Device parametrized
Name, Date, Signature

Device parametrization checked


Name, Date, Signature

Device parametrization released


Name, Date, Signature

08/01/2020 10:42 SA Page 2


Prowirl 200

Prowirl 200

Prowirl 200
Access status tooling Service

Operation / Operation
Language English Locking status ---
Access status display Maintenance

Operation / Display / Display


Format display 2 values Display interval 5s
Contrast display 40 %

Operation / Totalizer handling / Totalizer


Control Totalizer 1 Totalize Preset value 2 0.00000 kg
Preset value 1 0.00000 t Control Totalizer 3 Totalize
Control Totalizer 2 Totalize Preset value 3 0.00000 Nm³

Operation / Totalizer handling / Operation


Reset all totalizers Cancel

Setup / Setup
Device tag Prowirl

Setup / Commissioning / HART info


HART short tag PROWIRL HART address 0
Device tag Prowirl

Setup / Commissioning / System units


Volume flow unit m³/h Mass flow unit kg/h

Setup / Commissioning / Medium property


Select medium Liquid Fixed temperature 20.0000 °C
Fixed density 965.0000 kg/m³

Setup / Commissioning / HART output


Assign PV Mass flow Assign TV Totalizer 1
Assign SV Temperature Assign QV Totalizer 2

08/01/2020 10:42 SA Page 3


Prowirl 200

Setup / Commissioning / Current output 1


Assign current output 1 Mass flow 20 mA value 100000.000 kg/h
Current span 4...20 mA NAMUR Failure mode Max.
4 mA value 0.000 kg/h

Setup / Commissioning / Totalizer 1


Assign process variable Mass flow Unit totalizer 1 t
Failure mode Stop

Setup / Medium selection / Medium selection


Select medium Liquid

Setup / System units / System units


Volume flow unit m³/h Temperature unit °C
Volume unit m³ Velocity unit m/s
Mass flow unit kg/h Density unit kg/m³
Mass unit kg Dynamic viscosity unit cP
Corrected volume flow unit Nm³/h Length unit mm
Corrected volume unit Nm³

Setup / Current output 1 / Current output


Assign current output 1 Mass flow 20 mA value 100000.000 kg/h
Current span 4...20 mA NAMUR Damping output 1 2.0 s
4 mA value 0.000 kg/h Failure mode Max.

Setup / Pulse/frequency/switch output 1 / Pulse/frequency/switch output


Operating mode Pulse Pulse width 100.00 ms
Assign pulse output 1 Mass flow Failure mode No pulses
Value per pulse 10.00000 kg Invert output signal No

Setup / Low flow cut off / Low flow cut off


Sensitivity 5.0 Turn down 100 %

Setup / Advanced setup / Advanced setup


Enter access code 0

Setup / Advanced setup / Medium properties / Medium properties


Reference density 965.0000 kg/m³ Dynamic viscosity 1.0000 cP

Setup / Advanced setup / External compensation / External compensation


Fixed density 965.0000 kg/m³ Fixed temperature 20.0000 °C

Setup / Advanced setup / Sensor adjustment / Sensor adjustment


Inlet configuration Off Mating pipe diameter 0.0000 mm
Inlet run 285.0000 mm Installation factor 1.0000

08/01/2020 10:42 SA Page 4


Prowirl 200

Setup / Advanced setup / Totalizer 1 / Totalizer 1


Assign process variable Mass flow Failure mode Stop
Unit totalizer 1 t

Setup / Advanced setup / Totalizer 2 / Totalizer 2


Assign process variable Mass flow Failure mode Stop
Unit totalizer 2 kg

Setup / Advanced setup / Totalizer 3 / Totalizer 3


Assign process variable Corrected volume flow Failure mode Stop
Unit totalizer 3 Nm³

Setup / Advanced setup / Display / Display 1


Format display 2 values Value 2 display Totalizer 1
Value 1 display Mass flow Decimal places 2 x.xx
0% bargraph value 1 0.0000 kg/h Value 3 display None
100% bargraph value 1 270000.0000 kg/h Value 4 display None
Decimal places 1 x.xx Language English

Setup / Advanced setup / Display / Display 2


Display interval 5s Header Device tag
Display damping 0.0 s Separator .

