
reflected from this smaller mirror falls on the center of the ground-
glass scale when the instrument circuits are open. This allows any
shifts in the mechanical zero to be readily detected.
Before designing a senes resistor for the voltage circuit of the watt¬
meter, it was necessary to know the maximum allowable moving-coil
current and the wattmeter sensitivity. The wattmeter was therefore
set up in its operating position and the optical system adjusted until
the spot of light from the %-in. mirror fell on the center of the ground-
glass scale when the moving coils were in their position of zero mutual
inductance with the fixed coils. The position of zero mutual induc¬
tance is readily determined by passing an alternating current through
the fixed coils and shorting across the moving-coil terminals. If the
moving coils are in the zero mutual position, there will be no deflection
of the moving system. If they are not at the position of zero mutual,
there will be an emf induced in them and a current will flow when their
terminals are shorted, causing a deflection towards the position of
zero mutual. This test can be made as sensitive as desired by increas¬
ing the frequency of the current in the fixed coils.
A preliminary value of the sensitivity of the wattmeter with the
moving coils in the position of zero mutual was determined by passing
rated direct current (2.5 amp per wire) through the fixed coils ana
measuring the direct current through the moving coils necessary to
obtain a deflection of about 10 cm with a scale distance of 2 m. The
sensitivity with style A fixed coils and the smaller moving coils was
found to be 34 cm/ma in the moving coil. The sensitivity for style B
fixed coils and the larger moving coils was found to be 125 cm/ma in
the moving coils. These latter were therefore permanently mounted
in the instrument. This latter sensitivity might alternatively be
380/554
Park,]
LewisJ Standard Electrodynamic Wattmeter
expressed as 2.4 milliradians per ampere turn in the fixed s}Tstem per
ampere turn in the moving system.
The maximum allowable current through the moving coils is limited
by the heating due to this current. A heating test was made with 2.5
amp through the fixed coils (all connected in series) and 0.1 amp
through the moving coils. Under these conditions the moving coils
showed a temperature rise of 9.5° C at the end of 2y2 hours. A current
of 0.1 amp is therefore considered a safe value for the moving system.
Higher current values might be used, but their use is not anticipated
at this time.
(e) SERIES RESISTOR
381/555
Journal of Research of the National Bureau of Standards [Vot.ts
382/556
Park,!
Lewis J Standard Electrodynamic Wattmeter
(f) SHIELDING
383/557
Journal oj Research of the National Bureau of Standards [Voi.ts
shielded lead with rubber insulation over the shield (the capacitance
between conductor and shield being 710 g/if). At the wattmeter end.
the shield of this lead is connected through a binding post to the
lower moving-coil suspension, which is also tied to the wattmeter
shield. The central wire of this lead is connected through a binding
post to the upper moving-coil suspension. At the multiplier end,
the shield of this lead is connected to the lower moving-coil “MC”
binding post (the “±” supply voltage lead to which the electro¬
static tie between current and voltage circuits is attached is also
connected to this binding post). The central wire of the shielded
lead is connected to the upper moving-coil binding post. The multi¬
plier metal case is tied to the lower moving-coil binding post and
thus acts as a shield for the multiplier.
(g) WATTMETER CONSTANTS
384/558
Park, 1
Lewis J Standard Electrodynamic Wattmeter
Before the ac-dc errors were determined, the accuracy of the in¬
strument assuming no ac-dc errors, that is d-c accuracy, was investi¬
gated. The factors effecting the d-c accuracy are (1) drift of de¬
flection with time for a constant value of power, (2) variation of
wattmeter sensitivity for small changes in deflection, and (3) imper¬
fection of astatic construction.
With a constant value of current in the fixed coils and a constant
voltage across the moving-coil circuit, the deflection of the wattmeter
was noted during a period of 2 hours. The drift in deflection was
found to be very nearly a linear function of time and equal to 0.1
mm/min. In ac-dc transfer tests the effect of this drift is cancelled
in the average a-c and d-c readings by the sequence in which the
readings are taken. When the instrument is used as a standard
wattmeter, it is calibrated on direct current before and after the a-c
readings are taken so that here also the effects of drift tend to cancel.
Even if the drift were not cancelled by the sequence of taking readings,
the maximum error caused by it would be only 0.025 percent in 10
minutes.
In an ac-dc transfer test of a wattmeter that is neither magnetically
shielded nor of astatic construction, the reading of the standard
instrument may be several percent different for one direction of
current and voltage than for the reversed direction; also, the a-c
reading may be slightly different from the d-c readings. Since the
scale of the standard instrument is uniform, if its sensitivity is differ¬
ent for the various readings of the standard instrument in one set of
data, an error will be introduced which may be as large as the per¬
centage difference in sensitivity.
The sensitivity of the wattmeter can be derived from the following
quantities:
G= torque on moving system in dyne-centimeters per milliampere in
the moving coils with rated current in the fixed coils.
a=angular deflection of moving system from the mechanical zero
position as measured by the travel in centimeters of the working
spot of light at a 2-m radius from the mirror.
Z7=restoring torque of suspension in dyne-centimeters per unit of
angular deflection, a.
i=moving-coil current in milliamperes.
Since the summation of the torques on the moving system is zero, for
any fixed position of the moving system
iG—aU— 0 or a~ijjy
385/559
Journal of Research of the National Bureau of Standards [voi.gs
from which
ko_(Xg aL
U iB ii,
__Go-
°k) jj^o-
The total deflection at any other point on the scale was found by
adding the difference between the scale reading at this point and the
scale reading at the point of zero mutual to the total deflection at the
point of zero mutual. This gave total relative deflections for various
points on the scale which were as accurate as the scale could be read.
From these data the relative wattmeter sensitivity at any point on
the scale was computed by dividing the total deflection at the point
by the moving-coil current required to-deflect the wattmeter to this
point. The relative sensitivity this obtained is in deflection (in centi¬
meters) per milliampere in the moving coils with rated current in the
fixed coils. A plot of this sensitivity against scale reading is shown in
figure 6. This plot shows that there is a range of 10 cm near the point
of zero mutual within which the sensitivity is constant to within 0.01
percent. Thus no corrections for wattmeter sensitivity are required
for readings within this range. For readings outside this range cor¬
rections 6 can be applied.
As seen from the curve of figure 6, the point of zero mutual induct¬
ance between the fixed and moving coils (as indicated by the arrow)
6 For a computation of these corrections see appendix I.
386/560
Park, 1
Lewis] Standard Electro dynamic Wattmeter
0 5 to 15 20 25 30 35 40 45 50
SCALE READING IN CM
Figure 6.—Relative wattmeter sensitivity for various positions of working spot on
the ground-glass scale in centimeters of deflections per milliampere in the moving
coils with rated current in the fixed,coils.
The absolute wattmeter sensitivity is equal to the relative sensitivity to within 0.3 percent.
elusion is that the flux density is not strictly uniform in the present
instrument, and this is probably the case in most instruments of this
type.
Imperfection of the astatic construction of the wattmeter results in
a small deflection of the moving system with direct current through the
moving coils only, because of interaction with the earth’s magnetic
field. This deflection was measured with 60 pia (the maximum value
to be used) through the moving coils for various positions of the mov¬
ing system near the point of zero mutual inductance between the fixed
and moving coils. These deflections as measured by the movement
of the light beam on the ground-glass scale were found to be +0.07
cm for one direction of moving-coil current and —0.07 cm for the other
direction, and they were the same for all positions of the light beam
on the scale within ±15 cm of the point of zero mutual inductance.
Thus in using the instrument for an ae-de transfer test, even though
387/561
Journal oj Research of the National Bureau of Standards \voi.u
the two d-c readings are quite different, the small deflection due to
lack of astaticism will be cancelled out in the average d-c reading.
and the difference between the average d-c deflection and the a-c
deflection is
388/562
Park, 1
Lewis J Standard Electrodynamic Wattmeter
389/563
Journal oj Research oj the National Bureau oj Standards Woi. *5
The component parts of the moving system (1, moving coils; 2>
mirror; 3, collars and clamps for holding mirrors; 4, collar with
adjustable weights for balancing the system; 5, damping vane; and
6, Bakelite tube upon which all other parts were mounted) were all
made of materials that are normally considered to be nonmagnetic.
With the moving system assembled it was impracticable to determine
which of the component parts contained magnetic impurities. A new
moving system was constructed and each component part was tested
for magnetic impurities before assembly. This test consisted in
suspending the part, with a small mirror attached to it for reflecting
a beam of light onto a scale, in the field of a strong electromagnet and
noting the deflection of the beam of light when the coil of the electro¬
magnet was energized. These tests showed that a Bakelite tube,
obtained from the same source as the tube used in the first moving
system, contained considerable magnetic impurities which appeared
to be concentrated in a plane containing the central axis of the tube.
This would account for the results obtained with the first moving
system in the wattmeter. In searching for a substitute for this
tube, samples of glass, porcelain, clear Bakelite (no filler), and other
resins were tested for magnetic impurities. It was found that these
samples contained only very slight magnetic impurities in comparison
with a sample of the Bakelite tube used in constructing the first moving
system. Thus it was concluded that the filler used in making the
Bakelite tube and not the Bakelite itself contained the magnetic
impurity. Four samples of Bakelite tube were obtained from separate
sources and tested for magnetic impurities. Of these four, one was
chosen which contained about the same amount of magnetic impurity
as the glass or porcelain, and a new moving system was constructed
using a tube of this material for mounting the moving-system parts.
Before mounting each component part on the new moving system,
the system was suspended in the wattmeter and its deflection was
noted with current through the fixed coils only. At each stage of
construction this deflection was just barely noticeable with rated
current through the fixed coils. When the moving system was com¬
pleted, twice-rated current was passed through the fixed coils and
measurable deflections of the moving system were obtained. The
moving coils were then twisted around on the Bakelite tube until the
position of zero mutual inductance between the fixed and moving coils
(the position near which the moving coils are held when the wattmeter
is in use) coincided with the position for minimum deflection of the
moving system with current through the fixed coils only. The deflec¬
tion of the moving system at this position was found to be about 0.2
mm (0.005 percent of normal wattmeter deflection) with twice-rated
current through the fixed coils. Thus it was concluded that the error
due to magnetic impurities would be negligible with rated current
through the fixed coils.
390/564
Lewi's] Standard Electrodynamic Wattmeter
fixed coils with the moving coils open-circuited and noting any deflec¬
tion of the moving system. When rated current at a frequency of
2,000 c/s was passed through the fixed coils, deflections of the moving
system as large as 1 cm were obtained. These deflections are the
result of a combination of the following effects: (1) eddy currents
within the thickness of the wires forming the individual turns of the
moving coils, (2) eddy currents in other parts of the moving system,
such as the damping vane, and (3) capacitance currents flowing in
some of the moving-coil turns and through the capacitance existing
between the fixed and moving coils.
The summation of the torques produced by eddy currents in indi¬
vidual turns of the moving coils tends toward zero and would be
exactly zero with the moving coils in the position of zero mutual
inductance with the fixed coils if the fixed-coil flux were uniform
throughout the area occupied by the moving coils. Any torque pro¬
duced in the present wattmeter by the eddy currents in the moving
coils may be combined with the torques produced by eddy currents in
other moving parts of the wattmeter, since the torques due to all these
eddy currents are similar functions of the same variables.
The three effects causing the deflections noted above may now be
considered as two: (1) eddy currents in moving system, and (2)
capacitance currents flowing in some of the moving-coil turns. The
deflection due to eddy currents will (a) increase approximately as the
square of the frequency of the fixed-coil current, (b) increase as the
square of the current through the fixed coils; but it will be independent
of (a) the efficacy of the shield between the fixed and moving coils,
(b) the location of the electrostatic tie between the fixed and moving
coils, and (c) the connection of the fixed coils (series, series-parallel,
or parallel). The deflection due to capacitance currents will (a)
increase approximately as the square of the frequency of the fixed-
coil current, (b) increase as the square of the current through the
fixed coils, (c) depend upon the efficacy of the shield between the
fixed and moving coils, (d) be approximately unchanged in magnitude
but reversed in direction when the electrostatic tie between fixed and
moving coils is changed from one terminal of the fixed coils to the
other terminal, and (e) depend upon the fixed-coil connections, being
largest for series and least for parallel. The total deflection caused by
eddy currents and capacitance currents may be separated into its
two components by use of relation (d). For a given magnitude and
frequency of fixed-coil current the deflection of the moving system
was noted with (1) the ± or b terminal of the moving coils connected
to terminal a (see fig. 5) of the fixed coils, (2) the 6 terminal of the
moving coils connected to terminal b of the fixed coils. The algebraic
mean of these two deflections is the deflection caused by eddy currents,
and one-half the algebraic difference between these two deflections is
the deflection caused by capacitance currents.
From tests performed on the wattmeter with rated current at a
frequency of 2,000 c/s through the fixed coils, the deflections due to
eddy currents in the moving system and capacity currents in the mov¬
ing coils were determined and the results are shown in table 1. These
tests were performed with the shield between the fixed and moving
coils tied to the ± side of the moving-coil circuit as shown in figure 5.
With this shield “floating,” the deflections due to capacitance currents
were found to be about three times those given in the table.
391/565
Journal oj Research of the National Bureau of Standards Woi. ts
Deflection
Deflection
Total watt¬ due to
due to
Fixed coil-connections Ic MC terminal b tied to meter de¬ capaci¬
eddy cur¬
flection tance
rents
currents
amp cm cm cm
fFC a __ 0. 47
Series. 2.5 } 0.76 0.28
\FC b .... 1.04
fFC a _ 0. 50
Series-parallel 5 } .68 .18
\FC 6___ .86
392/566
Park, ]
Lewis J Standard Electrodynamic Wattmeter
errors can be separated from the total ac-dc errors measured in the
experiments to be described.
4. ERRORS DEPENDING UPON El—TEST AT UNITY POWER-FACTOR
(a) METHOD
572772 0 - 61 - 26
393/567
Journal oj Research oj the National Bureau oj Standards [Volts
switch (see fig. 7). The series resistor in the moving-coil circuit of
the wattmeter was adjusted to bring the wattmeter deflection to
normal full-scale position (this required about 1,300 ohms in the
moving-coil circuit), and the current was left on for one-half hour
to allow both instruments to come to temperature equilibrium. The
test data were taken in the following order: (1) with alternating
current of the desired frequency on, the value of current was adjusted
to give full-scale deflection on the hot-wire ammeter and the de¬
flection of the wattmeter was read;.(2) with direct current on and
adjusted to duplicate the setting on the hot-wire ammeter, the deflec¬
tion of the wattmeter was read; (3) with direct current reversed,
another reading of wattmeter deflection was obtained; and (4) with
alternating current on, the first step was repeated. The difference
between the average a-c (steps 1 and 4 above) and the average d-c
8 A quadrant electrometer connected to measure power could be used as a standard Instrument for this
purpose, but a considerable amount of time and patience would be required to set up and calibrate an instru¬
ment of this type so that the desired accuracy could be attained.
394/568
Lewu] Standard Electrodynamic Wattmeter
For the purpose of comparison the results of these tests have been
entered in figure 8. As already noted, the curve in this figure is a
plot of the wattmeter deflection “Aa” against frequency, obtained
on a previous test with 5 amp through the wattmeter fixed coils con¬
nected in seiies-parallel and the moving-coil circuit open (this is
the deflection due to eddy currents in the moving system and capaci¬
tance currents in the moving coils). The plotted points on this
figure represent the values of ac-dc errors of the wattmeter at unity
power-factor obtained by the tests just described. The agreement
of the plotted points with the curve (with the fixed coils connected
in parallel only one-half of rated current was used and the ac-dc
error is correspondingly less) indicates that for frequencies up to
3,000 c/s the ac-dc errors of the wattmeter at unity power-factor
are due only to eddy currents in the moving system and capacitance
currents in the moving coils. These errors enter into eq 1 as the term
“Aa”. The error designated as “Am” in this equation is thus found
to be negligible (less than 0.01 percent) for frequencies up to 3,000 c/s.
(1) Method.
The source of current and voltage consisted of two alternafors
driven by a direct-current motor, the shafts of all these machines being
coupled together. The motor was supplied from storage batteries, and
the speed was adjusted to give a 60-cycle output from each generator.
The output voltage of each alternator and the phase angle between
their voltages could be varied independently. One alternator was
used to supply the voltage circuit of the standard wattmeter, and this
395/569
Journal oj Research oj the National Bureau oj Standards [voi. t&
same voltage was connected between the case and the needle of the
quadrant electrometer. The other alternator was used to supply the
current through the fixed coils of the wattmeter, and by also passing
this current through a shunt and then stepping-up the drop across
the shunt by means of a voltage transformer whose secondary was
connected to the quadrants of the electrometer, a voltage was applied
to the quadrants which bore a fixed relation to the current through
the wattmeter fixed coils.
The voltage, E, from one of the alternators and the current, I, from
the other alternator correspond to a value of power El cos 0 which
Figube 9.—Electric connections for comparing the wattmeter with a quadrant elec¬
trometer at zero power-factor.
i?=resistance of shunt. ineffective inductance of transformer primary
£=inductance of shunt. winding.
r-= resistance of shunt potential leads. h =transformer primary current.
f?i=effective resistance of transformer primary Ei = transformer primary voltage,
winding. Uo = transformer secondary voltage from a to c.
Eb=transformer secondary voltage from a to b.
396/570
Standard Electrodynamic Wattmeter
397/571
Journal of Research of the National Bureau of Standards [voi.tt
Average_ 0.16
Average.... 0.34
.2 / 7.19 0.41
160 10 .do... 9,000 s . 6.78
\ 7.00 . 22
Average.... 0.32
7.22 0.05
7.22 .05
160 6 Series-parallel. 6,000 s. .4 7.17
7.14 -.03
7.07 -.10
Average_ -0.01
f 8.70 -0.01
8.74 -.03
160 2.5 Series__ 6,000 s. .8 8.71
1 8.58 —. 13
l 8.58 -.13
Average.... -0.06
* A positive value of rw indicates that the moving coil current leads the supply voltage.
398/572
Par A,]
Lewis J Standard Electrodynamic Wattmeter
was primarily dependent upon the phase angle of the potential trans¬
former. The phase angle of this transformer had an uncertainty of
±0.1 minute, probably due to magnetic instability of the core. The
experimental error in comparing the wattmeter and the electrometer
was ±0.1 minute. Thus the total uncertainty in tw was ±0.2
minute, which is sufficient to account for the variations of the phase
defect of the wattmeter as indicated in the last colunm of table 2.
The conclusion is that the phase defect of the wattmeter at a frequency
of 60 c/s is very probably less than 0.4 minute for any fixed-coil con¬
nection and for any arrangement of the multiplier circuit as normally
used.
(b) AT HIGH FREQUENCIES, USING CAPACITANCE LOAD
399/573
Journal oj Research of the National Bureau of Standards \voi. es
Figure 10.—Electric connections for measuring wattmeter phase defect at high fre¬
quencies using a capacitance load.
C=mica capacitor. rv=resistance of voltage divider for electrostatlo
i?=adjustable resistor. voltmeter.
rmresistance of wattmeter voltage circuit. E= voltage applied to wattmeter voltage circuit.
and voltage with S2 in the left-hand position and adjust the resistor,
R, until the wattmeter deflection was the same as just read. The
wattmeter phase defect can then be deduced from the value of the
resistor, R, capacitor, C, and phase defect, rc, of capacitor C.
(2) Derivation of Equation.
By referring to the wiring diagram in figure 10 and using the nota¬
tion indicated on the diagram, two equations for the quantity measured
by the wattmeter can be written. When the capacitor, C, is used as
a load on the wattmeter (switch S2 to the right) the quantity is
7?2 rr 2
Pc=wCE2 sin (re+0+—+—■
rm rv
When the resistor, R, is used as a load on the wattmeter (switch S2
to the left), this quantity is
E2 , E2 , E2
Pb= T+T+TT
400/574
Park, 1
Lewis] Standard Electrodynamic Wattmeter
During the tests, the voltage, E, and the deflection of the wattmeter
were held constant for the two loads on the wattmeter. Thus Pc= PR,
and the two equations give
wC sin (tc+t„,)=^;
IV. SUMMARY
The electrodynamic instrument described in this paper can be used
as a standard wattmeter or ammeter and as a standard ac-dc transfer
instrument in testing wattmeters or ammeters. The current range is
from 0.25 to 10 amp. The voltage range is from 25 to 300 v. The
10 This value is an average of all the data taken at 60 c/s.
401/575
Journal of Research of the National Bureau of Standards [vw.««
results obtained from this instrument are accurate within 0.01 percent
when it is used at power frequencies. The only correction which must
be applied to obtain this accuracy is the one due to the phase defect
FREQUENCY IN C/S
Figure 11.—Plot of wattmeter phase defect against frequency.
Results obtained from tests using a capacitance load.
402/576
Park, 1
LewisJ Standard Electrodynamic Wattmeter
V. APPENDIX I
The power being measured is proportional to Ii, and I (rated) is constant for a
given connection of the fixed coils. Thus the deflection may be expressed as
where K is a constant and W is the power being measured. If aa, Sa, and Wa de¬
note values with alternating current and voltage; ady Sd, Wd denote values with
direct current and voltage; aT, S,, WT denote values with direct current and volt¬
age in the reverse direction
«a=KSaWay or Wa=gtJ
ar = K'5rITr, or =
The relation to be obtained is the ratio of^prarto the average d-c power in terms of
the a’s and S’s.
2Wa . Wg—j(Wd-\- Wr)
Wd+wr T \{Wd+WT)
Substituting values of a and S, 8 becomes
an _ 1_/ad, aA
x_KSa 2K\SESr/[
1 ( “dljoA
2 K\SE Sr)
Since Sd and Sr differ but slightly from Sa, by putting Sf=Sa+sd and Sr =
Sa + Sr the above equation simplifies to
, _aj + ar
5=
~~2~
Qd + Or
2
One hundred times the first term on the right side of the above equation gives the
percentage excess of a-c power over average d-c power as normally computed
assuming constant sensitivity. One hundred times the second term gives the
correction in percentage to be applied when the sensitivity for ad and aT are differ¬
ent from the sensitivity for <*„. Values of Sa, sr, and sd are obtained from the
curves of sensitivity against deflection in figure 6.
403/577
Journal oj Research oj the National Bureau oj Standards ivoi. ts
VI. APPENDIX II
The currents in this equation are all complex quantities, but 7 may be assumed
to have no imaginary component. By equating the real and imaginary compo¬
nents of eq 3 and solving for the two components of 7,, these may be substituted in
Ei — ItiRt+juLt)
to obtain:
For the present work the approximate values of the constants in the above
equation are 12 = 0.4 ohm, 72,^400 ohms, r = 0.03 ohm, L = 30X10~fl h, u 7,,2r600
ohms.
Thus with sufficient accuracy the equation may be reduced to
and putting
g=(2r + R) _R,_
R i T to27i?
and
5 sm *[ R +(R + 2r^Ri + J2 J’
Substituting the value of Ex given in eq 4 and neglecting the products of the sines
of the small angles 5, 0C, and 0b,
Ec = ncIR[l — g+j (sin 5 +sin 0C)1 (5)
and
Eb = nbIR[l — g+j (sin 5 +sin /36)]- (6)
For the connections of the quadrant electrometer shown in figure 9, if Vn is the
voltage from the case to needle, Vc from case to quadrant c, and Vb from case to
quadrant b, the general equation for the torque is
T=\V'n+peV'c + pbVl+yeVnVe+ybVnVb+r1VeVb+vnVn-\- VcVcA VbVb-)- l>0,
404/578
Park,
Ltwlt ] Standard Electrodynamic Wattmeter
where X, pc> Pb»tc, 7i» v> vn» vc vb, and vb are constants of the electrometer. Take
the mechanical zero to be the reading when c, b, and n are all tied to the case, that
is, with current circuit open and »Sj to the right. This makes vo = 0. For a-c
work vn, ve, and vb can also be ignored, because the mean values of Vn, Ffl, and
Vb over any number of complete cycles is zero. The electrical zero is adjusted
by tilting or rotating the needle until no deflection occurs when <Sj is thrown to
the left. This makes X=0. With both current and voltage on, the torque then
becomes
T^pcVl+phVl+y'VnV'+yhVWb+vV'Vb.
The phase angle, 0, between the current and the voltage is adjusted until the
average torque becomes zero, that is, until the electrometer reading is the same
as its electrical zero, then
0=pcnlR2I2+pbn\R2I2 + ycnc EIR[cos (0—5 — pe) ]
— ybnbEIR [cos {Q—b— ^b)] — rrncnbR2I2 (7)
With switch S2 reversed, that is Ee and Eb reversed, 0 is again adjusted until the
average torque becomes zero. Then
0=pcn2cR2I2+pbnlR2I2 — yencEIR [cos (©' —5— /3C)]
-\-ybnbE!R [cos (©' — 5 — /3fc) — jjncni>JK2/2. (8)
Subtract eq 8 from eq 7
0 = ycncEIR [cos (0 — 5 — /3C) + cos (0'—5 — /3C)]
— ybnbEIR[cos (0—5 —/St,) + cos (0'—5— /3b)].
Expanding the cosines in this equation and neglecting the departure from unity
of cos (6+flc), cos (5-r/3b), sin 0, and sin ©' since 5 + /3« or S + /3b <10 minutes
and 0 or 0' is within 10 minutes of v/2
ycnc [cos 0 + sin (5 + (3c) + cos 0' + sin (S+&)]
— ybnb [cos 0 + sin (5+ £(,) +cos 0' + sin (5 + /3fc)]
Since y0 very nearly equals —yb and n„ nearly equals nb, —ycne (l + /i) may be
substituted for ybnb; then
cos 0 + cos 0(1 + /i) + cos 0' + cos 0'(l + /i) = —2 sin (5+(3c) — 2 [sin (5 + (36)l(l + h),
and this reduces to
cos 0 + cos 0'=—[sin (5 + (3c)]^l——[sin (5+ 1 + -^)’
Since f3e nearly equals pb the h terms cancel, and dividing by 2 to get the average
cos 0
cos 0 + cos O'
—sin (25 +/3C +/36). (9)
2
This equation yields the average value of the cosine of the angle between current
I and voltage E, for the two positions of switch S2, in terms of the constants of
the apparatus in the electrometer circuit. The cosine of this angle can also be
obtained from the expression
W— El cos 0,
where W is the average power indicated by the wattmeter for the two positions
of switch S2. Any difference between the two values of cos O thus obtained is
considered to be caused by the wattmeter phase defect, tw.
405/579
Journal of Research of"the National Bureau of Standards Vol. 53, No. 2, August 1954 Research Paper 2521
406/95
The two original portable standard, meters were
individually housed in separate temperature-con¬
trolled enclosures, each enclosure consisting essen¬
tially of a Bakelite housing completely surrounding
the aluminum ease of the meter, a mercury thermo-
regulator to control the temperature, and four small
lamps distributed inside the aluminum case to serve
as heaters. The temperature is regulated at approxi¬
mately 35° C to better than ±0.2 cleg C. The upper
end of the meter shaft projects through the top of
the Bakelite enclosure. The mirror for the light -
beam pointer is attached to this end of the shaft at
an angle of 45° to the axis of the shaft. Each meter
enclosure is assembled on a brass base plate to make
a complete unit for interchangeable mounting in the
housing that holds the phosphorescent scale. The
particular portable watthour meter that is used with
the circular scale is referred to in this discussion as
the Standard Watthour Meter, as distinguished from
the Comparison Standard Watthour Meters.
The cast-aluminum base of the circular-scale Fini’RE L Standard Watthour Meter with circular phosphor-
housing is equipped with leveling screws. The cir¬ escenl scale used as a reference standard for the testing of
watthour meters by a comparison method.
cular scale, approximately 26 inches in diameter, is
assembled on brackets attached to the base of the The light beam “pointer” is produced by a mercury vapor lamp and lamp
combination located above the Standard Watthour Meter.
housing. The meter enclosure rests on screws for a, Mercury vapor lamp housing; b, lens box; c, mirror; d, Standard Watthour
leveling the meter and adjusting its height for proper Meter; e, phosphorescent scale; f, engraved plastic scale; g, lamps.
407/96
the power circuit. Just before the end of the .preset
time interval (or preset number of revolutions of a
meter) the pulse counter again completes the circuit,
so that the last or stopping pulse fires the thyratron.
Thus the flashing is entirely automatic; the timing of
the flash depends on the signal pulse, which in the
case of the seconds signals is accurate to better than
a part in a million.
The third function of the accessory electronic
equipment is that of counting the disk revolutions
of the Standard Watthour Meter. This, too, .is
automatic. During a run the constant light actuates
the- phototube behind the slit in the scale. This
operates an electromagnetic counter by means of an
amplifier and a small thyratron. An electronic
switch opens the cathode circuit of the thyratron
tube to prevent counting except during the run.
The pulse that flashes the mercury lamp is also fed
to the grids of the electronic-switch tubes. The first,
or starting, pidse that flashes the lamp also enables
conduction in the counter thyratron and allows the leads. It is also important that the voltage
counting to start. The second, or stopping, pulse supplies be connected to these same terminals,
that flashes the lamp again at the end of the preset otherwise a phase-angle error in the watthour-mctcr-
timing or counting interval, disables the thyratron potential circuit might result from an IR drop in
and stops the counting. the leads. Lead resistances arc chocked and adjusted,
if necessary, just before each calibration. Xote also
3. Testing in figure 2 that the moving-coil side of the Standard
Wattmeter potential circuit is grounded, and that
The accuracy of a test of a watthour meter, using the current circuit is connected to ground through a
the apparatus described is directly dependent upon rectifier-type voltmeter, a 300-v instrument having
the latest previous calibration of the- reference a total resistance of 300,000 ohms. This serves as
standard, the stability of which is substantiated by an electrostatic tie to ground, which minimizes
more recent intercomparison tests. Phantom loading electrostatic forces between fixed and moving coils
is employed in all tests. Correct circuit connections of the wattmeter, and provides an indication of the
are made rapidly by means of links on an otherwise presence of leakage current and its magnitude.
permanently wired circuit board. In calibrating the Standard Watthour Meter the
procedure is as follows. First, temperature equi¬
3.1. Calibration of the Standard Watthour Meter librium of the entire equipment is established by
applying a-c power adjusted (in the case of the
The word “calibration” usually implies not only wattmeter and wattliour-meter circuits) to approxi¬
a carefully made test, but also an initial careful mately the values required for the test and then
adjustment or marking of the apparatus under test. waiting for an hour or two before proceeding.
For the purpose of this discussion, however, a After temperature equilibrium has been attained,
calibration of the Standard Watthour Meter implies a the Standard Wattmeter is calibrated on reversed
test to determine accurately its percentage registra¬ ylirect current to indicate the desired value of power.
tion and possibly, but not necessarily, an adjustment Alternating-current power is restored to the Standard
of its mechanism in order to bring the percentage Wattmeter and Standard Watthour Meter circuits
registration within certain limits. and adjusted to give the same deflection of the
The Standard Watthour Meter is periodically Standard Wattmeter as that obtained on direct
calibrated with 120 v, 5 amp applied, at unity power current. The a-c power is held substantially
factor, and at 0.5 power factor with the current constant throughout the run.
lagging the voltage. During calibration, a measured The pulse-counting circuit is then energized. The
value of a-c power is maintained constant for an next seconds signal flashes the mercury-vapor lamp
accurately determined time interval. and starts the run. The position of the spot on the
The testing circuit is shown in figure 2. In the phosphorescent scale is recorded to the nearest divi¬
potential circuit it is important that the effective sion. At the end of the preset interval (usually
terminals of the Standard Wattmeter be located 100 seconds) the final seconds pulse again flashes the
precisely at the Standard Watthour Meter terminals mercury-vapor lamp, thereby ending the run. The
on alternating current, and at the volt-box terminals position of this last spot is recorded as well as the
on direct current. Hence the lead resistance from reading of the revolution counter, which indicates
the a-c—d-c- switch to the Standard Watthour the number of complete revolutions. Thus, the
Meter potential terminals and to the volt-box number of revolutions of the Standard Watthour
terminals must be equal to prevent errors that would Meter’s disk equals the reading of the revolution
otherwise be caused by a difference in IR drops in counter plus the scale reading at the end of the run
408/97
minus the scale reading at the beginning of the run. A complete derivation of this equation appears in
For example, if the reading at the start of the run the appendix.
were 872 divisions, and at the end of the run, 493
divisions, and the revolution counter read 10, the 3.2. Tests of the Comparison Standard Watthour
number of revolutions would be 10 + 0.493 — 0.872 = Meters
9.621. After the a-c run, two additional readings of
the Standard Wattmeter on reversed d-c power are The watthour meters used as Comparison Stand¬
obtained in the same manner as at the start, giving ards are tested with 120 v and 5 amp applied, at
a total of four d-c readings. Care is taken to insure unity power factor, and at 0.5 power factor, with
a minimum of delay between the d-c readings and the current lagging the voltage. The method of
the run with alternating current. If this is done, testing a Comparison Standard Watthour Meter
any small linear drift that may be present in the involves a comparison with the Standard Watthour
deflection of the Standard Wattmeter causes no Meter immediately following or during its calibration
significant error in the result. The d-c readings are and periodically between calibrations. Figure 3 is
averaged and a correction applied for the difference a diagram of the circuit involved in the intercom¬
between the a-c setting and the average d-c reading. parison of the standard watthour meters.
Corrections are also applied for known errors in the The similarity between the circuit shown in figure
resistance standard, volt box, wattmeter (phase- 3 and the alternating-current portion of figure 2
defect angle), and for deviations from normal of the should be noticed. When a calibration of the Stand¬
average frequency of the source, if significant. ard Watthour Meter is in progress the Comparison
The formula used in this calibration for computing Standard Watthour Meters shown in figure 3 are
percentage registration is incorporated in the circuit of figure 2, although in
the interest of simplicity, this modification is not
'KRsn?A)()0 | Ddc Dac
^=100
. EpNvbEct + Ddc
cs Cpb~\~Cf Tw tan 0
(1)
] shown in the figure. With the circuit revised in
this manner, the Standard Watthour Meter, the
Comparison Standard Watthour Meters, and the
where Standard Wattmeter are energized simultaneously,
ix = disk constant of the watthour meter, in and the calibration runs on the reference standard
watt-hours per revolution. can be performed alternately with intercomparison
7?s = nominal resistance of resistance standard, test runs, thereby insuring nearly identical conditions
in ohms. for both tests. Truly simultaneous testing of the
ri=number of disk revolutions of watthour reference standard (having the circular phosphores¬
meter. cent scale) and the comparison standards is possible
Ep=potentiometer setting, in volts, when ad¬ with the addition of more equipment, but it is be¬
justing the d-c voltage. lieved that any improvement that might accrue
Nvb=nominal volt box ratio. from simultaneous testing would be too small to
Ee = potentiometer setting, in volts, when ad¬ detect with certainty.
justing the d-c current. There is a phototube-lamp combination as¬
/—time of run in seconds. sociated with each Comparison Standard, but only
Ddc — average of the four deflections of the one of these combinations is energized at a time.
Standard Wattmeter, in centimeters, ob¬ The phototube produces a pulse for each revolution
served during the d-c calibrations before of the disk of the Comparison Standard under test.
and after each run. This pulse is amplified and used in conjunction with
Dac — average deflection of Standard Wattmeter, the pulse-counting circuit to start and stop the run.
in centimeters, maintained during the To start the run, a pulse from the phototube
a-c run. associated with the Comparison Standard Watthour
c,,=correction to the resistance standard, in
parts per unit.
cvb = correction to the volt box, in parts per unit.
C/=frequency correction to the Standard
Watthour Meter, determined from the
reading of the synchronous timer, the
tune duration of the run, and a frequency
correction factor obtained in a separate
test of the Standard Watthour Meter.
Tw= phase defect angle of the Standard Watt¬
meter, in radians, positive if the moving
coil current leads the supply voltage.
0 = power factor angle, which is the angular
phase difference between the voltage, E,
applied to the Standard Wattmeter, and
the current, I, in the Standard Watt¬
meter current coils, and is positive if the
.current lags the voltage. Figure 3. Intercomparison of Standard Watthour Meters.
able an observer to set the secondary voltage, cur¬ in radians, considered positive when the
rent, and power factor. reversed secondary-voltage vector leads
A phototube-lamp combination is arranged to pro¬ the primary-voltage vector.
duce a pulse at each revolution of the disk of the 0S = power-factor angle at the Standard Watt¬
watthour meter under test if this is conveniently hour Meter, considered positive when
possible. The pulses are amplified, and, in conjunc¬ the current vector lags the voltage
tion with the pulse counter, are used to determine the vector.
duration of the run. A complete derivation of this equation appears in
An alternative method of testing meters, particu¬ the appendix.
410/99
tematic in nature and have been minimized in the
present work by carefully calibrating each piece of
apparatus for which a correction must be applied in
terms of the National Electrical Standards, which
are readily available. The largest systematic errors
of this type result from the uncertainty in the ratio
and phase-angle corrections of the current and poten¬
tial transformers. Other residual errors of this type,
such as those associated with the standard cell,
potentiometer, and volt box, occur in the primary
calibrations of the Standard Watthour Meter, and
are rather small. The history of the components used
in the measurements -circuit is well known. The
maximum net change in value of the components
between their periodic tests, for the last several years,
is listed in table 1.
The third type of systematic error is, in a sense,
similar to the second type, in that it arises from the
Figure 4. Testing watthour meters, using the standard imperfect application of corrections, but in this case
Watthour Meter.
the corrections are for the compensation of fluctu¬
ating errors. Such errors may vary with certain
4. Calibration and Testing Schedule parameters in a definite though unknown maimer.
On the other hand, the fact that an error exists and
The periodic primary calibrations of the Standard that it is a function of a certain parameter may be
Watthour Meter serve to accurately ascertain its known, but its effect may be negligibly small (and a
percentage registration. Primary calibrations per¬
formed at monthly intervals for a period of time have Table 1. Observed maximum net change in the value of the
indicated a long-time stability of the reference stand¬ comyo'nents necessary for the accurate standardization of the
ard that will permit an increase in the interval be¬ Standard Electrodynamic Wattmeter
tween calibrations to 6 months or longer. A graphical
record is kept of the intercomparison tests, which are Maximum net
Component change in value
performed at more frequent intervals, to detect any
appreciable drifts or erratic changes in registration of
ppm
any one of the group relative to the others, and pos¬ Standard cell.. . _ ... . .. 10
sibly, to form a basis for a revision of the registration Potentiometer_ .. . . _ _ 10
Volt box... _ .... 11
assigned to the Standard Watthour Meter if an ob¬
Total. . . . 31
served change is not large enough to warrant a pri¬
mary calibration. When a revision of the registration Standard cell. __. . 10
Potentiometer ... .. _ . _ 10
of reference standard is warranted on the basis of Resistance standard. . __ _ ... 6
intercomparison tests and the evidence does not indi¬ Total__... . . .... 26
cate gross instability in any one particular meter, the Grand total . . _ . .. .. __ _ 57
computation of registration is performed on the
assumption that the mean registration of all meters
remains unchanged between calibrations. With this correction impractical to apply) compared with that
assumption, the entire group of meters is useful in of the other errors. Varying systematic errors are
maintaining the unit of energy, although transfer of evident in the Standard Watthour Meter as well as
the unit to a watthour meter under test is accom¬ the Comparison Standard Watthour Meters and, in
plished through only one member of the group. fact, sharply limit the accuracy of which any watt¬
hour meter is capable. Reference is here made to
5. Precision and Accuracy Attainable the effect on the percentage registration of a watt¬
hour meter of such factors as incorrect magnitude of
Both systematic and random errors limit the ac¬ applied voltage and current, incorrect phase angle,
curacy of test results obtained with the Standard incorrect frequency, or the presence of harmonics in
Watthour Meter and its associated apparatus. Sys¬ the applied voltage or current. Studies have been
tematic errors, those which cannot be considered made of the effect of each of these variables on the
accidental, can be further classified into several types. meters used as standards. It has been found feasible
First, there are the errors that have escaped the atten¬ to use a good quality voltmeter, ammeter, and watt¬
tion of the observers. After a careful analysis of all meter, to facilitate setting the voltage, current,
possible sources of error and the comparison of the and phase angle. The power supplies are electron¬
results of tests made by several methods, it is believed ical^ regulated to maintain the initial adjustment of
that unknown systematic errors of magnitudes larger voltage and current, and are operated and loaded
than about 0.01 percent are highly improbable. in such a maimer as to minimize harmonics.
Second, there are residual errors from the applica¬ A third harmonic of as little as 1 percent of the
tion of imperfect corrections. These errors are sys¬ fundamental in the current circuit of a watthour
411/100
meter may cause a registration error of several only 0.001 percent if the watthour meter is operated
hundredths of a percent if certain phase relation¬ for 100 revolutions, as is usually done.
ships exist. This error arises from the existence of a Some of the residual errors thus far discussed will
third harmonic flux in the air gap of the electromagnet be positive, others will be negative, and the magni¬
even if a purely sinusoidal voltage is applied to the tude of many of these systematic errors will appear to
potential coil. It is important to realize that even vary in a random manner as more and more tests are
if a watthour meter is tested using alternating current performed. To this must be added the truly random
of practically sine wave form it will register inac¬ errors, the distribution of which is entirely by chance,
curately if used in a circuit in which the wave form is and to which the theory of errors may be applied.
nonsinusoidal. Although it would be unfair to The values listed in table 2 include an estimate of the
impute errors due to nonsinusoidal wave form to standard deviation of random errors based upon
watthour meters, which are designed for use on a numerous test data, and half the maximum syste¬
sinusoidal waveform, the possibility of an appre¬ matic error that it is believed may possibly be present
ciable error arising from this cause should not be in .any one measurement of energy. It is recognized
overlooked if accurate measurements are desired. that the errors listed do not represent truly random
Odd harmonics which are present in the wave form variations, yet their individual contribution (some
of the current and voltage supplied to the Standard being positive and some negative) to the final result
Watthour Meter and associated apparatus at the partakes of randomness and probably justifies the
time of this writing are, under most load conditions, computation of the total propagated error from the
considerably less than 1 percent, much less than square root of the sum of the squares of the com¬
usually found in commercial power sources. ponent errors. There is one exception. The error
During calibrations of the Standard Watthour in the standard cell and the error in balancing the
Meter, average frequency of the applied power during potentiometer against the electromotive force of the
a run is measured and a correction applied for any standard cell occur twice in a single measurement of
deviation from 60 c/s. In most other tests a meas¬ d-c power; hence, these component errors are not
urement of average frequency, and a correction for independent, and the propagation of these errors is
errors therein, is not conveniently possible, and this by simple addition. The potentiometer error also
probably contributes the greatest share of systematic occurs twice, but these are independent, since differ¬
errors that vary in a definite, but unknown, manner. ent resistance sections of the potentiometer are used
However, it has been determined that errors arising for the measurement of voltage and current. The
from customarily observed deviations in frequency total propagated error provides an estimate of ex¬
are of the order of a few thousandths of a percent. pected accuracy with which the applied energy is
The effect of temperature changes on the watthour known, when the Standard Watthour Meter and its
meter standards is reduced to a small value by associated apparatus is used to test other meters. It
keeping all meters in temperature-controlled en¬ is not to be inferred that the other meters are capable
closures. The standards are not removed from the of calibration or maintaining their calibration to the
position in which they are calibrated. Throughout degree of accuracy indicated at the bottom of the table.
all tests the Standard Watthour Meter is operated A test was made to determine the repeatability of
continuously, thus eliminating those errors asso¬ results in the comparison testing of the Comparison
ciated with starting and stopping the disk. Standard Watthour Meters against the Standard
Since the scale of the Standard Watthour Meter is Watthour Meter. Individual runs of about 100
marked with 1,000 small divisions, any small varia¬ seconds duration were made about 10 minutes apart
tions in the angular speed of the disk which would be until a group of 10 runs had been made. This was
undetectable in other meters may very likely be repeated the next day, and again 4 days later. The
easily discernible. To determine the nature and the relative percentage registrations of the two Compari¬
magnitude of errors due to angular speed variations, son Standards are shown in table 3. The values
the standard was centered as accurately as feasible listed are the means of the groups of 10 runs, and the
in its scale and carefully levelled. It was operated measure of precision is the standard deviation of a
at a reduced speed, and seconds pulses were used to single run of the group.
flash the mercury vapor lamp once every second. The precision indices listed in table 3 result from
The applied power was held constant at a value suffi¬ a combination of random errors inherent to (1) the
cient to cause successive spots from the mercury- Comparison Standard Watthour Meter being tested,
vapor lamp to appear on the phosphorescent scale at and (2) the Standard Watthour Meter, including ob¬
four points as the disk revolved. Scale readings servational errors in reading its scale. If it is as¬
corresponding to spot positions indicated a regular sumed that each meter contributes equally to the
(nearly sinusoidal) variation of disk speed during total error, but that the contributions are random,
each revolution. The variations were such as to then an index of precision for each meter may be
cause a maximum shift in scale reading of about 0.005 computed. If the total error is considered as 0.012
revolution. A method was devised for aligning the percent, the individual errors which produce this
watthour meter in its scale to reduce this error to less error are each 0.012/^/2=0.008 percent. From these
than 0.001 revolution, thereby partially compen¬ data it is estimated that the precision of each of the
sating for the angular speed variations. An error of watthour meters in the setup over a period of a few
0.001 revolution would cause a scale reading error of hours is about 0.008 percent.
412/101
Table 2. Effect of residual errors on the test of a watthour meter
Estimated resi¬
Type of measurement and components
dual error
413/102
varying systematic errors, discussed previously, 7. Appendix
which depend upon other parameters. A compari¬
son of tables 3 and 4 shows that the stability over The derivations of eq (1), (2), and (3) proceed
several months is very nearly the same magnitude most easily from the definition of percentage regis¬
as that over a period of an hour or two. This implies tration, which is the ratio of the indicated energy as
that any varying systematic errors that may exist obtained from the reading of the watthour meter to
vary rapidly and in such a manner that their effect the true energy, expressed in percent.
on the mean of a number of observations of per¬
centage registration taken several hours or more J=100 located energy
apart is negligibly small. Accordingly, when ■cali¬ true energy
(4)
brations of the Standard Watthour Meter are per¬
formed, the individual runs are taken at time inter¬ In a test of the Standard Watthour Meter, using the
vals sufficient to extend the duration of the test to Standard Wattmeter, the indicated energy in watt-
several hours. hours is Kn. The true energy is determined by the
When the apparatus is used for testing other watt- readings and settings of the Standard Wattmeter.
hour meters it is believed that the measurement of Because this instrument is very nearly astatic, its
the energy applied to the test watthour meter is sensitivity, s, will be the same for both directions of
valid and can be relied upon to better than 0.06 the direct current, and the average deflection is
percent.
Ddc — Slide, (5)
6. Conclusion
where Wdc is the true power impressed on the Stand¬
At the National Bureau of Standards a set of sev¬
ard Wattmeter.4 The true power in terms of the
eral carefully selected and prepared watthour meters
potentiometer settings and circuit constants is
has been arranged as a' standard group for testing
alternating-current watthour meters by a comparison
method. Wm=E’!%*E’+c (6)
Although the initial cost of designing, constructing,
and assembling this apparatus would ordinarily deter
the general adoption of such an elaborate system of where c represents the correction terms for the circuit
components. The average a-c deflection of the
testing watthour meters, the use of the apparatus
Standard Wattmeter is
described has yielded a definite saving in time and
labor, without significant loss of accuracy when com¬
pared with the primary method used in the past.
Dac=sEI cos(d+Tw), (7)
Accurate initial calibration of the apparatus is
where E is the a-c voltage between the Standard
accomplished by averaging the results of a number Wattmeter potential terminals causing a potential
of primary calibrations of the Standard Watthour
coil current, Ip, see figure 6. It is the magnetic field
Meter, using (1) a transfer wattmeter and potenti¬
from this current that reacts with the magnetic field
ometer for measurements in terms of the fundamental’ from the current, /, in the fixed coils of the Standard
electrical units of electromotive force and resistance, Wattmeter to produce the deflection, Dac.
and (2) standard seconds signals for time measure¬ The wattmeter sensitivity, s, will be the same on
ments. Assurance of the continued accuracy of the alternating current as on direct current provided all
reference standard watthour meter is obtained by readings are grouped closely on the Standard
intercornparison tests with other members of the
Wattmeter’s scale.
standard group. The reference standard, supple¬
Dividing (5) by (7)
mented by calibrated instrument transformers,
accurately measures the energy supplied to a watt¬ E)dc_ s]Vdc /p\
hour meter under test. Dac~sEI cos (9ETW)' K )
Solving for El
J. B. Dempsey initiated the early work in setting
up selected watthour meters for secondary standards.
Dn
EI=Wdc (9)
F. Silsbee proposed the light-beam pointer and Ddc cos (d+T„)
phosphorescent scale. F. J. Gross carried out the
developmental work and the original design of the The a-c power impressed on the Standard Watt¬
comparison apparatus. Most of the mechanical con¬ hour Meter is
struction of this apparatus was done by C. H. Wae=EI cos 0 (10)
Hochgesang of the Bureau’s Instrument Shop.
Robert W. Balcom assembled most of the electronic and substituting for El from eq (9)
equipment and provided many valuable suggestions.
Thomas W. Cushing carried the project forward for Dn, cos 6
a time and designed and supervised the construction Wni.=Wdl (ID
Ddc cos (d+Tw)
of the constant-temperature cabinet for the com¬
4 The symbols not specifically defined in the appendix have been previously
parison standards. defined in the text.
414/103
percentage registration of the Comparison Standard
Watthour Meter is
E
IC.c'tl'S
^c=ioo (15)
True energy
Ksns
(16)
True energy
100
W„=Wic [l +D“d®“+Tw tan el (14) EsIst cos ds (18)
or
When Wdc from eq (6) is substituted and the result 1 __j%s cos ds
is multiplied by f/3600 to obtain energy in watthours, (19)
EsIst 100 Ksiis
the true energy thus obtained may be used in eq (4),
and the results, showing all corrections as correction
terms, is eq (1) shown in section 3. and substituting eq (19) in eq (17)
The derivation of eq (2), which is used in inter-
comparison tests, follows logically from eq (4). Ktntf%s cos d$
(20)
The indicated energy equals Kcnc. Hence the 1 KsN'vtN'ctns cosd'
415/104
In this equation cos ds/cos 6 is a correction factor for 1
the phase angles of the current and voltage trans¬ (23)
1 — (J3—7) tan 6S
formers. To simplify this correction the denomina¬
tor may be expanded and when the indicated division is performed,
neglecting the extremely small second-order terms,
cos 6S _ cos d., the result is
cos (0s+/3—7) cos ds cos (|8—■7) —sin ds sin (fi—r)
(21) 1 + 03—7) tan es. (24)
Dividing both numerator and denominator by cos Thus the correction term for the phase angle in
6S the correction factor becomes parts per unit is c^ = (/3—7) tan 6S, and when the
correction terms for the current transformer ratio,
-^—7-(22) potential transformer ratio, and phase angles are
cos (|8—7)—sm (J3—7) tan ds shown, the final form of eq (20) is that of eq (3) in
part 3.3.
Since (J3—7) is a small angle, cos. (/3—7) is very
nearly 1 and sin (/3—7) is very nearly (/3—7), hence
the correction factor is Washington, March 5, 1954.
416/105
Preprint of paper to be delivered at the Eighth National Conference of the
Instrument Society of America, September 21-25, 1953, Hotel Morrison, Chicago, Ill.
Discussion sent to the ISA will he considered if received by October 1, 1953.
Paper No. 53-2-3
By FRANCIS L. HERMACH*
Abstract. This paper describes the basic These standard cells may be compared deflection potentiometer, and the newer
methods and procedures for testing elec¬ with others by means of potentiometers, forms of the portable potentiometer.
trical Instruments at power and audio fre¬
quencies, discusses the standards used and and the unit of voltage may thus be The general-purpose potentiometer is
their tests, and the power sources and con¬ disseminated. Similarly, by means of more accurate than required for instru¬
trol consoles. It also outlines the facilities bridges, resistors may be intercompared, ment testing and is somewhat slow and
and testing services of the National Bureau potentiometers, and indeed bridges
of Standards to science and industry in
cumbersome to use, but it is very stable,
this field, and describes some new develop¬ themselves, standardized. The potenti¬ flexible and reasonable in cost. Although
ments which may be useful in instument ometer, which serves the same purpose it is less flexible and stable, the deflec¬
testing. for voltage comparisons as a chemist’s tion potentiometer overcomes some of
balance serves for mass comparisons, these disadvantages and is well-suited
thus becomes the basic instrument for for instrument testing. However, it re¬
417/1
desirable to specify two or three extra switching only the standard to direct well as unsuspected changes due to
scale divisions at the top of the scale current, adjusting the current or volt¬ accidental misuse. Such tests are of two
of each standard instrument; i.e., a 153- age for the same deflection of the stand¬ types: frequent single-point checks and
volt instead of a 150-volt instrument. ard and then measuring it with a d-c intercomparisons to guard against
potentiometer in the usual way/1 The major change, and less-frequent, formal
direct current should then be reversed, tests which may in many cases have to
TESTS OF A-C reset and remeasured. The true value be performed in another laboratory. The
INSTRUMENTS of alternating voltage or current is frequency with which such tests should
Since the basic electrical standards then simply Qac = Qjr(l+S„), where be made is difficult to specify,*2 since it
can be used only for direct-current meas¬ Qdc is the average value of the meas¬ depends on the type of instruments, the
urements, all a-c measurements of volt¬ ured d-c voltage or current and So the accuracy sought, and the skill and care
age, current, and power depend funda¬ proportional ac-dc difference of the of the users." A fundamental rule, which
mentally on certain types of instruments standard at the frequency used. should not be violated, is that instru¬
which ideally have the same response on ments which are used as standards
Reversed D-C Tests
alternating and direct current. Such should not be used for general measure¬
D-c tests of a-c instruments may be ments or any other purpose.
instruments can be calibrated on direct
made with the potentiometer as already
current and then used for alternating- Standard cells are usually purchased
described, using the circuits of Fig. I
current measurements. There are three with a factory or NBS certificate stating
with the addition of a reversing switch
basic types of such instruments: elec¬ their emf. They are readily susceptible
in the input leads to the potentiometer.
trodynamic instruments, which depend to abuse, so that 2 or preferably 3, cells
on the forces between current-carrying Two readings should be taken at each
should be purchased. The working cell
conductors; electrostatic instruments, scale point to be tested with the supply
should be compared with a reserve cell
voltage reversed between readings, and
which depend on the force between weekly by connecting them in opposition
charged conductors; and electrothermic the average of the potentiometer read¬
and to the low range of the potentiom¬
instruments, which depend on the heat¬ ings should be multiplied by the volt-box
eter to measure their difference voltage,
ing of a current-carrying conductor. ratio or shunt resistance to obtain the
which should agree with the calculated
true voltage or current. Such a reversed
There are three distinct types of tests value. One cell should be sent to a stand¬
d-c test greatly reduces effects, such as
of a-c instruments; an a-c test, a re¬ ardizing laboratory yearly.
residual magnetism in the shield of the
versed d-c test, and a transfer test. An Shunts or resistors for current meas¬
test instrument, which depend on the
a-c test is a direct comparison of the urements are much less liable to abuse.
direction of the current. Wattmeters
instrument with another a-c instrument They can readily be intercompared in
are tested on reversed direct current by
of known accuracy. Alternatively, in¬ the laboratory with the potentiometer
the circuit of Id. Here, the separate
struments of the types mentioned can by connecting two, of adjacent ranges,
sources for the voltage and current cir¬
be tested on reversed direct current and in series and measuring their voltage
cuits are independently adjustable and
then given a transfer test to determine drops in succession at a convenient
need supply only the power taken by
their ac-dc difference. Since this dif¬ steady current. The potentiometer can
the instruments, not the fictitious power
ference is relatively permanent and gen¬ be checked at one point by balancing
indicated by the wattmeter, if the
erally small, a transfer test need ordi¬ its standard cell circuit in the usual
sources are stable, one potentiometer
narily be made only once for each in¬ way and then connecting the cell, with¬
may be used, as shown, to measure volt¬
strument. Thus, this latter proceduie out moving it,*3 to the X terminals of
age and current in succession. The indi¬
not only separates the errors but also the potentiometer and measuring its
cated power is the product of the two,
makes it feasible to confine subsequent emf, which should agree with its certifi¬
and no correction for instrument losses
periodic tests to direct current. Moving cate value. Resistors and the potentiom¬
iron, rectifier and electronic instruments is necessary.
eter should be sent to a standardizing
must be tested on alternating current Transfer Tests laboratory perhaps every five years.
only In a transfer test to directly deter¬ High-grade d-c instruments and elec¬
A-C Tests mine the ac-dc difference, the test instru¬ trodynamic a-c instruments, which are
Comparison tests on alternating cur¬ ment and a standard instrument of used as standards, should be tested on
rent are similar to those already dis¬ known or negligible ac-dc difference at d-c with the potentiometer perhaps
cussed for direct current. Ordinarily, the frequencies to be tested are con¬ every six months. If a potentiometer is
electro-dynamic instruments of the 0.1 nected in series or parallel to sec the not available, a valuable partial check,
or 0.25 per cent accuracy classes are same quantity. They are then in rapid to verify that major changes have not
used as standards at power frequencies succession connected to (a) alternating, occurred, can be made by comparing the
and can now be compensated for fre¬ (b) direct, (c) reversed direct and (d) full-scale deflection of the highest range
quencies up to about 2000 c/s. Multi¬ alternating current, which is in each of each instrument with the resulting
range current transformers of high ac¬ case adjusted to give the same deflec¬ deflection of the lowest range of the next
curacy can be obtained with primary tion of the test instrument after which highest instrument (for example, the
ranges from 0.25 to 100 amperes, and the response of the standard is re¬ 5-ampere range of a 2.5/5-ampere am¬
special three-winding voltage tians- corded. From the difference between the meter with the 10-ampere range of a
formers with ranges from 1.5 to 300 average of the a-c readings, Da<) and of 10/20-ampere instrument). Additional
volts, so that one ammeter and one volt¬ the d-c readings, D,u, of the standard, checks can be made by comparing the
meter can suffice for most measurements the proportional or per-unit ac-dc dif¬ a-c instruments with the d-c instru¬
ference of the test instrument, S can be ments of corresponding range. Such d-c
at power frequencies.
computed by the formula: checks of thermocouple instruments
For tests at higher frequencies, multi¬
Dac-Dae should be made frequently. If possible,
range thermocouple instruments are
S =- + So, the ac-dc differences of all a-c standards
preferred. However, they have unusu¬
ally large temperature influences (fre¬ D„c should be determined initially at a suit-
quently 0.2 per cent per deg. C) and are where So is the known proportional ac-
not as stable as electrodynamic instru¬ dc difference of the standard. *2A suggested schedule lor testing stand¬
ards used primarily in a meter laboratory
ments. is given in reference 2. Much valuable de¬
If the ac-dc difference of the standard TESTS OF STANDARDS tailed information on setting up a stand¬
instrument is known and the self-heat¬ ardizing laboratory and on the potenti¬
ing error is negligible, an a-c test may It is obvious that the standards used ometer and oilier measuring apparatus is
given in this reference.
be made with greatly increased accuracy in any instrument testing laboratory
by observing the response of the stand¬ should themselves be tested periodically *sTho emf of a cell may be changed tem¬
ard under a low-power microscope, to guard against errors due to drifts as porarily if the cell is moved.
418/2
ably equipped standardizing laboratory.
A recent- commercial development is
a line of self-checking d-c and thermo¬
couple instruments in which a single
point check and adjustment of the basic
dc or ac milliammeter in the instrument
can'readily be made in terms of an in¬
ternal standard cell each time the in¬
strument is used.
Instrument transformers are so stable
that a test at a standardizing laboratory
every five years should be sufficient.
419/3
ods have been developed for using1 them
to measure current or voltage over wide
ranges at audio frequencies.12
All instruments sent in to NBS for
standardization are tested with a null
potentiometer and its accessories, and
a-c tests are made with these transfer
instruments, which have ac-dc difference
of 0.01 per cent or less. The NBS in¬
struments do not have marked scales in
the usual sense and must be calibrated
with the potentiometer for every a-c
measurement. However, they are pri¬ Fill. 6. Elementary Diagram of NBS 1/2%
Thermocouple Volt-Ammeter
marily designed and used for rapidly
and conveniently determining ac-dc dif¬
ferences of other instruments.
a conventional 0.5 per cent thermocouple
instrument which can manually be com¬
NEW-INSTRUMENT pensated for temperature and other
DEVELOPMENTS changes.
420/4
Testing Electrical Instruments
By F. D. WEAVER
National Bureau of Standards
Washington 25, I). C.
Reference Standards
421/1
ever, for measurements of the highest
accuracy, it may be desirable to have
these differences determined in a stand¬
ardizing laboratory such as the Na¬
tional Bureau of Standards. Initially,
the instrument should be tested at each
of the cardinal points on the scale for
one range and at two of the cardinal
points on each of the other ranges.
Thereafter, an annual test at five of
the cardinal points on the base range
and at the same two cardinal points on
each of the other ranges is sufficient.
Transformers, potentiometers, shunts,
volt boxes, and reference watthour me¬
ters should be sent at intervals to a
standardizing laboratory for highly ac¬
curate tests. Shunts and volt boxes
should be initially tested at 10 percent
and 100 percent of rated current or
Fig. 2. Arrangement for testing voltmeters on direct voltage and reversed direct voltage. voltage; infrequent tests at 10-percent
load are ordinarily sufficient thereafter.
with 2.5- and 5-ampere ranges—of the Reference For Watthour Meters Ratio factor and phase-angle correc¬
0.25-percent accuracy class, are sug¬ A rotating standard watthour meter tions should be determined for the
gested as reference instruments for having voltage ranges of 120 and 240 transformers. However, phase-angle
testing other a-c. voltmeters and am¬ volts and current ranges of 1, 5, 12.5, corrections are not necessary unless
meters. This type of instrument is pre¬ and 50 amperes serves as a useful the transformers are used as acces¬
ferred because its indication is the same standard for energy measurements and sories in power and energy measure¬
on direct current as on alternating cur¬ as a reference meter for testing other ments.
rent. Each may be tested (corrections watthour meters. Several meter manu¬
accurately determined) on reversed d-c. facturers make suitable standards. The Tests of D-c. Ammeters
with the potentiometer and the above- percentage registration errors for 0.5
The diagram of Fig. 1 shows an ar¬
mentioned accessories, and then used and 1.0 power factors at 10, 50, 100,
as a reference instrument for a-c. test¬ and 150 percent rated current on any rangement for obtaining both fine and
range should be less than 0.5 percent. coarse control for adjusting currents
ing. In the selection of reference instru¬
ments the important consideration is in testing d-c. ammeters. The ammeter
that they have the needed ranges and Testing Reference Standards to be tested is connected in series with
A recommended schedule for testing an appropriate section of the multi¬
accuracy. They are most likely to fulfill
these requirements if they conform to standards is given on page 295 of Elec¬ range current shunt. The controls are
the specifications outlined for their ac¬ trical Metermen’s Handbook, Sixth Edi¬ adjusted until the pointer of the am¬
curacy class in ASA C39.1—1961 (see tion, 1950 (see reference 6). The time meter under test is deflected to the de¬
intervals elapsing between standard¬ sired position. The observer then meas¬
reference 1). If corrections are known
izing tests depend on the type of instru¬ ures the JR drop across the shunt with
for each of the cardinal points on the
scale and the instrument is used on ment or equipment, the accuracy de¬ the potentiometer. The current neces¬
sired, the frequency of use, the care of sary to deflect the pointer to the indi¬
only the upper third of its scale, then
the instrument while in service, and cated position is determined by the
the accuracy obtainable from a 0.25-
the stability of components. Ordinarily, formula 1 = E/R, where E is the volt¬
percent instrument (see reference 8)
is limited mainly by the ability of the instruments and meters should be tested age at the X terminals of the potenti¬
at least once a year, and potentiometers, ometer and R is the actual shunt re¬
observer to read pointer indications.
shunts, volt boxes, and transformers at sistance.
A three-winding potential transform¬
er—with tapped windings for 1.6, 3, less frequent intervals. The a-c. refer¬ If possible, a section of the multi¬
7.6, 15, and 30 volts on one secondary, a ence instruments recommended here range current shunt that has a slightly
160-volt winding for the reference volt¬ may be tested on reversed d-c. as out¬ higher or the same current rating as
meter on the other secondary, and a lined later in this article. The a-c. and the ammeter range to be tested should
120-volt primary winding—is available d-c. difference of this type is ordinarily be chosen. For example, a 10- or 15-
for testing a-c. voltmeters for the lower rather small at power frequencies. How- ampere instrument will likely be tested
voltage ranges (see reference 6). This D-C Voltoge Supply D-C Current Supply
type is recommended because losses as
a result of current in the supply wind¬
ing, which affect the accuracy of the
usual potential transformer, have no
effect on the ratios between the two
secondary windings that are compared
in the normal use of this transformer.
A multi-range current transformer—
with primary ranges of 0.25, 0.5, 1,
1.5, 2, 3, 4, 5, 10, 15, 20, 25, 30, 40, 50,
75, and 100 amperes and a 5-ampere
secondary to which the reference am¬
meter is connected—is satisfactory for
testing a-c. ammeters. The errors for
these two standard transformers on
any range for the burden imposed by
the usual instrument should be less
than 0.1 percent in ratio and 5 minutes
in phase angle. It is possible that errors
will be so small that corrections need
not be applied. Fig. 3. Arrangement for testing wattmeters on reversed d.c.
422/2
thermocouple types on reversed direct
current. In such a test, a reading is
taken for current in one direction and
is averaged with a second reading for
current in the opposite direction, there¬
by minimizing the effects of residual
and stray magnetic fields. The direction
of current in the instrument is reversed
after each reading by means of a re¬
versing switch. In general, this method
which makes use of a potentiometer
gives more accurate results than a sim¬
ple a-c. test.
A phantom-loading method is em¬
ployed in testing wattmeters on re¬
versed direct current. In this method
separate sources of voltage serve to
energize the voltage and current cir¬
cuits of the instrument (see Fig. 3).
Since the two sources are required to
supply only the losses in the instrument
and test equipment, they need not be
large sources of power. The voltage
coil of the instrument and the volt box
in parallel are connected to the direct-
voltage terminals. The current coil and
an appropriate range of the multi-range
shunt in series are connected to the
direct-current terminals. The potenti¬
ometer in conjunction with the volt
box and the voltage controls are used
to set the voltage to the desired value.
Then the current is adjusted by means
of the current controls to give the de¬
sired pointer deflection of the watt¬
meter, and the same potentiometer is
switched over to measure the 1R drop
across the shunt. If the wattmeter has
a compensating coil, the switch on the
instrument should be in the uncom¬
pensated position when the instrument
is tested. Care must be taken to prevent
overloading either the voltage or cur¬
rent coil. The precaution should be
0-120 V 0-30 V 0- 150 V 0- 100 Amps 5 Amp' taken to connect the voltage and
Fig. 4. Arrangement for alternating-voltage and alternating-current testing.
current terminals on the instrument
marked ± to the respective ± terminals
of the voltage and current supply as
with a 15-ampere 0.01-ohm or 15-am¬ or one having a range of 150 volts in shown in the diagram. To avoid the
pere 0.1-ohm section. Some commer- parallel with the 150-volt range of the errors caused by electrostatic effects
cially-available shunts have a drop of volt box, so that at full-scale deflection resulting from differences in poten¬
0.15 volt and others of 1.5 volts between of the instrument pointer the potenti¬ tial of the two circuits of a wattmeter
the potential terminals when carrying ometer will read approximately 1.0 or without a common connection, the ±
rated current. This is one of the rea¬ 1.5 volts respectively. A correspond¬ voltage terminal should be connected
sons for selecting a two-range 0.161- ingly desirable selection should be made through a voltmeter to the ± current
volt and 1.61-volt potentiometer. for other instrument and volt-box terminal.
As variable resistance at contacts ranges. For each point tested the volt¬ Values of power for each test point
causes annoying fluctuations in low- age V indicated by the instrument point¬ are determined by means of the follow¬
voltage, low-resistance, or high-current er will be determined by the formula ing formula: W = /V EvEl/R, where
circuits, it is essential that all electrical V = NE, where N is the actual ratio N is the actual ratio of the volt box,
contacts be kept clean and be of the and E the voltage at the X terminals R is the corrected resistance of the
best grade of material. of the potentiometer. sh: it, Ev is the voltage at low-side
For voltage ranges less than 1.5 terminals of the volt box and ET is the
Test of D-c. Voltmeters
volts the instrument with its leads voltage at the shunt potential termi¬
A voltage ratio box (volt box) in and the X terminals of the potentiom¬ nals.
conjunction with a potentiometer is eter should be connected in parallel Tests of A-c. Ammeters
necessary for testing d-c. voltmeters across a low-voltage source. This source
having ranges higher than 1.5 volts. To make a test of an a-c ammeter,
may be the voltage drop across a re¬
A definite and known fraction of the the ammeter to be tested is connected
sistor in which the current is con¬
total voltage applied to the terminals to the terminals marked C T Primary
trolled by the direct-current controls
of the instrument is measured at the (see Fig. 4). It is then in series with
shown in Fig. 1. In this case V — E
low side of the volt box by means of the primary of the multi-range current
provided that no corrections need be
the potentiometer. transformer. A primary range should
applied.
A suggested direct-voltage arrange¬ be selected that has the same current
ment is shown in Fig. 2. Notice that Tests on Reversed Direct Current rating as the ammeter to be tested.
the high side of the volt box is con¬ The direct-current and direct-voltage The 5-ampere electrodynamic ammeter
nected in parallel with the instrument. circuits illustrated in Figs. 1, 2, and 3 mentioned previously is connected to
It is desirable to connect a voltmeter serve for testing ammeters, voltmeters, the C T Secondary terminals. The con¬
under test having a range of 100 volts and wattmeters of electrodynamic or trols are then adjusted so that the point-
423/3
er of the ammeter under test is de¬ known accuracy is connected to respond resistance in the wattmeter potential
flected to the desired scale mark. The to the same value of power as seen by circuit may be used as a power-factor
reference instrument is then read in the watthour meter under test. The test.7
divisions. The current necessary to de¬ registrations of the two are observed Once the phase that gives 0.5 power
flect the pointer of the test instrument for the same time interval. This is factor, current lagging the voltage,
to any test point is 7pri == /3ec xN X RF. generally a much simpler method than is determined, the external power leads
Corrections for the reference ammeter (1) since power does not have to be held to the setup should remain fixed. The
and for the current transformer, if constant and the time interval need not selected leads may then be wired di¬
significant, should be applied. /sec is be accurately known. Method (2) is rectly to the test circuit and 0.5 power
the corrected reading of the reference usually to be preferred because of its factor can be obtained by shifting the
ammeter, N is the marked ratio of the simplicity. Standard watthour meters switch illustrated in the upper left-
current transformer and RF is the of suitable accuracy and stability are hand corner of Fig. 4. An adjustment
ratio factor of the transformer. now commercially available. It is neces¬ of the resistance in the primary of the
sary, of course, to determine correc¬ supply transformer mentioned above
Tests of A-c Voltmeters
tions for the standard at suitable in¬ may serve to give more nearly a 0.5
The three-winding potential trans¬ tervals by either method (1) or (2) power factor.
former mentioned previously is con¬ or to have such corrections determined Either a housetype or a portable
venient for testing voltmeters of 30 at a standardizing laboratory. meter may be compared with the ro¬
volts or less. It also serves as a supply tating standard. The potential termi¬
If there are several rotating stand¬
transformer for voltages up to 300 volts nals of the two are connected in parallel
ards in one laboratory, or in one lo¬
when all windings are connected in to the AV terminals indicated in the
cality, it may be advisable to main¬
series (see reference 6). This is ac¬ diagram. The current coil of the meter
tain one or more of these as reference
complished between terminal AV 1 and under test is connected to the CT
meters to be used only in the testing
AV 6 as illustrated in Fig. 4 by clos¬ Primary terminals and the standard
laboratory. This reference meter could
ing the switch marked 0-300V SW. to the CT Secondary terminals. The
be sent to a standardizing laboratory
A test voltmeter having ranges of current transformer should be con¬
for accurate tests and then used only
30 volts or less is connected to termi¬ nected for the proper range. A volt¬
in the laboratory for comparison tests
nals marked AV 3 and 4, Fig. 4, with meter and ammeter are used with the
with other rotating standards. This
the range switch on the appropriate tap controls to set the voltage and current.
could be a cooperative project with
and the reference voltmeter is con¬ The two meters may be started and
several companies in the same area
nected to terminals marked AV 5 and stopped by use of the panel voltage
participating. Since stable and accurate
6. The supply to the 120-volt primary on-off switch indicated on Fig. 4. When
watthour meters are now available,
is controlled by the usual AV controls a correction is to be applied for the
they would in this way be assured of
on the panel. Voltmeters having ranges readings of the standard, the percent¬
energy measurements of high accuracy.
greater than 30 volts are tested by age registration of the test meter is
comparison with the reference volt¬ Meter testing in accordance with the determined as follows:
meter on the 75-, 150- or 300-volt range proposed wiring diagram follows the Percentage Registration =
using terminals AV 1 and 6. The volt¬ plan of method (2). The voltage circuit khX r X A/Kb X R
age for any given indication of the of the standard is energized for a where r equals the revolutions of the
instrument under test (ranges of 30 time corresponding to a desired num¬ test meter, R equals revolutions of the
volts and less) is computed by multi¬ ber of revolutions of the test meter standard, kh equals the watthour con¬
plying the corrected reading of the (house type) as counted by visual ob¬ stant of the test meter, Kh equals the
reference instrument by the corrected servation. Tests are made at unity watthour constant of the standard and
value of the transformer ratio for the power factor, and 0.5 power factor, A equals the percentage registration
selected range. current lagging the voltage, using of the standard.
If the instrument tested is of a phantom loading. A 3-phase 3- or 4-
If transformer corrections are to be
rectifier or electronic type, there is a wire source is convenient for testing applied, formulae may be found in
possibility of error unless the test volt¬ at the two power factors mentioned. Electrical Engineer's Handbook.
age has almost a pure sine-wave form. Unity power factor is obtained by con¬
The recommended reference instrument necting the voltage circuits of the two (1) "American Standard for Electrical In¬
responds to rms. values of voltage and meters to the same phase as that dicating Instruments, C39.1—1951." American
is not appreciably affected by wave¬ energizing the current coils as illus¬ Standards Association, New York 17, N. Y.
(2) H. B. Brooks "The Accuracy of Com¬
form distortion. The observer should trated in Fig. 4. A resistance load in
mercial Electrical Measurements," Transactions
make sure that the output wave form the primary of the supply transformer of the American Institute of Electrical Engi¬
of the setup has a total harmonic dis¬ of the current circuit will tend to elimi¬ neers. Vol. 39, February 1920, pages 495-015.
nate any phase difference between the (3) -I. B. Dowden. "Organizing an Electrical
tortion of less than 2 percent.
Instrument Standardizing Laboratory," Engi¬
voltage and current circuits at near
neering l\otes, (Weston Electrical Instrument
Watthour Meter Testing unity power factor. A power factor Corp., Newark, N. J.) Vol. 2, June 1947, pages
A watthour meter can be tested by of closely 0.5, current lagging the volt¬ 6-8; Vol. 4, February 1949, pages 3-5.
age, is obtained by connecting the volt¬ (4) Electrical Engineers Handbook. by Fen¬
either of two methods: (1) using a der and Del Mar, Part IV, Electrical Power,
wattmeter, and (2) using a watthour age coils to a selected other phase. John Wiley & Sons, Inc., New York, 1936,
meter. In the first method, a wattmeter pages 5-02 to 5-112.
A wattmeter may conveniently be (5) Electrical Metermen's Handbook, Edison
is connected to respond to the same
used with a voltmeter and ammeter Electnal Institute, 420 Lexington Avenue, New
value of power as seen by the watt¬ York, N. Y., sixth edition, 1950.
to check the power factor in watthour
hour meter under test. The power (6) R. Estoppey, "Potential and Current
meter tests. If the wattmeter has a
(average active power on a.c.) is held Transformers for Standardizing A-C Voltmeters
double voltage range, 75 and 150 volts, and Ammeters,” Engineering Notes, (Weston
at a constant value as indicated by the
a capacitor may be used to determine Electrical Instrument Corp., Newark, N. J.)
pointer of a wattmeter for an ac¬ Vol. 3, October 1948, Pages 7-8.
whether the load current leads or lags
curately measured time interval, after (7) Frank A. Laws, Electrical Measurements.
the voltage. To check the phase of the
which the registration of the meter is McGraw-Hill Book Co., Inc., New York, 1938,
current with respect to the voltage with pages 624-639.
observed. The percentage registration
the 150-volt range in the test circuit (8) Frank D. Weaver, "Notes on the Care
is computed as and Use of Electrical Instruments,” Instruments
of such a meter, a 1-microfarad capaci¬
registered energy (Instruments Publishing Co., Pittsburgh 12,
tor of good quality is connected between Pa.) Vol. 23, No. 12, December 1950, pages
100 X-, the 75- and 150-volt posts. If the 1236-1239.
measured energy
pointer of the wattmeter goes upscale (9) Frank D. Weaver, "An Easily Assembled
Console for the Rapid Testing of Electrical In¬
where the measured energy is the prod¬ the load current is leading, if the
dicating Instruments," Instruments (Instru¬
uct of the measured power and time. pointer goes downscale it is lagging. ments Publishing Co., Pittsburgh 12, Pa.) Vol.
In the second method, a meter of Alternatively, a small inductance of low 22, No. 5, May 1949, pages 396-399.
424/4
Scale and Reading Errors of
ELECTRICAL INDICATORS
Scale and reading errors limit the accuracy of even the best electrical
indicating instrument to about 0.1 percent. Sources of errors are
analyzed, and the NBS standard calibration methods are presented.
Reprinted from
November 1954, Instruments and Automation, Vol. 27, No. II
425/1
572772 0-61-28
Fig. 2. Enlarged portion of middle scale of
Fig. 1.
426/2
marks themselves and at the midpoint 0.5-percent-accuracy class, and an a-c. estimated 0.40), and for 0.55, 0.65.
of the division, (2) an individual will and d-c. 150-v. voltmeter in ^-percent- 0.75, 0.85, and 0.95 positions he most
make estimations of tenths of a divi¬ accuracy class were selected for the frequently chose the 0.1 above the
sion in much the same manner over test. The interval between the 80 and value set (for 0.85 he estimated 0.90).
long time intervals. (3) the minimum 81 division marks on the d-c. voltmeter In further tests, a flat-field x2 lens.
error occurred when the line width was and the 120 and 121 division marks on 1% inches in diameter, was used, and
equal to the pointer width, and each the a-c. and d-c. voltmeter w'as selected each observer was asked to estimate to
was roughly 0.1 that of a division for the experiment. Each instrument the nearest l/20th of the scale interval.
width, and (4) the average error of was carefully tested at its two chosen For most observers, this lens materially
estimation decreased with increasing reference division marks with a po¬ aided estimational accuracy.
division width. There was little gain tentiometer. The obtained values then The results indicate that an observer
in increasing the width beyond about were used as a basis for dividing the generally can estimate on a good scale
2 mm. intervening space into twenty equal in¬ to within 0.1 scale division. Out of 945
In the ordinary procedure of read¬ tervals. An observer at the potentiom¬ estimations only 25 were in error by
ing an instrument scale an observer eter adjusted the voltage to set the as much as 0.15 division, and 123
will estimate the pointer position to the pointer in random order to the various (about 13 percent) were in error by as
nearest 0.1 scale division. Backstronfs tenths or twentieths of a division. Most much as 0.1 division. The greatest un¬
study and our experience indicate that of the subjects had considerable pre¬ certainty of estimation appears to be
the accuracy with which one is able to vious experience leading instruments. in the area of 0.3 and 0.7 division.
estimate depends on a number of the As indicated in the table, each was
factors. It is assumed that the observer asked to estimate to the nearest tenth
lias normal vision, that the instrument (and in other tests to the nearest twen¬ Conclusion
has a good scale and pointer, and that tieth) of a division. The subject was
The data indicate that some instru¬
the scale is well illuminated. The in¬ not aware of the setting being made
ments in the better-accuracy class have
fluencing factors are then (1) width at the potentiometer nor was he told
inaccurately marked scales. If these in¬
of scale marks and width of pointer, how close his estimated values were to
struments are tested at five of the cardi¬
(2) distance between scale marks, (3) the true values until the test was
nal points, values obtained by applying
the particular tenth which is being es¬ finished.
corrections to the scale indications at
timated, (4) fatigue and physical con¬ In obtaining the data for the table,
intermediate points may be expected
dition of the observer, (5) color com¬ the pointer was set to each 0.05 posi¬ to be in error by several tenths of a
binations for scale, pointer and lines, tion and the observer was asked to division when the corrections are ob¬
and (6) personal bias. estimate to the nearest 0.1. For the tained by linear interpolation between
We tested the ability of our labora¬ 0.05, 0.15. 0.25, 0.35, and 0.45 posi¬
values reported at the cardinal points.
tory personnel to estimate pointer posi¬ tions he most frequently estimated the It is possible to draw two conclu¬
tions. A 100-v. d-c. voltmeter in the 0.1 below the value set (for 0.45 he
sions: (I) If the corrections at five of
Table.— Estimation of pointer position of a-c. and d-c. the cardinal points are plotted on a
voltmeter to nearest 0.1 division, graph, as in the figures, and the broken
line connecting all plotted points is
14 observers, no lens used.
abnormally irregular, it is unlikely thai
Pointer Position
Set in Divisions Estimations interpolation between two adjacent
cardinal points will give correct values.
