AAdvanc..

ed Instrumentations

SL/FL Series Operating Lamps

Operation & Installation Manual


I
I Table of Co ntent s

I User Gui de--------------------------------------------------------------------------- 3

' Eq ui pmen I I ntrod uc ti 011------------------------------------------------------------ 4

Equipment F eatw-es----------------------------------------------------------------- 4

Primary Technical Specifications------------------------------------------------- S

Product Schematics----------------------------------------------------------------- 6-8

Installation Instructions------------------------------------------------------------ 9- l 0
. Instructlons--------------------------------------------------------------
Operatmg . ll

Cleaning and Sterilization---------------------------------------------------------- 12

Precautions--------------------------------------------------------------------------- 13

Transportation and Storage-------------------------------------------------------- 13

Equipment Handling---------------------------------------------------------------- 14
,

User Guide

or
Tha nk y o u ru1d co ngratu lati ons o n your purch ase our Adva nced ln strume ntati.on SL/FL Series
Operoting Lamps. In order to operate the equipment correctly, you need to read the operati ng
instructions prior to operat ion.

In any case. the operator must use the equipment on the basis of comprehending thoroughly and
corTectly the contents of this manual and strictly follow the request of the manual in the course of
operating the equipment.

Warning: The equipment must not be used close to flammable gases, flammable explosive or an
environment containing anesthetics.

Manufacturer's Limited Warranty

We shall be liable for the product's safety, reliability and performance only in the following
cases:
I. Certified operators carry out all assembly, extension, adjustment, improvement and
repaIT.
II. Location of installation complies with the national electrical safety standard.
III. All operations abide by the instructions within specified applications.

* This unit has a one-year warranty issued at the time the product is unpacked. Within the
warranty term Advanced Instrumentation covers all repairs, excluding man-made damage.
' l<:quipmcnC: lntroducCion

SI 'FL St.·ri cs Opc..·mting Lu mp is u new product m anuluc turcd by Advanced In strumentatio ns.
Jnc. The, on: dt·sig m:·d ns u mul1i -pris m rc ll cc lor, a uni ve rsa l ha ng in g device and an e nc losed
s leek Jar;, p ho lder tha1 ca n mee t the reques t or
illumina ti o n, ste rili zati o n a nd la mina r fl ow in a ll
kinds o f o pera tio ns .

T he sec urity class ificati o n of SL/FL Seri es Operating Lamps is a class I (gro und) , whi c h is
shockproo f rype R indirect contact with the human body.

S L/FL Series Operating Lamps are classified as operating lamps, wall operating lamps, wall
interventional lamps, stand operating lamps, and stand interventional lamps, which can be used
fo r all kinds of operations.

Equipment Features

1. The computer-aided design (CAD) multi-prism and multi-angle reflector can produce
deep illumination of 60 beams, which has an excellent shadow less effect. The spotlight
can produce a color temperature luminance of more than 60000 Ix. 3000k, which reflects
the color of the human body, and meets the illumination request of all kinds of
operations. The specially designed color temperature calibration system may provide an
approximate color temperature to natural daylight, and so reflect the original colors of the
bodies tissue, which allows doctor's diagnosis to be performed correctly.

2. Brightness adjustment and assistant lamp test may be conducted with the power supply
switch, which has functions of 10-band brightness selection and brightness memory. A
LED progress bar for brightness indication indicates the brightness needed for operation.
There is a failure indicator of the main lamp under the board, which alerts for bulb
replacement. A wide voltage allows for a l 80V-250V AC with a strong resistance for
interference.

3. Two bulbs with the same power can automatically and instantaneously switch on by an
intellect circuit ensuring reliable continuous lighting.

4. A universal hanging device designed to provide maximal region of accommodation could


work in any operating room. The structure is light and handy, easy to handle with precise
positioning.