Setup / Advanced setup / Heartbeat setup / Heartbeat base settings / Heartbeat base settings
Plant operator -------------------------------- Location --------------------------------

Setup / Advanced setup / Configuration backup display / Configuration backup display


Operating time 13d17h06m31s Backup state None
Last backup Comparison result No backup available
Configuration management Cancel

Setup / Advanced setup / Administration / Administration


Define access code 0 Device reset Cancel

Diagnostics / Diagnostics
Actual diagnostics -------- Timestamp 13d17h03m04s
Timestamp 0d00h00m00s Operating time from restart 0d00h03m47s
Previous diagnostics S350 Pre-amplifier defective Operating time 13d17h06m31s

Diagnostics / Diagnostic list / Diagnostic list


Diagnostics 1 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 4 --------
Diagnostics 2 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 5 --------
Diagnostics 3 -------- Timestamp 0d00h00m00s

08/01/2020 10:42 SA Page 5


Prowirl 200

Diagnostics / Event logbook / Event logbook


Filter options All Clear event list Cancel

Diagnostics / Device information / Device information 1


Device tag Prowirl Device name Prowirl 200
Serial number P61BDD19000 Order code 7F2C25-2363/0
Firmware version 01.03.01 ENP version 2.02.00

Diagnostics / Device information / Device information 1 / Extended order code


Extended order code 1 7F2C25-AABCCAAAAAAD2 Extended order code 3
Extended order code 2 SKA1+AA

Diagnostics / Device information / Device information 2


Device revision 0x4 Device type 0x1138
Device ID 0x4a06a6 Manufacturer ID 0x11

Diagnostics / Measured values / Process variables / Process variables


Volume flow 0.0000 m³/h Flow velocity 0.0000 m/s
Corrected volume flow 0.0000 Nm³/h Temperature 20.0000 °C
Mass flow 0.0000 kg/h

Diagnostics / Measured values / Totalizer / Totalizer


Totalizer value 1 49.89087 t Totalizer overflow 2 0
Totalizer overflow 1 0 Totalizer value 3 52.24295 Nm³
Totalizer value 2 50433.43000 kg Totalizer overflow 3 0

Diagnostics / Measured values / Output values / Output 1


Output current 1 4.00 mA Terminal voltage 1 24 V
Measured current 1 4.00 mA

Diagnostics / Measured values / Output values / Output 2


Pulse output 0.000 Hz

Diagnostics / Data logging / Data logging


Assign channel 1 Off Assign channel 4 Off
Assign channel 2 Off Logging interval 1.0 s
Assign channel 3 Off Clear logging data Cancel

Diagnostics / Heartbeat / Performing verification / Performing verification


Year 10 Verification mode Internal verification
Month January Start verification Cancel
Day 1d Progress 100 %
Hour 12 h Status Not done
Minute 0 min Overall result Not done

08/01/2020 10:42 SA Page 6


Prowirl 200

Diagnostics / Heartbeat / Verification results / Verification results


Overall result Not done

Diagnostics / Simulation / Process value simulation


Assign simulation process variable Off

Diagnostics / Simulation / Output simulation


Current output 1 simulation Off Pulse output simulation Off

Diagnostics / Simulation / Diagnostic event simulation


Device alarm simulation Off Diagnostic event simulation Off
Diagnostic event simulation 65533

Expert / Expert
Locking status --- Access status tooling Service
Access status display Maintenance Enter access code 0

Expert / System / Display / Display 1


Language English Decimal places 1 x.xx
Format display 2 values Value 2 display Totalizer 1
Value 1 display Mass flow Decimal places 2 x.xx
0% bargraph value 1 0.0000 kg/h Value 3 display None
100% bargraph value 1 270000.0000 kg/h Value 4 display None

Expert / System / Display / Display 2


Display interval 5s Separator .
Display damping 0.0 s Contrast display 40 %
Header Device tag Access status display Maintenance

Expert / System / Configuration backup display / Configuration backup display


Operating time 13d17h06m33s Backup state None
Last backup Comparison result No backup available
Configuration management Cancel