-.25 -.20 -.15 -.10 -.05 Correct + .05 + .10 + .15 +.20 (2) If the spacings between adjacent
120.00 14 division marks when viewed with the
eye appear to be of unequal widths it
.05 9 5
is also unlikely that an interpolation
.10 4 10 value will be valid.
.15 1 6 7 If an instrument is tested at every
five divisions on its scale and the sepa¬
.20 6 8 ration of intermediate marks appear
.25 6 8 normal the interpolation procedure
should give values that are reasonably
.30 4 10
correct. Thereafter it should be neces¬
.35 2 5 7 sary to test it only occasionally at five
.40 2 11 1 of the cardinal points. If the scale does
not meet the necessary accuracy re¬
.45 1 8 5 quirements (adjacent intermediate di¬
.50 2 9 2 1 vision marks are irregularly spaced or
cardinal points are incorrectly placed)
.55 1 2 10 1
it may be advisable to return the in¬
.60 1 7 5 1 strument to the manufacturer for a
.65 1 1 10 2 new scale.
The procedure of estimating tenths
.70 1 1 7 5 of a division is not the weakest link
.75 1 2 10 1 in electrical measurements. However,
this factor combined with a large num¬
.80 1 6 7
ber of other possible sources of error
.85 3 9 o limit the accuracy of even the best in¬
.90 9 5 strument under ideal conditions to
about 0.1 percent. If greater accuracy
.95 2 12 is desired one should resort to poten-
121.00 14 tiometric measurements.
427/3
Notes on the Care and Use of
Electrical Instruments
By FRANK D. WEAVER, Electrical Instruments Section, National Bureau of Standards, Washington, D. C.
This article presents notes on pre¬ pressure contacting surfaces a high- age of the supply circuit. One multi¬
cautions that should he observed by grade light oil, such as Marcol, may range current transformer with one
users of electrical instruments to en¬ be used with beneficial results if the ammeter can be used to replace a num¬
able them to achieve greater accuracy contacts are cleaned periodically and ber of separate a-c. ammeters. The
in measurements and to extend the only a small amount of oil is applied. transformer should be tested to deter¬
useful life of their equipment. In case the contact pressures are so mine the ratio factor for its various
high or edges so sharp that undue wear¬ ranges with its ammeter as a burden
F I. Introduction
428/1
the supply voltage to a low value be¬ closed, the cold-state resistance may be Instruments that are being prepared
fore throwing the supply switch or low and the initial surge of current for shipment should first be wrapped
circuit breaker. Remove conductors or rather high. An ammeter placed in in heavy paper to exclude dust and
leads to equipment one at a time, mak¬ such a circuit to measure the steady- dirt, surrounded by a cushioning ma¬
ing the first break at the power source state current may be overloaded by terial such as several inches of excel¬
terminal and next remove the end at¬ these initial high currents and result sior, and carefully and securely packed
tached to the equipment. This latter in a bent pointer. It may be advisable in a strong, properly labelled box. The
precaution applies when unplugging to short out the ammeter during the top of a wooden packing box should
from a distribution board to a sub¬ period of high initial currents.- In a be fastened with screws (not nails) to
panel as well as when disconnecting an similar way the insulation of the coils avoid damage resulting from blows
instrument, such as a voltmeter, from of a voltmeter may be damaged by with a hammer or nail puller. It may
a line, or a standard cell from a poten¬ voltage surges or by distorted waves be advisable to invert the instrument
tiometer. The reverse steps should be that have a high peak value, say as in the shipping container so that the
taken when placing equipment in a cir¬ much as five times as great as the upper instead of the lower jewel and
cuit. Some careless observers, when rms. value to which the instrument pivot receive any damaging blows that
taking an instrument out of a circuit, responds. This latter situation is es¬ occur in shipment. If this is done, the
disconnect the leads at the instrument pecially dangerous for electrostatic box should carry a “This Side Up”
terminals, remove the instrument, leave voltmeters. A protective resistor should label. These precautions aid in pro¬
the free ends of the two leads dangling be placed in series with the high side tecting delicate lower jewels and pivots
while the other two ends are connected of an electrostatic voltmeter to pre¬ from injurious mechanical shocks dur¬
to a source of power. This is a danger¬ vent “burning up” the instrument ing transit.
ous procedure. Power circuits are or¬ spring in case of flashover between its Precautions should be taken to pre¬
dinarily protected with fuses or cir¬ plates. vent marring or scratching the finish
cuit breakers and switches, but these In the laboratory one frequently of an instrument lid or case. Such
protective devices do not guarantee uses a suspension-type d-c. galvanom¬ defacing may occur if an instrument
that instruments and other equipment eter in connection with bridge and po¬ is placed near a hot rheostat or an
will not be damaged if the circuit is tentiometer measurements. Ordinarily inverted lid is carelessly pulled along
shorted by the careless handling of these sensitive instruments are used to a rough surface. An occasional clean¬
leads or switches. determine voltage balances between ing of the surface of an instrument
Laboratory personnel should be par¬ the arms of a bridge or potentiometer case is advisable in order to restore
ticularly careful when working with and do not directly measure electrical good appearance, as the better the ap¬
equipment from which there is a chance magnitudes. Such an instrument should pearance the more respect it is likely
have the sensitivity necessary for the to receive.
of electrical shocks. Currents as small
as 10 or 20 milliamperes can be dan¬ precision desired, should have a short A relative humidity of less than 65
period, and should be slightly under¬ percent and a temperature of 25 C.
gerous to life, depending upon such
damped in its circuit. If its sensitivity are desirable for the air of a room
factors as the current path through
is higher or its period longer than nec¬ where instruments are to be used. If
the body and physical condition of the
essary, time will be wasted in getting the air is too humid, moisture may
victim. The value of current will de¬
balances. Since galvanometers are condense on insulating surfaces be¬
pend upon the resistance of the body made of delicate parts they should be tween resistors and act as a high-
and the applied voltage. Under cer¬ handled with great care and be pro¬ resistance shunt or in effect reduce
tain conditions one may receive a fatal tected in a circuit by suitable switches the total resistance of that part of
shock from equipment that operates and resistors. the circuit. This effect is ordinarily
from the usual 110-volt circuit. If the
Make it a practice to handle all in¬ more noticeable in high ohms-per-volt
hands or feet are damp when contact
struments with care; treat those used instruments. For extremes of humidity,
is made, the body resistance may be
as laboratory reference standards with moisture may be absorbed by or evap¬
low enough (11,000 ohms or less) to
special respect. Pivoted instruments orated from the insulating fibers sur¬
permit a dangerous or fatal current.
should never be placed or left where rounding fine resistance wire, causing
Such an electrical shock may be re¬
they may be exposed to vibrations. the threads to change dimensions, thus
ceived if one touches two separate
One severe shock, such as a hammer exerting pressures and stresses on the
metal components of a piece of equip¬
blow on a bench or table where instru¬ fine wire and altering the resistance.
ment, or one metal component and a
ments are resting, can permanently It is a good practice to compare work¬
grounded object so that a portion of
damage their pivots and jewels. ing instruments against reference in¬
the body acts as a parallel path to the
In making adjustments or repairs struments at least once a year. If high
electrical circuit. Exposed metal parts
on instruments, bear in mind that only precision of measurements is needed,
should be insulated from the wiring
the manufacturer is really well quali¬ the instruments should be checked
of a piece of electrical equipment
fied to make major repairs. If it is against reliable reference standards or
(motors, drills, soldering irons, etc.)
necessary to open the case, it should sent to a measurements laboratory
but sometimes, because of worn or de¬
be done in a room where there is a where precision tests may be made.
fective insulation, the conductor comes
minimum of dust and in an area where Such tests will reveal changes in the
into electrical contact with the exposed
there are no iron filings. accuracy of instruments.
metal. The cases of some equipment
When instruments are not in use, Most instrument companies occasion¬
are provided with ground terminals
they should be stored in cases or cabi¬ ally issue instruction manuals, bulletins
and if properly connected will protect
nets free from dust, acid fumes, ex¬ and catalogs for the benefit of their
the operator from the hazards of elec¬
cessive heat, moisture and vibration. customers. These ordinarily give valu¬
trical shock.
Shorting the terminals of a galvan¬ able information about their instru¬
B. The Instrument ometer or microammeter by means of ments, with diagrams showing how to
Multi-range instruments are desir¬ a small conductor before such instru¬ connect them in a circuit. It is well
able for some applications. Many such ments are moved heavily damps the to consult such manuals in addition to
instruments show up for repairs with motion of the moving parts and helps handbooks, textbooks and other useful
bent pointers and damaged coils be¬ to prevent damage. Locking devices sources of information.
cause users have not been careful to are sometimes provided for moving
set the range switch in the proper po¬ systems to protect pivots and jewels III. Sources of Error (Instruments
and Associated Apparatus)
sition before closing the circuit. When or suspensions from shock while the
some circuits (for example a bank of instrument is in transit; make proper A knowledge of the common sources
tungsten filament lamps) are initially use of them. of error in instruments and associated
429/2
apparatus may be helpful in eliminat¬ Poor electrical contacts at the one arising from rubbing the glass
ing certain errors of measurement and switches or in any part of a measure¬ with a cleaning cloth, may usually be
increasing the accuracy of results. ment circuit are a source of consider¬ eliminated by blowing one’s breath on
. Some of the conditions that may in¬ able annoyance to an observer. In in¬ the affected area. Trouble from the
terfere with the normal turning of the stances where this condition cannot accumulation of static charges will be
moving system of an instrument are: be alleviated ' by cleaning and oiling more frequent at low humidities be¬
a warped or loosened paper scale, thin the contacts, it may be necessary to cause of increased insulation resistance.
fibers projecting upward from a paper have the contacting surfaces plated Insofar as possible, one should avoid
scale, iron filings in the air gap, a with nickel, silver or gold. For rugged locating instruments in an area where
bent pointer, a bent damping vane, a service, where high currents and great¬ there is a strong magnetic field or
flat pivot, a broken jewel, or a moving er wear due to higher contact pressures adjacent to large masses of metal.
system too tight or too loose in its are involved, a silver inlay in the con¬ Strong fields may be encountered near
jewels. Each of these possible sources tacting areas has proved highly satis¬ current-carrying conductors, trans¬
of error should be corrected and the factory. formers, Variacs, and even other in¬
instrument retested where necessary An electrical instrument may give struments. Pointer indications of some
before use if highly accurate results inaccurate values because of errors unshielded portable d-c. instruments
are desired. Obstructions can usually in the original scale calibration. In may be in error by as much as 2 per¬
be detected by a “jumping of the point¬ other words, the cardinal points on the cent when the instrument is read while
er” as current through the instrument scale may have been improperly placed resting on a table having a sheet-iron
is slowly increased or decreased to de¬ by reason of uncorrected errors in the top. Errors resulting from stray fields,
flect the pointer .upscale or downscale. test equipment or in procedures used eddy currents and the shunting effect
Occasionally during this test, set the to locate lines on the scales. An in¬ of nearby sheets of iron, are generally
pointer to a scale mark and tap the spection of the scales of some instru¬ much less for shielded instruments
case lightly: pivot friction will be ments reveals a noticeable inequality than for unshielded ones.
revealed by a slight change in pointer in the spacing of adjacent subdivision A panelboard instrument should not
position. marks. An observer should not place be inserted in a steel panel unless it
much trust in the indications of an has been calibrated by the manufac¬
Tapping the case of an instrument
instrument that has an erratic and turer for use in such a panel.
while it is being read is usually un¬
carelessly-drawn scale.
necessary if the instrument is properly There is a tendency for the magnets
constructed and in good condition. How¬ To provide the greatest reading ac¬ of d-c. instruments to become weaker
ever, when slight friction is present curacy, a portable instrument should with time if they have not been prop¬
from any cause, its effect can usually have a good scale, a knife-edge pointer erly aged or conditioned, thereby weak¬
be temporarily minimized by gently and an anti-parallax mirror. The knife- ening the flux in the air gap. Such a
tapping the case. Hard tapping may edge width should be about equal to tendency is indicated when periodic
result in damaged pivots and jewels. that of the scale division marks, and tests show that an increasing value of
not greater than one-tenth of the dis¬ the measured magnitude is required
Errors resulting from friction and
tance between successive scale divi¬ to deflect the pointer to a given scale
from mechanical unbalance may be
sions. For such an instrument an ex¬ mark, other factors remaining con¬
expected if a portable instrument, de¬
perienced observer should be able to stant.
signed to be operated with the axis of
estimate the pointer indications to bet¬
the moving coil in a vertical position, In making measurements with am¬
ter than ±0.2 scale division. With a
is read while the instrument is so meters, voltmeters, and wattmeters, it
laboratory standard having the major
placed that this axis is inclined from is well to remember that there may be
scale divisions subdivided into fifths
the vertical or is in a horizontal posi¬ considerable power losses and voltage
by means of diagonal lines intersect¬
tion. In general it is good practice to drops in the instruments as well as in
ing six concentric arcs, one would be
use an instrument with its moving- their leads. These must be taken into
expected to estimate to about 0.05 of
coil axis only in the position for which consideration in precise measurements.
the major divisions. Instruments that
it was designed. Voltmeters requiring 0.1, 1, 10 and
do not have anti-parallax mirrors are
Parts of moving systems, including subject to rather large reading errors. 100 milliamperes for full-scale deflec¬
pointers of instruments, tend to vi-“ tion are in common use. Before a volt¬
When a wattmeter is initially in¬
brate at certain frequencies where they meter is placed in a circuit, one should
serted in a circuit for the measurement
are mechanically resonant. Ordinarily, consider the effect of the current losses
of a-c. power, it is a good practice to
at the frequency for which the instru¬ of the instrument upon the circuit.
put an ammeter in series with the cur¬
ment is designed, this should not be Voltmeters having current losses of
rent coil and a voltmeter in parallel
expected as the maker tries to avoid 0.1 milliampere are desirable for some
with the voltage coil. This precaution
this in his design and construction. applications; however, one that re¬
is advisable because it is possible for
This phenomenon is revealed by er¬ quires 10 milliamperes for full-scale
either or both coils to be damaged by
ratic pointer movements, pointer wid¬ deflection is likely to have more rugged
overload and yet the pointer may not
ening or blurring, and sometimes by a components and therefore is preferable
indicate a full-scale reading since a
rattling noise. Errors should be ex¬ where current loss is not an item of
wattmeter measures watts (VI cos 6).
pected if the instrument is read under importance.
To avoid the errors caused by electro¬
these conditions. It is possible to change static effects resulting from differences When carrying rated current, shunts
the resonance frequency by changing in potential of the two circuits of a and resistors in some instruments are
the balance weights and rebalancing wattmeter without a common connec¬ required to dissipate a considerable
the moving system. amount of heat. Unless the resistance
tion, the ± voltage terminal should be
Some instruments show inherently material was properly annealed by.the
connected through a voltmeter to the
large zero shifts: After the moving manufacturer and has a low tempera¬
± current terminal.
system is deflected up-scale for a few ture coefficient of resistivity, changes
Errors may result from (1) electro¬ in resistance with use may be expected.
minutes, the pointer does not return to
static forces, and (2) leakage currents, These changes, accelerated by heating
its initially set zero position when the
circuit is opened. The observed shift if the moving system of an instrument which occurs when the resistor carries
may generally be explained on the basis is at a different voltage from that of rated current, may produce significant
of inelastic yield in the springs. Such its case or panel. Correct electrostatic errors in instrument indication. Ade¬
shifts constitute a source of error for shielding will eliminate this source of quate provisions have not been made
which it is not practical to apply a error. Static charges on the glass win¬ in some instruments and equipment
correction. The only remedy if the dow or case of an instrument may also for dissipating their large heat losses,
shifts are large is to have the spring cause errors in pointer indications. and this may result in excessively high
replaced by the manufacturer. The effect of a static charge, such as temperature when in use.
430/3
Some types of instruments not prop¬ The unsaturated type of cells are gen¬ if the formula 1 + C = T (indicated
erly compensated for changes in tem¬ erally used in laboratories. Even a reading plus correction equals true
perature may be in error by as much small current drain for a short period value) is used in applying corrections.
as 0.4 percent per degree Centigrade. of time. (100 microamperes for 5 min¬ The true value (in divisions) may then
An instrument will normally be at utes or more) ma^ temporarily impair be converted to measured units.
least partly temperature compensated, the usefulness of such cells. If used Assuming that an observer can esti¬
so the coefficient is more likely to be in potentiometers where highly accur¬ mate a pointer deflection on a 100-di¬
0.1 percent or less per degree Centi¬ ate measurements are desired the cells vision scale to 0.2 division on any part
grade. Since most instruments are should ^>e tested periodically. It is of the scale, a reading error of 0.2
calibrated for use at 25 C., for precise sometimes advisable to enclose them in division at the 100-division point will
work it is best to allow them to reach thermally insulated boxes to eliminate be 0.2 percent of the measured magni¬
temperature equilibrium in a room errors that arise from sudden changes tude, whereas a reading error of 0.2
maintained at this temperature. This in temperature or differences in tem¬ division at the midpoint of the scale
is especially true for thermocouple in¬ perature "between the two legs of the will be 0.4 percent of the measured
struments because they have rather cell. magnitude. On the same basis the error
large temperature coefficients. If this . A millivoltmeter in combination with at 20 divisions will be 1 percent of the
precaution is taken, it will not be nec¬ a pair of leads and an external shunt measured magnitude. It is apparent
essary to apply temperature correc¬ (or a group of shunts to provide a then that if one is to minimize reading
tions to pointer indications. It may number of ampere ranges) is some¬ errors, only about the upper half of
also insure against errors arising from times used for measuring direct cur¬ a scale should be used. For lower values
an unsuspected difference in tempera¬ rents. To get reliable results, it is nec¬ of the measured magnitude, one should
ture of the components of an instru¬ essary to use the particular leads (or make use of a lower range instrument.
ment. leads having equivalent resistance) The laboratory and its equipment
Lamps used for scale illumination with which the millivoltmeter and its should be restored to order after com¬
radiate considerable heat and, if placed shunts were tested. Since the milli¬ pletion of each job. Instruments and
too near the coil or spring, may in¬ voltmeter may have a resistance of other equipment should be returned
crease the temperature of these com¬ only a few ohms, it is apparent that to their assigned cabinet or storage
ponents and thereby produce consider¬ a small change in resistance of the space where they will be protected and
able error in the pointer indication. leads or their contacts may cause an can be readily located by the next user.
The values indicated by moving-iron appreciable error in measurement. An instrument that has been damaged
and rectifier-type instruments may be or found to be defective should be
in error by several percent if used on IV. Measurement Procedures properly labelled “To be repaired,”
direct current or if used on alternating and should not be returned to its usual
When making a test, record on the shelf space. The person in charge
current at a frequency other than that
data sheets all pertinent information should be notified and steps taken to
for which they were designed by the
about the test and the equipment used. repair or replace the damaged part.
manufacturer. Rectifier instruments
A wiring diagram of the test setup,
are generally calibrated for use on a
date, humidity of air, room tempera¬
pure sine wave and their readings may V. Conclusion
ture, duration of warm-up period,
be considerably in error if used for In conclusion then, to get the most
makes, models, ranges and serial num¬
measurements in which the wave-form out of an instrument one should take
bers of instruments and equipment
is distorted. advantage of the information in manu¬
used should be listed. Unusual tem¬
Electrodynamic (except those having perature rises as well as defective or facturers’ instruction manuals and any
iron cores or built-in transformers), unsatisfactory performance of any other good sources. The proper connec¬
electrostatic, and thermocouple instru¬ equipment should be noted. The rec¬ tions, correct applications and limita¬
ments are designed for use either on ord should be sufficiently detailed to tions should be known for each piece of
direct current or on alternating cur¬ permit an accurate appraisal of the equipment. Such factors as frequency,
rent within specified frequency ranges. test results or a later duplication of wave-form, temperature, and stray
The mean of values corresponding to the setup to permit a satisfactory ex¬ fields, must be given careful considera¬
the two directions of current should perimental check. Test records should tion. The instruments should be tested
be used when measuring direct cur¬ be complete, in ink, and on a good grade periodically, and when in use precau¬
rent with either of these three types. of paper so that they may be filed for tions should be taken to eliminate or
When alternating current is measured, future reference. The records should to correct for all possible sources of
the frequency should lie within the be made unless the observer is certain error. The life of an instrument will
range for which the scale was cali¬ that they will have no future use. In¬ be materially increased if a little fore¬
brated. Ac.-dc. differences for these adequate test data may be a source of thought and respect are given at all
three types are ordinarily rather small genuine embarrassment at some later times in its care and use.
at power frequencies. However, for date.
measurements of the highest accuracy,
Before an observer reads an instru¬ SELECTED REFERENCES
it may be desirable to have these dif¬
ment he should make sure that the
ferences determined. 1. K. Edgecumbe and F. E. J. Ockenden.
scale is properly illuminated. This is
‘‘Industrial Electrical Measuring Instruments"
Switchboard shunts of large current of considerable importance for precise
(book), Pitman Publishing Corp., New York,
rating (1000 amperes or more) should readings or if errors resulting from 1933.
be tested in position with the current fatigue are to be minimized where
2. W. H. Lawes, ‘‘Precautions in the Use of
and potential terminal connections nor¬ consecutive readings are to be made Standard Instruments,” Journal of the Institu¬
mally used. A test in any other posi¬ over a period of time. A hand lens hav¬ tion of Electrical Engineers, Vol. 67, 1929, pages
tion is not likely to duplicate the cur¬ ing a magnifying power of 5 is a help¬ 541-548.
rent distribution through or the heat¬ ful aid in getting accurate pointer 3. E. S. Lincoln, “Electrical Measurements
ing of the shunt encountered in normal and Meter Testing in the Power Station” (book),
readings.
Weston Electrical Instrument Corp., Newark,
use and may give values of current It is a good practice for an observer New Jersey, 1929.
that differ from those for normal con¬ to read and record the pointer indi¬ 4. A. E. Peterson, “Performance of Portable
nection by as much as 2 percent. cations of an instrument in divisions Electrical Instruments in Magnetic Fields,”
Standard cells must be used with estimated to tenths (on the appropriate Transactions of the American Institute of Elec¬
trical Engineers, Vol. 67, August 1948, pages
care. Unlike electrical instruments scale of a multi-range instrument)
1228-1231.
they give no visible signs of electrical rather than in measured units. The
5. -Tames Spencer, “Maintenance and Serv¬
abuse. They are designed for use as correction table for a given instrument
icing of Electrical Instruments” (book), The
standards of emf. only and should be should also be expressed in divisions. Instruments Publishing Company, Inc., Pitts¬
properly connected to a potentiometer. There will be a minimum of confusion burgh, Pa., 1944.
431/4
A CAMERA OBSCURA FOR INSTRUMENT READING
1. INTRODUCTION
2. CONSIDERATIONS
432/1
3. THE CAMERA
The hood is part of a cone 14" long with a diameter at the viewing
end of about 6-3/4" The ground glass screen is located inside the hood
about 4" from the viewing end. The distance from the scale to the lens is
about 4", lens to center of mirror 6", and from mirror to screen 12".
The instrument to be tested is placed below the lens head so that the
instrument scale faces away from the observer. When the instrument is
placed properly for observations one sees on the vertical ground glass
screen an inverted and enlarged image of a portion of the instrument scale
and pointer.
433/2
The arrangement of components is shown in Figure 3. Note that in this
Figure only one of the 7 lamps is shown and that the hood is turned through
an angle of 180 degrees from its normal position. All the machine work was
done in the shops at the National Bureau of Standards. All the metal parts
are made of non-magnetic material: duralumin, brass or bronze. The upright
post which supports the camera, and the other bulky metal parts are so
located that, when the test instrument is in its proper position, no eddy-
current error can be detected, even in an unshielded moving-iron instrument.
The central portion of the shaft on which the boom slides is threaded.
A threaded nut for focusing the lens on an instrument scale fits into an
opening in the boom frame. By manually turning this nut the boom may be
raised or lowered to accommodate instruments of various heights.
The lens head is supported from a horizontal arm attached to the end
of the boom through a hinge joint. The lens head may be moved along the
boom arm by a rack and pinion arrangement. If the axis of the moving
element of an instrument is placed directly beneath the pivot of the boom
arm and the lens is directly above the instrument scale, it is possible to
follow the pointer over the scale by manually turning the boom arm on its
pivot. The lens moves along the arc of the instrument scale.
434/3
It may be noted that the range of mean values determined by the various
observers, as well as the total spread in their individual observations is
about twice as large with the unaided eye as with the camera. If we assume
that the true value of voltage corresponding to the chosen scale mark is 60.10
it may be concluded that the average for any of the observers will be within
0.01 division of the true value when the camera is used. Using the unaided
eye, the average error in setting may be expected to be as much as 0.03
division.
4. CONCLUSIONS
There are of course factors other than the nature and quality of the
scale that have to be considered when one is setting a pointer to a mark or
estimating pointer position on a scale. Pivot friction, spring fatigue,
self heating, and mechanical unbalance are all possible sources of error
that the manufacturer attempts to eliminate through good design and work¬
manship. Errors of observation can be minimized by the use of a reading
aid such as the one we have described.
Table 1
435/4
FIG. I
## f
FIG. 2
436/5
FIG. 4
437/6
R-oo.ISA, V. (I, 19S(a
supplies, each consisting of two stabilizers energized An abrupt change in the supply voltage is in¬
from a 3-phase 208-volt distribution system through itially practically unattenuated by most types of
suitable phase-shift and control circuits, have stabilizers. Subsequently, the change in output
♦Electrical Engineer, Electrical Instruments Section, Electricity and Electronics Division, National
Bureau of Standards
+Regulation here refers to the effect of load changes on output voltage (with fixed input voltage) and
stabilization to the effect of input-voltage changes on output voltage (with fixed load impedance).
438/1
voltage is roughly exponentially reduced to nearly in rms values for a sinusoidal wave. These factors in
zero by the control action of the stabilizer. The turn depend on the phase angle as well as the magni¬
recovery time of this action must be so short that tude of each harmonic. For a voltage wave having a
the ballistic kick of the instrument resulting from single odd harmonic of amplitude, Vfc = h^Vf, the re¬
the abrupt change is negligible. sultant error in per cent may in the worst case be as
As shown in Appendix A, the ratio, under these large as 100for a peak reading instrument and
conditions, of the maximum change in deflection, Ad, lOOhfc/k ' for an average reading instrument.
to the steady state deflection, D, of a critically
voltage and tc is the time constant of recovery of the From the power equation, P = VI cos 0, it is
stabilizer. Thus if a 1 per cent change in voltage is evident that the two voltage sources used in testing
to result in less than 0.1 per cent momentary change wattmeters by phantom loading should be stabilized
in deflection of an instrument having a period of 1 for changes in phase angle as well as magnitude. Un¬
second, the time constant t of the stabilizer must fortunately, even though the output voltage remains
c
not be more than 0.04 sec. This is about 2.5 cycles constant, a change in input voltage causes a shift in
for a 60 cps supply, a rather severe requirement. phase of output voltage of most types of stabilizers,
of approximately er -
v
2.
2
e^h^, where hfc is the ratio The slight deviations in frequency (up to 0.4
of the kth harmonic to the fundamental. (The approxi¬ cps) which occur in many power systems will not affect
mation is reasonably close for each e^ < 20 per cent the accuracy of commercial indicating instruments,
and hfc < 0.2). For example, an instrument which is but may be significant in watthour meter testing if
correct at the fundamental and has an error of 10 per accuracies better than 0.1 per cent are desired. Quite
cent at the 5th harmonic will measure the rms value apart from this frequency influence on the meter,
of a wave having 10 per cent of that harmonic with an fluctuations in output voltage of the stabilizer
error of only 0.1 per cent. The frequency influence caused by these changes in frequency should be negli¬
for most a-c instruments is very small at power fre¬ gible.
quencies and for many types increases roughly as the
Adjustability and Range
square of the frequency. Thus harmonics in a 60 cycle
supply for testing only rms instruments should not be An instrument testing source should be adjustable
greater than perhaps 20/k per cent where k is the in increments of 0.02 per cent or less from zero to
order of the harmonic. rated voltage. In addition for wattmeter and watthour
When the peak or average reading instrument (such meter testing, the phase angle between the two channels
as an electronic or rectifier instrument) is tested should be adjustable in increments of about 1 minute
with an rms instrument, the effect of harmonics de¬ or less over at least 270°, if tests are to be made
pends on the crest and form factors* of the wave, at any power factor from zero lead to zero lag with
since the scales of such instruments are always marked the usual range of instrument loads.
♦Crest-factor is the crest (peak) value divided by the rms value, and form-factor is the rms value
(1) Numbers in parentheses refer to similarly number references in the bibliography at the end of this
paper.
439/2
Transformers in the output lines should provide tudes of the output voltages and of the phase angle
voltages in several ranges up to perhaps 750 volts between them. The booster transformer, T3, keeps the
and currents (at low voltage) up to about 100 amperes. input voltage to stabilizer VSj within the proper
To minimize possible wave-form distortion caused by range. Two impedance matching transformers (not shown
magnetizing current, each transformer should be opera¬ in Fig. 3) having rated input voltages of 300 volts,
ted below the knee of its iron-saturation curve. provide seven voltage ranges from 12 to 750 volts and
paper (2) a stray field of only 0.05 gauss may affect available electronic a-c stabilizers are used, having
some types of moving iron instruments by 0.1 per cent. 1000 v-a ratings for power supply A, and 250 v-a
Thus the stray field from thfe power supply at the ratings for power supply B.
Extensive tests have verified the
instrument under test must not exceed this value.
claims by the manufacturers of 0.1 per cent regula¬
DESCRIPTION OF NBS POWER SUPPLIES tion and stabilization, with developed harmonics of
circuits of the test instrument or meter, Pt and the Measurements of the regulation and stabilization
standard instrument, Ps. If an ammeter or voltmeter of the electronic voltage stabilizers used in these
is to be tested, only one of the two output channels power supplies were made under carefully controlled
A complete schematic diagram of this type of show excellent stabilization and regulation. The
power supply is shown in Fig. 3, and the ratings of difference in the output voltages for the two tests
the components are given in Table I. Two such supplies, at 25 per cent load and 210 volts are an indication
differing essentially only in power rating, have been of slow drifts, probably due to self heating and
assembled and rack-mounted. Supply B is shown in Fig. ambient temperature changes, which make such stabi¬
4. The adjustable auto-transformers, A4, A5 and Ag, lizers unreliable for use as long-time standards of
and the 120/5 volt transformers, T4, T5, and Tg, voltage, but in no way affect their suitability in
shown in Fig. 3, provide fine control of the magni¬ power supplies such as these.
♦This use of an adjustable transformer for shifting phase in phantom-loading circuits was first described
440/3
Recovery Time reversed, to evaluate any assymmetrica1 “turnover
A number of oscillograms were made of the effect” because of the second harmonic component of
output voltage of one of the 250 va stabilizers with the voltage of the power supply.* The percentage
10 per cent change in the input voltage. In most of ±0.3 per cent) was found to be +0.3 per cent for the
these oscillograms the time constant of recovery was rectifier voltmeter (average-reading instrument) and
not more than 1/60 second (one cycle). At low input —0.8 per cent and —1.1 per cent for the two connections
voltage a marked hunting effect persisted that de¬ of the electronic voltmeter (crest-reading) instrument.
cayed at a much slower rate but would not affect an Phase Shi ft
ins t rument. The phase shifts in output voltage of each stabi¬
An rms instrument having a one-second period lizer resulting from input voltage changes, were
was connected to the output, and a number of obser¬ measured with an electrodynamic wattmeter, voltmeter
vations were made of the momentary changes in deflec¬ and ammeter. A circuit similar to Fig. 1 was used,
tion produced when the input voltage was suddenly with transformer Tj removed and A2 connected ahead
changed by 10 per cent. These changes did not exceed of the stabilizer, VS,, under test. With the power
0.3 per cent, in good agreement with the value of factor seen by the wattmeter near zero and with maxi¬
0.4 per cent computed from the measured time constant mum rated current in the wattmeter current coils, the
of the stabilizer by means of the equation developed input voltage, , to stabilizer VSj was changed in
were repeated after the leads to the instrument were No tests were made of the effect of slight fre-
^An ordinary low-power-factor wattmeter with its zero shifted upscale several divisions by means of the
zero corrector was used in place of the special center-zero instrument described in reference 4.
♦Instruments which respond to the average or rms value are not subject to turnover effects, provided (for
♦♦Similar tests with the voltage circuit of the wattmeter connected ahead of the stabilizer verified that
441/4
572772 0 - 61 - 29
quency changes upon these electronic stabilizers be- where (J and C are the restoration and motor constants
cause such effects would be expected to be negligible respectively, and is the deflection of the instru-
The stray field around the 1000-va power supply If tc < < ta, the period of the instrument, the
was measured with a search coil of known area-turns deflection remains essentially unchanged during the
and an electronic voltmeter. (2) The field at table impulse, and we have
able.
ticularly
The method
simple,
of measuring phase
The
par¬
in¬
P©' =J Tdt (5)
formation given should make some of the tests un¬
results in
Appendix A
by a small amount nlj_, which is exponentially reduced At any time thereafter, the differential equation
to zero by a stabilizer having a time constant tc. governing the motion of the instrument is
The rms current during this time is then
i = i (, 1 + ne
-t/t
«) (l) pe" + a©1 + ue = gi. (7)
2
2nGI,t GM
(2) 1 c
U61 = 0 = 0., , and 0' =
O 1 o P~
442/5
and where A is the damping constant and 6" the Thus the maximum ballistic kick or change in
AD 2jtnt (11)
P0" + A0' + U0 = 0 (7a) -- = 2.3 nt /t
D et c a
If the instrument is critically damped, i.e., E. S. Williams, who constructed both power supplies,
2
if A = 4PU, the solution to this equation for these and T. W. Cushing, who made a number of the measure¬
e \ 4jt t -2*t/t
(8)
Re ferences
where t =2 TtVp/U, and is undamped period of the Voltmeter Indications,” R. F. Estoppey, Weston
p. 445.
if ^<<1
2.0 JD
443/6
Table 1
Ratings of Components
Load Resistors R-^ & R2 1500 ohm, lOOw 1500 ohm, lOOw
Table 2
444/7
a
445/8
Fig. 5 Phase Shift in Output Voltage of 1000 v-a
Stabilizer, VSj
446/9
October I960 List of Publications LP38
Electrical Units, Instruments, and Measurements
1. General Information
44 7/1
1. A determination of the absolute ohmr using an improved self
inductor. H. L. Curtis, C. Moon, and C. M. Sparks. J.
Research NBS _21, 375 (1938) RP1137.
448/2
13. Nuclear magnetic resonance and the measurement of magnetic
fields. R. L. Driscoll. Natl. Phys. Lab. Symp. of
Precision Electrical Measurements (1955).
3. Standard Cells
449/3
4. Standards of electromotive force. G. W. Vinal, D. N. Craig,
and L. H. Brickwedde. Trans. Electrochem. Soc. 68, 139
(1935). —
4. Resistors
5. Inductors
6. Capacitors
451/5
3. Standards for low values of direct capacitance. C. Moon and
C. M. Sparks. J. Research NBS 41, 497 (1948) RP1935.
452/6
8. Lindeck potentiometer. D. W. Oliver. Rev. Sci. Instr. 26,
No. 11, 1078 (1955).
10. The use of an a-c bridge to measure core loss at high inductions.
I. L. Cooter, W. P. Harris. Am. Inst. Elec. Engrs. Trans.
Paper 56-26 (1956).
453/7
11. Thermal converters as ac-dc transfer standards for current
and voltage measurements at audio frequencies. F. L.
Hermach. J. Research NBS 48, 121 (1952) RP2296.
23. The definition and measurement of the time constant and response
time of thermal converters. F. L. Hermach. Trans. Am.
Inst. Elec. Engrs. 77» 277 (1958).
454/8
25. Thermal voltage converters for accurate voltage measurements
to 30 megacycles. F. L. Hermach and E. S. Williams.
Communications & Electronics, Am. Inst. Elec. Engrs. ,
Paper 60-135 (I960).
9. Transformers
10. Galvanometers
455/9
2. General design of critically damped galvanometers. F. Wenner.
Bui. BS_13, 211 (1916-17) S273.
456/10
12. Dielectric Measurements
457/11
572772 0-61-30
8. A feedback amplifier with negative output resistance for
magnetic measurements. I. L. Cotter and W, P. Harris.
IRE Natl. Conv. Record, Pt. 5, 217 (1958).
14. Miscellaneous
11. Phase angle master standard for 400 cycles per second.
J. H. Park and H. N. Cones. J. Research NBS 64C3 , 229
(I960).
458/12
15. References to Books and Standards on
Electrical Instruments and Measurements
459/13
15 . Principles of electrical measurements, H. Buckingham jk
E. M. Price (Philosophical Library).
460/14
Selected Papers on
Electronics
(Contents on page V)
High-Frequency Standards of the Electronic
Calibration Center, NBSBL*
M. C. SELBYf
maintained at the NBS Radio Standards Laboratory needed and the minimum required accuracies. It is be¬
and laboratories throughout the country, and to do this lieved that these findings will be useful as a guide for
at minimum possible cost to the public. further development and improvement of the NBS
The reference standards and techniques to be used in standardization equipment now available for RF power,
this link must meet two basic requirements. The first voltage, attenuation, and field strength and that equip¬
is to retain the accuracy of the NBS primary standards ment for other quantities may also be made available.
For example, pulse power, pulsed voltage, balanced
as closely as possible. The second is to reduce the time
necessary for accurate calibrations of “unknowns” to a voltages, and balanced attenuation must be standard¬
reasonable minimum. It is, of course, clear that the ized, whereas the equipment and techniques developed
major consideration in developing a primary standard so far are generally applicable only to CW unbalanced
and technique for a certain quantity is the realization conditions. Other quantities such as current, peak
power, phase shift, thermal noise, modulation, antenna
of a magnitude as closely as possible to its true or “ab¬
gain, RF interference and, indeed, any measurable RF
solute” value. The time element, convenience to user,
and other considerations are of secondary importance quantity need standardization for the simple reason
and are given weight only when choosing one of two that the “measurements” without a stated accuracy are
methods rendering equal basic accuracy and reliability. utterly useless.
That is why primary standards and associated tech¬ Power Standards
niques are, in many cases, not practical for calibrating
At the time of this writing, three different types of
even limited numbers of reference standards. This is
high-frequency power reference standards have been in¬
particularly true as the radio frequency involved in¬
stalled at the Electronic Calibration Center: the low-
creases. At the time of initiation of this undertaking the
level thermistor bridge, the dry static calorimeter, and
calibration, for example, of an RF voltmeter at points
the liquid flow calorimeter. The first covers the approxi¬
up to several hundred megacycles was a job requiring
mate range of 100 juw to 100 mw; the second, 20 mw to
a number of weeks and at a nearly prohibitive cost; this
12 watts; and the third, 10 to 350 watts. The dry static
was particularly discouraging in view of the limited
calorimeter is fully described by Hudson and Allred1
ability of the “unknowns” to retain their calibrations
and will, therefore, not be rnentioned further here.
for a reasonable length of time. The new reference
standards and techniques incorporated, as described Bolometer Bridge
below, permit calibration of the same voltmeter in a
The operation of the power-measuring bridge is con¬
matter of days with relatively little loss in accuracy as
ventional and is based on the equivalent heating effects
compared directly with the previous calibration in
of dc power and RF power when each is in turn dissi¬
terms of the primary standard.
pated in a purely resistive load. A thermistor forms one
There seems to be little doubt that the ranges and
arm of a Wheatstone bridge circuit which is biased with
accuracies quoted below are inadequate to meet present
dc power until the bridge is balanced. When RF power
needs, in view of the considerable deterioration in ac¬
is fed into the thermistor simultaneously with the dc
curacy from the initial calibrated instrument to the
bias power, the bridge becomes unbalanced because of
final one because the equipment is calibrated through
the change in the resistance of the thermistor; a change
several steps starting with the National Bureau of
in the dc power must then be made to restore the initial
Standards and ending with the working bench equip¬
balance. The quantity of dc power withdrawn is
ment. The present minimum NBS accuracy require¬
equated to the RF power present (for the case of an
ments of many RF quantities are still largely unknown.
equal arm bridge) as follows:
* Manuscript received by the PGI, August 7, 1958. 1 P. A. Hudson and C. M. Allred, “A dry, static calorimeter for
t Boulder Labs., National Bureau of Standards, Boulder, Colo. RF power measurement,” this issue, pp. 292-296.
463/262
1958 Selby: High Frequency Standards of the Electronic Calibration Center, NBSBL
TO
POTENTIOMETER
(b)
resistor whose value is about 100 times that of the bridge Power levels in the range 10 to 350 watts to be meas¬
resistance. The maximum change in the total current ured in the Electronic Calibration Center will be done
due to load (bridge) impedance changes will then be only with a liquid flow type calorimeter which employs sili¬
about 1 per cent. Accuracies of the order of 5 per cent cone oil as the coolant. The RF power to be measured
464/263
IRE TRANSACTIONS ON INSTRUMENTATION December
is dissipated in a cylindrical resistor of the evaporated
thin film type, which serves as the inner conductor in
a coaxial system. The outer conductor is a thin-walled
tractorial taper2 jacket made of copper. This load has a
VSWR of less than 1.15 at all frequencies from 0 to 2500
me.
Operation of the calorimeter may best be understood
by reference to the block diagram of the equipment
shown in Fig. 3. Oil is pumped from the constant tem¬
perature reservoir through a flow meter to an input con¬
nection on the load where it passes -through an outer
copper jacket surrounding the load. Inside the load
heat is transferred from the cylindrical resistor to the
oil which is returned to the reservoir. In the reservoir
means are provided to cool the oil to its initial tempera¬
ture. The temperature rise of the oil, caused by its pas¬
sage through the load, is measured by a 10-junction dif¬
ferential thermopile, the cold junctions of which are
located in the oil stream immediately ahead of the load.
Fig. 3—Block diagram of liquid flow type calorimetric RF power
The hot junctions are located at the point where the
meter. Frequency range: 0 to 2500 me, power range: 10 w to
oil leaves the load, and mixing orifices are inserted in 350 w, RF accuracy: +1 per cent or better, maximum VSWR:
below 400 me—1.05, below 2500 me—1.16.
the line ahead of both sets of junctions to insure that the
true average temperature of the oil stream is the
quantity measured. The RF load and thermopile are ranges from 5 to 10 minutes depending on the power
enclosed in a dewar flask to prevent heat interchange level. This relatively short-time constant is realized
with the surroundings. The oil bath temperature is kept because the oil enters the load at a fixed temperature.
at or near ambient temperature. Thus, only the load need reach equilibrium rather than
In operation, RF power is fed into the load through the entire system as would be the case were not the
the type-N connector and the flow rate F is adjusted to reservoir temperature kept constant. Another advan¬
give a steady-state differential temperature AT of from tage in using a constant reference temperature is obvi¬
3°C to 6°C. When thermal steady state is reached in the ous from (2). The power is computed in terms of mass
system, the values of AT and F are observed and the flow, whereas the flow meter readings are given in terms
power absorbed is calculated by means of of volume flow. Since, for a fixed volume, the mass is a
function of density which is in turn a function of abso¬
P = C X F X AT (2)
lute temperature, computations are simplified by main¬
where C is the specific heat of the oil and F is in units taining the reference temperature constant.
of oil mass flow per second. Thus, the wattmeter does A photograph of the equipment is shown in Fig. 4.
not require calibration since power is measured in terms The upper half of the left-hand console houses the reser¬
of fundamental quantities and the known properties of voir and flowmeter while in the lower section is the
a material. The performance of the calorimeter has, dewar flask containing the load. The right-hand console
however, been checked, using known values of dc or houses the potentiometer used in determining AT.
low-frequency power.
Measurement accuracy of the wattmeter is limited RF Voltage Standards
primarily by the accuracy of the flowmeter which is The purpose of the RF voltmeter consoles is to pro¬
+ 0.5 per cent over a dynamic range of 10 to 1. Higher vide as rapidly as possible means of calibrating all types
accuracies ( + 0.25 per cent) are obtainable if the range of voltmeters in discrete steps ranging from 30 kc to
of the flowmeter were limited to say 2 to 1. Some sacri¬ 700 me with the maximum accuracy available at the
fice, however, would then be necessary in the dynamic present time. Two major steps made the above possible.
range of the wattmeter. The value of C is known to an The first was the use of the AT voltmeters;3 the second
accuracy of +0.1 per cent while AT is measured to an was the decision to calibrate at discrete frequencies. The
accuracy of +0.2 per cent. Thus the over-all accuracy latter decision is justified because the frequency response
of the wattmeter is better than +1 per cent; this esti¬ curves for present-day RF voltmeters seldom have se¬
mate has been checked and confirmed using dc power. rious discontinuities. However, frequencies other than
The time necessary for the calorimeter to reach those selected may be necessary. At the time of this
thermal equilibrium after the RF power is turned on writing two of these consoles have been completed and
2 C. T. Kohn, ‘‘Radio frequency coaxial resistor using tractorial 3 “Stable radiofrequency voltmeters,” NBS Tech. News Bull., vol.
jacket,” Proc. IRE, vol. 3, pp. 951-960; August, 1955. 40, pp. 29-30; February, 1956.
465/264
1958 Selby: High-Frequency Standards of the Electronic Calibration Center, NBSBL
(V, ) i'v, 1
\_/
1-5 V 5-20V
FILTER AAS
0LR
DC POWER MANUAL
SUPPLY LEVEL
CONTROL
466/265
IRE TRANSACTIONS ON INSTRUMENTATION December
Rapid connection of the “unknown” voltmeter to a
reference standard is achieved with the RF voltmeter
coupling plate, Fig. 7. Reference standards, seven for
each half of a console, are mounted on the under side of
each plate. To connect the “unknown” in parallel with
a standard, th‘6 operator attaches a suitable adaptor
ring to the probe of the “unknown,” inserts this ring in a
sliding carriage on the top side of the plate, centers it
over the reference standard to be used and clamps it in
place. The “unknown” and the standard are now con¬
nected in parallel through a T connection of approxi¬
mately i^-inch long, with the RF voltage applied mid¬ COAXIAL CABLE
FOR RF INPUT
way between the two. Such a T exists at each of the
seven standards. The operator then rotates the seven- Fig. 7 —Cross section of voltmeter coupling plate.
Note: carriage not sectioned.
position switch until the number below the index corre¬
sponds to the position of the standard in the plate and
closes the switch. The RF level is adjusted with the
manual level controls and the dc thermocouple output
in millivolts, as indicated on the panel meter, is con¬
verted immediately into actual applied RF voltage with
the aid of the line chart displayed in the window.
As already indicated above, fixed type AT voltmeters
have been chosen in preference to the adjustable type
used in the laboratory because of their simplicity and
superior stability. The mechanism required to operate
the piston of the continuously adjustable type increases
their complexity of construction and operation. Three
reference standards, having dissipative fixed attenuator
input pads, cover the voltage range 0.2 to 20 volts in
ranges of 0.2-1.0, 1.0-5.0, and 5.0-20 volts, respec¬
tively. These are used at both frequencies provided in
Fig. 8—Block diagram of the duo-channel null
the part of the console in which they are located. The balance RF attenuation system.
voltage ranges of 20-100 and 100-500 volts at each fre¬
quency are covered by single-frequency standards using
Attenuation Standards
fixed “capacitive” attenuators, bringing the total num¬
ber of reference standards in each half of a console to The present plans call for 8 or 9 independent attenua¬
seven. The series reactance of this type of “capacitive” tion consoles, one each for 9 discrete frequencies. At
attenuator is many times larger than the resistance of least 6 and possibly 7 will employ new precision, 3-inch
the thermoelement heater terminating the attenuator. ID, waveguide-below-cutoff attenuators. The other 2,
Therefore, in analyzing the effect of a change in heater for 30 and 100 kc, will probably employ dissipative at¬
resistance with a given RF voltage input to the volt¬ tenuators. One ol these consoles, for 30 me, is approach¬
meter on heater current, the attenuator can be repre¬ ing completion. The block diagram, Fig. 8, shows the
sented at its output terminals by a constant current general method and basic equipment which will most
source. Changing the heater resistance even by as much likely be used in all cases.
as 50 per cent has no effect on heater current. The system consists of a series substitution branch
Consequently, once an RF voltage calibration of a where the standard and unknown attenuators are con¬
given reference standard has been made, replacement nected in the conventional way. This branch is shunted
of its thermoelement does not necessitate an RF re¬ by another branch incorporating an impedance phase
calibration. Only a dc calibration to determine the rela¬ shifter. The vectoral sum of the outputs of the two
tionship of heater current to dc millivolts output need branches is detected by a common monitor or null de¬
be made, and if the characteristics of the replaced couple tector. The procedure, as usual, is to adjust the attenu¬
match those of the original closely enough, even this ator and phase shifter for a null balance both before and
calibration is unnecessary. When matching can be after inserting the unknown into the circuit. The differ¬
achieved, the same calibration chart can remain in serv¬ ence between the two positions of the attenuator and
ice. For the voltmeters using dissipative attenuators, phase shifter gives the results desired. For maximum
however, a complete recalibration of the voltmeter is sensitivity and minimum noise it was found best to
required when thermoelements are replaced. The over¬ work with output levels of each channel as high as pos¬
all accuracy is + 3 per cent or better. sible.
467/266
1958 Selby: High Frequency Standards of the Electronic Calibration Center, NBSBL
468/267
IRE TRANSACTIONS ON INSTRUMENTATION December
micrometer projector to an accuracy of +0.0001 inch standard free-space quasi-static magnetic field produced
in 6 inches. by a single-turn, unshielded, balanced, transmitting
The attenuation range of this console at 30 me, based loop of known radius and with a known current flowing
on preliminary measurements of provisional setups and with a typical unit as shown in Fig. 13. The receiver loop
design analysis, is expected to be 0.0 to 80 db at an esti¬ must, of course, either be shielded or balanced to ground
mated accuracy of +(0.005 db + 0.03 per cent of the to eliminate pickup from the electric component of the
indicated value in db). The maximum sensitivity of the field. In calibrating a field-strength meter the setup is
experimental setup (for unknowns of up to 10 db) was made in a cleared space such that a distance of at least
of the order of 0.00001 db. two or three times the loop spacing, d, exists to the near¬
est sizable metallioobjects and to the ground.
Field-Strength Standards
The magnitude of the field strength produced by a
The National Bureau of Standards offers a calibra¬ single-turn circular transmitting loop is given by (3) for
tion service for radio field-strength meters employing the case of coaxial transmitting and receiving loops as
loop-type antennas at frequencies from 30 c to 30 me shown in Fig. 13. The actual value of the quasi-static
and for meters employing dipole-type antennas from 30 magnetic field, H, produced by the loop is expressed in
to 300 me. The calibration consists of measuring the terms of the equivalent electric component, E, that
over-all linearity of the instrument at one or more fre¬ would exist in a free-space radiation field. The relation¬
quencies and power levels, measuring the internal at¬ ship used is E = ZH, where Z is the impedance of free
tenuator ratios at one or more frequencies and measur¬ space (Z = 376.7 ohms).
ing the so-called antenna coefficient or correlation factor
relating field strength to the output meter reading under 60trrxH
E _ (3)
specified conditions of set gain at specified frequencies.
_ (<72 + rx2 + r22)3/2
All measurements are made in terms of sinusoidal volt¬
ages or currents. where
The portions of the calibrations involving only RF E = equivalent free-space electric field strength in
voltage ratios are made in terms of precision dissipative
rms volts per meter,
step attenuators which comprise “Pi” or “T” coaxial rx = radius of transmitting loop in meters,
sections of evaporated or deposited metal film for fre¬ r2 = radius of receiving loop in meters,
quencies up to 300 me. d = axial spacing (meters) between coaxial loops
Since the over-all linearity of a field-strength meter (make d > 7rx and d > 7r2).
will, in general, be a function of both frequency and RF / = transmitting loop current in rms amperes,
input level, it is necessary to determine this dependence. X = free-space wavelength in meters.
Fortunately, this effect is usually limited to the highest
attenuator steps in those types of sets having inter¬ The value of field strength used for calibration is of the
mediate-frequency attenuators only. The departure order of 0.1 v/m, (rx = 0.1 m, / = 0.1 a).
from linearity has been found, in general, to be less than The “antenna coefficient,” K, is evaluated by the
2 per cent if the signal voltage applied to the mixer grid relation
does not exceed about 5 per cent of the injected local
Ei
oscillator voltage. For practical and economical reasons K = -X-■» (4)
it is necessary to limit linearity checks on the attenuator +x M1
steps in question to one or two frequencies as specified
where
by the user.
For those sets also employing RF attenuators in the E\ = standard field given by (3) usually expressed in
antenna input circuit it is necessary to insure that these microvolts per meter,
ratios are measured under actual operating conditions. Ai = true attenuator ratio used,
Those RF attenuator steps which may be affected by Mi = output meter reading corrected from linearity
the changing source impedance of the antenna, or the data,
changing input admittance of the first RF stage with /= frequency, usually in kc or me.
frequency, are measured with the antenna placed in a In using the field-strength meter later for actual
field which can be attenuated in steps of known ratios. measurements, the unknown field in microvolts per
The attenuation linearity is certified to an accuracy meter is given by
of + (0.1 db + 0.3 per cent of ratio in db). Fig. 12 shows
the equipment installed for these measurements. KM2A2
At the present time the National Bureau of Standards E =-— > (5)
certifies only field-strength meters using loop antennas
below about 30 me. The calibration is made in terms of a where K is given by (4), and
469/268
1958 Selby: High-Frequency Standards of the Electronic Calibration Center, NBSBL
f= frequency in same units as used in (4). Owing to the lack of selectivity in this method as
used, it does not lend itself to calibration work in the
The accuracy of antenna coefficients is usually certi¬ vicinity of strong interfering fields that may often be
fied to + 3 per cent below 5 me and + 5 per cent between encountered.
5 and 30 me for sets using loop antennas. While it is possible to use either the standard-field or
At the frequencies in the VHF band, the National the standard-antenna method for determining the an¬
Bureau of Standards certifies only field-strength meters tenna coefficients, the latter method is used at the Na¬
using dipole antennas. The determination of antenna tional Bureau of Standards at the present time. Field
coefficients is made in terms of a radiation field by the tests have shown the excellent agreement possible be¬
standard-antenna method using half-wave dipoles and tween the two methods throughout the VHF band. The
horizontally polarized transmission. choice was largely one of convenience, as it was found
The field strength of a locally generated field is de¬ somewhat more convenient to provide an accurate cali¬
termined by measuring directly the induced EMF in a bration of the silicon crystal voltmeter than to insure an
standard receiving dipole oriented for maximum re¬ accurate measurement of the transmitting antenna cur¬
sponse and calculated from the relation: rent throughout the VFIF band.
The “antenna coefficient” of the field-strength meter
| EI = e/l.H, (6)
under test is determined by placing its dipole antenna
where at a specified height, usually 10 feet, in the radiation
e = induced EMF' in the receiving dipole in volts, field previously measured using the standard receiving
dipole and the relationship given by (6). Use of the
Iji = effective length of the dipole in meters, (la = \/ir
meter at a later time at other antenna heights or over
meters for a half-wave dipole).
ground having appreciably different constants may re¬
The EMF is measured directly by means of a relatively quire corrections for heights less than about one wave¬
high impedance silicon crystal voltmeter built into the length.
gap at the center of the antenna. This eliminates the The accuracy of the antenna coefficients certified by
necessity tor a separate measurement of the antenna the National Bureau of Standards depends largely upon
470/269
IRE TRANSACTIONS ON INSTRUMENTATION December
the quality of the individual meter under test. In gen¬ jects.4-8 The individuals to be credited with the de¬
eral, the.se coefficients are certified at present to ± 10 velopment, design, and construction of the equipment
per cent for frequencies between 30 and 150 me and to are the project leaders, C. M. Allred, L. F. Behrent,
+ 15 per cent for frequencies between 150 and 300 me. F. M. Greene, P. A. Hudson, A. H. Morgan, M. C. Selby,
and E. C. Wolzien, and every member of the High-
Conclusion
Frequency Electrical Standards Section including those
The equipment described above embodies techniques assigned to the Electronic Calibration Center as well as
and design features arrived at during the last several a number of machine shop personnel who contributed
years. Because requirements for better accuracy are valuable ideas and skill throughout this activity.
already in evidence, there is a strong likelihood that con¬
siderable modifications in the above units may be intro¬ 4 F. M. Greene, “Calibration of commercial radio field-strength
meters at the National Bureau of Standards,” NBS Circular 517;
duced and perhaps entire setups may be replaced within
December, 1951.
the near future. The technical phase of the Electronic 6 C. C. Cook, “Calibration of field-strength meters,” Tele-Tech.,
vol. 11, pp. 44-46, 96, 99-101; October, 1952.
Calibration Center should of necessity remain fluid in
6 F. M. Greene and M. Solow, “Development of very high-fre¬
order to keep services, as closely as possible, up to date. quency field-intensity standards,” J. Res. NBS, vol. 44, pp. 527-547;
May, 1950. RP2100f
Complete papers are expected to be published on indi¬
7 C. M. Allred, “Precision piston attenuators,” NBS Rep. 5078;
vidual equipments as soon as possible. However, addi¬ May 1, 1957.
tional information may be obtained from previous Na¬ 8 M. C. Selby, L. F. Behrent, and F. X. Ries, “RF-voltage calibra¬
tion consoles,” 1958 IRE National Convention Record, pt. 5,
tional Bureau of Standards publications on these sub¬ pp. 251-257.
471/270
National Standards of Time and some cases as to origin also. The situation
is analogous to using a measuring rod of
Frequency in the United States* different length than the standard rod but
Time of known calibration. For example, UTO is
the astronomically observed Universal Time
Time is one of the independent quantities or mean solar time, uncorrected for polar
chosen as a basis for the measuring system of
variation and annual fluctuation in the
science. To establish a time scale, that is, to earth’s speed of rotation; UTl is UTO cor¬
specify a time coordinate, we must first rected for polar variation; UT2 is UTl
establish an origin and a constant unit of
corrected for annual variation. None of the
time and lay off the unit at least as far as
UT seconds is constant. They are variously
all points of interest, in exact analogy to the
determined by several nations. Various other
specification of position. The orbital motion
nominal “seconds” are defined by the dura¬
of the earth about the sun provides a time
tion of a specified number of periods of vari¬
scale, called Ephemeris Time (ET) which is
ous oscillators or resonators, such as quartz
suitable in these respects, and also consti¬ or atomic devices. These all have the out¬
tutes a standard from which the unit may
standing characteristic that they are more
always be obtained. In October, 1956, the
readily observable than ET. The atomic
International Committee on Weights and
time unit, furthermore, is assumed for the
Measures, with representation from the
present to be as constant as the unit of ET.
United States, effectively adopted the sec¬
ond of ET as the fundamental unit of time The United States Frequency
by resolving that “the second is the fraction Standard
1/31,556,925.9747 of the tropical year for
January 0, 1900 at 12 hours Ephemeris A national standard of frequency called
Time.”1 The tropical year for any given the United States Frequency Standard
epoch (or moment) is the interval, taken (USFS) is maintained at the Boulder Lab¬
symmetrically about that epoch, necessary oratories, National Bureau of Standards,
for the mean longitude of the sun to increase for the purpose of making immediately and
by 360°, as measured along the ecliptic from continuously available, through the Stand¬
the vernal equinox. This definition of the ard Frequency Broadcasts discussed below,
second has been one of many attempts to a provisional time scale with which to make
define a unit of time 1) which remains con¬ measurements. This provisional scale is suf¬
stant with epoch, 2) which is physically ficiently accurate for all civil and most
realizable to high precision, and 3) which is scientific uses. Nevertheless, for the highest
permanently available for observation. possible accuracy, this scale must be con¬
Ephemeris Time is obtained in practice tinually corrected to or calibrated against
by observations on the position of the moon the national time scale as finally deter¬
and reference to tables giving this position mined, and it must be relatable to other fre¬
as a function of ET. As the fundamental unit, quency and time scales in common use.
the second of ET replaced the second of The USFS in general consists of the
Universal Time (UT), a time scale based weighted value of the outputs (reduced of
on the rotation of the earth. The delay in course to a common basis) of several actual
the determination of UT from the astro¬ oscillators and resonators maintained at the
nomical observations is of the order of a Boulder Laboratories, these devices being
month; the delay in the determination of among the best available at any current
ET to any useful degree of accuracy is of state of the frequency control art. In par¬
the order of several years. ticular, the LJSFS is presently stable to 2
The maintaining of a national standard parts in 1010 or better over intervals from
of time consists of the process of determining about 1 to 103 minutes; over longer inter¬
the time scale in the United States from the vals, its value has been maintained as con¬
international standard and making this in¬ stant as possible prior to October 9, 1957
formation available to the user. Often, how¬ with respect to the UT2 second as deter¬
ever, as with position, it is only time differ¬ mined by the U. S. Naval Observatory, and
ences or time intervals which are of interest. since October 9, 1957 wdth respect to atomic
This need for convenient interval deter¬ frequency standards. These atomic stand¬
mination has led to the common use of other ards have been compared with other atomic
time scales with easier realizability than ET. standards via a network comparison. Such
Measurements often have been, and still atomic standards have been shown to be in
are, expressed in terms of time scales other agreement and to remain constant with
than ET. These other scales are all related respect to each other to 5 parts in 10'° or
or relatable to ET and to each other through better.2'3
conversion factors with various degrees of
precision as to size of unit at least, and in 2 Essen, Parry, Holloway, Mainburger, Reder, and
Winkler, “Comparison of cesium frequency standards
of different construction,” Nature, vol. 182, pp. 41-42;
* Received by the IRE, October 15, 1959. July, 1958.
1 Comite International des Poids et Mesures, tMockler, Beehler, and Barnes, “An evaluation
“Proces-Verbaux de Seances,” 1956 Session, Ser. 12, of a cesium beam frequency standard,” Symposium
vol. 25, Gauthier-Villars, Paris; 1957. Record, Office of Naval Research Symposium on
Quantum Electronics, Bloomingburg, N. Y.; Sep¬
tember 14-16, 1959.
472/1
Standard Frequency Broadcasts USFS and the Standard Frequency Broad¬
casts to the national time scale has changed
The United States Frequency Standard
over the years as improvements in the meas¬
is distributed to interested users by means
urement of time and frequency have been
of standard broadcasts of Radio Stations
made.
WWV and WWVH. The frequencies of these
Fig. 1 shows schematically the relation
stations are kept in agreement with respect
to each other and have been maintained as between the frequencies {i.e., cycles per
constant as possible with respect to the common unit of time) of hypothetical oscil¬
United States Frequency Standard since De¬ lators which oscillate such as to mark off the
cember 1, 1957. The nominal broadcast unit of the time scales indicated after ex¬
frequencies should, for the purpose of highly actly the same number of cycles in each
accurate scientific measurements, for estab¬ case. The frequencies are shown relative to
lishing high uniformity among frequencies, the value of the hypothetical oscillator
or for removing unavoidable variations in the marking off the defined unit of ET. The
broadcast frequencies, be corrected either figure is equivalently interpreted as the re¬
to the United States Frequency Standard, ciprocal of the size of the units of the indi¬
as indicated in the monthly tables published cated time scale relative to the unit of ET.
in the IRE,4 or to a particular time scale as The Standard Frequency Broadcasts and
determined by the Naval Observatory, with Time Signals have been and are related to
adequate limits assigned for propagation the UT and A.l time scales as explained
errors. above. From these observations, frequency
corrections of the Standard Frequency
Time Signals Broadcasts to these scales are easily made;
Time signals, sufficiently accurate for all and in addition, conversion from the UT
civil and most scientific uses, are also carried scale at a given epoch to the ET scale is
by the standard broadcasts. The WWV and available after reduction of astronomical ob¬
WWVH time signals are also kept in agree¬ servations.7-9
ment with each other. They are locked to the Prior to October 9, 1957, the final value
nominal frequency of the transmissions, and of the USFS was assigned retrospectively
consequently may depart continuously from on the UT scale from the determination of
other time scales such as UT2. Corrections the frequency of the Standard Frequency
expressed as the times of reception of the Broadcasts on this scale and the known rela¬
WWV time signals on th$ UT2 scale are de¬ tion between the Standard Frequency
termined and distributed by the U. S. Naval Broadcasts and the USFS. These values
Observatory.5 Recently, the Observatory have not been published, but are a matter of
has established a particular atomic time scientific record at the National Bureau of
scale, A.l, with unit determined by 9,192,- Standards.
631,770 periods of Cs at zero held and origin With improvement in stability and con¬
coinciding with UT2 on January 1, 1958, and stancy realized in the maintenance of the
has published times of reception of WWV on USFS by the availability of atomic fre¬
this scale also.6 Agreement with time on the quency standards, monthly publications of
UT2 scale, or simply agreement with UT2, corrections of the Standard Frequency
within + 30 msec at all times, has been main¬ Broadcasts to the USFS were made from
tained by making step adjustments in time October 9, 1957, to realize the advantage of
of precisely plus or minus twenty millisec¬ rapid correction to a more constant fre¬
onds on Wednesdays at 1900 UT when quency than that based on the unit of UT2.
necessary. Beginning January 1, 1960, the It is thus also necessary to relate the
broadcast frequencies will be offset from the USFS and Standard Frequency Broadcasts
United States frequency standard by a dif¬ to frequencies provided by atomic frequency
ferent amount than heretofore in order to standards such as cesium. Since the relation
establish a unit in substantial agreement of an atomic time unit defined by a given
with the current value of the unit of UT2. number of periods, N, of an atomic oscillation
Thus the time signals, locked to the broad¬ to the unit of ET has not yet been adopted
cast frequency, will require less frequent ad¬ internationally, we are required to state ex¬
justment than in the past. plicitly an assumed value of N when com¬
paring frequencies or times to an atomic
Corrections of the USFS and the scale. In comparing the USFS to an atomic
Standard Frequency Broadcasts time scale, the following equality was
The method of final correction of the adopted as of October 9, 1957, and holds to
4 W. D. George, “WWV standard frequency trans¬ a precision of 1 part in 1010; 100,000.000 . . .
missions,” Proc. IRE, vo!. 46, pp. 910-911; May, periods of the USFS equals Ni periods of
1958, and subsequent months.
5 “U. S. Naval Observatory Time Signals, I. Pre¬ the zero-field (4, 0)^(3, 0) Cs transition,
liminary Times of Reception, UT2,” Bulletin B, where AJ equals 9,192,631,838.
U. S. Naval Observatory, Washington, D. C. (un¬
published). If No is the number of Cs oscillations per
“U. S. Naval Observatory Time Signals, Final unit of ET, then the ratio No/N\ may be
Times of Reception, UT2,” Bulletin A, U. S. Naval
Observatory, Washington, D. C. (unpublished). used to convert the USF'S to the ET scale
6 “Time Service Notice No. 6,” U. S. Naval Ob¬ as soon as No is determined. Markowitz,
servatory, Washington, D. C., January 1, 1959 (un¬
published). et al.8 have measured No to be 9,192,631,770
473/2
572772 0 - 61 - 31
+ 20 as the number of Cs oscillations per In summary, to express a given time in¬
unit of ET, providing the best and only terval or frequency in terms of the inter¬
value expected for several years. Effective national standard, it must be referred to a
January 1, 1960, the value of the USFS will standard broadcast, or some atomic stand¬
be corrected so that iVi equals 9,192,631,770, ard with proper regard for the precision of
and the published corrections to the Stand¬ this comparison, and the latter standard
ard Frequency Broadcasts will thus be given must be referred to the ephemeris second as
with respect to the unit of ET as realized by indicated herein with proper regard for the
atomic standards to a precision of the inter¬ uncertainties involved.
comparison of atomic standards (a few parts National Bureau of Standards
in 1010) and to an accuracy determined by Boulder, Colo.
the work of Markowitz, et al. (±22 parts
in 1010). 7 Essen, Parry, Markowitz, and Hall, “Variation
in the speed of rotation of the earth since June, 1955,”
The final correction of the Standard Fre¬ Nature, vol. 181, p. 1054; April, 1958.
8 Markowitz, Hall, Essen, and Parry, “Frequency
quency Broadcasts, and hence the USFS, to of cesium in terms of Ephemeris Time,” Phys. Rev.
the national time scale, will be effected by Letters, vol. 1, pp. 105-106; August, 1958.
9 Brouwer, “A study of the changes in the rate of
the Naval Observatory by the observation of rotation of the earth,” Aslron. J., vol. 57, pp. 125-
the times of reception as described. 146; September, 1952.
ET ADOPTEO Cs DETERMINED
474/3
UNITED STATES DEPARTMENT OF COMMERCE • Frederick H. Mueller, Secretary
NATIONAL BUREAU OF STANDARDS • A. V. Astin, Director
Contents
Page
1. Technical services and related information_ 1
1.1. Standard radio frequencies_ 1
1.2. Standard audio frequencies_ 2
1.3. Standard time intervals_ 2
1.4. Standard musical pitch_ 2
1.5. Time signals_ 2
1.6. Radio propagation forecasts_ 2
1.7. Geophysical alerts_ 4
1.8. Radiated power, transmitting antennas, modulation_ 5
1.9. Other standard frequency and radio time signal services_ 5
For sale by the Superintendent of Documents, U.S. Government Printing Office, Washington 25, D.C.
Price 10 cents
475/1
Standard Frequencies and Time Signals From NBS
Stations WWV and WWVH
Detailed descriptions are given of six technical services broadcast by National Bureau
of Standards radio stations WWV and WWVH. The services include 1, standard radio
frequencies; 2, standard audio frequencies; 3, standard time intervals; 4, standard musical
pitch; 5, time signals; and 6, radio propagation forecasts. Other domestic and foreign
standard frequency and time signal broadcasts are tabulated.
476/1
1.2. Standard Audio Frequencies 1.4. Standard Musical Pitch
(a) Program The frequency 440 cps for the note A above
middle C is the standard in the music industry in
Two standard audio frequencies, 440 cps and
many countries and has been in the United States
600 cps, are broadcast on each radio carrier fre¬
since 1925. The radio broadcast of this standard
quency. The audio frequencies are given alter¬
was commenced by the National Bureau of Stand¬
nately at 5-min intervals starting on the hour with
ards in 1937. The periods of transmission of 440
600 cps, as shown in figure 1. At WWV, the first
cps from WWV and WWVH are shown in figure 1.
tone period (600 cps) of each hour is of 3-min With this broadcast the standard pitch is main¬
duration; the remaining periods are of 2-min
tained, and musical instruments are manufactured
duration. At WWVH, all tone periods are of 3- and adjusted in terms of an unvarying standard.
min duration.
Practically no instruments are manufactured
which cannot be tuned to 440 cps. Listeners of
(b) Accuracy
music are benefited because there are fewer instru¬
ments not in tune.
The accuracy of the audio frequencies, as
transmitted, is the same as for the carrier. The
frequency offset mentioned under 1.1(c) applies.
1.5. Time Signals
Changes in the propagation medium (causing
(a) Program
Doppler effect, etc.) result at times in fluctuations
in the audio frequencies as received. The audio frequencies are interrupted at pre¬
cisely 3 min before each hour at WWV, and 2 min
1.3. Standard Time Intervals before each hour at WWVH. They are resumed
on the hour at WWV and at 5- and 10-minute
(a) Program intervals at both stations as shown in figure 1.
Universal Time (referenced to the zero meridian)
Seconds pulses at precise intervals are derived is announced in International Morse Code each 5
from the same oscillator which controls the radio min from WWV and WWVH. This provides a
carrier frequencies, e.g., they commence at intervals quick reference to correct time where a timepiece
of 5,000,000 cycles of the nominal 5 Mc/s carrier. may be in error by a few minutes. The 0-to-24-hr
They are given by means of double-sideband system is used starting with 0000 at midnight at
amplitude-modulation on each radio carrier fre¬ longitude zero. The first two figures give the hour
quency. Intervals of 1 min are marked by omit¬ and the last two figures give the number of minutes
ting the pulse at the beginning of the last second past the hour when the tone returns. For example,
of every minute and by commencing each minute at 1655 UT, four figures (1, 6, 5, and 5) are broad¬
with two pulses spaced by 0.1 sec; the first pulse is cast in code. The time announcement refers to
used to mark the beginning of the minute. The the end of an announcement interval, i.e., to the
2-min, 3-min, and 5-min intervals are synchro¬ time when the audio frequencies are resumed.
nized with the seconds pulses and are marked by At station WWV a voice announcement of
the beginning or ending of the periods when the Eastern Standard Time is given before and after
audio frequencies are off. The pulse duration is each International Morse Code announcement.
0.005 sec. The pulse wave form is shown in For example, at 9:10 a.m., EST, the voice an¬
figure 2. At WWV each pulse consists of five nouncement in English is: “National Bureau of
cycles of a 1,000-cps frequency. At WWVH Standards, WWV; when the tone returns, Eastern
each pulse consists of six cycles of a 1,200-cps Standard Time is 9:10 a.m.”
frequency. The pulse spectrum is composed of
(b) Corrections
discrete frequency components at intervals of 1.0
cps. The components have maximum amplitudes
The time signals are kept in close agreement with
at approximately 995 cps and 1,194 cps for the
UT2 by making step adjustments of precisely 50
WWV and WWVH pulses, respectively. The
msec when necessary. Such adjustment will be
tone is interrupted 0.040 sec for each seconds
made on the first of the month following the month
pulse. The pulse starts 0.010 sec after commence¬
in which the transmitted time departs from UT2
ment of the interruption. by more than 50 msec. Corrections to the time
signals are published periodically by the U.S.
(b) Accuracy Naval Observatory.
The accuracy of the time intervals, as trans¬ 1.6. Radio Propagation Forecasts
mitted, is the same as for the carrier with an addi¬
tional limitation of ±1 /^sec. The frequency A forecast of radio propagation conditions is
offset mentioned under 1.1(c) applies. (See note broadcast in International Morse Code on each of
concerning time adjustments under section 1.5(b), the standard radio carrier frequencies; from WWV
Time Signals.) at approximately 19.5 and 49.5 min past each hour,
477/2
and from WWVH at approximately 9.4 and 39.4 the forecasts are first broadcast at 0639 and 1839
min past each hour, as shown in figure 1. Propa¬ UT, respectively. These are the short-term fore¬
gation notices were first broadcast from WWV in casts prepared by NBS-CRPL North Pacific
1946; the present type of announcement has been Radio Warning Service, Box 1119, Anchorage,
broadcast from WWV since July, 1952, and from Alaska.
WWVH since January, 1954. The forecast is broadcast as a letter and a digit.
The forecast announcement tells users the con¬ The letter portion of the announcement identifies
dition of the ionosphere at the regular time the the radio quality at the time the forecast is made.
forecast is made and how good or bad communi¬ The letters denoting quality are “N,” “U,” and
cation conditions are expected to be in the suc¬ “W,” signifying that radio propagation conditions
ceeding six or more hours. The NBS forecasts are normal, unsettled, or disturbed. The digit
are based on information obtained from a world¬ portion is the forecast of the radio propagation
wide network of geophysical and solar observa¬ quality on a typical North Atlantic (from WWV)
tories, including radio soundings of the upper or a typical North Pacific (from WWVH) trans¬
atmosphere, short wave reception data, and similar mission path during the 6 or more hours after the
information. Trained forecasters digest the infor¬ forecast is made. Quality is graded in steps
mation and formulate the predictions. ranging from 1 (useless) to 9 (excellent) as follows:
From WWV the forecasts refer only to North
Atlantic radio paths, such as Washington to Disturbed, Unsettled Normal
London or New York to Berlin. The times of grades (W) grade (U) grades (N)
issue are 0500, 1200 (1100 in summer), 1700, 2300
UT. These are the short-term forecasts prepared 1— useless 5—fair 6— fair-to-good
by NBS-CRPL North Atlantic Radio Warning 2— very poor 7— good
Service, Box 178, Ft. Belvoir, Va. 3— poor 8— very good
From WWVH the forecasts are for North 4— poor-to-fair 9— excellent
Pacific radio paths, such as Seattle to Tokyo, or
Anchorage to San Francisco. As of March 1, If, for example, propagation conditions at the time
1959, the times of issue are 0600 and 1800 UT, and the forecast is made are normal but are expected
RADIO PROPAGATION |
FORECAST (CODE)'
0 10 20 30 40 50 0 10 20 30 40 50 0 10 20 30 40 50 60 SECONDS
1 19 5 AND 49 5 MINUTES PAST EACH HOUR 3 43 AND 34.3 MINUTES PAST EACH HOUR
2 9 4 AND 394 MINUTES PAST EACH HOUR 4 14.4 AND 44 4 MINUTES PAST EACH HOUR
WWV
START
EACH
HOUR
WWVH
478/5
to be only “poor-to-fair” within the next 6 or alert, the symbol AGI AAAA is broadcast for
more hours, the announcement would be broad¬ 24 hr. This means that a significant magnetic
cast as N4 in International Morse code. storm has started, with A’-index reaching 5 or
higher at a midlatitude station. It can also mean
1.7. Geophysical Alerts that an outstanding auroral display has been
reported (or inferred, because the /V-index has
A symbol indicating the geophysical “state of reached at least 7) or that an outstanding increase
warning” as declared under the international pro¬ in cosmic ray flux has been observed. The IWDS
gram of the International Council of Scientific
Warning Message itself, distributed through other
Unions, is broadcast in very slow International
channels including the meteorological communi¬
Morse Code on each of the standard radio carrier
frequencies; from WWV at approximately 4.5 and cation networks coordinated by the World Mete¬
34.5 min past each hour, and from WWVH at orological Organization, specifies whether the
approximately 14.4 and 44.4 min past each hour, alert is Magnetic Storm, Aurora, or Cosmic Ray;
as shown in figure 1. Such notices were first the “state of warning” symbol broadcast from
broadcast during the International Geophysical WWV and WWVH does not make these distinc¬
Year 1957-58 and are continuing under the tions at the present time.
similarly organized program, International World A Special World Interval in progress is indicated
Day Service (1WDS). by the symbol AGI and three extra long dashes.