5. Tot~lly enclosed sleek lamp holder meets the request of illumination, sterilization and
!amm~r flow. Advanced multi-direct reflection system makes the consumption of light
mtens1ty the lowest, and the luminance may be above 150000K at minimal light spot.
Primary Technica l Spec ifications

SL & FL Operation L am p
I
Primary Technica l Suspended La mp Wall Lamp Stand Lamp
Paramete rs
Nominal lumina11ce 60000IX 45000lx 20000 lx 45000\ x 20000 \x
Main vo ltage 220Y± 22Y S0Hz.t \H z 50H z±IH z
220Y± 22Y 220Y± 22Y 50Hz-LIH z
Bulb voltage 24 V 12V 12V
Bulb power 15OW l lOW SOW 11 OW SOW \

5
l'rochu·t Schcmntkl'I

I lw prim:ipnl s1ru~·1u1\.: or /\d \'1111~~d l11stru11H.:11t 111ion SI ./I· I . Sc.: ri cH ( >rc r:itin g I .at11p'I arc
~,)1npri~t·d l,r' n li g. ht. h:il1111l·1.: d unn . llnri 1n11 t1 d 11rn1 . li xcd hol de r, suspe nd ed arm . ~,nd tru H.., arm
:-\\ i1d1 h,)~.

FIJ,?un· I: S11spl'ntlc1I Op(1rnU111,! Lnmp Schcmntlc

1-Light, 2-Balanced Arm, 3-Horizontal Arm, 4-Fixed Holder, 5-Suspended Arm, 6-Truss Arm , 7-
Switch Box, 8-Central Sterilizing Handle

Figure 2: Wall and Operating Lamp Schematic

100·
0 ,

:-Ligl~t. 2-Switch. 3-Suspendcd Arm, 4-Balanced Ann I-Central Sterilizing I landle. 2-Lamp Holder,
)-J-lonzontal Arm. 6-Fixeu Holder. 7-Wall 3- Panel of Switch Box, 4- Suspended Arm.
5-Balance an11, 6- Horizontal Ann, 7- Fixed Holder,
8-Wal\
I
F igure 3 : Opcraling Lump Shand Schcmalic

1-Balanced Arm, 2-Suspended Arm, 3-Switch, 4- Light, 5-Main Switch,


6-Stand Pole, 7-Fixed Holder, 8-Central Sterilizing Handle, 9- Fixed Holder

7
11111

'
, Table 2
t\ction l{angc Param eter

1
A ction Actio n range -1

Suspended W a ll lamp Stand lamp


lamp
Horizontal rotating arm circ ling 360° 180'
fixed hand le
Balanced rotating a rm circ ling 360° 180°
horizontal arm
Suspended rotating am1 circling 360° 360° 360°
balanced am1
Rotating truss arm circling 240° 240° 240°
suspended arm
Balanced rotating arm up/down 40° 40° 40°
Rotating light circling truss arm 180°

Figure 4: Schematic of operation panel

Auxiliary Lamp Indicator Primary Lamp Indicator On/Off indicator

o- b o·

. Advanced

Main/Auxiliary Light Switch Subdued Light Highlight On/Off


I nsh1lluforn I nst.ru cf'ion ,"

I. Open the prn.: kagc :md chee k ii' nll lhc purls un.; uvail uhl c accord ing to Lhc part-, li sl.

.., Chec k if the opera tion room con firm s with the instull ation position according to condition3.
The condition or su ppo rl.ing lh c we ight is nnt less thun 500kg; the height of the fl oor to the
ceiling should be between 2.8 and 3.2 meters. I I' the roof" is fl at, user mwit drill through
ceiling to install the fi xed bourd with bo lls. U type framework , which is machined, has to fi x
the girder if there is a girde r amo ng th e operat ion room.

3. Drill a hole, then fi x the U type framework on the girder or ceiling wi.th the expanding bolt
Ml2. Atter. fi x the linked board to the U type framework with 4 bolts MI 2.