Expert / System / Diagnostic handling / Diagnostic handling


Alarm delay 0s

Expert / System / Diagnostic handling / Diagnostic behavior / Diagnostic behavior


Assign behavior of diagnostic no. 350 Alarm Assign behavior of diagnostic no. 832 Warning
Assign behavior of diagnostic no. 371 Warning Assign behavior of diagnostic no. 833 Warning
Assign behavior of diagnostic no. 441 Warning Assign behavior of diagnostic no. 834 Warning
Assign behavior of diagnostic no. 442 Warning Assign behavior of diagnostic no. 835 Warning
Assign behavior of diagnostic no. 443 Warning Assign behavior of diagnostic no. 841 Warning
Assign behavior of diagnostic no. 801 Alarm Assign behavior of diagnostic no. 844 Warning
Assign behavior of diagnostic no. 828 Warning Assign behavior of diagnostic no. 870 Warning
Assign behavior of diagnostic no. 829 Warning Assign behavior of diagnostic no. 947 Alarm

08/01/2020 10:42 SA Page 7


Prowirl 200

Expert / System / Administration / Administration


Define access code 0 Software option overview Extended HistoROM,
Device reset Cancel Heartbeat Verification
Activate SW option 32771170

Expert / Sensor / Measured values / Process variables / Process variables


Volume flow 0.0000 m³/h Flow velocity 0.0000 m/s
Corrected volume flow 0.0000 Nm³/h Temperature 20.0000 °C
Mass flow 0.0000 kg/h Vortex frequency 0.0000 Hz

Expert / Sensor / Measured values / System values / System values


Vortex amplitude 0.0016 DSC sensor temperature PT1 NaN °C
Gap capacity 54.3303 pF DSC sensor temperature PT2 NaN °C
Gap capacity difference -2.5632 pF Sensor leakage current 0.4803 %

Expert / Sensor / Measured values / Totalizer / Totalizer


Totalizer value 1 49.89087 t Totalizer overflow 2 0
Totalizer overflow 1 0 Totalizer value 3 52.24295 Nm³
Totalizer value 2 50433.43000 kg Totalizer overflow 3 0

Expert / Sensor / Measured values / Output values / Output 1


Output current 1 4.00 mA Terminal voltage 1 24 V
Measured current 1 4.00 mA

Expert / Sensor / Measured values / Output values / Output 2


Pulse output 0.000 Hz

Expert / Sensor / System units / System units


Volume flow unit m³/h Temperature unit °C
Volume unit m³ Velocity unit m/s
Mass flow unit kg/h Density unit kg/m³
Mass unit kg Dynamic viscosity unit cP
Corrected volume flow unit Nm³/h Length unit mm
Corrected volume unit Nm³ Date/time format dd.mm.yy hh:mm

Expert / Sensor / System units / User-specific units / User-specific units


User volume text User vol. User corrected volume factor 1.0000
User volume factor 1.0000 User density text User dens.
User mass text User mass User density offset 0.0000
User mass offset 0.0000 User density factor 1.0000
User mass factor 1.0000 User specific-enthalpy text User enth.
User corrected volume text UserCrVol. User specific-enthalpy offset 0.0000
User corrected volume offset 0.0000 User specific-enthalpy factor 1.0000000

Expert / Sensor / Process parameters / Process parameters


Flow override Off Flow damping 1.00 s

08/01/2020 10:42 SA Page 8


Prowirl 200

Expert / Sensor / Process parameters / Low flow cut off / Low flow cut off
Sensitivity 5.0 Assign process variable Volume flow
Calculated min. flow coefficient 11.0000 On value low flow cutoff 0.0000 m³/h
Turn down 100 % Off value low flow cutoff 0%
Calculated min. vortex frequency liquid 5.1000 Hz

Expert / Sensor / Measurement mode / Measurement mode


Select medium Liquid

Expert / Sensor / Measurement mode / Medium properties / Medium properties


Reference density 965.0000 kg/m³ Dynamic viscosity 1.0000 cP

Expert / Sensor / External compensation / External compensation


Fixed density 965.0000 kg/m³ Fixed temperature 20.0000 °C

Expert / Sensor / Sensor adjustment / Installation settings


Inlet configuration Off Mating pipe diameter 0.0000 mm
Inlet run 285.0000 mm Installation factor 1.0000