The “state of warning” symbol indicates to This means that an alert, as above, has been
experimenters in radio, geophysical, and solar declared and, further, that the geophysical
sciences the content of the IWDS Warning activity is of sufficient, interest to warrant special
Message issued at 1600 UT by the World Warning attention and intensified observations by experi¬
Agency on days when outstanding geophysical menters throughout the world. A Special World
events have occurred in the preceding 24 hr. Interval will usually last two or three days. The
The information in each such message is first Interval is considered to extend until 2359 UT on
broadcast from WWV at 1604.5 UT and from the final day even though the symbol “Special
WWVH at 1714.4 UT. World Interval in progress” is not broadcast after
If the IWDS Warning Message declares an 1600 UT (WWV) or 1700 UT (WWVH).
FREQUENCY DEVIATION OF RADIO STATION WWV (WITH REFERENCE TO THE ATOMIC STANDARDS SINCE DEC.1,1957)
15
10
Figure 2. Sample characteristics of time and frequency broadcasts from NBS stations WWV and WWVH.
479/4
When the “state of warning” is neither alert (c) Modulation
nor Special World Interval in progress, the symbol v
broadcast is AGI EEEEE. 1 he amplitude modulation, double sideband, is:
1.8. Radiated Power, Transmitting Antennas, Audio frequencies 440 or 600 cps—75 percent,
Modulation voice and seconds pulses, peak—100 percent.
kc cps kw
WWVB > Boulder, Colorado 60 0. 0015
DCF77 Federal German Republic 77. 5 1,200, 440 12
CIIU Ottawa, Canada 3330, 7335, 14670 1 0. 3, 3, 5
OMA Czechoslovakia 50 5
GBR Rugby, England 16 300
MSF Rugby, England .... 60 1, 1000 10
NBA U.S. Navy, Canal Zone ... __ 18 1 100
1 National Bureau of Standards station WWVB (formerly KK2XEI) broadcasts continuously and the call letters are keyed approximately on the hour
and each 20 min thereafter. The frequency is normally controlled by an atomic standard. Present plans are to Increase the radiated power to about 3 kw as
soon as possible.
480/5
2.2. Other Broadcasts foreign radio time signals is given in chapters 3
and 9 of Radio Navigational Aids, Hydrographic
The U.S. Naval Observatory, Department of Office Publication No. 117 (formerly H.O. Pub.
the Navy, broadcasts time signals continuously No. 205), for sale by local and foreign authorized
or at regular intervals from a number of stations sales agents of the U.S. Navy Hydrographic
including NSS (Annapolis, Md.), NPG (Mare Office or direct from the U.S. Navy Hydrographic
Island, Calif.), NPM (Pearl Harbor, Hawaii), Office, Washington 25, D.C. Price of the publi¬
and NBA (Balboa, C.Z.). Detailed information cation, with a binder, is $5.00; contents only
may be obtained from the U.S. Naval Observa¬ (without binder), $3.50.
tory, Washington 25, D.C.
A comprehensive list of United States and Boulder, Colo.
Summary—High-precision frequency and time Equipment and Methods in General Use of the polyphase electrostatic or electro¬
standards are becoming more and more widely used magnetic type is desirable in the divider
in many technical fields. The basic equipment used by A high-precision frequency standard may
be defined as one whose changes in frequency chain, preferably at the 1,000-cycle stage,
the Central Radio Propagation Laboratory of the Na¬
tional Bureau of Standards in providing and broad¬ are less than 1 part in 108 per day. This de¬ to permit the clock or seconds signals to be
casting standard frequency and time signals is dis¬ gree of precision now requires continuous synchronized. This device also provides a
cussed. Mention is made of the manner in which it is operation of the standard oscillator, with convenient means of measuring small daily
adjusted and the corrections which may be applied the control unit and other critical elements differences in time kept by different stand¬
to make use of the ultimate accuracy that may be ex¬
operating in a temperature-stabilized com¬ ard-oscillator clocks, including time signals
pected in using these signals for measuring and cali¬
partment. Very careful shielding, filtering of received by radio. Measurements of time
brating similar equipment. Several methods are given differences may be made by adjusting the
for checking the frequency of precision oscillators and
battery supply leads, and use of buffer am¬
plifier stages are also necessary, especially if phase shifter dial until the pulses coincide
precision clock performance. Suggestions are given as
more than one oscillator unit is operated at on an oscilloscope screen when alternately or
to methods of recording and evaluating performance
data for such standards. Expected improvements in a single location. Such standards now gener¬ simultaneously connected.
constancy and accuracy and possible future changes ally employ a GT-cut quartz-crystal1 Other methods for measuring and record¬
in the types of standards used in physical time meas¬ operated' in a bridge-stabilized oscillator ing time differences may employ a spark
urements are considered.
circuit arrangement.2 chronograph, or a polyphase modulator and
A standard oscillator in wide use operates integrating phasemeter.3
H
Introduction
at 100 kc and has a multiposition switch for A harmonic generator and mixer unit
IGH-PRECISION frequency and coarse frequency adjustments of approxi¬ may be conveniently used in rapid fre¬
time standards are becoming more mately 4 parts in 1011 per step. A precision quency intercomparison. This device may be
and more widely used in technical gear-driven capacitor with a drum dial ar¬ provided with two inputs and means to con¬
■and scientific fields. Among their uses are: rangement of 5,000 dial divisions provides trol the inputs which feed into the crystal
control sources for standard-frequency and for a control of frequency to approximately diode harmonic generators. The common
time broadcasting: time standards for as¬ 1 part in 109 per division. A dial with such an output is connected to a radio receiver for
tronomical observatories; synchronization expanded scale is of great advantage in mak¬ counting the difference beats between two
of master and slave stations for pulse navi¬ ing precise frequency adjustments and inter¬ local standards. Two frequency multiplier
gation systems; investigation of long-dis¬ polations. units having outputs of 500 and 2,500 kc or
tance radio transmission phenomena by A minimum of three standard oscillators other convenient frequencies, are recom¬
pulse techniques; as the heart of frequency is recommended for those installations re¬ mended to facilitate frequency measure¬
synthesizers in large communications sys¬ quiring a reasonable maximum of reliability ments at higher harmonics.
tems; and in physical research laboratories and continuity of service. By intercomparing
for precise calibrations and measurements, three standards locally, either once daily or Standard Frequency and Time Broadcasts
such as time-rate phenomena and micro- continuously, short-time stability may be from WWV
wave spectroscopy. determined to a much higher order than pos¬ The radio and audio frequencies as
While this paper deals primarily with the sible through radio transmission compari¬ transmitted from WWV, near Washington,
adjustment and performance of high-pre¬ sons. The most reliable standard may thus D. C.4 are accurate (with reference to mean
cision frequency and time standards, many be determined and used to supply the de¬ solar time) within 1 part in 50 million. The
of the principles involved and techniques dis¬ sired need, with the other units available for time signals broadcast by WWV are main¬
cussed should prove helpful to those using standby duty. tained in agreement with U. S. Naval Ob¬
the WWV frequencies or time signals in ad¬ For checking frequencies by the time- servatory time within several hundredths of
justing or calibrating other high or medium comparison method, and for supplying the a second. This is done by setting the WWV
precision equipment. desired audio frequencies and time intervals, control oscillator frequency slightly higher
The accuracies presently required by each standard oscillator should be operated or lower than exactly 100 kc by an amount
some users of frequency and time services continuously, and two or more such oscil¬ ordinarily not greater than 1 part in 108, to
are approximately as shown in Table I. lators should be provided with frequency- advance or retard gradually the broadcast
dividing and synchronous-clock equipments. time. For this reason the time broadcast by
TABLE I The frequency dividers may be pulse count¬ WWV is uniform, changing by less than
ers, fractional frequency generators, locked 0.001-second average or 0.002-second maxi¬
oscillators, or multivibrators. The choice mum per day. Fig. 1 shows the frequency
Accu¬
racy Sec/day of type should be influenced by the number and time deviations of the WWV transmis¬
Service of output frequencies and wave form de¬ sions for the years 1947 and 1948.
(fre¬ (time)
quency) sired, simplicity of adjustment, and reli¬ The present oscillators at WWV have a
ability of operation required. If seconds in¬ gradual and fairly constant drift to a higher
Physical research tervals are desired for laboratory use, a frequency of from 0.6 to 1.2 parts in 109 per
Astronomers 1X10"8 ±0.001 system similar to that used at WWV, where¬ day. The control standard at WWV was
Monitoring stations in the pulse trains are generated electroni¬ adjusted on an average of every twelve days
Survevors lxio-7 ±0.01
cally and selected mechanically by a rotating during the past year. Adjustments are usu¬
Radio broadcasting
Astro-navigators 1 xio-6 ±0.1 cam, offers certain advantages. An alternate ally to a lower frequency but are occasion¬
Commercial method is to divide the standard frequency ally made to a higher frequency because of
communication 1X10-5 down to 1 cps by counter methods, employ¬ slight discrepancies in the assessment of
Musical instruments 1 X10-4 ing an all-electronic system. average drift rate in terms of the Observa¬
Commercial power A continuously adjustable phase shifter tory’s time determinations. The adjustments
distribution 1X10-3 (±5)
1W. P. Mason, “A new quartz crystal plate, 3 W. A. Marrison, “Evolution of quartz crystal
designated the GT, which produces a very constant clock,” Bell Sys. Tech. Jour., vol. 27, pp. 510-588;
* Decimal classification: R214. Original manu¬ frequency over a wide temperature range,” Proc. July, 1948.
script received by the Institute. September 27, 1949. I.R.E., vol. 28, pp. 220-223; May, 1940. 4 “Technical radio broadcast services, radio station
t National Bureau of Standards, Washington, 2 L. A. Meacham, “The bridge stabilized oscillator, ” WWV.” Letter circular LC886, obtainable from Na¬
D. C. Proc. I.R.E.. vol. 26, pp. 1278-1294; October. 1938. tional Bureau of Standards, Washington 25, D. C.
482/6
1950 Shaull: High-Precision Frequency and Time Standards
o N O
o o
o
> o .T)
s 3 5
Fig, 2—Chart recording of the beat frequencies at 80 Me of the three WWV standard oscillators in terms
of the monitoring reference oscillator at CRPL.
483/7
PROCEEDINGS OF THE I.R.E. Januar
N>
JV>
6
b o
o >
£ z
LU
o
>
o
59
Fig. 3 Chart recording showing short-time variations in the beat frequency between two frequency standards at the
National Bureau of Standards, on different days, at 80 Me (top) and 1,280 Me (bottom).
484/8
1950 Shaull: High-Precision Frequency and Time Standards
Z
2 z
o
UJ
O tt
UJ <
cr cl
Fig. 5 Illustration of Doppler variation of the WWV received frequencies in terms of two frequency standards at Western
Electric’s Hawthorne, Chicago, Laboratory. Dotted points are WWV transmitted frequencies.
Experience has shown that it is often by taking the average of several counts of measurement period, which make the re¬
quite difficult to adjust a local compared beats at a single offset point at approxi¬ ceived frequency slightly lower or higher
oscillator directly to zero beat frequency mately 1 part in 107 from the zero-beat than that transmitted. This well-established
with WWV at remote locations by either point. The average of these counts in seconds Doppler effect could be readily computed
aural or S-meter means. This is generally per beat is converted to beats per second for radio propagation if sufficient informa¬
caused by numerous rapid variations in which represents the amount the oscillator is tion on changes in the medium were avail¬
signal level, or by propagational flutter in then off zero beat at the standard frequency able. Its bearing on frequency measurements
the received frequency. Tests at the National to which the receiver is tuned. This is then may be outlined as follows. The ionospheric
Bureau of Standards where a ground-wave converted to parts-in-109 deviation from zero reflecting layers vary in height considerably
signal from WWV prevails, showed no diffi¬ beat by dividing the beat in cycles per second with frequency, time of day, season, geo¬
culty in reproducing zero-beat settings of an by the standard frequency in cycles per graphical location, and phase of sunspot
oscillator dial to ± 1 part in 109. second and multiplying by 109. To return cycle. Average conditions are predictable,
the oscillator to zero beat it is only necessary but conditions on a given day may depart
Midtiple-Setting Interpolation Method to advance or retard the dial the indicated greatly from the mean. Changes in effec¬
number of parts in 109 with allowance for tive layer height will cause the frequency
Multiple-setting interpolation is best the incremental dial correction factor for the of a received carrier to differ from that
carried out by setting the fine-frequency dial region in which it is operating. The dial cor¬ transmitted by a fractional amount equal
of the compared oscillator slightly off fre¬ rection factor (number of parts in 109 per to the rate of change in equivalent path
quency so as to obtain beats with WWV of dial division) is most readily obtained by length on kilometers per second divided
from 0.1 to 2 beats per second. Several set¬ comparing the standard with a similar local by the propagation velocity in kilometers
tings should be made on each side of zero standard at several dial settings, and thus per second. For example, if one assumes
beat, choosing values which are readily dis¬ avoiding radio transmission difficulties and E2-layer propagation and that the virtual
tinguished from fading, and a number of possible errors. height where reflections occur changes from
beat periods measured by means of a stop This method has the advantage of re¬ 300 to 400 km in 2 hours, then for a “3-hop”
watch. Values thus obtained are then con¬ quiring only one offset adjustment. With transmission front WWV, Washington, to
verted to beat frequencies in cycles per sec¬ either method the amount of offset should WWVH, Maui, (7,700-km surface distance)
ond and are plotted on rectangular co-or¬ be made small as any adjustments may an effective change in frequency of the re¬
dinate paper as ordinates with dial settings temporarily disturb the normal drift rate to ceived signal of -3.2 parts in 10s for each
as abscissa. By plotting the beat frequencies some extent, even though the dial is returned hop can be computed. On a similar basis the
obtained below zero beat negative and the to the original setting. If several units are computation for “4-hop” transmission (at a
ones above zero beat positive, and drawing a available in the frequency standard arrange¬ higher angle of departure) gives -3.7 parts
straight line through the average of these ment, the oscillator which is off frequency in 108 per hop so affected. Of course, it is
points, an accurate zero beat dial setting will by a small amount and infrequently ad¬ unlikely that changes of this magnitude will
be indicated where this line crosses the x axis. justed may be used to make the comparison occur simultaneously at all reflection points
For this method an adjustment dial having with the transmitted frequency. The others of the path considered; thus the error caused
good linearity, very little backlash and one may then be evaluated in terms of this by the Doppler effect will usually be less
dial division equal to about 1 part in 109 is standard so that no changes in the others are than the above figures multiplied by the
required. necessary. The offset standard should be number of hops.
sufficiently off-frequency so as to be certain The WWV frequencies as received at
Offset Method that it is higher or lower than the received Maui, T. H., in October, 1948, were gener¬
frequency. ally slightly higher in the morning and
The multiple-setting method is somewhat
slightly lower in the afternoon. Measure¬
cumbersome in that it requires a number of
Variations in Received Frequency Arising ments based on reception at a time when
separate settings of the dial and computa¬
from the Doppler Effect noon occurred about halfway between trans¬
tions and involves plotting the data, prefer¬
mitter and receiver showed consistent agree¬
ably on rather large graph paper. As the dial The accuracy of any method using direct ment to about 1 part in 108, with occasional
setting versus frequency is generally quite comparison of a local frequency standard discrepancies as great as 2 or 3 parts in 108.
linear over a small frequency range, the de¬ with a received standard frequency will be Observations madein England on WWV’s
sired zero-beat setting may be more easily affected by fluctuations in transit time of the 15-Mc received frequencies in December,
obtained to about the same degree of accu¬ received signals. This arises from changes in 1945, showed errors ranging from 2 to —7
racy by using an offset method. This is done the radio propagation medium during the parts in 10s, w'ith an average variation of — 3
485/9
PROCEEDINGS OF THE I.R.E. January
parts in 108 when compared with very stable Interval-Derived Frequency Determinations Clock Acceleration
primary frequency standards.6 Similar meas¬ In order to determine the amount of time
The computation of average frequency
urements reported by Booth and Gregory a crystal clock will gain or lose over a long
by time comparisons may be explained by
showed slightly greater variations.6 Fig. 5 period it is necessary to consider the effect
the following equations:
shows the variation in WWV’s 5-Mc received
of the change in frequency (and thus the
frequency at Western Electric’s Hawthorne,
change in rate of the clock) over the period.
Chicago, Laboratory in terms of two pre¬ CD
J- 2 ““ i 1 For very long intervals it is even necessary
cision frequency standards. It is evident
to consider the changein the drift rate of the
that high-accuracy measurements of fre¬ where /o is the nominal frequency of the os¬
oscillator (change in the change of rate of
cillator, i.e., that frequency for which the
quency in terms of WWV should be made the clock).
when ionospheric layer heights are likely to frequency division ratios were chosen; h, tz
The time a crystal clock will indicate at
be most stable, i.e., with noon or midnight and Ti, Tz represent the time indicated by
some future time Tx at / days distant is:
prevailing at about halfway between trans¬ the clock, and the correct time represented
by the time signals respectively at the be¬ T1 = To + / + at + bt2 + ct3 (3)
mitter and receiver locations. Long-distance
ginning and end of the period being averaged. where at represents the “rate” term, bit the
north-south transmission may be expected
Equation (1) is usually simplified in acceleration term, and ct3 the change in ac¬
to show greater variations in the morning
actual use by referring all measurements to celeration term.
and afternoon where all of the reflection
days or multiples thereof and by using The terms T0 and t are always positive.
points are subject to changing conditions at
measured errors in time indicated by a Coefficient ‘a may be positive or negative,
the same time. Long-distance comparisons
crystal clock. For example, the average and is the principal factor in determining
should, in most cases, prove most reliable
frequency of a 100-kc standard oscillator the overall rate of the clock over short pe¬
when lower angles and fewer modes of propa¬
during interval T is riods. Coefficient ‘b’ is generally positive, and
gation prevail. However, one should par¬
ticularly avoid use of a frequency and time / Ah - A/A must be considered for periods of more than
/av=106(l+ ---J, (2) a few days. Coefficient ‘c is very small and
where a dominant mode of transmission is
very near the maximum usable frequency at is generally negative. It may be neglected in
the propagation angle for any of the reflec¬ where Ah and A/2 are positive or negative all but very long-time computations.
tion points, as the effective layer height errors in time at beginning and end of the The above equation may be more readily
changes very rapidly under these conditions. measurement period. The value obtained understood if times T0, TXl Tt, etc., are con¬
It should be mentioned that errors other represents the instantaneous frequency for sidered as marking off an absolute, continu¬
than those attributed to the Doppler effect the center of the period, if uniform drift is ous time scale, with the expression t-\-at-\-
were occasionally found possible. assumed, or the value for the entire period bt2-\-ct3 marking off the distance covered by
Publications of the National Bureau of if no change in frequency during the entire the clock on this time scale in / days. This
Standards are available which are useful in period is assumed. Instantaneous departures may be recognized as the well-known equa¬
determining the optimum reception frequen¬ from the average values are generally deter¬ tion for linear motion. Usually t+at is com¬
cies at a given time and location.7'8 mined by daily comparisons with other os¬ bined and considered as the velocity term,
cillators, involving an averaging based on but for horological purposes it is more con¬
their predicted daily drifts. This is justified venient to keep these two terms separate.
Measurement of Precision Standard by the probability that the average of the Horologists express the “rate” of a clock as
Oscillators by Daily Time drifts of a number of selected oscillators will the seconds gained or lost per day.
Comparisons with WWV be more nearly linear than the drift of any Fig. 6 shows the computed daily change
one unit. in time of a crystal clock for a uniform daily
The average frequency of a standard
Daily measurements of time differences change in frequency of 1 part in 109 per day.
oscillator may be determined by successively
or errors between standard clocks and time The clocks are usually set slightly fast in
comparing the number of cycles generated
signals may be accomplished by adjustment such manner as to lose time, and become
(time kept by its synchronously-operated
of a calibrated phase shifter dial for pulse slightly slow as the oscillator passes through
clock) with any sufficiently reliable time
coincidence on an oscilloscope screen or by correct frequency. As the oscillator becomes
signals. If the standard's drift rate or grad¬
use of calibrated circular or linear sweeps on high in frequency, this time is regained to
ual change in frequency is assumed to be
the oscilloscope. A recording chronograph approximately the original setting when the
constant, its average frequency during the
may also be used to get a continuous record oscillator frequency is about 1 part in 107
period will be the same as the instantaneous
of the time differences of several standard high. The oscillator is then adjusted low
frequency for the center of the period con¬
clocks and time signals. again and the cycle repeated, this adjust-
sidered. Departures from a constant or uni¬
form frequency drift can be detected by this
method only by successive determinations FREQUENCY = 99,999.9900 +
486/10
1950 Shaull: High-Precision Frequency and Time Standards
487/11
PROCEEDINGS OF THE I.R.E. January
quency is amplified and supplied to the 58 to taking such data on a number of standards ance. If continuity of service is only of mod¬
62 cps frequency recorder so that the 60- is to determine the most suitable one for erate importance, two equipments may be
cycle point represents zero difference fre¬ use, and to determine its short-time stability infrequently adjusted and only the working
quency. This gives a continuous record of in terms of the other available standards. unit held within very precise limits. The os¬
the difference frequency to a sensitivity of a When such a standard is used to control cillator and time equipment least adjusted
few .parts in 1010 with a range of ± 2 parts in standard-frequency transmissions or to should be generally used as the reference or
108. The other oscillators can be switched make high-precision calibrations, it should one by which frequency and time determina¬
on successively in rotation to record the be monitored continuously against one or tions are made with respect to WWV. For
deviation of each one in terms of the refer¬ more similar standards. most purposes an adjustment of the working
ence oscillator. As a result of experience gained in the standard so as to hold its frequency within
A variation of this method which is cur¬ operation of frequency-standard oscillators 1 or 2 parts in 108 of WWV’s average fre¬
rently used at National Bureau of Stand¬ at the National Bureau of Standards and quency should prove satisfactory, with the
ards is to set the reference oscillator approxi¬ in particular at the WWVH, Maui, T. H., standby and spare oscillators being held to a
mately 60 parts low in 100 million, which transmitting station, it may be helpful to frequency tolerance of 1 part in 107.
gives a recordable difference frequency outline maintenance, procedures and tech¬ Daily phase shifter (time) adjustments
without the use of the phase-shifting mech¬ niques that have been developed and found may be made if time synchronization is de¬
anism. This method has the disadvantage of useful in this work. sired, and these daily changes used in com¬
putting the reference off frequency by an puting interval-derived frequencies. Allow¬
amount making it unusable for many pur¬ ance should be made for transmission time
Logging and Plotting of Data
poses. lag in setting or using the local time pulses,
The difference frequency between two At the NBS it has been found necessary if extremely precise time synchronization is
standard oscillators may be determined by to keep a daily record which includes all important.
accurately measuring the time required to frequency and time readings and measure¬
complete one beat at the fundamental or a ments, as well as the final results of all com¬ Short-Interval Performance
multiplied harmonic of each standard. A putations of frequency and time. Changes in
method for doing this has been described control standard, methods, or equipment are High-precision standard-frequency os¬
by H. B. Law, and is used by the British always noted. These data are conveniently cillators now in use have short-period sta¬
Post Office and the National Physical Lab¬ kept in a record book having a number of bilities (intervals of one minute or so) rang¬
oratory for the comparison of precision fre¬ ruled columns, or on specially prepared ing from 1 part in 109 to 1 part in 1010. In
quency standards.10 A balanced phase dis¬ mineographed sheets carrying the desired addition to the gradual ageing or drift,
criminator is supplied with two 100-kc fre¬ notations for a particular installation. various causes will contribute to these short-
quencies and arranged to trigger a counter The daily computed frequencies for each time frequency fluctuations. Mechanical
chronometer at the beginning and ending of oscillator are plotted on continuous cross- shock may cause an instantaneous effect
one beat. The chronometer counts a conven¬ section paper in a manner as shown in Fig. through easing of stresses or displacements
ient standard frequency (100 or 10 kc) and 4. These data are of considerable aid in in the crystal unit or associated components.
thus indicates the elapsed time of one beat. studying the relative performance of the A slow recovery in frequency may or may
The instrument accuracy under ideal con¬ individual standards, and in graphically not take place. A change in supply voltage,
ditions is estimated to be 1 part in 10n. extrapolating predicted values during peri¬ produces an immediate effect followed by
Two electronic counter chronometers ods when reception conditions do not permit additional changes as thermal equilibrium is
may be used in a number of ways to obtain making the daily frequency or time checks. restored. Changes in load impedance will
accurate comparison data between standard It has been found desirable to keep peri¬ cause frequency variations unless proper
oscillators. A general method uses one in¬ odic (weekly) instrument readings which decoupling is employed. Unless proper ther¬
strument to count the difference or beat include the readings of all battery chargers mal lagging is provided for the crystal com¬
frequency between high harmonics of the and power supplies, voltmeter readings on partment cyclic operations of its heaters will
two compared frequencies, while a second each switch position for the frequency stand¬ also cause variations in the output frequency.
instrument counts cycles of appropriate ards and frequency dividers, temperature The oscillator circuits must be protected
standard frequency. One unit may be ad¬ readings of each standard’s oven where ap¬ from direct or electromagnetic mutual
justed to serve as a predetermined counter plicable, and room temperature at time of coupling to better than 100 db to reduce
to stop the other unit after a definite count readings. These readings are useful in anti¬ frequency “pulling” or tendency to syn¬
and thus increase the accuracy somewhat. cipating tube or equipment failure in that chronize. Other changes in frequency, often
This method of frequency measurement is adjustment or repair can often be made be¬ of unpredictable source, may be caused by
very" flexible and can be used to measure fore complete breakdown occurs. imperfect connections, faulty components
frequencies which fall outside the limits of or erratic vacuum tubes. These causes of
other very sensitive measurement devices. poor frequency stability can generally be
Adjustments in Frequency and Time
A very simple but accurate method of eliminated only by a slow process of substi¬
recording low beat frequencies such as ob¬ Periodic adjustments in frequency and tution and observation.
tained from the harmonic mixer and receiver time of each standard oscillator and time Where average frequencies are computed
comparison method is to use a recording equipment are required to keep within the for periods of more than a few days, the daily
milliammeter or recording oscillograph. The desired accuracy tolerance and to simplify values may be determined by calculating
beats are then evaluated in terms of uniform the measuring and plotting requirements. the departure of the reference standard from
chart speed or time markers recorded simul¬ The frequency and range of these adjust¬ its average curve in terms of the other reli¬
taneously. This method offers comparative ments are influenced considerably by the able standards available. The daily values
simplicity for temporary or experimental type of service and standard. For greatest for the other standards may then be deter¬
use, but is quite laborious where a number constancy of drift rate and greatest ease in mined by adding the daily beat frequencies
of readings are required over a long period computation, frequency adjustments should for each standard to the computed value for
of time. be held to a minimum. For many uses the the reference. The mean values of relative
calculation of and allowance for the slight frequencies thus obtained represent a
Recording and Evaluation of Data error in frequency or time presents no great weighted extrapolation in terms of the se¬
difficulty. In other cases, where numerous lected group of reliable standards. Consid¬
The amount of statistical data taken on measurements are being made and services erations involving three independent reli¬
each standard should be kept to the mini¬ given, it is helpful to establish high initial able standards should give a relative daily
mum necessary to determine the probable accuracy so as to eliminate corrections. accuracy within 2 parts in 10°. The inclusion
performance of each individual unit over a Where the working standard (one used of six or more standards, as is done in the
continuous period. The chief reason for or distributed) is held to very close accuracy CRPL primary standard of frequency, re¬
tolerance, the continuity of service required sults in a relative daily determination of
10 H. B. Law, “An instrument for short-period fre¬ will determine if the second or standby performance of each standard oscillator to
quency comparisons of great accuracy,” Jour. I EE
vol. 94, part III, pp. 38-41; January, 1947. standard need be held to a similar close toler¬ within 1 part in 109. Very-short period or
488/12
1950 Shaull: High-Precision Frequency and Time Standards
48&/13
572772 0 - 61 - 32
PROCEEDINGS OF THE I.R.E. January
490/14
1950 Shaull: High-Precision Frequency and Time Standards
multipliers, auxiliary frequency-modulated solar time and Newtonian time would from one or both of the standards, is possible
search oscillator, and pulse discriminator amount to about three hours. to a precision of a few parts in 10s through
circuits. The 100 kc is then divided down to comparisons of average frequencies over
audio frequencies in conventional manner Conclusions
periods of 6 or more days.
and used to operate a synchronous-motor Frequency and time standards, using The development of atomic or molecular-
clock as a time standard. A constancy in fre¬ high-precision quartz crystals, are now avail¬ resonance standards of high constancy and
quency of 5 parts in 10s has been obtained able which are capable of supplying fre¬ absolute accuracy may greatly simplify the
for periods of several days with this experi¬ quencies and time intervals constant to maintenance of precise frequency and time
mental clock when compared with WWV considerably better than 1 part in 108 per standards. The practical realization of such
frequency standards. Improvement of this day. In order to achieve an accuracy ap¬ standards, which seems reasonably probable
type of standard and the development of proaching this order, these standards must in the near future, will eliminate the neces¬
more constant types of atomic resonance be frequently checked in terms of standard sity of making highly precise physical meas¬
controlled oscillators are to be expected. frequency or time broadcasts. An accuracy urements in terms of the earth’s variable
It is shown in the references cited that of 1 part in 10s represents about the limit rate of rotation. Such a standard would
the sharpness of resonance within individ¬ obtainable in terms of the earth’s mean rate supply a means of studying more precisely
ual oscillating molecules is extremely great. of rotation over^ 100-day period. Longer the motions of the earth and other astro¬
Because of collisions of molecules with each periods of averaging can not be expected to nomical bodies.
other and with the gas cell walls, and the give greatly improved accuracy, because of Considerable work is being done to im¬
Doppler broadening attributable to natural the possibility of slight oscillator frequency prove the constancy of quartz-crystal fre¬
thermal agitation, the practical working Q deviations and uncertainties in the uniform¬ quency standards, especially with regard to
of an ammonia gas absorption line ranges be¬ ity of the determinations of the earth’s mean aging or frequency drift and improved tem¬
tween 50,000 and 500,000. This compares rate. perature control methods.
favorably with the Q of quartz crystals used By using ordinary zero beating methods, It is probable that frequency and time
in frequency standards which ranges from a remote frequency standard may be ad¬ standards, which have improved by a factor
100,000 to 1,000,000. It is thus reasonable to justed within 1 part in 107 to WWV’s re¬ of ten or more per decade in the last thirty
hope that a constancy of 1 part in 108 to 1 ceived frequency. Special offset techniques years, will continue to reach new orders of
part in 109 may be obtained by proper refine¬ permit this setting error to be reduced to accuracy and constancy. However, these im¬
ments in circuitry and technique. Whether less than 1 part in 108. Changes in the radio provements in accuracy will probably be re¬
or not this degree of constancy of absolute propagation medium may cause the received ferred to a new kind of standard, rather than
value can be maintained without or even frequency to differ from that transmitted by to the mean solar second.
with precise temperature and pressure regu¬ as much as several parts in 107. By averaging
lation remains to be investigated. a number of determinations made when Acknowledgments
Civil time will, no doubt, continue to be noon or midnight prevails about halfway be¬
defined in terms of the mean solar second, tween transmitter and receiver, long-dis¬ The author wishes to acknowledge his in¬
as would ordinary frequency designations. tance frequency comparisons can generally debtedness for the valuable advice and as¬
After about 2,000 years, if the earth con¬ be made with a precision of better than 1 sistance of W. D. George, Chief of the High-
tinues to slow down at its present rate, the part in 108. Frequency Standards Section of the Na¬
mean solar second would be about 1 part in The intercomparison of two remote os¬ tional Bureau of Standards, and the many
3 million longer than at present and the ac¬ cillators, constant to 1 or 2 parts in 109 per suggestions of V. E. Heaton and E. L. Hall
cumulated time difference between mean day, by means of transmitted time pulses of the same organization.
491/15
Reprinted from The Review of Scientific Instruments, Vol. 31, No. 4, 410-413, April, 1960
Printed in U. S. A.
Accurate and convenient microwave cavity wavemeters are described that are suitable for many precise physical
measurements. High Q’s were attained by refined construction techniques. At 9000 Me a precision of 0.02 Me was
obtained. The absolute accuracy also may be very high after strains in the metal have stabilized. A calibration
table containing 104 entries, easily formed by means of a high speed computer, makes the wavemeters convenient
to use. The curve fitting method, accurate to one in 10fi, is described.
scientific experiments. For example, the present wave¬ investigation of this subject is in progress.
meters have been used to observe dielectric, magnetic, and
II. DESCRIPTION OF WAVEMETERS
other cavity perturbations, and they have been very valua¬
ble. They have conveniently replaced more complicated A good choice for the mode of oscillation of a wavemeter
methods of frequency measurements that use heterodyning is the TEoin mode of a right circular cylinder, as was indi¬
and frequency multiplication. The precision achieved with cated by Wilson, Kinzer, and Schramm.2 The wavemeters
the wavemeters is 1 or 2 parts in 106, as estimated both described here operate in the TE0n mode. All were trans¬
from routine use and from the repeatability of settings mission type cavities. Table I summarizes several char¬
during original calibration. The absolute accuracy depends acteristics of these wavemeters, designated as A, B, and C,
of course on the long time mechanical stability, the atmos¬ respectively. Figure 1 is representative of cavities A and B.
pheric conditions, and the VSWR of the waveguide arms. Figure 2 gives dimensions for cavity B, but the perturba¬
It was found, for example, that after a year one wavemeter tion plunger was only used in A. Figure 3 shows cavity C.
had changed frequency by 0.25 Me out of 9000 Me. This Tuning was accomplished in the usual way by moving
change was probably due to the relief of strains. It is ex¬ one end plate with a micrometer movement. The microm¬
pected that henceforth the accuracy will be a few parts per eter3 had a nonrotating spindle which by a small modifica¬
million, provided atmospheric corrections are made. tion could be keyed so as to be positively nonrotating. The
This paper summarizes the principles used in the design end plate in general is not quite normal to the micrometer
of these instruments. The chief features of an accurate
wavemeter are a fine micrometer for tuning, an accurate
calibration, a convenient and accurate calibration table,
a high Q, the temperature compensation scheme, and the
transmission or absorption coefficient. These will be
discussed.
Table I.
Freq. Diam la r
Cavity Me in. in. in. R db 0/.
A 8650-10 000 1.88 0.8-1.5 0.2101' 81% 12 19 000
B 8500-9500 1.79 1.17-2.17 0.201h 85% 16 23 000
Fig. 1. Invar wavemeter (8.6 to 10 kMc) with a high quality mi¬
C 2800-3200 6.42 2.5-3.5 •• ■ 83% 15 33 000
crometer to move end plate and an ordinary micrometer with needle
plunger for small frequency increments. Short waveguide sections
a Definitions: L is the length; 2a is the iris diameter; R is the ratio of the ob¬ were added later to provide isolation from variable screw tension
served unloaded Q to the theoretical Q expected from silver with a resistivity of during flange attachment.
1.63 X10-6 ohm cm; T is the transmission loss through the resonator. Ql is the
observed loaded Q.
b The thickness at the irises was 0.030 in. 21. G. Wilson, C. W. Schramm, and J. I1. Kinzer, Bell System
Tech. 1. 25, 408 (1946). They also give convenient tables and graphs
1 For another paper on accurate wavemeters see J. C. Van Den useful for cavity design.
Bosch and F. Bruin, Physica 19, 705 (1953). 3 Arizona Tool and Die Company, Tucson, Arizona.
492/410
M ICRO W A V E W A VEMETERS
493/411
H. E. BUSSEY AND A. J. ESTIN
be corrected for temperature and insulated to decrease the (d) deposit 0.007 in. of silver, and (e) remove about 0.005-
temperature changes. in. silver with sharp diamond tool in a lathe adjusted for
For absolute frequency determinations as in spectros¬ best possible finish. Steps d and e may, however, be re¬
copy a wavemeter should be sealed. Convenient seals for placed by simply depositing, say, 0.0004 in. of silver on a
50-ohm coaxial leads to a cavity were described by Bussey base that is as smooth as possible.
and Birnbaum,10 and seals for irises also were discussed. The Q also depends on the precision of construction, i.e.,
For observations on unsealed resonant cavities perturbed the geometrical perfection of the shape. For example, a
by dielectric or magnetic solid samples, however, the results tilted end plate or almost any distortion will couple the
may be more accurate if the wavemeter is left open since TEon mode to its companion TMm mode which would
pressure changes will affect both cavities in the same way lower the Q. The gap around the moving end plate may
and gradual humidity changes of the laboratory will diffuse split this degeneracy sufficiently, but this depends on the
into the two cavities, and thus compensate. Corrections for degree of imperfection present. A concave end plate can
atmospheric changes can be applied.11 introduce sufficient splitting to decouple the two modes
Recalibration of one of the Invar wavemeters after a even in severe cases.13,10 The Q did not improve when con¬
year indicated that the cavity dimensions had decreased cave end plates were tried, which proved that the construc¬
resulting in a frequency increase of 0.15 to 0.3 Me. The tion was sufficiently precise. While design with a concave
reason for this change is not definitely known. A possible end plate might reduce machining tolerances, some in¬
cause is dimensional changes due to creep. Invar for meter crease in the labor of fitting the calibration would occur
bars can exhibit a creep of 1 or more parts in 106/year.12 because the frequency formula could become more
The heat treatment may enhance this effect.8 The room complicated.
was air conditioned which eliminated ambient conditions
as a significant factor. V. COUPLING
There are other systematic errors. Small variable strains
The adjustable loops of cavity C provide a functional
are introduced by waveguide connections and by routine
design that allows the coupling to be adjusted easily to any
handling. The Invar cavities had 0.5 in. wall thickness to
desired value, which is thereafter left fixed. The coupling
reduce such effects, nevertheless, when waveguide flanges factor of an iris is given in reference 2. Starting with the
(perhaps distorted) were screwed very tightly to cavity A
iris sizes given in Table I for A or B, changes in coupling
(Fig. 1) a small frequency change was observed. Therefore
may be made on the basis that the power transmitted by
permanent waveguide sections were attached to cavities A
each iris is proportional to the sixth power of the iris
and B to provide mechanical isolation. After this no fre¬
diameter.
quency change was observed due to loosening or tightening
a flange. A bending moment on the waveguide of about VI. CALIBRATION TABLES
250 oz-in. changed the frequency by less than 0.01 Me.
The strains are not so variable with cavity C because the After constructing a wavemeter which could resolve over
connecting cables are flexible. Thinner walls, such as the its range some 105 individual frequencies it was, for reasons
of economy, only calibrated14 at 21 equally spaced fre¬
|-in. walls used, are then satisfactory.
quency points over the band. (The needle plunger was held
IV. Q FACTOR stationary for this calibration.) In order to use the instru¬
ment conveniently, however, it is necessary to construct a
The percentages of theoretical unloaded Q, 81 to 85%, calibration table containing, perhaps, some 104 points.
indicated in Table I, are fairly high. A few auxiliary ex¬ Such a table may be constructed easily with a high speed
periments at 9200 Me have shown that our particular silver
computer from any expression which fits the calibration
plating facility can furnish a deposit which attains up to
points sufficiently closely.
90% of the theoretical Q of bulk silver.
. The fitting of the calibration points may proceed as
The Q or conductivity of the silver depends upon the
follows: The observed frequency f at any micrometer
density of the deposit, which was beyond our control, and reading M may be considered to be made up of several
the smoothness of the finish, which was here obtained by
contributions as follows,
means of a sharp diamond tool. The machining and finish¬
ing proceeds approximately as follows: (a) rough out to f=Mb,L0,M)+XAf(M)+R(f)+S, (1)
within 0.01 in., (b) heat treat and anneal, (c) finish the where /o is the theoretical frequency associated with the
Invar by grinding and the brass with a diamond tool,
13 A. Banos, “Design of an improved X-band echo box,” MIT RL
10 H. E. Bussey and G. Birnbaum, Rev. Sci. Instr. 30, 800 (1959). Rept. 631, December 7, 1944. See also footnote 10 for experience with
nf\U\=fiU2 where / is frequency and u is refractive index, given, this method.
for example, in footnote 7. 14 Calibration was performed by the Electronic Calibration Center
12 B. L. Page, J. Research NBS, 54, 1 (1955). of the National Bureau of Standards, Boulder, Colorado.
494/412
MICROWAVE WAVEMETERS
MIC -MC/THOUS
dimensions, "52 Af represents the sum of all known pertur¬ 0 1 2 3
F R £ 0 U
4
E N C t
5
bations, R represents the residual deviations remaining • 675 .3293 2938.10 2938.06 2938.03 2938.00 2937.97 2937.r
• 676 .3290 2937.77 2937.73 2937.70 2937.67 2937.64 2937
after/0 and A/ have been fitted, and 5 represents stochastic • 677 • 3288 2937.44 2937.41 2937.37 2937.34 2937.31 29'
• 678 .3285 2937.11 2937.08 2937.04 2937.01 2936.98 2
deviations that remain unexplained. The theoretical ex¬ .679 • 3283 2936.78 2936.75 2936.72 2936.68 2936.65
pression for /o as a function of length L and radius b for a • 680 • 3280 2936.45 2936.42 2936.39 2936.35 2936»
• 681 .3277. 2936.12 2936. 09 2936.06 2936.03 293'
right circular cylinder is • 682 .3275 2935.80 2935.76 2935.73 2935.70 29'
• 683 • 3272 2935.47 2935.44 2935.40 2935.37 2‘
• 684 • 3270 2935.14 2935.11 2935.08 2935.04 2
(2tr fo/c)2 = 0mv/LY+ (rLm/bY, (2) • 665 .3267 2934.82 2934.78 2934.75 2934.72
• 686 • 3264 2934.49 2934.45 2934.42 2934.39
• 687 .3262 2934.16 2934.13 2934.10 2934.r
where c is the velocity of light, n is the number of half • 688 .3259 2933.84 2933.80 2933.77 293"*
>689 • 3257 2933.51 2933.48 2933.44 2r
wavelength along the axis, and rLm is 3.8317 for TE0in
• 690 .3254 2933.18 2933.15 2933
modes. The micrometer reading M enters from the fact • 691 • 3251 2 932•86 2932. 83 2°'
• 692 .3249 2932,53 2932.50
that L = L0+M where Z0 is the length when M is zero. .693 • 3246 2932.21 2932. I7
• 694 .3244 2931.88 2931-
The residuals for cavities A and B could not be fitted to
• 695 .3241 2931.56 29
high accuracy even with a high-degree polynomial when • 696
• 697
.3239 2931.23 7
• 3236 2930.91
perturbations were neglected. With perturbations taken • 696
• 699
.3233
.3231
2930,59
2930,2'
495/413
PROCEEDINGS OF THE IRE - Vo/. 47, p I,782 (Oct. J979)
496/1
AN EVALUATION OF A CESIUM
BEAM FREQUENCY STANDARD
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
National Bureau of Standards, Boulder, Colorado
497/127
R C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
line width of 300 cps (at 9192.631 Me), and provides a signal-
to-noise ratio in the range 100 to 400. The atomic beam is pro¬
duced by heating pure cesium metal to 150°C and allowing it to
effuse from the oven through a channel 0.038 inch long with a
cross section of 0.003 x 0.100 inch2. The beam is detected by
a surface ionization detector. The hot wire is made of a plati¬
num-iridium alloy (80 percent Pt; 20 percent Ir). The ion cur¬
rent is measured with an electrometer. Typical values of the
undeflected beam current fall in the range 1 to 3 x 10“11 amperes.
The beam excitation is derived from a quartz crystal oscillator
(10.317... Me). The output of this oscillator is multiplied in
frequency up to the cesium transition by the scheme shown sche¬
matically in Fig. 1. The exciting radiation is swept in frequency
over the width of the spectral line by tuning the 10.317... Me
crystal oscillator by means of voltage sensitive capacitors in
the crystal circuit. The detected beam signal is applied to the
y-axis and the analog output of a frequency counter is applied
to the a>axis of a x-y plotter. A Ramsey line shape plotted in
this manner is shown in Fig. 2. The frequency scale on the x-
axis is linear. This allows the position of the peak of the line
to be determined by taking the average of the frequency of two
points on opposite sides of the central peak, both points having
the same value of y. Several averages of this sort are made for
each line trace. The line is swept in both directions and a linear
interpolation is made. This compensates for delays that occur
in the detecting circuit.
The Ramsey excitation structure consists of a long electro-
formed resonant cavity of rectangular cross section bent into
the shape of a U. The Q of this cavity is about 6000. It is sym¬
metric about the coupling iris and operates in the TEj ^ mode.
The beam passes through the two ends of the cavity—which are
separated by 56 cm—and just grazes the end surfaces. Fre¬
quency shifts incurred through frequency “pulling” of the cavity
are given approximately by
where SvR is the shift in the peak of the atomic resonance re-
498/128
TO EXCITATION STRUCTURE
A CESIUM BEAM FREQUENCY STANDARD
499/129
* C^>
*-1
Block diagram of frequency measuring system
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
500/130
A CESIUM BEAM FREQUENCY STANDARD
ACCURACY
The accuracy of the machine depends upon the precision to
which certain parameters can be controlled (and/or measured).
The uncertainty in the measured magnitude of the“C” field is
+ 0.003 oersted. This corresponds to an uncertainty in the micro-
wave frequency measurements of 2 x 10~n—well within the pre¬
cision.
There is an uncertainty introduced in the measured frequency
if the phase relation between the two oscillating fields of the
Ramsey exciting structure is not precisely known. In the ma¬
chine described here, a single resonant cavity is used. Con¬
sequently, the phases are precisely the same at the two cavity
ends. The beam passes through these two regions of identical
phase. As further evidence of phase identity, the cavity can be
rotated 180° (except for the shorted ends) and the two measured
transition frequencies compared. A lack of phase identity in the
two oscillating field regions can also be detected by observing
the symmetry of the line shape. The absence of perfect line
symmetry implies unequal phases in the two regions. The degree
to which the inaccuracy can be determined by this kind of ob¬
servation is limited by the signal-to-noise ratio to a greater ex¬
tent than by comparing the frequencies for one orientation of the
waveguide structure and the inverted orientation. However, if
the line breadth is 300 cps and the signal-to-noise ratio is 400,
this method permits the inaccuracy in the frequency to be speci¬
fied within about 1.4 x 10~10. No asymmetry is observable in
the line traces.
Any frequency shift caused by electric fields has been shown
by Haum and Zacharias(1) to be negligible for magnitudes of
the electric field intensity that would be expected in the “C"
field region. They have found that the frequency shift in the
cesium line is
501/131
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
in cesium for which this selection rule applies. It can be shown
that these are sufficiently well resolved at .080 oersted so that
the resonant peak of interest is not shifted beyond the present
uncertainties in measurement by the overlapping of neighboring
lines.
Uncertainties can be introduced by the exciting radiation.
If this radiation consists of more than one signal and if these
signals are not symmetrical in frequency and amplitude about
a frequency that is an integer multiple of the primary crystal
oscillator driving the multiplier chain, then erroneous line
frequencies will obtain.(2) It is not unusual to find unsym-
metric power spectra of this sort from frequency multiplier
chains and large errors could result. The square root of the
power spectrum of an oscillator identical to that used in NBS-1
is shown in Fig. 3. Notice that the spectrum has no sidebands
and any asymmetry of the central peak will be of no consequence
502/132
A CESIUM BEAM FREQUENCY STANDARD
since this peak is much narrower than the spectral line breadth.
Other tests were made on the NBS-1 oscillator to show that its
spectrum was similar to that of its twin (Fig. 3)—in particular,
no low-frequency sidebands were detected in the multiplier chain
output. This spectral character will provide accurate frequency
comparisons. The usual assumption that the median frequency
of a frequency multiplier is an exact integer multiple of the pri¬
mary oscillator is an accurate assumption for the sort of spec¬
trum displayed in Fig. 3.
Considering the various sources of inaccuracy and the results
of their study, we believe—with some confidence—that the ac¬
curacy of the machine falls within the precision or + 8.5 x 10-11.
THE MEASUREMENTS
The reproducibility of the cesium standard (NBS-1) was de¬
termined by comparison with a 10 Me quartz crystal oscillator in
which the quartz crystal and associated crystal circuit are im¬
mersed in liquid helium. The stability of this oscillator over
periods of several days is 3 to 5 x 10-n*—a figure determined
from a continuous comparison with an Atomichron (106) during
periods of good behavior. This Atomichron, even though large
excursions in frequency occur in short time intervals (Fig. 4b),
has excellent stability when averaged over time intervals of 1
minute or more. Periods during which the helium-cooled oscil¬
lator and Atomichron (106) showed no significant variations rel¬
ative to each other were considered best to determine a repro¬
ducibility figure (precision) for NBS-1.
The helium oscillator has also been demonstrated to be stable
to 2-4 x 10-11 over periods of ~ sec to 6 hours by comparison
with a maser stabilized frequency multiplier chain (Fig. 4a).
In handling the data, the mean of a single day’s measurements
is considered one piece of data. An estimate of the standard
deviation was obtained from successive differences. Let
L—i n
i= 1
503/133
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
Fig. 4,.
(a) Maser stabilized chain compared with 10 Me helium-cooled
crystal oscillator. The recorder plots the analog output of a fre¬
quency counter versus time. The counting period is 1 sec and the
display time is 1 sec.
(b) Maser stabilized chain compared with 5 Me Atomichron crystal
oscillator. In this trace, the difference signal is measured with
a frequency meter and not with a counter. The time constant for
the meter is about ^ sec. Note the change in frequency scale be¬
tween (a) and (b).
where x. is the frequency measurement for the ith day and xi+x
is the frequency measurement for the (i + 1) day; n is the total
number of data points (or days). The estimate of the standard
deviation by successive differences is given by
(s) - Vr
For our set of data 8 = 1.1 cps and a(8) = 0.77 cps which is
8.5 x 10-11 of the cesium frequency. Successive differences are
likely to be random even thorugh certain systematic changes may
occur. These systematic changes are small when data is taken
on succeeding days but nevertheless still exist. It is impossible
to attribute these variations to the Atomichron, the helium-cooled
oscillator, or NBS-1, or to a combination when the variations
are small. Actually the standard deviation estimated from suc-
f *R. H. Kent and J. VonNeumann have shown that ct2(S) = (a')2 where
o is the true standard deviation.
504/134
A CESIUM BEAM FREQUENCY STANDARD
cessive differences is almost the same as that estimated in the
usual way. In fact, this standard deviation,
is .5 cps or 8 x 10-11.
The data show Atomichron (106) to be lower than NBS-1 by
3.6 cps for the week of August 30. The mean of all the data
taken over the months of July and August show about the same
difference. Data of one day was deleted from the average be¬
cause the Atomichron (106) made a large excursion, 8 x 10-10,
just prior to a general retuning. The excursion was verified by
NBS-1, the helium-cooled oscillator, the masers, and Cruft Lab¬
oratory data.
In addition to the tests carried on at the Boulder Laboratories
to determine precision and accuracy, frequency comparisons
were made between NBS-1 and other cesium standards in the
United States and England. The comparisons were made through
propagated signals between the different locations.
In the discussion that follows, the following designations will
be used to identify the various cesium standards:
D esignation Location
572772 0-61-33
505/135
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
It should be emphasized that this comparison assumed that 106
had the same mean frequency during July and August 1959 as it
had during the 6-month period September, 1958, to March, 1959.
The accumulated data over the past two years indicate that the
Atomichrons tend to wander in frequency. In fact, Atomichrons
112, 106, 110 were in rather good agreement during March 1959—
differing by probably less than 2 x 10“10. This comparison was
made by a single Atomichron transported by air to each of the
locations to provide a more direct comparison between the dif¬
ferent units.* The relative frequencies of Atomichrons 112, 106,
110, 109 have changed considerably between March and July ac¬
cording to the propagation data of July. During the month of
July, 1959, 112 differed from 106 by about 1 x 10“9 and it dif¬
fered from 110 by several parts in 1010. Other variations of this
nature are evident from the various data including the direct
comparison data made at Boulder between 106, NBS-1, and the
helium-cooled oscillator.
*The experiment was performed under the auspices of the U.S. Sig¬
nal Corps by Dr. J. H. Holloway from the National Company.
506/136
A CESIUM BEAM FREQUENCY STANDARD
sary to investigate the power spectra of the radiation from fre¬
quency multiplier chains used in such measurements. This is
especially important in atomic frequency standards where a
quartz oscillator is compared with an atomic resonance through
a frequency multiplication process. The multiplier chain meas¬
ured output frequency would be some sort of average—depending
upon the method of measurement. If the power spectrum were
unsymmetric, this average would not be an exact integral multi¬
ple of the primary crystal oscillator frequency.
In the method of observing the power spectrum used here, the
output signal of the multiplier chain to be investigated is mixed
with an essentially monochromatic signal. This relatively pure
signal is obtained from an ammonia maser stabilized chain (Fig.
11). In this fashion the power spectrum is shifted to low fre¬
quencies—in fact, audio frequencies—where it can be conven¬
iently examined with a variable frequency narrow band filter
507/137
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
508/138
A CESIUM BEAM FREQUENCY STANDARD
an amplitude-modulated signal to be symmetric in their amplitude
about the central peak. This is not necessarily true of frequency-
modulated signals—provided that the primary signal is modulated
with two or more modulating signals. Thus unsymmetric power
spectra are expected—under certain circumstances—when fre¬
quency modulation occurs in frequency multipliers. Suppose that
the output frequency of such a frequency-modulated chain were
measured. Suppose this is done by beating it with a known and
relatively monochromatic signal, and the beat note is measured
with a counter. The counter would measure the center of gravity
of the power spectrum and this frequency will not be an integer
multiple of the basic quartz oscillator (unless the power spec¬
trum is symmetric). Furthermore, if this radiation were used to
excite the atomic transition in an atomic beam frequency stand¬
ard, the measured frequency of the spectral line would be dif¬
ferent than that measured if the multiplier chain output were
monochromatic. In general, the frequency measured by the spec¬
tral line will be different than that measured by the counter under
509/139
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
the unfavorable—but not unusual—conditions discussed above.
Elimination of the sidebands and the reduction in the frequency
multiplying factor is the best cure—and also a possible cure for
these difficulties.* If the power spectrum is symmetric, regard¬
less of whether the frequency is measured by a spectral line or
a counter, the measured frequencies will be the same so long
as the receiver amplifier is not extremely narrow banded.
*See Fig. 3.
510/140
A CESIUM BEAM FREQUENCY STANDARD
have a maximum deviation in frequency of about 2 x 10“11 over
a six-hour interval (by comparison with the maser stabilized
chain) during periods of a fixed power spectrum. Variations of
this sort in the power spectrum would undoubtedly produce a
corresponding change in the “fixed” frequency of an atomic fre¬
quency standard.
511/141
R. C. MOCK.LER, R E. BEEHLER, AND J. A. BARNES
512/142
A CESIUM BEAM FREQUENCY STANDARD
513/143
Fig. 11. Diagram of the scheme used to plot power spectra
R. C. MOCKLER, R. E. BEEHLER, AND J. A. BARNES
helium-cooled crystal oscillator used in the frequency compari¬
son with NBS-1. The spectrum is quite complex and is also
unsymmetric. It provides a superb signal for frequency com¬
parisons (Fig. 4a) but would not be useful as a signal source to
induce the atomic transition.(4) The square root of the power
spectrum of the 5 Me oscillator belonging to the NBS Atom-
ichron (106) is shown in Fig. 10. The spectrum is symmetric.
In spite of its complex character it provides a uniform averaged
frequency over long periods of time.
In establishing the performance of an atomic standard it is
important to know the spectral character of the excitation. A
clean spectrum is likely to give the most reliable behavior.
CONCLUSION
The agreement between the NBS cesium atomic standard fre¬
quency and a mean frequency of four Atomichrons (M4) is rather
good—within ±1 x 10-10. This, agreement may be fortuitous,
however, because there is considerable disagreement among the
Atomichrons used in determining the mean frequency M-4. The
comparison between the NBS atomic standard and that of England
[NPL-(NBS-l) = +1.8] may be more meaningful.
The variations between atomic standards seem to be quite
real and remain unexplained. The source of disagreement can
only be determined by test experiments on the individual ma¬
chines.
At the present time a new cesium resonator is being put into
operation at the Boulder Laboratories. This machine will have
a line breadth of 90-100 cps.
514/144
A CESIUM BEAM FREQUENCY STANDARD
REFERENCES
515/145
PRECISE TIME SYNCHRONIZATION
SUMMARY
astronomy, etc. , which are being intensely investigated, the needs for
the need has been greatly increased for more precise determination of
group of widely separated clocks to precisely the same time and keeping
gation theory pertinent to two of the methods and a few sets of measure¬
WWV to WWVH are given. Several graphs and tables are included to
516/iv
Precise Time Synchronization
Alvin H. Morgan
I. INTRODUCTION
rate clocks, which are widely spaced from each other, has increased
etc. Only a few years ago the highest precision required was of the
required.
At the present time quartz crystal clocks are the best working
able to run reliably for long periods of time. However, they are in¬
spaced clocks to the same time and keep them in very close agreement,
other.
517/1
II. METHODS OF SYNCHRONIZING
when they are widely separated. However, most of these methods are
clock around to each of the slave clocks, and set each one to it, in
timing signal from the master clock to each slave clock and they in
turn would immediately retransmit the timing signal back to the master
clock. This would enable one to determine very accurately the signal
propagation time and the great circle distance to each clock, without
exact longitude and latitude and that of the master clock, and the calcu¬
lation of the great circle distance to each from the master clock. Using
this information calculations are made of the delay time of the signal
from the master to each slave clock, which is used to determine their
corrections. This system may employ either ground wave or sky wave
518/2
III. TRANSPORTING A MASTER
a. General
time the master clock is initially calibrated and the time it is used to
correct the time of each of the slave clocks. Unless the master clock
keeps time over long intervals to a precision better than that desired
this, the initial rate of the master clock must be known to a high preci¬
interval, i. e. , from the setting of the first to the last slave clock.
b. Clock Errors
^ + AP (1)
519/3
OC = accel. of clock, psec/day/day.
clock, both the r and oc terms would (so far as is known now) be equal
o
to zero, regardless of the length of the time interval.
first and the last slave clocks without the necessity of making any time
T = A + P + r P + 4- (2)
o o Z o
which is usually negative, may be neglected in all but very, very long
520/4
IV. TWO-WAY TRANSMISSION OF TIME SIGNALS
a. HF Timing Signals
First, because time signals are transmitted both ways over the given
without prior knowledge of the precise length of the path. Second, from
At - 6t
t = (3)
P
where: 5t is the delay time a time pulse undergoes from its reception at
the slave clock until its retransmission, and At is the delay time,
return to the master clock. From this, the path length is, obviously:
d = t c (4)
P P
from the master to the slave clock, and c is the velocity of light. Also,
it may be easily shown, Equation (13), Appendix B, that the great circle
d = (5)
g
7 V
sin <9/^'W(i
e/2 (1 + 7)
Z s in 9 / 2 J
572772 0 - 61 - 34
521/5
where: the symbols are as defined in Appendix B. Obviously, it is not
below.
522/6
V. ONE-WAY TRANSMISSION OF TIME SIGNALS
use sharp pulses and, therefore, have very good time resolution. Thes
but they can be used to obtain very stable timing signals. By gating to
accept the ground wave part of the signal at the receiving end, it is
precisely, the distances from the master to each slave clock. This
agation delay time for a given great circle distance, a chart is included
c. HF Propagation
a precise knowledge of the signal path lengths between the master and
each slave clock, which may be calculated from other known data. The
longitude and latitude at the master and each slave clock must be accu¬
rately known along with the average ionospheric layer height at the
523/7
time of day at which the calculation is to be made. In addition, the
mode of the signal being propagated between the master and each slave
clock is required.
distance between the two points on the earth for which the signal prop¬
rather simple they are quite lengthy, and are best explained by means
t = d / c (6)
g g/
where: d
is the great circle distance and c is the velocity of the
g
ground wave and may be taken equal to that of light. The next step is
sky wave signal and a ground wave signal over the path in question. In
At = (1+7) + (?)
2 sin 0/2
d
£
where: 0 = and y = h'/r. This may be further simplified (e. g. ,
Appendix B) to:
At = (8)
524/8
sirt 0/2 7
where: , h = (1 f 7) and p
0/2 2 sin 0/2 J
Either equation (7) or (8) would normally be evaluated for the given path
in terms of various ionospheric layer heights, h®, and angles (0/2), and
directly. Each chart has a given fixed layer height, h®, and the curves
in Figs. 10-15, for a given path and mode, it is added to the ground
(a) long term stability or constancy of the master clock, and (b) reso-
the error due to the long term clock stability is negligible. In case the
good: from 0.001 psec to 0.01 psec is not too difficult to achieve with
525/9
system used. With care and very high quality apparatus and rapid
(1) General
above, but they are obviously more convenient, especially when the
nature of the problems, the hi and the ground wave timing signals will
be considered separately.
observer: (a) time of day that the measurements are made, (b) method
or a’ peak of the signal is used, (c) the extent to which the antenna
the propagation mode with the least delay time is used, (e) whether or
not the measurements are related to a precise slave clock at the re¬
ceiving site so that advantage may be taken of the average (over several
days), (f) the actual frequency used, i. e. , in general, the higher the
frequency (but still below the MUF) the more stable the delay time, and
Other factors which may affect the stability of the relative delay
times of the signals are: (h) epoch of the sunspot cycle; (i) whether or
not the radio path passes near or through one of the auroral zones,
(j) whether or not the path is all in darkness or all in daylight at the
526/10
time of the measurement, (k) signal-to~noise ratio at the receiving
site, and (1) number of hops involved in the propagation of the signal.
(a) time of day for observation near noon or midnight at the center
of the path.
(b) part of signal to measure zero crossing time of the least de¬
layed pulses-
(d) propagation mode to use the one with Jeast, consistent delay
time.
(i) location of radio path well away from both auroral zones
this method obviously then depends on how near to ideal conditions the
accuracy may be better than ±0.1 millisecond, even for very long paths;
more.
527/11
At the other extreme, for example if the path passed near or
through an auroral zone, or was so long that it was never in all dark¬
ness or all daylight, etc. , the accuracy might be only about ± 10 milli¬
seconds, or worse.
time, but it is hoped that further measurements in the future will make
this possible.
The method of calculating the path delay time from the measure¬
quite apparent that the actual conditions of the measurements will not
quite good results. This is also apparent from the data given in Appen¬
dix D.
(a) One-Way
528/12
synchronization of widely separated clocks. Within the distance range
mine the great circle path length between the master and slave clock to
to receive only the ground-wave component of the timing pulses; (Z) pre¬
cise measurements of the signal delay in the transponder; and (3) use of
a. Introduction
the ionospheric layers and the surface of the earth. One reflection
and so forth.
529/13
Only two of the reflecting layers will be considered here, the
F-2 and E-layers, because other layers (F~l and Es) may usually be
periods of time. Signals reflected by the F-2 or E-layer are called F-2
c. One-Hop Modes
or E-layer. Remembering that both the earth's surface and the iono¬
sphere are curved, it is quite clear that there will be a limiting dis¬
tance for a one-hop mode. For frequencies equal to, or lower than,
the MUF (explained above) for the path, the limiting distance may range
from 3500 to 5000 km (2200 to 3000 mi. ) for the F-2 layer mode, de¬
distance of about 2400 km (1500 mi. ) for the E-layer mode. However,
it has been found in practice that the average limiting distance for the
F-2 layer mode is about 4000 km (2500 mi. ), which will be the value
used here.
Thus, for distances less than 4000 km (2500 mi. ) but greater
than 2400 km (1500 mi. ), the F-2 layer single-hop mode would be pre¬
could exist, if the frequency used were equal to, or below, the MUF
530/14
d. Multihop Modes
For great circle paths longer than about 4000 km, it is obvious
that more than one reflection from the ionospheric layer usually must
greater than the quotient of the total great circle distance expressed in
time the signal frequency is below the MUF for the path.
It is quite obvious that two hops of maximum limit (4000 km each), laid
end-to-end, would not cover this distance. Therefore, the simple geo¬
Over very long paths, the MUF for the signal may be estimated
it point "a", and the other a distance nd" from the receiver, which will
be called point nk,(. The distance "d" is one-half the limiting distance
for a one-hop mode (i. e. , one-half of 4000 km), or 2000 km (1250 mi. )
for F-2 propagation. The MUF for the path is taken to be the lower of
and seasonal variations are associated with the movement of the earth
in its orbit around the sun. In addition, there are longer period vari¬
ations that have been correlated quite well with the sunspot cycle.
531/15
VIII. VARIATIONS IN PROPAGATION DELAY OF HF SIGNALS
a. Introduction
mine the maximum variations that are likely to occur at the given time
its relatively complex nature. Among the many factors affecting the
This is the variation in the virtual height of the F-2 ionospheric layer.
Besides its quite regular diurnal and seasonal variations, the F-2 layer
ances. The ion density of the F-2 layer usually decreases and then
affect high precision work. Data taken during such periods are, ob¬
here, and should be discarded; only data taken during normal conditions
should be used.
layer may be minimized by: (a) measuring the timing signals at the
same time every day, and (b) selecting this time of day so that the
whole path, from the master clock to the slave clock, is all in daylight
532/16
or all in darkness. In other words, neither the sunrise nor sunset
period should occur anywhere over the path during the time the measure¬
ments are being made. On fairly short paths, for instance, the meas¬
the ionospheric height versus time at the reflecting points of the given
in timing problems.
virtual height on the propagation delay time, for a given path, might be
Fig. 5 was devised. This is a plot of virtual height, h', (in km) versus
along with a third parameter which is the great circle path length, d
&
divided by n. To simplify its use, only five discrete path lengths are
plotted; 500, 1000, 1500, 2000 and 3000 km; the latter curve being
agation delay time (At) and the nominal virtual height (h1 ) for the path.
For example, say h1 is 300 km, the desired propagation delay time
stability (At/n) is 10 psec. , the great circle path length (d /n) is 3000
©
km and n = 1 (one-hop mode) (see Fig. 5). Then from the curve, the
533/17
IX. CHART FOR DETERMINING TOTAL
KF PROPAGATION DELAY
a. Introduction
in Fig. 10 through Fig. 15 along with the chart in Fig. 1 is not necessary,
b. Use of Chart
This chart gives the total propagation delay time, per hop, for a
given path. That is, the delay time in Fig. 6 includes the ground wave-
delay titne in Fig. 1 plus the additional delay time in Figs. 10-15 caused
time was that of one hop multiplied by the number of hops. Thus, the
then assuming a three-hop path, (d^/n) would be 3300 km, and the path
delay time, as read on the chart, assuming a layer height of 350 km,
a. Introduction
Until very recently the only unit of time available was that de¬
rived from the movements of the earth and it was called solar time.
day" throughout the year, the mean solar day was used as the time
standard; it is the mean of all the solar days throughout a given solar
year.
534/18
In 1955 a more uniform astronomical time was adopted as the
by the orbital motion of the earth about the sun but in practice it is
determined from the orbital motion of the moon about the earth.
forces. Several standards are now in use employing either the cesium
The unit of mean solar time is the mean solar second and it is
A mean solar day is defined as the mean of all the solar days in
a solar year; there are 365 solar days in one solar year.
(a) UT-0
(b) UT- 1
(c) UT-2
, 8
c. Ephemeris Time
The Ephemeris Second is defined as the fraction 1/31, 556, 925. 9747
of the tropical year for 1900 January 0 at 1200 hours Ephemeris Time.
535/19
d. Atomic Time
frequency would keep "atomic time", its main merit being its uniformity.
WWVH, Maui, Hawaii), JJY (Tokyo, Japan), and MSF (Rugby, England).
any timing problems and then when necessary switch to another one
tracking and timing problem arises wherein, say, the time signals from
WWVH are used initially and then later a switch is made to the JJY
XII. ACKNOWLEDGMENTS
The data- used in plotting the curves in Fig. 10 to Fig. 15, inclu¬
sive, was obtained from Mr. A. G. Jean, who kindly consented to their
use in this paper. Also, many helpful discussions were held with
Mr. T. N. Gautier.
536/20
XIII. REFERENCES AND BIBLIOGRAPHY
572772 0 - 61 - 35
537/21
XIV. APPENDICES
APPENDIX A
4
Great Circle Distance (Z) Calculations
a. Equations Used
cot (c/Z)
y + x
tan cot (c/2)
2
y + x
sin
Z
tan tan
y - x
sin
2
538/23
/ L« - L \ / JL + L
. c ( B A ] , ( B A
b. Calculation oi ---J and \---
L - L L + La
B A r L + L B A
lb - la v B A
Sites
B , A 4° 02' 49, 97" 2° 01* 24, 98" 73° 56' 16.35" 36° 58' 08. 17"
2° 01.416* 36° 58. 14'
/ L + LA
( --—— ) = 36° 58. 14* 0.79896 0. 60138
y - x
= tan” 1 (0. 11799) = 6° 43. 76* (6)
539/24
•B
0. 9755Z
tan 0. 03533 = 0.2942 (11)
0. 11703
- 1
tan (0. 29429) = 16° 23. 6' (12)
540/25
APPENDIX B
1. Definitions (Figure 8)
Y
d® = 4r sin (9/2) s] (1 + y) + ( —
p
n/71
. .
V 2 sm 0/2
(3)
d®
f = -e =
p
P cc
iLjom Vd + 7) a 2
c
Y
sin 0/2
(4)
Now, the ground wave propagation delay time for the great circle
d®
_g
t® (5)
g c
541/27
Now, from trigonometry, we know that:
d» = 2r 0 (6)
g
so,
t- = 212 (7)
g c
t'
P
sin 0/2
0/2 "
J (1 + 7) + 2 sin 0/2
7
(8)
great circle distance delay time. For more than one hop, say n-hops
/ sin 0/2 7
At = 4 (1 + 7) + -1 (9)
lg\ e/2 2 sin 0/2
If we let:
(1 + 7) = h, (2 sin 0/2
j / sin 0/2 \
= p- and ( -072“ ) = y’
(10)
At=t
8
( y vhl pu -1 (id
542/28
4. Great Circle Distance Determined from Path Distance
t (0/2)
P
t (12)
g
sin y
(1 + 7) +
sin 0/2
TABLE I
(See Fig. 2)
d
g versus
n
444.4 1 0. 01745
888. 9 2 0.03491
1333.3 3 0.05236
1777.8 4 0.06981
2222.2 5 0.08727
2666. 7 6 0. 10472
3111. 5 7 0. 12217
3555. 5 8 0. 13963
4000.0 9 0. 15708
543/29
TABLE II
(y versus 0/2)
TABLE III
(See Fig. 3)
(h versus h1)
h*
X h where:
544/30
APPENDIX C
Now
t + = t + t
o o p
and:
t = t, + 6t = t + 2+ + 6t
2 1 op
so:
At = t_ - t = t + 2t + 6t - t = 2t + 6t
-5 o o p o p
Therefore:
t
P ( )
a nd:
d = ct
P P
545/32
APPENDIX D
(a) Acknowledgments
6
The measurements of the propagation delay time of the hf signals
from WWV to WWVH and return, and at Boulder, Colorado, were made
tion.
ave. 27. 16 0. 19
Path
WWV to WWVH 5 - -
(or vice versa) 10 Z7. 24 0. 20
(at 10 Me) 15 - -
27. 24 0. 20
547/35
(5) Calculations of Actual Propagation Delay Times,
Based on Above Measurements, at 10 Me
t , ms
Path Freq. , Me P
WWV to Boulder 10 8. 14
WWVH to Boulder 10 19. 10
WWV to WWVH 10 27. 24
d t
g t P
Path km g ms n
Meas. Calc.
Difference
t , ms t , ms
Path P P ms n
548/36
Max. dev.
Ave. Min.
UT No. from ave. ,
t t , ms
Fig. Date T ime Observ. p, ms ms P <r
Ave. of Min.
Max. Probabl e Ave. Probable
ave. t , ms t , ms
P cr, ms P n V h«. km
549/37
(2) Average, Actual Propagation Delay Times (June 20)
Ave.
UT Ave.
t j ms
Pa th Time P cr, ms
Ave.
UT Ave.
t , ms
Path Time P cr, ms
(3 hour)
(3 hour)
(3 hour) ave.path
Ave. t , ms
Day Path P Ave. cr, ms length, km
550/38
(6) Conclusions from Above and Figure 16
Ave. Ave.
Ave. Probable Approx,
t , ms d , km
P c, ms n P h', km
Min. Ave. _ .
Prob. Approx, Prob. Approx,
t , ms t , ms
Month P n h', km p n h*, km
551/39
b) Summary of Data (Over a 9-Month Period)
552/40
continuous
second
4-
+. 1 micro¬
Vert. dip.
2.5; 5; 10;
35= =E O
10x10"y
15; 20; 25
CD
WASHINGTON
CD lo
co O </>—«.
CD -— C3 QOO
CD LO 1 to to -»• CD ax-**
U.S.A.
S CNJ ^ "o i/)
01
o O to CD cr- r— <*> e
CD CO • CO UD
+
oo r— to >N
+
CD CD
CO CD CNJ
U_J
e 4
»— CO
0.02
None
None
33 CO
•— -<r cnj to CD
Vert.
UJ
i — - r- CNJ • CD
d rS o o CNl CNJ a
CD <o CD -4- o 3T 1
LO Z2Z +1
=> CO
MAIN CHARACTERISTICS OF STANDARD FREQUENCY AND TIME SIGNAL STATIONS
CD
Z LxJ CO 1 “O
5 in each
CNl 07 O C
>% u. • co o 1 u o
By steps
CO
CO CD CO • CL
- - CD 03 CD O O
qc ro • a>—v .— •—
20
CD <3- CD CO LO CD CD CD
O "O - CD -4* CNl x e to
J— — O o « CO ,- CD
to c— u CO .- CNl .—
10
CD
o +1
nr to +1 +1
r- 07
CNJ CNJ
to
]000
ss UJ
4 in each 5
By steps of
10 ms
to LO tO
continuous
CNl 1 “O
>- CNJ CO CNJ 07 O C
CNl -- “5 r -4- 1 U o
-4" CO *”D - CNJ CNJ CD CD O O
Tokyo
Jaoan
• T to CNJ —— <D
(30);
O o 1^- CNJ OO x e
+.20
u>
uo cr» L_ '’ CNJ CD
co co CD to- CNJ ,—
> • <o
CNJ .—
1
+1 +1
4-
ac 3= crT o
-o
5 in each
o
5 in each
i
CL CNJ 07 t_ d wco
CVJ .— CNl 1 O O QCNJ
MSF
LO
England
15
15
lO CD • CD
CNJ
+\ -^1 CO to
07
to
5 in each
U.S,S,R,
30 (19)
CL to r— CO
MOSKVA
Osl to CNJ CD
t 1 1 8 ~a — a CNJ 1 1
• CD LO
u
+i
o
or
CO UJ
JOHANNESBURG | LOWER HUTT
South Africa N, Zealand
co
<j- to
to o»
None
i— to
0.03
None
ZLFS
100
1 CO •
Nil
— - CNJ o
O o 1
»— -4* 2T 2=
-4- e*-
+.
p. (9>
H—
continuous
CO UJ 9 “O O
CD CD O C
I t- o CO CNl
.— CD CD CD O O n n—
onz
0.1
nor.di
a-
— CD -4" LO CNJ -- • — 03 03
— — CNJ
~
x e to to
O o Z2Z
+1 CD co e
to OO CNJ
Cvl cvi
H
+1 +1
^8
By steps of
ISLANDS, USA
'M.
Vert. dip.
continuous
second
1 micro¬
lOxlO'9
DC to LO CD _C
HAWAIIAN
to CO > --r o
CO <r 03
20 ms °
o- CNl 3C CNJ CO CO r-
3C CNJ <D
O o eT • CD
LO
01 +
CD tO
OJ tO oo s
+.
oo in
+
LO
BUENOS AIRES
second
+. 1 micro¬
CD LO
4- in each
+. 20xl0'9
CO 3E ,- CNl CD -C
Arqentine
o X
CNl •4* TO to o
r— »— LO CD CD 03 £= *-
LOL
O o LO • *-
-O' CO • to
ro to CNJ «- -4* LD
-+I
(minutes)__J_
transmissions (minutes)
Duration of time signal
in parts in 10y
Type of antenna
Latitude and
adjustment
Call Signs
Lonaitude
peso
SU0ISS |1U seiouenbejj
used
-SUBJ j pjepun$
CO -4* CO CO t— CO CD CD •— CNJ CO^TLO tO
572772 0-61-36
553/42
DAILY SCHEDULES
1. Weekdays.
9. Maximum radiation N - S.
11. Pulses of 10 cycles of 1000 c/s tone; the first pulse of each min¬
ute is prolonged (500 ms).
13. Tuesdays.
18. Signals A1 keyed. Duration of each signal 100 ms; the first signal
of each minute is prolonged.
21. Pulses of 5 cycles of 1000 c/s tone; the first pulse of each minute
is prolonged (100 ms).
23. Adjustments are made the first day of the month, when necessary.
554/43
24. Two half-wave dipoles on 1 5 Mc/s; one half-dipole on 2. 5 and
5 Mc/s.
33. Pulses of 5 cycles of 1000 c/s tone; the first pulse of each minute
is repeated 7 times at intervals of 10 ms.
36. Pulses of 5 cycles of 1000 c/s tone; no 59th pulse of each minute.
The first pulse of each minute is repeated 100 ms later.
555/44
soNOoasmiw
‘6t
556/45
awn Nouvovdoad
o
o
o
o
o
o
o
o
O
9 0.15708
8 0.13963
7 0.12217
6 0.10472
RADIANS
5 0.08727
0/2,
4 0.06981
3 0.05236
2 0.03491
I 0.01745
0 0.0000
1000 2000 3000 4000
dg /n, Km
557/46
PLOT OF TABLE HI
FIG. 3
558/47
FIG. 4 ILLUSTRATION OF DELAY TIMES,
WWV AND WWVH
Then:
(1) <T. - T + T ) = T ,
1 H 2 v
and
(2) T + T = T T
1 2 v H
therefore, T + T
Tl + T2 v H
(3) T =
P
559/48
( ji SECONDS)
FIG. 5
<
O
400
O O o o
m
ro
O m o in
ro OJ cv
m)
560/49
561/50
SKY WAVE PROPAGATION TIME
N
562/51
FIG.8 GEOMETRY OF THREE-HOP MODE
563/52
FIG.9 GEOMETRY OF n-HOP MODE
564/3
APPENDIX IV
CURVES OF FIXED h': WITH At(/lSEC) VERSUS DISTANCE dg (KM)
10,000
9000
8000
7000
6000
5000
4000
3000
2000
1000
4000
565/54
4000
566/55
1000 2000 3000 4000 5000 6000 7000 8000 9000 10,000
4000
567/56
FIG. 12
4000
568/57
o
o
o
FIG. 14
4000
572772 0-61-37
569/58
4000
570/59
FIG. 15
dp , KILOMETERS
KILOMETERS
tp, MILLISECONDS
350
300
250
200
150
tp , MILLISECONDS
572/61
dp , KILOMETERS
5100 5400 5700 6000
573/62
574/63
FIG. 19 SECOND TO SECOND MEASUREMENT OF DELAY TIME WWV TO WWVH
I
W 6Q 48 01
w 8S M 8
uu 65 US 01
575/64
W8Q 45
UJ6Q M2 01
W8S M2
SECONDS
CO
ID
CD
ID
<r
WWV TO WWVH
oo
LO
cn
PATH
cr
CL
c
>-
or
<z
MEASURED
ID
cr
co
>-
cr
FIG. 21
<c
<c
cn co r-~ co
C\J CVJ C\J C\J
SQN003SmiW
576/65
ERRATA SHEET
for
3. Page 13. Line 4, paragraph (b): should read *'100 feet" not
M10 feet."
4. Page 23. Last line: should be "Angle c" not "Angle C."
6. Page 30. Table III, right column: should read "h = (1 + y)."
577/66
Reprinted from the PROCEEDINGS OF THE I.R.E.