4. Fix the fi xed set to the linl<ed board with bolt MI 2, check with a leveler.

5. The circumference axis of the level arm is inserted to the fixed board; fix the level arm with
8 sinking head stainless steel bolts M8x 15(Note: the wire has to be clamped when installing
the level arm).

6. In order to install the balance arm the input of balance arm and the output socket of level
arm suspender are connected together. The level arm is installed on the level arm suspender
with 4 sinking head stainless steel bolts M6* 12.

7. In order to install the lamp head, take out the safety block guard ring, set the balance arm
tie-in; the input socket of lamp head and the output socket of the balance arm are connected
together. Fix the lamp head with 5mm thin arc safety block. The safety block guard rings
with 1 sinking head stainless steel bolt M3><8

8. In order to install the lamp holder, first install the 24V/l 50w bulb in the lamp holder, then
put the lamp holder in the lamp head (the plastic orientation pole of the lamp holder aims at
the square gap), fix tightly the 3 fixed bolts M5*25 , finally install the plastic mixed handle.

9. In order to install the handles take them out of the box, the drilled hole on the handles goes
at the bullet position of the lamp holder, push the two together.

I 0. To perform the power supply installation cut the power off. The power switch are in series,
so the blue wire is for the level ann which is connected to the positive end on the switch,
and the black wire is connected to the negative end. Connect L and N on the switch, then
tum on the power, press the on button on the power switch. Check if the light bulb is on. If
it is not working well , first check if there is 220V power. If there is 220V power, then check
if there is 24V DC. If there is 24V DC, then check if all connections are in order; if all
connections are in order, check light bulb.

9
I
I I . The opnating lamp hcud ean·r he s topped at o pos ition ,.vhen try in g to adjus t i r there is no
force ;1pplicd. /\ s crewuri vcr c.:on he used 10 adjus t the res istance holt.

12. To adjus t the balance ann and body when the lamp head s ways up and down. adjust the
res is tance bolt located w here the lamp head moves around the balance ann. Adjust the
res is tance bolt when the movement or the lamp head a rou nd the fixed board is unbalanced .
Ir the s pan is too big between th e balance arm and lamp head , use a 10mm board to adjust
the bolt of balance am, .

13. To avo id the lamp ·s covers to fall off, take out two covers after adjusting, use IO tape ring
screw MJ to fix the m , then use I cover ring with inner hexagon bolt M5 ><25 to fix covers .

* Note:
• The lamp 's connected parts should be checked regularly to avoid fallen parts.

• . Only trained technicians should perform installation and connection of the


equipment.

• Shut the power off when the equipment 1s not m use. The transformer could be
damaged if left on too long.

• Damaged bulbs should be replaced in time after operation.

• Only a trained technician should repair operating lamps.

• The lamp should be cleaned with a dry or damp cloth. Do not use a wet cloth to
clean.
Operating f n.1i1tructions

I. To position the operating lamps use the cantilever sys tem, which can be c1 utomatic..dly
positioned when lamp holder is adjusted. ft can also be positioned hy holding th<.: trus:-;
am1. The wall stand and intervcntional lump can be positioned by holding tht: suspended
ann.

2. To turn on the general power switch, which is located on the vertical pole, the red
indicator lamp on the operation panel will light up.

3. Press on/off button on operation panel to turn on the lamp, the green indicator light on
the operation panel will light up, and the light beam will illuminate the operation table.

4. To adjust and focus the beam, hold the central sterilizing handle, position the angle of
the lamp holder until the beam illuminates the operation area. Then rotate the central
sterilizing handle clockwise or counter-clockwise to adjust the spot size.

5. To adjust the brightness, press the lighted button on the operation panel and the
brightness will adjust gradually.

6. There are main/auxiliary lamp bulbs inside the lamp. When the main lamp works, the
indicator light on the operation panel lights up. When the main lamp fails, the circuit wilI
automatically switch to the auxiliary lamp, and the indicator light on the operation panel
will light at the same time. If the system can't automatically switch to the auxiliary lamp,
press the button of the main/auxiliary lamp switch to manual. When finished with
operation, the broken bulb should be replaced immediately.