Expert / Sensor / Sensor adjustment / Additional settings


Min. flow coefficient 11.0000 Sensor adjustment Cancel
Min. vortex frequency liquid 5.1000 Hz DSP mode ---
Max. vortex frequency liquid 330.0000 Hz

Expert / Sensor / Calibration / Calibration


Calibration factor 80645.1400000 1/m³ Sensor diameter 24.3000 mm
Meter body properties DN25F-PN40 Calibration date/time 18.06.19 11:34
Tube inner diameter 28.5000 mm

Expert / Sensor / Calibration / Recalibration / Recalibration


Calibration factor 1.000 1/m³ Hour 12 h
Year 10 Minute 0 min
Month January Confirm settings No
Day 1d

Expert / Sensor / Calibration / Logbook / Logbook


Calibration counter 1 Timestamp 18.06.19 11:34
Select calibration log entry 1 Calibration factor 80645.140 1/m³

Expert / Sensor / Testpoints / Testpoints


Signal quality 0.0000 Temperature reference 1 266.3888 °C
Resonance frequency NaN Hz Temperature reference 2 130.5168 °C
Reference frequency drift 0.0250 % Pre-amplifier temperature 47.1 °C
Overall reference capacity 85.5830 pF

08/01/2020 10:42 SA Page 9


Prowirl 200

Expert / Output / Current output 1 / Current output 1-1


Assign current output 1 Mass flow 20 mA value 100000.000 kg/h
Current span 4...20 mA NAMUR Damping output 1 2.0 s
4 mA value 0.000 kg/h Response time 3.0000 s

Expert / Output / Current output 1 / Current output 1-2


Failure mode Max. Measured current 1 4.00 mA
Output current 1 4.00 mA Terminal voltage 1 24 V
Start-up mode Min.

Expert / Output / Current output 1 / Current output 1-3


Trim Off Status 128

Expert / Output / Pulse/frequency/switch output / Pulse/frequency/switch output


Operating mode Pulse

Expert / Output / Pulse/frequency/switch output / Settings


Assign pulse output 1 Mass flow Failure mode No pulses
Value per pulse 10.00000 kg Pulse output 0.000 Hz
Pulse width 100.00 ms

Expert / Output / Pulse/frequency/switch output / Additional settings


Invert output signal No

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 350 Out of specification (S) Event category 371 Maintenance required (M)

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 441 Out of specification (S) Event category 443 Out of specification (S)
Event category 442 Out of specification (S)

Expert / Communication / Diagnostic configuration / Diagnostic configuration


Event category 801 Failure (F) Event category 835 Out of specification (S)
Event category 828 Out of specification (S) Event category 841 Out of specification (S)
Event category 829 Out of specification (S) Event category 844 Out of specification (S)
Event category 832 Out of specification (S) Event category 870 Out of specification (S)
Event category 833 Out of specification (S) Event category 947 Out of specification (S)
Event category 834 Out of specification (S)

Expert / Communication / HART input / Configuration / Configuration


Capture mode Off

Expert / Communication / HART input / Input / Input


Value 0.0000 Status Bad

08/01/2020 10:42 SA Page 10


Prowirl 200

Expert / Communication / HART output / Configuration / Configuration


HART short tag PROWIRL HART address 0
Device tag Prowirl No. of preambles 5

Expert / Communication / HART output / Burst configuration / Burst configuration 1 / Burst configuration
1 / Burst message 1
Burst mode 1 Off Burst trigger mode Continuous
Burst command 1 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Burst configuration / Burst configuration 2 / Burst configuration
1 / Burst message 2
Burst mode 2 Off Burst trigger mode Continuous
Burst command 2 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Burst configuration / Burst configuration 3 / Burst configuration
1 / Burst message 3
Burst mode 3 Off Burst trigger mode Continuous
Burst command 3 Command 2 Min. update period 1000 ms