VOL. 41, NO. 1, JANUARY, 1953
578/104
1953 Ellenwood and Ryan: UHF and Microwave Matching Termination
which are moved back and forth in a waveguide or dielectric load but also the complete termination after
coaxial line by means of bakelite rods parallel to the the transformer has been tuned. A brass slug beyond
axis of the guide.1 The two low-loss dielectrics operate the'end of the conductors (not shown in Fig. 1) with
together as a double-slug transformer to cancel out the holes through which the rods may slide and screws to
multiple reflections, and the lossy dielectric which is tighten on the rods, was used to fix the relative positions
backed by a brass short absorbs the electromagnetic of the slugs once the transformer was tuned. The other
energy. The termination (used with a fixdd probe) is two pairs of rods, which must be dielectric (bakelite
adjusted for an impedance match by varying the was chosen for mechanical strength), extend through
distances between the three dielectrics with the rods holes in the load and screw into their respective trans¬
which extend out the end of the guide. A match is indi¬ former slugs. The pair of rods attached to the slug furth¬
cated if no standing wave is observed as the three di¬ est from the load passes also through 0.125-inch di¬
electrics are slid back and forth together with the spac¬ ameter holes in the other transformer slug.
ing between them fixed. Because of the presence of these bakelite drive rods
The dimensions of the components of the termination and the loose fit of the slugs in the transmission line,
are not critical and most of them were chosen arbitrarily. accurate calculations of the electrical properties of the
The major requirement is that both transformer slugs load are impractical. It is simpler and more accurate to
be a quarter wavelength long at some suitable fre¬ determine the vswr or reflection coefficient of the termi¬
quency, as is explained later. nation by direct measurement. The general approxima¬
tion for the attenuation of a lossy dielectric in a coaxial
II. Coaxial Termination
transmission line is2
This design was adopted especially for a 7/8-inch
&.686ir\/e tan <5
coaxial termination (Fig. 1) for use in the uhf range a = —-db per inch,
from 300 to 3,000 me. The 7/8-inch coaxial model is X
579/105
PROCEEDINGS OF THE I.R.E. January
The vswr’s can be matched to the transmission line With this coaxial termination it is possible to obtain a
by a double-slug transformer, a type which is effective vswr measured with a differential vacuum-tube volt¬
over wide frequency ranges for fairly large vswr’s. This meter less than 1.004 (reflection coefficient <0.002)
transformer can match any impedance whose vswr does without any difficulty. This small limitation in making
not exceed e2 for the dielectric of which the slugs are a perfect match is apparently caused by a slight warp¬
made. Though the transformer will match a vswr of this ing of the conductors so that, as the load moves along
magnitude only at frequencies for which the slugs are the line, the conductors go in and out of concentric posi¬
an odd number of quarter-wavelengths long, it is suffi¬ tions, thus changing the characteristic impedance of the
ciently “broad-band” to match any vswr approaching line.
this values at other frequencies according to the follow¬ At the lower frequencies the lossy termination results
ing equation,3 in larger vswr’s. In order to match a vswr of the order of
6 to 10, it is necessary to use slugs of such material as
b / b2
VSWTmax = 14-b \/ b 4- bakelite or Dielectene, which have dielectric constants of
2 V 4 3 or more. The more critical adjustment of these high
dielectric-constant slugs is partially compensated for
where
by the fact that in this case they are being employed at
the longer wavelengths. Of course, some decrease in the
b = 1/2 (1 — cos 2Pah) COS (30k
(-t) vswr at low frequencies can be obtained by the use of
longer pieces of Catalin.
+ As the first set of polystyrene slugs could not match
0-t)-sin 2(3ak sin (30k any appreciable vswr’s near 2,000 me, a second pair
was made which was a quarter wavelength at 2,000 me
and
to cover that frequency range. Thus, with two or three
PJa = electrical length of each slug,
pairs of transformer slugs the entire uhf band can be
do/o = electrical distance between the slugs,
covered. With only one pair a match can be obtained
e = dielectric constant of the slugs.
over most of the range. In order that the sets may be
The curves marked e = 2.5 and e = 3.0 in Fig. 2(a) are easily changed, the slugs have threaded holes into
plots of the above equation for slugs a quarter-wave- which their respective drive rods may be screwed.
length long at 1,000 me and indicate the maximum
III. Waveguide Models
vswr’s that can be matched at each frequency with
slugs having these dielectric constants. It can be seen Though this termination was designed primarily for
that, if the vswr’s are not large, there is little increase in use with coaxial transmission lines, waveguide models
the matching band to be gained by using transformer were also built for the purpose of determining their
slugs of large dielectric constant. Moreover, high di¬ performance at W-band (8,200-12,400 me) and at
electric-constant slugs were found to be more difficult i£-band (18,000-26,500 me).
to adjust for match than low dielectric-constant slugs,
such as polystyrene, whose dielectric-constant is about
2.5. Transformer slugs of polystyrene are capable of
matching the vswr of the Catalin termination over all
but the lower end of the uhf range. The measured vswr
of the 13-inch Catalin termination is also shown in
Fig. 2(a).
The experimental performance of the adjustable
coaxial termination is shown in Fig. 2(b). With the
polystyrene slugs (e = 2.5) of Fig. 2(a) which were 1.87
inches long it was possible to obtain a match over more
than 80 per cent of the uhf range. It is seen that the fre¬
quencies at which one could not obtain a match with
this set of slugs (that is, where the vswr of the Catalin
load exceeds the vswr matching capability of the trans¬
former) are somewhat lower than the curves of Fig. 2(a) The waveguide model which is shown in Fig. 3 uses
predict. This is probably caused by an effective increase 2-inch lengths of Catalin for the lossy terminations in
in the dielectric constant of the transformer slugs due to both the 0.900- by 0.400-inch AT-band guide and the
the presence of the bakelite rods, so that the slugs were 0.420- by 0.170-inch iCband guide. These lengths are
a quarter wavelength long at a frequency somewhat less sufficiently lossy to reduce the vswr in the waveguide
than 1,000 me. to a value less than three. At these frequencies this
vswr is due mostly to the impedance mismatch at the
’ R. C. Ellenwood and E. C. Hurlburt, “The determination of air-Catalin interface. The Catalin in the waveguide is
impedance with a double-slug transformer,” to be published in a
forthcoming issue of the Proc. I.R.E. fastened to a 1.5-inch sliding brass block which serves
580/106
1953 Ellenwood and Ryan: UIJF and Microwave Matching Termination
as a short circuit and as a centering device. The slugs to a vswr of 1.004 with two pair of slugs over the entire
and two 0.113-inch diameter drive rods in both wave¬ uhf range. Similarly waveguide models will provide
guide models are made of polystyrene. In the V-band matches to 1.01 over their particular frequency ranges.
model the slugs are 0.675 inch long, three-quarters Not only can these terminations provide a match, but
wavelength at 9,380 me, and in the iv-band model they may also be adjusted to given values of vswr of the
0.475 inch long, five-quarters wavelength at 23,000 order of one to ten when for various reasons such a
me, in order to improve their rigidity and stability vswr is desired.
within the guide. The two thumb screws shown in Fig. The outstanding features of this termination are its
3 are tightened on the drive rods to anchor the slugs in simplicity of construction and ease of tuning. There are
position once they have been properly adjusted. In the no critical dimensions and the essential requirements
waveguides a single drive rod was sufficient for mechan¬ are three dielectric slugs (one lossy) to fit loosely within
ical control of each transformer slug. With each wave¬ a transmission line and the rods by which they may be
guide termination it was found that a match could be moved. It takes but a short time to reduce the vswr to
obtained toavswrof 1.01 (reflection coefficient <0.005). less than 1.01. The vswr could be reduced still further if
closer tolerances and more refined components were
IV. Conclusion
used, such as a gearing arrangement to adjust the posi¬
This instrument is capable of covering wide frequency tion of the slugs, with a sacrifice, of course, of some of the
ranges. The coaxial termination will provide a match simplicity of construction.
581/107
National Bureau of Standards Circular 481
Issued September 1, 1949
582/1
able for voltage measurements because of their The general expression for the r-f voltage is
relatively small physical dimensions. Precautions
must be observed in mounting the bolometer so E=
that the voltmeter under calibration is connected Rt
directly across an r-f current-carrying circuit (RrERb) (1 + a)
element having negligible series reactance, since a (i)
fundamental assumption of this method is that
the r-f resistance of a particular bolometer is equal and for thermistors matched to have identical
to its d-c resistance. resistances and characteristics,
Figure 1 shows an elementary circuit diagram
of a bridge employing two thermistors. In order R
to determine the value of a radio-frequency e=2irEw)
(2)
for a portion of the power initially entirely supplied 2. Matched bolometers, unequal-arm bridge, rela¬
to the thermistors from the battery source. tively high Vo¬
Since Ra and Rb are effectively the same for both lt is sometimes advisable to maintain bridge-
balances, the following equations arising from the arm resistances of the same order of magnitude,
bridge constants for the two balanced conditions thereby assuring high bridge sensitivity. V0 is
permit an evaluation of E. This method is usually not continuously adjustable. R\^0. In
applicable to several hundred and possibly to this case, eq 2 may be used.
1,000 Me over an approximate range of 0.02 to 3. Single bolometer (#c=0, C' = 0); unequal-arm
1.5 v. bridge, Rx = 0:
frequency ranges are limited and are interdepend¬ than a half-period of the highest frequency
ent. This effect takes place when the phase of to be used. The error caused by this effect is
the deflecting voltage begins to reverse before the usually expressed as a ratio, C, of dynamic to
electrons in the beam have sufficient time to cross static sensitivity, the sensitivity being the linear
the deflecting field. The transit time thus reduces deflection per volt. This ratio is expressed as
the magnitude of the deflection unless it is smaller [6, 7, 8]
where
o}=2irj
sin ./2
(10) 0.01 o.i I io 100
0/2
Figure 3 shows C as a function of ./2, given by Figure 3. Effect of electron transit time on sensitivity of
eq 10 [9, 10]. cathode-ray tube [5].
572772 0-61-38
Another source of error is the effect on the deflecting plates of an oscillosope by a section of
deflection field of the lead inductance and deflec¬ a two-wire transmission line (in a plane normal
tion plate capacity. The voltage between the to the axis of the tube) extending through the
plates is larger than the voltage applied to the tube [11], The beam passing through the spacing
outside terminals of the cathode-ray tube (where between the conductors is deflected in proportion
the voltmeter under calibration is connected), as to the voltage at this point of the line. The
given by the following expression [5]: voltmeter under calibration is placed either as
close as possible to the tube or a distance X/2
1 away along the extended transmission line, where
El=E
1 -Lf/fr)2’
(ID the' voltage across the line is the same as at the
cathode ray. This method is shown in figure 4.
where The deflection angle 6 is given by
E1 — deflecting voltage
E— applied voltage EmSLX TV
f— operating frequency (12a)
2Va cosh'1 (d/2r)
jr— plate and lead series resonance frequency.
For large values of djr
The design and construction of present-day
commercially available cathode-ray tubes is such E 7r
e^ 2Va Inidjr)'
(12b)
that they may be used with negligible frequency
error up to 75 Me for voltages ranging from 5 to
several hundred volts. Their suitability and where d and r are the separation and wire radius,
range is dependent upon electrode and lead respectively.
arrangement, spacing, and insulation. The effective deflecting field is largely concen¬
Deflection sensitivity may be increased by trated within a space between the wires approxi¬
reducing accelerating voltage as is evident from mately equal in width to the wire separation; the
eq 6, permitting satisfactory measurements at distance responsible for a transit time error is
less than 5v radio frequency but only at the ex¬ therefore approximately equal to the wire separa¬
pense of a reduction of the upper limit of frequency, tion.
and an increase in spurious effects from magnetic A line having a 3-mm wire diameter and a
fields and normally minor voltage fluctuations. 6-mm center-to-center separation results in a
Similarly, the upper frequency limit of satisfac¬ sensitivity of 0.3 mm/v for Va= 1,000 v at a
tory response may be increased by increasing the screen distance of 25 cm; this may be compared
accelerating voltage (within its rated limits) at with a sensitivity of approximately 0.63 mm
the expense of a reduction in deflection sensitivity. under the same conditions with deflecting plates
Modified designs of oscilloscope tubes, with having a transit time distance nearly five times
proper precautions in circuit connections, may be as large.
used for frequencies above 75 Me. One modi¬ Experimental work, now in progress at the
fication is the replacement of the conventional National Bureau of Standards, with these types
SHORT-FOCUS ELECTROSTATIC
APPROXIMATE OUTPUT OF
TYPE CIRCUITS PRINCIPLE OF OPERATION BASIC FORMULAE
INPUT IMPEDANCE COMPLEX WAVEFORM
I Diode Detection In circuits (a) and (b) C charges For linear diode characteristics In general it is a function of R E peak
to E peak. R acts to discharge C. V(jc = KE peak. Equivalent input and amplitude of E. There is a
r k4 4 i 13.15,16,17.48,49,
In circuit (b), R‘ and C1 act as a
filter to keep r-f out of the d-c
resistance =
R
IpR
_
2
R
value of R at which the input
impedance is practically indepen-
aent of E. The order of magni¬
(a)
measuring .circuit. 2r~ tude is the equivalent of 1 to
c peak
25 megohms resistance shunted
approximately unity as R with an input capacity of 3-10
E r and E increase.
For square-law type diode
is a function of E. I_3, [5.
MM f
I05 ohms at
possible. [3
300 Me. is
(b)
note: NOTE
Numbers refer to list of refer¬ See list of nomenclature
ences .
3. Plate Detection Approximate parabolic lower Alp — K(E ,2 + e| +e|+ ---.) to a Approximately I07 ohms resis¬ (E? + El + El+---.)
Full wave, square law curved portion of Ip-Ec charac¬ first approximation tive at frequencies up to a few
teristic is used Average plate AT - 1 (99mYr2a. me shunted by (Cgp +Cgc) May
lc K current is higher than the p 4Mogm0 (rP + RL)^EgAE| drop to I04 or I03 at 300 Me.
E,r t ,,4 quiescent plate current. Biased depending on the tube. 5
for ip>o throughout the cycle. + El + E|+ •.) 40
R - 1
Rl (except of current meter) is 9 KgmfZ7Z ±!,35,6i.
usually omitted 60.
4. Plate Detection Same Same as above except tube is For relatively large values of. Same as above Eave if plate current
Half wave, square law. biased to cutoff. For large RL Rl and E, IP^KEave.For small characteristic is linear.
and C=0, i.p is nearly propor¬ r-f voltages and low values of RLl Erms if plate current
tional to eg during positive Lp = k(e?+e|+e|+ •••.) characteristic is parabolic.
half cycles
5 Plate Detection Same Tube is biased appreciably be¬ Ip =. KEpeah At negligible transit time ef¬ Not recommended. Error
Peak yond cutoff. fect R g — Jp. 41.44. might be appreciable.
Cg = (Cgp + Ggc )
6. Grid-detection Operation takes place along AIp = gmRAIg over the linear Relatively low Erms or Eoeak depen¬
the lower curved portion of the portion of the Ip-Eg character¬ ding on input level and
grid-current grid-voltage charac¬ istic. operating voltages.
■±£3> teristic and over the straight or
or the curved portions of the Ip -
Eg characteristic. XC«R.
°c —1
7. Slide Back D-C bias adjusted to obtain The peak of the positive half 1 E peak of positive half
same plate current with r. f. as cycle = Ema)<= K(V,-V0). K ap- C°(cgp *cgc) cycle.
V0V,^ A was obtained without r. f. (i e proaches unity as E increases Input resistance is approaching
nftf T <k with input terminals shorted)
Ip = Ip0=a few microamperes
and as the sharpness of cutoff
increases. It may be as low as
ieakage resistance across input
terminals. 34, 6[
34 0.2 and lower depending on the
tube characteristic and on the
magnitude of E.
V0 = d-c voltage at E=0
Vi = d-c voltages at other values
of E.
8 Inverted Triode Ig is reduced when an r-f Epeak—VP required to produce Resistance of the order of E peak
voltage is applied to the input the same I g 1000 megohms shunted by
terminals. Vp is negative 38 Amplification factor^ -p- Cep + Cpg . 38
588/6a
Table 1. Elementary circuits and characteristics of vacuum-tube voltmeters - continued
I Diode Detection The source impedance must The upper limit depends on Upper limit is affected by: Good. Depends on the This circuit, followed by a self-biased d c ampli¬
be negligible at all harmonics, tube rating. With sensitive (1) Series resonance of in¬ constancy of filament fier seems to be the prefered type most suit¬
and the level of harmonics d-cvoltmeter or with d-c am¬ put L and C anode to cath¬ voltage and emission able for the widest frequency range
must be low otherwise error plifier the lower limit is a ode r-f voltage ^ May require yearly cali¬ To eliminate low-frequency discrimination ,
may be as large as the per¬ fraction of a volt. Range brations . the product of RC should equal at least 100
centage harmonics present. [3 depends on frequency; the at the lowest value of frequency. The effect
E^t)2
higher the frequency, the of series resonance of the input to the tube
narrower the range; correction 2% for a type 955 vacuum is to:
may be applied. tube at 100 Me. (1) Increase the apparent input voltage at the
(2) Transit time error = fundamental frequency
AV _ K da (2) Emphasize; harmonics more than the fun¬
V A s/E peak damental. For high voltages, triodes con¬
3% at 100 Me with 955 nected as diodes or power rectifiers are
"acorn" at IOv. Range is used.
therefore a function of vol¬
tage applied. Correction
curves may be used to ex¬
tend the range. Overall ±
12% is obtainable without
correction forE>0.5v. 16, J5.
2. Diode Rectification Not subject to turnover. 45 A fraction of a volt to a Should be calibrated at Same R(_ may vary from 0 to 1 megohm. For RL
(i.e. no error is caused by re¬ few hundred volts and high¬ the operating frequency - 100,000 .fl, error caused by slight curvature
versing input teminals even in er depending on the value when used in the neighbor¬ of static tube characteristic is negligible. 5L
the case of unsymetrical of Rl and tube voltage rat¬ hood of 1 Me and higher.
wave-form consisting only of ings Probable useful range up
fundamental and its harmonics.) to several Me.
3. Plate Detection In practice Alp will depend Fraction of a volt to top With present commercial Poor as a result of Noise output can be corrected for by subtrac¬
Fult wave, square law. to some degree upon wave limit within square law range type tubes , a low-frequency tube ageing and vari¬ ting it from total output, 40
form Theoretically there is no of tube (a few volts for calibration will hold within ations in the d-c volt¬ i.e. A Ix = A Ijoial ~ A I^oise
turnover. Phase of harmonic commercial tubes) 5-10% to 20 or 30 Me. At ages .
has no effect. higher frequencies calibra¬
tion at each frequency is
necessary. 39
4. Plate Detection Subject to turnover and phase Fraction of a volt to a Same Same
Half wave, square law. of harmonics. value of E causing grid cur¬
rent flow
5. Plate Detection Subject to turnover and From Emax to values Same Same
Peak phase of harmonics. causing flow of grid currents.
6. Grid-detection Error may be appreciable. Fraction of a volt to a Approximately to 10 Me. Very poor. When plate rectification takes place in addi¬
few volts with receiving type tion to grid rectification, Alp may equal
tubes. zero at a certain level of E 5_l
7. Slide Back Subject to turnover. Froction of a volt to a few Approximately to 10 or Good . Practically inde¬ Sharp cut off is obtained with pentodes
hundred volts. Calibration is 20 Me depending on input pendant of ageing and connected as triodes with screen grid used as
indispensible especially for capacity. operating voltage vari¬ the control element. 37
voltages below approximately ations.
10 volts; calibration should be
made for a given Ip.
8. Inverted Triode Subject to turnover Large voltages, depending Possibly to 10 Me.Theo¬ Probably good. No
on tube design. retically limited by the in¬ experimental data avail¬
put capacity No experimen¬ able.
tal data available.
589/6b
2. High-input resistance. their application without frequency correction for
3. Short-input terminals. frequencies well above several hundred mega¬
4. High series-resonance frequency of input- cycles.
lead inductance and capacity. Table 1 lists fundamental detecting circuit ele¬
5. Freedom from transit-time correction. ments of vacuum-tube voltmeters and their major
6. Calibration must not be affected by ordinary functional characteristics. Associated circuits in¬
line-voltage variations, aging and temperature cluding regular and feedback amplifiers, current
and humidity changes, and must have negligible balancing circuits, voltage dividers, voltage stab¬
zero-setting drift. Vacuum-tube voltmeters must ilizing elements,, etc., are equally important in
not generate disturbing voltages and must give determining sensitivity, linearity, stability, and
steady indications. range of the meter [14, 16, 17, 18, 19]. Voltage
7. Calibration must hold over a reasonable dividers, specially constructed to fit given mechan¬
length of time and shall not be affected by tube ical and electrical requirements, may be used to
replacement. measure high voltages at high frequencies. One
8. Maximum voltage range with minimum arrangement is shown in figure 6 [20].
auxiliary equipment like amplifiers and voltage
dividers.
9. Peak voltage calibration for nonsinusoidal
waves; rms for sinusoidal waves.
10. Linear scale or large number of overlapping
scales for square-law indications.
Some relative merits of triodes versus diodes for
voltmeter applications are listed below.
Triodes are preferable at frequencies below
approximately 20 Me.
1. For high sensitivity to small applied voltages,
2. For lower loading effect on circuit being
measured,
3. For greater reliability of calibration at a
power frequency.
Their major disadvantages are:
1. Input voltage is usually limited to values low
enough to keep the grid from going positive.
2. The d-c plate current has to be stably bal¬
anced out to obtain maximum sensitivity.
3. Null point shifts as a result of supply voltage
variations, aging and warm-up period required.
4. Triodes may have a shorter life and may
require more frequent calibrations as compared
with diodes.
5. The input resistance at frequencies of about
100 Me is lower than that of a diode by a factor of
10 [5],
6. It is difficult to construct a triode having the
small interelectrode spacing required to keep
transit time and resonance errors to a minimum
[5, 15, 23],
A special diode construction was reported em¬
ploying an indirectly heated cathode having the
shape of a cylinder with a sealed-off end. A similar
indirectly heated sealed-off cylinder is used as an
anode. Both cylinders are lined up end-to-end in
such a manner that the two sealed ends face each
other, and the gap between them forms the electron
transit distance from cathode to anode. The
width of this gap, and consequently the transit
time, is controlled and is held at a minimum by
varying the heating current fed to the anode, Figure 6. Voltage divider for high r-f voltage measure¬
thereb}T causing its expansion or contraction [54]. ments
Diodes were recently developed with an inter- C is the terminal connected to a calibrated low voltage meter. The high
voltage is coupled through the capacity between parts B and C. Various
electrode spacing and element structure permitting sizes of part C are provided to obtain different voltage ratios.
resistance, frequency, and current density. This A, 0.200 disk; B, 0.130 disk; C, 0.008 disk.
type of rectifier is manufactured in all sizes down
to a small fraction of an inch in cross-sectional Whereas copper-oxide rectifiers are applicable
area for currents of a few milliamperes. as voltage indicators only at the lowest frequencies
Copper-oxide rectifiers are preferable to sele¬ considered here, modern crystal rectifiers are useful
nium types for instrument application because of up to 300 Me and higher.
their lower resistance. Selenium types may be The crystals most commonly used at present arc
operated up to about 10 v per element as against silicon and germanium.
about 2 v for copper oxide.
Table 2 shows the chemical constituents of some
of these crystals [26]. Other useful crystals are
A/Wv\AA- galena, iron-pyrites and carborundum. In meas¬
R, urement applications involving present-day dry
AAAAAA- rectifiers, one should recognize the following dif¬
ference in design: the contact area, and thus
power-handling capacity, of crystal diodes is much
smaller than for copper-oxide rectifiers.
Figure 7. Approximate equivalent circuit of copper-oxide
rectifier in the 50 kc to 5 Me range.
Table 2. Composition of crystal rectifiers
B\ represents the resistance of the body of the oxide. i?2 represents the
resistance of the oxide-copper interface and is a function of applied voltage.
Impurities added
VOLTS
Diode peak voltmeters are generally used as (a) Heterodyne Method of Extending the
convenient moderate-precision indicators [30, 31]. Voltage Range [32]
However, the discrepancy between the voltmeter
reading and true peak value may be very large. This principle is illustrated in figure 16, A
Figure 15 shows the response of a diode-type
vacuum-tube voltmeter as a function of pulse-
repetition frequency for a rectangular pulse shape
and pulse duration of 3.5 microseconds. An
approximate expression is given by Easton [30]
for the rectangular pulse peak in terms of the d-c
diode output as follows:
e0^Edc^l+-(13)
where
Figure 16. Heterodyne principle of calibrating r-f volt¬
meters.
e0 = the peak voltage of a rectangular pulse
Edc=the d-c voltage across R2
T=the duration, of the discharging interval linear diode frequency changer, D, mixes voltages,
h = the duration of the charging interval E0, supplied from an auxiliary source, with ElX, a
i?i = the total resistance during charging fraction of the unknown voltage Ex used to cali¬
R2 = the total resistance during discharging. brate the voltmeter. The intermediate-frequency
One of the curves of figure 15 shows values output voltage is directly proportional to Ex for
computed on the basis of this expression. The large ratios of E0/ExX. This proportionality holds for
effective input impedance of this type of voltmeter values of Exl not exceeding the order of magnitude
for pulse voltages may be very low. The direct- of 1 volt. The capacitor, C, is therefore indicated
voltage output (measured across R2) decreases as an element of a linear voltage dividing network
with increasing rate of pulse repetition, with in¬ used to maintain Exl considerably below Ex. The
creasing values of R2 and with decreasing values interelectrode capacity of the diode, shunted by
of B\. Rl is a function of the combined source and other circuit capacities (not shown in the figure),
diode resistances at the particular operating con¬ constitutes the second element of the voltage
ditions. Improved performance may be obtained divider. One can thus use a single calibration
by means of auxiliary circuits like cathode follow¬ point of the same voltmeter determined at a low
ers [30] and automatic slideback arrangements voltage (for example, 1 volt obtained by means of
[31]. the bolometer or any of the other standard meth¬
ods listed above) and proceed with calibrating it
at higher voltages. The advantage of this method
is that it can be used to calibrate high voltage
levels at high frequencies in terms of a standard
attenuator used at a relatively low intermediate
frequency. Let, for example, C be chosen at such
a value that Ex.t, corresponding to the calibrated
Exi (of, let us say, 1 volt), is some small value
(say Xx mv). Ex is then increased until a new
value of E[.i (say X2 mv) is obtained; the new
higher value of Ex is then equal the previous times
the ratio of these i-f voltages (that is, Ex2 —
(X2/Xi)Exi volts). This latter ratio can be
accurately determined with a standard attenuator
placed in series with the i-f output terminals.
The use of a low transit-time diode may eliminate
Figure 15. Response of a diode-type FTVM as a function
of pulse-repetition frequency for a charging-interval dura¬ the transit time error at frequencies up to several
tion of 3.5 microseconds and a peak of 26 v. hundred megacycles.
Consoles to calibrate rf voltmeters for sci¬ weeks and a nearly prohibitive cost; this was
ence and industry were developed at the National particularly discouraging in view of the limited
Bureau of Standards Boulder Laboratories. Very ability of the "unknowns" to retain their calibra¬
accurate calibrations will be performed in a frac¬ tion for a reasonable length of time. The new
tion of the time heretofore required at the Bureau reference standards and techniques incorporated
at any practical voltage level starting with 0.2 in the consoles described below enable calibra¬
volt at twelve discrete frequencies between 30 kc tion of the same voltmeter in a matter of days
and 700 Me. New type AT voltmeters, the most with relatively little loss in accuracy as com¬
stable xf-voltage reference standards known to pared with a calibration directly in terms of pri¬
date are used. These voltmeters can reproduce mary standards.
calibration data to ± 1% or better over a period
of one year or longer. The major components, NBS Primary Standards
in addition to the reference voltmeters, are ex¬
tremely stable rf sources, automatic and manual The NBS primary standards of rf voltages
level controls and protective and indicating for levels currently used with popular commercial
circuits. voltmeters are a cathode-ray tube in a "slide-
back" gystem and a special rf-voltage thermistor
Introduction bridge .
Since 1952 the National Bureau of Standards The cathode-ray tube is used at frequencies
has been engaged in planning and installation of a from 30 kc to 40 Me for all voltage levels between
Calibration Center to house at one location elec¬ 5 and 300 volts. The voltmeter under test is
trical measurement standards covering the range connected directly across the vertical deflection
from dc to the millimeter wave region. The pur¬ plates; the cathode ray is used as a balance indica¬
pose is to provide a necessary link between the tor having equal deflection sensitivity to dc and rf
national standards, maintained at the NBS, and within the frequency limits specified. In opera¬
the laboratories throughout the country, and to tion, a fixed edge of the fluorescent spot is first
do this at a minimum possible cost to the public. set to a reference point on the screen with the aid
The reference standards and techniques to be of a microscope. Sinusoidal rf voltage of approx¬
used in this link must meet two basic require¬ imately the desired amplitude is then applied to
ments. The first is to retain the accuracy of the the plates and one end of the resulting trace is
NBS primary standards as closely as possible. returned to the same reference point with a dc
The second is to reduce the time necessary for "slide-back" voltage. The dc voltage is measured
accurate calibrations of "unknowns" to a reason¬ with a potentiometer in terms of a standard cell,
able minimum. It is, of course, clear that the its magnitude being equal to the peak value of the
major consideration in developing a primary rf applied in the absence of resonance or transit¬
standard and technique for a certain quantity is time effects.
the realization of a magnitude as closely as possi¬
ble to its true or "absolute" value. Time ele¬ The other primary voltage standard, the
ment, convenience and other considerations are bolometer bridge , uses two matched thermistors
of-secondary importance and are given weight as bolometer elements in one arm of the bridge.
only when comparing two methods rendering The voltmeter under test is connected across
equal basic accuracy and reliability. That is these thermistors in a special mount, illustrated
why primary standards and associated techniques in figure 1, and the rf power applied to them is
are in many cases not practical for calibrating measured by dc power substitution. Because of
even limited numbers of reference standards. the small physical dimensions and the special
This is increasingly the case as the radio fre¬ manner of mounting the thermistors their rf
quency involved increases. It thus turned out at resistances are essentially equal to their dc
the time of initiation of this undertaking, that the values and the series reactances are negligible at
calibration of an rf voltmeter to several hundred ^•11 frequencies to 1000 Me. Any other imped¬
megacycles was a job requiring a number of ances shunting the thermistors do not affect the
598/251
value of the voltage appearing across the therm¬ All AT voltmeters used as reference standards
istors. Consequently, knowing the rf power and with the rf consoles described below are of the
the resistance of the thermistors, the rf voltage high and low impedance "fixed" type because
may be calculated. In this manner voltage levels these are rendering highest efficiency of operation
from about 20 millivolts to 1.5 volts can be ob¬ and maximum calibration stability.
tained at all frequencies from dc up to 1000 Me.
The voltage range is extended to higher levels by Description of the Consoles
inserting accurately known amounts of rf attenua¬
tion between the voltmeter under test and the The purpose of the rf-voltmeter calibrating
bridge. consoles is to provide means of calibrating all
types of voltmeters from 30 kc to 700 Me and
The accuracy of the voltages obtained direct¬ higher frequencies with maximum accuracy avail¬
ly with the cathode-ray tube and the bridge as able at present and in the shortest possible time.
checked in terms of each other and several inde¬ Two major steps made this possible; the first
pendent methods is ± 1% or better. The attenua¬ was the application of the AT voltmeters; the
tor is introducing another possible error of ± 1% second was the decision to calibrate the meters
or le s s . at discrete frequencies. The latter decision is
justified because the frequency response curves
Reference Standards for present day rf voltmeters seldom have seri¬
ous discontinuities. However, frequencies other
To make efficient use of the above primary than those selected may be necessary. At the
standards it was necessary to develop a reference time of this writing two of these consoles have
instrument capable of fairly quick response in been completed and placed in service, one for 30,
indicating a given value of voltage and above all, 100, 300 and 1000 kc and the other for 5, 10, 30
capable of reproducing the same indication over and 100 Me. The third partially finished unit will
long periods of time. The latter critical quality operate at 300, 400, 500 and 700 Me. The place¬
may be referred to as "calibration stability". ment of components and controls is essentially
After a long search for suitable components, the same for all consoles to present a uniform
waveguide below-cutoff attenuators operating in appearance, to facilitate servicing and for effi¬
the TM mode and capacitive type attenuators in cient operation.
combination with thermoelements proved most
promising. Attenuator-thermoelement (AT) volt¬ Figure 3 illustrates the arrangement of the
meters are, therefore, currently being used as controls and indicators of one of the consoles. A
rf voltage reference standards. After calibration specially designed coupling plate for connecting
in terms of the primary standards, they retain the voltmeter under calibration in parallel with
their calibrations to ± 1% for more than a year. the reference standards is mounted at the front of
each of the two pedestals of the console. The
There are two general types of AT voltmeters, indicator lights, in the narrow panels above the
"fixed" types having voltage ranges of about 5 to 1 pedestals, are for the overload protective cir¬
and adjustable types. Figure 2 illustrates the cuits, the crystal oven heaters and the ac power
basic configuration of the adjustable types, having lines. The meters above the lights are in the dc
continuously adjustable attenuators. Two such power supply circuits to monitor the operation of
voltmeters were developed at the NBS for voltages the rf sources. The millivoltmeters in the adja¬
from 20 to 1000 volts at frequencies between 0. 1 cent panels indicate the thermocouple output from
and 1000 Me. One of them is using a wave-guide- the AT voltmeters, and the line charts, displayed
below-cutoff capacitive type attenuator with disk in the windows of the storage drawers beneath
electrodes for frequencies from 10 to 1000 Me. them, are used to convert these indications quick¬
The other, having conical electrodes as the ad¬ ly into the actual rf voltages applied. Controls
justable elements in a capacitive voltage divider, for operating the corresponding half of the console
covers the low-frequency end of the range. An¬ are located on a panel below each storage drawer.
other type employs a fixed attenuator of high input The controls in the center of the panel, figure 3,
impedance. Voltages between 0.2 and 20 volts at are for the adjustment of the rf voltage level.
all frequencies are measured with a third "fixed" Referring to figure 4, a seven position switch,
type, 50 ohm AT voltmeter; fixed interchangeable CSj, connects the rf source to the appropriate
attenuation pads are used to obtain voltage ranges reference standard and also connects the milli¬
of 0.2 -1.0, 1.0 -5.0, and 5-20 volts, respec¬ voltmeter to its thermocouple output through
tively. The adjustable AT voltmeters have input switch Sj . The remaining control, switch S^,
impedances comparable to that of vacuum-tube connects the dc high voltage to the rf voltage
voltmeter s. source of the appropriate frequency
599/252
and by means of a solenoid-operated coaxial sole in which they are located. The voltage
switch, CS , connects all the reference stand¬ ranges of 20 - 100 and 100 - 500 volts at each
ards to tha£ rf voltage source. These controls frequency are covered by single-frequency stand¬
and the four crystal controlled rf voltage sources ards using fixed "capacitive" attenuators, bring¬
--two in each half of the console--are perman¬ ing the total number of reference standards in
ently installed together with low-pass rf filters, each half of a console to seven. The series
impedance matching networks and overload pro¬ reactance of this type of "capacitive" attenuator
tective circuits. All of the rf components are is many times larger than the resistance of the
carefully shielded to eliminate stray radiation. thermoelement heater terminating the attenuator.
Therefore, in analyzing the effect on heater
Each half of a console may be operated inde¬ current of a change in heater resistance with a
pendently of the other, permitting the simultane¬ given rf voltage input to the voltmeter, the attenu¬
ous calibration of rf voltmeters at two frequen¬ ator can be represented at its output terminals by
cies per console. In this manner a large volume a constant current source. Changing the heater
of calibration work can be accomodated. resistance even by as much as 50% has no effect
on heater current.
The time required to calibrate a voltmeter,
including every step in performing the calibration Consequently, once an rf voltage calibration
of a voltage point on a typical rf voltmeter, has of a given reference standard has been made,
been analyzed. Then the components and their replacement of its thermoelement does not neces¬
arrangement in the consoles were designed to sitate an rf recalibration. Only a dc calibration
reduce the time wherever possible. Rapid con¬ to determine the relationship of heater current to
nection of the "unknown" voltmeter to a reference dc millivolts output, need be made, and if the
standard is achieved with the RF Voltmeter characteristics of the replaced couple match
Coupling Plate, figures 5 and 6. Reference those of the original closely enough, even this
standards, seven for each half of a console, are calibration is unnecessary. When matching can
mounted on the under side of each plate. To con¬ be achieved, the same calibration chare can re¬
nect the "unknown" in parallel with a standard, main in service. For the voltmeters using dissi¬
the operator attaches a suitable adaptor ring to pative attenuators, however, a complete recali¬
the probe of the "unknown", inserts this ring in a bration of the voltmeter is required when thermo¬
sliding carriage on the top side of the plate, cen¬ elements are replaced.
ters it over the reference standard to be used and
clamps it in place. The "unknown" and the stand¬ The designing of special labor saving equip¬
ard are now connected in parallel through a "T" ment and the streamlining of procedures to re¬
connection of approximately 3/16 inch length, duce time is of little value if the amplitude of the
with the rf voltage applied midway between the applied rf voltage does not remain stable to at
two. Such a "T" exists at each of the seven least ± 0. 1% during the course of a measurement.
standards. The operator then rotates the seven- It was found, from experimental tests, that
position switch until the number below the index regulation of the dc screen and plate voltages
corresponds to the position of the standard in the applied to the rf voltage sources produces suffi¬
plate, and closes the switch. The rf level is ad¬ cient output-amplitude stability, except in the
justed with the manual level controls and the dc case of the 100 M'c source, where additional
thermocouple output in millivolts indicated on the stabilization has been provided through an auto¬
panel meter is converted immediately into actual matic amplitude stabilizing circuit . Figure 7
applied rf voltage with the line chart displayed in shows one of the circuits used. is a tempera¬
the window. ture limited diode. Its plate current is very
sensitive to cathode temperature variations.
As already indicated above, fixed type AT Raising the temperature of the cathode increases
voltmeters have been chosen in preference to the the plate current, causing the drop through R^ to
adjustable type used in the laboratory because of increase and reducing the dc voltage at the diode
their simplicity and superior stability. The plate. As this voltage decreases, the positive
mechanism required to operate the piston of the voltage on the grid of decreases, resulting in
continuously adjustable type increases their com¬ an increased voltage drop through V^. The volt¬
plexity of construction and operation. Three age drop across added to the constant voltage
reference standards having dissipative fixed drop through determines the screen voltage on
attenuator input pads cover the voltage range 0.2 the final stage of the rf source. Therefore, if a
to 20 volts in ranges of 0.2 -1.0, 1.0 -5.0 and sample of the rf output power is fed to the diode,
5.0-20 volts, respectively. These are used at an increase in the rf output level causes an auto¬
both frequencies provided in the part of the con¬ matic decrease in the screen voltage of the final
600/253
rf stage which i^i turn reduces rf output. With an heat or "hot spots" which might adversely effect
open circuit gain of approximately 2000, the rf components or cause thermal emfs to appear in
voltage amplitude stability is considerably im¬ metering circuits, equipment containing vacuum-
proved. The setting of C determines the value tubes, which generate heat, has been carefully
of the rf output level. is adjusted during located where there is adequate ventilation.
installation of the equipment so that at any opera¬
ting level, i.e. at any setting of C^, the diode is Accuracy
never extinguished. From experimental observa¬
tions, the amplitude remains stable to better than Since in most cases, rf voltmeters calibra¬
0. 1% for a period much longer than is required ted by means of these consoles are to be used as
to make a measurement. primary standards in a laboratory outside the
NBS, it is imperative that the error in measure¬
For the 100 Me circuit, smooth manual ad¬ ments be as low as possible. With the calibra¬
justment of the rf voltage amplitude is obtained tion of the AT standards good to ± 1 or 2%, the
by varying an element in the circuit of the auto¬ other factors effecting measurement accuracy
matic amplitude stabilizing circuit, described have to be very carefully studied. As the thermo¬
below, and by varying the plate and screen volt¬ elements are used in the AT voltmeters, the dc
ages on the rf signal source. Final adjustment output from the thermocouples range approxi¬
of rf voltage level is made by using the dc voltage mately from 200 pv to 7 millivolts. Even a
control as the vernier element. For the other small amount of thermal emf in the circuit from
frequencies, all manual adjustment of rf voltage the couple to the millivoltmeter introduces error.
level is achieved simply by varying the dc plate Consequently, all wiring of the metering circuits
and screen voltages. is located well away from heat sources and junc¬
tions of dissimilar metals have been eliminated
Any of the reference standards might acciden¬ wherever possible.
tally be subjected to voltages considerably beyond
its maximum rating, damaging the standard and The millivoltmeter selected to indicate the
necessitating its removal for repair and recali¬ thermocouple output has an internal resistance of
bration. Protective circuits, therefore, were the same magnitude as the junction of the thermo¬
incorporated. Figure 8 shows, in block diagram couple itself (10 ohms) to eliminate sluggishness
form, the method of protection used. A crystal in the meter due to overdamping and to get maxi¬
diode associated with each standard (except the mum meter deflection. Since each AT voltmeter
one for the top voltage levels) monitors the rf is calibrated individually, reproducability may
voltage applied. When the level reaches 120% of seem to be more important than the accuracy of
a standard's maximum rated level, relays are the thermocouple output indication. However, in
activated to remove the high dc voltage from the case of replacing a thermoelement or if it is
rf voltage source, thus, instantaneously reducing more expedient to check an AT voltmeter with a
the rf voltage at the voltmeter to zero. Returning millivoltmeter other than the one used in the
the manual level control to zero operates a reset console, accuracy is as important as reproduci¬
switch, which restores the relays to their original bility. Using a multiplier, the millivoltmeter
condition. Attenuator sections A , A , A^q, indicates thermocouple output in two ranges. The
A^» A^* A , and A^ are used to enable calibration of actual applied rf voltage at the input
operation of the rf sources within certain mini¬ to the AT voltmeter vs, the millivoltmeter indica¬
mum and maximum levels with maximum stability tion for both ranges is presented on separate line
and at safe plate dissipations of the rf-source charts mounted back to back in the same slide
output tubes. Since the output is reduced by holder. Splitting each of these charts, a total
dropping the plate and screen operating voltages, scale length of 24 inches is obtained which ena¬
the latter must be kept above a certain minimum bles expansion of the scales for more precise
value; operation at lower values introduces insta¬ reading.
bilities in the frequency and amplitude.
Conclusion
In addition to protection against overload
damage to the reference standards, every possi¬ The development of the consoles has resulted
ble precaution has been taken in designing the in reduction in the time to calibrate rf voltmeters
system to minimize the man-hours required to to a fraction of that previously required at the
keep the consoles functioning properly. All com¬ NBS. RF voltmeters have been calibrated using
ponents and equipment used in a console are both the consoles and primary standards; the
operated well within the ratings specified by the agreement between the calibrations ranged from
manufacturers. To prevent the accumulation of 1 to 3%. If the workload should increase to the
572772 0 - 61 - 39
601/254
point where the system needs to be expanded, the 2. M. C. Selby and L. F. Behrent, A bolom¬
proper solution would be to duplicate the present eter bridge for standardizing radio-frequency
consoles. Some improvement in measurement voltmeters, J. Research, NBS 44, pp. 15 - 30,
accuracy may be possible by substituting a more (Jan. 1950).
accurate instrument such as a potentiometer for
the panel millivoltmeters . However, this may
result in a sacrifice in efficiency. 3. Stable radiofrequency voltmeters, NBS
Tech. News Bui. 40, No. 2, 29 - 30, (Feb. 1956).
Reference s
602/255
(FOR CALIBRATION ONLY) MN - MATCHING NETWORK
STANDARD
Vx - VOLTMETER UNDER
OL - OVERLOAD CIRCUIT CALIBRATION
RF VOLTAGE
OA - OSCILLATOR-AMPLIFIER, INDICATOR S - SWITCH
RF SOURCE CS - COAX SWITCH
OLR - OVERLOAD RELAY AT- ATTENUATOR - THERMO-
THERMOCOUPLE
Fig. 3 RF voltage calibration console. Fig. 4 Block diagram of half of rf voltage console.
603/256
COAXIAL CABLE
FOR RF INPUT
V, - MAY BE KALOTRON 2ASI5 OR 2SMI5 OR
SUPERIOR ELEC. TYPE I236C
NOTE : CARRIAGE NOT SECTIONED TEMP LIMITED DIODE
V2 - VR TUBE
604/257
U. S. Department of Commerce Research Paper RP2055
National Bureau of Standards Volume 44, January 1950
In the course of work on r-f standardization projects of the National Bureau of Standards,
practical equipment and techniques were developed for measurement of 20 millivolts to
1.9 volts at radio frequencies below 700 megacycles, with further broadening of these ranges
still in progress. The devices employed do not require frequency corrections, so that
results are based on direct-current measurements, which can be made quite accurately.
Reproducibility and agreement with independent methods was found to be ±1 percent or
better. The primary function of the equipment is to make available a standard voltage
of any desired value and frequency within the above range that would be required for a
highly accurate and at the same time practical voltmeter-calibration procedure. In addi¬
tion, the equipment may be used for direct measurement of r-f generator voltages; as a
known standard of r-f impedance; and for accurate power measurements of 20 microwatts
to 100 milliwatts. An analytical study of the range-limiting factors of this technique is
presented, appropriate curves and illustrations are shown, and major steps of procedure
are listed.
An independent method and appropriately designed equipment for accurate r-f voltage
determinations based on measurements of direct current and of linear dimensions along a
transmission line of known characteristic impedance are also described.
V=\[VR2(2V«-VRi)]K (4)
Figure 9 shows the drift of F0 (under the present The complete arrangement of the two-thermis¬
load) to be approximately 0.001 percent per hour. tor bridge is shown in figures 10 and 11. Figure 10
To evaluate this quantitatively one may assume is a view of the experimental setup including a
that the drift is equivalent to a readjustment of vacuum-tube voltmeter under calibration. A
Fr2 required to restore bridge balance during a final working arrangement could no doubt be
time interval of a single r-f voltage measurement. consolidated to a great extent. Figure 11 is a
The application of a special r-f and d-c switch combination schematic and block diagram in
(described in the next section) reduces this interval which the designations of the components corre¬
essentially to the time required for a measurement spond with those of figure 10. The source of r-f
of VR2. It may be seen from the curves given voltage consisting of a generator, filter, and match¬
and the indicated drift that VR2 for a F of 20 mv ing networks is not shown in figure 10.
will have to be determined in about 40 sec if the The major components of the equipment were:
drift is to be neglected. The steps leading to this (a) Resistance boxes comprising the legs RT
value are as follows: The actual drift in F0=10 v and Rb of the bridge.
is 100 pv/hr (fig. 9). From figure 7 a voltage (b) Resistance boxes and a helical type single-
Fr2 of about 550 pv corresponds to a F of 20 mv. wire adjustable resistance comprising the
The permissible error in this value of VR2 for a resistance R in series with the battery
- BRIDGE
SHORTING RELAY
galvanometer
R SLIDE-WIRE ELEMENT
UNBALANCE
METERS
STANDARD CELL
VTVM DC POTENTIOMETER
PROBE
INTERLOCKED RF'DG
COAXIAL SWITCH THERMISTOR MOUNT
V0 used to vary the voltage applied to bridge while the r-f power is removed
the bridge. Voltage drops Vm and VR2 from the bridge.
(used in eq 1, 2, 3, and 4) are measured (f) The thermistor mount.
across this resistance. (g) High-resistance d-c voltmeters to indicate
(c) D-c potentiometer, associated resistors, Rn the ratio of RT1 to RT2.
and RP2 to increase its voltage range, a (li) Switch S0 used to connect the battery V0 to
plug box used for conveniently connecting the bridge.
the potentiometer for measurement of (i) Switch SD used to connect a standby load
V0, VR1, VR2, and of (V0— TV) when de¬ Rd to the battery V0 in place of the bridge.
sirable, and a standard cell. This load was necessary to prevent a
(d) An interlocked, r-f, d-c switch used to re¬ slight increase in battery voltage when
place instantaneously the r-f power fed standing idle for several hours. On
to the bridge by its equivalent d-c power connecting a load to an idle battery the
or vice versa without upsetting the bridge rate at which the terminal voltage de¬
balance. creases is relatively high during the ini¬
(e) A shorting relay controlled by the inter¬ tial discharging period.
locked r-f, d-c switch, mentioned in item The operational procedure was briefly as follows.
(d); this relay shorts, the resistor R in Referring to figures 10 and 11, with no voltage
series with the bridge battery, thereby applied to the bridge, RT and both Rb arms were
increasing the voltage applied to the set to predetermined values that would approx-
INTERLOCKED R — F
AND D — C SWITCH
Fictjre 11. Operational schematic and block diagram of the r-f voltage thermistor-bridge equipment.
C,C'=mica 0.01 microfarad condensers; Rt, i?t=bridge ratio arms; i?=series dropping resistors; Rpi, Rp2=d-c voltage divider used with d-c potenti¬
ometer; Uo=galvanometer sensitivity control; *So=galvanometer switch.
imatcly balance the bridge at the specific d-c Cross sections of the thermistor mount arc shown
voltage to be applied. With the shorting relay figures 12 and 13. The plate, A, and the pin, L
closed, the interlocked r-f, d-c switch in position (fig. 12) are removable, as is illustrated in figure 14,
B (fig. 11), and the sensitivity of the galvanometer to permit means of connecting various probes of
reduced, switch £0 was closed. Both Bb arms voltmeters under calibration to the mount. The
were adjusted simultaneously until a rough line drawing in the lower corner of figure 14 is an
balance was obtained. After increasing galva¬ enlargement of the gap in which the thermistors
nometer sensitivity by reducing Rg, further adjust¬
ments in Bb were made until a satisfactory balance
resulted. R-f was then applied in small increments
after closing the coaxial switch and equilibrium
restored by increasing the value of B. Several
operations of the interlock switch were necessary
until a desirable balance was observed. In cali¬
brating a vacuum-tube voltmeter, the amount of
r-f applied depended upon the meter deflection
desired. The amount of d-c voltage removed
from the bridge to restore balance was then Figure 12. Cross-sectional drawing showing the construc¬
measured by the potentiometer. tion of the thermistor mount.
tions (7?r=200 ohms; Rb=300 ohms) was found Measurement of cur¬ 1 100 kc to 30 Me.. None.
to be about 25 pv and remained stable to within rent through a
known resistance.
the measuring accuracy of the potentiometer used.
The voltages V0, VR2, and VRX were measured Transmission-line 1 100 Me to 600 Me.. Do.
voltage-distribu¬
by the d-c potentiometer, P, in terms of the tion.
standard cell (fig. 11) at points x, y, and 2. To
Single-thermistor 1 100 kc to 30 Me.. Do.
reduce the resistance of leads, wherever advisable, bridge.
wide low-resistance straps were employed for
572772 0 - 61 - 40
tor bridge at frequencies considerably above 800
Me with accuracies equal to those at the lower
frequencies.
A
Let Rti — oiRt2, where a> 1 and remains constant for all
Substituting for F and V\ in terms of a
values of applied r-f voltage. It can be shown that eq 14
holds in this case. OL1 ( 1 -j- ^
--1 = 10-3. (25)
Then Rr = [<x/ (1 -f-c*)]/??^ or Rl— [a/(l (a-|-Aa')I(l-[-a-|-Aa) 1
V=
Rt (-+1^ = 0- and — = y.
(Rt^T Rb)( 1 + Oc)
"\j<x(VH2 Ir«i) (2F0 — V R2 — Fft]). (17) \a / a
Then -ferr^ = 1.002. (26)
Appendix II. Error in V When a — RTl/RT2
Is Neglected Solving for y in terms of a
In eq 1 let (2a - 1,002a2) ± V(2cr - 1,002a2)2 + 0.008a-
y=- (27)
F
As a increases, F decreases. ^>1, where Fi = thecal- ^=0.1%.
v 1
619/29
Voltages for r-f Standardization
= ratio of the voltage X/2 from the load to the maximum
Appendix VI. Error in V Caused by
voltage in the line.
Inaccuracy of Measuring VR1
Then
From eq 33
P - V*2 (43)
RF KJpZ0'
AVR1=AV ~^p = AV K^y^-yoy (36) and
VA = K7ZaPRFP)i (44)
AVri AF V2
(37)
VR1 ~ V KVR1(VR1-V0)' The ratio ymax/VA was determined in two steps. First,
when the line was loaded, the probe on the line was located
In the case under consideration VqS>VRi, because y0==10 where the voltage was a maximum. The probe output
v and Vrfl1 = 25X10-6 v. VRX can be measured to 1 p\ or Dm was recorded. The probe was then moved to the
AVR1/VR1=±1!25 position corresponding to X/2 from the load, and the
probe output DA was noted. The line was then shorted
AFS1 . AV V2
(38) and the probe moved to where the voltage was a maximum
vR\ ~ v 10kvr1'
in the line. The r-f level was adjusted until a y,Smax corre¬
sponding to the maximum indication, Dm, was reproduced
Neglecting the minus sign because the error can be posi¬
(Vs designates shorted line conditions). Then two
tive or negative and assuming Ay/y=10-3 we have
positions were found on the line that produced the indi¬
1 i /wi V2 cation Da, and the distance in centimeters between these
25 = 1° 3 10/v X25X 10-6’ points (both on the same side of the maximum) was
recorded. This distance was called As. The distance
K = 0.04 between a maximum and a minimum corresponded to
A/4 (tt/2 radians). Since the voltage distribution in the
y = 2 X 10_2v = 20mv
line on short circuit was very nearly sinusoidal, DA = DM
cos 6, where 6 was the angle in radians between DM and DA.
Appendix VII. Relation of the R-F Load
But 2#/7T=Z/(A/4), where l is the distance in centimeters
Voltage to the Power and VSWR in a from Irgmax t° Fa on the line. I also equaled (A/4—As/2).
Slotted Transmission Line Substituting and solving for 6 yields d=ir (0.5—As/A).
Therefore
3
y
_ v max * min
y .5
RF —-7}-> (39) The VSWR is determined in a similar manner, i. e., by
reproducing the probe outputs corresponding to Vm&x and
y„ ymin at two positions, one of them at a voltage maximum
p—V ■
(40)
y min of the probe with the line shorted. The distance between
these two positions in electrical degrees is then the angle,
Substituting for ymin in eq 39 yields
the cosine of which is equal to p. Sequence of steps
suitable for various standing-wave ratios, as well as a
(41) general analysis of this method of measuring voltage
pZ o
Let ratios along a transmission line is presented by Winzemer.6
_Va = A’i (42) 6 A. M. Winzemer, Method lor obtaining the voltage standing-wave
vma ratio on transmission lines independently of the detector characteristics.
Thesis presented to the Polytechnic Institute of Brooklyn (May 1948).
! Principles of radar, by the staff of the MIT Kadar School, pp. 8 to 45
(The Technology Press, Massachusetts Institute of Technology, Cambridge,
Mass., 1944). Washington, April 18, 1949.
37
JANUARY I960
APPLICATION OF RF MICROPOTENTIOMETERS
FOR
CALIBRATION OF SIGNAL GENERATORS TO 1000 Me
by
L. F. Behrent
SUMMARY
621/ i
APPLICATION OF RF MICROPOTENTIOMETERS FOR CALIBRATION
L. F. Behrent
1. INTRODUCTION
1
The RF Micropotentiometer is a simple and accurate standard
2. CALIBRATION PROCEDURE
receiver, for comparing the rf output voltage of the unknown with the
reference standard.
622/1
The level of the standard is adiusted until the detector shows its
ranges.
3. PRECAUTIONS
levels as high as 10 kpv errors greater than 100 percent have been
fore, the metal box containing the reference standard must be well
is used to provide large contact surfaces between the box and the
removable cover, the body of the coaxial rf input connector, and the
623/2
be obtained by mounting button type mica capacitors in recesses in the
each of the signal generators and the rf detector. The shielding will
the screened boxes. All other connections into the cabinets required
when external braid was placed over each of the rf cables, with the
624/3
Any receiver so used should be tested to determine the optimum oper¬
the output, the amplitude of the rf voltage would be the same at all
points along the line and it would be immaterial how long a line section
large changes in the amplitude of the output voltages may result from
changing the length of the coaxial connection between the output and
the load even by a fraction of an inch. The plane through the rf con¬
would be the plane in which the resistive annular ring lies, a-a
625/4
plane would pass through the connector essentially at the tip of the
female center pin (b-b Figure 5). Where other types of output con¬
few microvolts applied to the receiver input, and the receiver tuned
626/5
4. CONCLUSIONS
5. REFERENCES
APPENDIX I
Resistance Standardized
Thermoelement (RF Micropotentiometer), rf voltage,
rating, ma milliohms microvolts
5 1 1 - 5
25 1 5-25
100 1 20 - 100
50 10 100 - 500
50 50 500 - 2500
50 100 1000 - 5000
25 1000 5000 - 25000
100 1000 20000 - 100000
627/6
the rf generator to be calibrated.
D is the rf detector.
Figure 1
628/7
Figure 2
629/8
a-a is the plane in which the standardized reference voltage exists.
b-b is the plane in which the rf voltage is compared with the primary
rf voltage standard.
Figure 3
630/9
Figure
631/10
CD
£uo
nj <D
' CD CD
-Q Jh U
Figure
632/11
xO
Figure
<D
3
cti
u
f-H
6
o u
U O
v-t
X
?H
G
a, o xl
G G G
,rH ^ G
vn G 4->
^ G w
633/12
572772 0 - 61 - 41
PrcDrintcd from
TRANSACTIONS of the American
Institute of Electrical Engineers
33 West 39th St., New York, N. Y.
Synopsis: The questionable accuracy of the accuracy and stability of devices Basic Requirements and Principle
radio-frequency microvoltages has been of monitoring the input voltages to the Operation
great concern to the radio field for many
years. There is an urgent need for a simple, attenuators. In searching for a source of accurate
yet reliable, source of microvolts for meas¬ Considerable progress was apparently and reliable r-f microvolts the following
urements in general and for radio receiver made as time went on, and a large sec¬ basic requirements seemed indispensable
sensitivity determinations in particular. tion of the field has managed to attain or highly desirable:
Extremely simple devices, which seem to
high accuracies despite these difficulties.
satisfy that need most adequately, were re¬ 1. The output voltages of the source had
cently developed. They are sources of A recent poll of the field, instigated by to be known irrespective of loading condi¬
potential drop obtained across a known the National Bureau of Standards in tions, that is, a constant voltage source was
resistance through which known currents 1951 to determine the urgency of this necessary.
flow. These devices provide constant problem, revealed the following status.
voltage sources of accurate microvolts over 2. Freedom from frequency corrections was
a range of 1 to 105 and wider, at all fre¬ Of about 70 standard voltage-generator essential at least over reasonable frequency
quencies to 300 megacycles and higher. manufacturing and applying laboratories ranges.
They are adaptable for balanced as well as (six of them in United Kingdom and 3. A reliable, simple, and rugged physical
for unbalanced sources. Their electrical France) about one-half believed they had construction with a very minimum of com¬
constants are simply determined by using ponent parts was most desirable.
available 1 microvolt at frequencies to
known direct voltages and currents. Basic
1,000 megacycles with an absolute ac¬ 4. The simplest and a very minimum of
principles, design features, and applications
calibration requirements were essential.
are discussed. curacy of 30 per cent or better. The
indicated accuracy was naturally better All of these requirements seem to be
at lower frequencies. The other half satisfactorily achieved with the Micro¬
T HE RADIO and electronics field has
been facing the problem of accuracy
was uncertain of their values in various
degrees, in many cases exceeding 100 and
potentiometer. In basic principle it is a
source of potential drops obtained across
of radio-frequency (r-f) microvoltages in some 200 per cent. About one-half a known resistance through which known
since sensitivity of radio receivers came of them desired absolute accuracies of currents flow. These resistances are of
into prominence as a competitive index 0.5 to 6 per cent for all frequencies to the order of milliohms (for microvolt
of performance. The reasons for the 1,000 megacycles; the rest desired ac¬ levels), and therefore constitute an essen¬
continued existence of this problem are too curacies of 50 per cent or better. tially zero source impedance (constant
well known. There is no need to discuss The critical need of reliable tools to voltage source) for all practical present-
them other than perhaps to indicate the supply or measure microvolts accurately day needs. A cardinal requirement of
major ones: namely, 1. the uncertainty was therefore still very much in evidence.
Paper 53-22, recommended by the AIEE Instru¬
in voltage source and load (receiver input) Devices apparently fulfilling this need ments and Measurements Committee and approved
impedance values, 2. the uncertainty in most satisfactorily were recently devel¬ by the AIEE Committee on Technical Operations
for presentation at the AIEE Winter General
performance of attenuators over wide oped. They seem to meet the most Meeting, New York, N. Y., January 19-23, 1953.
Manuscript submitted August 29, 1952; made
frequency ranges and attenuations of up persistently demanded feature of a stand¬
available for printing November 10, 1952.
to 100 decibels and in many cases of ard of microvolts, namely, extreme
Myron C. Selby is with the National Bureau
higher values, and 3. the uncertainty in simplicity and reliability. These devices, of Standards. Washington, D. C.
634/1
INPUT COAXIAL OUTPUT COAXIAL Figure 1 (left). Equi¬
TRANSMISSION LINE TRANSMISSION LINE valent circuit diagram
of system employing
a Micropotentiom¬
eter
TO SOURCE
OF R-F
Figure 2 (right).
ENERGY
Section of Micro¬
potentiometer show¬
ing position of
annular element in
a low dielectric
coaxial line
OF THICKNESS
d (.l-l/a)
this impedance, as previously implied, Let em, Mm, and -qm designate re¬
was that its reactive component be negli¬ spectively the permittivity, permeability,
gible compared with its resistive com¬ conductivity, and intrinsic impedance of Using well-known transmision line theory,
ponent. This suggested the annular an infinite metallic medium. Under the the output voltage V of a line section
type of resistance element construction. preceding assumptions equivalent to the annulus terminated by
The equivalent circuit of the Micro¬ an impedance Z, is given in terms of the
OJ€m<^.<C(TTn
potentiometer together with the input input current 7i by
and output circuits is that of a coaxial where
Ii
line shunted by a conductive film in a Vr =
w = 2wj (8)
plane normal to the axis of the line, as — cosh 77+— sinh yd,
ZT Zq
shown in Figure 1. A source of r-f For copper at 1,000 megacycles
energy supplies current to the Micro¬ and the transfer impedance is
oi6ffl= 0.17 mhos per meter
potentiometer section of die system
through a coaxial line and the voltage as against 7
vr 1
output V, from the Micropotentiometer, 1 <f|
= 5.7X10? mhos per meter — cosh 1+j); +
available across the annulus, is fed to the zT L sj
receiver either directly or through an Therefore3 1 sinh [(1 +j)d/8]
output line. To a very good approxima¬ ~Fm (1+/)<*/«
tion the output voltage V is simply the (9)
^=(1 +J)\I~ (1)
product of the input line current entering 1 2<rm
In this application Zr»Rm, therefore
the annulus and the d-c resistance of the
The propagation constant in the same for all practical purposes
annulus.
medium is
To show that this is true the transfer Zm = Rm{l -\-j){d/b) csch [(l+j)d/S] (101
impedance of a thin conducting film placed
7m=(l+j)-W—— <2)
or taking the first three terms of the ex¬
as already indicated (see Figure 2) will
pansion4
be considered under the following assump¬
The characteristic impedance of a
tions: i d2 7 d<
coaxial line with a solid conductor as a = Rm 1 —j-- T (ID
352 90 54
1. The characteristic impedance of the propagation medium is
coaxial lines is of practical magnitude, for
The absolute value of the transfer
example, any value between 20 and 200 1 (coAm)1/2
ohms. Zo Vrn ~ in -=(i+i) impedance is therefore equal to the d-c
Ztt r2 2<rm
resistance of the annulus to better than
2. The bridging film has a finite high con¬ (3)
ductivity approaching that of silver, copper,
platinum, and so forth. Perfect contact is where r\ and r2 are respectively the large
assumed between the film and the coaxial and small radii of the conductors.
conductors. The d-c resistance of the annulus of
3. The diameter values of the coaxial lines thickness d is
are very small compared with a wave length
4. The coaxial line terminating impedance Rm = - (4)
(that is, the receiver input impedance) is of n2wd
such a value that the impedances along the
line, beyond the film, remain fairly high, and the depth of penetration is
that is, 1 ohm or higher.
(5)
The general field-theory approach2-3
treats the solid metallic disk (or annulus)
Therefore
as a section of a coaxial line with an in¬
trinsic impedance corresponding to that
Zo=(l-\-j)Rm - (6)
of theparticular metallic medium. Under 5
the preceding assumptions the fields in
and
all the coaxial line sections are that of
the dominant transverse electromagnetic Figure 3. Radial conductor structure ap¬
Ym=(l+7); (7)
(TEM) plane wave mode of propagation. o proaching that of a solid annular ring
635/2
Figure 4 (left). Inductance of a system
of tin (Sn) radial filaments short-circuit¬
ing the output end of a coaxial trans¬
mission line
1 per cent for d/5 < 0.5, and ble value of the ring inductance evaluated inductance of all the elements in parallel,
by computing the effective inductance or it may be considerably less than this
| VT\ =Ii\Zn\=I\Rm (12)
of the system of radial conductors. Figure value because of the additional parallel
That this is true can also be shown in a 3 shows four of these radial conductors elements not accounted for, which com¬
simpler but less rigorous manner, as starting at the periphery of the center plete the solid annular ring. Thus
follows. conductor and ending at the outer con¬
d
The input impedance Zm' of the line ductor of a coaxial line. The coaxial La^~ (La'-\-Maa1) (16)
2irr
section (Figure 1) having a metal as a conductors are at right angles to the
dielectric and terminated in an open radial elements and therefore have no The self-inductance of a square bar is5
circuit or in an impedance Zr»Z0 is mutual inductance with them. The
2/ 0.447d
return path for the current in the annulus La' 0.0021 log 2 _|-
Zo _ Rm(l+j)d/s 0.447c? /
(13) is assumed far enough away to avoid any
m — tgh ymd tgh [(l+i)d/5l microhenrys (17)
influence on the radial elements. Com¬
For sufficiently small values of d/5 plete penetration of the current, that is, where d = thickness of the square bar in
therefore uniform current density in these elements centimeters and / = its length in centi¬
is assumed at all frequencies. Conse¬ meters.
Zm’=Rn (14)
quently all deductions based on low- The mutual inductance between ele¬
Since for small values of d/5 the current frequency analysis may be used at the ments a and a\ will have a negative sign,
in the annulus is uniform over its thick¬ higher frequencies at which current den¬ because they are connected in series
ness to better than 1 per cent, the annulus sity remains essentially uniform. bucking, and is given by6
can be treated as a resistance element is Each radial element has self-inductance
and mutual inductance with all the other Maai = ~ 0.002 [(2/+2r) log (2/+2A+2-- X
ordinarily treated at direct current, and
log 2r — 2(/+2r) log (/+2r)]
the voltage drop across it is simply the elements. Assuming an even number of
microhenrys (18)
product of its d-c resistance and h. elements, for each element b on one side
As was already mentioned, the success of a given element a there is another
Computations for a typical annular
or failure of this device depended on the element b\ located at an angular symmetry
ring for resistances of the order of one
requirement of negligible reactance in the which will cancel the mutual effect of
milliohm, having
annulus. Therefore all possible substan¬ b with a. Thus the effective inductance
tiating evidence was investigated both of a will be 7 = 0.1 centimeter
analytically and experimentally. d =2.5X 10-4 centimeters
La = La' Mab~\- MaC-\- ■ ■ ■ + ^061 +
In applying the lumped-circuit ap¬ r = 0.5 centimeter, result in
M ar, T ■ • • T-I/aa, = La' Alaa, (15)
proach, the top limiting value of the induc¬ Z,a=0.45 millimicrohenry
tance of an annular conductor carrying where Lal is the self-inductance of ele¬ Maa,=8 micromicrohenrys
radial currents was derived as follows. ment a and Maax is the mutual inductance La is0.035 micromicrohenry
The annular ring can be looked upon as between element a and m located dia¬
The series reactance of this inductance
consisting of a large number of equal metrically opposite a.
is about 22 microhms at 100 megacycles
radial conductors of finite length and Let d be the thickness of the element,
which seems entirely negligible even at
thickness, all connected in parallel. A and r be the radius of the inner coaxial
considerably higher frequencies, it being
square cross section may be chosen for conductor. Then the total inductance
in quadrature with the resistive compo¬
these conductors and a maximum possi¬ of the annulus, La, will be equal to the
nent. The actual series inductive reac¬
tance may be much lower, as already
indicated, and seems still less important
VOLTMETER Vg as the annular resistance is increased.
The validity of the lumped circuit
Rr> to vr
Figure 5. Schematic representa¬ approach was verified experimentally
R-F SOURCE Equations similar to 16, 17, and 18 were
tion of "coaxial" type Micro¬
potentiometer used to compute the inductances of
63 6/3
symmetrical systems of radial tin (Sn) ances desired to a given top frequency. continuous contact with the outside and
foil filaments of a centimeter in length One might call attention to the fact that inside conductors and must remain
and approximate width and thickness the resistance is a function of diameter mechanically rugged and stable. Other
of 0.11 centimeter and 10 microns re¬ ratios and not of diameter values. Thus requirements are of the ordinary variety
spectively, shorting the output end of a annular resistances of a fraction of a which can be readily met. One of these
slotted measuring transmission line. The milliohm to 1 ohm and even higher may requirements concerns the proximity of
inductances were measured at 200 and be used over very wide frequency ranges. the current monitoring element to the
400 megacycles. Figure 4 shows a The voltage range of a given resistance annulus and is not critical because there
fairly good agreement between computed element is clearly a function of the power is a current loop at the annulus; the
and measured values of inductances of it can dissipate in a particular physical current falls off as a cosine function with
these conductors, with a tendency of the arrangement. However, the power levels distance towards the current monitor
measured values to diminish somewhat in question are generally of such a low (or r-f power source). This feature tends
faster than the computed, with increas¬ level and the annular films are in such to equalize the current distribution along
ing number of filaments. close proximity to large and efficient heat heaters of thermoelements as well, and
The only conclusive experimental evi¬ conductors that the voltage range of a thus extends the frequency range. For
dence verifying the preceding analytical Micropotentiometer with a given resist¬ example, a distance of 2 centimeters from
predictions was obtained by comparing ance element seems limited at the pres¬ the resistive film to the current indicator
voltage drops across annular elements ent time by the current monitoring may cause a 1-per-cent error at 300
with those of a voltage standardization elements rather than by power dissipa¬ megacycles. With thermistors as cur¬
bolometer bridge.7 These results will be tion of the resistance films. No appre¬ rent monitors this question seems of no
discussed later. Some further support¬ ciable resistance changes were observed consequence, even over the entire ultra¬
ing evidence was found in the apparent on 1- to 10-milliohm silver elements for high-frequency range since thermistor
failure to detect inductance in shorting currents up to an ampere. beads having diameters of 15 mils can be
disks of precision slotted transmission Any means to indicate current accu¬ placed right at the annulus.
lines. None seemed to have been re¬ rately and independently of frequency are Another noncritical requirement con¬
ported in the past anywhere in the field usable with the Micropotentiometers cerns the length of the output coaxial
of slotted lines, resonance lines, nor in provided they can be physically located connector of the Micropotentiometers.
coaxial resonant cavities. However the in such a manner that the same current is This connector may be considered an
impedance values here under considera¬ passing through the annular resistor as integral part of the feeder supplying the
tion require a precision of measurement through the current monitor. This con¬ standardizing voltage to a receiver or
of voltage-node displacements of the dition becomes more difficult to fulfill any other monitor. For best results the
order of a micron or better; such a pre¬ as frequency is increased. However, annular resistance film should be placed
cision was hardly available to anyone to the fact that the annular resistances are in the output plane of the Micropotenti¬
date. very low renders the application of con¬ ometer and the connectors to the monitor
ventional r-f current indicators con¬ and to the equipment being calibrated
Ranges of Voltage and Frequency
siderably less critical than in their applica¬ in terms of the Micropotentiometers
tion with high resistance elements. Thus should be mechanically matched. In
The general conclusion that can be
thermoelements and thermistors were case this cannot be done, then the annular
drawn from the preceding is that annular
successfully used, as will be shown later, resistance film should be placed as close
resistance elements of the type described
and other bolometer type and thermo¬ as possible to the output plane. The
can be used without frequency corrections
electric elements are applicable for fre¬ accuracy will then depend on the length
over the entire ultrahigh-frequency range
quencies approaching 1,000 megacycles of the Micropotentiometer connectors
and higher, and indeed up to the fre¬
and perhaps higher. and the input impedance to the monitor¬
quency for which the metal-film thick¬
The low limit of the voltage range is ing feeder. To be more specific, the
ness required is insufficient to maintain a
controlled by the smallest practical output of the Micropotentiometer may be
homogeneous, continous metallic medium.
diameter ratios of the annular elements, affected in two ways. The first is the
This limit (about 0.05 of a micron) indi¬
by the conductivity of the materials, actual increase of the shunt admittance
cates that the top frequency of the Micro¬
by the accuracy with which low d-c resist¬ across the annulus; since the resistance
potentiometer application is dictated
ance values can be determined as well of the latter is very low the probability
by the current indicators and higher
as by the accuracy with which low cur¬ of trouble from this cause is remote.
modes of propagation rather than by the
rent values can be measured. For The second is the transformation action
resistance elements.
frequencies up to the very-high-frequency of the short connector from the resistance
At any frequency the resistance of the
range the low limit to date seemed of the film to the physical output plane which
annulus will remain the same as at direct
order of 0.1 of a microvolt; for higher may contribute an error of several per
current to within 1 per cent or better as
frequencies it was of the order of 1 micro¬ cent in the ultrahigh-frequency range,
long as the thickness of the conductive
volt. depending on the monitor-feeder im¬
film does not exceed one-half the depth
of penetration at that frequency for a pedance. In the extreme case, if that
thick conductor of the same material. General Design Features impedance is infinite, there will be a
If the film thickness is equal to the full In designing assemblies of components cosinusoidal reduction of the voltage
depth of penetration, there may be an and housing of Micropotentiometers along the short connector as one travels
error of about 3 per cent in the r-f output only one somewhat critical requirement from the output plane towards the annular
voltage. It is necessary to select mate¬ has to be fulfilled. The resistance films film; for example, at 300 megacycles a
rials of proper conductivity and choose comprising the annular elements and connector 1.5 centimeters in length will
annular diameters of proper ratios and having various thicknesses down to a have a voltage about 1 per cent higher
practical values to obtain various resist¬ micron and lower have to be in perfect at the output plane as compared with
637/4
Figure 7 (left). Cross
section of interchange¬
able Micropotentiom¬
eter resistance element
assembly. Solid black
line indicates resistance
ring clamped between
electrodes
638/5
for a given air or solid dielectric line, (depositedon bakelite) disks were clamped vary the internal impedance of these sources
one can compute and plot Vr versus fre¬ in a solid coaxial assembly; silver painted for various requirements with a high degree
of certainty in the values of this impedance.
quency for Vs = 1 volt to facilitate the rings were used as conducting electrodes A compact assembly incorporating an
determination of Vr over wide frequency for the disk at the clamped surfaces. oscillator and Micropotentiometer may be
and voltage ranges. A further simplifi¬ One modification of the low resistance constructed as a standard voltage (signal)
cation is obtained by using check fre¬ elements made use of commercial Kovar- generator.
quencies for which the line length is an glass sealed' terminals; metallic films fired 3. Calibration of attenuators directly in
odd multiple of X0/4, in which case on, or evaporated and plated over one terms of voltage ratios.
side of the terminals formed the resist¬ 4. Determination of performance of cur¬
Vr=^Vs (21) ance elements; the terminals were then rent indicators, such as of various types of
thermoelements.
soldered into an appropriate housing.
It can be shown that equation 21 holds Figures 6, 7, and 8 show cross-sectional 5. Determination of output impedances of
r-f sources. Resistance elements of the
for any value of Rr, thus a wide range of and photographic views of these elements
Micropotentiometers may be used to short-
voltages is available at the output of and Figure 9 shows an assembled unit circuit the output of a source; the voltage
this type of Micropotentiometer; for employing a thermoelement. output of these elements is then a measure
example, with Z0 = 50 ohms, F,v = 0.1 Once a Micropotentiometer is assem¬ of the short-circuited current of the source
bled, all that is necessary in order to deter¬ and consequently of its internal impedance.
volt and interchangeable Rr elements of
0.5 to 100 milliohms, one can obtain Vr mine its voltage output is to measure the 6. Calibration of modulation meters. This
d-c resistance of the annular film and to application is common to all devices having
values of 1 to 200 microvolts.
an accurately known correlation between
calibrate the current indicating element
output and percent modulation, for example,
Practical Micropotentiometers and on direct current. However, in the up-to- the output of the couple of the thermo¬
Verification of Performance date stages of this development, it was element. Attention may be called here to
necessary to verify the results in terms the fact that presence of harmonics in the
r-f carrier will normally have a negligible
The ideal Micropotentiometer should of other independent methods at all
effect on the couple output, whereas the
have a construction whereby the annular frequencies. Reliability of new units effect may be very appreciable on a vacuum-
resistance element constitutes an inte¬ can, of course, in the future be checked tube voltmeter usually employed with con¬
gral part of the inner or outer surface of against older units with little difficulty. ventional voltage generators.
the relatively heavy metallic enclosure. Agreement tests were conducted against 7. Applications where voltages in the
There are numerous ways to approach the voltage standardizing bolometer microvolt range are desired without the
bridge, mentioned previously7 and pre¬ usual Johnson noise of higher impedance
this ideal and many of them were tried
sources present.
with various degrees of success. Details cision wave guide below cutoff attenua¬
tors. These tests indicated agreement well 8. Applications where constant voltage
of construction and fabrication processes
sources are required, for example for Q-
are beyond the scope of this discussion. within over-all experimental errors, that
measurement circuitry.
However, three different satisfactory is ± 1 per cent to about 50 megacycles and
9. Applications where a single device to
methods will be indicated on the basis of ± 3 per cent to 300 megacycles. Measure¬
cover the entire frequency range from zero
up-to-date experience. Low resistance ments at higher frequencies have been to 300 megacycles and higher is essential.
elements, of the order of 0.2 to several conducted so far only on the clamped 1-
hundred milliohms, were constructed ohm units and resulted in agreements of
in one of two ways. First, ceramic or ± 5 per cent to 900 megacycles. References
glass cylinders were sealed as insulators 1. Radio-Frequency Micropotentiometers.
between the two concentric conductors. Application of Micropotentiometers Technical News Bulletin, National Bureau of
Standards (Washington, D. C.), volume 35, Mareli
For this sealing, metallic paints or pastes 1951, pages 33-34.
(silver, platinum, gold) were fired on Though the primary objective of these 2. Electromagnetic Waves (book), S. A. Schel-
the cylindrical surfaces of these insulators; Micropotentiometers is to eradicate the kunoff. D. Van Nostrand Company, Inc., New
York, N. Y., 1943, page 280.
the actual sealing was accomplished by wide uncertainty in the absolute values of
3. Networks, Lines and Fields (book), J. D.
soldering. An annular resistance ele¬ voltages in the microvolt range, they can Ryder. Prentice Hall, Inc., New York, N. Y.,
ment was then formed at one end of the be used for numerous other purposes. chapters 8 and 10, 1949.
coaxial assembly by evaporating and The more obvious applications may be 4. Tables of Complex Hyperbolic and Circu¬
lar Functions, A. E. Kennelly. Harvard Univer¬
plating a given metal over its entire briefly indicated as follows: sity Press (Cambridge, Mass.), second edition,
surface. Second, the same process was 1927, page 231.