7. Press the on/off button on the operation panel to turn off the operating lamp and shut off
general power switch. The operating lamp stand should be in a safe position to avoid
injuring personnel or damaging equipment.

ll
('lcnnin~ und Sterilization

Cleaning and sterili zation process is :.1s rollows:

I. Depress the hundk lo take it out. and then soa k il in f<Jrmal<lchydc for 2 0 minutes. A bo,
ultra v iolet irrndintion cun be uppl icd for steri Iizalion of the handle .

The gloss bell must be cleaned and sterilized prior to operation, po li sh the g lass s urface
with a son cotton cloth by dipping it in formaldehyde or sterile water.

3. The switch box and operation paneJ must be cleaned and sterilized pr1or to operation.
Clean and polish the glass surface with a soft cotton cloth by dipping it in formaldehyde
or medical alcohol. Cotton cloth for sterilization cannot contain too much water in order
to avoid a circuit failure.

4. Lampshade and other parts should be sterilized periodically by cleaning and polishing
these parts with a cotton cloth by dipping it in formaldehyde or medical alcohol, avoid
the cotton cloth containing too much water.

*Note: The power switch must be off prior to cleaning and sterilizing the operating lamp.
Prccnutions

1. It is im po rta nt to chec k w hether ma in/auxili ary lamp wo rks properl y pri or to t·


b • ·
Y p 1ess111g t I1e marn/a
·
ux d- ia
• .
ry lamp switch button on operation panel. O pera 10n,

2. ~ tte r o~era ti o n fini s hes. every powe r s witch s ho uld be turned off according to operation
mstructi o ns.

3 . Una uthorized pe rsonne l must not repair the switch box.

4 . The equipment must not be used under a gaseous, flammable or explosive environment.

5. Maintenance should be conducted twice a year, which should be recorded. Advanced


Instrumentations will not assume the responsibility of failure caused poor maintenance.

Transportation and Storage

1. The equipment should be stored in its original box with appropriate padding.

2. The equipment should be protected from humidity, rain, being inverted, over-pressed and
rolled while being transported.

3. Temperature should be 10°C - 40°C, and r~lative humidity should be less than 80%, with
good ventilation, while avoiding direct sunlight.

13
..:,,u ir,m cnt I I ,wdlin'-'

,--
S,111prom _ ,_ I >i :. 11;11m i, l<c..of uti,,n
.j Pre<.auti,m•, -- -
Check whclhcr th e
gcncrn / power .'! witc h i5
Indicator /amp on '/ um on general p<Jv1c-r
turn ed on and w hether
opera ti on pane/ isn't 5witch, then tum on the::
the switch box on
working s witch on operati on panel
opera tion panel is tu rned
on. I

Loosen three screws in


central handle; take off
General power sv;itch
handle, then loosen the
Main or auxiliary A uxi Jiary lamp doesn;t must be shut off and
screw fixing the bulb;
lamp isn't working work bulb must cool prior
replace the bad bulb; and
to operation.
install handle at lamp
holder. I
Operation panel is Info rm biomedical
Unauthorized I
Failure in switch box damaged due to long- engineer or local
personnel should not
on operation panel term use or it has a maintenance center for
open the S\\:itch box
circuit failure . reparr.

*If the failure cannot be solved according to instructions inform Advanced Instrumentations or
a 1o ca 1 maintenance center. Advanced Instrumentations does not take responsibility for the
repair by unauthorized personnel.

14
AAdvanced Instrumentations

6800 N.W. 77th Court


Miami, Fl. 33166
Phone: (305) 477-6331
Fax: (305) 477-5331
Website: www.advanced-inst.com
Email: info@advanced-inst.com