Expert / Communication / HART output / Information / HART info


Device revision 0x4 HART descriptor PROWIRL
Device ID 0x4a06a6 HART message PROWIRL
Device type 0x1138 Hardware revision 1
Manufacturer ID 0x11 Software revision 4
HART revision 7 HART date code 2009-07-20

Expert / Communication / HART output / Output / HART output


Assign PV Mass flow Assign TV Totalizer 1
Primary variable (PV) 0.000 kg/h Tertiary variable (TV) 49.891 t
Assign SV Temperature Assign QV Totalizer 2
Secondary variable (SV) 20.000 °C Quaternary variable (QV) 50433.430 kg

Expert / Application / Application


Reset all totalizers Cancel

Expert / Application / Totalizer 1 / Totalizer 1


Assign process variable Mass flow Preset value 1 0.00000 t
Unit totalizer 1 t Failure mode Stop
Control Totalizer 1 Totalize

Expert / Application / Totalizer 2 / Totalizer 2


Assign process variable Mass flow Preset value 2 0.00000 kg
Unit totalizer 2 kg Failure mode Stop
Control Totalizer 2 Totalize

08/01/2020 10:42 SA Page 11


Prowirl 200

Expert / Application / Totalizer 3 / Totalizer 3


Assign process variable Corrected volume flow Preset value 3 0.00000 Nm³
Unit totalizer 3 Nm³ Failure mode Stop
Control Totalizer 3 Totalize

Expert / Application / Inventory counter / Inventory counter


Totalizer value 52226.8900 l Unit l
Totalizer overflow 0

Expert / Diagnostics / Diagnostics


Actual diagnostics -------- Timestamp 13d17h03m04s
Timestamp 0d00h00m00s Operating time from restart 0d00h03m54s
Previous diagnostics S350 Pre-amplifier defective Operating time 13d17h06m39s

Expert / Diagnostics / Diagnostic list / Diagnostic list


Diagnostics 1 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 4 --------
Diagnostics 2 -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Diagnostics 5 --------
Diagnostics 3 -------- Timestamp 0d00h00m00s

Expert / Diagnostics / Event logbook / Event logbook


Filter options All Clear event list Cancel

Expert / Diagnostics / Device information / Device information


Device tag Prowirl Order code 7F2C25-2363/0
Serial number P61BDD19000 Configuration counter 149
Firmware version 01.03.01 ENP version 2.02.00
Device name Prowirl 200

Expert / Diagnostics / Device information / Device information / Extended order code


Extended order code 1 7F2C25-AABCCAAAAAAD2 Extended order code 3
Extended order code 2 SKA1+AA

Expert / Diagnostics / Sensor information / Sensor information


DSC sensor serial number @f1902850141 Meter body ID 71333061
DSC ID 71345561 Pressure cell serial number
Meter body serial number 4156879 Pressure cell ID

Expert / Diagnostics / Pre-amplifier module / Pre-amplifier module


Hardware revision Serial number
Module name

Expert / Diagnostics / Mainboard module / Mainboard module


Software revision 10301 Hardware revision V1.01.00
Build no. software 2 Module name V308-HART

08/01/2020 10:42 SA Page 12


Prowirl 200

Expert / Diagnostics / Mainboard module / Mainboard module


Serial number 0154703031 Bootloader revision 701

Expert / Diagnostics / I/O module / I/O module


Software revision 20005 Module name IO211_212_SB
Build no. software 2124 Serial number 0154763905
Hardware revision V2.00.01 Bootloader revision 641

Expert / Diagnostics / Display module / Display module


Software revision 10204 Module name SD02
Build no. software 1 Serial number 0155510911
Hardware revision V2.00.00 Bootloader revision 630

Expert / Diagnostics / Data logging / Data logging


Assign channel 1 Off Assign channel 4 Off
Assign channel 2 Off Logging interval 1.0 s
Assign channel 3 Off Clear logging data Cancel

Expert / Diagnostics / Min/max values / Min/max values


Reset min/max values Cancel

Expert / Diagnostics / Min/max values / Terminal voltage / Terminal voltage


Minimum value 24 V Average value 24 V
Maximum value 29 V

Expert / Diagnostics / Min/max values / IO module temperature / IO module temperature