1. Reference standard for accurate volt¬
5. Inductance of Linear Inductors, E. B.
used as in the first case except that the ages in the range previously indicated for Rosa. Bulletin, National Bureau of Standards
metallic paints or pastes were also fired calibration of voltmeters, signal generators, (Washington, D. C.), volume 4, number 2, 1907-08,
on one of the ends of the glass or ceramic field intensity meters, and so forth. The page 313.
voltage and frequency range seems to be 6. Circular C74, National Bureau of Standards
insulators to form the annular resistor;
limited primarily by current-indicating (Washington, D. C.), 1937. page 270.
evaporation and plating were thereby facilities. 7. A Bolometer Bridge for Standardizing
eliminated. A third method was ap¬ Radio-Frequency Voltmeters, M. C. Selby,
2. Sources of accurate voltages for direct L. F. Behrent. RP 2055, Journal of Research,
plied successfully only to elements of use in place of conventional signal gener¬ National Bureau of Standards (Washington, D. C ),
the order of 1 ohms in resistance; carbon ators. The user has complete freedom to volume 44, 1950, page 15.
-♦-
639/6
suitably free of inductance. method can be devised which insures perfect of this development. Unfortunately we at
The temperature coefficient of the com¬ symmetry, it will be necessary to test each the National Bureau of Standards have had
bination of an annular resistor and a thermo¬ resistor individually. no opportunity as yet to correlate degree of
couple was found to be 0.11 per cent per departure from symmetry with accuracy.
degree centigrade Annular elements of superior symmetry
One point which has perhaps not been Myron C. Selby: The author wishes to were fabricated by means of evaporation
sufficiently emphasized by Mr. Selby is that thank Dr. Hill for his encouraging discus¬ and plating. However, there was no evi¬
in order for the inductance to be negligibly sion. One can hardly question Dr. Hill’s dence to date that this fabrication resulted
low the resistive film must have complete emphasis of the axial symmetry of the re¬ in units of higher average accuracy com¬
axial symmetry. Until a manufacturing sistive film because it is a basic requirement pared with others.
640/7
U. S. Department of Commerce Research Paper RP2100
National Bureau of Standards Volume 44, May 1950
*A summary of this paper was presented at the joint meeting of the Inter ational Scientific Radio Union and the Institute of Radio Engineers at Wash-
ington, D. C , on May 2. 1949.
641/527
Field-Intensity Standards
mcnt were purposely made as . elementary as but is related to it by the voltage-transfer ratio,
possible for this reason. It is believed that this
enables the achievement of the maximum accuracy" _\VL\_\ ZL \
possible or practicable at the present time. This "-iKrfcw (1)
of necessity results in some limitation of the range
where ZL is the load impedance, in ohms, connected
of measurement and flexibility of use, but the
to the antenna terminals, and ZA the antenna
premium paid results in increased accuracy.
input impedance in ohms.
The standards or methods used to establish
The magnitude of the electric component of
accurately known values of YHF field-intensity
field-intensity (in volts per meter) is related to
are (a) the standard-antenna method and (b) the
the induced voltage by
standard-field method.
It is the purpose of this paper to describe two
such experimental standards in some detail, the |ei=V‘. (2)
lH
techniques involved for measuring the antenna
voltage or current, as well as the propagation where lH is the effective length of the antenna in
tests made to intercompare the field-intensity meters.1
values obtained by both methods. The accuracy
2. Effective Length of the Antenna
and limitations of the two methods are also
discussed. Assuming a sinusoidal current distribution on
This work done by the authors represents one the antenna, the effective length (in meters) is
phase of a program for the development of accurate
VIIF field-intensity calibration standards. Al¬
Ih—~ tan (3)
though these standards were developed to cover
the entire VHF band, the initial propagation
where X is the wavelength in meters, and l the
tests described herein were made at a frequency
antenna half-length in meters.
of 100 Me, using horizontal polarization only;
For a half-wave dipole the effective length
and the standard antennas were horizontal half¬
(in meters) reduces to
wave self-resonant dipoles in all cases.
No claims are made as to the originality of the
individual methods or techniques used, as most
of these have been reported elsewhere. However,
the standards discussed here arc novel at least As is known, the current distribution on an
in their simplicity and in the accuracy obtainable. antenna is not exactly sinusoidal except for an
Practical rationalized inks units are used through¬ infinitely thin filament. For cylindrical antennas
out this paper. of radius a and half-length l, the departure from
a sinusoidal distribution becomes progressively
II. Standard-Antenna Method greater as the ratio 2l/a decreases (i. e., as the
antenna gets fatter).
1. Relation Between Field-Intensity and Induced
An approximate solution for the free-space cur¬
Voltage
rent distribution on a cylindrical transmitting
The magnitude of the electric component of antenna 2 has been obtained by Schelkunoff [1] 3
VHF field-intensity existing at a given point in among others, and is given by the sum of cq 34
space may be determined in terms of the voltage, and 35 in the appendix. The relative distribution
Voc, induced in a standard receiving dipole immersed
1 It Is assumed here (and our results seem to indicate) that the relative
in the field, together with the antenna geometry. current distribution and hence the effective length is not (to a first approxi¬
It will be assumed in the following that the mation) a function of either the height of the antenna above the ground or
the terminating impedance, Zi„
antenna is a horizontal half-wave dipole oriented 2 How this is handled in the case of the receiving antenna has not been made
too clear in the literature, but the difference between the two cases is appar¬
for maximum response, and that the voltage is
ently small for moderately thin antennas. The reader is referred, for instance,
referred to the center terminals. In case the to J. C. Slater, Microwave transmission, pp. 219 to 230 (McGraw-Hill Book
Co., Inc., New York, N. Y., 1942).
antenna is not open-circuited, the voltage meas¬
2 Figures in brackets indicate the literature references at the end of this
ured, VL, is of course not the induced voltage paper.
642/528
Journal of Research
at the end of the antenna, r/l—1.0. The contri¬
bution from the neglected terms in this series is
appreciable only near the ends of the antenna, so
that this approximation does not appreciably
affect our results here.
The effective length of the antenna (in meters)
may be determined from the relative current
distribution by the relation
la^f- i
m
[7(b)
dr (6)
W)=m,K°)’ (5)
where
several values of K0. Because of the increasing Figure 2. Percentage increase in effective length of a
complexity of the higher order Bessel functions cylindrical half-wave dipole (over that of an infinitely thin
filament) versus the average characteristic impedance, K0,
in the infinite series of eq 35, only the first two
as determined by substituting the sum of eq 34 and 35 in 6.
terms of this series were evaluated. This ac¬
counts for the fact that the current as shown in 4 The percentage increase in In for the case of thin self-resonant dipple
antennas was found to be very nearly the same as that for full half-wave
figure 1 does not reduce quite to zero as it should dipoles for the values of A'« usually encountered.
The standard receiving antenna used was a self¬ Figure 3. Percentage shortening of a cylindrical half-wcn c
dipole required for self-resonant operation versus the
resonant (half-wave) dipole shortened approxi¬
average characteristic impedance, Ko, as determined from
mately 4 percent from a full half-wavelength (J=
cq 7.
100.0 Me) (as determined from fig. 3) to make it
self-resonant. Dural tubing in. in diameter The effective shunt capacity of the crystal
was used, giving a ratio 2 l/a^600, and an average diode was less than 0.75 ppi. However, since the
characteristic impedance A0=120 [loge(2Z/a) — 1] resulting reactance of some 2,000 ohms was in
^650 ohms. The induced voltage referred to the quadrature with the radiation resistance of the
center terminals of the antenna was measured antenna, the resulting shunting error from this
directly by means of a relatively high-impedance source was a small fraction of 1 percent.
balanced voltmeter connected across the gap at In selecting the crystal rectifier it was necessary
the center. In this manner the necessity for a to consider both its r-f (input) resistance and its
separate measurement of the antenna input im¬
pedance was eliminated, which greatly simplified,
the problem. The voltmeter consisted of a modi¬
fied type 1N28 silicon-crystal rectifier built into
the gap as shown in figure 4. The crystal output
was filtered by means of a balanced resistance- A B
a:
was measured directly against a standard cell on
a special precision slide-wire potentiometer. The
central portion of an experimental receiving
antenna assembly minus the antenna rods is c
shown in figure 5. The input impedance of the
Figure 4. Method used to mount the type 1N28 silicon
resistance-capacitance filter was approximately crystal rectifier in the gap at the center of the half-wave
10,000 ohms in shunt with 0.05 ppi. High- standard receiving antenna.
frequency resistors were employed, which were
A, Western Electric 1X28 silicon crystal rectifier; B, 1X28 as modified; C,
available commercially. modified 1X28 as mounted in the antenna gap.
o
/1
T\
low value of d-c-output resistance. Ixl \
8
O
A satisfactory compromise was obtained by z \
<
\ \
H \
selecting a crystal having an r-f resistance between (/)
\
CO
lx) 8
8,000 and 10,000 ohms when measured at an r-f CC \
X
level of 0.1 v and self-biased. This resulted in a v
APPLIED R F VOLTS
to meeting these requirements.
The variation of both the r-f resistance and the Figure 6. Variation of the r-f input resistance and dynamic
d-c-output resistance versus applied r-f voltage at d-c output resistance as a function of the applied r-f
voltage level for a selected type IN28 silicon crystal rectifier.
100 Me is shown in figure 6 for the 1X28 crystal
The measurements were made at 100 Me with full self d-c bias applied to
selected when operated self-biased. These crystal
the crystal. These curves are not representative of the type 1N28 crystals in
characteristics are not representative of the 1X28 general. O, d-c output resistance; #, r-f resistance.
Zo=y^c” (11)
where
Calibrations “A” and “B” were made 2 days apart at approximately the
same ambient temperature. , Calibration A; O, calibration B. /-100 Me.
648/534
Journal of Research
diameter of the individual points represents ap¬ somewhat and for the accuracy desired are quite
proximately 5 percent of the voltage, so that any justified. This range of distances is usually of
point touching the curve is in agreement to within little interest to the propagationist, at these fre¬
i 2/2 percent. quencies, and many of the transmission vagaries
It was usually possible to repeat a crystal cali¬ of concern to him are ignored here. At these
bration to within 1 percent if made within a few smaller separations a more exact knowledge of the
hours at the same ambient temperature. In no actual angle of incidence and ground reflection
case was the calibration relied on for more than a coefficient is usually needed in order to produce
day. the required degree of accuracy in predicting the
In making the field-intensity measurements de¬ actual value of field intensity existing. The
scribed in section IV, using this standard-antenna, geometry of the transmission system is shown in
the crystal voltmeter was always calibrated in the figure 10.
field within an hour of the time the tests were The root-mean-square value of the electric
completed. A record of the temperature was kept component of field intensity produced by a hori¬
and usually did not vary more than 2 deg C during zontal transmitting dipole at distances greater
any test. Thus all that was required of the crystal than about 2\ over plane homogeneous earth is
voltmeter was a good short-time stability. [8, 9, 10] (in volts per meter)
Although the effect of variations in temperature
and the accompanying hysteresis are known to . 6O7rlHI
E^—j
contribute largely to the instability in a crystal
Direct Ground- Surface
voltmeter, no data were accumulated on the exact wave reflected wave
wave
magnitude of these effects. We have so far avoided
(12)
this by operating under as small a temperature
in which
variation as possible. Thus it would probably
not be feasible to calibrate the crystal in a labora¬
lH = effective length of the antenna in
meters
tory and use it for measurements outside at a
\=wavelength in meters
temperature as much as 10 or 15 deg C different.
The response of the antenna crystal voltmeter
Ri = direct-ray path length in meters
/?2=ground-reflected-ray path-length in
was found to be essentially independent of fre¬
meters
quency up to 200 Me with a rising response beyond
T = pe~J,p = complex plane-wave reflec¬
(output indication high) due to series resonance
tion coefficient (horizontal polar¬
within the crystal and mount. The series-resonant
ization)
frequency for this voltmeter was of the order of
2,000 Me,
A(R) = complex surface-wave attenuation
factor
III. Standard-Field Method I—rms current in amperes at the center
of the transmitting antenna
1. Theory k = 2irj\
A predetermined value of field intensity may be j= V—i
established at a given point in space in terms of The first term of eq 12 represents the field
the current distribution in a transmitting antenna, intensity that would exist if the transmitting and
the effect of the ground, and the geometry in¬ receiving antennas were located in free space.
volved. The remaining terms take into account the pres¬
In the development of the standards being ence of the earth. The second term represents the
reported here, only horizontally polarized trans¬ ground-reflected wave which, when added vec-
mission over plane homogeneous earth having torially to the first term of eq 12, comprises the
finite values of relative dielectric constant, eT, and space wave. The third term of eq 12 represents
conductivity, a, was considered. The distance of the surface wave associated with the r-f currents
separation used between the transmitting and actually flowing in the ground.
receiving antennas was from 2 to 20 wavelengths. The total field given by eq 12 is usually referred
These conditions simplify the resulting formulas to as the ground wave. Propagation via the
572772 0 - 61 - 42
TRANSMITTING
ANTENNA ground may be treated as low-loss dielectrics over
a large portion of the VHF band as far as their
reflecting properties are concerned. This is par¬
ticularly true at the higher frequencies above 50
or 75 Me.
In this case the phase shift on reflection <f>, is
very nearly 180° for horizontal polarization, so
that the value of the reflection coefficient is given
approximately hv
r=s—p. (14)
Figure 10. Fay-path diagram showing: (a) direct ray 60irlHI 1 . Te 2 i?i)|
along R\; (6) ground-reflected ray along R2 making an \E\ = - X
----
R^ R2 (15)
angle 0 = tan_1 (hi-{-hi)Id with the earth.
Heights of the transmitting and receiving antennas are h\ and hi, respec¬ Equation 15 may he expanded without further
tively, and d is the horizontal distance of separation.
approximation, giving (in volts per meter)
Journal of Research
650/536
Tp — DEGREES
0.2
0.4
0.5
0.6
0.7
0.8
0.9
0.92
0.94
0.95
0.96
0.97
0.98
0.99
0.992
0.994
0.995
0.996
0.997
0.998
0.999
Figure 11. Variation of the magitude, p, of the complex plane-wave reflection coefficient, F, versus tan f — (hi — h>) /d for
several values of relative dielectric constant, er.
The curves apply to low-loss dielectrics, (<r/«.u <30), and horizontal polarization only and were determined from eq 13.
heights hi or h2 above are varied, this maximum since sin 9^9 to within the stated accuracy for
will occur when hih2/\d—0.25. 6< 1/4 radian.
When 2w hjijhd^1/4 the field intensity i^ given When attempting to establish values of standard
(in volts per meter) to within 1 percent by field intensity, however, one should rely on eq 16
i jp\ 2407rdjjl f hihf\ or 17 until the assumptions applying to eq 18 and
(19)
1J&I= (/- V x- / 19 are found to be fully justified.
Antenna
SWR p €r
Values of the magnitude of eq 26 are plotted
versus er in figure 15 for low-loss dielectrics,
2. 67 0. 648 21. 9
1. 85 . 635 20. 1 (a/e co<Cl), remembering that p for this
1. 42 . 636 20. 2 condition.
Average_ 640 20. 7 Determining the value of the reflection coeffi¬
cient, T, of the ground from the observed data of
figure 13 actually required only relative values of
The values ‘of p and eT measured are determined the measured field intensity. With the value of
largely by the moisture content of the ground. T now accurately determined, the only remaining
Consequently, their values fluctuate somewhat unknown in eq 20 is the effective length, lH, of
from day to day with changing moisture content. the transmitting antenna. Since the standard
Four values of eT obtained in a similar manner over transmitting and receiving antennas used are
a period of a few months are shown in figure 14, geometrically identical, the assumption is made
together with values of daily rainfall. It can be that their effective lengths are equal. There has
seen that the variation in the dielectric constant been some indication in the works of King [13]
is quite large. Normally, this is not of much con¬ that such is not excatly the case, and further that
cern to the propagationist, who is usually inter¬ the value of lH is also somewhat dependent upon
ested in grazing incidence, but it is apt to affect the the antenna termination. However, it is shown
accuracy of standard-field measurements at very that for self-resonant antennas these differences
short distances if the incident angle is large. Near are small. Shelkunoff does not elaborate on
the surface the moisture content and consequently these points, but our experimental values of
Figure 14. Chart showing four measured values of the apparent relative dielectric
constant, eT, of the ground at one particular site, as affected by rainfall. f= 100.0 Me.
(17)
where
lH — 0.96 m
/= 0.100 amp
X=3.00 m
k = 2ir/\
R,=[{hx-h2y+d^
R2—[ (^i 4“ ^2)2 T d“]1 / 2 •
Values of the magnitude, p, of the reflection
coefficient were determined by taking the mag¬
nitude of eq 13, assuming low-loss dielectrics,
(afeu><Cl). Values of p may also be taken from
figure 11.
I E\=V±, (2) Figure 18. Variation of the magnitude of the electric com¬
ponent of field intensity, \E\ in volts per meter as deter¬
mined by the standard-antenna and standard-field
where Voc is the measured value of open-circuit methods vs the height, h2 in meters, of the standard receiv¬
voltage. ing antenna.
The value of field determined by the standard- Fixed heights fti=1.58 m (A), and 3.05 m (B) of the transmitting antenna
were used, and d =30.5 m. /=100.0Mc,/=0.100amp, €.=l5,(r=0. Horizontal
field method was obtained (in volts per meter)
self-resonant half-wave dipoles were used having effective lengths of 0.960 m.
from eq 17, O. Standard-antenna method;-, standard-field method.
4. Variable-Distance Runs
DISTANCE d , METERS
7 (r) 60 4m+ 3
[772m + l (^Z) -\- jj2m+\(&l)]J2m + \(Pr) ■ (35)
70 Ko,ms
=0 (m+ 1) (2m+ 1)
/o=/(0). J2m+i(.fll) and N2m+i(Pl) are special functions [7] Nathand Marchand, Transmission-line conversion
related to J and N Bessel functions of order n + 1/2 transformers, Electronics p. 142 to 145 (Dec. 1944).
[1, P- 52], [8] K. A. Norton, The propagation of radio waves over
The total relative current on the antenna is given by the earth and in the upper atmosphere, Proc. IRE
the sum of eq 34 and 35. For an antenna composed of 25, 1217 (Sept. 1937).
an infinitely thin filament (A'o=c°) the relative current [9] K. A. Norton, The calculation of ground-wave field
distribution reduces to cos /3r, as is usually assumed for intensity over a finitely-conducting spherical earth,
such a case. Proc. IRE 29, 623 to 639 (Dec. 1941).
VII. References [10] C. R. Burrows, Radio propagation over plane earth—
Field strength curves, Bell System Tech. J. 16,
[1] S. A. Schelkunoff, Electromagnetic waves, p. 290 to 45 to 75 (Jan. 1937).
292 and 441 to 479 (D. Van Nostrand Co., Inc.,
[11] K. Bullington, Radio propagation at frequencies above
New York, N. Y., 1943).
30 megacycles, Proc. IRE 35, 1122 to 1136 (Oct.
[2] S. A. Schelkunoff, Theory of antennas of arbitrary
1947).
size and shape, Proc. IRE 29, 508 (Sept. 1941).
[12] J. A. Stratton, Electromagnetic theory, p. 493 (Mc¬
[3] R. King, Graphical representation of the characteris¬
Graw-Hill Book Co., Inc., New York, N. Y., 1941).
tics of cylindrical antennas, Cruft Laboratory,
Tech. Rept. No. 20 (ONR) (Oct. 1, 1947). [13] R. King, H. R. Mimno, A. H. Wing, Transmission
[4] L. S. Nergaard, A survey of ultra-high-frequency lines, antennas, and waveguides, p. 160 (McGraw-
measurements, RCA Rev. 3, 183 (Oct. 1938). Hill Book Co., Inc., New York, N. Y., 1945).
[5] MIT Radiation Laboratory Series—Crystal rectifiers, [14] Frank M. Greene, The influence of the ground on the
Louis N. Ridenour, editor, p. 333 to 360 (McGraw- calibration and use of VHF field-intensity meters,
Hill Book Co., New York, N. Y., 1948). J. Research NBS 44, 123 (1950) RP2062.
[6] Ernst A. Guillemin, Communication Networks II,
64 (John Wiley & Sons Inc., New York, N. Y.,
1935). Washington, October 3, 1949.
One type of error known to be present in VHF field-intensity meters (30 to 300 Me) is
caused by the influence of the ground on the value of the antenna voltage-transfer ratio. This
is a result of fluctuation of the receiving-antenna input impedance with height and changing
ground conditions. An approximate method is presented for calculating the input impedance of
horizontal dipole antennas over earth having finite values of dielectric constant and con¬
ductivity. The effect of both changes in ground conditions and antenna terminating im¬
pedance on the above error is calculated as a function of the antenna height. Measured
values are presented in support of the above method, and the results are discussed.
662/123
method is applicable to horizontal antennas of k'—Ii(Zu Z12) -p/3 {Z13 'Z14)A
any length, the results will be evaluated only for 0)
0—7i(Z31 Z32) +73(Z33 Z34),)
the case of parallel horizontal half-wave dipoles.
Their locations and the geometry involved are where V is the impressed einf at the center of t he
shown in figure 1. transmitting antenna, and 74, and 73, are the re¬
spective currents at the centers of the transmitting
TRANSMITTING
ANTENNA and receiving antennas. The terms Zn and Z33
are the free-space self impedances, respectively,
of the transmitting and receiving antennas referred
to the center terminals. Also
7 _7 ___
V
v run /q\
'Jmn — r.m ~ T > \6)
*n
Journal of Research
663/124
ef -relative dielectric constant of the ground Values of eq 5 are shown plotted in figures 11 and
(referred to frce-space as unity), 12 of the reference given in footnote 5 for spacings
e— e^r, where e„ is the permittivity of evacu¬ from 0 to 7.5 wavelengths.
ated free-space, It is possible to derive a more simple expression
than eq 5, valid for distances of separation in
,y,, X10-9 farads/meter, excess of about 2X. At this distance from an
O07T
antenna, only the radiation component of the
<7=ground conductivity is mhos/meter, electric field-intensity usually need be considered.
co 27rf, For a half-wave transmitting dipole in free-space,
oriented normal to a line from its center to the
X wave length in meters,
point of observation, the field intensity is (in volts/
j= V— i. meter)
■;27tR
664/125
III. Relations Existing in the Receiving the higher frequencies above 50 or 75 Me. Under
Antenna these conditions the phase shift on reflection, <p, is
very nearly 180°, so that px.
1. Antenna Current
3. Voltage Relations
Equations 3 may now be solved for the current
/3 at the center of the receiving antenna. The The terminal voltage of the receiving antenna
problem will be simplified if it is assumed that the terminated in an impedance ZL is, from eq. 9
distance between transmitting and receiving an¬
T7- (^3i T r2Z32) ZLIX . .
tennas is sufficiently large that the presence of the
(Zl + Zss + ^Zs,)’
receiving antenna does not measurably affect the
current flowing in the transmitting antenna. This and the open-circuit voltage is, letting ZL—> °°,
is usually the case in practice, and the assumption
is certainly justified if the spacing is at least several u0C=-(Z31+r2z32)/1. (13)
wavelengths. In this case the current in the re¬
ceiving antenna terminated at its center in a load If the receiving antenna is sufficiently high above
impedance ZL is, from eq 3: the ground, Z34 may be considered negligible com¬
pared to (ZL-j-Z33), in which case the terminal
j (Z3i+r2Z32)/i /q\ voltage will be, from eq 12,
ZL T Z33 -f- r4 Z34
t/-_(Z3i t r2Z32)ZLIX , .
'l y i 7 '
2. Input Impedance ^33
The numerator of eq 9 is the induced emf in The "true value of the electric component of
the receiving antenna, and the denominator is the field intensity at any antenna height, h2, above
input impedance (in the presence of the ground) the ground is, from eq 7 and 13:
plus the terminating impedance ZL connected at
7? _Voc. (Z3i+r2Z32)/i
the center, the input impedance being (15)
l~ U ~ In
Z*=Z33-f-r,1Z34. (10)
The value of field intensity that would be indi¬
The effect of the ground in the immediate cated by a field-intensity meter previously cali¬
vicinity of the receiving antenna is accounted for brated in the presence of the ground is, from eq 12:
by the second term on the right of eq 10, r1Z34.
K (Z,1 T r 2Z32) ZJX
Zu is the mutual impedance that would exist Ei=KVL=
between the receiving antenna and its image if
(•^l+z +r,z34)
33 (10)
the ground were perfectly conducting, and is K may be defined as the antenna constant and
given by eq 5 or eq. 8 upon substituting R=2h2. may be evaluated at any desired height of the
Tj is of course the actual reflection coefficient of receiving antenna. If a height h2 is chosen such
the ground for normal incidence obtained from that Z34 <C (ZL+Z33), K might then be termed
eq. 4 by placing ^=7r/2, which gives the free-space antenna constant and in such a
case its value would be
(17)
r.(-£r>
since Et ^ Et at this height. The antenna con¬
Values of the magnitude of eq. 11, p1; are shown stant is seen to be the reciprocal of the product
plotted vs er in figure 2 for low-loss dielectrics, of the effective length, lH, and the voltage-transfer
(a/ew<(<Cl). Many types of ground may be ratio, TV/Voc=ZL/ (ZL+Z33).
treated as low-loss dielectrics over a large portion The assumption is made here that the relative
of the VHF band as far as their reflecting proper¬ current distribution and hence the effective length
ties are concerned. This is particularly true at of the half-wave receiving dipole is not (to a
Journal o£ Research
572772 0 - 61 - 43 665/126
first approximation) a function of either the ter¬ in the case of low-loss dielectrics, from figure 2,
minating impedance, ZL, or the height of the since F, ^ — pt.
antenna above ground. Although a complete
solution to the general problem of the receiving
antenna has unfortunately not yet been achieved,
this assumption is supported by a number of om
measurements.7
666/127
(Zl—73J2) calibrated under free-space conditions
may indicate values of field intensity that are in
error by as much as 10 percent for values of h2/\
near 0.3, and 7.5 percent for values of h2/\ near
0.6. If this error is to be held to values less than
5 percent, antenna heights greater than about 0.65
wavelength should be used for field-intensity
measurements under these conditions.
It is somewhat doubtful at the present state of
the art just what maximum values of measurement
error of this type should be permitted. One
method of reducing the error, obviously, is to
increase the value of the antenna terminating
0 1.0 2.0 3JO
ANTENNA HEIGHT, WAVELENGTHS impedance, ZL.
Figure 3.—Calculated percentage difference (\Ei/Et\ — l)X
100 vs receiving antenna height, h2/\, in wavelengths.
Ei is the true value of field intensity, and Ei is the value indicated by a
field-intensity meter with a previously determined free space value of antenna
constant (horizontal polarization). Curves are shown for four values of
ground constants: (A) <r= (B) t,=30; (C) €, = 15; and (D) &=9 (for low-loss
dielectrics cr/ea><<l). Antenna length Z=X/2. The free-space antenna input
impedance is taken as £33=73.2+.?42.5, and the terminating impedance
Zt — Ti+j 0 ohms.
667/128
polarization) is shown vs height, h2, in meters over
ground having a measured relative dielectric
constant er^15, (o-/eco<!<Cl) for: (A) antenna
“open-circuited”; and (B) antenna terminated in
Zl=62 f-j 0 ohms. The measured percentage dif¬
ference shown in the upper curve of figure 7 was
determined from the data with the aid of eq 18
which, upon substituting eq 15 and 16, gives
(in percent)
5= -l)xi00. (21)
+ 30
9
20
\
\
k P \
0 / k / / - '"-O-
>
/ V "cr 'tNr
- 10
s\ >
p \
0.30 \
1
A /
A
i
. /
Figure 6. View of the various -pieces of transmitting and,
/
receiving equipment used in obtaining the measured data /
of figure 7. /
In the background is shown the ladder-mast and carriage for the receiving /
/
dipole. The location is the Beltsville, Md., airport. /
d-
/
/
7
self-resonant dipole, Z33=65Tj 0 ohms, and ) B
/
antenna actually used for obtaining the measured / /
V
points of figure 5. The terminating impedance, / Ar
Z£, = 62+j 0 ohms, was the closest value to 65 —V
/ /
ohms available at the time the measurements d /
/
were made. As might be expected, there is no Id
substantial difference between the calculated ° i
values of the measurement error for the full X/2 4 6 10
ANTENNA HEIGHT, TT1
dipole and for the self-resonant dipole. The
measured points support the theory reasonably
Figure 7.—Measured values of receiving antenna terminal-
well. The difference does not exceed 3 percent
voltage (horizontal polarization) vs height in meters over
for antenna heights above 0.1 wavelength. Vari¬ ground having a measured relative dielectric constant
ous pieces of the transmitting and receiving- €r== 15, (<r/eco<[<C 1) for: (A) antenna “open-circuited
equipment used in making these measurements and (B) antenna terminated in Zi = 62+i 0 ohms
668/129
The free-space value of the factor, Kla=(ZL-\- This error8 in measurement caused by the
Zzz)lZL= VoJVl, used, was estimated from the ground may be reduced by increasing the value of
data, and is the reciprocal of the voltage-transfer ZL. The error vs height is shown in figure 4 for
ratio previously mentioned. three values of ZL, viz, 73, 150, and 300 ohms.
For the open-circuited condition referred to The error approaches zero as ZL approaches
above, the receiving antenna was actually ter¬ infinity.
minated in a special balanced voltmeter of the In figure 5, measured values of the error are
silicon crystal-rectifier type. This crystal rectifier, compared with theoretical values calculated as
together with the balanced RC network used to previously described. The agreement is reason¬
take off the direct-current output voltage, pre¬ ably good for antenna heights above 0.1 wave¬
sented a resistance of approximately 4,000 ohms length.
in shunt with 0.75 ju^f across the gap at the center In view of the approximations involved it is felt
of the antenna. This accounts for the slight oscil¬ that the curves shown in figures 3, 4, and 5 prob¬
lation of the points around the averaging curve, ably should not be used for actually applying
but introduced an error of less than 1 percent in corrections to field-intensity measurements.
the final results, as the shunting was present dur¬ Rather they might be used to estimate the maxi¬
ing both the open-circuited and terminated runs. mum probable error (due to ground effect) existing
in measurements made below a given antenna
VI. Conclusions height.
An approximate method has been presented for Figure 3 shows the variations in the error with
determining the effect of finitely conducting ground antenna height occurring over perfectly conducting
beneath a horizontal receiving dipole on the value ground.9 The error is appreciably larger in this
of the antenna constant as used for measuring case than for finitely conducting ground. This
would seem to indicate the inadvisability of using
VHF field intensity. Three variables are mainly
involved in this effect: (a) the antenna height, in or calibrating a VHF field-intensity meter over a
wavelengths, A2/X, (b) the ground constants er and perfectly conducting plane unless the antenna
a; (c) the antenna terminating impedance ZL. heights were carefully chosen so as to result in a
low value of error.
Changes in antenna height probably have the
greatest effect on the antenna constant, as can be 8 The error, as previously defined, is the percentage-difference between the
true value of field intensity existing at a given antenna height, in wavelengths,
seen from figure 3, and are of primary concern here. hi/X and that indicated by a field-intensity meter whose antenna constant
Normal variations in the ground constants en¬ was determined’ under free-space conditions.
9 Perfectly conducting ground and a solid metallic ground plane would have
countered in practice apparently have only a minor essentially the same reflecting properties for the present purpose.
effect. Under most conditions and to within the
probable accuracy of this method, these variations
can probably be neglected. Washington, October 18, 1949.
Journal of Research
669/130
National Bureau of Standards Circular 517
1. General
The National Bureau of Standards offers a specified conditions of set gain at specified fre¬
calibration service1 for certain types of radio quencies. All measurements are made in terms
field-strength meters in the frequency range 10 kc of sinusoidal voltages or currents.
to 300 Me. The calibration consists in measuring In developing the equations used in this paper
the over-all linearity of the instrument at one or for determining field strength, the practical
more frequencies and radio-frequency (r-f) levels, rationalized inks units have been used. This is
measuring the internal attenuator ratios at one or in agreement with recent international action.
more frequencies, and measuring the so-called
antenna coefficient or correlation factor relating i For further information and calibration fees, write to Director, National
field strength to the output meter reading under Bureau of Standards, Washington 25, D. C.
Those quantities involving only r-f voltage ratios This type of attenuator2 serves as a primary
are measured in terms of precision dissipative step reference standard, since the attenuation is a func¬
attenuators or precision mutual-inductance (wave- tion only of its linear dimensions which can be
guide-below-cutoff) attenuators, depending on precisely determined. The TE x,i mode of excita¬
frequency. Use is made of two types of dissipa¬ tion is used, the higher order unwanted modes
tive attenuators comprising “Pi” or “T” sections, being reduced by means of a strip filter'. Errors
one employing wire-wound resistance elements for resulting from the finite conductivity of the
frequencies below 1 or 2 Me, the other using- cylinder wall, and from the proximity of the op¬
coaxial sections of evaporated or deposited metal erating frequency to the cut-off frequency are less
film for frequencies up to 300 Me. These step than 0.1 db in a 60-db range in the HF band
attenuators are standardized on direct current and (3 to 30 Me) and the VHF band (30 to 300 Me) and
their ratios determined to within 0.1 percent. can usually be neglected.
Frequency corrections are determined by compar¬ Since the over-all linearity of a field-strength
ing one standard attenuator used at signal fre¬ meter will in general be a function of both fre¬
quency against another used at the intermediate quency and r-f input level, it is necessary to de¬
frequency in a receiver. Cross checks are made termine this dependence. Fortunately, this effect
against the standard mutual-inductance attenua¬ sr.E. Grantham and J. J. Freeman, A standard of attenuation for micro-
tor at frequencies above about five megacycles. wave measurements, Trans. AIEE 67, 535 (1948).
670/1
frequency attenuators only. The departure from
linearity has been found in general to be less than
2 percent if the signal voltage applied to the mixer
grid does not exceed about 5 percent of the in¬
jected local oscillator voltage. For practical and
economical reasons it is necessary to limit line¬
arity checks on the attenuator steps in question
to one or two frequencies as specified by the use)-.
For those sets also employing r-f attenuators in
the antenna input circuit it is necessary to ensure
that these ratios are measured under actual
operating conditions. Those r-f attenuator steps
which may be affected by the changing source
impedance of the antenna, or the changing input
admittance of the first r-f stage with frequency,
are measured with the antenna placed in a field
which can be attenuated in known ratios.
Figure 1. Calibration of the internal attenuators and The voltage linearity and attenuator ratio
lineaiity of a commercial field-strength meter.
measurements made on field-strength meters are
is usually limited to the highest attenuator steps usually certified to ±2 percent. The equipment
in those types of sets having intermediate- used is shown in figure 1.
GOtI'II /t , /2trd\2
\E\^ (1)
(rf2+
+r*+rf)3/2V +V X / ’
where
7?= equivalent free-space electric field strength
in rms volts per meter
/’i—radius of transmitting loop, meters. To
keep a uniform current in the transmit¬
ting loop make 27rr1<(X/8
3 This refers to a four-terminal unit having an r-f heater electrically insulated
from the thermocouple.
671/2
2ttA
(3)
672/3
4. Measurement of Antenna Coefficients From 30 to 300 Me
At the frequencies in the VHF band, the Na¬
tional Bureau of Standards certifies only field-
strength meters using dipole antennas. The de¬
termination of antenna coefficients is made in
terms of a radiation field by either the standard-
field method or the standard-antenna method
using half-wave dipoles and horizontally polarized
transmission in either case. These two methods
were intercompared and excellent agreement was
obtained over most of the range of antenna heights
and distances of separation used between trans¬
mitting and receiving antennas.* 6
In using the VHF standard-field method it was
necessary to select a flat graded site free of any
reflecting objects such as trees, buildings or wires
within a radius of at least several hundred feet
around the transmitting and receiving antennas.
portion of the VHF band. In this case, eq (6)
With these conditions approaching the ideal, the
absolute magnitude of the field at the receiving may be written
antenna was accurately computed at these fre¬
quencies by considering only the direct and
ground-reflected waves from the transmitting
antenna, as indicated in figure 4. Both horizontal
half-wave dipoles are assumed oriented normal to
a line joining their centers, and spaced a horizontal
distance, d, greater than 2\. The resulting rms
field strength 7 in volts per meter at the receiving Unless (h1Jrh2)/d<^0.05, it may be necessary to
antenna is determine accurately the actual dielectric constant
or reflection coefficient of the ground. For
Q0tIhI instance, if (hi-{-h2)/d=0.2, the magnitude of the
\E\=* reflection coefficient may have any value in the
4p
r,r2
sin2
^k(R2-Rl) + ^ J (6)
range 0.85 to 0.93, depending upon the moisture
content of the top layers of the soil at the measure¬
ment site. If (hiJrho)!d=0.1, this range may be
roughly 0.93 to 0.96. For (/il-\-h2)/d<^'0.05,
where the reflection coefficient will usually be greater
iH = effective length of antenna in meters than 0.96. For this condition, and if further
^\/tt meters for a half-wave dipole 2ttM2/VK1/4, eq (7) will reduce to
\ = wavelength in meters
Ri — [(hi — A2)2+(i2]I=direct-ray path-length in 240trlHI fhih2\
meters
(8)
d2 V X2 /
R2—[(hi-\-h2)2+d2]i=ground -reflected - ray'
path-length in meters giving the field to within ±2 percent of the value
d=horizontal distance of separation, obtained from eq (7) for these conditions.
meters In using the standard-antenna method the
hi and h2— the heights in meters above ground of field strength of a locally generated field is deter¬
the transmitting and receiving an¬ mined by measuring directly the induced emf
tennas, respectively in a standard receiving dipole oriented for maxi¬
mum response from the relation
p=magnitude of the plane-wave reflection
coefficient of the ground (horizontal
polarization), and <t> is the angle of
\E\=e/lH, (9)
where
phase lag on reflection
e=induced emf in the receiving dipole in
I— rms current in amperes at the center volts
of the transmitting antenna lH =effective length of the dipole, meters,
k= lH = ^l7T meters for a half-wave dipole.
The phase shift on reflection, 0, is very nearly
180 degrees for many types of ground over a large The emf is measured directly by means of a
relatively high impedance silicon crystal voltmeter
8 F. M. Greene and M. Solow, Development of very high-frequency field- built into the gap at the center of the antenna.
intensity standards, J. Research NBS 44, 527 (1950) RP2100.
7 Based on equation derived from reference 4 in mks units. This eliminates the necessity for a separate meas-
673/4
u remold, of the antenna impedanee, which greatly The “antenna coefficient” of the field-strength
simplifies (lie problem, improving the over-all meter under test is determined by placing its
accuracy of the measurement. dipole antenna at a specified height, usually 10
Owing to the lack of selectivity in this method as feet, in the radiation field previously measured
used, il does not lend itself to calibration work in the using the standard receiving dipole and the rela¬
vicinity of strong interfering fields that may tionship given by eq (9). Equations (4) and (5)
often be encountered. The site used for VHF are used as at lower frequencies, except that the
calibration work at the National Bureau of frequency term,/, is usually omitted. Esc of the
Standards is in a partially' shielded valley some meter later at other antenna heights or over
15 miles from the nearest interfering station. ground having appreciably different constants
While it is possible to use either the standard- may require corrections for heights less than about
field or the standard-antenna method for determin¬ one wavelength.8
ing the antenna coefficients, the latter method is The accuracy of the antenna coefficients certi¬
used at the National Bureau of Standards at the fied by the National Bureau of Standards depends
present time. Field tests have shown the excel¬ largely upon the merits of the individual meter
lent agreement possible between the two methods under test. In general, these coefficients are
throughout the VHF band. The choice was certified at present to ±10 percent for frequencies
largely one of convenience, as it was found some¬ between 30 and 150 Me. and to ±15 percent for
what more convenient to provide an accurate frequencies between 150 and 300 Me.
calibration of the silicon crystal voltmeter than
to ensure an accurate measurement of the trans¬ Washington, May 9, 1951.
mitting antenna current throughout the VHF
8 F. 3U. Greene, Influence of the ground on the calibration and use of VHF
band. field-intensity meters, J. Research NBS 44,123-130 (1950) RI’2062.
674/5
Reprinted from IRE TRANSACTIONS
ON INSTRUMENTATION
Volume 1-7, Numbers 3 & 4, December, 19SS
675/270
1958 Powell, Jickling, and Hess: High-Frequency Impedance Standards at the NBS
been multiplied by the same factor in determining our SMALLEST MEASURABLE INCREMENT OF LENGTH
limits of L and C. This is important in that it shows that -MAXIMUM AND MINIMUM IMPEDANCE MEASURABLE TO A PART IN A
with an impedance equal to the characteristic imped¬ RANCE: OF COMMERCIAL 0.1% TWO TERMINAL INSTRUMENTS
676/271
IRE TRANSACTIONS ON INSTRUMENTATION December
ONE - PICOFARAD
INCREMENTAL CAPACITOR
677/272
1958 Powell, Jickling, and Hess: High-Frequency Impedance Standards at the NBS
678/273
IRE TRANSACTIONS ON INSTRUMENTATION December
tions of the working standards are measured on the in¬ terminals. There are hundreds of different types of
strument so that the error in each dial reading of the in¬ terminals in use today and it would be impractical to
strument can be plotted as a function of the readings of maintain impedance standards with each type; there¬
all dials on the instrument as shown in Fig. 8. This fore, all equipment in this group is constructed with a
method makes it unnecessary for the calibration per¬ terminal that will allow the highest accuracy of the pri¬
sonnel to completely understand the construction of all mary standard while giving the least loss of accuracy in
models of all instruments made and also gives the opera¬ transfer to calibrated instruments. This terminal is a
tor a set of calibration charts which can be quickly used 50-ohm air dielectric coaxial line with an outer conduc¬
with a minimum of calculation. tor inside diameter of f inch, the faces being flat and
For comparing standards submitted for calibration, perpendicular to the axis. In order to accommodate all
the Center is being equipped with an admittance and an types of terminals, adapters are designed and calibrated
impedance bridge at each of nine fixed frequencies. to give minimum loss of accuracy to a defined plane in
These 18 bridges are adjusted to be direct reading to the commercial connectors.
0.1 per cent so that comparison with the working stand¬ All of the dozens of instruments and hundreds of
ards can be done with minimum loss of accuracy. Such standards used in this system are newly constructed,
adjustment is possible since each bridge is operated in and it will not be possible to certify maximum accuracy
conjunction with a crystal-controlled generator and until a certain amount of history is obtained on each
receiver so that the frequency does not deviate more unit.
than one part per million from nominal. Since high accuracy in radio frequency impedance
With such a system nearly every existing radio fre¬ measurements is relatively new, and much technique is
quency impedance instrument or standard can be cali¬ required to utilize it effectively, there is still much work
brated efficiently to better than its rated accuracy. Still, to be done, and we hope that the limits we have placed
each piece of equipment presents its own individual on possible impedances and accuracies as a guide for
problems which are too numerous to detail in this paper. designing the iacilities in the Center will serve as a chal¬
The most important of these, however, is the matter of lenge rather than the barrier it is implied to be.
679/274
Recently Developed Microwave Impedance Standards
and Methods of Measurement*
R. W. BEATTYf and D. M. KERNSf
of Standards in the field of microwave impedance faces. The VSWR (cre) of a short-circuited quarter wave¬
standards and measurements will be briefly described. length section of waveguide operating in its dominant
The majority of the experimental work has been done TEi.o mode can be shown to be approximated closely by
in WR-90 (A-band) rectangular waveguide systems al¬ 1
though the results are applicable to any size of wave¬ VSWR «-=-— (1)
guide and with increased difficulty, perhaps, to coaxial
systems. sVt4t)
These developments cannot be described in detail in where /x and nm are, respectively, the permeabilities of
a paper of this length, but it is expected that the details the medium within the waveguide and the metal of
will be published elsewhere in separate papers. which the waveguide is constructed. 5 is the skin depth
in the metal; Ag is the guide wavelength; a is the attenu¬
Standards of Impedance for ation4 per unit length of the waveguide.
Rectangular Waveguide For example, such a short-circuited section of WR-90
Fixed impedance, or reflection, standards consist of (A-band) copper waveguide has a calculated VSWR at
terminated waveguides,1 having practically no reflec¬ 10 kmc of approximately 5680 corresponding to a volt¬
tion; short-circuits, having almost total reflection; and age reflection coefficient having a magnitude of approxi¬
discontinuities2,3 of fairly simple geometry, having a mately 0.99965.
reflection calculable from the essential dimensions and Even though the effective conductivity may not be
wavelength. known to great accuracy, the difference between the
An adjustable sliding termination, having a VSWR actual and calculated reflection coefficient magnitude
(voltage standing-wave ratio) of less than 1.0002, or a will probably show up only in the 4th or 5th significant
return loss greater than 80 db, has been constructed figure. This will cause errors of less than 0.01 per cent
using a well-known principle.1 In order to provide fine in measured reflection coefficient magnitudes when the
adjustment and minimum variation in reflection as the short-circuited section is used as a reference.
termination is slid, a number of refinements were neces¬ Inductive half-round obstacles have been investi¬
sary. These included construction of a uniform wave¬ gated for use as impedance standards in rectangular
guide section with a dimensional tolerance of approxi¬ waveguide. These obstacles consist of either one or two
mately 0.0001 inch in the cross-sectional dimensions, a opposed semicircular cylindrical indentations extending
spring-loaded ball bearing alignment plunger, a resistive across the narrow sides of the waveguide. Standards
strip tapered from both sides toward the center of the employing half-round obstacles may be fabricated
waveguide, and independent fine controls for the re¬ by electroforming or by machining, do not require un¬
quired mechanical movements. duly close tolerances, and avoid a change in the wave¬
Short-circuited sections of waveguide have been con¬ guide cross section. They seem especially suitable for
structed in which the input flange is a quarter wave¬ use as independent, absolute standards of reflection or
length from the short circuit. Flange losses are mini¬ waveguide impedance. Accurate values have been cal¬
mized since there is no longitudinal current in the wave- culated for the lowest-mode lumped-element represen¬
tation using well-known methods to obtain stationary
* Manuscript received by the PGI, September 5, 1958. expressions for the desired reactance elements as func¬
f Natl. Bur. of Standards Boulder Labs., Boulder, Colo. tionals of the obstacle currents. The Rayleigh-Ritz
1 R. W. Beatty, “An adjustable sliding termination for rectangu¬
lar waveguide,” IRE Trans, on Microwave Theory and Tech¬ process with one-term, two-term, and three-term kourier
niques, vol. MTT-5, pp. 192-194; July, 1957. expansions for the obstacle current as an “extremaliz-
2 N. Marcuvitz (ed.), “Waveguide Handbook,” M.I.T. Rad. Lab.
Ser., McGraw-Hill Book Co., Inc., New York, N.Y., vol. 10; 1951.
3 D. M. Kerns, “Half-round inductive obstacles in rectangular 4 See, e.g., E. C. Jordan, “Electromagnetic Waves and Radiating
waveguide as standards of impedance,” (in preparation). Systems,” Prentice-Hall, Inc., New York, N. Y., p. 290; 1950.
680/319
IRE TRANSACTIONS ON INSTRUMENTATION December
ing” sequence yields rapidly convergent numerical re¬ denotes the reflection coefficient6 corresponding to the
sults. Accuracy of the results is estimated at five figures internal impedance of the equivalent generator con¬
or better for VSWR’s up to three. Tables were calcu¬ nected directly to the load.
lated of VSWR and reflection coefficient. Values are ob¬ In impedance measuring systems such as the slotted
tained for 15 obstacle sizes and for 41 frequencies within line, it is approximately true that \ K\ =1 and r2i = 0.
the recommended operating range of 'the waveguide. For these systems,
The results are based upon normalized parameters and
| Z>3-| = C | 1 + | IT | | , (4)
are applicable to any size of rectangular waveguide.
Further theoretical and numerical work is being done
and as the phase \J/l of Tl varies,
to enable corrections to be made for finite conductivity
and for dimensional imperfections in the half-round ^3 I max 1 ~f~ I rL
or
If a variable standard were available, one could ad¬
Sz\ 1 + KIV just it until bz = 0, then Fc/ = r,s. If only fixed standards
bz — bG (3) are available, the following procedure may be used to
(1—SAIT) SAIT i-r2trL
obtain the magnitude of IT.
SziT0 (I-S33IT)
where terms of the form Sm,n are the scattering coef¬ (1 — 5nrc) Sn S13C1)
ficients of the junction, and the voltage reflection co¬ ShTg S22 SmTd
efficients of the generator, detector, and load are repre¬ _ SuTg S32 (1 — ShTd) I
sented by rG, Td, and Tl, respectively. The term r2i I (1 — ^uTg) SuTd I
I S31IV (I — •SWEd) I
5 For background material, see A. C. Macpherson and D. M. 7 E. F. Barnett, “More about the hp precision directional cou
Kerns, “A new technique for the measurement of microwave stand¬ piers,” Hewlett-Packard J., vol. 4, no. 5-6; January/Februa'
ing-wave ratios,” Proc. IRE, vol. 44, pp. 1024-1030; August, 1956. 1953.
572772 0 - 61 - 44
681/320
1958 Beatty and Kerns: Recently Developed Microwave Impedance Standards
If one varies the phase of Ts (a line stretcher, not alternately connects the unknown and the standard to
necessarily of the constant impedance type, may be the output. In this case
used), the ratio of maximum to minimum response is
\ b3 \U IV , . . .
\ b3\ = ba ---• IT
(1 - 6nr0)(l - 633IV) Acknowledgment
= C' I Tl I . (9) The following personnel of the NBS Boulder Labora¬
tories contributed to the work described above: D. F.
With this type of response it is not necessary to vary
Wait, W. W. Longley, Jr., P. F. Wacker, D. Nash,
the phase of Tz,, but merely to observe |&3| when one
W. Anson, E. Risley, R. Manka, E. Niesen, R. Blixt,
8 G. F. Engen, private communication. and W. E. McNaney.
682/321
Journal of Research of the National Bureau of Standards Vol. 52, No. 3, March 1954 Research Paper 2480
683/121
ratio (VSWR) introduced on a uniform transmission
line feeding the connector when the latter is termi¬
nated in a load equal to the characteristic impedance
of the line. This convention can only be used to
estimate the maximum error in a measured VSWR
caused by connector reflections. It does not furnish
necessary corrections for all impedance values.
The approach in this paper is to treat the connector
as a component part of a transformer; the figures
of merit of this transformer are two transformation
corrections for the components of its terminating
impedances. The problem is to establish relation¬
ships between the connector input and output
impedances or admittances. Admittance relation¬
ships prove to be more convenient for computation
simplicity.
Let, in figure 2, a uniform transmission line
having a characteristic impedance Z01 be joined
with another uniform line, Z02, by means of a con¬
nector terminated in the planes A', at its input end,
and S0, at its output end. Also, let T0 be the location
of a voltage node when a short circuit is placed at Sa¬
lt can be shown [2] that simple expressions inter¬
relate admittances on opposite side of the con¬
nector, e. g., the admittances YA = GA-\-jBA and
Figure 1. Typical connector showing one of terminating
planes. YA. — GA>-\-jBA. at planes A and A', where A and
A' are effectively an integral number of half wave¬
lengths apart. Two planes in a transmission-line
system are effectively an integral number of half wave¬
lengths apart when a voltage node placed in one of
these planes (closer to the load) will cause a voltage
node to appear in the other plane irrespective of
what the physical distance is between them. A sup¬
plementary line section, l-2, is added to fulfill the
above requirement. The relationships of these
admittance components (fig. 2), as shown by Oliver
[2], are given by
G( — a?Gi (1)
and
To Sq
Bi^afBi+aX, (2)
Figure 2. Representation of connector and connecting trans¬
formers.
where a2 is a dimensionless constant, and axbx is an
additive constant in mhos. Similarly,
connector is shown in figure 1. The transmission
lines are assumed to have uniform distributed G'A=a*GA (3)
and
constants up to the connector. To complete the
definition of a connector, it is necessary to fix the
B'A^a2BAYaAbA. (4)
minimum allowable distance between its terminating
If desired, one can treat the connector shown in
planes and its adjacent discontinuities (e. g., h, fig. 2). figure 2 between A' and S0 (minus the supplementary
This distance depends on the particular configuration line section 4) as a transformer by expressing the
of the transmission lines and the attenuation rate of input admittance in terms of the output admittance.
the higher modes generated at the discontinuities. From well-known interrelations
For coaxial lines, h (fig. 2) should be equal to or
larger than the diameter of the outer conductor [l].1 ^(1+COt2 flaZa) m
When used in this sense, connectors can be readily 9' (cot fok-bAy+gi w
treated as four-terminal networks and their constants
defined and measured. _ bA cot2 /S24+(l — gl— bA) cot 132l2—bA .
Another accepted convention is to indicate the (cot M-bA)'+gl ’ W
quality of a connector by the voltage standing-wave
where: /S2 is the phase constant and the admittance
1 Figures in brackets indicate the literature references at the end of this paper. components are normalized with respect to Y^.
684/122
The termination admittance of the connector may To summarize, the definitions of connector, con¬
also be obtained, using various charts and diagrams necting transformer, and connector quality, as
and thus avoiding computation of eq (5) and (6). applied here, are as follows: (a) A connector is a
However, there is no great advantage in knowing lossless linear bilateral passive four-terminal net¬
the admittance at the plane S0 in preference to that work terminated at each end in a uniform line section
at any other plane. The additional computation of a minimum length required to attenuate higher
work involved is therefore seldom justified. modes likely to appear at these ends. These line
The definition of a “connecting transformer” may sections are identical electrically and mechanically
here be introduced as follows: to the respective lines of the interconnected system,
(b) A connecting transformer is a connector as
A connecting transformer is any linear bilateral defined in (a), terminated at one end in a supple¬
passive four-terminal lossless network terminated at
both ends in uniform line sections; the latter are mentary lossless line section. This supplementary
electrically and mechanically identical to the corres¬ line section is less than half a wavelength long and is
ponding adjacent lines of the system. When a short identical electrically and mechanically to that at the
circuit is placed at either end of the connecting trans¬ connector end. When a short circuit is placed at
former, a voltage node appears at its other end.
either end of a connecting transformer (with r-f
Thus the connecting transformer consists of the power fed from the opposite end), a voltage node
connector and 1 or 2 supplementary line sections. will appear at the other end of the transformer.
However, there is generally no practical advantage With perfect connectors a connecting transformer
of having more than one supplementary section; the is an ideal transformer; the normalized input admit¬
disadvantage is, of course, the need of two (instead tance components of this transformer are equal to
of one) admittance transformations when one is its normalized output admittance components.
interested in the actual input and output admittance (Normalization is with reference to input and output
of the connector. A connecting transformer, as lines, respectively.) (c) With imperfect connectors
defined above, therefore consists of the connector one must apply corrections to obtain accurate
proper and one supplementary line section at its termination admittances. These “transformation
output or at its input end. In figure 2, lx is the corrections” a2 and ab' indicate the quality of the
input-end supplementary line section of the con¬ connectors. The simplest form of the corrections
necting transformer correlating Y[ and Yx through are given in eq (1) to (4). These transformation
a set of corrections a\ and axb‘i. If the output corrections hold only for the chosen combination of
supplementary line section l2 is chosen, a set of cor¬ connector and supplementary line section.
rections, aA and aAb'A, is required, and the correla¬
tion between Y'A and YA is established. It is con¬ 3. Determination of Connector Quality
venient to use l2 in preference to lx when the quality
of connectors is studied. On the other hand, lx is 3.1. Analytical Background
used when the connector is an integral part of an
admittance measuring system used to determine Yx The inadequacy of specifying the VSWR of a
with optimum accuracy. The justification for these “matched” connector as its quality index has been
-preferences will be further clarified below when the pointed out above. For systems of equal char¬
methods of obtaining the transformation corrections acteristic impedances the effective characteristic
are discussed. Special problems may indicate a impedance of the connector is frequently taken as the
different choice of supplementary sections and may square root of the product of the measured open and
justify the use of two sections simultaneously (i. e., shorted input impedances (i. e., Z0=-v/ZocZsc). This
one at the input, the other at the output). How¬ yields, strictly speaking, an inaccurate parameter,
ever, the two major groups of problems anticipated because the approach assumes infinite and zero
at present are impedance and connector measure¬ impedances at open- and short-circuited conditions,
ments; therefore the discussion below will be limited respectively, and uniformly distributed constants,
to connecting transformers using either lx or l2. none of which are true in practice. In addition, this
The interpretation of a2 and ah' in the case of approach neglects likely discontinuities at the
impedance measurements is straightforward. For termination planes of the connector as the latter is
connector measurements the interpretation is as conventionally referred to. Therefore, the value
follows: With a perfect connector (referring to fig. 2) of Z0 so obtained cannot be used as a simple trans¬
normalized admittance YA/Y01 will always be equal formation factor. A reliable way to find quantitative
to the normalized YA/Y02; otherwise the values of values of the transformation corrections of connectors
a2 and ab' will indicate the errors introduced by the is the one employing the node-shift method [2, 3, 4, 5].
connector discontinuities, i. e., the relative quality of Frequency-variation methods can be used only to
various connectors for all values of admittances.
find qualitative data.
For a connector mounted on the panel of an enclosed
network like an attenuator, voltage generator, etc., The node-shift (n-s) method is well described in
one may simply consider the connector as a part of the references cited and consists briefly in measuring
the enclosed network with the reference plane at corresponding shifts of voltage-minima at both ends
A'. Otherwise, required reference impedance-planes of the connecting transformer. The original appli¬
like S0 and A may be assumed inside the network. cation of this method involved a less efficient and
685/123
less accurate handling of measurement data, re¬ sarv to permit measurements along a uniform line
quiring the determination of tangents to curves of section. With the nodes in these positions,
uncertain regularity. An improvement, introduced
by Oliver, largely eliminated this difficulty, because Z/=jZ01 tan (3d (ID
the transformation corrections are obtained from a
curve that is analytically shown to be a straight Z, =jZ02 tan fts (12)
line; measurement errors can be readily detected
in the plot of a straight line. Techniques developed where (3=2t/\ is the phase constant of the respective
for the practical application of this method in systems lines.
employing equal or unequal characteristic impedances
Using eq (9),
are described in section 3.3 Several variations in
these techniques may be used, depending on condi¬
tions and requirements. tan ,3,1= 2 (13)
The essential analytical steps underlying this
Z02 az-\-jabZ02 tan (32s
method of approach are as follows: Let a, b, c, d
since the network is lossless, b is an imaginary quan¬
represent, at one frequency, linear parameters [2]
of a four-terminal network equivalent to either one tity [2] and one may substitute —b' iov jb. Rewrit¬
ing the terms,
of the two connecting transformers shown in figure
2. As the elements of the network are assumed
Z,
passive and linear, and a zero termination impedance cot (3d=y— (a2 cot (32s — ab'Z02). (14)
is transformed into a zero input imedpance, ^02
therefore,
t-H
e
(7)
II
1
686/124
"reversed” sets of corrections, obviating the necessity
of additional measurements. It can be shown [6]
that
af=- (19)
ar
and
K=b’r=b', (20)
afb' = (21)
b arb —n
arb =—— ■J-r= . -r- (23)
ar aj Z02(m +1)
687/125
where B= — lfx, B' = — l/x'; x and x' are the respec¬ that the distribution of cot 02s will be fairly evenly
tive equivalent shunt reactances of the unknown and spaced from —2 to +2. These limits are selected
measured impedances at the ends of the connecting because usable data, for the rapidly changing cotan¬
transformer. gent function outside this interval, require equipment
Equation (26) thus gives the errors in the measured precision which is usually beyond practical achieve¬
shunt reactances, which can be plotted versus x as ment. When the magnitude of the s values is lim¬
straight lines with shunt capacitive reactances ited by the usable length of the reactor, lengths of s
equivalent to (—ah') mhos as slopes, and (a2—1) as corresponding to cot /32s values from 0 to +2 may be
intercepts. Representative values usually encoun¬ adequate. Each of the S and T nodes are located by
tered in practice at frequencies to 1,000 Me/s are determining the midpoint between two positions of
shown in figure 5. equal detector response one on each side of the mini¬
mum. However, when structural defects in the
slotted line produce nonuniformity in the voltage
3.3. Techniques and Procedures distribution along the line, errors may be caused in
the determination of the nodal positions. To mini¬
The node-shift method of connector-quality deter¬ mize these errors, it has been shown [7] that the opti¬
mination may utilize one of the following reactors, mum equal-response voltage in the determination of
depending on the problem and facilities at hand: (a) a nodal position should be approximately 3 db
An adjustable stub, (b) fixed line sections, (c) a (within ±0.5 db) above the minimum voltage when
slotted line, or (d) a perforated line. Steps of pro¬ the VSWR is above 7. In addition to structural
cedure and precautions that are common to the use defects, the presence of losses in the measuring sys¬
of all of these reactors will first be discussed; then tem may introduce errors in making accurate nodal
features of each device and its associated technique measurements. When such is the case, one may
will be described individually. minimize these errors by following the procedure
outlined later in section g.
a. Equipment The difference, cot dd—cot d2s, is then plotted
versus cot /S2s and the resulting straight line inter¬
^ In general, the equipment requirements are these: preted as shown in section 3.1. The values of a2
The known reactor, connected as shown in figure 3 and ah' thus obtained are the transformation correc¬
at S0 (the end of the unknown connecting trans¬ tions of the unknown connecting transformer at the
former) must be lossless, uniform, and adjustable or test frequency.
variable over at least half a wavelength. The junc¬ As the magnitudes of a2 and ah' are functions of
tion at S0 must be free of discontinuities; thus, for the relative positions of the discontinuities in the
best results, the cross-sectional dimensions and di¬ connecting transformer, and as these positions are
electric material of the reactor and the connecting scattered and unknown, the effect of these discon¬
transformer must be the same at this plane: In tinuities may cancel at some frequencies and increase
addition, this uniformity must be maintained on the at others. Therefore, the number of frequencies at
input side of S0 for an axial distance equal to at least which measurements must be made in order to find
the diameter of the outer conductor. Sometimes, the quality of a connector over a given frequency
as mentioned in section g, a small amount of mis¬ range will be determined by the extent of the fre¬
match may be tolerated without invalidating the
quency range and the accuracy desired. Measure¬
measured data. Good mechanical alinement and
ments indicated that the transformation corrections
contact at this plane are essential. All discontinui¬
ties should lie between So and K, the latter being a for commercial connectors varied rather slowly with
plane of reference on the measuring line, effectively frequency up to 1,000 Mc/s, and that measurements
an integral number of half wavelengths from S0. at 100-Mc/s intervals in this range would be sufficient
to determine the frequency characteristics of a given
connector.
b. General Procedure
688/126
c. Connectors in Tandem To determine values of s, relative stub positions
may be measured with the required precision by
When the measuring line does not match the un¬ means of large vernier calipers, gage blocks, or a
known connector in size or characteristic impedance, carefully engraved scale attached to the adjustable
or both, an adapter is required between them.. Then
stub. An advantage of this type of reactor is that
it is necessary to separate the corrections of the one stub of sufficient length can be used to obtain
unknown connector from those of the adapter ac¬
convenient values of s over a wide frequency range.
cording to the procedure described above in section
3.2. In this situation S02 (see fig. 6) must be so In the construction of the stub it is essential to
use the least amount of solid support for the center
placed that a voltage node appears at S0i- This
condition can be met by first placing a short circuit conductor near the input end. Whatever support
at Soi, the output end of connecting transformer No. may be required should be small in mass and have
1, and by locating a corresponding T0i on the meas¬ the lowest possible dielectric constant. When it is
uring line. Then the short circuit is replaced by the possible to support the center conductor by the con¬
unknown connector and adjustable reactor, and the nector being measured, the omission of any additional
latter is varied until a voltage node again appears at support is desirable. Another critical construction
Toi. Thus a voltage node is also assured at Nm. The feature is that the movable shorting device should
values of s and t are then obtained as described locate a nearly lossless short circuit in positively
previously. In some cases the voltage node at S02 defined planes. For small-sized connectors it is very
cannot be moved continuously because of the type difficult to build a stub long enough for use in low-
of variable reactor being employed or because con¬ frequency n-s measurements without an objection¬
necting transformer No. 2 has a fixed length. The able sag in the center conductor. An adjustable
procedure to place a voltage node at N01 under such stub is unsuitable for measurements of solid-dielectric
conditions will be described in section li. connectors, because of unavoidable discontinuities
of the junction of the two dielectric media.
d. Curvature in Plot of Data
f. Use of Fixed-Line Sections as Reactors
It was occasionally observed that the plot of meas¬
urement data substituted in eq. (15) resulted in a The use of accurately constructed line sections for
curve rather than a straight line. The following reactors simplifies n s measurements. First, a short
precautions were found to prevent such curvature. circuit is placed at N0; this is then replaced by
(1) The phase constants of the measuring line and the shorted fixed-line sections of different lengths corre¬
reactor must be known to an accuracy of ±0.01per- sponding to selected s values. Alternatively, sec¬
cent. This required that the corresponding fre¬ tions of equal length open at both ends may be con¬
quency stability be maintained to an accuracy of at nected in tandem; a short circuit is then placed at
least ±0.01 percent. (2) The positions of No and T0 the output end of the combination. In both cases
should be measured, for example, to ±0.1 mm at 300 it is essential that a good mechanical contact exist
Mc/s for lines with dielectrics having constants of 1 at N0 and that no discontinuities be introduced in
to 3. The presence of any appreciable discontinuity this plane. In using combined sections, these condi¬
at these planes, such as a poor contact at N0, may tions must also be maintained at the junctions.
cause an excessive error. (3) When the voltage Otherwise, a curvature or excessive random scatter
distribution near a node is nonsymmetrical because will result in the final plot of the data.
of losses in the measuring system, proper precautions Although the procedure followed with this type of
must be taken to locate the positions of N and T cor¬ reactor is simple, a number of disadvantages are
rectly. These precautions are described in section g.. involved. First, the careful matching and alinement
(4) The accuracy of finding voltage-node positions referred to above requires precision machining.
may be improved by keeping the input voltage high Thus when n-s measurements are to be made over
and the probe penetration deep (for low input volt¬ a wide frequency range, numerous costly precision
ages), probe reflections being negligible near voltage sections are required. Second, if an adapter is
nodes, by filtering out harmonics from the input required between the unknown connector and the
voltage, and by operating probe output indicators measuring line, it may be awkward to use fixed
at the most sensitive positions of the meters and cir¬ sections as reactors (particularly at low frequencies)
cuit action. Means of minimizing errors caused by because of the requirement that the ends of the
structural defects in slotted lines were indicated unknown connecting transformer be an effective half
under section b. wavelength apart (see section 3.2). Third, although
reactors of this type can be constructed to match
e. Use of an Adjustable Stub as a Reactor any present-day air-dielectric connector, this can not
easily be done for solid-dielectric components because
This type -of reactor is the most obvious one to it is difficult to determine with sufficient accuracy
select for use in n-s measurements of air-dielectric the phase constant of solid-dielectric sections. In
elements and is quite simple to operate. A 5-ft 50- coaxial lines having solid insulation the phase con¬
ohm stub of this type was found particularly useful stant is very closely equal to 27r/(&)1/2/(2.998X108)
in measuring %-in. connectors in the VHF range. radians/meter, where k is the dielectric constant of
689/127
the insulation. Variations from 1 to 3 percent may
be found in the dielectric constants of commonly
used insulators, depending on the batch or manufac¬
turer. In addition, these constants may be affected
by a change in volume (i. e., density) caused during STRETCHER
fabrication steps and by the application of excessive
heat. The phase-constant accuracy is also limited Figure 7. System for node-shift measurements using a slotted
by the accuracy (±0.1 percent) to which the capacity line as a reactor.
of line sections can be measured, by effects of fre¬
quency on the dielectric constant, and by dimensional
as that described in section b. It is not essential
tolerances of the conductors. Thus it appears that
that the voltage node be moved to the exact S’
the best prospect of obtaining the required accuracy
location previously selected; it is possible to locate
of ±0.01 percent in the phase constant is to measure
the position of the voltage minimum very accurately
the electrical lengths directly in the finished line
and thus know the corresponding value of s.
sections. For example, a section of line can be
resonated by varying the frequency within the range At higher frequencies the presence of losses in
in question and observing the output of a fixed probe the measuring system may make accurate measure¬
ments of S and T more difficult. As the voltage
located at a precisely known distance approximately
a half wavelength from the shorted end. Assuming curve in the region of a minimum closely approaches
that the dielectric and dimensional uniformity was a parabola, the voltage nodes may be located ac¬
curately bv joining the midpoints of three horiziontal
maintained, this line section can then be cut into a
number of short sections. This fixed probe method chords through each selected trough of the voltage
standing-wave [8]. However, this procedure is
suggests the application of a perforated solid-dielec¬
quite tedious; in most cases the following simpler
tric line; it also suggests an accurate way to measure
method may be employed, with results of sufficient
the dielectric constant of a coaxially fabricated
accuracy; (a) place a short circuit at S0 and find the
material.
true T0; (b) connect the slotted-line reactor at S0;
A type of reactor combining the features of the this will introduce losses (especially when a solid-
adjustable shorting stub and fixed-line sections dielectric line is used), reducing the VSWR and
might be used to alleviate some of the disadvantages upsetting the symmetry of the voltage distribution
mentioned above. First, an air-dielectric stub is in the measuring line particularly; (c) locate a node
required, in which the shorting device can be removed at T0 as follows: Adjust the auxiliary variable reactor
from the coaxial line. Coaxial sections of any de¬ to obtain a node near T0, using a voltage increment,
sired solid insulator are then inserted in the coaxial AE,, about 3 db above the minimum; readjust the
line, and the shorting device is replaced so as to variable reactor to move the node toward the desired
produce a short circuit at the end of the newly location; repeat these steps until a node is obtained
formed solid-dielectric coaxial line. It is necessary exactly at the true T0; (d) then locate the correspond¬
that each of the insulator sections be accurately ing minimum for S'0 (an effective half wavelength
machined to fit the coaxial line and to have end from S0 toward the short circuit). Use a certain
planes perpendicular to its axis. The lengths of AEs approximately 3 db above the voltage minimum,
these sections depend on the desired values of cot (e) Finally, locate the selected S and corresponding
jS2s. Advantages of this type of reactor are (a) T positions, using the same AEs and AEt in each case,
the conductors are not broken; (b) numerous s as was used to locate the Sr0 and T0, respectively.
values are readily available, because it is simpler
Relative merits of this technique are: (a) This
to machine any desired number of insulators rather
type of reactor is useful wherever a slotted line is
than to construct an equal number of individual
available to match the connector being measured.
line sections; (c) the terminiating plane of the
The technique was found particularly applicable in
unknown connecting transformer (solid-dielectric
determining the quality of %-in. connectors, the size
type) can easily bo placed at any approximate point
in which solid-dielectric slotted fines of fairly good
in the reactor, when required; (d) any number of
precision have been made, (b) Only one slotted line
different dielectric materials may be used with the
is needed as a reactor for n-s measurements over a
same air stub for various desired Z02 values; (e) the
wide frequency range, e. g., from 100 to 1,000
phase constants may be determined with an auxiliary
Mc/s. (c) Because of the precision with which
fixed probe, as described above.
voltage nodes can be located, the velocity of propaga¬
g. Use of a Slotted Line as a Variable Reactor
tion, and hence the phase constant of this type of
reactor, can be very accurately determined, (d) The
In employing a slotted line as a known reactor, an use of this reactor is more convenient than the others
auxiliary line stretcher or adjustable stub is connected for the. n-s measurement of connectors in tandem,
at the output end of the slotted line to move a voltage because of the ease with which S02 can be located
minimum along the slotted line where it is detected whenever it must be placed in the reactor to meet
by a probe (see fig. 7). The method recommended the half-wavelength requirement referred to
for accurately locating this voltage node is the same previously.
690/128
Ill turn, this technique possesses a few disad¬
vantages. Measuring the s values in a slotted line
is somewhat more time consuming that adjusting
them, as described in the previous methods. A
slotted line is considerably more expensive than
reactors of the other types for use over the same
frequency range. Because of the slot, it is impossible
to match perfectly a slotted line to a closed line, a
compromise necessarily being required either in the
conductor dimensions or in the characteristic im¬
pedances. However, the use of a slotted line as a
reactor is practical if the resulting discontinuity at
the junction of the line and the unknown connector
is kept as small as possible. For example, satis¬
factory data were obtained by using such a slotted
line to measure the quality of adapters on precision
slotted lines.
h. Use of a Perforated Line as a Variable Reactor Figure 8. Details of a perforated line used as a variable
reac*or in node-shift measurements.
Although applicable to making n-s measurements
of connectors in all sizes, this technique is especially
recommended for use in determining the quality of circuited line stretcher was used at the output end to
large-sized connectors because of the economy made move the voltage node along the perforated line. To
possible in the construction of the variable reactor. facilitate placing the voltage node at a preselected
As an example, a 50-ohm 3}8-in. perforated air- hole, a line stretcher that can be moved by very
dielectric line, shown in figure 8, was constructed for small increments in a positive fashion is desirable. If
the quality measurements of connectors and lines difficulty is experienced in this operation (as indi¬
having that diameter and characteristic impedance. cated by a lack of reproducibility), the following
The line consisted of a 4-ft brass outer conductor method may be used: (a) Shift the voltage node
with a duralumin center conductor supported at the slightly to one side of the selected hole. Let the
output end by a Teflon disc % in. thick. At the probe output voltage at this hole be Vx and the
input end no disc was used; the center conductor was position of a corresponding voltage node on the
supported bv the connector being measured so as to measuring line be Tx. (b) Move the voltage node to
minimize the discontinuity at the junction of the the other side of the selected hole so that the new
connector and reactor. Stock tubing was used for probe voltage V2 equals Vx. Let the position of the
the conductors. Probe holes, % in. in diameter, corresponding voltage node on the measuring line be
spaced 1 in. apart, were bored in the outer conductor. T2. (c) Then the true T for a voltage node placed
This spacing permitted the choice of convenient s at the desired hole will be midway between Tx and
values over a wide frequency range. Surrounding T2. This procedure assumes that the voltage curves
each hole was a milled “flat” to help center the probe. at S and T are not distorted by excessive losses.
The probe was a stiff copper wire. 0.020 in. in diam¬ This is not always true at T. If such is the case, the
eter, soldered to a BNC connector, and was shielded precautions discussed in section g must be observed
by a short cylinder that fitted the probe hole. An in finding T. In all cases, when the movable short
external adjustable stub was used to tune the probe circuit is being adjusted, it is advisable to have the
(see fig. 9). As with the slotted line, a short- source impedance equal to Z0 of the line connected
Figure 9. Perforated line used as a variable reactor to measure the corrections of a transformer.
691/129
to it; this assures a constancy of voltage amplitudes 4. Typical Measurement Results
along the system.
When it is necessary to use an adapter between the In all the techniques described above, a precision
unknown connector and the measuring line, the slotted line has been used to measure the input
corrections of the two connectors must be separated, reactance variations of the connecting transformer
as previously explained. Because of the fixed probe corresponding to known output-reactance variations.
positions of the perforated line, a special step must Obviously, impedance-measuring devices other than
usually be employed here to determine the exact slotted lines may be used for this purpose; e. g.,
frequency at which the unknown connecting trans¬ bridges, Q-meters, instruments employing directional
former is effectively equal to a half wavelength. A couplers, admittance comparators, etc. However,
short circuit is first placed at the end of the adapter present experience indicates a need of precision in
to the measuring line, and three values of T0l are measuring input-reactance variations considerably
found for three different frequencies around the greater than that provided by currently available
estimated frequency at which the unknown connect¬ commercial instruments other than slotted lines.
ing transformer meets the length requirement. Also, when these other impedance-measuring devices
Then the short circuit is removed, and the unknown are employed for this purpose, the steps of proce¬
connector and perforated line are attached, the probe dure in the n-s method must be modified. A dis¬
of the perforated line is placed in a selected hole, and cussion of the use of such devices is reserved for
three values at T02 are found for three similar fre¬ another treatment of the subject.
quencies. With these data two curves of frequency The results of some n-s measurements of typical
versus T0 are plotted. The intersection of these connecting transformers, using a precision slotted
curves determines the desired frequency. If the line as the measuring instrument, are compiled in
effective discontinuity of the adapter is small (so table 1. The transformation corrections given do
that the transformation corrections remain essen¬ not necessarily represent average values for the
tially the same for small displacements of S01), it is particular types of commercial units listed, but
permissible to. use an approximate frequency, e. g., rather demonstrate the typical magnitudes of these
the integral number of megacycles per second corrections. Each connecting transformer, of course,
nearest the intersection of the above curves. Then terminates in uniform line sections and, when re¬
the original corrections of the adapter, predetermined quired, includes mating connectors. For type N
at a slightly different frequency, can still be used in the connectors and adapters the mating connectors are
separation procedure. When it is necessary to know assembled from UG-21B/U plugs and UG-23B/U
the desired frequency more precisely, the above pro¬ jacks, combined with 1%-in. uniform sections of 50-
cedure is repeated over a more narrow frequency ohm Teflon-dielectric rigid line. For a particular
range bracketing the frequency located above. The connector the transformation corrections vary, de¬
frequency used in this technique must be known and pending on the position of the connector along the
kept stable to 0.01 percent. The corrections of the transmission-line type of transformer. The cor¬
measuring-line adapter may, of course, be remeasured rections for the cable connectors and adapters, Nos.
at the new exact frequency if optimum accuracy is 3 to 11, in table 1 represent the parameters of each
to be assured. transformer, with the test unit placed at the input
Connecting transformer
Frequency a2 ah' Variable reactor
Number Components
Mcjs Micromhos
1 M6-in.-diam line; taper to 2H-in.-diam solid-dielectric precision slotted 300 1.010 -410
line. 600 1.028 -260
15 S'/i-in.-diam, 140-cm partial-air-dielectric line. . _ __ ... 1 / 0.996 130 lSJ-S-in.-diam air-dielectric perforated
16 3}4-in.-diam, 150-cm air-dielectric line... _ _ _ \ .999 70 / line. Zo=50.9 ohms.
692/130
and T0. The movable short circuit is then used to
place a voltage node at selected S' locations, and
the corresponding values of s and t are accurately
measured. The data are then plotted and the values
of a2 and ab' calculated, by the usual procedure.
These resulting corrections represent the total
transformation produced by all discontinuities be¬
tween S0 and T0. Usually it is found that the mag¬
nitude of these corrections is less than would be de¬
duced from the nonuniformity of the probe output
voltage, indicated as the probe is moved along the
perfectly terminated line. The choice of whether
to use the n-s technique or the voltage uniformity
method to determine the quality of a slotted line
depends on the information desired. Variations in
-2X> -1.5 -1.0 -0.5 0 0.5 IjD L5 2.0 probe coupling do not necessarily affect the accuracy
cot^2s
with which nodal positions can be located, but do
Figure 10. Measurement results. result in probe output irregularities along the line.
