Documenti di Didattica
Documenti di Professioni
Documenti di Cultura
User Information
User Name
Address
Tel
Fax
E-Mail
Notes
Device parametrized
Name, Date, Signature
Promass 100
Promass 100
Access status tooling Service
Operation / Operation
Access status tooling Service Locking status ---
Setup / Setup
Device tag Promass 100
Setup / Advanced setup / Calculated values / Corrected volume flow calculation / Corrected volume flow
calculation
Corrected volume flow calculation Linear expansion coefficient 0.0000
Calculated reference density Square expansion coefficient 0.0000
Reference temperature 20.0000 °C
Setup / Advanced setup / Sensor adjustment / Zero point adjustment / Zero point adjustment
Zero point adjustment control Cancel Progress 100 %
Setup / Advanced setup / Heartbeat setup / Heartbeat base settings / Heartbeat base settings
Record application reference data Cancel Plant operator --------------------------------
Progress 100 % Location --------------------------------
Status Not done
Diagnostics / Diagnostics
Actual diagnostics -------- Timestamp 0d00h00m00s
Timestamp 0d00h00m00s Operating time from restart 0d18h13m49s
Previous diagnostics -------- Operating time 746d18h33m07s
Expert / Expert
Locking status --- Enter access code 0
Access status tooling Service
Expert / Sensor / Process parameters / Low flow cut off / Low flow cut off
Assign process variable Mass flow Off value low flow cutoff 50 %
On value low flow cutoff 10.0000 kg/min Pressure shock suppression 0.1 s
Expert / Sensor / Process parameters / Partially filled pipe detection / Partially filled pipe detection
Assign process variable Off
Expert / Sensor / Calculated values / Corrected volume flow calculation / Corrected volume flow
calculation
Corrected volume flow calculation Linear expansion coefficient 0.0000
Calculated reference density Square expansion coefficient 0.0000
Reference temperature 20.0000 °C
Expert / Sensor / Sensor adjustment / Zero point adjustment / Zero point adjustment
Zero point adjustment control Cancel Progress 100 %
Expert / Sensor / Sensor adjustment / Process variable adjustment / Process variable adjustment
Mass flow offset 0.0000 kg/s Mass flow factor 1.0000
Expert / Sensor / Sensor adjustment / Process variable adjustment / Process variable adjustment
Volume flow offset 0.0000 m³/s Corrected volume flow factor 1.0000
Volume flow factor 1.0000 Reference density offset 0.0000 kg/Nm³
Density offset 0.0000 kg/m³ Reference density factor 1.0000
Density factor 1.0000 Temperature offset 0.0000 K
Corrected volume flow offset 0.0000 Nm³/s Temperature factor 1.0000
Event List
# Date - Time Operation Hours Event Description
(dddd:hh:mm:ss)
1 18-02-2020 9:19:06 0746:18:33:14 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
2 18-02-2020 9:19:06 0746:18:33:14 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
3 18-02-2020 9:19:00 0746:18:33:08 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
4 18-02-2020 9:19:00 0746:18:33:08 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
5 18-02-2020 9:18:28 0746:18:32:36 I1398 CDI: access status
AM changed
6 18-02-2020 9:17:36 0746:18:31:44 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
7 18-02-2020 9:17:36 0746:18:31:44 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
8 18-02-2020 9:17:25 0746:18:31:33 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
9 18-02-2020 9:17:25 0746:18:31:33 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
10 18-02-2020 9:16:41 0746:18:30:49 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
11 18-02-2020 9:16:41 0746:18:30:49 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
12 18-02-2020 9:16:41 0746:18:30:49 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
13 18-02-2020 9:16:41 0746:18:30:49 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
14 18-02-2020 9:16:24 0746:18:30:32 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
15 18-02-2020 9:16:24 0746:18:30:32 S842 Process limit Low flow cut off active!
AM 1. Check low flow cut off configuration
(Service ID:145)
Promass F
Calibration type: Standard calibration (Field density adjustment: No )
Liquid/Application: Defined liquid Aquaeous glucose
Table data of concentration range [%] Low: 0 High: 100
Table data of temperature range [°C] Low: 10 High: 45
Quantity of table data:
Measuring uncertainty of density: +/- 0.01 kg/l
Coefficients overview
Calculated coefficients Calculated coefficients
A0 B1 E-3
A1 B2 E-6
A2 B3 E-9
A3
A4
[%]
2.00
0.00
-2.00
-4.00
Calculation model
Compensation coefficient X1 NaN
Compensation coefficient X2 NaN
Reference temperature NaN °C
Temperature unit °C
Dynamic viscosity unit/ cP
Kinematic viscosity unit
0.40
0.30
0.20
0.10
0.00
10.00 20.00 30.00 40.00 50.00 60.00 70.00 [°C]
measured viscosity.
temperature corrected viscosity
reference point