Minimum value 25.0 °C Average value 37.7 °C
Maximum value 46.0 °C

Expert / Diagnostics / Min/max values / Pre-amplifier temperature / Pre-amplifier temperature


Minimum value 23.5 °C Maximum value 49.5 °C

Expert / Diagnostics / Min/max values / Flow velocity / Flow velocity


Maximum value 6.8 m/s

Expert / Diagnostics / Min/max values / Gap capacity / Gap capacity


Minimum value 53.9683 pF Maximum value 54.4642 pF

Expert / Diagnostics / Min/max values / Gap capacity difference / Gap capacity difference
Minimum value -2.8142 pF Maximum value -2.4785 pF

Expert / Diagnostics / Min/max values / DSC sensor temperature PT1 / DSC sensor temperature PT1
Minimum value 100000000000000000000 Maximum value -
00000000000000.0 °C 100000000000000000000
00000000000000.0 °C

08/01/2020 10:42 SA Page 13


Prowirl 200

Expert / Diagnostics / Min/max values / DSC sensor temperature PT2 / DSC sensor temperature PT2
Minimum value 100000000000000000000
00000000000000.0 °C
Maximum value -
100000000000000000000
00000000000000.0 °C

Expert / Diagnostics / Min/max values / Vortex frequency / Vortex frequency


Maximum value 339.8042 Hz

Expert / Diagnostics / Heartbeat / Heartbeat base settings / Heartbeat base settings


Plant operator -------------------------------- Location --------------------------------

Expert / Diagnostics / Heartbeat / Performing verification / Performing verification


Year 10 Verification mode Internal verification
Month January Start verification Cancel
Day 1d Progress 100 %
Hour 12 h Status Not done
Minute 0 min Overall result Not done

Expert / Diagnostics / Heartbeat / Verification results / Verification results


Overall result Not done

Expert / Diagnostics / Simulation / Process value simulation


Assign simulation process variable Off

Expert / Diagnostics / Simulation / Output simulation


Current output 1 simulation Off Pulse output simulation Off

Expert / Diagnostics / Simulation / Diagnostic event simulation


Device alarm simulation Off Diagnostic event simulation Off
Diagnostic event simulation 65533

08/01/2020 10:42 SA Page 14


Extended HistoROM

0005:23:49:54 0007:03:36:34 0008:07:23:14 0009:11:09:54 0010:14:56:34 0011:18:43:14

Diagram Legend.
Overall statistics

Number of values recorded per channel: 0


Storage interval: s
First chronological measured value:
Last chronological measured value:

Channel statistics

Off Off Off Off

Mean value:
Standard deviation:
Minimum:
Maximum:
Range:

08/01/2020 10:42 SA Page 15


Extended HistoROM
# Date - Time Operating hours Diagnostic information Additional Information
(dddd:hh:mm:ss)
1 22/01/2020 3:47:36 SA 0013:17:04:55 I1091 Configuration changed
2 22/01/2020 3:47:25 SA 0013:17:04:44 I1091 Configuration changed
3 22/01/2020 3:47:06 SA 0013:17:04:25 I1398 CDI: access status changed
4 22/01/2020 3:46:50 SA 0013:17:04:09 I1091 Configuration changed
5 22/01/2020 3:46:43 SA 0013:17:04:02 I1091 Configuration changed
6 22/01/2020 3:46:31 SA 0013:17:03:50 I1091 Configuration changed
7 22/01/2020 3:45:45 SA 0013:17:03:04 S350 Pre-amplifier defective Change pre-amplifier
(Service ID:490)
8 22/01/2020 3:45:45 SA 0013:17:03:04 S350 Pre-amplifier defective Change pre-amplifier
(Service ID:490)
9 22/01/2020 3:45:37 SA 0013:17:02:56 I1398 CDI: access status changed
10 22/01/2020 3:45:22 SA 0013:17:02:41 I1089 Power on
11 - 0013:17:01:39 I1091 Configuration changed
12 - 0013:17:01:11 I1091 Configuration changed
13 - 0013:17:00:52 I1091 Configuration changed
14 - 0013:17:00:42 I1091 Configuration changed
15 - 0013:17:00:32 I1091 Configuration changed
16 - 0013:16:58:35 I1091 Configuration changed
17 - 0013:16:58:01 I1398 CDI: access status changed
18 - 0013:16:57:49 I1398 CDI: access status changed
19 - 0013:16:57:01 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
20 - 0013:16:56:57 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
21 - 0013:14:31:13 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
22 - 0013:14:31:11 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
23 - 0013:14:30:58 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
24 - 0013:14:30:56 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
25 - 0013:14:30:48 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
26 - 0013:14:30:38 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
27 - 0013:14:30:03 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)