Hence, the n-s method (depending only on accurate
Plots of equation (15) for connecting transformers listed in table 1: A, trans¬
former 1; B, transformer 7 in tandem with transformer 1; and C, transformer 8 in nodal measurements) determines rather accurately
tandem with transformer 1.
the electrical uniformity of a slotted line, and voltage
uniformity measurements (depending on line quality
end. The magnitude of correction that can be and probe coupling) determine the combined effect
produced by a given transformer may be estimated of probe penetration variations and electrical non-
by considering, for example, transformer 8; a2=0.981, uniformity and establish the error limitations of
ab'= — 250 micromhos. Using eq (17) and (18), it complete impedance measurements. -Thus, the n-s
can be shown that a measured impedance of method offers means of separating the two major
300— jlO.O ohms at the input end of this connecting causes of slotted-line errors. It also follows that
transformer corresponds to an actual load impedance slotted lines, which are poor as impedance-measuring
of 291— 7*31.5 ohms at its output end. Thus the instruments, often may serve satisfactorily as vari¬
error in the resistive component is 3 percent, and the able reactors.
reactive component is in error by 3 to 1. A way in which the uniformity of a slotted line
The accuracy with which the values of a2 and ab' can be qualitatively evaluated is by using it as a
can be determined depends first on the absence of variable reactor in the n-s measurement of some
curvature in the plot of eq (15) as represented in discontinuity placed outside this line. Fixed dis¬
figure 10. With no observable curvature, the prer continuities in the line, which change the electrical
cision with which the straight line can be drawn lengths of the s values, will cause curvature in the
depends on the scatter of the plotted points. The plotted results; when no curvature is obtained, the
scatter, in turn depends on the accuracy with which line is free from serious discontinuities.
the s and t values are .measured. The accuracy of
the transformation corrections was ±0.5 percent or 6. Conclusions
better for a2, and ± (5 percent + 0.01 micromicro-
farad) for ab'. The work described above was limited to lossless
connectors. Transmission-line sections may also be
considered connectors as long as their losses are
5. Measurement of the Quality of Slotted negligible. A long lossless line may, of course, be
Lines looked upon as consisting of several connecting trans¬
formers in tandem; one can then obtain the trans¬
The n-s method may advantageously be applied formation corrections of representative sections and
to determine the uniformity of a slotted line and calculate the transformations likely to result from
thus indicate the order of its utility, either as a the long line.
measuring line or as a variable reactor. The aux¬
The measurements were made on connectors
iliary equipment required in performing this quality
having ideal junctions with uniform rigid-line sec¬
measurement is an adjustable stub or a short-cir¬
tions; therefore, some of the transformation cor¬
cuited line stretcher connected to the output end of
rections cited are probably better than those for
the slotted line. This stub need not be of high ac¬
connectors joined to commercial cables having
curacy; its purpose is only to move a voltage node
braided outer conductors. One may, of course,
along the slotted line. The procedure follows the
measure a number of the latter samples and get
general, n-s technique. At a half wavelength from
comparative data for various types and makes of
input and output ends of slotted line, voltage nodes connectors, cables, and adapters, as well as for any
are selected to represent, respectively, S0 and T0. combinations of these. The transformation cor¬
It is desirable that the measurement frequency be rections are obtained for discrete frequencies; it is
high enough so that most of the line lies between S0 pointed out, however, that in cases of rather small
693/131
distributed discontinuities these corrections do not [4] Albert Weisfloch, Ein Transformationssatz liber ver-
lustlose Vierpole und seine Amvendung auf die experi-
vary sharply with frequency. Thus, measurements mentelle Untersuchung von Dezimeter und Zenti-
at several well-selected frequencies may be sufficient meterwellen-Schaltungen, Hochfrequenztechnik und
to cover a wide frequency range. Elektroakustik 60, 67 (Sept. 1942).
[5] E. Feenberg, The relation between nodal position and
standing wave ratio in a composite transmission system,
7. References J. Appl. Phys. 17, 530-532. (June 1946).
[6] A. T. Starr, Electric circuits and wave filters, p. 171, 2d ed.
[1] J. R. Whinnery, M. W. Jamieson, and Theo. Eloise (Pitman & Sons Ltd., London, 1948).
Robbins, Coaxial line discontinuities, Proc. Inst. Radio
[7] II. E. Sorrows, W. E. Ryan, and R. C. Ellenwood, Evalua¬
Engrs. 32, 695-709 (Nov. 1944). tion of coaxial slotted-line impedance measurements,
[2] M. H. Oliver, Discontinuities in concentric-line impedance Proc. Inst. Radio Engrs. [2] 39, 162-168 (Feb. 1951).
measuring apparatus, Proc. Inst. Elec. Eng. 97, part
[8] F. D Bennett, P. D. Coleman, and A. S. Meier, The design
III, p. 29-38 (Jan. 1950); also discussion, p. 242 (July
of broad-band aircraft antenna systems, Proc. Inst.
1950). Radio Engis. p. 671-700 (Oct. 1945).
[3] R. W. King, Absolute method for measuring dielectric
constants of fluids at ultra-high frequencies, Rev. Sci.
Instr. 8, 201 (June 1937). Washington, July 21, 1953.
694/132
11 August* 1958
TABLES OF
B.ECTANGULAR WAVEGUIDE
by
695/1
The following graph and tables pertain to four half-round inductive
They were prepared primarily for the use of the NBS Electronic
0.0720 inch and 0. 1080 inch. The other two standards each use two half-
round obstacles as shown in the sketch below. The radii of these pairs
Although not all of the standards give useful reflections at all X-band
the calculations is probably beyond the 5th significant figure in both VSWR
are in the process of calculation at this time (August, 1958). The accur¬
The personnel responsible for the preparation of these tables are the
following:
696/2
572772 0 - 61 - 45
697/3
o
CO
co
Oi
o
o
o
O
o
FREQUENCY (KMC/SEC)
FREQUENCY DEPENDENCE OF VSWR OF HALF-ROUND
INDUCTIVE OBSTACLE IMPEDANCE STANDARDS FOR
WR90(X-BAND) RECTANGULAR WAVEGUIDE
SINGLE HALF-ROUND INDUCTIVE OBSTACLE
IN WR 90 RECTANGULAR WAVEGUIDE
— / df
'requency (f) |r
o
| = Is VSWR ( <r )
s
(KMc/s)
IN WR 90 RECTANGULAR WAVEGUIDE
Radius = 0.1080 inch
- / U1
'requency (f) VSWR ( o- )
(KMc/s)
lrJ = lsnl s V T
8.242 1 0. 173 12 1.4 1873
8.3464 0. 16619 1.39864 1. 098
8.4508 0.15977 1.38030
8.5551 0.15378 1. 36345 0. 976
8.6594 0.14817 1.34789
8.7638 0. 14291 1. 33347 0. 878
8.8681 0.13794 1.32003
8.9724 0.13325 1.30747 0. 800
IN WR 90 RECTANGULAR WAVEGUIDE
IN WR 90 RECTANGULAR WAVEGUIDE
Radius = 0. 1080 inch
Frequency (f) |r | = |S | VSWR { a )
(KMc/s) s II s
8.2421
8.3464
8.4508
8.5551
8.6594
8.7638
8.8681
8.9724
^ X
9.0768
9. 1811
/ l/aL
9. 2854 SINGLE DOUBLE
9.3897
9.494 1
HALF-ROUND INDUCTIVE OBSTACLES
9. 5984
9. 7027
9.807 1
9.9114
10.0157
10. 1201
10.2244
10.3287 0.22021 1.56481
10.4330 0.21492 1. 547 52 1.093
10.5374 0.20979 1. 53099
10.6417 0. 20482 1.51515 1. 044
10.7460 0. 19999 1.49996
10. 8504 0.19529 1.48538 i. 000
10.9547 0.19073 1.47137
TABLES OF
RECTANGULAR WAVEGUIDE
by
702/1
The following graph and tables describe the reflections from four
These calculations are based on ideal conditions and the only correc¬
tions made were in the index of refraction of air using 23°C, 50% relative
beyond the 5th significant figure in both VSWR and jrj . Corrections for
The personnel responsible for the preparation of these tables are the
following:
703/2
^ x
SINGLE DOUBLE
704/3
<x>
<*— dMS A
705/4
FREQUENCY DEPENDENCE OF VSWR OF HALF-ROUND INDUCTIVE OBSTACLE IMPEDANCE STANDARDS FOR RECTANGULAR
WAVEGUIDE WR- 284
SINGLE HALF-ROUND INDUCTIVE OBSTACLE
706/5
SINGLE HALF-ROUND INDUCTIVE OBSTACLE
requency (f) |r
1 s
1 = Is
(KMc/s)
707/6
DOUBLE HALF-ROUND INDUCTIVE OBSTACLE
kJ ft.
Frequency (f) VSWR (a )
s
(KMc/s)
709/8
December 17, 1958
TABLES OF
RECTANGULAR WAVEGUIDE
by
710/1
The following graph and tables describe the reflections from four
These calculations are based on ideal conditions and the only correc-
o
tions made were for the index of refraction of air using 23 C, 50% rela¬
beyond the 5th significant figure in both VSWR and | rj . Corrections for
The personnel responsible for the preparation of these tables are the
following:
711/2
IZZZIZZZV
SINGLE DOUBLE
712/3
<x>
572772 O - 61 - 46
ir~>
"3-
ro
— UMSA
713/4
OJ
FREQUENCY DEPENDENCE OF VSWR OF HALF-ROUND INDUCTIVE OBSTACLE IMPEDANCE STANDARDS FOR RECTANGULAR
WAVEGUIDE WR- 187
SINGLE HALF-ROUND INDUCTIVE OBSTACLE
mJ CL
requency (f) |r | = |S.J VSWR (a )
1 s' 1 111 s
(KMc/s)
'requency (£)
(KMc/s)
lrJ = lsnl
VSWR (a ) c
o - A
3.9626 0. 17312 1.41873
►-Jo.
requency (f) l1r s'1 = |s. J VSWR (0- ) a /
1 11' s
(KMc/s)
3.9626 0.38621 2.25846
Formulas are derived for the accurate calculation of the Jowest-mode, lumped-element
representation of perfectly conducting half-round inductive obstacles in rectangular wave¬
guide. These obstacles consist of either one or two opposed semicircular cylindrical indenta¬
tions extending across the narrow sides of the waveguide. They seem especially suitable for
use as precise calculable standards of reflection or impedance in waveguide. Schwinger’s
integral equation approach [1,2] 1 is used to obtain stationary expressions for the desired
parameters as functionals of the surface currents on the obstacles. Upper bounds are ob¬
tained for one of the two parameters. Explicit formulas are derived for the values of the parameters
under the assumption of n-term Fourier sine-series expansions for the obstacle currents. Rapid
convergence is indicated by numerical evaluations for n=l, 2, and 3. In the process of
obtaining expressions suitable for numerical calculation, an expansion (believed to be new)
of the Green’s function of the problem is obtained and the sums of certain infinite series of
Bessel’s functions occurring in this expansion are expressed in terms of definite integrals.
A brief numerical table of these sums, sufficient for the evaluation of the n = 1 approximation,
is included.
1. Introduction
In this paper formulas are derived for the accurate calculation of lowest-mode, lumped-
clement parameters for what may be identified as "single half-round” and "double half-round”
inductive obstacles in rectangular waveguide. As shown in figure 1, the obstacles consist of
semicircular cylindrical indentations extending across the narrow sides of the waveguides.
The particular geometry considered seems especially suitable for obstacles to be used as
calculable standards of reflection or impedance in waveguide. The geometry is well suited to
electroforming, so that obstacles may be fabricated by this process as well as by machining.
For obstacles producing standing-wave ratios of moderate values, the obstacle radius is large
compared to high-standard machining tolerances, so that unduly close tolerances are not
required. A change in waveguide cross section before and after the obstacle is avoided, elim¬
inating the need for waveguide components in odd sizes and permitting maximum flexibility
of interconnection.
In the present work the obstacle and waveguide surfaces are assumed perfectly conducting.
The Waveguide Handbook [3] contains approximate formulas for semi-elliptical obstacles,2
which can be specialized to apply to the present problem. However, for the contemplated use
in standards work, formulas with very much greater accuracy (103 or 104 times greater) are
wanted.
Brief descriptions of the present and some related work have been given previously [4].
2. Formulation of Problem
We employ the customary complex electric and magnetic field vectors E—E(r), H=H(r),
which satisfy Maxwell’s equations in the form
VXE=-junH,
(2.1)
V X H=ju>eE-\- J,
1 Figures in brackets indicate the literature references at the end of this paper.
2 The pertinent formulas are afflicted with a number of misprints; correct forms of these results for half-round obstacles are given later in this
paper.
718/113
under the assumption of time-dependence represented by the (omitted) factor exp (jut). Here
t is the time, co/(27t) is the frequency, j is the imaginary unit, J is the complex current-density
vector, and n and e are respectively the permeability and the permittivity of the homogeneous,
isotropic, nondissipative medium in the waveguide. The MKS system of units is employed.
As shown in figure 1, we choose a rectangular coordinate system Oxyz such that the in¬
terior of the waveguide is the space 0<fx<ja, 0<jy<fb, — co<2<^co and the surfaces of the
obstacles are given by the loci
R2=x2+z2, 0<x<R,
R2=(a—x)2-\-z2,
-
0<a—x<Rj
- r (2.2)
The fields incident on the obstacles are, by hypothesis, to be in the TEi0 mode, Traveling-
wave forms for this mode may be written
E±=sin (ttx/a)e^zev, ^
(2.4)
H±=[zfr] sin (Trx/a)exJrj'rr(una)~1 cos (-7rx/a)eJc=F^z'j
The upper and lower signs refer respectively to waves going in the positive and the negative
2-directions; the phase constant 13 is equal to [^2— (7r/a)2]1/2, where k = co(jue)1/2; the wave-ad¬
mittance r/=)8/(coiu); and ex, ev, ez are the unit vectors of the system Oxyz. It is assumed that
Tv<fka<C2ir, so that the TEi0 mode, but no higher mode of the type TE„0, will have real propaga¬
tion.
We define “voltage” and “current” vm(zm),im(zm) for the TE]0 mode by means of the
equations
Here the indices m=l,2 refer respectively to the “left-hand” side (zx<f — R) and the “right-
hand”. (z2^>R) of the obstacle considered, and Emv, Hmx are components (identified by the
subscripts) of the TEi0-mode part of whatever total electromagnetic field may be present in
the waveguide at z=zm. From (2.6) and (2.4) it follows that the waveguide characteristic
impedance, defined as the value of vm(zm)/im(zm) for a pure traveling wave incident on the side
m of the obstacle, is equal to unity for m=l,2.
719/114
The desired lowest-mode, lumped-element description of an obstacle is contained in its
impedance matrix Z, which characterizes the linear relations
imposed by the obstacle on the terminal variables vm,im. The values of the Zmn depend upon
the location of the terminal planes z—zx and z=z2; in the present problem it turns out to be
convenient to extrapolate these planes to z1 = z2 = 0.
Now, the equality Zn = Z22 follows from the structural symmetry (and the symmetrical
disposition of the terminal planes) with respect to the plane ~=0; Z12 = Z21 is assured by both
symmetry and reciprocity; and, since losslessness has been assumed, Z is pure imaginary.
Thus Z is of the form
where we have added a subscript distinguishing the “even” case. In tin' antisymmetric
(“odd”) case, r,(0) = — r2(0), q(0) = — ?‘2(0), and the impedance may be written
Zoo=«i,o(0)Ai,o(0), (2.9I>)
follow directly from (2.7), (2.8), and (2.9). (it may be noted that the symmetric and anti¬
symmetric field distributions correspond to the eigenvectors of the matrix (2.8) and the quanti¬
ties Zee and Z00 are the eigenvalues of this matrix.)
An equivalent network for the obstacles is of interest and is presented here for convenient
reference. For a T-network, using the sign conventions shown in figure 2, we find
where Zx and Z2 are respectively the series anti the shunt elements of the symmetrical T. It
may be remarked that the obstacles considered are called “inductive” because for small radii
the important element, Z_>, is a positive reactance.
720/115
3. Analysis
w hich must be specialized appropriately for the present problem. E is moreover subject to
the boundary condition that its tangential components vanish on perfectly conducting surfaces.
Iu the present problem both the exciting field and the waveguide-obstacle structure are
independent of y in the range 0<//<6, and it follows that the whole field, secondary as well as
primary, will be independent of y in the same range. Further, since the x- and '•-components
of E must vanish in particular on the y—0 surface of the waveguide, these components must
vanish for all values of y involved. Thus the electric field is of the form E=(f)(x, z)eif; eq (3.1)
(with J=0) reduces to
V'd + Fd = d in S,
whore S is the cross section and C is the boundary of the structure in a plane //^constant (fig. 3).
0 z
Figure 3. Relevant geometry and coordinates for the single
and double half-round problems.
K=-(ju}p.)~1n-\/(t>ey (3.2b)
evaluated on ('. where n is the outward normal unit vector on (\ Thus K has only a //-com¬
ponent and may be written as K(x,z)ey.
Inasmuch as the whole problem can be dealt with in terms of </> and K, the problem is es¬
sentially a two-dimensional scalar one. In what follows it will be convenient to refer to 4>
simply as ‘‘the field”
721/116
3.2. Integral Equation Formulation
We now introduce a Green’s function r=T(x, 2, x', z') such that T is the field that would
be produced in the unperturbed waveguide at the point (x, z) by a unit current filament par¬
allel to ey at the point (x', z'). As a function of (x, z), T satisfies the differential equation
obtained from (3.1) with J=8(x—x') 8(z—z')ey, where 8 denotes the Dirac delta-function.
T further satisfies the boundary condition r = 0 for x=0, a and represents outgoing waves
for jz—z'\—>00 . An explicit form for r is [5]
where an=-\] (mr/a)2—k2 for n>2 and the other symbols are as already defined.
It is convenient to put
T=—ap(F+jG), (3.5)
where F and G are real. We observe that F can be written in the form
The functions \f/e and \J/Q are respectively symmetric and antisymmetric with respect to re¬
flection in the plane 2=0.
With the aid of the Green’s function we can set up the fundamental equations of the prob¬
lem in integral-equation form. Let \f/=^(x, 2) denote the exciting field—j.e., the field, due to
remote sources, that would exist in the absence of obstacles. Then, with an obstacle present,
the total field 4> can be written as the sum of the exciting field and the secondary field due to
the current on the obstacle:
<i>(x,z)=\p(x,z)-\- f
Jc0
T(x,z,x',z')K(x',z')dC', (3.7)
where K is the surface current on the obstacle and the integral is a line integral going over the
contour C0 of the obstacle (C0 consists o.f two parts in the case of the double half-rounds). If
the point (x, 2) is taken on C0, 0 must vanish and we have
This is an inhomogeneous integral equation of the first kind determining K for a given 0. The
equation reduces to slightly different forms in the symmetric and antisymmetric cases, which
we now consider.
a. Symmetric Case
Let the exciting field be A\J/e(x, 2), where A is an arbitrary amplitude and 0e, as defined in
(3.6), is a symmetric function of 2. The surface current on the obstacle will then also be a
symmetric function of 2; we denote it by Ke. We now examine (3.7) for z<^<^z', say, to find
the lowest-mode component and thus to find vx,e(z) and ii,e(z). For 2<<V, (3.7) becomes
0(2:,2)=A cos 02 sin (7rx/a) —(0a)_1wM sin (irxla)eiffz \f/eKe dC, (3.9)
J Co
722/117
since integrals of the type \poKe dC vanish because of the mutual orthogonality of sym-
J co
metric and antisymmetric functions. By comparison with (2.6) it is seen that
Equations (3.10) and (3.11a) hold directly from the definitions for 2+ —i? and by extrapolation
for 2=0. In particular (3.11a) yields
Next, in the integral eq (3.8), we separate the Green’s function into real and imaginary parts as
in (3.5), use the orthogonality property of even and odd functions, and thus find
This equation and (3.11b) together furnish a definitive mathematical statement of the sym¬
metric part of our problem.
b. Antisymmetric Case
In this case we let the exciting field be A\f/0(x, z), where A is again an arbitrary constant and
i/'o, defined in (3.6), is an antisymmetric function of 2. The current, K0, on the obstacle will
then also be antisymmetric with respect to 2. In the same manner as in the symmetric case
one finds, for 2<0,
In particular,
F urther,
This equation and (3.14b) together furnish a definitive statement of the antisymmetric cases
of our problem.
In the case of the double half-rounds the above integrations go over a two-part contour
consisting of an upper semicircle Cu and a lower semicircle CY An equivalent problem involving
only one of the parts of the contours, say Ch can be formulated using the fact that the exciting
field as well as the structure is symmetric with respect to the plane x=a/2, so that the current
on the obstacles must also be symmetric with respect to this plane. This symmetry is quite
723/118
independent of the symmetries with respect1 to 2 = 0 distinguished by the subscripts “e” and
“o,” and so these subscripts will be dropped for the moment. Thus, it is clear that
F urther,
Equations (3.IS) and (3.19) enable the desired restatement of the double half-round problems.
In what follows, superscripts 1,2 will be used when it is desired to distinguish quantities
associated with the single and the double half-round problems, respectively (this means in
particular G=Gil)). Integrals will be indicated by means of the convenient scalar product
notation, e.g.,
where h= 1 or 2 for the single or the double half-rounds, respectively, and Xee, Xo0 are the
reactances corresponding to the (pure imaginary) impedances Zee, Z00 defined in (2.9).
In each of the problems specified by (3.22) and (3.23), an unknown reactance or suscept-
ance is proportional to a scalar product
M=(ip,-I), (4.1a)
GI=\p, (4.1b)
in which ^ and G are given and G is symmetric. From these equations one may easily con¬
struct the “stationary representation” [1, 2]
724/119
for M. This constitutes the starting point for the method of Rayleigh and Ritz [6], which
we wish to use to obtain an approximate expression for M. One assumes in (4.2) an approxi¬
mation for the surface current in the form
In = Xifi, (4.3)
where the a, are coefficients to be determined, the/, are members of a suitable set of basis
functions (to be chosen explicitly), and summation from 1 to n over repeated indices i,j, , . .,
is understood. Expression (4.2) becomes an ordinary function of tlie xt and is to be subjected
to the conditions of stationarity dM/dx,=0, i— 1, 2, . . ., n. This leads to the system of
equations
GijXj {x kC k j 1V[ n~) C i
where
Ci=WJt), Qu^Ut, Of.)), (4.4)
and Mn denotes the now-determinate approximate value for M. If the arbitrary normalization
of the Xj is chosen so that xkck=Mn, then
(6?
where (G~1)ij denotes an element of the inverse of the nXft matrix of the Gi}, and it immediately
follows that
Mn=(G~l)ijcicj.
0 Cl . . . Cn
Cl Gn ■ ■ ■ Gm
n
Gn 1 . . . ^ 71 n
Gn • • Gm
Gnl ■ •
Thus Mn is expressed in terms of the known quantities (4.4); if the objective is only to calculate
the reactances, it is not necessary to calculate the xt.
Although in our problem G is not a definite kernel, it does have definite character with
respect to a suitably restricted class of admissible functions. In fact, G is positive definite
with respect to functions that are antisymmetric with respect to 2 and vanish for ]^>Xg/4,
where A^=27r/f3 is the “guide wavelength” of the single propagating mode.3 In terms of the
integrals of interest here, this means that if 70 is any assumed surface current distribution,
continuous, not identically equal to zero, and antisymmetric with respect to 2, then
725/120
where / is an admissible approximation for the surface current, with the true value given by
(4.1a) (as well as by (4.2)) when 7
satisfies (4.1b). From (4.1b) it follows that ( , \f/) = (l, GI); 7
the Schwarz inequality for positive-definite symmetric transformations [7] gives
7
( . Giy<(l, GI) (/, GI).
M<M. (4.7)
Moreover the sign of equality holds in the Schwarz inequality and hence in this result if and
only if I is proportional to /.
Clearly, (4.7) is applicable when G is merely positive definite with respect to all functions
that need be admitted; by (4.6) this will be the case in the antisymmetric parts of our problem
provided R<C\S/4. Hence, subject to this inequality, in the antisymmetric cases we shall have
Mn<CM (we do not anticipate an exact result for finite n); this in turn implies that numerical
values for A"00 will be (algebraic) upper bounds for the true values—assuming, of course, that
numerical evaluations involved are sufficiently precise.
Unfortunately one can not obtain a bound for Xee in this simple manner. Furthermore,
derived quantities of practical interest (such as (4.17)) will usually depend upon both Xee and
W00. Thus in general no bounds for such derived quantities are determined.
where R is the obstacle radius and O<0<7r. On this path the surface current I=I(R sin 9,
R cos 9) becomes a function of 9, which we denote simply by 1(6). We observe that sym¬
metries with respect to 2=0 are equivalent to symmetries with respect to 9= -k/2.
As basis functions we take
for the antisymmetric cases (expansion of /0). The two sets of functions are complete for the
expansion of symmetric and antisymmetric functions, respectively, in the interval (0,7r). The
completeness gives good assurance that for sufficiently large n, Mn will approach arbitrarily
close to the true M. For practical numerical calculation it is vital that the convergence of Mn
be rapid. In the present instance rapid convergence might be anticipated on the grounds that
the true Ie and /0 may be expected to be smooth (possibly infinitely smooth) functions of 6
vanishing at 0 and 7r, so that their Fourier sine series should converge rapidly.5
5 The points (R, 0) and (R, ir) are the vertices of the corners formed where the half-round centered at 0 meets the waveguide wall. The van¬
ishing of the surface current in these vertices is equivalent, by (3.2), to the vanishing of the components of the surface magnetic field normal to
the vertices. That these field components should vanish in the limit as a vertex is approached follows from properties of Sommerfeld’s solution
[14] of the problem of diffraction by a wedge of arbitrary angle.
726/121
4.4. Evaluation of Integrals
Integrals of the type (ip,ft) may be obtained relatively easily with the aid of the well-known
expansion [8]
eJSaln a=1tjn($)eina,
71 — oo
wherein Jn is the nth order Bessel function of the first kind. In this expansion we put £ — kR
and a=±0—x+(7t/2), with k sin x=W®> k cos x=j3, and employ (4.9). The expansion
is then easily obtained. The integrations yielding the desired quantities are now elementary;
one obtains
for the antisymmetric cases. (It is encouraging that these quantities decrease rapidly with
increasing index.)
The evaluation of the integrals of the tj^pe (Ju GfJ) = Gij is considerably more complicated
than the evaluation of the cu and the task has been relegated to the appendix The results
obtained may be summarized conveniently in the following form, which comprehends four
cases:
QW
w 1Q _L_
[p P(kR)8ptY Gpl (ka)\ (4.14)
CiCj 4 sp.sa
4.S. Results
Pid_0'n) • • CTih)
al,2n-l
•h piY^iV
727/122
(In this and in the following equation the arguments kR arid ka are understood.) Similarly,
(3.23), (4.5), and (4.14) yield
0 s2 ■ ■ S2n
s2 P2~\~^(lV • • 02(2 n
/A 1 _(/?)
P'2n "I 0 2 n, 2 n
The determinantal form of these expressions seems convenient enough for practical calcula¬
tions with n at least as large as 3. However, it may be remarked that, because of the special
form of the diagonal elements, the determinants can be expanded in the same way as certain
determinants occurring in the Fredholm theory of integral equations. This expansion may
be convenient for the consideration of arbitrarily large n, but is not needed here.
Once the reactances Xee and -,Y00 are determined, other quantities or parameters associated
with the representation of an obstacle are of course also determined. In particular, the “volt¬
age” standing-wave ratio seen on one side of an obstacle when the other side is terminated
in a matched (reflectionless) load is a familiar quantity having immediate physical significance,
and it is convenient to discuss some of the results in terms of this quantity. Recalling that
waveguide characteristic impedances were chosen equal to unity and referring to the equivalent
network, figure 2, it is easily found that the standing-wave ratio in question is given by
(4.17b)
In order to get some idea of the behavior of the sequence of approximations, the reactance
elements and the related VSWR (4.17) have been calculated for n — 1, 2, and 3 and for a range
of values of ka and kR. Table 1 presents more or less typical results in terms of the values
of T'S'IlTf. The convergence of the sequence of approximations, judged on the basis of
numerical results exemplified in the table, appears to be very rapid; by the same token, the
first approximation furnishes a rather good result for VSWR’s up to about 2.
Thus far it has not proved feasible to determine the manner of convergence of the sequence
of approximations by theoretical means. In view of the character of the particular problem
at hand, it seems probable that the numerical evidence may be relied upon.
( 1 1.0370008 1.0776492
0. 2 1.0370970 1.0776499
1 3 1. 0370970 1.0770499
| 1 1.4547068 2. 8411581
U 7 2 1.4554655 2.8416266
1 3 1. 4554055 2.8416268
728/123
Further calculations have been made regarding ka and R/a as independent variables.
This choice of variables corresponds to the practical situation where one has a given obstacle
(having a given value of R/a) that is to be operated at various frequencies. Figures 4 and 5
show the matched-termination VSWR (4.17) resulting for various values of R/a and jjfc,
where f is the operating frequency and jc is the cut-off frequency for the TEi() mode (and
Mc = kalx). (Precise tables (based on the third approximations) giving basic parameters and
possibly some derived quantities are being prepared for separate publication.)
The first approximations appear to be of useful accuracy for some purposes and represent
an appreciable extension of previously available results. Consequently, a brief tabulation
of the cr-functions, sufficient for the calculation of first approximations, is presented as table 2.
For convenience of reference we write down (4.15) and (4.16) for the special case n— 1,
X',V=-~\pAkB)+<\ka)], (4.18)
4h
WW + 'fftf*)]-1, (4-19)
X o-SjI’fx)
<dV(z) o-uW
729/124
572772 0 - 61 - 47
For sufficiently small kR, the expressions just given may be simplified still further. For
small kR, the quantities px(kR) and p2(kR) become large, in accordance with the formula
(2>D,
and thus determine the qualitative behavior of Xee and Ar00. Using this approximation for
the p’s, neglecting the a’s, and introducing the guide wavelength A1,=27r//3, one obtains
In order to arrive conveniently at the respective types of Green’s functions needed for
the single and the double half-round problems, we consider a waveguide of width w (instead
of a), bounded by the lines ;r = 0 and x=w; and, in addition to the electric boundary condition
at z = 0, we consider both electric and magnetic boundary conditions at x=w (by “electric”
or “magnetic” boundary conditions it is meant that- the Green’s function or its normal deriva¬
tive, respectively, vanishes on the boundary). The radius vectors of the source-point, and the
field-point will be denoted by r' and r, respectively.
Now, it is well known that the two-dimensional “free-space” Green’s function, satisfying
(3.3) and representing outgoing waves at- infinity, is — (up/4i)H0(k\r' —r|), where H0 denotes
the Hankel function of the second kind (as is appropriate for exp (jut) time dependence).
It is also well known that the desired Green’s function, <t>, satisfying (3.3) and the required
boundary conditions at j:=0, x=w, and at 2=±°°, can be obtained in the form of a sum of
free-space Green’s functions by the method of images [9]. In fact
where P±= (2nw±x')ex-\-z'ez and s= + l or — 1 according t-o whether the boundary condition
at x—w is electric or magnetic. By inspection of the array of sources (fig. 6) it can be seen
that this function has the proper symmetry to satisfy the boundary conditions: it is anti¬
symmetric with respect to x=0 and it- is antisymmetric or symmetric with respect to x—w
according to whether s= + l or —1.
Equation (5.1) can be transformed into a sum of elementary wave-functions of polar
coordinates by repeated application of Graf’s addition theorem for Bessel functions [10].
We introduce polar coordinates (r, 6), (r', O'), and (pn, dn) for r, r', and p„ in the manner shown in
(4.8), and we assume r and r' both to be less than w, so that pnj>w for n ^ 0. Considering
the triangles with sides pn, r, and pn—r and applying the addition theorem, we obtain
-4(WM)-i$= it
P= — CD
Jp(kr)Hp(kr')
Jp(kr')Hp(kr)
gjpO^g-jpO'_g^fl')
+ Z)' S JP(kr)e^[Hp(kp+n)e-^n-Hp(kp-n)e-^-n],
730/125
Relative signs of sources are indicated by the adjacent symbols +, —,
+s, and —s; the role of s is defined in connection with (5.1).
where the terms arising from n=0 appear in the first line and are to be omitted from the primed
summation in the second line; the upper and lower alternatives in the first line apply according
as r<Cr' or r^>r', respectively. Again, by the addition theorem,
731/126
Combining the last two equations and rearranging terms, one finds that (5.1) can be written
in the form
Jp(Jcr)Hp(kr')
Vo
■{
4>=—co/i Xj ]C siR sin qQ
2>=1 9=1 Jp(kr')Hp(Jcr)
If now one puts w=a and s= + l, this expression becomes the desired representation of T of
the text; similarly, if w=a/2 and s~ — 1, (5.2) becomes the desired representation of a Green’s
function T(2) related to G(2) in the same way as T is related to G (cf. 3.5). Thus, employing the
relation (3.5), the relation Hp=Jp—jYp, the notation (3.20), the basis functions (4.10, 4.11),
and carrying out the integrations with respect to 6, 6', one obtains
(sin p#, Gih) sin qd') Jp{kR)Yv{kR)8pq-Jp{kR)Jq{kR) cos [(tt/2) (p-q)W»(ka), (5.3)
We note that the above expressions are needed only for positive integer indices p, q that
are both even or both odd. We observe also that the form of the expressions is such that
those with unequal subscript indices can be evaluated simply in terms of those with equal
subscript indices. Further, in view of the easily-verified relation
values or formulas for the tr’s “of the second kind” can be derived readily from the same for
the cr’s “of the first kind.” Thus, the evaluation of the Gp]a, which, on the face of it, originally
involved double integrals with ka, kR, p, q, and h as independent parameters, has been reduced
co
essentially to the evaluation of sums of the form X; Y2s(nx) with 7r<x<27r or 27r<a;<(4r and
71=1
Two methods of evaluating the series defined in (5.4) will be discussed briefly here.
An evaluation of the desired series may be obtained in terms of definite integrals as follows.
We start with the series of Hankel functions (of the second kind)
S\(x)=itl (5.6)
71 = 1
where x is real, positive, and not an integer multiple of 2-k. For the Hankel functions we
use the integral representation [11]
Hx(x)=-
* Jc
f
e~ixaln t+Wd£
732/127
where £=£-\-jv and C is a suitable path, to be chosen in a moment, between the limits — jco
and Tr-\-jc° . With the aid of the well-known expression for the sum of a finite number of terms
of a geometric series, one finds that the M-term partial sum of (5.6) can be written
M
jC H\(nx)
=1 r e**dt 1 r e-w*sln *• ,r,
(5.7)
n= l Trjceix slnf-l 7rJce^slnf-l s
Among the infinitely many poles of the integrands in these integrals, we note in particular
and £m=7r—$-m, where is the principal value of sin ~1(2rmr/x), m = 0, 1,2, . . ., Q, and Q
is the largest integer such that 2(?7t/x<1. The path C is now chosen to comprise (except for
indentations) the negative imaginary axis, the real axis between 0 and tt, and the upper half of
the line r-j-jy; indentations, of radius p, are made at each of the above identified poles in the
manner shown in figure 7, which illustrates C for the case Q=l. Taken on this path, the
second integral in (5.7) approaches 0 as M increases indefinitely; as a detailed examination
shows, this happens because on the various parts of the path the integrand approaches 0 or it
becomes increasingly oscillatory (or both) as . Hence we have
This result can be expressed in terms of real integrations plus contributions from the
indentations. In the limit as p^>0, the indentations yield
(5.8a)
*'°=h
Here the notation corresponds to that used for the respective poles and the factor 1/ir standing
in front of the integral is included. Provided X is an even integer, the straight-line segments of
the path of integration yield
where 80\ is the Kronecker delta. It should be noted that when the integrand in the finite
integral is singular, the Cauchy principal value is to be taken, as indicated by the symbol P
(the indicated limit as p^O essentially is also a Cauchy value).
Upon comparing (5.4) and (5.6) it is seen that
733/128
It happens that for \=2p (where p=0, 1,2,. . .) the net imaginary contribution of (5.8) to
S2p vanishes. Consequently (5.9) yields
7T/2
r sinh1 2 prjdr]
r(D
(ka)
8
ttJo
^2ka sinh tj ^ --P
7r 1 sin2p£ cot (ka sin £) (5.10)
wherein the integrands are now well-behaved at their lower limits and thus we were permitted
to put p — 0. The corresponding result for apv is
(2)
sinh2 prjdri 4 „ P*72
a VP gka slnh,+ 1—Jq sin2 p£ tan (5.11)
where p = 0, 1,2,. . .; x and Q are as specified above. This result was obtained by Ignatowsky
[12] by a different method.)
Thus far numerical results for the o's have been obtained primarily from (5.10). The
integrals were evaluated numerically by Paul F. Wacker and William W. Longley, Jr., using
Legendre-Gaussian quadrature formulas programed for- machine computation.6 Values
for <r® were calculated from values of o-® using (5.5). Results for the two lowest values of
p are given in table 2. (More extensive tables are planned for the previously-mentioned
separate publication.)
2k Q 1 CO /I
Y0(nx)=log^+S^+ S (L =\ (5.12a)
T n=i a m=q-)-iV71 Vn2 k2~\)
1 ^ (2K)2»(p+p-l)\B2tl
■K £ Y2p(nx)=~
71=1 *P 2 (2M)\(p-M)\
(nn—^rdk2— \)2p
■k J2 sin sin 1(nK)]_(_)Py (5.12b)
77=1 v \—n2K n=Q+l Vrt2*2-!
8 The estimates of accuracy of computed quantities given in this paper were made by Mr. Longley, who was responsible for most of the detail
of the calculations.
734/129
[13].) The series for a[f obtained from (5.12a) and (5.12b) has been duplicated by an in¬
dependent mode of derivation; thife represents a partial check of the complicated expression
(5.13b). Thus far in the present work, Ignatowsky’s series have been used only to spot check
some of the numerical results obtained from the definite integrals. The series are especially
suitable for either estimation or calculation of the <r’s when p is large.
The author has benefited from the discussion of many points with Paul F. Wacker and is
indebted to David F. Wait for assistance with some of the theory and computations.
6. References
[1] D. S. Saxon, Notes on lectures by Julian Schwinger: Discontinuities in waveguide, Mass. Inst. Tech.
Rad. Lab. report 43-7/4/44.
[2] F. E. Borgnis and C. H. Papas, Randwertprobleme der Mikrowellenphysik (Springer-Verlag, 1955).
[3] N. Marcuvitz, ed., Mass. Inst. Tech. Rad. Lab. Series 10, 255-257 (McGraw-Hill Book Co., New York,
N.Y. 1951).
[4] R. W. Beatty and D. M. Kerns, Recently developed microwave impedance standards and methods of
measurement,, IRE Trans, on Instrumentation 1 7-, 319 (1958).
[5] See [2], p. 24.
[6] R. Courant and D. Hilbert, Methods of mathematical physics, 1st Eng. Ed., p. 175 (Interscience, N.Y.,
1953).
[7] F. Riesz - B. Sz.-Nagy, Functional analysis, p. 262 (Frederick Ungar Publishing Co., New York, N.Y.,
1955).
[8] See [6], p. 474.
[9] P. M. Morse and H. Feshback, Methods of theoretical physics, p. 814 (McGraw-Hill Book Co., Inc.,
New York, N.Y., 1953).
[10] G. N. Watson, A treatise on the theory of Bessel functions, p. 361 (The University Press, Cambridge,
England, 1944).
[11] See [6], p. 468.
[12] W. von Ignatowsky, Uber Reihen mit Zylinderfunktionen nach dem Vielfachen des Argumentes, Arch.
der Math. u. Phvs. 23, 193 (1914).
[13] V. G. Smith, L. Infeld, and W. J. Chien, On some series of Bessel functions, J. Math, and Phys. 20, 22
(1947).
[14] P. Frank and R. v. Mises, Die Differential- und Integralgleichungen der Mechanik und Physik, Vol. 2,
p. 808ff (Mary Rosenberg, New York, N.Y., 1943).
735/130
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES July
Summary—A new adjustable sliding termination for rectangular cuit can be varied by rotating6 the strip. With inde¬
waveguide has been developed. The termination is of simple design
pendent control of these two motions, complete cancel¬
and can easily be adjusted to have reflection coefficients from zero
to nearly unity in magnitude and any desired phase. In addition to the lation of reflections can be obtained. On the other hand,
usual applications of adjustable sliding terminations for rectangular with the strip surface perpendicular to the electric field,
waveguide, it provides a suitable design for an adjustable transfer or minimum losses occur in the strip and almost perfect
secondary standard of impedance for rectangular waveguide systems. reflection is obtained. It is possible to adjust the termi-
na ion to any intermediate condition, then to slide the
Introduction
entire assembly, to obtain almost any reflection coef¬
ADJUSTABLE sliding terminations for rectangular ficient desirable.
f—\ waveguide are useful for a variety of purposes,
"^including calibration of microwave impedance
measuring devices, obtaining the directivity of direc¬
tional couplers, and determining the parameters of
waveguide junctions.
An adjustable sliding termination for rectangular
waveguide was described by Grantham in 1951.1 Sev¬
eral years later, a similar termination was produced
commercially. This termination2 differed from its prede¬
cessor in details of construction of the dissipative ele¬
ment and the reflecting antenna. Both of these termina¬
tions were designed primarily to provide minimum re¬ Theory
flection and could not be adjusted over a wide range of Apparently, the analysis of the fields inside a rec¬
vswr (voltage standing-wave ratio). The commercial tangular waveguide containing an arbitrarily positioned
termination has a vswr range of 1.005 to 1.15. The dissipative strip has not yet been published. Lacking
principle of the double slug tuner was used3 to obtain this basic information, a rigorous analysis of the action
an adjustable sliding termination with a somewhat of this termination has not been attempted. An ap¬
greater range of adjustment. An adjustable sliding proximate study based upon microwave circuit theory is
termination having a wide range of vswr was described given below.
by Kato and Sakai,4 but it was not possible to adjust As shown in Fig. 2, the termination can be regarded
this termination for cancellation of reflections. as an attenuator terminated in a short-circuited line of
variable length. In terms of the scattering coefficients6
Principle of Operation
•Su, -Si2, and S22 of the attenuator, the input voltage re¬
As shown in the diagram of Fig. 1, this termination flection coefficient is
slides inside a rectangular waveguide and consists of a
short-circuiting piston to which is attached a dissipative ra Su
= (1)
strip supported by a dielectric rod, which can rotate and S22 + e 12B‘
slide relative to the piston. The phase of the reflection
where (3 = 2tt/\g, and \<j equals the wavelength in the
from the strip can be varied by sliding the strip, while
waveguide. The condition for cancellation of reflections
the magnitude of the net reflection from the short cir-
(Ti = 0) is
* Manuscript received by the PGMTT, November 19, 1956.
t Natl. Bur. of Standards, Boulder, Colo.
1 R. E. Grantham, “A reflectionless wave-guide termination,”
Si22 = Sue^il +Sne-M‘). (2)
Rev. Sci. Instr., vol. 22, pp. 828-834; November, 1951.
2 W. A. Andrews, U. S. Patent No. 2,701,861.
3 R. C. Ellenwood and W. E. Ryan, “A uhf and microwave 6 G. C. Southworth, “Principles and Applications of Waveguide
matching termination,” Proc. IRE, vol. 41, pp. 104-107; January, Transmission,” D. Van Nostrand Co., Inc., New York, N. Y., pp.
1953. 374—376; 1950. An attenuator employing a rotating strip is described.
4 N. Kato and T. Sakai, “Waveguide Type Variable Impedance 6 Scattering coefficients are the elements of the scattering matrix.
Circuit and its Application for Rieke Diagram,” Rep. of the Micro- See C. G. Montgomery, R. H. Dicke, and E. M. Purcell, “Principles
wave Communication Res. Committee in Japan, p. 7; December, of Microwave Circuits,” M.I.T. Rad. Lab. Ser., McGraw-Hill Book
1955. Co., Inc., New York, N. Y., vol. 8, pp. 146-149; 1948.
736/192
1957 Beatty: Sliding Termination for Rectangular Waveguide
Normally, |An | and [ A221 are much less than unity, and
the approximate condition for cancellation of reflections
is
737/193
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES Ji/fy
Acknowledgment
738/194
Reprinted jrom the PROCEEDINGS OF THE I.R.E.
VOL. 44, NO. 8, AUGUST, 1956
Summary—A new microwave standing-wave-ratio (or reflection other high-precision work, is described. An earlier stage
coefficient) measurement technique, apparently suitable for stand¬
in the development of this technique has been reported
ards and other high precision work, is described. The technique re¬
previously by the authors.1
quires that the phase angle of the unknown reflection coefficient be
subjected to arbitrary, known variations—as is possible with a sliding The technique requires that the phase angle of the
load, for example. Accurate measurement of suitable sliding loads unknown reflection coefficient be subjected to arbitrary,
furnishes standards, and with the aid of comparison techniques, known variations. (A reflection coefficient whose phase
enables indirect measurement of arbitrary unknowns.
can be subjected to such variations will be termed
Generator and detector are connected to two arms of a three-
“phasable.”) It is clear that this phasability require¬
arm waveguide junction. Tuning elements in the junction are (de¬
sirably) adjusted for small reflection looking into the third arm, to ment would be no limitation if an ideal phase-shift de¬
which the unknown connects. Observation of the detector response vice were available.
vs variation of phase of the unknown yields a curve, similar to a Thus far the technique has been used primarily with
standing-wave pattern, from which the unknown is determinable by
specially constructed “sliding-loads” (terminating struc¬
procedures that are given.
The technique has the advantages of 1) being amenable to
tures arranged to slide in the waveguide in which used-
rigorous theoretical analysis, 2) enabling the attainment of heavy see Fig. 5). One other phasing method has been tried and
coupling to the detector without simultaneous severe distortion of will be mentioned. However, the sliding-load technique
the response pattern, and 3) being, in a sense explained in the text, in itself offers a solution to most problems of standing-
fundamentally simpler than the conventional slotted-line technique.
wave-ratio (swr) standardization. A suitable sliding
load whose swr has been accurately measured consti¬
A Introduction
739/1024
1956 Macpherson and Kerns: Measurement of Microwave Standing-Wave Ratios
The experimental work done thus far has been con¬ is hardly restrictive. Reciprocity is not used in the main
cerned only with the determination of the magnitude of part of the analysis. It may be well to mention explicitly
reflection coefficients. There is no reason why angles that it is not required that the detector or generator2
cannot also be determined, however, and the necessary present matched loads to the junction and that', by the
theory and equations for this will be given.. same token, the joints in the waveguides connecting
The basic arrangement (see Fig. 1) consists essentially detector and generator to the junction are not required
of a detector and a generator connected to two of the to be perfect.
arms of a three-arm waveguide junction. This arrange¬ The idea of using a fixed detector and a phasable load
ment (generator and detector included) constitutes an is not in itself new.3 It is felt that the present work
instrument for the measurement of any phasable reflec¬ represents an advance in three main respects: 1) In
tion coefficient presented at the reference plane in the significantly increased generality and rigor. Apparently
third arm of the junction. The operating procedure it has not been realized that a much more thorough
consists essentially of observation of detector response treatment could be carried out with very little additional
as a function of the variation of phase of the unknown complication in the results. 2) In enabling the attain¬
reflection coefficient. This yields a curve, similar to a ment of heavy coupling to the detector without simul¬
standing-wave pattern, from which the unknown reflec¬ taneous distortion of the response pattern. In the pres¬
tion coefficient is determinable by procedures to be ent work the heavy coupling permitted the use of a
given below. bolometric substitution method of power measurement
as the detector, virtually eliminating the question of
detector linearity. Probably the principal advantage of
having small distortion is that it permits simplification
of the general procedure for the determination of the
unknown reflection coefficient from the observed data;
it may also enable increased accuracy. 3) In the applica¬
tion of an analysis (developed in a different context by
Altar, Marshall, and Hunter4) which derives the desired
information frofn the response pattern whether or not
the pattern is highly distorted. It is remarkable that
the analysis applies more rigorously in the present
problem than it does in its original context.
Fig. 1—Basic arrangement. The characteristic equation of the instrument—the
analytical expression of detector power as a function of
The junction is provided with tuning elements that phase of the unknown reflection coefficient—is obtained
are adjusted for zero reflection coefficient as seen looking by a rigorous analysis under the above hypotheses. The
into the load arm of the junction (with the detector and characteristic equation is of a simple form and contains
inactive generator connected). This matching operation effectively just two complex constants. In general both
is not critical; the technique does not assume that a zero constants must be evaluated or accounted for empiri¬
or negligible reflection coefficient is actually obtained. cally; the necessary procedures are relatively simple.
The matching need be performed only once for a given The new technique should also be compared with the
setup operating within a limited frequency range. The conventional slotted-line standing-wave measurement
matching operation reduces (or may practically elimi¬ technique. In Appendix I the slotted-line instrument is
nate) distortion of the response pattern and enables one analyzed with rigor comparable to that applied in the
to have heavy coupling to the detector without at the text to the new technique, and it is shown that the char¬
same time having a highly distorted response pattern. acteristic equation for the slotted-line instrument
It should be emphasized that the waveguide junction (neglecting slot-end effects) contains four complex con¬
used in this technique is not subject to special require¬ stants and is of a fairly complicated form. It is con¬
ments, such as symmetry or losslessness. The junction sidered that the new technique has a fundamental and
may indeed be of a rather general type, in that the important advantage in the relative simplicity of its
waveguide leads may individually be of arbitrary cross characteristic equation.
section and one or more of them may be two-conductor
Theory
systems, such as coaxial line. The main assumptions are
that the electromagnetic field in a portion of each wave¬ Our first theoretical task is to obtain and discuss the
guide lead consists of a single waveguide mode and that characteristic equation of the system.
the variation of the phase of the unknown is accom¬ 2 Here and subsequently the “generator” is to be understood as
plished without affecting its magnitude. These two the equivalent generator as seen at the junction reference plane.
3 C. G. Montgomery, “Technique of Microwave Measurements,”
assumptions are only mildly unrealistic. The further M. I.T. Rad. Lab. Ser., The McGraw-Hill Book Co., Inc., New York,
requirement that the system (apart from the rf source N. Y., Vol. 11, p. 507; 1947.
4 Wm. Altar, F. B. Marshall, and L. P. Hunter, “Probe error in
itself, which must be well padded) be linear and passive standing wave detectors,” Proc. IRE, vol. 34, p, 33, January, 1946.
740/1025
PROCEEDINGS OF THE IRE August
Let reference planes Ti, T2, and r3be chosen in the M13 — (721(732 — (731G22 (5)
generator, load, and detector waveguide leads, respec¬
Mii = (Gn Sg)(G3 3 — Sd) — G31G13. (6)
tively, as indicated in Fig. 1. For the purpose of analy-'
sis, the definitions of the “junction” and the “termina¬ The power P can be expressed in the form
tions” are, of course, relative to the choice of reference
P = C | 1 + y + Y(KS, 9) | -2 7
( )
planes. (It should be noted that any discontinuities
introduced by flanged joints in the generator and detector where C is a constant of proportionality,
leads will be taken care of implicitly by inclusion either
in the junction or in the terminations.) It is convenient
K = G31/M\3
to characterize the junction (as defined by the choice - 2(M22/A)
of reference planes) by means of the inverse of its (8)
K + (M22/A)
scattering matrix. Letting ax, a2, a3 measure the ampli¬
1 — KS exp (— 2j9)
tudes of the incident traveling-wave components of the Y(KS, 9)
field on the respective terminal surfaces, and letting 1 + KS exp (— 2j9)
6 1, b2, 63 similarly measure the amplitudes of the emer¬
gent traveling-wave components, we then have the last being a functional notation. The significance of
expressing K and y in terms of the quantities (4), (5),
ai = G\\b\ -(- Gnb-i T G1363 and (6) will become apparent. The “constants” G, y, and
a2 = Gnbi + G2262 + G2363 (1) K are of course frequency-dependent. Eq. (.7) is the
desired characteristic equation.
a3 = (73161 + G3262 + (73363
The value of the constant C is immaterial as far as the
where the G;/ are the elements of the inverse scattering analysis of the response curves is concerned, inasmuch
matrix G (G will be called the “gathering matrix” of the as only relative values of P will be involved. Thus the
junction). The relations imposed at the terminal sur¬ equation contains effectively just two parameters, y and
faces by the terminations we express in the form K, that characterize the instrument. Of course y and K
are in general complex so that four real parameters are
d\ = Sgb\ + ba
involved.
(Pi = Se~2ieb2 (2) We observe that the characteristic constant K and the
a3 — Sdb3 unknown reflection coefficient 5 appear in the character¬
istic equation only in the combination KS. In what
Here the reflection coefficient Sg and the fixed wave- follows it will be convenient to deal with the product KS
amplitude bg characterize the equivalent generator at as an intermediate unknown and to postpone the ques¬
the reference plane T\, S is the unknown reflection co¬ tion of determining 5 itself. The notation S' = KS will
efficient, — 20 is the phase shift imposed upon S (so that be used.
Se~2,e is the reflection coefficient presented at reference
As a function of S' and 6, the characteristic equation is
plane T2), and Sd is the reflection coefficient of the detec¬
of the same form as the highly simplified equation for
tor at T3. (In the case of a sliding-load unknown, 6 may
the slotted-line standing-wave machine given in Ap¬
be interpreted as the variable electrical distance be¬
pendix II. (This is the basis of our application of the
tween T2 and the reference plane of the unknown.) Of
Altar, Marshall, and Hunter analysis; see below and
these quantities only 5 and 6 will be explicitly involved
Appendix II.) Thus it is recognized that in general the
in the final results.
graph of the characteristic equation will have the form
Combining (1) and (2) one obtains a set of three
of a distorted standing-wave pattern, the distortion
equations in three unknowns,
being introduced by the presence of the quantity y and
bg = (Gn — Sg)bi + G1262 + (71363, vanishing for y = 0. An experimentally obtained graph
in which the distortion is rather small is shown in Fig. 2.
0 = (72161 T (G22 — Se~2’e)b2 T (P2363, (3)
It is interesting to note that the quantity y can be re¬
0 = G3161 -f- G3262 + ((733 — Sd)b3 garded as a thoroughgoing generalization of the “probe
admittance” yp appearing in (16); y depends on Sg,
which determines 61, 62, and 63. The desired quantity,
Sd, and on all the elements of the matrix G and will in
the power P delivered to the detector, is proportional
general be nonzero whether or not the system actually
to the square of the magnitude of 63. We introduce the
contains a probe.
determinant
If an ordinary waveguide junction (for which reciproc¬
Gil — Sg G12 G13 ity holds and in which dissipation is small) is employed,
A = G21 G22 G23 (4) qualitative statements can be made about the magni¬
tudes of K and y:
G31 G32 G33 — sd
1) The magnitude of K will be approximately equal
(note that this is not exactly the determinant of the to unity. This follows from the fact that under the as¬
above system of equations) and the two minors sumed conditions the gathering matrix will be symmetric
741/1026
1956 Macpherson and Kerns: Measurement of Microwave Standing-Wave Ratios
Su = M11/ A
y = ~ 2SU/(K + S2i).
Now, although K as well as Su will be affected by the
matching operation, the magnitude of K will remain
approximately equal to unity (as shown in the preced¬
ing paragraph). Thus it is seen that making Sn small
will make y small.
A small or vanishing value for the aisioi non term y Further, let
is not incompatible with heavy coupling to the detector.
To establish this it is sufficient to consider an ideal case. V ^min
742/1027
PROCEEDINGS OF THE IRE August
The angle a' of S' may be determined as follows. The and 5 appear in the characteristic equation only in the
shift of the minima corresponding to the presence of combination S' = KS, the ’ empirical evaluation of K
the term y in the characteristic equation may be de¬ requires that S' be determined for at least one phasable
fined as load whose reflection coefficient (S = Ss, say) is known
5 = 0 . — f)i/min
. (« independently (K being then given simply by K = S'/Ss).
u '/mm
Probably the type of load for which the reflection co¬
where 0min((\ the “true” minimum position, is the value efficient can best be determined independently is a
that 0min would have if Im(y) were zero. It may easily “short.” A sliding short was used by the authors and
be determined that the means of evaluating it are described briefly in the
following section.
a' = 2(0m;n — S) + tt (10a)
As soon as K is known, the value of any 5 is of course
apart from multiples of 27r. The value of 5 is given in at once obtainable from the corresponding value of S'
terms of rj and e by"
Experimental
1 1
— sin-1 -ij. (10b) For purposes of illustration, some of the equipment
2 2 and practices employed in the experimental part of the
present work will be described. The arrangement used
was designed with the idea of obtaining the highest
Here the inverse sine is to be taken in the 1st or 2nd
possible accuracy. Much attention was given to prob¬
quadrant according to whether 77 is in the 1st or 2nd
lems of generator stability, detector linearity, detector
quadrant.
noise and drift, and mechanical precision. The complete
It will be observed that the above formulas do not
set-up is shown in Fig. 4. The rf source is an X-21
involve the characteristic constant y explicitly and are
klystron in cw operation. This tube yields about 5
applicable whether or not y is small. However, when y
watts output, which easily permits the use of adequate
is small, certain approximate formulas involving the
padding, and in addition is very stable in power level
real and imaginary parts of y explicitly may be pref¬
and frequency. For ordinary work the X-21 has the dis¬
erable. The value of y is obtainable from an expression
advantage of being essentially nontunable. (An ar-
given by AMH (p. 38P):
2 (\ & \
=-j-H — e* - 1 ) (ID
(1 - S7 a)\ So /
Here S' is as already defined, |S0| is defined in terms
of the directly observed vswr p0 = v'-Pmax/-Pmin by | S0|
= (p0 — 1)/(po + 1), and \p (which is to be taken between
— 7t/2 and +7t/2) is given by tan \p = esc e cot 77-(cos
e —cos 77). Let us denote the real and the imaginary parts
of y by g and b, respectively. It can be shown that, to
the first order in g and b, the intermediate vswr is given
by10
po — 1
p' ~ Po + Pog-;—7 (12)
Po+1
- b
(13)
5 ~ (po + l)2 ‘
743/1028
1956 Macpherson and Kerns: Measurement' of Microwave Standing- Wave Ratios
744/1029
PROCEEDINGS OF THE IRE August'
tional “noise” associated with the mechanics of the increases and that on the other decreases in such a way
phase shifting process. Some measurements made on a that 0i+02 remains constant.
sliding load of nominal vswr 1.01 give an idea of the Solving for b3 in the problem just set up yields a
mechanical stability possible. Six vswr values were result expressible in the form
_c exp ( jdi)_
1 + y + mSa exp (- 2j6i) + [l + nS0 exp (- '2j0x)]Y(KS, 02)
computed from readings of maxima and minima with where c, y, m, n, and K are constants independent of
the load at 7 different positions in the waveguide; the 0i and 02, and Y(KS, 02) is the function defined in (8).
mean was 1.0166 and the greatest deviation from the The characteristic equation, which is an expression for
mean was 0.0003. With loads of higher vswr, and with the rf power P delivered to a detector on arm 3, may be
the “line-stretcher” phase shifter, the reproducibility written
in terms of vswr was poorer in general. Nevertheless,
P = C|l+y + w' exp (— 2j9i)
the mechanical problem for a sliding load seems con¬
siderably easier than for a slotted line. + [1 + n' exp (- 2jd\) ] Y(KS, 02) |~2, (14)
where we have put m'= mSQ and n' = nSg. The four
Appendix I
characteristic constants are y, m', and K. This K
Characteristic Equation for Standing-Wave is the same function of the elements of G as the K of
Machine the text and thus shares the property |i£| =1 for a
We give a brief but rigorous derivation of the char¬ lossless junction fulfilling the reciprocity condition. It is
acteristic equation for a conventional standing-wave worth noting that if the generator is matched (5’(7 = 0),
machine. The general approach and much of the nota¬ then m' and n' vanish and (14) becomes formally identi¬
tion used in the second section will be applied her^. cal to (7), as it should.
Let reference planes T\, Tz be chosen on the genera¬
Appendix II
tor side and on the load side of the probe assembly,
respectively (see Fig. 8). These two reference planes, Problem Treated by AMH
1
1
1 )
tacit), and 2) that.the effects of the probe assembly may
(1 1 -
be represented by a pure shunt element in a lossless
©
1 1
1 1 1 1
transmission line. These assumptions lead at once to
GENERATOR*—! 1-► LOAD an expression for the voltage V across the equivalent
h—»i—1 1*-ez-1
transmission line at the probe position,
Fig. 8—Notation for standing-wave machine; coaxial line for
probe output is indicated in the sketch.
V =--- (15)
1 + yp + 7(5, 0)
which are to move with the probe assembly, plus a
where c is a constant, yv is the admittance associated
third one T3 in the rf probe output, define a 3-arm wave¬
with the probe, and 7(5, 0) is the normalized load-
guide junction. We denote the gathering matrix of this
admittance referred to the probe position. If it is as¬
'junction by G and use the notation aX} a2, 03, and bx, bz,
sumed that the power delivered to a detector is propor-
b3 for the incident and emergent wave amplitudes, as in
tionl to | U|2, (15) leads to the characteristic equation
the second section. The relations imposed by the termi¬
nations are P = C| 1+ yP+ 7(5,0) |~2. (16)
572772 0-61-48
745/1030
Magnified and Squared VSWR Responses for Microwave
Reflection Coefficient Measurements*
R. W. BEATTYf
Summary—In conventional microwave impedance’ measuring Other radically different types of response are ob¬
instruments, the measured ratio of maximum to minimum detector
tainable. The two responses to be discussed in this paper
signal level is ideally equal to the voltage standing-wave ratio
(VSWR) of the termination. In this paper, it is shown how radically
have been called magnified and squared VSWR re¬
different types of response are obtainable in which the observed ratio sponses for reasons which will become apparent.
may approximately equal the square of the VSWR or may be magni¬ A simplified explanation will first be given, followed
fied any desired amount. Theory is given enabling accurate-measure¬ by a more complete mathematical description.
ments by interesting techniques. Accuracies of 0.1 per cent in
The differences among responses are shown in Fig. 1,
VSWR to 2.0 have been achieved using magnified response tech¬
three response curves calculated for the same termi¬
niques.
nation.
Introduction
Simplified Explanations
N most microwave impedance measuring instru¬
ments, such as the idealized slotted line, the reso¬ Squared VSWR Response
nance line, and rotary standing-wave indicators, the A simplified explanation can be given for one system
ratio of the maximum to the minimum amplitude of the yielding squared VSWR response. Other systems which
output to the detector is ideally equal to the voltage have been devised apparently do not permit simplified
standing-wave ratio (VSWR) of the termination sub¬ explanations and will not be thoroughly analyzed.
jected to measurement. Enough theory will be given however, to permit their
use as measurement systems.
* Manuscript received by the PGMTT, February 20, 1959.
f Radio Standards Lab., Nat. Bur. of Standards, Boulder, Colo. The system shown in the diagram in Fig. 2 consists
746/346
1959 Beatty: Magnified and Squared VSWR Response*
1
E = e- igZv — iG (1)
Zo + Zp 1 1 Fig. 1—Response curves of three measurement systems with termina¬
tion having a VSWR of 1.26;
Z7* Z7 (a) idealized slotted line,
If the short circuit is located' A<?/4 from the probes, d= c[l + Krie-w]
(b) squared VSWR response,
Z's — 00 , and
r 1 + Turwi-i
d=
E ~ IgZl' — i-G
1 + TLe-Ml
(2) L 1 - rLe~WlJ ’
1 - TLe~™1 (c) magnified response,
<7 A
etia)' <?L
r
other symbols used above should become clear upon DETECTOR
Magnified Response Fig. 2—Diagram and simplified model of one system yielding squared
VSWR response; (a) diagram of system, (b) simplified model.
A system yielding magnified response is shown in Fig.
3. A directional coupler is connected to respond mainly to
to the wave reflected from a-phasable or sliding ter¬
mination whose VSWR is to be measured. For simplicity,
it is assumed that the generator and detector do not pro¬
duce reflections (re = r.D = 0), and that no reflections
are produced in arm 2 by the directional coupler (S22 = 0).
The signal coupled to the detector has two compo¬
nents. One is fixed and exists because the directivity is
not infinite. The other is from the load reflection and
varies in phase as the load is slid inside the waveguide.
Fig. 3—Diagram of simplified system to illustrate
As the relative phase of the two components vary, the magnified response.
magnitude of the resultant varies. If the components are
of approximately equal magnitudes, the range of varia¬ approximately equals the directivity ratio of the direc¬
tion of the resultant may be large even though the re¬ tional coupler), one may consider Tl to be magnified by
flection from the termination may be small. the factor | K | , leading to the term magnified response.
Inspection of the diagram of Fig. 3 leads to the fol¬ It should be noted however that the response variation
lowing equations describing the response.1 will increase as one increases [ K\ up to the point where
[ KTL\ = 1, and then will decrease with further increasfe
*3 - bG(Sn + S2iS32TLe-Ml) = 6g53i(1 + KTLe~Ml). (4)
in\K\.
The response is of the same form as that of the ideal¬ Analysis
ized slotted line (see Fig. 1) excepting that Ti, is multi¬
Both magnified and squared VSWR responses may
plied by the factor K. Since | K| may be very large (it
be analyzed by considering the generalized treatment2
1 In this equation, b represents a wave amplitude, S a scattering 2 A. C. MacPherson and D. M. Kerns, “A new technique for the
coefficient of the directional coupler, and TV the voltage reflection measurement of microwave standing-wave ratios,” Proc. IRE, vol.
co.efficient of the load. 44, pp. 1024-1030, August, 1956.
747/347
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES July
748/348
7959 Beatty: Magnified and Squared VSWR Responses
r2l- to have the same phase, and this is obtained by some Ur——— 0
tuning device, such as the adjustable short-circuit in nir
DETECTOR
CT
(a)
Fig. 2. There may be some difficulty in obtaining the ib) a,EC,w
desired response in some cases, because it is not always
possible to obtain the correct phase relationship, but
the junction forms represented in Fig. 5(a) to (d) have
all been found experimentally to permit a close ap¬
proach to squared VSWR response by proper adjust¬
ment of the tuner. The arrangement of 5(e) should also
permit squared VSWR response, but has not been con¬
structed or tested.
Magnified Response occurs upon making | K \ greater
than unity while r2i = 0. It evidently cannot be ob¬
tained with a lossless junction, for then |if| =1. If it
is assumed that we can always make Fg = 0, then
r2i would equal S22, and this would vanish, so that
K = S21S32/S31. The directional coupler connected as
shown in Fig. 3 evidently permits magnified response MOTE: AFTER CORRECT POSITION OF TUNER IS FOUND,
IT MOVES WITH C0UPUN6 LOOPS INSIDE COAXIAL
UNE .
since | ^32/^311 is the directivity ratio and may be quite
large while [ S2i| is usually between 0.7 and 1.0. The use Fig. 5—Schematic drawings of junctions permitting squared VSWR
response, (a), (b), (c), (d), (e).
of auxiliary tuners4 with a directional coupler permits
greater versatility since one may adjust.the directivity
ratio upwards or downwards with one tuner, then ad¬
-f'T3
just the other tuner to make r2, = 0. These adjustments
are independent only if made in the order described.
Referring to Fig. 6, the tuner in arm 2 is adjusted first 1 TTTTT V 1 TTTTT 1
1 —i—
in order to obtain the desired value of ] jKT| , then the i
i
tuner in arm 1 is adj usted to make r2l- = 0. DIRECTIONAL COUPLER t.
Tl
TUNER TUNER
Measurements Using Squared VSWR Response TO ADJUST TO ADJUST
r2i 3 0 DIRECTIVITY
Any of the junctions of Fig. 5 or their equivalents may
be used if the unknown is phasable. This requirement Fig. 6—Directional coupler with auxiliary tuners.
is satisfied if the unknown termination slides inside the
waveguide. In principle, a phase shifter or line stretcher
may also be used to provide the phase variation, but in been made, the response curves will be symmetrical
practice, they are less than perfect, leading to additional about the maxima and minima.
errors in measurement. If the unknown termination A correction may be made for deviations from the
does not slide within the waveguide, the arrangements ideal conditions |K| «1, |r2i| *»1, using the methods
of Fig. 5(d) and (e) may be used.' Either flexible cables indicated2 by MacPherson and Kerns. Instead of ana¬
must be used to couple the generator and detector to lyzing the response curve to obtain the parameter y re¬
the moving junction, or the generator and detector may quired for the correction, a shorter method is as follows.
be arranged to move with the junction. If the termina¬ Only the real part (g) of y is needed for a correction to
tion is not too large, it and the waveguide section could the VSWR. It can be shown that to a good approxima¬
be moved, keeping everything else fixed. The arrange¬ tion (to the first order in g and b),
ment shown for rectangular waveguide is not readily
1 + KTl
adaptable to operation with coaxial lines. However the err — y/a a + d* l)(a4 1), (7)
arrangement of Fig. 5(e) should be satisfactory for 1 - KTl
operation with coaxial line.
where
The correct adjustment of the tuner is made with
Tl = 0 and corresponds to maximum detector output 1 b31 (rL = o)
(g+ 1)
for Figs. 2 and 5(a) and to minimum detector output bs |max( | PL | 1)
for the others shown. When the correct adjustment has
It is still necessary to determine | K\ in order to obtain
749/349
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES July
The adjustments of the tuners preceding the measure¬ Because of the special condition r2, = 0, the line
ment is as follows.4 One adjusts the tuner in arm 2 until stretcher need not be of the constant impedance type,
no variation is observed in \b3\ as one slides a termina¬ since reflections that it may introduce may be cancelled
tion of low reflection inside the output waveguide. by reflections from the tuner in arm 1. Also, the refer¬
Then the tuner in arm 1 is adjusted until no variation ence plane for arm 2 may be located in the uniform
in [ &31 is observed as one slides a termination of high waveguide section of the line stretcher between the
reflection inside the output waveguide. If necessary, the source of its reflections and the load. As shown in Fig.
above operations are repeated in sequence until no vari¬ 7, the reference plane T2 remains fixed although the out¬
ation in | &3| is observed as either termination is slid. put waveguide and load move. The line stretcher must
In the second method, the tuner in arm 2 is adjusted be stable however, so that the parameters 5u, Su, and
(with the unknown connected to arm 2) until the de¬ S22 with respect to the reference planes Ti and T2 do not
tector output is zero. Then KYu~— 1. The tuner in vary as it operates.
arm 1 is then adjusted until r2, = 0. A reflection standard With the addition of the line stretcher, the second
of known | Ts| is then connected to arm 2, and the phase method described above may be called a magnified dif¬
of T's is varied. Substitution of the above conditions ference method, since the smaller the difference between
into (5) leads to |rs| and [ rc/1 , the greater the variation in |&3| as the
phase is changed (9).
I ^Imax | rp| + | rs|
aA \b3\min I I To-1 - I r51 I w Evaluation
In the event that |rs| =1, (approximately true for a It is too early to make a conclusive evaluation of
the worth of the responses described above and their
sliding short-circuit), then ga=0u■
Note that it is unnecessary to vary the phase of IV, applications in measurement systems. However, the
the reflection coefficient of the unknown termination in techniques employing magnified response give promise
either method. In the second method one needs to vary of increased accuracy in the measurement of low and
the phase of Ts, but this is easily done if a sliding short- intermediate VSWR. Accuracies of approximately 0.1
per cent in VSWR to 2.0 have been achieved, and
circuit is used.
perhaps an order of magnitude better than that is pos¬
Alternatively a fixed reflection standard may be used
if a suitable line stretcher is incorporated into arm 2 of sible.
the measuring instrument. Acknowledgment
750/350
Reprinted from IRE TRANSACTIONS
ON MICROWAVE THEORY AND TECHNIQUES
Volume MTT-7, Number 3, July, 1959
T
Fig. 1—Basic reflectometer.
Introduction
HE microwave reflectometer in its usual form where G is the voltage reflection coefficient of the termi¬
consists of a pair of directional couplers so ar¬ nation on arm 2, and K is a constant whose value must
ranged that one couples to the forward, and the be determined.
other to the reverse wave. The ratio of the sidearm out¬ A mathematical treatment of the four arm junction
puts is, in the ideal case, equal or at least proportional and detectors shows (assuming linearity) that the re¬
to the magnitude of the reflection coefficient of the sponse will in general be of the form:
termination, from which one can calculate the standing
wave ratio. b% A Ti + B
— = -- (1)
In practice, this relationship is only approximately bA CG + D
realized because of imperfections in the directional
where the A, B, C, and D are functions of the parameters
couplers and other factors. However, duectional cou¬
of the four arm junction and detectors.
plers having high directivity (40 db or more) and low
A particularly simple and convenient group of ex¬
main guide VSWR (less than 1.05) are commercially
pressions for the terms A, B, C and D, may be obtained
available which permit good accuracy to be realized
by the following procedure. The performance of the re¬
over a large (1.5 to 1) frequency range, while at a given
flectometer of Fig. 1 is characterized by the functional
frequency, further improvements may ,be realized by
relationships imposed upon the terminal variables ay,
the use of auxiliary tuning.1
by, ai, and 62, which represent the incident and emergent
The use of auxiliary, tuners has not been fully ex¬
voltage wave amplitudes at arms 1 and 2 respectively
ploited or completely treated however, and in this paper
and the responses of the detection systems (usually
a more general and rigorous analysis of the microwave
power) employed at arms 3 and 4. The main interest is
reflectometer will be presented leading to the introduc¬
in the relationship between the detector responses and
tion of additional tuning elements. Procedures for the
the ratio
adjustment of these transformers will be described and
a particularly simple form of the reflectometer devel¬ a2
oped. A quantitative treatment of the errors in these
techniques will be presented.
as given in (1). A variety of equivalent circuit repre¬
General Theory sentation's may be substituted for that shown in Fig. 1
The basic form of the reflectometer is shown in Fig. provided the relationships of interest among the ter¬
1. If &3 and bi represent the voltage amplitudes of the minal variables are preserved. A convenient choice for
signals at the respective detectors, the desired response the present purpose is as follows:
is Assuming that the detector impedance is constant (as
will be true for example of a barretter operated at a
b3 constant resistance) mathematical models for the de¬
G or G|
G tectors may be constructed of lossless and matched de¬
tectors preceded by lossy fourpoles of the required param¬
eters to produce the externally observed behavior.
* Manuscript received by the PGMTT, January 13, 1959; re¬
vised manuscript received, February 20, 1959. Reference planes in arms 3 and 4 are then chosen be¬
f Radio Standards Lab., Nat. Bur. of Standards, Boulder, Colo.
tween these ideal detectors and the lossy discontinuities
1 J. K. Hunton and N. L. Pappas, “The -hp- microwave reflectom-
eters,” Hewlett-Packard J., vol. 6; September-October, 1954. such that the latter become part of the four arm junc-
751/531
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES July
B = S,i
C = S21S42 — S 41.S 22
D = Sn Fig. 3—Reflectometer with auxiliary tuners.
where the Sm>n are the scattering coefficients of the four justed to produce a null in arm 4. If TV is now adjusted
arm junction comprised of the directional couplers and such that the reflection coefficient
lossy fourpoles.
b*
It is evident that the desired response will be realized r2i = —
if B = C — 0. For ideal couplers of infinite directivity and a2
a main guide VSWR of unity (and matched detectors)
observed at arm 2 vanishes, it can be shown2 that
the terms Sn, S42, and S22 are all zero, and
B 21*5* 4 2 — ‘S'416'22 = 0,
bz A
= — Tt and the desired operating conditions have been realized.3
bi D
It will be noted2 that the first adjustment (of Tx) is
as required. independent of the second, while the converse is not
In practice, the failure of the directional couplers and true, thus the adjustments should be made in the order
detectors to meet these criteria may be compensated or indicated. In addition, while Tz is employed in the sec¬
corrected for by the introduction of tuners at positions ond tuning operation, once the proper adjustment of
X and Y as shown in Fig. 3 (the tuner at Z serves an Ty has been realized, Tz may be readjusted (as will sub¬
auxiliary role to be described later). The adjustment sequently prove desirable) without affecting this result
of Tx and Ty to produce the conditions B = C= 0 may obtained by adjustment of Ty.4
be carried out in a variety of ways, two of which will be The alternative procedure will now be described. Re¬
described. The first of these tuning procedures perhaps ferring again to (1), it will be evident that even before
gives one a better feel for the conditions under which the introduction of tuning transformers, the terms B
the adjustments may be physically realized, and is in¬ and C (for a reflectometer assembled from commercial
cluded for the sake of completeness, while the second
method, the one which is recommended, is more con¬ 2 The proof of this statement and several of those to follow may
be effected by solving the scattering equations for the indicated
venient and potentially more accurate. quantities under the stated conditions, or by obtaining the scattering
parameters of the individual components, i.e., tuners and directional
couplers. The proofs are generally straightforward but tedious.
Adjustment of Tuners 3 The minimum requirements which must be satisfied by the two
couplers in order that these adjustments may be physically realized
In the first method, arm 2 is terminated in a matched have not been determined at the time of this writing, but.would
appear to be well satisfied by commercially available components.
load-and Tx adjusted for a null in arm 3 (with arm 1 4 The expression (S21 Sn — Sn S22) may also be written
connected to the generator). This produces the condi¬
tion £31 = 0 (by definition of the scattering coefficient).
The generator is then connected to arm 2; a load (not
*(^-*)
It can be shown (see note 2) that the second factor is invariant to
necessarily matched) connected to arm 1, and Tz ad¬ the adjustment of tuner Tz.
752/532
7959 Engen and Beatty: Microwave Reflectometer Techniques
components) are quite small with regard to A and D re¬ would be quite small, so as a first step (where only the
spectively. Consider the system response to a phasable phase of Tz is adjustable), it is usually desirable to first
or sliding load of such magnitude that |^Hh| ~|S|. adjust for a null in arm 3 for an arbitrary position of the
From inspection it is evident that the numerator of (1) load and then adjust for a constant ratio
will undergo marked changes in amplitude, while the
denominator remains relatively constant as the phase h
of the load is varied; and if Tx is adjusted to minimize b4
the variation in the ratio
as the phase is varied. It will be noted that the magni¬
h tude of 63/64 depends upon the magnitude of Ti.
b4
It is thus of interest to note that the technique re¬
as the phase of T/ is varied, the condition S3i = 0 will be quires neither a perfectly matched load or ideal short,
approximately realized. The sliding load of low VSWR but only two phasable loads of different reflection co¬
is then replaced by one of large VSWR (a sliding short). efficient magnitudes. The more closely these ideals are
Variations in output as the position of the short is realized however, the more rapidly will a series of these
changed will now be predominantly due to variations operations converge to the desired conditions, which in
in the denominator, and if Ty is adjusted such that practice can usually be realized to the required accuracy
| 63/641 is again constant, the condition S21S42 — S41S22 = 0 by only one adjustment each of Tx and Ty. It is also of
will also be very nearly realized. interest to note that if a perfectly matched load (Tz = 0),
A more complete mathematical treatment of the an ideal sliding short (Ti = eie), and a dissipation free
above procedure yields three solutions for a constant transformer at X are assumed, the two tuning opera¬
magnitude of 63/64 as the phase of 1/ is varied. The first tions, in this method, are completely independent of
solution, A/C = B/D is trivial since it gives for 63/64 a one another. That is, the adjustment of Tx to yield the
value which is entirely independent of IV The second condition 5 = 0, as noted earlier, is independent of the
solution, the one of interest, is: adjustment of Ty, while the adjustment of Ty for the
condition B/A = (C/D)* is independent of Tx 3
Having completed these adjustments, the response
becomes:
753/353
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES July
A variety of techniques are available for measuring will be noted that the reference planes have been chosen
the ratio to coincide with the terminal surfaces of the directional
coupler, thus exhibiting the dependence on the de¬
h
tector impedance explicitly.
bi Eq. (5) is of the same form as (1) and the adjustment
of TV and TV to make F and G vanish may be carried
In general, both bz and &4 will change with generator
out in the manner already described except that in step
output and load impedance, requiring a ratio type meter
two of the first method Tzi = bi/is made to vanish by
such that
adjustment of TV with arm 1 terminated in the gen¬
bz erator impedance, instead of the procedure described
b4 earlier.
Measurement of the ratio
is indicated directly, or a pair of instruments to deter¬
mine | 631 and |&4| individually. Alternatively, a feed¬ | ^31 u j ru |
back servo loop may be employed to keep | bt\ constant, | b3 |s | r* |
Analysis of Errors6
~(1 - SnTg) S12 s i*r d
The sources of error, in determining |r„| and <ru by
G = - SziTg S22 S2 zTd these methods to be discussed in this section include:
SziTg Sz2 -(1 - SzzTd) a) incorrect measurement of r, and uncertainty in value
of j r,| , and b) improper adjustment of the tuners, such
(i - suTg) s13rd
that B and C do not vanish.
■SaiT,, (1 — VaTi) The consideration of other sources of error is outside
the scope of this present paper, but will be treated more
ba is the equivalent generator voltage wave amplitude,
and Tg and Td are the generator and detector reflection
6 The analysis also applies to the modified reflectometer if the
coefficients at reference planes 1 and 3 respectively. It quantities E, F, G, H are substituted for A, B, C, D.
754/354
1959 Engen and Beatty: Microwave Refiectometer Techniques
fully in a subsequent paper (in preparation) on the ap¬ tion coefficient of the load of small VSWR, and a reflec¬
plication of the technique to measurement of bolometer tion coefficient of unity has been assumed for the load of
mount efficiency. large VSWR.
The error due to a) may be determined by inspection. Except for the presence of the factor lb, these equa¬
If the equation for | Tu| is written in the form: tions for K\ and K2 are of the same form, and values for
| B/A | and | C/D | may be obtained from Fig. 5 where
h the value of j C/D| is taken from the line | Fq = 1. It will
I I u | I be noted that the evaluation of the right hand side of
r« | = I r. | —— = | r. | r, (6)
(7) further presupposes a knowledge of but for the
03
present purpose of assigning a limit of error, an accurate
^4 8
value is not required.
it is evident that the fractional error in |r„| will equal
the sum of the fractional errors in | T,| and r if the latter
are small.
With regard to the second item b) it will be recalled
that the condition for constant output as the phase of
the load is varied is:
B
Ti 2
A
b4 | CIh+ D\
As an example, if a sliding load of VSWR = 1.005 is
are due entirely to variations in the numerator or de¬ employed in the first tuning operation, and the varia¬
nominator as the loads of small and large VSWR are tion in output is reduced to 1 db, |5/d.| will have a
employed respectively. value of approximately 1.6 X10-4. If the variation in
A first order correction to (6) may be written as fol¬ the second step with the sliding short is reduced to 0.02
lows: db, the value for | C/D | will be approximately 1.2 X 10~3.
r„ - ru Assuming that the unknown load has a value |ru| —0.2,
(cr ~ 1.5), |r,| = 1.000 (corresponding to a short-circuit)
r3r„
and assuming the terms in the right hand factor of (7)
c
+ — (r, - r„) + ■ • • .
combine in the worst phase, values of 6 and 1.2 for
(7
r, - rj
The ratios | B/A | and C/D | may be determined from rsw I
the expressions:
and |r,-r„| obtain respectively, for a total error of
B -
K\ = 20 log + (1.6 X 10-4 X 6 + 1.2 X 10-3 X 1.2) » ±2.4X10~3, or
1 + 2
a r, J + 0.24 per cent.
and Acknowledgment
755/355
NATIONAL PHYSICAL LABORATORY
PAPER 4-3
by
(40305)
756/1&2
PRECISION MILLIMETER WAVE INTERFEROMETRY
AT THE U.S. NATIONAL BUREAU OF STANDARDS
by
ABSTRACT
1. INTRODUCTION
(40305)
757/3
hence radiate a relatively wide angular spectrum of plane waves, Instead
of the Ideal single plane wave along the axis of the Interferometer. The
angular spectrum of plane waves radiated by a given aperture may be deter¬
mined by measurements of the distant field, as In antenna pattern
measurements.
Wavelengths measured on a microwave Interferometer are In general too
large due to the resultant Interference effects of plane waves radiated by
the aperture, and to modifications of these by the finite- size of
reflectors used. The magnitude of the error depends on the path lengths
used In the Interferometer, and on the aperture and reflector dimensions.
When these are some 15 wavelengths In extent, the error may be as much as
a few parts In 104 In a microwave Mlchelson Interferometer (ref.l).
A similar remark applies to the microwave Fhbry-Perot Interferometer,
(ref.2), although further work Is necessary to determine the magnitude of
the error.
In the application of precision microwave Interferometry to the deter¬
mination of 'c', and to metrology, a diffraction correction to the wave¬
length measured on the Interferometer must be considered. Froome (refs. 3,
14, 5) In the precision determinations of ' c* by variants of the Mlchelson
(40305) 758/4
Flg.l. Microwave Interferometer Reflector and Mount
balls running In two V. grooves on the base plate. This rests on a sub¬
stantial granite block to give stability to the system. A displacement of
the reflector of the order of a meter can be made, and the meter bar,
which Is supported above by ball bushings on the steel rod, can be swung
Into position to measure the displacement. The small differences In length
between an Integral number of fringes and the meter bar, will be measured
by a suitable gauge such as an optlmeter. This also has sufficient
measuring range to allow for path length changes due to expected varia¬
tions In refractive Index of the air. Facilities for accurate lining up
of the measuring anvils and the reflector are provided, and auto colli¬
mators will be used to see that the orientation of the reflector and the
meter bar alignment are maintained during the course of the experiment.
Figure 2 Is a diagram of the complete Interferometer. Two alternative
radiating apertures 30 cm and 60 cm square respectively will be used,
and are each formed In the usual way by tapering up from rectangular
waveguide. Matched polystyrene lenses are used In the apertures in order
to reduce the beam widths radiated, which are around 2 degrees and 1 degree
respectively for the present operating frequency of 47,730 Me. Radiation
(40305) 759/5
at this frequency is obtained by multiplying up from a 5 Me crystal con¬
trolled oscillator, a power output of 1 - 2 mw being finally obtained,
o
with a frequency stability of a few parts in 10' . The microwave radiation
is divided by the hybrid junction shown in figure 2\ half of the energy
passes down the reference waveguide arm, and half passes through the
radiating horn out into the open, where it is ultimately returned after
reflection from the movable reflector. Hie returning waves from the
reference arm and reflector are then combined in the detector arm, and
suitable adjustments on the reference level can be made to equalize the
amplitudes of the two signals. Hie fringes are thus due to the inter¬
ference of sinusoidal wave trains which have traversed different paths
and apart from the effects of diffraction, the shape of the recorded
fringes will depend on the type of detector employed. Sharp minima at
displacement intervals of k/Z may be obtained by the use of linear detec¬
tion with large amplification and precise balancing of the beams, the
fringe shape then varying as |cos 4>/Z |, where 4> is the phase angle
between the two interfering beams. Hie balance of the beams changes with
distance ' d' due to diffraction, and if the minima are not sharp enough
over the region of operation used, some adjustment for this must be used.
(40305) 760/6
Operation in the Fresnel region has the advantage that the beams are
well collimated In this region, and the effects of stray radiation, which
are troublesome In distant field operation, are considerably reduced.
Also the attenuation In the open arm Is less In this region. The diffrac¬
tion correction, however. Is more difficult, especially If the reflector
size must be considered. We require the effective reflection coefficient
of the open arm when the reflector Is In the Fresnel region of the radiat¬
ing aperture, and hence the manner In which the amplitude and phase of
this vary with the distance ' d» of the reflector from the radiator.
Deviations from the Ideal phase change with displacement, viz 2 ltd radians
where k - 2rr/\, can then be found and corrections applied. By thus
correcting for diffraction, and also measuring the refractive Index of
the medium surrounding the Interferometer, the wavelength In vacuo may be
found from the number of fringes In a given displacement. The refractive
7
Index can be observed to an accuracy of around 1 part In 10 by observing
the shift In the resonant frequency of a high Q microwave cavity, when
evacuated and when filled with the medium (refs. 6,7). The velocity of
electromagnetic waves In vacuo Is then determined- from the known frequency
and measured wavelength.
An outline of a derivation of the diffraction correction may be given
as follows*. The field at any point In front of the antenna Is due to the
superposition of the angular spectrum of plane waves radiated by It, and
this leads to the representation of an antenna In terms of a scattering
bn
(40305) 761/7
572772 0 -61-49
bl S11 j S12 S13 ~ 3ln~j
—
1 —1 1-1 1- -1
1 b2 1 1 a21 | 1 S22 S23 ~ ” S2n|
1 1 | | 1
1 1
1 c 1 S33 _____
1 b3 1 -
| S32 s3n 1
1 31 1
1 1 1 (1)
1 1 1 l
11 1
1 1 1 1
1 1
1 1 S-nl 1 |_Sn2 sn3 "* ” SnnJ
l_ S' i__ J
(40305) 762/8
and then recombined on entering the horn again. The appropriate Impedances
to use are those for propagation In the z direction and are given by
(ref. 1l).
and
_ COfJL
ri sec 6 -~z— for E perp. to plane of Incidence (4)
Kz
ZoJ_* (5)
sij sij "Z
ol
(40305) 763/9
solve for b1/a1 One obtains
b _1
1 - S11 - S^2 exp (-210) [l + SJ>2 exp (-21ip)]~ S 21 (7)
a1
where \p - k^d. Here the primes denote the column vectors, row vectors,
etc., corresponding to the partitioning of the scattering matrix shown In
equation (1). The term Sg2 Is quite complicated and gives rise to multiple
reflections between antenna and reflector; this effect will usually be
small In a well designed interferometer, and can be reduced by the Inser¬
tion of attenuation In the beam. It will therefore be assumed negligible.
The field distribution In the horns to be used approximates that of
the H10 rectangular waveguide mode, whence the radiated angular spectrum
Is
cos(k_a/2) sln(kyb/2)
where 'a' and 'b• are the aperture dimensions, C1 Is a constant, and the
electric field In the aperture Is In the y direction. Resolving this
field into components In and perpendicular to the plane of Incidence,
using the reciprocity relation given by equation (6), and In the limit
of a continuous distribution Identifying
1
al
- K exp(-12kd) SS y[g(kx, ky)]2 exp [l2kd (1 - cos 6)] dkx dky (9)
2 2 2 2
Y = (ky coe/kz +^kz/cu/x) / (kx + ky) (10)
(40505) 764/10
[g(kx,ky) gC-k^.-ky)] when g(kx,ky) Is not symmetrical. In a similar way
the effect of mirror tilt can be considered by Inserting the value of
g(-k^,-k-p for the returning waves where k^.,account for the tilt. Hie
Integral In equation (9) may be written as
+ 10(d,a)
| f (d, a) |e (H)
for a square aperture, where for large apertures and hence sharp radia¬
tion patterns, \p(d,a) Is a small phase change due to diffraction, and
|f (d,a) | corresponds to the modulus of the reflection coefficient and
Is nearly constant for distances d' well within the Fresnel region.
In the operation of the Interferometer we note the reflector displacement
D = d2 - d^ for which the phase of the reflection coefficient changes by
2h77, and hence
where and correspond to the value of i//at the Initial and final
distances ' d' used. Equations (9) and (12) give the diffraction correc¬
tion to the simple Interferometer equation. Thble I shows results of
some calculations made using equation (9), the apertures being square
with sides of dimensions 'a', 1d' Is the distance between horn and
reflector and k the wavelength.
(40305)
765/11
Equation (9) is applicable when the reflector may be considered as
infinite in extent, so that the Incident angular spectrum is reflected
unchanged. If its dimensions are finite compared with the wavelength, the
radiated spectrum is modified on reflection, and this must be considered.
Let k^, k^ be components of a propagation vector for the radiation from
the reflector, then neglecting edge effects the incident angular spectrum
g(kY,
a
kv)
y is modified according to the formula
sin(K^b/2) sln(K^b/2)
°(k^,kp gfk^ky) K, exp(-lkzd) dkx dk^ (13)
where = (k^. - kx), K^. = (k^ - ky), • b* is the side of the square
reflector, d the distance from the radiating aperture, and integration is
over the radiated spectrum g(kx,k ). Equation (13) is obtained by applying
the relation corresponding to equation (2) to the distribution of Induced
electric field over the reflector face due to the incident field. The
consequence of this is that equation (13) replaces a factor g(k_x, k^)
exp(-ikr7d) in equation (9). It is thus advantageous to use a large re flee-
tor when operating in the Fresnel region, since the diffraction correction
becomes quite complicated otherwise.
(40305) 766/12
X/4 DIELECTRIC SHEETS
/ \
COMPOSITE REFLECTORS
n 1 2 3 4 5 6 7 8
0.4378 0.7340 0.8861 0.9528 0.9806 0.9913 0. 9961 0. 9977
rn
(40305) 767/13
Figure 5 shows the basic element of such a structure, the particular
grating being represented by a discontinuity transmission matrix symmetri¬
cally located on a length '11 of parallel plate transmission line. The
matrix representing the discontinuity may be derived from the equivalent
circuit of the grating, and the reflectivity from a finite number of such
elements must be determined. Only TEM mode interaction between the grat¬
ings is considered, and the spacing between the grating elements is less
than k/2 so that all other possible modes on the structure are highly
attenuated.
1/2 1/2
GRATING ELEMENT
Neglecting losses the equation for the basic element shown in figure 5
may be written
Vi A B
Vn
Xn+1 C A :n
,
where A = s cos 9 + [i (w+t)/2] sin 9 B = [(w+t) /2] cos 9 + is sin 9 -
(w-t)/2, C = [(w+t)/2] cos 9 + is sin 9 + (w-t)/2, and 9 = 2771/A.. The
reflectivity from a number of such elements, and the spacing '1' for
optimum reflection can then be determined from the effective propagation
constant P per section, and characteristic impedance Z of the element.
These are given by (ref.iu)
(40305)
768/14
and may be determined from equation (14) for any particular structure.
Hie overall matrix for n such structures in cascade is then given by
— —
*n 1 slnh-nT, cosh nP
Z
_ _
and since for purely reactive elements cosh P must be real, the stop
bands of the structure are given by sin jB = 0, and the pass bands by
slnh a = 0, where T = a + 1/3. For the pass bands
(Z2 - 1) sinh na
rn - P- (2D
2Z cosh na + (Z + 1) slnh na
For spaclngs such that 9 lies within the pass band, the reflection co¬
efficient rn will oscillate with increasing n, and becomes zero when
sin n/3 - 0. When '1' corresponds to a stop band, the reflectivity will
Increase monotonically with n, and it can be shown that the spacing for
maximum reflectivity corresponds to the value of 9 which gives the maxi¬
mum value of cosh a= |A| within the stop band.
As an example consider the capacitive rod grating shown in figure 6,
the electric vector being perpendicular to the rods. Values of the appro¬
priate reactances are available (ref. 15) and for rods of diameter
0.063 in., a wavelength of 6.28 mm, and a spacing of a/k - 0.417, we
(40305) 769/15
1 ~kJ kj kj
a
1 -C\ r\
E
l—f J-J—I
O O O CEIy
Zb zb
O O O O -K- -)l—o- o
O O O O h. -o
n 1 2 3 4 5 6
0. 48542 0.87963 0.97699 0. 9 9 583 0. 99925 0.99986
U2
I IP
Thble III. Calculated reflectivity |r | from a number n of
stacked capacitive rod gratings at k/2 spaclngs.
Rod diameter 0.063 In., spacing 'a' 0.103 In.,
k = 6.28 mm.
(40305) 770/16
Fig.7. Pass and Stop Bands for Thick Gratings
where C2 and are constants , g(kx,ky) Is again the radiated and received
angular spectra of the apertures, here assumed Identical, z1 and z0 are
(40305) 771/17
Fig.8. Fringes Recorded on Microwave Fabry-Perot Interferometer
with Capacitive Rod Grating Reflectors. A. - 6.28 mm.
the distances of the radiating and receiving apertures from the reflec¬
tors, and rT and tT are the amplitude reflection and transmission co-
efficients of the Interferometer section shown In figure y. Assuming no
losses In the reflector elements, these are given by
(40305) 772/18
for the Michelson interferometer, the only change Is that the reflection
and transmission coefficients of the reflector system are now Introduced.
The shape of the reflected and transmitted fringes thus depends on the
radiation pattern gfk^, k^), and provided the reflectivity achieved Is
high enough, this Is effectively scanned by the filtering action of the
reflector system, as the distance 'd' varies. Also for high reflectivity
and discrimination in the reflector system, the reflected power will
always be quite high, even at the spacings for optimum transmission,
since only a few of the Incident plane waves will be transmitted by the
reflectors at a given spacing.
The setting accuracy In this inteferometer thus depends on the reflec¬
tivity, and on the radiation pattern, and these must be such that the
fringes become very sharp, especially for precision work. The maximum
transmitted fringe intensity will tend to occur when 'd' is such that
the portion of the radiated angular spectrum centered about the maximum
amplitude is transmitted, and for very large reflectors the diffraction
correction should be small If the reflectivity is extremely high so that
only a small portion of the radiated angular spectrum is effective
between the reflectors. Further work is required to establish the pre¬
cision possible with this Interferometer, and to determine the magnitude
of the diffraction correction for various reflectivities. This Is
especially so when the reflectors must be considered as finite compared
with the wavelength, as the radiated angular spectrum will be modified
in a complicated way by the multiple reflections, and this will affect
the diffraction correction.
4. CONCLUSIONS
The theory given for the diffraction correction In the Michelson inter¬
ferometer operating in the Fhesnel region represents the solution of the
problem for a reflector of infinite size. Table I, which applies for the
stated aperture field distribution, shows that for apertures some
hundred wavelengths in extent, the correction amounts to around two parts
in 10 . Thus if it Is known to 1% the results should be accurate to a
7
few parts in 10 . In practice the antenna radiation pattern will have to
be measured quite accurately, and the measured pattern used to determine
the diffraction correction. Both the amplitude and phase of the angular
spectrum g(kx,ky) are required, and measurements of the antenna pattern
should thus Include phase as well as amplitude measurements, or else the
phase must be deduced from Information on the phase distribution In the
aperture. As Indicated, for these aperture sizes the scalar diffraction
theory approximation is adequate to at least this order of accuracy.
(40305) 773/19
These results are valuable, since applications of microwave Interferometry
to metrology would most likely require operation In the Fresnel region of
large radiating apertures.
Advantages of operation In this region, are that the radiation Is well
collimated, and effects of stray radiation are less troublesome. Also the
attenuation In the open arm of the Interferometer Is considerably less,
with a consequent Increase In available signal power; an extremely
Important consideration at millimeter wavelengths. To alleviate multiple
reflection troubles In the Interferometer, some attenuation In this arm
may be necessary, but It Is not expected that large values will be
required. The main disadvantage Is that for precise work large radiators
and reflectors are required for the wavelengths presently available, and
this leads to some mechanical problems as Indicated by the Michelson
Interferometer now under consideration. The work also represents a
different approach to the diffraction correction than that used by Froome,
and Is Important for this reason. As In his work. It Is expected that the
accuracy In the determination of ‘C will be limited by the length
measurements, and by the determination of the refractive Index of the
air.
The Fhbry-Perot Interferometer with Its sharp fringes and basic sim¬
plicity, represents a very promising Instrument for application to the
determination of ' c', and the reflector designs discussed should ade¬
quately meet the problem of obtaining high reflectivity In the microwave
region. The diffraction correction Is more difficult especially If the
reflectors are finite compared with the wavelengths, but calculations
assuming Infinite reflectors will give an Indication of the magnitude
of this. There are also possibilities of using evacuated "etalons" In the
microwave region, with the consequent elimination of the difficult
measurement of refractive Index at short wavelengths. The high wavelength
resolution due to sharp radiation patterns and high reflectivity, could
also prove useful In the wavelength analysis of millimeter wavelength
sources too high in frequency to be measured conventionally.
Finally, the general complexity of the formula Involved In the diffrac¬
tion correction for the Fresnel region, especially In the microwave Fhbry-
Perot Interferometer, Indicates the great advantages of using very short
wavelengths, preferably In the one millimeter region or below. At these
wavelengths apertures of dimensions approaching 1000 times the wavelength
could be used, with a considerable reduction In diffraction effects.
(40305) 774/20
5. ACKNOWLEDGMENT
The authors would like to acknowledge the assistance derived from the
Initial work of Dr. F. S. Dayhoff on the diffraction correction In the
Fresnel region, and axso the work of William W. Longley, Jr., and
Dr. Paul F. Wacker on the computation of the Integrals representing the
diffraction correction.
REFERENCES
(40305) 775/21
DS 40311/2/1230 300 2/59 TCL
ing a previously calibrated r-f attenuator,
A Standard of Attenuation for together with the heterodyne method,
extends the heterodyne attenuation range
to that of the direct substitution method,
Microwave Measurements although with a decrease in accuracy.
General Description unknown is then equal to the increase of 0.05 per cent (0.002 db) change in power
attenuation of the standard. level. Variable attenuator A, in Figure 1,
Because of the extremely wide fre¬ This method assumes that the fre¬ is used to calibrate the power monitor.
quency range over which attenuators quency converter is linear, that is, that The r-f generator klystron is followed
must be calibrated, the construction of a the i-f power from the converter is pro¬ by buffer attenuator B which prevents
series of standard microwave attenuators portional to the input microwave power. changes in impedance caused by inserting
to cover the entire spectrum, and the use For small enough input power this is true and removing the unknown attenuator
for a crystal converter, and the operating from changing the load impedance pre¬
Paper 48-50, recommended by the AIEE instru¬ conditions for a prescribed departure from sented to the klystron. Thus the power
ments and measurements committee and approved
by the technical program committee for presenta¬ linearity may be determined experiment¬ output and frequency of the r-f generator
tion at the AIEE winter general meeting, Pitts¬ ally. are maintained constant throughout a
burgh, Pa., January 26-30, 1948. Manuscript sub¬
mitted November 12, 1947; made available for The range of attenuation which may be measurement. A twenty-db attenuator
printing December 15, 1947. provides an adequate amount of buffing,
measured by the heterodyne method is
R. E. Grantham and J. J. Freeman are both with as evidenced by the fact that the intro¬
less than that measurable by direct sub¬
the National Bureau of Standards, Washington,
D. C. stitution, since in the heterodyne method duction of a 10 to 1 voltage standing-
The authors wish to express appreciation for the the maximum permissible power input wave ratio at transformer A resulted in
valuable services of R. H. McCracken who con¬
structed much of the electronic equipment and into the converter is limited by the ad¬ no change of generator output, as indi¬
performed many of the experiments, and A. A. vent of converter nonlinearity. However, cated by the power monitor.
Feldmann who designed the mechanical features
of the wave-guide-below-cutoff attenuator. a combination of direct substitution, us¬ When measuring the insertion loss of
776/1
Figure 1. Block diagram of the intermediate-frequency method of attenuation measurement
LOCAL
OSCILLATOR
ADJUSTABLE VARIABLE
BUFFER CRYSTAL
transformei
'ATTENUATOR
MIXER
B C
an attenuator, it is necessary to insert the local oscillator may occur up to about 200
attenuator in a transmission line which is kilocycles per second without causing ap¬
matched looking both ways from the preciable variations in the output differ¬
point of insertion. To meet this require¬ ential voltmeter. If the local oscillator tenuators, frequency and amplitude in¬
ment it is necessary to use adjustable were operated c-w, the beat frequency stability and errors in the i-f standard at¬
transformers A and B to match attenua¬ could not vary more than about plus or tenuator. Errors of the i-f standard will
tors B and C respectively as accurately as minus 3 kilocycles per second for the same be discussed in another paragraph, and
possible to the characteristic impedance variations in output. the question of errors due to mismatch is
of the wave guide. too lengthy and detailed to consider at
Buffer attenuator C is needed to iso¬ Range of Measurement and Errors this time. If we neglect the last two
late the local oscillator from the changes of Heterodyne Method sources of error, the over-all accuracy of
in the impedance of its load which would attenuation measurements with the equip¬
occur when the unknown attenuator is The attenuation range measurable by ment described herein is estimated to be
inserted and removed. The conversion the heterodyne method is determined by within plus or minus 0.02 db in the 0-10 db
efficiency of the crystal changes appre¬ the maximum input power consistent with range, and plus or minus 0.2 per cent of
ciably with local-oscillator power. Hence, crystal linearity, and by the crystal noise the attenuation value in db for the 10-50
the local-oscillator power output must be power. The range of crystal linearity was db range.
maintained as constant as possible. determined by measuring a fixed value of
Twenty db is the minimum attenuation attenuation as a function of crystal input Theoretical Factors Affecting the
used for buffer attenuator C. power, using the apparatus indicated in Choice and Design of the
The repeller voltage of the local-oscil¬ Figure 3. The measured values of attenu¬ Attenuator
lator klystron is swept by a linear saw¬ ation of a 30-db pad as a function of r-f
tooth voltage so that the beat-frequency buffing, and hence crystal input power, To be suitable as a standard, the i-f at¬
output of the mixer sweeps over about are depicted in Figure 4. For 100 milli¬ tenuator must be able to yield known
0.5 megacycles per second. The i-f am¬ watts of r-f input power, it appears that values of attenuation in terms of an ac¬
plifier band width is about 50 kilocycles 40 db of buffing attenuation (correspond¬ curately measurable parameter, such as
per second, so the i-f response curve is ing to a crystal input power of 10 micro¬ length, or it must be capable of calibra¬
plotted on the oscilloscope as in the usual watts) suffices for accuracies of plus or tion in terms of d-c measurements. First
spectrum analyzer, but with much more minus 0.02 db. designed by Harnett and Case,3 the cutoff
than the usual frequency dispersion. The crystal noise power was measured attenuator, Figure 6, which consists of a
The peak voltage of the pulse on the os¬ to be less than one micromicrowatt. Ac¬ hollow tube excited at one end below its
cilloscope is measured by a differential cordingly, the maximum range of attenu¬ cutoff frequency, and a coil or condenser
vacuum-tube voltmeter, which has a ation measurement corresponding to an which picks up the attenuated field at the
sensitivity of one scale division per 0.2 error of 0.05 db due to crystal noise, is 50 other end, fulfills the former requirement.
per cent (0.01 db) change in power level. db. Since the generated field falls off expo¬
The reason for sweeping the local-oscil¬ In addition to possible crystal nonlin¬ nentially with distance from the exciting
lator frequency rather than operating it earity, errors in the heterodyne method source, and since the attenuation constant
continuous wave (c-w) is that small shifts may occur through r-f or i-f leakage, mis¬ may be computed from the dimensions of
in the frequency of the r-f generator or the match on either side of the r-f or i-f at¬ the tube, the ratio of any two voltages, or
777/2
572772 0 - 61 - 50
the attenuation introduced by a given 0; of these the mode with the lowest at¬ TEU mode, and as e~a°'z for the TM0
length of tube, is reduced to a measure¬ tenuation constant is the TMoi- The mode, where z is the separation between
ment of length. danger with this type of attenuator is that exciter and receiver. Accordingly, a
any asymmetry will produce a TEn modp further advantage of the TEn mode over
Cross-Section of Tube
which has a lower attenuation constant, the TMoi mode attenuator is the smaller
and hence, for a large enough value of at¬ interaction between exciting and receiv¬
However, it is well known that the elec¬
tenuation, can lead to appreciable error. ing circuits for a given attenuation.
tromagnetic field generated within the
Rectangular Cross Section. The first
tube consists of the superposition of an Skin Effect
infinite number of modes, each of which four columns of Table I enumerate the
is attenuated along the axis of the tube as ratios of attenuation constants, anm, of the For a perfectly conducting circular
e-anmZ' where a„m is a function of the first few modes to that of the lowest mode, tube, of radius a,
geometry of cross section, and the particu¬ for several ratios of rectangular cross-
1.8411"
lar mode. The higher the mode, the section dimensions, a and b. It will be no¬ an =-
ticed that for a = b/2, or less, the un¬
a L
greater is its attenuation constant anm, so
for the TEu mode (1)
that only for large enough values of z do wanted modes decay much faster than
for circular cross-section. Also, if the ex¬
the higher modes become negligible. Ob¬
citing-current distribution is chosen so
viously, then, the accuracy of the attenu¬
that to each element of current corre¬
ator increases as the ratios of the ampli for the TMoi mode (2)
sponds its negative mirror image with re¬
tudes of the higher modes to that of the
lowest one decrease, and as the ratios of spect to the plane x = a/2, and its positive
If one takes into account the effect of finite
mirror image with respect to the plane
the attenuation constants of the higher conductivity, o, following the method of
y = b/2, then it may be shown that only
modes to that of the lowest one increase. Stratton,6 one gets easily, for the attenu¬
those modes are excited for which m is
The comparison of these factors for tubes ation constans.
even and n is odd. Although this type of
of circular and rectangular cross section
symmetry is difficult to achieve at lower
is given in Table I.
frequencies, using coil structures, at mi¬
Circular Cross Section. The fifth col¬ crowave frequencies this type of sym¬
for the TEu mode (3)
umn of Table I enumerates the ratio of metry should be easily obtainable using
attenuation constants, anm, of the first symmetric windows as the exciting struc¬
few modes to that of the lowest mode, for oi'oi
tures. a L \2.405 X/ \ a) J
circular cross-section. f 7 S I, T i /I n
tance of the coil makes the achievement Mode a T Eoi ot T Eoi a T Eoi a T Eo\ a TEn Mode
778/3
Experimental Evaluation of Cutoff
Attenuator
Mechanical Description
A buffing attenu¬
Edb —10 logio
ation of 40-db cor¬
[1 + r2 exp { — 2bso) — 2r exp ( — bzd) cos f]
responds to a crystal
(12)
25 30 35 40 45 50 55 input power of 10
BUFFING ATTENUATION—OB microwatts The value of Edb for a given minimum
779/4
^8ubraax> a third mode must be present.
Using the measured values of 5submai
and Bmin as defined herein, the values for
r and f may be determined from the fol¬
lowing equations, as derived by Griesh-
eimer.5
cqs (20)
(i+J)
F2aubmax
V2 max
and
VK
V2max
V2
•8 — 10 logi, (13)
F2m»,
V2
v 81
8submax — 10 log (14)
V2
VKun
-^min — 10 logi (15)
1 F2max
UJ8ubmax
= h2
F2 max
V2
- = /32 (16)
V2 Figure 8. Calcu¬
lated insertion loss
Figure 7 is a typical plot of B as a func¬
of barium titanate
tion of angle between the axes of the coils,
disk for TM0i mode
and illustrates the power coupled into the versus thickness of
receiving coil as a function of angle of disk
780/5
in attenuation was 0.08 db. The latter
error was not detected by measuring the
5-db step of the Leeds and Northrup at¬
tenuator because the error was spread over
such a wide range of attenuation that the
amount included in a 5-db interval was
smaller than the experimental error.
U-nmj
Z nm (22)
2ir/a£i
aui^fj
Z nm (23)
V nm
781/6
Conclusion the heterodyne method is a waveguide- 2. Radio-Frequency Attenuators, S. G. Sydo-
riak. Radiation Laboratory Report 404 (Massa¬
below-cutoff attenuator operated at an chusetts Institute of Technology (Cambridge,
The heterodyne or i-f substitution Mass.), September 7, 1943.
intermediate frequency. A preliminary
method of calibrating microwave attenu¬ 3. The Design and Testing of Multirange
model of an i-f standard cutoff attenuator Receivers, D. E. Harnett, N. P. Case. Proceed¬
ators has been investigated, and prelimi¬ has been built, tested, and improved until ings, Institute of Radio Engineers (New York,
nary equipment has been built to utilize N. Y.), volume 23, number 6, June 1935.
its accuracy is plus or minus 0.01 db per
the method. The accuracy obtained is 4. Capacitance Type Radio-Frequency Attenu¬
20 db unit of attenuation. ators, W. O. Smith. Radiation Laboratory Re¬
estimated to be within plus or minus 0.02 port 995, Massachusetts Institute of Technology
(Cambridge, Mass.), January 18, 1946.
db in the 0-10 db attenuation range, and
plus or minus 0.2 per cent of the attenua¬ References 5. Notes on TEu Mode Cutoff Attenuators,
R. N. Griesheimer. Radiation Laboratory Report
tion value in db for the 10-50 db range, 55, Massachusetts Institute of Technology (Cam¬
1. A Method of Calibrating Standard-Signal
bridge, Mass.), November 16, 1944.
assuming unity standing-wave ratio pre¬ Generators and Radio-Frequency Attenuators,
sented to both sides of the attenuator un¬ G. F. Gainsborough. Journal, Institution of Elec¬ 6. Electromagnetic Theory (book), J. A.
trical Engineers (London, England), volume 94, Stratton. McGraw-Hill Book Company, New
der calibration. The basic standard of part 3, number 29, May 1947. York, N. Y., 1941, page 543.
Summary—Heretofore, the determination of attenuation from the method may be used, if measured impedances are
impedance measurements has been applied to cases in which the re¬ converted to reflection coefficients.
flections from the attenuator terminals were negligibly small. In
the proposed method no restrictions are placed upon the attenuator II. Theory
except the requirement that it be a linear, passive, four-terminal
network. Attenuation is defined as the insertion loss which oc¬
The dissipative and reflective components of attenuation Ad curs when an attenuator is placed in a matched system.*
and Ar are measured separately. Ad is shown to be a function of In the matched transmission-line system shown in Fig.
the efficiency of the attenuator. The efficiency is determined from
1, the generator and load impedances are equal to the
reflection coefficient measurements of the short-circuited attenua¬
tor. Ar is determined from a single voltage-standing-wave ratio characteristic impedance of the line. The generator
measurement of the attenuator when terminated in a matched load.
Experimental data show close agreement with an independent Zo
method of determining attenuation.
I. Introduction
782/895
PROCEEDINGS OF THE I.R.E. August
1 1
Ad = 10 logio — =10 logio — • (5)
Vm Rl
783/896
1950 Beatty: Determination of Attenuation from Impedance Measurements
_ _ II ii ^ l.O’max 0min7
2i?2 = | r |max - I r|min = —---- (9)
(o’raax + 1) (u'min + 1)
MEASURED ATTENUATION
At From
Nominal
Input Power Difference
Value of A Ad A
VSWR Ratio
Fig. 4—Measured reflection coefficients of 20-db attenuator, Attenuation
A.d~ 19.74 db, 1,000 Me, i?2 = 0.0106. Method
3 db 1.070 .005 db 2.87 db 2.88 db £85 db .03 db
decibels. The observed data in the measurement of Ad 6 db 1.235 .048 5.74 5.79 5.75 db .04 db
are shown in Figs. 3 and 4. (Lack of space prohibits 10 db 1.180 .030 9.50 9.53 9.53 db .00 db
showing more data.) It is apparent that the measured 20 db 1.240 .050 19.74 19.79 19.81 db .02 db
circles either enclose the origin or lie completely outside.
Fig. 6—Observed attenuation data.
In each case, R2 may be expressed in terms of the mini¬
mum and maximum value of voltage-standing-wave
The total attenuation At was also measured at 1,000
ratio measured as the position of the short-circuit
Me by a power-ratio method using a bolometer detec¬
changes.
tor8 which permits an accuracy better than +0.15 db.
The results of both methods are shown in Fig. 6.
-9 V. Discussion of Errors
784/897
JOURNAL OF RESEARCH of the National Bureau of Standards—C. Engineering and Instrumentation
Vol. 64C, No. 2, April-June 1960
1. Introduction
The following is a report on some recent work at the Radio Standards Laboratories of the
National Bureau of Standards at Boulder, Colo. It is believed that this work represents per¬
haps the most accurate measurements of microwave attenuation yet made.
In the measurement of microwave attenuation, the stability or resolution of the associated
measuring system places a limit on the accuracy with which small values of attentuation may
be measured, because the reduction of other sources of error (such as mismatch) will tend to
make the total error approach this value. In practice this limit has been typically of the order
of ±0.01 db.
The application of certain power stabilization 1 and measurement2 techniques to the
problem of attenuation measurement has yielded a measurement system with a nominal
hundredfold increase in resolution and stability. A practical application for demonstrating
the capabilities of this technique was recently provided in the calibration of a rotary vane type
of variable microwave attenuator which is cluiracterized, in part, by a high degree of resolution
for small values of attenuation. Thus when a recent application called for this characteristic
in a variable attenuator, and when a preliminary study of the mismatch and stability character¬
istics of an attenuator of this type indicated that such a step was warranted, it proved desirable
to experimentally confirm the mathematically predicted values of attenuation which were
engraved on the attenuator dial. This proved to be by far the most exacting practical appli¬
cation of the cited techniques encountered to date.
In order to take complete advantage of the improved stability, refined techniques were
applied to the evaluation and reduction of the mismatch error. The capability of the system
is indicated by the tabulated results of the attenuation calibration, and the estimate of the
limits of error is supported by an analytic and experimental treatment.
1 Glenn F. Engen, Amplitude stabilization of a microwave signal source, IRE Trans, on Microwave Theory & Tech. MTT-6, No. 2, 202-206
(Apr. 1958).
2 Glenn F. Engen, A self-balancing direct-current bridge for accurate bolometric power measurements, J. Research XBS 59, No. 2, 101-105
(Aug. 1957) RP2776.
785/139
2. The Calibration System
A simplified diagram of the calibration system is shown in figure 1. The attenuator under
test is placed between an amplitude stabilized microwave signal source 1 and a bolometer
mount and power meter. The power meter consists of a self-balancing d-c bolometer bridge 2
with provisions for measuring and recording the d-c bias power required to maintain the bolom¬
eter at its operating resistance of 200 ohms. The two bolometer mounts M1 and M2 shown
in the temperature stabilized water bath are for power measurement and signal source stabiliza¬
tion, respectively. A reasonable amount of care was exercised to obtain good performance
from each item of equipment, with a resultant system performance as shown in figures 2a and
2b. Figure 2a shows the stability and repeatability with the attenuator alternately set at
the 0.00- and 0.01-db positions. It will be noted that the stability and repeatability are better
than 0.0001 db (10 microbels). A recording of the long term system stability is given in figure
2b where the maximum variation is of the order of ±10 microbels.
WATER - BATH
The absolute accuracy of measurement of small changes in d-c power at this level is
estimated to be of the order of 0.02 pw. This indicates that further improvement could be
expected in the results if the system stability could be improved in some manner, perhaps by
use of a frequency-stabilized signal source.
3. Theory of Measurement
786/140
In making such a measurement, one assumes that the microwave power P absorbed by the
bolometer element is proportional3 to the d-c power IT withdrawn in order to keep the bolom¬
eter resistance constant. Letting ir0 represent the d-c power required to bias the bolometer
at its operating resistance when P equals zero,
A= 10 log10
TLo-Uq
(2)
W0-Wa’
where 11 x and Ilk are the d-c bias powers corresponding to Pi and P2. (For the bolometer used,
U o~15mw and Pl ~ lOmw.)
The apparatus employed permitted direct measurements of differences in d-c power, a
procedure permitting greater accuracy and convenience than calculation of differences from
separate measurements. The changes in d-c power level during an attenuation measurement
are shown in figure 3.
3 The constant cf proportionality is determined by the substitution error of the bolometer which is known from other experimental data to
be independent of power level.
787/141
<r
*
2
o
o
TIME
I
I
/
ATTENUATOR
s
LESS THAN
For attenuations less than approximately 3 db, the power difference W2—M\ was meas¬
ured directly. If we let W2l = W2—Wi, and TT0i= Ho— Wi, eq (2) may be written
For attenuations greater than approximately 3 db, the power difference 1F0— IF2 or TT02
was measured directly. Equation (2) becomes
A= 10 log10 Wm (4)
wnm
It is estimated that the error in measuring d-c power differences is within 0.1 percent
+ 0.1 pw. When W2~Wl is measured, it can be shown that the limit of error in determining
the attenuation is
1
6=10 log10 (6)
O.Imw+0.001 11+
if02
5. Mismatch Errors
The mismatch error4 in calibrating a variable attenuator depends upon the reflections from
the system in which the attenuator is placed and upon the changes in characteristics of the atten¬
uator as its dial is moved from the reference position. The graph of figure 4 shows calculated
1 R- W. Beatty, Mismatch errors in the measurement of ultrahigh-lrequency and microwave variable attenuators, J. Research XBS 52,
No. 1, 7-9 (1954) RP2465.
788/142
.01 0.1 I 10 100
ATTENUATOR DIAL READING, DECIBELS
limits of error 5. for the attenuator used, based upon measurements of the magnitude of the
changes in the scattering coefficients Sn and S22 of the attenuator. It was assumed that there
was no phase change in $21. The mismatch error is below 0.0001 db for attenuator settings up
to 0.1 db, and remains below 0.001 db for higher settings.
6. Results
The calibration data taken at 9.3897 kMc is shown in table 1. Three sets of data are
shown in order to give an idea of the reset Lability of the attenuator, and more significant figures
are given than one can normally use when interpolating between the marked dial divisions. It
should be noted that the calibration is not necessarily representative of this type of attenuator,
because only one such attenuator was calibrated by this procedure, and this was done at only
one frequency. It should also be noted that this type of calibration is not available on a routine
basis, because there are as yet no commercially available attenuators which can be read and are
repeatable to 0.0001 db.
The estimated limit of error for the complete range of the attenuator, as determined from,
figure 4, is also shown in table 1. Above 20 db, the calibration was made in two parts: Measure¬
ment of the 20 db step, and measurement of the additional attenuation referred to this step.
For these values, the quoted limit of error is the sum of the errors in the individual steps. The
accuracy of setting the attenuator dial on the marks is not as good as the accuracy of the
measurements.
It is noted that at the low end, the estimated accuracy of the measurement is also better
than the repeatability of setting the attenuator on the mark. One concludes that further
improvements in attenuators need to be made if we are to take full advantage of the most
accurate measurement techniques.
5 See appendix.
789/143
Table 1. Results of three sets of measurements and calculated errors in single measurements
db db db db db db db
0.01 0.01077 0.01083 0.01044 0.0107 0.0003 0.000055
.02 .02123 .02145 .02145 .0214 .0002 .000068
.03 .03015 . 03035 .03038 .0303 . 0001 .000080
.04 .04043 .04088 .04074 .0407 .0003 . 000091
.05 .05181 .05216 .05226 .0521 .0003 .00010
7. Appendix
The analysis (see footnote 4) of mismatch errors in the calibration of variable attenuators
yielded an equation for the error in terms of the scattering coefficients of two fourpoles corre¬
sponding to two settings of the attenuator dial, and the reflection coefficients of the system in
which the attenuator was inserted. The measurement of all these quantities may be tedious
or difficult, and an approximate method has been developed. One obtains reasonably close
limits within which the error lies from a fairly simple experimental procedure.
The complete expression for the mismatch error is
where scattering coefficients are denoted by Smn, and rG, VL represent, respectively, the re¬
flection coefficients of the system “looking towards” the generator and load. Primes are
used to designate a setting of the attenuator other than the zero or reference setting.
For small reflections, the following expression is derived from (la):
If the attenuator VSWR is small, the products S'uS'22 and Sn S22 may be neglected. Then
(2a) becomes
It is convenient to determine the magnitudes of the individual terms but not then1 phases,
so that the limit of error, allowing random phase variations in Sn and S22, but none in S2i is
790/144
The quantity \Sn—is determined as follows. With the attenuator under test con¬
nected as shown in figure 5, and set to its zero or reference position, tuner A is adjusted for a
detector null, and tuner B is then adjusted until the reflection coefficient V2i of the equivalent
generator at terminal plane 2 vanishes. (This condition may be recognized by means of an
auxiliary reflectometer.)
Movement of the attenuator dial to some other setting will then give an observable
detector output. Using the theory6 of a directional coupler with auxiliary tuners, one can
obtain \Sn—/SRI in the following way.
The amplitude of the output power from the sidearm of the directional coupler in figure 5
\+KYl
has the magnitude |&3|=& where k is a constant for a given stable generator oper¬
i-r2irL
ating level.
With the attenuator set first on zero, TL=Sn- Adjusting tuner A for |63| = 0 changes K
so that KSii= —1. Adjusting tuner B for r2i=0 makes the dependence of |63| on IT take the
simple form:
Suppose the attenuator dial is moved to a new position such that VL now equals S'u; then
S'n,_ k
i\Sn-S'n\ The factor &/|/S'n| is obtained by replacing the attenuator by a
'Snl I Su
waveguide section containing a sliding short. Upon-sliding the short, ;63| goes through
small variations so that one may observe |&3|max and |&3|mIn. It is easily shown that
One can assume that |&3|max= l^lmin and employ a fixed short circuit with negligible error if
the VSWR corresponding to Sn is less than 1.15. It may be that this error is tolerable for
higher VSWR’s since it is not important to know \Sn—SRI to great accuracy.
The quantity |*S'22—S22| is found in the same way as above with the attenuator turned
end for end.
The authors acknowledge the assistance of James E. Gilbert and Morris E. Harvey in
making the attenuation measurements and of Wilbur J. Anson and Ronald Ridge in matching
the system and making measurements to evaluate the limits of mismatch error. G. E. Schafer
read the manuscript and provided information on phase shifts in variable attenuators.
6 R. W. Beatty and D. M. Kerrs, Recently developed microwave impedance standards and methods of measurement, IRE Trans, on Instr.,
1-7, Nos. 3 A 4, 319-321 (Dec. 1958).
791/145
Journal of Research of the National Bureau of Standards Vol. 52, No. 1, January 1954 Research Paper 2465
(i - r;rG)(i — &2.,rL)
2. Expression for Mismatch Error € = 20 logio (5)
(i-r1rG)(i-/S,22rj
A change in a variable attenuator from the
reference position to another position is equivalent where Id and Ti are the input-voltage reflection
to removing one attenuator and inserting another coefficients of the attenuator terminated in a load
attenuator in the circuit. The insertion loss, in having a voltage reflection coefficient rL.
decibels, of the attenuator2 corresponding to the The mismatch error in the measurement of a
reference position of the variable attenuator is single attenuator can be obtained as a special case
of eq (5). The reference attenuator vanishes in
(i -sll rG)(i—S22rL)—S22rGrL this case, changing Id to rL, Sv2 to unity, and Su and
A = 20 logio (i) So* to zero. Substituting these values for Id and
s12(i-rGrL)
S22 into eq (5) yields
The voltage-reflection coefficients rG and rL refer
to the generator and load, respectively, and are (l—r;rG)(i-s;2rL)|
measured at the terminals where the attenuator is € = 20 logio
i—rGrL r (6)
inserted. The scattering coefficients3 Sn, Sl2, and
1 C. G. Montgomery, Technique of microwave measurements, p. 824 (McGraw- which corresponds to eq (24), page 826 of the refer¬
Hill Book Co.. Inc., New York, N. Y., 1947).
2 See Appendix. ence of footnote 1.
2 The scattering coefficients So, S12, and Sn are in effect defined by eq 12 and
figure 2. See also, C. G. Montgomery, R. H. Dicke, and E. XI. Purcell, Prin
ciples of microwave circuits, p. 146-151 (McGraw-Hill Book Co., Inc., New * Substitute rj = Sii+and rys'n+.^'j in eq (3).
York, N. Y„ 1948). 1—0221 L 1 —0J31 L
792/7
3. Evaluation of the Mismatch Errors 4. Appendix. Derivation of Expression for
Insertion Loss
It. is possible in principle to evaluate the mismatch
error by measuring the voltage reflection coefficients The insertion loss of a four-terminal network is
rG, I'l, rt, Ti, S22, and S'22, and substituting them defined by the equation
into eq (5).
In many cases, the magnitudes of the reflection
coefficients can be determined, but their phases Z,= 10logl„£s (9)
cannot be conveniently determined because of the
limitations of a particular measuring apparatus. where PL denotes the power delivered directly to a
Equation (5) can then be used to find the limits of load, and P2 denotes the power delivered to that
the mismatch error, permitting the phases of the load when the network is inserted between the load
reflection coefficients to have all possible values.
and the generator. It is assumed that the generator
The limit of error can be expressed in the form is stable and unaffected by changes in loading.
The insertion loss of a four-terminal network can
(i ±|r,1rG|)(i±ls;2rL|) be expressed in terms of its scattering coefficients
€ limit—20 logio (7)
aT|ihrG|)(iT|S22rL|)’ Su, Su, and S22, and the voltage reflection coefficients
rG and rL of the generator and the load.
and the corresponding limit of error for single Referring to figures 1 and 2, the insertion loss is
attenuators is
are 1.1, 1.1, 1.2, 1.5, 1.2, and 1.5, the limits of ZG-Z0 z, ~Zt
error for the initial attenuator calculated from eq (8) l<5‘ 2r+Z0
are approximately ±0.095 decibel. The corre¬
sponding limits of error for the final attenuator are
approximately ±0.185 decibel. The limits of error^ e U-rG)(i + rL)
for the change in attenuation calculated from eq (7) EiL = e zG+zL ' 2 U-rsrL)
are approximately ± 0.242 decibel. It is seen that
the mismatch error for the change in attenuation is
less than the sum of the mismatch errors in measuring Ik _i_
Il= zl 2Z0 (i- rGrL)
each attenuator individually.
It is often possible to determine the phases of
T,, rG S22, and S22, enabling one to calculate the
mismatch error within narrower limits. The error Pu 1 EL* 1L - 4Zf
1-fcfT
in measuring voltage standing-wave ratios and phase
angles produces an error in determining the limits of Figure 1. Load connected directly to generator.
793/8
572772 0 - 61 - 51
|A,|2-|B,|2 | A, |'
F|= E, -I, = -Ini'
I ®21 lA2l B-
p2= e2-(-i2) = MrL
0 0
Figure 2. Network inserted between generator and load.
Pl |l-rGEi 1 —lrL
(15)
Pi |i-rGrL 1 —Ird
794/9
Mismatch Errors in Cascade-Connected
Variable Attenuators*
G. E. SCHAFER and A. Y. RUMFELTf
Summary—The* treatment of mismatch errors is extended to ously calibrated or standard) attenuator. The total in¬
cover variable attenuators cascade-connected in a system which is sertion loss of the series-connected pair of attenuators
not free from reflections. The method of analysis is applicable to any
is adjusted to be the same at two different settings. The
number of cascaded attenuators, but only the analysis of two and
three variable attenuators in cascade is presented. Graphs are given relative attenuation of the test attenuator is taken to be
to aid in estimating the limits of mismatch error. equal in magnitude to the relative attenuation of the
In an example, which is considered representative of rigid rec¬ reference attenuator. This is in error because the stand¬
tangular waveguide systems, the limits of error are: for two attenu¬ ard attenuator is connected to a mismatched attenu¬
ators in cascade, 0.19 db in a 3-db measurement, and 0.17 db in a
ator, and also because the pair of attenuators are in¬
40-db measurement; and for three attenuators in cascade, 0.25 db
in a 40-db measurement, and 0.23 db in a 75-db measurement. serted in a mismatched system.
An expression can be derived for the insertion loss of
I Introduction
795/497
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES October
nected attenuators considered as a single attenuator smaller limits of mismatch error than the addition of the
(called the combination attenuator in the remainder separate limits of mismatch error for attenuation at the
of the paper). These may differ from the sums of the initial and final settings.
individual attenuations by €„• and ecj, respectively, and The total error in cascading variable* * * 6 attenuators in
therefore one may write a mismatched system is then €i+e2, where 61 is the error
Af A tf Arf + 6C/ caused by interactions of reflections from the generator,
= + (5)
combination attenuator, and the load, and 62 is the
At = A ti + Ari + 6 d. (6) error caused by interactions of reflections from the indi¬
vidual attenuators.
Substituting (5) and (6) into (4) yields
The scattering matrix of the combination attenuator
Af — Ai + e2 = €t, (7) is used to evaluate the component ex of the mismatch
error according to (10). This is readily obtained in terms
where e2 has been written for ec; —ef/. The component €2
of the scattering matrices of the individual attenuators
of the mismatch error is the difference between the at¬
through use of the T matrices.7 The T matrix for a com¬
tenuation of the combination attenuator and the sums
bination of n cascade-connected attenuators may be
of the attenuations of the individual attenuators. This
written as
component may be evaluated by an extension of the re¬
sults of Ref. 3 to include the case of variable attenuators k=i
connected in cascade, and may be written as
T=\\Tk, (11)
k=n
1 - WSn"<”Sn'
62 = 20 logxo (8) where the Tk are the matrices for the individual at¬
1 - (i).sr7 tenuators. This T matrix is the inverse of the A matrix
used in Ref. 3. The scattering matrix, S, for the two-arm
where the front superscripts (i) and (f) refer to the initial
junction is related to the T matrix as follows:
and final values, respectively; the primed S’s are ele¬
ments of the scattering matrices6 of the individual at¬ 1 ~Tn 1 1 S’122 —S,22S'xx S22
tenuators, and the number of primes indicate the posi¬ — and T-- , (12)
t22 1 T12 S' 12 Sn 1
tion of the attenuator as counted from the generator.
Eq. (7) may be written as when reciprocity in the form Spg = Sqp is assumed. The
Af — Ai = er — 62 = €1. (9) necessary characteristics of the combination of attenu¬
ators needed to evaluate the error by (10) are Tx and
This form is equivalent to considering the combination S22. These may be obtained by use of (11) and (12), and
attenuator as a single attenuator at two different at¬ the expression
tenuation settings, and 61 can therefore be evaluated as
s122rL
the mismatch error in a relative attenuation measure¬ Ti - Si 1 + (13)
ment with a single attenuator. This component of mis¬ 1 —
match error has been treated by Beatty1 and for this
The results for two attenuators in cascade are
application may be written as
ri_ U + (l-52/,Ti)[(l-522'51/0(l-522'%/'0-^22'(51/025ll,,1-(5x2,,02ri[522,'(l-5i1'%2,)+-S'22,(5x2,/)2] ’ ^
and
cient of the equivalent generator, and 622 is an element 6 The same analysis applies to cascaded-fixed, or a combination of
fixed and variable attenuators in a mismatched system by using the
of the scattering matrix, S, of the combination attenu¬ appropriate forms of (8) and (10).
ator. It should be recalled that this expression gives 7 C. G. Montgomery, R. H. Dicke, and E. M. Purcell, “Principles
of Microwave Circuits,” Mass. Inst. Tech. Rad. Lab. Ser., McGraw-
6 C. G. Montgomery, op. cit., Ch. 14. Hill Book Co., Inc., New York, N. Y., vol. 8, Ch. 5; 1948.
796/448
1959 Schafer and Rumfelt: Mismatch Errors in Cascade-Connected Variable Attenuators
where the primed A’s are the scattering coefficients of mate of the limits of the misrnatch error. For this pur¬
the individual attenuators, where the number of primes pose, certain assumptions to be discussed were made
indicates the position of the attenuator as counted from concerning the scattering coefficients of the attenuators.
the generator end where Tx is the equivalent reflection Since many commercially available attenuators have
coefficient of the detector, and where reciprocity in the similar voltage-standing-wave ratio (VSWR) character¬
form Spg = Sgp has been assumed. istics, it will become evident that these assumptions
Substitution of (14) and (15) into (10) gives the ex¬ caused very little loss in the generality of application.
pression for the component ei of the mismatch error for The following equalities were assumed :
two variable attenuators in cascade. Similarly, substi¬
tution of (16) and (17) gives the corresponding expres¬
1 1 = \Su"\ = | Sn" | — | -S'22r I
sion for the case of three variable attenuators in cascade. | S22" | = Isv'l
Assuming that the phases of the individual coefficients and
of (8) and (10) are not known and that they can have
I rG| = | it | (19)
any possible value, the actual values of the error com¬
ponents cannot be determined, but limits may be Furthermore, it was assumed that the phases of each
found. Such a limit is a conservative figure, and a closer coefficient in (14)—(17) and the SA’s took on values at
estimate of the actual error may be obtained if one de¬ the initial and final settings of the attenuators which
termines the phases of the individual coefficients. The would give the maximum possible mismatch error (lim¬
limits of ei and ti may be added to obtain limits of the its of mismatch error). This assumption yields a con¬
mismatch error, er, since the phases of the coefficients servative estimate of the mismatch error.
in (8) and (10) can take on values so that each com¬ The range of the limits of the mismatch error, er, for
ponent of mismatch error is simultaneously at its maxi¬ two attenuators in cascade and for three attenuators in
mum possible value. cascade are shown in Figs. 1 and 2, respectively. The
sectors indicate how the limits of error vary with the
Graphical Presentation of Results
number of decibels to be measured and with the VSWR
One of the factors taken into consideration, in con¬ of the attenuators for a number of specified equivalent
structing the- graphs, was the ease of obtaining an esti¬ generator and detector mismatches. Eq. (18) implies
Fig. 1—Variation with range of measurement in limits of mis¬ Fig. 2—Variation with range of measurement in limits of mis¬
match error, er, for two variable attenuators in cascade. match error, er, for three variable attenuators in cascade.
797/449
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES October
co
_j
UJ
CD
C_>
UJ
Q
cr
o
cr
cr
UJ
Fig. 3—Limits of component ei of mismatch error for two attenuators Fig. 4—Limits of component ei of mismatch error for two attenuators
with equal magnitudes of 5n and S22 cascaded in a system where with equal magnitudes of 5n and S22 cascaded in a system where
aa — OL— 1.01. oq = <tl — 1.05.
Fig. 5—Limits of component «i of mismatch error for two attenuators Fig. 6—Limits of component ei of mismatch error for two attenuators
with equal magnitudes of 5n and Sn cascaded in a system where with equal magnitudes of 5n and S22 cascaded in a system where
<ro — —1.10. CQ = <JL ~ 1.15.
798/450
1959 Schafer and Rumfelt: Mismatch Errors in Cascade-Connected Variable Attenuators
cn
O
cn
cc.
UJ
x
<_>
<5
Fig. 7—Limits of component «i of mismatch error for three attenua¬ Fig. 8—Limits of component ei of mismatch error for three attenua¬
tors with equal magnitudes of 5n and 522 cascaded in a system tors with equal magnitudes of 5n and 522 cascaded in a system
where cq = <xr = 1.01. where co = <tl = 1.05.
Fig. 9—Limits of component «i of mismatch error for three attenua- Fig. 10—Limits of component ei of mismatch error for three attenua¬
tors with equal magnitudes of 5n and 522 cascaded in a system tors with equal magnitudes of 5n and 522 cascaded in a system
where <ra = aL — 1.10. where c<? — cl = 1.15.
799/451
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES October
that the VSWR corresponding to Sn or S22 of the at¬ the equivalent generator and detector VSWR’s are 1.03
tenuators, which is the abscissa, has been assumed to be and 1.04, respectively; 2) the three attenuators have
the same and equal for the input and output of each input and output VSWR’s of 1.15 or less; and 3) the
attenuator. Each sector is labeled with the appropriate desired attenuation measurements are 3 db, 40 db, and
values of the VSWR associated with the equivalent 75 db, and the reference attenuators are calibrated up
generator and detector reflection coefficients which were to 40 db. The values of the generator and detector
assumed to be equal in (19). The ordinates are the limits VSWR’s are used to determine which of the graphs of ei
of mismatch error. These figures are presented to illus¬ is to be used. One selects the graph with the nearest
trate the range of limits of mismatch error to be ex¬ available value of aG( = crL) which is equal to, or greater
pected in the assumed situations. than, the larger value of the actual load or generator
Figs. 3-6 are a series of graphs of the limits of the VSWR. This will give a conservative estimate of the
component ei of mismatch error for two attenuators in limits of error. In this example, the actual VSWR’s of
cascade with a a ( = ol) of 1.01, 1.05, 1.10. and 1.15, re¬ the generator and the detector are 1.03 and 1.04, re¬
spectively. crg and al are the VSWR’s associated with spectively, and for the case of two attenuators in cas¬
the equivalent generator and detector reflection coeffi¬ cade, Fig. 4 would be selected, since it is constructed
cients, respectively. The VSWR’s used as the abscissas on the assumption that both of these are 1.05. Having
are the input and output VSWR’s of the variable at¬ selected the graph, the largest value of the VSWR at the
tenuators. It is assumed that these VSWR’s are all input or output of the attenuators is the abscissa. Typ¬
equal. The parameter for the family of curves is the ical commercially-available attenuators for rectangular
number of decibels to be measured by this technique. waveguide systems have maximum VSWR of 1.15 over
Figs. 7-10 are an equivalent series of graphs of the their entire frequency and attenuation ranges. For a
limits of the component e\ of the mismatch error for the conservative estimate, then, one could use an abscissa
case of three attenuators in cascade. These are appli¬ of 1.15 unless the actual values of the VSWR are known
cable for measurements of attenuation of 40 dbor more. to be different. The value of ei is different for different
Figs. 11 and 12 are graphs of the limits of the com¬ values of attenuation. For a 3-db measurement it has
ponent 62 of the mismatch error for the cases of two and the value 0.11 db, and for a 40-db measurement it is
three attenuators in cascade. The abscissas are the 0.09 db. Note that for measurements of attenuation of
VSWR’s of the attenuators. The limits of error in this 20 db or greater the component ei of the limits of error
case are independent of the load and generator mis¬ does not change within the resolution of the graphs.
matches. For the case of three attenuators in cascade, One determines e2 from Fig. 11, and it depends only
Fig. 12 is constructed on the assumption that at least 20 on the VSWR’s of the attenuators, and not on the
db attenuation is in the middle attenuator at one of the amount of attenuation or on the generator and load
attenuation settings, either the initial or the final. mismatches. For a VSWR of 1.15, e2 is 0.08 db. Addi¬
tion of these two components, as determined from Figs.
Use of the Graphs
4 and 11, yields the limits of mismatch error, tr, of 0.19
In order to use the graphs of Figs. 3-12 to estimate
the limit of error, one must know the VSWR’s of the
attenuators and of the equivalent generator and de¬
tector, and the approximate number of decibels to be
measured. For an example, consider a case where 1)
Fig. 11—Limits of component £2 of mismatch error for two at¬ Fig. 12—-Limits of component 62 of mismatch error for three
tenuators with equal magnitudes of S11 and .S22. attenuators with equal magnitudes of Sn and S22.
800/452
1959 Schafer and Rumfelt: Mismatch Errors in Cascade-Connected Variable Attenuators
801/453
A Method for Measuring the Directivity
of Directional Couplers*
G. E. SCHAFERf and R. W. BEATTYf
phase of the termination. The inherent errors of the method are ana¬
lyzed and found to be within the limits—0.01 to 0.00 db in a specific
example.
Fig. 1—Arrangement of equipment.
A Introduction
802/419
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES October
Theory
b = Sa, (1)
and
(where A is the scattering matrix, and b and a refer
respectively to outgoing and incoming wave ampli¬
tudes), and solving for the above ratios. The elements
| b31 min ( 5)
of a and b in (1) are displayed in Fig. 2.
The substitution of the appropriate elements into
matrix (1) yields a solution for the amplitude of the
emergent wave from arm 3, The average value of these two responses may be written
— SiaTa 1 — — •SmF.d
| bz |av
(3) Dai — 20 logio
| bz |o
Y + & + r exp (j4>)
M
| y+2
where - 20 logio- 2 (8>
y
2 r2-
1 ~i+R
R = r = | KTl I R,
1 - | ktl\*’
In procedure 2), it is convenient to express the re¬
sponse as
5 A similar technique is described by H. C. Poulter in “A note on
measuring coaxial coupler directivity,” Hewlett-Packard J., vol. 8,
pp. 1-4; May-June, 1957. h = ba ( 9)
6 A. C. MacPherson and D. M. Kerns, “A new technique for the
measurement of microwave standing-wave ratios,” Proc. IRE, vol. y +
44, pp. 1024-1030; August, 1956. 1 + KTLe-**1
803/420
1958 Schafer and Beatty: A Method for Measuring the Directivity of Directional Couplers
where j3l is the electrical length between reference Fig. 4—Limit of error calculated from (17a) (Term 2).
planes of the load and the junction. One adjusts |.Tx,|
and l to give a null response and it is apparent from (9) y = 0, which reduces the ratios in (8) and (11)
that KTLe~2iffl = —1. In deriving the equation for maxi¬
mum response as / is varied, one considers that the locus to | KTl | ; (14)
2 I Sjs]
: 0, S12 = 1, which reduce (15) to -j-r- • (16)
1 + KTLe~2^1 ’ 1 -Sis
when | KTL\ =1, is a straight line parallel to the imagi¬ It is assumed that departures from these conditions
nary axis through the point (1, 0) in the complex plane. will be small and are therefore considered as individual
The maximum response may be written as cases.
Case I
h\T = 1~~, (10)
g +1 S22?^0 and 5i2?^l, but other conditions are satisfied.
where g is the real part of y. Therefore, the apparent Since DAi and DA2 reduce to 20 logio |-K| in Case I,
2 Us lav
Da.2 = 20 logio
I bz |r
M
— 20 logio (g + 1) y
2 (ID
r2 -
i+«
Error Analysis One can separate (17) into two terms. If the coupling
in decibels is 10 db or more, £12 may be eliminated
The sources of error can be evaluated by considering
from the second term with little loss of accuracy:
the true directivity, defined by
804/421
IRE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES October
Case ll
y + 0, but q^her conditions are satisfied. Considerable
manipulation of (12) minus (8) and use of the approxi¬
mation
L £!l
2 K
Tu
1 - 2 -4 r„
K
201ogi + tu ,i
Tu
i +
K
Fig. 5—Limit of error calculated from (18) and (19).
Tu
1 + 2 - 41 r2j |5 By use of (16) and implications of (17), the error may
K
+■ 20 logio (18) be written
Tu cm = — 20 logio | IT) . (20)
K
This is of the same form as the first term of (17a) and
where r2l- is the reflection coefficient of arm 2, with the is included in Fig. 3 as an alternate abscissa. This for¬
inactive generator and detector connected to arms 1 and mula applies to both procedures.
3. For the second procedure the limits of error after similar The following example shows the limits of error for a
manipulation of (12) minus (11) may be expressed as case that might be considered typical. For a well-con¬
structed, high directivity, 20-db coupler (|523| =0.1),
Tu_ the absolute values of the other coefficients are usually
1 - 2 - 4 r2i
K
20 log! ^ £n,i | Sii | < 0.025, | Si21 ~ 0.995,
Tu_
| S221 < 0.025, | Sul < 0.001.
K
| S331 < 0.025,
Tu
1 + 2 - 4 r2; . 2 With the preliminary adjustment of |r2t-| to less
K than 0.01, and use of a short circuit with a reflection
< 20 logio (19)
Tu coefficient of magnitude greater than 0.995, the limits
1 - 2 of error from the above sources are —0.01 to +0.08.
K
By measuring |Si2| and | Fz, |, and applying correc¬
This expression differs from en, 1 only in the denominator. tions for their departures from unity, the limits of error
Fig. 5 shows the limits of error for both procedures 1) are reduced to —0.01 to 0.00 db. One must also con¬
and 2) as a function <r2,-, (the VSWR corresponding to sider the error made in measuring the ratio
r2i). The solid lines indicate the limits of error for | ^3 | av | b<S | av
procedure 1) and the dashed lines the limits of error
for procedure 2). The limiting values for the error as I |0 \bz\T
the directivity becomes infinite are indicated on the Using IF substitution techniques with a below-cutoff
right-hand side of the graph. standard attenuator, the error may be held to +0.05
db with normal precautions. Unless this error is reduced
Case III to +0.01 db or less, the error in measuring directivity
An imperfect short circuit, ] Tx,| + 1, but other condi¬ in this example would be limited by the error in'measur¬
tions are satisfied. ing the above ratios.
805/422
Recent Developments in the Field of Microwave
Power Measurements at tne National
Bureau of Standards*
GLENN F. ENGENf
806/304
1958 Engen: Recent Developments in the Field of Microwave Measurements at the NBS
accuracy which could be achieved by this technique was where the r„, Te, and Tm are the reflection coefficients of
rather severely limited by the state of the impedance the generator, standard power meter, and meter under
measuring art, and this led in turn to the development calibration at the point or plane of connection respec¬
of a number of modifications with the objective of im¬ tively.
proving the over-all accuracy. Beatty,4 for example, pro¬ It will be noted that a determination of the middle
posed a modification based on certain mathematical ap¬ factor requires, in general, a knowledge of the reflection
proximations and restrictions in generality which pro¬ coefficients in both magnitude and phase. A considerable
vided improved accuracy, but the associated operating simplification in the expression may be effected by ad¬
procedures were nonetheless time-consuming and exact¬ justing the generator for an impedance match so that
ing, and the accessory instrumentation was never de¬ vanishes. Under these conditions the expression re¬
veloped or refined to the point where a great deal of duces to the form:
confidence could be placed in the results. Further refine¬
ments or modifications of the technique have been pro¬ (2)
Pm = P*
posed by Weinschel,6 Ginzton,6 Lane,7 and perhaps
others, but it is probably safe to say that none of these
requiring a knowledge of only the magnitudes of T,
proposals have yet come into widespread use.
and Tm.
Another variation of the impedance method has been
It has been shown9 that a generator match may be
'recently developed at the Boulder Laboratories, em¬
simulated by means of a high directivity directional
ploying directional coupler techniques, which provides
coupler with appropriate sidearm monitor or detector
improved accuracy and simplified operational proce¬
and suitable auxiliary apparatus. While the equivalent
dures, but unlike the earlier modifications, involves
generator match was achieved by Engen9 by means of
neither mathematical approximations nor restrictions in
a servo loop which held the sidearm power at a constant
generality, although still limited to barretter type bo¬
level, the same result may also be effectively realized
lometers. A particularly attractive feature of the new
with a stable generator by a manual attenuation ad¬
method is its independence (assuming dissipationless
justment or by simply noting the change in sidearm
connectors) of an impedance discontinuity at the input
power and making an appropriate correction.
connector, which has been a particularly troublesome
Viewed in this manner, the directional coupler and
source of error in coaxial systems. Another advantage
sidearm power detector assembly becomes an output
is its direct applicability to either matched or un¬
monitor for the signal source in addition to providing
matched mounts. A preliminary implementation of the
the generator impedance match. Thus, in practice, one
method in waveguide has given agreement of one half
determines a calibration factor K for the directional
of one per cent and better with results obtained by the
coupler such that
microcalorimetric method. The details of this technique
will be presented in a later paper.
K = (3)
Comparison of Power Meters
The mismatch or impedance measurement problems where Pc is the power indicated by the sidearm detector,
involved in intercomparing power meters have been and the other terms have been previously defined.
treated in detail by Beatty and Macpherson.8 One of Then
the simpler intercomparison techniques is that of alter¬
Pm = KPc{ l-|rm|2), (4)
nate connection to a stable generator, and it was shown
that the power, Pm, dissipated in the meter to be cali¬ which gives the power, Pm, dissipated in the load of re¬
brated is given in terms of the power, Ps, absorbed by flection coefficient, Tmi in terms of the sidearm power,
the standard meter by the expression: Pc, |rm| , and the measured constant K.
Eqs. (3) and (4) have been obtained from (2) which
i - rffr5 2 l - 1 rm|
assumes a matched generator. In general:
l - rsrm l - 1 rs|!
1r gLr s | 2
4 R. W. Beatty and F. Reggia, “An improved method of measur¬ K = ( 5)
ing efficiencies of ultra-high-frequency and microwave bolometer Pc 1 -
mounts,” J. Res. NBS, vol. 54, R.P. 2594, pp. 321-327; June, 1955.
6 B. O. Weinschel, “Phase Engineering Rep. No. 2 on Standard and
Signal Generator Output Power, Voltage, and Attenuation Calibra¬
tion Assembly,” Contract AF33 (600)-25238, E.O. No. C-30047 l - rm| 2
SR-6J3, pp. 32-43; May, 1956. KPc (6)
6 E. L. Ginzton, “Microwave Measurements,” McGraw-Hill 11 - :rffrm |2
Book Co., Inc., New York, N. Y., pp. 185-186; 1957.
7 J. A. Lane, “Measurements of efficiency of bolometer and
thermistor mounts by impedance methods,” Proc. I EE, pt. B, no. 17, The procedure for determining the equivalent Tff
pp. 485-486; September, 1957.
8 R. W. Beatty and A. C. Macpherson, “Mismatch errors in mi¬ 9 G. F. Engen, “Amplitude stabilization of a microwave signal
crowave power measurements,” Proc. IRE, vol. 41, pp. 1112-1119; source,” IRE Trans, on Microwave Theory and Techniques,
September, 1953. vol. MTT-6, pp. 202-206; April, 1958.
807/305
IRE TRANSACTIONS ON INSTRUMENTATION December
achieved by this technique is discussed by Engen9 agreement achieved in the microcalorimetric and im¬
who shows that, in practice, a broad-band match of pedance methods of bolometer mount evaluation, in
|r„| in the range of 0.025 to 0.035 may be readily many cases to 0.2 per cent or so, implies that this error
realized. If the meter under calibration and standard is probably no greater than a few tenths of a per cent.
are also nearly matched, the additional factors in (5) (The difference between these two determinations
and (6) are small and their simpler counterparts may be should in principle give the substitution error.) Even if
used, while for larger generator or load mismatches the larger value of 1-2 per cent is assumed, the cooling
these more complete expressions should be employed if process is still approximately linear, and thus even if a
accurate results are to be achieved. strong dependence of the substitution error upon this
Provided that the sources of error associated with the cooling were postulated, a few tenths of a per cent would
bolometric measurement as discussed earlier, and the still appear to be a conservative estimate of the limit
properties of the coupler are independent of power of this error.
level, the technique provides a convenient and accurate Confirmation of this estimate is rather strongly im¬
method of comparing power meters of different as well plied by the results of an experiment in which a bar¬
as equal power levels. For example, if it were required retter mount was connected to one arm and a thermistor
to compare a 100-mw meter against a 10-mw standard, mount to the other arm of a 3-db coupler. The ratio of
a nominal 10-db coupler would be chosen and the meas¬ power indicated by the barretter to that of the thermis¬
urement required for (3) carried out at nominal values tor was found to be constant, within the limit (0.1 per
of 1 and 10 mw for Pc and Ps, respectively. The power cent) of experimental error, over a 20-db range. This
input to the coupler would then be increased by 10 db, result may be interpreted as implying that there is no
yielding nominal, values of 10 and 100 mw for Pc and change in substitution error within this limit and over
Pm in (4). If calibration of a 1-mw meter were required, this range, or if there is a change, it is at least the same
the procedure is similar except that the connections to for both elements. Because of the marked difference in
the main and sidearms are reversed. the nature of the barretter and thermistor elements,
It should be noted that this technique of extending the first implication is strongly favored. In another ex¬
the operating range is different from, and potentially periment the effective efficiency of a thermistor mount
more accurate than the usual method where the param¬ was determined by the microcalorimetric method using
eters of the individual components are measured. The first unmodulated CW microwave energy, and then
present method treats the coupler and power detector pulse energy in which the peak power exceeded the
as a unit assembly and measures a single parameter for average by approximately 40 db. Again the results were
the entire system. (A second measurement is required equivalent within the limits of experimental error, lend¬
if the equivalent generator impedance is desired.) The ing further support to the above conclusion.
extension of range is thus based upon the linearity of the A directional coupler with appropriate sidearm bolo¬
system and the ability to make accurate bolometric metric detector thus provides a useful technique for
type measurements over a wide dynamic range. The realizing a calibrated and matched signal source, and
bolometric measurement may be easily carried out over permits an extension of the range (in both directions)
a 10-db range and can be extended to 20 db with little of a given power standard by 20 db or so. For example,
loss in accuracy if suitable precautions are taken. Re¬ accurate signals in the range 1 microwatt-1 watt may
garding the assumed linearity, it may be noted that the be derived from a calibrated bolometer mount which
parameters of interest in the coupler and associated operates in the 100-microwatt-10-milliwatt range. In
bolometer mount detector depend primarily upon the order to realize these results, however, care must be
geometry and surface impedance of the waveguide used in the associated bolometric measurements, and
structure which are substantially, if not entirely, inde¬ the techniques described by Engen1 are suggested.
pendent of power level.
The other potential source of nonlinearity is in the
Acknowledgment
dc-RF substitution error, but here it will be noted that
a conservative upper limit of 1-2 per cent has been es¬ The author extends his thanks to the members of the
tablished analytically10 for this error, and the close Microwave Power Standards Project, particularly
10 H. J. Carlin and M. Sucher, “Accuracy of bolometric power
Morris Harvey and William Case, for their contribu¬
measurements,” Proc. IRE, vol.40, pp. 1042-1048; September, 1952. tions to the developments discussed in this paper.
808/306
A Dry, Static Calorimeter for RF
Power Measurement*
P. A. HUDSON/ AND C. M. ALLREDf
Introduction Theory
HE measurement of RF power at the National The calorimetric method of measuring power involves
Bureau of Standards at frequencies below about the determination of the rate of conversion of electrical
500 me with accuracies of + 1 per cent or better energy into heat in a resistive load. The load may con¬
has been chiefly limited, in the past, to the dynamic sist of an extended lossy medium or a finite resistive ele¬
range which can be measured with bolometer or ther¬ ment. The heat generated may be measured in various
mistor bridges. At the present time such bridges have ways, such as in a liquid flow calorimeter, where the
an upper limit of about 100 mw. A thermistor bridge temperature rise of a coolant fluid is observed. The heat
with a range of 100 jaw to 100 mw and a dc accuracy of may be absorbed in the fluid by means of a heat ex¬
+ (0.05 per cent + 2 /aw) has recently been constructed changer or the fluid may absorb the energy directly if it
at the National Bureau of Standards Boulder Labora¬ is sufficiently lossy. The phenomena of change of state
tories and is presently being used as a standard.1 may also, be utilized. A third method is the use of a
To extend the power range upward, at least three pos¬ system in which the dissipating medium or element is
sibilities were considered at the National Bureau of fixed in position and in addition there is no change of
Standards. First, the range of the 100-mw thermistor state. That is, the system is static and the temperature
bridge could be extended by the use of suitable attenua¬ rise of some part of the system is observed. Calorimeters
tor pads or with calibrated directional couplers used as operating with the latter properties are termed “static
power dividers. The second alternative was to develop calorimeters.” They are usually simply instrumented,
a bolometer bridge with a higher power range than the may have high accuracy, bu't are usually slow in re¬
present model. Thirdly, a power standard operating on sponse.
a principle different from that of the thermistor bridge In the static calorimeter where high accuracy is de¬
could be developed. Actually, all three of these alterna¬ sired, the temperature rise is usually measured at condi¬
tives have been pursued simultaneously. For example, tions of thermal steady state. Since a measurement of
high-power, high-quality attenuator pads have been temperature rise only is desired, the steady-state tem¬
constructed in the laboratory and their use with the perature Ti may be measured with respect to some fixed
thermistor bridge has been partially investigated. Also, arbitrarily chosen reference temperature T0.
a bolometer bridge capable of measuring RF power up In this method is aTunction of power input as well
to 8 watts has recently been built and a report covering as the degree of thermal isolation of the body from its
its design and performance is planned for the near surroundings. Ideally, would be independent of fre¬
future. quency, and the calorimeter could conveniently be cali¬
In pursuing the third alternative, that of developing brated with dc power and subsequently used to measure
a standard independent of the bolometer bridge method, power at any frequency. Thus the calorimeter is ac¬
a dry, static calorimeter-type wattmeter was chosen as tually a transfer standard measuring RF power in terms
being the most desirable from the standpoint of accu¬ of accurately known dc power. The validity of the as¬
racy, high sensitivity, and wide dynamic range.2 Because sumption that the equilibrium temperature (Ti) is inde¬
intercomparison of independent methods of measure¬ pendent of frequency will be discussed later.
ment is highly desirable in standards work, it was de¬ An insight into the physical phenomena involved as
cided that the range of the calorimeter should extend well as the prediction of the approximate system be¬
well below 100 mw and up to about 10 watts to permit havior may be gained by representing the calorimeter
its comparison with both the thermistor bridge and a by its electrical analog shown in Fig. 1. It should be
liquid flow type calorimetric power meter currently noted here that the accuracy of the calorimeter is not
under development. The power range of the latter ex¬ dependent on this analogy which is, of course, only
tends from 10 to 350 watts. approximate.
When power is fed into the input of the calorimeter, at
* Manuscript received by the PGI, August 23, 1958. a fixed level, heat is generated in the load resistor at the
t National Bureau of Standards, Boulder, Colo.