08/01/2020 10:42 SA Page 16


Extended HistoROM
28 - 0013:14:29:58 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
29 - 0013:14:29:43 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
30 - 0013:14:29:28 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
31 - 0013:01:48:25 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
32 - 0013:01:48:23 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
33 - 0013:01:47:30 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
34 - 0013:01:47:25 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
35 - 0013:01:46:59 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
36 - 0013:01:46:52 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
37 - 0013:01:46:30 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
38 - 0013:01:46:26 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
39 - 0013:01:46:09 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
40 - 0013:01:46:01 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
41 - 0013:01:45:55 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
42 - 0013:01:45:49 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
43 - 0013:01:45:45 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
44 - 0013:01:45:25 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
45 - 0013:01:45:17 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)

08/01/2020 10:42 SA Page 17


Extended HistoROM
46 - 0013:01:45:06 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
47 - 0013:01:44:59 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
48 - 0013:01:44:41 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
49 - 0013:01:44:29 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
50 - 0013:01:44:13 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
51 - 0012:19:01:22 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
52 - 0012:19:01:20 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
53 - 0012:19:00:20 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
54 - 0012:19:00:18 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
55 - 0012:18:51:29 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
56 - 0012:18:51:27 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
57 - 0012:15:03:10 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
58 - 0012:15:03:06 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
59 - 0012:15:02:56 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
60 - 0012:15:02:53 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
61 - 0012:15:02:49 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
62 - 0012:15:02:41 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
63 - 0012:11:40:25 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)

08/01/2020 10:42 SA Page 18


Extended HistoROM
64 - 0012:11:40:16 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
65 - 0010:09:54:31 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
66 - 0010:09:54:25 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
67 - 0010:09:54:22 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
68 - 0010:09:53:45 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
69 - 0010:09:53:40 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
70 - 0010:09:53:34 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
71 - 0009:21:53:35 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
72 - 0009:21:53:23 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
73 - 0009:21:53:19 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
74 - 0009:21:53:15 S870 Measuring inaccuracy increased 1. Check process
2. Increase flow volume
(Service ID:441)
75 - 0009:01:49:32 I1089 Power on
76 - 0008:13:01:10 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
77 - 0008:13:01:04 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
78 - 0008:12:59:10 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
79 - 0008:12:58:53 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
80 - 0008:12:58:40 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)

08/01/2020 10:42 SA Page 19


Extended HistoROM
81 - 0008:12:58:16 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
82 - 0008:12:58:10 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
83 - 0008:12:57:55 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
84 - 0008:12:55:40 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
85 - 0008:12:55:35 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
86 - 0008:12:55:24 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
87 - 0008:12:55:16 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
88 - 0008:12:53:40 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
89 - 0008:12:53:32 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
90 - 0008:12:53:25 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
91 - 0008:12:53:15 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
92 - 0008:10:24:40 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
93 - 0008:10:24:24 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
94 - 0008:10:24:10 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)

08/01/2020 10:42 SA Page 20


Extended HistoROM
95 - 0008:10:23:53 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
96 - 0008:10:23:40 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
97 - 0008:10:23:33 S948 Signal quality bad 1. Check process conditions: wet gas,
pulsation
2. Check installation: vibration
(Service ID:917)
98 - 0006:01:40:22 S350 Pre-amplifier defective Change pre-amplifier
(Service ID:490)
99 - 0006:01:40:21 S350 Pre-amplifier defective Change pre-amplifier
(Service ID:490)
100 - 0006:01:39:58 I1089 Power on

08/01/2020 10:42 SA Page 21