1 “Improved thermistor bridge for RF power measurements,” same rate as the absorption of the electrical energy. The
NBS Tech. News Bull., vol. 40, pp. 134-135; September, 1956. heat is first manifested in a temperature rise of the load
2 M. Wind, ed., “Handbook of Electronic Measurements,” vol.
I, pp. 3-35 to 3-41; 1956. resistor which has a certain thermal capacitance C. Sub-
where Q = charge and / = time, in seconds; also up to at least 500 me, skin effect is negligible, and at
temperatures up to 100°C the resistance changes by
only 0.2 per cent. The change in reflected power due to
such a small change in resistance is negligible in com¬
parison to other uncertainties in the system.
dV The load resistor and its coaxial mount are housed in¬
I = ic + ii + *2 + H = C-b G\{Vi — Fo) side a thick-walled (f-inch) cylindrical aluminum con¬
dt
tainer 4 inches in diameter by 6 inches in length. The
+ Gt(Yi- Fo)1-» + G3(F14- Fo4). (3) entire assembly is placed inside a thermally-insulated
Now if Gi and G3 can be neglected, then chamber containing a refrigeration coil over which air
is continually recirculated. The temperature inside the
dV chamber is 0°C and is kept constant to within + 0.002°C
I — C —-b VGi. (4)
dt by means of a sensitive electronic control circuit. Thus,
the thick-walled aluminum container is kept at the
Eq. (4) is a linear differential equation with constant
temperature of the chamber which serves as the refer¬
coefficients whose solution is well known. It is satisfied
ence temperature T0. In the absence of power input the
by
equilibrium temperature of the entire calorimeter will,
of course, be T0 also.
8 M. W. Zemansky, “Heat and Thermodynamics,” McGraw-Hill
Book Co., Inc., New York, N. Y., pp. 94-95; 1951. A measure of the load resistor temperature rise, which
810/293
IRE TRANSACTIONS ON INSTRUMENTATION December
811/294
1958 Hudson and Allred: A Dry, Static Calorimeter for RF Power Measurement
812/295
IRE TRANSACTIONS ON INSTRUMENTATION Decemi
mver
813/296
Journal of Research of the National Bureau of Standards Vol. 59, No. 2, August 1957 Research Paper 2776
Until recently, the most accurate microwave power measurements of the bolometric
type have required the use of a manual d-c bridge. A self-balancing d-c bridge has been
developed that preserves the inherent accuracy of the manual bridge, extends the dynamic
range of operation, and greatly simplifies the operating procedure. A general descriptioi’i
of the equipment and operating techniques is given, followed by a comprehensive survey of
the sources of error accompanying the method and the accuracy achieved.
Pr,=(1)
3. Self-Balancing Bridge
where ix and i2 fire die total currents required to
balance the bridge without and with r-f power The self-balancing d-c bolometer bridge employs a
present, respectively. A precision potentiometer- d-c amplifier in a feedback loop to balance the
standard resistor combination is used to measure bolometer bridge. The amplifier is connected in
ix and i2. If it is assumed that this measurement such a manner that the bridge unbalance signal is
amplified to provide the bridge current (fig. 2). The
1 R. W. Beatty and Frank Reggia, An improved method of measuring efficien¬ use of this self-balancing circuit eliminates many of
cies of ultra-high-frequency and microwave bolometer mounts, J. Research
NBS 54, 321 (1955) RP2594. the time-consuming features of the technique, but
2 A. C. Macpherson and D. M. Kerns, A microwave microcalorimeter, Rev.
Sci. Instr. 26, 27 (1955). the power-measurement problem is essentially the
3 H. .1. Carlin and Max Sucher, Accuracy of bolometric power measurements, same. This problem, at low power levels, is basically
Proc. Inst. Radio Engrs. 40, 1042 (Sept. 1952).
1 D. M. Kerns, Determination of efficiency of microwave bolometer mounts one of trying to accurately measure a small change
from impedance data, J. Research NBS 42, 579 (1949) RP1995.
5 More correctly, “the retracted d-c power”. in a comparatively large d-c power.
814/101
ometer and standard resistor are employed in place
of Ax and A2 to achieve high accuracy.
Alternatively, a voltmeter, V, connected as shown,
will read (r0/4) (ii+i2), and the power is given by the
product of V and A2. These operations may also be
combined in a single instrument of the dynamom¬
eter type to achieve a direct reading of the power.
The degree of success obtained with this technique
is, of course, contingent upon the degree of stability
achieved in the constant-current generator. In this
case the change in generator output for a complete
cutoff of bridge current is less than 1 part in 1CF,
resulting in a negligible error from this source.
Figure 2. Basic circuit of self-balancing bridge.
4. Performance
815/102
Figure 5. Prototype model of self-balancing bridge, front and rear views.
816/103
obtain:
(current amplification)
s
0<A/Pi<2X10-5
)
o<A<io-7 ^(amplifier zero stability).
I
M<io- J
balanced_ .02 percent. The failure of the expression for power to have a
Measurement of A1 + A2_ • 03 percent. first-order dependence on A' or Y is explained as
Measurement of f&i—it2_ ■ 03 percent. follows: A prescribed, small value for A" or Y will
Bolometer lead resistance_ .00 percent.
Amplifier noise and instability cause a first-order variation in bolometer resistance
(experimentally observed)-_ 0.1 juw. r, as indicated in eq (7), but also produces an unequal
Total_ . 09 percent+ 0.1 jiw. division of current in the two sides of the bridge.
These two sources of error cancel to a first approxi¬
6. Appendix mation. The only general requirement on the upper
bridge arms is that they be equal, and if they have a
From figure 6 the power may be expressed in nominal value of r0, it can be shown that they may
the form differ from each other by as much as 2 percent
and produce an error in the power measurement
of only 0.01 percent.
PTf—^(4*1 t52)+/(di,T >Z,ig,e,ri) (5) At a constant ambient temperature and assuming
equivalence of d-c and r-f heating,6 the power, P,
dissipated in the bolometer will lie a function of its
where X, Y, and Z are the deviations of each of the
resistance, r:
bridge arms from, the value r0, ig and e are the
detector current and thermal (or other) emf, re¬ P=j (r).
spectively, and rx the resistance of the bolometer
leads. Because the bolometer is nearly a square-law
Assuming a knowledge of the bolometer resistance detector, one may in the neighborhood of an ar¬
law, an exact expression for the second term could bitrarily selected operating resistance r0, write
be derived, but the1 labor required suggests the fol¬
lowing approximate treatment. By hypothesis, the
variables A" . . . 77 and f(X . . . rt) are small, and P=Po~\~‘^7 (r—A)>
f{X . . . Vi) may, to a good approximation, be written
as the sum of the first-order corrections for each of
these quantities, where the correction for each, of where P0 is the power required to bring the bolometer
the variables is determined with the other variables to the resistance r0 at the given ambient temperature,
equal to zero. P is the sum of the d-c and r-f power, P=Pdc-\- PTf,
Application of this technique to the circuit in and dP/dr is the reciprocal! of the “ohms-per-watt
figure 6 yields the following first-order expressions
for the d-c power dissipated in the bolometer, and
6 The r-f-d-c substitution error is assumed to be negligible in this discussion.
the bolometer resistance For a treatment of this problem, see footnotes 2 and 3.
817/104
bolometer coefficient, i. e., (dP/'dr) = (1/7). Then is subject to change with the application of r-f power.
Substituting eq (6) and (7) in eq (8) gives the desired
result
P<ic+Pr/ = Po + “ if — r0).
Tl 2e
Pr}e*~ (ih—ih) [l + Z-
P0 may be determined by measurements in the Tq r oC&iTC:
hi) C 7^61^62/
818/105
Journal of Research of the National Bureau of Standards—C. Engineering and Instrumentation
Vol. 63C, No. 1, July-September 1959
The microcalorimetric method for evaluation of the efficiency and substitution error of
a bolometer mount proposed by Macpherson and Kerns has been the object of further
study and refinement at the Boulder Laboratories of the National Bureau of Standards,
and an improved instrument based on this technique has been recently placed in operation.
The new microcalorimeter design features are: (1) Greatly improved ambient tempera¬
ture control, permitting higher sensitivity and resolution; (2) improved d-c instrumentation;
(3) improved mechanical construction giving better repeatability; (4) relocation of the
thermopile so that it no longer is attached directly to the bolometer mount, thus providing
flexibility in the choice of termination; and (5) a more comprehensive treatment of the
calorimetric substitution or equivalence error. These features permit the determination
of the effective efficiency of a bolometer mount to an accuracy of better than 0.2 percent.
819/77
The thermopile consists of 34 eonstantan-
chrbmel-P junctions with an approximate senstivity j
of 2,000 mv/° C. The instrument is designed to j
yield a nominal output at equilibrium of 100 /xv
corresponding to a temperature rise of 0.05° C. |
Since this potential may be measured by the asso- |
dated instrumentation to an accuracy of a few parts !
in l(ff, a temperature background which is stable to i
a few millionths of a degree is indicated for the '
application.
In order to achieve this background, the twin
Joule arrangement is employed as shown in figures
1 and 2, along with a temperature-regulated water
bath. The advantage of this method stems from the j
fact that if symmetry exists with respect to a plane
normal to the plane of the illustration as shown, and
if no thermal gradients exist, within the thermal
shields, then the actual temperature of these shields
may fluctuate without producing a temperature
difference between the bolometer mounts. Thus,
by using a triple thermal shield (the innermost one
of which is a heavy copper casting), a symmetrical
waveguide input system, and suitable thermal
shunts, a degree of symmetry is achieved such that
the dependence of the temperature difference be¬
tween the bolometer mounts upon the thermal head
existing between the inner thermal shield and the
external water bath, is only a factor of approximately Figure 2. Microcalorimeter and thermal shields.
820/78
and keeping the length short. The design should
also provide a high value for the efficiency as this
will also reduce the error from calorimetric non-
THERMO-JUNCTIONS (34) equivalence. Some of these problems will be dis¬
cussed in greater detail in later sections, but it will
be recognized from the present discussion that the
design must effect a compromise among a variety of
conflicting requirements.
3. Operational Procedure
The operational procedure associated with the
instrument is comparatively simple and straight¬
forward. Using the terminology introduced by
Macpherson and Kerns [3] an effective efficiency,
pe, may be defined,
Substituted or retracted bias power
as measured by the bolometric technique , .
c Total microwave power dissipated within
Figure 3. Detail of thermopile construction. bolometer mount
821/79
where k and g are constants, and P2 and Prf are
the d-e power dissipated in the bolometer element
and the total microwave power dissipated in the
bolometer mount, respectively. Equation (3) ex¬
presses the fact that a given amount of microwave
power will, in general, produce a different thermopile
response than an equal amount of d-c power through
the factor g.
The constant k may be determined in the manner
already discussed so that k=exIPldc. Substituting
this result and solving eq (3) for Prf gives,
and
20 40 60 80 100 120
TIME , MINUTES pxd -ph 1 —(f2/ii)2
Ve=g (4)
Figure 6.—Response curve for determination of effective e2 zge2/ex— (?2/b)2
efficiency. Pid-P*.dc
Let ix and i2 represent the bolometer bridge cur¬
The more complete analysis thus gives a factor g
rents without and with microwave energy present
by which the right-hand side of eq (2) is to be
in the bolometer and, respectively, ex and e2 the cor¬
multiplied in determining r]e. The error in deter¬
responding thermopile responses. Then for . the
mining rjc thus consists of the uncertainty in the
equal arm bridge in common use, the bolometric
value of g, and the experimental error in deter¬
power is given by the usual formula,
mining the value of this second factor, ——ll]} .
e2/ei — (i2/h)2
(H-il) These will be referred to as the calorimetric equiva¬
lence and instrumentation errors, respectively.
where r0 is the resistance of the bolometer clement The instrumentation error may be evaluated with
and bridge arms. little difficulty. If the second factor is differentiated
The total power input to the bolometer mount with respect to ix and i2, it is found that for typical
(microwave and d-c) during the microwave “on” values of the ratio e2\ex, i.e., l^Co/e^l.OS, the error
interval ideally will be proportional to the thermopile in r]e is only 5 percent or less of the error in q or i2. In
response, e2, the proportionality factor being r0iy4ex. practice, ix and i2 are measured by a precision poten¬
The microwave power is the difference between this tiometer-standard resistor combination to an accuracy
total input and the d-c component, thus of about 0.03 percent; thus, the total error due to the
uncertainty in q or i2 is only of the order of 0.01
p _'o percent and may be neglected.
lrf~4 The error due to the uncertainty in e2/ex may be
and found in a similar manner, and if the ratio of i2 to q
is % or less, the error in ye is in tlie range 1.0 to 1.3
times the error in e2/ex.
e2/ex— (i2/ii)2 In order for eq (3) to be valid, the thermopile re¬
which gives the effective efficiency (rq) in terms of sponse must be a linear function of the d-c and
the thermopile response and bolometer-bridge cur¬ microwave powers. The error in the ratio c2/c, may
rents. The prime is used to indicate that the equa¬ conveniently be defined to include, in addition to
tion is based on the assumptions that the thermopile the instrumentation error in measuring the thermopile
response is linear and independent of the distribution output, the error due to nonideal performance of the
of the power dissipation within the mount. The calorimet er caused by such t liings as nonlinear t hermo¬
validity of these assumptions will be examined below. pile response and cooling, thermal-background drift,
etc., which might affect the validity of eq (3). The
4. Error Analysis error in measuring the ratio e2/ex was determined in
the following manner. The measurement procedure
Equation (2) for rj'c has been obtained under outlined in the foregoing section was followed except
certain simplifying assumptions; a more general that instead of applying microwave power at the
approach to the problem will now be presented. end of the first 45-min interval, the d-c power
Within the limits of experimental error (to be dis¬ applied to. the bolometer was increased by a nominal
cussed below), the thermopile response and cooling 10 percent. In this way the actual operating con¬
mechanisms are both linear, and the thermopile ditions were simulated, and the measured value of
response e2 is of the form: this second d-c power was compared with that
predicted on the basis of a linear response from the
e2=k(P 2dc-\-gPr/) (3) observation of ex, e2 and the first value of d-c power.
822/80
In ten measurements of this type, the maximum for a convection shield or barrier could be eliminated,
error observed was 0.04 percent, with an average but these results clearly demonstrated the need for
error of 0.013 percent. A figure of 0.1 percent, cor¬ such a shield. In order to provide a convection
responding to an error in e2/ej of 0.07 percent, has barrier, in which the microwave dissipation would
been arbitrarily assigned as a conservative limit to be negligible, a series of eight 0.002 in. Teflon mem¬
the error from this source, in which the probable branes spaced at 0.05 in. intervals and transverse to
error is perhaps 0.02 percent or less. the waveguide axis was arranged at the waveguide
With regard to the equivalence error, g is a number input. This effectively eliminated the convection
which expresses the relative effectiveness of micro- problem and gave repeated determinations of k in a
wave power as compared with d-c power in produc¬ refined bolometer mount which agreed to 0.05 per¬
ing a thermopile response, that is, for a given thermo¬ cent, which is about the limit of experimental error
pile output, g—Pdc/PTf. Ideally, g— 1 but in for this particular measurement. Since the power
practice differs from this value due to the following distributions in the barretter and thermistor are
phenomena: markedly different, it may be safely assumed that
1. The d-c and microwave powers take different the small difference in distribution of the d-c and
distributions within the bolometer element yielding microwave powers which occurs in practice when the
different thermopile responses. same element is employed, will have a negligible
2. Some of the microwave energy will be dissipated effect.
in places other than the bolometer element and the With regard to the degree of equivalence realized
thermopile response will again differ from that pro¬ in the thermopile output for power dissipated in the
duced by an equal amount of d-c power in the bolometer mount walls as compared with an equal
bolometer. amount in the bolometer element, the problem,
3. The thermopile will respond to the dissipation once again, is primarily that of attempting to pre¬
of microwave energy in the waveguide lead-in vent the escape of heat from the mount via air
which does not comprise part of the bolometer convection. The effectiveness of the convection
mount. barrier described in the previous paragraph was
It is just these first and second phenomena, of demonstrated by temporarily inserting two other
course, which are responsible for the dc-rf substitution heat sources within the mount such that they were
error and mount inefficiency. in good thermal contact with the side walls. The
With reference to the first of these, the tempera¬ first of these was located just inside the mount before
ture-sensing device in the calorimeter is relatively the teflon barrier, the second on the other side of
far removed from the main heat source and it might the barrier, close to the bolometer element as
be estimated conservatively that the error in the shown in figure 7. The thermopile response for
calorimetric method due to this phenomena would be equal amounts of power in the two positions was
at least an order of magnitude or more below the then compared with that for the bolometer element,
bolometric substitution error which according to the and it was found that the heat sources at positions
information available [7], does not exceed a few 1 and 2 were approximately 1.5 and 0.5 percent
percent.2 more effective respectively. A conservative figure
In order to check the validity of this assumption for the increased effectiveness of the power dissi¬
experimentally, the d-c calibration factor, k, was pated in the walls is thus 1 percent with an error
determined for several different bolometer elements of ± 1 percent. It should be noted that this figure
(barretter and thermistor). The initial results of applies only to that fraction of the power which
this experiment gave vTallies which differed from one is so dissipated. In practice a lower limit of 97
another by as much as 2 percent with a repeatability percent may be taken for the efficiency of the
of 0.1 percent. This anomaly at first appeared to mounts designed for use in this calorimeter, which
challenge the validity of the above assumption, but leaves only 3 percent of the energy to be dissipated
further investigation showed that the insertion of a other than in the bolometer element.
0.05 in. copper plate at the bolometer-mount input
flange (thus blocking the waveguide input), would
eliminate this difference. This was interpreted to
mean that air convection was playing a significant
role in cooling the bolometer capsule (not to be con¬
fused with the cooling of the bolometer element
within the capsule), and the variations in the value
for k were ascribed to the variable role, occasioned
by the differences in the capsules associated with the
bolometers, played by air convection in conveying
this heat from the capsule to the external heat sink
via the interior of the waveguide.
By inverting the bolometer mounts such that the
closed end was up, it had been hoped that the need
2 Recent advances in the impedance method of measuring bolometer mount
efficiency have demonstrated that the substitution error at frequencies near 10
kMc is probably no greater than a few tenths of a percent. Figure 7. Bolometer mount with auxiliary heat sources.
823/81
Finally, as noted, the thermopile will respond to a limited range, serve the same function as a move-
the dissipation of microwave energy in the thermal able back plate, while the other three stubs serve
isolation section which is not part of the bolometer the usual purpose.
mount. If an electrical conductivity of 50 percent . In the five mounts which have been built to this
of the d-c value is assumed for this % in. length of general pattern, effective efficiencies in the vicinity
copper waveguide, the theoretical value of power of 98 to 99 percent have been realized. The results
dissipation will be approximately 0.05 percent, of of additional measurements to determine the place of
which the thermopile will probably measure some¬ this residual dissipation seem to imply that most
what more than half. A value of 0.04 percent ±0.04 of it occurs in the plastic capsule surrounding the
percent thus represents a conservative estimate element and/or in the microwave bypass mechanism,
for this contribution. at the d-c lead-in to the bolometer. While addi¬
The results of this section are summarized in tional research will be required to completely
table 1. resolve these questions, it appears likely, on the
basis of present information, that effective effi-
Table 1
ciences of 99.5 percent and perhaps higher may be
Total Contribution ] ; realized by proper design and fabrication techniques.
Place of heat dissipation power to thermopile! Error Finally, table 2 shows the results of intercom¬
response a j
parisons with commercially available mounts.
er
.0 % ,0
Bolometer element__- . .. 97.00 ! 0. 00
Waveguide walls_
Isolation section..__ _ . _
% 3.03 !
0.04 |
±0. 03
±0. 04
Table 2
5. Bolometer Mount Design and Experi¬ The author acknowledges the assistance of Morris
E. Harvey, to whose care in making the experimental
mental Results
observations a large amount of credit for the success
As may be inferred from the foregoing description, of this project is due, and extends his thanks to
the design of the bolometer mounts employed in R. W. Beatty and D. M. Kerns for their helpful
the microcalorimeter has undergone a series of tran¬ suggestions in the preparation of the manuscript.
sitions as required to cope with the various engineer¬
ing problems encountered in the evaluation of the 6. References
device. While no claim is made as to the extent of
their applicability, certain features of the construc¬ [1] G. F. Engen, A self-balancing d-c bridge for accurate
tion are believed to be of enough interest to merit bolometric power measurements, J. Research XBS 59,
further description. 101 (1957) RP2776.
[2] R. W. Beatty and A. C. Macpherson, Mismatch errors in
The general features of the construction are microwave power measurements, Proc. IRE 41, 1112
shown in figures 4 and 7. The mount is machined (1953).
in two halves from a solid piece of tellurium copper, [3] A. C. Macpherson and D. M. Kerns, A microwave micro¬
and the two halves harcl-soldered together along calorimeter, Rev. Sci. Instr. 26, 27 (1955).
[4] M. Sucher and H. J. Carlin, Broad-band calorimeters for
the center of the wider side. The brass flange is the measurement of low and medium level microwave
then attached by soft solder. Except for these two power. I. Analysis and design, IRE Trans, on Microw.
applications, the use of solder has been completely Theory Tech. 6, 188 (1958).
avoided in the construction. The incentive for this [5] A. V. James and L. O. Sweet, Broad-band calorimeters
for the measurement of low and medium level micro-
method of fabrication was provided by the observa¬ wave power. II. Construction and performance, IRE
tion of efficiencies of the order of 85 percent for Trans, on Microw. Theory Tech. 6, 195 (1958).
mounts in which the back plate was fastened with [6] W. M. Sharpless, A calorimeter for power measurements
soft solder, while the use of hard solder increased at millimeter wavelengths, IRE Trans, on Microw.
Theory Tech. 2, 45 (1954).
these values to 95—98 percent. [7] H. J. Carlin and M. Sucher, Accuracy of bolometric power
The design employs five tuning stubs as shown, measurements, Proc. IRE 40, 1042 (1952).
three in front of the bolometer and two at the side.
These side stubs provide additional susceptance at
the plane of the bolometer element, and thus, over Boulder, Colo. (Paper 63C1-9)
824/82
2,883,620
United States Patent Office Patented Apr. 21, 1959
1 2
improved R-F power measuring bridge in which a single
thermo-sensitive unit may be used to measure R-F powers
2,883,620
over a range of approximately 30 decibels.
HIGH FREQUENCY POWER MEASURING A better understanding of the invention can be had
BRIDGE CIRCUIT 5 with reference to the single figure of the drawing show¬
ing the improved R-F power measuring bridge circuit of
Myron C. Selby, Charles M. Allred, Paul A. Hudson, and
the present invention.
Ira S. Berry, all of Boulder, Colo., assignors to the
United States of, America as represented by the Sec¬ Referring to the drawing, a conventional Wheatstone
retary of Commerce bridge indicated generally at 1 includes equal resistance
10 arms 2, 3, and 4. The fourth arm of bridge 1 contains
Application September 6, 1957, Serial No. 682,560 a thermo-sensitive element 6 which may be in the form
3 Claims. (Cl. 324—106) of a thermistor or barretter mount. An R-F generator
7 supplies high-frequency energy to element 6 through a
coupling capacitor 8. Bridge balance may be detected
This invention relates to R-F or microwave power 15 by a galvanometer 9 connected across two arms of the
measurements and more particularly is concerned with bridge.
a bridge including a temperature-sensitive thermistor or Connected in parallel with bridge 1 are a pair of series
bolometer mount for measuring low-level R-F power. resistors 11 and 12. Adjustment of the variable resis¬
Resistance-type bridges have been used almost exclu¬ tor 11 controls the amount of current shunted around
sively for the measurement of low-level R-F powers for 20 bridge 1. A switch 13 provides means for disconnecting
the past several years. As compared with calorimetry resistors 11 and 12 from the remainder of the circuit.
equipment, bridges have the advantage of high sensitivity, The magnitude of the shunt current flowing through resis¬
fast response and higher accuracy. Prior art measuring tors 11 and 12 may be measured by voltmeter 14 con¬
circuits have almost invariably been of the Wheatstone nected across measuring resistor 12.
bridge type in one arm of which is included a temperature- 25 Bias current for bridge 1 is derived from a constant
sensitive resistor such as a thermistor or barretter. current D.C. supply 16. Source 16 maintains a steady
Thermistors have negative temperature coefficients while flow of constant magnitude current at all times. The
barretters have generally positive ones. The prior art magnitude of this current is measured by means of a volt¬
bridges include a source of low frequency- or D.C. bias meter 17 connected across a second measuring resistor 8
current having means for varying the magnitude of the 30 in series with the D.C. supply source.
current. Since this bias source must be extremely stable, In operation the bridge is first balanced by varying the
batteries operated under conditions of constant current current T from the constant current source 16. When
drain have usually been used. To meet the condition bridge 1 is balanced, the current from source 16 is set
of constant current drain, constant impedance attenua¬ and remains constant throughout the remainder of the
tors are generally placed in series with the bridge to 35 measuring period. During this original balancing opera¬
vary the bias power. The computation of R-F power is tion switch 13 in the shunt circuit is open so that the
then obtained from the equation Prf—V4(l12—I22)R for shunt current Al! is zero. The R-F power to be measured
the case of an equal arm bridge, where R is the bridge from R-F generator 7 is then fed into the temperature-
resistance and 1^ and I2 are the total bridge currents be¬ sensitive element 6 with switch 13 in the closed posi¬
fore and after the application of R-F power, respectively. 40 tion. Shunt resistor 11 is then adjusted until the bridge
The validity of this equation is based on the equivalent is rebalanced. The two currents Ii and Alx may be meas¬
heating effect of D.C. power and R-F power when both ured on voltmeters 17 and 14, respectively, and the R-F
are dissipated in a purely resistive load. power may then be simply computed.
Batteries as used in the prior art bridges need fre¬ By using a constant current bias source 16 the shunt
quent recharging and/or replacement and constant im¬ current Alj. is exactly equal to the differential current
45
pedance attenuators are both bulky and expensive. The (/j—/2) in the equation as given above for computing
chief disadvantage, however, is that if the R-F power is R-F power. This shunt current may be measured di¬
small then I1—I2 may well be of the same order of mag¬ rectly to the desired degree of accuracy regardless of the
nitude as the accuracy with which It and I2 are meas¬ magnitude of the current In The power equation
ured. Such a condition can lead to large errors and the 60 PTf=V4(h2-h2)R
only alternative is to use special high sensitivity, low
power capacity barretters or thermistors for low power is used in the factored form Pr)=V4 (2lx—AI^A^R where
level; thus several elements are necessary to cover the A/1=(/1—12). The shunt current Alx can be substituted
range from microwatts to milliwatts. directly into the equation. In this way a single high-power
The bridge circuit of the present invention avoids the thermistor, for example, can be used to cover the range
55
abovementioned difficulties by employing a variable re¬ from about 10 microwatts to around 100 milliwatts of
sistance in parallel across the bridge. A conventional radio frequency power.
constant current D.C. supply source provides a single¬ Constant current supply source 16 may be any of the
valued total current at all times. The bridge is rebalanced well-known constant current D.C. generators. A special
by shunting a portion of the bridge current through the electronically regulated power supply has been constructed
60
parallel variable resistor. The change in current neces¬ having a regulation of one part in 105 for load variations
sary to rebalance the bridge can then be directly measured from 0 to 100 ohms with a stability of plus or minus one
in the shunt circuit. part in 105 per hour. With this source and using a
It is therefore one object of this invention to provide thermistor for temperature-sensitive element 6, R-F
an improved R-F power measuring bridge. lower levels from 100 milliwatts down to 100 microwatts
65
Another object of this invention is to provide an R-F have been measured with respective accuracies of 0.05
power measuring bridge in which the change in bridge to 5%.
current may be measured directly. A low-cost bridge with reduced accuracy may be con¬
Still another object of this invention is to provide a structed by substituting a resistor having a value of about
simple, inexpensive R-F power measuring bridge having 100 times that of the bridge resistance for the constant
70
improved accuracy and greater range. current control circuit. In such a modification a large
An additional object of this invention is to provide an resistor is placed in series with the bridge and fed from
572772 0 - 61 - 53 825/1
2,883,620
3 4
a conventional voltage regulated power supply. In this ing an electrical balancing bridge having a variable im¬
case changes in the shunt resistance from zero to infinity pedance in one arm sensitive to high-frequency energy,
will result in a maximum change of total bridge current a constant direct-current source connected to said bridge
Ii of only 1%. for supplying a bias current of fixed amplitude, means
As can be seen, the present invention provides a novel 5 for applying the high-frequency energy the power of
inexpensive bridge circuit for measuring low-level, high- which is to be measured to said sensitive impedance to
frequency power limited in frequency range only by the produce unbalance in said bridge, means for varying
frequency limitations of a thermistor or bolometer mount. the amount of said fixed-amplitude bias current applied
Expensive batteries and constant impedance attenuators to said bridge comprising variable shunt impedance means
are eliminated through the use of a shunt circuit making 10 connected across said bridge and said constant direct-
it possible to measure directly the differential current in¬ current source for restoring the balance of said bridge,
volved in rebalancing the bridge. The design of this and means for measuring the current flow through said
bridge circuit readily permits incorporating a self-balanc¬ shunt impedance means.
ing feature by having the amplified output of the bridge 2. The invention of claim 1 in which said balancing
control the amount of shunt current. Improvement over 15 bridge is a Wheatstone bridge and said variable impedance
present day commercial bridges is about 10 to 1 in power arm comprises a thermal-resistive resistor.
range and as much as 100 to 1 in percentage accuracy 3. The invention of claim 2 in which said variable
with the greatest improvement at low power levels. impedance arm is a thermistor.
It will be apparent that the embodiment shown is only
20 References Cited in the file of this patent
exemplary and that various modifications can be made
in construction and arrangement within the scope of in¬ UNITED STATES PATENTS
vention as defined in the appended claims. 2,415,823 Houghton_Feb. 18, 1947
What is claimed is: 2,417,820 Ginzton-Mar. 25, 1947
1. A high-frequency power measuring circuit compris- 25 2,611,795 Carlin-Sept. 23, 1952
826/2
April 21, 1959 m. c. selby etal 2,883,620
HIGH FREQUENCY POWER MEASURING BRIDGE CIRCUIT
INVENTORS
Myron C. Selby
Charles M AH red
Paul 4. Hudson
BY Ira S Berry
ATTORNEYS
827/3
A Technique for Reducing Errors in core. It thus follows that errors in perme¬ DEPENDENCE OF PERMEABILITY
ability measurements should subtract out if ON INDUCTANCE FORMULA
Permeability Measurements Lm and L., are both measured and the results
with Coils* are substituted in an equation of the follow¬
ing form:
A recent note1 gave a summary of the
results of an investigation carried on at the , Lm - Ls
National Bureau of Standards by Kostyshyn M =—-+1- (3)
and Haas2 for the purpose of studying the
sources of error that are associated with the
This equation follows readily from (1) if we
measurement of initial permeability using let AL be the error in the measured value of
coils wound on toroidal samples of rectangu¬
both Lm and Ls. Since AL is independent of
lar cross section. It was pointed out that
permeability, we should write for the cor¬
various laboratories are, in general, using
rect values of Lm and Ls:
one of the following formulas for evaluating
the permeability of a magnetic material. I*m M La ^L (4)
M = (1)
La Ls = La AL. (5)
828/1
Supplementary References C. M. Allred, Precision piston attenuators (unpublished
report, May 1957).
Self-calibrating method of measuring insertion ratio, NBS
Electricity, d-c and Low-Frequency Tech. News Bui. 41, 147 (1957).
A guide to the operation of microwave attenuation measure¬
F. K. Harris, Electrical measurement (John Wiley & Sons, ment systems developed for the Electronic Calibration
Inc., New York, N.Y., 1952). Center (Unpubl. rep’t. by the staff, Microwave Attenuation
G. W. Vinal, Standard cells, standards of electromotive force, Standards, SM-1008, Nov. 1958).
ch. 6 in book, Primary Batteries (John Wiley & Sons, A. Y. Rumfelt and R. J. Como, Rapid insertion device for
Inc., New York, N.Y., 1958). coaxial attenuators, Rev. Sci. Instr. 30, 687-688 (1959).
G. D. Vincent, The construction and characteristics of Ultraprecise attenuation measurement, NBS Tech. News
standard cells, IRE Trans. Instr. 1-7, 22 (1958). Bui. 43, 230-231 (1959).
C. Peterson, Alloys for precision resistors (National Physical A supplement to a guide to the operation of microwave
Laboratories Symp. Precision Electrical Measurement, attenuation measurement systems developed for the
1955) . Electronic Calibration Center (unpublished report by the
F. R. Kotter, Location of the galvanometer branch for staff of Project 84183, 1959).
maximum sensitivity of the Wheatstone bridge, J. Research
NBS 40, 401-404 (1948) RP1884. Calibration Services
A. M. Thompson, The precise measurement of small capaci¬
tances, IRE Trans. Instr. 1-7, 245 (1958).
Facilities and services of the Electronic Calibration Center
J. F. Hersh, A close look at connection errors in capacitance
(unpublished report by the staff of the Electronic Calibra¬
measurements, General Radio Experimenter 33, No. 7
tion Center, Aug. 1958; rev. Jan. 1959).
(1959).
W. H. Wood, Low-cost microvolt potentiometer, Rev. Sci.
Instr. 28, 202-203 (1957). Dielectrics
F. B. Silsbee, R. L. Smith, et al., Equipment for testing
current transformers, NBS J. Research 11, 93-122 (1933) J. L. Preston and E. L. Hall, A study of the seasonal variation
RP580. of radio frequency phase difference of laminated phenolic
B. L. Dunfee, An a-c Kelvin bridge for the audio-frequency insulating materials, Tech. Paper BS 19, 225-235 (1925).
range, Comm, and Electronics (AIEE, Paper No. 56-25, A. B. Lewis, E. L. Hall, and F. R. Caldwell, Some electrical
1956) . properties of foreign and domestic micas and the effect of
elevated temperatures on micas, BS J. Research 7, 403-
Amplifiers and Mixers 418 (1931).
G. Birnbaum and J. Franeau, Measurement of the dielectric
G. F. Montgomery, Intermediate-frequency gain stabilization constant and loss of solids and liquids by a cavity pertur¬
with inverse feedback, Proc. IRE 38, 662-667 (1950). bation method, J. Appl. Phys. 20, 816-818 (1949).
G. E. Boggs, Improvement in gain stability of the super¬ G. Birnbaum, A recording microwave refractometer, Rev.
heterodyne mixer through the application of negative feed¬ Sci. Instr. 21, 169-176 (1950).
back, Proc IRE 40, 202-207 (1952). J. L. Dalke and R. C. Powell, Measuring power factor of low-
N. C. Hekimian, Chart speeds design of feedback amplifiers, loss dielectrics, Electronics 24, 224 (1951).
Electronics 25, 153-155 (1952). G. Birnbaum and S. K. Chatterjee, The dielectric constant
N. C. Hekimian, Feedback amplifiers with stabilized output of water vapor in the microwave region, J. Appl. Phys. 23,
impedances, Tele-Tech 12, 103-105, 148-149 (1953). 220-223 (1952).
G. E. Boggs, Gain stable mixers and amplifiers with current G. Birnbaum, S. J. Kryder, and Id. Lyons, Microwave
feedback, IRE Conv. Record 1953, Part 4, 140-145; Proc. measurements of the dielectric properties of gases, J. Appl.
IRE 42, 1144-1147 (1954). Phys. 22, 95-102 (1951).
G. E. Schafer, A modulator for microwave mixers, IRE .T. H. Beardsley, A variable length re-entrant cavity for
Trans, on Microwave Theory Tech. MTT-6, 333-334 (1958). dielectric measurements from 100 to 400 Me, Rev. Sci.
Instr. 24, 180-181 (1953).
Attenuation, RF to Microwave Dielectric research, NBS Tech. News Bui. 43, 168 (1959).
H. E. Bussey and G. Birnbaum, Cavity resonators for di¬
J. J. Freeman, Theory and design of a cavity attenuator, J. electric spectroscopy of compressed gases, Rev. Sci. Instr.
Research NBS 40, 235-243 (1948) RP1868; Theory and 30, 800-804 (1959).'
design of a waveguide below cutoff attenuator, CRPL-9-2 H. E. Bussey, Cavity resonator dielectric measurements of
(Mar. 1948). rod samples, Insulation, 26-27 (Nov. 1959); Electr. Mfgr.
R. W. Beatty, Cascade-connected attenuators, J. Research 65-66 (Jan. 1960).
NBS 45, 231-235 (1950); Proc. IRE 38, 1190 (1950); CRPL
Preprint 50-17. Field-Strength, RF to Microwave
NBS magnetic attenuator, NBS Tech. News Bui. 35, 109-111
(1951). F. M. Greene, A source of error in the measurement of radiated
F. Reggia, Design details of a linear magnetic uhf oscillator harmonics, Proc. IRE 40, 486-487 (1952); CRPL Preprint
attenuator, FM-TV Radio Comm. 11, 16 (1951). 50-23.
C. M. Allred, Chart for the TEn mode piston attenuator, C. C. Cook, Calibration of commercial field-strength meters,
•T. Research NBS 48, 109-110 (1952); Electronics 26, Tele-Tech 11, 44 (1952).
No. 1 (1953). D. M. Kerns and E. S. Dayhoff, Theory of diffraction in
F. Reggia and R. W. Beatty, Characteristics of the magnetic microwave interferometry, J. Research NBS 64B, 1 (1960).
attenuator of uhf, Proc. IRE 41, 93-100 (19531.
R. W. Beatty and G. E. Huber, Lossv cable attenuator pads,
( Rev. Sci. Instr. 24, 1002-1003 (1953). Frequency
C. M. Allred, Precision piston attenuator (second and third
quarters) (unpublished report, Jan. 1954). P. G. Sulzer, Modified locked-oscillator frequency dividers,
G. E. Schafer and R. R. Bowman, Preliminary report on a Proc IRE 39, 1535-1537 (1951).
modulated subcarrier technique of measuring microwave L. J. Rueger, A. E. Wilson, The microwave frequency stand¬
attenuation (unpublished report, April 28. I960). ard, Radio and Television News 49, 5-7, 40-41 (1953).
829
J. M. Shaull, High precision automatic frequency comparator A. C. MacPherson and D. M. Kerns, Accuracy with which
and recorder, Tele-Tech 14, 58-59 (1955). two loads can be matched on a magic tee, Electronics 23,
P. G. Sulzer, One-megacycle high-stability frequency stand¬ 190-192 (1950).
ard, Radio-Elec. Engr. 24, 14, 38 (1955). H. E. Sorrows, W. E. Ryan, and R. C. Ellenwood, Evaluation
J. M. Shaull, Frequency multipliers and converters for of coaxial slotted-line impedance measurements, Proc
measurement and control, Tele-Tech 14, 86-89, 120-122, IRE 39, 162-168 (1951).
142-146, 148, 159-160 (1955). R. E. Grantham, a reflectionless waveguide termination, Rev-
P. G. Sulzer, High stability frequency standard, Tele-Tech Sci. Instr. 22, 828-834 (1951).
14, 108-109, 420-423 (1955).
R. C. Ellenwood and E. H. Hurlburt, The determination of
M. C. Thompson, M. J. Vetter, and D. M. Waters, SHF
frequency standard uses double conversion, Electronics impedance with a double-slug transformer, Proc. IRE 40,
1690-1693 (1952).
31, 100-101 (1958).
W. J. Anson, A guide to the use of the modified reflectometer
techniques of VSWR measurement (unpublished report,
Frequency and Time, Atomic April 13, 1960).
C. H. Townes, Atomic clock and frequency stabilization in
microwave spectral lines, J. Appl. Phys. 22, 1365-1372
(1951). Magnetism, Low-Frequency
H. Lyons, Spectral lines as frequency standards, Ann. N.Y.
Acad. Sci. 55, 831-871 (1952). Methods of testing magnetic materials, ASTM A 34-55.
J. M. Richardson, Experimental evaluation of the oxygen Methods of test for normal induction and hysteresis of mag¬
microwave absorption as a possible atomic frequency netic materials, ASTM A 341-55.
standard, J. Appl. Phys. 29, 137-145 (1958). Methods of test for permability of feeblv magnetic materials,
R. C. Mockler, J. A. Barnes, R. E. Beehler, H. F. Salazar, ASTM A 342-53.
and R. L. Fey, The ammonia maser as an atomic frequency Methods of test for alternating current core loss and permea-
and time standard, IRE Trans, on Instr. 1-7, 201-202, ability of magnetic materials, ASTM A 343-54.
311-312 (1958).
Methods of test for electrical and mechanical properties of
R. C. Mockler, R. E. Beehler, and J. A. Barnes, A practical magnetic materials, ASTM A 344-52.
limitation to the length of an atomic beam machine
R. L. Sanford, Magnetic testing, NBS Circ. 456 (1946).
(unpublished report, May 25, 1959).
N. Tepley, Ultimate noise limitation of electron multipliers W. P. Harris and I. L. Cooter, Improved bridge method for
and vacuum tube electrometers as used for the measure¬ the measurement of core losses in ferromagnetic materials
ment of beam current in a cesium clock (unpublished at high flux densities, J. Research NBS 60, 509 (1958)
report, Sept. 29, 1959). RP2865.
W. Culshaw and R. C. Mockler, A millimeter wave Fabry-
Perot maser (unpublished report, Nov. 16, 1959). Magnetism, RF to Microwave
Frequency and Time Broadcasts A. A. Feldmann, The machining of powdered iron materials
and ferromagnetic ferrites (unpublished report, Mar. 27,
Experimental standard frequency broadcast on 60 kilocycles, 1952).
NBS Tech. News Bui. 41, 99-100 (1957). High frequency calibration of magnetic materials, NBS Tech.
A. D. Watt and W. W. Brown, Anticipated coverage of a vlf News Bui. 36, 158-160 (1952).
standard frequency broadcasting station located near Fort P. II. Haas, A radio-frequency permeameter, J. Research
Collins, Colo, (unpublished report, Aug. 1959). NBS 51, 221-228 (1953) RP2454.
A. D. Watt, R. W. Plush, W. W. Brown, and A. H. Morgan, B. Kostyshyn and P. H. Haas, Discussion of current-sheet
World-wide standard frequency and time signal broad¬ approximations in reference to high-frequency magnetic
casting (unpublished report, Mar. 8, 1960). measurements, J. Research NBS 52, 279-287 (1954)
RP2501.
Generators, Oscillators, Detectors, and Receivers P. H. Haas, R. D. Harrington, R. C. Powell, and A. L. Ras¬
mussen, Instruments for magnetic high frequency measure¬
P. G, Sulzer, Frequency and amplitude stability of the ments, WADC Tech. Report 54-238 (1954).
cathode-coupled oscillator, Proc. IRE 38, 540-542 (1950). R. D. Harrington, Cavity techniques for permeability
P. G. Sulzer, Wide-range R-C oscillator, Electronics 23, measurements in the vhf region, 6th Electronics Compo¬
88-89 (1950). nents Conf. Proc., 27-28 (1955).
P. G. Sulzer, Precision transistor oscillator, Radio and R. D. Harrington, R. C. Powell, and P. H. Haas, A re-entrant
Television News, Radio Engineering Sec. 49, 18-19 (1953). cavity for measurement of complex permeability in the very-
P. G. Sulzer, High-stability, bridge balancing oscillator, Proc. high-frequency region, J. Research NBS 56, 129-134
IRE 43, 701-707 (1955). (1956).
P. G. Sulzer, An oscillator with direct-feedback frequency R. D. Harrington, Characteristics and present requirements
correction (unpublished report, Feb. 10, 1956). of ferrites, Proc. 13th Ann. Meeting Metal Powder Assoc.,
A. Y. Rumfelt., List of commercially available klystron Chicago, pp. 177-188 (1957).
oscillator tubes, backward wave oscillator tubes and R. D. Harrington and R. C. Powell, High-frequency magnetic
traveling wave amplifier tubes (unpublished report, May permeability measurements using toroidal coils, Proc.
1957). IRE 46, 784 (1958).
G. F. Engen, Amplitude stabilization of a microwave signal A. L. Rasmussen and A. E. Hess, Rf permeameter techniques
source, IRE Trans, on Microwave Theory Tech. MTT-6, for testing ferrite cores, Elec. Mfg. 61, 86-91, 308 (1958).
^ 202-206 (1958). A. L. Rasmussen, A. W. Enfield, and A. Hess, Advances in
E. Niesen, R. W. Beatty, and W. J. Anson, Water-cooling the design and application of the radiofrequency permeam¬
of low-power klystrons used in the laboratory, Rev. Sci. eter, J. Research NBS 56, 261-268 (1956).
Instr. 29, 791-792 (1958). H. E. Bussey and L. A. Steinert, Exact solution for a gyro-
A. H. Morgan and P. A. Hudson, Vhf and uhf power magnetic sample and measurements of a ferrite, IRE
generators for rf instrumentation (unpublished report, Trans, on Microwave Theory Tech. MTT-6, 72-76 (1958).
Mar. 17, 1960). V. E. Bottom, Dynamic measurements of the magnetoelastic
properties of ferrites, NBS Tech. Note 49 (1960).
Impedance, RF to Microwave H. E. Bussey, Tensor permeability data for seven ferrites at
9200 Me, (unpublished report, Dec. 29, 1958).
E. L. Hall, Resistance of conductors of various types and R. D. Harrington and A. L. Rasmussen, Initial and remanent
sizes of windings of single-layer coils at 150 to 6000 kilo¬ permeability spectra of yttrium iron garnet, Proc. IRE
cycles, Tech. Paper BS 21, 109-119 (1926). (Jan. 1959).
G. R. Waite, F. G. Brickwedde, and E. L. Hall, Electrical A. L. Rasmussen, R. D. Harrington, R. C. Powell, and J. L.
resistance and magnetic permeability of iron wire at radio Dalke, A summary of the investigation of ferromagnetic
frequencies, Phys. Rev. 32, 967-973 (1928). materials—ferrites (unpublished report, June 1959).
830
L. A. Steinert, Geometrical anistropy of magnetic materials R. A. Schrack, Radio-frequencv power measurements, NBS
in wave guides and cavities, J. Appl. Phys. 30, 1109 (1959). Circ. 536 (1953).
Magnetic materials; measurement and evaluation laboratory R. W. Beatty and A. C. MacPherson, Mismatch errors in
(unpublished report by the radio and microwave materials microwave power measurements, Proc. IRE 41, 1112-1119
section, NBS, Sept. 1959). (1953).
H. E. Bussey, Tensor permeability measurements on commer¬ A. C. MacPherson and D. M. Kerns, A microwave micro-
cial ferrites (unpublished report, Sept. 30, 1959). calorimeter, Rev. Sci. Instr. 26, 27-33 (1955).
Electromagnetic materials work at NBSBL (unpublished re¬ R. W. Beatty and F. Reggia, An improved method of measur¬
port by the radio and microwave materials section, NBS, ing efficiencies of ultra-high-frequency and microwave
Oct. 22, 1959). bolometer mounts, J. Research NBS 54,'321-327 (1955).
Microwave Optics Improved thermistor bridge for rf power measurements, NBS
Tech. News Bui. 40, 134-135 (1956).
Applications of optical techniques of millimeter wavelengths, H. A. Thomas, Microwave power measurements employing
NBS Tech. News Bui. 41, 24-28 (1957). electron beam techniques, Proc. IRE 45, 205-211 (1957).
W. Culshaw, Reflectors for a microwave Fabry-Perot inter¬ A. C. MacPherson, Calorimetric measurements of microwave
ferometer, IRE Trans, on Microwave Theory Tech. power (unpublished report, April 1, 1957).
MTT-7 (April 1959). Recent international comparisons of microwave power
W. Culshaw, Diffraction corrections in microwave optics standards, NBS Tech. News Bui. 43, 155 (1959).
(unpublished report, July 1, 1958). G. F. Engen, A bolometer mount efficiency measurement
J. M. Richardson, Provisional alinement requirement for technique (unpublished report, June 20, 1960).
Michelson microwave interferometer carriage (unpublished
report, Sept. 1958). Quartz
W. Culshaw, High resolution millimeter wave Fabry-Perot
interferometer, IRE Trans. (Mar. 1960). W. D. George, M. C. Selby, and R. Scolnik, Electrical
D. M. Kerns and E. S. Dayhoff, Theory of diffraction in characteristics of quartz-crystal units and their measure¬
microwave interferometrv, J. Research NBS 64B, 1-13 ment, J. Research NBS 38,'309-328 (1947) RP1774.
(1960). L. T. Sogn and C. Barclay, Fundamental techniques in the
W. Culshaw and R. D. Mockler, A millimeter wave Fabry- frequency adjustment of quartz crystals, NBS Circ. 480
Perot maser (unpublished report, Nov. 16, 1959). (1949).
W. D. George, M. C. Selby, and R. Scolnik, Precision meas¬
Microwave Theory urement of electrical characteristics of quartz crystal units,
Proc. IRE 36, 1122-1131 (1948).
D. M. Kerns, Basis of the application of network equations C. Barclay and L. T. Sogn, Reference data for orienting
to waveguide problems, J. Research NBS 42, 515-540 quartz plates by X-ray diffraction, NBS Circ. 543 (1953).
(1949) RP1990. C. Barclay, Quartz X-ray standards (tentative specifications
J. J. Freeman, The field generated by an arbitrary current for, and notes on their use) (unpublished report, March
distribution within a waveguide, J. Research NBS 44, 1958).
193-198 (1950) RP2070. P. A. Simpson and A. H. Morgan, Investigation of stability
D. M. Kerns, Analysis of symmetrical waveguide junctions, of quartz resonators at low temperatures (final report)
.J. Research NBS 46, 267-282 (1951). (unpublished report, Oct. 1959).
R. W. Hedberg and D. M. Kerns, Propagation constant in S. A. Basri, A method for the dynamic determination of the
rectangular wave guide of finite conductivity, J. Appl elastic, dielectric and piezo-electric constants of quartz,
Phys. 25, 1550-1551 (1954). NBS Mono. 9 (1960).
H. E. Bussey and L. A. Steinert, An exact solution for a
cylindrical cavity containing a gyromagnetic material, RF and Microwave Standards and Measurements,
Proc. IRE 45, 693-694 (1957).
D. M. Kerns and E. S. Dayhoff, Theory of diffraction in General
microwave interferometry, J. Research NBS 64B, 1 (1960).
Radio instruments and measurements, 1st ed. (1918), 2d
Power ed. (1924, reprinted 1937).
Electromagnetic materials work at NBSBL (unpublished
D. M. Kerns, Determination of efficiency of microwave report by the radio and microwave materials section, NBS,
bolometer mounts from impedance data, J. Research NBS Oct. 22, 1959).
42, 579-585 (1949). Standards and calibration in radio and electronics, NBS
D. M. Kerns, Analysis of the direct-current bolometer bridge, Tech. News Bui. 43, 226-230 (1959).
J. Research NBS 43, 581-589 (1949).
831
Subject Index
Combined subject index for the three volumes of NBS Handbook 77
A Volume
Volume and page
and page Calibration of liquid-in-glass thermometers^_ 11—153
Absolute temperatures below 1° K_ 11-10 Calibration of proving rings_ 11-573
Absolute viscosity of water at 20° C_ 11-523 Calibration of pyrometric cones_ 11-131
Absorption of X-rays in air_ III-900 Calibration of signal generators, application of RF
Accurate radio frequency microvoltages_ 1-634 micropotentiometers_ 1-621
Acoustic impedance of a right circular cylindrical Calibration of thimble-chamber, soft X-rays_ III-907
enclosure_ 11-858 Calibration of vibration pickups by the reciprocity
Acoustic method for measurement of vibration method_ 11-621
amplitudes_ 11—802 Calibration, pressure, of condenser microphones
Acoustics, list of publications_ 11-952 above 10000 cps_ 11-862
Adiabatic demagnetization experiments in low tem¬ Calibration techniques of liquid-flowmeters_ 11-648
perature investigations_ 11-22 Calorimeter, adiabatic, range 30° to 500° C_ 11-228
Adjustment of high-precision frequency and time Calorimeter, bunsen-type_ 11-222
standards_ 1-482 Calorimeter, dry, static, for RF power measurement- 1-809
Airplane camera calibration, sources of error_ III-747 Calorimeter for measuring heat capacity and heat
Ammonia vapor, super heated, specific heat of_ 11-460 of vaporization of water, 0° to 100° C_ 11-277
Angle blocks, assembled polygon for the calibration Calorimeter for precise measurement of heat con¬
of_ III—318 tent and heat capacity_ 11-241
Attenuation, microwave measurements, standard Calorimeter (micro), microwave, X-band_ 1-819
for_ 1-776 Calorimetry, precision adiabatic, automatic tem¬
Attenuators, variable, UHF and microwave, mis¬ perature regulation and recording_ 11-236
match errors in the measurements of__ 1-792 Calorimetry, temperature correction methods_ 11-212
Audiometers, calibration of_ 11-852 Camera, airplane, errors in calibration_ III-747
Camera obscura, for instrument reading_ 1-432
B Capacitance measurement, absolute_ 1-296
Capacitor, calibration by step-up methods_ 1-305
Balance, density, for liquid hydrocarbons, assembly
Capacity standards_ III-38
and calibration_ 11-691
Carbon dioxide, gaseous, heat capacity of_ 11-352
Balance, quick-weighing, testing_ III-740
Cavity ionization as a function of wall material_ III—933
Benzoic acid, heat of combustion and density_ 11-410
Cells, standard, effect of temperature_ 1-252
Beryllium oxide, thermal conductivity, 40° to 750°
Cells, standard, effect of temperature inequality in
C_ 11-508
the limbs_ 1—254
Bibliography, electrical instruments and measure¬
Cells, standard, hysteresis effects_ 1-259
ments_ 1-83, 1-447
Cells, standard, international comparisons- 1-244
Blocks, angle, calibration of_ III-381
Cells, standard, modifications_ 1-241
Bolometer bridge for standardizing radio frequency
Cells, standard, unit of electromotive force- 1-236
voltmeters_ 1-605
Chamber, ionization, standard, requirements for
Bolomet.ric power measurements, accurate, self-
balancing, d-c, bridge for_ 1-814 250- to 500-kilovolt X-rays_ III-926
Chamber, reverberation, study of the sound power
Bomb calorimeter measurements_ 11—361
Branching ratio in the decay of polonium-210_III-1014 output of subsonic air jets_ 11-825
Bridge, bolometer, for standardizing RF volt¬ Chambers, ionization, free-air, design of_ III-870
meters_ 1-605 Chambers, ionization, free-air, soft X-ray region,
Bridge, circuit, power measuring, high-frequency_ 1-825 design of_ III—889
Chromic methylammonium alum as a thermometric
Bridge equations_ 1-162
Bridge, establishment of known ratios_ 1-168 substance_ 11-10
Bridge, insulation_ 1-184 Color temperature, lamp standards_ HI—587
Bridge, NBS precision_ 1-158 Colorimetry_ III-470
Bridge, precision and accuracy_ 1-172 Colorimetry and spectrophotometry, list of publica¬
Bridge, self-balancing, d-c, for accurate bolometric tions__ III—792
power measurements_ 1-814 Colorimetry, photoelectric tristimulus with three
Bridge, thermoelectromotive forces_ 1-182 filters__ III-529
Bridge, Thomson_ 1-153 Comparison of four national radium standards- III-993
Bridge, Thomson, adjustments of_ 1-171 Comparisons of national radium standards- III-1002
Bridge, Wagner branches_ 1-188 Converter, volt-ampere, wide-range_ 1-352
Bridge, Wheatstone_ 1-147 Converters, thermal, as a-c-d-c transfer standards-. 1-317
Bridges, index to subjects_ 1-210 Converters, voltage, theory, construction, tests- 1-336,
1-339
Bunsen-type calorimeter_ 11-222
1,3-butadiene, thermodynamic properties_ 11-413 Converters, voltage, thermal_ 1-335
Couplers, directional, method of measuring the
directivity of_ 1-802
C Current, absolute measurement of_ 1-41
Calibrating consoles, voltmeter, RF_ 1-598 Current and voltage measurements at audio fre¬
Calibrating testing machines with proving rings_ 11-573 quencies_ 1-317
Current balance, NBS_ 1-99
Calibration and use of VHF field-intensity meters, Current measurement_ 1-89, 1-99
influence of the ground on_ 1-662
Calibration Center, Electronic, high frequency D
standards NBSBL_ 1-463
Calibration of a density balance for liquid hydro¬
carbons___ 11—691 Dead-weight machines, 111,000 and 10,000 pound
Calibration of audiometers_ 11-852 capacities___.-
Calibration of commercial radio field-strength Density and heat of combustion of benzoic acid- 11-410
meters at the National Bureau of Standards_ 1-670 Density and volumetric determinations- III-719
832
Volume Volume
and page and page
Density balance for liquid hydrocarbons, assembly High frequency impedance standards at the National
and calibration-...- 11-691 Bureau of Standards_ 1-675
Density of solids and liquids- 11-659 High frequency standards of the Electronic Calibra¬
Design of free-air ionization chambers - .- 111—87 0 tion Center, NBSBL_ 1-463
Design of free-air ionization chambers for the soft High-frequency voltage measurement_ 1-582
X-ray region (20-100 kv)- 111-889 High-precision frequency and time standards,
Directional couplers, method of measuring the adjustment of_ 1-482
directivity of_.- - - - 1-802 High temperature materials, physical properties_ 11-241
Double freezing-point method of determining Humidity, measuring_ 11-698
styrene purity- 11-195 Humidity-pressure apparatus_ 11-788
Humidity (relative)-temperature relationships_ 11-794
E Humidity test apparatus, low-temperature_ 11-780
Hydrocarbons, determination of purity by measure¬
Earphones, response in ears and couplers- 11-869 ment of freezing points_ 11-176
Electric hygrometers- 11-718 Hydrometers, testing of_ III-707
Electrical and magnetic units, extension and dis¬ Hygrometers, electric_ 11-718
semination of, by NBS- 1-53 Hygrometers, testing_ 11-698, 11-741
Electrical and magnetic units, international rela¬ Hygrometers, testing, recirculation apparatus for. 11-774
tions _ 1-82
Electrical instruments, care and use- 1-43
Electrical units, derivation- 1-58 I
Electrical units, establishment and maintenance. _ 1-12
Electrodynamometer, Pellat-type- 1-89 Impedance standards, high frequency, at the
Electrometer, Brooks attracted disk- 1-283, 1-286 National Bureau of Standards_ 1-675
Electrometer, compressed gas- 1-284 Impedance standards, microwave, methods of meas¬
Electrometers_ 1—277 urement_ 1-680
Electromotive force; history of USA unit- 1-237 Impedance tube method of measuring sound
Electrothermic transfer standards, development and absorption coefficient_ 11-812
description_ 1-318 Indicators, electrical, scale and reading errors of_ 1-425
Engineering mechanics, list of publications on- 11-913 Indices of refraction, precision, rapid determination
Expansion, thermal, of solids- III-349 of_ 111—577
Influence of the ground on the calibration and use of
VHF field-intensity meters_ 1-662
F Infrared high-resolution grating spectrometer_ III—835
Facility for the evaluation of resistance strain gages Infrared measurements with a small grating from
at elevated temperatures_ 11-638 100 to 300 microns_ III—838
Field intensity meters, calibration and use of, Instrument reading, a camera obscura for_ 1-432
influence of the ground on_ J-662 Instruments, electrical, testing_ 1-421
Field-intensity standards, VHF, development of- 1-641 Instruments, spectrophotometric_ III-381
Field strength meters, commercial, radio, calibration Interferometer for large surface testing_ III-467
of__-_----- 1-670 Interferometer, parallel testing_ III-459
Free-air ionization chambers for soft X-ray region.. III-889 Interferometer, stroboscopic for vibration measure¬
Freezing points of hydrocarbons, a measurement of ment_ 11-808
purity_ 11-168 Interferometry, precision millimeter wave, at the
Frequencies and time signals, standard- 1-475 National Bureau of Standards_ 1-756
Frequency and time, national standards in the U.S. 1-472 International Bureau of Weights and Measures_ III—28
Frequency and time standards, high precision, ad¬ International temperature scale of 1948_ II—1
justment of_ 1-482 Ionization (cavity) as a function of wall material III—933
Frequency, VSWR and |T| for certain impedance Ionization chamber, standard, requirements for 250-
standards, tables_ 1-695, 1-702, 1-710 to 500-kilovolt X-rays_ III-926
Ionization chambers, free-air, design of_ III-870
Ionization chambers, free-air, soft X-ray region- 111—889
G
H
M
Heat capacity and heat of vaporization of water in
the range 6° to 100° C, measurements_ 11-277 Machines, dead-weight, 111,000 and 10,000 pound
Heat capacity of gaseous carbon dioxide_ 11-352 capacities_ 11-596
Heat capacity of sodium between 0° and 900° C, the Magnetic and electrical units, extension and dis¬
triple point and heat of fusion_ 11-341 semination of, by NBS_— 1-53
Heat content and heat capacity, precise measure¬ Magnetic thermometer to examine the helium
ment, 0° to 1500° C__ 11-241 vapor-pressure scale of temperature- 11-34
Heat of combustion and density of benzoic acid_ 11-410 Mass standards_ III-31, 111—34, 111—615
Heat of combustion of benzoic acid, standardization Mass standards, precision_ 111-591
of bomb calorimeters_ 11-386 Matching termination, UHF and microwave- 1-578
Heat of combustion, precise measurement with bomb Measurement, methods, microwave impedance
calorimeter_ 11-361 standards_ 1-680
Heat of fusion, triple point, and heat capacity of Measurement of absorption line half-widths and
sodium between 0° and 900° C_I_ 11-341 intensities, slit function effects- 11-148
833
\
Volume P
and page
Volume
Measurement of correlation coefficients in reverber¬ and page
ant sound fields_ 11-846 Parallelism_ III-459
Measurement of microwave SWR, new technique,-- 1-739 Photoelectric tristimulus colorimetry_ III-529
Measurement of UHF and microwave variable Photographic lenses, testing_;_ III-761
attenuators, mismatch errors in the_ 1-792 Planeness of optical flats_ III-336
Measurement of vibration amplitudes, acoustic Polonium-210, branching ratio in the decay of_III-1014
method_ 11-802 Potentiometer, lindeck_ 1-218
Measurement of vibration by stroboscopic inter¬ Power measurements, RF, calorimeter for_ 1-809
ferometer_ 11-808 Power measurements, microwave, at NBS_ 1-806
Measurement, power, RF, calorimeter for_ 1-809 Power measuring bridge circuit, high frequency_ 1-825
Measurement, thickness_ III-80 Power supplies for meters and instruments, 60 cycle
Measurement, voltage, high-frequency_ 1-582 tests-._ 1-438
Measurements, electrical, transfer from d-c to a-c,_ 1-65 Practices, suggested, for electrical standardizing
Measurements, infrared_ III-838 laboratories_ I-l
Measurements, microwave reflection coefficients, Pressure-humidity apparatus_ 11-788
magnified and squared, VSWR responses_ 1-746 Pressure measurements, list of publications_ 11-939
Measurements, microwave, standard of attenuation Proving rings for calibrating testing machines_ 11-573
for_ 1-776 Proving rings, temperature coefficients for_ 11-612
Measurements of acoustic impedance to determine Publications, list, optical instruments_ III-778
reverberant sound absorption coefficients_ 11-814 Publications on electrical units, instruments and
Measurements of heat of combustion, precise, with measurements_ 1-447
bomb calorimeter_ 11-241 Pyrometric cones, calibration of_ 11-131
Measurements of pressure, list of publications_ 11-939
Measurements, power, self-balancing d-c bridge_ 1-814 R
Measurements, power, microwave, at NBS_ 1-806 Radiation, cavity ionization, wall material_ III-933
Measurements, precise, with Bingham and Cannon Radiation physics, list of publications_ III-1018
master viscometers_ 11-554 Radioactivity, standards of, preparation, mainte¬
Measurements, temperature, list of publications_ 11-200 nance and application_ III-942
Measuring, bridge circuit, power, high-frequency_ 1-825 Radium, national standards, comparison of four_ III-993
Measuring humidity_ 11-698 Radium standards, national, comparisons of_ III-1002
Measuring sound absorption coefficient_ 11-812 Radium standards, preparation of new solutions. _ III-1008
Measuring system, the basis of our_ III—14 Reflection coefficient measurements VSWR re¬
Measuring the directivity of directional couplers_ 1-802 sponses_ 1-746
Mechanics, list of publications_ 11-913 Reflectometer techniques, microwave_ 1-751
Meter bar #27, United States Prototype_ III—53 Refraction, indices of, and dispersion by immersion,
Meter line standards, calibration at NBS_III-53, III-67 precision, determination_ III-577
Meters, electrical, reading errors of_ 1-425 Refractometry, prism, standard conditions for_ III-432
Method for measurement of E'l' in the reciprocity Relative humidity-temperature relationships of some
calibration of condenser microphones_ 11-893 saturated salt solutions_ 11-794
Methods of calibration and testing of hygrom¬ Resistance, index to subjects_ 1-210
eters_ 11-698, 11-741 Resistance thermometry, precision, and fixed points. 11-40
Methods of measuring humidity and testing hygrom¬ Resistors, commercial, aging and voltage of_ 1-234
eters_ 11-698 Resistors, load coefficients_ 1-155
Micrographs, interference, interpretation of_ III-328 Resistors, multimegohm, measurement_ 1-230
Microinterferometer, description_ III-324 Resistors, precision, measurement_ 1-108
Microphones, calibration of_ 11-894 Resistors, standard, precision_ 1-143
Microphones, method of measurement_ 11-893 Response of earphones in ears and couplers_ 11-869
Microphones, condenser, pressure calibration of_ 11-862 Reverberation chamber method of measuring sound
Micropotentiometers, RF, for calibration of signal absorption coefficient_ 11-843
generators to 1000 Me_ 1-621 Reverberation chamber, output of a sound source
Microwave impedance standards, methods of meas¬ in _ 11-833
urement_ 1-680 RF micropotentiometers for calibration of signal
Microwave measurements, standard of attenuation generators to 1000 Me_ 1-621
for_ 1-776 RF microvoltages, accurate_ 1-634
Microwave microcalorimeter, X-band_ 1-819 RF voltmeter calibrating consoles_ 1-598
Microwave power measurements at NBS_ 1-806
Microwave reflection coefficient measurements mag¬ S
nified and squared VSWR responses_ 1-746
Microwave reflectometer techniques_ 1-751 Scintillation spectrometry of low-energy brems-
Microwave SWR, new technique for the measure¬ strahlung_ III—913
ment of_ 1-739 Slit function effects in the direct measurement of
Mismatch errors in the measurement of UHF and absorption line half-widths and intensities_ 11-148
microwave variable attenuators_ 1-792 Solids and liquids, density of_ 11-659
Sound absorption coefficient, measuring_ 11-812, 11-843
N Sound absorption (reverberant) coefficients deter¬
mined by acoustic measurements_ 11-814
National radium standards, comparison of four_111-993 Sound power output of subsonic air jets_ 11-825
National radium standards, comparisons of_III-1002 Sound source, output of, in a reverberation chamber. 11-833
National standards of time and frequency in the Specific and latent heat of water, 0° to 100° C_ 11-277
United States_ 1-472 Specific heat of superheated ammonia vapor_ 11-460
New technique for the measurement of microwave Spectral absorption method for determining popu¬
SWR_ 1-739 lation “temperatures” in hot gases_ 11-139
Nitrogen, thermal condueitivity, 50° to 500° C, 1 to Spectral radiant intensities, procedure_ III-458
100 atm_ 11-515 Spectral standards_ III-381
O Spectrometer, infrared, high resolution_ III-835
Spectrometers_ III-390
Ohm, maintenance of, at NBS_ 1-20 Spectrometers, wavelength calibration of, applica¬
Optical flats, determination of planeness and bind¬ tion of vibration-rotation structure_ III-841
ing- II1—336 Spectrometry, scintillation, low energy bremsstrah-
Optical instruments, list of publications on_ III-778 lung___ III—913
Optical surfaces, large, testing_ III-343 Spectrophotometers_ III-381
834
Volume Volume
and page and page
Spectrophotometry, 200 to 1000 millimicrons_ III—381 Thermometers, liquid-in-glass, calibration_ 11-153
Spectrophotometry and colorimetry, list of publi¬ Thermometry, precision resistance_ 11-40
cations_ III-792 Thermometry, thermoelectric_ 11-68
Standard conditions for refractometry_ III-432 Thickness measurement_ III-80
Standard frequencies and time signals_ 1-475 Thimble-chamber calibration on soft X-rays_ III-907
Standard ionization-chamber requirements for 250- Thorium-halide lamps, wavelengths from_ III-861
to 500-kilovolt X-rays_ III-926 Time and frequency, national standards in the U.S.. 1-472
Standard of attenuation for microwave measure¬ Time and frequency standards, high precision,
ments_ 1-776 adjustment of_ 1-482
Standardization of bomb calorimeters, heat of Time signals and frequency, standard_ 1-475
combustion of benzoic acid_ 11-386 Time standards_ 1-472
Standards, high frequency impedance, at the Transfer instrument, ac-dc_ 1-346, 1-369
National Bureau of Standards_ 1-675
Standards, high frequency, of the Electronic Cali¬ U
bration Center, NBSBL_ 1-463 Units and systems of weights and measures, their
Standards, impedance, microwave, methods of origin, development, and status_ III—21
measurement_ 1-680 Units, electrical, capacitance, derivation_ 1-59
Standards, lamp, color temperature_ III-587 Units, electrical, current, power, energy, derived_ 1-62
Standards, length, commercial_ III-73 Units, electrical, establishment and maintenance_ 1-12
Standards, length, reference_ III-72 Units, electrical, inductance, derived_ 1-61
Standards, length, working_ III-72 Units, magnetic, derived_ 1-63
Standards, national of time and frequency in the Units, measurement, tables of equivalents_ III-49
U.S_ 1-472 Units of measurement, interrelation_ III-45
Standards, national radium, comparisons of_ III-1002
Standards, national radium, comparison of four_ III-993 V
Standards, national radium, statistical analysis_ III-999
Standards of mass, design and test_ III-615 Vapor-pressure scale of temperature, helium, using
Standards of mass, precision_ III-591 a magnetic thermometer_ 11-34
Standards of radioactivity, preparation, mainte¬ VHF field-intensity standards, development of_ 1-641
nance and application of_ III-942 Vibration amplitudes, measurement of, acoustic
Standards or radium, preparation of new solutions.- III-1008 method_ 11-802
Standards, spectral energy_ III-424 Vibration measurement, stroboscopic interferom¬
Standards, spectral reflection_ III-427 eter for_ 11-808
Standards, spectral transmission_ III-425 Vibration pickups, calibration of_ 11-621
Standards, threshold, probe tube method for the Vibration-rotation structure in absorption bands
transfer of_ 11-865 for the calibration of spectrometers from 2 to 16
Standards, VHF field intensity, development of_ 1-641 microns_ III—841
Strain gages, evaluation of elevated temperatures 11-638 Viscometers, Bingham and Cannon master, precise
Styrene, determination of purity by double freezing measurements with_ 11-554
point method_ 11-195 Viscosity, precise measurements_ 11-554
Surface roughness, measurement_ 111-324 Volt box, standard_ 1-214
SWR, microwave, new technique for the measure¬ Volt boxes, testing apd performance_ 1-211
ment of_ 1-739 Volt, maintenance of, at NBS_ 1-23
Voltage and current measurements at audio fre¬
T quencies_ 1-317
Tables of frequency, VSWRand |r| for selected half- Voltage converters, thermal_ 1-335
round inductive obstacle impedance standards_ 1-695, Voltage, high, measurement of_ 1-262
1-702, 1-710 Voltage, high, measurements by transformer meth¬
Tapes, measuring, calibration of_ III-67 ods_ 1-272
Technique (new) for the measurement of microwave Voltage measurement, high-frequency_ 1-582
SWR_ 1-739 Voltage, measurement of crest and surge_ 1-79
Temperature coefficients for proving rings_ 11-612 Voltage measurements, to 30 Me_ 1-335
Temperature, effects on refractive media_ III-446 Voltage ratio, measurement, audio frequency_ 1-290
Temperature measurements, list of publications_ 11-200 Voltages, micro, RF, accurate_ 1-634
Temperature measurements of hot gases spectral Voltmeter consoles, calibrating, RF_ 1-598
absorption method_ 11-139 Voltmeter, electrostatic, compressed gas_ 1-282
Temperature regulation and recording, automatic Voltmeter, ellipsoidal_ 1-280
in precision adiabatic calorimetry._ 11-236 Voltmeters, electrostatic_1—277, 1-279
Temperature scale, international, 1948_ II—l Voltmeters, RF, bolometer bridge for standardizing. 1-605
Termination, matching, UHF and microwave_ 1-578 Volumetric and density determinations_ III—719
Test apparatus, humidity, low-temperature_ 11-780 Volumetric apparatus, glass, testing_ III-723
Testing and calibration of hygrometers_ 11-698, 11-741 VSWR, frequency and |r|, tables for impedance
835
Author Index
Combined author index for the three volumes of NBS Handbook 77
83G
Volume and page Volume and page Volume and page
Saunders, J. B_... Ill—216, Stimson, H. F_II—1, 11-40, Waterhouse, R. V .. 11-825,11-833,
III—343, 11-277, 11-846
111-459, II- 460 Weaver, Frank D .. 1-421, 1-425,
111—467 Stockmann, L. L_ III-1002, 1-428, 1-432
Schafer, G. E _ 1-795, 1-802 III- 1008 Wenner, Frank 1-143
Schmidt, V. A _ 11-808 Strang, A. G_ 111-249, West, E. D .. 11-228, 11-236
Schwebel, A ... Ill—1008 111-318, Wexler, Arnold ... 11-698,11-718,
Scott, Arnold H . - ... 1-230 111-324, 11-741,
Scott, R. B 11-413 III-349 11-774,
Selby, M. C 1-463,1-582, Streiff, Anton J_ 11-176 11-780,
1-598,1-605, Sugg, R. E_ III-225 11-788,
1-634,1-683, Swindells, James F_ 11-153,11-155, 11-794
1-825 11-523, Williams, E. S . . 1-335, 1-352
Seliger, H. H __ II1-942, II- 554 Wilson, B. L .. 11-573,11-596,
III-945 Sze, Wilbur C_ 1-290 11-612
Shafer, M. R — 11-648
1-482 Wolzein, E. C. ... 1-683
Shaull, John M_
Tate, D. R_ 11-573, 11-596, Wvckoff, H. 0__ .. III—870,
Shoaf, J. H __ 1-475
• 11-612 111-926
Silsbee, Francis B __ 1-1,1-12,1-15,
Taylor, L. S_ 111-900
1-53, 1-211
Thomas, James L_1-108
Simmons, B. D - 11-862 Youden, W. J .. 111—993,
Thompson, M. C., Jr__ 11-846
Sligh, T. S., Jr 11-460 III-999,
Tidwell, E. D_ 111—841
Smith, C. C . 111-993 111—1002,
Tilton, Leroy W_ III-432
Smith, E. R _ _ 11-808 III—1008
Snyder, W. F_ __ 11-852
1-641 Victory, F. C_ 111-243 Young, T. R_ _ .. III-208
Solow, Max
Souder, Wilmer _ . 111-349
Spinks, A. W 1-406 Washer, Francis E_ III—747, Zapf, Thomas L. . _ 1-305, 1-310,
Stanley, R. W_ _ 111-861 III- 761 1-406
837
PUBLICATIONS PROGRAM OF THE NATIONAL BUREAU
OF STANDARDS
Results of fundamental and applied research at the National Bureau of
Standards in physics, chemistry, engineering, and mathematics are published
in the Bureau’s own series of publications. Papers are also published in the
journals of professional and scientific societies and in technical and trade
publications.
JOURNAL OF RESEARCH
reports National Bureau of Standards research and development in physics,
mathematics, chemistry, and engineering. Comprehensive scientific papers
give complete details of the work, including laboratory data, experimental
procedures, and theoretical and mathematical analyses. Illustrated with
photographs, drawings, and charts.
The Journal presents review articles by recognized authorities and com¬
pilations of information on subjects closely related to the Bureau’s technical
program.
Selected NBS nonperiodical publications and articles bv the Bureau staff
in professional journals are abstracted in the appropriate section of the Journal.
In addition, each section carries a complete listing of all Bureau publications
that are not abstracted in that section.
D. RADIO PROPAGATION
Reporting research in radio propagation, communications, and upper
atmospheric physics. Topics covered include propagation in
ionized media, scattering by turbulence, effect of irregular terrain on
propagation, diffraction and scattering by solid obstacles, propaga¬
tion through time-varying media, surface waves, and antennas.
Issued six times a year. Annual subscription: Domestic, $4.00;
foreign, $4.75.*
* Difference in price is due to extra cost of foreign mailing.
838
OTHER PERIODICALS
The following periodicals may be obtained on a 1-, 2-, or 3-year subscrip¬
tion basis (although no reduction in rates can be made) from the Superintendent
of Documents, U.S. Government Printing Office, Washington 25, D.C.:
Technical News Bulletin. Summaries of current research at the National
Bureau of Standards are published in the Bulletin. The articles are brief, with
emphasis on the results of research, chosen on the basis of their scientific or
technologic importance. All publications by Bureau staff are announced in the
Bulletin. Annual subscription: Domestic, $1.50; foreign, $2.25.
Basic Radio Propagation Predictions. The Predictions provide the information
necessary for calculating the best frequencies for communication between any
two points in the world at any time during the given month. The data are
important to users of long-range radio communication and navigation, including
broadcasting, airline, steamship, and wireless services, and to investigators of
radio propagation and ionosphere. Each issue, covering a period of 1 month,
is released 3 months in advance. Annual subscription: Domestic, $1.50;
foreign, $2.00.
NONPERIODICALS
Applied Mathematics Series. Mathematical tables, manuals, and studies.
Monographs. Major contributions to the technical literature on various sub¬
jects related to the Bureau’s scientific and technical activities. These comprise
much of the type of material previously published in the larger National Bureau
of Standards Circulars.
Handbooks. Recommended codes of engineering and industrial practice
(including safety codes) developed in cooperation with interested industries,
professional organizations, and regulatory bodies.
Miscellaneous Publications. Charts, administrative pamphlets, annual reports
of the Bureau, conference reports, etc.
Technical Notes. This series consists of communications and reports (covering
both other agency and NBS-sponsored work) of limited or transitory interest.
NBS Technical Notes are designed to supplement the Bureau’s regular publi¬
cations program. They are for sale by the Office of Technical Services, U.S.
Department of Commerce, Washington 25, D.C.
839
Monthly Catalog of United States Government Publications. Issued by the
Superintendent of Documents. $3.00 a year (foreign $4.50).
Government Printing Office Price Lists. Lists of Government publications on
particular subjects. Free from the Superintendent of Documents.
In addition, many of the technical journals carry notices of new Bureau
publications of interest in their respective fields.
Purchase Procedure
Most NBS publications are for sale ONLY by the Superintendent of Docu¬
ments, Government Printing Office, Washington 25, D.C. (PLEASE DO
NOT SEND ORDERS TO THE NATIONAL BUREAU OF STANDARDS.)
Payment in advance is required, and may be made by coupons sold in sets of
20 for $1.00 and good until used, postal money order, express order, check, or
by currency (at sender’s risk). A discount of 25 percent is allowed to pur¬
chasers of 100 or more copies of a single publication to be mailed to one address.
Postage is not required for orders mailed to the United States and possessions.
Remittances from foreign countries should be made by international money
order or draft on an American bank, and should include an additional one-fourth
of the purchase price for postage.
Some purchasers of Government publications prefer to use the “deposit
account” service provided by the Superintendent of Documents. Deposits of
$10 or more are accepted, against which orders may be placed without making
individual remittances or first, obtaining quotations. Order blanks are provided
by the Superintendent of Documents for the convenience of users of the deposit
account arrangement. Orders on these blanks are returned with notations
showing the publications supplied, explanations regarding those not sent, the
amount charged, and the balance on deposit.
Out of print NBS publications are available for reference use in many
scientific, technical, and public libraries. The Bureau’s Catalog of Publi¬
cations, NBS Circular 460, contains a list of Government Depository Libraries
that receive selected Bureau publications. Some libraries are equipped to
provide photostat or microfilm copies of out of print documents.
840
U.S. GOVERNMENT PRINTING OFFICE: 1961 O - 572772